Rigol MSO4000, DS4000 User Manual

Page 1
RIGOL
User’s Guide
MSO4000/DS4000 Series Digital Oscilloscope
Oct. 2013
RIGOL Technologies, Inc.
Page 2
Page 3
RIGOL
I

Guaranty and Declaration

Copyright
© 2013 RIGOL Technologies, Inc. All Right s Re s erved.
Trademark Information
RIGOL is a registered trademark of RIGOL Technologies, Inc.
Publication Number
UGA15103-1110
Notices
RIGOL products are protected by patent law in and outside of P.R.C. RIGOL reserves the right to modify or change parts of or all the specifications
and pricing policies at company’s sole decision.
Information in this publication replaces all previously corresponding material. RIGOL shall not be liable for losses caused by either incidental o r consequenti al
in connection with the fu r nis hing , use or perfo rman ce of this manual as well as any information contained.
Any part of this document is forbidden to be copied or phot ocopied or
rearranged without prior written approval of RIGOL.
Product Certification
RIGOL guar antees this pr oduct confo rms to the national and industrial standards in China as well as the ISO9001: 2008 standard and the ISO14001: 2004 standar d. Other international standard conformance certifications are in progress.
Contact Us
If you have any problem or requirement when using our products or this manual, please contact RIGOL. E-mail: service@rigol.com Websites: www.rigol.com
Page 4
RIGOL
II

Safety Requirement

General Safety Summary

Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injuries or damages to the instrument and any product connecte d to it. To prevent potential haza rds, please use the instrument as specified in this manual.
Use Proper Power Cord.
Only use the power cord designed for the instrument and authorized for use within the local country where the product is being used.
Ground The Instrument.
The instrument is grounded through the Protective Earth lead of the power cord. To avoid ele ctric sh ock, it is e ssential to conne ct the ear th terminal of power co rd to the Protective Earth terminal before any inputs or outputs.
Connect The Probe Properly.
Do not connect the ground lead to high voltage.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratin gs and mark ers on the inst rument and check your manual for more information about ratings before connecting.
Use Proper Overvoltage Protection.
Make sure that n o ov erv oltage (such as that produced by a thunderstorm) can reach the product, or else the operator might expose to danger of electrical shock.
Do Not Operate Withou t Co v e rs.
Do not operate the instrument with covers or panels removed.
Change The Power Fuse.
If the power fuse needs to be changed, please return the instrument back to our factory and the RIGOL authorized operator will change it for you.
MSO4000/DS4000 User’s Guide
Page 5
RIGOL
III
Avoid Circuit or Wire Exposure.
Do not touch exposed junctions and components when the unit is powered.
Do Not Operate With Suspected Failures.
If you suspect damage occurs to the instrument, have it inspected by qualified service personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
Keep Well Ventilation.
Inadequate ventilation may cause increased temperature or damage to the device. So please keep well ventilated and inspect the intake and fan regularly.
Do Not Operate In Wet Conditions.
In order to av oid short circuiting t o the interi or of the device or electric shock, please do not operate in a humid environment.
Do Not Operate in an Explosive Atmosphere.
In order t o a voi d da mage s to t he de vi ce or pers ona l injury, it is important to operate the device away fr om an exp losive a tmosphere.
Keep Product Surfaces Clean and Dry.
To avoid the influence of dust and/or moisture in air, please keep the s urface of the device clean and dry.
Electrostatic Prevention.
Minimize electrostatic discharge to avoid damage induced by static discharges. Always ground both the internal and external condu ctors of cables to neutralize static before making connections.
Handling Safety.
Please handle with care during transportation to avoid damage to buttons, knob interfaces and other parts on the panels.
MSO4000/DS4000 User’s Guide
Page 6
RIGOL
IV
CAUTION
indicates a potential dama ge to the inst rument or ot her property mi ght
Hazardous
Protective
Chassis
Test

Safety Terms and Symbols

Terms in this Ma n ual. The following terms may appear in this manual:
WARNING
Warning statements indicate conditions or practices that could result in injury or loss of life.
Caution statements indicate conditions or practices that could result in damage to this product o r other property.
Terms on the Product. These terms may appear on the product:
DANGER WARNING indicates an injury or hazard may occur. CAUTION
Symbols on the Product. These symbols may appear on the product:
indicates an injury or hazard may immediately happen.
occur.
Voltage
Safety Warning
Earth Terminal
Ground
MSO4000/DS4000 User’s Guide
Ground
Page 7
RIGOL
V

Allgemeine Sicherheits Informationen

Überprüfen Sie diefolgenden Sicherheitshinweise sorgfältigumPersonenschädenoderSchäden am Gerätundan damit verbundenen weiteren Gerätenzu vermeiden.Zur Vermeidung vonGefahren, nutzen Sie bitte das Gerät nur so, wiein diesem Handbuchangegeben.
Um Feuer oder Verletzungen zu vermeiden, verwenden Sie ein ordnungsgemäßes Netzkabel.
Verwenden Sie für dieses Gerät nur das für ihr Land zugelass ene u nd genehm igte Netzkabel.
Erden des Gerätes.
Das Gerät ist durch den Schutzleiter im Netzkabel geerdet. Um Gefahren durch elektrischen Schlag zu vermeiden , ist es unerlässlich, die Erdung durchzuführen. Erst dann dürfen weitere Ein- oder Aus gä nge verbunden werden.
Anschluss einesTastkopfes.
Die Erdungsklemmen der Sonden sindauf dem gleichen Spannungspegel des Instruments geerdet. SchließenSie die Erdungsklemmen an keine hohe Spannung an.
Beachten Sie alle Anschlüsse.
Zur Vermeidung von Feuer oder Stromschlag, beachten Sie alle Bemerkungen und Markierungen auf dem Instrument. Bef olgen Sie die Bedienun gsanleitung für weitere Informationen, bevor Sie weitere Anschlüsse an das Instrument legen.
Verwenden Sie einen geeigneten Überspannungsschutz
Stellen Sie sicher, daß keinerlei Überspannung (wie z.B. durch Gewitter verursacht) das Gerät erreichen kann. Andernfallsbestehtfür den Anwender die GefahreinesStromschlages.
Nicht ohne Abdeckung einschalten.
Betreiben Sie das Gerät nicht mit entfernten Gehäuse-Abdeckungen.
Betreiben Sie das Gerät nicht geöffnet
Der Betrieb mit offenen oder entfernten Gehäuseteilen ist nicht zulässig. Nichts in entsprechende Öffnungen stecken (Lüfter z.B.)
Passende Sicherung verwenden
Setzen Sie nur die spezifikationsgemäßen Sicherungen ein.
Vermeiden Sie ungeschützte Verbindungen
Berühren Sie keine unisolierten Verbindungen oder Baugruppen, während das Gerät
MSO4000/DS4000 User’s Guide
Page 8
RIGOL
VI
in Betrieb ist.
Betreiben Sie das Gerät nicht i m Fehlerfall
Wenn Sie am Gerät einen Defekt vermuten, sorgen Sie dafür, bevor Sie das Gerät wieder betreiben, dass eine Untersuchu ng durch qual ifiziertes Kundendienstpersonal durchgeführt wird.Jedwede Wartung, Einstellarbeiten oder Austausch von Teilen am Gerät, sowie am Zubehör dürfen nur von RIGOL autorisiertem Personal durchgeführt werden.
Belüftung sicherstellen
Unzureichende Belüftung kann zu Temperaturanstiegen und somit zu thermischen Schäden am Gerät führen. Stellen Sie deswegen die Belüftung sicher und kontrollieren regelmäßig Lüfter und Belüftungsöffnungen.
Nicht in feuc h te r Um g ebung betreiben
Zur Vermeidun g von Kurzschluß im Geräteinne ren und Stromschlag betreiben Sie das Gerät bitte niemals in feuchter Umgebung.
Nicht in explosiver Atmosphäre betreiben
Zur Ve rm e idung von Person e n - u nd Sachs chä den ist es unumgä ngli ch, das Ger ät ausschließlich fernab jedweder explosiven At mosphäre zu betreiben.
Geräteoberflächen sauber und trocken halten
Um den Einfluß von Staub und Feuchtigkeit aus der Luft auszuschließen, halten Sie bitte die Geräteoberflächen sauber und trocken.
Schutz gegen elektrostatische Entladung (ESD)
Sorgen Sie für eine elektrostatisch geschützte Umgebung, um somit Schäden und Funktionsstörungen durch ESD zu vermeiden. Erden Sie vor dem Anschluß immer Innen- und Außenleiter der V erbindungsleitung, um st atische Aufladung zu entladen.
Sicherer Transport
Transportieren Sie das Gerät sorgfältig (Verpackung!), um Schäden an Bedienelementen, Anschlüssen und anderen Teilen zu vermeiden.
MSO4000/DS4000 User’s Guide
Page 9
RIGOL
VII
WARNING
Schäden oder den Tod von Personen zur Folge haben können.
CAUTION
Schäden am Gerät hervorrufen können.
DANGER
weist auf eine Verletzung ode r Gefäh r dun g hin, die s ofort geschehen kann.
WARNING
weist auf eine Verletzung oder Gefäh rdung hin, die möglicherweise nicht sofort geschehen.
CAUTION
bedeutet, dass eine mögliche Beschädigung des Instruments oder anderer Gegenstände auftreten kann.

Sicherheits Begriffe und Symbole

Begriffe in diesem Guide. Diese Begriffe können in diesem Handbuch auftauchen:
Die Kennzeichnung WARNING beschreibt Gefahrenquellen die leibliche
Die Kennzeichnung Caution (Vorsicht) beschreibt Gefahrenquellen die
Begriffe auf dem Produkt. Diese Bedingungen können auf dem Produkt erscheinen:
Symbole auf dem Produkt. Diese Symbole können auf dem Produkt erscheinen:
GefährlicheS pannung
Sicherheits­Hinweis
Schutz-erde Gehäusemasse Erde
MSO4000/DS4000 User’s Guide
Page 10
RIGOL
VIII
WARNING

Measurement Category

Measurement Category
The MSO4000/DS4000 series digital oscilloscopes can make measurements in Measurement Category I.
This oscilloscope can only be used for measurements within its specified measurement categories.
Measurement Category Definitions
Measurement category I is for measurements performed on circuits not directly connected to M AINS. Examples are me a s urements on circuits no t de rived from MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable. For that reason, the transient limit capability of the equipment is made known to the user.
Measurement category II is for measurements performed on circuits directly connected to a low voltage installation. Examples are measurements on household appliances, portable tools and similar equipment.
Measuremen t c ategory III is for measurements performed in building installations. Examples are measurements on distribution boards, circuit-breakers, w iring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, equipment for industrial use and Stationary motors with permanent connection to a fixed installation.
Measurement category IV is for measurements performed at the source of a low-voltage installation. Examples are electricity meters and measurements on primary overcurr e n t prot ec ti on devices and ripple control units.
MSO4000/DS4000 User’s Guide
Page 11
RIGOL
IX
WARNING

Ventilation Requirement

This oscilloscope uses f ans to force cooling. Please make su re t hat the ai r int ak e a nd exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
Inadequate ventilation may cause a temperature increase which can damage the instrument. So please keep the instrument well ventilated during operation and inspect the intake and fan regularly.
MSO4000/DS4000 User’s Guide
Page 12
RIGOL
X
WARNING

