Guaranty and Declaration
Copyr
ight
© 2022 RIGOL TECHNOLOGIES CO., LTD. All Rights Reserved.
Trademark Information
RIGOL®is the trademark of RIGOL TECHNOLOGIES CO., LTD.
Notices
RIGOL products are covered by P.R.C. and foreign patents, issued and pending.
RIGOL reserves the right to modify or change parts of or all the specifications and pricing
policies at the company's sole decision.
Information in this publication replaces all previously released materials.
Information in this publication is subject to change without notice.
RIGOL shall not be liable for either incidental or consequential losses in connection with the
furnishing, use, or performance of this manual, as well as any information contained.
Any part of this document is forbidden to be copied, photocopied, or rearranged without prior
written approval of RIGOL.
Product Certification
RIGOL guarantees that this product conforms to the national and industrial standards in China as well as the ISO9001:2015 standard and the ISO14001:2015 standard. Other international standard conformance certifications are in progress.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Section Description Page
List of Figures..............................................................................................................................IV
List of Tables................................................................................................................................VI
1 Safety Requirement.....................................................................................................................1
1.1 General Safety Summary................................................................................................... 1
1.2 Safety Notices and Symbols.............................................................................................3
1.3 Measurement Category..................................................................................................... 3
1.4 Ventilation Requirement....................................................................................................4
1.5 Working Environment.........................................................................................................4
1.6 Care and Cleaning................................................................................................................6
1.7 Environmental Considerations.........................................................................................6
2 Product Overview........................................................................................................................ 8
3 Document Overview.................................................................................................................10
4 Quick Start................................................................................................................................... 12
4.1 General Inspection.............................................................................................................12
4.2 Appearance and Dimensions.........................................................................................13
4.3 Front Panel...........................................................................................................................13
4.4 Rear Panel.............................................................................................................................17
4.5 User Interface......................................................................................................................18
4.6 Connecting to Power........................................................................................................20
4.7 Power-on Inspection.........................................................................................................21
4.8 Connecting the Outputs..................................................................................................22
4.9 Replacing the Fuse............................................................................................................ 22
4.10 Using the Built-in Help System..................................................................................... 23
5 Using the Protection Function..............................................................................................25
6 Constant Voltage Output........................................................................................................27
7 Constant Current Output........................................................................................................30
8 Series/Parallel Connections....................................................................................................31
8.1 Series Connections............................................................................................................31
8.2 Parallel Connections..........................................................................................................33
9 Arb Function................................................................................................................................36
9.1 Setting Arb Properties......................................................................................................37
9.2 Arb Editor..............................................................................................................................39
9.2.1 Inserting a Single Point........................................................................................... 40
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 4
9.2.2 Editing Waveform Templates.................................................................................40
9.2.3 Delete.............................................................................................................................46
9.3 Import and Export............................................................................................................. 46
9.4 Enabling the Arb Output.................................................................................................47
10 Analyzer........................................................................................................................................ 48
10.1 Common Analysis..............................................................................................................48
10.1.1 Selecting Analysis Object........................................................................................49
10.1.2 Selecting Statistical Method..................................................................................50
10.2 Pulse Current Analysis......................................................................................................51
10.3 Data Logging.......................................................................................................................52
10.4 Print Screen..........................................................................................................................53
10.5 Label Control.......................................................................................................................53
11 Trigger........................................................................................................................................... 54
11.1 Trigger Input........................................................................................................................55
11.2 Trigger Output.................................................................................................................... 57
12 Store and Recall..........................................................................................................................59
12.1 Select......................................................................................................................................60
12.2 Save.........................................................................................................................................60
12.3 Read........................................................................................................................................61
12.4 Delete.....................................................................................................................................61
12.5 Copy and Paste...................................................................................................................62
13 Utility..............................................................................................................................................63
13.1 System Settings..................................................................................................................63
13.1.1 Instrument Settings.................................................................................................. 63
13.1.2 Display Options..........................................................................................................66
13.2 Output Settings..................................................................................................................67
13.2.1 Track Mode.................................................................................................................. 68
13.2.2 Safe Mode....................................................................................................................68
13.2.3 Output Connection...................................................................................................69
13.2.4 Output Off Mode.......................................................................................................69
13.3 Interface Settings...............................................................................................................70
13.3.1 LXI Status......................................................................................................................70
13.3.2 LXI Setting....................................................................................................................72
13.3.3 USB Configuration.....................................................................................................74
13.4 Option Settings...................................................................................................................75
14 Locking/Unlocking the Front Panel.....................................................................................78
15 Remote Control..........................................................................................................................80
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 5
15.1 Remote Control via USB..................................................................................................80
15.2 Remote Control via LAN..................................................................................................81
16 Troubleshooting.........................................................................................................................83
17 Specifications..............................................................................................................................85
18 Appendix...................................................................................................................................... 91
18.1 Appendix A: Accessories and Options........................................................................91
18.2 Appendix B: Warranty.......................................................................................................92
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 6

List of Figures

Figure 4.1 Front View ..............................................................................................................13
Figure 4.2 Side View ................................................................................................................13
Figure 4.3 DP900 Front Panel .............................................................................................. 14
Figure 4.4 DP900 Output Terminals .................................................................................. 16
Figure 4.5 DP900 Rear Panel ................................................................................................17
Figure 4.6 User Interface .......................................................................................................19
Figure 4.7 Help Interface .......................................................................................................24
Figure 5.1 CH1 Output Settings Interface .......................................................................25
Figure 8.1 Series Output Interface .....................................................................................32
Figure 8.2 External Series Circuit ........................................................................................33
Figure 8.3 Parallel Output Interface .................................................................................. 34
Figure 8.4 External Parallel Circuit ......................................................................................35
Figure 9.1 Arb Main Interface ..............................................................................................36
Figure 9.2 Arb Settings Interface ........................................................................................37
Figure 9.3 Arb Editor Interface ............................................................................................39
Figure 9.4 Template Editing Menu .....................................................................................40
Figure 9.5 Arb Output Interface ......................................................................................... 47
Figure 10.1 Analyzer Main Interface ................................................................................. 48
Figure 10.2 Common Analysis Setting Interface ...........................................................49
Figure 10.3 Analysis Object Setting Menu ......................................................................49
Figure 10.4 Pulse Current Analysis Setting Interface .................................................. 51
Figure 10.5 Pulse Current Analysis Interface ..................................................................52
Figure 10.6 Log Setting Interface .......................................................................................52
Figure 11.1 Digital Port Connections ................................................................................54
Figure 11.2 Trigger Setting Interface ................................................................................ 55
Figure 11.3 Trigger Input Setting Interface .....................................................................56
Figure 11.4 Trigger Output Setting Interface .................................................................57
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 7
Figure 12.1 Disk Interface ..................................................................................................... 59
Figure 13.1 Utility Interface ..................................................................................................63
Figure 13.2 Instrument Settings Interface .......................................................................64
Figure 13.3 Display Options Interface .............................................................................. 67
Figure 13.4 Output Settings Interface ..............................................................................68
Figure 13.5 Interface Settings ..............................................................................................70
Figure 13.6 LXI Status Interface .......................................................................................... 71
Figure 13.7 LXI Settings Menu ............................................................................................ 72
Figure 13.8 USB Configuration Interface .........................................................................75
Figure 13.9 Option Settings Interface .............................................................................. 76
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 8

List of Tables

Table 4.1 DP900 Rear Panel Description ..........................................................................17
Table 4.2 User Interface Description ................................................................................. 19
Table 4.3 AC input power specifications (including AC selector setting) .............20
Table 4.4 Fuse Rating ..............................................................................................................23
Table 9.1 Arb Parameters (Template) ................................................................................ 44
Table 13.1 Factory Default Values ...................................................................................... 64
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 9

Safety Requirement

1
Safety Requirement

1.1 General Safety Summary

Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the instrument properly.
Use Proper Power Cord.
Only the exclusive power cord designed for the instrument and authorized for use within the destination country could be used.
Ground the Instrument.
The instrument is grounded through the Protective Earth lead of the power cord. To avoid electric shock, it is essential to connect the earth terminal of the power cord to the Protective Earth terminal before connecting any inputs or outputs.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and markers on the instrument and check your manual for more information about ratings before connecting the instrument.
Use Proper Overvoltage Protection.
Ensure that no overvoltage (such as that caused by a bolt of lightning) can reach the product. Otherwise, the operator might be exposed to the danger of an electric shock.
Do Not Operate Without Covers.
Do not operate the instrument with covers or panels removed.
Do Not Insert Objects Into the Air Outlet.
Do not insert objects into the air outlet, as doing so may cause damage to the instrument.
Use the Proper Fuse.
Please use the specified fuses.
Avoid circuit or wire exposure.
Do not touch exposed junctions and components when the instrument is powered on.
Do Not Operate With Suspected Failures.
If you suspect damage occurs to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 10
Safety Requirement
Keep Well Ventilation.
Do Not Operate in Wet Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Instrument Surfaces Clean and Dry.
replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
Inadequate ventilation may cause an increase of temperature in the instrument, which would cause damage to the instrument. So please keep the instrument well ventilated and inspect the air outlet and the fan regularly.
To avoid short circuit inside the instrument or electric shock, never operate the instrument in a humid environment.
To avoid personal injuries or damage to the instrument, never operate the instrument in an explosive atmosphere.
To avoid dust or moisture from affecting the performance of the instrument, keep the surfaces of the instrument clean and dry.
Prevent Electrostatic Impact.
Operate the instrument in an electrostatic discharge protective environment to avoid damage induced by static discharges. Always ground both the internal and external conductors of cables to release static before making connections.
Use the Battery Properly.
Do not expose the battery (if available) to high temperature or fire. Keep it out of the reach of children. Improper change of a battery (lithium battery) may cause an explosion. Use the RIGOL specified battery only.
Handle with Caution.
Please handle with care during transportation to avoid damage to keys, knobs, interfaces, and other parts on the panels.
Do not use this instrument to provide power for the active load.
The backflow current may cause the power control loop to be out of control, which could further damage the devices that receives the power supply from this instrument. Therefore, this instrument is only allowed to provide power for the pure load that does not have the current output function.
WARNING
Equipment meeting Class A requirements may not offer adequate protection to broadcast
services within residential environment.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 11

1.2 Safety Notices and Symbols

Safety Notices in this Manual:
WARNING
Indicates a potentially hazardous situation or practice which, if not avoided, will result in
serious injury or death.
CAUTION
Indicates a potentially hazardous situation or practice which, if not avoided, could result
in damage to the product or loss of important data.
Safety Symbols on the Product:
DANGER
Safety Requirement
It calls attention to an operation, if not correctly performed, could result in injury or hazard immediately.
WARNING
It calls attention to an operation, if not correctly performed, could result in potential injury or hazard.
CAUTION
It calls attention to an operation, if not correctly performed, could result in damage to the product or other devices connected to the product.
Safety Symbols on the Product:
Hazardous
Voltage
Safety Warning Protective Earth

1.3 Measurement Category

Chassis Ground Test Ground
Terminal
Measurement Category
This instrument can make measurements in Measurement Category I.
WARNING
This instrument can only be used for measurements within its specified measurement
categories.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 12
Safety Requirement
Measurement Category Definitions
Measurement category I is for measurements performed on circuits not directly
Measurement category II is for measurements performed on circuits directly
Measurement category III is for measurements performed in the building
connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable. Thus, you must know the transient withstand capability of the equipment.
connected to low voltage installation. Examples are measurements on household appliances, portable tools and similar equipment.
installation. Examples are measurements on distribution boards, circuit-breakers, wiring (including cables, bus-bars, junction boxes, switches and socket-outlets) in the fixed installation, and equipment for industrial use and some other equipment. For example, stationary motors with permanent connection to a fixed installation.
Measurement category IV is for measurements performed at the source of a low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.

1.4 Ventilation Requirement

This instrument uses a fan to force cooling. Please make sure that the air inlet and outlet areas are free from obstructions and have free air. When using the instrument in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
CAUTION
Inadequate ventilation may cause an increase of temperature in the instrument, which
would cause damage to the instrument. So please keep the instrument well ventilated and
inspect the air outlet and the fan regularly.

