RIGOL is a registered trademark of RIGOL TECHNOLOGIES, INC.
Publication Number
QGA22102-1110
Notices
RIGOL products are co vere d by P.R.C. and foreign patents, issued and
pending.
RIGOL reserves the right to modify or change parts of or all the
specifications and pricing policies at the company’s sole decision.
Information in this publication replaces all previously released materials.
Information in this publication is subject to change without notice.
RIGOL shall not be liable for either incidental or consequential losses in
connection with the furnishing, use, or perform a nce of this manual, as
well as any information contained.
Any part of this document is forbidden to be copied, photocopied, or
rearranged without prior written approval of RIGOL.
Product Certification
RIGOL guarantees that this product conforms to the national and industrial
standards in China as well as the ISO9001:2008 standard and the
ISO14001:2004 standard. Other international standard conformance
certifications are in progress.
Contact Us
If you have any problem or requirement when using our products or this
manual, please contact RIGOL.
E-mail: service@rigol.com
Website: www.rigol.com
MSO7000/DS7000 Quick Guide 1
RIGOL
English
1. Only the exclusive power cord
7. Use the proper fuse.
WARNING
avoided, will result in serious injury or death.
CAUTION
important data.
DANGER
It calls attention to an operation, if not correctly performed,
could result in injury or hazard immediately.
WARNING
It calls attention to an operation, if not correctly performed,
could result in potential injury or hazard.
CAUTION
It calls attention to an operation, if not correctly performed,
connected to the product.
General Safety Summary
designed for the instrument
and authorized for use within
the local country could be
used.
2. Ensure that the instrument is
safely grounded.
3. Observe all terminal ratings.
4. Use proper overvoltage
protection.
5. Do not operate without covers.
6. Do not insert objects into the
air outlet.
8. Avoid circuit or wire exposure.
9. Do not operate the instrument
with suspected failures.
10. Provide adequate vent il at io n .
11. Do not operate in wet conditions.
12. Do not operate in an explosive
atmosphere.
13. Keep instrument surfaces clean
and dry.
14. Prevent electrostatic impact.
15. Handle with caution.
Safety Notices and Symbols
Safety Notices in this Manual:
Indicates a potentially hazardous situation or pr actice which, if not
Indicates a potentially hazardous situation or pr actice which, if not
avoided, could result in damage to the product or loss of
Safety Terms on the Product:
could result in damage to the product or other devices
2MSO7000/DS7000 Quick Guide
English
Safety Symbols on the Product:
Hazardous
Safety
Protective Earth
Chassis Ground
Test Ground
CAUTION
liquids.
WARNING
to the power supply.
Tip
of RIGOL (www.rigol.com).
RIGOL
Voltage
Warning
Terminal
Care and Cleaning
Care
Do not store or leave the instrument where it may be exposed to direct sunlight
for long periods of time.
Cleaning
Clean the instrument regularly according to its operating conditions.
1. Disconnect the instrument from all power sources.
2. Clean the external su r fac es of th e inst ru ment with a soft cloth dampened
with mild detergent or water. When cleaning the LCD, take care to avoid
scarifying it.
To avoid damage to the instrument, do no t expose it to cau stic
To avoid sh ort-circuit resulting from moisture or personal injuries,
ensure that the in strum en t is c om pletely dry b efore c onn ect ing it
Document Overview
This manual gives you a quick review about the front and rear panel of
MSO7000/DS 7 0 0 0 series, the user interface, and th e basic operation method.
For the latest version of this manual, download it from the official website
MSO7000/DS7000 Quick Guide 3
RIGOL
English
Label
Knob
Label
Knob
Horizontal
SCALE
Horizontal Time
Base Knob
Vertical
SCALE
Vertical Scale
Knob
Horizontal
POSITION
Horizontal
Position Knob
Vertical
OFFSET
Vertical Offset
Knob
Wave Vertical
SCALE
Waveform Vertical
Scale Knob
Trigger
LEVEL
Trigger Level
Knob
Wave Vertical
POSITION
Waveform Vertical
Position Knob
Multifunction
Knob
Model
Analog
Bandwidth
No. of
Analog
Channels
No. of Function/Arbitrary
Waveform Generator
Channels
No. of
Digital
Channels
DS7054
500 MHz
4
——
——
DS7034
350 MHz
4
——
——
DS7024
200 MHz
4
——
——
DS7014
100 MHz
4
——
——
MSO7054
500 MHz
4
2, Opt.