Working Environment

Temperature
Operating: 0 to +50 Non-operating: -40 to +70
Humidity
0 to +30: ≤95 relative humidity +30 to +40: 75 relative humidity +40 to +50: 45 relative humidity
WARNING
To avoid short circuits or electric shock, please do not operate in humid environment.
Altitude
Operating: less than 3 km Non-operating: less than 15 km
Installation (overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
Make sure that no overvoltage (such as that caused by a thunderstorm) can reach the product, or else the operator might be exposed to the danger of electric shock.
Installation (overvoltage) Categor y Definitions
Installation (overvoltage) category I refers to signal level which is applicable to equipment measurement terminals connected to the source circuit. In these terminals, precautions are done to limit the transient voltage to the corresponding low level. Installation (overvoltage) category II refers to the local power distribution level which is applicable to equipment connected to the AC line (AC power).
MSO4000/DS4000 User’s Guide
Page 13
RIGOL
XI
Pollution Degree
Degree 2
Pollution Degree Definitions
Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. For example: a clean room or air-conditioned office environment. Pollution degree 2: Normally only dry, non-conductive pollution occurs. An occasional and temporary conductive environment may caused by condensation. Fo r example, a general indoor environment. Pollution degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due to condensation. For example, a Sheltered outdoor environment. Pollution degree 4: Pollution that generates persistent conductivity through conductive dust, rain, or snow. Fo r example, outdoor locations.
Safety Class
Class 1 – Grounded Product
MSO4000/DS4000 User’s Guide
Page 14
RIGOL
XII

General Care and Cleaning

General Care:
Do not store or leave the instrument at places where the instrument will be exposed to direct sunlight for long periods of time.
Cleaning:
Clean the instrument regularly according to its operating conditions. To clean the exterior surface:
1. Disconnect the instrument from all power sources.
2. Clean the loose dust on the outside of t he instrument with a lint - free cloth (with
mild detergent or water). When cleaning the LCD, take care to avoid scratching it.
CAUTION
To avoid damage to th e instrument, do no t e x pose to corrosive liquids.
WARNING
To avoid injury resulting from short circuit, make sure the instrument is completely dry before reconnecting it to a pow e r s ou r ce.
MSO4000/DS4000 User’s Guide
Page 15
RIGOL
XIII

Environmental Consideratio ns

The following symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002/96/E C on waste electrical and electronic equipment (WEEE).
Product End-of-Life Handling
The equipment may contain substances that could b e harmf ul t o the envi ronm ent or human health. In order to avoid release of such substances into the environment and harm to human health, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately. Please contact your local authorities for disposal or recycling information.
MSO4000/DS4000 User’s Guide
Page 16
RIGOL
XIV

MSO4000/DS4000 Series Overview

MSO4000/DS4000 is a multifunctional and high performance digital oscilloscope developed using of the RIGOL Ultra Vision chipset. MSO4000 series is a high-performance mixed-signal oscilloscope which is optimized for the debugging in embedded design and test areas that require simultaneous measurement of both analog and digital signals. The MSO4000 provides superb specifications, convenient functions, and easy-to-use design which help users fulfill their tasks (such as measu rement and remote control) more quickly.
Main feature s:
500 MHz, 350 MHz, 200 MHz and 100 MHz bandwidth 2 +16 chann el or 4 +16 channel MSO4000 series digital oscilloscope Dual-channel or 4-channel model DS 4000 series digital oscilloscope Analog channel sample rates up to 4 GSa/s, 140 Mpts maximum memory depth
(standard). Digital channel sample rates up to 1 GSa/s, 28 Mpts maximum memory depth (per channel)
110,000 wfms/s (digital channels are turned off) maximum waveform refresh
rate or 85,000 wfms/s (digital channels are turned on, dots displa y) maximum waveform refresh rate
Ultra Vision technology 9.0 inch, WVGA (800*480) 160,000 color TFT LCD , wit h ultra-wide screen, vivid
picture, low power consumption and long service life
Automatic probe identification Adjustable brightness of analog channel waveform Auto setting of waveform display (AUTO) Various trigger functions including multiple protocol triggers and supporting
digital channels as trigger source
Support standard parallel decoding, multiple serial decoding options and mixed
decoding for analog and digital channels
Auto measurements of 24 waveform parameters and measurement functions
with statistics
Real-time waveform recording, waveform playback, record open (constant on)
and waveform analysis, supporting digital channels recor ding and play back
Precise delayed sweep function Built-in FFT function
MSO4000/DS4000 User’s Guide
Page 17
RIGOL
XV
Built-in logic analyzer function Pass/Fail test function Multiple waveform math o peration functions Standard configuration i n t e rfaces: USB Device, dual US B Host, LAN and GPIB
(optional)
Support USB storage device and printer Conforms to LXI-C instrument standards which enable quick, economic and
efficient creation and reconfig uratio n of test system
Supports remote command control Embedded help enables easier information access Supports multiple languages and Chinese/English input Supports one-key measurement, storage and print Convenient digital channel grouping and operating functions
MSO4000/DS4000 User’s Guide
Page 18
RIGOL
XVI

