All other brand or product names within this publication are trademarks or
registered trademarks of their respective companies.
All information contained within this publication is, to the best of our knowledge, complete and
accurate at the time of publication. Rice Lake Weighing Systems reserves the right to make
changes to the technology, features, specifications and design of the equipment without notice.
The most current version of this publication, software, firmware and all other product
updates can be found on our website:
www.ricelake.com
Page 3
Contents
Technical training seminars are available through Rice Lake Weighing Systems.
Course descriptions and dates can be viewed at www.ricelake.com/training
or obtained by calling 715-234-9171 and asking for the training department.
The purpose of this manual is to help the user understand the SCT-1100 functioning modes, key functions, and display
indications. This manual applies to indicators using version 8.00 or higher of the STC-1100 firmware. Configuration and
calibration of the indicator can be accomplished by pressing the indicator front panel keys, the serial command set or RLTools
utility. The SCT-1100 is designed to work with up to 4 load cells in either a dependent or independent channel mode.
Manuals and additional resources are available from the Rice Lake Weighing Systems website at
Warranty information can be found on the website at www.ricelake.com/warranties
1.1Safety
Safety Signal Definitions:
Indicates an imminently hazardous situation that, if not avoided, will result in death or serious injury. Includes
hazards that are exposed when guards are removed.
Indicates a potentially hazardous situation that, if not avoided, could result in serious injury or death. Includes
hazards that are exposed when guards are removed.
Indicates a potentially hazardous situation that, if not avoided, could result in minor or moderate injury.
Indicates information about procedures that, if not observed, could result in damage to equipment or corruption
to and loss of data.
www.ricelake.com
General Safety
Do not operate or work on this equipment unless this manual has been read and all instructions are understood.
Failure to follow the instructions or heed the warnings could result in injury or death. Contact any Rice Lake
Weighing Systems dealer for replacement manuals.
Failure to heed could result in serious injury or death.
Electric shock hazard!
There are no user serviceable parts. Refer to qualified service personnel for service.
The unit has no power switch, to completely remove DC power from the unit, disconnect the DC power cable from the main
socket.
For pluggable equipment the socket outlet must be installed near the equipment and must be easily accessible.
Always disconnect from main power before performing any work on the device.
Do not allow minors (children) or inexperienced persons to operate this unit.
Do not operate without all shields and guards in place.
Do not use for purposes other then weighing applications.
Do not place fingers into slots or possible pinch points.
Do not use this product if any of the components are cracked.
Do not make alterations or modifications to the unit.
Failure to follow could result in damage to equipment or corruption to and loss of data.
Keep away from heat sources and direct sunlight.
Protect the instrument from environmental factors: rain, snow, dust, etc.
Do not wash, dip in water or spill liquid on the instrument.
Do not use solvents to clean the instrument.
Do not install in areas subject to explosion hazard.
Always mount the instrument and platform in a vibration free setting.
All instrument connections must be made with respect to local zone and environment standards.
1.2Options
Features Include:
• 5-key, dual function, tactile feel keypad
• 6-digit LED display, 0.50" (13mm) high
• (6) red LED Annunciators
• NEMA type 1 plastic enclosure
• 12 - 24 VDC power
• Mountable to a DIN 35mm rail
• (1) 6-wire load cell connection with Remote Sense
• (3) 4-wire load cell connections
• Two configurable digital inputs and two configurable digital outputs
• Analog Output
• 0-20 mA, 4-20 mA (Maximum 350,000 ohm)
• 0-5 VDC, 0-10 VDC (Minimum 10,000 ohm)
• (1) RS-485 bidirectional port configurable for connection to a PC/PLC
• (1) RS-232 bidirectional port for connection to a printer or PC
• Unit of measure conversion
• Switching of net/gross weight setpoint on the gross weight/net weight/pieces,
• Alibi memory
• Peak detector
• Weight or Theoretical calibration with up to 3 linearity points
• Diagnostic information via serial port, message display, printing, simulation of key pressure
• Reading of the net, gross and tare weights. Clearing and entering tare, scale switch, setting of output values
2Visit our website www.RiceLake.com
Page 7
1.3Overview
8
9
2.9
Ǝ
Ǝ
Ǝ
Ǝ
Ǝ
Ǝ
Ǝ
2.28Ǝ
3
2
1
456
7
Note
The indicator has a plastic case with external dimensions as shown in Figure 1-1.
Introduction
Figure 1-1. SCT-1100 Dimensions and Components
No.Description
1(+) 12-24 Vdc power supply input
2GND power supply input
3Digital I/O
4Analog output
5Connection for serial line RS-485
6Connection for serial line RS-232
7Connections for load cell
Table 1-1. SCT-1100 Components
For detailed call out of pin designations, See Section 2.5 on page 7
The front panel of the SCT-1100 consists of a display with six digits that are 0.5 in (13 mm) high, six LED annunciators and a
five-key keyboard.
Figure 1-2. SCT-1100 Front Panel
Item No. SymbolDescription
1lbUnits – lb is printed on the instrument; kg, Ton, g, stickers are included for changing the units on the overlay
2>0<Illuminates when the weighing system is within ±1/4 division of zero
3~ (tilde) Illuminates when the weight is unstable
4NETIlluminates when a tare is established, measuring net weight
5FIlluminates:
• when the specification function of the instrument is active (set in F.Mo d EF Un Ct parameter) See Section 3.7 on page 14
• when a key is pressed
Turns off:
• when the specification function of the instrument is disabled
• with an active function (a key is released)
Blinking means the instrument function is active for five seconds
6W1Indicates the activation of the first output (Sp1)
7W2Indicates the activation of the second output (Sp2)
8ZERO – Clears the displayed gross weight of up to ± 2% of the total capacity; Cancels tare
At power up: Momentary press during startup displays current settings. See Section 3.1.1 on page 10
In setup: scroll through parameters
In numeric input: decreases the digit to be modified
9TARE – Momentary press executes semiautomatic tare; Cancels tare
At power up: Momentary press during startup displays setup mode. See Section 4.1 on page 17
Long press allows for entering a manual tare from the keyboard
In setup: scroll through the parameters
In numeric input: increases the digit to be modified
10MODE – Executes a specific function (set in the setup mode) See Section 3.7 on page 14
Long press allows for toggling the displayed channel (if configured in independent channels mode (Ind . C h )
At power up: Momentary press during startup displays quick setup menu. See Section 4.2 on page 18
In setup: enter into a parameter or to confirm a setting
In numeric input: confirms the entry made
11PRINT – Executes a specific function (set in the setup mode) See Section 4.4.1 on page 26.
Executes a printout or transmission of data from the serial port dedicated to the printer
In setup: enter into a parameter or to confirm a setting
In numeric input: confirms the entry made
12ON/OFF – Turns the instrument on and off
C
In setup: press multiple times to display SA v E? and/or press to exit a step without confirming the setting
In numeric input: momentary press clears the present value
Long press beyond -o ff -: Displays information of the scale (capacity, division, minimum weight for each configured range, gravi-
tational acceleration value, number of configured channels)
Table 1-2. SCT-1100 Front Panel
4Visit our website www.RiceLake.com
Page 9
Installation
IMPORTANT
IMPORTANT
2.0Installation
Rice Lake Weighing Systems recommends the instrument and the platform (transducer) be installed on a flat level surface, that
is stable and vibration free.
2.1Location Selection
The following should be considered when selecting a location for the equipment:
• Dust-free
• Free of strong breezes or vapors
• Moderate temperature and humidity (59 to 86°F and 40-70%)
• Use waterproof conduit and couplings in order to protect the load cell cables
• Use a waterproof junction box to connect the cells
• Avoid welding with load cells installed
2.2Electrical Precautionary Measures
The following electrical precautionary measures must be considered when installing this equipment:
• Main power supply must be maintained within ± 10% of the rated voltage
• Electrical best practices must be observed by the installing technician
• Follow recommended minimal separation distances given for cable categories, See Section 2.3
• The extension leads of the load cells, or signal amplifiers connecting to the serial ports and analog output, must be within stated maximum
lengths, See Section 2.3
• It is recommended that load cell cables are shielded and run in conduit at an acceptable distance from power transmission lines to avoid
signal interference and signal noise
• All cable not in conduit or otherwise shielded should be of minimal length and terminated as close to conduit exit as possible to avoid
extraneous signal noise
• If the instrument is situated inside an electric panel, the power supply cable must be shielded and as short as possible, separate from every
coil supply cable, inverter, electromotive force, and others. In addition, provide dedicated power supply to the instrument
• Install RC filters on the contact coils, solenoid valves and all devices producing electric fields
• It is recommended to leave the instrument powered on at all times to avoid condensation forming on the inside of the instrument
2.3Maximum Cable Length
Load Cell Cable
The maximum length of a standard load cell cable with sense wires is:
• 150’ at 30#AWG
• 300’ at 24#AWG
RS-232 Cable
The maximum length of the RS-232 cable is 50' with a maximum baud rate of 19200.
RS-485 Cable
The maximum RS-485 cable length is 4000', See
Analog Output Cable
The maximum length of the analog current output cable at 4-20mA is 300'.
The maximum length of the analog voltage output cable at 0-10Vdc is 150'.
A centrally located, single point ground, such as the ground bar of the electric panel, must be created and/or identified for
proper grounding and functioning of the system. The ground must be sized so that the total resistance of grounding is lower
than 1Ω. Connect grounding points of all instrumentation, load cells, and weighing structure to this single point ground.
2.4.1Load Cells and Junction Box
When the load cells are connected to the instrument through a junction box, the shielding of the load cell cables and the
instrument must be connected to the junction box grounding.
When the load cells are directly connected to the instrument, the load cell cable shielding must be connected to the single point
ground.
System cabling should be kept as short as possible to minimize noise potential. After exiting conduit or other shielding, a ferrite
device should be used prior to conductor termination.
After platform and the load cell are properly grounded, connect the shield from the load cell cable to the instrument ground. See
Figure 1-1 on page 3.
Non weighing
Weighing
Grounding
Strap
Junction
box
Figure 2-1. Grounding Example
Procedures not expressly described in this manual are considered improper use of the equipment.
Ensure the platform is level or the loading cells are shimmed evenly.
All connections must meet all local zone and environment standards.
Follow the recommended electrical precautionary measures described in Section 2.2 on page 5.
Make sure that the grounding is made correctly, See Section 2.4.
