RigExpert AA-2000 ZOOM Instructions manual

AA-2000 ZOOM
Antenna and cable analyzers
RigExpert
®
User’s manual
For latest manuals and sof tware updates,
please visit
http://rigexpert.com
Table of contents
Introduction Operating the AA-2000 ZOOM
First time use Main menu Multifunctional keys Connecting to your antenna SWR chart Chart ZOOM Data screen Frequency and range entr y Return loss char t R,X chart Smith chart Memory operation SWR mode Display all parameters MultiSWR mode
Applications
Antennas Coaxial lines Measurement of other elements
Annexes
Annex 1: Specifications Annex 2: Precautions Annex 3: Tools menu Annex 4: Setup menu Annex 5: TDR mode Annex 6: Calibration Annex 7: Dummy loads
4 5 5 5 6 6 7 7 8 8 8 9 9 10 10 11 12 13 13 14 21 24 24 25 26 31 32 36 38
Introduction
Thank you for purchasing a RigExpert AA-2000 ZOOM Antenna an d Ca bl e Analyz er ! We
did our best to make it powerful yet easy to use.
The analyzer is designed for measuring SWR (standing wave ratio), return loss, cable loss, as well as other parameters of cable and antenna systems in the range of 100 kHz to 2000 MHz. A built-in ZOOM capability makes graphical measurements especially effective. An integrated Time Domain Reflectometer mode can be used to locate a fault within the feedline system. The analyzer is equipped
Bluetooth Low Energy
with a wireless connection with your laptop, tablet or smartphone.
The following tasks are easily accomplished by using this analyzer:
• Rapid check-out of an antenna
• Tuning an antenna to resonance
• Comparing characteristics of an antenna before and after specific event (rain, hurricane, etc.)
• Making coaxial stubs or measuring their parameters
• Cable testing and fault location, measuring cable loss and characteristic impedance
• Measuring capacitance or inductance of reactive loads
module for a
1
2
3
4
1. Antenna connector
2. Color display
3. Keypad
4. USB connector
4
User’s manual
First time use
Operating the
AA-2000 ZOOM
Please insert three AA sized batteries (alkaline , Ni-MH or LiPO accumulators) into the battery compartment of the analy zer, watching the polarity.
Press the on the analyzer. After displaying the initial message (showing a firmware version and a serial number of the instrument), a Main menu appears on the screen.
(Power) key located at the bottom-right corner of the keypad to turn
The analyzer will be turned off automatically if not in use for too long.
Main menu
The Main menu acts as a starting point from where dif ferent tasks may be launched.
(Cursor up) and (Cur sor
Use down) keys to scroll through the menu, then press (OK) to select an item. For your convenience, a battery indicator is shown at the top-left corner of the sc reen. T hi s in dicat or is rep la ce d with a USB icon when the analyzer is connected to your computer.
You may use hot keys for the quick access to certain tasks. For instance, press
the
(SWR chart) button to open the SWR chart screen immediately.
RigExpert AA-2000 ZOOM
5
Multifunctional keys
Most keys on the analyzer’s keypad perform several functions.
For instance, numbers (1) are used to enter frequency and other numerical parameters. Ma in func ti ons (2) pr ovide quick ac cess to most common tasks. Alternative functions (3) are executed if the user holds the key. For the convenience, alternative functions are marked with yellow.
You may press the (Help) key to open a help screen listing all active hot keys.
(Functional)
3
2
1
Connecting to your antenna
Plug the cable to your analyzer’s antenna connector, and then tighten the rotating sleeve. The rest of the connector, as well as the cable, should remain stationary.
If you twist other parts of the connector when tightening or loosening, damage may
easily occur. Twisting is not allowed by design of the N-connector.
6
User’s manual
SWR chart
Once your antenna is connected to the analyzer, it is time to measure its characteristics. Press the (SWR chart) key to open the SWR chart screen, then press (OK) to start a new measurement.
A few moments later, the result will be displayed on the analyzer’s screen.
Press the run a continuous sweep.
