The undersigned, representing the following manufacturer
Ohaus Corporation
19A Chapin Road
PO Box 2033
Pine Brook, NJ 07058 USA
hereby declares that the following products are in conformity with the EEC directives listed below (including any and all modifications).
Balance models: EP12001, EP12001C, EP22001, EP22001C, EP32001, EP32001C
Marked with:DirectiveStandard
Gekennzeichnet mit:RichtlinieNorm
Munis de la mention:DirectiveNorme
Contrassegnati con laDirectivaNorma
marcatura:
Con el distintivo:DirettivaNorma
EU 73/23/EEC IEC 1010 -1:1990 + A1: 92 + A2: 95
xx
Low Voltage
Niederspannung
Basse tension
Baja tensión
Bassa tensione
Safety requirements for electrical equipment for measurement,
control and laboratory use – Part 1: General requirements
Sicherheitsbestimmungen für elektrische Meß-, Steuer-, Regel- und
Laborgeräte — Teil 1: Allgemeine Anforderungen
Règles de sécurité pour appareils électriques de mesurage, de
régulation et de laboratoire – Partie 1: Prescriptions générales
Requisitos de seguridad de equipos eléctricos de medida, control y
uso en laboratorio – Parte 1: Requisitos generales
Prescrizioni di sicurezza per apparecchi elettrici di misura, controllo e
per utilizzo in laboratorio — Parte 1: Prescrizioni generali
EN61326: 1997 + A1: 1998
Electrical equipment for measurement, control and laboratory use
(Class B)
Elektrische Betriebsmittel für Leittechnik und Laboreinsatz —EMVAnforderungen (Class B)
Matériels électriques de mesure, de commande et de laboratoire —
Prescriptions relatives à la CEM (Class B)
Equipo eléctrico de medida, control y uso en laboratorio — Requisitos
de compatibilidad electromagnética (Class B)
Apparecchi elettrici di misura, controllo e laboratorio — Prescrizioni
di compatibilità elettromagnetica (Class B)
EN45501:1992
Non-automatic weighing instruments
Nichtautomatische Wiegevorrichtungen
Instruments de pesage à fonctionnement non automatique
Instrumentos de pesaje de funcionamiento no automático
Strumenti per pesare a funzionamento non automatico
ISO 9001 Registration for Ohaus Corporation
tional, BVQI, and was awarded ISO 9001 registration. This certifies that Ohaus Corporation, USA, has a quality system that conforms with the international
standards for quality management and quality assurance (ISO 9000 series). Repeat audits are carried out by BVQI at intervals to check that the quality
system is operated in the proper manner.
Ted Xia
President
Ohaus Corporation
Pine Brook, NJ USA
Date: July 29, 2003
. .
.
Ohaus Corporation, USA, was examined and evaluated in 1994 by the Bureau Veritas Quality Interna-
. .
Urs Müller
General Manager
Ohaus Europe
Greifensee, Switzerland
Additional Standards
CAN/CSA-C22.2 No. 1010.1-92; UL Std. No. 3101-1
Safety requirements for Electrical Equip. for measurement, Control and Laboratory Use, Part 1; General Requirements
FCC
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the
interference at his own expense.
This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada
3.1Overview of Controls............................................................................................................................. EN-7
3.2Overview of Display Indicator................................................................................................................ EN-9
3.4.11 Statistical Quality Control (SQC) ............................................................................................... EN-28
3.4.12 Density ................................................................................................................................... EN-30
4.3Error Codes List ................................................................................................................................. EN-46
4.4Service Information ............................................................................................................................ EN-47
4.5Replacement Parts ............................................................................................................................. EN-47
5. TECHNICAL DATA ..................................................................................................................................... EN-48
Thank you for deciding to purchase a Explorer® Pro Balance
from Ohaus. Explorer Pro’s software design allows direct
access to 11 application modes and all menus. The backlit
dot matrix display, is capable of displaying either limited or
multiple data fields at each application. Explorer Pro’s
standard AutoCal™ automatically calibrates the balance due
to temperature changes that may affect calibration. Panel
controls and soft key functions on the display clearly indicate
functions and data. Operation is extremely simplified, you
enter a menu, select an item, modify the item according to
menu selections and exit.
Capacities from 12,000 grams to 32,000 grams are available.
To ensure you make full use of the possibilities offered by
your Explorer Pro balance, we advise you to read through
these operating instructions.
2.INSTALLATION
2.1 Unpacking
Open the package and remove the instrument and the
accessories. Check the completeness of the delivery. The
following accessories are part of the standard equipment of your
new Explorer Pro balance.
• AC Power
• Instruction Manual
• Warranty Card
• Weigh Below Hook
Remove packing material from the instrument.
Check the instrument for transport damage. Immediately
inform your Ohaus dealer if you have complaints or parts
are missing.
Store all parts of the packaging. This packaging guarantees
the best possible protection for the transport of your
instrument.
1.2 Features
• Ready to weigh without complicated setup
• 3 soft keys are application mode specific.
• Operating languages include English, Spanish, French,
German and Italian.
• Simplified menu navigation and balance setup.
• Capacity Bar guide and Fill bar guide.
• Flexibility to display either simple weighing results or
more advanced results including piece count, average piece
weight, sample size, tare weight, weight, under and over.
• Dot Matrix display with backlight
• Weighing, Parts Counting, Animal Weighing, Percent
Weighing, Check Weighing, Gross/Net/Tare Weighing and
Filling.
• Do not operate the balance around corrosive fumes.
• Only use the adapter provided with the balance.
• Do not try to service the Explorer Pro balance.
EN-4 Explorer
2.2Selecting the Location
2.2.1Balance Location
The balance should always be used in an environment which is
free from excessive air currents, corrosives, vibration, and
temperature or humidity extremes. These factors will affect
displayed weight readings.
DO NOT install the balance:
• Next to open windows or doors causing drafts or rapid
temperature changes.
• Near air conditioning or heat vents.
• Near vibrating, rotating or reciprocating equipment.
• Near magnetic fields or equipment that generates magnetic
fields.
• On an unlevel work surface.
• Allow sufficient space around the instrument for ease of
operation and keep away from radiating heat sources.
Pro
2.2.2Leveling the Balance
Exact horizontal positioning and stable installation are
prerequisites for repeatable results. To compensate for small
irregularities or inclinations at the location, the instrument can
be leveled.
For exact horizontal positioning, the balance is equipped with a
level indicator located at the front on the control panel and two
leveling feet located at the rear of the balance.
Position the balance in the intended operating location. Adjust
the leveling feet at the rear of the balance until the air bubble in
the indicator is centered.
NOTE: The instrument should be leveled each time its location is
changed.
Correct Incorrect
Leveling
Indicator In
Rear
Leveling
Feet
WEIGHING
UNITS
*
01/01/03 03:05:45: PM
2nd UNITS
0g
0.00
g
0.00g
Explorer
Pro
2.3Connecting Power and Communications
2.3.1AC Adapter Installation
Connect the AC Adapter supplied to the three pin connector located
at the rear of the balance.
The balance is now ready for operation.
2.3.2RS232 Interface
Explorer Pro balances are equipped with a bidirectional RS232
compatible interface for communication with serial printers and
computers. When the balance is connected directly to a printer,
displayed data can be output at any time by simply pressing the
Print button, or by using the Auto Print feature. Connecting the
balance to a computer enables you to operate the balance from
the computer, as well as receive data such as displayed weight,
weighing mode, stability status, etc.
EN-5
AC Adapter Connection
Left Side of Balance
12345
6789
Male Connector
The following sections describe the hardware and software
provided with the balance
Hardware
On the rear of the balance, the right-hand, 9-pin male subminiature
“D” connector is provided for interfacing to other devices. The
pinout and pin connections are shown in the adjacent illustration.
Refer to paragraph 3.5.6 for setup.
The balance is equipped with hardware handshaking, it will not
output any data unless pin 5 (CTS) is held in a high state (+3 to
+15 V dc). Interfaces not utilizing the CTS handshake may tie pin
5 to pin 6 to defeat it.
Output Formats
Data output can be initiated in one of three ways: 1) By pressing
PRINT; 2) Using the Auto Print feature; 3) Sending a print
command (“P”) from a computer.
1 N/C
2 Data Out (TXD)
3 Data In (RXD)
4N/C
5 Clear To Send (CTS)
6 Data Terminal Ready (DTR)
7 Ground
8 Request To Send (RTS)
9N/C
Additional information is located in Section 5, Technical Data
which contains the RS232 Command Table.
EN-6 Explorer
2.3.3 Initial Calibration
Calibrating from the Weighing Screen
When the balance is first turned on, three soft keys appear on
the weighing screen. The CALIBRATE soft key is highlighted.
This permits calibration immediately without entering the
CALIBRATION menu.
WEIGHING 01/01/03 03:05:45: PM
UNITS
2nd UNITS
CALIBRATE
Pro
*
0g
Press the Enter button with the CALIBRATE soft key highlighted.
0.00
0.00g
g
6100g
Explorer
Pro
3OPERATION
3.1Overview of Controls
EN-7
1
WEIGHING 01/01/03 03:05:45: PM
UNITS
12
*
11
10
9
0g
8
No.DesignationFunction
1 IPower On/ Off button.
2nd UNITS
CALIBRATE
0.00
0.00g
g
6100g
7
10
2
3
4
5
9
6
7
2 buttonWhen pressed in Menu mode, single press moves menu selection bar in an up direction and
highlights the activated field or changes the settings of a selected field in increasing order.
When at an alphanumeric field, then the number or letter is incremented.
3 buttonWhen pressed in Menu mode, single press moves the cursor within a alphanumeric field
to the left (example date 07/02/2003). Selection does not wrap at end.
In application mode, single press moves to the next soft key selection to the left.
4 buttonWhen pressed in Menu mode, single press moves the cursor within a alphanumeric field
to the right (example date 07/02/2003). Selection does not wrap at end.
In application mode, single press moves to the next soft key selection to the right.
EN-8 Explorer
No.Designation Function
5 buttonWhen pressed In Menu mode, single press moves menu selection bar in a down
direction and highlights the selected field or changes the settings of a selected field in
decreasing order.
When at an alphanumeric field, then the number or letter is decremented.
6Enter buttonWhen pressed in application modes, functions as an “Enter” button to accept Soft-key selection.
When pressed in menu, functions as an “Accept” button for accepting and saving
parameter of selected items.
7 & 8 >O/T< buttonsWhen pressed, performs tare function or zero function.
9LCD DisplayBacklit LCD display provides all necessary indications for operation.
Pro
10Mode buttonWhen pressed, causes the balance to cycle through all active application modes.
11Print buttonWhen pressed in any application mode, causes a print function to occur.
12Menu buttonWhen pressed in any application mode, will bring up the Menu screen. Pressing the
button within a Menu screen will return to the application mode.
Explorer
Pro
3.2Overview of Display Indicator
EN-9
12
WEIGHING 01/01/03 03:05:45: PM
UNITS
Message Line
13
12
11
No.DesignationFunction
1ApplicationIndicates the active application by function or library name.
2DateIndicates current date when properly set.
*
-
>0<
Result Line Result Line
Result Line Result Line
Result Line Result Line
0g
200.0000
2nd UNITSCALIBRATE
10
3
dwt
6100g
4
5
6
7
8
9
3TimeIndicates current time when properly set.
4Soft KeysThree Soft-keys are displayed in the upper display area. Depending on the application mode
selected, the functionality changes. A Soft-key can be selected by using the left and right
arrow buttons. Pressing the Enter button on a highlighted soft-key will activate the
function.
5Message LineContains instructional messages.
6Numerical DisplayIndicates primary weight/load.
7Unit of measureIndicates active unit of measure.
8Result Lines 1-6Six lines that contain balance results.
9CapacityIndicates full balance capacity.
10Bar GraphShaded area indicates capacity used for, under, accept, over for check weighing application,
or target values for filling application.
11->0<-Indicates center of zero (only active in legal for trade).
12Indicates negative value.
13Stability indicator.
*
EN-10 Explorer
Pro
3.3Menu
3.3.1Menu Structure
The Explorer Pro balance utilizes a menu structure that permits
entering various menus by using a dedicated Menu button.
Pressing the Menu button allows access to additional sub
menus. See illustration below.
When the balance is first turned on, the main weighing screen is
displayed. To enter the menu and to change parameters, the
following controls are used:
To select a highlighted menu, press the Enter button. Each menu
item contains a display that allows setting balance parameters
to individual requirements. These are described in detail in section
3.5 Balance Settings.
Section 3.5.1 describes Calibration procedures and section 3.4
describes the Applications.
