Rice Lake Minebea Intec Combics 3 Digital Weight Indicator Manual

98648-018-29

Operating Instructions

Sartorius Combics 3

Models CAISL3 | CAIS3 Indicators
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Symbols and Signs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Warning and Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . 4
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
IP Protection Rating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Use in Legal Metrology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Device Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General View of the Equipment . . . . . . . . . . . . . . . . . . . . . . . 6
Storage and Shipping Conditions . . . . . . . . . . . . . . . . . . . . . . . 7
Installation Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Checking Package Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Acclimatizing the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Equipment Downtime . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Connecting a Weighing Platform . . . . . . . . . . . . . . . . . . . . . . . 7
Connecting a Weighing Platform to WP 1 . . . . . . . . . . . . . . . . . 8
Connecting a Weighing Platform to WP 2 . . . . . . . . . . . . . . . . . 9
Interface Pin Assignments COM1, COM2
and PS2 with Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Interface Pin Assignment Chart COM1 . . . . . . . . . . . . . . . . . . . 11
Connecting a PC via Interface COM1 . . . . . . . . . . . . . . . . . . . . 12
Interface Pin Assignment Chart COM2 . . . . . . . . . . . . . . . . . . . 13
Connecting a PC via Interface COM2 . . . . . . . . . . . . . . . . . . . . 14
Pin Assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Interface Pin Assignment Chart PS2 . . . . . . . . . . . . . . . . . . . . . 15
Connecting a Barcode Scanner via the PS2 Interface . . . . . . . . 15
Closing the Combics Indicator . . . . . . . . . . . . . . . . . . . . . . . . . 16
Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Warm-up Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Using a Verified Device in Legal Metrology . . . . . . . . . . . . . . . 16
Configuring Weighing Platforms . . . . . . . . . . . . . . . . . . . . . . 17
Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Overview of the Setup Menu in Service Mode . . . . . . . . . . . . . . 18
Setup Menu for ADC Configuration . . . . . . . . . . . . . . . . . . . . . 20
Analog/Digital Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Setting Parameters for ADC Configuration . . . . . . . . . . . . . . . . 21
Analog/Digital Converter (ADC) Configuration . . . . . . . . . . . . . 23
Entering Geographical Data for Use in Legal Metrology . . . . . . 27
Entering Adjustment and Linearization Weights . . . . . . . . . . . . 32
Function Allocation of the J Key for Calibration/Adjustment 33 External Calibration/Adjustment with Factory-Defined Weight
(Default Weight) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
External Calibration/Adjustment with a User-Defined Weight . . 36
Internal Calibration/Adjustment . . . . . . . . . . . . . . . . . . . . . . . . 37
Adjustment Without Weights . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Function Allocation of the J Key for Linearization and
Setting/Deleting the Preload . . . . . . . . . . . . . . . . . . . . . . . . . . 40
External Linearization with Factory-Defined Weights
(Default Weights) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
External Linearization with User-Defined Weights . . . . . . . . . . 42
Setting the Preload . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Deleting the Preload . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Operating Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Weighing Mode Operating Design . . . . . . . . . . . . . . . . . . . . . 46
Saving Settings in Weighing Mode . . . . . . . . . . . . . . . . . . . . . 48
Applying the Tare Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Input Through the Digital Control Port . . . . . . . . . . . . . . . . . 48
Display in Weighing Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Menu Operating Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Menu Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Saving Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Setup Overview (Parameters) . . . . . . . . . . . . . . . . . . . . . . . . . 56
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Weighing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Adjustment/Configuration Counter for Standard Scales . . . . . 72
Device Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Calibration and Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Configuration for Use in Legal Metrology . . . . . . . . . . . . . . . 79
External Calibration and Manual Adjustment with Default
Weights (Weighing Parameters: Factory Settings) . . . . . . . . . . 81
SQmin Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
SQmin Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Data ID Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Using individual ID codes . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Data Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Connection Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Configuring the Data Interface as a COM Port . . . . . . . . . . . . 91
Setting the SBI Data Output . . . . . . . . . . . . . . . . . . . . . . . . . 92
Data Input Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Data Output Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
External Keyboard Functions (PC Keyboard) . . . . . . . . . . . . . . 97
Configuring the Data Interface as a Printer Port . . . . . . . . . . . 98
Printout Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
GMP-compliant Printouts . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Sample Printouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105
Cleaning Stainless Steel Surfaces . . . . . . . . . . . . . . . . . . . . . . .106
Replacing the Dust Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Safety Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
ADC Scale Interface 2*3000e (Option A8) . . . . . . . . . . . . . . . . 108
ADC Scale Interface 10,000e (Option A10) . . . . . . . . . . . . . . . 109
Device Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Documents and Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Documents List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Sartorius Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Declaration of Conformities . . . . . . . . . . . . . . . . . . . . . . . . . 115
EC type-approval Certificate . . . . . . . . . . . . . . . . . . . . . . . . 118
Test Certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Plates and Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Appendix: Guide to Verification of Weighing Instruments . . 124
Appendix: General Password . . . . . . . . . . . . . . . . . . . . . . . . 127
2

Contents

3

Notes on Using this Manual

t Please read this entire manual carefully and completely before using the device. t Read the safety precautions carefully. t This manual is part of the product. Keep it in a safe and easily accessible location. t If the manual should be lost or misplaced, please contact Sartorius for a replacement or
download the latest manual from our website: www.sartorius.com