+ key combination to
A small triangle at the bottom of the chart corresponds to a point at which the SWR
reaches its minimum.
Chart ZOOM
Use the arrow keys to increase or decrease the center frequency or the scanning range.
Watch the chart zooming in or out, or changing its position. Use the (Functional
key) and (Cursor up) or (Cursor down) key combination to zoom the vertical
scale of the char t.
Do not forget to press the (OK) key for the new measurement to start.
Press
RigExpert AA-2000 ZOOM
(Functional key) and to quickly choose a radio amateur band.
7
Data screen
The data screen is available in all chart modes. Press the display various parameters of a load at cursor.
(Data) key to
Frequency and range entry
To enter the center frequency or the sweep range, press the (Frequency, Range) key.
Use cursor keys to navigate, or the to keys to en ter val ues. Do not forg et to press the (OK) key to apply.
Press (Up) or (Down) cursor keys while holding
the (Functional) key to quickly choose a radio amateur band.
Return loss chart
The return loss (RL) chart, which is very similar to the SWR chart, is activated by pressing the (Functional key) and (RL chart) key combination in the Main
menu.
8
User’s manual
R,X chart
Press the (R,X chart) key in the Main menu to acc es s the R,X cha rt mode.
Positive values of reactance (X) correspond to inductive load, while negative values correspond to capacitive load.
The chart will display R and X for series or parallel models of a load. Press (Functional key) and to switch between these models.
The marker at the bottom of the screen shows a resonant
frequency closest to the center of the scan.
Smith chart
The (Smith chart) key opens a screen where the reflection coefficient is plotted on the Smith char t.
A small marker is used to indicate the center frequency.
RigExpert AA-2000 ZOOM
For a list of hot keys, press the (Help) key, as usual.
9
Memory operation
Device has 250 unified slots for storing your records. Record operations are linked to relevant data screens - SWR, RX, RL, Smith charts mark their records with ‘Z’ slot identificator, TDR char t marks with ‘T’ identificator.
Use , , or and respectively for
Load, Save or Edit Name operations in relevant data screens.
SWR mode
To watch the SWR at a single frequency, press the (SWR) key. Do not forget to press the (OK) key to start or stop the measurement. Change the frequency with (Left) or (Right) cursor keys, or press the (Frequency) key to enter a new frequency.
Use and to switch between meter and trend view.
10
User’s manual
Display all parameters
To display various parameters of a load on a single screen, press the (All) key.
Do not be confused by negative values of L or C. This can be useful for
This screen displays values for
experienced users.
series
as well as
parallel
of impedance of a load.
series
• In the expressed as resistance and reactance connected in series:
model, impedance is
• In the
parallel
expressed as resistance and reactance connected in parallel:
model, impedance is
R
Z = R + jX
Z = R ||+ jX
X
RigExpert AA-2000 ZOOM
models
R X
11
MultiSWR mode
Press the (Functional key) and (Multi) key combination to see the SWR at up to five different frequencies. This mode may be useful for tuning multi­band antennas.
(Up) and (Down) cursor key s to select a frequency to be set or changed,
Use
then press the (Frequency) key to enter a new value. Do not forget to press the
(OK) key to start the measurement.
12
User’s manual
Applications
Antennas
Checking the antenna
It is a good idea to check an antenna before connecting it to the receiving or transmitting equipment. The SWR chart mode is good for this purpose. Also you can use Multé tool located in Too ls . Multé will check your rig against all HAM bands defined in your region and rate all bands where SWR lower than 3.0 in range of 1...5 stars. Select band for detailed evaluation and press to run SWR char t for this band.
The picture on the lef t shows the result for one band real antenna.
The nex t screen shot shows result for 50 Ohm calibrated load, making in such way ideal all-band antenna (in mathematical terms only).
RigExpert AA-2000 ZOOM
13
Adjusting the antenna
When the measurement diagnoses that the antenna is off the desired frequency, the analy zer can help in adjusting it. Physical dimensions of a simple antenna (such as a dipole) can be adjusted knowing the actual resonant frequency and the desired one. Other types of antennas may contain more than one element to adjust (including coils, filters, etc.), so this method will not work. Instead, you may use the SWR mode, the All parameters mode or the Smith chart mode to continuously see the results while adjusting various parameters of the antenna.