3.3.3Turning on the Balance
The Explorer Pro balance is ready to operate after the
installation procedures are performed. When the balance is
first turned on and it completes its checks, it can be used to
weigh or tare materials without setting the menus.
It is recommended that you read this manual carefully and set
the balance to operate to your specific applications before using.
Power On/Off
To turn the balance ON, press the On/Off button. To turn OFF,
press the button again.
Menu button - When pressed, enters the Menu.
Enter button- When pressed, accepts the menu field selections.
- Scrolls up through the menus.
-Scrolls down through the menus.
- Moves the cursor within the field to the left.
- Moves the cursor within the field to the right.
There are two ways to exit the Menu; one is to scroll to Exit and
press the Enter button, the other is to press the Menu button. All
changes will be saved.
Stabilization
Before initially using the balance, allow time for it to adjust to its
new environment. The balance only requires to be plugged in to
warm up.
Recommended warm up period is twenty (20) minutes. Analytical
Class I balances require at least 2 hours.
The internal circuits of the balance are powered whenever it is
plugged into a power source.
Explorer
Pro
EN-11
3.4Applications
The Explorer Pro balance contains Weighing, Parts Counting,
Animal Weighing, Percent Weighing, Check Weighing, G/N/T,
Statistics, SQC, Formulation, Differential Weighing and Filling
applications. Before using any of these applications, they
must be turned on first before they can be accessed. A
choice can be made to turn on or off as many of these
applications as required. The balance default setting has the
weighing mode turned on and all other modes are off.
To select applications to be turned on or off, press the Menu
button and using the arrow buttons, scroll to Application Modes,
then press the Enter button.
The Explorer Pro balance is shipped with grams only enabled.
The balance can be used immediately after calibration has
been performed. When the balance is to be used with other
units of measure, the desired unit must be enabled.
USE
Zero the balance. Place objects or material to be weighed on the
pan. Example indicates a 200 gram weight on a 6100g balance.
Wait for the stability indicator to appear before reading the weight.
WEIGHING 01/01/03 03:05:45: PM
UNITS
*
0g
2nd UNITS
CALIBRATE
200.00
0.00g
g
6100g
APPLICATION MODES
Weighing: On
Parts Counting: Off
Animal Weighing: Off
Percent Weighing: Off
Check Weighing: Off
G/N/T: Off
Filling: Off
Differential: Off
Formulation: Off
Statistics: Off
SQC: Off
Density: Off
By pressing the Enter button and using the arrow buttons, each
item on the Application Modes screen can be turned on or off.
When finished, press the Menu button, the balance returns to
weighing mode. All settings are saved.
To use any of the modes that have been turned on, press the
Mode button repeatedly to cycle through all of the application
modes.
WEIGHING WITH A CONTAINER
With no load on the pan, zero the balance.
Place an empty container on the pan. Its weight is displayed.
Tare the balance. The container’s weight is stored in memory.
Add material to the container. As material is added, its net weight
is displayed.
Removing the container and material from the pan will cause the
balance to display the container’s weight as a negative number.
The tared container weight will remain in memory until the balance
is zeroed again or the balance is turned off.
EN-12 Explorer
Pro
ADJUSTMENTS
Three soft-keys at the top of the display screen labeled UNITS,
2nd UNITS and CALIBRATE are accessed by using the arrow
buttons and selected by pressing the Enter button.
CALIBRATE
When the CALIBRATE soft key is highlighted, and the Enter button
is pressed, calibration can be performed.
2ND UNITS
When 2nd UNITS is highlighted, each press of the Enter button
cycles through the measuring units that are turned On in the
Units menu. The last option when cycling through the units is
Off.
UNITS
When UNITS is selected and highlighted, each press of the Enter
button cycles through the measuring units that are On and will
appear adjacent to the large numerals.
PARTS COUNTING 01/01/03 03:05:45 PM
SAMPLE SIZE
Add 10 Pieces, Press Enter
APW: 0.00g
Sample Size: 10
Tare: 0.00g
0g
Tare the balance if required. With the NEW COUNT soft key
highlighted, press the Enter button. Place 10 pieces on the pan
as instructed by the blinking text on the display. Press the Enter
button.
The display now indicates the Average Piece Weight (APW) and
the sample size.
PARTS COUNTING 01/01/03 03:05:45 PM
SAMPLE SIZE
SETUP
SETUP
NEW COUNT
0.00 g
6100g
NEW COUNT
3.4.2Parts Counting
Three different Parts Counting modes can be selected, Count,
Check or Fill in the Parts Counting Setup. Each mode is described
separately.
USE
PARTS COUNTING
In the Parts Counting mode, the balance displays the quantity of
parts you place on the pan. Since the balance determines the
quantity based on the average weight of a single part, all parts
must be reasonably uniform in weight.
The balance has a default setting of 10 pieces. This permits
starting parts counting immediately without setting up the balance.
The following display indicates the first time operation of parts
counting.
*
APW: 2.00g
Sample Size: 10
Tare: 0.00g
0g
Place parts to be counted on the pan. Balance displays number
of pieces.
10
20.00 g
Pcs
6100g
ADJUSTMENTS
The Parts Counting display contains three soft keys at the top of
the screen SAMPLE SIZE, SETUP and NEW COUNT.
NEW COUNT
When NEW COUNT soft key is selected, follow the screen
instructions “Add 10 Pieces, Press Enter “. After pressing the
Enter button, the balance calculates the APW and the display
indicates the number of pieces.
SETUP
Select the SETUP soft key and press the Enter button. The PARTS
COUNTING SETUP is shown.
Explorer
DISPLAY SETTINGS
APW: On
Sample Size: On
Tare: On
Weight: On
Exit
Pro
EN-13
PARTS COUNTING SETUP
Library Name: PARTS COUNTING
Mode: Count
Unit: g
Sample Size: 10
APW: Off
Tare: 0.00
Auto Optimize: Off
Display Settings: Custom
Size/APW Key: Unlocked
New Count Key: Unlocked
Exit
Using the arrow keys and Enter button, each item on the display
can be entered and modified as required. The following
information describes each entry on the screen.
Library Name:
Allows storage and recovery of parameters and results. When
a library is recovered, the library name will replace the
application name in the application mode screen. The library
name has maximum of 18 characters.
Mode: Count, Check, Fill (default is Count)
Three different modes are available, Count, Check or Fill. The
applications for Check and Fill are described in detail after this
section.
For higher quantities, the APW will not be recalculated. During
the auto optimization process, the message “Auto Optimizing,
Please wait” will be displayed in the message bar on the screen.
Display Settings: Custom, Default (default is Default)
Selecting Display Settings to Custom will bring up the DISPLAY
SETTINGS screen that will allow individual display items to be
turned on or off. Selecting Exit will return balance to previous
screen.
Size/APW Key: Unlocked, Locked (default is Unlocked)
Allows locking and unlocking the soft key function in the Parts
Counting screen. If locked, the message “Key Is Locked” will
be displayed in the message bar.
Unit: g, kg, etc... (default is g)
Allows the selection of all activated units.
Sample Size : 0-99 (default is 10)
Defines the amount of pieces used to calculate APW (average
piece weight). Setting the sample size will force the APW field
to Off and change the APW soft key to SAMPLE SIZE.
APW: 0-999999999 (default is Off)
Defines the average piece weight. Setting the APW will force
the Sample Size field to Off and change the SAMPLE SIZE soft
key to APW.
Tare: 0-99999999 (default is 0)
Defines the weight of the container being used.
Auto Optimize: On,Off (default is Off)
Selecting Auto Optimize On will recalculate the APW value
automatically during the parts counting process up to double
the amount of parts.
New Count Key: Unlocked, Locked (default is Unlocked)
Allows locking and unlocking the soft key function in the Parts
Counting screen. If locked, the message “Key Is Locked” will
be displayed in the message bar.
Exit:
When selected, balance returns to the Parts Counting Mode.
SAMPLE SIZE/APW
With SAMPLE SIZE soft key highlighted,press the Enter button.
The PARTS COUNTING display is shown with the value of the
sample size highlighted. Using the arrow buttons and the Enter
button, the sample size value can be changed.
After the Enter button is pressed, and a new sample size has
been entered, remove the sample and place parts to be counted
on the balance using the new sample size.
APW
With the APW soft key highlighted, press the Enter button. The
APW is highlighted at the lower left of the screen. Using the
arrow buttons and the Enter button, the APW value can be
changed.
EN-14 Explorer
Pro
USE
PARTS COUNTING-CHECK
This feature permits establishing a set quantity of pieces as a
criteria for similar items that can be quickly checked against the
sample. In the Parts Counting Check mode, the displayed data
includes, present count (Pcs), APW, Sample Size, Tare, Weight
and a bar graph indicating UNDER, ACCEPT (blinking) and
OVER.
Enter the PARTS COUNTING SETUP menu and change Mode to
Check.
The PC Check display contains three soft keys at the top of the
screen SAMPLE SIZE, SETUP and NEW COUNT.
NEW COUNT
When NEW COUNT soft key is selected, follow the screen
instructions “Add 10 Pieces, Press Enter “. After pressing the
Enter button, the balance calculates the APW and the display
indicates the number of pieces.
SETUP
Using the left arrow key, select SETUP soft key and press the
Enter button. The PARTS COUNTING SETUP is shown.
PARTS COUNTING SETUP
Library: PARTS COUNTING
Mode: Check
Unit: g
Sample Size: 10
APW: Off
Tare: 0.00
Auto Optimize: Off
Over: 10
Under: 5
Display Settings: Custom
Size/APW Key: Unlocked
New Count Key: Unlocked
The balance indicates to add 10 pieces and has an Under value
of 5 and an Over value of 10 set up as default values and an
average piece weight has not been set. Place 10 pieces on the
pan and press the Enter button.
An average piece weight has been set by placing the pieces on
the pan. To use the PC Check, the Over and Under values must
be set as well as other display setup parameters. Refer to
adjustments and set the balance parameters to fit specific
requirements.
Two additional entries have been added to the PARTS COUNTING
SETUP, they are Over and Under. Refer to the Adjustments section
of parts counting for a description of all other settings.
Over 10: 0-9999 (default is 10)
Defines the over limit in pieces.
Under 5: 0-9999 (default is 5)
Defines the under limit in pieces.
Display Settings: Custom, Default (default is Default)
Selecting Display Settings to Custom will bring up the DISPLAY
SETTINGS screen that will allow individual display items to be
turned on or off. Sample Size, Tare, Weight are described under
parts counting Display Settings.
Explorer
Pro
EN-15
Three new entries Weight, Over and Under appear on this screen.
Selecting Exit will return balance to previous screen.
Weight: On,Off (default is On)
Can be turned on or off.
Over: On, Off (default is On)
Can be turned on or off.
Under: On, Off (default is On)
Can be turned on or off.
SAMPLE SIZE/APW
Select SAMPLE SIZE soft key and enter the sample size desired.
APW
With the APW soft key highlighted, press the Enter button.
The APW is highlighted at the lower left of the screen. Using
the arrow buttons and the Enter button, the APW value can be
changed. Follow screen instructions and place sample on pan.
Remove the sample and place items on pan. In the following
example, the sample size was 10, the APW was 5.00g, the
under value was set at 99 and the over value was set at 101.
The acceptable value was 100 pieces.
NOTE: The values appearing on the screen initially are from
previous entries in parts counting.
Placing an item on the pan at this time will only respond to the
previous settings for sample size and APW.
To use PC FILLING, the sample size and other parameters must
be set first. Refer to adjustments and set the balance parameters
to fit specific requirements.
50%
SETUP
NEW COUNT
0
90%
100%
0.00 g
Pcs
110%
ADJUSTMENTS
The PC FILLING display contains three soft keys at the top of the
screen SAMPLE SIZE, SETUP and NEW COUNT.
NEW COUNT
When NEW COUNT soft key is selected, follow the screen
instructions “Add 10 Pieces, Press Enter “is displayed After
pressing the Enter button, the balance calculates the APW and
the display indicates the number of pieces.
PARTS COUNTING-FILL
This feature permits establishing a set quantity of pieces that can
be shown as a percentage when pieces are added. A split bar
display indicates 0% to 90% on the first bar and 90% to 110%
on the second bar.
Enter the PARTS COUNTING SETUP menu and change Mode to
Fill.
The balance is now in the PC FILLING mode.
SETUP
Select SETUP soft key and press the Enter button. The PARTS
COUNTING SETUP is shown.
PARTS COUNTING SETUP
Library Name: PARTS COUNTING
Mode: Fill
Unit: g
Sample Size: 10
APW: Off
Tare: 0.00
Auto Optimize: Off
Target:1000
Display Settings: Default
Size/APW Key: Unlocked
New Count Key: Unlocked
Exit
EN-16 Explorer
g
g
Pro
One additional entry has been added to the PARTS COUNTING
SETUP, that is Target. Refer to the Adjustments section of parts
counting for a description of all other settings.