Symbols and Signs

The following symbols are used in this manual:
Warning symbol for various types of dangers.
2
These symbols are explained in more detail in the “Safety Instructions" section.
h
This symbol indicates useful information and tips.
This symbol indicates notes on use in legal metrology within the scope of validity of Council Directive No. 90/384/EEC, replaced by 2009/23/EC (models MS...-.CE...).
e, 1, This and similar symbols mean that the respective key should be pressed.
T T ..., This means that this key must be pressed more than once.
t Indicates a required action
y Describes the result of an action
1.
If a procedure has multiple steps...
2. ... the steps are numbered consecutively.
Indicates an item in a list
h
Technical advice/hotline:
Phone: +49.551.308.4440 Fax: +49.551.308.4440

Safety

Combics indicators comply with the European Council Directives as well as international regulations and standards for electrical equipment, electro magnetic compatibility, and the stipulated safety requirements. Improper use or handling can, however, result in damage and/or injury.
t Read these operating instructions
carefully before use. This will prevent damage to the equipment.
! Do not use this device in hazardous
areas.
! The device may only be opened by
trained service technicians.
! Disconnect the device from power
before connecting or disconnecting peripheral devices.
! If you operate the device under
ambient conditions subject to higher safety standards, you must comply with any applicable installation regulations.

Installation

! Caution when using pre-wired
RS-232 connecting cables: RS-232 cables purchased from other manufacturers often have pin assignments that are incompatible with Sartorius products. Be sure to check the pin assignments against the chart in this manual before connecting the cable, and disconnect any lines identified differently from those specified by Sartorius.
! Use only standard cables that have
protective grounding conductors. The protective conductor must not be disconnected for any reason.
! If there is visible damage to the device
or power cord: unplug the device and secure it against further use.
! Connect only Sartorius accessories
and options, as these are optimally designed for use with your device. Therefore, do not use any proprietary solutions. The operator shall be solely responsible for installation and testing of any modifications to Sartorius equipment, including connection of cables or equipment not supplied by Sartorius. Information on operational quality (in line with norms pertaining to immunity) is available on request.
t If you have any problems with your
device, contact your local Sartorius office, dealer or service center.

IP Protection Rating

CAISL models are rated to IP44 (with
option L1: IP65) – CAIS models are rated to IP69K. – The IP65/IP69K protection rating is
ensured only if the rubber gasket is
installed and all connections are
fastened securely (including the caps
on unused sockets).
Weighing platforms must be installed
and tested by a certified technician. – If you install an interface port or
battery connection after setting up
your indicator, keep the protective
cap in a safe place for future use.
The cap protects the interface
connector from vapors, moisture
and dust or dirt.

Use in Legal Metrology

When the indicator is connected to a
weighing platform and this equipment
is to be verified, ensure that the
applicable regulations regarding
verification are observed. Please read
and observe the “Guide to Verification”
on the enclosed CD. – When connecting Sartorius platforms,
observe the permitted weighing range
as listed in the “Guide to Verification
of Weighing Instruments" and the
Declaration of Conformity. – A sticker with the “Sartorius" logo was
affixed to the indicator as a control
seal following verification.
This seal will be irreparably damaged
if you attempt to remove it. This will
nullify the verification's validity. In this
case, re-verification would be required
in compliance with all relevant national
regulations and laws.
4

Warnings and Safety Instructions

5

Device Description

Description

Combics 3:
– Is robust and durable, thanks to its stainless steel housing. – Is easy to clean and disinfect. – Is easy to operate, thanks to the following features: – Large, backlit, fully graphical dot-matrix display – Large keys with positive click action – Alphanumeric character entry – Plain-text user guidance
– Can be operated independently of the weighing platform location. – Has a range of interfaces for flexible use. – Has optional password protection for operating parameters.
Combics 3 speeds up your routine procedures with:
– Automatic initialization when the scale is switched on. – Fast response times. – Automatic taring when a load is placed on the weighing platform. – Independence from location of platform installation. – Designation of weight values with up to 4 lines of alphanumeric text. – Flexibility afforded by diversity of interfaces. – Security through password protection. – Can be controlled via two external computers using various protocols. – Barcode scanner connection option for entering tare value or IDs (6 units). – Ability to input tare values via the number keys. – LED for measurement range identification. – Connection option for a second weighing platform. – Alibi memory. – Configurable printout. – Flex print

Intended Use

The Combics 3 indicator is a robust indicator for daily quality control in industrial applications. It was designed for suitable scales or weighing platforms that correspond to the described technical specifications. Any other use beyond this is considered improper.
Display and Keypad 1 Alphanumeric keypad 2 LEDs
(for checkweighing and classifica-
tion)
3 Toggle key for alphabetic input 4 Configuration:
Access to the product data memory
5 Delete key 6 Configuration:
Access to Setup 7 Toggles the display between gross
value (net value plus tare) and net
value (gross value minus tare) 8 Toggles between application
program and application-specific
information 9 Toggles the display between normal
and 10-fold higher resolution
10 Start data output 11 Start calibration or adjustment 12 Display of 2nd weight unit or SQmin
(depending on the setup)
13 Tare 14 Zeroing 15 Toggle between weighing platforms 16 On/Off 17 Function keys 18 Graphic-capable dot-matrix display
Back panel
19 “UNICOM"
Optional:
– RS-232|RS-485 or 422 interface,
e.g. for: – PC connection
– Printer connection – Digital I/O – 4 to 20 mA – Profibus DP – Ethernet TCP/IP|ModBus TCP
20 RS-232 interface “COM1" e.g. for:
– PC connection – Printer connection – Digital input – Stop light output
21 “PS2" e.g. for:
– Keyboard connection – Barcode scanner connection
22 Power cord with country-specific
plug
23 Input for menu access switch
(standard or legal-for-trade mode) for WP 1
24 Weighing platform WP 1 connection
for analog scales, optional for “xBPI"|"SBI" scales
25 Weighing platform WP 2 connection,
optional for analog scales or “xBPI"|“SBI" scales
26 Input for menu access switch
(standard or legal-for-trade mode) for WP 2
27 RS-232 interface “COM2" e.g. for:
– PC connection – Printer connection
28 Vent valve: 1.5 Nm
6

General View of the Equipment

Display and keypad
Rear view: Model type CAISL
Rear view: Model type CAIS
19
20
21
20
22
17 16
11
25
23
19
27
25
23
15 14 13
1
10
6 7 8 9
2
18
12
5
4
3
27 26
24
22
24
26
28
PS
2
7

Installation

When an indicator is ordered with special equipment, the desired options come pre-installed from the factory.