Document Overview

1
Quick Star t
Provides information about preparations before using the instrument and a brief introduction of the instrument.
To Set the Vertical System
2
Introduces the functions of the vertical system of the oscilloscope.
To Set the Horizontal System
3
Introduces the functions of the horizontal system of the oscilloscope.
To Set the Sample System
4
Introduces the functions of the sample system of the oscilloscope.
To Trigger the Oscilloscope
5
Introduces the trigger mode, trigger coupling, trigger holdoff, external trigger and various trigger types of the oscilloscope.
To Make Measurements
6
Introduces how to make math oper a t i on, cursor and auto measurements.
Logic Analyzer
7
Introduces how to use the logic analyzer function.
Protocol Decoding
8
Introduces how to decode the input signal using those common protocols.
Reference Waveform
9
Introduces how to compare the input waveform with the reference waveform.
Pass/Fail Test
10
Introduces how to monitor the input signal using the Pass/Fail test.
Waveform Record
11
Introduces how to analyze the input signal using waveform re co rding.
Display Control
12
Introduces how to control the display of the oscilloscope.
MSO4000/DS4000 User’s Guide
Page 19
RIGOL
XVII
Logo
Knob
Logo
Knob
Multi-function Knob
Vertical Scale Knob
Navigation POSITION
Vertical Position SCALE
Horizontal
Trigger Level POSITION
Horizontal
13 Store and Recall
Introduces how to store and recall the measurement result and the setting of the oscilloscope.
System Function Setting
14
Introduces ho w to set the remote interface and system-related functions.
Remote Control
15
Introduces how to control the oscilloscope remotely.
Troubleshooting
16
Introduces how to deal with common failures of the oscilloscope.
Specifications
17
Lists the specifications and general specifications of the oscilloscope.
Appendix
18
Provides common information such as options and accessories.
Format Conventions in this Manual:
Front panel key: denoted by the format of “Text Box + Button Name (Bold)”. For example, Storage. Menu softkey: denoted by the format of “Character Shading + Menu Word (Bold)”. For example, Storage. Operation steps: denoted by the arrow “”, for example, Storage Storage. Link: denoted by the format of “Blue Characters (Bold + Underline d)”. For example,
Stor e and Recall.
Knob:
MSO4000/DS4000 User’s Guide
HORIZONTAL
HORIZONTAL
Knob
Scale Knob
Position Knob
VERTICAL SCALE VERTICAL
TRIGGER LEVEL
——
Knob
Knob
——
Page 20
RIGOL
XVIII
Content Con v entions in this Manual:
This manual takes the MSO4054 for all examples and the descriptions here contains the functions and performance of other models. The MSO4000 and DS4000 series includes the following models:
Model
MSO4054 500MHz 4 250MHz 16 MSO4052 500MHz 2 250MHz 16 MSO4034 350MHz 4 250MHz 16 MSO4032 MSO4024 MSO4022 MSO4014 100MHz 4 250MHz 16 MSO4012 100MHz 2 250MHz 16 DS4054 500MHz 4 DS4052 500MHz 2 DS4034 350MHz 4 DS4032 350MHz 2 DS4024 200MHz 4 DS4022 200MHz DS4014 100MHz DS4012 100MHz
Analog Bandwidth
350MHz 200MHz 200MHz
Analog Channels
2 4 2
2 4 2
Digital Channel Bandwidth
250MHz 250MHz 250MHz
—— —— —— —— —— —— —— —— —— ——
—— —— ——
Digital Channels
16 16 16
—— —— ——
Note: The content related to the logic analyzer (digital channels) in this document only applies to MSO4000 mixed signal digital oscilloscope.
MSO4000/DS4000 User’s Guide
Page 21
RIGOL
XIX
Contents
Guaranty and Declaration ......................................................................... I
Safety Requirement ................................................................................ II
General Safety Summary ........................................................................... II
Safety Terms and Symbols ....................................................................... IV
Allgemeine Sicherheits Informationen ......................................................... V
Sicherheits Begriffe und Symbole ............................................................. VII
Measurement Category .......................................................................... VIII
Ventilation Requirement ........................................................................... IX
Working Enviro nment ............................................................................... X
General Care and Cleaning ...................................................................... XII
Environmental Considerations ................................................................. XIII
MSO4000/DS4000 Seri es Over vie w ..................................................... XIV
Document Overview ............................................................................. XVI
1 Quick Start ......................................................................................... 1-1
General Inspection ................................................................................ 1-2
Appearance and Dim ensions ................................................................... 1-3
To Prepare the Oscilloscope for Use ......................................................... 1-4
To Remove the Cover ...................................................................... 1-4
To Adjust the Supporting Legs .......................................................... 1-5
To Connect to AC Power Supply ........................................................ 1-6
Power-on Inspection ....................................................................... 1-7
To Connect the Probe ...................................................................... 1-8
Function Inspection ......................................................................... 1-9
Probe Compensation ...................................................................... 1-11
To Connect the Logic Probe ............................................................ 1-12
Front Panel Overview ............................................................................ 1-13
Rear Panel Overview ............................................................................. 1-14
Front Panel Function Overview............................................................... 1-16
VERTICAL ..................................................................................... 1-16
HORIZONTAL ................................................................................ 1-17
TRIGGER ...................................................................................... 1-18
CLEAR .......................................................................................... 1-18
MSO4000/DS4000 User’s Guide
Page 22
RIGOL
XX
AUTO ........................................................................................... 1-19
RUN/STOP .................................................................................... 1-19
SINGLE ......................................................................................... 1-19
Multi-function Knob ........................................................................ 1-20
Navigation Knob ............................................................................ 1-20
Default ......................................................................................... 1-21
Print ............................................................................................. 1-21
Function Keys ................................................................................ 1-21
Record .......................................................................................... 1-22
LOGIC ANALYZER .......................................................................... 1-23
User Interface ...................................................................................... 1-24
To Use the Security Lock ....................................................................... 1-29
To Use the Built-in Help System ............................................................. 1-30
2 To Set the Vertical System ................................................................. 2-1
To Enable the Analog Channel ................................................................. 2-2
Channel Coupling ................................................................................... 2-3
Bandwidth Limit ..................................................................................... 2-4
Probe .................................................................................................... 2-5
Input Impe da nce ................................................................................... 2-7
Waveform Invert .................................................................................... 2-7
Vertical Scale ......................................................................................... 2-8
Vertical Expansion .................................................................................. 2-9
Amplitude Unit ....................................................................................... 2-9
Channel Label ...................................................................................... 2-10
Delay Calibration .................................................................................. 2-11
3 To Set the Horizontal System ............................................................ 3-1
Delayed Sweep ...................................................................................... 3-2
Time Base Mode .................................................................................... 3-4
Y-T Mode ........................................................................................ 3-4
X-Y Mode ........................................................................................ 3-5
Roll Mode ....................................................................................... 3-8
Horizontal Scale ..................................................................................... 3-9
Horizontal Reference ............................................................................ 3-10
4 To Set the Sample System ................................................................. 4-1
Acquisition Mode .................................................................................... 4-2
Normal ........................................................................................... 4-2
MSO4000/DS4000 User’s Guide
Page 23
RIGOL
XXI
Average ......................................................................................... 4-2
Peak Detect .................................................................................... 4-4
High Resolution .............................................................................. 4-4
Sample Mode ........................................................................................ 4-5
Sample Rate ......................................................................................... 4-6
LA Sample Rate ..................................................................................... 4-8
Memory Depth ...................................................................................... 4-9
Antialiasing .......................................................................................... 4-10
5 To Trigger the Oscilloscope ................................................................ 5-1
Trigger Sou rce ...................................................................................... 5-2
Trigger Mode ........................................................................................ 5-3
Trigger Coupling .................................................................................... 5-5
Trigger Holdoff ...................................................................................... 5-6
Noise Rejection ..................................................................................... 5-7
Trigger Type ......................................................................................... 5-8
Edge Trigger ................................................................................... 5-9
Pulse Trigger ................................................................................. 5-11
Runt Trigger .................................................................................. 5-13
Nth Edge Trigger............................................................................ 5-15
Slope Trigger ................................................................................. 5-17
Video Trigger ................................................................................. 5-20
Pattern Trigger .............................................................................. 5-22
RS232 Trigger ................................................................................ 5-24
I2C Trigger .................................................................................... 5-26
SPI Trigger .................................................................................... 5-29
CAN Trigger ................................................................................... 5-31
FlexRay Trigger .............................................................................. 5-34
USB Trigger ................................................................................... 5-36
Trigger Output Connector ...................................................................... 5-38
6 To Mak e Measurements ..................................................................... 6-1
Math Operation ..................................................................................... 6-2
Addition ......................................................................................... 6-2
Substraction ................................................................................... 6-3
Multiplication .................................................................................. 6-3
Division .......................................................................................... 6-4
FFT ............................................................................................... 6-5
MSO4000/DS4000 User’s Guide
Page 24
RIGOL
XXII
Logic Operation ............................................................................... 6-8
Advanced Operation ....................................................................... 6-10
Auto Measurement ............................................................................... 6-13
Quick Measurement after AUTO ...................................................... 6-13
One-key Measurement of 24 Parameters .......................................... 6-14
Frequency Counter Measurement .................................................... 6-19
Measurement Setting ..................................................................... 6-20
To Clear the Measurement .............................................................. 6-22
All Measurement ............................................................................ 6-23
Statistic Function ........................................................................... 6-24
Measurement History ..................................................................... 6-25
Cursor Measurement ............................................................................ 6-26
Manual Mode ................................................................................ 6-27
Track Mode ................................................................................... 6-31
Auto Mode .................................................................................... 6-34
X-Y Mode ...................................................................................... 6-35
7 Logic Analyzer ................................................................................... 7-1
To Select the Logic Channel .................................................................... 7-2
To Turn on/off the Logic Channel ............................................................. 7-2
Group Setting ........................................................................................ 7-3
To Set the Waveform Size ....................................................................... 7-3
Automatic Arrangement Setting ............................................................... 7-4
To Set the Threshold .............................................................................. 7-4
To Use the Digital Bus ............................................................................ 7-5
To Set the Label ..................................................................................... 7-6
Delay Calibration .................................................................................... 7-6
8 Protocol Decoding ............................................................................. 8-1
Parallel Decoding ................................................................................... 8-2
RS232 Decodi n g (Option) ....................................................................... 8-5
I2C Decoding (Option) .......................................................................... 8-10
SPI Decoding (Option) .......................................................................... 8-13
CAN Decoding (Option) ......................................................................... 8-16
FlexRay Decoding (Option) .................................................................... 8-19
9 Reference Waveform ......................................................................... 9-1
To Enable REF Function .......................................................................... 9-2
To Set the Color ..................................................................................... 9-3
MSO4000/DS4000 User’s Guide
Page 25
RIGOL
XXIII
To Select REF Source ............................................................................. 9-3
To Save to Internal Memory ................................................................... 9-3
To Adjust REF Waveform Display ............................................................. 9-4
To Export to Internal or External Memory ................................................ 9-4
To Import fro m In t e rnal or External Memory ............................................ 9-4
10 Pass/Fail Test ................................................................................ 10-1
To Enable Pass/Fail Test ........................................................................ 10-2
To Select Source................................................................................... 10-3
To Create Mask .................................................................................... 10-3
Test and Ouput .................................................................................... 10-4
To Save the Test Mask .......................................................................... 10-5
To Load the Test Mask .......................................................................... 10-5
11 Waveform Record .......................................................................... 11-1
Waveform Record ................................................................................. 11-2
Record Constant O n.............................................................................. 11-4
Waveform Playback .............................................................................. 11-5
Waveform Analysis ............................................................................... 11-7
Analysis Based on Trace ............................................................... 11-10
Analysis Based on Pass/Fail Mask................................................... 11-11
12 Display Control ............................................................................... 12-1
To Select the Display Type ..................................................................... 12-2
To Set the Persistence Time .................................................................. 12-3
To Set the Waveform Brightness ............................................................ 12-5
To Set the Screen Grid .......................................................................... 12-5
To Set the Grid Brightness ..................................................................... 12-5
To Set the Menu Display ........................................................................ 12-5
13 Store and Recall ............................................................................. 13-1
Storage System .................................................................................... 13-2
Storage Type ....................................................................................... 13-3
Internal Storage and Recall ................................................................... 13-5
External Storage and Recall ................................................................... 13-8
Disk Management ............................................................................... 13-12
To Select File Type ....................................................................... 13-13
To Create a New File or Folder ...................................................... 13-14
To Delete a File or Folder .............................................................. 13-17
MSO4000/DS4000 User’s Guide
Page 26
RIGOL
XXIV
To Rename a File or Folder ........................................................... 13-18
To Clear the Local Memory ........................................................... 13-18
Factory.............................................................................................. 13-19
14 System Function Setting ................................................................ 14-1
Remote Int e rface Configuration ............................................................. 14-2
LAN Setting ................................................................................... 14-2
USB Device ................................................................................... 14-6
To Set the GPIB Address ................................................................ 14-6
System-related .................................................................................... 14-7
Sound .......................................................................................... 14-7
Language ..................................................................................... 14-7
System Informat i on ....................................................................... 14-8
Power-off Recall ............................................................................ 14-8
System Time ................................................................................. 14-9
Self-test Information .................................................................... 14-10
Screen ........................................................................................ 14-10
Error Informa t ion ......................................................................... 14-11
Self-calibration ............................................................................ 14-12
Power Status ............................................................................... 14-13
External Trigger Impedance .......................................................... 14-13
Aux Output ................................................................................. 14-14
Reference Clock ........................................................................... 14-15
Option Management .................................................................... 14-16
15 Remote Control ............................................................................. 15-1
Remote Control via USB ........................................................................ 15-2
Remote Control via LAN ........................................................................ 15-6
Remote Control via GPIB ....................................................................... 15-9
16 Troubleshooting ............................................................................ 16-1
17 Specifications ................................................................................ 17-1
18 Appendix ....................................................................................... 18-1
Appendix A: Accessories and Options ..................................................... 18-1
Append i x B: Warranty ........................................................................... 18-3
Index ........................................................................................................ 1
MSO4000/DS4000 User’s Guide
Page 27
1 Quick Start RIGOL
1-1
1 Quick Start
This chapter introduces the oscilloscope, the front panel, rear panel and user interface as well as the built-in help system.
The contents of this chapter:
General Inspection Appearance and Dimensions To Prepare the Oscilloscope for Use Front Panel Overview  Rear Panel Overview Front Panel Function Overview User Interface To Use the Security Lock To Use the Built-in Help System
Page 28
RIGOL 1 Quick Start
1-2

General Inspection

1. Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both ele ctrica l and mechanical tests.
The consigner or carrier shall be liable for the damage to instrument resulting from shipment. RIGOL woul d n ot be responsible for free maintenance/rework or replacement of the unit.
2. Inspect the instrument.
In case of any damage, or defect, or failure, notify your RIGOL sales representative.
3. Check the Accessories
Please check the accessories according to the packing l is t s. If t he acces sories are incomplete or damaged, please con t act your RIGOL sales representative.
MSO4000/DS4000 User’s Guide
Page 29
1 Quick Start RIGOL
1-3

Appearance and Dimensions

Figure 1-1 Front View Unit: mm
Figure 1-2 Side View Unit: mm
MSO4000/DS4000 User’s Guide
Page 30
RIGOL 1 Quick Start
1-4

To Prepare the Oscilloscope for Use

To Remove the Cover

Before using the oscilloscope, remove the front panel cover by grasping the tabs on each side and pull them in the arrow directions as shown in the figure below.
Figure 1-3 To Remove the Cover
MSO4000/DS4000 User’s Guide
Page 31
1 Quick Start RIGOL
1-5

To Adjust the Supporting Legs

For bette r observ at ion, a djust the supporting legs. Unfold or fold the supporting le gs in the arrow directions as shown in the figures below.
Figure 1-4 To Adjust the Supporting Legs
MSO4000/DS4000 User’s Guide
Page 32
RIGOL 1 Quick Start
1-6
AC Power Input Terminal

To Connect to AC Power Supply

This oscilloscope accepts two kinds of AC power supplies: 100-127V, 45-440Hz and 100-240V, 45-65Hz. Please use the power cord supplied with the accessories to connect the oscilloscope to the power source as shown in the figure b elow. Turn on the power switch under the power plug. At this point, the oscilloscope is energized and the power key
at the lower-left corner of the front panel will pulse.
Figure 1-5 To Connect to AC Power Supply
MSO4000/DS4000 User’s Guide
Page 33
1 Quick Start RIGOL
1-7

Power-on Inspection

When the oscilloscope is energized, press the power key corner of the front panel to start the oscilloscope. During the start-up process, the oscilloscope performs a series of self-tests and you can hear the sounds of relay switching. After the self-test is finished, the welcome screen is dis played and you can view the Option Type, Option Name, Option Edition and the remaining trail time available in the “Installed Options” pop-up dialog box on the screen if the trail versions of options a re cu rrently installed on the inst rument. W hen the inst rument is shipped, a trial version of all options are provided with about 2000 minutes of on-time. Press Utility System SelfTestInfo to view the self-test results.
a t the low er-left
MSO4000/DS4000 User’s Guide
Page 34
RIGOL 1 Quick Start
1-8

To Connect the Probe

RIGOL provides passive and active probes for the MSO4000/DS4000 series oscilloscopes. For detailed technical information, please refer to the corresponding Probe User’s Guide. The following a re the pro bes rec ommended f or this oscillos cope.
Model Description
RP3500 500 MHz, passive probe, standard, auto detection RP7150 1.5 GHz, active probe, optional, auto detection
Connect the Probe:
1. Connect the BNC terminal of the probe to a channel BNC connector of the oscilloscope at the front panel.
2. Connect the probe tip to the circuit point to be tested and connect the ground alligator clip of the probe to the circuit ground terminal.
Figure 1-6 To Connect the Probe
MSO4000/DS4000 User’s Guide
Page 35
1 Quick Start RIGOL
1-9