Target: 0-9999 (default is 1000)
Defines the target pieces for the 100% limit.
Display Settings: Custom, Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Parts Counting mode can be individually turned on or off. If
Display Setting is set to default, all the information is turned
on.
Two new entries Target and Difference appear on this screen.
Selecting Exit will return balance to previous screen.
Target: On, Off (default is On)
Can be turned on or off.
3.4.3Animal Weighing
Animal weighing permits you to weigh small animals and filters
out animal movements. Choices of manual, semi automatic
and automatic operation is possible.
The display data includes the filtered weight of the animal in
selected unit of measurement, and unfiltered weight of the animal
in 2nd unit of measurement, filter level, mode of operation and
a bar graph indicating present used capacity of the balance.
ANIMAL WEIGHING 01/01/03 03:05:45 PM
FILTER
Place animal, Press Enter
Filter: Low
Mode: Manual
0g
SETUP
START
0.00 g
6100
Difference: On, Off (default is On)
Can be turned on or off.
SAMPLE SIZE/APW
Select SAMPLE SIZE soft key and enter the sample size desired.
APW
With the APW soft key highlighted, press the Enter button. The
APW is highlighted at the lower left of the screen. Using the
arrow buttons and the Enter button, the APW value can be
changed. Follow screen instructions and place sample on pan.
Remove the sample and add items on pan until 100% is
highlighted at the bottom of the screen as shown. In the following
example, the sample size was 10, the APW was 2.00g, the Target
value was set at 100 pieces.
ANIMAL WEIGHING
Place the animal on the pan. With the START soft key highlighted,
press the Enter button. The display performs a countdown to
average out the weight. The weight is indicated as shown and
remains on the display.
ANIMAL WEIGHING 01/01/03 03:05:45 PM
FILTER
*
2038.73
Filter: Low
Mode: Manual
0g
Remove the animal from the pan. With the CLEAR soft key
highlighted, press the Enter button to clear the weight. The
balance is now ready for another measurement. To change
the mode of operation from manual to semi automatic or
automatic and change filtering levels, refer to Adjustments.
SETUP
CLEAR
2038.73 g
g
6100
Explorer
g
Pro
EN-17
ADJUSTMENTS
The ANIMAL WEIGHING display contains three soft keys at the
top of the screen FILTER, SETUP and START/CLEAR .
START/CLEAR
Select the START soft key and press the enter button to start the
animal weighing process. The soft key is changed to CLEAR
and allows the weight to be cleared when the Enter button is
pressed.
SETUP
Select SETUP soft key and press the Enter button. The ANIMAL
WEIGHING SETUP display is shown.
Unit: g, kg, etc...(default is g)
Allows the selection of all activated units.
2nd Unit: g, kg, etc...(default is g)
Allows the selection of all activated units.
Filter: Low, Medium, High (default is Low)
Can be set to one of three filtering levels. A higher level filter
will provide more accurate measurement.
Automatic Mode
Weighing process starts automatically as soon as the animal
is placed on the pan. The weight is displayed until the animal
is removed. The balance is ready for another animal weighing.
Display Settings: Custom, Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Animal weighing mode can be individually turned on or off. If
Display Setting is set to default, all the information is turned
on.
Filter Key: Locked, Unlocked (default is Unlocked)
This feature allows to lock or unlock the Soft key function.
Exit
When selected, returns to Animal Weighing mode.
FILTER
Select FILTER soft key and press the Enter button. The display
is shown with the Filter: Low highlighted. Pressing the up or
down arrow buttons, either Low, Medium or High can be
selected, then press the Enter button. Low is the default setting.
Display is shown. After selection is made, the screen display
returns to an Animal Weighing Mode. A higher filter level will
provide a more accurate measurement.
ANIMAL WEIGHING 01/01/03 03:05:45 PM
FILTER
SETUP
START
*
Mode: Manual, Semi, Auto (default is Manual)
One of three modes can be set.
Manual Mode
Weighing process starts manually. After placing animal on
pan with soft key START highlighted, press the Enter button.
Screen display retains reading and is manually cleared by
pressing the Enter button when soft key CLEAR is highlighted.
Semi Automatic mode
Weighing process starts automatically as soon as the animal
is placed on the pan. When the animal is removed, the display
retains the weight reading. The balance must be cleared
manually by pressing the Enter button.
Filter: Low
Mode: Manual
0g
0.00 g
6100
EN-18 Explorer
g
g
g
Pro
3.4.4Percent Weighing
Percent weighing allows placing a reference load on the balance,
then viewing other loads as a percentage of the reference. The
load placed on the pan as a reference is displayed as 100%.
Subsequent loads are displayed as a percentage of the reference.
The maximum limit is the capacity of the balance. The minimum
limit is 100d. The display data includes reference weight in
unit of measurement, difference value in percent, difference value
in measuring unit and a bar graph indicating present used
capacity of the balance.
PERCENT WEIGHING 01/01/03 03:05:45 PM
REFERENCE
Add Reference Weight, Press Enter
SETUP
NEW REF
*
Ref: 0.00g
Diff: 0.00 %
Diff: 0.00g
0g
0.00 g
6100
The reference weight is removed and a second weight is placed
on the pan. The example illustrates a 50 g weight.
PERCENT WEIGHING 01/01/03 03:05:45 PM
REFERENCE
*
Ref: 200.00g
Diff: -75.00 %
Diff: -150.00g
0g
SETUP
NEW REF
25.00
50.00 g
%
6100
ADJUSTMENTS
The PERCENT WEIGHING display contains three soft keys at the
top of the screen REFERENCE, SETUP and NEW REF.
NEW REF
When NEW REF soft key is selected, follow the screen instructions
“Add Reference Weight, Press Enter”. This establishes a new
reference weight.
USE
PERCENT WEIGHING
With the NEW REF soft key highlighted, press the Enter button.
Place the reference weight on the pan and press the Enter
button again. The sample shown indicates a 200g weight
was placed on the pan.
PERCENT WEIGHING 01/01/03 03:05:45 PM
REFERENCE
Add Reference Weight, Press Enter
*
Ref: 200.00g
Diff: 0.00 %
Diff: 0.00g
0g
SETUP
NEW REF
100.00
200.00 g
%
6100
SETUP
Select SETUP soft key and press the Enter button. The PERCENT
Over: 15.00 g
Target: 10.00 gDiff: -10.00 g
Under: 5.00 gDiff: -100.00 %
O.00
g
*
UnderAccept Over
g
Unit: g, kg, etc...(default g).
Allows the selection of all activated units.
Reference: 0-99999999 (default is 10)
Defines the reference weight at 100%.
Display Settings: Custom, Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Percent Weighing mode can be individually turned on or off. If
Display Setting is set to default, all the information is turned
on.
Reference Key: Locked, Unlocked (default is Unlocked)
Allows locking or unlocking the REFERENCE soft key in the
percent Weighing screen. Message “Key is locked” when Locked
is selected.
New Reference Key: Locked, Unlocked (default is Unlocked)
Allows locking or unlocking the NEW REF soft key in the Percent
Weighing screen. Message “Key is locked” when Locked is
selected.
Pro
EN-19
USE
CHECK WEIGHING
Before using the Check Weighing feature, the Over, Target and
Under limits must be set.
ADJUSTMENTS
The CHECK WEIGHING display contains three soft keys at the top
of the screen UNITS, SETUP and CALIBRATE.
CALIBRATE
When the CALIBRATE soft key is highlighted, and the Enter
button is pressed, calibration can be performed.
Exit:
When selected, balance returns to the Percent Weighing Mode.
REFERENCE
Select REFERENCE soft key and press the Enter button. The
display screen changes with Ref field highlighted. Reference
weight can be specified.
PERCENT WEIGHING 01/01/03 03:05:45 PM
REFERENCE
SETUP
*
Ref: 0.00
Diff:
Diff:
0g
3.4.5Check Weighing
Check weighing is used when items are checked against preset
balance parameters. This feature permits you to weigh an item,
set balance parameters such as the over weight, target weight
and under weight. A bar graph at the bottom of the Check
Weighing screen indicates UNDER, ACCEPT and OVER for items
being checked.
NEW REF
0.00 g
SETUP
Select the SETUP soft key and press the Enter button. The
CHECK WEIGHING SETUP display is shown.
Library Name:
Allows storage and recovery of parameters and results. When
a library is recovered, the library name will replace the
application name in the application mode screen. The library
name has maximum of 18 characters.
Unit: g, kg, etc...(default is g)
Allows the selection of all activated units.
Over: 0-99999999 (default is 15)
Defines over limit value.
Target: 0-99999999 (default is 10).
Defines target value.
Under: 0-99999999 (default is 5)
Defines under limit value.
Display Settings: Custom, Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Check Weighing mode can be individually turned on or off. If
Display Setting is set to default, all the information is turned
on.
Reading allows a choice of either weight readings to be
displayed in large numeral or message that displays OVER,
ACCEPT and UNDER in large letters. Display below indicates
message format.
CHECK WEIGHING 01/01/03 03:05:45 P
UNITS
SETUP
CALIBRATE
3.4.6Filling
Filling permits entering a target reference weight. Material can
be placed on the pan and by monitoring the Difference weight
and the bar display, an accurate fill can be achieved. A split
bar display indicates 0% to 90% on the first bar and 90% to
110% on the second bar.
USE
FILLING
A sample (target weight) is placed on the pan and the Enter
button is pressed. The balance stores this weight as shown on
the display as Target. The sample is removed from the pan and
the material is added to the pan. The display provides Target
Weight, Difference weight, Percentage and the split bar at the
bottom of the screen displays the percentage of the sample
compared against the target weight. When a container is used,
zero the balance before placing the actual sample on the pan.
See Adjustments to set other balance parameters.
ADJUSTMENTS
The FILLING display contains three soft keys at the top of the
screen TARGET, SETUP and NEW TARGET.
*
UNDER
Over: 15.00 g
Target: 10.00 gDiff: -10.00 g
Under: 5.00 gDiff: -100.00 %
Exit:
When selected, returns to Check Weighing mode.
UNITS
When the UNITS soft key is highlighted, each press of the Enter
button cycles through the measuring units that are On.
NEW TARGET
With the NEW TARGET soft key highlighted, press the Enter button.
0.00g
Follow display instructions to “Add Target Weight, Press Enter”.
This establishes a new target weight and is displayed on the
screen.
SETUP
When SETUP soft Key is highlighted, press the Enter button. The
FILLING SETUP display is shown.
Explorer
g
g
Pro
EN-21
FILLING SETUP
Unit: g
2 nd unit: g
Target Weight: 0000
Display Settings: Custom
Target Key: Unlocked
New Target Key: Unlocked
Exit
Units: g, kg, etc...(default is g)
Allows the selection of all activated units.
2nd UNITS: g, kg, etc...(default is g)
Allows the selection of all activated units.
Target Weight: 0-99999999 (default is 0)
Allows entering a specific target filling weight.
Display Settings: Custom, Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Filling mode can be individually turned on or off. If Display
Setting is set to default, all the information is turned on.
G/N/T WEIGHING 01/01/03 03:05:45 PM
UNITS
2nd UNITS
TARE
*
0.00
Gross: 0.00 g
Net: 0.00 g
Tare: 0.00 g
0g
USE
G/N/T WEIGHING
Press the O/T button to zero the balance. In this application,
the O/T button functions as a Zero and not as a Tare.
Place a container on the pan. With the TARE soft key
highlighted, press the Enter button. The container’s weight is
stored in memory in the balance. Place the material in the
container. The balance immediately displays the Gross, Net
and Tare weights. The net weight is displayed as large numerals.
Example shown represents a container weight of 50g and
material of 200g. The gross weight is displayed as 250g.
Before a new container is used, zero the balance by using the
O/T button.
g
0.00 g
6100
Target Key: Locked, Unlocked (default is Unlocked)
This feature allows to lock or unlock the soft key function.
New Target Key: Locked, Unlocked (default is Unlocked)
This feature allows to lock or unlock the soft key function.
Exit:
When selected, balance returns to the Filling Mode.
TARGET
With the TARGET soft key highlighted, press the Enter button.
The FILLING display is shown with the value of the Target
highlighted. Using the arrow buttons and the Enter button, the
Target value can be changed.
FILLING 01/01/03 03:05:45 PM
TARGET
*
Target: 0.00 g
Diff: -0.00 g
0%
50%
SETUP
NEW TARGET
0.00
90%
100%
0.00 g
g
110%
G/N/T WEIGHING 01/01/03 03:05:45 PM
UNITS
2nd UNITS
TARE
*
200.00
Gross: 250.00 g
Net: 200.00 g
Tare: 50.00 g
0g
ADJUSTMENTS
The G/N/T WEIGHING display contains three soft keys at the top
of the screen UNITS, 2nd UNITS and TARE.