16
MSO7034
350 MHz
4
2, Opt.
16
MSO7024
200 MHz
4
2, Opt.
16
MSO7014
100 MHz
4
2, Opt.
16
Format Conventions in this Manual
1. Key
The key on the front panel is denoted by the format of "Key Name (Bold)
+ Text Box" in the manual. For example, Utility denotes the "Utility" key.
2. Menu
The menu items are denoted by the format of "Menu Word (Bold) +
Character Shading". For example, System denotes the "System" menu
item under Utility.
3. Operation Procedures
denotes the next step of operation. For example, Utility System
denotes that first press Utility, and then press the System softkey.
4. Knob
Content Conventions in this Manual
MSO7000/DS7000 series includes the following models. Unless otherwise
specified, this manual takes MSO7054 as an example to illustrate the functions
and operation methods of MSO7000/DS7000 series digital oscilloscope.
4MSO7000/DS7000 Quick Guide
RIGOL
English
General Inspection
1. Inspect the packaging
If the packaging has been damaged, do not dispose the damaged
packaging or cushioning materials until the shipment has been checked
for completeness and has passed both electrical and mechanical tests.
The consigner or carrier shall be liable for the damage to the instrument
resulting from shipment. RIGOL would not be responsible for free
maintenance/rework or replacement of the instrument.
2. Inspect the instrument
In case of any mechanical damage, missing parts, or failure in passing the
electrical and me chanical tests, contact your RIGOL sales
representative.
3. Check the accessories
Please check the accessories according to the packing lists. If the
accessories are damaged or incomplete, please contact your RIGOL
sales representative.
Product Over v iew
MSO7000/DS7000 series is a multifunctional and high-performance digital
oscilloscope designed on the basis of the Ultra Vision II technology developed
by RIGOL. Integrating 6 independent instruments into one, including digital
oscilloscope, 16-channel logic analyzer, dual-channel function/arbitrary
waveform generator, d igital voltmeter, 6-digit counter and totalizer, and
protocol analyzer , the MSO7000/DS7000 series is equipped with super high
sample bandwidth ratio, extremely high memory depth, clear display, excellent
waveform capture rate, and powerful data analysis functions. Many of its
specifications have reached the top level in the industry. It provides sound
solutions for mainframes, optional&accessories, and application software. It
has aroused great attention from customers in the areas such as industrial
control, power supply, and automotive electronics.
For descriptions of the front panel, refer to Figure 1 and Table 1; for
descriptions of the rear panel, refer to Figure 2 and Table 2; and for
descriptions of the main interface (display screen), refer to Figure 3 and Table
3.
MSO7000/DS7000 Quick Guide 5
RIGOL
English
Horizontal
OFFSET
Zero
OFFSET
Zero
Intensity
OFFSET
Zero
50%
Wave
Vertical
Menu
T
rigger
Naviga te
Digital Oscilloscope
500MHz 10GSa/s
GⅠGⅡ
For output only
Default
Touch
Lock
Menu
Mode
Force
Math L A
CLEAR
Cursor
Quick
Utility
Display
Zoom
Ref
12
GGⅠⅡ
34
Decode
StorageAcquire
Measure
AUTO
RUN
STOP
SINGLE
Off
Back
Menu
Search
No.
Description
No.
Description
1
Capacitive Touch Screen
11
RUN/STOP Key
2
Function Menu Operation Keys
12
Single Trigger Control Key
3
Quick Key (Shortcut Key)
13
Trigger Control System
4
Multifunction Knob
14
Vertical Control System
5
Common Operation Keys
15
Analog Channel Input Terminals
6
Touch Lock Key
16
Probe Compensation Signal Output
Terminal/Ground Terminal
7
Horizontal Control System
17
Digital Channel Input Interface
[1]
8
CLEAR Key
18
Dual-channel Function/Arbitrary
Waveform Generator Output
Terminals
[2]
9
Auto Waveform Display Key
19
USB HOST Interface
10
Waveform Control System
20
Power Key
1 2 34 5 6 7 8 910111213
20 19 18 17 16 15 14
Figure 1 Front Panel
Table 1 Front Panel Description
[1]
Note:
This function is only available for the MSO7000 model.