6Visit our website www.RiceLake.com
Page 11
2.5Wiring Schematic
I/O
BOOT
J1
Load Cell 1
RS-232
RS-485
Analog Output
12/24 VDC
Load Cell 2, 3, 4
Load
Cell 2
Load
Cell 3
Load
Cell 4
SIG +
SIG –
EXC +
EXC –
SIG +
SIG –
EXC +
EXC –
SIG +
SIG –
EXC +
EXC –
SIG +
SIG –
EXC +
EXC –
SEN +
SEN –
181920212223242526272829303132333435
V
I
Note
Installation
Figure 2-2. SCT-1100 Wiring Schematic
The CELL1 terminal board of the indicator can be connected to a six-wire load receiver (wiring must be jumpered if connecting
to 4-wire load cell. See Figure 2-3 on page 8); CELL2, CELL3 and CELL4 are only for four-wire connection. See Figure 2-2.
The load cell 1 terminal board of the SCT-1100 must be connected to the 6-wire load cell; if using a 4-wire load in the load cell 1
terminal board, cell excitation must jumper to sense, See Figure 2-3. Load cells 2, 3, and 4 must be connected to 4-wire load
cells, See Figure 2-4.
Sense is always enabled and, when not using 6-wire load cell, the sense terminals must be jumpered to the
same polarity excitation wires.
The sense compensates for drops in voltage along the cable that connects the instrument to the load cell. Voltage is
lost when the instrument and the load cell are greater than 30ft apart. A cable is typically provided with a load cell.
When exceeding the length of the provided load cell cable, six wires must be used to compensate for voltage drop. It
is, however, recommended to never cut or shorten the load cell cable.
Power must be provided to the SCT-1100 through an external AC/DC charger that supplies a stabilized voltage of 12 VDC to 24
VDC while connected to an appropriately rated AC power supply. Connect the two power supply wires to the appropriate
terminals on the side of the instrument, See Figure 1-1 on page 3.
3.1.1Turn on the Instrument
Press until the instrument turns on, then release. The instrument executes a start up procedure and displays the installed
software version.
An auto zero function zeros the instrument at startup if the weight detected on the scale is ± 10% of the capacity. If the weight
is not within this tolerance the instrument displays ZE RO and, after 10 seconds, the current weight is displayed.
The auto zero function at start-up can be disabled in the setup mode.
SE tu P C o n F i G PAr AM A u to-0 in Table 4-6 on page 24.
See
To view the following settings momentarily press while the display self-check is running:
• XX .Y Y – X X is the software release and YY is the sub release
• XX .Y Y – X X indicates the type of instrument, YY indicates the software version
• XX .Y Y.ZZ – the installed software version
• XX X X X X – the name of the installed software
• XX X . XX X – capacity and division of channel 1
• h i r ES – displayed together with the gravitational acceleration value of the area of use
The instrument then executes a self-check.
3.1.2Standby
Press until OFF displays. The LED at the left of the display remains on.
3.1.3Power Off the Instrument
To completely power off the unit, remove the power supplied voltage.
3.1.4Zero
Press to zero a gross weight within ± 2% of the total capacity (or as set). Weight value displays as 0 and the relative
annunciators illuminate.
10Visit our website www.RiceLake.com
Page 15
Operation
C
Note
C
C
3.1.5Tare
Semiautomatic Tare
Press to tare the weight value on the scale. tArE displays momentarily and then 0 (net weight). The relative annunciators
illuminate.
Manual Tare
Press for a few seconds. -tM- displays and then 0 0 00 0 0 . Enter the desired tare value with the following keys:
selects the digit to be modified (moves left to right, selected digit blinks)
•
• decreases the blinking digit
• increases the blinking digit
•
momentary press clears the present value, long press returns to the weigh mode without saving changes
• confirms the entered tare value
The tare value is subtracted from the weight on the scale and the relative annunciators illuminate.
The entered tare will be rounded off to the nearest division.
Cancel a Tare
A tare value can be manually canceled in multiple ways:
• Unload the scale and press or
• Press without unloading the scale
• Enter a manual tare equal to zero
Locked or Unlocked Tare
When a tare value is entered manually, automatically or from storage the tare value displays with a negative sign when the
scale is unloaded. This is known as a locked tare. An unlocked tare is automatically canceled each time the scale is unloaded.
To set the tare type:
1.Turn on the instrument and press while the firmware version displays. F .M o d e displays.
2.Press to enter the menu.
3.Press or until tArE displays. Press to select.
4.Press or to scroll through options.
• Lo C K – locked tare
• u n Lo CK – unlocked tare
• di SA b – disable tare
5.Press
6.Slowly press multiple times until SA V E? displays.
• Press to confirm and store to the instrument memory
• Press any other key to cancel and exit without saving
The multi-range function allows for subdividing the scale capacity in two, each up to 3000 divisions.
Example: with a 10 kg cell platform it is possible to approve the weighing system with:
Single range: 6 kg capacity and 2 g division (3000 div.)
Dual range: 6/3 kg capacity and 2/1 g division (3000 + 3000 div.)
Multi-range functioning is indicated by illumination of the relative LED identifying the operating range; when the
weight on the scale enters into the second range, the division of the second range is enabled. The first range division
is restored only when the weight on the scale goes below the gross zero of the scale.
The selection of the range number with multi-range functioning is made during the instrument’s calibration,
See Section 5.0 on page 32.
3.3Display Configuration Data
The INFO function makes it possible to view the configuration data, such as:.
• First range capacity, first range minimum weight, first range division
• Second range capacity, second range minimum weight, second range division (if set)
• Gravitational Acceleration Value
The minimum weight corresponds to 20 net weight divisions
The data of the second range appears only if a range is configured
To view the configuration data:
1.Press and hold until inF O displays.
2.Release . The capacity value of the first range displays. Press or to scroll forward or back through the
following data.
• First range capacity Ch1. MA X
• First range minimum weight Ch1.Min
• First range division Ch 1.E
• Second range division Ch1. MA X
• Second range minimum weight Ch1. Mi n
• Second range division Ch1. E
• Gravitational Acceleration Value GrAVit
• Number of Configured Channels ConF .C h
3.Press to return to the weigh mode.
Pressing when information of the currently active channel is displayed allows for the data of the other configured channels
to be viewed. This selects and pauses the information for each channel, otherwise the information automatically cycles through
all channels.
For example, if channels 2 and 3 are configured, for the maximum capacity of the 1st range:
1st range capacity channel 1 (1 C h 1 .MAX ). Press . 1st range capacity channel 2 (Ch2 .MAX ).
Press . 1st range capacity channel 3 (Ch3. MA X).
3.4Selecting the Channel to be Displayed
When indicator has multiple scale channels configured in tra n s M type mode or i nd .Ch type mode, it is possible to select the
channel to be displayed using the key. See Section 4.3.1 on page 21 for configuring type parameters and Section 4.4 on
page 24 for configuring the number of channels.
1.Press and hold . The currently selected channel displays first followed by ChAn momentarily. A menu of available
chanels displays.
2.Select the channel to be displayed. Press to confirm.
12Visit our website www.RiceLake.com
Page 17
Operation
Note
C
Note
C
3.5Simultaneous Transmitter Mode
The tr a n sM mode makes it possible to have simultaneous transmission of the values on each single channel via the serial
line. In addition, through the optional alibi memory board, the transmitted weight values can be stored in a computer for data
processing and/or integration. The filed values can then be recalled from the PC serial line or directly on the instrument’s
display. See Section 4.3.1 on page 21 for configuring type parameters and Section 4.4 on page 24 for configuring the number
of channels.
In this mode the zero tracking and the scale keys , and are disabled. It is not possible to set functioning
modes and the alibi mode is set automatically.
To set the transmitter mode:
1.Turn on the instrument and press while the firmware version displays. tY P E displays. Press to enter the
menu.
2.Navigate to select t r An S M displays. Press to select.
3.Slowly press multiple times until SA V E? displays. Press
to confirm.
Operation
In the simultaneous transmitter mode it is possible to view the weighing information. See Section 3.7.2 on page 14 regarding
Reviewing Stored Weigh Information.
• Press to switch channels, if the scale is configured as multichannel
• Only functions which can be enabled through serial commands can be performed; not all serial commands can be
performed; See Section 6.3 on page 42; it is not possible to carry out other operations in the simultaneous
transmitter mode.
• The list of usable serial commands includes:
DINT, PCOK, STAT, KEYP, KEYR, KEYEE,KEYED
Pressing simulation of the scale keys through the KEYP and KEYR commands allows for management of the
functions linked to the
In addition to the standard weighing mode, the instrument can be set to carry out four alternative operational functions. Each
operating mode activates certain LEDs.
To set the operating mode:
1.Turn on the instrument and press while the firmware version displays. F. Mo dE displays. Press to enter the
menu.
2.Navigate to Fun C t. Press to enter the menu.
3.Select operating mode:
• Co nver – convert displayed value to a calculated value, See Section 3.7.1
• Al i b i - Alibi memory, See Section 3.7.2
• Vi SS – sensitivity times ten, See Section 3.7.3 on page 15
• PE A k – peak hold detector, See Section 3.7.4 on page 16
4.Press to confirm selection.
5.Slowly press multiple times until SA V E? displays.
• Press to confirm and store to the instrument memory
• Press any other key to cancel and exit without saving
3.7.1Conversion
This function toggles the displayed weight between the scale unit of measure and an alternative unit measure.
• Long press to set the conversion factor
• Short press to toggle between the unit of measures
• Press to save the conversion value
3.7.2Alibi Memory
The alibi memory allows for transmitted weight values to be filed in the PC for data processing and/or integration. The filed
values can then be recalled from the PC serial line or directly on the instrument’s display for a following check.
Storage of a weight value occurs following the reception of the serial command or by pressing . The instrument transmits
the gross and tare weights and an ID on the serial port.
The ID has the format: <Rewriting number>-<Weigh number>
• The rewriting number is a five digit number from 00000-00255; it indicates the number of complete rewritings of the alibi
memory
• Weigh number is a six digit number from 000000-131072; it indicates the weigh number in the current rewriting of the
alibi memory. The weigh number is increased by 000001 with each weigh storage. Once the value reaches 131072, it
restarts from 000000.
The storage of a weigh value occurs only if the gross weight is greater than or equal to zero, it is stable and valid (not in underload or overload). Depending on how F.M od E r EACt has been configured in the technical set up, the storage of a weight by
pressing a key is possible only if the condition is met (weight exceeds zero, weight instability or always).
14Visit our website www.RiceLake.com
Page 19
Reviewing Stored Weigh Information
C
Note
C
To review stored information:
1.Press . r E W.id displays.