TARE
When the TARE soft key is selected, pressing the Enter button
will store the current display reading as the tare value.
2nd UNITS
When 2nd UNITS soft key is highlighted, each press of the
Enter button changes the measuring units that are On and will
appear under the large numerals. The last option when cycling
through the units is Off.
200.00 g
g
6100
3.4.7Gross / Net / Tare Weighing
Gross/Net/Tare (G/N/T) application allows to display Gross
(sample plus container weight), NET (sample weight) and TARE
(container weight) simultaneously.
UNITS
When the UNITS soft key is highlighted, each press of the Enter
button cycles through all active measuring units on the Gross,
Net and Tare displays.
EN-22 Explorer
g
Pro
3.4.8Differential Weighing
Differential weighing stores tare and weight values so samples
can be dried or processed and the difference in weight calculated
at a later time. Up to 80 samples can be stored. The balance
has the capability to work with one or two different containers
or no container at all. Samples can be added to the library or
extracted by name using the previously stored data.
USE
DIFFERENTIAL 01/01/03 03:05:45 PM
RESULTS
SETUP
START
*
Tare: 1 0.00 g
0.00
Initial: 0.00 gDiff: 0.00 g
Final: 0.00 g Diff: 0.00 %
0g
Before using the Differential Weighing feature, the Differential
Setup should be reviewed. Highlight the SETUP soft key and
press the Enter button. Assign a Library Name. Review the
settings and change as necessary.
DIFFERENTIAL SETUP
Library Name: DIFFERENTIAL
Mode: Weight
Unit: g
Number Of Samples: 3
Tare Required: Single
Auto Sample: Off
Clear All Data ? : No
Display Settings: Default
View Results
Exit
g
6100
Tare Wt Init Wt Tare Wt Final Wt
00.00 0.00 0.00 0.00
00.00 0.00 0.00 0.00
00.00 0.00 0.00 0.00
Sample # 1
Continue Resample Delete
With CONTINUE highlighted, press the Enter button and follow
the screen instructions and add the Initial samples. During
this mode, the first screen now contains an EDIT soft key. The
EDIT soft key permits entering any Initial or Final sample and
rerun.
The screen shown above will now indicate the initial sample
weights. At this point, you can select any one of the samples
and with the RESAMPLE soft key highlighted and perform a
resample.
With the DELETE soft key highlighted, any one sample can be
selected and deleted. With CONTINUE highlighted, press the
Enter button and follow the screen instructions and add the
Final samples.
After the last Final sample is entered, the screen above indicates
all tare weights, Initial weights and Final weights. With CONTINUE
highlighted, pressing the Enter button brings up one final
screen that contains the Sample #, Initial Weight, Final Weight,
Weight Difference and Totals.
With the START key highlighted, press the Enter button. Follow
the instructions on the screen and place the first container on
the pan and press Enter. The START key changes to ACCEPT.
Continue until all containers have been weighed. The following
screen is displayed indicating the container weights.
All test samples and results can be reviewed by entering
Differential Setup and selecting the Library Name. On the first
screen, select the RESULTS soft key, press the Enter button and
the test results are displayed. A total of 80 samples can be
stored.
Explorer
Pro
ADJUSTMENTS
START/ACCEPT
When the START key is selected, pressing the Enter button
will start differential weighing application. The screen displays
“Place Initial Container X, Press Enter” message and the
Differential Weighing process has started. When a new
process starts all the data from previous Differential Weighing
is erased. At this point, the soft key has changed into an
ACCEPT key. When the ACCEPT key is selected, the weight
value has been accepted as valid data.
SETUP
When executed, the DIFFERENTIAL WEIGHING SETUP screen
will be displayed.
EDIT
When executed, an edit screen will display that will allow an
user to re-sample a sample weight, or to delete an undesirable
sample, or to continue.
Library Name:
Allows storage and recovery of parameters and results.
When a library is recovered, the library name will replace
the application name in the application mode screen. The
library name has maximum of 8 characters.
Mode: Weight, Percent, % Retention (default is Weight)
Allows the selection of the various calculations in the final
results.
Units: g, kg, etc...(default is g)
Allows the selection of all activated units.
EN-23
Tare Required: Off, Single, Dual (default is Single)
Allows selection of Tare Required. When Off is selected, the
balance will not prompt for container. When Single is selected,
the balance will prompt for the initial container weight. When
Dual is selected, the balance will prompt for the initial
container weight and final container weight.
Auto Sample: Off, On (default is Off)
Allows Auto Sample to be turned on or off . When Auto Sample
is turned on, the Differential Weighing process is automatic.
Clear All Data: No, Yes (default is No)
All the data from the Differential Weighing process will be
erased when Yes is selected.
Display settings: Custom/Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Differential Weighing mode can be individually turned on or
off. If Display Setting is set to default, all the information is
turned on.
View Results:
Allows the calculated results of the Differential Weighing
process to be viewed.
Exit:
In the sub menu, will return to the previous menu.
In the main men, will return to the last application mode.
Number of Samples: xx (default is 10)
Allows manual entry for number of samples. Only integer
values up to 80 are valid.
Total Weight: xxx.xx (default is 10.00)
Retention Mode only, allows manual entry for the Total
Weight value. Total Weight has to be less than the balance
capacity.
EN-24 Explorer
Pro
3.4.9Formulation
Formulations can be named and have from 2 to 10 components
specified. Names are limited to 18 characters. Once named,
they may be recalled and used at any time. Each component
of a given formulation can be specified as to its weight or
percentage. Each element of a formulation is shown on a dual
bar graph as a percentage and displays the desired weight.
Thus, each component may be placed on the pan until 100%
is indicated.
USE
FORMULATION 01/01/03 03:05:45 PM
RESULTS
SETUP
START
*
Item: 1
0.00
Target: 0.00 g
Diff: 00.00 g Pct:0.00 %
50%
0%
90%
100%
g
110%
Add the first component to the container. The dual bar graph
at the bottom of the screen indicates the percentage of the
component added. The target weight, difference and percent
are also displayed. Add each component until the specified
Target weight is reached (100%). When adding an individual
component, make sure the balance has stabilized and indicates
the specified weight. Continue until all components have been
added.
The next screen automatically appears and indicates the Target,
Result and Difference weights of each component of the formula
as well as the totals. Pressing the PRINT button will send the
screen results to an externally connected printer or computer.
Press the Enter button to exit and return to the starting screen.
Results of any formulation may be displayed by selecting the
RESULTS soft key at the top of the screen. The actual formulation
is selected by entering the FORMULA SETUP screen and scrolling
through the Library Names.
Before using the Formulation feature, the Formulation Setup
should be reviewed. Highlight the SETUP soft key and press
the Enter button. Assign a Library Name. Review the settings
and change as necessary. Enter the Mode type, Weight or
Percent. Enter the number of Items.
FORMULATION SETUP
Library Name: FORMULATION
Mode: Weight
Unit: g
Number Of Items: 2
Display Settings: Default
Items Setup
View Results
Exit
Exit the Formulation Setup. The initial screen appears with the
name of the formulation appearing at the upper left-hand corner
of the screen.
Place a container on the pan and press the >O/T< button to
tare the container weight.
With the START key highlighted, press the Enter button. Follow
the instructions on the screen. The START key changes to ACCEPT.
To start a new formulation, repeat the entire process.
ADJUSTMENTS
START/ACCEPT
When the START key is selected, pressing the Enter button will
start formulation application. The screen displays “Add Item
“name field”, Press Enter” message and the Formulation process
has started. When a new process starts, all the data from
previous Formulation is erased. At this point, the soft key has
changed into an ACCEPT key. When the ACCEPT key is selected,
the weight value has been accepted as valid data.
SETUP
When selected, the FORMULA SETUP screen will be displayed.
RESULTS
When selected, a screen will display all calculated results.
Library Name:
Allows storage and recovery of parameters and results.
When a library is recovered, the library name will replace the
application name in the application mode screen. The library
name has maximum of 18 characters.
Explorer
Pro
EN-25
Mode: Weight, Percent (default is Weight)
Allows the selection of the various calculations in the final
results.
Unit: g, kg, etc... (default is g)
Allows the selection of all activated units.
Number of Items: xx (default is 2)
Allows manual entry for number of items. Only integer values
up to 10 are valid.
Target Weight: xxx.xx (default is 5.00)
Allows manual entry for the Target Weight value. Target
Weight has to be less than the balance capacity.
Display settings: Custom/ Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Formulation mode can be individually turned on or off. If Display
Setting is set to default, all the information is turned on.
Item Setup:
Displays a setup menu that will allow a name and a value to
be assigned to each item.
With Items Setup selected, specify the names for each component
of the formula (limit 9 characters) and either the weights or
percentages.
3.4.10Statistics
Statistics are used when it is desired to compare a number of
samples and examine the relative deviation of the samples
along with other statistical data. A minimum of three samples
is required in this program. Statistics contains menu options
which include: number of samples, maximum, minimum,
difference, sum, mean, standard deviation, relative deviation,
auto sample, and auto print. Most of these can be set on or off
except sample size which can be set for a particular number.
When a printer or computer is connected to the balance, all
statistical information can be observed and printed. WEIGHING,
ANIMAL WEIGHING, CHECK WEIGHING and FILLING are the
functions for providing Statistical Data.
View Results:
Allows the calculated results of the Formulation process to be
viewed.
Exit:
In the sub menu, will return to the previous menu.
In the main men, will return to the last application mode.
*
0.00g
Max: 0.00 g Sum: 0.00 g
Min: 0.00 g Mean: 0.00 g
Target: 10.00 g
50%
0%
(Used for Filling)
90%
100%
110%
EN-26 Explorer
Pro
STATISTICS 01/01/03 03:05:45 PM
RESULTS
SETUP
START
*
0.00g
Max: 0.00 g Sum: 0.00 g
Min: 0.00 g Mean: 0.00 g
Under: 5.00 g Over: 15.00 g
Under
Accept OverUnder
Accept Over
(Used for Check Weighing)
Before using the Statistics feature, the Statistics Setup should
be reviewed. Highlight the SETUP soft key and press the Enter
button. Assign a Library Name.
NOTE: All Statistics settings made with a Library Name will be
directly associated with the Library name. When a Library Name
is called up, the Mode and all settings will be retained with the
Library Name.
Review the settings and change as necessary.
Place a container on the pan and press the >O/T< button to
tare the container weight.
With the START key highlighted, press the Enter button. Follow
the instructions on the screen. The START key changes to
ACCEPT. Add the sample to the container. Repeat for all
samples.
Weighing/Animal Modes
When Weighing Mode or Animal Weighing Mode is selected,
The bar graph at the bottom of the screen indicates the weight
of each sample when added. The Maximum, Minimum, Sum
and Mean weights are also displayed on the screen.
Filling Mode
When the Fill Mode is selected, the Target Weight that was
entered on the STATISTICS SETUP screen will display the sample
weight on the bar graph as a percentage. The Maximum,
Minimum, Sum , Mean and Target weights are also displayed
on the screen.
STATISTICS SETUP
Library Name: STATISTICS
Mode: Weighing
Unit: g
Number Of Samples: 3
Maximum: On
Minimum: On
Difference: On
Sum: On
Mean: On
Std. Deviation: On
Rel. Deviation: On
Auto Sample: On
To review remaining items on the screen, scroll down.
STATISTICS SETUP
Maximum: On
Minimum: On
Difference: On
Sum: On
mean; On
Std. Deviation: On
Rel. Deviation: On
Auto Sample: On
Auto Print: On
Display Settings: Default
View Results
Exit
Check Weighing Mode
When Check Weighing Mode is selected, the display at the
bottom of the screen changes to Under/Accept/Over. As each
sample is placed on the pan, the large numerals on the display
will indicate the weight of the sample and the bar graph at the
bottom of the screen indicates either Under, Accept or Over
based on the Over and Under values added in the STATISTICS
SETUP menu. The Maximum, Minimum, Sum, Mean Under
and Over weights are also displayed on the screen.
Explorer
Pro
EN-27
The STATISTICAL RESULTS screen automatically appears and
indicates the Number of samples, Unit, Mean, Maximum,
Minimum, Standard Deviation, Relative Deviation, Sum and
Difference weights. Pressing the PRINT button will send the
screen results to an externally connected printer or computer.
