Rice Lake MSI6260cs User Manual

MSI6260cs
RF Digital Crane Scale
Calibration Guide
INTRODUCTION
The MSI-6260CS Series crane scales are best calibrated using a remote indicator such as the MSI­3750CS, MSI-9850, or the MSI-9750A Handheld. Calibration procedures using these devices are covered in their respective User Guides. If above listed remote devices are not in use with the MSI-6260CS, the only way to calibrate it is through the Comm. Port using the Terminal Access Mode. MSI recommends having test weights of at least 50% capacity to properly calibrate the Crane Scale. The calibration procedures will operate with as little as 12.5% of capacity, though rated accuracy cannot be guaranteed. For details of Multi-Point, Certified, or other specialized calibrations, refer to full version of MSI-6260CS User’s Guide.
The following procedure describes how to perform a standard, single point calibration for a single load cell MSI-6260CS Crane Scale. Some menu items may appear slightly different than shown, depending upon actual software application version. However, those relative to Scale Calibration are accurate.
REQUIRED EQUIPMENT:
1) PC or terminal with terminal program and RS-232 serial port.
2) Serial Communications cable, PN: 501705-0001.
3) Known calibration test weight between12.5 and 100% of the scale’s capacity.
ACCESSING EMBEDDED CELL SCALE
The MSI-6260CS Crane Scale is controlled by an embedded M SI-9000 CellScale. All scale functions & operating parameters are stored within the MSI-9000.
1) Set the terminal program to emulate an ASCII terminal.
2) Set the terminal programs port protocol to 9600 Baud, no parity, 8 data bits and 1 stop bit.
3) Disable flow control in the terminal program.
4). Tur n sc ale ON.
5). Connect the MSI-6260CS comport to the PC serial port with the 501705-0001 serial cable.
6). From PC, press ENTER. Main Cell Scale Manu will appear. Select Channels/Calibration press 1.
---------- MSI 9000 - CellScale ------------------------
-(c) 2005 Measurement Systems International -
---------------------------------------------------------------­13:48:06 01/01/08
(1) Channels/Cali brat ion
(2) Communications (3) Display (4) Feature Locks/Passwords (5) Product IDs (6) Set Points (7) System (8) About...
Enter choice:
Page 1 of 7
CALIBRATION PROCEDURE
.
1) Select Channel Calibration press 1.
***** Channel Setup *****
(1) Channel Calibration
(2) Data backup (3) Filter (software) (4) Chann el S c an lis t fu nc t i ons (5) Motion Compensation (Accelerometer option required) (6) Moti on D et ec t i on (7) Tare (8) Total (9) Units Lock (a) Zero (b) Standard Mode: Industry (c) 4 to 20mA (d) Temperature Sensor Installed: No (ESC) Exit
2) Start the Calibration procedure by pressing 1.
***** Choose Channel Parameters *****
(1) Continue. These settings are correct
(2) Choose the channel to calibrate: 1-1 (3) Calibration timeout: None (ESC) Exit
3) Press 1 to perform a completely new calibration.
***** Channel Calibration ***** (1) Perform a completely new calibration
(2) Re-do the current calibration (3) Manually adjust current cal points (4) Reset all cal parameters to defaults (5) Change display resolution countby (ESC) Exit
Calibrate channel #: 1-1
The channel name is: ....................
Calibration Capacity: 0. Calibration Countby: 1. Calibration Units: Display Resolution Countby: 1. Display Units:
4) The MSI-9000 CellScale prompts for the calibration units of mass. Press number value corresponding to desired units.
***** Choose Calibration Units ***** (1) Pounds
(2) Kilogr am s (3) Tons (4) Metric Tons (5) Ounces (6) Grams (7) Deka-Newtons
Type a key. ..