Rigol MSO2000A Series, MSO2102A, DS2000A Series, MSO2102A-S, 100 Quick Manual

...
Page 1
RIGOL
Quick Guide
Lambda Photometrics Limited
Lambda House Batford Mill Harpenden Herts AL5 5BZ United Kingdom
E: info@lambdaphoto.co.uk W: www.lambdaphoto.co.uk T: +44 (0)1582 764334
F: +44 (0)1582 712084
MSO2000A/DS2000A Series
Digital Oscilloscope
May 2019
RIGOL(SUZHOU) TECHNOLOGIES INC.
Page 2
Page 3
RIGOL
I
Guaranty and Declaration
Copyright
© 2013 RIGOL (SUZHOU) TECHNOLOGIES INC.All Rights Reserved.
Trademark Information
RIGOL is a registered trademark of RIGOL (SUZHOU) TECHNOLOGIES INC.
Publication Number
QGA18104-1110
Notices
z RIGOL products are covered by P.R.C. and foreign patents, issued and pending. z RIGOL reserves the right to modify or change parts of or all the specifications
and pricing policies at the company’s sole decision.
z Information in this publication replaces all previously released materials. z Information in this publication is subject to change without notice. z RIGOLshall not be liable for either incidental or consequential losses in
connection with the furnishing, use, or performance of this manual, as well as any information contained.
z Any part of this document is forbidden to be copied, photocopied, or rearranged
without prior written approval of RIGOL.
Product Certification
RIGOL guarantees that this product conforms to the national and industrial standards in China as well as the ISO9001:2015 standard and the ISO14001:2015 standard. Other international standard conformance certifications are in progress.
Contact Us
If you have any problem or requirement when using our products or this manual, please contact RIGOL. E-mail: service@rigol.com Website: www.rigol.com
MSO2000A/DS2000A Quick Guide
Page 4
RIGOL
II
1. Only the exclusive power cord
authorized for use within the local
2. Ensure that the instrument is safely
3.
4.
5. Use proper overvoltage protection.
6.
7.
8. Use the proper fuse.
9.
10. Do not operate the instrument with
11.
12.
13.
14. Keep instrument surfaces clean and
15.
16.
WARNING
Indicates a potentially hazardous situation or practice which, if not avoided, will result in serious injury or death.
CAUTION
Indicates a potentially hazardous situation or practice which, if not avoided, could result in damage to the product or loss of important data.
DANGER
It calls attention to an operation, if not correctly performed, could result in injury or hazard immediately.
WARNING
It calls attention to an operation, if not correctly performed, could result inpotential injury or hazard.
CAUTION
It calls attention to an operation, if not correctly performed, could result indamage to the product or other devices connected to the product.
Hazardous Voltage
Safety Warning
Protective Earth Terminal
Chassis Ground
Test Ground
Safety Requirement
General Safety Summary
designed for the instrument and
country could be used.
Avoid circuit or wire exposure.
suspected failures.
Provide adequate ventilation.
grounded.
Connect the probe correctly. Observe all terminal ratings.
Do not operate without covers. Do not insert objects into the air
outlet.
Do not operate in wet conditions. Do not operate in an explosive
atmosphere.
dry.
Prevent electrostatic impact. Handle with caution.
Safety Notices and Symbols
Safety Notices in this Manual:
Safety Terms on the Product:
Safety Symbols on the Product:
MSO2000A/DS2000A Quick Guide
Page 5
RIGOL
III
WARNING
This oscilloscope can only be used for measurements within its specified measurement categories.
WARNING
Inadequate ventilation may cause damage the instrument during operation and inspect the intake and fan regularly.
Measurement Category
Measurement Category
MSO2000A/DS2000A series digital oscilloscopes can make measurements in Measurement Category I.
Measurement Category Definitions
Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable; for that reason, the transient withstand capability of the equipment is made known to the user.
Measurement category II is for measurements performed on circuits directly connected to the low voltage installation. Examples are measurements on household appliances, portable tools and similar equipment.
Measurement category III is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit-breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example. Stationary motors with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of the low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
Ventilation Requirement
This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
temperature increase which would
. So please keep the instrument well ventilated
MSO2000A/DS2000A Quick Guide
Page 6
RIGOL
IV
WARNING
T
avoid short circuit inside the instrument or electric shock, please do not
operate in humid environment.
WARNING
Make sure that no overvoltage (such as that caused by thunder
) can reach the product, or else the operator might expose to danger of electric shock.
Working Environment
Temperature
Operating: 0ć to +50ć Non-operating: -40ć to +70ć
Humidity
0ć to +30ć: ≤95ˁ relative humidity +30ć to +40ć: ≤75ˁ relative humidity +40ć to +50ć: ≤45ˁ relative humility
o
Altitude
Operating: less than 3 km Non-operating: less than 15 km
Installation (overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
bolt
Installation (overvoltage) Category Definitions
Installation (overvoltage) category I refers to signal level which is applicable to equipment measurement terminals connected to the source circuit. In these terminals, precautions are done to limit the transient voltage to the corresponding low level.
