This manual contains installation, operation and maintenance instructions for the Valor™ 7000
Series. Please read the manual completely before using the scale.
1.1 Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and
warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument,
malfunctions and false results.
Signal Words
CAUTION for a hazardous situation with low risk, resulting in damage to the device or
the property or in loss of data, or injuries if not avoided.
Note (No symbol)
For useful information about the product
Warning Symbols
Attention Symbol
Alternating Current
1.2 Safety Precautions
Please follow these safety precautions:
Verify that the AC input voltage printed on the data label matches the local AC power
supply.
Do not drop loads on the platform.
Make sure that the power cord does not pose a potential obstacle or tripping hazard.
Do not position the scale in a way that makes it difficult to pull the power plug.
Use only approved accessories and peripherals.
Operate the scale only under ambient conditions specified in these instructions.
Disconnect the scale from the power supply when cleaning.
Do not operate the scale in hazardous or unstable environments.
Do not immerse the scale in water or other liquids.
Do not place the scale upside down on the platform.
Only use weights within the scale’s capacity as specified in these instructions.
Service should be performed only by authorized personnel.
2. INSTALLATION
2.1 Package Contents
EN-2 Valor™ 7000 Series
Level Indicator
Adjustable feet
2.2 Installing Components
Install the sub-platform and metal pan as shown below. Press to lock the sub-platform into
place.
Figure 2-1. Install the sub-platform and metal pan
2.3 Selecting the Location
Use the scale on a firm, steady surface. Avoid locations with excessive air current, vibrations,
heat sources, or rapid temperature changes. Allow sufficient space around the scale.
2.4 Leveling the Equipment
The Valor Series has a level indicator as a reminder that the scale should be leveled for
accurate weighing. To level the scale, adjust the feet so the bubble is centered in the circle.
Be sure the equipment is level each time its location is changed.
Figure 2-2. Level indicator
Valor™ 7000
Figure 2-3A Connect the AC
power plug to the input jack on
the back of scale.
Figure 2-3B Connect AC power plug to
the proper AC supply.
Symbol
Charge level
0 to 10 % Remaining
11 to 40 % Remaining
41 to 70 % Remaining
71 to 100 % Remaining
Series EN-3
2.5 Connecting Power
AC power is used to power the scale when battery power is not needed. First, connect the AC
power cord (supplied) to the power input jack then connect the AC plug to an electrical outlet.
2.5.1 Battery Power:
The scale can be used on AC power immediately. Allow the battery to charge for 12 hours
before using the scale on battery power. The Scale will automatically switch to battery
operation if there is a power failure or the power cord is removed. With AC power, the scale
is constantly charging, so the battery charge indicator (see item 11 in figure 3-2) will remain
lit. The scale can be operated during charging, and the battery is protected against
overcharging.
For maximum operating time, the battery should be charged at room temperature.
During battery operation, the battery symbol indicates the battery charge level remaining.
The indicator will automatically turn off when the batteries are fully charged.
TABLE 2-1
Notes:
When battery symbol blinks fast, approximately 30 minutes working time is left.
When [lo.bat] is displayed, the scale will shut off.
Charging the scale must be performed in a dry environment.
CAUTION: Battery is to be replaced only by an authorized Ohaus service
dealer. Risk of explosion can occur if the rechargeable battery is replaced with
the wrong type or if it is not properly connected. Dispose of the lead acid
battery according to local laws and regulations.
EN-4 Valor™ 7000 Series
Button
Functions
On/Zero
Off
3
Yes
Short Press1 (when on): Sets display to zero
(when off): Turns scale on
Long Press2 (when on): Turns the scale off
Short Press (in Menu): Selects/accepts displayed setting
Print
Units
No
Short Press: See Interface Manual for operation description.
Long Press: Toggles through active weighing units
Short Press (in Menu): Toggles through available settings
Function
Mode
Back
Short Press: Selects function setting
Long Press: Selects active Mode
Short Press (in Menu): returns to previous settings
M+
Menu
Exit
Short Press: Accumulates the weight and displays the accumulated
information at 0 load.
Long Press: Enters User Menu
Short Press (in Menu): Quickly exit User Menu
Tare
Short Press: Enter/clear a tare value. Clears the accumulation when
the accumulation information is displayed.
The IR Sensor can be programmed to act as “touchless” buttons.
See section 4.3 for the available settings.
3. OPERATION
3.1 Controls
Figure 3-1. Valor 7000 Front Control Panel with LCD Display.
TABLE 3-1
Notes: 1 Short Press: Press less than 1 seconds.