Storage and Shipping Conditions

3
Unpacked devices can lose their precision if subject to extreme vibrations. Excessive vibrations may compromise the safety of the equipment.
Do not expose the equipment to unnecessarily extreme temperatures, moisture, shocks,
blows or vibration.
Permissible storage temperature: –10 ... +40 °C

Installation Location

Avoid adverse influences at the place of installation: – Extreme temperatures (operating temperature: –10 ... +40 °C) – Aggressive chemical vapors – Extreme moisture
(according to IP protection class)

Unpacking

t After unpacking the device, check it for any visible damage as a result of rough handling
during shipment.
y If you detect any damage, proceed as directed in the chapter entitled “Care and
Maintenance" under “Safety Inspection."
t Save the original packaging for any future transport.
Unplug all connected cables before packing the equipment.

Checking Package Contents

Indicator – Operating instructions – Options (special accessories) as listed on the bill of delivery

Acclimatizing the Device

Condensation can form on the surfaces of a cold device when it is brought into a substantially warmer area.
t Allow the device to acclimatize for about 2 hours at room temperature, leaving it
unplugged from AC power.

Equipment Downtime

Switch off the equipment when not in use.

Connecting a Weighing Platform

See the chapter entitled “Getting Started."
3
Make absolutely sure that the device is unplugged from the power supply before connect­ing/disconnecting any peripheral device (e.g. printer, PC) to or from the data interface.
Steps
1.) Connect weighing platform to the indicator.
2.) Configure the analog/digital converter (ADC): see chapter “Configuring Weighing Platforms, section “Setting Parameters for ADC Configuration."
3.) Carry out a calibration/adjustment: Adjustment/Calibration, see chapter “Configuring Weighing Platforms", section “Eternal Calibration/Adjustment" and linearization, see chapter “Configuring Weighing Platforms", section “External Linearization".
4.) Connect peripheral devices, e .g. printer to the COM1 or UNICOM interface: see the chapter entitled “Data Interfaces."

Connecting Weighing Platforms to WP1

An analog Sartorius platform (CAPP, CAPS, IU or IF) or a commercially-available DMS load cell can be connected to the Combics indicator WP1 input.
3
The load cell should be connected by a certified technician who has received specialized training from Sartorius. Any installation work that does not conform to the instructions in this manual results in forfeiture of all claims under the manufacturer’s warranty.
3
Peripheral devices should be connected by a certified technician who has received specialized training from Sartorius. Any installation work that does not conform to the instructions in this manual results in forfeiture of all claims under the manufacturer’s warranty.
3
Disconnect the equipment from the power supply before starting connection work.
t Set up the weighing platform (see Operating instructions for the weighing platform). t Place the cable from the weighing platform next to the indicator. t Open the Combics indicator:
Loosen the 10 cap nuts on the front panel. Remove the front panel.
Installing Connection and Interface Cables
3
The cable gland (IP69K protection) is pre-mounted on the indicator. Please use extreme caution when performing any work on the equipment that affects this cable gland. You must use a torque wrench. The torque for this cable gland is 5 Nm.
Preparing the Cable
t Strip approx. 14 cm from the end of the cable. t Shorten the shielding to approx. 2 cm and pull back over the insulation. t Strip approximately 5 mm of the insulation from the wires of the connecting cable and
affix ferrules to the wire ends.
8
Getting Started
Attaching the Cable Entry
3
Please use extreme caution when performing any work on the equipment that affects this cable gland. You must use a torque wrench. The torque for this cable gland is 5 Nm.
t Remove the protective cap from the bore hole on the indicator. t Insert the included cable gland through the bore hole and secure from the inside using
the locknut (1).
t Insert the cable through the cable gland until the shielding (2) comes into contact with
the clamps (3). Tighten the screw-down nut (4) until the gasket (5) inserted between the screw-down nut and cable forms a small beaded rim.
t Check the shielding and clamps.
t Securely connect the wires of the connecting cable in accordance with the terminal
assignments.
t After you close the housing again, use a pressure gauge to check the integrity of the
IP69K protection. For details, contact the Sartorius Service Center.
Connecting Cables
t Insert all cable wires through the ferrite case, wind them around the ferrite case and then
reinsert back through the ferrite case.
t Screw the wires tightly into the clamps.
See the following pages for pin assignment
t Refer to the data sheet or operating instructions of the weighing platform for details
on the assignment of wire colors/signals. Ensure any lines that are not assigned are insulated correctly.
t When connecting a load receptor that uses 4-conductor technology (the cable of the
weighing platform to be connected only has 4 lines), connect clamp pairs 1 and 2 (EXC+ und SENSE+), and 5 and 6 (SENSE- und EXC-) with a wire jumper.