STATISTICAL RESULTS
Number Of Samples: 3
Unit: g
Mean: 0.00
Maximum: 0.00
Minimum: 0.00
Std. Deviation: 0.00
Relative Deviation: 0.00
Sum: 0.00
Difference: 0.00
Press <Enter> To Exit
Press the Enter button to exit and return to the starting screen.
Results may be displayed by selecting the RESULTS soft key at
the top of the screen.
To start a new test, repeat the entire process.
ADJUSTMENTS
START/ACCEPT
When the START key is selected, pressing the Enter button will
start statistics application. The screen displays “Add Sample
#1, Press Enter” message and the Statistics process has started.
When started, all the data from previous statistics is erased. At
this point, the soft key has changed into an ACCEPT key. When
the ACCEPT key is selected, the weight value has been accepted
as valid data.
SETUP
When selected, the STATISTICS SETUP screen will be displayed.
RESULTS
When selected, a screen will display all calculated results.
Library Name:
Allows storage and recovery of parameters and results.
When a library is recovered, the library name will replace
the application name in the application mode screen. The
library name has maximum of 18 characters.
Mode: Weighing, Animal, Check, Filling (default is Weighing)
Allows the selection of the various calculations in the final
results.
Unit: g, kg, etc... (default is g)
Allows the selection of all activated units.
Number of Samples: xx (default is 3)
Allows manual entry for number of items. Only integer values
up to 100 are valid.
Filter: Low, Medium, High (default is Low)-Animal mode only
Can be set to one of three filtering levels. A higher level will
provide more accurate measurement.
Over, Under: Check mode only
Allows manual entries for the Over limit and the Under limit
values. Eight decimal places are allowed for these values.
Target: Check mode and Filling mode
Allows manual entry for the Target Weight value. Eight
decimal places are allowed for target Weight value.
Maximum: On, Off (default is On)
Allows Maximum to be turned On or Off. When Maximum is
turned on, the maximum stored sample is included into the
final results.
Minimum: On, Off (default is On)
Allows Minimum to be turned On or Off. When Minimum is
turned on, the minimum stored sample is included into the
final results.
Difference: On, Off (default is On)
Allows Difference to be turned On or Off. When Difference is
turned on, the difference calculation between the maximum
stored sample and the minimum stored sample is included
into the final results.
Sum: On, Off (default is On)
Allows Sum to be turned On or Off. When Sum is turned on,
the sum calculation of the samples is included into the final
results.
Mean: On, Off (default is On)
Allows Mean to be turned On or Off. When Mean is turned on,
the mean calculation of the samples is included into the final.
EN-28 Explorer
Pro
Std. Deviation: On, Off (default is On)
Allows Std. Deviation to be turned On or Off. When Std. Deviation
is turned on, the standard deviation calculation of the samples
is included into the final results.
Rel. Deviation: On, Off (default is On)
Allows Relative Deviation to be turned On or Off. When Relative
Deviation is turned on, the relative deviation calculation of the
samples is included into the final results.
Auto Sample: Off, On (default is Off)
Allows Auto Sample to be turned On or Off . When Auto Sample
is turned on, the Statistics process is automatic.
Auto Print: Off, On (default is Off)
Allows Auto Print to be turned On or Off . When Auto Print is
turned on, the final calculated results are automatically printed
out at the end of the Statistics.
Display settings: Custom/ Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Statistics mode can be individually turned on or off. If Display
Setting is set to default, all the information is turned on.
These weight limits are identified as +Tl, +T2, NOMINAL and TI, -T2. As samples are weighed and stored in the balance, a
trend analysis is developed and displayed on the balance. Up
to 25 samples in a batch with up to ten batches are visible on
a trend screen for quality control purposes. Each batch of
samples is shown on the display which indicates the maximum/
minimum standard deviation and mean values for each batch.
An on going examination of the relative deviation of the samples
along with other statistical data can be viewed and is stored.
By observing the results of the VIEW TRACE screen, you can
effectively monitor the filling process operation. Setup parameters
can be stored in the library and up to 5 products with statistical
history can be stored in memory. All SQC information can be
printed.
USE
SQC 01/01/03 03:05:45 PM
RESULTS
SETUP
START
*
0.00
Tare: 0.00 g
Max: 0.00 g Sum: 0.00 g
Min: 0.00 g Mean: 0.00 g
0g
g
6100g
Exit:
In the sub menu, will return to the previous menu.
In the main men, will return to the last application mode.
View Results:
Allows the calculated results of the Statistics process to be
viewed.
3.4.11Statistical Quality Control
The Statistical Quality Control (SQC) feature is extremely useful
during various types of process filling operations when it is
desired to monitor and control the process to eliminate under
and over filling. Provisions are made in the balance to
accommodate the weight of various packaging methods. During
operation, parameters of the packaged product are set into the
balance such as packaging weight, acceptable weight limits
and non-acceptable weight limits of the product.
Before using the SQC feature, the SQC Setup should be reviewed.
Highlight the SETUP soft key and press the Enter button. Assign
a Library Name.
NOTE: All SQC settings made with a Library Name will be directly
associated with the Library name. When a Library Name is
called up, all settings will be retained with the Library Name.
Review the settings and change as necessary. Enter the Unit,
Number of Samples, Tolerance (either weighing or percentage),
Tare, Auto Sample, Auto Print, Display Settings (Default or
Custom), and Batch Name.
SQC SETUP
Library Name: SQC
Unit: g
Number Of Samples: 3
Tolerance: Weighing
Tare Required: Off
Auto Sample: Off
Auto Print: Off
Clear All Data: No
Display Settings: Default
Batch Name: Batch 1
View Trace
View History
Explorer
Pro
EN-29
Defining Tolerances
When Tolerance is selected a choice of either Weighing or
Percent is displayed. Define either the weight or percentage for
your product starting with +T2, +T1, Nominal, -T1 and -T2.
Weight Tolerance
+ Tolerance 2: 0.00
+ Tolerance 1: 0.00
Nominal: 0.00
- Tolerance 1: 0.00
- Tolerance 2: 0.00
Exit
After all entries have been made, exit the SQC SETUP menu.
Pressing the Enter button will start SQC application. Continue
to place samples on the pan until all samples have been entered.
The following sample screens are for 5 batches. A Nominal
weight of 50.00g with T1 set at 5 g above and below and T2
set at 10g above and below. A sample history of 5 batches is
displayed.
Pressing the Enter button will return the balance to the Start
Screen. Highlight the SETUP soft key and press the Enter button.
Assign a new Batch Name. Continue until all batches have
been processed.
After you have taken the required number of data samples, you
can view the sample View Trace and/or the View History. The
View Trace screen below indicates 5 batches were processed.
NOTE: Once a batch with all samples has been completed, you
cannot make changes to the batch. A new batch must be run.
Return to the SETUP soft key and rerun the Batch.
Definitions for the symbols are shown below.
MAXIMUM LIMIT
STANDARD
MEAN VALUE OF SQC SET
DEVIATION
MINIMUM LIMIT
ADJUSTMENTS
START/ACCEPT
When the START key is selected, pressing the Enter button will
start SQC application. The screen displays “Add Sample #1,
Press Enter” message and the SQC process has started. At this
point, the soft key has changed into an ACCEPT key. When the
ACCEPT key is selected, the weight value has been accepted as
valid data.
Return to the SETUP soft key and press the Enter button. Scroll
down to View Trace and or View History and press the Enter
button.
SETUP
When selected, the SQC SETUP screen will be displayed.
RESULTS
When selected, a screen will display all calculated results of
the SQC batch.
EN-30 Explorer
Pro
Library Name:
Allows storage and recovery of parameters and results.
When a library is recovered, the library name will replace the
application name in the application mode screen. The library
name has maximum of 18 characters.
Unit: g, kg, etc... (default is g)
Allows the selection of all activated units.
Number of Samples: xx (default is 10)
Allows manual entry for the number of samples. Only integer
values up to 80 are valid.
Tolerance: Weight, Percentage (default is Weight)
Displays a screen that will allow manual editing of the
tolerances. Selecting Weight or Percentage will determine the
entries for tolerances which are actual weight or percentage of
nominal weight.
Tare Required: Off, Mean Tare, Individual, Add w/ Mean Tare
(default is Off)
Allows selection of Tare Required. When Off is selected, the
balance will not prompt for a container. When Mean Tare is
selected, manual entry of Mean Tare is allowed . When Individual
is selected, the balance will prompt for a container weight for
each sample. When Add w/ Mean Tare is selected, the entry of
Mean Tare is allowed. Also, the sample weight is added and
stored into the balance then the balance tares the sample weight
and prompts for the next sample weight to be added and stored.
Clear All Data: No, Yes (default is No)
All the data from the SQC process will be erased when Yes is
selected.
Display settings: Custom/ Default (default is default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
SQC mode can be individually turned on or off. If Display Setting
is set to default, all the information is turned on.
Exit:
In the sub menu, will return to the previous menu.
In the main men, will return to the last application mode.
View Trace:
Displays a graphical representation of the last 10 batches of
the SQC process to be viewed.
View History:
Displays all the calculated results for the batches of the SQC
process to be viewed.
3.4.12 Density
Four methods of density determinations can be made with
the Explorer Pro balance. These are:
1. solids more dense than water,
2. solids less dense than water,
3. liquid density,
4. porous material (impregnated with oil).
Mean Tare: Off, 00.000 (default is Off)
When Tare Required is set to Mean Tare or Add w/ Mean Tare,
this menu selection will allow manual entry of the Mean Tare
value.
Auto Sample: Off, On (default = Off)
Allows Auto Sample to be turned On or Off . When Auto Sample
is turned on, the Statistics process is automatic.
Auto Print: Off, On (default is Off)
Allows Auto Print to be turned On or Off . When Auto Print is
turned on, the final calculated results are automatically printed
out at the end of the SQC process.
A Density Determination Kit Part Number 470007-010 is designed
to be used with Ohaus Explorer Pro balances. Illustrations
in this procedure refer to the density kit, however, you may
use whatever lab apparatus that will suit the requirements
for density measurements. A built in reference density table
for water at temperatures between 10ºC and 30ºC is included
in the balance software.
When making density measurements, the material should weigh
at least 10.0 mg on an analytical balance and 100 mg on a
precision balance.
Explorer
Pro
USE
Balance Preparation for Density Measurement
The Explorer
hook which can be attached to the bottom of the balance.
When making density measurements, it is necessary to support the balance on both sides allowing enough clearance
to for a beaker to fit under the balance, a platform or laboratory jacks can be used. Remove the protective adhesive
cap from the bottom of the balance as shown in the figure
below. Screw the weigh below hook into the bottom of the
balance.
Pro
Balance is equipped with a weigh below
EN-31
BOTTOM OF BALANCE
PLATE
PROTECTIVE
COVER
WEIGH BELOW
HOOK INSTALLATION
LOCATION
WEIGH BELOW HOOK
BALANCE SUPPORT
EN-32 Explorer
Pro
Solid Density Determinations for items More Dense Than
Water
DENSITY 01/01/03 03:05:45 PM
TEMP
SETUP
START
*
0.00
Temp: 25.0 Air: 0.00 g
Aux D: 0.9971 Liquid: 0.00 g
Volume: 0.00 g
0g
Before using the Density feature, the Density Setup should be
reviewed. Highlight the SETUP soft key and press the Enter
button. Assign a Library Name.
NOTE: All Density settings made with a Library Name will be
directly associated with the Library name. When a Library Name
is called up, all settings will be retained with the Library Name.
Review the settings and change as necessary.
DENSITY SETUP
Library Name: DENSITY
Mode: Solid
Auxiliary Liquid: H2O
Auxiliary Density: 0.9971
Temperature (C): 25.00
Porous Material: Off
Auto Sample: Off
Auto Print: Off
Display Settings: Default
Exit
g
6100g
The density
Q
is the quotient of the mass
m
Q
=
m
and the volume V.
V
Density determinations are performed by using Archimedes’
principle
in a fluid loses weight by an amount equal to that of the fluid it
displaces. The density table for water is included in the
Explorer balance software.
The density of a solid is determined with the aid of a liquid
whose density, Qo,
liquid). The solid is weighed in air
liquid (B). The density Q
weighings as follows:
The balance allows direct determination of the buoyancy
P
simplified:
. .
. This principle states that every solid body immersed
. .
is known (water is used as an auxiliary
((
) )
(
A
) and then in the auxiliary
((
) )
can be calculated from the two
A
•
A -B
•
Q
0
Q
0
(P =A - B
Q =
) and consequently the above formula can be
A
Q =
P
Q = Density of the solid
A
= Weight of the solid in air
B = Weight of the solid in the auxiliary liquid
Q
= Density of the auxiliary liquid at a given tempera-
0
ture (this value de
pends on the temperature). The
density table for water is included in Explorer balances.