Installation (overvoltage) category II refers to the local power distribution level which is applicable to equipment connected to the AC line(AC power).
Pollution Degree
Degree 2
Pollution Degree Definitions
Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. For example: a clean room or air-conditioned office environment.
MSO2000A/DS2000A Quick Guide
Page 7
RIGOL
V
CAUTION
To avoid damage to the instrument, do
WARNING
To avoid short
circuit resulting from moisture or personal injuries, ensure that the instrument is completely dry before connecting it to the power supply.
Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. For example: general indoor environment. Pollution degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due to condensation which is expected. For example: Sheltered outdoor environment. Pollution degree 4: Pollution that generates persistent conductivity through conductive dust, rain, or snow. For example: outdoorlocations.
Safety Class
Class 1 – Grounded Product
General Care and Cleaning
Care:
Do not store or leave the instrument in where the instrument will be exposed to direct sunlight for long periods of time.
Cleaning:
Clean the instrument regularly according to its operating conditions.
1. Disconnect the instrument from all power sources.
2. Clean the external surfaces of the instrument with a soft cloth dampened with mild detergent or water. Avoid having any water or other objects into the chassis via the heat dissipation hole. When cleaning the LCD, take care to avoid scarifying it.
not expose it to caustic liquids.
-
MSO2000A/DS2000A Quick Guide
Page 8
RIGOL
VI
Label
Knob
HORIZONTAL SCALE
Horizontal Scale Knob
HORIZONTAL POSITION
Horizontal Position Knob
VERTICAL SCALE
Vertical Scale Knob
VERTICAL POSITION
Vertical Position Knob
TRIGGER LEVEL
Trigger Level Knob
Environmental Considerations
The following symbol indicates that this product complies with the WEEE Directive 2002/96/EC.
Product End-of-Life Handling
The equipment may contain substances that could be harmful to the environment or human health. To avoid the release of such substances into the environment and avoid harm to human health, we recommend you to recycle this product appropriately to ensure that most materials are reused or recycled properly. Please contact your local authorities for disposal or recycling information.
Document Overview
This manual is used to guide users to quickly get familiar with the front panel, rear panel, user interface and basic operation method of MSO2000A/DS2000A series digital oscilloscope. You can download the newest version of the manual from RIGOLofficial website (www.rigol.com).
Format Conventions in this Manual:
1. Button
The front panel keys are denoted by the format of "Button Name (Bold) + Text Box". For example,Utilitydenotes the "Utility"key.
2. Menu
The menu softkeys are denoted by the format of "Menu Word (Bold) + Character Shading". For example,Systemdenotes the "System"menu under
Utility.
3. Operation Step
The next step of operation is denoted by an arrow "Æ". For example,UtilityÆSystemdenotesthat first press Utility on the front panel and then press System.
4. Knob
MSO2000A/DS2000A Quick Guide
Page 9
RIGOL
VII
Model
Analog Bandwidth
Number of Analog Channels
Number of Source Channels
MSO2102A/DS2102A
100 MHz
2
--
MSO2102A-S
100 MHz
2
2
MSO2202A/DS2202A
200 MHz
2
--
MSO2202A-S
200 MHz
2
2
MSO2302A/DS2302A
300 MHz
2
--
MSO2302A-S
300 MHz
2
2
Content Conventions in this Manual:
MSO2000A/DS2000A series includes the following models. This manual takes MSO2302A-S as an example to illustrate the functions and operation methods of MSO2000A/DS2000A series.
MSO2000A/DS2000A Quick Guide
Page 10
RIGOL
VIII
Contents
Guaranty and Declaration ......................................................................... I
Safety Requirement ................................................................................. II
General Safety Summary........................................................................... II
Safety Notices and Symbols ...................................................................... II
Measurement Category ............................................................................ III
Ventilation Requirement ........................................................................... III
Working Environment .............................................................................. IV
General Care and Cleaning ........................................................................ V
Environmental Considerations ................................................................... VI
Document Overview ............................................................................... VI
Quick Start ............................................................................................... 1
General Inspection .................................................................................... 1
Appearance and Dimensions ...................................................................... 2
To Prepare for Operation ............................................................................ 3
To Adjust the Supporting Legs ............................................................. 3
To Connect to AC Power Supply ........................................................... 3
Turn-on Checkout ............................................................................... 4
To Connect the Probe .......................................................................... 4
Function Inspection ............................................................................ 6
Probe Compensation ........................................................................... 7
Front Panel Overview ................................................................................ 8
Rear Panel Overview ................................................................................. 9
Front Panel Function Overview ................................................................. 11
VERTICAL ........................................................................................ 11
HORIZONTAL ................................................................................... 12
TRIGGER ......................................................................................... 13
CLEAR ............................................................................................. 14
RUN/STOP ....................................................................................... 14
SINGLE ............................................................................................ 14
AUTO .............................................................................................. 14
Knob ............................................................................................... 15
Navigation Knob ............................................................................... 15
MENU .............................................................................................. 16
Signal Source ................................................................................... 16
Record ............................................................................................. 17
Print ................................................................................................ 17
Logic Analyzer .................................................................................. 18
User Interface ......................................................................................... 19
To Use the Security Lock .......................................................................... 24
To Use the Built-in Help System ................................................................ 25
Troubleshooting ..................................................................................... 26
MSO2000A/DS2000A Quick Guide
Page 11
1
Quick Start
General Inspection
1. Inspect the packaging
If the packaging has been damaged, do not dispose the damaged packaging or cushioning materials until the shipment has been checked for completeness and has passed both electrical and mechanical tests.