For multi-band antennas, use the MultiSWR mode. You can easily see how changing one of the adjustment elements (trimming capacitor, coil, or physical length of an aerial) affects SWR at up to five dif ferent frequencies.
Coaxial lines
Open- and short-circuited cables
The pictures on the right show R and X charts for a piece of cable with
open- and shor t-circuited far end. A
resonant frequency
(reactance) equals to zero:
• In the open-circuited case, resonant frequencies correspond to (left to right) 1/4, 3/4, 5/4, etc. of the wavelength in this cable;
• For the short-circuited cable, these points are located at 1/2, 1, 3/2, etc. of the wavelength.
is a point at which X
14
User’s manual
Cable length measurement
Resonant frequencies
velocity factor
A the wave in the cable compared to vacuum. The speed of wave (or light) in vacuum is known as the per second.
Each type of cable has different velocity factor: for instance, for RG-58 it is 0.66. Notice that this parameter may vary depending on the manufacturing process and materials the cable is made of.
To measure the physical length of a cable,
1. Locate a resonant frequency by using the R ,X char t.
of a cable depend on its length as well as on the velocity factor.
is a parameter which characterizes the slowdown of the speed of
electromagnetic constant: c
=299,792,458 meters (or 983,571,056 feet)
Example:
The 1/4-wave resonant frequency of a piece of open-circuited RG-58 cable is 4370 kHz.
2. Knowing the
constant
and the particular type of cable, find the speed of electromagnetic wave in this cable.
electromagnetic
velocity factor
of the
RigExpert AA-2000 ZOOM
299,792,458 × 0.66 = 197,863,022 meters per second
- or -
983,571,056 649,156,897 feet per second
×
0.66 =
15
3. Calculate the physical length of the cable by dividing the above speed by the resonant frequency (in Hz) and multiplying the result by the number which corresponds to the location of this resonant frequency (1/4, 1/2, 3/4, 1, 5/4, etc.)
Velocity factor measurement
For a known
velocity factor
resonant frequency
can be easily measured:
and physical length of a cable, the actual value of the
197,863,022 / 4,370,000 ×(1/4) =
11.32 meters
- or -
649,156,897 / 4,370,000
37.14 feet
×
(1/4) =
1. Locate a described above.
2. Calculate the speed of electromagnetic wave in this cable. Divide the length by 1/4, 1/2, 3/4, etc. (depending on the location of the resonant frequency), then multiply by the resonant frequency (in Hz).
3. Finally, find the Just divide the above speed by the electromagnetic constant.
resonant frequency
16
as
velocity factor
Example:
5 meters (16.4 feet) of open-circuited cable. Resonant frequency is 9400 kHz at the 1/4-wave point.
5 / (1/4) × 9,400,000 = 188,000,000 meters per second
- or -
16.4 / (1/4) 616,640,000 feet per second
.
188,000,000 / 299,792,458 = 0.63
- or -
616,640,000 / 983,571,056 = 0.63
×
9,400,000 =
User’s manual
Cable fault location
To locate the position of a probable fault in a cable, just use the same method as when measuring its length. Watch the behavior of the reactive component (X) near the zero frequency:
• If the value of X is moving from –∞ to 0, the cable is open-circuited:
• If the value of X is moving from 0 to +
RigExpert AA-2000 ZOOM
, the cable is short-circuited:
17
Making 1/4-λ, 1/2-λ and other coaxial stubs
Pieces of cable of cer tain electrical length are often used as components of baluns (balancing units), transmission line transformers or delay lines. To make a stub of the predetermined electrical length,
1. Calculate the physical length. Divide the electromagnetic constant by the required frequency (in Hz). Multiply the result by the velocit y factor of the cable, then multiply by the desired ratio (in respect to λ).
2. Cut a piece of cable slightly longer than this value. Connect it to the analyzer. The cable must be open-circuited at the far end for 1/4-λ, 3/4-λ, etc. stubs, and short-circuited for 1/2-λ, λ, 3/2-λ, etc. ones.
