QB Series Limited Warranty................................................................................................................... 28
QB Series Operation Manual
Page 5
About This Manual
The QB Series precision balance is an advanced precision weighing instrument incorporating a
highly-sophisticated electronic tuning fork sensor assembly capable of providing quick, reliable, and stable
weight measurement. This manual is intended to give users guidelines and specifications to ensure proper use
and care of the device.
1.0Introduction
The QB Series balance contains several features to provide optimum performance.
Standard features include:
•Twelve program-selectable units of measure, including: grams (g), kilograms (kg), pounds (lb), carats,
decimal oz, troy ounces, pennyweights, grain, Hong Kong tael, Sin tael, Taiwanese tael, or momme
•Automatic shut-off (used with battery operation only)
•Front panel controls
•Automatic sleep mode (when used with internal rechargeable battery pack only)
•Large weighing platform
"
to indicate unit of measure other than g, kg, or lb.
Options include:
•Glass breeze break (for nominal capacities of 620g and below)
•RS-232C for interfacing with a printer or other peripheral equipment
•Internal rechargeable battery (operates for 32 hours on one charge)
NOTE: There are many AC adapters available in the marketplace. If you should need to replace the AC adapter provided
with this unit, we recommend that you choose the RLWS AC adapter PN 42566 for 115 VAC or PN 42168 for 230 VAC
use. Use of any other AC adapter can void the warranty.
1.1Unpacking and Inspection
Each QB Series balance comes with the following components (see Figure 1-1 on page 2).
•AC adapter
•Weighing platform (round or rectangle, model dependent)
•Weighing platform support (size is model dependent)
Be sure all components have come with your QB Series balance.
NOTE: Store the packing material and box in a safe, easily accessible place. Should you ever need to transport the
balance across any substantial distance, using the original packing material is the first step to protecting your
investment.
Introduction
1
Page 6
2
Weighing Platform
(both styles)
Platform Support
Adaptor
Connector
AC Adaptor
Level Indicator
Display
Leveling Leg (4)
QB
RICE LAKE
OVER
ACCEPT
UNDER
ON
OFF
T
F
P
Expansion Bay
Figure 1-1. Main Component Parts of the QB Series
1.2General Care and Maintenance
To keep your QB Series balance operating properly, the device needs to be maintained and cared for. Keep the
housing and platform clean by wiping them off with a soft damp cloth daily. If necessary, use a slight solution of
water and dish soap to clean away stains and dirt. Check under the platform support area for any build up of dirt
and debris that could interfere with the mechanism. Keep your calibration masses in a safe, dry area. Unplug the
AC adapter when the device is not in use. For long term storage, remove the internal battery (if equipped).
1.3Installation Guidelines
For optimum performance, we recommend that you use your QB Series device in a clean, stable environment.
Avoid the following:
•Excessive drafts
•Significant temperature and/or humidity changes
•Strong magnetic fields exist or equipment that may generate a magnetic field
•Excessive dust
•Unstable or vibrating work surface
•Corrosive gases or large amounts of dust
•Long-term exposure to direct sunlight
•Disassembling or modifying the unit in any way
•Moving the balance when an item is on the weighing platform
•Handling your balance with wet hands
•Shock loading or intentionally overloading the device
•Leaving the device exposed to direct sunlight
•Using any solvents on the device
QB Series Operation Manual
Page 7
1.4Installation Steps
Use the following steps to properly install the QB Series balance:
1. Inspect the leveling legs and ensure that they are all completely retracted.
2. Place the balance down on a firm, level mounting surface.
3. Using the bubble level to the upper left of the display panel as a guide, adjust the leveling legs as
necessary until the bubble is centered inside the bulls-eye of the bubble level (see Figure 1-2).
BALANCE IS LEVEL
BALANCE IS NOT LEVEL
Blue
circle
Leveling leg
Bubble
Figure 1-2. Level Adjustment
4. Plug the AC adapter into the back of the unit (see Figure 1-3).
Figure 1-3. AC Adapter Connection
5. Secure the weighing pan support onto the device. Tighten the adjusting knob until it is finger tight. Do
not use a screwdriver (see Figure 1-4).
6. Place the platform cover down over the weighing pan support (see Figure 1-1 on page 2 for exploded
view).
7. Plug the AC adapter into a suitable wall outlet.
Adjusting
Knob
QB
RICE LAKE
R
E
V
O
T
P
E
C
C
A
R
E
D
N
U
N
O
F
F
O
T
F
P
Figure 1-4. Attach Weighing Pan Support
Introduction
3
Page 8
4
1.5Special Information
The QB Series balance is a sensitive, precision weighing instrument and should always be treated with care.
While every attempt is made prior to shipping to ensure that your balance arrives ready-to-use, uncontrollable
events during the shipping process may make it necessary to calibrate the device before being used.
While not absolutely necessary , we recommend that you perform a span calibration prior to using the balance for
the first time. Even though your QB Series balance was carefully calibrated at the factory, shipping and the
effects of specific gravity can ha v e a ne gati v e impact on accurac y. Review Section 3.0 on page 14 for instructions
on this feature.
In the event that the device fails to operate properly, contact your Rice Lake Weighing Systems dealer for
assistance.
1.6Front Panel
Figure 1-6 shows the QB Series keypad and key functions. Figure 1-7 on page 5 shows the LCD display
segments display.
Tare key. In weighing mode, used
to tare weight. In calibration mode,
Print key, or data transmission key.
used to increment/decrement
parameter values.
ON
OFF
Turns scale on or off.
OVER
ACCEPT
UNDER
Not used.
PF
Function key. In weighing mode, used
to select display unit of measure. In
calibration mode, used to select operational
parameters.
Figure 1-5. Front Panel Keypad
Memory function activated for values used in
limit or sampling functions.
Indicates the battery charge is low.
LOW
M
Indicates a negative
weighing value.
Limit mode judgement results.
T
Indicates weighing mode in
lb or oz (dependent on model)
See Table 2-1.
lb
P
Indicates default weighing mode in grams.
Indicates supplemental
unit of measure.
Indicates count mode.
QB Series Operation Manual
Figure 1-6. QB Series Display Segments
Page 9
1.7Powering Up the QB Series Balance
Press and release the
1. All display segments momentarily illuminate.
ON/OFF
key to turn on the QB Series balance.
M
2. All display segments momentarily flash.
3. The device activ ates the def ault weighing mode and displays the
display to indicate that weighments are displayed in grams.
1.8Powering Down the QB Series Balance
To turn the balance off, press and release the
ON/OFF
key.
lb
P
k
g
symbol in the lower right corner of the
Introduction
5
Page 10
6
2.0Weighing Operations
The following section describes weighing operations of the QB Series balance, including: units of measure,
general weighing, weighing with a tare, batching, comparator, and piece counting.
2.1Units of Measure
Table 2-1 shows the various units of measure (maximum capacity and minimum increment) available on your
QB Series balance.