1.5 Working Environment

Temperature
Operating: 0℃ to +40℃
Non-operating: -40℃ to +60℃
Humidity
Operating:
Below +30℃: ≤90% RH (without condensation)
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 13
Safety Requirement
+30℃ to +40℃: ≤75% RH (without condensation)
Non-operating:
Below +65℃: ≤90%RH (without condensation)
WARNING
To avoid short circuit inside the instrument or electric shock, never operate the
instrument in a humid environment.
Altitude
Operating: below 3 km
Non-operating: below 15 km
Protection Level Against Electric Shock
ESD ±8kV
Installation (Overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
WARNING
Ensure that no overvoltage (such as that caused by a bolt of lightning) can reach the
product. Otherwise, the operator might be exposed to the danger of an electric shock.
Installation (Overvoltage) Category Definitions
Installation (overvoltage) category I refers to signal level which is applicable to equipment measurement terminals connected to the source circuit. Among these terminals, precautions are done to limit the transient voltage to a low level.
Installation (overvoltage) category II refers to the local power distribution level which is applicable to equipment connected to the AC line (AC power).
Pollution Degree
Pollution Degree 2
Pollution Degree Definition
Pollution Degree 1: No pollution or only dry, nonconductive pollution occurs.
The pollution has no effect. For example, a clean room or air-conditioned office environment.
Pollution Degree 2: Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected. For example, indoor environment.
Pollution Degree 3: Conductive pollution or dry nonconductive pollution that becomes conductive due to condensation occurs. To be found in industrial
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 14
Safety Requirement
environment or construction sites (harsh environments). For example, sheltered outdoor environment.
Pollution Degree 4: The pollution generates persistent conductivity caused by conductive dust, rain, or snow. For example, outdoor areas.
Safety Class
Class 1–Grounded Product

1.6 Care and Cleaning

Care
Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time.
Cleaning
Clean the instrument regularly according to its operating conditions.
1. Disconnect the instrument from all power sources.
2. Clean the external surfaces of the instrument with a soft cloth dampened with mild
detergent or water. Avoid having any water or other objects into the chassis via the heat dissipation hole. When cleaning the LCD, take care to avoid scarifying it.
CAUTION
To avoid damage to the instrument, do not expose it to caustic liquids.
WARNING
To avoid short-circuit resulting from moisture or personal injuries, ensure that the
instrument is completely dry before connecting it to the power supply.

1.7 Environmental Considerations

The following symbol indicates that this product complies with the WEEE Directive 2002/96/EC.
The equipment may contain substances that could be harmful to the environment or human health. To avoid the release of such substances into the environment and avoid harm to human health, we recommend you to recycle this product appropriately to ensure that most materials are reused or recycled properly. Please contact your local authorities for disposal or recycling information.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 15

Product Overview

2
Product Overview
Product Features
Three models available in the series:
- DP932A (Standard): 32 V/3 A || 32 V/3 A || 6 V/3 A
- DP932U (University-with safety sockets): 32 V/3 A || 32 V/3 A || 6 V/3 A
- DP932E (E-commerce): 30 V/3 A || 30 V/3 A || 6 V/3 A
3 electrically isolated independent channels with a maximum total power of up
to 210 W
4.3-inch LCD color touch screen
Front-panel safety sockets available on some models
Internal series/parallel connections for CH1 and CH2
Excellent programming/readback accuracy
Transient response time <50 μs
Low output ripple and noise <350 μV
Command processing time <10 ms
Three rack-units (3U), 1/2-rack form factor
rms
/2 mV
pp
Software control
Timing output, data logging and analysis
A maximum of 512 arbitrary points with dwell time down to 100 ms; various
built-in basic waveforms
Over voltage, over current, and over temperature protection
Various interfaces available: USB, LAN, and Digital I/O
Comparison of the Features Available in Each Model
DP932A DP932U DP932E
Model Standard University E-commerce
Display resolution
Minimum dwell time between 2 arbitrary points
1 mV/1 mA
100 ms
10 mV/1 mA (upgradable)
1000 ms (upgradable)
10 mV/10 mA (upgradable)
Not available
USB Device
Communication interfaces
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
USB Host
LAN
www.GlobalTestSupply.com
USB Device
USB Host
LAN
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
USB Device
USB Host
LAN
Page 16
Comparison of the Features Available in Each Model
Digital IO Digital IO (optional)
Product Overview
Safety sockets at front panel
Options available
Not available Available Not available
Full-featured, needless of options
DP900-ARB
DP900-HIRES
DP900-DIGITALIO
DP900-HIRES
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 17

Document Overview

3
Document Overview
This manual gives you a quick overview of the front and rear panel, user interface,
asic operation methods of DP900 Series Programmable DC Power Supply.
and b
TIP
For the lat
Publication Number
UGH08100-1110
Software Version
00.00.01
Software upgrade might change or add product features. Please acquire the latest version of the manual from RIGOL website or contact RIGOL to upgrade the software.
Format Conventions in this Manual
est version of this manual, download it from the official website of RIGOL (
1. Key
The front-panel key is denoted by the menu key icon. For example, the indicates the “Utility” shortcut key.
Menu
2.
The menu function key is denoted by the format of "Menu Name (Bold) + Character Shading" in the manual. For example, menu option in the operation interface. Tap function menu.
3. Operation Procedures
The “>” denotes the next step of the operation. For example,
indicat
Cont
The following DP900 series power supply models are available. Unless otherwise specified, this manual takes DP932A as an example to illustrate the functions and operation methods of DP900 series power supply.
es first tapping
ent Conventions in this Manual
, and then tapping Store.
System indicat
System t
o access the “System”
es the “System”
>Store
Model
DP932A 3 32 V/3 A, 32 V/3 A, 6 V/3 A
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Number of Channels
Output Voltage/Current
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 18
Document Overview
Model
DP932U 3 32 V/3 A, 32 V/3 A, 6 V/3 A
DP932E 3 30 V/3 A, 30 V/3 A, 6 V/3 A
Number of Channels
Output Voltage/Current
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 19

Quick Start

4
Quick Start

4.1 General Inspection

1. Inspect the packaging
If the packaging has been damaged, do not dispose the damaged packaging or cushioning materials until the shipment has been checked for completeness and has passed both electrical and mechanical tests.
The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument.
2. Inspect the instrument
In case of any mechanical damage,missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative.
3. Check the accessories
Please check the accessories according to the packing lists. If the accessories are damaged or incomplete, please contact your RIGOL sales representative.
Recommended Calibration Interval
RIGOL suggests that the instrument should be calibrated every 12 months.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 20

4.2 Appearance and Dimensions

Quick Start

4.3 Front Panel

This section introduces the front panel of the DP900 series by taking DP932A (as shown in the figure below) as an example. The model-specific features will be introduced in particular.
Figure 4.1 Front View
Figure 4.2 Side View
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 21
Quick Start
Figure 4.3 DP900 Front Panel
1. LCD
4.3-inch TFT touch screen to display the the system parameter settings, output state, menu items, prompt messages, etc.
2. Channel Selection Keys and Output On/Off Keys
The keys are channel selection keys and output on/off keys.
Press this key to select CH1 to control. You can set the parameters of this
channel, such as voltage, current, and overvoltage/overcurrent protection.
Press this key to select CH2 to control. You can set the parameters of this
channel, such as voltage, current, and overvoltage/overcurrent protection.
Press this key to select CH3 to control. You can set the parameters of this
channel, such as voltage, current, and overvoltage/overcurrent protection.
Pressing this key can turn the corresponding channel on or off.
Press this key and a prompt message will be displayed asking whether to turn all channels on. Tap OK to turn all channels on. Pressing this key again can turn all channels off.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 22
Quick Start
3. Parameter Input Area
As shown in the figure below, the parameter input area includes the arrow keys (unit selection keys), numeric keypad, and knob.
- Arrow keys/unit selection keys
Arrow keys: In normal mode, you can use the keys to navigate through the menu items; in parameter editing mode, you can use the left/right key to select the digit place or the up/down key to increment or decrement the value at the selected digit place. You can press the knob to enter or quit the parameter editing mode.
Unit selection keys: When setting parameters with front-panel numeric keypad, you can also use the arrow keys to select the unit of voltage (V/mV) or the unit of current (A/mA).
- Numeric keypad
Ring-type numeric keypad: The keypad is composed of numeric values from 0 to 9 and the decimal point. Pressing the corresponding key can directly input the number or decimal point.
- Rotrary Knob
Press: When in normal mode, pressing the knob can switch to the editing mode; pressing the knob again can disable the editing mode.
Rotate: You can rotate the knob either clockwise or counter-clockwise. In normal mode, rotating the knob can navigate through the menu or menu items; after selecting an item to control, you can rotate the knob to increment or decrement the value or select options from a drop-down button.
4. Preset Key
Press the key to restore the instrument to factory default (refer to
Default Values
).
Factory
5. Enter Key
Press the key to confirm your value.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 23
Quick Start
Long pressing the key can lock the touch screen, for which you cannot configure the instrument using the touch screen. Long pressing the key again can unlock the touch screen.
6. Back Key
Press the key to delete the character currently before the cursor.
When the instrument is in remote control, pressing this key can return to local (front panel) operation.
7. Output terminals
Figure 4.4 DP900 Output Terminals
- Channel output terminals: used to output voltage and current of the channel.
Please refer to of the model.
- Earth ground terminal: be connected to the instrument chassis and earth
ground (the ground terminal of the power cord) and is grounded.
- The output terminals of DP932U are safety sockets.
8. Function Menu
Press this key to enter the Arb interface, in which you can set Arb
parameters, and enable/disable the Arb function.
Press this key to set the Analyzer interface, in which you can select the
object to analyze and turn on/off the Logger.
Press this key to enter the Utility setting interface, in which you can set parameters related to remote interface, system, outputs, and options. Besides, you can also set the display.
Document Overview
for available output voltage/current range
Press this key to enter the Store and Recall interface, in which you can save, read, delete, copy, or paste files. The file types available for storage include state file, log file, Arb file, calibration file, option installation verification file, and bitmap file. You are allowed to store a file to internal or external memory and also recall the file from internal or external memory.
9. USB HOST Port
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 24
The instrument only supports the flash memory USB storage device of FAT32 format.
10. Power Switch Key
Press this key to power on or off the instrument.

4.4 Rear Panel

This section introduces the rear panel of DP900 series by taking DP932A (as shown in the figure below) as an example.
Quick Start
Figure 4.5 DP900 Rear Panel
Table 4.1 DP900 Rear Panel Description
No. Name Description
1 LAN port
2 USB DEVICE
3
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Digital I/O
www.GlobalTestSupply.com
[1]
The instrument is connected to the LAN network via RJ45 interface.
Connect the instrument (as "slave" device) to external USB device (such as PC).
Digital I/O port
DP900 User Guide
Page 25
Quick Start
No. Name Description
Connect the instrument (as "host" device)
4 USB HOST
5 AC selector
6 AC power inlet socket AC inlet power connector
7 Fuse
to external USB device (such as USB storage device).
Select the specification of the input voltage: 100, 120, 220, or 230 (please refer to
The fuse rating is related to the instrument model and actual input voltage (please refer to
Rating
Connecting to Power
Table 4.4 Fuse
).
).
8
9 Ground terminal -
10 Fan ventilation hole -
Note[1]: Digital I/O is optional for DP932U and not available for DP932E
AC input power requirement

4.5 User Interface

Tap > Display to assess the interface as shown in the figure below. This section introduces the meter-view display of the power supply.
The table describes the relationship between the frequency and voltage of the input power, and fuse rating.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 26
Quick Start
Figure 4.6 User Interface
Table 4.2 User Interface Description
No. Description
1 Channel identifier
2 Actual output voltage
3 Actual output current
Output mode
CV: The output is in constant voltage mode.
4
CC: The output is in constant current mode.
UR: The output is unregulated.
5
2-Wire sense connection
6 Actual output power
7 Navigation icon
8 Voltage setting value
9 Over-voltage protection (OVP) status indicator, indicating the
present OVP on/off state
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 27
Quick Start
No. Description
10 Current setting value
11 Over-current protection (OCP) status indicator, indicating the
present OCP on/off state
The following icons indicate the system status.
: CH1 and CH2 are independent of each other.
: CH1 and CH2 are connected in series.
: CH1 and CH2 are connected in parallel.
: over-temperature protection is tripped.
12
: the screen is locked.
: network connected.
: a USB device is detected.
: the instrument is in remote control.
: the beeper is turned off.
: the beeper is turned on.
: downloading the firmware upgrade file.