Function Inspection

1. Press Default to restore the oscilloscope to its default configuration.
2. Connect the ground alligator cli p of the probe to the “Ground Terminal” under the probe compensation signal output terminal.
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the “Compensation Signal Output Terminal” of the probe.
Compensation Signal Output Terminal Ground Terminal
Figure 1-7 To Use the Compensation Signal
4. Press AUTO.
5. Observe the wa vef orm o n the dis play. In normal condition, the display s houl d be a square waveform as shown in the figure below:
Figure 1-8 Square Waveform
6. Use the same method to test the other channels. If the square waveforms actually shown do not match that in the figure above, please perform
Probe
Compensation in the next section.
WARNING
To avoid electric shock, pl ease make su re that the insulate d wire o f the probe is in good condition and do not touch the metallic part of the probe when the probe is connected to a high vol t ag e source.
MSO4000/DS4000 User’s Guide
Page 36
RIGOL 1 Quick Start
1-10
Tip
probe compensation adjustment and can not be used for calibration.
The signal output from the probe compensation connector can only be used for
MSO4000/DS4000 User’s Guide
Page 37
1 Quick Start RIGOL
1-11

Probe Compensation

When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensa ted or poorly compensate d probes may increase measurement error. The probe compensation procedures are as follows:
1. Perform steps 1, 2, 3 and 4 of “
2. Check the wavefor ms displayed and compare them with the following.
Over compensated Perfectly compensated Under compensated
Figure 1-9 Probe Compensation
3. Use a nonmetallic driver to adjust the variable capacitor on the probe until the waveform displayed is as the “Perfectly compensated” in the figure above.
Function Inspection” in the previous section.
MSO4000/DS4000 User’s Guide
Page 38
RIGOL 1 Quick Start
1-12

To Connect the Logic Probe

RIGOL provides a logic probe for the MSO40 00 series digital oscilloscope. For detailed technical information of the logic probe, please refer to the corresponding logic probe User’s Guide.
Connect the logic probe single head to the digital channel input connector [LOGIC D0-D15] at the front panel of the MSO4000 series digital oscilloscope. Note: connect the logic probe adaptor to the corresponding branch header (called channel group) before connecting the logic probe to the device under test.
Figure 1-10 To Connect the Logic Probe
MSO4000/DS4000 User’s Guide
Page 39
1 Quick Start RIGOL
1-13

Front Panel Overview

1 2 3 4 5 6 7 8 9 10
11
12
13 14 15 16 17 18 19 20 21 22
Figure 1-11 Front Panel Overview
Table 1-1 Front Panel Description
No. Description No. Description
1 Menu 12 Record 2 LCD 13 Power Key 3 Multi-function Knob 14 USB HOST 4 Navigation Knob 15 Digital Channel Input Terminal 5 HORIZONTAL 16 Function Setting Menu Softkeys 6 CLEAR 17 VERTICAL 7 AUTO 18 Analog Channel Input Terminals 8 RUN/STOP 19 Function Menu Keys 9 SINGLE 20 TRIGGER 10 Default&Print 21 EXT TRIG Input Terminal 11
LOGIC ANALYZER
22 Probe Compensation Signal Output
Terminal/Ground Terminal
MSO4000/DS4000 User’s Guide
Page 40
RIGOL 1 Quick Start
1-14

Rear Panel Overview

1 2 3 4 5 6 7 8
Figure 1-12 Rear Panel Overview
1. Trig Out/Ca libration
Various kinds of signals can be output from this connector (press Utility AuxOutput to select the desired output type).
1) TrigOut: the oscill oscope sources a signal which has a frequency
proportional to the trigger rate of the instrument. Connect the signal to the waveform display device and measure the frequency of the signal, the measurement result is the same as the current capture rate.
2) Fast: the oscilloscope out puts a fast edge signal with 500 ps rise time which can be used in the self-calibration of the oscilloscope.
3) GND: the oscilloscope ground voltage level.
4) PassFail: the oscilloscope outputs a TTL high level when a failed waveform
is detected and low level at the beginning of the next waveform. This signal can be transmitted to other control system for easy view of the test result.
2. Reference Clock
Provides a more precise sample clock signal for the oscilloscope. It can also
MSO4000/DS4000 User’s Guide
Page 41
1 Quick Start RIGOL
1-15
synchronize two or more oscilloscope clocks.
3. Video Output
Through this interface, t he os cillosco pe can be connected to an external monito r for clearer waveform display. Note that the display of the oscilloscope is still operational when connected to an external monitor.
4. LAN
Through this interface, the oscilloscope can be connected to the local area network for remote control.
5. USB DEVICE
Through this interface, t he oscilloscope can be co nnected to pri nter or PC to print data waveforms or to control the oscilloscope using remote PC software.
6. USB HOST
Through this interface, the oscilloscope can be connected to printer or USB storage device to print waveform dat a or store waveform files. GPIB interface communication can be realized using the USB-GPIB interface con verter provide d by RIGOL. Note that there is also a USB HOST interface at the front panel.
7. Lock Hole
Use a security lock (available separately) to lock the oscilloscope to a fixed location.
8. AC Input/Switch
AC power input terminal. The power supply require ments of this oscillos cope are 100-127 V, 45-440 Hz and 100-240 V, 45-65 Hz. Please connect the oscilloscope to AC power using the power cord provided with the accessories. Then, turn on the AC power switch to energize the oscilloscope. Press the power ke y at the front panel to start the oscilloscope.
MSO4000/DS4000 User’s Guide
Page 42
RIGOL 1 Quick Start
1-16

Front Panel Function Overview

VERTICAL

The MSO4000/DS4000 p ro vides in dependent ve rtic al cont rol systems f or each of the four analog input channels.
CH1, CH2, CH3, CH4: analog input channels. The four channels are marked
by differe nt colo rs which are als o use d to ma rk the c orrespon ding w ave forms on the screen and the channel input connectors. Press any key to open the corresponding channel menu and press again to turn off the channel. AC: illuminated when AC channel coupling is enabled. 50: illuminated when the channel input impedance is 50Ω. BW: illuminated when bandwidth limit is enabled.
MATH: press this key to open the math operation menu under which (add,
subtract, multiply, divide) oper ati o ns, FFT opera tion, logic operation and advanced operation are provided.
REF: press this key to enable the reference waveform function to compare the
waveform actually tested with the referenc e wav eform.
VERTICAL POSITION: vertical position knobs. During the modification,
the position label at the lowe r left corner of the sc reen would change accordingly. Turn clockwise to increase the position and counterclockwise to decrease. Press down the knob to quickly reset the vertical position.
VERTICAL SCALE: vertical scale knobs. During the modification, the scale
label at the bottom of the screen would change accordingly. Turn clockwise to
MSO4000/DS4000 User’s Guide
Page 43
1 Quick Start RIGOL
1-17
decrease the scale with a step of 5-2-1 and turn counterclockwise to increase with a step of 1-2-5. Press down the knob to switch the vertical scale adjustment mode between “Coarse” and “Fine”.
Decode1 and Decode2: decoding function keys. Press the corresponding key
to open the decoding function menu. MSO4000/DS4000 supports parallel decoding and p rotocol decoding (for more details, refer to “
Protocol
Decoding”).

HORIZONTAL

MENU: press this key to open the horizo ntal control menu with which users can
turn on or off the delayed sweep function, switch between different time base modes, switch between “Coarse” and “Fine” adjustments of scale as well as modify the horizontal reference setting.
HORIZONTAL SCALE: modify the horizontal time base. Turn clockwise to
reduce the time base and turn counterclockwise to increase. During the modification, waveforms of all the channels will be displayed in expanded or compressed mode and the time base message (such as upper side of the screen would change accordingly. Press down the knob to quickly switch to delayed sweep state.
HORIZONTAL
POSITION: modify the trigger position. The trigger po i n t
would move left or right relative to the center of the screen when you turn the knob. During the modification, waveforms of all the channels would move left or right and the trigger position message (such as upper-right cor ner of the screen w ould change accor dingly . P ress down the knob to quickly reset the trigger position (or the delayed sweep position).
) at the
) at the
MSO4000/DS4000 User’s Guide
Page 44
RIGOL 1 Quick Start
1-18
MODE: press this key to switch the trigger mode to Auto,

TRIGGER

Normal or Single and the state backlight of the current
trigger mode would be illuminated.
TRIGGER
clockwise to increase the level and turn counterclockwise to reduce. During the modif ic ation, the trigger level line moves up and down and the value in the trigger level
message box (such as corner of the screen would change accordingly. Press down the knob to quickly reset the trigger level to zero point.
MENU: press this key to open the trigger o peration menu. This oscilloscope provides various trigger types.
FORCE: in Normal and Single trigger modes, press this key to generate a trigger signal forcefully.
50%: press this key to set the trigger level to the vertical midpoint of the trigger signal amplitude.
LEVEL: modify the trigger level. Turn

CLEAR

Press this key to clear all the waveforms on the screen. If the oscilloscope is in “RUN” state, waveforms will still be displayed.
) at the lower-left
MSO4000/DS4000 User’s Guide
Page 45
1 Quick Start RIGOL
1-19

AUTO

Press this key to enable the waveform auto setting function. The oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input s ignal to realize optimum w aveform display. Besides, the quick parameter measurement function is also provided (for more details, refer to “
Measurement after AUTO”).
Note: Auto setting requires that the frequency of the signal under test should be no l ower t han 50 H z, the dut y cycle be greater than 1% and the amplitude be at le ast 20 mVpp. If the parameters exceed these limits, “Can’t detect any signal” would be displayed after pressing this key and the quick parameter measurement menu might not be displayed.
Quick

RUN/STOP

SINGLE

Press this key to set the state of the oscilloscope to “RUN” or “STOP”. In “RUN” state, the key is illuminated in yellow and red in “STOP” state.
Press this key to set the trigger mode to “Single”. In single trigger mode, the oscilloscope generates a trigger when the trigger conditions are met and then stops.
MSO4000/DS4000 User’s Guide
Page 46
RIGOL 1 Quick Start
1-20

Multi-function Knob

This knob can be used to adjust the value of the parameters, adjust the waveform brightness and select the desired menu (the backlight turns on during operation). During parameter input, turn the knob clockwi s e to increase the p ar ameter and
count ercloc kwise to decreas e.
When the menu is hidden, rotate the knob to adjust the waveform brightness.
The adjustable range is from 0% to 100%. Turn the knob clockwise to increase the waveform brightness and counterclockwise to decrease. Press down the knob to reset the wavef orm brightness to 50%. Use rs can also p ress Display WaveIntensity and then rotate the knob to adjust the waveform brightness.
During menu operation , after p ressing a menu softkey, rotate the knob to switch
to the desired sub-menu under that menu and press down the knob to select that sub-menu. In addition, the kno b can also be used to modif y para meters and input filenames.

Navigation K n ob

This knob provides quick Adjust/Locate function for numerical parameters with relatively larger settable ranges. Turn th e kn ob clockwise (counterclockwise) to increase (reduce) the value. Note that the inner knob is used for fine adjustment and the outer knob for coarse adjustment. The larger the rotati o n ampli tu d e of the outer knob is, the fatster the change speed of the figure will be. For example, this knob can be used to quickly locate the waveform frame (Current Frame) to be played back in waveform p layback. Similar menus include trigger holdoff, pulse width setting, slope time etc.
MSO4000/DS4000 User’s Guide
Page 47
1 Quick Start RIGOL
1-21

Default

Press this key and then press OK to restore the oscil loscope
to its default configuration.

Print

Press this key to execute print function or save the screen in the USB storage device. If the oscilloscope is currently connected to a printer and the printer is in idle state, press this key to execute print function. If no printer but a USB storage device is currently connected, press this key to save the screen to the USB storage device in “.bmp” format. When printer and USB storage device are connected at the same time, the printer enjoys higher priority.