[2]
Note:
This function is only available for the MSO7000 model installed with the Function/Arbitrary Waveform
Generator option.
6MSO7000/DS7000 Quick Guide
English
1
Handle
2
LAN Interface
3
USB DEVICE Interface
4
5
6
7
8
9
10
1
2 3 4 5
10 9 8 7 6
RIGOL
Figure 2 Rear Panel
Table 2 Rear Panel Description
No. Description
USB HOST Interface
HDMI Video Output
EXT Trigger Input
Trigger Output Interface
Lock Hole
Fuse
AC Power Cord Connector
MSO7000/DS7000 Quick Guide 7
RIGOL
English
No.
Description
No.
Description
1
Analog Channel Label and Waveform
13
Operation Menu
2
Operating Status
14
Notification Area
3
Horizontal Time Base
15
Arbitrary Waveform Generator 2
Waveform Label
4
Sample Rate and Memory Depth
16
Arbitrary Waveform Generator 1
Waveform Label
5
Auto Measurement Label
17
Digital Channel Status Area
6
Waveform Memory
18
CH4 Status Label
7
Trigger Position
19
Message Box
8
RUN/STOP Label
20
CH3 Status Label
9
Horizontal Position
21
CH2 Status Label
10
Trigger Type
22
CH1 Status Label
11
Trigger Source
23
Digital Channel Label and Waveform
12
Trigger Level/Threshold Level
24
Function Navigation
24 23 22 21 20 1918 17 16 15 14 13
1 2 3 4 5 6 7 8 9 1011 12
Table 3 User Interface Icons
Figure 3 User Interface
8MSO7000/DS7000 Quick Guide
RIGOL
English
(a) To unfold the supporting legs
(b) T o fold th e suppo rting legs
Power Cord Connector
Supporting Legs
To Prepare for Us e
To Adjust the Supporting Legs
Adjust the supporting legs properly to use them as stands to tilt the oscilloscope
upwards for stable placem en t of the oscilloscope as well as better operation
and observation, as shown in Figure 4.
To Connect to AC Power
The input AC power requirements of the oscilloscope are 100~240 V, 45~440
Hz. Please use the power cord provided in the accessories to connect the
oscilloscope to the AC power source via the power cord connector , as shown in
Figure 5.
MSO7000/DS7000 Quick Guide 9
Figure 4 To Adjust the Supporting Legs
Figure 5 To Connect to AC Power
RIGOL
English
Turn-on Checkout
When the oscillo scope is connected to power, press the Power key
lower-left corner of the front panel to start the oscilloscope. (You can also press
Utility System Power status, and select "Switch On". After the
instrument is connected to po wer source, it will start directly. You can also
enable the to uch screen to perform th e above menu operation.) During the
start-up process, the oscilloscope performs a series of self-tests. After the
self-test, the welcome screen is displayed.
at the
To Set the System Language
MSO7000/DS7000 series oscilloscope supports multiple system languages. You
can press Utility Language to set the system language. You can also refer
to descriptions in "To Use the Built-in Help System" to open the "Help"
menu, and then press Language to set the system language. You can also
enable the touch screen to perform the above menu operation.
To Connect the Probe
RIGOL provides the passive probe, the active probe, and the logic probe for
MSO7000/DS7000 series. For specific probe models, please refer to
MSO7000/DS7000 Series Datasheet
probes, please refer to the specified Probe User’s Guide.
Connect the passive probe:
1. Connect the BNC terminal of the probe to an analog channel input
terminal of the oscilloscope on the front panel, as shown in Figure 6.
2. Connect the ground alligator clip or spring of the probe to the circuit
ground terminal, and then connect the probe tip to the circuit point to be
tested.