4.6 Connecting to Power

DP900 series power supply supports various AC power inputs. The AC selector setting on the rear panel differs when input power is different, as shown in the table below.
Table 4.3 AC input power specifications (including AC selector setting)
AC input power AC selector
100 Vac±10%, 50 Hz to 60 Hz 100
120 Vac±10%, 50 Hz to 60 Hz 120
220 Vac±10%, 50 Hz to 60 Hz 220
230 Vac±10% (Max. 250 Vac), 50 Hz to 60 Hz 230
Please connect the power strictly following the steps below.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 28
1. Check the Input Power
Quick Start
Please make sure that the input AC power fulfills the requirements in
input power specifications (including AC selector setting)
2. Check the AC Voltage Selector on the Rear Panel
Please make sure that the AC voltage selector (100, 120, 220, or 230) on the rear panel is set to match the actual input voltage. To change the input AC voltage selector on the power supply, use the two AC selector switches on the rear panel of the power supply as shown below.
Please set the switches following the diagram above. For example, to select 100 V, slide both switches to the left; to select 220 V, slide the upper switch to the left and the lower switch to the right.
.
Table 4.3 AC
3. Check the Fuse
The instrument has been equipped with the specified fuse when it leaves factory. Please make sure that the fuse matches the actual input voltage by referring to “Input power requirement” or
4.
Connect to AC Power
Please connect the instrument to AC power using the power cord provided in the accessories.
WARNING
To avoid electric shock, please make sure that the instrument is correctly grounded.

4.7 Power-on Inspection

Press the power key on the front panel to switch on the instrument. During the start­up process, the instrument performs a series of self-tests. If the instrument passes the self-test, the welcome interface will be displayed; otherwise, the corresponding self­test failure information will be displayed.
Replacing the Fuse
.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 29
Quick Start
TIP
If you want to power on the instrument again after powering it off, please make sure that the
time interval between the two power-on operations is greater than 5 s.

4.8 Connecting the Outputs

This series power supply provides three-channel outputs, which share the same connection method. Please follow the instructions below to connect the outputs.
Method 1: Connect wires to the front of the terminals as shown in (A).
Method 2: Rotate the terminal block screws counter-clockwise and connect wires to the terminals in location (B). Then rotate the screws clockwise to tighten the wires. This method can help avoid errors caused by the terminal resistance.
TIP
As the output terminals of DP932U are safety sockets, “Method 2” is not supported.
CAUTION
Turn off AC power before making front panel connections. All wires and straps must be
properly connected to prevent currents from damaging the loads.

4.9 Replacing the Fuse

The fuse rating required depends on the instrument model and actual input voltage, as shown in the table below. You can also refer to
specifications (including AC selector setting)
Table 4.3 AC input power
.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 30
Quick Start
Table 4.4 Fuse Rating
AC Input Voltage Fuse Rating
100 Vac/120 Vac T6.3 A 250 Vac
220 Vac/230 Vac T3.15 A 250 Vac
If you need to replace the fuse, please follow the steps below.
1. Power off the instrument and remove the power cord.
2. Insert a small straight screwdriver into the slot at the power socket and pry out the
fuse holder gently.
3. If necessary, please adjust the AC selector setting manually to select the voltage that matches the actual input voltage (please refer to
4. Remove the original fuse and insert the proper fuse into the fuse holder (please refer to the “Input power requirement” or
5. Re-insert the fuse holder into the power socket (please pay attention to the direction).
WARNING
To avoid personal injury, please power off the power supply before replacing the fuse; to
avoid electric shock or fire, select the input power that matches the actual input voltage
and use the correct fuse before connecting to power.
Replacing the Fuse

4.10 Using the Built-in Help System

Tap > Help to enter the help interface as shown in the figure below. The built-in help system can help you learn about the instrument, obtain a list of help topics, and update instrument firmware.
Connecting to Power
).
).
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 31
Quick Start
Figure 4.7 Help Interface
About
Tap About to view the system information of the instrument, including the instrument model number, serial number, firmware version, and calibration time.
Help
Tap Help to view the help information for front-panel keys or menu keys. You can tap a function key topic at the left side to obtain a detailed description of its function.
Upgrade
Tap Upgrade to access the software upgrade interface. You can read the USB storage device and tap instrument to network and then tap Remote to download the firmware upgrade file
( is displayed in the notification area in the process). After the file is downloaded, the instrument restarts in 2 seconds and starts to upgrade.
Others
Tap Others to enter the interface in which you can obtain the following help topics:
Upgrade to update instrument firmware. You can also connect the
View the last displayed message
View remote command error queue
Contact RIGOL Technical Support
Open source statement
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 32

Using the Protection Function

5
Using the Protection Function
Each channel output has independent over-voltage protection (OVP) and over-current protection (OCP) functions. The “OVP”/“OCP” status indicator is lit when a protection function is enabled.
Overvoltage Protection (OVP)
When OVP is enabled, the output will be disabled when the output voltage reaches the voltage limit setting.
Tap > Display to enter the meter view for multiple outputs. Then you can tap one of the channel labels at the bottom to open the meter view for a single channel. For example, you can configure CH1 output settings in its meter view, as shown in the figure below.
Figure 5.1 CH1 Output Settings Interface
You can set the OVP value in the following ways.
Tap the “OVP” input field and set the limit value with the pop-up numeric keypad.
Rotate the knob or use the arrow keys to select the “OVP” input field. Then press the knob to enter the editing mode. After that, you can either use the front panel numeric keypad to directly enter the value or use the up/down arrow keys to increment or decrement the value in the specified digit place. Press the knob again to disable the editing mode.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 33
Using the Protection Function
After finishing the voltage limit setting, you can click or tap the “OVP” on/off button to turn ON or OFF the OVP function.
Overcurrent Protection (OCP)
When OCP is enabled, the output will be disabled when the output current reaches the current limit setting.
You can tap > Display to enter the meter view for multiple outputs. Then you can tap one of the channel labels at the bottom to open the meter view for a single channel. For example, you can configure CH1 output settings in its meter view, as shown in
You can set the OCP value in the following ways.
Tap the “OCP” input field and set the value with the pop-up numeric keypad.
Rotate the knob or use the arrow keys to select the “OCP” input field. Then
Figure 5.1
press the knob to enter the editing mode. After that, you can either use the front panel numeric keypad to directly enter the value or use the up/down arrow keys to increment or decrement the value in the specified digit place. Press the knob again to disable the editing mode.
.
After finishing the current limit setting, you can click or tap the “OCP” on/off button to turn ON or OFF the OCP function.
OCP Delay
The power supply may momentarily has a peak current value beyond the OCP value when it is turned on, when an output value is programmed, or when the output load is connected. In most cases, these temporary conditions would not be considered an over-current protection fault, and you do not need to disable the output. The series power supply ignores this status bit during the specified delay period (10 ms by default). Once the OCP delay time has expired and if the current is still beyond the limit, the output will shut down.
You can also set the OCP delay period by sending :OUTPut:OCP:DELAy [CH1|CH2| CH3], {<value>|MINimum|MAXimum} SCPI command (refer to the series
Programming Guide
). The delay can be programmed from 0 to 1000 ms.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 34

Constant Voltage Output

6
Constant Voltage Output
The series power supply provides three output modes: constant voltage (CV), constant current (CC), and unregulated (UR). In CV mode, the output voltage equals to the voltage setting value, and the output current is determined by the load; whereas in CC mode, the output current equals to the current setting value, and the output voltage is determined by the load. UR mode is the unregulated state between CV and CC mode. This chapter introduces how to make the instrument operate in CV mode.
Operation Procedures:
1.
Connect the load to the channel output terminals (please refer to
Outputs
CAUTION
While making connections, pay attention to the polarity to avoid damaging the
instrument or the devices connected to the instrument.
2. Press the power key to start the instrument and the meter view appears.
3.
Set the voltage
).
Connecting the
Method 1:
You can use the front-panel parameter input area to set the voltage.
a. Use the knob or arrow keys to select the voltage field in meter view or in
channel output settings interface.
b. Press the knob and the input box turns blue, indicating that the selected field
can be edited.
c. Use the left/right arrow keys to select the digit place and then rotate the knob
or use the up/down arrow keys to adjust the value in the selected place. You can also directly set the voltage by using the front-panel numeric keypad. The default unit is V.
d. Press the knob again to disable the editing mode.
TIP
In meter view, if you select a current "Set" field to edit, you can still set the voltage by using
the front-panel numeric keypad to set the value and then pressing the unit selection key
/ .
Method 2:
You can also use the touch screen function to set the voltage.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 35
Constant Voltage Output
In the channel output settings interface, tap the voltage “Set” field and use the pop-up numeric keypad to enter the value and unit.
You can tap on the virtual keypad or press on the front panel to delete the character before the cursor.
4.
Set the current
Method 1:
You can use the front-panel parameter input area to set the current.
a. Use the knob or arrow keys to select the current field in meter view or in
channel output settings interface.
b. Press the knob and the input box turns blue, indicating that the selected field
can be edited.
c. Use the left/right arrow keys to select the digit place and then rotate the knob
or use the up/down arrow keys to adjust the value in the selected place. You can also directly set the current by using the front panel numeric keypad. The default unit is A.
d. Press the knob again to disable the editing mode.
TIP
In meter view, if you select a voltage "Set" field to edit, you can still set the current by using
the front panel numeric keypad to set the value and then pressing the unit selection key
/ .
Method 2:
You can also use the touch screen function to set the current.
In the channel output settings interface, tap the current “Set” field and use the pop-up numeric keypad to enter the value and unit.
You can tap on the virtual keypad or press on the front panel to delete the character before the cursor.
5. Set the OCP
Set the OCP value (refer to
Using the Protection Function
) and then tap OCP on/off button to enable the OCP function. The output will be disabled automatically when the actual output current reaches the OCP value.
6.
Enable the output
Pressing the channel on/off key
can turn on the respective channel output.
When a channel is turned on, its on/off key is illuminated. Rather, when the
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 36
Constant Voltage Output
channel is turned off, the light will be off. You can press to turn all outputs on or off simultaneously.
WARNING
To avoid electric shock, please turn on the output after the output terminals are
correctly connected.
7. Check the output mode
In CV mode, the displayed output mode is "CV"; if "CC" is displayed, increase the current value properly and the power supply will switch to CV mode automatically.
TIP
In CV output mode, when the load current is greater than the current setting value, the power
supply will switch to CC mode automatically. At this time, the output current equals to the
current setting value, and the output voltage can be calculated by the current times the load
impedance.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 37

Constant Current Output

7
Constant Current Output
In CC mode, the output current equals to the current setting value, and the output voltage is determined by the load. This chapter introduces how to make the instrument operate in CC mode.
Operation Procedures:
Connect output terminals
1.
Please connect the load to the channel output terminals on the front panel (please refer to
2. Press the power switch to start the instrument.
3. Set the voltage
Set your desired voltage value (refer to
4. Set the current
Set your desired current value (refer to
5. Set the OVP
Connecting the Outputs
).
Constant Voltage Output
Constant Voltage Output
).
).
Set the proper OVP value (refer to the OVP function by tapping OVP on/off button. The output will be disabled automatically when the actual output voltage reaches the OVP value.
6.
Enable the output
Please turn on the output (refer to
WARNING
To avoid electric shock, please turn on the output after the output terminals are
correctly connected.
7. Check the output mode
In CC mode, the output mode displayed should be "CC"; if "CV" is displayed, increase the voltage value properly and the power supply will switch to CC mode automatically.
TIP
In CC mode, when the load voltage is greater than the voltage setting value, the power supply
will switch to CV output mode automatically. At this time, the output voltage equals to the
voltage setting value, and the output current can be calculated by the voltage divided by the
load impedance.
Using the Protection Function
Constant Voltage Output
).
) and then enable
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 38

Series/Parallel Connections

8
Series/Parallel Connections
Connecting two or more isolated channels in series provides a greater voltage capability while connecting two or more isolated channels in parallel provides a greater current capability. The series power supply supports both internal and external series/parallel connections.
TIP
The three channels of the power supply are electrically isolated with independent
outputs. For a single power supply, any two of the three channels can be externally
connected in series/parallel.
Isolated channels of different power supplies can be externally connected in series/
parallel.
CH1 and CH2 can be internally connected in series/parallel.
When in internal series mode, CH1 and CH2 cannot be externally connected in parallel;
when in internal parallel connection mode, CH1 and CH2 cannot be externally connected
in series.
The parameter settings for series and parallel connections must comply with the safety
requirements.