3. Switch the analyzer to the All parameters measure ment mode . Set the frequency the stub is designed for.
4. Cut little pieces (1/10 to 1/5 of the margin) from the far end of the cable until the X value falls to zero (or changes its sign). Do not forget to restore the open­circuit, if needed.
Example:
1/4-
λ
stub for 28.2 MHz, cable is RG-
58 (velocity factor is 0.66)
299,792,458 / 28,200,000 × 0.66 × (1/4) = 1.75 meters
- or -
983,571,056 / 28,200,000 (1/4) = 5.75 feet
A piece of 1.85 m (6.07 ft) was cut. The margin is 10 cm (0.33 ft). The cable is open-circuited at the far end.
28,200 kHz was set.
11 cm (0.36 ft) were cut off.
×
0.66 ×
18
User’s manual
Measuring the characteristic impedance
The
characteristic impedance
Usually, its value is printed on the cable by the manufacturer. However, in certain cases the exact value of the characteristic impedance is unknown or is in question.
To measure the characteristic impedance of a cable,
is one of the main parameters of any coaxial cable.
1. Connect a non-inductive resistor to the far end of the cable. The exact value of this resistor is not impor tant. However, it is recommended to use 50 to 100 Ohm resistors.
2. Enter the R,X chart mode and make measurement in a reasonably large frequency range (for instance, 0 to 200 MHz).
Example 1:
50-Ohm cable
RigExpert AA-2000 ZOOM
Example 1: 50-Ohm cable with 75 Ohm resistor at the far end.
Example 2: Unknown cable with 50 Ohm resistor at the far end.
Example 2:
Unknown cable
19
3. Changing the display range and performing additional scans, find a frequency where R (resistance) reaches its maximum, and another frequency with minimum. At these points, X (reactance) will cross the zero line.
Example 1:
4.46 MHz – min., 8.92 MHz – max.
Example 2:
3.47 MHz – max., 7.16 MHz – min.
4. Switch to the Data at cursor screen by pressing the find values of R at previously found frequencies.
5. Calculate the square root of the product of these two values.
(Data) key and
Example 1:
36.2 Ohm – min., 71.6 Ohm– max.
Example 2:
114.0 Ohm– max, 50.0 Ohm – min.
Example 1: square root of (36.2 × 71.6) =
50.9 Ohm
Example 2: square root of (114.0 × 50.0) =
75.5 Ohm
The Tools menu (see page 26) contains several automated tools for coaxial line
calculations.
20
User’s manual
Measurement of other elements
Although RigExpert A A-2000 ZOOM is designed for use with antennas and antenna­feeder paths, it may be successfully used to measure parameters of other RF elements.
Capacitors and inductors
The analy zer can measure capacitance from a few pF to about 0.1 μF as well as inductance from a few nH to about 100 μH. Since measuring of capacitance and inductance is not a main purpose of RigExpert analyzers, the user will have to gain some experience in such measurements.
Be sure to place the capacitor or the inductor as close as possible to the RF connector of the analyzer.
1. Enter the R,X chart mode and select a reasonably large scanning range. Perform a scan.
Example 1:
Unknown capacitor
RigExpert AA-2000 ZOOM
Example 2:
Unknown inductor
21
2. By using lef t and right arrow keys, scroll to the frequency where X is -25…-100 Ohm for capacitors or 25…100 Ohm for inductors. Change the scanning range and perform additional scans, if needed.
3. Switch to the Data at cursor screen by pressing the key and read the value of capacitance or inductance.
Example 1:
Unknown capacitor
Example 2:
Unknown inductor
Transformers
RigExpert analyzers can also be used for checking RF transformers. Connect a 50 Ohm resistor to the secondary coil (for 1:1 transformers) and use SWR chart, R,X chart or Smith chart modes to check the frequency response of the transformer. Similarly, use resistors with other values for non-1:1 transformers.
