Technical training seminars are available through Rice Lake Weighing Systems.
Course descriptions and dates can be viewed at www.ricelake.com/training
or obtained by calling 715-234-9171 and asking for the training department.
Rice Lake Weighing Systems is an ISO 9001 registered company.
Specifications subject to change without notice.
Version 2 12/2011
Page 4
Rice Lake continually offers web-based video training on a growing selection
of product-related topics at no cost. Visit www.ricelake.com/webinars.
II
Page 5
About This Manual
This manual contains operating procedures for the Versa-portion RLP-S/SM multi-purpose bench scale.
Caution
Procedures that require working inside
the scale base are to be performed by
qualified service personnel only.
This manual can be viewed or
downloaded from the RLWS website at
www.ricelake.com/manuals.
1.0Introduction
The Versa-portion RLP-S/SM is a multi-purpose bench scale that offers practical solutions for a full range of
applications. Its internal resolution of 1/90,000 gives you maximum precision and accuracy. Its backlit LCD
display enables operators to easily see weights. When portability is required, choose the optional rechargeable
battery for over 60 hours of continuous usage. The Versa-portion RLP-S/SM is designed to reliably operate in
many environments and withstand transport from one area of the plant to another or from one business to the
next.
1.1Front Panel
Table 1-1 lists the indications and front-panel buttons available on the Versa-portion RLP-S/SM.
IndicationDescription
On when the weight display is in ounces
*(not available in Versa-portion RLP-SM)
On when the weight display is in pounds
*(not available in Versa-portion RLP-SM)
On when the weight display is in grams
*
*
*
(kg)
*(Versa-portion RLP-SM value is kg, display is in kilograms)
On when the zero point is adjusted and the weight is stable
On when tare subtraction is performed
On (blinking) when the battery is weak and needs to be recharged
Front-Panel ButtonDescription
Resets the weight display to zero
Sets or clears the tare value
Turns the display on/off
Changes measurement units between lb, oz, g, kg
(not available in Versa-portion RLP-SM)
Table 1-1. Versa-portion RLP-S/SM Indications and Front-Panel Buttons
Introduction1
Page 6
2.0Operation
2.1On/Off
1. Press the POWER key. The display will blink
8.8.8.8.8.8. and show 0.000 when ready to
weigh.
2.2Tare
Tare values can be used for up to half of scale
capacity.
Performing a Tare
1. Place the item to be tared (typically a
container) on the platter.
2. Press the
3. The display will show
TARE key.
0.000 with the tared
item on the platter.
Removing a Stored Tare
1. Remove the tared item from the platter.
2. Press the
3. The display will show
TARE key.
0.000.
2.3Automatic Shut-Off Timer
The Versa-portion RLP-S/SM has a built-in automatic
shut-off timer. This timer can be set to turn the unit off
after three minutes of inactivity or 10 minutes of
inactivity.
Setting the Timer for Three Minutes
1. If the unit is powered on, press the POWER
key until it shuts down.
2. With the unit powered off, press and hold the
POWER key for three seconds.
The display will show 3.
3. Release the
to turn the unit off after three minutes of
inactivity.
Setting the Timer for 10 Minutes
1. If the unit is powered on, press the POWER
key until it powers off.
2. With the unit powered off, press and hold the
POWER key for six seconds.
The display will show 3 after a period of three
seconds. It will show 10 after six seconds.
3. Once the display shows
POWER key. The timer is now set to turn the
unit off after 10 minutes of inactivity.
POWER key. The timer is now set
10, release the
Disabling the Timer
1. If the unit is powered on, press the POWER
key until it shuts down.
2. With the unit powered off, press and hold the
POWER key for nine seconds.
The display will show 3 after a period of three
seconds and 6 after six seconds. It will show
999 after nine seconds.
3. Once the display shows
POWER key. The timer is now disabled.
999, release the
2.4Display Brightness
The Versa-portion RLP-S/SM display brightness can
be set to one of three levels.
Changing the brightness setting
1. Hold the ZERO key and press Tare three times.
2. Press the
ZERO key to increase the dimmer
level and the Power button to decrease the
dimmer level.
3. The display will show
1, 2, 3, or 4 depending
on which dimmer level is being displayed.