2.Enter the rewriting number (from 00000-00255).
3.Press . displays.
4.Enter the weigh number (from 000000-131072).
5.Press . The weigh information displays.
6.Press or to view the weigh information.
• ch .x - x is the scale number (from 1-4)
• u myy - y y is the unit of measure (lb, kG, G )
• G r oS S momentarily displays and then the gross weight value
• tA r E orTa r E Pt (manual tare) momentarily displays and then the tare weight value
7.Press to return to the weigh mode.
Operation
If the alibi memory is empty and is pressed,
mode.If the entered ID is not valid,
displays momentarily and the instrument returns to the weigh
displays and the instrument returns to the weigh mode.
Clearing the Alibi Memory
The alibi memory can be cleared directly on the instrument in the SE T UP i n i. ALparameter.
1.Turn on the instrument and press while the firmware version displays. F. Mo dE displays.
2.Press until SEt Up displays. Press to enter the menu.
3.Press or until in i. AL displays. Press . i. AL ib .?displays.
4.Press to clear the alibi memory or any other key to cancel.
• AL .OK displays if the operation is successful
• AL .ER R displays if the memory was not successfully cleared (repeat procedure)
5.Press to return to the weigh mode.
It is not possible to clear an individual weigh record.
3.7.3Sensitivity Times Ten
This mode converts the weight to sensitivity times ten for display and is used for testing during calibration.
Press to toggle the weight display between standard sensitivity and sensitivity times ten.The last digit on the right of the
display has a sensitivity equal to the scale’s division divided by 10.
This mode can be used to store the maximum (peak) weight value measured during the weighment.
Operation
If Peak Hold Weight Detection has been set as the functioning mode, the following functions are enabled while in weigh mode:
1.Press to enable the peak weight detection. -PEAK- displays alternately with the maximum weight value reached
up to that point.
2.Press again to terminate the peak weight detection. Peak weight detection also terminates if the weight surpasses
the maximum capacity of the instrument. In both cases, PEAK.o F displays and then the current weight on the scale
displays. The detected weight value will be:
• The maximum before a rapid decrease of the weight (measurement of the highest weight)
• The maximum and persistent weight detected on the scale
Exit the peak mode to toggle from one scale to another when multiple scales are connected to the instrument.
Setting Sampling Time
To set the minimum sample time of the peak weight detection while in weigh mode:
1.Press and hold .
2.Select PI C . tl M.displays followed by a number which corresponds to the minimum time length of the impulse
expressed in hundredths of seconds.
3.Press or until the desired value displays. See Table 3-1 for a list of settable values.
4.Press to confirm. The instrument returns to the weigh mode
Time in
1/100 sec
140011
220011
310011
410042
55042
102542
201242
50642
100682
1276122
Table 3-1. Sample Times in Peak Weight Detection
Sample per
Second
Acquired
Values
Mediated
Values
The peak detection function sensitivity is dependent on the parameter set. The higher the number of samplings, the greater the
sensitivity. If an unexpected peak is detected immediately, decrease the sensitivity.
Example: If 0.000 lb is on the load cell of 20 lb (350 Ω), and the sampling time is equal to 1, when the peak
function is enabled, 0.005 lb displays.
16Visit our website www.RiceLake.com
Page 21
Setup Mode
C
4.0Setup Mode
The setup mode is used to set the functioning parameters of the instrument. There are two setup menus in the SCT-1100.
• The Quick Setup Menu is a limited menu that includes settings essential to basic scale configuration such as quick
calibration and communication. To enter the quick setup menu, pressto turn the instrument on. Press as the
firmware version displays. See Section 4.2 on page 18 for more information on the quick setup menu
• The Setup mode Menu is a more in-depth menu that incorporates all configuration settings. To enter the setup mode,
pressto turn the instrument on. Press as the firmware version displays. See Section 4.3 on page 20 for more
C
information on the Setup mode menu
4.1Setup Mode Navigation
Use the keys on the front panel of the indicator to navigate trough the menu options as follows:
KeyFunction
Scroll through parameters
In numeric input: decreases the digit to be modified
Scroll through the parameters
In numeric input: increases the digit to be modified
Quickly position at the first step of a menu
In numeric input: selects the digit to be modified, from left to right
Enter into a parameter or confirm a setting
In numeric input: confirms the entry made
Exit a step without confirming the setting
In setup: press multiple times to display SA v E? prompt and/or press to exit a
step without confirming the setting
Press , and to set the total capacity of the load cells.
Press , and to set the mV/V sensitivity of the load cell.
Press , and to set the dead load present on the load cell.
Press , and to set the scale capacity.
Apply calibration weight. Press , and to enter calibration weight that was applied.
Set the analog output on net or gross weight.
Select channel to be configured.
Set the maximum value of the analog output.
Set the analog output when the scale shows zero weight.
Set the minimum value of the analog output.
Set the key to be emulated by the input.
Press and to select number of channels.
Press and to select channel to calibrate.
Press and to set channel type ind.ch, transm, or dep.ch.
For new acquisition of zero, with scale empty. Establishes new zero reference.
Disabled.
Setpoint on the gross weight.
Setpoint on the net weight.
Press , and to set number of additional units.
Press and to set baud rate.
Set the output normal state.
Quick calibration of zero.
Check the load cell signal in millivolts.
FUNC
set.Add
baud
OUT.01
No/NC
485
oUTPUT
Ao ZER
MODE
chan
Ao MAX
INP.02
Ao MIN
INP.01
OUT.02
nChan
Chan
type
CEL.CAP
CEL.SEN
DEAD.LD
CAPAC
0.CALIB
ZERo
SPAN
ADC.uV
AN.oUT
INPUTS
NoNE
GRoS
NET
DIV.DEC
Note
C
4.2Quick Setup Menu
When settings are complete pres
weigh mode. Pressing any other key exits the setup and discards changes.
Figure 4-1. Quick Setup Menu
s until the indicator displays SAVE?. Press to save set up and return to
18Visit our website www.RiceLake.com
Page 23
Setup Mode
C
4.2.1Default Factory Calibration
The instrument is shipped with the following default calibration settings: See Table 4-4 on page 22 to return indicator to default
settings.
• Capacity – 10,000 lb
• Load cell sensitivity – 2.000 mV/V
• Division – 1
4.2.2Quick Scale Setup
Use this procedure to set the scale(s). Cells may need to be rimmed and summed with a junction box if more than one is
connected. See Section 5.2 on page 33 for full calibration procedures
1.Restart indicator. Momentarily press during startup to display quick setup menu. tyPE displays. Press to set
parameter.
2.Navigate to select desired channel type. Press . nChandisplays. Press to set parameter.
3.Navigate to select number of channels to be used. Press
to set parameter.
4.Navigate to select a channel to configure. Press
5.Navigate to select decimal and minimum division settings
6.Set the total capacity of the scale. Press to confirm. CEL.CAP displays. Press to set parameter.
7.Set the total capacity of the load cells. Press to confirm. CEL.sen displays. Press to set parameter.
8.Set the mV/V sensitivity of the load cells. Press to confirm. dead.Ld displays. Press to set parameter.
9.Set the dead load. If unknown, enter all zeros. Press to confirm.
10. If configuring multiple channels, navigate to return to Chan. Repeat steps 6-18 for each channel to be configured.
div.deC displays. Press to set parameter.
. If multiple channels are chosen, Chan displays. Press
. Press . CAPAC displays. Press to set parameter.
11. Press
12. Press to confirm. Store displays momentarily and the instrument reboots.
until the instrument displays SAVE?.
4.2.3Analog Output
See Section 4.6 on page 30 for Analog Output Settings.
4.2.4Inputs
See Section 4.5.1 on page 27 For Input setup parameters.
• scales with independent channels simultaneously viewable on the PC.
SettingsDescription
in d .C hInstrument connected to 1, 2, 3 or 4 independent scales; to set channels, see Section 3.4 on page 12.
de p.ChInstrument connected to a scale with 2, 3, or 4 dependent load cells (could be digitally summed)
tr ansMIndependent channel; transmits values read by each chanel through the serial line; See Section 3.5 on page 13
Table 4-2. Type Parameter
4.3.2F.Mode Parameters
The function mode parameters set the functionality of the scale. The function mode parameters set four operations that can be
accessed with the > key. It also sets the functionality of printing, taring and zeroing.
Figure 4-3. F.Mode Menu
ParameterSettingsDescription
Fu nc tFunctioning Mode
NOTE: For the details of the operating modes, see Section 3.7 on page 14. Once the functioning mode is selected, if a printer is
configured, the printout is automatically enabled. This parameter is not displayed if trA nSM is set in the tY PE parameter.
Co nverConvert the current displayed value to a calculated value; if the Mode key is pressed for 1 second, the conversion value
can be edited, See Section 3.7.1 on page 14
Al ibiAlibi memory, See Section 3.7.2 on page 14
vi ssSensitivity times ten when the mode key is pressed, See Section 3.7.3 on page 15
PE AKPeak hold detector displays PEAK and alternates with displaying the highest captured value after the mode key is
pressed, See Section 3.7.4 on page 16
reaC tSets the re-enable function of printout based on this criteria: rearm at zero, rearm when weight is unstable, or always print when the print
button is pressed, See Section 3.6 on page 13
ze roRearms print function after weight returns to zero; only prints once after rearming
in stAlways prints when print key is pressed
AL WA YSInstability rearms the print function when the weight becomes unstable; only prints once after rearming and weight
becomes stable
ta reWhen a tare value is entered manually, automatically or from storage the tare value displays with a negative sign when the scale is unloaded.This is
known as a locked tare. An unlocked tare is automatically canceled each time the scale is unloaded, See Section 3.1.5 on page 11
Lo CkRetains tare value until manually cleared
di sa bTare value cannot be entered
Un loC kTare value is automatically cleared when gross weight is zero
rs .z eroEnables restoring the last captured zero after a power cycle
di sa bDisables restore zero after power cycle
En abEnables restore zero after power cycleNOTE: This parameter does not display if TrA nS M is selected in the TY P E parameter.