2
Long Press: Press and hold for more than 2 seconds.
3
Press and hold the Off button until OFF is displayed, then release the button.
Valor™ 7000
IR Sensor
(back)
Item
Description
Item
Description
1
NET symbol
9
Calibration / Menu Mode symbol
2
Center of Zero symbol
10
Dynamic symbol (not used)
3
Negative symbol
11
Battery charge symbol
4
Stable weight symbol
12
Pound, Ounce, Pound:Ounce symbols
5
Preset Tare, Tare symbols
13
Percent symbol
6
Pointer symbols
14
Kilogram, gram symbols
7
IR Sensor symbol
15
Pieces symbol, tonne symbol (both not used)
8
Accumulation symbol
16
Scale symbol (not used)
(Red) Loads > Upper limit
(Green) Loads ≥ Lower limit and ≤ Upper limit
(Yellow) Loads < Lower limit
Figure 3-2. LCD Display
1
2
1
3
4
5
6
8
9
10
11
12
13
14
15
16
LED Indicators
Series EN-5
TABLE 3-2
The colored LED indicators on the left side of the control panel are used in Check mode (section
3.5) and will light up according to the following rules:
Figure 3-3. Valor 7000 back LCD Display
EN-6 Valor™ 7000 Series
Power Input
Weigh Below
Access
Lock Switch
IR Sensor
(front)
RS232 and
Option
Interface Port
(below cover)
IR Sensor (front)
Figure 3-4. Below view of Valor 7000
IR Sensors
The Valor 7000 is equipped with two IR sensors, one at the front (see figure 3-5) and one at the
rear (see figure 3-3). The front and rear sensors perform the same function. The sensors can be
activated by a hand or other object that is placed at a specified distance in front of the sensor
locations. The sensor activation distance will vary based on the reflective nature of the object. If
unwanted activations occur due to unique situations the sensor can be turned off. See section
4.3 Setup Menu for available settings.
Figure 3-5. Front view of Valor 7000
Valor™ 7000
Series EN-7
3.2 Turning Scale On/Off
To turn the scale on, press and hold the On/Zero Off button for 1 second. The scale performs a
display test, momentarily displays the software version, and then enters the active weighing
mode.
To turn the scale off, press and hold the On/Zero Off button until OFF is displayed.
3.3 Weigh Mode
This mode is the factory default setting.
1. If needed, press and hold Mode until [wJeIGH] (Weigh) is displayed.
2. If required, place an empty container on the pan and press Tare.
3. Add sample to the pan or container. The display shows the weight of the sample.
3.4 Percent Mode
This mode measures the weight of a sample as a percentage of a reference weight.
1. If required place an empty container on the pan and press Tare.
2. Press and hold Mode until [Percnt] is displayed. [Clr.ref] (clear reference) will then
display. If no reference weight exists, the scale will display [Put.ref], proceed to step 5.
3. Press No to use the stored reference weight and proceed to step 6.
4. Press Yes to establish a new reference. Scale will now display [Put.ref].
5. Add the desired reference material to the pan or container. Press Yes to store the
reference weight. The display shows 100%.
6. Replace the reference material with the sample material. The display shows the
percentage of the sample compared to reference weight.
7. To clear the stored reference press and hold Mode until [Percnt] is displayed. Press Yes
when [Clr.ref] is displayed.
Note: Press Function to view the current reference weight.
EN-8 Valor™ 7000 Series
3.5 Check Mode
Use this mode to compare the Weight or Percent of items to a target range. The scale supports
positive, negative and zero check weighing.
Refer to section 4.5 to set the desired Check sub-mode.
3.5.1 Check Weighing
Set Checkweigh to Weight in the Mode menu. Use this mode to compare the weight of items to
a target weight range.
1. Press and hold Mode until [CHeCk] (Check) is displayed. [CL.lImM] (clear check limits)
will then display.
2. Press No to use the stored check limits and proceed to step 5.
3. Press Yes to establish new check limits. The scale will then display [Set. Lo]. Press Yesto view the “Low” limit value. Press Yes to accept or Noto edit the “Low” limit value. The
stored value then displays with the first digit highlighted [000.000 kg]. Repeatedly press
No until the desired number appears. Press Yes to accept and highlight the next digit.
Repeat until all the digits are correct. Press Yesto accept the “low” limit value, [Set. Hi]
will be displayed.
4. Repeat the same procedure to accept or edit the “high” value.
5. If required, place an empty container on the pan and press Tare.
6. Place sample material on the pan or in the container. If the sample weight is under the
target weight range, the yellow LED will light. If the sample is within the target weight
range, the green LED will light. If the sample is over the target weight range, the red LED
will light.