P = Buoyancy of the solid in the auxiliary liquid (corresponds to
A -B).
In the event that a different liquid is to be used, provisions are made to enter the density of the desired
liquid and enter its name into a library. The following
procedure uses water.
Explorer
Pro
EN-33
Place a beaker underneath the balance (not supplied)
and suspend a precision calibrated thermometer 0 °C to
30 °C on the edge of the beaker.
• Record and store the temperature in the Setup Menu.
• Tare the balance.
• Press Enter button to start the Density application.
• Follow directions of the prompts
• “Place weight in Air, Press Enter”. Place the solid sample
on the pan and press the Enter button.
• “Place weight in Liquid, Press Enter”. Suspend the solid
sample from the Weigh Below Hook into the beaker. Use
a fine wire or thread. Fill the beaker with auxiliary liquid
(liquid of known density
ethanol). Insure that the liquid will cover the sample by
at least 1 cm after immersion.Ensure that no air bubbles
adhere to the immersed part of the supporting wire. Remove air bubbles by moving the wire or by means of a
fine brush.
Weigh the solid (buoyancy P ) by pressing the Enter
button.
• The display indicates the density in grams/ cc.
Q
' usually distilled water or
0
A typical final display is shown which indicates all of the
parameters and values.
DENSITY 01/01/03 03:05:45 PM
TEMP
*
Temp: 25.0 Air: 100.19 g
Aux D: 0.9971 Liquid: 95.17 g
0g
SETUP
19.90
Volume: 5.03 g
START
g/cc
6100g
EN-34 Explorer
Pro
Improving the Accuracy of the Result of Solid Density
The following tips should help you improve the accuracy of
the results in the density determination of solids.
Temperature
Solids are generally so insensitive to temperature fluctuations
that the corresponding density changes are of no consequence.
However, as work is performed with an auxiliary liquid in the
density determination of solids, their temperature must be taken
into account as the temperature has a greater effect with liquids
and causes density changes in the order of magnitude 0.I to
1% per °C. This effect is already apparent in the third decimal
place of the result.
To obtain accurate results, we recommend that you always
take the temperature of the auxiliary liquid into account on all
density determinations.
Air Buoyancy
1 1
1 CM3 of air weighs approximately 1.2 mg (depending on the
1 1
physical condition). As a consequence, in the weighing in air,
each solid experiences a buoyancy of this magnitude (the socalled “air buoyancy”) per cm3 of its volume.
Liquid Density Determinations
The density of a liquid can be made using a sinker of known
volume. The sinker is weighed in air and then in the liquid
whose density is to be determined, The density, Q, can be
determined from the two weighings as follows:
Q
=
Density of the liquid
A
Q =
A - B
V
= Weight of the sinker in air
B
=
Weight of the sinker in liquid
V
=
Volume of the sinker
P
=
Buoyancy of the sinker in the liquid
(
P = A-B)
In DENSITY SETUP, set Mode to Liquid Density and enter sinker
volume in cc’s.
After weighing the sinker in air and then weighing the sinker
immersed in liquid, the balance calculates the density of the
liquid and is displayed in grams/cc.
However, the air buoyancy must be taken into account only
when a result is required with an accuracy of 3 to 4 decimal
places. To correct for this, the air buoyancy (0.0012 g per cm
volume of the body) is added to the calculated result:
Calculated density + 0.0012 g/cm
3
air buoyancy =
effective density
Surface tension of the auxiliary liquid
Adhesion of the liquid to the Weigh Below Hook causes an
apparent weight increase of up 3 mg.
As the Weigh Below Hook is immersed in the auxiliary liquid in
both weighings of the solid (in air and in the auxiliary liquid),
the influence of the apparent weight increase can be neglected
because the balance is tared before every measurement.
To reduce the effect of air bubbles and to ensure the greatest
possible accuracy, use a few drops of a wetting agent (not
supplied) and add them to the auxiliary liquid.
3
Explorer
Pro
EN-35
Porous Material Density Determinations
The density of a porous (oil impregnated part) can be made
with the balance. Weigh the part (dry) prior to oil impregnation
and
record its weight
. You must also know the density value
of the oil to be used in immersing the part before starting. In
this procedure, you will follow the method for solid density
measurements using water.
When Porous Material is selected and the Enter
button is
pressed, a second display POROUS MATERIAL SETUP is shown.
Enter the dry weight of the porous material and the density of
oil used to impregnate the part.
POROUS MATERIAL SETUP
Dry Weight: 0.000
Density Of Oil: 0.000
Exit
ADJUSTMENTS
START/ACCEPT
When the START key is selected, pressing the Enter button will
start the Density application. The screen displays “Place Weight
In Air, Press Enter” message and the Density process has
started. At this point, the soft key has changed into an ACCEPT
key. When the ACCEPT key is selected, the weight value has
been accepted as valid data.
SETUP
When selected, the DENSITY SETUP screen will be displayed.
TEMP/AUX D/SINKER
Depending on the setting, this will allow manual editing to the
Temp field, the Aux. D field, or the Sinker field.
Library Name:
Allows storage and recovery of parameters and results. When
a library is recovered, the library name will replace the
application name in the application mode screen. The library
name has maximum of 8 characters.
Exit the POROUS MATERIAL SETUP display. The DENSITY SETUP
screen re-appears.
To Determine Wet Density
Wet density of the sample can be calculated by following the
normal Solid Density procedure using the oil impregnated part.
When in DENSITY SETUP menu, turn off Porous Material, then
follow previous solid density measuring procedure.
Mode: Solid Density, Liquid Density (default is Solid Density)
Allows selection of mass for density calculations.
Auxiliary Liquid: H2O, Other (default is H2O)
Allows selection of type of liquid for density calculation. If water
is selected, the liquid density is fixed to 0.9971 g/cc and the
temperature of the water can be manually entered. If other is
selected, the density of the liquid can be manually entered.
Auxiliary Density: 00.000
When Auxiliary Liquid is set to water, Auxiliary Density is fixed
to 0.9971 g/cc. When Auxiliary Liquid is set to other, the density
of liquid can be manually entered.
Temperature: 00.00
When Auxiliary Liquid is set to water, the temperature of the
water can be manually entered.
EN-36 Explorer
Porous Material: Off, On (default is Off)
Allows Porous Material to be turned On or Off . When Porous
Material is turned on, a screen is displayed to allow the manual
entry of Dry Weight and Density Of Oil.
Sinker Volume: 000.00
Allows the volume of the sinker to be manually entered.
Auto Sample: Off, On (default is Off)
Allows Auto Sample to be turned On or off . When Auto Sample
is turned on, the Density process is automatic.
Auto Print: Off, On (default is Off)
Allows Auto Print to be turned On or off . When Auto Print is
turned on, the final calculated results are automatically printed
out at the end of the Density process.
Display settings: Custom/ Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Settings is set to custom, information on the result lines of the
Density mode can be individually turned on or off. If Display
Setting is set to default, all the information is turned on.
Pro
Exit:
In the sub menu, will return to the previous menu.
In the main menu, will return to the last application mode.
Explorer
Pro
EN-37
3.5Balance Settings
The Explorer Pro balance contains eleven submenus that
are accessible from the Main Menu. The submenus are
Calibration, Balance Options, Readout, Application Modes,
Units, Interface, Print Options, GLP Print Options, Library,
Lockout and Factory Reset.
Each of the submenus contain settings that will affect the operation
of the balance. Please review all submenus settings to obtain
the best performance from the balance. Make the necessary
settings to suit specific needs.
3.5.1Calibration
Press the Menu button and select CALIBRATION. Press the
button, the CALIBRATION screen is displayed. Explorer Pro
balances offers a choice of five calibration methods: Internal
Calibration, Autocal Calibration, Span Calibration, Linearity
Calibration and User Calibration,
Linearity, Span and User calibration are disabled for Type
Approved/LFT balances Class II and Class III.
CALIBRATION
Internal
AutoCal : Off
AutoCal Delta (d): 0
Span
Linearity
User
User Cal Weight (g): 500.00
Cal Test
Exit
IMPORTANT !
DO NOT DISTURB THE BALANCE
DURING CALIBRATION.
Internal Calibration
Calibration is accomplished using the internal calibration mass.
Internal calibration can be performed at any time providing the
balance has warmed up to operating temperature and is level.
EnterEnter
Enter
EnterEnter
AutoCal: On, Off (default is On)
When AutoCal is set ON, the balance performs a self calibration
when a measured predefined delta temperature change occurs.
After calibration, the display returns to the last application mode.
Environmental conditions must be met for the AutoCal to be
successfully completed.
AutoCal Delta: -100 - +100 (default is 0)
AutoCal Delta allows the internal calibration mass value to be
adjusted. This permits calibrating the balance using an external
mass which is traceable to a certified standard.
Perform the internal calibration procedure. After the calibration is
completed, zero the balance.
Place a certified mass equal to the
span calibration valuespan calibration value
span calibration value
span calibration valuespan calibration value
of the balance.
Compare the reading on the balance to the expected weight
being used. If the reading is above the expected weight, the
delta entered in digits is negative. If the reading is below the
expected weight, the delta entered in digits is positive.
See following example:
Actual Weight Reading:200.0014
Expected Weight Reading: 200.0000
Delta Weight (d): 0.0014
Delta weight in Digits:-14
Recalibrate using internal calibration. After calibration, place
the certified mass on the pan and see if the mass agrees with
the displayed value. If not, repeat procedure until internal
calibration reading agrees with the certified mass.
Span Calibration:
Span calibration utilizes two calibration points, one at zero
load and the other at specified full span. See the following
table.
EN-38 Explorer
Pro
CALIBRATION MASSES
LINEARITY SPAN ONLY
CAPACITY MASSES MASSES
12000 g 5000g/10000 g 10000 g
22000 g 10000g/20000 g 20000 g
32000 g 15000g/30000 g 30000 g
It is recommended that masses must meet or exceed
ASTM Class 1 Tolerance. Calibration masses are
available as accessories.
Additional calibration values to be used are shown on the display
screen. The best accuracy is achieved using the mass closest
to the full span value. After calibration, the display returns to the
last application mode.
The display indicates the difference in weight between calibration
mass placed on the pan and the previous weight value which
was stored in the balance. After Cal Test, the display returns to
to the last application mode.
3.5.2Balance Options
Press the Menu button, and select Balance Options.
Press the Enter button. The BALANCE OPTIONS screen is
displayed.
BALANCE OPTIONS
Language: English
Display Date/Time: Date/Time
Date Format: DD/MM/YY
Date: 00/00/00
Time Format: 12 Hour
Time: 00:00:00
Contrast:
Brightness:
Audible Signal: Off
Exit
Languages: English, Spanish, etc...(default is English)
Allows the selection of English, Spanish, French, German or
Italian as the operating language to be displayed.
Linearity Calibration:
Linearity calibration utilizes three calibration points, one at zero
load and two at specified calibration masses. This method
minimizes deviation between actual and displayed weights
within the balance's weighing range. After calibration, the
display returns to the last application mode.
User Calibration:
User calibration is used when it is desired to calibrate the balance
using a user defined mass. The user defined mass value has
to be entered in the User Cal Weight. After calibration, the
display returns to the last application mode.
User Calibration Weight: 25%-100% of capacity (default is
Span)
Defines the mass value for User calibration. The user calibration
may now be done with the mass selected.
Cal Test:
Calibration test feature allows a check of a known calibration
mass against the last stored calibration information in the
balance.
Display Date/Time: Time, date, etc...(default is Date/Time)
Allows display selection of Time, Date, Date/Time or Off in the
upper right-hand corner of the Application Mode screen.
Date Format: MM/DD/YY, YY/MM/DD, ETC...(default is MM/DD/
YY)
Allows setting of one of six date formats: MM/DD/YY,
YY/MM/DD , DD/MM/YY, DD/YY/MM, MM/YY/DD or YY/DD/MM.
Date: OO/00/00 (default is none)
Allows setting the present date.
Time Format: 12 Hour, 24 hour(default is 12 Hour)
Allows setting the time format of either 12 hours or 24 hours.
Time: 00:00:00 (default is none)
Allows setting the present time of hours, minutes and AM/PM.
Valid numbers change with time format from 1->12 to 0->23.
When in 24 hour mode, no AM/PM flag appears.
Contrast:
Allows adjusting the contrast level of the display.
Explorer
Pro
EN-39
Brightness:
Allows adjusting the brightness level of the display.
Audible Signal: On, Off (default is Off)
With audible signal set On, each button press will generate a
tone.
Exit:
When selected, returns to previous screen.
3.5.3Readout
Press the Menu button, and select Readout.
Press the Enter button. The READOUT screen is displayed.
