Rice Lake Static Weighing User Manual

CW-90/90X
Firmware Version 1.06
Technical Manual
Checkweigher
PN 105942 Rev D

Contents

Technical training seminars are available through Rice Lake Weighing Systems. Course descriptions and dates can be viewed at www.ricelake.com/training or obtained by calling 715-234-9171 and asking for the training department.
1.0 Introduction..................................................................................................................................... 1
2.0 Installation ...................................................................................................................................... 6
2.7.1 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.7.2 Digital I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.10 Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.11 Installing Option Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.12 Replacement Parts and Assembly Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.0 Configuration ................................................................................................................................ 15
3.1 Menu Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2 Audit Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.3 Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4 Setup Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4.1 SCALE Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4.2 FEATUR Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.4.3 FEATUR Menu, REGULA / INDUST Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.4.4 Serial, PFORMT, DIG I/O, VERS Submenus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.5 Test Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.6 Time and Date Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.7 ACCUM Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.8 BRIGHT Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.9 ID Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.0 Calibration .................................................................................................................................... 36
4.1 Front Panel Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2 EDP Command Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.3 Revolution® Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.4 More About Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.4.1 Adjusting Final Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
© Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America.
Specifications subject to change without notice.
Rice Lake Weighing Systems is an ISO 9001 registered company. Version 1.06
December 17 , 2014
Contents 1
5.0 Operation....................................................................................................................................... 39
Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost. Visit www.ricelake.com/webinars.
5.1 Range Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.1.1 Value Set: Keyed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.1.2 Value Set: Push . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.2 Target Weight Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.2.1 Value Set: Keyed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.2.2 Value Set: Push . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.3 Target Percent Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.3.1 Value Set: Keyed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.3.2 Value Set: Push . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.4 IDs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.4.1 Setting an ID. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.4.2 Using a Stored ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.4.3 WeighVault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.5 Negative Checkweighing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.0 Serial Commands ......................................................................................................................... 48
6.1 The Serial Command Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.1.1 Key Press Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.1.2 ID Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.1.3 Reporting Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.1.4 Clear and Reset Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.1.5 Parameter Setting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.1.6 Normal Mode Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.1.7 Unique Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.2 Custom Stream Formatting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.0 Print Formatting ............................................................................................................................ 56
7.1 Print Formatting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.2 Customizing Print Formats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.2.1 Using the EDP Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.2.2 Using the Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.2.3 Using Revolution®. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.0 WLAN Option Card......................................................................................................................... 59
8.1 WLAN Card Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
9.0 Appendix ....................................................................................................................................... 60
9.1 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
9.2 Using the XE EDP Command. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
9.3 Status Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
9.3.1 Using the P EDP Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
9.3.2 Using the ZZ EDP Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
9.4 Continuous Output (Stream) Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
9.5 Digital Filtering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
9.5.1 DIGFLx Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
9.5.2 DFSENS and DFTHRH Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
9.5.3 Setting the Digital Filter Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
9.5.4 Audit Trail Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
9.6 Regulatory Mode Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
9.7 Updating CW-90/90X Firmware. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
9.8 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
CW-90/90X Limited Warranty ................................................................................................................... 70
For More Information ................................................................................................................................ 71
2 CW-90/90X Checkweigher

1.0 Introduction

WARNING
Important
CAUTION
WARNING
WARNING
This manual is intended for use by qualified service technicians responsible for installing and servicing the CW-90/CW-90X checkweighing scale.
Manuals can viewed or downloaded from the Rice Lake Weighing Systems website at
www.ricelake.com/manuals.

1. 1 S af et y

Safety Symbol Definitions:
Indicates a potentially hazardous situation that, if not avoided could result in serious injury or death, and includes hazards that are exposed when guards are removed.
Indicates a potentially hazardous situation
Indicates information about procedures that, if not obser corruption to and loss of data.
that, if not avoided may result in minor or moderate injury.
Safety Precautions
Do not operate or work on this equipment unless you have read and understand the instructions and warnings in this manual. Failure to follow the instructions o death. Contact any Rice Lake Weighing Systems dealer for replacement manuals. Proper care is your responsibility.
Some procedures described in this manual require work inside the indi are to be performed by qualified service personnel only.
General Safety
Failure to heed may result in serious injury of death. DO NOT allow minors (children) or inexper DO NOT operate without all shields and guards in place. DO NOT step on the unit. DO NOT jump on the scale. DO NOT use for purposes other then weight taking. DO NOT place fingers into slots or possible pinch points. DO NOT use any load bearing component that is DO NOT use this product if any of the DO NOT exceed the rated load limit of the unit. DO NOT make alterations or modifications to the unit. DO NOT remove or obscure warning labels. Before opening the unit, ensure the power cord Keep hands, feet and loose clothing away from moving parts.
ienced persons to operate this unit.
worn beyond 5% of the original dimension.
components are cracked.
is disconnected from the outlet.
ved, could result in damage to equipment or
r heed the warnings could result in injury or
cator enclosure. These procedures
Introduction 1