8.1 Series Connections

Connecting outputs in series provides a greater voltage capability than can be obtained from a single output. The total voltage is the sum of the individual voltages of all channels connected in series. To connect outputs in series, you need to ensure identical current ratings for each channel.
WARNING
To avoid shock hazard, please do not touch the output terminals when the output voltage
exceeds 60 V in series mode.
Internal Series Connections
This series power supply supports internal series connection for CH1 and CH2. In internal series connection mode, the output voltage (up to 64 V) is twice the voltage setting. The output voltage and current are displayed in CH1 meter view, as is shown in the figure below.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 39
Series/Parallel Connections
The figure below shows how to connect front output terminals to a single load in internal series mode. Please refer to connection.
Figure 8.1 Series Output Interface
Output Connection
for internal parallel
TIP
In internal series/parallel connections, CH1 and CH2 always have identical voltage and
current settings.
The output voltage displayed in CH1 meter view is twice the voltage setting. CH2 has no
reading.
In internal series mode, the output voltage at the terminal is twice the voltage setting. To
avoid damage to the load, please set proper voltage values
External Series Connections
This series power supply supports external series connection for multiple channels (from a single power supply or multiple power supplies). The figure below illustrates how to connect two channels when in external series mode.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 40
Series/Parallel Connections
Figure 8.2 External Series Circuit
Operation Procedures:
1. Connect the power supply to load as shown in the figure above. Pay attention to
the polarity when making connections.
2. Set proper voltage, current, and over-current protection value for each channel
(refer to
channels connected in series should be set to CV mode and have identical current
setting value and OCP value.
CAUTION
Please make sure that all the channels connected in series operate in CV mode. One of the
channels operating in CC mode can cause unregulated output mode for the other
channels.
Constant Voltage Output

8.2 Parallel Connections

Connecting outputs in parallel provides a greater current capability than can be obtained from a single output. The total current is the sum of the individual currents of all channels connected in parallel. To connect outputs in parallel externally, you need to ensure identical voltage value and OVP value for each channel.
) and turn on the channel outputs. All the
Internal Parallel Connections
This series power supply supports internal parallel connection for CH1 and CH2. In internal parallel connection mode, the output current (up to 6 A) is twice the current setting. The output voltage and current are displayed in CH1 meter view, as is shown in the figure below.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 41
Series/Parallel Connections
Figure 8.3 Parallel Output Interface
The figure below shows how to connect front output terminals to a single load in internal parallel mode. Please refer to
Output Connection
for channel connection.
TIP
In internal series/parallel connections, CH1 and CH2 always have identical voltage and current
setting value.
External Parallel Connections
This series power supply supports external parallel connection for multiple channels (from a single power supply or multiple power supplies). The figure below illustrates how to connect two outputs in external parallel.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 42
Series/Parallel Connections
Figure 8.4 External Parallel Circuit
Operation Procedures:
1. Connect the power supply to load as shown in the figure above. Pay attention to the polarity when making connections.
2. Set proper parameters for each channel (refer to
Constant Current Output
work in CV mode or CC mode.
CAUTION
All the channels can work in CV or CC mode according to the actual need of the load.
) and turn on the output of each channel. All channels can
Constant Voltage Output
and
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 43

Arb Function

9
Arb Function
The Arbitrary function of DP900 (not available for DP932E) series power supply enables you to generate freely programmable waveforms which can be reproduced within the limit settings for voltage and current. You can set the repetition cycle for the arbitrary waveform as well as the output voltage/current and time for each group of data. Moreover, the instrument has multiple built-in waveform templates for you to select and edit. You can set your waveform parameters based on those templates. The instrument will output waveforms according to the parameters set.
Press or tap > Arb to enter the Arb main interface, as shown in the figure below. Then tap the channel selection drop-down button at the lower-left corner of the interface to select a channel to control. The menu keys and their functions are as follows:
Run/Stop:
Turn on or off the arbitrary waveform generator.
Setting:
Open the Arb settings interface to set the output properties for the arbitrary waveform.
Import:
Read Arb file.
Export:
Save the Arb parameters edited.
Edit:
Access the Arb editor to edit the data.
Figure 9.1 Arb Main Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 44

9.1 Setting Arb Properties

After entering the Arb main interface, tap Setting to enter the Arb settings interface as shown in the figure below. The selected channel is highlighted at the lower-left corner of the interface.
Arb Function
Figure 9.2 Arb Settings Interface
In the interface, you can set the following output properties for the arbitrary waveform.
Repetition Cycle
The repetition cycle is the number of times that the arbitrary waveform is repeated based on the preset voltage/current and time. You can tap the Rep. input field to set the desired repetition cycle by using the touch screen or the front-panel numeric keypad. The repetition cycle can be set from 1 to 99999. Entering “0” in the field will set the repetition cycle to “infinity”.
End State
End state is the instrument state after the output sequence completes the output of all voltage/current values when the repetition cycle is set to a finite value. You can tap
EndState drop-down button to select “Off” or “Last”.
the
Off: The output of the selected channel is turned off automatically after the output sequence completes.
Last: The instrument remains at the last voltage and current values after the output sequence completes.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 45
Arb Function
TIP
If the repetition cycle is set to “Infinity”, the end state is not available.
Run Mode
Run mode is the output state of the selected channel once the generator is turned on. You can tap the
RunMode drop-down button to select “Continuous” or “Single”.
When the run mode is set to “Continuous”, tap Run(Cont) in the Arb main
interface and the instrument will repeat the sequence continuously based on the number of data groups and repetition cycle currently set.
When the run mode is set to “Single”, tap Run(Sgl) and the button for running
a single arbitrary point will be displayed, as shown in the figure below. Each
time you tap , the instrument will output a single group of data in order. The current output group number and repetition cycle are displayed at the bottom of the interface.
Trigger Source
Trigger source specifies the way of starting the output of the arbitrary sequence. You can tap the TrigSrc drop-down button to select “Manual” or “BUS”.
Manual:
- Selects the Run/Stop key as a trigger source. When the run mode is set to
“Continuous”, tap Run(Cont) and the instrument will repeat the sequence continuously based on the number of data groups and repetition cycle currently set.
- When the run mode is set to “Single”, tap Run(Sgl) and then tap to
output a single group of data.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 46
Arb Function
BUS: Selects a remote command as a trigger source. You can send SCPI commands to control the Arb sequence output. In the Arb main interface, tap Run(Cont) or Run(Sgl) and a prompt message “Wait for Trigger signal...” will be displayed. The instrument will start or stop the waveform output according to the received SCPI command. Please refer to
DC Power Supply Programming Guide
Apply or Back
Tap Apply to save the current settings and return to the Arb main interface. Otherwise, tap Back to cancel the current settings and return to the main interface.
TIP
The settings are valid only for the selected channel. You can change the selected channel in
Figure 9.2
select the channel in
through pressing the specified channel selection key on the front panel. You can also
Figure 9.1
using the touch screen.
DP900 Series Programmable Linear
for available SCPI commands.

9.2 Arb Editor

After entering the Arb main interface, you can tap Edit to access the Arb editor interface, as shown in the figure below. The Arb editor allows you to create each single point of a waveform and freely edit individual points. You can also edit the Arb parameters based on the instrument's built-in waveform templates.
Figure 9.3 Arb Editor Interface
As shown in the figure above, the table displays 4 groups of data per page. You can view and edit data of other groups in the following ways.
Tap Prev or Next to turn pages.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 47
Arb Function
Tap the page input field and input the page number to jump to by using the pop-up virtual keypad or the front-panel numeric keypad.
Tap the “No.” cell at the upper-left of the table and input the number to jump to the specified row by using the pop-up virtual keypad or the front-panel numeric keypad.

9.2.1 Inserting a Single Point

In the Arb editor interface, you can tap Single to insert one group of data in the currently selected row (1 V default voltage, 1 A default current, and 1 s default time). You can set the voltage/current/time for each group by using the touch screen or the numeric keypad on the front panel.
TIP
If the currently selected row is empty and the row above the selected row has data
inserted, tap Single and one group of data will be inserted in the selected row.
If the currently selected row has data inserted, tap Single and a group of data will be
inserted in the selected row with the original group of data moved to the next row.

