Rice Lake Weighing Systems SCT20 Operating Manual

Page 1
SCT Weight Transmitter
20 Series
Installation & Operator’s Manual
131128 Rev C
Page 2
Page 3
Content
Technical training seminars are available through Rice Lake Weighing Systems. Course descriptions and dates can be viewed at www.ricelake.com/training or obtained by calling 715-234-9171 and asking for the training department.
1.0 Introduction........................................................................................................................................ 1
1.1 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Equipment Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Load Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.2 Load Cell Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.5 Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.6 LED and Key Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.7 Instrument Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.7.1 If The Instrument Has Not Been Calibrated . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.0 Configuration ................................................................................................................................... 10
2.1 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.1.1 Theoretical Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.1.2 Maximum Capacity (NASS ). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.1.3 Zero Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.1.4 Zero Value Manual Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.1.5 Weight (Span) Calibration (With Test Weights) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.1.6 Setting Units of Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.1.7 Display Coefficient. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.2 Filter On The Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.3 Zero Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4 SCT20-DN (DeviceNet) Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.5 Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.6 Profibus Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.7 SCT20-IP (EtherNet/IP) Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.8 SCT20-IP (Ethernet TCP/IP) Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
2.9 Serial Communication Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.9.1 RS-485 Serial Communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.10 Outputs And Inputs Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.11 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.12 Setpoints Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2.13 Reserved For The Installer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
2.13.1 Menu Locking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
2.13.2 Menu Unlocking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
2.13.3 Temporary Menu Unlocking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
2.13.4 Default Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
2.13.5 Program Selection - Reverse: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.13.6 Program Selection - Not Legal: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.13.7 Keypad Or Display Locking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.0 Operation.......................................................................................................................................... 33
3.1 Semi-automatic Tare (Net/Gross) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.2 Preset Tare (Subtractive Tare Device) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.3 Semi-automatic Zero (Weight Zero-setting For Small Variations) . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.4 Peak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
© Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America.
Specifications subject to change without notice.
Rice Lake Weighing Systems is an ISO 9001 registered company.
April 3, 2014
Page 4
Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost. Visit www.ricelake.com/webinars.
3.5 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.6 SCT20-DN (DeviceNet) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.7 Profibus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
3.8 SCT20-IP (EtherNet/IP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
3.9 SCT20-TCP/IP (Ethernet TCP/IP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.10 Modbus/TCP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
3.11 Modbus-RTU Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
3.12 ASCII Bidirectional Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.13 Fast Continuous Transmission Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
3.14 Continuous Transmission Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
3.15 Interface to Remote Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
3.16 Communication Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
SCT Weight Transmitter Limited Warranty ............................................................................................... 62
For More Information ................................................................................................................................ 63
ii SCT Weight Transmitter Operator’s Manual
Page 5
1.0 Introduction
WARNING
Important
WARNING
Manuals can be viewed and downloaded from the Rice Lake Weighing Systems website at
www.ricelake.com
1.1 Safety
Safety Symbol Definitions:
Indicates a potentially hazardous situation that, if not avoided could result in death or serious injury, and includes hazards that are exposed when guards are removed.
Indicates information about procedures that, if not observed, could result in damage to equipment or corruption to and loss of data.
Do not operate or work on this equipment unless you have read and understand the instructions and warnings in this Manual. Contact any Rice Lake Weighing Systems dealer for replacement manuals. Proper care is your responsibility.
.
General Safety
Failure to heed may result in serious injury or death.
Risk of electrical shock. No user serviceable parts. Refer to qualified service personnel for service.
The unit has no power switch, to completely remove D/C power from the unit, disconnect the D/C power cable from the main socket.
DO NOT allow minors (children) or inexperienced persons to operate this unit.
DO NOT operate without all shields and guards in place.
DO NOT use for purposes other then weighing applications.
DO NOT place fingers into slots or possible pinch points.
DO NOT use this product if any of the components are cracked.
DO NOT make alterations or modifications to the unit.
DO NOT remove or obscure warning labels.
DO NOT use near water.
Introduction 1
Page 6
1.2 Equipment Recommendations
Important
Failure to follow the installation recommendations will be considered a misuse of the equipment.
To Avoid Equipment Damage
• Keep away from heat sources and direct sunlight.
• Protect the instrument from rain.
• Do not wash, dip in water or spill liquid on the instrument.
• Do not use solvents to clean the instrument.
• Do not install in areas subject to explosion hazard.
1.2.1 Correct Installation Of Weighing Instruments
• The terminals indicated on the instrument’s wiring diagram to be connected to earth must have the same potential as the scale structure (ground). If you are unable to ensure this condition, connect a ground wire between the instrument and the scale structure.
• The load cell cable must be run separately to the instrument input and not share a conduit with other cables. A shielded connection must be continuous without a splice.
• Use “RC” filters (quench-arcs) on the instrument-driven solenoid valve and remote control switch coils.
• Avoid electrical noise in the instrument panel; if inevitable, use special filters or sheet metal partitions to isolate.
• The panel installer must provide electrical protection for the instruments (fuses, door lock switch, etc.).
• It is advisable to leave equipment always switched on to prevent the formation of condensation.
• Maximum Cable Lengths:
- RS-485: 1000 meters with AWG24, shielded and twisted cables
- RS-232: 15 meters for baud rates up to 19200
1.2.2 Correct Installation Of The Load Cells
Installing Load Cells:
The load cells must be placed on rigid, stable structures within 0.5% of plumb and level. It is important to use mounting modules for load cells to compensate for misalignment of the support surfaces.
Protection Of The Load Cell Cable:
Use water-proof sheaths and joints in order to protect the cables of the load cells.
Mechanical Restraints (pipes, etc.):
When pipes are present, we recommend the use of hoses, flexible couplings and rubber skirted joints. In case of rigid conduit and pipes, place the pipe support or anchor bracket as far as possible from the weighed structure (at a distance at least 40 times the diameter of the pipe).
Connecting Several Load Cells In Parallel:
Connect several load cells in parallel by using, if necessary, a watertight junction box with terminal box. The load cell connection extension cables must be shielded, run individually into their piping or conduit and laid as far as possible from the high voltage cables.
Welding:
Avoid welding with the load cells already installed. If this cannot be avoided, place the welder ground clamp close to the required welding point to prevent sending current through the load cell body.
2 SCT Weight Transmitter Operator’s Manual
Page 7
Windy Conditions - Shocks - Vibrations:
Uses ground plate to
continue ground.
Uses structure to continue ground.
The use of weigh modules is strongly recommended for all load cells to compensate for misalignment of the support surfaces. The system designer must ensure that the scale is protected against lateral shifting and tipping relating to shocks and vibration, windy conditions, seismic conditions and stability of the support structure.
Grounding The Weighed Structure:
By means of a 10ga solid or braided wire or braided grounding strap, connect the load cell upper support plate with the lower support plate, then connect all the lower plates to a single earth ground. Once installed electrostatic charges accumulated are discharged to the ground without going through or damaging the load cells. Failure to implement a proper grounding system might not affect the operation of the weighing system; this, however, does not rule out the possibility that the load cells and connected instrument may become damaged by ESD. It is forbidden to ensure grounding system continuity by using metal parts contained in the weighed structure.(see Figure 1-1.)
Figure 1-1. Installation Recommendations
1.3 Load Cells
1.3.1 Load Cell Input Test (Quick Access)
1. From the weight display, press for three seconds.
2. The display will read NU-CEL. Press .
3. The response signal of the load cell is displayed, expressed in mV with four decimals. Press three times
to exit test mode.
1.3.2 Load Cell Testing
Load Cell Resistance Measurement (Use A Digital Multimeter):
• Disconnect the load cells from the instrument and check that there is no moisture in the load cell junction box caused by condensation or water infiltration. If so, drain the system or replace it if necessary.
• The value between the positive signal wire and the negative signal wire must be equal or similar to the one indicated in the load cell data sheet (output resistance).
• The value between the positive excitation wire and the negative excitation wire must be equal or similar to the one indicated in the load cell data sheet (input resistance).
• The insulation value between the shield and any other load cell wire and between any other load cell wire and the body of the load cell must be higher than 20 Mohm (mega ohms).
Introduction 3
Page 8
Load Cell Voltage Measurement (Use A Digital Multimeter):
Important
ANALOG OUTPUT
(Current and Voltage)
PLC
PLC or FIELD SIGNALS
2 DIGITAL INPUTS
(Optoisolated,
Externally supplied)
JUNCTION BOX
3 RELAY
MAX 8 LOAD CELLS IN
PARALLEL
RS-485 RS-232 Modbus RTU
SERIAL PORT
DC power
supplier
(12-24 Volt )
Other Options
Devicenet
Profibus
Ethernet
OUTPUTS
• Remove weight of scale from load cell to be tested.
• Make sure that the excitation wires of the load cell connected to the instrument is 5 Vdc +/- 3%.
• Measure the millivolt signal between the positive and the negative signal wires by directly connecting them to the multi-meter, and make sure it reads between 0 and 0.5 mV (thousandths of a Volt).
• Apply load to the load cell and make sure that there is a signal increment.
If one of the above conditions is not met, please contact the technical assistance service.
1.4 Specifications
• Weight indicator and transmitter for Omega/DIN rail mounting suitable for back panel; space saving vertical shape. Six-digit semi alphanumeric display (18mm h), 7 segment. Four-key keyboard. Dimensions: 25x115x120 mm.
• Displays the gross weight; with an external contact capable of remote zeroing and gross/net switching.
• Peak weight function.
• Analog Output Option - Transmits the gross or net weight via opto-isolated analog output 16 bit, current 0­20mA, 4-20mA or voltage 0-10V, 0-5V (±0V / ±5V by closing a soldered in jumper).
DeviceNet Option – It has a DeviceNet DP V0 Slave port that allows to exchange the main system parameters
with a DeviceNet master.
Profibus Option – It has a Profibus DP V0 Slave port that allows to exchange the main system parameters
with a Profibus master.
Ethernet Option – It has an Ethernet/IP device port that allows to exchange the main system parameters with
an Ethernet/IP scanner.
Modbus/TCP Option – It has a Modbus/TCP device port that allows to exchange the main system parameters
with a Modbus/TCP master.
• Transmits the gross or net weight via RS-485 serial port, by means of protocols:
- Modbus RTU
- ASCII bidirectional protocol
- Continuous transmission
4 SCT Weight Transmitter Operator’s Manual
Page 9
Power Supply and Consumption (VDC) 12 - 24 VDC +/- 10% ; 5 W
No. of Load Cells in Parallel And Supply max 8 ( 350 ohm ) ; 5VDC/120mA
Linearity / Analog Output Linearity < 0.01% F.S. / < 0.01% F.S.
Thermal Drift < 0.0005 % F.S. /°C ;
< 0.003 % F.S./°C (Analog Only)
A/D Converter 24 bit (16.000.000 points)
Max Divisions (With Measurement Range: +/-10mv = Sens. 2mv/v) +/- 999999
Measurement Range +/- 39 mV
Max Sensitivity of Usable Load Cells +/-7mV/V
Max Conversions Per Second 300 conversions/second
Display Range - 999999 ; + 999999
No. of Decimals / Display Increments 0 - 4 / x 1 x 2 x 5 x 10 x 20 x 50 x 100
Digital Filter / Readings Per Second 0.060–7sec/5-300Hz (Analog,DeviceNet)
0.012-7sec/5-300Hz (Ethernet,Profibus)
Relay Logic Outputs N.3 - max 115 VAC ; 150mA
Logic Inputs N.2 - optoisolated 5 - 24 VDC PNP
Serial Ports RS-485 (RS-232)
Baud Rate 2400, 4800, 9600, 19200, 38400, 115200
Humidity (Non Condensing) 85 %
Storage Temperature -22° to 176°F (- 30° to + 80°C)
Working Temperature -4° to 140°F (- 20° to + 60°C)
Optoisolated Analog Output 16 Bit - 65535 Divisions 0-20 mA; 4-20 mA (max 300 ohm); 0-10 VDC; 0-5 VDC;
+/- 10 VDC; +/- 5 VDC (min 10 kohm)
Ethernet/IP Port RJ45 10Base-T or 100Base-TX (auto-sensing)
Profibus Port: Baud Rate Profibus Port Addresses
DeviceNet Port: Baud Rate DeviceNet Port: Addresses
Modbus/TCP Port RJ45 10Base-T or 100Base-TX (auto-sensing)
to 12 Mbit/s 1-99
125kbit/s, 250kbit/s, 500kbit/s
1.63
File Number:E151461
Table 1-1. Specifications
Introduction 5
Page 10
1.5 Electrical Connections
+ EXCITATION
- EXCITATION
4-wire load cell
connection
SHIELD
+ SIGNAL
- SIGNAL
1
2
345
6
RS485
Relay
Outputs
115Vac / 150mA
12-24Vdc
- SIGNAL
+ EXCITATION
SHIELD
- EXCITATION
- SENSE
+ SENSE
Inputs
5-24Vdc
+ SIGNAL
+
-
Current output: max load 300 ohm Voltage output: min. load 10 kohm
6-wire load cell
connection
Analog Output shown See below for Devicenet, Profibus Ethernet/IP or Modbus/TCP
RS485 Termination
POWER
EX
-
REF
-
1
EX +
2
REF
+ 3
SIG
+6IN17IN28IN COM9mA10V11mA-V COM
12
4
SIG
­5
OUT1
OUT2
OUT3
OUT COM
+ RS485
- RS485
+ 12-24
0 VDC
131415
181920
16
17
SCT
20-AN
DeviceNet
CAN V­10
11
121314
CAN L
CAN H
CAN V+
EX-REF
-
1
EX +
2
REF +
3
SIG +6IN17IN28IN COM
9
4
SIG
­5
Profibus DP
EX-REF
-
1
EX +
2
REF +
3
SIG
+6IN17IN28IN COM
9
4
SIG
­5
Ethernet/IP
EX-REF
-
1
EX +
2
REF +
3
SIG +6IN17IN28IN COM
9
4
SIG
­5
EtherNet/IP RJ-45
Profibus DP Socket DB-9
Modbus/TCP
EX-REF
-
1
EX +
2
REF +
3
SIG +6IN17IN28IN COM
9
4
SIG
­5
Modbus/TCP RJ-45
Basic Information
• It is recommended that the negative side of the power supply be grounded.