The consigner or carrier shall be liable for the damage to the instrumentresulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument.
2. Inspect the instrument
RIGOL
In case of any mechanical damage, missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative.
3. Check the accessories
Please check the accessories according to the packing lists. If the accessories are damaged or incomplete, please contact yourRIGOL sales representative.
MSO2000A/DS2000A Quick Guide
Page 12
RIGOL
2
361.6
179.6
Appearance and Dimensions
Figure 1 Front View Unit: mm
Figure 2 TopView Unit: mm
MSO2000A/DS2000A Quick Guide
Page 13
RIGOL
3
Power Cord Connector
To Prepare for Operation
To Adjust the Supporting Legs
Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as betteroperation and observation.
Figure 3 To Adjust the Supporting Legs
To Connect to AC Power Supply
This oscilloscope can accept100-240V, 45-440 Hz AC power supply. Please use the power cord supplied with the accessories to connect the oscilloscope to the power supplyas shown in the figure below. Whenthe oscilloscope is powered onand the Power key at the lower-left corner of the front panel is blinking.
MSO2000A/DS2000A Quick Guide
Figure 4 To Connect to AC Power Supply
Page 14
RIGOL
4
Turn-on Checkout
When the oscilloscope is connected to power, press the Power key at the lower-left corner of the front panel to start the oscilloscope. During the start-up process, the oscilloscope performs a series of self-test items and you can hear the sound of relay switching. After the self-test, the welcome screen is displayed. The instrument is installed with the trial versions of the options before leaving factory and the remaining time is about 2,000 minutes. The "Current Options"dialog box will be displayed if your instrument has currently installedthe trial versions of options. From this dialog box you can view the types, names, versions and the remaining time of the options currently installed.
To Connect the Probe
RIGOLprovides passive probes for DS2000A, and provides passive probes and logic probes for MSO2000A. For the model of the probes, please refer to
MSO2000A&DS2000A Series Datasheet
probes, please refer to corresponding Probe User’s Guide.
Connect the Passive Probe:
1. Connect the BNC terminal of the probe to an analog channel inputof the oscilloscope onthe front panel.
2. Connect the ground alligator clip or spring of the probeto the circuit ground terminal and then connect the probe tip to the circuit point under test.
. For detailed technical information of the
Figure 5 To Connect the Passive Probe
MSO2000A/DS2000A Quick Guide
Page 15
RIGOL
5
After you connect the passive probe, check the probe function and probe compensation adjustment before making measurements. For detailed procedures, refer to "Function Inspection" and "Probe Compensation" sections in this manual.
Connect the Logic Probe:
1. Connect the single-wire terminal of the logic probe to the [LOGIC D0-D15] digital channel terminalonthe front panel of MSO2000A in the correct direction.
2. Connect the other terminal of the logic probe tothe signal under test. RIGOL provides MSO2000A with the standard RPL2316 logic probe. To cater to different application scenarios, RPL2316 provides three connection methods for the signal under test. For details, please refer to the
Figure 6 To Connect the Logic Probe
RPL2316 Logic Probe User’s Guide
.
Note: The digital channel input terminal does not support hot plugging. Please do
not insert or pull out the logic probe when the instrument is in power-on state.
MSO2000A/DS2000A Quick Guide
Page 16
RIGOL
6
WARNING
To avoid electric shock during the use of probe, please make sure that the insulated wire of the probe is in good condition
o not touch the metallic
part of the probe when the probe is connected to high voltage source.
Compensation Signal Output Terminal
Ground Terminal
Function Inspection
1. Press StorageÆDefault to restore the oscilloscope to its default configuration.
2. Connect the groundalligator clip of the probe to the "GroundTerminal"as shown in the figure below.
3. Use the probe to connect the input terminal of CH1 and the "Compensation Signal Output Terminal" of the oscilloscope.
Figure 7 To Use the Compensation Signal
4. Set the attenuationon the probe to 10X. ThenpressAUTO.
5. Observe the waveform on the display. In normal condition, the display should be a square waveform as shown in the figure below:
Figure 8 Square Waveform
6. Use the same method to test the other channels. If the square waveforms actually shown do not match that in the figure above, please perform "Probe Compensation".