9.2.2 Editing Waveform Templates

The instrument provides multiple built-in output templates based on which you can create and edit arbitrary waveforms.
Tap Templet to open the template editing menu, as shown in the figure below. You can edit and create your desired waveform in the following steps.
Figure 9.4 Template Editing Menu
1.
Select Waveform Templates
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 48
Arb Function
Tap the Wave drop-down button to select a waveform template. Available waveform templates include Sine, Pulse, Ramp, Stair UP, Stair Dn, StairUpDn, Exp Rise, and Exp Fall.
- Sine
The Sine waveform is as shown in the figure below. Setting the maximum and minimum value creates the amplitude. Setting the period (T) and time interval creates the total number of points (P) in a complete period. Thus, a Sine waveform is created. Setting the inserted groups (G) determines the number of groups to be inserted in the Arb editor.
- Pulse
The Pulse waveform is as shown in the figure below. Setting the maximum and minimum value creates the amplitude. Setting the positive pulse width (t) determines the duration of high levels. The currently set period (T) minus the positive pulse width (t) is the duration of low levels. Setting the inserted groups (G) determines the number of groups to be inserted in the Arb editor.
- Ramp
The Ramp waveform is as shown in the figure below. Setting the maximum and minimum value creates the amplitude. Setting the period (T) and time interval creates the total number of points (P) in a complete period. Setting the symmetry (Sym) creates the Ramp waveform. The rising edge of the Ramp
waveform has int (P*Sym)
[1]
points and the falling edge has P-int (P*Sym)
points.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 49
Arb Function
[1]
Note
: int (P*Sym) refers to rounding P*Sym (to discard the decimal part).
- Stair Up
The Stair Up waveform is as shown in the figure below. Setting the maximum (MAX), minimum (MIN), period, and total number of steps (N) creates the Stair Up waveform. The number of steps (N) ranges from MIN to MAX. The step size is ( MAX-MIN )/(N-1). Interval=Period/Number of Steps.
- Stair Dn
The Stair Dn waveform is as shown in the figure below. Setting the maximum (MAX), minimum (MIN), period, and total number of steps (N) creates the Stair Dn waveform. The number of steps (N) ranges from MIN to MAX. The step size is (MAX-MIN )/(N-1). Interval=Period/Number of Steps.
- Stair UpDn
The Stair UpDn waveform is as shown in the figure below. Setting the maximum (MAX), minimum (MIN), period (T), and total number of steps (N) creates the Stair UpDn waveform. When N is odd, the waveform rises from MIN to MAX by step size of (MAX-MIN )/int(N/2) and then drops to MIN by the same step size. When N is even, the waveform rises from MIN to MAX by step size of (MAX-
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 50
Arb Function
MIN)/{(N/2)-1} and then drops to MIN by step size of (MAX-MIN)/(N/2). Interval=Period/Number of Steps.
- Exp Rise
The Exp Rise waveform is as shown in the figure below. Setting the maximum (MAX), minimum (MIN), inserted groups (total number of points, denoted by P), interval, and rise index (denoted by RiseIndex) creates the Exp Rise waveform.
The waveform function is (MAX-MIN)*(1-e
-i*RiseIndex/P
); wherein, “i” is
independent variable ranging from 0 to (P-1).
- Exp Fall
The Exp Fall waveform is as shown in the figure below. Setting the maximum (MAX), minimum (MIN), inserted groups (total number of points, denoted by P), interval, and fall index (denoted by FallIndex) creates the Exp Fall waveform. The
waveform function is (MAX-MIN)*e
-i*FallIndex/P
; wherein, “i” is independent
variable ranging from 0 to (P-1).
TIP
When the template selected is Exp Rise or Exp Fall, the maximum or minimum value cannot
be reached due to the characteristic of the exponential function itself. The available range
depends on the rise/fall index currently set. The larger the rise/fall index, the wider the
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 51
Arb Function
range, as shown in the figures below (maximum=8, minimum=0, total number of
points=100, rise index/fall index: 1, 4, 7 and 10).
2. Select Voltage/Current to Edit
Tap the Object drop-down button and select “Voltage” or “Current” to edit.
- Voltage: When voltage is selected, the output current of all groups is fixed. You
can tap the Current input field to set the current value using the touch screen
or the parameter input area on the front panel.
- Current: When current is selected, the output voltage of all groups is fixed. You
can tap the Voltage input field to set the voltage value using the touch screen
or the parameter input area on the front panel.
3. Edit Arb Parameters
When selecting different waveform templates, you need to edit different parameters. Please refer to the table below for details.
Table 9.1 Arb Parameters (Template)
Template Parameters
Sine Max, Min, Period, Interval, Groups, Invert On/Off
Pulse Max, Min, Period, Positive Pulse Width, Groups, Invert On/Off
Ramp Max, Min, Period, Interval, Groups, Symmetry, Invert On/Off
Stair Up Max, Min, Period, Steps, Groups
Stair Dn Max, Min, Period, Steps, Groups
StairUpDn Max, Min, Period, Steps, Groups
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 52
Template Parameters
Exp Rise Max, Min, Groups, Interval, Exp
Exp Fall Max, Min, Groups, Interval, Exp
- Max
Set the maximum voltage/current of the template currently selected. The available range is related to the channel currently selected. The max value should be greater than or equal to the min value currently set.
- Min
Set the minimum voltage/current of the template currently selected. The available range is related to the channel currently selected. The min value should be smaller than or equal to the max value currently set.
Arb Function
- Period
Set the duration of a cycle for the waveform currently selected. The max period is 3600 s.
- Interval
Set the time interval between two points (the time required for the instrument to output each group of data) of the template currently selected. The maximum interval is 3600 s.
- Groups
Set the number of points selected from the waveform at equal intervals. The available range is from 1 to 512.
- Invert
When the template currently selected is Sine, Pulse, or Ramp, if Invert is turned on, the instrument will first turn the preset waveform upside down and then create the output waveform.
- Positive Pulse Width
When the template currently selected is Pulse, you can set the positive pulse width (namely the duration of high levels within a period). The maximum pulse width is 3600 s. The actual available positive pulse width depends on the period currently set.
- Symmetry
When the template currently selected is Ramp, you can set the symmetry (namely the ratio of the duration of the rising edge within a period to the whole period). The range is from 0% to 100%.
- Steps
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 53
Arb Function
When the template currently selected is Stair Up, Stair Dn, or Stair UpDn, you can set the total number of points within a complete period. The available range is from 3 to 99999.
- Exp
When the template currently selected is Exp Rise, you can set the rise index which ranges from 0 to 10.
When the template currently selected is Exp Fall, you can set the fall index which ranges from 0 to 10.
4. Apply Settings
After editing the Arb parameters (template), tap Apply to create the arbitrary waveform. Its Arb parameters will be displayed in the Arb editor as shown in
9.3
.
Figure

9.2.3 Delete

In the Arb editor interface ( group of data. If the selected row has no data, a prompt message “No data in the unit” will be displayed.
Tap Clear and a prompt message “Clear all groups?” will be displayed. Tapping Yes will remove all data in the table.
Figure 9.3

9.3 Import and Export

You can save the Arb parameters to internal or external memory and recall them when required.
Export
After editing the parameters, you can tap Export to access the store and recall interface. Select the desired storage path and then tap Save. Set the filename in the pop-up filename input field. A prompt message “Arb file is exported successfully” will be displayed. The file type is fixed to “.RTF”.
), you can tap Delete to delete the selected
Import
Tap Import to access the store and recall interface. Select the file that you want to read and then tap Read. A prompt message “Arb file is imported successfully” is displayed and the instrument goes back to the Arb main interface. You can edit the imported Arb file.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 54

9.4 Enabling the Arb Output

After editing the Arb parameters, you can go back to the Arb main interface and turn on waveform output based on settings in Arb output interface.
Trigger Source
Arb Function
. The figure below shows the
Figure 9.5 Arb Output Interface
TIP
Enabling the output will change the channel output value. Please make sure that the
change of the output value will not affect the devices connected to the power supply
before turning on the output.
When the generator is turned on, you need to turn on the corresponding channel to
ensure valid waveform output.
When the output is initiated, all the Arb settings and parameters cannot be configured.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 55

Analyzer

10
Analyzer
DP900 series power supply provides the Analyzer function for you to analyze the quality of output signals. Available analysis functions for the 3 channels include common analysis, pulse current analysis (only for CH1 and CH2), and data logging.
You can press the front-panel or tap > Analyzer to access the Analyzer main interface, as shown in the figure below. The label at the upper-left of the interface indicates the currently selected analysis function (common analysis/pulse current analysis). The four function keys at the right side allow you to easily configure the Analyzer settings, execute analysis, print screen, and control labels.
Figure 10.1 Analyzer Main Interface

10.1 Common Analysis

In the Analyzer main interface, tap QicSet to access the quick setup interface. Then tap Common to enter the common analysis setting interface, as shown in the figure below. In this interface, you can select the object to analyze and statistical method.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 56
Figure 10.2 Common Analysis Setting Interface
Analyzer

10.1.1 Selecting Analysis Object

After entering the common analysis interface, you can tap Object to access the analysis object setting menu, as shown in the figure below. Select one to three items from the “Current”, “Voltage”, and “Power” items of CH1, CH2, and CH3.
Then tap OK to go back to the analyzer main interface. Tap Run and the instrument will execute analysis based on the your selections.
Figure 10.3 Analysis Object Setting Menu
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 57
Analyzer

10.1.2 Selecting Statistical Method

After entering the common analysis interface, tap Method to pop up the statistical method setting menu. You can select RT Plot or Trend Plot.
RT Plot
It displays the analysis object currently selected in real-time plot. In RT Plot mode, the waveform of the collected signals slowly scrolls from right to left. The latest data is displayed on the screen.
Trend Plot
It displays the analysis object currently selected in trend plot. In Trend Plot mode, the waveform plot area is divided into two parts: the left displays trend waveform while the right displays real-time data. Real-time data will be drawn point by point. When the data is out of the screen, it will be compressed and displayed proportionally in the trend plot. Then, real-time data will be redrawn from left to right.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 58

10.2 Pulse Current Analysis

The pulse current analysis function is only available for CH1 and CH2 of this series power supply. After entering the quick setup interface, you can tap Pulse Current to enter the pulse current analysis setting interface, as shown in the figure below.
Analyzer
Figure 10.4 Pulse Current Analysis Setting Interface
1. First select the channel you want to analyze.
2. Tap Pos pulse or Neg pulse to select the pulse current to analyze.
- Positive pulse: Select Pos pulse, and set the upper threshold level in the input
field. In the analysis process, the Analyzer records the number of pulses above the threshold and the most recent pulse width beyond this upper threshold.
- Negative pulse: Select Neg pulse, and set the lower threshold level in the input field. In the analysis process, the Analyzer records the number of pulses below the threshold and the most recent pulse width below this threshold.
3. Then tap OK to return to the Analyzer main interface. Tap Run and the instrument will execute analysis based on your selections. The analysis interface is shown in the figure below. The analysis results are at the lower part of the interface.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 59
Analyzer
Figure 10.5 Pulse Current Analysis Interface

10.3 Data Logging

The data logging function allows you to record waveform data collected in real time. The data is stored to a *.ROF file in the predefined target folder.
After entering the quick setup interface, you can tap Log to enter the interface as shown in the figure below. You can enable/disable the data logging function and set the target location to save the data.
Figure 10.6 Log Setting Interface
Storage Path
Tap Save to input field to access the Store and Recall interface. Select the desired storage path and then tap Save. Set the filename in the pop-up filename input field. The data is stored to a file with .ROF extension.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 60
Analyzer
TIP
You can store files in the instrument's internal memory or an external USB device. The
maximum size of log file saved in internal memory is 20M. The size of the file saved in the
external USB device depends on the memory size of the device.
Logger On/Off
Tap the Logger on/off toggle switch to turn on/off the logger.
Select ON to enable the data logging function. When turning on the Analyzer
(through tapping Run in the Analyzer main interface), the instrument stores the
waveform data collected in real time at the current sample rate. The logging
finishes as the analysis stops. The file is saved in the specified location.
Select OFF to disable the data logging function. When the Analyzer is on, the
system will not perform any data logging.
TIP
If the storage location is not changed, the data logged at different times will be saved to
the same file.
Record at least 1 point to save the file.

10.4 Print Screen

In the analyzer main interface, you can tap PrtScn to capture and save the current screen. By default, the screen images are stored in internal memory of the instrument. If a USB storage device is detected, the images will be saved to the USB device under the target folder. A dialog window will pop up indicating the file location. The saved file is in .BMP format.
The first screen image is named “Analyzer0.BMP” by default. The number next to “Analyzer” in the name increments by 1 each time a new image is captured and saved under the current path. For example, the second screen image is named “Analyzer1.BMP”.

10.5 Label Control

Tap label Control in the Analyzer main interface to hide or display the labels of analysis objects.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 61

Trigger

11
Trigger
DP900 series power supply supports trigger input and trigger output via the rear­panel Digital I/O interface (optional for DP932U and not available for DP932E).
Trigger Input:
The data lines of the digital I/O interface receive external trigger signal. The source under control (namely the selected output channel) turns on/off the output or inverts the output state when the preset trigger condition is met.
Trigger Output:
The data lines of the digital I/O interface output a high/low level signal when the control source (namely the selected output channel) turns on the output.
The 4 data lines of the digital I/O interface are mutually independent and can be used for trigger input or trigger output separately.
The digital port connections are shown in the figure below.
Figure 11.1 Digital Port Connections
Operation Procedures:
1. Connect the wires to the terminal block connector according to the figure above
(pay attention to the corresponding relations).
2. Insert the terminal block connector into the digital I/O terminal port on the rear
panel (pay attention to the corresponding relations).
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 62
Trigger
Tap >Trigger to access the trigger setting interface, as shown in the figure below. You can tap Trigger Input or Trigger Output to access the corresponding interface.

11.1 Trigger Input

When the specified data line receives input signal that meets the current the specified source under control will turn on/off the output, or toggle the output state according to the setting in
Tap > Trigger at the lower-left of the screen. Then select Trigger Input to open the trigger input setting interface, as is shown in the figure below.
Figure 11.2 Trigger Setting Interface
Output Response
.
Trigger Type
,
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 63
Trigger
Figure 11.3 Trigger Input Setting Interface
Data Line
Select a data line (D1, D2, D3, or D4) you wish to configure. The data line label is lit after selected. You can configure the trigger settings for the four data lines respectively.
Trigger Input On/Off
Tap the TrigIn on/off button to enable or disable the respective trigger input settings for the selected data line.
Source Under Control
In the Channel item, you can select one or more of CH1, CH2, and CH3 as the source under control by ticking the box.
Trigger Type
Tap the TrigType drop-down button to select to trigger on the rising edge, falling edge, high level, or low level of the input signal.
TIP
For the input signal, high level ranges from 2.5 V to 3.3 V, low level from 0 V to 0.8 V, and the
noise tolerance is 0.4 V.
Output Response
Tap the Response drop-down menu to set the trigger response type to "OutOn", "OutOff”, or “OutFlip”.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 64
OutOn: Turns on the output of the channel currently selected as the source
under control when the trigger condition is met.
OutOff: Turns off the output of the channel currently selected as the source
under control when the trigger condition is met.
OutFlip: Toggles the channel output state. That is, turns off the channel when the
current channel is on, or turns on the channel when the current channel is off.