READOUT
Filter Level: Medium
Stability Level (d): 1.0
Auto Zero (d): 0.5
Auto Tare: Off
Legal for Trade: Off
Exit
Auto Tare: On, Off (default is Off)
Sets the Auto Tare function to On or Off. This feature provides
auto tare in all application modes. When Auto Tare is On, the
balance waits for the container/load to be placed on the pan
and automatically tares it. This function is repeated for each
new container/load.
Legal for Trade: On, Off (default is Off)
Sets the Legal for Trade (LFT) function either On or Off. Refer to
paragraph paragraph 3.5.12 for additional information.
3.5.4Application Modes
Press the Menu button, and select Application Modes.
Press the Enter button APPLICATION MODES screen is displayed.
APPLICATION MODES
Weighing: On
Parts Counting: Off
Animal Weighing: Off
Percent Weighing: Off
Check Weighing: Off
G/N/T: Off
Filling: Off
Differential: Off
Formulation: Off
Statistics: Off
SQC: Off
Density: Off
Filter Level: Low, Medium, High (default is Medium)
Sets the balance reading averaging level to a value of Low,
Medium or High. A higher filter level will provide a more repeatable
measurement.
Stability Level: 0.5, 1.0, etc... (default is 1.0)
Sets the balance stability level for the stability indicator to either
0.5, 1.0, 2.0 or 5.0. A 0.5 setting is equivalent to end result of
.5 display digit. A 5.0 setting is equivalent to end result of 5
display digits.
Auto Zero: Off, 0.5, etc... (default is 0.5)
Sets the balance auto zero level to either Off, 0.5, 1.0, 2.0 or
5.0. Auto Zero tracking compensates for the drift at zero load.
0.5d means 0.5 display digit drift compensation per second.
5.0d means 5 display digit drift compensation per second.
All application modes are displayed on this screen. Each mode
may be set on or off. These modes are cycled through when the
Mode button is pressed.
3.5.5Units
Press the Menu button, and select Units. The UNITS screen
displays a list of the available weighing units. Depending on
the model, some units are not available.
UNITS
Milligram (mg): Off
Gram: On
Kilogram: Off
Oz avdp (oz): Off
Pound avdp (lb): Off
Carat (ct): Off
Penny weight (dwt): Off
Troy OZ (ozt): Off
Grain (GN): Off
HK Tael (Hkt): Off
SG Tael (SGt): Off
ROC Tael (ROt): Off
Scroll down using arrow button to view remaining units.
EN-40 Explorer
Pro
UNITS
Carat (ct): Off
Penny weight (dwt): Off
Troy OZ (ozt): Off
Grain (GN): Off
HK Tael (Hkt): Off
SG Tael (SGt): Off
ROC Tael (ROt): Off
Newton (N): Off
Momme (m): Off
Tical (ti): Off
Custom (Cst): Off
Exit
Milligram (mg): On, Off (default is Off)
Unit (mg) = g x 1000, displayed readability by 1.
Gram (mg): On, Off (default is On)
Unit (g) = g x 1, displayed readability by 1.
Kilogram (kg): On, Off (default is Off)
Unit (kg) = g x .001, displayed readability by 1.
OZ avdp (oz): On, Off (default is Off)
Unit Ounces (oz) = g x .03527396, displayed readability by 5.
Newton (N): On, Off (default is Off)
Unit (N) = g x 0.00980665, displayed readability by 1.
MOMME (m): On, Off (default is Off)
Unit (m) = g x 0.2666667, displayed readability by 5.
Tical (ti): On, Off (default is Off)
Unit (ti) = g x 0.0612395, displayed readability by 1.
Custom Unit (Cst): On, Off (default is Off)
When the Custom Unit is set On, the CUSTOM UNIT screen will
appear.
This feature can be used to create a custom weighing unit. It
permits entering a conversion factor which the balance will use
to convert grams to the desired unit of measure.
Conversion Factor x Grams = Custom Unit
Unit (Cst) = g x user set custom factor, displayed resolution can
not exceed resolution in gram.
Pound avdp (lb): On, Off (default is Off)
Unit Pounds (lb) = g x .002204623, displayed readability by 5.
Carats (ct): On, Off (default is Off)
Unit (ct) = g x 5, displayed readability by 5.
Pennyweight (dwt): On, Off (default is Off)
Unit (dwt) = g x .6430149, displayed readability by 1.
Troy OZ (ozt): On, Off (default is Off)
Unit (ozt) = g x .03215075, displayed readability by 5.
Grain (GN): On, Off (default is Off)
Unit (GN) = g x 15.43236, displayed readability by 2.
Hong Kong Tael (HKt): On, Off (default is Off)
Unit (HKt) = g x 0.02671725, displayed readability by 5.
Singapore Tael (SGt): On, Off (default is Off)
Unit (SGt) = g x 0.02645547, displayed readability by 5.
ROC Tael (ROt): On, Off (default is Off)
Unit (ROt) = g x 0.02666667, displayed readability by 5.
Conversion factors are expressed in scientific notation
and entered into the balance in three parts:
• Mantissa (0.1 and 1.999999)
• Exponent (10E)
• Least Significant Digit (LSD)
CUSTOM UNIT
Factor: 1.0
Exponents: E1
LSD: 1
Exit
SCIENTIFIC NOTATIONSCIENTIFIC NOTATION
SCIENTIFIC NOTATION
SCIENTIFIC NOTATIONSCIENTIFIC NOTATION
MantissaMantissa
Mantissa
MantissaMantissa
BetweenBetween
Between
Conv.Conv.
Conv.
Conv.Conv.
FactorFactor
Factor
FactorFactor
123.4 = .1234 x 1000 =.1234 x 10
12.34= .1234 x 100 = .1234 x 10
1.234= .1234 x 10 = .1234 x 10
.1234= .1234 x1 = .1234 x 10
.01234 = .1234 x .1 = .1234 x 10
.001234 = .1234 x .01 = .1234 x 10
.000123 = .123 x .001 = .123 x 10
BetweenBetween
0.1 and Exponent 0.1 and Exponent
0.1 and Exponent
0.1 and Exponent 0.1 and Exponent
1.9999991.999999
1.999999
1.9999991.999999
tissa (10 tissa (10
tissa (10
tissa (10 tissa (10
Man-Man-
Man-
Man-Man-
Exp.Exp.
Exp.
Exp.Exp.
EE
E
EE
))
)
))
3
2
1
0
-1
-2
-3
Explorer
Pro
EN-41
EXPONENTSEXPONENTS
EXPONENTS
EXPONENTSEXPONENTS
E-3Moves decimal point 3 places to the left.
E-2Moves decimal point 2 places to the left.
E-1Moves decimal point 1 place to the left.
E0E0
E0
E0E0
E1Moves decimal point 1 place to the right.
E2Moves decimal point 2 places to the right.
E3Moves decimal point 3 places to the right.
LSD .5Adds one decimal place display counts by 5’s.
LSD 1Display counts by 1’s.
LSD 2Display counts by 2’s.
LSD 5Display counts by 5’s.
LSD 10Display counts by 10’s.
LSD 100Display counts by 100’s.
Leaves decimal point in normalLeaves decimal point in normal
Leaves decimal point in normal
Leaves decimal point in normalLeaves decimal point in normal
position.position.
position.
position.position.
LSD’sLSD’s
LSD’s
LSD’sLSD’s
3.5.6Interface
Press the Menu button, and select Interface.
Press the Enter button, The INTERFACE screen is displayed.
3.5.7Print Options
Press the Menu button, and select Print Options.
Press the Enter button PRINT OPTIONS screen is displayed.
PRINT OPTIONS
Auto Print: Off
Print Interval: Off
Stable Data: Off
Numeric Data: Off
Exit
Auto Print: Off, interval, etc... (default is Off)
When enabled, the Auto Print feature causes the balance to
automatically output display data in one of three ways:
continuously, at user specified time intervals, or upon stability.
OFFTurns off the auto print feature.
IntervalProvides a user specified printing interval
StableProvides printed data each time a stable
reading is achieved.
ContinuousOutputs data continuously.
INTERFACE
Baud Rate: 2400
Parity: None
Data Bits: 7
Stop Bits: 2
Exit
Baud Rate: 300, 1200, etc...(default is 2400)
Baud Rate is selectable between 300, 1200, 2400, 4800 or
9600 BPS.
Parity Bit: None, Odd, Even (default is None)
Parity Bit is selectable between None, Odd or Even.
Data Bits: 7, 8 (default is 7)
Data Bits is selectable between 7 or 8 data bits.
Stop Bits: 1, 2 (default is 2)
Stop Bits is selectable between 1 or 2 stop bits.
Print Interval: 1-3600 (default is 1)
When Auto Print is set to Interval, a specified printing interval
between 1 and 3600 seconds can be set.
Stable Data: Load, Load & Zero (default is Load)
When Auto Print is set to Stable, an option of Load or Load &
Zero can be set.
LoadWill print stable load data only.
Load & ZeroWill print stable load data and stable zero
data.
Numeric Data: On, Off (default is Off)
When set On, will print numeric data.
3.5.8GLP Print Options
GLP stands for Good Laboratory Practice. The GLP Print Options
enables the printing of a Date & Time, Balance ID, Project Name,
User Name, Calibration, Reference, Application Mode and Result
lines. Press the Menu button, and select GLP Print Options.
Exit:
Will return to previous display.
Press the Enter button GLP PRINT OPTIONS screen is displayed.
EN-42 Explorer
Pro
GLP PRINT OPTIONS
Project Name: OHAUS
User Name: OHAUS
Date & Time: Off
Balance ID: Off
Project Name: Off
User Name: Off
Calibration: Off
Reference: Off
Application Mode: Off
Result Line 1: Off
Result Line 2: Off
Result Line 3: Off
Project Name: (default is OHAUS)
A project name of up to 8 characters may be entered.
User Name: (default is OHAUS)
A user name of up to 8 characters may be entered.
Date & Time: On, Off (default is Off)
When set On, will output date and time.
Balance ID: On, Off (default is Off)
When set On, will output Balance ID.
Project Name: On, Off (default is Off)
When set On, will output project name.
Result Lines: On, Off (Default is Off)
Each of the six Result Lines can be set independently. When set
to On, the Result Line information will be printed
Exit:
When selected, returns to previous menu.
3.5.9Library
Seven application modes in the balance have provisions for
storing a library name, they are: Parts Counting, Check
Weighing, Differential Weighing, Formulation, Statistics,
Statistical Quality Control (SQC), Density. When a library
name is selected, the associated function is also displayed.
A Library menu is provided which allows the selected library
name and function to be run, deleted or delete all entries. If
you have accessed the library and do not want to run or
delete a name, an exit to weighing selection can be made
by pressing the Menu button which does not affect the library
Select LIBRARY function and press the Enter button. Display
advances to LIBRARY menu with RUN, DELETE and DELETE
ALL options shown.
User Name: On, Off (default is Off)
When set On, will output User name.
Calibration: On, Off (default is Off)
When set On, prints out the calibration data after the completion
of every calibration process. See sample of span calibration.
SAMPLE PRINTOUT
---------------SPAN CAL-------------Cal: 400.000 g
Old:399.445 g
Diff:-0.555 g
Wt. ref______________________
Name_______________________
-----------------END--------------------
Reference: On, Off (default is Off)
When the Reference function is set ON, it will output the value
of weight used as a reference in either Percent Weighing mode
or Parts Counting mode.
LIBRARY
Run
Delete
Delete All?: No
Exit
Select either RUN, DELETE, or DELETE ALL. When RUN is selected,
a screen showing all stored libraries is displayed. Any stored
library can be recalled, enabled and run. All parameters from
a previously run library are retained.
When a particular name is selected and the DELETE selection
is made, that particular name and function with all parameters
is removed from the library.
DELETE ALL, when selected, removes the entire contents of the
library.
Application Mode: On, Off (Default is Off)
When set to On, the Application name will be printed during a
print operation.
Explorer
Pro
EN-43
3.5.10Lockout
Provides the capability to lock the Menu settings individually to
protect selected parameters against modifications. Locked means
the items can be viewed but not changed.
Press the Menu button, and select Lockout.
Press the Enter button LOCKOUT screen is displayed.
The default setting for all items on the LOCKOUT screen are
Unlocked.
3.5.11Factory Reset
The FACTORY RESET allows parameters to be set to the default
values. Each menu item can be reset by setting to Yes. After
accepting the changes, the balance will reset the selected menus
items. Global Reset will change all menus to a factory setting in
one step.
Press the Menu button, and select Factory Reset.
Press the Enter button FACTORY RESET screen is displayed.