NOTE: There is no push-button conversion from one unit to another. The QB Series balance is capable of showing only
one unit of measure during a weighing operation. Switching to another unit of measure requires a change in the set up
0.05
2300 x 2 46000 x 292000 x 595000 x 2180000 x 200.0647989
40 x
0.002
39 x
0.002
40 x
0.002
400 x
0.02
3000 x
0.1
3 x
0.0001
15000 x
0.5
100 x
0.005
6.6 x
0.0002
96 x
0.005
1900 x
0.05
80 x
0.002
79 x
0.002
80 x
0.002
800 x
0.02
6000 x
0.2
6 x
0.0002
3000 x 1 31000 x
210 x
0.005
13 x
0.0005
190 x
0.005
3800 x
0.1
160 x
0.005
150 x
0.005
160 x
0.005
1600 x
0.05
6200 x
0.1
6.2 x
0.0001
0.5
210 x
0.005
13 x
0.0002
190 x
0.005
3900 x
0.05
160 x
0.002
160 x
0.002
160 x
0.002
1600 x
0.02
12000 x 11
12 x
0.001
60000 x 50.2
420 x
0.05
26 x
0.002
380 x
0.05
7700 x
0.5
320 x
0.02
310 x
0.02
320 x
0.02
3200 x
0.2
Table 2-1. Unit Capacities
Conversion
Factor/g
0.001
28.349527
453.5924
31.103481
1.5551740
37.428932
37.799466
37.5
3.75
2.2Changing Units of Measure
After the balance has been powered up (see Section 1.7 on page 5), all display segments momentarily illuminate,
followed by two flashes of all displays. The device then displays default weighing mode. If the balance does not
show a
value. The default unit of measure from the factory is
unit of measure may already be activated. To change the active unit of measure, use the following steps:
0.00
value after the normal start sequence, press and release the
1. Ensure that there is nothing on the platform and that the weight reading is
2. Press and hold the
3. Release the
F
key until the word
key. The display changes to show either
F
4. Repeatedly press and release the
measure is currently
). If the active unit of measure is another v alue,
g
one of the following values:
5. Refer to Table 2-1 above and Table 4-1 on page 15 to determine the setting required for the desired unit
of measure.
6. Repeatedly press and release the
key to confirm the setting.
7. Press and release the
QB Series Operation Manual
P
g
(grams). If the balance has already been used, a different
Func
is displayed.
1.SET.1
F
key to change the display, until it reads
2, 3, 4, 5, 6, 7, 8, 9, A, b,
T
key until the appropriate value is displayed.
or
or
.
C
T
key. The balance returns to a
.
0.00
1.SET.2
.
7.un.1
is displayed, followed by any
7.un.
(if the active unit of
0.00
Page 11
8. The display returns to a
0.00
value for the appropriate unit of measure.
To change to a different unit of measure, repeat steps 1 – 9 above.
2.3General Weighing
The following section provides a step-by-step procedure on general weighing using the QB Series balance.
1. Power up the balance (see Section 1.7 on page 5).
2. All display segments momentarily illuminate, followed by two flashes of all displays. The device then
displays default weighing mode. To change the default unit of measure, please refer to Section 2.2 on
page 6. If the balance does not show a
key. The device returns to a
0.00
value.
0.00
value after the normal start sequence, press and release the
3. Gently place the item to be weighed on the platform. The unit of measure indicator begins to flash
indicating the balance is registering the weight of the commodity. The value display begins to update.
When the unit of measure indicator ceases flashing, the device is stable and you can read the weight
value.
2.4Weighing with a Tare
The following section provides a step-by-step procedure on weighing with a tare using the QB Series balance.
NOTE: Whenever the tare function is activated, the effective weighing range of the device is reduced by the value of the
tare. For example, on a QB-600 with a total capacity of 600g, if a tare of 100g is entered into the balance, the total
effective NET weighing range of the device is only 500g (total rated capacity [600g] - tare value [100g] = 500g. Any
attempt to weigh over the effective weighing range results in a “o-Err” error message.
1. Power up the balance (see Section 1.7 on page 5).
2. The balance displays default weighing mode. If the balance does not show a
start sequence, press and release the
T
key.
3. The default unit of measure is displayed. To change the unit of measure, see Section 2.2.
4. Gently place the empty container on the platform. The unit of measure indicator begins to flash,
indicating that the balance is registering the weight of the commodity. The value display begins to
update. When the unit of measure indicator ceases flashing, the balance is stable and you can read the
weight value.
5. Press the
NOTE: The value of the original tare must be
value is
T button can be activated. The tare function is not available. To disable this “auto-zero” feature, please
refer to Section 4.0 on page 15.
T
key. The zero display (
≤
+/- 3 divisions from zero, the device automatically re-establishes a new zero setting before the
0.00
) appears.
≥
+/- 4 displayed divisions from zero. If the original tare
6. Gently place the product to be weighed in the previously tared container. The unit of measure indicator
begins to flash indicating that the balance is registering the weight of the commodity. The displayed
value begins to update. Continue adding commodity until the display indicates the desired weight value.
When the unit of measure indicator ceases flashing, the device is stable and you can read the weight
value.
NOTE: To view the gross weight of the commodity and the tare, remove the container with its contents
from the weighing platform. The total weight is displayed as a negative number. To return to normal
weighing, replace the container with its contents on the platform.
7. To weigh another commodity and tare or to re-use the current tare and weigh additional samples, remove
the commodity and container from the platform and press the
T
key to have the display return to a
reading, then repeat steps 1– 6.
value after the normal
0.0
0.00
T
2.5Batching
The following section provides a step-by-step procedure on using the batching feature of the QB Series balance.
NOTE: Whenever the tare function is activated on a QB balance, the effective weighing range of the device is reduced by
the value of the tare. For example, on a QB-600 with a total capacity of 600g, if a tare of 100g is entered into the balance,
the total effective NET weighing range of the balance is only 500g (total rated capacity [600g] - tare value [100g] = 500g).
Any attempt to weigh over the effective weighing range results in a “o-Err” error message.
1. Power up the balance (see Section 1.7 on page 5).
Weighing Operations
7
Page 12
8
2. The balance displays default weighing mode. If the balance does not show a
start sequence, press and release the
key.
T
0.0
value after the normal
3. The default unit of measure is displayed. To change the unit of measure, see Section 2.2.
4. Gently place the empty container on the platform. The unit of measure indicator begins to flash
indicating the device is registering the weight of the commodity. The value display begins to update.
When the unit of measure indicator ceases flashing, the balance is stable and you can read the weight
value.
5. Press the
NOTE: The tare value must be ≥ ± 4 display divisions from zero. If the tare value is ≤ ± 3 divisions from
zero, the device automatically re-establishes a new zero setting before the T button can be activated.
The tare function is not available. To disable this auto-zero feature, please refer to Section 4.0 on
page 15.
key. The zero display (
T
0.00) appears.