11.2 Trigger Output

The specified data line outputs a high/low level signal when the control source turns on the output.
Trigger
Tap > Trigger and select Trigger Output to access the trigger output setting interface, as shown in the figure below.
Figure 11.4 Trigger Output Setting Interface
Data Line
Select a data line (D1, D2, D3, or D4) you wish to configure. The data line label is lit after selected. You can configure the trigger settings for the four data lines respectively.
Trigger Output On/Off
Tap the TrigOut on/off button to enable or disable the respective trigger output settings for the selected data line.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 65
Trigger
Control Source
Tap the
Output Response
Tap the Response drop-down button to set the trigger response type to “HighLevel” or “LowLevel”.
High Level: The selected data line outputs 3.3 V high level when the control
Low Level: The selected data line outputs low level signal (CMOS level) when the
Channel drop-down button to select CH1, Ch2, or CH3 as the control source.
source is turned on.
control source is turned on.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 66

Store and Recall

12
Store and Recall
You can store various types of files in internal or external memory, and recall the stored files when required. Besides, you can also copy, paste, or delete files of the specified type in internal or external memory.
This series power supply provides an internal non-volatile memory and two USB HOST interfaces (one on the front panel and the other on the rear panel), which can used to connect USB storage devices for external storage. The internal memory is C disk, and the external memory is D disk and E disk.
TIP
The instrument only supports the flash memory USB storage device of FAT32 format.
Press the front-panel or tap > Store to enter the interface as shown in the figure below.
Figure 12.1 Disk Interface
C Disk
Folders cannot be stored in C disk. The file types available for storage include state file (.RSF), log file (.ROF), Arb file (.RTF), bitmap file (.BMP), calibration file (.CLF), and option installation verification file (.LIC).
TIP
- Do not copy unsupported file type to C disk, or a prompt message “Filetype not
supported.” will be displayed.
- The C disk supports file locking. You can lock/unlock files via the SCPI
command :MEMory:LOCK <name>,<bool>. A lock icon will appear at the upper-
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 67
Store and Recall
D Disk and E Disk
The D and E disks are available only when a USB storage device (FAT32 format, flash memory) is detected at the USB HOST port. You can save state file, log file, Arb file, and bitmap file to the USB storage device as well as copy the files in C disk to the USB storage device. The number of files that can be stored depends on the memory size of the USB storage device.

12.1 Select

First select a specified file. Tap the file and it is displayed as . Tapping the file again can unselect the file.
right of the locked file which cannot be copied or deleted (but can be safe cleared).
For more details about the SCPI commands and programming, please refer to the
Programming Guide
of the series.
The series power supply allows you to select multiple files at once: tap at the
upper-right corner to select all files under the folder; tap operation.

12.2 Save

1. Press the front-panel or tap > Store to open the disk interface and
select the desired path.
2. Tap Save to pop up the filename setting interface, as is shown in the figure below.
again to cancel the
Enter your desired filename in the “FileName” input field. Then tap
3.
complete the entry and save the file, or tap
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
to cancel the operation. This
to
Page 68
Store and Recall
instrument supports both Chinese input method (T9) and English input method (uppercase/lowercase). In the filename setting interface, you can do the following to select your desired input method.
- Tap “CH” key to switch to the Chinese input method.
- Tap “1#” key in the keypad to enter numbers or symbols.
- Tap “ABC” key to enter upper-case English letters or tap “abc” key to switch
to the lower-case mode.
The save operation in the store/recall function is available only when the file is a “state file”.
To save log file, please refer to
To save Arb file, please refer to
To save bitmap file, please refer to
To save option installation verification file, please refer to

12.3 Read

1. Press the front-panel or tap > Store to access the disk interface. Enter
the target directory to select the desired file.
The read operation is available only when the file is “state file”. To read Arb file, please refer to
2. Then tap Read, and the file currently selected will be read. After the file is successfully read, a prompt message is displayed. The instrument will load the system parameters in the state file and reset the instrument with the loaded instrument settings.
Import and Export
Data Logging
.
Import and Export
Print Screen
.
.
.
Option Settings
.

12.4 Delete

1. Press the front-panel or tap > Store to access the disk interface. Enter
the target directory to select the file that you want to delete.
2. Tap Delete, and a prompt message “Delete selected files?” is displayed. Tap Yes
to delete the file.
TIP
You can also click or tap SafeClear in the C disk to delete all files (including locked files) in it.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 69
Store and Recall

12.5 Copy and Paste

1. Press the front-panel or tap > Store to access the disk interface. Enter
the target directory to select the desired file.
2. Tap Copy to copy the file currently selected.
3. Tap at the upper-left side of the screen to go back to the disk interface. Select
the directory to which you want to paste the file.
4. After entering the target folder, tap Paste to complete the paste operation. If the
file already exists in the current folder, a prompt message “File exists.” will pop
up.
You can copy files in the three disks to each other, except the following situations.
You are not allowed to copy folders.
You are not allowed to copy the file locked in C disk.
C disk has 2GB memory size. When it is running out of space, a prompt message
“Low disk space.” will be displayed.
The file copied to C disk cannot be larger than 2MB.
Unsupported file type cannot be copied to C disk.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 70

Utility

13
Utility
You can press the front-panel or tap > Utility to enter the interface as shown in the figure below. You can configure system setting, output setting, interface parameter setting, and option setting.
System: Configure system parameters, display, and so on.
Output: Set track mode and output connections.
Interface: Configure the LAN as well as view the USB information.
Option: View the current option installation information and install the option.

13.1 System Settings

Press > System or tap > Utility > System to enter the system settings interface.
Instrument Settings
Set the system parameters, such as Preset key, power-on settings, and beeper.
Display Options
Set the brightness and system language.

13.1.1 Instrument Settings

Figure 13.1 Utility Interface
Press > System > Instrument Settings or tap > Utility > System > Instrument Settings to access the interface as shown in the figure below. The
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 71
Utility
Instrument Settings interface allows you to configure the front-panel Preset key, beeper, and power-on settings.
Figure 13.2 Instrument Settings Interface
Beeper
Tap Beeper switch to toggle between ON and OFF to enable or disable the beeper sound. When it is turned on, the click heard is enabled when the touch screen is used or the front-panel keys and knob are used.
Power-on Setting
You can tap the Power On drop-down button to select the state that will be loaded at power-up. This can be either “Default” or “Last”. The default setting is "Default”.
Default: Uses the factory setting, except for the parameters that are not affected by factory reset.
Last: Loads the system state at the last power-off.
Preset Key Function
Tap Preset to define the function of the front-panel “Preset” key. You can select
“Default”. Press on the front panel and a prompt message “Restore the instrument to factory default settings?” is displayed. Tap Yes to restore to factory default.
Table 13.1 Factory Default Values
Channel Settings
DP932A, DP932U
Voltage/Current Settings CH1/CH2/CH3: 0 V/0.1 A
Voltage/Current Limit Settings CH1: 33.6 V/3.3 A
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 72
Channel Settings
CH2: 33.6 V/3.3 A
CH3: 6.3 V/3.3 A
OVP/OCP On/Off CH1/CH2/CH3: OFF/OFF Output On/Off CH1/CH2/CH3: OFF
DP932E
Voltage/Current Settings CH1/CH2/CH3: 0 V/0.1 A
CH1: 31.5 V/3.3 A
Utility
Voltage/Current Limit Settings
OVP/OCP On/Off CH1/CH2/CH3: OFF/OFF Output On/Off CH1/CH2/CH3: OFF
Arb
Repetition Cycle 1 End State Last Run Mode Continuous Trigger Source Manual Triggered Channel CH1 Group Number 0 Waveform Template Sine Object Voltage Constant Current 1 A Max 1 V Min 0 V Groups 50 Interval 1 s Period
CH2: 31.5 V/3.3 A
CH3: 6.3 V/3.3 A
(Sine, Ramp, StairUp, StairDn, StairUpDn)
Invert OFF Positive Pulse Width 25 s Symmetry 50% Steps 25 Exp 10
Analyzer
Display Empty Object NA Channel NA Method RT Plot
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
50 s
DP900 User Guide
Page 73
Utility
Analyzer
Channel (Pulse Current Analysis) NA Positive/Negative Pulse NA Positive Pulse 1 A Negative Pulse 0.1 A Logger OFF Save Path NA
Trigger
Trigger Input
Trigger Input On/Off OFF Source Under Control NA Trigger Type Rising edge Output Response Output off
Trigger Output
Trigger Output On/Off OFF Control Source CH1 Output Response High Level
Utility
Preset Key Function Default Power-on Setting Default Beeper ON Language Chinese Brightness 50% Tracking OFF Safe Mode ON CH-Off Mode 0 V Channel Connection Independent DHCP ON AUTO ON MANUAL OFF
Store
Current Directory C:/

13.1.2 Display Options

Press > System > Display Options or tap > Utility > System > Display Options to enter the interface as shown in the figure below.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 74
Figure 13.3 Display Options Interface
Utility
Language
The series power supply supports multiple languages. You can tap the Language drop-down button to select the system language.
Brightness
Use the slider of Brightness to adjust the brightness of the display. The range is from 1% to 100%, and the default is 50%. This setting is non-volatile; it will not be changed by factory reset.

13.2 Output Settings

The Output Settings interface allows you to set the track mode, output off status, safe mode, and channel connection status.
Press > Output or tap > Utility > Output to enter the interface as shown in the figure below.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 75
Utility
Figure 13.4 Output Settings Interface

13.2.1 Track Mode

The tracking function is usually used to provide symmetrical voltages needed by operational amplifiers and other circuits. For this series power supply, the tracking function is available for CH1 and CH2. You can enable or disable the track mode for those two channels.
Tap the Tracking on/off button to toggle between ON and OFF to enable or disable the track mode.
ON: Enables the track mode. For the two channels (CH1 and CH2) that support
this mode, changes made on one channel (voltage/current setting value,
OVP/OCP level and on/off status) is applied to the other channel.
OFF: Disables the track mode. For the two channels (CH1 and CH2) that support
the track function, changes made on one channel will not affect the other.
TIP
The tracking function only tracks the voltage/current setting value. The actual output voltage/
current will not be affected.

13.2.2 Safe Mode

Tap the SafeMode button to toggle between ON and OFF to enable or disable the safe mode.
ON: Enables the safe mode. When CH1 and CH2 are connected in series (internal) and the sum of their voltages is higher than 60 V, a prompt message
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 76
“Over safety voltage, sure set it?” will be displayed. Tap Yes to ignore the message or tap No to prevent the output voltage from exceeding 60 V.
OFF: Disables the safe mode. You are free to set the voltage.
WARNING
When the output voltage exceeds 60 V in series connection, please do not touch the
output terminals.

13.2.3 Output Connection

This series power supply supports internal series or parallel connection for CH1 and CH2. Tap the CH-Connection drop-down button to select “Series”, “Parallel”, or “IND” to control the channel connection state.
When “IND” is selected, CH1 and CH2 are independent of each other.
Utility
When “Series” is selected, CH1 and CH2 are connected in series. The output voltage of CH1 is the sum of both CH1 and CH2 voltages, but CH2 is no longer working. For connections in internal series mode, please refer to
Connections
When “Parallel” is selected, CH1 and CH2 are connected in parallel. The output current of CH1 is the sum of both CH1 and CH2 currents, but CH2 is no longer working. For connections in internal parallel mode, please refer to
Connections
TIP
In internal series/parallel connections, CH1 and CH2 always have identical voltage and current
setting value.
.
.