Note: Press Function to view the low and high check limits.
3.5.2 Check Percent
Set Checkweigh to Percent in the Mode menu. Use this mode to compare the percentage of
items to a target percent range.
1. Press and hold Mode until [CHeCk] (Check) is displayed. [CLr.ref] (clear reference) will
then display.
If no reference weight exists, the scale will display [Put.ref], proceed to step 4.
2. Press No to use the stored reference weight and proceed to step 5.
3. Press Yes to establish a new reference. Scale will now display [Put.ref].
4. Add the desired reference material to the pan or container. Press Yes to store the
reference weight.
5. The scale will display [CL.lImM] (clear Check limits).
6. Press No to use the stored check limits and proceed to step 9.
7. Press Yes to establish new check limits. The scale will then display [Set. Lo]. Press Yesto view the “Low” limit value. Press Yes to accept or Noto edit the “Low” limit value. The
stored value then displays with the first digit highlighted [000.000 %]. Repeatedly press
No until the desired number appears. Press Yes to accept and highlight the next digit.
Repeat until all the digits are correct. Press Yesto accept the “low” limit value, [Set. Hi]
will be displayed.
8. Repeat the same procedure to accept or edit the “high” value.
Valor™ 7000
9. If required, place an empty container on the pan and press Tare.
10. Place sample material on the pan or in the container. If the sample weight is under the
target percentage range, the yellow LED will light. If the sample is within the target
percentage range, the green LED will light. If the sample is over the target percentage
range, the red LED will light.
Note: Press Function to view the low and high check limits.
Positive Check
Positive check is used to determine when the material added to the scale is within the target
range. In this case the UNDER and OVER limits must be positive values. (The OVER limit must
be greater than the UNDER limit.)
Add material to the scale until it is within the ACCEPT (green) range.
Negative Check
Negative check is used to determine when the material removed from the scale is within the
target range. In this case the UNDER and OVER limits are both negative values.
The UNDER limit must be greater than the OVER limit (for example: UNDER= -10/OVER= -15).
Place the item to be weighed on the scale and press Tare.
Remove a portion of the item until it is within the ACCEPT range.
Zero Check
Zero check is used when comparing subsequent samples to an initial reference sample. In this
case, the UNDER limit must be a negative value and the OVER limit must be a positive value.
Place the reference item on the scale and press Tare. Remove the reference sample and place
the item to be compared on the scale to determine if it is within the ACCEPT range.
Series EN-9
EN-10 Valor™ 7000 Series
In weighing mode; sub menu ACCUmM setting, select mMaNU:
If required, place an empty container on the pan and press Tare.
Step i: Put the weight (0.04 kg) on the pan and press M+ key. The Σ
pointer indicator will keep flashing until the weight is removed.
Step ii: Remove the weight from the pan. Put another
weight (0.03 kg) and press M+ key:
Step iii: Remove the weight from the pan:
Step iv: Press M+ key to display the statistical information:
Step v: If required, follow section 3.6.2 to clear the accumulation data.
mMaNU
*
0.040 kg
*
0.030 kg
*
0.000 kg
N 2
tOtal
*
0.070 kg
mMIN
0.030 kg
0.040 kg
3.6 Accumulation and Statistics
The Accumulation feature enables manual or automatic totalizing of displayed values. Statistical
data (total accumulated, min and max weight or %, and total number of weighments) is stored in
memory for review and printing. Accumulation works together with each application mode
except Dynamic.
3.6.1 Accumulating Displayed Values
With ACCUMULATE set to MANUAL, place the item on the scale and press the M+ key to add
the weight to accumulation data. The ∑ pointer will keep flashing until the weight is removed.
With ACCUMULATE set to AUTO, place the item on the scale. The displayed value is
accumulated automatically. The ∑ pointer will keep flashing until the weight is removed.
3.6.2 Viewing and Clearing Statistical Data
When the pan is cleared, press the M+ key to display the statistical information. To clear the
accumulation data press the Tare key while the statistical information is displayed. The display
shows [Clr.aCC]. Press the Yes key to clear the stored data and return to current mode.
Notes: The item must be removed from the pan before the next item can be accumulated.
Only stable weights are stored to accumulation data.
Changing the mode will clear the accumulation data.
When Legal for Trade is ON, the display must return to 0 gross. Otherwise, the ∑
pointer will continue flashing. Gross loads and net loads cannot be added to the
same total. If the first load is a gross weight, future loads must also be gross
weights. If the first load is a net weight, future loads must also be net weights.
Example:
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.