22
User’s manual
Traps
A
trap
is usually a resonant L-C network used in multi-band antennas. By using a
simple one-turn wire coil, a resonant frequency of a trap may be measured.
Example:
A coaxial trap constructed of 5 turns of TV cable (coil diameter is 6 cm) was measured.
A one-turn coil (about 10 cm in diameter) connected to the analyzer was placed, co-axially, a few centimeters away from the measured trap. The SWR chart shows a visible dip near 10.3 MHz, which is a resonant frequency of the trap.
RigExpert AA-2000 ZOOM
23
Annex 1
Specifications
Frequency range: 0.1 to 2000 MHz
Frequency entry: 1 kHz resolution
Output impedance: 50 Ohm
Directivity: 43db(calibrated)
SWR measurement range:
• 1 to 100 in numerical modes
• 1 to 10 in chart modes
SWR system impedance: -selectable
25, 50, 75, 100, 150, 200, 300, 450, 600
Display modes:
• SWR at single or multiple frequencies
• SWR, return loss, R, X, Z, L, C at single frequency
• SWR chart
• R, X char t
• Smith chart
• Return loss char t
• TDR chart (Time Domain Reflectometery)
• Cable tools (stub tuner, length & velocity factor, cable loss and characteristic impedance
measurement, LC meter, Multé)
Factory OSL: Yes User OSL: 10 profiles available Records: 250 unified memory slots
RF output:
• Connector type: N
• Output signal shape: square
• Output power: -10 dBm (at 50 Ohm load)
Power:
• Three 1.5V alkaline batteries, type AA
• Three 1.2V Ni-MH bat teries, type A A
• Three 3.7V LiPo batteries, 14200
• Max. 4 hours of continuous measurement, max. 2 days in stand­by mode when fully charged batteries are used
• When the analyzer is connected to a PC or a DC adapter with USB socket, it takes power from these sources
Interface:
• 800×480 color Blanview® display
• 6x3 keys on the water-proof keypad
• Multilingual menus and help screens
• USB connection to a personal computer
• Bluetooth Specification v.4.2, LE
Dimensions: 98 × 230 × 53 mm
(3.9 × 9.0 × 2.0 in)
Operating temperature: 0…40 °C
(32…104 °F)
Weight: 442 g (15.6 oz)
Warranty: 2 years
Made in Ukraine.
24
User’s manual
Annex 2
Precautions
Never connect the analyzer to your antenna in thunderstorms.
Lightning strikes as well as static discharge may kill the operator.
Never leave the analyzer connected to your antenna after you fin­ished operating it. Occasional lightning strikes or nearby trans­mitters may permanently damage it.
Never inject RF signal or DC voltage into the antenna connec tor of the analyzer. Do not connect it to your antenna if you have active transmitters nearby.
Avoid static discharge while connecting a cable to the analyzer. It is recommended to ground the cable before connecting it.
Do not leave the analyzer in active measurement mode when you are not actually using it. This may cause interference to nearby receivers.
If using a personal computer, first connect the cabl e to the antenna connector of the analyzer, then plug the analyzer to the computer USB port. This will protect the analyzer from static discharges.
RigExpert AA-2000 ZOOM
25
Annex 3
Tools menu
For the quick access to the the Tools menu, press the + key combination.
Stub tuner
The Stub tuner mode is designed to
help making or checking 1/4-λ or 1/2-λ coaxial stubs.
Connect either open or short circuited
cable to the analyzer and press (OK)
to start.
The analyzer will immediately show
resonant frequencies for both
quarterwave and halfwave stubs.
Longer cables have lower resonant
frequency.
26
User’s manual
Length & VF
Knowing the
length of a cable can be easily calculated.
Press the (Up) but ton and edit the
value of the velocity factor, then press
To find the
cable, press the (Down) key and
enter the physical length, then press
The velocity factor depends on a type of your transmission line.
For instance, RG-58 cable with polyethylene insulator has VF=0.66.
velocity factor
(OK) to start measurement.
velocity factor
(OK).
, a physical
of an unknown
RigExpert AA-2000 ZOOM
27
Cable loss
To measure the
connect a piece of a cable to the antenna
connector of the analyzer. Make sure
the far end of the cable is open circuited.