13.2.4 Output Off Mode

Tap the CH-Off Mode drop-down menu to select the output state when channels are off. Available selections include “0 V”, “IMM”, and “Delay.”
0 V: Outputs 0 V voltage when channels are off.
Series
Parallel
IMM: Clears the circuit immediately and turn off outputs when channels are off.
Delay: Uses the turn-off delay when channels are off.
TIP
When "IMM" is selected, the voltage fall time is not guaranteed.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 77
Utility

13.3 Interface Settings

DP900 series supports remote interface communication over interfaces including USB and LAN. You can control the instrument remotely via those interfaces.
Press > Interface or tap > Utility > Interface to enter the interface and set the I/O parameters for remote operations.

13.3.1 LXI Status

Before using the LAN interface, please use the network cable to connect the instrument to your PC or to the network of the PC.
In the Interface Settings menu, tap LXI Status to access the interface as shown in the figure below. You can tap LXI Status again to collapse the current menu to revert to the Interface Settings menu.
Figure 13.5 Interface Settings
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 78
Figure 13.6 LXI Status Interface
Utility
The interface displays the network status, IP source, IP address, MAC address, and VISA.
LXI Status
- Network access
- Network is not connected
- No network adapter
- Network initialization process
- Network configuration is successful
- Failed to obtain IP from DHCP server
- The IP assigned from DHCP is lost
- IP address conflict
- Invalid IP address
- DNS service detection success
IP Source
It displays the IP configuration information: IP not obtained, DHCP, or AUTO.
You can set the IP source in
IP address
It displays the IP address of the instrument.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
LXI Setting
interface.
DP900 User Guide
Page 79
Utility
MAC address
It displays the MAC address of the instrument.
The Media Access Control (MAC) address, also called the hardware address, is used to define the location of the network device. For a power supply, the MAC address is always unique. When attributing IP address for the instrument, the MAC address is usually used to identify the instrument. The MAC address (48 bits, namely 6 bytes) is usually expressed in hexadecimal form, for example, 00-2A-A0-AA-E0-56.
VISA
Virtual Instrument Software Architecture (VISA), an advanced application programming interface developed by National Instrument (NI), is used to communicate with various instrument buses. It uses the same method to communicate with the instrument regardless of the type of the instrument interface (GPIB, USB, LAN/Ethernet, or RS232). The instrument communicating with it via the GPIB, USB, LAN/Ethernet, or RS232 interface is called "Resource".
VISA descriptor is the resource name and describes the accurate name and location of the VISA resource. If the LAN interface is currently used to communicate with the instrument, the VISA descriptor is TCPIP::172.18.8.41::INSTR.

13.3.2 LXI Setting

In the Interface Settings menu, tap LXI Config to access the menu as shown in the figure below. You can tap LXI Config again to collapse the current menu to revert to the Interface Settings menu.
Figure 13.7 LXI Settings Menu
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 80
IP Configuration
The TCP/IP address information for the instrument can be obtained in different ways, including DHCP, Auto IP, and Manual IP. You can tap the DHCP, AUTO, or MANUAL on/off switch to enable or disable the configuration. When ON is selected, the corresponding configuration mode is enabled. When OFF is selected, the corresponding configuration mode is disabled.
- DHCP: The Dynamic Host Configuration Protocol (DHCP) can automatically
assign a dynamic IP address to the instrument from a DHCP server. By default, the instrument is configured to use DHCP.
- Auto IP: When “AUTO” is “ON”, it obtains the IP address (169.254.0.1
to 169.254.255.254) and subnet mask (255.255.0.0) depending on the current network configuration.
- Manual IP: When “MANUAL” is “ON”, you need to set the network
parameters (such as IP address) manually.
Utility
TIP
- When all the three configuration modes are set to "ON", the priority of parameter
configuration is "DHCP", "Auto IP", and "Manual IP". Therefore, to set the address
manually, DHCP and Auto IP should be set to "OFF".
- They cannot be "OFF” at the same time.
IP Address
Contact your network administrator for a valid IP address and set the address manually. The format of the IP address is nnn.nnn.nnn.nnn; wherein, the first nnn ranges from 1 to 223 (excluding127), and the other three range from 0 to 255.
Enter the desired IP address using the front-panel parameter input area or the virtual keypad. This setting is non-volatile. The instrument will load your IP address automatically at the next power-on if “DHCP” and “Auto IP” are set to "OFF".
Gateway
Contact your network administrator for a valid gateway address and set the default gateway manually. The format of the address is nnn.nnn.nnn.nnn; wherein, the first nnn ranges from 1 to 223 (excluding 127), and the other three range from 0 to 255.
Enter the desired gateway address using the front-panel parameter input area or the virtual keypad. This setting is non-volatile. The instrument will load your gateway address automatically at the next power-on if “DHCP” and “Auto IP” are set to "OFF".
Subnet Mask
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 81
Utility
Contact your network administrator for a valid subnet mask and set it manually. The format of the subnet mask address is nnn.nnn.nnn.nnn; wherein, the range of the nnn is from 0 to 255.
Enter the desired subnet mask using the front-panel parameter input area or the virtual keypad. This setting is non-volatile. The instrument will load your subnet mask automatically at the next power-on if “DHCP” and “Auto IP” are set to "OFF".
DNS
Contact your network administrator for a valid DNS address and set the address manually. The format of the DNS address is nnn.nnn.nnn.nnn; wherein, the first nnn ranges from 1 to 223 (excluding 127), and the other three range from 0 to
255.
Enter the desired DNS address using the front-panel parameter input area or the virtual keypad. This setting is non-volatile. The instrument will load your DNS address automatically at the next power-on if “DHCP” and “Auto IP” are set to "OFF".
Default Settings
Tap Default and a prompt message “Restore to default network settings” is displayed. This operation enables DHCP and Auto IP and disables Manual IP.
Apply
Tap Apply to apply the current settings and revert to the Interface Settings menu.
TIP
Note that you must tap Apply to save and apply the network settings.

13.3.3 USB Configuration

In the Interface Settings menu, tap USB Config to enter the interface as shown in the figure below. The interface displays the USB information.
You can tap USB Config again to collapse the current menu to revert to the Interface Settings menu.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 82
Figure 13.8 USB Configuration Interface
Utility

13.4 Option Settings

This series power supply provides multiple options to fulfill your measurement requirements. If you need any of these options, please order the option according to the Order No. available in referring to this section. Besides, you can also view the options currently installed on the instrument or activate the newly purchased option.
View the Installed Option
Press > Option or tap > Utility > Option to enter the interface as shown in the figure below. You can view the installation information of all options in the interface.
Appendix A: Accessories and Options
, and then install it by
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 83
Utility
Figure 13.9 Option Settings Interface
Install the option
The option license is a string o format, with the file extension “.lic”. After you purchase an option, you will obtain a key (used for obtaining the desired option license). Please install the option following the steps below.
1. Obtain an option license
a. Log in to the RIGOL official website ( ), and click
CENTRE > License Activation t
interface.
b. In the software license registration interface, enter the correct key, serial
number (tap code. Then click need to use the file, please click the link to download the file to the root directory of the USB storage device.
> Help > About t
Generate t
f fixed characters. The option file should be in specific
SERVICE
o enter the software license registration
o obtain the serial number), and verification
o obtain the option license file download link. If you
2. Install the option
a. Connect the USB storage device to the instrument. Press the front-panel
or tap > Store t option settings int
b. Select the desired file and tap
c. After installation, a prompt message "Option activated successfully" is
displayed. After the option has been installed, it is recommended to restart the instrument.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
o enter the disk interface. You can also tap
erface to access the disk interface.
Read t
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
o install the option.
Install in the
Page 84
TIP
During the installation process, you are not allowed to power off the instrument or
unplug the USB storage device.
You can send SCPI commands to install options (please refer to
Guide
). Installing options by entering the license code manually is not supported.
DP900 Programming
Utility
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 85

Locking/Unlocking the Front Panel

14
Locking/Unlocking the Front Panel
To avoid any adverse effect caused by misoperation, you can long press the front-
panel to lock the touch screen. After the touch screen is locked, a lock icon
appears in the status area at the lower-right of the screen. You can long press again to disable the touch lock.
You can also send SCPI commands to lock the display or front-panel keys.
Command Description
:SYSTem:TLOCk {ON|OFF|0|1} /*Enables or disables the touch lock.*/
:SYSTem:TLOCk? /*Queries whether the display is locked.*/
:SYSTem:KLOCk:STATe {ON|OFF|0|1} /*Enables or disables the remote lock mode.*/
:SYSTem:TLOCk? /*Queries whether the remote lock is enabled.*/
In remote lock mode, all front-panel keys are disabled except for the output on/off
keys , all output on/off key , and the power switch key . At this point, you can press the output on/off key only to disable channel output.
Three Methods to Lock/Unlock the Keys Using SCPI Commands
• Edit a command file that meets the following requirements and store it to a
USB device; read this file in the instrument to lock or unlock the specified
key.
The content and format requirements of the command file are as follows.
- The filename suffix must be ".scpi".
- The filename cannot be longer than 20 characters.
- Content requirements:
The first line: the command used to lock or unlock the specified key.
The second line: blank (cannot be omitted).
TIP
The command file can contain multiple lines (for example, N lines). The first N-1 lines are
the commands used to lock or unlock the specified keys and the Nth line is blank. At this
point, you can execute locking /unlocking operations N-1 times by reading the file in the
instrument.
Operation Procedures:
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 86
Locking/Unlocking the Front Panel
1. Edit a command file that meets the above requirements and store it to a USB
storage device.
2. Start the instrument and insert the USB storage device. Press the front-panel
or tap >Store to open the store and recall interface.
3. In that interface, read the command file according to
Send command via Ultra Sigma to lock or unlock the specified key
- Set up communication between the instrument and PC.
- Run Ultra Sigma and search for the instrument resources.
- Open the remote command control panel to send commands.
Program and send command to lock or unlock the specified key.
- Set up communication between the instrument and PC.
- Use a programming language (e.g. LabVIEW, C#) to program and send
command.
Store and Recall
.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 87

Remote Control

15
Remote Control
The instrument can be remotely controlled by the following three methods:
User-defined Programming
You can program and control the instrument by using the SCPI (Standard Commands for Programmable Instruments) commands. For details about the SCPI commands and programming, please refer to the the series.
PC Software
You can use the PC software to send commands to control the instrument remotely. Ultra Sigma of RIGOL is recommended. You can log in to RIGOL official website ( ) to download the software.
Operation Procedures:
- Set up communication between the instrument and PC.
- Run Ultra Sigma and search for the instrument resources.
Programming Guide
of
- Open the remote command control panel to send commands.
Web Control
This instrument supports Web Control. When connected to the network, the Web Control will display the real-time interface displayed on the instrument screen. Through the Web Control method, you can migrant the device control to the control terminals (e.g. PC, Mobile, iPad, and other smart terminals) to realize remote control of the instrument. When you first log in to the Web Control, the user name is "admin" and password is "rigol".
This instrument can be connected to the PC via the USB or LAN interface to set up communication and realize remote control through the PC. The remote control is realized via SCPI commands.
This chapter will illustrate how to use the RIGOL Ultra Sigma software to remotely control the instrument via various interfaces.
TIP
e setting up communication, please turn off the instrument to avoid causing damage to
Befor
the communication interfaces.

15.1 Remote Control via USB

1. Connect the device
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 88
Remote Control
Use the USB cable to connect the rear-panel USB DEVICE interface of the instrument to the USB HOST interface of the PC.
2. Search for the device resource
Start up Ultra Sigma and the software will automatically search for the resource currently connected to the PC via the USB interface. You can also click USB-TMC to search for the resource.
3. View the device resource
The resources found will appear under the "RIGOL Online Resource" directory, and the model number and USB interface information of the instrument will also be displayed.
4. Control the instrument remotely
Right-click the device resource name and select "SCPI Panel Control" to open the remotely command control panel. Then you can send commands and read data through the panel. For details about the SCPI commands and programming, refer to the Programming Guide of this instrument.