. D
MSO2000A/DS2000A Quick Guide
Page 17
RIGOL
7
Tip
The signal output from the probe compensation connector can only be used for probe compensation adjustment and cannot be used for calibration.
Over compensatedPerfectly compensatedUnder compensated
Probe Compensation
When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy or errors. The probe compensation proceduresare as follows:
1. Perform Step 1, 2, 3 and 4 specified in"Function Inspection".
2. Check the displayed waveforms and compare them with the waveforms in the following figure.
Figure 9 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment hole on the probe until the displayed waveform is displayed as the "Perfectlycompensated"in the figure above.
MSO2000A/DS2000A Quick Guide
Page 18
RIGOL
8
No.
Description
No.
Description
1 Measurement Menu Softkeys
13
Power Key
2 LCD 14
USB HOST Interface
3 Logic Analysis C
15
Digital Channel Input Interface
4 Multifunction
16
HORIZONTAL Control Area
5 Function Menu
17
Function Menu Softkeys
6 Signal
18
VERTICAL Control Area
7 Navigation Knob
19
Analog Channel Input Area
8 CLEAR
20
Waveform
9 AUTO
21
TRIGGER Control Area
10
RUN/STOP
22
EXT TRIG Input Terminal
11
SINGLE
23
Probe Compensation Signal Output
Terminal/ Ground Terminal
12
Help/Print
--
--
13 14 15 16 17 18 19 20 21 22 23
Front Panel Overview
1 2
3 4 5 6 7 8 9 10 11 12
Table 1 Front Panel Description
ontrol Key
Knob
Keys
Source
[2]
[1]
Note
Note
: Only applicable to MSO2000A and MSO2000A-S models.
[2]
: Only applicable to the MSO2000A-S model.
Figure 10 Front Panel Overview
[1]
[1]
Record/Playback Control Keys
MSO2000A/DS2000A Quick Guide
Page 19
9
Rear Panel Overview
ˊ
ˊ
ˊ
ˊ
1
RIGOL
2 3
4 5 6 7 8
Figure 11 Rear Panel Overview
Handle
1ˊ
Pull up the handle vertically for easy carrying of the instrument. When you do not need the handle, press it down.
LAN
2
Connect the instrument to the network via this interface for remote control. This oscilloscope conforms to the LXI CORE 2011 DEVICE class instrument standards and can quickly build test system with other instruments.
USB DEVICE
3
PictBridge printer or PC can be connected via this interface to print waveform data or control the instrument using PC software.
Fuse
4
If a new fuse is required, please use the specified fuse (250V, T2A). The operationmethod is as follows:
MSO2000A/DS2000A Quick Guide
Page 20
RIGOL
10
ˊ
ˊ
ˊ
ˊ
(1) Turn off the instrument and remove the power cord. (2) Insert a small straight screwdriverinto the slot at the power cord
connectorand pry out the fuse holder gently.
(3) Take out the fuse and replace it with a specified fuse. Then,install the fuse
holder.
AC PowerCord Connector
5ˊ
AC power input terminal. The power requirements of this oscilloscope are100-240 V, 45-440 Hz. Use the power cord provided with the accessories to connect the instrument to AC power. Then, you can press the Power key onthe front panel to start the instrument.
Lock Hole
6
You can lock the instrument to a fixed location by using the security lock (please purchaseit yourself) via the lock hole.
Trigger Out
7
The oscilloscope can output a signal that can reflect the current capture rate of the oscilloscope at each trigger via this interface. Connect the signal to awaveform display device and measure the frequency of the signal. The measurement result is the same asthe current capture rate. The instrument can also outputa signalwhen a failed waveform is detected during the pass/fail test.
Source Output
8
When Source1 or Source2 is enabled, thesignal currently set can beoutputthrough the [Source1]or[Source2]connectoron the rear panel to the analog input terminal of the oscilloscope or external devices connected to the connector. This function is only available for the MSO2000A-S model.
MSO2000A/DS2000A Quick Guide
Page 21
RIGOL
11
CH1, CH2: analog input channels. The 2 channels are marked by different colors which are also used to mark both the corresponding waveforms on the screen and the channel input connectors. corresponding channel menu and press again to turn off the channe
Front Panel Function Overview
VERTICAL
Press any key to open the
l.
MATH: press this key to open the math operation menu under which add, subtract, multiply, divide, FFT, logic and advanced operations are provided.
REF: press this key to enable the reference waveform function to compare the waveform actually tested with the reference waveform.
Vertical POSITION
waveform. Turn clockwise to increase the position and turn counterclockwise to decrease.During the modification, the waveform would move up and down and
theposition message (e.g. )at the lower-left corner of the screen would
change accordingly. Press down this knob to quickly reset the vertical position to zero.