Press (OK) to start.
Next, short circuit the far end of the cable
and press (OK) to continue.
Once the analy zer finishes the
measurement, you will see the Loss
versus frequency chart. Use (Left)
and (Right) cursor keys to change
frequency and watch the loss value in
decibels at the bottom of the analyzer’s
screen.
To see the list of other keypad shortcuts,
press the (Help) key.
loss
in a coaxial cable,
28
User’s manual
Cable impedance
To measure the
use a piece of an open circuited cable:
a half of a meter (or a foot) or longer
should be fine. Press (OK) to start.
Next, the far end of the cable should
be short circuited. Press (OK) to
continue.
There are several reasons why the
resulting char t does not look smooth,
so we need to use (Left) and
(Right) cursor keys to find the location
where the impedance is stable. The result
is shown at the bottom left corner of the
screen.
Use
(Down) key combinations to change the
scale, if needed.
characteristic impedance
+ (Up) and +
,
RigExpert AA-2000 ZOOM
29
LC Meter
Use to choose one of six modes (Inductance L, Q, tan δ and Capacitance C, Q, tan δ), press to sta rt measure,
evaluate result, tune range to most
reliable values and restar t measure, if
required.
Be aware of many factors that will impact result of measure -stray capacitance and
inductance, nonuniformities of element connection, more complex behaviour of
component at higher frequency.
Avoid pick values at series and parallel
resonances (reactance change sign) and
adjacent areas, avoid fast changing
slopes. Pick values from more stable,
persistent areas of chart. In most cases
measures at lower frequncies are more
reliable.
Self tests
There are several built-in self tests in the
AA-2000 ZOOM analyzer, which can be
run by the user to make sure the analyzer
is working properly. Press to start
the test and then follow instructions
given on the display.
30
User’s manual
Annex 4
Setup menu
For the quick access to the Setup menu,
press the
There are several settings in the Setup menu:
• Language – select a language for the analyzer’s menus
• Palette – choose a color scheme
• Battery – select a power consumption scheme
• Sound – select sound volume
• Bluetooth – controls bluetooth module power
• Sys. imp. – select sys tem impeda nce (25, 50, 75, 100 , 150, 200, 300, 45 0, 600 Ohm)
which affects SWR and return loss readings.
• Units – select metric (meters) or imperial (feet) units
• Bands – select region for highlighting of radio amateur bands
• Cable vel. factor – choose a velocity factor of the coaxial cable for the TDR mode
• Freq. corr. – frequency correction of the analyzer’s oscillator
• Reset settings – reset the analyzer to factory defaults
• Clear saved charts – clear all memory slots
+ key combination.
RigExpert AA-2000 ZOOM
31
Annex 5
TDR mode
Theory
Time domain reflectometers (TDR) are electronic instruments used for locating faults
in transmission lines.
A
short electrical pulse
the line, and then a reflected pulse is
obser ved. By knowing the delay between
two pulses, the speed of light and the
cable velocity factor, the
to-fault) is calculated. The amplitude
and the shape of the reflected pulse give
the operator idea about the nature of the
fault.
is sent over
DTF
(distance-
In st ea d of a sho rt pulse, a “
may be sent over the cable.
32
step
” function
User’s manual
Unlike many other commercially-available reflectometers, RigExper t AA-2000 ZOOM
does not send pulses into the cable. Instead, another technique is used. First, R and
X (the real and the imaginary part of the impedance) are measured over the whole
frequency range (up to 2000 MHz). Then, the
applied to the data. As a result,
impulse response
IFFT
(Inverse Fast Fourier Transform) is
and
step response
are calculated.
This method is often called a “
is used in this document since all calculations are made internally and the user can only
see the final result.
The vertical axis of the resulting chart displays the
load, 0 for matched impedance load (ZLoad=Z0), or +1 for open load. By knowing the
cable velocity factor, the horizontal axis is shown in the units of length.