6. Gently place the commodity to be weighed in the previously tared container.
7. The unit of measure indicator begins to flash indicating the device is registering the weight of the
commodity.
8. The value display begins to update.
9. Continue adding commodity until the display indicates the desired weight value.
10. When the unit of measure indicator ceases flashing, the device is stable and you can read the weight
value.
11. Without removing the product, press the
T key .
12. Gently add the next product quantity. Only the net weight of the new commodity is displayed.
13. Repeat steps 11 and 12 as necessary.
NOTE: To view the gross weight of the commodity and the tare, remove the container with its contents
from the weighing platform. The total weight is displayed as a negative number. To return to normal
weighing, replace the container with its contents on the platform.
2.6Removing the Tare
Use the following steps to remove a tare value from the QB balance:
1. Remove the container (empty or full) from the weighing platform.
2. The display shows the weight of the container and the commodity, if any, as a negative value.
3. Press and release the
4. The display returns to a
T key .
0.0 value.
2.7Weighing with the Comparator Function
The QB Series balance is equipped with an onboard comparator function that determines how the weight of a
commodity relates to a pre-determined over/under zone. This feature can help alleviate weighing errors and
speed up the judgement-making process. Rather than reading a weight value and making a mental calculation as
to whether or not the value is acceptable, the operator can simply concentrate on the
indicators to determine whether a weight value is acceptable or not. The comparator function is not available in
the counting mode.
NOTE: When the counting mode is activated, the comparator function is automatically deactivated. Should you desire to
return to the comparator function, all of the most recently entered parameters are automatically recalled when the
counting function is deactivated (see Section 2.9 on page 12).
2.7.1Comparator Function Activation
Prior to using the comparator function, the parameters for the operation must be programmed into the device.
Use the following steps to program the parameters for the comparator function.
1. Power up the balance (see Section 1.7 on page 5).
2. The balance displays default weighing mode. If the balance does not show a
start sequence, press and release the T key .
3. The default unit of measure is displayed. To change the unit of measure, see Section 2.2.
4. Press and hold the
F key until the word Func is displayed.
OVER, UNDER and ACCEPT
0.0 value after the normal
QB Series Operation Manual
Page 13
5. Release the F key. The display changes to show either 1.SET.1 or 1.SET.2. If the display shows 1.SET.1,
continue with step 6 below. If the display shows 1.SET.2, press and release the T key once to change the
value to 1.SET.1, then continue with step 6 below.
6. Press and release the
F key once. The display changes to read 2.SEL.1. This indicates that the comparator
function is currently inactive.
7. Press and release the
T key once. The display changes to read 2.SEL.2. This indicates that the comparator
function is active.
8. Press and release the
F key once. The display changes to read 21.Co.X, where “X” is a v alue of either 1 or
2. Continue with Section 2.7.2 to establish the basic parameter configuration of the comparator function.
2.7.2Basic Over/Under Parameter Selection
1. With the display reading 21.Co.X, where “X” is a v alue of either 1 or 2, (see step 8, Section 2.7.1 above),
determine if you wish to have the comparator function with only a stable weight reading (recommended)
or for all weight readings. If you wish to have a comparison done for only stable weight readings press
and release the T key until the display reads 21.Co.2. If you wish to have a comparison done for all
weight readings, press and release the T key until the display reads 21.Co.1.
2. Press and release the F key once. The display changes to read 22.Li.X where “X” is a value of either 0 or 1.
3. Determine if you wish to have a comparison performed for the entire weighing range of the balance, or if
you wish to exclude comparisons for values close to zero and for negative values (recommended). If you
wish to have comparisons performed for the entire weighing range of the balance, press and release the T
key until the display reads 22.Li.1. If you wish to have comparisons excluded for values close to zero and
for negative weight values (recommended), press and release the T key until the display reads 22.Li.0.
4. Press and release the
1, 2, 3, 4, 5, or 6.
F key once. The display changes to read 23.bu.X, where “X” is a value of either 0,
5. Determine which of the following visual indications you would prefer for the various over/under/accept
conditions.
Parameter ValueIndication
23.bu.0Solid arrow next to appropriate display legend
23.bu.1Flashing arrow next to UNDER display legend. Solid arrow next to ACCEPT and OVER display legends
23.bu.2Flashing arrow next to ACCEPT display legend. Solid arrow next to UNDER and OVER display legends
23.bu.3Flashing arrow next to OVER display legend. Solid arrow next to UNDER and ACCEPT display legends
23.bu.4Flashing arrow next to UNDER and ACCEPT display legends. Solid arrow next to OVER display legend
23.bu.5Flashing arrow next to ACCEPT and OVER display legends. Solid arrow next to UNDER display legend
23.bu.6Flashing arrow next to ACCEPT, UNDER, and OVER display legends
Table 2-2. Over/Under Parameter Values
6. Press and release the
7. Press and release the
legends
OVER, UNDER, and ACCEPT flash continuously indicating that upper and lower limits need to be
T key repeatedly until the appropriate parameter value from Table 2-2 is selected.
P key once. The display returns to a 0.00 reading and the arro ws next to the display
established. Continue with Section 2.8.3 on page 11.
NOTE: If the OVER, UNDER and ACCEPT arrows are not flashing, one of two conditions exist. Either the
comparator function has not been turned on correctly, or there are UPPER and LOWER comparison
limits already programmed into the unit. Verify that the comparator function is properly activated (see
Section 2.7.1 on page 8), then continue with Section 2.8 below.
2.8Setting the Upper and Lower Comparison Limits
There are two methods for establishing the upper and lower comparison limits – the reference sample method
and the digital reference method. Section 2.8.2 on page 10 discusses the reference sample method where samples
of commodity that exactly represent the upper and lower limits are weighed and memorized by the balance.
Section 2.8.2 on page 10 discusses the digital reference method where the upper and lower limits are known
values and can therefore be digitally keyed into the balance.
Weighing Operations9
Page 14
2.8.1Reference Sample Method
Use the following steps to establish upper and lower comparison limits in the QB Series balance using the
reference sample method.
1. Press and hold the
2. The following indicators flash simultaneously – unit of measure,
entered lower limit value, under arrow, and
P key until the display reads L Set, then release the P key.
0.00 weight value or the previously
M.
3. Place a commodity on the weighing platform that represents the minimum acceptable weight value.
4. Press and release the
F key once.
5. The display continues to flash showing the unit of measure, minimum acceptable weight value, under
arrow, and
6. Press and release the
7. The display momentarily indicates
8. The following indicators then flash simultaneously: unit of measure,
M (see Figure 1-7 on page 5).
and
M (see Figure 1-7 on page 5). Write down these values for future reference.
P key once.
H Set.
0.00 weight value, over arrow (#)
9. Place a commodity on the weighing platform that represents the maximum acceptable weight value.
10. Press and release the
F key once.