Co nfi GConfiguration Parameter -See Table 4-6 on page 24
ds p, rfSets the speed of the display refresh
no r mThe function is disabled
20 hz20 display update / sec refresh rate
10 hz10 display update / sec refresh rate
5 h z5 display update / sec refresh rate
2. 5 h z2.5 display update / sec refresh rate
1 h z1 display update / sec refresh rate
Seri a LSerial Communications Setup, See Table 6-3 on page 40
in i. a L
in put sInput Configuration – sets the function of each input
ou tp utOutput Configuration; See Table 4-9 on page 28
An .o u tAnalog Output Configuration; See Table 4-10 on page 30
De FA U
Initialize alibi memory – The initialization cancels all the data stored in the alibi memory; press to enter the operation, theni.A Lib.?
displays; press again to confirm or any other key to cancel; AL. OK
displays if the operation is successful; if not, A L. ERR displays; the
parameter displays only if the alibi functioning mode is selected
Default settings – restores instrument default settings; press ; de fa u? displays; press to confirm or exit by pressing any other key
NOTE: Returning the instrument to default settings cancels the present calibration.
Table 4-4. Setup Parameters
22Visit our website www.RiceLake.com
Page 27
4.3.4Diagnostic Menu
See the Figure 4-2 on page 20 for the diagnostic (diAG) menu structure.
SettingsDescription
Pr G. ver
di v. i nt
ad C.Uv
ad C.Pnt
WeiG h t
CA L.Pt s
di sP LA
Ke yb
Ser
Cts. S t
ou t PU t
in PUt s
An OutAnalog Output Test – provides a basic test to verify correlation of weight and analog output; See Section 4.6 on page 30
SEr. n uMSerial Number – displays the instrument’s serial number
S. ra dio
Press to display the software version
Press to display the calibration internal divisions; The parameter displays only in a primary instrument
Press to display the microvolts relative to the weight on the scale; use
configured channel of the scale; in the dEP .C h mode it is also possible to view the sum of the microvolts of the configured channels;
Su M displays briefly; if the instrument displays the message ER RO R check the connection of the SCT to the junction box and load
cells; the parameter displays only in a primary instrument
NOTE: The maximum input voltage the instrument accepts is 30 mV (30000 µV); the scale system is powered by the
instrument at 5 Vdc; in a properly operating system there will be less than 30000 µV with full capacity on the scale system
Press to display the A/D converter points relative to the weight on the scale; press
points for each configured channel; in the deP . CH mode the sum of the microvolts of the configured channels can be viewed. (C S uM
displays momentarily)
Press to display the weight on the scale; press
Press to alternately display the A/D converter points and the corresponding weight value; use
display to each calibration point and relative weight value
Display Test – press to turn on display segments one at a time; continue pressing ; The instrument turns on the display
segments, one at a time, then exits automatically from this step
Keyboard Test – press , then 0000 displays; press the keys on the keyboard, one at a time, to display related codes; press any
key three times to exit
RS-232 Serial Port Test – press , then XY; X displays, in which Y indicates the status of the PC serial port
• – Serial port is not working
• – Serial port is working
Press
CTS Status Test – press to view the CTS signal status of the printer connected to the PRN serial port
Output Test – press , then rEL.1 displays and output 1 is enabled; press
Input status – press , then 1. b x- Y; X displays, in which y indicates the input status
Press
Press , then 00000 displays. Enter a value between 00000 and 65535 and confirm by pressing ; the instrument assigns
the corresponding analog value in output
Press on the same entered value
Press to select the desired radio channel; OKdisplays if the configuration is successful, E R R o R displays if not successful
Selection of active channel; 1 to 4 in scales with non dependent channels functioning mode (ind . Ch /tr an sm )
NOTE: The parameter is not displayed if dEP .C h is set in the
channel application, SE t uP
Enable or disable the 50Hz filter
Table 4-6. Config Menu Parameters and Settings
Figure 4-5. Configuration Menu
Co nFi G nCh An.
tY PE. parameter or in the event of a single
24Visit our website www.RiceLake.com
Page 29
ParameterSettingsDescription
PA ra MMetrologic parameters
St abiLSelect and set the type and degree of filtering; See Section 4.4.1 on page 26
FL t 0 - 3 – filter for simple weighing
Cu St oM – customizable filter for manufacturer use
F. F. 200 .1 - 3 – filter at 200 Hz
F. F. 50. 1 - 3 – filter at 50 Hz
dy n. 0 - 3 – filter for crane scale
h. r. 0 - 6 – filter for high resolution
F. F. 100 .1 - 4 – filter at 100 Hz
F. F. 40 0 – filter at 400 Hz
NOTE: The F. F. 20 0.1 -3 and F .F .4 00 filters cannot be used in the dE P. Ch 2, 3 and 4 channel functioning
modes.
With a Legal for Trade instrument, only the FL t 0 , FL t 1, FL t 2, F Lt 3 parameters can be selected.
Au to -0Automatic acquisition of the gross zero at startup (default is 2% of capacity)
di SA b – disabled
En Ab – enabled on scale 1
CY CL E – Executed cyclically on all the present scales. This parameter is not visible if there is only one scale
NOTE: If auto zero parameter is enabled, 0.PEr C displays and a value between 1 and 50 as a percentage of
the capacity of the auto zero must be set.
0- PErc
0. tr aCKZero tracking – This menu allows setting the zero tracking (compensation parameter of the scale’s thermal drift); the
di v. S t bDivisions by stability – enter the number of divisions by which the instrument detects the weight stability; a higher
Gr aV,Gravity acceleration – See Section 5.7 on page 37
Ca li bScale calibration – See Section 5.0 on page 32
0. C a L ibZero calibration – See Section 5.0 on page 32
Zero capacity – This menu allows to set (0-50%) of capacity that can be zeroed by pressing zero key ( ); Entering
0% disables the zero key ( )
set value corresponds to the number of divisions tracked off in 1 second
tr 1/2– ± half division
tr 1/4– ± one fourth of a division
tr No – tracking disabled
tr 10– ± ten divisions
tr 8– ± eight divisions
tr 6– ± six divisions
tr 4– ± four divisions
tr 2– ± two divisions
tr 1 – ± one division
number of divisions makes stability more easily detected; Settable values are 0 (weight always stable) to 99
Table 4-6. Config Menu Parameters and Settings (Continued)
Standard digital filtering uses mathematical averaging to compensate for the noise that the A/D converter sends periodically
because of external vibration. This filter compensation makes data less susceptible to a DC signal bias error for some signals.
Each of the filters of the SCT-1100 are intended to compensate for different types and intensities of digital noise and vibration.
Below, find general usage for each family of filters and the A/D Rate and Window for each filter within those families:
NameA/D Rate Window
Static Weighing on platform
FLT 32524
FLT 22516
FLT 12512
FLT 0258
CU ST OMFor manufacturer use
only
High Speed Weight Capture
F. F. 4 0040024
Filling or Dosing
F. 200.320030
F. 200.220032
F. 200.120032
Instability, motion or vibration
F. 50 . 35020
F. 50 . 25022
F. 50 . 15022
Table 4-7. Filter Values
NameA/D Rate Window
Suspended and oscillating loads
DY N.3612
DY N.2612
DY N.1612
DY N.0612
High Resolution Weighing
H. R .6632
H. R .5624
H. R .4624
H. R .3612
H. R .2612
H. R .1610
H. R .068
Filling or Dosing
F. 100.410026
F. 100.310024
F. 100.210020
F. 100.110010
26Visit our website www.RiceLake.com
Page 31
Setup Mode
Note
setup
output
inputs
out. 1
out. 2
fun C1 GRoS
0 Non e
30.net.t
29 err
28.k.tar
27.k.zer
26 k.C
25. k.m o d
23 k. pr
6 moti
5net.0
4 Gro.0
2 net
nC
no
DRCT
STBL
iST. o n
ist.off
xxx.x
xxx.x
no/nC
oNSTAT
RL.iST
ENAB.TM
DELAY
inp.01
inp.02
zero
none
di s. k e y
off
C
pri n t
mode
tare
4.5Input/Output Functions
The instrument is fitted with two opto-isolated inputs and two dry contact outputs. See specifications in Section 2.5 on page 7.
Some of functioning modes of the outputs are relative to the specific functioning modes of the instrument; see the
following descriptions for the details.
Figure 4-6. Output Menu
4.5.1Input Functions
The input configuration menu sets the function of each of the inputs
Parameter SettingsDescription
in pu tsin p .0 1
in p. 0 2
Input 1 or Input 2
NOTE: In the event two inputs are simultaneously enabled, only the lowest number input will activate.
The parameters of each of the outputs is set in the setup outPut menu. See Figure 4-6 on page 27.
Some of functioning modes of the outputs are relative to the specific functioning modes of the instrument; see the
following descriptions for the details.
ParameterSettingsDescription
FU nC
Define the functionality of each output
1 Gr o SSetpoint based on the gross weight (Default)
Functioning with hysteresis ( parameter set at )
Setpoint based on gross weight; Two setpoints for each output must be set; one which disables the output when the
gross weight falls below that setpoint; and one which enables the output when the gross weight is equal or greater than
that setpoint
1. Press and hold to enter the setpoint values for each configured output
2. Select InP . St P. S . 1 on displays (output 1 – this enables the setpoint). Press .
3. Enter the weight value. See Section 4.1 on page 17 for key function. Press to confirm. S.1 o FF
displays (output 1 – this disables setpoint) Press .
4. Enter the weight value. Press to confirm
5. Repeat steps 1 to 8 for all outputs
6. Slowly press multiple times until SAV E? displays.
- Press to confirm and store to the instrument memory
- Press any other key to cancel and exit without saving
Functioning without hysteresis (rL . iSt parameter set at i St . oF F)
Enables output function on gross weight; One setpoint for each output is set
1.Press and hold to enter the setpoint values for each configured output
2.Select InP .St P. S.1 o n displays (output 1 – enabling setpoint) Press .
3.Enter the weight value; See Section 4.1 on page 17 for key function. Press to confirm.
4.Repeat steps 1 to 5 for all outputs
5.Slowly press multiple times until SA VE ?displays.