1. 2 O ve rv i ew

Note
Note
Note
The CW-90/CW-90X is a high-speed digital weight indicator and scale base programmed to compare weight readings with predetermined tolerance limits defining an “accept” band.
If the current weight reading is within the acceptable range, the green “ACCEPT” LED is illuminated. If the current weight reading is less than the acceptable range, one or more of the red “UNDER” segments is illuminat
ed. If the current weight reading is greater than the acceptable range, one or more of the yellow “OVER” segments is illuminated. Features include:
L arge .8” (20.3 mm) LED display with “under,” “accept,” and “over” bands
Full numeric keypad (CW-90 only)
Four configurable digital inputs/outputs
Two independent communication ports (2) full duplex, (1) unidirectional active 20
mA current loop

1.3 Operating Modes

The CW-90/CW-90X has two modes of operation:
Normal (Weighing) Mode
Normal mode is the “production” mode o on whether a tare has been entered. LED annunciators indicate the type of weight value.
Menu mode
Most of the procedures described in this manual, including configuration and calibration, require in Menu mode.To enter Menu mode, press the indicator displays “AUDIT.” If the Audit jumper is in the “OFF” position, remove the Insert a screwdriver into the access hole and press the setup switch once to enter Menu mode.
f the indicator. The indicator displays the gross or net weight, depending
the indicator to be
MENU key (if the Audit jumper is in the “ON” position).The
large fillister head screw from the bottom of the enclosure.

1.4 Keypad Functions

Key Function
Turns t h e CW-90/CW-90X unit on/off.
If the PC jumper is set to SW, the POWER button must be used to turn the unit on and off. If the PC jumper is set to ON, the unit will automatically power on only way to turn it off is to unplug power.
Enters Menu mode, allowing configuration if the Audit jumper i key in Menu mode.
Sets the current gross weight to zero, provided the amount of weight to be removed or added is within the specified zero range and the scale is not in motion. The zero band is defaulted to 1.9% of full scale, but can be configured for up to 100% of full scale.
1.9% is required for legal for trade applications.
Switches the weight display to an alter kg, g, lb, oz, or lb/oz. Conversions of the weight reading, the Tare value, the Over value, and the Under value occur when the unit of measure is changed with the Units key.
When working in lb/oz, any entered values need to be entered in ounces. This applies when entering over, under, target, and tare values.
nate unit. The alternate unit is defined in the Setup menu, and could be
when it’s plugged in and the
s in the “ON” position. Also used as an “escape”
Sends “on-demand” serial information out the serial port, provided the conditions for standstill are met. PRINT will be displayed while the unit prints.
Table 1-1. Keypad Functi ons
2 CW-90/90X Checkweigher
OVER
UNDER
TARGET
ID
ENTER
Key Function
Performs one of several predetermined Tare functions dependent on the mode of operation selected in the TAREFN parameter (See Section 3.4.1 on page 18). To view a stored tare, see Section 1.5 on page 3.
Allows the display of the current “over tolerance” value, or allows setting the “over tolerance” value.
Allows the display of the current “under tolerance” value, or allows setting the current “under tolerance” value.
When the MODE parameter is set to TARGWT or TARG% (see Section on page 19), this key is used to acquire a weight value and assign it as th over/under values using the target value and the over/under tolerance settings defined during the weighing operation.
Selects a particular over/under/tare/units register set to be retrieved and used.
Used to accept entries and move down to below parameters in Menu mode.
e desired “target” value. The CW-90/CW-90X then computes the final
CW-90 only. Clear key. Used to backspace on entries
Numeric keypad (see Figure 1-1)
CW-90 only. Used to enter values directly.
Table 1-1. Keypad Functions (Continued)
Figure 1-1. CW-90 Numeric Keypad

1.5 Tare Recall

When a stored tare is displayed, the Gross and Net annunciators will be turned off and the PT annunciator will be turned on. To display a stored tare,
1. Press the
2. Press the The tare value will be displayed for 10 seconds. If there is no tare in the system, pressing the TAR E key will have no effect.
MENU key. TARE key.
Introduction 3

1.6 Annunciators LEDs

0
LED Description
Zero (Center of Zero) LED
While in gross weight display mode, this LED indicates that the current displayed weight reading is within +/-0.25 display divisions of the acquired zero, or is within the center of zero band. When in the net weight display mode, it divisions of the center of net zero. A display division is the resolution of the displayed weight value, or the smallest incremental increase or decrease that can be displayed or printed.
G/B LED
Gross weight mode (or B
N LED
Net weight display mode.
T LED
Indicates that a tare has been acquired and stored by the system.
PT LED
Indicates that a preset tare weight has bee
Displays which unit of measure is being used.
% LED
Percent mode
kg LED
Kilograms
g LED
Grams
lb LED
Pounds
oz LED
Ounces
rutto in OIML mode)
indicates that the current net weight reading is within +/-0.25 display
n keyed in or entered via the EDP serial port.
Table 1-2. Annunciator LEDs
4 CW-90/90X Checkweigher