4. Press the
TARE key to save the setting and
return to weigh mode.
2.5Default Primary Unit
The Units bu t ton c ycl e s thr o ug h g , lb, a n d oz.
(Versa-portion RLP-SM is only available in kg)
Setting the Default Primary Unit
1. Hold the ZERO key and press the UNITS key
three times.
2. Press the
POWER key to cycle through
available units.
3. When the desired default unit is displayed,
s the TARE key to save the setting and
pres
return to weigh mode.
2.6G, lb and oz Conversion
(Not available in Versa-portion RLP-SM)
G, lb and oz conversion is available only when
SPEC08=011 (see Section 3.4).
1. Switch between units by pressing the
key.
(G) uppercase is the abbreviation for gross.
Note
(g) lowercase is the correct abbreviation for
grams.
UNITS
2Versa-portion RLP-S/SM Operation Manual
Page 7
3.0Calibration
To enter calibration mode, remove the screw on the bottom of the scale as shown in Figure 3-1. Use a small tool
to press the span setup switch located under the screw. Setup will time out to verify the switch is set to the
position. Then, perform a span switch state check (see Section 3.1).
Remove Screw
ON
Figure 3-1. Setup switch
3.1Software Version and Span State Check
1. With the unit in weigh mode, press and hold the ZERO key.
2. With the
ZERO key held, press and release the POWER key three times.
3. The display will show the software version for two seconds, then show the current span switch state
S-ON or S-OFF).
(
After three seconds, the unit will return to weigh mode.
3.2Performing a Calibration
The Versa-portion RLP-S/SM must be calibrated in grams or kilograms, and SPEC 01 and 02 must be set for the
gram value (or kg in the SM version). See Section 3.4.
1. With the unit in weigh mode, press and hold the
2. With the
ZERO key held, press and release the POWER key twice, then press and release the TARE key.
The display will show CAL00.
Note
To exit calibration at any time, press the TARE key.
3. Ensure there is no weight on the platter, then press the ZERO key to calibrate the zero point.
The display will show dashes, then CALSP.
4. Place the scale’s full capacity weight on the platter and press the
The display will show dashes, then the weight value.
ZERO key.
ZERO key to calibrate span.
Note
Calibration weight is full grams capacity. Full capacity values are as follows: RLP-6S = 3000 g (3 kg);
RLP-15S = 6000 g (6 kg); RLP-30S = 15000 g (15 kg); RLP-60S = 30000 g (30 kg).
Calibration3
Page 8
5. The scale will return to weigh mode automatically. Power down the unit.
6. Calibration is complete.
3.3Enclosure Security
After an NTEP inspector has examined the unit, he/she will install security cables through three bolts on the
underside of the scale. These cables prevent the unit from being tampered with by an unauthorized individual. If
the cable is removed, NTEP certification will become void.
Figure 3-2. Legal for trade seal; cable passes through three bolts
3.4Setting the SPEC Value
The SPEC value is used to specify decimal point position, minimum display, net/gross multi-interval, resolution,
weight single or multi-interval, decimal point figure, negative weight display mask, IR mode protection (SPAN
switch must be turned on to allow calibration), start range (amount of weight that can be on the scale when it is
turned on), re-zero range, manual tare cancellation, tare subtraction, tare accumulation, auto tare, weight reset,
tare auto clear, zero tracking, re-zero, and the
1. Press the setup switch (see Figure 3-1).
2. With the unit in weigh mode, press and hold the
3. With the
ZERO key held, press and release the TARE key, then press and release the POWER key twice.
The SPEC number is displayed.
4. To increment the SPEC setting value, press the
5. Press the
6. Press the
7. Press the
SPEC NumberBit 2Bit 1Bit 0
ZERO key to save data and increase the SPEC count.
ZERO key to advance to the next SPEC.
TARE key to store SPECs and return to weigh mode.
Table 3-1. SPEC number and bit explaination (defaults are bold)
4Versa-portion RLP-S/SM Operation Manual
Page 9
SPEC NumberBit 2Bit 1Bit 0
Note
02Resolution Selection
00 - 1/2000
01 - 1/3000
10 - 1/6000
11 - 1/7500
03Decimal point figure for numeral
0 - Period
1 - Comma
04Start range (amount of weight that can be on the scale when it is
turned on; if there is more weight than allowed, the display will show
all 8s)
00 - ±10% F.S. 10 - ±3%F.S.