11. The display momentarily blanks, followed by a flashing display showing the unit of measure, maximum
acceptable weight value, over arrow and
M (see Figure 1-7 on page 5). Write down this value for future
reference.
12. Press and release the
P key once. The OVER indicator flashes and the weight display shows the value
used in step 9.
13. Remove the sample from the weighing platform. The display returns to a
0.00 value and the over arrow
(#) disappears.
2.8.2Digital Reference Method
Use the following steps to establish upper and lower comparison limits in the QB Series balance using the digital
reference method.
1. Ensure that you have followed the instructions in Sections 2.7.1 and 2.7.2 on page 9, then continue
below.
2. Determine and write down the values you need to use as the lower limit and upper limit values.
3. The display shows a
0.00 reading and the arrows next to the display legends OVER, UNDER, and
ACCEPT flash continuously indicating that over and under limits need to be established. Continue with
Section 2.8.3 on page 11.
NOTE: If the OVER, UNDER, and ACCEPT arrows are not flashing one of two conditions exists. Either the
comparator function has not been turned on correctly (see Section 2.7.1 on page 8), or there are OVER
and UNDER comparison limits already programmed into the unit. Verify that the comparator function is
properly activated (see Section 2.7.1 on page 8). Continue with Section 2.8.4 on page 11.
4. Press and hold the P key until the display reads L Set, then release the P key.
5. The following indicators flash simultaneously: unit of measure,
0.00 weight value or the previously
entered lower limit value, under arrow, and M (see Figure 1-7 on page 5).
6. Press and release the
T key once. With the exception of the least significant digit, the display stops
flashing, and indicates all zeros, or the last lower limit value entered.
7. Press and release the
8. Continue to press and release the
9. Press and release the
T key once. The flashing digit changes by a value of one.
T key until the flashing digit reads the value you desire.
F key once. The value selected in step 8 above is set. The digit stops flashing and
the next most significant digit begins to flash.
10. Repeat steps 7 – 9 as necessary until the display reads the value you desire as the lower limit value.
11. If you need to change a digit once it has been set, repeatedly press and release the
F key until the value
you want to change is flashing. Then return to steps 7 – 9.
12. Once the balance reads the correct lo wer limit value, press and release the
P key once. The entire display
flashes showing the value you have selected as the lower limit value.
13. Press and release the
P key again. The display momentarily reads H Set.
10QB Series Operation Manual
Page 15
14. The following indicators flash simultaneously: unit of measure, 0.00 weight value or the previously
entered upper limit value, over arrow (#), and M (see Figure 1-7 on page 5).
15. Press and release the
T key once. With the exception of the least significant digit, the display stops
flashing, and indicates all zeros, or the last upper limit value entered.
16. Press and release the
17. Continue to press and release the
18. Press and release the
T key once. The flashing display changes by a value of one.
T key until the flashing display reads the value you desire.
F key once. The value selected in step 17 above is set. The display stops flashing
and the next most significant digit begins to flash.
19. Repeat steps 17 and 18 as necessary until the display reads the value you desire as the upper limit value.
20. If you need to change a digit once it has been set, repeatedly press and release the
F key until the value
you want to change is flashing. Then return to steps 17 and 18.
21. Once the balance reads the correct upper limit v alue, press and release the
P key once. The entire display
flashes showing the value you have selected as the lower limit value.
22. Press and release the
2.8.3Weighing with the Comparator Function
P key again. The display returns to the weighing mode showing a 0.00 value.
Once the comparator function has been activated (see Section 2.7.1 on page 8) and upper and lower limit values
have been set (see Section 2.8 on page 9), use with the following steps to weigh with the comparator function:
1. Place a commodity on the weighing platform that is exactly equal to the lower limit value. The
ACCEPT
indicator should illuminate confirming that any sample ≥ the lower limit value is acceptable. If not,
return to Section 2.8.1 or 2.8.2 on page 10 to re-enter the lower limit value.
2. Place a commodity on the weighing platform that is exactly equal to the upper limit value. The
ACCEPT
indicator should illuminate confirming that any sample ≤ the upper limit value is acceptable. If not,
return to Section 2.8.1 or 2.8.2 on page 10 to re-enter the upper limit value.
3. Once the lower and upper limits are functioning correctly, any sample that is ≤ the upper limit value and
≥ the lower limit value is acceptable.
4. Remove the test sample from the platform.
5. If the balance does not show a
0.00 value, press and release the T key. The balance returns to a 0.00
value.
6. If applicable, place an empty container on the platform and press the
T key. The balance returns to a 0.00
reading (see Section 2.4 on page 7)
7. Gently place the item to be weighed on the platform. The unit of measure indicator begins to flash
indicating the balance is registering the weight of the commodity. The displayed value begins to update.
When the unit of measure indicator ceases flashing, the balance is stable and displays the weight value
along with the appropriate OVER, UNDER or ACCEPT indicator illuminated.
2.8.4Deactivating the Comparator Function
Use the following steps to deactivate the comparator function in the QB Series balance.
1. Press and hold the
2. Release the
3. Press and release the
F key until the word Func is displayed.
F key. The display changes to show 1.SET.1.
F key once. The display changes to read 2.SEL.2. This indicates that the comparator
function is currently active.
4. Press and release the
T key once. The display changes to read 2.SEL.1. This indicates that the comparator
function is inactive. Any existing over/under parameters remain in memory, but are inactive.
5. Press and hold the
F key for approximately four seconds. The display returns the conventional weighing
mode and the comparator function is inoperative.
Weighing Operations11
Page 16
2.9Operating the Counting Function
The QB Series balance is capable of functioning as a counting device, enabling it to determine quantity as
opposed to weight. In order to perform this function, the balance must be programmed with the average piece
weight of the sample being evaluated. Once programmed, the balance performs the following calculation to
determine the total number of pieces to display:
Weight on platform - tare weight (if applicable) / average piece weight = displayed piece count
NOTE: When the counting mode is activated, the comparator function is automatically deactivated. Should you desire to
return to the comparator function, all of the most recently entered parameters are automatically recalled when the
counting function is deactivated. If you need to return to the counting mode at a later time, the last average piece weight
utilized is still active.
2.9.1Counting Function Activation
Prior to using the counting function, the parameters for the operation must be programmed into the balance.
1. Power up the balance (see Section 1.7 on page 5).
2. The balance displays the default weighing mode. If the balance does not show a
normal start sequence, press and release the
3. Press and hold the
4. Release the
F key until the word Func is displayed.
F key. The display changes to show either 1.SET.1 or 1.SET.2. If the display shows 1.SET.2,
continue below with step 5. If the display shows
1.SET.2 and then continue with step 5 below.
5. Press and release the
P key once. The display advances to the weighing mode with a 0 value and the P
T key.
1.SET.1, press the T key once to change the value to
(counting pieces) indicator illuminated.
2.9.2Programming the Average Piece Weight
To program the average piece weight, use the steps outlined below.