Function Keys

Measure: press this key to enter the measurement setting menu. Acquire: press this key to enter the sample setting menu. Here, y ou can set the
acquisition mode, the memory depth, and the antialiasing function of the oscilloscope.
Storage: press this key to enter the file store and recall interface. The file types
available include traces, waveforms, setups, pict ure and CSV. This i nstrument supports internal/external stora ge as well as disk management.
Cursor: press this key to enter the cursor measurement menu. The oscilloscope
provides three cursor modes (manual, track a nd auto).
MSO4000/DS4000 User’s Guide
Page 48
RIGOL 1 Quick Start
1-22
Stop Play/Pause Record
Display: press this key to enter the display setting menu to set the display t ype,
persistence time, wave intensity, grid type, grid brightness and menu display time o f t he waveform.
Utility: press this key to enter the system function setting menu to set the
system-related functions or parameters, such as I/O setting, sound and language. The oscilloscope also supports some advanced functions such as pass/fail test and print setting.
Record: press this key to enter the waveform record menu to set the waveform
record and open (record constant on) mode.
Help: press this key to open the help interface. For detailed info rmation, pleas e
refer to the introduction in To Use the Built-in Help Syste m”.

Record

Play/Pause: in stop or pause state, press this key to play the recorded or paused
waveform and press again to pause the play. The backlight is illuminated in yellow. Stop: press this key to stop the waveform in record or play back m ode. The backlight is illuminated in orange. Record: press this key to start recording the waveform. The backlight is illuminated in red. Besides, the backlight will a lso be illuminate d when the open (record constant on) mode is enabled.
MSO4000/DS4000 User’s Guide
Page 49
1-23
LA
1 Quick Start RIGOL

LOGIC ANALYZER

LA_ : this knob can be used to reposition the digital
: press this key to enter the logic analyzer setting interface. Y ou can enable or disable groups of channels or a single channel, change the display size and the logic threshold of the digital channels and group the 16 digital channels and display them as a bus. You c an also set a label for each digital channel.
channels. Press this knob to switch the digital channels currently enabled. The selected number of the digital channel (from small to large) will be highlighted in red. Rotate the knob to move the digital channel to the desired position and press this knob again to fix this channel.
MSO4000/DS4000 User’s Guide
Page 50
RIGOL 1 Quick Start
1-24

User Interface

The MSO4000/DS4000 oscilloscope features 9 inch, WVGA (800*480) 160,000 color TFT LCD. The display also features a 14-grid ultra-wide screen view that enables you to observe “longer” waveforms.
1 2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 17 18 19 20
Figure 1-13 User Interface
1. Measurement Menu
Provide 12 horizontal (HORIZO NTAL) and 12 vertical (VERTICAL) measurement parameters. Press the softkey at the left of the screen to activate the corresponding measurement item.
MSO4000/DS4000 User’s Guide
Page 51
1 Quick Start RIGOL
1-25
2. Analog Channel Label/Waveform
Each analog channel is col or c ode d t o match the input. For example, CH1 is yellow and CH2 is light blue.
3. Digital Ch an n el Label/Waveform
The logic high levels of th e digital wa vefo rms are di splayed in blue, the lo gic low levels are display ed in green whic h complies wit h the color of the digital channel label and the edges are displayed in white. Both the label and the waveform of the digital channel currently selected are displayed in red.
4. Status
Available states include RUN, STOP, T’D (triggered), WAIT and AUTO.
5. Horizontal Time Base
Represen t the tim e per grid on the horizontal axis on the screen. Use HORIZONTAL
SCALE to modify this parameter. The range
available is from 1.000 ns to 1000 s.
6. Sample Ra te/Memory Depth
Displays the current sampl e rate and me mory depth of the oscilloscope. Use HORIZONTAL
SCALE to modify this parameter.
7. Waveform Memory
Provides the schematic diagram of the memory position of the waveform currently on the screen.
waveform on the
screen
8. Trigger Position
Displays the trigger posi t ion of the waveform.
9. Horizontal Position
Use HORIZONTAL
POSITION to modify this parameter. Press down
the knob to automatically set the parameter to 0.
MSO4000/DS4000 User’s Guide
Page 52
RIGOL 1 Quick Start
1-26
10. Trigger Type
Displays the currently selected trigger type and trigger condition setting. A different label is displayed when a different trigger type is selected. For example,
represents triggering on the rising edge in “Edge” trigger.
11. Trigger Source
Displays the trigger source (CH1 to CH4, EXT, EXT/5, AC Line or D0-D15) currently selected. A different label is displayed when a different trigger source is selected and the color of the trigger parameter area will change accordingly. For example,
de notes that CH1 is selected as the trigger source.
12. Trigger Level/Threshold
The trigger level label
will be displayed at the right of the screen when the trigger source is set to CH1 to CH4 and the trigger level value is displayed at the upper-right corner of the screen. Note: In runt trig ger a nd s lope t rigger, there are two trigger level la bels (
and ). For detailed information, refer to “Vertical Window” on
page
5-14 and 5-18.
When using TRIGGER
trigger level value will change with the up and down of
LEVEL to modify the trigger level, the
.
There is no trigger level label displayed when the tri gger source is set to
D0-D15 and the trigger threshol d is displayed at the upper-right corner of the screen.
13. CH1 Vertica l Sc a le
Displays the voltage value per grid of the CH1 waveform. Use VIRTICAL
SCALE to modify this parameter.
The following labels will be provided according to the current channel
setting: Channel Coupling
(such as in AC coupling), Bandwidth
Limit (such as when bandwidth limit is enabled) and Input Impedance (such as when the input impedance is 50 Ω).
14. CH2 Vertica l Sc a le
Displays the voltage value per grid of the CH2 waveform. Use VIRTICAL
SCALE to modify this parameter.
The following labels will be provided according to the current channel
MSO4000/DS4000 User’s Guide
Page 53
1 Quick Start RIGOL
1-27
setting: Channel Coupling (such as in AC coupling), Bandwidth
Limit (such as when bandwidth limit is enabled) and Input Impedance (such a s when the input impedance is 50 Ω).
15. CH3 Vertical Sca le
Displays the voltage value per grid of the CH3 waveform. Use VIRTICAL
SCALE to modify this parameter.
The following labels will be provided according to the current channel
setting:
Channel Coupling (such as in AC coupling), Bandwidth Limit (such as when bandwidth limit is enabled) and Input Impedance (such a s when the input impedance is 50 Ω).
16. CH4 Vertica l Sc a le
Displays the voltage value per grid of the CH4 waveform. Use VIRTICAL
SCALE to modify this parameter.
The following labels will be provided according to the current channel
setting:
Channel Coupling (such as in AC coupling), Bandwidth Limit (such as when bandwidth limit is enabled) and iInput Impedance (such a s when the input impedance is 50 Ω).
17. Message Box
Displays prompt messages.
18. Digital Chann el Status Area
Displays the current state of the 16 di gital channels (D0 to D15 from right to left). The digital channels currently enabled are displayed in green and the digital channel currently selected is highlighted in red. Any digital channel that is turned off will be grayed out in the Digital Channel Status Area.
19. Notification Area
Displays system time, sound icon and USB storage device ico n. System Time: displayed in “hh:mm (hour:minute)” format. When
printing or storing the waveform, the output file will contain this time message. Press Utility System System Time to set the time in the following format: yyyy-mm-dd hh:mm:ss (year-month-date hour:minute:second)
MSO4000/DS4000 User’s Guide
Page 54
RIGOL 1 Quick Start
1-28
Sound Icon: when sound is enabled, is displayed. Press Utility 
Sound to turn the sound on or off.
USB Storage Device Icon: when the oscilloscope detects a USB storage
device, is displayed.
20. Operation Menu
Press any softkey to activate the corresponding menu. The following symbols might be displayed in the menu:
Denotes that items. The backlight of
at the front panel can be used to select parameter
turns on when the parameter selection is valid. Denotes that can be used to modify a parameter value. The backlight of
turns on when the parameter input is valid. Denotes that you can use the navigation knob to quickly adjust/locate parameters. Denotes that users can rotate and then press down
to select the desired parameter. Denotes that the current menu has several opti on s . Denotes that the current menu has a lower level menu. Press this key to return to the previous menu.
Note: The following direction keys mig ht a ppear in t he grid at the lowe r-left corner of the menu bar:
Denotes that you can open the next page menu. Denotes that you can return to the previous page menu.
MSO4000/DS4000 User’s Guide
Page 55
1 Quick Start RIGOL
1-29

To Use the Security Lock

If needed, you can use a security lock (available separately) to secure the oscilloscope to a fixed location. The method is as f ollows, align the lock with the lock hole and plug it into the lock hole vertically, turn the key clockwise to lock the oscilloscope and then pull the key out.
Figure 1-14 To Use the Security Lock
Note: Do not insert other articles into the security lock hole to avoid damaging the instrument.
Security Lock Hole
MSO4000/DS4000 User’s Guide
Page 56
RIGOL 1 Quick Start
1-30

To Use the Built-in Help System

The help system of this oscilloscope provides instructions for all the function keys (including menu keys) at t he front panel. Press Help to open the help interface and press again to close the interface. The help interface mainly consists of two parts. The left is “Help Options” and y ou can use “Button” or “Index” mode for selection. The right is “Help Display Area”.
Help Options Help Display Area
Figure 1-15 Help Information
Button:
Default mode. In this mode, you can press the keys (ex cept the power key knob the corresponding help information of that key in the “Help Display Area”. Use mode.
Index:
In this mode, use item currently selected is display ed with brown sh ading. Pre s s down the knob to get the corresponding help information in the “Help Display Area”. Use mode.
a n d menu page up/ d own key / ) a t th e front panel directly to get
to select “To Index” and then press down the knob to switch to Index
to se lect th e desired item (for example, “Band Width”). The
to select “To Button” and then press down the knob to s w itch to Button
MSO4000/DS4000 User’s Guide
, the
Page 57
2 To Set the Vertical System RIGOL
2-1
2 To Set the Vertical System
The contents of this chapter:
To Enable the Analog Channel Channel Coupling Bandwidth Limit Probe Input Impedance Waveform Invert Vertical Scale Vertical Expansion Amplitude Unit Channel Label Delay Calibration
Page 58
RIGOL 2 To Set the Vertical System
2-2

To Enable the Analog Channel

The MSO4000/DS4000 provides four analog input channels (CH1 to CH4) and provides independent vertical control system for each channel. As the vertical system setting methods of the four channels are c ompletely the s ame, this c hapter uses CH1 as the example.
Connect a signal to the ch annel connect or of CH 1 and the n p ress CH1 in the vertical control area (VERTICAL) at the fr ont pane l t o enable CH1. At this point:
Panel: The backlight of this key turns on. If the corresponding function has been enabled, the character “AC”, “50” or “BW” will also be illu minated. Note that the on /off state of the key light of “AC”, “50” or “BW” is not controlle d by the on/off state of the ch annel.
Screen:
The channel setting menu is display ed at the ri ght side of the screen an d the channel label at the bottom of the screen (as shown in the figure below) is highlighted. The information displayed in the channel label is related to the current channel setting.
After the channel is turned on, modify the paramete rs such as the vertical scale, the horizontal time base and the trig ger mo de according to the input signal to make t he waveform display easy to observe and measure.
MSO4000/DS4000 User’s Guide
Page 59
2 To Set the Vertical System RIGOL
2-3

Channel Coupling

Set the coupling mode to filter out the undesired signals. For example, the signal under test is a square waveform with DC offset. When the coupling mode is “DC”: the DC a nd AC co mponents of the signal under
test can both pass the channel.
When the coupling mode is “AC”: the DC components of the signal under test
are blo cked.
When the coupling mode is “GND”: the DC and AC components of the signal
under test are both blocked.
Press CH1 Coupling and use to select the desired coupling mode (the default is DC). The current coupling mode is displayed in the channel label at the bottom of the screen. When “AC” is selected, the character “AC” below CH1 at the front panel will be illuminated. You can also press Coupling continuously to switch the coupling mode.
MSO4000/DS4000 User’s Guide
Page 60
RIGOL 2 To Set the Vertical System
2-4