Connecting Weighing Platforms to WP2

An IS platform can be connected to the WP2 connection of the Combics indicator.
Features
IS weighing platforms process weighing data independently of the indicator. – Internal calibration/adjustment option – IS...-0CE models: have a separate approval number, printed on a tag that is affixed to
the cable.
Please observe the conditions described in the manual for the weighing platform you
connect.
4
1
5
2
3
9
Getting Started

Interface Pin Assignments COM1, COM2 and PS2 with Options

Digital PCB CAIS3 (IP69K)
COM1, COM2 and PS2 terminal assignments (applies to all PCBs)
PS2
1 LOAD_PRINTER 11 Clear to Send (CTS) 21 5 V switched 2 RESET_OUT 12 Data Terminal Ready (DTR) 22 PS2_Daten 3 GND 13 Data Input (RXD) 23 PS2_Timer 4 GND 14 Data Output (TXD) 24 GND 5 5V_OUT 15 GND COM2 6 5V switched 16 Universal In 31 CTS_COM2 7 GND 17 Control Output: “lighter" 32 DTR_COM2 8 GND 18 Control Output: “equal" 33 RXD_COM2 9 Schirm 19 Control Output: “heavier" 34 TXD_COM2
10 LINE_OUT 20 Control Output: “set" 35 GND
36 GND
A8 terminal assignments
1 EXC+ Bridge supply voltage (+) 2 SENSE+ Sense (+) for bridge supply voltage 3 OUT+ Measuring voltage positive 4 OUT- Measuring voltage negative 5 SENSE- Sense (-) for bridge supply voltage 6 EXC- Bridge supply voltage (–)
Keypad LED + Display
Interface PCB for RS-232/RS-485 for IS platforms (option A6/A7)
A6/A7 terminal assignments
1 CTS 11 TxD/RxD+ 2 DTR 12 TxD/RxD­3 RxD 13 LINE_OUT 4 TxD 14 LINE_OUT 5 GND 15 GND 6 Calibration Lock 16 GND
Keypad LED + Display
Interface PCB for ADC 10.000e (option A20)
A20 terminal assignments
1 EXC+ 2 SENSE+ 3 OUT+ 4 OUT­5 SENSE­6 EXC-
Keypad LED + Display
10
Getting Started
PS2 COM1 COM2 A8
Keys
LED + Display
A6/7
Keys
LED + Display
A20
Keys
LED + Display
PS2 COM1 COM2
PS2 COM1 COM2
Weighing platform Interface 2
Weighing platform Interface 1
Weighing platform Interface 2
11
Getting Started
Interface PCB for RS-232/RS-485 for IS platforms (option A62/A72)
A62/A72 terminal assignments
1 CTS 11 TxD/RxD+ 2 DTR 12 TxD/RxD­3 RxD 13 LINE_OUT 4 TxD 14 LINE_OUT 5 GND 15 GND 6 Calibration Lock 16 GND
Keypad LED + Display

Interface Pin Assignment Chart COM1

Model type CAISL (IP44 protection)
COM1 female connectors: 25-pin D-Submini female connector (DB25S) with screw lock hardware for cable gland
Recommended interface connector: 25-pin D-Submini (DB25) with shielded cable clamp assembly and shield plate (Amp type
826 985-1C) and fastening screws (Amp type 164868-1) Pin assignment: Pin 1 Shield
Pin 2 Data output (TxD) Pin 3 Data input (RxD) Pin 4 Internal ground (GND) Pin 5 Clear to Send (CTS) Pin 6 Not assigned Pin 7 Internal ground (GND) Pin 8 Internal ground (GND) Pin 9 Not assigned Pin 10 Not assigned Pin 11 +12 V for printer Pin 12 \RES_OUT Pin 13 +5 V Switch Pin 14 Internal ground (GND) Pin 15 Universal switch Pin 16 Control output: “lighter" Pin 17 Control output: “equal" Pin 18 Control output: “heavier" Pin 19 Control output: “set” Pin 20 Data Terminal Ready (DTR) Pin 21 Ground power supply (GND) Pin 22 Not assigned Pin 23 Not assigned Pin 24 Power supply +15...25 V (peripherals) Pin 25 +5 V
A62/72
Keys
LED + Display
PS2 COM1 COM2
Weighing platform Interface 1

Connecting a PC via Interface COM1

Use the following cables to connect a PC to the indicator in accordance with the RS-232C/V24 standard (max. cable length 15 m):
Model type CAISL: connecting cable 7357312 Model type CAIS: connecting cable YCC02-D9F6
Pin Assignment
Pin assignments for the cable from the indicator to an RS-232 PC interface (COM1).
Indicator side PC side
Model type CAISL D-Sub plug D-Sub socket 25-pin D-Sub male connector 9-pin or 25-pin
>1 Sgn GND >7 5 GND >7 GND TxD >2 2 RxD >3 RxD RxD >3 3 TxD >2 TxD DTR 20 8 CTS >5 CTS CTS >5 4 DTR 20 DTR
Model type CAIS D-Sub plug D-Sub socket Open cable end 9-pin or 25-pin
Sgn GND 15 5 GND >7 GND TxD 14 2 RxD >3 RxD RxD 13 3 TxD >2 TxD DTR 12 8 CTS >5 CTS CTS 11 4 DTR 20 DTR
12
Getting Started
RS232
13
Getting Started

Interface Pin Assignment Chart COM2

Model type CAISL (IP-44 protection)
COM2 female connectors: 9-pin D-Submini female connector (DB9S) with screw lock hardware for cable gland
Recommended interface connector: 9-pin D-Submini (DB9) with shielded cable clamp assembly and shield plate and
fastening screws (Amp type 164868-1) Pin assignment: Pin 1 +5 V out
Pin 2 Data output (TxD) Pin 3 Data input (RxD) Pin 4 Clear to Send (CTS) Pin 5 Internal ground (GND) Pin 6 Reset Pin 7 Not assigned Pin 8 Data Terminal Ready (DTR) Pin 9 Load Printer
5
9
1
6