VERTICAL SCALE
clockwise to decrease the scale and turn counterclockwise to increase. During the modification, the display amplitude of the waveform would enlarge or reduce but the
: modify the vertical position of the current channel
: modify the vertical scale of the current channel. Turn
actual amplitude remains unchanged.The scale information(e.g. ) at the lower side of the screen would change accordingly. Press down this knob to quickly switch the vertical scale adjustment modes between "Coarse" and "Fine".
Decode1, Decode2: decoding function keys. Press the corresponding key to open the decoding function menu. MSO2000A/DS2000A support parallel decoding and protocol decoding.
MSO2000A/DS2000A Quick Guide
Page 22
RIGOL
12
MENU: press this key to open the horizontal control menu under which to turn on or off the delayed sweep function, switch between different time base modes, switch between
"Coarse" and "Fine"adjustment of scale as well as modify the horizontal reference setting.
HORIZONTAL SCALE: modify the horizontal time base. Turn clockwise to reduce the time base and turn counterclockwise to
the time base. During the modification, waveforms of all the channels will be displayed in expanded or compressed mode and the time base message
at the upper side of
the screen would change accordingly. Press
the
delayed sweep state.
HORI
rigger point
would
when you turn the knob.
During
and
the trigger
right corner of the
screen would change accordingly. Press
the trigger
position
HORIZONTAL
increase
(e.g. )
down this knob to quickly switch to
ZONTAL POSITION: modify the horizontalposition. The t
move left or right relative to the center of the screen
the modification, waveforms of all the channels would moveleft or right
position message (e.g. ) at the upper-
down this knob to quickly reset
(or the delayed sweep position).
MSO2000A/DS2000A Quick Guide
Page 23
13
TRIGGER
MODE: press this key to switch the trigger mode to A
state
would be
illuminated.
TRIGGER
T counterclockwise to reduce the level.
the
modification, the trigger level line
up
and down and the value in the trigger level message box
at the
lower
ress down the knob to
quickly reset the trigger level to zero
MENU: press this key to open the trigger operation menu. This oscilloscope provides
FORCE
trigger modes, press this key to generate a trigger
signal forcefully.
RIGOL
uto, Normal or Singleand thecorresponding
backlightof the current trigger mode
LEVEL: modify the trigger level.
urn clockwise to increase the level and turn
During
would move
(e.g. )
-left corner of the screen would change accordingly. P point.
various trigger types.
: in Normal and Single
MSO2000A/DS2000A Quick Guide
Page 24
RIGOL
14
Press this key to clear all the waveforms on the screen. If the oscilloscope is in illuminated displayed.
Press this key to set the state of the oscilloscope to "RUN In In the "STOP" state, the key is illuminated in red.
Press this key to set the trigger mode to "Single", the key is illuminated in orange FORCEto generate a trigger signal immediately.
Press this key to enable the waveform auto setting function.
he oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input signal to realize optimum waveform display.
auto setting requires that the frequency of the signal under test should be no lower than detect any signal
be displayed after pressing this key and the quick parameter measurement menu might not be displayed.
CLEAR
the "RUN" state (the key is
in yellow), new waveforms will still be
RUN/STOP
SINGLE
AUTO
" or "STOP".
the "RUN" state, the key is illuminated in yellow.
. In single trigger mode, press
T
Note:
MSO2000A/DS2000A Quick Guide
25Hz. If the parameter exceed the limit, "Can’t
!"will
Page 25
15
Knob
Adjust waveform brightness: In non
turn this
knob to adjust the brightness of waveform display.
he
adj
urn clockwise to increase the brightness and counterclockwise to reduce. P Y
and use
the knob to adjust the waveform brightness.
Multifunction(the backlight goes on during operation):
In menu operation, press any submenus under this menu and then press down submenu.
n addition,
for
menu softkey and then press the knob you
This knob provides quick Adjust/Locate function for numerical parameters with relatively large settable (counter value. The inner knob is used for fine adjustment and the outer knob for coarse adjust
For example, this knob can be used to quickly locate the waveform frame ("Current Frame
. Similar menus
includetrigger holdoff, pulse width setting, slope time etc.
RIGOL
-menu-operation mode (menu is hidden), T
ustable range is from 0% to 100%. T
ress down this knob to reset the brightness to 50%.
ou can also press DisplayÆWaveIntensity
menu softkey and turn the knob to select the
the knob to select the current
It can also be used to modify parameters and input filename.I
MSO2000A-S, in the source interface, press the corresponding
; the numeric keyboard will be displayed on the screen and
can input the desired parameter value and unit directly using this knob.
Navigation Knob
range. Turn clockwise
clockwise) to increase (reduce) the
ment.
" menu) to be played back in the waveform playback function
MSO2000A/DS2000A Quick Guide
Page 26
RIGOL
16
Measure: press this key to open the measurement setting menu. You can set the measurement setting, all measure, statistic function etc. Press
at the left of the screen
to open the measurement menus of 29 waveform parameters. Then, press down the
key measurement and the
measurement result will be displayed at the bottom of the screen.