Single or multiple discontinuities can be displayed on these charts. While the Impulse
Response chart is suitable for measuring distance, the Step Response chart helps in
finding the cause of a fault.
See the examples of typical Step Response charts on the next page.
Frequency Domain Reflectometry
reflection coefficient
RigExpert AA-2000 ZOOM
”, but the “
: Γ=-1 for short
TDR
” term
33
34
User’s manual
Practice
Use for quick setup popup, choose smallest range approriate for your cable length (shorter distances have better length resolution), tune velocity factor fast. Change of velocity factor corrects range distances.
The result of measure can be viewed in
three modes: impulse response, step
response, impedance response. To
swi tch view press
for numerical representation all data at
marker. Three pictures there show real
measure of short open stub in different
view modes. Explore built-in help for
other buttons functionality.
+ , use
RigExpert AA-2000 ZOOM
35
Annex 6
Calibration
RigExpert AA-2000 ZOOM is designed for high performance and uses factory
short-load
are 10 user OSL profiles, that can be applied above factory OSL.
The
especially important for high frequencies (100 MHz and upper). Three different
calibration standards should be used: an “open”, a “short” and a “load” (usually, a
50-Ohm resistor). A place where these standards are connected during calibration is
called a
parameters of this line will be subtracted from measurement results and the analy zer
will display “true” parameters of a load.
(OSL) calibration. To make usage more flexible for expirienced users there
standards
used for calibration should be of high quality. This requirement is
reference plane
. If the calibration is done at the far end of a transmission line,
open-
36
User’s manual
Use
calibration profiles managment screen.
Select profile and press to make
profile active, combination of +
Key starts calibration sequence for
selected profile. Follow instructions on
screen to perform 3 step calibration.
+
in main menu to go in
allows to edit existing profile name.
You may connect calibration standards
to the far end of a cable, so the cable
will be “nulled”.
To apply calibration, press the
+
measurement mode. The small “CALn
mark will appear on screen, where n is
number of active profile(0..9)
key combination in any
RigExpert AA-2000 ZOOM
37
Annex 7
Dummy loads
50 Ohm dummy loads are not all equal.
Amphenol
202109-10
1 Watt
terminator plug
Low SWR
calibration
For
low-power RF terminators which provide
low SWR over the wide frequency range.
High-power terminators, with cable or
not, are suitable neither for
purposes (page 32), nor for analyzer
self-tests
(see page 32), please use
calibration
(page 30).
38
Bird 8201
500 Watt oil-cooled termination
High SWR
User’s manual
RigExpert AA-2000 ZOOM
39
For pr ivate househol ds: Info rmatio n on Disposal for User s of WEEE
This symbol on th e product(s) an d / or accompa nying documents means that u sed elec trical a nd electronic equipment (W EEE) should not be mixe d with g eneral h ousehol d waste. For p roper trea tment, recovery a nd recycling, plea se take th is produc t(s) to designa ted collec tion point s where it w ill be acce pted fre e of charge.
Alte rnatively, in some co untries, y ou may be able to r eturn you r pro ducts to your l ocal ret ailer upo n purchase of an equivalent new product. Disposing of this product correctly will help save valuable reso urces an d preven t any poten tial neg ative ef fects o n human he alth and th e environm ent, whi ch could o therwis e arise from i nappropriate waste handling.
Please cont act your local au thority for further details of your n earest designated col lectio n point . Penal ties may be applic able for incorrec t dispos al of t his waste, in ac cordance with your nati onal legislation.
For pr ofessi onal users i n the Euro pean Unio n
If you wish to d iscard electrical and electr onic equipmen t (EEE), please cont act y our dealer or suppl ier for fur ther info rmation.
For dis posal in countries o utside of t he European Union
This symbol is on ly valid in the Europ ean Union ( EU). If you wish to discar d this produ ct, plea se conta ct your loca l authori ties or dealer and ask for the co rrect method of di sposal.
http://www.rigexpert.com
Copyright © 2020 Rig Expert Ukraine Ltd.
“RigExpert” is a registered trademark of Rig Expert Ukraine Ltd.
Doc. date: 23-June-2020
Loading...