Connecting a PC via Interface COM2

Use the following cables to connect a PC to the indicator in accordance with the RS-232C/V24 standard (max. cable length 15 m):
Model type CAISL: connecting cable 7357312 Model type CAIS: connecting cable YCC02-D9F6

Pin Assignment

Pin assignments for the cable from the indicator to an RS-232 PC interface (COM2).
Indicator side PC side
Model type CAISL D-Sub plug D-Sub socket 9-pin D-Sub male connector 9-pin or 25-pin
Sgn GND 5 5 GND 7 GND TxD 2 2 RxD 3 RxD RxD 3 3 TxD 2 TxD DTR 8 8 CTS 5 CTS CTS 4 4 DTR 20 DTR
Model type CAIS D-Sub plug D-Sub socket Open cable end 9-pin or 25-pin
Sgn GND 35 5 GND 7 GND TxD 34 2 RxD 3 RxD RxD 33 3 TxD 2 TxD DTR 32 8 CTS 5 CTS CTS 31 4 DTR 20 DTR
14
Getting Started
RS232
15
Getting Started

Interface Pin Assignment Chart PS2

Model type CAISL (IP-44 protection)
PS2 female connector: 6-pin miniature socket PS2 (Mini-DIN)
Recommended interface connector: 6-pin miniature socket PS2 with integrated shielded cable clamp assembly
Pin Assignments: Pin 1 Keyboard data (data interface cable)
Pin 2 Not assigned Pin 3 Internal ground (GND) Pin 4 +5 V switched Pin 5 Keyboard clock Pin 6 Not assigned

Connecting a Barcode Scanner via the PS2 Interface

Accessory YBR02CISL
t Disconnect the indicator from AC power (unplug the AC adapter).
For model type CAISL:
t Connect the barcode scanner via PS/2.
For model type CAIS:
t Pin assignment, see “Interface Connection Assignments COM1, COM2 and PS2"
(implemented via the YCC02-BR02 connecting cable or as option M8).
3
1
4 2
6
5

Closing the Combics Indicator

t Re-attach the front panel and secure it with the 10 cap nuts.

Connecting the Device to AC Power

The indicator is powered through the pre-installed power cord. The power supply is integrated into the indicator. The device can be operated with a supply voltage of 100 to 240 V.
3
The power connection must be made in accordance with the regulations applicable in your country.
Make sure that the voltage rating printed on the manufacturer's ID label is identical to that of your local line voltage. If the voltage specified on the label or the plug design of the AC adapter do not match the rating or standard you use, please contact your Sartorius office or dealer.
t Check the voltage rating and plug design. t The device must be plugged into a properly installed wall outlet.
Protection Class 1 Device
t The device must be plugged into a properly installed wall outlet which has a protective
grounding conductor (PE).

Safety Precautions

3
If you use an electrical outlet that does not have a protective grounding conductor, ensure that an equivalent protective conductor is installed by a certified electrician (as specified in the applicable regulations for installation in your country). The protective effect must not be negated by using an extension cord without a protective grounding conductor.
Before using for the first time, any superstructure parts must be completely installed. Avoid connecting the equipment to lines that have a heavy electrical load, e.g. compressors, large machinery, etc.

Warm-up Time

To deliver exact results, the device must warm up for at least 30 minutes after connection to AC power. Only after this time will the device have reached the required operating temperature.

Using a Verified Device in Legal Metrology

Ensure that there is a warm-up time of at least 24 hours after connection to the power supply.
16
Getting Started
17

Configuring Weighing Platforms

Service Mode

Purpose
The Service mode enables access to additional menu items in the Setup menu which are not displayed when the Service mode is not active. The most important calibration and adjustment work for the indicator and for the connected weighing platform can be carried out in the Service menu, e.g. ADC configuration.
When the Service mode is active, an “S” is shown in the top right-hand corner of the display. To deactivate the Service mode, restart the indicator (turn the indicator off and back on again).
Activating the Service Mode
t Press e to turn on the device.
y When turned on the scale is in an application program.
t Enter the service password (see Appendix General User Password) and press M to
confirm.
y The device in now is Service mode. An “S" appears in the top right-hand corner of the
display.
t Press the “q" soft key several times to select the “Device parameters" line.
t Press the “O" soft key.
y The “Device" submenu will open.
t Select and open the corresponding submenu. Repeat this procedure until the desired
menu item in the lowest menu level can be opened.
t View or change the menu item (confirm with “l") and use “o" to return to the previous
menu.
t Press M or “oo" to exit the Setup menu.
Exiting Service Mode
Turn the device off and then on again to return to the normal application mode. If you exit the Setup menu without changing settings by confirming with M or the “
oo"
soft key, the Service mode will remain active. Press the M key to re-open the Setup menu.