- Press to confirm and store to the instrument memory
- Press any other key to cancel and exit without saving
• The configuration of setpoints cannot be accessed if all outputs are set in the N O NE functioning mode, or if the
selected functioning mode does not require entry of a setpoint value
• With the instrument off or in standby outputs are normally open
• The disabling setpoint must be equal to or less than the enabling setpoint; if the disabling setpoint is set at a value
greater than the enabling setpoint the instrument sets the setpoint to zero until a valid value is entered
• If the enabling setpoint is set at a value lower than the disabling setpoint, the enabling setpoint is entered and
accepted, however, the disabling setpoint will be set to zero
• A zero value is valid on both the enabling and disabling setpoints
• A setpoint value remains active while modifying the setpoint until the new value is confirmed
• The tare operations are active
• These outputs are enabled by pressing a key (
two seconds the output is disabled and remains disabled until the following pressing of the key
, , , or ); if the key press time is greater than
IMPORTANT: the weight thresholds set with are common to all connected scales; it is not possible to have
different thresholds for different scales. The value assumes the unit of measure and decimals of the selected
channel; for example, if 1000 is the set value, the setpoint values will be the following:
ChannelUnit of MeasureDecimalssetpoint Value
1kg31.000 kg
2g01000 g
3kg210.00 kg
4g1100.0 g
Table 4-9. Output Functions
28Visit our website www.RiceLake.com
Page 33
Setup Mode
C
ParameterSettingsDescription
FU nC0 no neNo function, the output is inactive
2 ne tSetpoint based on net weight. Setpoints are set in the same manner as gross weight; See 1 G ros in Table 4-9 on
page 28. In addition setpoints can be set and activated on a negative weight
• Positive weight (SiG n set at P O S it )
• Negative weight (SiG n set at n E GA t)
30 . net .tSetpoint based on the net weight with tare activated; selecting this mode the function of the output on the net weight is
activated if a tare is entered
29 E rrError indication. Function of the output is enabled on an invalid weight (overload/underload), or without the signal
coming from the cell (disconnected cell)
NOTES:
• In the ind . Ch and trAn SM mode the output is enabled only when the condition takes place on the selected
channel.
• In the dEP .C h mode the output is enabled when the condition takes place on any of the set channels.
28 k .t arTare Key – function is enabled when TARE key is pressed
27 k . ze rZero Key – function is enabled when ZERO key is pressed
26 k. c
25 k. M odMode Key – function is enabled when MODE key is pressed
23 k. PrPrint Key – function is enabled when PRINT key is pressed
6 Not iInstability – Output is on for an unstable weight
5 n E t .oSetpoint based on net weight being at zero
4 G r o. 0Setpoint based on gross weight being at zero
no /n cNO/NC Contacts
• n o – output normally opened
• n C – output normally closed
on St atSwitching condition
• d rC t – the output is activated when the weight reaches the set threshold, (independently from the stability) and is disabled when
the weight goes below the set disabling threshold
• S tb L – the output is activated when the weight, after reaching the set activation thresholds, becomes stable; the output is disabled
when the weight, after going below the set disabling threshold, becomes stable
rL .i stHysteresis
• i St .o FF – hysteresis disabled
• i St .o n – hysteresis enabled
En Ab.t MEnabling time – enter the length of time the output is enabled in seconds (4 digits with a decimal); the output is disabled once the set
time has passed, starting from the moment of the activation (see deL ay ); by setting 000.0 the output remains always active
NOTE: The delay time is only evaluated when a setpoint on gross weight, setpoint on net weight or is selected as an output
function.
de La yEnables delay period – enter the enabling delay period in seconds (4 digits with a decimal);
The output is enabled once the set time has passed, starting from the moment the condition takes place; By setting 000.0 the output is
enabled when the enabling condition takes place
NOTE: The output is enabled only if the enabling condition takes place for the length of time set. The delay is valid only for the
enabling of the output. When the enabling condition no longer takes place the output is disabled.
In a switching condition with stability, the output is enabled only when the weight is stable. The delay time is evaluated when a
setpoint on gross weight or setpoint on net weight is selected as an output function.
An optional analog output is configurable at 0-20mA, 4-20mA, 0-10Vdc, 0-5Vdc; with minimum and maximum settable values.
The output voltage and the current from the interface are proportional to the gross weight or net weight present on the scale.
The analog output is updated every 20ms and takes on the value corresponding to the weight converted in that instant;
therefore, as the filter is increased, the analog output update rate will slow down.
Scale needs to be calibrated per Section 5 before analog output can be calibrated
To configure the parameters, enter the setup mode andSEtuPAn.out.
Figure 4-7. Analog Output Menu
ParameterDescription
ch anSelect the active channel 1-4 in the scale with non dependent channels mode (/ ); The parameter is not displayed
is set in the parameter or in a single channel application,
ModeSelect the type of analog output:
Ao no – analog output disabled
Ao Gro – analog output tracks gross weight
Ao nEt – analog output tracks net weight
Once the functioning mode is confirmed, set the values of the analog output; the digital/analog converter values are entered
(between 0-65535) which corresponds to an output value in voltage or in current
The instrument keys have the functions:
– decreases the selected digit (blinking)
– increases the selected digit (blinking)
– selects the digit (blinking) from left to right
– press once to enter a value, the corresponding output analog value is enabled; press a second time to confirm and exit the step
– press to quickly zero the present value to 000000
C
Ao MaxSet the maximum value of the analog output:
• With a positive weight – the value of the output when the weight is greater than or equal to the full scale capacity; also corresponds to
the overload condition
• With negative weight – the value of the output when the negative weight is greater than or equal to the full scale capacity, also
corresponds to the underload condition
The value can be between 0 and 65535 (values of the digital/analog converter); if a higher value is entered, the instrument zeros the value
Ao ZerSet the analog output value when the scale displays zero weight (supplied when the scale is in underload); this value can be
between 0-65535 (values of the digital/analog converter); if a higher value is entered, the instrument zeros it
Ao MinSet the minimum value of the analog output:
• With positive weight – the minimum value provided by the analog output, corresponding also to the underload condition
• With negative weight – the minimum value provided by the analog output, corresponding also to the overload condition
This value can be between 0 and 65535 (values of the digital/analog converter); If a higher value is entered, the instrument zeros it.
Table 4-10. Analog Output Parameters
30Visit our website www.RiceLake.com
Page 35
Setup Mode
Ao = Analog output
Calibrating Analog Output
The following calibration procedure requires a multimeter to measure voltage or current output from the analog output module.
1.Enter setup mode and go to the A n out menu (see Table 4-10 on page 30).
• Set ao M I N to lowest weight value to be tracked by the analog output
• Set ao M A X to highest weight value to be tracked by the analog output
2.Connect multimeter to analog output:
• For voltage output, connect voltmeter leads to pins 3 and 4
• For current output, connect ammeter leads to pins 1 and 2
3.Adjust zero calibration: Scroll to the ao ZE R parameter. Check voltage or current reading on multimeter. Press and
hold or to adjust the zero value up or down.
4.Return to normal mode. Analog output function can be verified using test weights.
Approximate Values Between The DA Converter and Analog Output
The type of calibration used is dependent on the type of application chosen for the instrument: independent channels (ind.Ch)
and dependent channels for (dEP.Ch), which can be digitally equalized. See the type parameterin Table 4-2 on page 21.
5.1Calibration Menu
Figure 5-1. Calibration Menu
32Visit our website www.RiceLake.com
Page 37
Calibration
Note
5.1.1Calibration Parameters
ParameterSettingsDecription
NC hanSelect number of channels. See Table 4-6 on page 24
Ch anSelect current channel to configure. See Table 4-6 on page 24
Gr aV,9,75 00 1-9 ,8 4999
default: 9, 80 655
de cDecimal Point Location – when combined with the decimal point location, specifies the location of the decimal point or dummy zeroed in the
unit display
u. n .Units – specifies units for displayed and printed weight
di vDisplay Divisions – selects the minimum division size for the displayed weight; scale capacity is determined by display division x graduations
RA NGE 1Maximum weight for first range or interval
RA NGE 2Maximum weight for second range or interval
CA LL I B. PCalibration
NT PNumber of calibration points
TP OSet weight value of unloaded scale
DD T 1Enter weight value of first sample weight
TP 1Add first sample weight and set calibration point
DD T 2Enter weight value of second sample weight
TP 2Add second sample weight and set calibration point
DD T 3Enter weight value of third sample weight
TP 3Add third sample weight and set calibration point
TH EO . CATheoretical Calibration
CE L. SE NCell sinsitivity in mV/V
CE L .C APCell Capacity in the configured unit of measure
DE AD . LDWeight of the structure bearing on the load cells
KN O. W G TKnown value of the sample weight
MA N. C ALManual Calibration - manually change the weight and ADC value of calibration points
MO D. PN THSelect calibration point to change
WE IG HTEnter or confirm weight value
PO IN TSEnter or confirm ADC value
0.calibPerforms a zero calibration
Gravity acceleration – select the acceleration value of calibration location and installation location of the instrument;
manual entry of the g value: the gravitational acceleration value may be manually entered; the minimum decimal
value is 9.75001m/s
2
; any decimal number that is not between 9.75001 and 9.84999 m/s2 (inclusive), is incorrect.
Table 5-1. Calibration Parameters
In the case that a number needs to be entered to set a parameter, press to select the digit to be modified and
or to increase or decrease the digit.
To navigate a menu to select an option, Press or .
5.2Procedure
Use this procedure to set the scale(s). To connect more than four cells, it is necessary to trim the cells.
1.Restart indicator. Momentarily press during startup to display setup mode. tyPE displays. Press .
2.Navigate to select channel type. Press when the desired channel type is displayed to select it.
3.Navigate to s e t u p
4.Navigate to set the number of channels (Ch1 - Ch4). Press to confirm the desired number of channels. FiLt.SO
displays.
5.Navigate to CALib. Press to set parameter. dECi displays. Press to set parameter.
6.Navigate menu to select the decimal point location
parameter.
Co nF iG nChan. Press to set parameter.
. Press to confirm selection. u.M. displays. Press to set
6.Navigate to set the weight unit. Press to confirm. diV displays. Press to set parameter.
7.Navigate to set the display divisions. Press
to confirm. rAnGE1 displays. Press .
8.Enter the total capacity of the scale or the first range in case of multi-range functioning. Press to confirm. rAnGE
2 displays. Press to set parameter.
9.Enter the second range or enter all zeros, if only one range. Press to confirm. CALib.P displays.
10. Navigate to thEo.CA displays. Press to confirm. CEL.SEn displays. Press to set parameter.
11. Enter cell sensitivity. Press to confirm
. CEL.CAPdisplays.Press to set parameter.
12. Enter cell capacity. Press to confirm. dEAd.Ld displays. Press to set parameter.
13. Enter the dead load. If unknown, enter all zeros. Press to confirm. Kno.WGT displays.
Cell Sensitivity - if several load cells are connected through a junction box enter the average sensitivity value of the
cells.
Cell Capacity - if several load cells are connected through a junction box, enter the sum of the load cells.
By setting the value to zero, the dead load is acquired.
14. Repeat steps 3-13 for each connected scale.
15. Press
until the instrument displays SAVE?. Press to confirm. Store displays momentarily and the instrument
C
reboots.
5.6.2Dependent Channels
Use this procedure to perform a theoretical calibration on a scale in the dEP.Ch functioning mode
1.Restart indicator. Momentarily press during startup to display setup mode. Type displays. Press to set
parameter.