1.7 Bar Graph LEDs

ACCEPT
UNDER
OVER
Greatly over
Moderately under
Slightly under
Acceptable weight
Moderately over
Slightly over
Greatly under
The bar graph LEDs provide you with a fast way of determining if a container is too heavy (OVER), too li ght
UNDER), or is within an acceptable weight range (ACCEPT).
(
Figure 1-2. Bar Graph LEDs
You can enable the UNDER and OVER lights to act as a graph as illustrated in Figure 1-2, or to appear as one a solid bar regardless to how far over/under a This is done by accessing the Feature menu (see Section 3.4.2 on page 23).
Red Segments
When illuminated, this indicates an underweight value. The leftmost red segments indicate that the container weight is far below the acceptable weight band (greatly under); the rightmost segments indicate that the container weight is almost in the acceptable weight band, but still under (slightly under). Illuminating the center segments indicates middle-ground, or moderately under.
weighment is. They can also be disabled so no illumiation is provided.
condition; the container weighs less than the lowest acceptable
Green Segment
Indicates an acceptable value. When lit, the green segment (one solid bar) indicates that the container
weight is
within the actual acceptable band of weight limits.
Yellow Segments
When illuminated, this indicates an overw
eight condition; the container weighs more than the highest acceptable value. The rightmost yellow segments indicate that the container weight is far above the acceptable weight band (greatly over); the leftmost segments indicate that the container weight is almost in the acceptable weight band, but still over (slightly over). Illuminating the center segments indicates middle-ground, or moderately over.
Introduction 5

2.0 Installation

1
6
9
4
2
5
7
10
Torque Pattern
8
3
Torqued screws may become less tight as the gasket is compressed during torque pattern, therefore a second torque is required using the
same pattern and torque value.
Important
Important
Indicator
Column
Column Support
Lock Washer
Cap Screw 1/4-20NC X 3/4
Foot
Coupling Nut
Cap Screw 1/4-20NC x 1-3/4
Lock Washer
This section contains instructions on unpacking and assembly, leveling, making power connections, load cell wiring, wiring standard serial port, optional network communications, wiring optional digital outputs, optional backup battery operation, board diagrams, and power-up sequence.
Do not pick up the scale by the “spider” assembly which supports the platter. Lifting by the spider may damage the load cell. Lift the scale from under the base removed, align the gasket holes carefully to prevent driving a screw through the gasket and causing a leak. Tighten screws to 15 in/lbs (1.7 N-m) in alternating patterns shown Figure 2-1.
Figure 2-1. Indicator Back Panel Screw Tightening Sequence

2.1 Unpacking and Assembly

The indicator head and support column or stand are shipped detached from the scale platform.
1. Remove all assemblies from the shipping ca load cell cable. This cable is correctly wired to the load cell terminal in the indicator head. Do not pull with excessive force on the connections at either end of this cable.
2. If mounting the head onto a column, remove the platter from the sc
3. Invert the platform to access t
he column mounting holes on the rear and bottom of the platform.
4. Position the column over the four mount in mounting holes on bottom side of the platform. Install (2) 1/4-20 x 1-3/4”cap screws, (2) lock washers, and (2) column supports in mounting holes on rear of platform. Install coupling nuts, counter bored end first, onto threaded stem of rubber feet. Install feet onto bottom side of the column.
5. Turn the CW
-90/CW-90X upright and replace the platter on the platform.
6. Attach indicator to the column with the two knobs and nylon between indicator enclosure and column mounting holes.
rton. Notice that the head and scale platform are joined by a
ing holes. Install (2) 1/4-20 x 3/4” cap screws and (2) lock washers
to move it. If the rear panel of the indicator is
ale platform and set aside.
washers provided. Position nylon washers
6 CW-90/90X Checkweigher
Figure 2-2. Mounting Column to Scale Platform

2.2 Leveling

Note
Note
WARNING
Note
Select a location for the CW-90/CW-90X that is reasonably level and free of vibrations and air currents. Adjust the four corner feet on the base and refer to the bubble level on the inside frame. The base should not rock and the feet should have solid contact with the surface. If using a column, adjust the two column feet until they make solid contact with the support surface.
Ensure the nut on each foot’s bolt is secured flush against the scale base.

2.3 Making Power Connections

The power source used for the CW-90 must be properly grounded to an acceptable earth ground. If the indicator is remotely mounted, the platform must be separately grounded from the chassis ground screw located on the bottom of the platform. Connect this screw with 18 gauge wire to the same earth ground system as the AC power source. Failure to ground the base may cause static buildup and incorrect weights.
The CW-90 must be installed near an easily accessible power outlet to allow for quick disconnect in case of emergency.

2.4 Enclosure Disassembly

The indicator enclosure must be opened to connect cables for load cells, communications, and digital inputs/outputs.
Before opening the unit, ensure the power cord is disconnected from the power outlet. The power outlet must be located near the indicator to allow the operator to
Ensure power to the indicator is disconnected, then place the indicator face-down on an antistatic mat. Remove the screws holding the backplate to the enclosure body, then lift the backplate away from the enclosure and set it aside.
easily disconnect power to the unit.