Acquire: press this key to enter the sample setting menu to set the acquisition mode, memory depth and antialiasin
S
he storable file types include traces, waveforms, setups, picture and CSV. The picture can be stored in management are also supported.
C
he oscilloscope
provides
mode is
only available
D
pe, persistence time, wave intensity, grid type, grid brightness and menu display time of the waveform.
U
setting menu to set the system Besides,
record and
print setting) are also supported.
Press this key to enter the source setting interface.You can enable or disable the output of the
on the rear panel, frequency, amplitude, offset and phase).
MENU
MENU
corresponding menu softkey to quickly realize one-
g function of the oscilloscope.
torage: press this key to enter the file store and recall interface. T
bmp, png, jpeg and tiff formats. Internal and external storage as well as disk
ursor: press this key to enter the cursor measurement menu. T
four cursor modes: manual, track, auto and X-Y. Note:X-Y cursor
forthe X-Ytime base mode.
isplay: press this key to enter the display setting menu to set the display ty
tility: press this key to enter the systemutility function
-related functions or parameters, such as the I/O, sound and language. some advanced functions (such as the pass/fail test, waveform
Signal Source
setthe output signal parameters (such as the
Note:This function is only available for MSO2000A-S.
MSO2000A/DS2000A Quick Guide
[Source1]or[Source2] connector
Page 27
RIGOL
17
Press this key to execute the print function or save the screen in the USB storage device. connected and the printer is in idle state, pressing this key will execute th
f no printer but a USB storage device is currently connected, pressing this key will save the screen to the USB storage device in
You can also save
the screen in
tiff)
by
then
press
hen printer and USB storage device are connected at the same time, the printer enjoys higher priority.
Stop Play/Pause Record
Record
Record: press this key to start recording the waveform. At this moment, the red
backlight of the key will start flashing. Besides, when the record constant on (Open) is enabled, the backlight will also keep flashing.
Play/Pause: in the stop or pause state, press this key to play back the waveform and press again to pausethe play. The backlight is illuminated in yellow.
Stop:press this key to stop the waveform being recorded or being played back. The backlight is illuminated in orange.
Print
If a PictBridge printer is currently
e print function. I
".png" format by default.
the specified picture format (bmp, png, jpeg and
pressing the Storagetoset the storage type as picture, and
ing the Pic Typeto select the desired format. W
MSO2000A/DS2000A Quick Guide
Page 28
RIGOL
18
Press this key to open the logic analyzer control menu. You can turn on or off any channel or channel group, modify the display size of the digital channel, modify the logic
of the digital channel as well as group the 16 digital channels and display them as a bus. You can also set a label for each digital channel.
Note:
MSO2000A-S.
Logic Analyzer
threshold
This function is only applicable to MSO2000A and
MSO2000A/DS2000A Quick Guide
Page 29
RIGOL
19
12 13 14 15 16 17 18 19 20
1 2 3 4 5 6 7 8 9 10 11
User Interface
MSO2000A/DS2000A provides 8.0 inch, WVGA (800*480) 160,000 color TFT LCD. What is worth mentioning is that the 14-grid ultra-wide screen enables you to view "longer" waveform.
Figure 12User Interface
1. Auto Measurement Items
Provide 16 horizontal (HORIZONTAL) and 13 vertical (VERTICAL) measurement parameters. Press the softkey at the left of the screen to activate the corresponding measurement item. Press MENU continuously to switch between the horizontal and vertical parameters.
MSO2000A/DS2000A Quick Guide
Page 30
RIGOL
20
2. Digital Channel Label/Waveform
The logic high level of the digital waveform is displayed in blue and the logic low level in green (correspond to the color of the channel label). Its edge is displayed in white. The label and waveform of the digital channel currently selected are displayed in red. Note:This function is onlyapplicable to MSO2000A and MSO2000A-S.
3. Status
Available states include RUN, STOP, T’D (triggered), WAIT and AUTO.
4. Horizontal Time Base
z Represent the time per grid on the horizontal axis on the screen. z UseHORIZONTAL SCALE
to modify this parameter. The range available is from 1.000 ns to 1.000 ks (for 200 MHz bandwidth oscilloscope, the range available is 2.000 ns to 1.000 ks; for 100 MHz and 70 MHz bandwidth oscilloscope, the range available is 5.000 ns to 1.000 ks).
5. Sample Rate/Memory Depth
z Display the current sample rate and memory depth of the oscilloscope. z Use HORIZONTAL SCALE
to modify this parameter.
6. Waveform Memory
Provide the schematic diagram of the memory position of the waveform currently on the screen.
waveform on the screen
7. Trigger Position
Display the trigger position of the waveform in the waveform memory and on the screen.
8. HorizontalPosition
Use HORIZONTAL POSITION
to modify this parameter. Press down the
knob to automatically set the parameter to zero.