• It is possible to power up to eight 350 ohm load cells or sixteen 700 ohm load cells.
• For 4-wire load cells, jumper between EX- and REF- and between EX+ and REF+.
• Connect terminal “0 VDC” to the RS-485 common when interfacing to A/C powered equipment so that there is an opto-isolated RS-485 connection.
• In case of an RS-485 network with several devices it is recommended to activate the 120 ohm termination resistance on the two devices located at the ends of the network, see Section 2.9.1 “RS-485 Serial Communication” on page 26.
Profibus Termination
Figure 1-2. Wiring Diagram
CAN Termination
CAN SHLD
3 outputs: configurable setpoints or remote output management via protocol. 2 inputs (Default: SEMI-AUTOMATIC ZERO input 1; NET/GROSS input 2) (see Section 2.10 “Outputs And
Inputs Configuration” on page 27).
6 SCT Weight Transmitter Operator’s Manual
Page 11
1.6 LED and Key Functions
0
LED Functions
LED Main function Secondary function *
NET Net weight LED: net weight display (semi-automatic tare or preset tare) no meaning
Zero LED (deviation from zero not more than +/- 0.25 divisions) LED lit: output 3 closed
Stability LED LED lit: output 2 closed
kg Unit of measure: kg LED lit: output 1 closed
g Unit of measure: g LED lit: input 2 closed
L Unit of measure: lb LED lit: input 1 closed
* To activate the secondary LED function, press and hold , then press .
DeviceNet Models
DeviceNet Status Blinks quickly: DeviceNet OK
Blinks slowly (about 0,5s): Initialization LED lighted: Slave not initialized LED off: DeviceNet KO
Profibus Models Only
Profibus Status Blinks quickly: Profibus OK
Blinks slowly (about 0.5s): Profibus error
Ethernet Models Only
Ethernet/IP Status Blinks quickly: Ethernet OK
Blinks slowly (about 0.5s): Ethernet/IP KO
RJ45 Connector Left side [Link LED]
RJ45 Connector Right side [Activity LED]
Modbus/TCP Status Blinks quickly: Modbus/TCP OK
RJ45 Connector Left side [Link LED]
RJ45 Connector Right side [Activity LED]
Off: no link Amber: 10 Mbps Green: 100 Mbps
Off: no activity Amber: half duplex Green: full duplex
Modbus/TCP Models Only
Off: no link Amber: 10 Mbps Green: 100 Mbps
Off: no activity Amber: half duplex Green: full duplex
Introduction 7
Page 12
Key Functions
Note
Note
KEY Short press
Escape
Captures Tare
Scroll/
Backspace
Gross Net
Long press
(3 sec)
Zero Setting Escape from a parameter or return to previous menu or operation mode.
Removes Tare Net Gross
Move to the previous parameter in a level or scroll to the next digit in a parameter value.
Into menus
Next/ Data
Entry
Enter
+
+
Save to alibi memory (if present)
Setting setpoints and hysteresis
Enter configuration for setting general parameters (press and hold then press to enter set-up menu.
Setting preset tare (press and hold then press to enter set-up menu.
The LEDs light up in sequence to indicate that a setting and not a weight is being viewed.
After pressing the first digit will flash and can be edited. All LEDs will be flashing, the value is a displayed weight.
For Numeric Entries:
Press to select desired digit.
s to increment digit.
Pres
mV load cell test
Move to the next parameter in a level or increment a value in a parameter.
Move to next level of configuration or select and edit a parameter.
8 SCT Weight Transmitter Operator’s Manual
Page 13
1.7 Instrument Commissioning
Note
1. Plug power cord in to outlet to turn on indicator, the display shows in sequence:
- SU followed by the software code (e.g.: SU S );
- r followed by the software version (e.g.: r 1.04.01 );
- HU followed by the hardware code (e.g.: HU 104 );
- The serial number (e.g.:1005 15 );
2. Check that the display shows the weight and that when loading the load cells there is an increase in weight.
3. If there is not, check and verify the connections and correct positioning of the load cells.
If instrument has NOT been calibrated, complete Section 2.1 on page 12 before proceeding to next step
4.Reset to zero. See Section 2.1.3 on page 14.
5. Check the calibration with test weights and correct the indicated weight if necessary. See Section 2.1.5 on
page 15.
6. Set the desired analog output type and the full scale value.
- Analog Output Model see Section 2.5 on page 19.
- DeviceNet Model see Section 2.4 on page 18.
- Ethernet Model see Section 2.7 on page 21.
- Profibus Model see Section 2.6 on page 20.
- Modbus/TCP see Section 3.10 on page 46.
7. If you use serial communication, set the related parameters (see Section 2.9 on page 24).
8. If setpoints are used, set the required weight values and the relevant parameters (see Section 2.12 on page
30 and Section 2.10 on page 27).
1.7.1 If The Instrument Has Not Been Calibrated
Missing plant system identification tag, proceed with calibration:
1. If load cells data are unknown, follow the procedure in Section 2.1.5 on page 15.
2. Enter the rated data of load cells following the procedure given in Section 2.1.1 on page 13.
Introduction 9
Page 14
2.0 Configuration
CALIb FILtEr PArA 0
dEUnEt
0 SEt AuTO 0 TrAC 0
bAUdAddr
tEStOut-In
000000
2 3 4 5
1
0
8 9
7
6
Enter #
6250H
6125H
6500H
In
Out
NU CEL
Enter #
ANALOG
DeviceNet
Models
P-tArE
SEtP1
SErIAL
PrOFI
Addr
Enter #
Profibus
Models
Analog
Models
EtHnEt
1 PAddrSUAP
YES
nO
Ethernet
Models
SUbnET
GAtUAY
C
b
A
d
C
b
A
d
C
b
A
d
See
Figure 2-2
See
Figure 2-3
See
Figure 2-7
Parameter Choices Description
CALIb FS-tEO
SEnS I b dI UI S NASS ZErO
Figure 2-1. Scale Menu Structure
Calibration – see Section 2.1 on page 12.
1 NP 0 WEIGHt
FILtEr 0-9
unIt COEFF
4 *
PArA 0 0 SEt
AutO 0 trAC 0
dEuNEt Addr
Filter on weight – see Section 2.2 on page 17.
Zero Parameters – see Section 2.3 on page 18.
DeviceNet Options for outputs and inputs configuration – see Section 2.10 on page 27.
bAud
10 SCT Weight Transmitter Operator’s Manual
Page 15
Parameter Choices Description
ANALOG tYPE
Analog Options for outputs and inputs configuration – see Section 2.10 on page 27.
NOdE ANA 0 ANA FS COr 0
COr FS PrOFI Addr EtHnEt SUAP
Profibus settings – see Section 2.6 on page 20.
SCT20-IP (EtherNet/IP) settings – see Section 2.7 on page 21.
1 PAddr
SUbnEt
GAtUAY SErIAL rS-485
Serial Communications settings – see Section 2.9 on page 24.
bAud
Addr
HErtZ
dELAY
PArItY
StOP Out-In Out 1
Outputs and Inputs configuration – See Section 2.10 on page 27.
Out 2
Out 3
In 1
In 2 tESt In
Test – see Section 2.11 on page 29.
Out
NU-CEL
*
- indicates default value.
Configuration 11
Page 16
2.1 Calibration
dI UI S
WEIGHt
unIt
COEFF
LItrE
bAr AtN
PI ECE
nEUton
lb
OtHEr
HI LOG
HI LO-N
nEU-N
G t
FS-tEO
Enter #
SEnS Lb
Enter #
NASS
Enter #
ZErO
Enter #
Enter # Enter #
I NP 0
Enter #
50
100
0.0001
0.0002
20
10
0.002
0.005
0.001
0.0005
2 5
0.02
0.05
0.01
0.2
0.5
0.1
1
FILtEr
PArA 0 SErIAL
tESt
Out-In
ANA LOG
000000
SEtP1
CALIb
PtArE
Note
Figure 2-2. Calibration Menu Structure
Parameter Choices Description
FS-tEO
Enter #
0 =dENO *
System Full Scale is determined by multiplying one load cell capacity by the number of load cells
used. Example of system full scale value calculation:
4 cells of 1000kg ---->
FULL SCALE = 1000 X 4 = 4000
The instrument is supplied with a theoretical full scale value deno corresponding to 10000. To restore factory values, set 0 as full scale.
SEnSI b
Enter #
0.50000
to
7.00000
Sensitivity is a load cell rated parameter expressed in mV/V. Set the average sensitivity value indicated on the load cells. Example of 4-cell system with sensitivity
2.00100, 2.00150, 2.00200, 2.00250; enter 2.00175,
calculated as (2.00100 + 2.00150 + 2.00200 + 2.00250) / 4.
2.00000 *
dI UI S 1
2 5
*
Division (resolution) - the weight increment (display division size) that the scale counts by. Selections are: 0.0001 and 100 with x1 x2 x5 x10 increments.
10
50
20
100
0.0001
12 SCT Weight Transmitter Operator’s Manual
0.0002
0.0005
0.001
0.002
0.005
0.01
0.02
0.05
0.1
0.2
0.5
Page 17
Parameter Choices Description
Note
Note
NASS
ZErO
I NP 0
Enter #
0 *
to max full scale
0
Enter #
0 to 999999
Maximum Capacity (Live Load/Product) that can be displayed. When the weight exceeds this value by 9 divisions, the display will go to dashes, indicating overload. Setting this value to 0 will disable the over capacity function.
Used to capture the deadload of the scale system. With the scale empty, the displayed value can be zeroed off. This menu may also be accessed directly from the weighing mode to compensate for zero changes or variations.
Press to display the accumulated deadload weight that has been zeroed off.
Estimated dead load value of the scale when a scale contains product that cannot be removed. The value entered is the dead load. This value will be replaced if the zero function is performed later.
0 *
UEIGHt
unIt G
Enter #
0 *
Weight (Span) Calibration - after the Theoretical Calibration has been completed and zero is set, the calibration can be adjusted with actual test weights by changing the displayed value in this parameter.
If changes are made to the theoretical Full Scale ( or Divisions ( the Theoretical Calibration is initiated and applied.
If the theoretical full scale ( (Span) Calibration ( different, the calibration in use is the Weight (Span) Calibration based on calibration weights.
If changes are made to the theoretical full scale ( divisions ( set to default values (setpoints, hysteresis, etc.).
Unit of Measure - select to determine what unit of measure is displayed and printed.
dI UI S
) parameters, all the system’s parameters containing a weight value will be
dI UI S
) parameters, the Weight (Span) Calibration is cancelled and
FS-tEO
WEIGHt
), then the calibration currently in use is theoretical; if they are
) and the capacity full scale (
FS-tEO
t lb
See Section 2.1.6 on page 16 for description of units.
nEUton lI trE bAr AtN PI ECE nEU-N HI lO-N OtHEr HI LOG *
COEFF
Enter #
0-99.9999
Multiplier Value entered will display an alternative unit of measure if the digital input is set for COEFF and is in a closed state.
0 *
*
- indicates default value.
To calibrate the instrument, the Theoretical calibration, Section 2.1.1 on page 13, must be completed first. After Theoretical calibration is set, the scale can be set with actual weights (see Section 2.1.5 on page 15).
FS-tEO
), the Sensitivity (
NASS
) are equal in Weight
), the capacity full scale (
NASS
SEnSIb
) or
)
2.1.1 Theoretical Calibration
This function allows load cell rated values to be set. To perform the theoretical calibration set FS-tEO, SEnS lb and dI UI S (refer to Figure 2-2) in sequence:
1. Press and hold , then press to enter set-up menu, CALIb will be displayed.
2. Press , FS-tEO is displayed. Press again.
3. Press or until total load cell capacity (system full scale) is displayed, press .
4. Press or until SEnS lb is displayed, press .
5. Press or until desired load cell mV/V is displayed, press .
6. Press or until dI UI S is displayed, press .
Configuration 13
Page 18
7. Press or until desired display division size is displayed, press .
Note
Note
Note
Important
8. This completes the Theoretical Calibration, press twice to exit set-up menu or continue to Section 2.1.2.
By modifying the theoretical full scale, the sensitivity or divisions, the Weight (Span) Calibration is cancelled and the Theoretical Calibration only is considered valid.
If the theoretical full scale and the recalculated full scale in Weight (Span) calibration (see Section 2.1.5 on page 15) are equal, this means that the calibration currently in use is Theoretical Calibration; if they are different, the calibration in use is the Weight (Span) Calibration based on test weights.
By modifying the theoretical full scale, the sensitivity or divisions and all the system’s parameters containing a weight value will be set to default values (setpoints, hysteresis, etc.).
2.1.2 Maximum Capacity (NASS )
Maximum capacity (live load/product) that can be displayed. When the weight exceeds this value by 9 divisions
------
the following is displayed ‘
1. Press and hold , then press to enter set-up menu, CALIb will be displayed.
2. Press , FS-tEO is displayed.
3. Press or until NASS is displayed, press .
4. Press or until desired capacity is displayed, press .
5. Press twice to exit set-up menu.
’, indicating overload. To disable this function, set to 0.
2.1.3 Zero Setting
Perform this procedure after having set the Theoretical calibration, see Section 2.1.1 on page 13.
This menu may also be accessed directly from the weight display, press and hold for 3 seconds.
1. Press and hold , then press to enter set-up menu, CALIb will be displayed.
2. Press , FS-tEO is displayed.
3. Press or until ZErO is displayed, press .
4. The weight value to be set to zero is displayed. In this phase all of the LEDs are flashing. Press , the weight is set to zero (the value is stored to the permanent memory).
5. Press twice to exit set-up menu.
Press to display the accumulated deadload that has been zeroed off by the instrument, displaying the sum of all of the previous zero settings.
2.1.4 Zero Value Manual Entry
Perform this procedure only if it is not possible to zero off the scale structure, for example because it contains product that can not be unloaded.
Enter the estimated structured dead load value that would be zeroed.
1. Press and hold , then press to enter set-up menu, CALIb will be displayed.
2. Press , FS-tEO is displayed.
3. Press or until INP 0 is displayed, press .
4. Press or until desired dead load value is displayed, press .
5. Press twice to exit set-up menu.
14 SCT Weight Transmitter Operator’s Manual
Page 19
2.1.5 Weight (Span) Calibration (With Test Weights)
Important
Note
After performing the Theoretical calibration (Section 2.1.1 on page 13) and the Zero setting (Section 2.1.3 on page
14), this function allows calibration to be done using test weights of known value. If adjustment is required, change the displayed value to display the test weight value.