Overview of the Setup Menu in Service Mode

o = Factory setting x = User-defined setting
Service password entry Setup Application parameters, please refer to the “Basic Application Programs” manual
Fn key “Setup Overview (Parameters)" Device parameters Info, see the “Setup Overview (Parameters)" section Language, see the “Setup Overview (Parameters)" section
Setup access with service password
Device parameters
WP1
RS-232
1
)
SBI standard SBI verifiable
o IS-232
ADC-232
RS-485 1)
o IS-485
ADC-485
Internal
ADC configuration (see the “Setup Menu for ADC Configuration" section) Calibration/Adjustment
CAL key function
o Ext. cal./adj.; factory-def. wt.
Ext. cal./adjust.; user-defined weight Ext. lineariz.; factory-def. wts Ext. lineariz.; user-def. wts Set preload Delete preload Key blocked
Cal./adj. sequence
Cal. then auto adjust
o Cal. then manual adjust
isoCAL function
3
)
o Off
Adjustment prompt
Activate external adjustment 2)
o Activated
Deactivated
Parameter for external weight
Cal./adj. weight: Lin. weight 1...4:
Adjust without weights
2
)
Input parameters
Nominal load: Resolution: Sensitivity 1…4:
Save parameters
Yes
o No
Geographical data 2)
Input parameters
Geographical latitude Altitude Gravitational acceleration
Save parameters
Yes
o No
1
) Equipment version: – then blocked internally
2
) = Not available on devices verified for use in legal metrology
3
) only when operated with Sartorius IS weighing platforms or an external ADC
18
Configuring Weighing Platforms
19
Configuring Weighing Platforms
Device Parameters
WP1
Internal
Calibration/Adjustment unit
Grams /g Kilograms /kg Tons /t
o Pounds /lb
Menu items “Adapt filter" - “Factory settings: only bal. func," see the “Setup Overview (Parameters)" section
Off COM-1 (when the WP is assigned to this interface) COM-2 (when the WP is assigned to this interface) UNICOM (only if available)
WP2, see the “Setup Overview (Parameters)" section
RS-232
1
) similar to “Internal" menu for WP1
RS-485
1
) similar to “Internal" menu for WP1
o Off
COM-1 similar to WP 1 COM-2 similar to WP 1
COM-1, see the “Setup Overview (Parameters)" section
COM-2, see the “Setup Overview (Parameters)" section
"Control I/O ports" - “Terminal data," see the “Setup Overview (Parameters)" section
SQmin
SQmin input
SQmin WP1
SQmin WP1 0.000 kg
SQmin WP2
SQmin WP2 0.000 kg
SQmin WP3
SQmin WP3 0.000 kg
Display
No
o Yes
GMP print
o No
Yes
Alibi memory
2
)
Clear alibi memory
Yes
o No
Alibi memory period
In days 90
1
) Equipment version: – then saved internally
2
) Only if alibi memory is available (option)

Setup Menu for A/D Converter Configuration

Setup access in Service mode
WP1 - Internal ­ADC configuration Standard configuration Ranges Single-range mode Scale interval d
Max. cap.
Multi-interval mode Scale interval d
Range 1 Range 2 Range 3 Max. cap.
Multiple-range mode Scale interval d
Range 1 Range 2 Range 3 Max. cap.
Available units User-definable /o
Grams /g Kilograms /kg Carats /ct ...
Save parameters Yes
oNo
oVerifiable Accuracy Class III/IIII
configuration class
Ranges Single-range mode
D: E: Min. capacity: Max. cap.:
Multi-interval mode
D: E: Min. capacity: Range 1: Range 2 Range 3: Max. cap.:
Multi-interval mode
D: E: Min. capacity: Range 1: Range 2 Range 3: Max. cap.:
Available units User-definable /o
oGrams /g oKilograms /kg
...
Save parameters
Yes
oNo
20
Configuring Weighing Platforms
21
Configuring Weighing Platforms

Analog/Digital Converter (ADC)

Purpose
Adjust the parameters of the analog/digital converter to the connected load cell or weighing platform. After ADC configuration, the ADC in connection with the load sensor is defined as a scale.
3
Once the ADC configuration has been locked, the indicator can no longer be used to influence weighing results. The scope of functions available in the weighing instrument is defined by the A/D converter. Weighing functions that can be activated include reading weight values, taring, adjustment, reading the tare value, saving/deleting the tare entry.
Setup information
ADC configuration is only possible when the menu access switch is open. Re-close the menu
access switch after ADC configuration.
Before ADC configuration, you must first set whether or not the weighing platform will be
used as a standard or verifiable weighing platform.
ADC configuration is carried out in the Service mode in the Setup menu under “
WP 1" for
the first weighing platform and “
WP 2" for the second.
Entries made in ADC configuration will not be affected by a menu reset (returning the setup
parameters to their factory settings). See also the overview in the “Setup Menu for A/D Converter Configuration" section.