. For detailed technical information of the
10MSO7000/DS7000 Quick Guide
RIGOL
English
Figure 6 To Connect the Passive Probe
After you connect the passive probe, check the probe function and probe
compensation adju s t m en t before making measurem en ts. For detailed
procedures, refer to "Function Inspection" in this manual and the "Probe Compensation" section introduced in
MSO7000/DS7000 User's Guide
.
Connect the active probe:
Take RP7080 (use the active differential probe head) as an example.
1. Connect the pro be head to th e preamp of the acti ve probe, as sh own in
Figure 7.
Figure 7 To Connect the P r o be Head to the Pr eamp of the Active Probe
2. Connect the other end of the preamp to an analog channel input terminal
of the oscilloscope on the front panel, as shown in Figure 8. Note that you
should push the probe to the end to ensure that it is tightly connected.
MSO7000/DS7000 Quick Guide 11
RIGOL
English
Figure 8 T o Connect the Active Probe
3. Use the probe auxiliary equipment to connect the probe front to the circuit
under test. For d etails about the pro bes, refer to
Probe User's Guide
.
RP7000 Series Active
After connecting the active probe, you can perform probe calibration and offset
voltage adjustment if necessary. For detailed procedures, refer to the
descriptions of the active probe in user's guide.
Connect the logic probe:
1. Connect the single-wire terminal of the logic probe to the digital channel
input terminal on the front panel of the oscilloscope in the correct direction,
as shown in Figure 9.
2. Connect the other terminal of the logic probe to the signal terminal under
test. RIGOL's MSO7000 series has a standard configuration of a logic
probe RPL2316. To apply to different application scenarios, RPL2316
provides three connection methods to connect the signal under test. For
details, refer to
RPL2316 Logic Probe User’s Guide
.
Figure 9 To Connect the Logic Probe
12MSO7000/DS7000 Quick Guide
RIGOL
English
Note:
Before you connect the logic probe to the signal under test, connect logic
probe's adapter provided in the accessories to its corresponding channel
group.
The digital channel input terminal does not support hot plugging. Please
do not insert or pull out the logic probe when the instrument is in
power-on state.
Function Inspection
1. Press Default on the front panel, and then a pro mpt message " Restore
default?" is displ ayed. Pre ss OK to restore the oscilloscope to its factory
default settings.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in Figure 10 below.
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and
the "Compensatio n Signal Output Terminal" of the probe, as shown in
Figure 10.
Compensation Signal
Output T e rminal
Ground Terminal
Figure 10 T o Use the Compensation Signal
4. Set the probe attenuation to 10X, and then press AUTO.
5. Observe the waveform on th e display. In normal conditio n, the square
waveform as shown in Figure 11 should be displayed.
MSO7000/DS7000 Quick Guide 13
RIGOL
English
WARNING
To avoid electric shock when using the probe, please make sure
that the insulated wire of the probe is in good condition. Do not
touch the metallic part of the probe when the probe is connected
to high voltage source.
Figure 11 Square Waveform Signal
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform
"Probe Compensation" introduced in
MSO7000/DS7000 User's Guide
Touch Screen Operation
MSO7000/DS7000 series provides a 10.1-inch super large capacitive touch
screen, which supports multi-touch and gesture operation. It has strong
waveform display capability and excellent user experience. It features great
convenience, high flexibility, and great sensitivity. The actions supported by the
touch screen controls include tapping, pinching&stretching, dragging as well as
rectangle drawing.
Tap
Use one finger to touch the symbol or characters on the screen slightly, as
shown in Figure 12. The actions supported by the tap action include:
Tap and operate the menus on the menu.
Tap the function navigation icon at the lower-left corner of the touch
screen to enable the function navigation.
Tap the displayed numeric keypad to set the parameters.
14MSO7000/DS7000 Quick Guide
.
RIGOL
English
Tap the virtual keypad to set the label name and the filename.
Tap close button at the upper-right corner of the message box to close it.
Tap and operate the windows on the screen.