2.5 Load Cell Wiring

On all complete units (indicator head and base), the load cell has been wired to the indicator’s CPU load cell terminal at the factory. If you have purchased just the indicator head, you must wire the load cell. Wire the cable to the CPU board’s J1 connector as shown in Table 2-1. Leave any excess cable outside of the indicator head during installation
If using a 4-wire connection, set JP1 and JP2 to ON. If using a 6-wire connection, set JP1 and JP2 to OFF.
J1 Pin Function
1 + Signal
2 - Signal
3 + Sense
4 - Sense
5 + Excitation
6 - Excitation
Table 2-1. J1 Pin Assignments

2.6 Wire Specifics

Wires connecting to J1, J2, J3, or J4 should adhere to the following specifications:
Wire Range Wire Strip Length
28~12 AWG stranded or solid wire 5~6 mm (3/16” ~ 1/4”)
Table 2-2. Wire Specifications for Connectors
Installation 7

2.7 Cable Grounding

Insulated cable
Foil (silver side out)
Grounding clamp
Shield wire (cut)
Length of foil before folding back on cable insulation
Cut insulation here for foil-shielded cables
Braid
Cut insulation here for braided cables
NOTE: Install lockwashers first, against backplate, under grounding clamp
Figure 2-3. Grounding Clamp Attachment for
Foil-Shielded and Braided Cabling
Except for the power cord, all cables routed through the cord grips should be grounded against the indicator enclosure. Do the following to ground shielded cables.
Use the lockwashers, clamps, and kep nuts provided in the parts kit to install grounding clamps on the enclosure studs adjacent to cord grips. Install grounding clamps only for cord grips that will be used; do not tighten nuts.
Route cables through cord grips and grounding clamps reach cable connectors. Mark cables to remove insulation and shield as described below:
For cables with foil shielding, strip insulation and foil from the cable half grounding clamp (see Figure 2-3). Fold the foil shield back on the cable where the cable passes throug side of foil is turned outward for contact with the grounding clamp.
For cables with braided shielding, strip cable insu of insulation only to expose the braid where the cable passes through the clamp (see Figure • on page 8).
F inish installation using cable mounts and ties to secure ca
to determine cable lengths required to
an inch (15 mm) past the
h the clamp. Ensure silver (conductive)
lation and braided shield from a point just past the grounding clamp. Strip another half inch (15 mm)
bles inside of indicator enclosure.

2.7.1 Serial Communications

Wire the serial communications cables to J2, which is Port 1 (5-wire RS-232 port). J3 is Port 2 (RS-232 and 20 mA). Connect communications cables to J2 and J3 as shown in Table 2-3.
Use cable ties to secure serial cables to Port 1 supports full duplex RS-232 communications only; Port 2
the inside of the enclosure.
provides either active 20 mA output or duplex
RS-232 transmission. Both ports are configured using the SERIAL menu. See Section 3.4.4 on page 29.
Connector Pin Signal Port
J2 1 Ground 1
2 Ground
3 Tx
4 Rx
5 DTR
6 RTS
J3 1 20mA+ 2
2 Ground
3 Tx
Table 2-3. J2 and J3 pin assignments
4 Rx
5 CTS
6 RTS
8 CW-90/90X Checkweigher

2.7.2 Digital I/O

Screws
Screw
The Digital I/O can be configured as either digital inputs or digital outputs as determined by the DIO menu (see Section 3.4.4 on page 29). The inputs are active (on) with low voltage (0 VDC) and can be driven by TTL or 5V logic without additional hardware. Use the inputs. LEDs on the CPU board
light when digital inputs are active (see Figure 2-4).
Digital outputs are typically used to
DIG I/O menu (see Section 3.4.4 on page 29) to configure the digital
control relays that drive other equipment. Outputs are designed to sink not source, switching current. Each output is a CMOS circuit, capable of sinking 24 mA when active. Digital outputs are wired to switch relays when the digital output is active (low, 0 VDC) with reference to 5 VDC supply. LEDs on the CPU board light up when the digital outputs are active (see Figure 2-4).
Screw
DIO LEDs
Figure 2-4. Digital I/O lights and screw locations
Connector Pin Signal
J4 1 +5V
2 Ground
3 DIG I/O 1
4 DIG I/0 2
5 DIG I/O 3
6 DIG I/O 4
Screw
Table 2-4. J4 Pin Assignments (Digital I/0)

2.8 Enclosure Reassembly

Once the cabling is complete, position the backplate over the enclosure and reinstall the screws. Use the torque pattern shown in Figure 2-1 on page 6 to prevent distortion. Torque to
15 in/lbs (1.7 N-m).

2.9 Board Removal

If you must remove the CW-90/CW-90X CPU board, use the following procedure:
1. Disconnect power to the indicator. Remove backpla
2. Disconnect power supply cable from connector J9 on the CW
3. Disconnect the wires at the following connectors: J1, J2,
4. There are two boards: the CPU board and the keypad disp the CPU board (see Figure 2-4), then lift the board out of the enclosure.
T o replace the CPU board, reverse the above procedure. Be sure to indicator enclosure.
te as described in Section 2.4 on page 7.
-90/CW-90X CPU board.
J3, J4.
lay board. Remove the five screws connecting
reinstall cable ties to secure all cables inside the
Installation 9

2.10 Battery Replacement

CAUTION
Note
CAUTION
Risk of explosion if battery is replaced with incorrect type. Dispose of batteries per manufacturer instructions.
The lithium battery on the CPU board maintains the real-time clock and protects data stored in the system RAM when the indicator is not connected to AC power. Data protected by this battery includes time and date, IDs, and configuration information. If any data is lost, the indicator configuration can be restored from the PC. Watch for the low battery warning on the LCD display and periodically check the battery voltage on the CPU board. Batteries should be replaced when the indicator low battery warning comes on, or when battery voltage falls to 2.2 VDC. Life expectancy of the battery is 10 years. Use Revolution to store a copy of the configuration before attempting to replace the battery.
For best results, replace the battery while in weigh mode a
nd with AC power applied. use care not to bend the
battery retaining spring.