MSO2000A/DS2000A Quick Guide
Page 31
21
9. Trigger Type
Display the currently selected trigger type and trigger condition setting. Different labels are displayed when different trigger types are selected. For example: represents triggering on the rising edge in "Edge" trigger.
10. Trigger Source
Display the trigger source currently selected (CH1, CH2, EXT, EXT/5,AC Line or D0-D15). Different labels are displayed when different trigger sources are selected and the color of the trigger parameter area will change accordingly. For example: denotes that CH1 is selected as the trigger source.Note:EXT/5 is onlyapplicable to MSO2000A and MSO2000A-S.
11. Trigger Level
z When the trigger source is set to CH1 or CH2, the trigger level label is
displayed at the right of the screen and the trigger level value is displayed at
RIGOL
the upper-right corner of the screen.When using TRIGGER LEVEL modify the trigger level, the trigger level value will change with the up and down of .
z When the trigger source is set to EXT or EXT/5, the trigger level value is
displayed at the upper-right corner of the screen. No trigger level label is displayed.
z When the trigger source is set to AC Line, no trigger level value and trigger
level label is displayed.
z When the trigger source is set to D0 to D15, the trigger threshold is
displayed at the upper-right corner of the screen. No trigger level label is displayed.
z In Runt trigger,Slope trigger and Windows trigger, two trigger level labels
( and ) are displayed.
12. CH1 Vertical Scale
z Display the voltage value per grid of CH1 waveform vertically. z Use VIRTICAL SCALE
of CH1 to modify this parameter.
to
z The following labels will be displayed according to the current channel
setting: channel coupling (e.g. ), input impedance (e.g. ) and bandwidth limit (e.g. ).
MSO2000A/DS2000A Quick Guide
Page 32
RIGOL
22
13. Analog Channel Label/Waveform
Different channels are marked with different colors and the colors of the channel label and waveform are the same.
14. CH2 Vertical Scale
z Displaythe voltage value per grid of CH2 waveform vertically. z Use VIRTICAL SCALE
of CH2 to modify this parameter.
z The following labels will be displayed according to the current channel
setting: channel coupling (e.g. ), input impedance (e.g. ) and bandwidth limit (e.g. ).
15. Digital Channel Status Area
Display the current status of the 16 digital channels (D0 to D15 from right to left). The digital channels currently turned on are displayed in green and the digital channel currently selected is displayed in red. The digital channels turned off are displayed in grey in the digital channel area.Note:This function is onlyapplicable to MSO2000A and MSO2000A-S.
16. Message Box
Display the prompt messages.
17. Source1 Waveform
z Display the type of waveform currently set for Source1. z When the impedance of source 1 is set to 50 Ω, will be displayed at the
right of the Source1 waveform.
z When the modulation of source 1 is enabled, will be displayed at the
right of the Source1 waveform.
z Only available for MSO2000A-S.
18. Source2 Waveform
z Display the type of waveform currently set for Source2. z When the impedance of source 2 is set to 50 Ω, will be displayed at the
right of the Source2 waveform.
z When the modulation of source 2 is enabled, will be displayed at the
right of the Source2 waveform.
z Only available for MSO2000A-S.
MSO2000A/DS2000A Quick Guide
Page 33
23
19. Notification Area
Denote that onthe front panel can be used to select the parameter items.
parameter
selection is valid.
Denote that can be used to modify the parameter values. The backlight of turns on when the parameter input is valid.
Denote that can be used to modify the parameter values and press
using the
pop
the
parameter input is valid.
Denote that you can use the "Navigation Knob" to quickly adjust/locate parameters.
Denote that you can use to adjust the parameter and then press down
is constant
on.
Denote that the current menu has several options.
Denote that the current menu has a lower level menu.
Press this key to return to the previous menu.
Note:The following direction keys might appear in the grid at the lower-left
corner of the menu bar:
Denote that you can open the next page menu.
Denote that you can open the previous page menu.
Display the system time, sound icon and USB storage device icon.
z System Time: displayed in "hh:mm (hour:minute)" format. When
printing or storing the waveform, the output file will contain this time message. Press UtilityÆSystemÆSystem TimeÆSystem Timeto set through the following format: yyyy-mm-ddhh:mm:ss (year-month-datehour:minute:second)
z Sound Icon: when the sound is enabled, will be displayed.
PressUtilityÆSoundto enable or disable the sound.
z USB Storage Device Icon: when a USB storage device is detected, will
be displayed.
20. Operation MENU
Press any softkey to activate the corresponding menu. The following symbols might be displayed in the menu:
RIGOL
The backlight of turns on when the
to input the desired paramrter values directly
-up numeric keyboard. The backlight of turns on when
to select the parameter. The backlight of
MSO2000A/DS2000A Quick Guide
Page 34
RIGOL
24
Security Lock Hole
To Use the Security Lock
If needed, you can use the security lock (please buy it yourself) to lock the oscilloscope to a fixed location. The method is as follows, align the lock with the lock hole and plug it into the lock hole vertically, turn the key clockwise to lock the oscilloscope and then pull the key out.