Figure 12 Tap Gesture
Pinch&Stretch
Pinch or stretch two points on the screen with two fingers to zoom in or out the
waveform. To zoom in the waveform, first pinch the two fingers and then
stretch the fingers; to zoom out the waveform, first stretch the two fingers, and
then pinch the fingers together, as shown in Figure 13. The actions supported
by the pinch action include:
Pinching&stretching in the horizontal dire cti on can adj ust the horizontal
time base of the waveform.
Pinching&stretching in the vertical direction can adjust the vertical scale of
the waveform.
Figure 13 Pinch&Stretch Gesture
Drag
Use one finger to select the object, and then drag the object to a destination
place, as shown in Figure 14. The actions supported by the drag action include:
Drag the waveform to change its position or offset.
Drag the window controls to change the positions of t he contr ol s (e.g.
numeric keypad).
Drag the marker to change the position of the marker.
MSO7000/DS7000 Quick Guide 15
RIGOL
English
Figure 14 Drag Gesture
Rectangle Drawing
Enable the function navigation, and then tap the "Draw rect" icon to switch to
the rectangle drawing mode. Drag to draw a rectangle on the screen, as shown
in Figure 15. The actions supported by the rectangle drawing include:
Select "Trigger zone A":
Draw the region for Trigger zone A;
Open Trigger zone A;
Open the "Zone trig" menu.
Select "Trigger zone B":
Draw the area for Trigger zone B;
Open Trigger zone B;
Open the "Zone trig" menu.
Select "Histogram":
Draw the region for the histogram;
Open the "Histogram" menu.
Select "Horizontal zoom": expands the waveforms in the horizontal
direction.
Select "Vertical zoom": expands the waveforms in the vertical direction.
Select "waveform zoom": expands the waveforms both in the horizontal
and vertical direction.
Figure 15 Rectangle DrawingGesture
16MSO7000/DS7000 Quick Guide
RIGOL
English
Help Options Help Display Area
To Use the Built-in Help System
The help system of this oscilloscope provides instructions for all the function
keys on the front panel and their corresponding menu keys. The steps for
opening the built-in help system are as follows:
1. Using the touch screen function, tap the function navigation icon at
the lower-left corner of the touch screen to enable the function
navigation.
2. Tap the "Help" icon, and then the "Help" menu is displayed on the screen.
3. Press Help on the front panel or touch the "Help" icon, and then the help
information is displayed on the screen, as shown in Figure 16. The help
interface mainly consists of two sections. The left section lists "Help
Options", and the right section is the "Help Display Area".
After opening the help interface, you can get its help information in the "Help
Display Area" through the following three methods:
Method 1:
For the keys, you can directly press the front-panel keys (except the Power key
, the Menu Off key and Back key) to obtain the help information. For the
buttons, you can rotate the knob or press down the knob to get the help
information (except the multifunction knob ). As the function of rotating the
knob is different from that of pressing down the knob, these two operations will
MSO7000/DS7000 Quick Guide 17
Figure 16 Help Information
RIGOL
English
Tip
Help information for other keys and buttons:
Power key : powers on/off the instrument.
Menu Off key: displays or hides menus. By default, menus are
displayed. If menus are hidden, pressing this key can display
the menus again.
Back key: returns to the previous menu or the last set function
menu.
Multifunction knob : In non-menu-operation mode, rotate
this knob to adjust the brightness of waveform display. When
you operate on the menu, this knob is illuminated. For the
menu item that has multiple parameters under it, when you
press the menu softkey, rotate the knob to select the parameter
item, then press down the knob to select it (sometimes, the
specified parameter item can be selected by rotating the knob).
It can also be used to modify parameters, input a filename, etc.
If the menu item is grayed out, you cannot press the corresponding
front-panel menu key to obtain the help information. What you can
do is only to follow the above Method 2 or 3 to get the help
information.
give you different help information.
Method 2:
In the "Help Options", rotate the multifunction knob to switch among the
help options to get the corresponding help information.
Method 3:
Enable the touch screen and tap the desired help option to get the
corresponding help information.