01 - ±5% F.S. 11 - ±2%F.S.
05
Not NTEP
Approved
06Auto tare clear when rezero
07Zero tracking when Tare
08Unit key (Not used on Versa-portion RLP-SM)
Manual tare cancellation
0 - Allow
1 - Prohibit
0 - Allow
1 - Prohibit
0 - Allow
1 - Prohibit
000 - No
001 - For g and lb conversion
010 - Not used
011 - G, lb and oz
100 - 111 Not used
Negative weight display mask
0 - Minus gross > 9e
1 - Minus gross weight
Tare subtraction
0 - Allow
1 - Prohibit
Weight reset when tare
0 - Allow
1 - Prohibit
Re-zero function
0 - Allow
1 - Prohibit
Weight single interval or multi-interval
0 - Single interval
IR (inter resolution) mode protected by
SPAN SW (SPAN switch must be
turned on for calibration)
0 - No
1 - Yes
Re-zero range
0 - ±2% F.S.
1 - ±3% F.S.
Tare accumulation
0 - Allow
1 - Prohibit
Tare auto clear
(Greater or equal to Gross 21e and
greater or equal toNet 5e)
0 - Allow
1 - Prohibit
Additional zero for weight when only
fixed zeros are displayed (V1.02 and
above)
0 - Allow (NTEP)
1 - Prohibit
Table 3-1. SPEC number and bit explaination (defaults are bold) (Continued)
3.5Internal Count and A/D Count Display
1. With the unit in weigh mode, press and hold the ZERO key.
2. With the
The display will show
3. Press the
4. To return to weigh mode, press the
ZERO key held, press and release the TA RE key twice, then press and release the POWER key.
,,,,,0,.
POWER key to display the A/D count.
TARE key.
(G) uppercase is the abbreviation for gross.
(g) lowercase is the correct abbreviation for grams.
Calibration5
Page 10
4.0Appendix
4.1DC Battery Pack
4.1.1Battery Pack Replacement/Installation
An optional DC battery pack (PN 108516) for the Versa-portion RLP-S/SM is available and can be purchased
from RLWS to ship with the scale or retrofit in the field. The rechargeable 6V 5.0 AH battery pack allows for up
to 60 hours of scale use without an AC power supply (with display set to Dimmer Level 2). It is located in the
bottom of the scale base. Use the following procedure to install or replace the battery pack.
1. Unplug the scale from power source.
2. Place scale its left side.
Caution
Do not turn the scale upside down. Always work with the scale on its side. Damage to the load cell can
occur if the scale is turned upside down.
3. Unscrew the Phillips-head screws holding the battery compartment door closed.
4. Unscrew the two screws holding on the battery holding bracket and remove the bracket.
5. If you are removing a battery previously installed, remove the battery from its compartment, then
disconnect the black (-) and red (+) electrical leads from the battery. If you are installing a new battery,
go to step 6.
6. Attach the red lead to the positive (+) terminal of the new DC battery pack, and the black lead to the
negative (-) terminal. Place new DC battery pack in battery compartment.
7. Replace the battery holding bracket and fasten it with its two screws.
8. Replace the battery compartment door and fasten it with the screws removed in Step 3.
9. Put the scale back upright on its feet.
10. Press the
4.1.2Battery Charging
POWER key located on the front of the scale.
A fully charged battery allows for approximately 60 hours of continuous use. When the battery is low, the battery
indicator light will light up. To charge the battery, plug in the AC power cord. Charge time is 8-10 hours.
4.2
Error Messages
DisplayDescriptionSolution
8 8 8 8 8 8The scale is not steady.Place the scale on a firm, flat base.
O FThe weight exceeds capacity + 9d, or something is on the platter
when it was powered on.
U FThe negative weight exceeds the display limit.Re-zero or cycle power.
E r r o rAn error has occured in maintenance mode.Repeat the operation.
E R R 01There is an A/D error.Contact the scale dealer or RLWS.