2.Navigate to select dEP.Ch. Press to confirm. F.Mode displays.
3.Navigate to s e t u p
4.Navigate to set the number of channels (Ch1 - Ch4). Press to confirm. FiLt.SO displays.
5.Navigate to CALib. Press . DECi displays. Press to set parameter.
6.Navigate to set the decimal place. Press to confirm. u.M. displays. Press to set parameter.
7.Navigate to set the weight unit. Press to confirm. diV displays. Press to set parameter.
8.Navigate to set divisions. Press
9.Enter the total capacity of the scale or the first range in case of multi-range functioning. Press to confirm. r AnGE2
displays. Press to set parameter.
10. Enter the second range or enter all zeros, if only one range. Press to confirm. CALib.P displays.
11. Navigate to thEo.CA. Press to enter menu. CEL.SEn displays. Press to set parameter.
12. Enter cell sensitivity.Press to confirm. CEL.CAP
13. Enter cell capacity. Press to confirm.dEAd.Ld displays. Press to set parameter.
14. Enter the dead load. If unknown, enter all zeros.
Cell Sensitivity - if several load cells are connected through a junction box enter the average sensitivity value of the
cells.
Cell Capacity - if several load cells are connected through a junction box, enter the sum of the load cells.
By setting the value to zero, the dead load is acquired.
15. Press to confirm. Kno.WGT displays.
16. Repeat Steps 3-15 for each connected scale.
Co nF iG nChan. Press to set parameter.
to confirm. rAnGE1 displays. Press to set parameter.
displays.
Press to set parameter.
17. Press
until the instrument displays SAVE?.
18. Press to confirm. Store displays momentarily and the instrument reboots.
36Visit our website www.RiceLake.com
Page 41
Calibration
Note
IMPORTANT
5.7Gravity Setting
Use this procedure to correct the weight error caused by a different gravitational value between the calibration zone and the
zone of use.
1.Calibrate the indicator. Restart.
2.Navigate to s e t u p
3.Enter the local gravity
Co nF iG Grav. Press to set parameter.
. Press . Default is 9.80390.
4.Slowly press
• Press
multiple times to exit the menus until SAVE? displays.
C
to confirm and return to the weigh mode
• Press any other key to cancel and exit without saving
The weight error caused by a different gravitational value between the calibration zone and the zone of use is automatically
corrected.
Press and hold when turning on the instrument. The g value relative to the gravitation zone of the user displays for a few
seconds, after the name and the installed software version.
To find the local gravity, enter the latitude and elevation into the International Gravity Formula.
Listed are links to websites that can be used to determine local latitude and altitude. Please note these website
address are provided for reference only and may change.
Map Coordinates uses Google maps to find latitude and elevation: www.mapcoordinates.net/
Once local latitude and altitude have been determined, use the following link to calculate local gravity
The gravity correction function has not been evaluated by an approvals agency, therefore it is up to the
authorized scale dealer to ensure the device is accurate at the intended point of use.
5.8Zero Dead Load A/D Counts
Table 5-2 lists the ideal A/D counts that result from input signals of 0Ð45 mV with zero deadload. Actual values will typically be
higher than the values shown in Table 5-2 but the ideal values can be used when calibrating the indicator with no attached
scale.
The instrument has two bidirectional serial outputs which are ASCII code compatible with most printers, remote displays, PCs
and other devices. See Section 2.5 on page 7 for connection information.
Transmission of data through the serial ports can be configured in the parameters PC SEL, PCModE and Pr.ModE in the setup
mode. See Section 6.2 on page 39
6.1.1COM1 Serial Port
The COM1 serial port is bi-directional (half duplex) and uses RS-485 for transmitting data. It is mainly used to connect PCs,
PLCs and additional remote displays. The transmission speed may be selected in the setup as: 1200, 2400, 4800, 9600,
19200, 38400, 57600, 115200 baud (bit/sec).
On the same RS-485 line, it is possible to connect up to 32 devices (instruments, RS-485/RS-232 signal converter).
Considerations when making RS-485 connections:
• Use a Shielded Twisted Pair cable to make the connection (twisted and shielded pair(s) with single shielding for each
pair through aluminum band and total shielding through external shielding)
• Using 2x24 AWG duplex cable with external shielding, the RS-485 cable should not exceed 4000' (1200 m); See
Section 2.3 on page 5
• With very long cables, cable capacity becomes a dominant factor in power consumption (normally near 50pF/m); cable
capacity decreases as length increases; capacity also decreases when speed is increased; the maximum distance
cannot be covered with the maximum possible speed
.
Baud Rate (bit/sec) Total Cable Capacity (pF)
1200400000
2400200000
4800100000
960050000
1920025000
3840012000
576008000
1152004000
Table 6-1. Cable Capacity
• Verify single point grounding on all equipment; See Section 2.4 on page 6
• Use correct single point grounding to avoid forming ground loops
• On the RS-485 network, two termination resistances equal to the impedance of the cable (typically 120 Ω) are normally
connected on the two devices at the ends of the cable. The terminal resistance is not supplied with the ports of the
instrument
• Consult the device product data sheet for all connected devices to ensure consistency in the markings
6.1.2COM2 Serial Port
The COM2 serial port is bi-directional (full duplex) and uses an RS-232 for transmitting data. It is mainly used to connect
printers, PCs, and PLCs. The transmission speed may be selected in the setup as: 1200, 2400, 4800, 9600, 19200, 38400,
57600, 115200 baud.
38Visit our website www.RiceLake.com
Page 43
6.2Serial Port Transmission Modes
Seri a L
pr- n o
PrPC. h k
repe.6
prpC.ex
prpC.st
aLL. ex t
ALL. s td
tpr
485
232
bi t
baud
PCMo DE
pWR.PRN
bi t. Pr
baud.pr
pC SEL
Co m. p C
Co m. p rn
Pr Mod e
9600
4800
2400
1200
11 5 2 0 0
57600
38400
19200
9600
4800
2400
1200
11 5 2 0 0
57600
38400
19200
Mod b us
FLD.b uS
aLL.MAX
ALL.STD
ALL.EXT
STAB.ST
STAB.EX
onde
repe.6
pri n.st
pri n t.ex
485
PWr.Ex t
ext.o ff
Prn.Ct s
CtS h
STSL
noCTS
EmuCts
Eng l
ital
esp a
fran
deut
PrConf
La n g
n-8-1
o-8-1
o-7-2
-7 -1
e-7-2
e-7-1
n-7-2
n-8-2
n-8-1
o-8-1
o-7-2
-7 -1
e-7-2
e-7-1
n-7-2
n-8-2
Communications
6.2.1PC Port Selection
It's possible to select the serial port to be used as a PC port. When a port is selected for PC port transmission, the other serial
port is selected as default for PRN port transmission.
This setting is made in theSEtuP
Select the RS-485 serial port as the PC PORT and the RS-232 serial port sets as the PRN PORT.
Select the RS-232 serial port as the PC PORT and the RS-485 port sets as the PRN PORT.
ParameterSettingsDescription
PC SeL Select the communication carrier for the ports
SEriALPCSEL step.
485Communication between the instrument and the PC takes place through the RS-485 port and
232Communication between the instrument and the PC takes place through the RS-232 port and transmission of data
transmission of data to the printer through the RS-232 port
This section describes the selectable serial weight transmission modes of the PRN serial port through the parameter
set in PR Mode
ParameterSettingsDescription
Com.PrnSerial Format for the printer port
Pr. ModeTransmission to serial printer
Pr- noTransmission disabled
PrPC.HK
rEPE.6The weight is displayed on the instrument and is transmitted to a 6 digit remote display
PrPC.EX
PrPC.St
ALL.EXt
ALL.Std
tPrEnables printing with ASCII compatible printer
NOTE: In the
For the protocol and transmission mode specifications, see Section 6.2 on page 39.
BAUD.PRSet baud rate - selection of the data transmission speed (baud = bit/second); (9600 default)
bit.prSet parity, word, stop bit
Pwr.PrnManufacturer Use Only
Prn.CtsManufacturer Use Only
Pr.ConfLangSelect Language of Printouts
protocol, the serial output is automatically set at 4800, N-8-1 but can be configured differently.
Transmission of the weight string by pressing
PrPC.EX allows for transmission of the extended string by pressing enter See Section 6.4.2 on page 47; (or multi-
scale string in the TRANSM mode)
PrPC.St allows for transmission of the standard string by pressing enter See Section 6.4.1 on page 47
Transmission when is pressed – the instrument transmits the weight data through the serial port when
is pressed; Transmission takes place if the weight is stable and the net weight is > 20 divisions;
re-enabling the transmission depends on how rEACt is set in the setup mode (passing by zero of the net weight,
weight instability or always)
Data is transmitted with the standard string Prin.St or the extended string; See Section 6.4 on page 47 for a
description of the strings; The transmission is confirmed whentrAnSM displays
Continuous Transmission for interfacing to the PC, remote displays and other devices which request a constant
stream of the data independently from the weight stability; the instrument transmits data with each A/D cycle:
•Baud rate at 9600 up to 10 transmissions per second are possible
•Baud rate at 115200 up to 16 transmissions per second are possible for the PC port and up to 12 for the
printer port
The data transmits both positive and negative values
•ALL.Std - The data is transmitted using the standard string
•ALL.EXt - The data is transmitted using the extended string; (or multi-scale string in the TRANSM mode)
- SeeSection 6.4 on page 47for a description of the strings
NOTE: Filter selection directly affects data transmission; To obtain 250TX/sec configure the filter F.F.400
(SEtuP
ConFiGPArAM.StAbiL)
Data is transmitted to the printer by pressing on the instrument; the print command is inhibited if the weight is
in motion and in all other circumstances in which the data is not valid
NOTE: Language selection only available if tpr is selected
enter
Table 6-3. Serial Menu Parameters and Settings
40Visit our website www.RiceLake.com
Page 45
6.2.3PC PORT
This section described the selectable serial weigh transmission modes of the PC serial port.
ParameterSettingsDescription
CoM. PCPC Serial
PCModeTransmission on the PC Serial
ModbusTransmission with the MODBUS protocol
•mod.typ displays; press ; select ascii or rtu; press
•Mod.add displays; press ; enter the address of the unit (0 to 98); press
•Enter baud rate; press
•Enter bit parameters; press
ALL.max
ALL.STD
ALL.EHT
STAB.ST
STAB.EH
ondeTransmission requested on demand, from an external device – the instrument waits for a command before
repe.6Transmission to 6 digit remote display / Reception of the “rEPE.6” string
prin.st
prin.eh
485Transmission in RS-485 serial mode; Protocol mimics the command that was sent; the instrument responds only
BAUDSet baud rate - selection of the data transmission speed (baud = bit/second)
bitSet parity, word, stop bit.