The balance default parameters are listed as follows:
Calibration:
Auto Cal: On
Autocal delta: 0
User Cal weight = Span weight
Balance Options:
Language: English
Display Date/Time: Date/Time
Date Format: MM/DD/YY
Time Format: 12 Hour
Readout:
Filter Level: Medium
Stability Level (d): 1.0
Auto Zero (d): 0.5
Auto Tare: Off
Legal for Trade: Off
Balance Options: No
Readout: No
Application Modes: No
Units: No
Interface: No
Print Options: No
GLP Print Options: No
Lockout: No
Global Reset: No
Accept Changes: No
Print Options:
Auto Print: Off
Print Interval: Off
Stable Data: Off
Numeric Data: Off
GLP Print Options:
Balance ID=OHAUS
User Name = OHAUS
All others Off
Lockout:
All Unlocked
EN-44 Explorer
UNLOCKED/OFF LOCKED/ON
3.5.12LFT Legal for Trade
Legal for Trade (LFT) is a software controlled option which can
be set ON in the Readout menu. When LFT is set ON, certain
items in the Calibration, Readout and Print menus are
automatically preset and locked to permit the balance to operate
in a legal for trade application and works in conjunction with a
Lockswitch. Default setting is OFF. See default table.
LFT DEFAULT TABLE
LFT and LockswitchDefault Value
Balance Menu
Readout
Stability LevelLocked to 1
Auto ZeroLimited to OFF & 0.5
Auto CalLocked to ON
Print Current SettingsBalance will only output stable data.
BOTTOM OF BALANCE
LFT SWITCH LOCATION
PLATE
SEALS OVER PLATE
Pro
When the balance is first turned ON and LFT has been previously
set ON, the initial display will indicate that LFT is ON.
The last digit on the weighing display will be highlighted in white
and is used to indicate the auxiliary digit.
Depending upon country regulations, additional settings will have
to be locked. Before sealing the balance, check with your local
Weights and Measure official.
3.5.13Hardware Lockswitch
Access to the various menus can be disabled setting the Lockswitch located on the PC board inside the balance to locked
position. The Lockswitch locks out all menus which have been
set to LOCKED. The default setting for the Lockswitch is UNLOCKED.
3.5.14Sealing the Balance
Certified balances have a securing sticker and additional
labeling applied at the factory. When subsequent verification
is carried out, they can be sealed either with a lead seal and
wire, or with a new securing sticker.
Example of Sealing Method
3.6Printing Data
Printing data requires that the Interface menu, Print Options and
GLP Print Options are set properly.
Pressing the Print button will initiate printing each time it is
pressed.
Sample printout is shown below with GLP Options turned on.
SAMPLE PRINTOUT
DATE: 19/02/03
09:56:16
9.989g
Ref:
Balance id:000001B9A925
Project Name:OHAUS
User Name: OHAUS
Explorer
Pro
EN-45
4.CARE AND MAINTENANCE
4.1 Cleaning
To keep the balance operating properly, the housing and pan should be kept clean and free from foreign material. If necessary, a
cloth dampened with a mild detergent may be used. Keep calibration masses in a safe dry place.
4.2Troubleshooting
SYMPTOMSYMPTOM
SYMPTOM
SYMPTOMSYMPTOM
Unit will not turn on.
Incorrect weight reading.
Cannot display weight in desired unit.
Cannot access desired application.
Unable to change menu settings.
RS232 interface not working.
PROBABLE CAUSE(S)PROBABLE CAUSE(S)
PROBABLE CAUSE(S)
PROBABLE CAUSE(S)PROBABLE CAUSE(S)
Power cord not plugged in or properly
connected to balance.
Balance was not tared before weighing.
Balance out of level.
Balance not properly calibrated.
Desired unit not enabled.
Desired application mode is not enabled.
Menu locked.
Interface parameters not properly set up.
REMEDYREMEDY
REMEDY
REMEDYREMEDY
Check power cord connections.
Press >O/T< with no weight on the
pan, then weigh item.
Adjust leveling feet.
Recalibrate correctly.
Enable units in Units menu.
Enable desired application in
Application Modes menu.
Verify that Lockswitch is in the Off
position, unlock menu in the
Lockout menu.
Verify interface settings in RS232
menu correspond to those of
peripheral device.
Unstable readings.
Error message display.
Incorrect calibration
Incorrect cable being used.
Cable connections.
Vibration on table surface.
Balance not tared.
Internal calibration not adjusted properly.
Refer to Accessory list for proper
cable.
Check cable connections are
installed properly. Check correct
cable end is plugged into the
balance.
Check environmental conditions.
Close draft shield doors.
Change averaging level to a higher
setting or place balance on a stable
surface
See Error Codes list.
Tare the balance.
Perform calibration adjustments.
EN-46 Explorer
4.3Error Codes List
Error Codes List
The following list describes the various error codes that can appear on the display and the suggested remedy.
Data Errors
1.0 Transient error (hardware error, probably static discharge). If error persists, the balance must be
serviced.
1.1 Balance temperature transducer hardware error.
1.2 No data from main board.
Tare Errors
2.0 Balance is unable to stabilize within time limit after taring. Environment is too hostile or balance needs
recalibration.
Calibration Errors
3.0 Incorrect or no calibration mass used for calibration. Recalibrate with correct masses.
RS232 Errors
4.4 RS232 buffer is full.
Pro
User Errors
7.0 User entry out of bounds.
7.2 Number outside of display capacity.
Over-Under Load Errors
8.0 Hardware error causing an internal weight signal which is too low. Check if pan is off. If not, the balance
must be serviced.
8.1 Hardware error caused by an internal weight signal which is too high. Check load on the pan which may
be excessive. If error persists, the balance must be serviced.
8.2 Power-on load out of specification (LFT only)
8.3 Rated capacity exceeded. Remove excessive weight from pan.
8.4 Underload condition on balance. Check that the proper pan is installed.
8.5 Auto Cal weight internal sensor indicated its weight on the pan.
CheckSum Errors
9.1 Bad factory checksum. If error persists, have the balance serviced.
9.2 Bad factory checksum. If error persists, have the balance serviced.
9.3 Bad factory checksum. If error persists, have the balance serviced.
9.4 Auto Cal data failed checksum. This failure will disable access to the Auto Cal feature (if installed).
9.5 Factory calibration data failed checksum.
9.6 Bad program checksum.
9.7 Bad CMOS checksum.
9.8 User calibration data failed checksum.
9.9 Factory temperature compensation data failed checksum.
Explorer
Pro
EN-47
4.4Service Information
If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized Ohaus Service
Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-free at (800) 526-0659. An
Ohaus Product Service Specialist will be available to help you. Global contact addresses and phone numbers are accessible on
www.ohaus.com.
4.5Replacement Parts
DescriptionU.S. Part No. Global Part No.
Power Pack, 100/120 V ac US Plug (Cord set part of power pack)490202-01021202536
Power Pack, (Cord set required for UK, European and Australian)490203-01021202537
Cord Set, 230 V ac, UK Plug 76448-00 89405
Cord Set, 230 V ac, European Plug 76212-00 87925
Cord Set, 230 V ac, Australian plug 76199-01 88751
In-Use Display Cover Kit8085004280850042
4.6Accessories
Description
Calibration Masses - ASTM Class 1 Tolerance:Calibration Masses - ASTM Class 1 Tolerance:
Calibration Masses - ASTM Class 1 Tolerance:
Calibration Masses - ASTM Class 1 Tolerance:Calibration Masses - ASTM Class 1 Tolerance:
1 kg 49016-1180780021
2 kg 49026-1180780024
4 kg 49046-1180780027
RS232 Interface Cable, Blunt end (user defined) AS017-0180850013
RS232 Interface Cable, IBM
RS232 Interface Cable, (connects impact printer)8050057080500570
RS232 Interface Cable, IBM
RS232 Interface Cable, Apple
Printer SF4280250933
® -
PC 25 Pin AS017-0280850014
® -
PC 9 Pin AS017-0980850015
®
llGS/Macintosh AS017-1080850072
EN-48 Explorer
5.TECHNICAL DATA
5.1 RS232 Commands
Command
Character Description
CBegin span calibration
xDSet 1 second print delay (set x = 0 for OFF, or x = 1 for ON)
PMApplication mode.
xFLSet Averaging Filter Level 1= Low, 2= Medium, 3= High
LBegin linearity calibration
Pro
PPrint displayed weight (stable or unstable)
TSame as pressing Tare key.
PVVersion: print name, software revision and LFT ON (if LFT is set ON)
xALSet Auto-Zero level to x. x = 0 for OFF, x = 1 for 0.5d, x=2 for 1.0d, x=3 for 2.0d, x=4 for
5.0d.
Esc RResets Setup and Print menus to factory defaults
OnTurns balance on
OffTurns balance off
x%Set % reference weight (x) in current unit
x#Set PC reference weight (x) in current unit
P#Print PC reference weight
P%Print percent reference weight
xMSet current Application mode to x. x = 1 for Weighing, x = 2 for Parts Counting, x = 3 for
Animal Weighing, x = 4 for Percent Weighing, x = 5 for Check Weighing, x = 6 for G/N/T,
x = 7 for Filling, x = 8 for Differential, x = 9 for Formulation, x = 10 for Statistical, x = 11
for SQC, 12 = for Density
xAWSet Animal Weigh Level to x. x = 1 for LOW, x = 2 for MEDIUM, x = 3 for HIGH.
Field: Weight Unit Stab CR LF
Length: Max 9 5 1 1 1
xAMSet animal mode. Where x is 1=Manual, 2=Semi and 3=Auto
SAWStart Animal cycle.
xTDownload Tare value in current unit. Sending 0T clears tare.
PIDPrint current user ID string
xIDProgram user ID string. 1-8 characters
ACAbort calibration
xUCSet user defined weight
UCInitiate user calibration
ICInitiate internal calibration
PTIMEPrint current time
mm/dd/yyDATE Set date
hh:mm:ssTIMESet time
PDATEPrint current date
Explorer
Pro
EN-49
5.2Specifications
Admissible Ambient Conditions
Use only in closed rooms
Ambient Temperature range:5 ºC to 40 ºC
Atmospheric humidity:80% rh @ to 30 ºC
Voltage fluctuations:–15% +10%
Installation category:II
Pollution degree:2
Power supply voltage:12 VAC, 50/60 Hz or 12 VDC, 1A
Displayed Information Settings, Selectable Environment Settings, Selectable
Auto-Print Settings, Integral Weigh Below Hook, Contrast and Brightness
Control, Protective In-Use Cover
Repeatability (Std. dev.) (g)0.1
Linearity (g) (+)0.4
Tare rangeFull capacity by subtraction
Safe overload capacity150 % of capacity
Stabilization time
Sensitivity drift PPM/°C 3
(10 °C - 30 °C)
Operating temperature range:
w/internal calibration10° to 40°C/50° to 104°F
w/o internal calibration10° to 30°C/50° to 86°F
Power requirementsExternal Adapter, 100-120 V ac, 220 V ac, 50/60 Hz
Plug configuration for US, Euro, UK, Japan & Australia
Display (in/cm)0.6/1.5
Pan size (in/cm)11 x 1 3/4 x 13/27.9 x 4.4 x 33
Dimensions
(WxHxD) (in/cm)14 x 5 1/4 x 16/35.5 x 13.3 x 40.6
<4 seconds
*** Units availability is country dependent.
EN-50 Explorer
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the
warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be
defective at no charge, provided that the product is returned, freight prepaid, to Ohaus.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive
materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than
Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the
authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for
any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for
further details.
Pro
Explorer
AA
A
AA
Pro
IndexIndex
Index
IndexIndex
EN-51
AC Adapter Installation 5
Accessories 47
Admissible Ambient Conditions 49
Ambient Temperature rang 49
Animal Weighing 16
Application Modes 39
Applications 11
Atmospheric humidity 49
Audible Signal 39
Auto Print 41
Auto Tare 39
Auto Zero 39
AutoCal 37
AutoCal Delta 37
Average Piece Weight 12
BB
B
BB
Balance ID 42
Balance Location 4
Balance Options 38
Balance Preparation for Density Measurements 31
Balance Settings 37
Baud Rate 41
Brightness 39
CC
C
CC
Cal Test 38
CALIBRATE 6, 37
CALIBRATION MASSES 38
Capacity 49
CARE AND MAINTENANCE 45
CHECK WEIGHING 19
Cleaning 45
Connecting Power and Communications 5
Contrast 38
Cord Set 47
Custom Unit (Cst) 40
DD
D
DD
Data Bits 41
Date 38
Date Format 38
Density 30
Differential Weighing 22
Display Date/Time 38
Features 3
FILLING 20
Filter Level 39
Formulation 24
GG
G
GG
G/N/T WEIGHING 21
Global Part No. 47
GLP Print Options 41, 42