E R R 02There is a data flash erase error.Contact the scale dealer or RLWS.
E R R 03There is a data flash program error.Contact the scale dealer or RLWS.
Remove any items from the platter and cycle power.
6Versa-portion RLP-S/SM Operation Manual
Table 4-1. Error Messages
Page 11
5.0Specifications
t
/
"
5
*
0
/
"
-
$
0
/
'
&
3
&
/
$
&
t
0
/
8
&
*
(
)
5
4
"
/
%
.
&
"
4
6
3
&
4
Measurement
Canada
Approved
Caution
Display
LED, 6 digits
Capacity
RLP-6S6.000 lb x 0.002 lb; 3000 g x 1 g;
60.00 oz x 0.02 oz
RLP-15S15.000 lb x 0.005 lb; 6000 g x 2 g;
150.00 oz x 0.05 oz
RLP-30S30.00 lb x 0.01 lb; 15000 g x 5 g ;
300.0 oz x 0.1 oz
RLP-60S60.00 lb x 0.02 lb; 30000 g x 10 g ;
600.0 oz x 0.2 oz
Versa- portion RLP- SM
RLP-3SM 3 kg x 1 g
RLP-6SM6 kg x 2 g
RLP-15SM 15 kg x 5 g
RLP-30SM 30 kg x 10 g
Display Indicators
Ounces, Grams, Pounds, NET, Center of Zero, Battery Usage
18W when using AC power
3W when using rechargeable battery
Battery life: 60 hours of continuous usage
Power Source
DC 12V/1.5A, DC 6V5Ah rechargeable battery (Optional)
Dimensions
251.6 mm x 289.7 mm x 134.6 mm
Package Weight
6.3 kg
Package Size
377 mm x 337 mm x 245 mm
Rating
IP68
Do not use any parts on this scale that
may contain lead.
Specifications7
Page 12
Versa-portion RLP-S/SM Limited Warranty
Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a
Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by
the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in
materials and workmanship for one year.
RLWS warrants that the equipment sold hereunder will conform to the current written specifications authorized
by RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipment
fails to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within the
warranty period subject to the following conditions:
•Upon discovery by Buyer of such nonconformity, RLWS will be given prompt written notice with a
detailed explanation of the alleged deficiencies.
•Individual electronic components returned to RLWS for warranty purposes must be packaged to
prevent
electrostatic
publication, Protecting Your Components From Static Damage in Shipment, available from RLWS
Equipment Return Department.
•Examination of such equipment by RLWS confirms that the nonconformity actually exists, and was
not
caused
by accident, misuse, neglect, alteration, improper installation, improper repair or
improper testing; RLWS shall be the sole judge of all alleged non-conformities.
•Such equipment has not been modified, altered, or changed by any person other than RLWS or its
duly authorized repair agents.
•RLWS will have a reasonable time to repair or replace the defective equipment. Buyer is responsible
for shipping charges both ways.
•In no event will RLWS be responsible for travel time or on-location repairs, including assembly or
disassembly of equipment, nor will RLWS be liable for the cost of any repairs made by others.
discharge (ESD) damage in shipment. Packaging requirements are listed in a
HESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WITHOUT
T
LIMITATION WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
. NEITHER
RLWS NOR DISTRIBUTOR WILL, IN ANY EVENT, BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
RLWS
REPAIR OR REPLACEMENT OF SUCH GOODS
ALL OTHER CLAIMS TO WARRANTY
S
WARRANTY CLAIMS
N
WARRANTY SHALL HAVE ANY LEGAL EFFECT UNLESS MADE IN WRITING AND SIGNED BY A CORPORATE
OFFICER OF
AND BUYER AGREE THAT RLWS’S SOLE AND EXCLUSIVE LIABILITY HEREUNDER IS LIMITED TO
. IN ACCEPTING THIS WARRANTY, THE BUYER WAIVES ANY AND
.
HOULD THE SELLER BE OTHER THAN RLWS, THE BUYER AGREES TO LOOK ONLY TO THE SELLER FOR
.
O TERMS, CONDITIONS, UNDERSTANDING, OR AGREEMENTS PURPORTING TO MODIFY THE TERMS OF THIS