NOTE: The QB series balance can determine the average piece weight by evaluating initial samples of 10, 30, 50, or 100
pieces. We strongly recommend that you always use a sample of 100 pieces to ensure the most uniform sample piece
calculation resulting in the most accurate counting operation possible based on the samples provided. It is important to
remember that an accurate counting function relies on a uniformity of parts being counted. If there is a large variance
from piece to piece, that error is introduced into the counting function with a resultant error in the value displayed by the
balance.
1. Press and release the T key once to ensure that the value on the display is 0.
2. If you desire to use a container to hold the parts to be counted, place the empty container on the weighing
platform.
3. The display may change to indicate a counting value. Press and release the
0 value.
a
NOTES:
•The tare value must be ≥ +/- 4 display divisions from zero. If the tare value is ≤ +/- 3 divisions from
zero, the balance automatically re-establishes a new zero setting before the T button can be
activated. The tare function is not available. To disable this “auto-zero” feature, please refer to
Section 4.0 on page 15.
•Whenever the tare function is activated on a QB balance, the effective weighing range of the device
is reduced by the value of the tare. For example, on a QB-600 with a total capacity of 600g, if a tare
of 100g is entered into the device, the total effective NET weighing range of the balance is only 500g
(total rated capacity [600g] - tare value [100g] = 500g. Any attempt to weigh over the effective
weighing range results in a “o-Err” error message.
T key to return the display to
4. Press and hold the F key until the display reads U Set. Then release the F key.
5. The display shows
on xxx where “xxx” is either 10, 30, 50, or 100 depending on what value was last used
to determine an average piece weight.
6. Repeatedly press and release the
T key until the display reads on 100 (recommended) or the quantity you
desire to use as a sample.
7. With the display flashing the number of pieces you have chosen to use as a sample, place those items on
the weighing platform or in the empty container.
8. Press and release the
F key once. The display momentarily blanks, except for the flashing M indicator
which confirms that the balance is memorizing the average piece weight.
0.00 value after the
12QB Series Operation Manual
Page 17
NOTES:
•It is possible to use a sample that is too light for the balance to accurately calculate an average
piece weight. If you notice that the display momentarily shows ADD followed by a display that
includes an illuminated UNDER annunciator, you have attempted to use a sample that is too light for
the device to accurately calculate an average piece weight. Continue adding additional samples and
pressing the F key until the UNDER annunciator goes out or simply return to step 4 and utilize a
larger sample.You can abort a sample evaluation at any time by pressing and releasing the P key
once. The display momentarily displays STOP followed by a counting value display.
•The minimum average piece weight must be ≥ the minimum increment (readability) of the balance
being used. If you notice that the display momentarily shows L-Err, the commodity in question can
not be counted with this balance.
9. The display then resumes flashing. To begin counting based on the current average piece weight
calculation, press and release the
P key once. The display stops flashing and the M indicator disappears.
The balance shows the current quantity of the commodity on the weighing platform. Proceed to
Section 2.9.4 on page 13.
10. To utilize the QB’s exclusive APWR (Automatic Piece Weight Refinement) program, continue with
Section 2.9.3 on page 13 (recommended).
2.9.3Using the APWR (Automatic Piece Weight Refinement) Program
The APWR program is an exclusive feature of the QB Series. When utilized, its unique capabilities refine the
average piece weight as often as desired to take into consideration the largest sample possible. As the average
piece weight of a commodity is refined, the overall counting results are dramatically improved. Within the net
weighing range of the balance, there is no limit as to how many samples are used.
1. Complete steps 1 – 8 in Section 2.9.2 on page 12.
2. With the display flashing the sample count,
M indicator and P indicator, add any number of additional
samples to those already being weighed.
3. Press and release the
F key once. The display momentarily blanks, except for the flashing M indicator
which confirms that the balance is calculating and updating the previously memorized average piece
weight. Repeat steps 2 and 3 as many times as desired.
4. When you are ready to exit the APWR, press and release the
M indicator disappears.
the
P key once. The display stops flashing the
2.9.4Using the Counting Mode
Use the following steps to perform piece count operations on the QB Series balance.
1. Complete the steps in Section 2.9.1 – 2.9.3 as appropriate.
2. Remove the test sample from the platform.
3. If the balance does not show a
4. If applicable, place an empty container on the platform and press the
0 value, press and release the T key. The balance returns to a 0 value.
T key. The balance returns to a 0
reading (see Section 2.4 on page 7).
5. Gently place the items to be counted on the weighing platform or in the empty container as appropriate.
6. The
P annunciator begins to flash as the display updates. When the P annunciator stops flashing, you can
read the quantity of commodity being evaluated.
2.9.5Deactivating the Counting Mode
Use the following steps to deactivate the count mode of the QB Series balance.
1. Press and hold the
2. Release the
3. Press and release the
4. Press and release the
F key until the word Func is displayed (U SET appears first, followed by Func).
F key. The display changes to show 1.SET.2.
T key once to change the value to 1.SET.1. This deactivates the counting mode.
P key once. The display advances to the weighing mode with a 0.00 value.
5. Even though the counting mode has been deactivated, the most recently used average piece weight
remains in memory and is recalled when the counting mode is reactivated.
Weighing Operations13
Page 18
3.0Span Calibration
To achieve optimum accuracy from your QB Series balance, it should be calibrated on a regular basis in the
environment in which it is normally used. If the device is moved to another physical environment, recalibrate the
device before using it. A few extra moments invested with this section will ensure virtually error-free
performance of the device.
After properly performing the installation guidelines (see Section 1.3 on page 2 and Section 1.4 on page 3),
follow these calibration steps.
1. Power up the balance (Section 1.7 on page 5).
2. Have ASTM Class 4 test weights available equal to 100% of the nominal capacity (in grams) of the
device. (Weights and weight sets are available from Rice Lake Weighing Systems.)
3. Ensure that there is no load on the platform.
4. Ensure that the device is level and on a stable surface, free from environmental changes. Pay special
attention to possible and sudden changes in humidity, temperature and moving air.
5. Ensure that the balance displays a
6. If the device is currently using any unit of measure other than grams, activate the grams unit of measure
(Section 2.2 on page 6).
7. If the device is currently in the counting mode, reactivate the conventional weighing mode (see
Section 2.9.5 on page 13).
8. Press and hold the
two seconds, the word
9. When the word
F key until the word CAL appears. As you press and hold the F key for approximately
FUNC appears. The word CAL appears after two additional seconds.
CAL appears, release the F key. If you hold the F key down too long, the unit returns to
the default weighing mode. If this occurs, start over at step 8.
10. While pressing and holding the
display shows
11. Press and release the
unit 1 where 1 represents the function code for the grams unit of measure as active.
F key .
12. The device momentarily displays a flashing
zero.
13. After zero calibration, the device automatically advances to read
14. Place a load equal to the nominal capacity of the device on the center of the platform.
15. The
on FS display flashes, indicating that the device is electronically calibrating a full scale load
equivalent to the nominal capacity of the device.