Figure 13To Use the Security Lock
Note:Please do not insert other articles into the security lock hole to avoid damaging
the instrument.
MSO2000A/DS2000A Quick Guide
Page 35
RIGOL
25
To Use the Built-in Help System
The help system of this oscilloscope provides instructions for all the function keys (including the menu keys) onthe front panel. Press Help to open the help interface and press again to close the interface. The help interface mainly consists of two parts. The left is "Help Options" and you can use "Button" or "Index"mode to select. The right is "Help Display Area".
Help Options Help Display Area
Figure 14Help Information
Button:
Default mode. In this mode, you can press the button(except the power key and the function menu page up/down key / at the right of the screen) or rotate themultifunction knob onthe front panel to select the button name to get the corresponding help information in the "Help Display Area". In addition, you can get the help information of the navigation knob by rotating the navigation knobdirectly or using to select "WaveSearch". Rotate to select "To Index" and then press the knob to switch to Index mode.
Index:
In this mode, use to select the item that needs to get help (for example, "BW"). The item currently selected is displayed in brown. Press the knob to get the corresponding help information in the "Help Display Area".Use to select "To Button" and then press the knob to switch to Button mode.
MSO2000A/DS2000A Quick Guide
Page 36
RIGOL
26
Troubleshooting
The commonly encountered failures and their solutions are listed below. When you encounter those failures, please solve them following the corresponding steps. If the problem remains still, please contact RIGOLand provide your device information (acquisition method: UtilityÆSystemÆSystemInfo).
1. The screen is still dark (no display) after you powered on the instrument.
(1) Check whether the power switch is turned on. (2) Check whetherthe power is correctly connected. (3) Check whetherthe fuse is burned out. If the fuse needs to be changed,
please use the specified fuse. (4) Restart the instrument after finishing the above inspections. (5) If it still persists, please contactRIGOL.
2. The signal is sampled but no waveform of the signal is displayed.
(1) Check whetherthe probe is correctly connected to the item under tested. (2) Check whethersignals are generated from the item under test (you can
connect the probe compensation signal to the problematic channel to locate
the problem). (3) Resample the signal.
3. The voltage amplitude measured is greater or smallerthan the actual value (note that this failure usually only occurs when probe is used).
Check whether the attenuation ratioof the channel complies with the attenuation ratio of the probe.
4. Waveforms are found to be displayed but not stable.
(1) Check the trigger signal source: press MENUofthe TRIGGER Control
Areaonthe front panel, then press Sourceto confirmwhether the setting complies with the signal channel actually used.
(2) Check the trigger type: general signals should use "Edge" trigger and video
signal should use "Video" trigger. Only when the proper trigger type is used, can the waveform be displayed stably.
(3) Check the trigger level: adjust the trigger level to the middle of the signal.
MSO2000A/DS2000A Quick Guide
Page 37
27
(4) Change the trigger holdoff setting.
5. No display is found after pressing RUN/STOP.
Check the TRIGGER Control Areaonthe front panel to confirm whether the trigger mode is "Normal"or "Single" and whether the trigger level exceeds the waveform range. If yes, set the trigger level to the middle or pressthe MODE to set the trigger mode to"Auto".
Note:Pressing AUTO could automatically finish the above setting.
6. The display of waveform is ladder-like.
(1) The horizontal time base might be too low. Increase the horizontal time
base to increase the horizontal resolution and improve the display.
(2) If the display Type is "Vectors", the lines between the sample points may
cause ladder-like display. Press DisplayÆType, set the display type to"Dots"to solve the problem.
RIGOL
7. I Failto connect PC or PictBridge through USB. (1) PressUtilityÆIO SettingÆUSB Device to check whether the
settingmatches the devicecurrently connected.
(2) Check whether the USB cable is correctly connected to the instrument and
PC.
(3) Check whether the USB cable is in good condition. If needed, restart the
oscilloscope.
8. The USBstoragedevice cannot be recognized.
(1) Check whetherthe USB storagedevice can work normally. (2) Check whether the USB storagedevice being used is flash type. This
oscilloscope does not support USB3.0 and hardware type USB storage device.
(3) Check whether the capacity of the USB storagedevice is too large. It is
recommended that the capacity of the USB storagedevice being used with
this oscilloscope should not exceed8GBytes. (4) Restart the instrument and then insert the USB storagedevice to check it. (5) If the USB storagedevice still cannot be used normally, please contact
RIGOL.
MSO2000A/DS2000A Quick Guide
Page 38
Characterisation, Measurement & Analysis
Lambda Photometrics Limited
Lambda House Batford Mill Harpenden Herts AL5 5BZ United Kingdom
E: info@lambdaphoto.co.uk W: www.lambdaphoto.co.uk T: +44 (0)1582 764334
F: +44 (0)1582 712084
Loading...