Continuous Transmission for interfacing to the PC, remote displays and other devices which request a constant
stream of the data independently from the weight stability;
the instrument transmits data with each A/D cycle:
•Baud rate at 9600 up to 10 transmissions per second are possible
•Baud rate at 115200 up to 16 transmissions per second are possible for the PC port and up to 12 for the
printer port
The data transmits both positive and negative values
•ALL.Std - The data is transmitted using the standard string
•ALL.EXt - The data is transmitted using the extended string (or multi-scale string in the TRANSM mode)
- SeeSection 6.4 on page 47for a description of the strings
•ALL MAX? - The weight is transmitted in hexadecimal format (for example: 03E8= 1000g), without decimal
point
- This transmission protocol is recommended for applications where a high nubmer of output
transmissions is required. (up to 250TX/sec. with baud rate equal to 115200)
NOTE: Filter selection directly affects data transmission; To obtain 250TX/sec configure the filter F.F.400
(SEtuP
ConFiGPArAM.StAbiL)
Transmission on stability – each time a weight on the scale becomes stable, a communication string is transmitted
on the PC port;
•The transmission takes place when the weight is stable and the net weight is greater than 10 display
divisions
•Re-enabling the transmission depends on how rEACt has been set in the setup mode; See Table 4-3 on
page 21
•The data is transmitted with the standard string StAb.St or the extended string StAb.EX; (or multi-scale
string in the TRANSM mode) See Section 6.4 on page 47for a description of the three strings
transmitting data; See Section 6.3 on page 42
With baud rate at 9600, up to 10-11 requests per second are possible through the READ command; with baud
rate at 115200, up to 16 requests per second are possible through the READ command;
The data transmits both positive and negative values
The weight display occurs both in the instrument and is transmitted to a 6 digit remote display
The instrument communicates the weight data through the serial port when is pressed
•Transmission takes place if the weight is stable and the net weight is > 20 divisions; re-enabling
transmission depends on how the rEACt is set in the setup mode (passing by zero of the NET weight,
weight instability or always)
•Prin.St - The data is transmitted using the standard string
•Prin.EX - The data is transmitted using the extended string
- SeeSection 6.4 on page 47for a description of the strings
•The transmission is confirmed when trAnSM displays
if its ID is the one requested (before the request the module ID must be input, i.e. 00READ<CRLF>);
If a broadcast address command (99) is received no answer is given; If the command is correct it is executed
[CC] o <II>Instrument ID, e.g. 00 (The ID is only used with RS-485 protocol)
<CR LF>Carriage Return plus Line Feed (ASCII character 13 and 10)
<ESC>ASCII character 27
<STX>ASCII character 02
BSpace character, ASCII character 32
Table 6-5. Serial Command Legend
Serial Errors
The instrument transmits a response string or it transmits one of the following indications with each serial command received:
Status ResponseDescription
OK<CR LF>Displays when a correct command is transmitted from the PC to the instrument; the OK does not imply that the instrument
executes the zero
ERR01<CR LF>Displays when a correct command is transmitted from the PC to the instrument, but the command is followed by unexpected
letters; for example READF, TARES instead of READ, TARE
ERR02<CR LF>Displayed when a correct command is transmitted from the PC to the instrument, but contains wrong data
ERR03<CR LF>Displayed when an incorrect command is received; when the command may not be used in the selected functioning mode; or
when the command is received while the keyboard buffer is already full
ERR04<CR LF>Displayed when a nonexistent command is received
Table 6-6. Serial Status Response
NOTE: The instrument does not transmit an indication with momentary commands; for example, those made up of
only one letter and then the parameter.
Version Reading Command
[CC]VER<CR LF>
Instrument response: [CC]VER,vvv,DGT1Sbbb<CR LF>
In which:
• vvv is the firmware version
• bis the space character, ASCII 32
Extended Weight Read Command
[CC]REXT<CR LF>
Instrument response in the V mode or in thedEP.Ch
Instrument response in the
If the instrument is in theind.Chmode (scale with independent channels) or in thedEP.Ch
dependent channels and digitally equalized) the weight value is read relative to the active channel; to read the values
of the other channels (if configured) switch to the desired channel. See Converter Channel Switching Command on
page 43.
If the instrument is in the
values for all the configured channels
trAnSMmode: multi-scale string. SeeSection 6.4.3 on page 48.
trAnSMmode (scale with independent channels) it is possible to simultaneously read the
mode: extended string. See Section 6.4.2 on page 47.
Weight Read Command
[CC]READ<CR LF>
Instrument response: standard string (Section 6.4.1 on page 47).
mode (scale with
Weight Reading Command With Sensitivity Times 10
[CC]GR10<CR LF>
Instrument response: standard string (Section 6.4.1 on page 47).
42Visit our website www.RiceLake.com
Page 47
Reading Command of MicroVolts Relative to the Weight
Note
[CC]MVOL<CR LF>
Instrument response in
Response of the instrument in
If the instrument is in theind.Chmode (scale with independent channels) or in thedEP.Ch
dependent channels and digitally equalized) the weight value is read relative to the active channel; to read the values
of the other channels (if configured) switch to the desired channel. See Converter Channel Switching Command.
ind.Chmode: standard string (Section 6.4.1 on page 47).
trAnSManddEP.Chmode: multi-scale string (Section 6.4.3 on page 48).
Communications
mode (scale with
If the instrument is in the
values for all the configured channels.
trAnSMmode (scale with independent channels) it is possible to read simultaneously the
Tare Command
[CC]TARE<CR LF> or [CC]T<CR LF>
Instrument response: [CC]OK<CR LF>
Zero Command
[CC]ZERO<CR LF> or [CC]Z<CR LF>
Instrument response: [CC]OK<CR LF>
Clear Command
[CC]CLEAR<CR LF> or [CC]C<CR LF>
Instrument response: [CC]OK<CR LF>
The command also works in the setup mode.
Converter Channel Switching Command
[CC]CGCHN<CR LF>
Instrument answer: [CC]OK<CR LF> if the CGCH command has been received.
In which: N is the number of the channel on which to position the instrument
Test Command
[CC]ECHO<CR LF>
Instrument response: [CC]ECHO<CR LF>
Print Command
[CC]PRNT<CR LF> or [CC]P <CR LF>
Instrument response: [CC]OK<CR LF> if the command has been received, no answer for the P command.
Tare Insertion Command
[CC]TMANVVVVVV<CR LF> or [CC]WVVVVVV <CR LF>
In which: VVVVVV is the manual tare value with the decimal point, from 1 to 6 characters; the non significant zeros can be
omitted.
Instrument response: [CC]OK<CR LF> if the command has been received; no answer for the W command.
Command for Displaying Temporary Message on an Instrument
[CC]DISPNNVVVVVV <CR LF>
In which:
• NN is the instrument display number, standard 00 (ASCII hex)
• V is the message:
• if present, it is shown on the NN display
• if not present, the command interrupts the possible visualization enabled with a previous DISP command, restoring
the visualization of the weight data
If the display shown in the command is numeric (for example the standard display 00) and in the transmitted
message there are two consecutive points, the message is stopped after the first of the two points. When the display
is showing a message transmitted serially through the DISP command, the instrument does not display those
messages usually shown in the scale status (ZERO, TARE, HOLD, etc.).
Instrument response: [CC]OK<CR LF>
The message remains for the time set through the DINT command
The ASCII characters having the decimal code greater than 31 are accepted.
Command for Setting Display Message Interval
[CC]DINTNNNN<CR LF>
In which: NNNN is the visualization interval (in milliseconds), expressed in ASCII hex character; for example, in order to set a
visualization time of 2 seconds (2000 milliseconds, which converted into hex it becomes 07D0), the command becomes
[CC]DINT07D0<CR><LF>.
By setting a time equal to zero, the message transmitted with the DISP command remains permanently shown on the display.
Instrument response: [CC]OK<CR LF>
PC Confirmation Command
[CC]PCOK<CR LF>
The instrument shows on the display the
message for about two seconds.
Instrument response: [CC]OK<CR LF>
Serial Command which Returns the Instrument Status
[CC]STAT<CR LF>
Instrument response: [CC]STATXX<CR LF>
In which: XX is a decimal value which returns the status of the instrument; the possible values are:
XXInstrument Status
00Normal scale status
01Normal scale status in input
02Instrument in technical setup
03Instrument in boot phase
04Instrument in rx/tx setup phase
05Instrument in test phase of the serial ports
06Instrument in print test
07Instrument in firmware update phase
08Instrument in standby
09Instrument in automatic zero phase
10Instrument in change channel
11Instrument in inputs test phase
Instrument response: [CC]OK<CR LF>: accepted command.
In the event the simulated key has two linked functions, key momentarily pressed or pressed at length, if the KEYP command is
followed by the release command (KEYR) within a maximum time of 1.5 seconds, the momentary key press is executed;
otherwise the key pressed at length is executed.
- ZERO key
- TARE key
- MODE key
- PRINT key
- C key
Release Key Press Simulation Command
[CC]KEYR<CR LF>
Instrument response: [CC]OK<CR LF>
The instrument does not respond OK to the following momentary commands (P, Q, T, W, X, Z).
OLOverload
STStability of the display
USInstability of the display
TLActive inclination input
BNumber of platform on which the totalization has been made
NNNNNNNUMNet weight with unit of measure
LLLLLLLUMGross weight with unit of measure
YYTare Type; Blank spaces if semi-automatic tare; PT If preset tare
•002 Setup menu
AAACounter of pressed keys
CCCCode of last key pressed
TTTNot used
XXXXXLast rewriting number stored in the alibi memory
YYYYYYLast weigh number stored in the alibi memory
[CC]STPTntxxxxxxtyyyyyy<CR LF>
In which: n indicates the setpoint number (1, 2)
F if the following weight value indicates that the setpoint will disable the outputs (OFF).
t
t
O if the following weight value indicates that the setpoint will enable the outputs (ON).
xxxxxx and yyyyyy represent the weight value of the setpoint that disables or enables the outputs: the digits must be entered
without the decimal point, omitting the non-significant zeros.