1. Load the test weight onto the scale, use as high a percentage of the maximum quantity to be weighed as possible.
2. Press and hold , then press to enter set-up menu, CALIb will be displayed.
3. Press , FS-tEO is displayed.
4. Press or until UEIGHt is displayed, press .
5. The value of the weight currently on the system will be flashing on the display. All of the LEDs are off. (If adjustment is not required, skip to step 8.)
6. Adjust the value on display to match weight loaded on the scale if necessary, by pressing or . The LEDs will begin scrolling.
7. Press , the new set weight will appear with all the LEDs flashing.
8. Press , UEIGHt will be displayed again.
9. Press twice to exit set-up menu.
Example: For a system of maximum capacity of 1000 kg and 1 kg division, two test weights are available, one 500 kg and one 300 kg.
Load both weights onto the system and correct the indicated weight to 800. Now remove the 300 kg weight, the system must show 500; remove the 500 kg weight, too; the system must read zero. If this does not happen, it means that there is a mechanical problem affecting the system linearity.
Identify and correct any mechanical problems before repeating the procedure.
If theoretical full scale and recalculated full scale in Weight (Span) Calibration are equal, it means that the Theoretical Calibration is currently in use; otherwise, the Weight (Span) Calibration based on test weights is in use.
If the correction made changes the previous full scale for more than 20%, all the parameters with settable weight values are reset to default values.
Linearization Option On Max 5 Points:
It is possible to perform a linearization of the weight repeating the above described procedure up to a maximum of
five points, using five different test weights. The procedure ends by pressing or after entering the fifth value; at this point it will no longer be possible to change the calibration value, but only to perform a new Weight
(Span) Calibration. To perform a new calibration, return to the weight display and then re-enter the calibration menu.
By pressing after having confirmed the test weight that has been set, the full scale appears, recalculated according to the value of the maximum test weight entered and making reference to the cell sensitivity set in the theoretical calibration (SEnSI b).
Configuration 15
Page 20
2.1.6 Setting Units of Measure
Note
Important
1. Press and hold , then press to enter set-up menu, CALIb will be displayed.
2. Press , FS-tEO is displayed.
3. Press or until unIt is displayed, press .
4. Press or until desired unit is displayed, press .
5. Press twice to exit set-up menu.
HI LOG G t Lb nEUton LI tre bAr
kilograms grams tons pounds* newton* litres* bar*
LI tre bAr AtN PI ECE nEU-N HI LO-N OtHEr
litres* bar* atmospheres* pieces* newton metres* kikgram metres* units of measure not included*
* Indicates it is possible to set the display coefficient. To use COEFF it is necessary to enable it, closing the COEFF
input. See Section 2.1.7 on page 16.
If the print function is enabled, the symbol of the selected unit of measure will be printed after the measured value.
2.1.7 Display Coefficient
By setting the coefficient the display is changed accordingly. If one of the inputs is set to COEFF mode (see Section 2.10 on page 27) when the input is closed the value will be
displayed modified according to the coefficient; when the input is opened the standard weight display will be restored.
1. Press and hold , then press to enter set-up menu, CALIb will be displayed.
2. Press , FS-tEO is displayed.
3. Press or until COEFF is displayed, press .
4. Press or until desired number is displayed, press .
5. Press twice to exit set-up menu.
HI LOG G t Lb nEUton
LI tre
bAr AtN PI ECE nEU-N HI LO-N
OtHEr
16 SCT Weight Transmitter Operator’s Manual
kilograms
grams
tons
pounds Value set in COEFF will be multiplied by the weight value currently displayed
newton Value set in COEFF will be multiplied by the weight value currently displayed
litres in COEFF set the specific weight in kg/l, assuming that the system is
bar Value set in COEFF will be multiplied by the weight value currently displayed
atmospheres Value set in COEFF will be multiplied by the weight value currently displayed
pieces in COEFF set the weight of one piece
newton metres Value set in COEFF will be multiplied by the weight value currently displayed
kikgram metres Value set in COEFF will be multiplied by the weight value currently displayed
other generic units of measure not included in list
All other settings (setpoints, hysteresis, calibration ...) are expressed in weight value. If you want to convert them to the new unit of measurement, perform one of the following procedures for changing the system calibration.
The parameter must remain set to 1.0000.
calibrated in kg
Value set in COEFF will be multiplied by the weight value currently displayed
Page 21
Theoretical Calibration For Other Units Of Measure
Note
Set in the parameter the F.SCALE value divided by the conversion coefficient from kg to the new unit of measure.
Example: The 4 load cells of 1000 kg are placed under a scale for olive oil, which has a specific gravity of 0.916 kg / l. Setting the F.SCALE = (4x1000) / 0916 = 4367, the system works in liters of olive oil. If you set the unit to litres, the system will display and print the symbol ‘l’ instead of ‘kg’. See Section 2.1.6 on page 16.
Weight (Span) Calibration For Other Units Of Measure
Load a known quantity of product litres on the scale (use as high a percentage of the maximum quantity to be weighed as possible) and enter in the parameter UEIGHt, the product loaded value in litres. If you set the units to litres, the system will display and print the symbol ‘l’ instead of ‘kg’. See Section 2.1.6 on page 16.
2.2 Filter On The Weight
The filtering selection is used to eliminate environment noise, and is typically a compromise between responsiveness and stability. The lower the number, the more responsive the display will be to weight changes. The filter is used to stabilize a weight as long as the variations are smaller than the corresponding “Response Time”. The filter setting is dependent on the type of application and the required update rate.
Setting the parameter allows a stable weight display to be obtained. To increase the effect (weight more stable), increase the value.
1. Press and hold , then press to enter set-up menu, CALIb will be displayed.
2. Press or until FILtEr is displayed, press . The currently programmed filter value is displayed.
3. Press or until desired filter value is displayed, press .
4. The current weight is displayed (all LED’s flashing) and the displayed stability can be experimentally
verified. Press .
5. If stability is not satisfactory, press , this returns indicator to FILtEr option and the filter may be modified again until an optimum result is achieved.
6. Press to exit set-up menu.
The filter enables to stabilize a weight as long as its variations are smaller than the corresponding “Response Time”. It is necessary to set this filter according to the type of application and to the full scale value set.
Response times
Filter Value
012 300
1150 100
2260 50
3425 25
4* 850 12.5
5 1700 12.5
6 2500 12.5
7 4000 10
8 6000 10
9 7000 5
* indicates default setting
[ms]
Display and serial port refresh frequency
[Hz]
Configuration 17
Page 22
2.3 Zero Parameters
1. Press and hold , then press to enter set-up menu, CALIb will be displayed.
2. Press or until  is displayed, press .
3. Press or until desired parameter (see Table 5 ) is displayed, press . The currently programmed value is displayed.
4. Press or until desired value is displayed, press .
5. Press twice to exit set-up menu.
Parameter Choices Description
0 SEt
Enter # 0-max full scale
300 *
Considered decimals: 300 – 30.0 – 3.00 –
0.300
AutO 0
Enter # 0 - max 20% of full scale
0 *
trAC 0 nOnE *
1-5
*
- indicates default value.
Maximum zero range
Indicates the maximum weight value that can be zeroed off by external contact, keypad or serial protocol
Automatically zeroes the scale at power-on
If the weight value is lower than the value set in this parameter, the scale will zero itself provided the weight does not exceed the value in the 0 SET parameter. To disable this function set to 0.
Zero tracking
Automatically zeroes the scale when within the range specified, as long as the weight is within the 0 SET parameter and the scale is at standstill for at least one second. To disable this function, set to none
S
Example: if the parameter dI UI will be automatically set to zero for variations smaller than or equal to 10 (dI UI S x trAC 0 ).
is set to 5 and trAC 0 is set to 2, the weight
2.4 SCT20-DN (DeviceNet) Settings
1. Press and hold and to enter set-up menu. CALIb will be displayed.
2. Press or until dEUnEt is displayed, press .
DeviceNet
3. Press or until desired parameter is displayed, press .
4. Press or until desired value is displayed, press .
5. Press twice to exit set-up menu.
Parameter Choices Description
Addr 1-63
1 *
bAud 6125H
6250H 6500H *
*
- indicates default value.
18 SCT Weight Transmitter Operator’s Manual
Set instrument address in the DeviceNet network.
Set instrument address in the DeviceNet network.
Page 23
2.5 Analog Output
),/W(U
3$U$ 6(U,$/
W(6W
2XW,Q
$1$/2*

3W$U(
6(W3
&$/,E
P2G(
W<3(
Enter #
X
P$
P$
$Q$ $Q$)6 &2U)6&2U
Enter #
Enter #
Enter #Enter #
X

X
*URVV
1HW
Note
Figure 2-4. Soldered Jumper
Figure 2-3. Analog Menu Structure
Parameter Choices Description
 4-20 mA *
0-20 mA 0-10 V 0-5 V
-10 +10 V
-5 +5 V

Enter #
 


Enter # Set the weight value for the minimum analog output value.
Enter # Set the weight value for the maximum analog output value; it must correspond to the value set in the


* Indicates default value.
Selects the analog output type.
See “Soldered Jumper” on page 19 See “Soldered Jumper” on page 19
Select mode to be tracked, gross or net. If the net function is not active, the analog output varies according to gross weight.
Only set a value different from zero to limit the analog output range.
For instance: for a full scale value of 10000 kg, a 4 mA signal at 5000 kg is required, and 20 mA at 10000 kg, in this case, instead of zero, set 5000 kg.
PLC program (default: calibration full scale).
E.g.: if using a 4-20 mA output and in the PLC program a 20 mA = 8000 kg is desired, set the parameter to 8000.
Analog output correction to zero: if necessary adjust the analog output, allowing the PLC to indicate 0. The sign ‘-‘ can be set for the last digit on the left.
E.g.: For a 4-20 mA output and a minimum analog setting, the PLC or tester reads 4.1 mA. Set the parameter to 3.9 to obtain 4.0 on the PLC or tester. (See “Analog Output Type Scale Corrections” on page 20)
Full scale analog output correction: if necessary adjust the analog output, allowing the PLC to indicate the value set in the parameter.
E.g. For a 4-20 mA output with the analog set to full scale and the PLC or tester reads 19.9 mA, set the parameter to 20.1 to obtain 20.0 on the PLC or tester. (See “Analog Output Type Scale Corrections” on page 20)
Soldered Jumper
For the output -10 +10 V and -5 +5 V the soldered jumper (SW4) must be closed:
Release the locking tabs, both sides, to open enclosure.
Locate the soldered jumper (SW4), on the circuit board. See Figure 2-4.
Close the jumper shorting the pads with a solder bridge.
Configuration 19
Page 24
Analog Output Type Scale Corrections
Note
Minimum and maximum values which can be set for the zero and full scale corrections Refer to COr 0 and COr FS.
Analog Output Type Minimum Maximum
0–10 V -0.15 10.2
0–5 V -0.15 5.5
-10 +10 V -10.3 10.2
-5 +5 V -5.5 5.5
0-20 mA -0.2 22
4-20 mA -0.2 22
The analog output may also be used in the opposite manner, i.e. the weight setting that corresponds to the analog zero may be greater than the weight set for the analog full scale. The analog output will increase towards full scale as the weight decreases; the analog output will decrease as the weight increases.
E.g.: analog output type having selected 0-10V ANA 0 = 10000 ANA FS = 0 Weight = 0 kg analog output = 10 V Weight = 5000 kg analog output = 5 V Weight = 10000 kg analog output = 0 V
2.6 Profibus Settings
Profibus
1. Press and hold and to enter set-up menu. CALIb will be displayed.
2. Press or until PrOFI is displayed, press .
3. Addr is displayed, press .
4. Press or until desired value is displayed, press .
5. Press twice to exit set-up menu.
Parameter Choices Description
Addr 1-99
1 *
*
- indicates default value.
Set instrument address in the Profibus network.
20 SCT Weight Transmitter Operator’s Manual
Page 25
2.7 SCT20-IP (EtherNet/IP) Settings
FILtEr
PArA 0 SErIAL
tESt
Out-In
EtHnEt
000000
P-tArE
SEtP1
CALIb
SUbnEt
1 PAddr
GAtuAY
NO
YES
SUAP
1. Press and hold and to enter set-up menu. CALIb will be displayed.
2. Press or until EtHnEt is displayed, press .
3. Press or until desired parameter is displayed, press .
4. Press or until desired value is displayed, press .
5. Press twice to exit set-up menu.
Figure 2-5. SCT20-IP Menu Structure
Parameter Choices Description
SUAP
nO * Little Endian YES Big Endian
1 PAddr A
b C d
192.8.0.141 *
SUbnEt A
b C d
255.255.255.0 *
GAtUAY A
b C d
192.8.0.111 *
*
- indicates default value.
Allows selection of reading/writing of bytes.
Set IP address in the Ethernet/IP network.
Set Subnet mask
Set the Gateway address
Ethernet
Configuration 21
Page 26
2.8 SCT20-IP (Ethernet TCP/IP) Settings
FILtEr
PArA 0 SErIAL
tESt
Out-In
EtHnEt
000000
P-tArE
SEtP1
CALIb
SUbnEt
1 PAddr
GAtuAY NodE
NOdbUS
UEb5ru
nOnE
HdrI P
rIP
ASCII
HdrI Pn
AddrbAud HErt2 dELAY PArItY StOP
Enter #
9600 19200 38400
115200
4800
2400
30 40 50 60
20
10
100 200 300
80
70
Enter #
EUEn
Odd
nOnE
2 3 4 5
1
0
8 9
7
6
Figure 2-6. SCT20-TCP/IP Menu Structure
Parameter Choices Description
1 PAddr ** A
SUbnEt ** A
GAtUAY ** A
NodE
(Comm Port)
b C d
192.8.0.141 *
b C d
255.255.255.0 *
b C d
192.8.0.111 * nOnE * UEbSru NOdbuS ASCII
COntIn
rI P
Set IP address in the Ethernet/IP network.