15.2 Remote Control via LAN

1. Connect the device
Use the network cable to connect the instrument to your local area network (LAN).
2. Configure network parameters
Configure the network parameters of the instrument according to
Settings
3. Search for the device resource
Start up Ultra Sigma and click LAN to open the window as shown in the figure below. Click Search and the software searches for the instrument resources currently connected to the LAN. The resources found are displayed at the right section of the window as shown in the figure below. Click OK to add it.
.
Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 89
Remote Control
Besides, you can input the IP address of the instrument manually into the text field under “Manual Input LAN Instrument IP”, and then click TEST. If the instrument passes the test, click ADD to add the instrument to the LAN instrument resource list at the right section of the window; if the instrument fails the test, please check whether the IP address that you input is correct, or use the auto search method to add the instrument resource.
4. View the device resource
The resources found will appear under the "RIGOL Online Resource" directory.
5. Control the instrument remotely
Right-click the device resource name and select “SCPI Panel Control” to open the remote command control panel. Then you can send commands and read data through the panel.
6. Load LXI webpage
As this instrument conforms to LXI CORE 2011 DEVICE standards, you can load LXI webpage through Ultra Sigma (right-click the instrument resource name and select "LXI-Web"). Various important information about the instrument (including the model, manufacturer, serial number, description, MAC address, and IP address) will be displayed on the webpage. You can also directly input the IP address of the instrument in the address bar of the PC browser to load the LXI webpage.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 90

Troubleshooting

16
Troubleshooting
The following failures may occur during the use of the instrument. Please first try to handle the problem according to the following methods; if the problem remains unsolved, please contact RIGOL and provide the device information of your
instrument (by tapping > Help > About).
1. The instrument does not start.
a. Check whether the power cord is correctly connected.
b. Check whether the power switch on the front panel is turned on.
c. Unplug the power cord, and check whether the voltage selector is at the proper
scale, whether the fuse specification is correct and the fuse is in good condition.
To replace the fuse, please refer to
d. If the problem still persists, please contact RIGOL.
2. The constant voltage output is abnormal.
Replacing the Fuse
a. Check whether the maximum output power of the selected meets the load
requirement. If the load requirement is met, go to the next step.
b. Check whether the cable connecting the load and power supply is short-
circuited and whether it is in good contact.
c. Check whether the load works normally.
d. Check whether the current setting value of this scale is proper; if it is too low,
increase the current setting value properly.
e. If the problem still persists, please contact RIGOL.
3. The constant current output is abnormal.
a. Check whether the maximum output power of the selected meets the load
requirement. If the load requirement is met, go to the next step.
b. Check whether the cable connecting the load and power supply is short-
circuited and whether it is in good contact.
c. Check whether the load works normally.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 91
Troubleshooting
4. The USB storage device cannot be recognized.
d. Check whether the voltage setting value of this scale is proper; if it is too low,
increase the voltage setting value properly.
e. If the problem still persists, please contact RIGOL.
a. Check whether the USB storage device can work normally.
b. Make sure that a flash memory USB storage device of FAT32 format is used, as
this instrument does not support hard disk USB storage device.
c. Restart the instrument and insert the USB storage device to check.
d. If the USB storage device still cannot work normally, please contact RIGOL.
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 92

Specifications

17
Specifications
All the specifications continuously for more than 30 minutes under the specified operation temperature.
Number of Channels
Model Number of Channels
DP932A 3
DP932U 3
DP932E 3
DC Output (0℃~40℃)
DC output (0℃~40℃)
[1]
can only be guaranteed when the instrument is operated
Voltage/Current OVP/OCP
CH1 0 to 32 V/0 to 3 A
1 mV to 35.2 V/1 mA to 3.3 A
DP932A/ DP932U
DP932E
Internal Series/Parallel Mode
Internal series/parallel mode
Series mode voltage 64 V
Parallel mode current 6 A
Load Regulation Rate
CH2 0 to 32 V/0 to 3 A
CH3 0 to 6 V/0 to 3 A 1 mV to 6.6 V/1 mA to 3.3 A
CH1 0 to 30 V/0 to 3 A 1 mV to 33 V/1 mA to 3.3 A
CH2 0 to 30 V/0 to 3 A 1 mV to 33 V/1 mA to 3.3 A
CH3 0 to 6 V/0 to 3 A 1 mV to 6.6 V/1 mA to 3.3 A
1 mV to 35.2 V/1 mA to 3.3 A
Load regulation rate, ± (% of output + offset)
Voltage
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
[2]
www.GlobalTestSupply.com
<0.01%+2 mV
DP900 User Guide
Page 93
Specifications
Load regulation rate, ± (% of output + offset)
Current <0.01%+250 μA
Line Regulation Rate
Line regulation rate, ± (% of output + offset)
Voltage <0.01%+2 mV
Current <0.01%+250 μA
Output Ripple and Noise
Output ripple and noise (20 Hz to 20 MHz)
Normal mode voltage <350 μV
Normal mode current <2 mA
Accuracy 12 months (25ºC±5ºC)
Accuracy 12 months (25ºC±5ºC)
Programming Readback Voltage Current Voltage Current
CH1 0.05%+10 mV 0.2%+5 mA 0.05%+10 mV
DP932A/DP932E
CH2 0.05%+10 mV 0.2%+5 mA 0.05%+10 mV
CH3 0.1%+5 mV 0.2%+5 mA 0.1%+5 mV
CH1 0.05%+20 mV 0.2%+5 mA 0.05%+20 mV
/2 mV
rms
rms
[3]
, ± (% of output + offset)
pp
0.15%+5 mA
0.15%+5 mA
0.15%+5 mA
0.15%+5 mA
DP932U
CH2 0.05%+20 mV 0.2%+5 mA 0.05%+20 mV
CH3 0.1%+5 mV 0.2%+5 mA 0.1%+5 mV
0.15%+5 mA
0.15%+5 mA
Resolution
Resolution
Programming Readback Display
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 94
Resolution
Specifications
Voltage Current Voltage Current Voltage Current
CH1 1 mV 1 mA 0.1 mV 0.1 mA 1 mV 1 mA
DP932A Standard
Standard
DP932U
With high resolution option
Standard
DP932E
With high resolution option
CH2 1 mV 1 mA 0.1 mV 0.1 mA 1 mV 1 mA
CH3 1 mV 1 mA 0.1 mV 0.1 mA 1 mV 1 mA
CH1 10 mV 1 mA 10 mV 1 mA 10 mV 1 mA
CH2 10 mV 1 mA 10 mV 1 mA 10 mV 1 mA
CH3 10 mV 1 mA 10 mV 1 mA 10 mV 1 mA
CH1 1 mV 1 mA 0.1 mV 0.1 mA 1 mV 1 mA
CH2 1 mV 1 mA 0.1 mV 0.1 mA 1 mV 1 mA
CH3 1 mV 1 mA 0.1 mV 0.1 mA 1 mV 1 mA
CH1 10 mV 10 mA 10 mV 10 mA 10 mV 10 mA
CH2 10 mV 10 mA 10 mV 10 mA 10 mV 10 mA
CH3 10 mV 10 mA 10 mV 10 mA 10 mV 10 mA
CH1 1 mV 1 mA 0.1 mV 0.1 mA 1 mV 1 mA
CH2 1 mV 1 mA 0.1 mV 0.1 mA 1 mV 1 mA
CH3 1 mV 1 mA 0.1 mV 0.1 mA 1 mV 1 mA
Transient Response Time
Transient response time
Less than 50 μs of time to recover to within the ±15 mV settling band following a load change from 50% to 100% or from 100% to 50% of full load.
Command Processing Time
Command processing time
[4]
<10 ms
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 95
Specifications
OVP/OCP Accuracy
OVP/OCP accuracy, ± (% of output + offset)
OVP accuracy, ± (% of output + offset)
OCP accuracy, ± (% of output + offset)
Voltage Programming Response Time (Within 99% of the Total Variation Range)
Voltage programming response time (within 99% of the total variation range)
Channel Full Load (Up) No Load (Up) Full Load (Down) No Load (Down)
CH1 <50 ms <40 ms <50 ms <400 ms
CH2 <50 ms <40 ms <50 ms <400 ms
CH3 <15 ms <14 ms <30 ms <100 ms
Temperature Coefficient Per ºC
Temperature coefficient per ºC, ± (% of output + offset)
Channel Voltage Current
0.2%+20 mV
0.5%+20 mA
CH1 0.01%+4 mV 0.01%+2 mA
CH2 0.01%+4 mV 0.01%+2 mA
CH3 0.01%+4 mV 0.01%+3 mA
Mechanical Characteristics
Mechanical characteristics
Dimension 239 mm (W) x 157 mm (H) x 419 mm (D)
Weight 9.15 kg
Rack mount kit Three rack-units (3U), 1/2-rack form factor
Power Source
Power source
100 Vac±10%
AC input (50 Hz to 60 Hz)
120 Vac±10%
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 96
Specifications
Power source
220 Vac±10%
230 Vac±10% (max. 250 Vac)
Maximum input power 650 VA
Interface
Interface
USB DEVICE 1
USB HOST 2 (1 for front panel and 1 for rear panel)
LAN 1
Digital IO
Environmental Conditions
Environmental conditions
Cooling Method Fan cooling
Operating Temperature 0°C to +40°C
Storage Temperature -40°C to +60°C
Humidity 5% to 80% relative humidity
Altitude Below 1500 meters
IP Rating IP20
Pollution Degree PD2
1 (optional for DP932U; not available for DP932E)
Overvoltage Categories OVC II
Operating Environment
Warranty and Calibration Interval
Warranty and calibration interval
Warranty 3 years (excluding the accessories)
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
For indoor use only and non-operating in humid environment.
DP900 User Guide
Page 97
Specifications
Warranty and calibration interval
Recommended calibration interval
Note[1]:
Unless otherwise stated, the specifications are applicable to all the channels of
the specified model.
Not applicable in series/parallel connection mode.
Note[2]: Limited by the terminal structure, the voltage load regulation rate cannot be
guaranteed for DP932U.
Note[3]: The accuracy parameters are acquired via calibration under 25℃ after 1­hour warm-up.
Note[4]: the time required for the output to change accordingly after receiving the APPLy and SOURce commands.
Every 12 months
Find Quality Products Online at: sales@GlobalTestSupply.com
DP900 User Guide
www.GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Page 98

Appendix

18
Appendix

18.1 Appendix A: Accessories and Options

Order Information Order No.
Base Unit
Programmable linear DC power supply, triple-output, high resolution
Programmable linear DC power supply, triple-output, University, safety terminals
Programmable linear DC power supply, triple-output, E­commerce
Standard Shipped Accessory
USB cable CB-USBA-USBB-FF-150
DP932A
DP932U
DP932E
One fuse — —
Power cord (based on destination country) — —
Two pairs of connecting wires (10 A) 10A-Testing-Cable
Optional Accessory
1 mA&1 mV high-resolution setting DP900-HIRES
Arbitrary function with the minimum dwell time of 100 ms
DP900-ARB
(available on DP932U only)
4-pin trigger in/out function
DP900-DIGITALIO
(available on DP932U only)
DP900 Rack Mount Kit (for a single instrument) RM-1-DP800
DP900 Rack Mount Kit (for two instruments side-by-side) RM-2-DP800
NOTE
For purchasing models, accessories, and options, please contact local RIGOL office.
Find Quality Products Online at: sales@GlobalTestSupply.com
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
www.GlobalTestSupply.com
DP900 User Guide
Page 99
Appendix

18.2 Appendix B: Warranty

RIGOL TECHNOLOGIES CO., LTD. (hereinafter referred to as RIGOL) warrants that the product mainframe and product accessories will be free from defects in materials and workmanship within the warranty period. If a product proves defective within the warranty period, RIGOL guarantees free replacement or repair for the defective product.
To get repair service, please contact your nearest RIGOL sales or service office.
There is no other warranty, expressed or implied, except such as is expressly set forth herein or other applicable warranty card. There is no implied warranty of merchantability or fitness for a particular purpose. Under no circumstances shall RIGOL be liable for any consequential, indirect, ensuing, or special damages for any breach of warranty in any case.