16. When calibration is complete, the device automatically displays the calibrated nominal capacity of the
balance.
17. Remove the test load from the balance.
18. The device automatically returns to a
In some rare cases, automatic calibration may not be possible. In the event of an error message during
calibration, please refer to the table below for the appropriate corrective action.
0 or 0.00 value. If not, press and release the T key once.
T key, press and hold the F key. Release both keys at the same time. The
on 0, indicating an automatic calibration of true mechanical
on F.S.
0.0 value and the default weighing mode.
Error MessageProblemSolution
O-ERR• Calibration load is in excess of the nominal capacity of the
device. The device is out of calibration and is sensing a
100% load being greater than the nominal capacity of the
device.
• The calibration test load is > 100% of the nominal
capacity.
1-ERRCalibration load is less than 1/2 of the nominal capacity of the
device.
2-ERRData error is in excess of 1% of the rated capacity.Contact your RLWS dealer.
Table 3-1. Possible Calibration Process Error Messages
14QB Series Operation Manual
• Change the test load so that it is
equal to 1/2 or 3/4 the nominal
capacity of the device. Then run a
second calibration at 100% nominal
capacity.
• Reduce the test load so that it is
equal to 100% of the nominal
capacity of the device.
Increase the test load so that it is equal to
the nominal capacity of the device.
Page 19
4.0Function Parameters
The parameter menu allows you to activate or deactivate certain features and functions of the QB Series balance.
The following section provides information on the functions of the QB and how to view and change those
functions. Table 4-1 below describes the various functions of the QB.
NOTE: Bold type indicates factory setting.
DisplaySettingFunctionDescription
1.SET
2.SEL
21.CO
22.LI
23BU
3.AO
4.AP
1Mode SelectionWeighing mode only
2Counting mode only
1Comparator
Function
2Comparator ON
1Conditional
Response
2Active when balance is stable ONLY (recommended)
Parameter
(available only when
comparator is set
to ON)
0Range Response
Parameter
1Active for the entire weighing range (all data)
(available only when
comparator is set
to ON)
0Response
Condition (available
1Flashing arrow next to UNDER display legend.
only when the
comparator is set
2Flashing arrow next to ACCEPT display legend.
to ON)
3Flashing arrow next to OVER display legend.
4Flashing arrow next to UNDER and ACCEPT display legends.
5Flashing arrow next to ACCEPT and OVER display legends.
6Flashing arrow next to ACCEPT, UNDER, and OVER display legends.
0Auto ZeroAuto zero function off
1Auto zero function on (± 3d)
0Sleep ModeSleep mode ON (available only when internal rechargeable battery installed).
1Sleep mode OFF
Comparator OFF
Active when balance is stable and in motion
Inactive for data near zero and for negative data (recommended)
Solid arrow next to appropriate display legend
Solid arrow next to ACCEPT and OVER display legends.
Solid arrow next to UNDER and OVER display legends.
Solid arrow next to UNDER and ACCEPT display legends
Solid arrow next to OVER display legend.
Solid arrow next to UNDER display legend.
Turns device off after three minutes of inactivity when running off of battery power.
5.rE
6.IF
1Response SpeedVery fast (extremely stable environment)
2Fast (stable environment)
2Continuous output when data is stable. No output when data is unstable.
3Data output x 1 each time P key is pressed and released in weighing mode
4Automatic output x 1 after data stablized
5Automatic output x 1 of stable data; output prohibited if data unstable
6Automatic output x 1 of stable data; continuous output of unstable data
7Data output x 1 each time P key is pressed and released with data stablized
1Baud rate (available
only when interface
22400 bps
option is set to ON)
34800 bps
0Parity bit selection
(avalable only when
1Odd
interface option is
2Even
set to ON)
1Unit of measuregram
2kilogram
3carat
4decimal ounce
5pound
6troy ounce
7pennyweight
8grain
9Hong Kong tael
ASingapore tael
bTaiwanese tael
Cmomme
Output prohibited (stop)
NOTE: When the RS-232C interface is installed, output interval of data in
constant transmission is 0.1 – 1 second, depending on the weighing condition
and response parameters of 5.rE.X above.
NOTE: When RS-232C interface installed, output interval of data in constant
transmission is 0.1 – 1 second, depending on the weighing condition and
response parameters of 5.rE.X above.
NOTE: Only one transmission is sent after a stabilization from an increasing load.
No additional output will be generated until the device returns to a 0.00 value and
resets for the next load.
1200 bps
NoneAvailable only when option 6.IF.2 is selected
(parity bit available only in 7-digit format)
16QB Series Operation Manual
Table 4-1. Function List (Continued)
Page 21
4.1Viewing Parameters
The following information provides a step-by-step procedure to view the parameters of the QB Series balance.
1. Turn the balance on (see Section 1.7 on page 5).
2. If the device does not show a
3. Turn the counting function off (see Section 2.9.5 on page 13).
4. Press and hold the
F key until the display reads FUNC.
5. Immediately release the
the P key to return to the default weighing mode and return to step 4.
NOTE: If you hold the F key down too long, the display changes from FUNC to CAL. Simply press and release
6. When you release the F key, the display reads 1.SEt.X where “X” is a value of either 1 or 2. (see
Table 4-1 on page 15 for a definition of these parameters).
7. Press and release the
F key again to advance to the next parameter display. Compare the information in
the display to the information in Table 4-1 on page 15.
8. Continue to press and release the
0.0 value after the normal start sequence, press and release the P key.
F key .
F key to scroll through each of the parameter function displays.
Weight Unit
1.SET. 1
ON
P
OFF
Zero Tracking
➩
F
T
1.SEL. 2
ON
OFF
F
P
➩
T
Auto-power Off
21.C . 2
ON
OFF
F
P
➩
T
Response Speed
22.L . 1
ON
OFF
F
P
T
➩
Response Speed
5. E. 3
ON
P
OFF
F
Auto-power Off
➩➩
4. AP. 1
ON
T
P
OFF
T
F
Response Speed
3. A.O 1
ON
P
OFF
F
➩
T
Response Speed
23. . 2
ON
OFF
F
P
T
➩
Weight Unit
6. IF. 1
ON
P
OFF
*Only applies when device
is RS-232 equipped.
T
F
Output Control*
6 1. . . 0
ON
OFF
F
P
62. .L. 1
T
Weight Unit
➩➩➩
ON
OFF
F
P
T
7. . 1
Figure 4-1. Function Parameters
9. When you press and release the F key after the last parameter, the device returns to the default weighing
mode.
Weight Unit
ON
OFF
F
P
T
4.2Changing Parameters
The following information provides a step-by-step procedure for changing the parameters of the QB Series
balance.
1. Turn the balance on (see Section 1.7 on page 5).
2. Ensure there is no load on the platform.
3. Ensure that the device is in the normal weighing mode, not the counting mode. To turn off the counting
mode and return to the normal weighing mode, see Section 2.9.5 on page 13.