2.11 Installing Option Cards

Option cards are not hot-pluggable. Disconnect power cord entirely before installing option cards.
Ethernet port is not suitable for connection to circuits used outside the building and is subject power faults.
to lightning or
Each option card is shipped with installation instructions specific to that card. For specific instructions on the WLAN card, refer to Section 8.0 on page 59. The general procedure for all option cards is:
1. Disconnect power cord from the indicator.
2. Remove the backplate as described in Section 2.4.
3. Install the plastic standoffs in the standof
4. Carefully align the option card connector with the J5 connector on
5. Press down firmly to seat the option
6. Make connections to the option ca
f holes (shown as black-shaded circles in Figure 2-5).
the CPU board.
card in the CPU board connector.
rd as required. Use cable ties to secure loose cables inside the enclosure.
When installation is complete, reassemble the enclosure as described in Section 2.8.
10 CW-90/90X Checkweigher
Figure 2-5. CW-90/CW-90X CPU Board
Note
Jumper Description
JP1/JP2 Jump excitation to sense. If using a 4-wir
and JP2 on. If using a 6-wire load cell cable, take JP1 and JP2 off. Default is ON.
JP3/JP4 Used when upgrading firmware. The jumpers
upgrading firmware and off when the update is complete.
PC Power control. If the jumper is set to SW, the POWER key can be used
turn the unit on/off. If set to ON, the unit will power on when
to plugged in and can only be powered off by unplugging.
JMP1 If set to Audit ON, calibration and conf
through the front keypad. If set to Audit OFF, calibration and configuration can only be accessed by removing the screw on the base of the unit and pressing the Setup switch with a screwdriver. Default is Audit ON.
e load cell cable, leave JP1
should be on when
iguration can be acessed
Table 2-5. Jumper Descriptions
If the RESET button on the CPU board is pressed, the indicator will perform a reboot.
Installation 11
13
30
31
22
21
8
5
18
17
15
34
28
27
19
16
26
24
25
9
7
6
3
20
4
10
12
11
10
10
11
11
1
33
29
2
14
32
23

2.12 Replacement Parts and Assembly Drawings

12 CW-90/90X Checkweigher
Figure 2-6. CW-90/CW-90X Assembly and Components
Item No. Part No. Description (Quantity)
1 Enclosure 1
2 Mounting Plate, Display 1
3 14621 Nut, Kep 6-32NC HEX 2
4 85202
85203
5 68600 Cord Grip, PG11 1
6 68599 Seal Ring, Nylon PG11 1
7 68601 Nut, PG11 1
8 58983 Cable Grip, SL-7 w/Nut 3
9 16892 Label, Earth Ground 1
10 15134 Washer, Lock NO 8 Type A 3
11 14626 Nut, Kep 8-32NC HEX 3
12 45043 Wire, Ground 4in W/No. 8 1
13 104914
104915
14 105850 Bracket, Power Supply 1
15 76556 Power Supply, Switching 1
16 14822 Screw, Mach 4-40NC x 1/4 6
17 14825 Screw, Mach 4-40NC x 1/4 4
18 105976 Cable Assy, Power Supply 1
19 102354 CPU 1
19 105741 Display 1
20 69291 Battery, 3v Coin Lithium 1
21 42640 Screw, Mach 1/4-28NF X 1/4 1
22 44676 Washer, Bonded Sealing 1
23 Backplate 1
24 84388 Gasket, Backplate 420 1
25 88733 Vent, Breather Sealed 1
26 88734 Nut, Breather Vent 1
27 14862 Screw, MACH 8-32NCx3/8 4
28 45042 Washer, Bonded Sealing SST 4
29 29635 Stand, Tilt SST 1
30 103988 Washer, Nylon .52 ID 2
31 103610 Knob, Black 1/4-20 2
32
33 53308 Label 1
34 16861 Label, Warning High 1
-- 105945 Parts Kit, CW-90/90X 1
-- 107476 Bench Scale Foot
-- 105555 Coupling Nut for Feet and Overload Stops
53307 Label 1
Power Cord Assy, 120VAC Power Cord Assy, 230VAC
Overlay, Membrane Swit Overlay, Piezo (CW-90X)
ch (CW-90)
Figure 2-7. CW-90/CW-90X Replacement Parts
1
1
Installation 13
Scale Capacity Load Cell Part Number
11.24
9.50
10.02
10.02
4x .281
.38
3.75
2.01
6.00
3.00
3.75
3.12
6.00
8.57
4.00
5 lb 107174
10 lb 107174
25 lb 107175
50 lb 107176
100 lb 107177
Table 2-6. CW-90X Load Cells
Scale Capacity Load Cell Part Number
5 lb 107756
10 lb 107757
25 lb 107758
50 lb 107759
100 lb 107760
Table 2-7. CW-90 Load Cells
Figure 2-8. CW-90/CW-90X Dimensions
14 CW-90/90X Checkweigher