Set Subnet mask.
Set the Gateway address.
Disables any type of communication.
Web Server
MODBUS-RTU protocol; possible addresses: from 1 to 99 (see Section 3.11)
ASCII bidirectional protocol; possible addresses: from 1 to 99 (see Section 3.12)
NOdU6-
NOd td
Continuous weight transmission protocol (see Section 3.13), at the frequency set in HERTZ parameter (from 10 to 300). NOd t(set: PARITY=none, STOP=1)
NOd td(set: PARITY=none, STOP=1)
Continuous weight transmission protocol, streams net and gross (see Section 3.14) (set: BAUD=9600,PARITY=none, STOP=1)
22 SCT Weight Transmitter Operator’s Manual
Page 27
Parameter Choices Description
Note
NodE
(Continued)
Hdr1 P
Hdr1 PN
bAud 2400
Continuous weight transmission protocol, streams net and gross including decimal. (see Section 3.14) (set: BAUD=9600,PARITY=none, STOP=1)
Continuous weight transmission protocol (see Section 3.14) (set: BAUD=9600,PARITY=none, STOP=1)
When the remote display is set to gross weight:
- if the instrument displays the gross weight, the remote display shows the gross weight.
- if the instrument shows the net weight the remote display shows the net weight alternated with the message “net”
Transmission speed.
4800 9600 *
19200
The baud rate must be the same as the baud rate setting in the Serial Communications Settings, see Section 2.9.
38400 115200
ADDR 1-99
Instruments address
1 *
HERTZ
0Hz *
1
20Hz
Maximum Transmission Frequency To be set when the CONTIN transmission protocol is selected. (see Figure 2-7)
Max setting with min 2400 baud rate
30Hz 40Hz
Max setting with min 4800 baud rate
50Hz 60Hz 70Hz
Max setting with min 9600 baud rate
Max setting with min 19200 baud rate
Max setting with min 38400 baud rate
Max setting with min 38400 baud rate
Delay in milliseconds which elapses before the instrument replies
dELAY
80Hz 100Hz 200Hz 300Hz
0-200 msec
0 *
PAritY nOnE *
EUEn Odd
StOP
1 *
parity none even parity odd parity
Stop bit
2
*
- indicates default value.
** - These settings do not function with Firmware 1.02.00 or lower.
Front Panel Configuration
1. Press and hold and to enter set-up menu. CALIb will be displayed.
2. Press or until EtHnEt is displayed, press .
3. Press or until desired parameter is displayed, press .
4. Press or until desired value is displayed, press .
5. Press twice to exit set-up menu.
Configuration 23
Page 28
2.9 Serial Communication Settings
rS485
nOnE
NodbUS
ASCII
Cont In
rI P
HdrI P
HdrI Pn
CALIb FILtEr PArA 0 SErIAL tEStOut-InANA LOG
AddrbAud HErt2 dELAY PArItY StOP
Enter #
9600 19200 38400
115200
4800
2400
30 40 50 60
20
10
100 200 300
80
70
Enter #
EUEn
Odd
nOnE
2 3 4 5
1
0
8 9
7
6
Figure 2-7. Serial Communications Menu Structure
Parameter Choices Description
rS-485
(Communication Port)
bAud 2400
24 SCT Weight Transmitter Operator’s Manual
ADDR 1-99
nOnE * NOdbuS
ASCII
COntIn
rI P
Hdr1 P
Hdr1 PN
4800 9600 *
19200
38400 115200
1 *
Disables any type of communication.
MODBUS-RTU protocol; possible addresses: from 1 to 99 (see Section
3.11)
ASCII bidirectional protocol; possible addresses: from 1 to 99 (see
Section 3.12)
NOdU6-
NOd td
Continuous weight transmission protocol (see Section 3.13), at the frequency set in HERTZ parameter (from 10 to 300). NOd t(set: PARITY=none, STOP=1)
NOd td(set: PARITY=none, STOP=1)
Continuous weight transmission protocol, streams net and gross (see Section 3.14) (set: BAUD=9600,PARITY=none, STOP=1)
Continuous weight transmission protocol, streams net and gross including decimal. (see Section 3.14) (set: BAUD=9600,PARITY=none, STOP=1)
Continuous weight transmission protocol (see Section 3.14) (set: BAUD=9600,PARITY=none, STOP=1)
When the remote display is set to gross weight:
- if the instrument displays the gross weight, the remote display shows the gross weight.
- if the instrument shows the net weight the remote display shows the net weight alternated with the message “net”
Transmission speed.
Instruments address
Page 29
Parameter Choices Description
HERTZ
0Hz *
1
20Hz
Maximum Transmission Frequency To be set when the CONTIN transmission protocol is selected.
Max setting with min 2400 baud rate
30Hz 40Hz
Max setting with min 4800 baud rate
50Hz 60Hz 70Hz
Max setting with min 9600 baud rate
Max setting with min 19200 baud rate
Max setting with min 38400 baud rate
Max setting with min 38400 baud rate
Delay in milliseconds which elapses before the instrument replies
dELAY
80Hz 100Hz 200Hz 300Hz
0-200 msec
0 *
PAritY nOnE *
EUEn Odd
StOP
1 *
parity none even parity odd parity
Stop bit
2
*
- indicates default value.
Configuration 25
Page 30
2.9.1 RS-485 Serial Communication
817181718171
SCTSCT
SCT
RS485 +
RS485 -
max 500 m
RS485 +
RS485 -
PC RS232
RX+
RX-
TX-
TX+
Converter
24 Vcc
+
-
0
TX
RX
VIN
RS485 +
RS485 -
RS485 termination
20 20
0 VDC
20
RS485 +
RS485 -
0 VDC
RS485 +
RS485 -
0 VDC
5 2 3
If your device is labeled A and B
• Connect A to RS485-
• Connect B to RS485+
Note
Figure 2-8. RS-485 Serial Communications
If the RS-485 network exceeds 100 metres in length or baud rate over 9600 are used, close the two jumpers, called "RS-485 termination", to activate two 120 ohm terminating resistors between the ‘+’ and ‘–’ terminals of the line, on the terminal strip of the furthest instrument. Should there be different instruments or converters, refer to the specific manuals to determine whether it is necessary to connect the above-mentioned resistors.
26 SCT Weight Transmitter Operator’s Manual
Page 31
Direct Connection Between RS-485 And RS-232 Without Converter
Note
OUt 1
OUt 2 OUt 3 In 1 IN 2
CLOSE
OPEn
SEt 1
NOdbUS
StAbLE
GrOSS
NEt
dEunEt
2ErO
PEAH
COntI n
PLC
nE-lO
COEFF
2Er0 PEAH
COntI n
PLC
nE-lO
COEFF
CLOSE
OPEn
SEt 2
NOdbUS
StAbLE
dEunEt
CLOSE
OPEn
SEt 3
NOdbUS
StAbLE
POS/nEG
POS nEG
dEunEt
OFF
On
CALIb
FILtEr
PArA 0
SErIAL
tESt
Out-In
ANA LOG
DeviceNet Options
SEt 1
StAbLE
PLC
SEt 2
StAbLE
PLC
SEt 3
StAbLE
PLC
Analog Options
SEt 1
NOdbUS
StAbLE
PrOFI
SEt 2
NOdbUS
StAbLE
PrOFI
SEt 3
NOdbUS
StAbLE
PrOFI
SEt 1
NOdbUS
StAbLE
EtHnEt
SEt 2
NOdbUS
StAbLE
EtHnEt
SEt 3
NOdbUS
StAbLE
EtHnEt
Profibus Options
Ethernet Options
Since a two-wire RS-485 output may be used directly on the RS-232 input of a PC or remote display, it is possible to implement instrument connection to an RS-232 port in the following manner:
Instrument RS-232
RS-485 - RXD
RS-485 + GND
This type of connection allows a SINGLE instrument to be used in a ONE WAY mode.
2.10 Outputs And Inputs Configuration
Figure 2-9. Outputs and Inputs Menu Structure
Configuration 27
Page 32
Parameter Choices Description
OUt 1 OUt 2 OUt 3
OUt 1 OUt 2 OUt 3
SEt 1 SEt 2 SEt 3
OPEn
CLOSE *
SEt 1 SEt 2 SEt 3
NOdbUS
dEunEt
PrOFI
EtHnEt
PLC
StAbLE POSnEG * POS NEG
Normally Open: the relay is de-energized and the contact is open when the weight is lower than the programmed setpoint value; it closes when the weight is higher than or equal to the programmed setpoint value.
Normally closed: the relay is energized and the contact is closed when the weight is lower than the programmed setpoint value; it opens when the weight is higher than or equal to the programmed setpoint value.
Number corresponds with OUT 1, 2 or 3. The contact will switch on the basis of weight, according to setpoints (see Section 2.12 “Setpoints Programming” on page 30) Select:
Gross (default) - the contact will switch on the basis of gross weight or Net - the contact will switch on the basis of net weight (If the net function is not active,
the contact will switch on the basis of gross weight).
The contact will not switch on the basis of weight, but is controlled by remote Modbus protocol commands. (DeviceNet, Profibus, Ethernet Only)
The contact will not switch on the basis of weight, but is controlled by DeviceNet. (DeviceNet Model Only)
The contact will not switch on the basis of weight, but is controlled by Profibus. (Profibus Model Only)
The contact will not switch on the basis of weight, but is controlled by Ethernet. (Ethernet Model Only)
The contact will not switch on the basis of weight, but is controlled by remote protocol commands. (Analog Only)
Relay switching occurs when the weight is stable.
Relay switching occurs for both positive and negative weight values.
Relay switching occurs for positive weight values only.
Relay switching occurs for negative weight values only.
SEt 1 SEt 2
OFF * On
SEt 3
In 1 In 2
nE-LO *
(In 2 default)
2ErO *
(In 1 default)
PEAH
PLC
COntIn
COEFF
*
- indicates default value.
Relay switching will not occur if the setpoint value is ‘0’.
Setpoint = ’0’ and nodbus=posneg, relay switching occurs when the weight is ‘0’; the
relay will switch again when the weight is different from zero, taking hysteresis into account (both for positive and for negative weights).
Setpoint = ’0’ and nodes=pos, relay switching occurs for a weight higher than or equal to
‘0’, the relay will switch again for values below ‘0’, taking hysteresis into account.
Setpoint = ’0’ and nodes=neg, relay switching occurs for a weight lower than or equal to
‘0’, the relay will switch again for values above ‘0’, taking hysteresis into account.
(NET/GROSS): by closing this input for less than one second, it performs a SEMI­AUTOMATIC TARE and the display will show the net weight. To display the gross weight again, hold the NET/GROSS input closed for 3 seconds.
By closing the input for less than one second, the weight is set to zero (see Section 3.3 “Semi-automatic Zero (Weight Zero-setting For Small Variations)” on page 33)
With the input closed the maximum weight value reached remains on display. Opening fthe input the current weight is displayed.
Closing the input no operation is performed, the input status may however be read remotely by way of the communication protocol.
Closing the input for less than one second the weight is transmitted via the serial connection according to the fast continuous transmission protocol one time only (only if contin is set in the item serial).
When the input is closed the weight is displayed based on the set coefficient (see Section
2.1.7 “Display Coefficient” on page 16), otherwise the weight is displayed.
28 SCT Weight Transmitter Operator’s Manual
Page 33
2.11 Test
1. Press and hold and to enter set-up menu. CALIb will be displayed.
2. Press or until tEst is displayed, press .
3. Press or until desired parameter is displayed, press .
4. For In and NU-CEL, current reading is displayed, press . For Out, press until corresponding value of the out you want to change is flashing, press to change
the value, press .
5. Press twice to exit set-up menu.
Parameter Choices Description
In
Out 0 *
ANALOG ANALOG Allows the analog signal to range between the minimum
NU-CEL
*
- indicates default value.
N/A Input Test - for each open input 0 is displayed, 1 is
displayed when the input is closed.
Output Test -
1
NA Current output test UOLt voltage output test
N/A Millivolt Test - displays the load cell response signal in
Setting 0 opens the corresponding output . Setting 1 closes the corresponding output .
and the maximum values starting from the minimum.
mV with four decimals.
Configuration 29
Page 34
2.12 Setpoints Programming
Note
1. Press to enter setpoints and hysteresis settings.
2. Press or until desired setpoint or hysteresis parameter is displayed, press to enter.
3. Press or until desired value is displayed, press to confirm.
4. Press to exit setpoints and hysteresis settings.
These values are set to zero if the calibration is changed significantly (see Section 2.1.1 “Theoretical Calibration” on page 13 and Section 2.1.5 “Weight (Span) Calibration (With Test Weights)” on page 15).
Parameter Choices Description
SEtP 1 SEtP 2
0-Full Scale
0 *
SEtP 3
HYStE 1 HYStE 2
0-Full Scale
0 *
HYStE 3
*
- indicates default value.
Setpoint; relay switching occurs when the weight exceed the value set in this parameter. The type of switching is settable (see Section 2.10 “Outputs And Inputs Configuration” on page 27).
Hysteresis, value to be subtracted from the setpoint to obtain contact switching for decreasing weight. For example with a setpoint at 100 and hysteresis at 10, the switching occurs at 90 for decreasing weight.
30 SCT Weight Transmitter Operator’s Manual
Page 35
2.13 Reserved For The Installer
Important
Note
2.13.1Menu Locking
Through this procedure, it is possible to block the access to any menu on the instrument.
1. Press and hold then press to enter set-up menu.
2. Press or until menu to be locked is displayed.
3. Press , and simultaneously for 3 seconds, the display shows C.ALIb (a decimal point appears between the first and second letter that indicates this menu is now locked). If the operator tries to enter this menu, access is denied and the display reads bLOC.
2.13.2Menu Unlocking
1. Press and hold then press to enter set-up menu.
2. Press or until menu to be un-locked is displayed.
3. Press , and simultaneously for 3 seconds, the display shows CALIb (the decimal point between the first and second letter is gone indicating the menu is un-locked).
2.13.3Temporary Menu Unlocking
1. Press and hold then press to enter set-up menu.
2. Press or until menu to be temporarily un-locked is displayed.
3. Press and simultaneously for 3 seconds, the display shows CALIb (the decimal point between the first and second letter is gone indicating the menu is un-locked).