Instrument responses: [CC]OK<CR LF> correct syntax and correct values have been recieved
[CC]NO<CR LF> correct syntax but wrong values have been recieved
Example of instrument with capacity 10.000 kg and division 1 g:
Command: STPT1F5000O6500 (Disabling first output at 5 kg and enabling at 6.5 kg)
Instrument response: OK
The ERR 02 code displays when:
One of the two entered values is greater than the capacity.
One of the two entered values has a minimum division that is inconsistent in comparison to the one set in the
instrument.
The disabling value is greater than that of enabling.
The transmitted values are valid until the instrument is turned off. To permanently save these on the instrument use
the saving command (CMDSAVE). To save various setpoints set all of them and at the end transmit the saving
command.
Setpoint Saving Command
[CC]CMDSAVE<CR LF>
Response: [CC]OK<CR LF>
Enable/Disable Keyboard
To enable the keyboard: [CC]KEYEE<CR LF>
Instrument response: [CC]OK<CR LF>
To disable the keyboard: [CC]KEYED<CR LF>
Instrument response: [CC]OK<CR LF>
46Visit our website www.RiceLake.com
Page 51
Communications
Note
6.4Transmission Protocols
The weight data transmission on the PC and PRN serial ports may take place in 3 formats: standard string, extended string or
multi-scale string.
6.4.1Standard String
String transmitted in the dependent or independent channel mode: [CC]hh,kk,pppppppp,uu <CR LF>
String transmitted in the TRANSM mode: [CC]hh,pppppppp,uu <CR LF>
In which:
CharactersDescription
[CC] The instrument ID as two ASCII decimal digits (RS-485 protocol)
hhULUnderload
OLOverload
STStability of the display
USInstability of the display
, Comma character
kkNT Net weight
GS Gross weight
GX Gross weight with sensitivity times 10
VLValue in microvolts relative to the weight
RZValue in converter points relative to the weight
, Comma character
pppppppp8 digits (including sign and decimal point) which identify the weight; the insignificant digits are filled with spaces; through the MVOL
and RAZF command the instrument transmits the relative value on 10 digits instead of 8
uuUnit of measurement kg, bg, bt, lb, mv (microvolts), vv (A/D counts); (b signifies blank)
<CR LF>Carriage Return + Line Feed (ASCII decimal character 13 and 10)
Table 6-10. Standard String Characters
The transmitted weight is gross weight (GS) if no tare weight has been entered; otherwise, net weight (NT) is
transmitted.
6.4.2Extended String
Without APW (any mode other than in response to the REXT command):
[CC]B,hh,NNNNNNNNNN,YYTTTTTTTTTT,PPPPPPPPPP,uu,(dd/mm/yybbhh:mm:ss|NO DATE TIME)<CR LF>
With APW (in response to the REXT command):
[CC]B,hh,NNNNNNNNNN,YYTTTTTTTTTT,PPPPPPPPPP,AAAA.AAAAA,uu<CR LF>
In which:
CharactersDescription
[CC]The instrument ID as two ASCII decimal digits (RS-485 protocol)
BScale number is always 1
,Comma character
hhULUnder load
OLOverload
STStability of display
USInstability of display
, Comma character
NNNNNNNNNN Net weight on 10 characters including possible sign and decimal point
,Comma character
YYPT if the tare is manual, if YY = two empty spaces display with semiautomatic tare
, Comma character
TTTTTTTTTTTare weight on 10 characters including possible sign and decimal point
, Comma character
PPPPPPPPPPAlways 0
, Comma character
uuUnit of measure Kg, bg, bt, lb; (b signifies blank)
, Comma character
dd/mm/yyDate in the dd/mm/yy format (only with REXD command)
bbTwo space characters, ASCII decimal 32 character (only with REXD command)
hh:mm:ssTime format (only with REXD command)
<CR LF>Carriage Return + Line Feed (ASCII decimal character 13 and 10)
The non significant digits of the net, tare, pieces and gross weights are filled with spaces (space characters, ASCII
decimal 32 character).
If the optional TIME DATE board has not been detected, in response to the REXD command, only the weight is
transmitted and not the date and time; in its place there is NO DATE TIME.
6.4.3Multi-Scale String
The string can vary depending on the configured channels:
• [CC]hh,pppppppp,uu, (dd/mm/yybbhh:mm:ss|NO DATE TIME)<CR LF>
• [CC]hh,pppppppp,uu,hh,pppppppp,uu, (dd/mm/yybbhh:mm:ss|NO DATE TIME)<CR LF>
• [CC]hh,pppppppp,uu,hh,pppppppp,uu,hh,pppppppp,uu, (dd/mm/yybbhh:mm:ss|NO DATE TIME)<CR LF>
• [CC]hh,pppppppp,uu,hh,pppppppp,uu,hh,pppppppp,uu,hh,pppppppp,uu, (dd/mm/yybbhh:mm:ss|NO DATE TIME)<CR
LF>
In which:
CharactersDescription
[CC] is the instrument code as two ASCII decimal digits (RS-485 protocol)
For each set channel:
hh•STStability of the display
,Comma character
pppppppp8 digits (including eventual sign and decimal point) which identify the weight. The insignificant digits are filled with spaces. Through the
,Comma character
uuUnit of measure (b signifies blank)
,Comma character
dd/mm/yyDate in the dd/mm/yy format (only with REXD command)
bb2 space characters, ASCII decimal 32 character (only with REXD command)
hh:mm:ssTime format (only with REXD command)
•USInstability of the display
•VLValue in microvolts relative to the weight
•RZValue in converter points relative to the weight
MVOL and RAZF commands the instrument transmits the relative value on 10 digits instead of 8.
•kg
•bg
•bt
•lb ,
•mv (microvolts)
•vv (converter points)
Table 6-12. Multi-Scale String Characters
48Visit our website www.RiceLake.com
Page 53
Communications
Note
C
6.4.4Secondary Mode Strings
Secondary Standard String
Standard string transmitted on the print port whenPR.MODE= ALL.STD or PRPC.ST; SS,NT,WWWWWWWW,UU<CR LF>
In which:
CharactersDescription
SSStatus:
•NV Weight not valid
,Comma character
NTSTStable data
USUnstable data
ULUnderload
OLOverload
,Comma character
WWWWWWWWWeight
,Comma character
UUUnit of measure
Table 6-13. Secondary Standard String Characters
Secondary Extended String
Extended string transmitted on the print port whenPR.MODE= ALL.EXT or PCPR.EX
C, SS,NT,WWWWWWWW,UU<CR LF>
In which:
CharactersDescription
CSecondary or sum:
•S, if the sum is sent
,Comma character
SSULUnderload (not transmitted in the TRANSM mode)
OLOverload (not transmitted in the TRANSM mode)
STStability of the display
USInstability of the display
,Comma character
WWWWWWWWweight
,Comma character
Table 6-14. Secondary Extended String Characters
When thePR.MODE
=
PcPR.HK is set, only the weight is transmitted on the printer port.
6.5Connection to a Remote Display
Use the following steps to connect the SCT-1100 to a remote display.
1.Restart indicator. Momentarily press during startup to display setup mode. tyPE displays.
2.Navigate to s e t u p
3.Press until the instrument displays
instrument reboots.
Co nF iG SerialCoM.PCPCModeALL.STD. Press to confirm selection.
Use the following table to troubleshoot error messages on the instrument.
MessageDescription
AL . Er rDisplays when not connected at start-up, if there are communication problems between the instrument and the board or when the alibi
memory operation is selected; The unit of measure conversion is automatically set, but not saved in the setup mode
bU SyPrinting - PRN serial port is occupied or the instrument is waiting to transmit a print job to a PC
Un S t A bTrying to print with an unstable weight
Un . Ov ErTrying to print with the weight in underload or in overload; with a weight of 9 divisions greater than the capacity or 100 divisions below the
gross zero
The weight is 9 divisions above the maximum capacity
The weight is under the gross zero (- capacity – 9 divisions)
Gr os .E rTrying to print with a negative gross weight (equal or less than 0)
NE t.ErrTrying to print with a negative net weight (equal or less than 0)
Lo WNet weight less than the minimum necessary for the printing or the totalization
no .0 .U n SWeight did not exceed net 0 or was not stable
Co nv.Trying to print while the instrument is converting the unit of measure
no i nSecond attempt to acquire the input weight (input/output mode, set as in .o ut )
no o u tSecond attempt to acquire the output weight (input/output mode, set as in .o u t)
no 1Second attempt to acquire the input weight (input/output mode, set as G. t .1 st.2n d)
no 2Second attempt to acquire the output weight (input/output mode, set as G. t .1 st .2 n d)
Pr ECDisplays when trying to calibrate a point without first having confirmed the number of calibration points
Er Mo tWeight is unstable during the acquisition of a point during calibration
ER Pn tDuring the acquisition of a calibration point a null value has been read by the converter
er-1 1Calibration error – the sample weight used was too small. Use a weight equal to at least half of the scale capacity
Er-1 2
Er -3 7Scale must be calibrated. Perform a technical default (dE F A u ) parameter, before proceeding; SeeTable 4-4 on page 22.
Er-3 9Scale must be calibrated. Perform a technical default (dEF Au ) parameter, before proceeding; SeeTable 4-4 on page 22.
C_ er_- 3 6During calibration some internal negative points have been calculated:
C_ er_- 3 7During the calibration some internal points less than the minimum value have been calculated:
HW -E r rHARDWARE ERROR: software not compatible with the installed hardware; the hardware expansion component which allows the soft-
Calibration error – the acquired calibration point
NOTE: Press to access the setup.
NOTE: Press to access the setup.
• The calibration point is less than the zero point
• the signal is negative (check the connections)
• The calibration point is equal to the zero point
• a capacity too high in relation to the division has been set
ware to function is missing
(tP 1 o t P 2 o tP3 ) is equal to the zero point ()
Power Supply12-24 VDC
Power Consumption70 mA min to 100 mA max
Excitation Voltage5 VDC, 120mA, 8 x 350Ω
Analog Signal Input±39 mV
Range
Analog Signal 0..03 µV/gradation minimum
Sensitivity0.3 µV/gradation recommended
Operator Interface
DisplayLED 6 digits 0.51" (13 mm high)
Keypad5-key, membrane panel, tactile feel
LED6 status indicator red LEDs
Enclosure
CasePlastic console suitable for mounting on DIN
rail or on the wall
NEMA Type 1 plastic pluggable connectors
Dimensions (W x H x D)4.17" x 3.54" x 2.28" (106mm x 90mm x 58mm)
Weight1lb (0.5kg)
Operation
Resolution:
Internal 1.5 million counts
Weight display800,000 minimum