Setting Parameters for ADC Configuration

Standard or Verifiable Configurations
In ADC configuration, you must first select whether the weighing platform should be configured as a standard or verifiable (for use in legal metrology) weighing platform.
Standard configuration “
Standard"
Verifiable configuration “
Verifiable"
Select using the “
Q"or “q" soft key. Press the “O" soft key to confirm the setting and open
the Configuration menu.
Accuracy class
This menu item is not shown when the Standard configuration is active. When the Verifiable configuration is active (for weighing platforms verified or verifiable for use in legal metrology), only Class l/m can be selected. Activate the “
Accuracy class"
menu item, select “
Class III/IIII" and confirm your selection using the “l"
soft key.
Ranges
The capacity of the weighing platform can be divided into multiple ranges:
–“
Single-range mode":
The entire weighing range is divided into scale intervals on the basis of the lowest interval d and the maximum load.
–“
Multi-interval mode":
The “Multi-interval mode” function divides the weighing capacity into as many as four ranges, each with a different readability. Corresponding changes take place automatically. Once the scale has been tared, the highest possible resolution is available even if the weighing platform is loaded with a higher weight. This is only permitted in the accuracy classes l/m for the verifiable configuration.
–“
Multiple-range mode":
Multiple-range mode has two or three weighing ranges. When the range limit is exceeded, the scale switches into the next highest weighing range (lower resolution) and remains there. The scale can be returned to the lower weighing range (higher resolution) only by pressing the ( key and then unloading the scale.
Select the desired configuration using the “
Q" or “q" soft key. Confirm your selection using
the “
O" key. Make additional settings in the submenu: scale interval d/verification scale
interval e, minimum load (Verifiable configuration only), range limits (Multi-interval or Multiple range mode only) and maximum capacity. Confirm using the “l" soft key or cancel using “
Esc."
Scale interval d
Scale interval d indicates the resolution of the weighing instrument. The scale interval d can be entered only in increments of 1, 2, 5, 10, 20, etc.
When using verifiable or verified weighing platforms, the scale interval d is the same as
the verification scale interval e.
Verification scale interval e
The verification scale interval e indicates the resolution of the weighing instrument in legal metrology. The verification scale interval e can be entered only in increments of 1, 2, 5, 10, 20, etc. For weighing instruments of accuracy class l or m, e = d. This is why the scale interval d does not need to be entered separately. When “Standard configuration” is used, this menu item is not displayed.
Minimum capacity (min. cap.)
When “Standard configuration” is used, this menu item is not displayed. The minimum load of the connected weighing platform is entered under this menu item. The minimum load for scales of accuracy class l is 20 e and 10 e for class m. Important Note: The function of the minimum load setting is to warn operators that below this limit, the summation of tolerances might lead to significant measurement errors. In Germany, for example, initial weights below the minimum load are not allowed.
Maximum capacity (Max. cap.)
The maximum capacity is the maximum load that may be placed on the weighing instrument. When heavier weights are used the weighing instrument displays overload “H.” The scale intervals of the weighing instrument are calculated using the maximum load and the scale interval d (e.g. max. capacity = 15.000 kg, smallest scale interval d = 0.005 kg yields 3000 scale intervals). In legal metrology the total number of intervals must be no more than 3000 e, and when using multi-interval scales there must not be more than 3000 e intervals per range. In standard operation, as opposed to legal metrology, you can define a “Super Range” weighing instrument of over 3000 intervals. These parameters, however, may be influenced by physical restrictions.
Range 1, Range 2, Range 3
The range limits are entered for the individual ranges. The accuracy changes when these limits are exceeded. The following applies when entering limits: range 1 < range 2 < range 3 < maximum capacity.
This means that the weighing range can be divided into a maximum of 4 ranges. The resolution changes at intervals of 1, 2, 5, 10, 20 etc., where the lowest resolution is the smallest scale interval entered. Set ranges that are not required for use to zero.
Available units
With this function, you can make particular weight units (weight unit x, x=1, 2) inaccessible during weighing. Available units are indicated by a * on the display (more than one can be selected).
To enable or disable a unit, select the unit by pressing the “
Q" or “q" soft key, and then
press the “l" soft key (toggle function).
Saving parameters
To save the ADC configuration, use the “
O" soft key to select “Yes" and confirm with
the “l" soft key. The device software is reset, and the scale returns to the normal weighing mode. To exit the menu without saving configuration changes, press the “
o"
soft key.
Once these parameters have been configured, the A/D converter in conjunction with the load cell(s) is defined as a weighing instrument. The A/D converter, in conjunction with the weighing platform, can now be used like any standard weighing platform. In addition, the weight unit must be defined and the weighing platform adjusted (calibration, adjustment and linearization must be performed). For a detailed description of these procedures, see the “Adjustment in Service Mode" section.
22
Configuring Weighing Platforms
23
Configuring Weighing Platforms