4. It is now possible to enter and modify all menus including those which are locked. By returning to the weight display, the menu lock is restored.
2.13.4Default Scale
Operation must only be performed after contacting technical assistance
1. With power off, press and hold , then power on. Display shows PrOG .
2. Press , bASE is displayed.
3. Press , PASSU is displayed.
4. Press , 000000 is displayed.
5. Press and to enter 6935.
6. Press , UAIt is displayed.
7. Instrument will reboot.
By confirming the displayed program, the system variables are set to default values.
Configuration 31
Page 36
2.13.5Program Selection - Reverse:
Note
Scale capacity is displayed when scale is empty. As weight is added display will count down.
1. With power off, press and hold , then power on. PrOG is displayed.
2. Press , bASE is displayed.
3. Press , rEuEr is displayed.
4. Press , UAIt is displayed.
5. Instrument will reboot.
2.13.6Program Selection - Not Legal:
Scale approval state is preset for not-legal. For other choices please contact technical assistance:
nOtLEG: Not Legal for Trade. LEGAL: Legal for Trade (OIML) NuLt: Legal for Trade, multi-interval (OIML)
1. With power off, press and hold , then power on. PrOG is displayed.
2. Press , bASE is displayed.
3. Press , NOtLEG is displayed.
4. Press , UAIt is displayed.
5. Instrument will reboot.
By pressing you will quit the program without introducing any changes and without deleting any of the set variables.
If you do not have a specific manual for the newly set program, you can request it from technical assistance.
2.13.7Keypad Or Display Locking
1. Press immediately followed by , hold them down for about 5 seconds (this operation is also possible via the MODBUS and ASCII protocols):
2. Press or until desired parameter is displayed, press .
Parameter Description
FrEE KEY
dI SP
no lock
keypad lock: if active, when key is pressed the message bLOC is displayed.
Keypad and Display lock: if active, the keypad is locked and the display shows the instrument model (weight is not displayed); by pressing a key the display shows bLOC for 3 seconds.
32 SCT Weight Transmitter Operator’s Manual
Page 37
3.0 Operation
Note
Note
Note
Note
Note
3.1 Semi-automatic Tare (Net/Gross)
The semi-automatic tare value is lost upon instrument power-off.
The semi-automatic tare operation is not allowed if the gross weight is zero.
1. To capture tare and weigh in net mode (SEMI-AUTOMATIC TARE), close the NET/GROSS input or press for 3 seconds. The instrument displays the net weight (zero) and the NET LED lights up.
2. To display the gross weight again, keep the NET/GROSS input closed or press for 3 seconds.
3. This operation can be repeated by the operator to allow the loading of several products.
Press and hold to display the gross weight temporarily. When is released, the net weight will be displayed again.
3.2 Preset Tare (Subtractive Tare Device)
It is possible to manually set a preset tare value to be subtracted from the display value provided that the P-tArE max capacity.
1. Press and hold and to display P-tArE, press .
2. Press or until desired value is displayed, press .
3. Press to exit P-tArE.
4. After setting the tare value, go back to the weight display, the display shows the net weight (subtracting the preset tare value) and the NET LED lights up to show that a tare has been entered.
Press and hold for 3 seconds to display the gross weight temporarily. When is released, the net weight will be displayed again.
To delete a preset tare and return to the gross weight display:
1. Press hold for 3 seconds or keep the NET/GROSS input (if any) closed for the same length of time (3 seconds). The preset tare value is set to zero. The NET LED is turned off when the gross weight is displayed once again.
If a semi-automatic tare (net) is entered, it is not possible to access the enter preset tare function.
If a preset tare is entered, it is still possible to access the semiautomatic tare (net) function. The two different types of tare are added.
All the semi-automatic tare (net) and preset tare functions will be lost when the instrument is turned off.
3.3 Semi-automatic Zero (Weight Zero-setting For Small Variations)
By closing the SEMI-AUTOMATIC ZERO input, the weight is set to zero. The zero setting will be lost when the instrument is turned off.
This function is only allowed if the weight is lower than the 0 set value (see 0 SET in Section 2.3 on page 18),
-----
t
otherwise the
alarm appears and the weight is not set to zero.
3.4 Peak
By keeping the input closed the maximum weight value reached remains displayed. Opening the input the current weight is displayed.
If you wish to use this input to view a sudden variation peak, set the FILTER ON THE WEIGHT to 0.
Operation 33
Page 38
3.5 Alarms
Note
Display Description
Load cell is not connected or is incorrectly connected; the load cell signal exceeds 39 mV; the conversion electronics
ErCEL
Er OL Er Ad
---------
Er OF
------
t
NAH-PU
Err0r
bLOC nOdl SP
DeviceNet Error (DeviceNet Models Only): A problem in the DeviceNet communication is marked by a lack of
flashing or a slow flashing (duration about 0.5 s) of the DeviceNet status LED.
Profibus Error (Profibus Models Only): A problem in the Profibus communication is marked by a lack of
flashing or a slow flashing (duration about 0.5 s) of the Profibus status LED.
Ethernet Error (Ethernet Models Only): A problem in the Ethernet communication is marked by a slow
flashing (duration about 0.5 s) of the Ethernet status LED.
Modbus/TCP Error (Modbus/TCP Models Only): A problem in the Modbus/TCP communication is marked by
a slow flashing (duration about 0.5 s) of the Modbus/TCP status LED.
(A/D converter) is malfunctioning; the load cell is a 4-wire and there are no jumpers between EX- and REF- and between EX+ and REF+.
Weight display exceeds 110% of the full scale.
Internal instrument converter failure; check load cell connections, if necessary contact Technical Assistance.
The weight exceeds the maximum weight by 9 divisions.
Maximum displayable value exceeded (value higher than 999999 or lower than -999999).
Weight too high: zero setting not possible.
This message appears in the test weight setting, in Weight (Span) Calibration, after the fifth test weight value has been entered.
The value set for the parameter is beyond the permitted values; press to quit the setting mode leaving the previous value unchanged.
Example: a number of decimals is selected for full scale which exceeds the instrument's display potential; value above the maximum setting value; the weight value set in test weight verification does not match the detected mV increase.
Lock is active on menu item, keypad or display.
It’s not possible to display the number properly because it is greater than 999999 or less than -999999.
Table 3-1. Alarm Descriptions
MODE ErCEL Er OL Er Ad --------- Er OF t-----
Bit LSB
Status Register
MODBUS RTU
ASCII __O-F_ __O-L_ __O-F_ __O-L_ __O-F_ &aa#CR
RIP * __O-F_ __O-L_ __O-F_ __O-L_ __O-F_ __O-F_
HDRIP-N _ERCEL _ER_OL _ER_AD ###### _ER_OF O__SET
CONTIN _ERCEL _ER_OL _ER_AD ^^^^^^ _ER_OF O__SET
76543210 xxxxxxx1
76543210 xxxx1xxx
76543210 xxxxxx1x
76543210 xxxxx1xx
76543210 On gross: xxx1xxxx On net: xx1xxxxx
The response to the zero command is a 'value not valid' error (error code 3)
Table 3-2. Serial Protocols Alarms
* For RIP remote displays, if the message exceeds 5 digits the display reads
In case of alarm the relays, not managed by Modbus, Modbus/TCP or DeviceNet, open.
------
.
In Analog Models only, if an alarm becomes active the relays open and the analog outputs go to the lowest
possible value according to the following table:
Range 0/20mA 4/20 mA 0/5 V 0/10 V -10/10 V -5/5 V
Output
Value
-0.2 mA 3.5 mA -0.5 V -0.5 V 0 V 0 V
34 SCT Weight Transmitter Operator’s Manual
Page 39
3.6 SCT20-DN (DeviceNet)
Note
Note
The instrument works as a slave in a DeviceNet network.
Load the included eds file (TLBDNT.eds) in the DeviceNet master’s development instrument.
The data exchanged by SCT20-DN are:
Outgoing Data from SCT20-DN (Read) Addresses
Gross Weight [4 bytes] 0x0000-0x0003 Net Weight [4bytes] 0x0004-0x0007 Setpoint [4 bytes] 0x0008-0x000B Status Register [2 bytes] 0x000C-0x000D Digital Inputs Status [1 byte] 0x000E Digital Outputs Status [1 byte] 0x000F
Input Data to SCT20-DN (Write) Addresses
Command Register [2 bytes] 0x0000-0x0001 Digital Outputs Command [2 bytes] 0x0002-0x0003 Setpoint [4 bytes] 0x0004-0x0007
Possible Command To Send To The Command Register
0 No command 90 Read setpoint 1
7 Semi-automatic tare (NET display) 91 Read setpoint 2
8 SEMI-AUTOMATIC ZERO 92 Read setpoint 3
9 GROSS display 93 Write setpoint 1
21 Keypad lock 94 Write setpoint 2
22 Keypad and display unlock 95 Write setpoint 3
23 Keypad and display lock 99 Save data in EEPROM
9999 Reset (reserved)
DeviceNet
If it is necessary to perform the same command twice in a row, send the command 0 between a command and the other.
Gross Weight, Net Weight:
The weight values are expressed as positive integer numbers, including decimal figures but without decimal point.
Read the “Status Register” to obtain information about sign and possible errors on the weight.
Setpoints:
The Setpoints are weight values expressed as positive integer numbers, including decimal figures but without decimal point.
• READING: send to the Command Register the reading command of the required setpoint (e.g. 90 for reading setpoint 1) and read the content of the "Setpoint" register.
• WRITING: write the weight value to be set in the “Setpoint” and send to the Command Register, the writing command of the required setpoint (e.g. 93 for writing setpoint 1).
Setpoints are stored to the RAM volatile memory and lost upon instrument power off. To save them permanently in the EEPROM memory, so that they are maintained upon the instrument power on, it is necessary to send the command 99 “Save data in EEPROM” of the Command Register.
Digital Inputs Status
Bit 0 INPUT 1 status Bit 4
Bit 1 INPUT 2 status Bit 5
Bit 2 Bit 6
Bit 3 Bit 7
Bit a 1: input high; Bit a 0: input is low
Operation 35
Page 40
Digital Outputs Status
Note
Bit 0 OUTPUT 1 status Bit 4
Bit 1 OUTPUT 2 status Bit 5
Bit 2 OUTPUT 3 status Bit 6
Bit 3 Bit 7
Digital Outputs Command
It allows to control the outputs status in dEUnEt mode (see Section 2.10 on page 27)
Bit 0 OUTPUT 1 status Bit 8
Bit 1 OUTPUT 2 status Bit 9
Bit 2 OUTPUT 3 status Bit 10
Bit 3 Bit 11
Bit 4 Bit 12
Bit 5 Bit 13
Bit 6 Bit 14
Bit 7 Bit 15 Force Outputs
Bit a 1: output is closed; Bit a 0: output is open.
Setting bit 15 to 1 on the PLC, DeviceNet takes control of all the outputs, even if they are in different modes.
Status Register
Bit 0 Cell Error
Bit 1 A/D Converter Malfunction
Bit 2 Maximum weight exceeded by 9 divisions
Bit 3 Gross weight higher than 110% of full scale
Bit 4 Gross weight beyond 999999 or less than -999999
Bit 5 Net weight beyond 999999 or less than -999999
Bit 6
Bit 7 Gross weight negative sign
Bit 8 Net weight negative sign
Bit 9 Peak weight negative sign
Bit 10 Net display mode
Bit 11 Weight stability
Bit 12 Weight within +/-¼ of a division around ZERO
Bit 13
Bit 14
Bit 15
3.7 Profibus
• The instrument works as a Profibus DP V0 slave in a Profibus DP network.
• Load the included gsd file (LAU_0C91.gsd) attached to the instrument in the development system of the Profibus master.
• The available modules are:
- R = the register can be read only
- W = the register can be written only
- R/W = the register can be both read and written
36 SCT Weight Transmitter Operator’s Manual
Page 41
MODULE Saving to EEPROM
Note
GROSS WEIGHT) 4 R
NET WEIGHT 4 R
PEAK WEIGHT 4 R
SETPOINT 1
SETPOINT 2 4 4 R/W
SETPOINT 3 4 4 R/W
HYSTERESIS 1 4 4 R/W
HYSTERESIS 2 4 4 R/W
HYSTERESIS 3 4 4 R/W
Divisions and Units of measure
Visualization Coefficient 4 R
INPUTS 2 R
OUTPUTS NO 22R/W
STATUS REGISTER 2 R
COMMAND REGISTER NO 2 W
Sample Weight for calibration
Only after command ‘99’
of the
COMMAND REGISTER
Use with command ‘101’ of the COMMAND REGISTER
N. Byte RN. Byte
44R/W
2 R
44R/W
W
ACCES
S
Profibus
GROSS WEIGHT – NET WEIGHT – PEAK
The weight values are expressed as positive integer numbers, include decimal figures but without decimal point.
Read the “Status Register” to obtain information about sign and possible errors on the weight.
To find out the decimal figures use the Division module value; E.g.: the net weight is 100000 and the division is
0.001 (three decimals), the real weight value is 100.000kg
Setpoints – Hysteresis
The weight values are expressed as positive integer numbers, include decimal figures but without decimal point.
• to set the value to 0; write the hexadecimal value hex 80000000 to the register (the most significant bit is 1 and all the other bits are 0).
• to set them correctly consider the module Division value; E.g. if you want to set a set-point to 100kg and the verification division value is 0.001, set the set-point value to 100000 (you remove the decimal points from the value that would be 100.000 with decimals).
• If from PLC you set a value out of the permitted interval (from 0 (not included) to full scale) the value is ignored and the bit 13 “Writing error” rises in the “Status Register”.
Setpoints and Hysteresis are stored to the RAM volatile memory and lost upon instrument power off. To save them permanently in the EEPROM memory, so that they are maintained upon the instrument power on, it is necessary to send the command 99 “Save data in EEPROM” of the Command Register.
Divisions And Units Measure Registry (40014)
This register contains the current setting of the divisions (parameter dI UI S) and of the units of measure (UnIt parameter).
H Byte L Byte
Units of measure division
Use this register together with the coefficient registers to calculate the value displayed by the instrument.