Bandwidth Limit

Set the bandwidth limit to reduce display noise. For example, the signal under test is a pulse with high frequency oscillations. When bandwidth limit is disabled, the high frequency components of the signal
under test can pass the channel.
Enable bandwidth limit and limit the bandwidth to 20 MHz, 100 MHz or 200 M Hz,
the high frequency components that exceed 20 MHz, 100 MHz or 200 MH z are attenuated.
Press CH1 BW Limit and use to enable or disable the bandwidth limit (the default is OFF). When the bandwidth limit (20 MHz, 100 MHz or 200 MHz) is enabled, the char acter “B” will be displayed in the channel label at the bottom of the screen. You can also press BW Limit continuously to switch between on and off of the bandwidth limit.
MSO4000/DS4000 User’s Guide
Page 61
2 To Set the Vertical System RIGOL
2-5
0.01X
1000 X
1:100
1000:1

Probe

This oscilloscope supports normal passive probes and active differential probes. It can automatically identify the type of the probe currently connected and the probe ratio. Press CH1 Probe to open the probe operation menu.
1. ProbeType
Read the type of the probe currently connected as “Nor-Probe” or “DiffProbe”. Note that when a 50 Ω “DiffProbe” is used, the channel is set to “50 Ω” automatically.
Normal Pr obe: such as RIGOL RP3500A. Differential Probe: such as RIGOL RP7150.
2. Ratio
For so me “Nor-Probe”s, the oscilloscope can identify the probe rat io automatically. If not, press this softkey to select the corresponding probe ratio. The values available are listed in the table below.
Table 2-1 Probe Attenuation Coefficient
Menu Attenuation coefficient
Input Impedance of the
0.02X
0.05X
0.1X
0.2X
0.5X 1X 2X 5X 10X
When “DiffProbe” is connected, the oscilloscope recognizes it automatically and the following menus are added.
MSO4000/DS4000 User’s Guide
20 X
50 X 100 X 200 X 500 X
1:50 1:20 1:10
1:5 1:2 1:1 2:1
5:1 10:1 20:1 50:1
100:1 200:1 500:1
Page 62
RIGOL 2 To Set the Vertical System
2-6
3. Front-End
RP7150 active probe provides “Single-end” and “Difference” probe heads. Press this softkey to select the desired probe head.
4. Probe-Cal
Connect the differential p robe to the channel in put terminal (such as CH 1) of the oscilloscope correctly, then connect the fast edge signal output from the [Trig Out/Calibration] connector at the rear panel of the instrument to the probe and CH2 respectivel y. At this point, press this softkey and the oscilloscope starts to calibrate the probe.
Note: If the probe is connected to CH1, CH3 or CH4, the fast edge signal should be connected to CH2; if the probe is connected to CH2, the fast edge signal should be connected to CH1.
5. Probe Info
Press this softkey to view information ab out the probe, su ch as the manufa cturer, model, serial number and the date of last calibration.
6. Bias Voltage
The oscilloscope provides a bias voltage for diffe rential probes. This bias voltag e is used to adjust the input dynamic range of the probe amplifier to an appropriate range to ensure accurate measurement of the signal under test. Press this softkey and use the knob to adjust the bias voltage. The range i s from
-12V to +12V.
MSO4000/DS4000 User’s Guide
Page 63
2 To Set the Vertical System RIGOL
2-7

Input Impedance

To reduce the ci rcuit load caused by the interaction between the oscilloscope and the circuit under test, the oscilloscope provides two input impedance modes: 1 MΩ (default) and 50 Ω. 1 MΩ: at this point, the input im pedance of the oscilloscope is v ery high an d the
current flows into the oscilloscope from the circuit under test can be ignored.
50 Ω: match the oscilloscope with devices with 50 Ω output impedance.
Press CH1 Input to set the input imped ance of the oscilloscope. When “50 Ω” is selected, the character “Ω” is displayed in the channel label at the bottom of the screen.
Note that after the probe is aut omatically ide ntified, the input impedance will also be automatically identified and you do not need to set it manually.

Waveform Invert

When waveform invert is enabled, the waveform display rotates 180 degree relative to the ground potential. When waveform invert is disabled, the waveform display is normal. Press CH1 Invert to enable or disable wave form in vert.
MSO4000/DS4000 User’s Guide
Page 64
RIGOL 2 To Set the Vertical System
2-8

Vertical Scale

The vertical scale can be adjusted in “Coarse” or “Fine” mode.
Press CH1 Volts/Div to select the desired mode. Rotate VERTICAL SCALE to adjust the vertical scale (clockwise to reduce the scale and counterclockwise to increase).
The scale information (such as of the screen will chan ge accordingly during the adjustment. The adjustable range of
the vertical scale is relat ed to the current probe ratio. By default, the probe ratio is 1X and the vertical scale range is from 1 mV/d i v t o 5 V/div.
Coarse adjustment (take counter clockwise a s an example): s et the v ertical s cale
in 1-2-5 step namely 1 mv/div, 2 mV/div, 5 mV/div, 10 mV/div…5 V/div.
Fine adjustment: enables fine adjustment to the vertical scale. If the amplitude
of the input waveform is a little bit greater than the full scale and the amplitude would be a little bit lower if the next scale is used, fine adjustment can be used
to improve the ampl i t ude of waveform display to v i ew signal details. Note: You can also press VERTICAL and “Fine” adjustments.
SCALE to quickly switch between “Coarse”
) in the channel label a t the bottom
MSO4000/DS4000 User’s Guide
Page 65
2 To Set the Vertical System RIGOL
2-9

Vertical Expansion

When using VERTICAL channel, you can choose to expand or compress the signal vertically arou nd the center of the screen or the ground point of the signal.
Press Utility System VerticalExp to select “Center” or “Ground” and the default is “Ground”. Center: when the vertical scale is modified, the waveform will expand or
compress around the center of the screen.
Ground: when the vertical scale is modified, the waveform ground level will
remain at the same point on the screen and the waveform will expand or compress around this point.
SCALE to change the vertical scale of the analog

Amplitude Unit

Select the amplitude displ ay unit f or the cur rent channel. The available units are W, A, V and U. When the unit is changed, the unit displayed in the channel label will change accordingly.
Press CH1 Unit to se lect the desired unit and the default is V.
MSO4000/DS4000 User’s Guide
Page 66
RIGOL 2 To Set the Vertical System
2-10

Channel Label

You can m odif y the labels used t o mark the analog c hannels (CH1 to CH 4) at t he left side of the screen. The label is the number of the channel (such as ) by default and the length of the label can not exceed 4 ch ar acters.
Press CH1 Label to enter the label modification interface as shown in the figure below.
Name Input Area Keyboard Upper/Lower Case Switch
For example, change “
” to “ ”.
Press Keyboard to select the “Keyboard” area. Select “Aa” using and press down
to switch it to “aA”. Select “C” using and press down to input the
character. Use the same method to input “hn1”.
T o modif y or delete the input char acter, press Name to select the “Name Input Area” and use
to se lect the character to be modified or deleted. Enter the desired
character or press Delete to delete the character selected.
After finishing the input, press OK to finish the modification and the channel label will change to “
”.
MSO4000/DS4000 User’s Guide
Page 67
2 To Set the Vertical System RIGOL
2-11

Delay Calibration

When using an oscilloscope for actual measurements, the transmission delay of the probe cable may bring greater error (zero offset). The MSO4000/DS4000 allows users to set a de lay time to c alibrate the zero offset of the co r respondi ng channel. Zero offset is defined as the off set of the crossing point of the waveform and trigger level line relative to the trigger position, as shown below.
Trigger Position
Trigger Level
Press CH1 Delay-Cal and use to select the desired delay time. The range available is from -100 ns to 100 ns. Note: This parameter is related to the current horizontal time base. For example, when the horizontal time base is 5 us, the step of the delay time setting is 100 ns. when the horizontal time base is 1 us, the step of the delay time setting is 20 ns. when the horizontal time base is 500 ns, the step of the delay time setting is 10 ns.
MSO4000/DS4000 User’s Guide
Zero Offset
Page 68
Page 69
3 To Set the Horizontal System RIGOL
3-1
3 To Set the Horizontal System
The contents of this chapter:
Delayed Sweep Time Base Mode Horizontal Scale Horizontal Reference
Page 70
RIGOL 3 To Set the Horizontal System
3-2

Delayed Sweep

Delayed sweep can be use d to enla rge a len gth of waveform horizontally to view the waveform details.
Press MENU in the horizontal control area (HORIZONTAL) and press Delayed to enable or disable delayed sweep. Note that to enable delay ed sweep, the curre nt time base mode must be “Y-T” and th e “Pass/Fail test” must be disabled.
MSO4000/DS4000 User’s Guide
Page 71
3 To Set the Horizontal System RIGOL
3-3
Tip
In delayed sweep mode, the screen is divided into t wo dis play a reas as show n in the figure below.
The waveform before
enlargement
Main time base
Delayed Sweep Time Base
The waveform after enlargement
The waveform before enlargement:
The waveform in the area that has not been covered by the subtransparent blue in the upper part of the screen is the waveform before enlargement. You can turn
HORIZONTAL HORIZONTAL
POSITION to move the area left and right or turn SCALE to enlarge or reduce this area.
The waveform after enlargement:
The waveform in the l ower part of the screen is the h orizontally expande d wa veform. Note that compared to the main tim e base, the delayed time base has increased waveform resolution (as shown in the figure above).
Note: The delayed time base should be less than or equal to the main time base.
When the current channel is CH1 to CH4, you can also press down HORIZONTAL
SCALE (delayed sweep shortcut key) to directly switch to delayed sweep mode.
MSO4000/DS4000 User’s Guide
Page 72
RIGOL 3 To Set the Horizontal System
3-4

Time Base Mode

Press MENU in the horizontal control area (HORIZONTAL) at the front panel and then press Time Base to select the time base mode of the oscilloscope and the default is Y-T.