4. If the device does not show a
5. Press and hold the
6. Immediately release the
7. If you hold the
to return to the default weighing mode and return to step 4.
8. When you release the
F key until the display reads FUNC.
F key down too long, the display changes from FUNC to CAL. Press and release the P key
F key, the display reads 1.SEt. 1.
0.0 value after the normal start sequence, press and release the T key.
F key .
Function Parameters17
Page 22
9. Using Table 4-1 on page 15 as a guide, scroll to the function you wish to change by repeatedly pressing
F key. If you scroll through all of the functions, the device automatically returns to the conventional
the
weighing mode. To return to the function mode, return to step 5.
10. Once the function you wish to change is displayed, repeatedly press the
T key until the parameter you
want to use is displayed.
11. To continue on to another function, return to step 9.
12. If you do not need to change another function, press the
P key to confirm your settings and return to the
conventional weighing mode.
18QB Series Operation Manual
Page 23
5.0Specifications
This section contains tables listing common specifications and model specific data for the QB Series balance.
Table 5-1 below provides specifications of each of the models in the QB Series of balances.
The QB Series may be equipped with an optional RS-232C feature (PN 42174), which is compatible for
communication with printers and computers. When the balance is connected directly to a printer, displayed data
can be output either via the
to Table 4-1 on page 15 for a list of parameters available.
NOTE: The optional battery pack (PN 42169) cannot be used simultaneously with the RS-232C option.
P key on the balance, or automatically when certain conditions have been met. Refer
5.1.1Pin Function
Pin NumberSignalOutput/InputFunction
1TareInputExternal tare input (1)
2DTROutputHigh level with scale power ON
3RXDInputReceiving data
4TXDOutputTransmission data
5GNDN/ASignal ground
Table 5-2. RS-232C Pin Function
Pin 2
Pin 4
Pin 1
Pin 5
Pin 3
Balance
NOTE: Printer side of connector
to show correct pin number
orientation.
side
Figure 5-1. RS-232 Connector Schematic
Printer
side
20QB Series Operation Manual
Page 25
5.1.2RS-232C Commands
One of the following two formats is selectable in the function setting (see Section 4.0 on page 15).
D1 ~ D7, 7 words with 6-digit format (see Table 5-3)
D1 ~ D8, 8 words with 7-digit format (see Table 5-4)
DCodeContents
0~930H ~ 39HData 0 ~ 9
Data
(Max 6 digits with 6-digit format or max 7 digits with
7-digit format)
.2EHDecimal point (floating)
sp20HSpace; leading zero suppression
Table 5-5. Data
Polarity: P1 = 1 word
P1CodeContents
+2BHData is 0 or positive
–2DHData is negative
sp20HData is 0 or positive
Table 5-6. Polarity
(U1, U2 = 2 words, based on ASCII codes)
U1U2ContentsSymbol
kgkilogramkg
pcquantity(pcs)
momommemom
lbpoundlb
dwpennyweightdwt
tltaeltl
Units
Table 5-7. Units
Specifications21
Page 26
(U1, U2 = 2 words, based on ASCII codes)
spggramg
ctcaratct
ozounceoz
ottroy ounceozt
gggraingr
Units
Table 5-7. Units (Continued)
Conditional Response Parameter For Comparator Function
S1CodeContents
l4CHLow (under)
g47HGood (accept)
h48HHigh (over)
(S1 = 1 word)
Table 5-8. Conditional Response Parameter for Comparator Function
Status of Data
(S2 = 1 word)
S2CodeContents
s53HData stabilized
u55HData unstabilized
e45HErroneous data; all data unreliable except S2 (device
display reads either “o-Err” or “u-Err”)
Table 5-9. Input Command, Data Reception
The balance can be controlled by transmitting commands from an external device and is independent of the
timing on the device due to the full duplex nature of the interface. If the command is properly received, the
balance transmits acknowledgement (ACK, code 06H). If the command is unacceptable, the balance transmits
non-acceptance (NAK, code 15H).
In the conventional display weighing mode (non-counting mode), the device responds with ACK or NAK within
one second of receipt of command.
When transmitting function code changes, such as unit of measure, conditional response parameter, or
calibration commands, ACK or NAK are transmitted after completion of the requested command process (e.g.,
when the new parameter has been set and confirmed). Commands that are sent before ACK or NAK are
transmitted from the device are ignored.
Input format command consists of four (4) words, including terminators (CR=0DH or LF=0AH) as follows:
1234
C1C2CRLF
External tare command is as follows:
C1C2CODEContents
Tsp54h 20HTare Command
22QB Series Operation Manual
Page 27
5.1.3Output Control Settings
C1C2Code 1Code 2Contents
00 4FH30HInterface off, no printing available
01 4FH31HAutomatic continuous printing of zero display, unstable and stable weighing values
02 4FH32H
03 4FH33H
04 4FH34HAuto-printing of stabilized weight value only
05 4FH35HAuto-printing x 1 of stabilized zero value and stabilized weight value
06 4FH36HAuto-printing x 1 of any weight stable display value
07 4FH37H
08 4FH38HAutomatic output x 1 of stable data; continuous output of unstable data
09 4FH39HData output x 1 each time P key is pressed and released with data stabilized
Automatic continuous printing of stabilized data only (for example, stable
stable weight value)
Pushbutton printing from balance P key of stable
Pushbutton printing from balance P key of stable
0 value and stable weight value
0 value and stable weight value
0 value and
Table 5-10. Output Control Settings
NOTES:
• The performance of the device can be controlled by output controls with commands 00-07 and by the parameter
settings (see Table 4-1 on page 15). Commands 08 and 09 are limited to an input command string.
• Once any command is initiated, that status is maintained until a new command is initiated. However , once the power is
turned off, the output control status returns to the setting previous to the last input command.
5.1.4Interface Specifications
ItemSpecification
Transmission methodRS-232C serial data, random access
Transmission speed1200/2400/4800 bps; user selectable; (see Table 4-1 on page 15)
Transmission codeASCII; 8-bit
Signal levelEIA RS-232C;
High Level - Data logic “0” + 5 VDC to +15 VDC
Low Level - Data logic “1” -5 VDC to -15 VDC
Contents of word8-bit word based on ASCII standard character codes; 1 start bit; 2 stop bits; 0/1 parity bit
Parity bitNull, even or odd parity
Table 5-11. RS-232C Interface Specifications
5.2Connecting to a Computer
Connecting the balance to a computer enables you to operate the balance from the computer as well as receive
data such as displayed weight and unit of measure. The following sections describe the hardware and output
signal formats provided with the balance.
5.2.1Hardware
A 5-pin DIN connector for interfacing to other devices (see Figure 5-2 on page 24) is located on the rear of the
QB Series balances equipped with the optional RS-232C interface.