Operation 37
Page 42
Least significant byte
(L Byte)
Most significant byte
(H Byte)
Units of
Division
value Divisor Decimals
0 100 0 0 Kilograms Does not intervene
1 50 0 1 Grams Does not intervene
2 20 0 2 Tons Does not intervene
3 10 0 3 Pounds Does not intervene
4 5 0 4 Newton Multiples
520 5 Litres Divides
6 1 0 6 Bar Multiples
70.51 7
8 0.2 1 8 Pieces Divides
90.11 9
10 0.05 2 10
11 0.02 2 11 Other Multiples
12 0.01 2
13 0.005 3
14 0.002 3
15 0.001 3
16 0.0005 4
17 0.0002 4
18 0.0001 4
measure
value
Units of
measure
description
Atmsphere s
Newton Meter
Kilogram Meter
Utilization of the Coefficient value with the different units of measure settings compared to
the gross weight detected
Multiples
Multiples
Multiples
Coefficient Display
The Profibus register contains the value of parameter COEFF, expressed as integer number, with 4 decimal figures but without decimal point.
E.g.: If COEFF is 1.2000, the Profibus register contains 12000
38 SCT Weight Transmitter Operator’s Manual
Page 43
Digital Inputs Status
Note
Bit 0 INPUT 1 status Bit 8
Bit 1 INPUT 2 status Bit 9
Bit 2 Bit 10
Bit 3 Bit 11
Bit 4 Bit 12
Bit 5 Bit 13
Bit 6 Bit 14
Bit 7 Bit 15
Bit a 1: input high; Bit a 0: input is low
Digital Outputs Command
It allows control to the outputs status in PrOFI (see Section 2.10 on page 27)
Bit 0 OUTPUT 1 status Bit 8
Bit 1 OUTPUT 2 status Bit 9
Bit 2 OUTPUT 3 status Bit 10
Bit 3 Bit 11
Bit 4 Bit 12
Bit 5 Bit 13
Bit 6 Bit 14
Bit 7 Bit 15 Force Outputs
Bit a 1: output is closed; Bit a 0: output is open.
Setting bit 15 to 1 on the PLC, Profibus takes control of all the outputs, even if they are in different modes.
Operation 39
Page 44
Status Register
Note
Bit 0 Cell Error
Bit 1 A/D Converter Malfunction
Bit 2 Maximum weight exceeded by 9 divisions
Bit 3 Gross weight higher than 110% of full scale
Bit 4 Gross weight beyond 999999 or less than -999999
Bit 5 Net weight beyond 999999 or less than -999999
Bit 6
Bit 7 Gross weight negative sign
Bit 8 Net weight negative sign
Bit 9 Peak weight negative sign
Bit 10 Net display mode
Bit 11 Weight stability
Bit 12 Weight within +/-¼ of a division around ZERO
Bit 13 Writing error
Bit 14
Bit 15
Bit 13 (writing error) rises if the PLC tries to write a value out of a parameter’s permitted interval. In this case the writing has no effect and the previous parameter value is maintained.
Possible Command To Send To The Command Register
0 No command 17 Reserved
7 NET display 18 Reserved
8 SEMI-AUTOMATIC ZERO 21 Keypad lock
10 Reserved 22 Keypad and display unlock
11 Reserved 23 Keypad and display lock
12 Reserved 99 Save data in EEPROM
13 Reserved 100 Zero-setting for calibration
14 Reserved 101
15 Reserved 9999 Reset (reserved)
16 Reserved
Commands are sent only once; to reassert it write the value 0 to the Command register and then the command again
Sample weight storage for calibration
Real Calibration Commands (With Sample Weight)
The instrument calibration can be changed via PROFIBUS.
To correctly set the sample weight, consider the value of the division module (e); e.g. if you want to set the value to 100kg and the division value is 0.001, set the register value to 100000 (you remove the decimal points form the value that would be 100.000 with decimals).
To perform the edit calibration procedure, it needs to unload the system and zero-setting the displayed weight value with the command 100 “Zero-setting for calibration” of Command Register. Then, put a sample weight onto the system, write the correct known weight value to the “Sample Weight for calibration” register; to save in memory the sample weight value, send the command 101 “Sample weight storage for calibration”.
40 SCT Weight Transmitter Operator’s Manual
Page 45
To verify the correct execution of the calibration command:
Note
• Write the sample weight value to the “Sample Weight for calibration” module;
• Write the value 0 to the “”Sample Weight for calibration” module;
• Send the command 101 “Sample weight storage for calibration” to the Command Register;
• The operation works correctly if the module value “Sample Weight for calibration” goes to zero.
Changing some instrument parameters via keyboard can take the Profibus network out of the data exchange status and put the PLC in stop status; when programming the instrument the plant must be in stand-by status
3.8 SCT20-IP (EtherNet/IP)
The instrument works as a device in a Ethernet/IP network.
Load the included eds file (TLBEIP.eds) in the Ethernet/IP scanner’s development instrument. Or you can open Class 1 I/O Connection with the following settings:
• Assembly Instance = 101; Size = 10
• Assembly Instance = 102; Size = 5
• Assembly Instance = 128; Size = 0
EtherNet/IP
The data exchanged by SCT20-IP are:
Outgoing Data from SCT20-IP (Read) Addresses
Internal Status [2 bytes] 0x0000-0x0001
Gross Weight [4 bytes] 0x0002-0x0005
Net Weight [4bytes] 0x0006-0x0009
Setpoint [4 bytes] 0x000A-0x000D
Status Register [2 bytes] 0x000E-0x000F
Digital Inputs Status [2 byte] 0x0010-0x0011
Digital Outputs Status [2 byte] 0x0012-0x0013
Input Data to SCT20-IP (Write) Addresses
Write Enable [2 bytes] 0x0000-0x0001 Command Register [2 bytes] 0x0002-0x0003 Digital Outputs Command [2bytes] 0x0004-0x0005 Setpoint [4 bytes] 0x0006-0x0009
Operation 41
Page 46
Possible Command To Send To The Command Register
Note
Note
0 No command 90 Read setpoint 1
7 Semi-automatic tare (NET display) 91 Read setpoint 2
8 SEMI-AUTOMATIC ZERO 92 Read setpoint 3
9 GROSS display 93 Write setpoint 1
21 Keypad lock 94 Write setpoint 2
22 Keypad and display unlock 95 Write setpoint 3
23 Keypad and display lock 99 Save data in EEPROM
9999 Reset (reserved)
If it is necessary to perform the same command twice in a row, send the command 0 between a command and the other.
Internal Status
When it is not 0 there is an internal error, so data from SCT20-IP are not reliable.
When it is 0 data from SCT20-IP are reliable.
Write Enable
Write 0x0000 in Write Enable register if you want that no data are written to SCT20-IP. Write 0xFFFF in Write Enable register if you want to enable that data are written to SCT20-IP.
Gross weight, Net weight:
The weight values are expressed as positive integer numbers, include decimal figures but without decimal point.
Read the “Status Register” to obtain informations about sign and possible errors on the weight.
Setpoints:
The Setpoints are weight values expressed as positive integer numbers, include decimal figures but without decimal point.
• READING: send to the Command Register the reading command of the required setpoint (e.g. 90 for reading setpoint 1) and read the content of the "Setpoint" register.
• WRITING: write the weight value to be set in the “Setpoint” and send to the Command Register, the writing command of the required setpoint (e.g. 93 for writing setpoint 1).
Setpoints are stored to the RAM volatile memory and lost upon instrument power off. To save them permanently in the EEPROM memory, so that they are maintained upon the instrument power on, it is necessary to send the command 99 “Save data in EEPROM” of the Command Register.
Digital Inputs Status
Bit 0 INPUT 1 status Bit 4
Bit 1 INPUT 2 status Bit 5
Bit 2 Bit 6
Bit 3 Bit 7
Bit a 1: input high; Bit a 0: input is low
42 SCT Weight Transmitter Operator’s Manual
Page 47
Digital Outputs Status
Note
Bit 0 OUTPUT 1 status Bit 4
Bit 1 OUTPUT 2 status Bit 5
Bit 2 OUTPUT 3 status Bit 6
Bit 3 Bit 7
Digital Outputs Command
It allows control to the outputs status in EtHnEt mode (see Section 2.10 on page 27)
Bit 0 OUTPUT 1 status Bit 8
Bit 1 OUTPUT 2 status Bit 9
Bit 2 OUTPUT 3 status Bit 10
Bit 3 Bit 11
Bit 4 Bit 12
Bit 5 Bit 13
Bit 6 Bit 14
Bit 7 Bit 15 Force Outputs
Bit a 1: output is closed; Bit a 0: output is open.
Setting bit 15 to 1 on the PLC, DeviceNet takes control of all the outputs, even if they are in different modes.
Status Register
Bit 0 Cell Error
Bit 1 A/D Converter Malfunction
Bit 2 Maximum weight exceeded by 9 divisions
Bit 3 Gross weight higher than 110% of full scale
Bit 4 Gross weight beyond 999999 or less than -999999
Bit 5 Net weight beyond 999999 or less than -999999
Bit 6
Bit 7 Gross weight negative sign
Bit 8 Net weight negative sign
Bit 9 Peak weight negative sign
Bit 10 Net display mode
Bit 11 Weight stability
Bit 12 Weight within +/-¼ of a division around ZERO
Bit 13
Bit 14
Bit 15
Operation 43
Page 48
3.9 SCT20-TCP/IP (Ethernet TCP/IP)
To configure Ethernet TCP/IP port of SCT20-TCP/IP, the Lantronix DeviceInstaller must be installed on a Windows PC (launch DevInst.exe on the CD to install it).
1. Connect the PC and SCT20-TCP/IP through a LAN (point-to-point or with hub/switch), and then launch
the Lantronix DeviceInstaller.
2. Select the
3. Select the device and then select the Telnet Configuration.
EtherNet TCP/IP
4. Select the
Search button and the application will find the SCT20-TCP/IP in the LAN.
Figure 3-1. Find SCT20-TCP/IP in LAN
Connect button and then press Enter.
5. Press
0 [Server] and change only IP Address four fields (don’t change other parameters: press Enter to
confirm). It is strongly recommended to set a fixed IP Address.
6. Press
1 [Channel 1] if you want to change serial BaudRate of the Ethernet Module to match the BaudRate
set in SCT20-TCP/IP Protocol Selection (default 9600) (don’t change other parameters: press confirm.
44 SCT Weight Transmitter Operator’s Manual
Enter to
Page 49
To connect to the SCT20-TCP/IP using a socket (for example Winsock), the port to use is 10001.
Note
To connect to SCT20-TCP/IP via Ethernet TCP/IP through a serial virtual port (virtual COM port of a PC) perform the following steps.
1. Run TCPIP.exe found on the CD
2. Then enter IP Address of SCT20-TCP/IP and press
ACTIVATE. It activates socket connection.
3. Type the ASCII command (without CR) and press
Figure 3-2. TCP/IP Socket Test Screen
SEND. You can see the answer in Received box.
Figure 3-3. TCP/IP Socket Test Screen (Examples)
Use the added COM to communicate with SCT20-TCP/IP using the selected protocol.
SCT20 TCP/IP.exe is included on the CD for testing commands, it only works with Port 10001.
Do not include the <CR> with the command.
Supports the same commands as ASCII, see Section 3.12.
Protocol Selection
To select the protocol to use on the Ethernet port, you can choose one of the following:
EthNet - Modelines of the list below (none, Modbus, ASCII, Contin, Rip, Hdrip, Hdripn). For details about
these protocols see Section 2.8.
IPAddr,SUbNET,GAtuAYare not used yet, IPAddress, Subnet Mask or Gatway address cannot be assigned directly through the keyboard: use configuration SW Lantronix DeviceInstaller.
UEbSrvprotocol is not implemented yet.
Operation 45
Page 50
3.10 Modbus/TCP
The instrument works as a slave in a Modbus/TCP network. TLBMODBUSTCP instrument is configured with DHCP (default). IP Address can be automatically assigned by DHCP or manually via Telnet. To manually set the IP address via PC, type "telnet <ipaddress> 9999" and press Enter to confirm. The following screen appears:
1. Type “1” if you want to manually configure IP Address, Default Gateway Address and Netmask.
2. Then type “S” to save.
Modbus/TCP commands and registers of TLBMODBUS/TCP are the same as ModbusRTU protocol (see Section 3.11).
3.11 Modbus-RTU Protocol
The MODBUS-RTU protocol enables management of the reading and writing of the registers listed here below according to the specifications contained in the reference document for this standard Modicon PI-MBUS-300.
To select the communication with MODBUS-RTU, refer to Section 2.9 on page 24.
When specifically indicated certain data will be written directly to EEPROM type memories. This memory has a limited number of writing operations (100.000), therefore unnecessary operations at said locations must be avoided. The instrument, in any case, ensures that no writing occurs if the value to be stored is equal to the stored value.
The numerical data listed below are expressed in decimal notation, or hexadecimal notation if preceded by 0x.
Modbus-RTU Data Format
The data received and transmitted via MODBUS-RTU protocol have the following characteristics:
- 1 start bit
- 8 data bits, least significant bit sent first
- Instrument settable parity bit
- Instrument settable stop bit
46 SCT Weight Transmitter Operator’s Manual
Page 51
Modbus Supported Functions
Among the commands available in the MODBUS-RTU protocol, only the following are used to manage communication with the instruments. Other commands may not be interpreted correctly and could generate errors or system shut-downs:
FUNCTIONS DESCRIPTION
03 (0x03) Read Holding Register (Programmable Register Reading)
16 (0x10) Preset Multiple Registers (Multiple Register Writing)
The interrogation frequency is linked with the preset communication rate (the instrument will stand by for at least 3 bytes before beginning to calculate a possible response to the query). The dELAY parameter (see Section 2.9 on page 24) allows for a further delay in the instrument response, and this directly influences the number of possible queries in the unit of time.