3.0 Configuration

AUDIT
Only displayed if
turned on
from Setup menu
Only displayed if
turned on
from Setup menu
CALIBR SETUP TEST EXIT
ID ACCUM BRIGHTT&D
OVER
UNDER
MENU
ID
ENTER
TARGET
Move UP/ Increment
Value
Move DOWN/
Decrement
Value
BackspaceEnter/Exit
Configuration
Menu
Enter VALUE/
Move DOWN
Move LEFT/
Previous
Move RIGHT/
Next
Note
Note
To set up and configure the CW-90/CW-90X checkweigher, you will either use the Menu key on the front panel or press the setup switch on the bottom of the indicator. The indicator is defaulted at the factory with the audit trail jumper (JMP1) in the ON position, allowing configuration access by pressing the will take you to the Audit menu selection. Use the selections. Use the
UNDER ( ) and OVER ( ) navigation key s to enter and exit menu selections, respectively.
ID ( ) and TARGE T ( ) navigation keys to move to other menu
Menu key. Pressing the Menu key
Figure 3-1.
Menu Function
AUDIT Audit Trail Displays the legally relevant (LR) firmware versi
CALIBR Calibrate Calibrates the scale.
SETUP Setup Accesses setup options for the scale, features, serial
and displays the version number.
TEST Te st Performs a basic test on the A/D, digital inputs and ou
keypad.
EXIT Exit Exits Configuration mode and r
T&D Time & Date Allows you to view and change time and date.
ACCUM Accumulator Displays the current accumulator
BRIGHT Brightness Adjusts display intensity on a scale fr
ID ID Allows you to setup or edit IDs.
Menu Mode Navigation Structure
on, configuration count, and calibration count.
output, print format, digital inputs and outputs,
tputs, communication ports, RAM, and
eturns to the weigh screen.
value.
om 0 (dimmest) to 7 (brightest).
Table 3-1 . Menu Menu Summary

3.1 Menu Navigation

Once you have entered the menus, you can use the front panel buttons to nav iga te. Note the loop ed me nu st ructur e shown in Figure 3-1. This provides a shortcut for a vice-versa for the opposite side of the men
u. This shortcut also holds true for sub-menu navigation.
ccessing the rightmost menu items by navigating to the left, and
Figure 3-2. Front Panel Key Functions in Menu Mode
During calibration, the ENTER key acts as a data entry confirmation key. It also acts as an EXECUTE key and accepts the value upon successful calibration. On the CW backspace.
-90, the ZERO or CLR key can be used for a
Various keys are used as directional keys to navigate through the menus (see Figure 3-2). The ID ( ) and TAR GE T ( ) keys scroll left and right (horizontally) on the same menu level; OVER ( ) and UNDER ( ) move up and down (vertically) to different menu levels. These
ENTER key has the same function as UNDER ( ) when navigating the menu; either will move down to access
The sub-categories of a main menu item. When addin keys will edit the blinking digit. The CW-90, whose numeric keys can be used to insert a digit to the right of the blinking digit. Press
keys have a directional symbol indicating its menu navigation function.
g or editing data using the CW -90X, the OVER ( ) and UNDER ( )
ID ( ) and TA RG ET ( ) keys navigate to another digit. This also applies to the
ENTER to accept
the data and return to the next menu item.
To exit Configuration and return to weighing, press the MENU key or navigate to the EXIT menu and press ENTER.
Configuration 15

3.2 Audit Menu

CALIB
Number
EXIT
EXIT T&DTESTSETUPAUDIT
XXXXXXX
CALIBR BRIGHTACCUM
Only displayed if
turned on
from Setup menu
ID
Only displayed if
turned on
from Setup menu
LR,V.
CONFIG Number
The audit menu accesses audit trail support. It provides tracking information for configuration and calibration events. To prevent potential misuse, all configuration and calibration changes are counted as change events. Audit information can be printed by pressing the menu.
AUDIT Menu
Parameter Description
LR.V. Legally relevant firmware version.
CALIB Number Displays total calibration events. The
four-digit number.
CONFIG
mber
Nu
EXIT Exits the audit trail and returns to weigh mode.
Displays total configuration events. The display alternates (in one-second intervals) between CFG and the four-digit number.
PRINT key while displaying the audit trail items beneath the AUDIT
Figure 3-3. Audit Menu
display alternates (in one-second intervals) between CALIB and the
Table 3-2. Audit Menu Parameters
16 CW-90/90X Checkweigher