18MSO7000/DS7000 Quick Guide
RIGOL
English
Parameter Input Field
Delete Key
Max. Value Key
Min. Value Key
Default Key
Clear Key
Parameter Setting Method
For the MSO7000/DS7000 series, you can use the multifunction knob or enable
the touch screen to set the parameters of MSO7000/DS7000 series. The
common parameter setting methods are as follows:
Method 1:
For the parameters with the sign , rotate the multifunction knob on the
front panel directly to select the parameter item or modify the parameter value.
Method 2:
For the parameters with the sign , rotate the multifunction knob on the
front panel and press it down to select the parameter item or modify the
parameter value.
Method 3:
For the parameters with displayed on the menu, rotate the multifunction
knob on the front panel directly to set the parameter, or press down the
multifunction knob or the specified menu softkey, and then the numeric
keypad is shown in Figure 17 below.
In the numeric keypad, rotate the multifunction knob to select the
parameter, then press down the knob to select it. Then input the value or the
unit. You can also enable the touch screen and tap the value or unit in the
numeric keypad to input them. After you input all the values and select the
desired units, the numeric keypad is turned off automatically. This indicates
that you have completed parameter setting. Besides, after you have input the
MSO7000/DS7000 Quick Guide 19
Figure 17 Numeric Keypad
RIGOL
English
Tip
relevant chapters of
MSO7000/DS7000 User's Guide
.
values, you can also press OK directly to close the numeric keypad. At this time,
the unit of the parameter is the default unit. In the numeric keypad, you can
also perform the following operations:
Delete the parameter value that has been input.
Set the parameter to a maximum or minimum value (sometimes, the
maximum or minimum value are the specified one for the current state).
Set the parameter to a default value.
Clear the parameter input field.
Method 4:
For the parameters without the above signs, press the desired menu softkey to
switch between the parameter items. This method is applicable to the
parameters with only two available options.
The above method is commonly used for the parameter settings of the
oscilloscope. For other methods of parameter settings, refer to details in
20MSO7000/DS7000 Quick Guide
RIGOL
English
WARNING
fuse to be used conforms to the fuse rating.
Fuse Holder
Fuse
Fuse Replacement
If you need to replace the fuse, use only the specified fuse (AC 250 V, T3.15 A;
5.2 mm×20 mm) and perform the following operations, as shown in Figure 18:
1.Turn off the instrument, cut off the power, and remove the power cord.
2.Use a small straight slotted screwdriver to pry out the fuse holder .
3.Take out the fuse.
4.Replace the old fuse with a specified fuse.
5.Install the fuse holder.
Figure 18 Fuse Replacement
To avoid electric shock, please ensure that the instrument has
been turned off, the power source has been cut off, and the
MSO7000/DS7000 Quick Guide 21
RIGOL
English
Remote Control
MSO7000/DS7000 series digital oscilloscope can be connected to the PC via the
USB, LAN, or GPIB interface to set up communication and realize remote control
through the PC. The remote control can be realized by using SCPI (Standard
Commands for Programmable Instruments) commands. MSO7000/DS7000
series digital oscilloscope supports two ways of remote control: user-defined
programming and PC software (e.g. RIGOL Ultra Sigma).
More Product Information
1. Obtain the device information
Press Utility System About to obtain the information of the
instrument, such as the manufacturer, model, serial number, hardware
version number, etc. You can also refer to descriptions in "To Use the Built-in Help System" to open the "Help" menu, and then press About
to obtain the device information. You can also enable the touch screen to
perform the above menu operation.
2. View the option information and the option installation
The instrument is installed with the trial versions of the options before
leaving factory. When you power on the instrument for the first time, the
trial time is about 2,160 minutes. Refer to descriptions in "
Built-in Help System" to open the "Help" menu, and then press Option
list to view the options currently installed on the oscilloscope and their
information. Press Option setup to install the option. For detailed setting
methods, refer to descriptions in
also enable the touch screen to perform the above menu operation.
For more information about this instrument, refer to the relevant manuals by
logging in to the official website of RIGOL (www.rigol.com) to download them.
MSO7000/DS7000 User's Guide
the operation methods, remote control methods, possible failures and solutions
in using the instrument, the technical specifications, and order information.
MSO7000/DS7000 Programming Guide
commands and programming examples of the instrument.