For additional information on this protocol, refer to the general technical specification PI_MBUS_300. In general, the query and response to and from a slave instrument are organized as follows:
Function 3: Read Holding Registers (Programmable Register Reading)
QUERY
Address Function Add. Reg. 1 No. register 2 bytes
A 0x03 0x0000 0x0002 CRC
Tot. bytes = 8
RESPONSE
Address Function No. bytes Register1 Register 2 2 bytes
A 0x03 0x04 0x0064 0x00C8 CRC
Tot. bytes = 3+2*No. registers+2
in which:
No. registers= number of Modbus register to be read, starting from the Address 1° register;
No. bytes = number of data bytes to follow;
Function 16: Preset Multiple Registers (Multiple Register Writing)
QUERY
Address Function Add. reg. 1 No. reg. No. bytes Val. reg.1 Val. reg.2 2 bytes
A 0x10 0x0000 0x0002 0x04 0x0000 0x0000 CRC
Tot. bytes = 7+2*No. registers+2
RESPONSE
Address Function Add. Reg. 1 No. register 2 bytes
A 0x10 0x0000 0x0002 CRC
Tot. bytes = 8
in which:
No. registers= number of Modbus register to be read, starting from the Address 1° register;
No. bytes = number of data bytes to follow;
Val. reg. 1 = register contents beginning from the first.
The response contains the number of records changed starting from the Address 1° register.
Operation 47
Page 52
Communication Error Management
The communication strings are controlled by CRC (Cyclical Redundancy Check).
In case of a communication error the slave will not respond with any string. The master must allow for a time-out before response reception. If no response is received it infers that a communication error has occurred.
In the event of a string received correctly but not executable, the slave responds with an EXCEPTIONAL RESPONSE. The "FUNCTION" field is transmitted with the msb at 1.
EXCEPTIONAL RESPONSE
Address Function Code 2 bytes
A Funct + 0x80 CRC
CODE DESCRIPTION
1
ILLEGAL FUNCTION (Function not valid or not supported)
2
ILLEGAL DATA ADDRESS (The specified data address is not available)
3
ILLEGAL DATA VALUE (The data received have no valid value)
List Of Usable Registers
The MODBUS-RTU protocol implemented on this instrument can manage a maximum of 32 registers read and written in a single query or response.
R = the register can be read only W = the register can be written only R/W = the register can be both read and written H = high half of the DOUBLE WORD forming the number L = low half of the DOUBLE WORD forming the number
REGISTER DESCRIPTION Saving to EEPROM ACCESS
40001 Firmware version - R
40002 Type of instrument - R
40003 Year of Production - R
40004 Serial Number - R
40005 Active program - R
40006 COMMAND REGISTER NO W
40007 STATUS REGISTER - R
40008 GROSS WEIGHT H - R
40009 GROSS WEIGHT L - R
40010 NET WEIGHT H - R
40011 NET WEIGHT L - R
40012 PEAK WEIGHT H - R
40013 PEAK WEIGHT L - R
40014 Divisions and Units of measure - R
40015 Coefficient H R
40016 Coefficient L R
48 SCT Weight Transmitter Operator’s Manual
Page 53
REGISTER DESCRIPTION Saving to EEPROM ACCESS
Important
40017 SETPOINT 1 H
40018 SETPOINT 1 L R/W
40019 SETPOINT 2 H R/W
40020 SETPOINT 2 L R/W
40021 SETPOINT 3 H R/W
40022 SETPOINT 3 L R/W
40023 HYSTERESIS 1 H R/W
40024 HYSTERESIS 1 L R/W
40025 HYSTERESIS 2 H R/W
40026 HYSTERESIS 2 L R/W
40027 HYSTERESIS 3 H R/W
40028 HYSTERESIS 3 L R/W
40029 INPUTS - R
40030 OUTPUTS NO R/W
40037 Test weight for calibration H Use with command ‘101’ of the
40038 Test weight for calibration L R/W
40043 Weight value corresponding to ZERO of the analog
output H
40044 Weight value corresponding to ZERO of the analog
output L
40045 Weight value corresponding to Full Scale of the analog
output H
40046 Weight value corresponding to Full Scale of the analog
output L
Only after command ‘99’ of the
COMMAND REGISTER
COMMAND REGISTER
Only after command ‘99’ of the
Command Register.
(Analog Models Only)
R/W
R/W
R/W
R/W
R/W
R/W
At the time of writing, the setpoints, hysteresis values are saved to the RAM and will be lost upon the next power-off; to store them permanently to the EEPROM so that they are maintained at power-on, the ‘99’ command of the Command Register must be sent.
Weight (Span) Calibration Commands (With Test Weights)
The instrument calibration can be changed via MODBUS. To carry out this procedure, the system must be unloaded and the weight value display reset to zero with the command ‘100’ of the Command Register. Then, a load must be placed on the system and the correct weight value must be sent to the registers 40037-40038; to save this value, send the control ‘101’ from the Command Register. If the operation is successfully completed, the two test weight registers are set to zero.
Analog Output Setting (Analog Models Only)
Write the weight in the registers “Weight value corresponding to the Full Scale of analog output H” (40045) and “Weight value corresponding to the Full Scale of analog output L” (40046) or write the weight in the registers “weight value corresponding to the ZERO of the analog output H” (40043) and “weight value corresponding to the ZERO of the analog output L” (40044). After writing the value, send the command 99 from the Command Register to save it to EEPROM memory.
Operation 49
Page 54
Status Register (40007)
Note
Bit 0 Cell Error
Bit 1 AD Convertor Malfunction
Bit 2 Maximum weight exceeded by 9 divisions
Bit 3 Gross weight higher than 110% of full scale
Bit 4 Gross weight beyond 999999 or less than -999999
Bit 5 Net weight beyond 999999 or less than -999999
Bit 6
Bit 7 Gross weight negative sign
Bit 8 Net weight negative sign
Bit 9 Peak weight negative sign
Bit 10 Net display mode
Bit 11 Weight stability
Bit 12 Weight within +/-¼ of a division around ZERO
Bit 13
Bit 14
Bit 15
INPUTS REGISTER (40029)
(Read Only)
Bit 0 INPUT 1 Status Bit 0 OUTPUT 1 Status Bit 1 INPUT 2 Status Bit 1 OUTPUT 2 Status Bit 2 Bit 2 OUTPUT 3 Status Bit 3 Bit 3 Bit 4 Bit 4 Bit 5 Bit 5 Bit 6 Bit 6 Bit 7 Bit 7 Bit 8 Bit 8 Bit 9 Bit 9 Bit 10 Bit 10 Bit 11 Bit 11 Bit 12 Bit 12 Bit 13 Bit 13 Bit 14 Bit 14 Bit 15 Bit 15
The output status can be read at any time but can be set (written) only if the output has been set as NOdbUS or dEUnEt (see Section 2.10 on page 27); otherwise, the outputs will be managed according to the current weight
status with respect to the relevant setpoints.
OUTPUTS REGISTER (40030)
(Read and Write)
50 SCT Weight Transmitter Operator’s Manual
Page 55
Divisions And Units Measure Registry (40014)
This register contains the current setting of the divisions (parameter dI UI S) and of the units of measure (UnIt parameter).
H Byte L Byte
Units of measure division
Use this register together with the Coefficient registers to calculate the value displayed by the instrument.
Least significant byte
(L Byte)
Division
value Divisor Decimals
0 100 0 0 Kilograms Does not intervene 1 50 0 1 Grams Does not intervene 2 20 0 2 Tons Does not intervene 3 10 0 3 Pounds Does not intervene 4 5 0 4 Newton Multiples 520 5Litres Divides 6 1 0 6 Bar Multiples 7 0.5 1 7 Atmspheres Multiples 8 0.2 1 8 Pieces Divides
9 0.1 1 9 Newton Meter Multiples 10 0.05 2 10 Kilogram Meter Multiples 11 0.02 2 11 Other Multiples 12 0.01 2 13 0.005 3 14 0.002 3 15 0.001 3 16 0.0005 4 17 0.0002 4 18 0.0001 4
Units of measure
value
Units of measure
description
Most significant byte
(H Byte)
Utilization of the Coefficient value with
the different units of measure settings
compared to the gross weight detected
Operation 51
Page 56
Possible Commands To Send To The Command Register (40006)
Note
0 No command 17 Reserved
1 18 Reserved
2 19
3 20
4 21 Keypad lock
5 22 Keypad and display unlock
6 23 Keypad and display lock
7 NET display 24
8 SEMI-AUTOMATIC ZERO 99 Save data in EEPROM
9 GROSS display 100 Zero-setting for calibration
10 Reserved 101 Test weight storage for calibration
11 Reserved
12 Reserved
13 Reserved
14 Reserved
15 Reserved
16 Reserved 9999 Reset (reserved)
3.12 ASCII Bidirectional Protocol
All the sample commands use address 1. If a different address is used it will need a different Check-Sum calculated.
The instrument replies to the requests sent from a PC/PLC. It is possible to set a delay time for the instrument before it transmits a response (see dELAY parameter in Section
2.9 on page 24).
The following communication modes available (see Section 2.9 on page 24):
• NOdU60:
NOd td:
Data Identifiers
$: Beginning of a request string (36 ASCII); & o &&: Beginning of a response string (38 ASCII); aa: 2 characters for instrument address (48 ÷ 57 ASCII); !: 1 character to indicate the correct reception(33 ASCII); ?: 1 character to indicate a reception error (63 ASCII); #: 1 character to indicate an error in the command execution (23 ASCII); ckck: 2 ASCII characters for Check-Sum (for further information, see “Check-Sum Calculation” on
page 56);
CR: 1 character for string end (13 ASCII); \: 1 character for separation (92 ASCII).
52 SCT Weight Transmitter Operator’s Manual
Page 57
Setpoint Values Setting:
Note
The PC transmits : $aaxxxxxxyckckCR
in which:
xxxxxx = 6 characters for the setpoint value (48 ? 57 ASCII); y = A (set the value in the Setpoint 1)$01010000A40CR y = B (set the value in the Setpoint 2)$01010000B42CR y = C (set the value in the Setpoint 3)$01010000C43CR
Possible instrument responses:
- correct reception: &&aa!\ckckCR
- incorrect reception: &&aa?\ckckCR
Setpoints Storage Into EEPROM Memory:
The setpoints value relevant to the two setpoints programmed via the PC are stored to the RAM volatile memory and lost upon instrument power off. It is necessary to send a special command to save them permanently in the EEPROM memory. Please note that the number of writes allowed in the EEPROM memory is limited (about
100000).
The PC transmits: $aaMEMckckCR $01MEM44CR
Possible instrument responses:
- correct reception: &&aa!\ckckCR
- incorrect reception: &&aa?\ckckCR
Reading Weight, The Setpoint And The Peak (If Present) From The Pc:
The PC transmits: $aajckckCR
in which:
j = a to read setpoint 1$01a60CR j = b to read setpoint 2$01b63CR j = c to read setpoint 3$01c62CR j = t to read gross weight$01t75CR j = n to read net weight$01n6FCR j = p to read the gross weight peak if the ASCII parameter is set as NOdU60; if, instead the ASCII parameter
is set on NOd td the gross weight will be read.
To read the points, set the FStE0 equal to 50000. $01p71CR
Possible instrument responses:
- correct reception: &aaxxxxxxj\ckckCR
- incorrect reception: &&aa?\ckckCR
- if the peak is not configured: &aa#CR
in which:
xxxxxx = 6 value characters of the required weight;
In case of negative weight, the first character on the left acquires the value « - » (minus sign - ASCII 45).
In case of weight value is under -99999, the minus sign (‘-‘) is sent alternated with the most significant figure.
Error messages:
In case of an instrument alarm for exceeding 110% of the full scale or 9 divisions above the value of the parameter NASS, the instrument sends the string:
&aassO-Lst
In case of faulty connection of the load cells or of another alarm, the instrument sends:
&aassO-Fst
\ckck
\ckck
Operation 53
Page 58
in which:
Important
s = 1 separator character (32 ASCII – space-).
Generally refer Section 3.5 “Alarms” on page 34.
Semi-Automatic Zero (Weight Zero-Setting For Small Variations)
The zero-setting will not be maintained after an instrument power-off.
The PC transmits: $aaZEROckckCR $01ZERO03CR
Possible instrument responses:
- correct reception: &&aa!
- incorrect reception: &&aa?
\ckckCR
\ckckCR
- the current weight is over the maximum value resettable: &aa#
Switching From Gross Weight To Net Weight
The PC transmits: $aaNETckckCR $01NET5ECR
Possible instrument responses:
- correct reception: &&aa!\
- incorrect reception: &&aa?
ckckCR
\ckckCR
Switching From Net Weight To Gross Weight
The PC transmits: $aaGROSSckckCR $01GROSS5BCR
Possible instrument responses:
- correct reception: &&aa!
- incorrect reception: &&aa?
\ckckCR
\ckckCR
CR
Reading Of Decimals And Number Of Divisions
The PC transmits: $aaDckckCR $01D45CR
Possible instrument responses:
- correct reception: &aaxy
- incorrect reception: &&aa?
in which:
x = number of decimals y = division value
The y field acquires the following values:
'3' for division value = 1;
'4' for division value = 2;
'5' for division value = 5;
'6' for division value = 10;
'7' for division value = 20;
'8' for division value = 50;
'9' for division value = 100;
\ckckCR
\ckckCR
54 SCT Weight Transmitter Operator’s Manual
Page 59
Tare Weight Zero Setting
Important
The PC transmit the following ASCII string containing the zeroing command:
ckckCR $01z7BCR
$aaz
in which:
z = weight zeroing command (122 ASCII)
Possible instrument responses:
- correct reception: &aaxxxxxxt
- incorrect reception: &&aa?
\ckckCR
\ckckCR
- If the instrument is not in gross weight displaying condition, the
response is: &aa#
CR
in which:
xxxxxx = 6 characters for the required weight value; t = weight identification code (116 ASCII).
Example: Weight zero setting for instrument with address 2:
For the calibration, make sure that the scale is empty and the instrument measures a corresponding mV signal.
query: $02z78(Cr) response: &02000000t\76(Cr)
In case of correct weight zero setting the read value (response) must be 0 (in the string “000000”).
The zero values are stored to the EEPROM memory, please note that the number of writes allowed is limited (about 100000). If it is necessary to reset the weight quite often, it is recommended to perform it by PC or PLC program, keeping in mind the weight deviation respect to the zero instrument.
Weight (Span) Calibration (With Test Weights)
After having performed the TARE WEIGHT ZERO SETTING, this function allows correct calibration to be done using test weights of known value and, if necessary, any deviations of the indicated value from the correct value to be corrected.