3.3 Calibration Menu

Note
WZERO
Calibrating, please wait
Calibrating, please wait
Display and edi
Press Enter
to calibrate zero
t
test weight value
WVAL WSPAN
Press Enter to
remove offset from
zero and span
calibrations
REZEROWLIN
POINT–> 1
POINT–> 2
POINT–> 3
POINT–> 4
POINT–> 5
Same as POINT-> 1
WGT 1
number
Calibrating, please wait
Press Enter
to calibrate span
EXIT T&DTESTSETUPAUDIT
XXXXXXX
CALIBR BRIGHT
Press Menu
to cancel
Press Menu
to cancel
Previous A/D raw
counts are shown
Previous A/D raw counts are shown
ACCUM
Only displayed if
turned on
from Setup menu
CAL1
ID
Only displayed if
turned on
from Setup menu
Note
Note
See Section 4.0 on page 36 for calibration procedures. The Calibration menu can be protected by assigning a password in the Feature menu.
When entering a floating point value, the decimal point will appear and flash on the far right (if it is not already on the screen). To move the decimal, press ENTER, then use the desired location.
The CW-90/CW-90X requires the WZERO and WSPAN points to be calibrated. The linearity points are optional, but must
NOT duplicate zero or span. During calibration, the ENTER key acts as a data entry confirmation key. It also acts as an EXECUTE key, and accepts the value if calibration was successful.
and keys to move the decimal to the
Figure 3-4. Calibration Menu
CALIBR Menu
Parameter Choices Description
Level 2 submenus
WZERO Press ENTER and the A/D raw counts will be displayed.
Press ENTER again to calibrate zero, or press MENU to cancel. “CALIBRATING, PLEASE
AIT” will be appear prior to automatically moving to WVAL.
W
WVAL Press ENTER to display and edit the test weigh
WSPAN.
WSPAN Press ENTER and the A/D raw count
press MENU to cancel. “CALIBRATING, PLEASE WAIT” will be appear prior to automatically moving to WLIN.
WLIN PT->1 — PT->5 WGT 1 allows you to display and edit the
CAL1 allows you to calibrate and display the raw A/D value. Pr
REZERO Press ENTER to remove an offset value from the
calibration. Press ENTER to will move to the next calibration point.
chains are being used during calibration).
The linearity points are optional, but must NOT duplicate zero or span. They must be between zero and span.
Use this parameter only after WZERO and WSPAN have been set. See Section 4.1 on page 36 for more information about using
s will be displayed. Press ENTER to calibrate the span or
test weight value. Press ENTER to edit the value.
zero and span calibrations (e.g. if hooks or
t value. Press ENTER again to move to
ess TARE to perform
this parameter.
Table 3-3. Calibrat ion Menu Parameters
Configuration 17

3.4 Setup Menu

Note
5000
number
OVRLOA
FS+2%
FS+9D
FS+1D
FS
MOTBAN
1
number
DFSENS
32OUT 64OUT
2OUT
8OUT
4OUT
16OUT
128OUT
DFTHRH
20D 50D
250D
5D
2D
10D
100D
NONE
200D
SMPRAT
30HZ
60HZ
7.5HZ 15HZ
120HZ
480HZ
240HZ
960HZ
PWRUPM
GO
DELAY
TAREFN
BOTH
PBTARE
NOTARE
KEYED
DFLTR1
64
128
4
16
8
32
256
1 2
ACCUM
OFF
ON
VISIBL
ON
OFF
GRADS
CALIBR
SCALE
FEATUR
SERIAL DIGI/OPFORMT VERS
SSTIME
number
10
THRESH
0
UNITS DECPNT DSPDIV
PRIMRY
LB/OZ *
KG
OZ
G
LB
888888
1D
5D
2D
* Not legal for trade
888880
8.88888
88.8888
8888.88
888.888
88888.8
SEC
LB/OZ *
LB
G
OZ
LB/OZ*
KG
OFF
TER
G
OFF
KG
LB OZ
LB/OZ*
DFLTR2
64
128
4
16
8
32
256
1 2
DFLTR3
64
128
4
16
8
32
256
1 2
Display and edit
test weight value
WVAL WLIN
WZERO
Calibrating, please wait
Press Enter
to calibrate zero
Press Menu
to cancel
Previous A/D raw counts are shown
WSPAN
Calibrating, please wait
Press Enter
to calibrate span
Press Menu
to cancel
Previous A/D raw
counts are shown
CONFIG
EXIT T&DTESTSETUPAUDIT
XXXXXXX
CALIBR BRIGHTACCUM
Only displayed if
turned on
from Setup menu
ID
Only displayed if
turned on
from Setup menu
Calibrating, please wait
Press Enter to
remove offset from
zero and span
calibrations
REZERO
ZTRKBN
number
0
ZRANGE
1.900000
number
FORMAT
POINT–> 1
POINT–> 2
POINT–> 3
POINT–> 4
POINT–> 5
Same as POINT-> 1
WGT 1
number
CAL1
The setup menu allows you to configure settings for the scale, features, serial port settings, print format, and digital inputs and outputs. You can also view the software and regulatory versions, and revert to default settings.
If RANGE is displayed after a setup value is entered, the value is too large or small for the given parameter.