Y-T Mode

This mode is the main time base mode and is applicable to CH1 to CH4. In this mode, the Y a xis rep resents volta ge and the X axis represents ti me. Note that only when this mode is enabled can
Delayed Sweep be tur ned on.
MSO4000/DS4000 User’s Guide
Page 73
3 To Set the Horizontal System RIGOL
3-5

X-Y Mode

In this mode, the oscilloscope automatically turns on all f our chan ne ls (CH1 to CH4) and the screen is divided into two coordinate areas. X1 and Y1 track the vo ltages of CH2 and CH1 and X2 and Y2 track the vo l t ages of CH4 and CH3. The phase deviation between two signals with the same frequency can be easily measured via the Lissajous method. The figure below shows the measurement schematic diagram of the phase deviation.
The signal must be centered horizontally
D
C
III IV
II I
II I
A
III IV
B
According to sin
θ=A/B or C/D (wherein, θ is the phase deviation angle between the two channels and the definitions of A, B, C and D are as shown in the figure above), the phase deviation angle is obtained, that is:
θ=±arcsin (A/B) or ±arcsin (C/D)
If the principal axis of the ellipse is within quadrant I and III, the phase deviation angle obtained should be within quadrant I and IV, namely within (0 to π/2) or (3π/2 to 2π). If the principal axis of the ellipse is within quadrant II and IV, the phase deviation angle obtained s hould be within quadrant II and III, namely within (π/2 to π) or (π to 3π/2).
MSO4000/DS4000 User’s Guide
Page 74
RIGOL 3 To Set the Horizontal System
3-6
The X-Y function can be used to measure the phase deviation when the signal under test passes through a circuit network. Connect the oscilloscope to the circuit to monitor the input and output signals of the circuit.
Application example: measure the phase deviation of the input signals of two channels.
Method 1: Use Lissajous method
1. Connect a sine signal to CH1 and then connect a sine signal with the same
frequency and amplitude but a 90° phase deviation to CH2.
2. Press AUTO and enable X-Y mode. Then, press XY1 to select “On”.
3. Roate VERTICAL POSITION of CH1 and CH2 to display the signals at the center of the screen and roate VERTICAL
SCALE to make the signals easy
to observe. At this point, the circle as shown in the figure below should be displayed.
4. As shown in the figure above, the distances from the crossing points of axis and the circle to the origin of the coordinates are approximately equal. Thus, the phase deviation angle
θ=±arcsin1=90°.
Note:
In Y -T mode, the oscilloscope could use any sample rate (within the guaranteed
range) to ca pture the waveform. The maximum sample rate of X-Y mode is 2. 0 GSa/s. Generally, reducing the sample rate can impr o v e the display ed Lissajous figure.
MSO4000/DS4000 User’s Guide
Page 75
3 To Set the Horizontal System RIGOL
3-7
When X-Y mode is enabled, Delayed Sweep will be disabled automatically. The following functions are not available in X-Y mode:
Auto measure, cursor measure, math operation, reference waveform, delayed sweep, vector display, HORIZONTAL
POSITION, trigger control, memory
depth, acquisition mode, Pass/Fail test and waveform record.
Method 2: Use the shortcut measurement function
Please refer to “Phase AB ” and “Phase AB ” measurement fun ctions of “Delay and Phase” on page
6-16.
MSO4000/DS4000 User’s Guide
Page 76
RIGOL 3 To Set the Horizontal System
3-8
hen the horizontal time base is set to 200

Roll Mode

In this mode, the waveform scrolls f r om the right to the left, much like a strip chart. This is useful fo r viewing slower signals a s the si gnal i s “drawn out” on the display as time elapses. The waveform horizontal position and tri gger controls are not available. The range of horizontal scale adjustment is from 200.0 ms to 50.00 s.
Note: When the Roll mode is enabled, “ “
Pass/Fail Test”, “Measurement Range”, “Waveform Record”, “To Set the
Persistence Time”, “To Trigger the Oscilloscope” and “Logic Analyzer” are
not available.
Slow Sweep
Another mode similar t o the Roll mode. W ms/div or slower , the instrument enters “slow sweep” mode in which the instrument fi r s t acq u ires the data to the left of the trigger point and then waits for trigger event. After the trigger occurs, the instrument continues to finish the waveform at the right of the trigger point. When the slow sweep mode is used to observe low frequency signal, it is recommended that the
Delayed Sweep”, “Protocol Decoding”,
Channel Coupling is set to “DC”.
MSO4000/DS4000 User’s Guide
Page 77
3 To Set the Horizontal System RIGOL
3-9

Horizontal Scale

Similar to “ “Fine” mode.
Press MENU ScaleAdjust in the horizontal control area (HORIZONTAL) at the front panel to select the desired mode. Turn HORIZONTAL SCALE to adjust the horizontal scale. Turn clockwise to reduce the horizontal scale and turn countercloc kwi se to increase.
Vertical Sc ale”, the horizontal scale can be adjusted in “Coarse” or
The scale information (such as will change accordingly during the adjustment . The range of horizontal scale
adjustment is from 1.00 0 ns to 1000 s.
Coarse (take counterclockwise as an example): sets the horizontal scale in 1-2-5
step namely 1 ns, 2 ns, 5 ns…1000 ns.
Fine: higher resolution steps allow fine adjustment of horizontal scale.
) at the upper left corner of the screen
MSO4000/DS4000 User’s Guide
Page 78
RIGOL 3 To Set the Horizontal System
3-10

Horizontal Reference

The horizontal reference is the reference position according to which the screen waveform expands and compresses horizontally when adjusting HORIZONTAL
SCALE. In Y-T mode (this function is not available in X-Y mode and Roll mode), press MENU HorRef in the horizontal control area (HO RIZONTAL) to select the des ired
reference mode and the default is “Center”.
1 The Center of the Screen
When changing the horizontal time base, the waveform expands or compress es horizontally around th e center of the screen.
2 Trigger Position
When changing the horizontal time base, the waveform expands or compresses horizontally around the trigger point.
3 User-defined
When changing the horizontal time base, the waveform expands or compresses horizontally around the user-defined reference position. In the horizontal direction, the screen can display a maximum of 700 points with the leftmost as 350 and the rightmost as -350.
MSO4000/DS4000 User’s Guide
Page 79
4 To Set the Sample System RIGOL
4-1
4 To Set the Sample System
The contents of this chapter:
Acquisition Mode Sample Mode Sample Rate LA Sample Rate Memory Depth Antialiasing
Page 80
RIGOL 4 To Set the Sample System
4-2

Acquisition Mode

The acquisition mode is used to control how to generate waveform points from sample points.
Press Acquire Acquisition in the function menu at the front panel and use to select the desired acquisition mode (the default is normal), then press down the knob to select this mode. You can also press Acquisition continuously to switch t he acquisition mode.

Normal

In this mode, the oscilloscope sam ples the signal at equal time int erval to re build th e waveform. For most of the waveforms, the best display effect can be obtained using this mode.

Average

In this mode, the oscilloscope averages the wavef orms from multiple samples to reduce the random noise of the i nput signal and im prove the v ertical resolution. More averages can lower the noise and may increase the vertical resolution but it will slow the overall response of the displayed waveform to input changes.
The available ran ge of the number of ave rages is from 2 to 8192 a nd the def ault is 2. When “Average” mode is selected, press Averages and use to set the desired number of averages as the power function of 2.
MSO4000/DS4000 User’s Guide
Page 81
4 To Set the Sample System RIGOL
4-3
The Waveform before Average:
The Waveform after 256 Averages:
MSO4000/DS4000 User’s Guide
Page 82
RIGOL 4 To Set the Sample System
4-4

Peak Detect

In this mode, the oscilloscope acquires the maximum and minimum values of the signal w ithin the sample in terval to get the env elope or the narrow pulse of the signal that might be lost. In this mode, signal confusion can be prevented but the noise displayed would be larger.
In this mode, the oscilloscope can displa y all the pulses with pulse widths at le ast as wide as the sample period.

High Resolution

This mode uses an over-sample technique to average the neighboring points of the sample waveform. This reduces the random noise on the input signal and generates a much smoother waveform on the screen. This is generally used when the sample rate of the digital converter is hi gher than the st orage r ate of the a cquisition memory.
Note: “Average” and “High Res” modes use dif fe rent a ver agi ng me tho ds. The former uses “Waveform Average” and the latter uses “Single-sample Average”.
MSO4000/DS4000 User’s Guide
Page 83
4 To Set the Sample System RIGOL
4-5
Tip
to stop the sample, the oscill osc o pe will hold the last display. At
zoom the waveform.

Sample Mode

This oscilloscope only supports real-ti me sampling. In this mode, the oscilloscope samples and displays waveforms within a trigger event. The maximum real-time sample rate of MSO4000/DS4000 is 4 GSa/s and the current sample rate is displayed under the Sa Rate menu.
Press RUN/STOP this point, you can still use the vertical control and horizontal control to pan and
MSO4000/DS4000 User’s Guide
Page 84
RIGOL 4 To Set the Sample System
4-6
t
/Δ
1
Rate Sample =

Sample Rate

Sample refers to the process that the oscilloscope converts analog signals to digital signals at a certain time interval and stores the data in order. Sampl e ra t e i s t he reciprocal of the time interval.
The analog channels sample rate of this oscilloscope is up to 4 GSa/s. Note that the sample rate is displayed in the status bar at the upper side of the screen and in the Sa Rate menu and can be ch anged by adjusting the horizontal time base (s/div) through HORIZONTAL
The influence on the waveform when the sample rate is too low:
1. Waveform Distortion: when the sample rate is too low, some waveform details are lost and the waveform displayed is rather different from the actual signal.
SCALE or modifying the “Memory Depth”.
MSO4000/DS4000 User’s Guide
Page 85
4 To Set the Sample System RIGOL
4-7
2. Waveform Confusion: when the sample rate is lower than twice the actual signal frequency (Nyquist Frequency), the frequency of the wave form rebuilt from the sample data is lower than the actual signal frequency. The most common aliasing is the jitter on a fast edge.
3. Waveform Leakage: when the sample rate is too low, the waveform rebuilt from the sample data does not reflect all the actual signal information.
Pulse disappeared
MSO4000/DS4000 User’s Guide
Page 86
RIGOL 4 To Set the Sample System
4-8

LA Sample Rate

LA sampling ref ers t o the process t hat the oscillosco pe sa mples the compare d digital signal at a certain time interval. LA sample rate is the reciprocal of the interval. For example, the oscilloscope with 1 GHz LA sample rate samples digital signals every 1 ns. T he digital signal ment ioned here refers to the signal generated by comparing the input signal with the threshold spe cified by users. The input signal is ju dged a s logi c high level when its amplitude g reater than the th reshold a nd lo gic low le vel w hen its amplitude is lower than the threshold.
The digital channels sample rate of this oscilloscope is up to 1 GSa/s (per channel). Note that the LA sample rate is displayed in the La SaRate menu and can be changed by a djusting the horizontal time base (s/div) through HORIZONTAL SCALE or modifying the “Memory Depth”.
MSO4000/DS4000 User’s Guide
Page 87
4 To Set the Sample System RIGOL
4-9
T
Trigger Point
Pre-
sample
Delayed Sample

Memory Depth

Memory Depth
Memory depth refers to the number of waveform points that the oscilloscope can store in a single trigger sample and it reflects the storage ability of the sample memory. MSO4000/DS4000 provides up to 140 M points memory depth.
The relation of memory depth, sample rate and wa ve form len gth ful fills the equation below:
Memory depth = sample rate (Sa/s) × waveform length (s/div × div)
Thus, higher memory depth can en s ure higher sample rate.
Press Acquire Mem Depth, use to switch to the desired memory depth (the default is auto) and then press down the knob to select the option. You can also press Mem Depth continuously to switch the memory depth.
When a single channel is enabled, the memory depths available include Auto, 14kPoints, 140kPoints, 1.4MPoints, 14MPoints and 140MPoints. In “Auto” mode, the oscilloscope selects the memory depth automati cally accordin g to the current sa mple rate.
When dual channels are enabled (CH1 and CH2 or CH3 and CH4), the memory depths available include Auto, 7kPoints, 70kPoints, 700kPoints, 7MPoints and 70MPoints. In “ Auto” mode, the oscillosco pe selects the memory de pth automatically according to the current sample rate.
MSO4000/DS4000 User’s Guide
Page 88
RIGOL 4 To Set the Sample System
4-10