The balance does not output any data unless pin 2 is held in an ON state (+5 VDC to +15 VDC). Interfaces
requiring the CTS handshake may tie pins 5, 4, and 8 on the computer side to defeat it.
Specifications23
Page 28
TXD2
RXD3
DB-25
Connection
GND7
DCD8
RTS4
CTS5
DSR6
DTR 20
Jump pins
8, 4, and 5
Jump pins
6 and 20
Figure 5-2. Connection to a Computer
5.3Accessories
RLWS Part NumberDescription
42566115 VAC adapter for QB/MB Series balances
42168230 VAC adapter for MB/QB/IB Series balances
42171Internal rechargeable battery for QB Series (incompatible with PN 42174 below)
42174RS-232C interface for QB Series (incompatible with PN 42171 above) includes DIN-5 pin connector
(cable not included)
42177Glass breeze break for QB Series (scales with round weighing pans only)
42181In-use cover
41555Limit output pack
3RXD
4TXD
5GND
1
2DTR
DIN-5
Balance
Connections
42194Additional operating manual
Table 5-12. MB Series Balance Accessories
24QB Series Operation Manual
Page 29
6.0Troubleshooting
The following section contains a troubleshooting chart for the QB Series scale. Refer to this chart to locate
causes and corrective actions to various conditions.
ConditionCauseCorrective Action
Display does not
stabilize
Displayed value is in
error
Weight value is drifting
Linearity error• Mechanical adjustment has changed• Consult your local Rice Lake Weighing
Display is blank• Adapter is not connected
Comparator limits can
not be established
•Wind or vibration is affecting scale• Set up balance elsewhere or under a
• Scale is set on unstable surface• Set the scale on a firm surface that is free
• Span is out of calibration• Level the scale
•Tare value entered incorrectly• Clear and re-enter tare value
• Balance is not set level• Set balance level
• Span is out of calibration• Perform span adjustment
• Interference with weigh pan or pan
support
• Battery needs recharging
• Auto Power OFF is activated.
• Device is not set to comparator mode
• Reference value is greater than the
nominal capacity of the device
• Lower limit value is ≥ upper limit value
protective hood
from vibrations
• Perform span and zero adjustment
• Check for foreign material under weigh pan
or weigh pan support
Systems Service distributor
• Connect AC adapter to the scale and 120
VAC outlet
• Recharge battery
•Press the ON/OFF key
•Turn on comparator mode
• Change reference value
• Change lower limit value
Balance does not weigh
to nominal capacity
b
-ERR
L-ERR
O-ERR
U-ERR
•Tare + sample weight exceeds capacity
•(Weight range = maximum capacity –
tare weight)
• Electronic malfunction due to
electro-static discharge (ESD) or
electro-magnetic interference (EMI)
• In counting mode, the unit weight of the
samples is less than the minimum
division of the device
• Capacity is exceeded• Use a test mass to ensure the balance can
• Interference with the weighing pan or
weighing pan support
• Add the individual weights of the tare and
the commodity to determine if together they
are more than the nominal capacity of the
device
• Consult your local Rice Lake Weighing
Systems distributor
• This commodity can not be counted with
this device
weigh to its rated capacity
• Ensure that the commodity being weighed
is not in excess to the nominal capacity of
the balance
• Consult your local Rice Lake Weighing
Systems distributor
• Remove interference
Table 6-1. Troubleshooting Chart
Troubleshooting25
Page 30
ConditionCauseCorrective Action
1 - ERR
2-ERR
3-ERR
4-ERR
U - ERR
• The calibration weight is less than 1/2
full capacity
• The data error exceeds 1%
• Defective weighing mechanism
• Calibration has started with a load on
the platter
• Material wedged underneath the
platform support plate
Table 6-1. Troubleshooting Chart (Continued)
• Reinitiate calibration with a test mass equal
to the nominal capacity of the balance
• Consult your local Rice Lake Weighing
Systems distributor
• Consult your local Rice Lake Weighing
Systems distributor
• Remove load and reinitiate calibration
process
• Consult your local Rice Lake Weighing
Systems distributor
• Remove platform and platform support
plate, then clean and reassemble
26QB Series Operation Manual
Page 31
7.0Dimensions
Figure 7-1 below shows the dimensions of the QB Series balance.
Ø140
310
208
71
(86)
Figure 7-1. QB Series Dimensions
190
310
208
71
(87)
Dimensions27
Page 32
QB Series Limited Warranty
Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems manufactured and sold by
RLWS and properly installed by an authorized RLWS Distributor or Original Equipment Manufacturer (OEM)
will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems
and components are warranted against defects in materials and workmanship for five years.
RLWS warrants that the equipment sold hereunder will conform to the current written specifications authorized
by RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipment
fails to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within the
warranty period subject to the following conditions:
•Upon discovery by Buyer of such nonconformity, RLWS will be given prompt written notice with a
detailed explanation of the alleged deficiencies.
•Individual electronic components returned to RLWS for warranty purposes must be packaged to
prevent electrostatic discharge (ESD) damage in shipment. Packaging requirements are listed in a
publication, Protecting Your Components From Static Damage in Shipment, available from RLWS
Equipment Return Department.
•Examination of such equipment by RLWS confirms that the nonconformity actually exists, and was
not caused by accident, misuse, neglect, alteration, improper installation, improper repair or
improper testing; RLWS shall be the sole judge of all alleged non-conformities.
•Such equipment has not been modified, altered, or changed by any person other than RLWS or its
duly authorized repair agents.
•RLWS will have a reasonable time to repair or replace the defectiv e equipment. Buyer is responsible
for shipping charges both ways.
•In no event will RLWS be responsible for travel time or on-location repairs, including assembly or
disassembly of equipment, nor will RLWS be liable for the cost of any repairs made by others.
THESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING WITHOUT LIMITATION WARRANTIES OF MERCHANTABILITY OR FITNESS FOR
A PARTICULAR PURPOSE. NEITHER RLWS NOR DISTRIBUTOR WILL, IN ANY EVENT, BE
LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
RLWS AND BUYER AGREE THAT RLWS’ SOLE AND EXCLUSIVE LIABILITY HEREUNDER IS
LIMITED TO REPAIR OR REPLACEMENT OF SUCH GOODS. IN ACCEPTING THIS WARRANTY,
THE BUYER W AIVES ANY AND ALL OTHER CLAIMS TO W ARRANTY .
SHOULD THE SELLER BE OTHER THAN RLWS, THE BUYER AGREES TO LOOK ONLY TO THE
SELLER FOR WARRANTY CLAIMS.
NO TERMS, CONDITIONS, UNDERSTANDING, OR AGREEMENTS PURPORTING TO MODIFY
THE TERMS OF THIS WARRANTY SHALL HAVE ANY LEGAL EFFECT UNLESS MADE IN
WRITING AND SIGNED BY A CORPORATE OFFICER OF RLWS AND THE BUYER.