3.4.1 SCALE Submenu

Calibration can be performed in two places within the menu: the CALIBR menu shown in Figure 3-5 is an in-depth scale setup and calibration. A “quick acc
ess” calibration is shown in Figure 4.1 on page 36.
18 CW-90/90X Checkweigher
Figure 3-5. Setup Menu: Scale Submenu
SCALE Menu
Note
Note
Note
Parameter Choices Description
Level 3 submenus
CALIBR WZERO
WVAL WSPAN WLIN REZERO
GRADS 10000
1–100000
See Section 3.3 on page 17 for descriptions, and see Section 4.0 on page 36 for calibration procedures.
Calibration can be perfor Figure 3-5 is an in-depth scale setup and calibration. A “quick access” calibration is shown in Figure 4.1 on page 36.
Specifies the number of full scale graduations.
The value entered must be in the range 1–100000 and should be consistent with legal
equirements and environmental limits on system resolution.
r
To calculate GRADS, use the formula: GRADS = Capacity / Display Divisions.
Display divisions are specified under
med in two places within the menu: the CALIBR menu shown in
the FORMAT submenu.
If RANGE is displayed after the value is entered, the value is too large or small for the given parameter.
FORMAT PRIMRY
SEC TER
ZTRKBN 0
number
ZRANGE 1.900000
number
MOTBAN 1
number
OVRLOA FS+2%
FS+1D FS+9D FS
SSTIME 10
number
SMPRAT 30HZ
7.5HZ 15HZ 60HZ 120HZ 240HZ 480HZ 960HZ
Select your primary, secondary, and tertiary units of me oz, and g. Secondary and tertiary units can also be set to lb/oz (lb/oz is not legal for trade), or off.
Automatically zeroes the scale when within the ZRANGE and scale is at standstill. Specify the zero tracking band in ± display divisions. Maximum legal value varies depending on local regulations.
For scales using linear calibration, do not set the zero tracking band to a value greater than that specified for the first linearization point.
Selects the range within which the scale can be zer
1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at standstill to zero the scale. Use the default value for legal-for-trade applications.
Sets the level, in display divisions, at which detected for 1 second or more, the standstill symbol lights. Some operations, including print, tare, and zero, require the scale to be at standstill. Maximum legal value varies depending on local regulations.
If this parameter is set to 0 the standstill annu requiring standstill (zero, tare, print) are performed regardless of scale motion. If 0 is selected, ZTRKBND must also be set to 0.
Overload. Determines the point at whic message is displayed. Maximum legal value varies depending on local regulations.
Specifies the length of time the scale must be ou the scale is considered to be at standstill. Values greater than 10 are not recommended.
Sample rate. Selects measurement rate, in samples per second, of the analog-to-digital converter
. Lower sample rate values provide greater signal noise immunity.
Settings of 120Hz or above may be too fast to provide the desired stability in some static weighing applications.
within the range specified, as long as the input is
scale motion is detected. If motion is not
nciator does not light; operations normally
h the display blanks and an out-of-range error
asure. Sub-choices include lb, kg,
oed. The 1.900000 default value is ±
t of motion, in 0.1-second intervals, before
Table 3-4. Scale Menu Parameters
Configuration 19
Note
SCALE Menu
Parameter Choices Description
DFLTR 1-3 4
8 16 32 64 128 256 1 2
DFSENS 2OUT
4OUT 8OUT 16OUT 32OUT 64OUT 128OUT
DFTHRH NONE
2D 5D 10D 20D 50D 100D 200D 250D
TAREFN BOTH
NOTARE PBTARE KEYED
PWRUPM GO
DELAY
Selects the digital filtering rate used to reduce the ef immediate area of the scale. The overall filtering effect can be expressed by adding the values assigned to the three filter stages:
DFLTR1 + DFLTR2 + DFLTR3 See Section 9.5 on page 62 for information on digital filtering.
Choices indicate the number of A/D displayed reading. A higher number gives a more accurate display by minimizing the effect of a few noisy readings, but slows down the settling rate of the indicator.
Digital filter cutout sensitivity. Specifies the number of c outside the filter threshold (DFTHRH parameter) before digital filtering is suspended.
Digital filter cutout threshold. Specifies the filter thr specified number of consecutive scale readings (DFSENS parameter) fall outside of this threshold, digital filtering is suspended. If NONE is selected, the filter is always enabled.
Enables or disables push-button and keyed tares. Possible values are:
BOTH: Both push-button and ke NOTARE: No tare allowed (gross mode only) PBTARE: Push-button tares enabled KEYED: Keyed tare enabled
Power up mode. In GO mode, the indicator goes into operation immediately after a brief power up display test.
fects of mechanical vibration from the
conversions per update that are averaged to obtain the
onsecutive readings that must fall
eshold, in display divisions. When a
yed tares are enabled.
In DELAY mode, the indicator performs a power up displa warm up period. If no motion is detected during the warm up period, the indicator becomes operational when the warm up period ends; if motion is detected, the delay timer is reset and the warm up period repeated.
ACCUM OFF
ON
VISIBL ON
OFF
THRESH 0 Enter a value to be used as display divisions. Zer
Accumulator. Specifies whether the sc occurs whenever a print operation is performed. Scale must return to zero to re-arm a new print.
Scale visibility. Specifies whether weight data is displayed. Status annunciators, under/
ccept/over lights remain on in weigh mode even if VISIBL is set to OFF.
a
or reset point where automatic printing functions reset themselves to be retriggered. When a display division value is entered, anything that would normally require the scale to reach zero before it happens/rearms (except for LFT parameters) now only needs to go below this display division value and then above it again.
If checkweighing is used, THRESH should be less than the under value or it can disable the digital I/O points from tripping.
ale accumulator is enabled. If enabled, accumulation
Table 3-4. Scale Menu Parameters (Continued)
y test, then enters a 30-second
o threshold allows you to select a threshold
20 CW-90/90X Checkweigher
Loading...
+ 54 hidden pages