INTRODUCTION
The MSI-6260CS Series crane scales are best calibrated using a remote indicator such as the MSI3750CS, MSI-9850, or the MSI-9750A Handheld. Calibration procedures using these devices are covered
in their respective User Guides. If above listed remote devices are not in use with the MSI-6260CS, the
only way to calibrate it is through the Comm. Port using the Terminal Access Mode. MSI recommends
having test weights of at least 50% capacity to properly calibrate the Crane Scale. The calibration
procedures will operate with as little as 12.5% of capacity, though rated accuracy cannot be guaranteed.
For details of Multi-Point, Certified, or other specialized calibrations, refer to full version of MSI-6260CS
User’s Guide.
The following procedure describes how to perform a standard, single point
calibration for a single load cell MSI-6260CS Crane Scale. Some menu items may
appear slightly different than shown, depending upon actual software application
version. However, those relative to Scale Calibration are accurate.
REQUIRED EQUIPMENT:
1) PC or terminal with terminal program and RS-232 serial port.
2) Serial Communications cable, PN: 501705-0001.
3) Known calibration test weight between12.5 and 100% of the scale’s capacity.
ACCESSING EMBEDDED CELL SCALE
The MSI-6260CS Crane Scale is controlled by an embedded M SI-9000 CellScale. All scale functions
& operating parameters are stored within the MSI-9000.
1) Set the terminal program to emulate an ASCII terminal.
2) Set the terminal programs port protocol to 9600 Baud, no parity, 8 data bits and 1 stop bit.
3) Disable flow control in the terminal program.
4). Tur n sc ale ON.
5). Connect the MSI-6260CS comport to the PC serial port with the 501705-0001 serial cable.
6). From PC, press ENTER. Main Cell Scale Manu will appear.
Select Channels/Calibration press 1.
---------- MSI 9000 - CellScale ------------------------
-(c) 2005 Measurement Systems International -
---------------------------------------------------------------13:48:06 01/01/08
(1) Channels/Cali brat ion
(2) Communications
(3) Display
(4) Feature Locks/Passwords
(5) Product IDs
(6) Set Points
(7) System
(8) About...
Enter choice:
Page 1 of 7
CALIBRATION PROCEDURE
.
1) Select Channel Calibration press 1.
***** Channel Setup *****
(1) Channel Calibration
(2) Data backup
(3) Filter (software)
(4) Chann el S c an lis t fu nc t i ons
(5) Motion Compensation (Accelerometer option required)
(6) Moti on D et ec t i on
(7) Tare
(8) Total
(9) Units Lock
(a) Zero
(b) Standard Mode: Industry
(c) 4 to 20mA
(d) Temperature Sensor Installed: No
(ESC) Exit
2) Start the Calibration procedure by pressing 1.
***** Choose Channel Parameters *****
(1) Continue. These settings are correct
(2) Choose the channel to calibrate: 1-1
(3) Calibration timeout: None
(ESC) Exit
3) Press 1 to perform a completely new calibration.
***** Channel Calibration *****
(1) Perform a completely new calibration
(2) Re-do the current calibration
(3) Manually adjust current cal points
(4) Reset all cal parameters to defaults
(5) Change display resolution countby
(ESC) Exit
Calibrate channel #: 1-1
The channel name is: ....................
Calibration Capacity: 0.
Calibration Countby: 1.
Calibration Units:
Display Resolution Countby: 1.
Display Units:
4) The MSI-9000 CellScale prompts for the calibration units of mass. Press number value
corresponding to desired units.
***** Choose Calibration Units *****
(1) Pounds
(2) Kilogr am s
(3) Tons
(4) Metric Tons
(5) Ounces
(6) Grams
(7) Deka-Newtons
Type a key. ..
Page 2 of 7