Load the test weight onto the scale, use as high a percentage of the maximum quantity to be weighed as possible. Otherwise make sure that the instrument measures a corresponding mV signal.
The PC sends the following ASCII string containing the calibration command:
$aasxxxxxx
in which:
s = calibration command (115 ASCII) xxxxxx = 6 characters for test weight value.
Possible instrument responses:
- correct reception: &aaxxxxxxt
- incorrect reception or full scale equal to zero: &&aa?
in which:
t = gross weight identification code (116 ASCII). xxxxxx = 6 characters to indicate the current weight value.
In case of correct calibration, the read value must be equal to test weight.
Example: Calibration for instrument with address 1 and test weight of 20000 kg:
query: $01s02000070(Cr) response: &01020000t\77(Cr)
In case of correct calibration the read value has to be “020000”.
ckckCR
\ckckCR
\ckckCR
Operation 55
Page 60
Keypad Lock (Access Protection To The Instrument)
The PC transmits: $aaKEYckckCR $01KEY56CR
Possible instrument responses:
- correct reception: &&aa!
- incorrect reception: &&aa?
\ckckCR
\ckckCR
Keypad Unlock
The PC transmits: $aaFREckckCR $01FRE50CR
Possible instrument responses:
- correct reception: &&aa!
- incorrect reception: &&aa?
\ckckCR
\ckckCR
Display And Keypad Lock
The PC transmits: $aaKDISckckCR $01KDIS14CR
Possible instrument responses:
- correct reception: &&aa!
- incorrect reception: &&aa?
\ckckCR
\ckckCR
Check-Sum Calculation
The two ASCII control characters (ckck) are the representation of a hexadecimal digit in ASCII characters. The
check digit is calculated by performing the operation XOR (exclusive or) 8-bit ASCII codes of the only part of the underlined string.
The procedure to calculate the check- sum is the following:
• Consider only the string characters highlighted with underlining;
• Calculate the EXCLUSIVE OR (XOR) of the ASCII codes for the characters;
Example:
Character Decimal ASCII Code Hexadecimal ASCII Code Binary ASCII Code
0 48 30 00110000
1 49 31 00110001
t 116 74 01110100
XOR = 117 75 01110101
• The result of the XOR operation expressed in hexadecimal notation is made up of 2 hexadecimal digits (numbers from 0 to 9 or letters from A to F). In this case the hexadecimal code is 0x75.
• The check-sum inserted in the strings transmitted is made up of the 2 characters which represent the result of the XOR operation in hexadecimal notation (in our example the character " 7 " and the character " 5)
56 SCT Weight Transmitter Operator’s Manual
Page 61
3.13 Fast Continuous Transmission Protocol
Note
This protocol allows for continuous serial output at high update frequencies. Up to 300 strings per second are transmitted (with a minimum transmission rate of 38400 baud). See Section 2.9 on page 24 for limitations.
Following communication modes available (see Section 2.9 on page 24):
NOd t: communication compatible with TX RS485 instruments;
NOd td: communication compatible with TD RS485 instruments.
• If NOd t is set, the following string is transmitted to PC/PLC: xxxxxxCRLF
in which:
xxxxxx = 6 ASCII characters for gross weight (48 ÷ 57 ASCII). CR = 1 character of carriage return (13 ASCII). LF = 1 character of line feed (10 ASCII).
In case of negative weight, the first character on the left acquires the value « - » (minus sign - ASCII 45).
In case of error or alarm, the 6 weight characters are replaced by the messages found in Table 3-1 on page 34.
• If NOd td is set, the following string is transmitted to PC/PLC: &TzzzzzzPzzzzzz\ckckCR
in which:
& = 1 character of string start (38 ASCII). T = reference character for gross weight. P = reference character for gross weight. zzzzzz = 6 ASCII characters for gross weight (48 ? 57 ASCII). \ = 1 character of separation (92 ASCII). ckck = 2 ASCII control characters calculated considering that the characters between & and \ are excluded. The
control value is obtained by carrying out the XOR (or exclusive) operation for the 8 bit ASCII codes of the characters considered. A character expressed in hexadecimal is thus obtained, with 2 digits which may acquire
values from “0” to “9” and from “A” to “F”. “ckck” is the ASCII code of the two hexadecimal digits. CR = 1 character for string end (13 ASCII).
In case of negative weight, the first character on the left acquires the value « - » (minus sign - ASCII 45).
In case of error or alarm, the 6 gross weight characters are replaced by the messages found in the table of the ALARMS.
Fast Transmission Via External Contact: A single string can be transmitted by closing a digital input, not exceeding 1 sec. (see Section 2.10 on page 27 and Section 2.9 on page 24).
3.14 Continuous Transmission Protocol
Using this protocol, the instrument transmits, in continuous mode, the weight to remote displays; the communication string is transmitted 10 times per second. Following communication modes are available (see Section 2.9 on page 24):
rL P: remote display shows the net or gross weight, depending on the remote display setting.
HdrL P: remote display shows the net or gross weight, depending on the remote display setting.
HdrL Pn:
See following page for more information.
The instrument sends the following string to the remote display:
&NxxxxxxLyyyyyy\ckckCR
in which:
& = 1 character of string start (38 ASCII).
N = 1 reference character for net weight. (78 ASCII).
Operation 57
Page 62
xxxxxx = 6 ASCII characters for net or peak weight if present (48 ÷ 57 ASCII).
Laser Light SCT
Connector Pin Connector
J8 (RS-232) 3 RS-485 RX-
5 RS-485 RX+
OR
J9 (RS-485) 1 RS-485 RX+
2 RS-485 RX-
L = 1 reference character for gross weight (76 ASCII).
yyyyyy = 6 ASCII characters for gross weight (48 ? 57 ASCII).
\ = 1 character for separation (92 ASCII).
ckck = 2 ASCII control characters calculated considering that the characters between “&” and “\” are
excluded. The control value is obtained by carrying out the XOR (or exclusive) operation for the 8 bit ASCII
codes of the characters considered. character expressed in hexadecimal is thus obtained, with 2 digits which
may acquire values from “0” to “9” and from ”A” to “F”. “ckck” is the ASCII code of the two hexadecimal
digits.
CR = 1 character for string end (13 ASCII).
In case of negative weight, the first character on the left acquires the value « - » (minus sign - ASCII 45). If the protocol on Hdrl P has been set, the decimal point at the position shown on the instrument's display can also
be transmitted. In this case, if the value exceeds 5 digits, only the 5 most significant digits are transmitted, while if the value is negative, no more than the 4 most significant digits are transmitted. In both cases, however, the decimal point shifts consistently with the value to display.
If Hdrl Pn has been set, in addition to what stated in Hdrl P protocol, the instrument transmits the prompt net every 4 seconds in the gross weight field, if the instrument is in net mode. (see Section 3.1 “Semi-automatic Tare (Net/ Gross)” on page 33).
In case weight value is under -99999, the minus sign (‘-‘) is sent alternated with the most significant figure.
In case of error or alarm, the 6 characters of the gross and net weight are replaced by the messages found in Table 3-1 on page 34.
3.15 Interface to Remote Display
Remote Display (Laser Light)
1. Press and hold , then press to enter set-up menu, CALIb is displayed.
2. Press or until SErIAL is displayed, press . rS485 is displayed.
3. Press , nOnE is displayed.
4. Press or until COntIn is displayed, press .
5. Press or until NOd td is displayed.
6. Press or until desired protocol is displayed, press .
7. Press twice to exit set-up menu.
Laser Light Setup
E CHAR = CR
LWPOS = 7
LENGTH = 19
58 SCT Weight Transmitter Operator’s Manual
Page 63
3.16 Communication Examples
The numerical data below are expressed in hexadecimal notation with prefix h.
Example 1
Command for multiple writing of registers (hexadecimal command 16, h10):
Assuming that we wish to write the value 0 to the register 40017 and the value 2000 to the register 40018, the string to generate must be:
h01 h10 h00 h10
The instrument will respond with the string:
h01 h10
h00 h10 h00 h02 h40 h0D
h00 h02 h04 h00 h00 h07 hD0 hF1 h0F
Query Field Name Hex Response Field Name Hex
Instrument Address h01 Instrument Address h01
Function h10 Function h10
Address of the first register H h00 Address of the first register H
Address of the first register L h10 Address of the first register L
Number of registers to send H
Number of registers to send L
Byte Count h04 CRC16 H
Datum 1 H
Datum 1 L
Datum 2 H
Datum 2 L
CRC16 H hF1
CRC16 L h0F
h00 Number of registers H h00
h02 Number of registers L h02
h00 CRC16 L h0D
h00
h07
hD0
h00
h10
h40
Operation 59
Page 64
Example 2
Command for multiple writing of registers (hexadecimal command 16, h10):
Assuming that we wish to write the two setpoint values on the instrument, at 2000 and 3000 respectively, the string must be sent:
h01 h10
hB0 hA2
The instrument will respond with the string:
h01 h10 h00 h10 h00 h04 hC0 h0F
h00 h10 h00 h04 h08 h00 h00 h07 hD0 h00 h00 h0B hB8
Query Field Name Hex Response Field Name Hex
Instrument Address h01 Instrument Address h01
Function h10 Function h10
Address of the first register H
Address of the first register L h10 Address of the first register L h10
Number of registers to send H h00 Number of registers H h00
Number of registers to send L h04 Number of registers L h04
Byte Count h08 CRC16 H
Datum 1 H
Datum 1 L
Datum 2 H
Datum 2 L
Datum 3 H
Datum 3 L
Datum 4 H
Datum 4 L
CRC16 H hB0
CRC16 L hA2
h00 Address of the first register H h00
hC0
h00 CRC16 L h0F
h00
h07
hD0
h00
h00
h0B
hB8
60 SCT Weight Transmitter Operator’s Manual
Page 65
Example 3
Multiple command reading for registers (hexadecimal command 3, h03):
Assuming that we wish to read the two gross weight values (in the example 4000) and net weight values (in the example 3000), reading from address 40008 to address 40011 must be performed by sending the following string:
H01 h03
The instrument will respond with the string:
H01 h03 h08 h00 h00 hF hA0 h00 h00 h0B hB8 h12 h73
h00 h07 h00 h04 hF5 hC8
Query Field Name Hex Response Field Name Hex
Instrument Address h01 Instrument Address h01
Function h03 Function h03
Address of the first register H
Address of the first register L
Number of registers to send H h00 Datum 1 H h00
Number of registers to send L h04 Datum 1 L h00
CRC16 H hF5 Datum 2 H h0F
CRC16 L hC8 Datum 2 L hA0
h00 Address of the first register H h08
h07 Address of the first register L h00
Datum 3 H
Datum 3 L
Datum 4 H
Datum 4 L
CRC16 H h12
CRC16 L h73
h00
h00
h0B
hB0
For additional examples regarding the generation of correct control characters (CRC16) refer to the manual
Modicon PI-MBUS-300.
Operation 61
Page 66
SCT Weight Transmitter Limited Warranty
Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for one year.
RLWS warrants that the equipment sold hereunder will conform to the current written specifications authorized by RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipment fails to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within the warranty period subject to the following conditions:
Upon discovery by Buyer of such nonconformity, RLWS will be given prompt written notice with a detailed explanation of the alleged deficiencies.
Individual electronic components returned to RLWS for warranty purposes must be packaged to prevent electrostatic discharge (ESD) damage in shipment. Packaging requirements are listed in a publication, Protecting Your Components From Static Damage in Shipment, available from RLWS Equipment Return Department.
Examination of such equipment by RLWS confirms that the nonconformity actually exists, and was not caused by accident, misuse, neglect, alteration, improper installation, improper repair or improper testing; RLWS shall be the sole judge of all alleged non-conformities.
Such equipment has not been modified, altered, or changed by any person other than RLWS or its duly authorized repair agents.
RLWS will have a reasonable time to repair or replace the defective equipment. Buyer is responsible for shipping charges both ways.
In no event will RLWS be responsible for travel time or on-location repairs, including assembly or disassembly of equipment, nor will RLWS be liable for the cost of any repairs made by others.
These warranties exclude all other warranties, expressed or implied, including without limitation warranties of merchantability or fitness for a particular purpose. Neither RLWS nor distributor will, in any event, be liable for incidental or consequential damages.
RLWS and buyer agree that RLWS’ sole and exclusive liability hereunder is limited to repair or replacement of such goods. In accepting this warranty, the buyer waives any and all other claims to warranty.
Should the seller be other than RLWS, the buyer agrees to look only to the seller for warranty claims. No terms, conditions, understanding, or agreements purporting to modify the terms of this warranty shall have any
legal effect unless made in writing and signed by a corporate officer of RLWS and the Buyer.
© Rice Lake Weighing Systems, Inc. Rice Lake, WI USA. All Rights Reserved.
RICE LAKE WEIGHING SYSTEMS • 230 WEST COLEMAN STREET
RICE LAKE, WISCONSIN 54868 • USA
62 SCT Weight Transmitter Operator’s Manual
Page 67
For More Information
Web Site
Frequently Asked Questions (FAQs) at
http://www.ricelake.com/faqs
Contact Information
Hours of Operation
Knowledgeable customer service representatives are available:
6:30 a.m. - 6:30 p.m. Monday through Friday
8 a.m. to 12 noon on Saturday. (CST)
Telephone
Sales/Technical Support 800-472-6703
Canadian and Mexican Customers 800-321-6703
International 715-234-9171
Immediate/Emergency Service
For immediate assistance call toll-free 1-800-472-6703 (Canadian and Mexican customers please call 1-800-321-
6703). If you are calling after standard business hours and have an urgent scale outage or emergency, press 1 to reach on-call personnel.
Fax
Fax Number 715-234-6967
E-mail US sales and product information at
prodinfo@ricelake.com
International (non-US) sales and product information at
intlsales@ricelake.com
Mailing Address
Rice Lake Weighing Systems 230 West Coleman Street Rice Lake, WI 54868 USA
For More Information 63
Page 68
64 SCT Weight Transmitter Operator’s Manual
Page 69
Page 70
230 W. Coleman St. • Rice Lake, WI 54868 • USA
U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0) 88 2349171
www.ricelake.com www.ricelake.mx www.ricelake.eu www.ricelake.co.in m.ricelake.com
© Rice Lake Weighing Systems March 12, 2014 PN 131128 Rev C
Loading...