Antialiasing

At slower sweep speeds, the sample rate is reduced and a dedicated display algorithm is used to minimize the possibility of aliasing.
Press Acquire Anti_aliasing to enable or disable the antialiasing function. By default, antialiasing is disa bled. The dis played wa veforms w ill be mo re susceptible to aliasing when this function is disabled.
MSO4000/DS4000 User’s Guide
Page 89
5 To Trigger the Oscilloscope RIGOL
5-1
5 To Trigger the Oscilloscope
The trigger system of an oscilloscope allows users to tailor data collection an d isolate particular events by providing a flexible control system to “select” certain events (rising edge, falling edge, etc.).You can set a trigger condition according to the requirement and when a wavef orm in the wa veform st ream meets this con dition, the oscilloscope captures this waveform as well as the neighbouring part and then displays the captured waveforms on the screen. The oscilloscope samples waveform data continuously, but only a stable trigger can guarantee stable waveform displays. The trigger circuit guarantees that each sweep or sample starts from the trigger condition defined by the user. Each sweep and sample are synchronous and the waveforms acquired overlap to display the stable waveform.
The trigger setting should be based on the features of the input signal, thus you nee d to have some knowledge of the signal under test to quickly capture the desired waveform. This os cilloscope provides abundant adv anced trigger f unctions which can help you to focus on the waveform details of interest.
The contents of this chapter:
Trigger Source Trigger Mode Trigger Coupling Trigger Holdoff Noise Rejection Trigger Type Trigger Output Connector
Page 90
RIGOL 5 To Trigger the Oscilloscope
5-2

Trigger Source

Press MENU Source in the trigger control area (TRIGGER) at the front panel to select the desired trigger source. Siganls input from CH1 to CH4, the [EXT TRIG] and [LOGIC D0-D15] connectors as well as the AC Line can all be used as trigger source.
Analog channel input:
Signals input fr om analog channels CH 1 to CH4 can all be used as the trigger sou rce . No matter whether the input of the channel selecte d is enabled, the channel can work normally.
External trigger input:
External trigger source can be used to trigger on the fifth channel when all the four channels are sampling data. The trigger signal (such as external clock and signal of the circuit to be tested) should be connected to EXT or EXT/5 trigger source via the [EXT TRIG] connector. You can set the trigger condition wit hin the r ange of trigger level (-0.8 V to +0.8 V).
Digital ch an n e l in p ut:
Signals input from digital channels D0 to D15 can all be used as the trigger source. No matter whether the input of the channel selected is enabled, the channel can work normally.
AC line:
The trigger signal is obtai n e d from the AC power i nput of the oscilloscope. This kind of signals can be used to display the relationship between the signal (suc h as illuminating device) and power (power supply devic e). For example, to stably trigger the waveform output from the transformer of a transformer substation, which is mainly used in related measurement of the power industry.
MSO4000/DS4000 User’s Guide
Page 91
5 To Trigger the Oscilloscope RIGOL
5-3

Trigger Mode

The trigger mode affects the way in which the oscilloscope searches for the trigger. The following is the schematic diagram of the acquisition memory. As shown in the figure below, the position of the trigger event in the acquisition memory is determined by the reference time point and the delay setting.
Pre-trigger Buffer Post-trigger Buffer
Acquisition memory
Pre-trigger/Delayed trigger:
This area is where you can view data acquired before/after the trigger event. The trigger position is usually at the hori zontal center of the screen. In full-screen display, seven-grid pre-trigger and delayed trigger information are displayed respectively. You can adjust the horizontal position through HORIZONTAL view more pre-trigger information, th rough which the signal information before/after the trigger (such as capture the glitch generated by the circuit and analyze the pre-trigger data to find out the reasons for glitch) can be obtained.
Press MODE in the trigger control area (TRIGGER) at the front panel or press MENU Sweep to select the desired trigger mode. The corresponding stat us light of the mode currently selected turns on.
Auto:
This trigger mode automatically triggers regardless of whether the trigger condition is met. A horizontal line is displayed when there is no input signal. In this mode, the oscilloscope operates by first filling the pre-trigger buffer. It starts searching for a trigger after the pre-trigger buffer is filled and continues to flow data through this buffer while it searches for the trigger. While searching for the trigger, the oscilloscope overflows the pre-trigger buff er and the first data put into the buffer is first pushed out (FIFO). When a trigger is found, the pre-trigger buffer contains
Trigger Event
POSITION to
MSO4000D/S4000 User’s Guide
Page 92
RIGOL 5 To Trigger the Oscilloscope
5-4
the data acquired just before the trigger. If no trigger is found, the oscilloscope will force a trigger. If the forced trigger is in v alid , the osc illoscope s till dis plays w av efo rm but th e wav eform will not be stable. If the forced trigger is valid, the oscilloscope displays a stable wav eform. This trigger mode is applicable to low-repetitive-rate signals and unknown signal levels. To display DC signals, you must use auto trigger mode. Note: When the horiz ontal time base is set to 50 ms/div or greater , thi s t rigge r m ode allows the absence of trigger signal.
Normal:
Displays a waveform when the trigger condition is met. Otherwise, the oscillosc ope holds the original waveform and waits for the next trigger. In this mode, the oscilloscope fills the pre-trigger buffer first and then search for a trigger while at the same time continues filling data. While searching for the trigger, the oscilloscope overflows the pre-trigger buff er and the first data put into the buffer is fir st pushed out (FIFO). When a trigger is found, the oscilloscope will fill the post-trigger buffer and display the a cquisitio n mem o ry. Use normal trigger mode for low-repetitive-rate signals or when auto trigger is not required. Note: In this mode, pressing FORCE can generate a forced trigger s ignal.
Single:
When this mode is selected, the backlight of the SINGLE key turns on. The oscilloscope waits for a trigger an d displa ys the wa ve form whe n the trigger c ondition is met and then stops. Note: In this mode, pressing FORCE can generate a forced trigger signal.
MSO4000/DS4000 User’s Guide
Page 93
5 To Trigger the Oscilloscope RIGOL
5-5

Trigger Coupling

The trigger coupling setting determines the type of signal components that will be transmitted to the trigger circuit. This is similar to the settings of “Channel
Coupling” but is specific to the triggered signal only.
DC: allow DC and AC components into the trigger path. AC: block all the DC components and attenuate signals lower than 8 Hz. LF reject: block the DC components and reject the low frequency components
(lower than 5 kHz).
HF reject: reject the high frequency components (higher than 50 kHz).
Press MENU Setting Coupling in the trigger control area (TRIGGER) at the front panel to select the desired coupling type (the defa u l t is DC).
Note: Trigger coupling is invalid in the video trigger mode.
MSO4000D/S4000 User’s Guide
Page 94
RIGOL 5 To Trigger the Oscilloscope
5-6
Trigger Position

Trigger Holdoff

Trigger holdoff can be used to stabilize the display of complex waveforms (such as pulse series). The holdoff ti me is the delay bef ore re -arming the trigger circuitry. The oscilloscope will not trigger until the holdoff time expires.
Holdoff Time
Press MENU Setting Holdoff in the trigger control area (TRIGGER) at the front panel and use to modify the holdoff time (the default is 100 ns) until the waveform triggers stably. The holdoff time has a range from 100 ns to 10 s.
MSO4000/DS4000 User’s Guide
Page 95
5 To Trigger the Oscilloscope RIGOL
5-7

Noise Rejection

Noise rejecti on minimizes the possibility of triggering on a noise signal reducing the trigger sensitivity. This decrease in sensitivity does require a greater amplitude waveform to trigger the oscilloscope.
Press MENU Setting Noise Reject in the trigger contro l area (TRIGGER) at the front panel to enable or disable noise rejection.
MSO4000D/S4000 User’s Guide
Page 96
RIGOL 5 To Trigger the Oscilloscope
5-8

Trigger Type

The MSO4000/DS4000 pro vides various tri gger functions, in cluding various serial bus triggers.
Edge Trigger Pulse Trigger Runt Trigger Nth Edge Trigger Slope Trigger Video Trigger Pattern Trigger RS232 Trigger I2C Trigger SPI Trigger CAN Trigger FlexRay Trigger USB Trigger
MSO4000/DS4000 User’s Guide
Page 97
5 To Trigger the Oscilloscope RIGOL
5-9

Edge Trigger

The edge trigger configures the instrument to capture data when the rising or falli ng edge of the input signal passes the threshold (trigger) voltage setting.
Trigger Type:
Press Type to select “Edge”. At this point, the trigge r setting inf ormation as shown i n the fi g ure below is displayed at the upper right corner of the screen.
Source Selection:
Press Source to select CH1 to CH4, EXT, EXT/5 or AC Line as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen. Note: Select the channel with the signal input as the trigger source to obtain stable trigger.
Edge Type:
Press Slope to select the kind of edge of the input signal on which the oscilloscope triggers. The current edge type is displayed at the upper right corner of the screen.
: trigger on t he rising edge of the input signal when the voltage level meets
the preset Trigger Level.
: trigger on the rising and falling edges of the input signal when the
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto, normal or single. The correspondin g status light o f the current tri gger mode turns on.
Trigger Setting:
Press Setting to set the trigger p aramete r s ( trigger coupling, trigg e r hol do ff and noise rejection) under this trigger type.
Trigger Level:
Use TRIGGER trigger mark “ the knob, while at the same time, the trigger level value (such as
: trigger on the f alling e dge of the inp ut signal when the v oltage level meets
the preset Trigger Level.
voltage level meets the preset Trigger Level.
LEVEL to modify the level. An orange trigger level line and the
” appear on the screen and move up and down with the rotation of
) at
MSO4000D/S4000 User’s Guide
Page 98
RIGOL 5 To Trigger the Oscilloscope
5-10
the lower left corner of the screen also changes accordingly. The trigger level line and the trigger mark will disappear in about 2 seconds if no input (rotation) is detected.
MSO4000/DS4000 User’s Guide
Page 99
5 To Trigger the Oscilloscope RIGOL
5-11

Pulse Trigger

Trigger on the positive or negative pulse with a specified width.
Trigger Type:
Press Type to select “Pulse”. At this point, the trigger sett ing information as s hown in the fi g ure below is displayed at the upper right corner of the screen.
Source Selection:
Press Source to select CH1 to CH4, EXT or D0-D15 as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen. Note: Select the channel with the signal input as the trigger source to obtain stable trigger.
Pulse Condition: Press When to select the desired pulse condition.
: trigger when the positive pulse width of the input signal is greater than
the specified Pulse Width Setting.
: trigger when the positive pulse wi dth of the input si gnal is lower than the
specified Pulse Width Setting.
: trigger when the positive pulse width of the input signal is greater than the specified Lower Limit of Pulse Width and lowe r than the Upper Limit of Pul se Width.
: trigger when the negativ e pulse wi dth of the in put s ignal is great er tha n the sp ecified Pulse Width Setting.
: trigger when the negative pulse width of the input signal is lower than
the specified Pulse Width Setting.
: trigger when the negativ e pulse wi dth of the in put s ignal is great er tha n the specified Lower Limit of Pulse Width and lower t han the Upper Limit of Pul se Width.
MSO4000D/S4000 User’s Guide
Page 100
RIGOL 5 To Trigger the Oscilloscope
5-12
Positive
A
B
Pulse Width Setting:
Positive pulse width is defined as the time difference between the two crossing points of the trigger level and positive pulse as shown in the figure below.
Trigger level
pulse width
When the Pulse Condition is set to
, , or , press Setting and use to input the desired value. The range av ailable is fr om 4 ns to 4 s.
When the Pulse Condition is set to or , press Upper Limit and
Lower Limit and use to input the desired values respectively. The range of
the upper limit is from 12 ns to 4 s. The range of the lower limit is from 4 ns to
3.99 s. Note that the lowe r limit of the pulse width m ust be lowe r than the upper limit.
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto , normal or single. The corresponding status light of the current tri gger mode tu rns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger coupli n g, tr i gger holdoff and noise rejection) under this trigger type.
Trigger Level:
Use TRIGGER LEVEL to modify the l e ve l. For details, please refer to the description of “
Trigger Level” on page 5-9.
MSO4000/DS4000 User’s Guide
Loading...