Analog/Digital Converter (ADC) Configuration

The weighing platform must already be connected.
Open the menu access switch
The menu access switch is located on the back of the indicator right next to the weighing platform connection.
t Remove the cap. t Slide the switch to the left (= “open" position).
Activating the Service Mode
t See “Service Mode" on page 17.
Configuration
t Select weighing platform “WP 1." t If the “Internal" setting is not already activated (marked by “o"), select the setting
using the “
Q" or “q" soft key and confirm with “O."
y The message “Function active" appears briefly. The “WP 1 internal" menu
will then open.
t Open the ADC Configuration menu. t Select the desired configuration data record using the “Q" or “q" soft key:
Standard" or “Verifiable" (verifiable configuration). The default setting
depends on the data record.
t Open the menu for configuring A/D converter parameters. In this example, the menu for
ADC configuration of a weighing platform for use in legal metrology is opened. If “
Standard" was selected previously, then the “Accuracy class" is not
displayed.
t Open the first menu item.
For a standard configuration, the “
Ranges" menu item, for a verifiable configuration,
the “
Accuracy class" menu item.
When the “Verifiable” configuration is active, always select the “
Accuracy
class" menu item first.
t Set accuracy class l/m. The “o" symbol should mark the setting, if required confirm
using the “l" soft key.
t Press the “o" soft key to exit the menu item.
Open the “
Ranges" menu item.
In the example shown here, “
Single range mode" has been selected (marked
by “
o").
t To select a different weighing range configuration, use the “Q" or “q" soft key to select
the corresponding line and open the selected menu using the “
O" soft key.
y The selected weighing range configuration is now active. When you return from the
input menu for entering the weighing range parameters, the new range configuration is marked by a “
o."
For more information about range configuration, please see “Setting Parameters for ADC Configuration."
The default values displayed depend on the data record loaded and might have to be changed. In the example shown here, the A/D configuration is set with a “Verifiable” data record in single-range mode.
Single-range mode
t Select the individual input fields using the “Q" or “q" soft key. t For numeric input: use the 0 ... 9 keys and the . key (decimal point). Make
corrections using c.
t Confirm using the “l" soft key.
If other parameters follow the one just entered, the highlight bar is automatically positioned on the next input field. To cancel numbers entered: press the “
ESC" soft key.
t Use the “o" soft key to go to the next menu level.
y This will apply all parameters.
t Press M or “oo" to exit the Setup menu.
In the example shown here, a single-range scale in “Verifiable” configuration with a maximum capacity of 6.000 kg is modified; the verification scale interval e is changed from 0.001 kg to 0.002 kg, in accordance with the maximum permitted value of 3000 verification scale intervals. Press the “l" soft key to confirm the changed value. The highlight bar is automatically positioned on the “
Min. cap." field.
The following values apply for the minimum load for verifiable scales:
For class l: Min. cap. = 20 e – For class m: Min. cap. = 10 e
A verification scale interval that is changed, therefore, also affects the minimum load. Changing the verification scale interval “e" is automatically applied to the “Min. cap." You can also change this value manually:
In the example, the minimum capacity must be increased to 0.04 kg for class l.
t Press the following keys in sequence: 0.040and confirm using the “l."
y The highlight bar is automatically positioned on the “Max. cap." field.
t The value for the maximum capacity is not changed. For this example, the input of
parameters for single-range mode in the “Verifiable” configuration is now concluded.
t Use the “o" soft key to go to the next menu level.
24
Configuring Weighing Platforms
0.002
25
Configuring Weighing Platforms
Multi-interval mode
The illustration shows an example of the input menu opened for multi-interval mode range configuration. This example shows the parameters for a scale in “Verifiable”
configuration, with 2 weighing ranges and a maximum capacity of 6.000 kg. – Range 1: 0…3.000 kg with e1 = 0.001 kg – Range 2: 3.002…6.000 kg with e2 = 0.002 kg
t Enter the verification scale interval for Range 1 in the “E" input field. The minimum
capacity for a class l scale must be set to 0.02 kg.
t Use the “o" soft key to go to the next menu level.
y The active range configuration is marked with “o."
t Use the “o" soft key to go back to the “Verifiable" menu.
Selecting units
t Use the “q" and “O" soft key to open the “Available units" menu item.
This menu lets you enable or disable the weight units available under “
Weight
unit 1," “Weight unit 2" and “Weight unit 3."
t Select the respective unit using the “Q" or “q" soft key and confirm using the “l"
soft key.
In most cases, you will not need to change defined values.
Available weight units are marked by a *.
The weight unit used for configuration of weighing ranges cannot be blocked.
t Use the “o" soft key to go back to the “Internal" menu.
t Use the “q" and “O" soft key to open the “Calibration/adjustment"
menu.
t Use the “q" and “O" soft key to open the “Calibration/adjustment
unit" menu item to define the weight unit for calibration and adjustment. In most
cases, you will not need to change defined values.
y All units are displayed in the menu that are activated in “Available units."
The current setting is marked by a “
o."
t To change the calibration/adjustment unit, select the unit using the “Q" or “q" soft key
and confirm using the “l" soft key.
t Use the “o" soft key to go back to the “Internal" menu. t Use the “Q" soft key to select the “ADC configuration" menu item. t Use the “O" and “q" soft key to select the “Save configuration data"
menu item.
t To save the configuration, use the “Q" soft key to select “Yes" and confirm using the
l" soft key.
y The message “Function activated" appears briefly. The program then
returns automatically to the regular weighing mode. To not save the configuration:
t Press the “o" soft key to exit the menu.
y The program returns to the next higher menu level.
To not save data: Press the “
o" soft key. The program returns to the
WP1:Internal.
t Slide the menu access switch to the right (= “closed" position) and reattach the cap.
y The device is now in normal weighing mode.
h
Once ADC configuration has been completed, an adjustment of the weighing platform (calibration/adjustment and linearization) must be carried out (see “Calibration/Adjustment without Weights" and “External Linearization").
The displays depicted in the next two illustrations on the left show data from a multi­interval scale configured as described above, or a similarly configured multiple-range scale.
If the A/D converter was configured with a “Verifiable” data record, the lines for display of metrological data (lines 1 and 2) show the data valid for use in legal metrology.
The current range (e.g. R1) is displayed top left under the weighing point for multiple­range scales.
ADC Configuration with Load Cell(s) Connected
Procedure:
1. Open the menu access switch, see “Analog/Digital Converter (ADC) Configuration."
2. Activate the Service mode, see “Service Mode."
3. Configure WP 1, see “Analog/Digital Converter (ADC) Configuration."
4. Set single-range mode, for example, see “Analog/Digital Converter (ADC) Configuration."
5. Select the units, see “Analog/Digital Converter (ADC) Configuration."
6. Adjust without weights, see “Adjust without weights."
7. Set/Delete the preload, see “Setting the Preload" and “Deleting the Preload."
26
Configuring Weighing Platforms
27
Configuring Weighing Platforms

Entering Geographical Data for Use in Legal Metrology

Purpose
Entering geographical data allows the external adjustment of weighing equipment at a place (e.g. at the manufacturer or vendor's place of business) that is not the same as the place of installation. If the weighing equipment is adjusted at the place of installation, it is not necessary to enter geographical data. The sensitivity of weighing equipment changes depending on the place of installation as it is dependent on the on-site gravitational force – or, more precisely, on gravitational acceleration. Saving geographical data makes it possible to change the place of installation of the weighing equipment after external adjustment has been carried out. The adjustment of weighing equipment is valid at the place of installation and within a specific tolerance zone. At 3000 e this zone extends ±100 km from the set geographical latitude and ±200 m from the set elevation above sea level.
Installation Location in Germany
An exception to this is the setting for “Germany (Zone D):” If during external adjustment of weighing equipment within Germany the geographical data
Geographical latitude: 51.00 degrees – 513 m elevation above sea level
are entered, the weighing equipment can be used throughout Germany. Gravitational acceleration for “Germany (Zone D)” is 9.810 m/s
2
. On delivery the geographical data for
“Germany (Zone D)” are entered in the output device. It is recommended to use the geographical data settings for “Germany (Zone D)” when
adjusting and delivering the weighing equipment within Germany. Entering exact geographical data will lead to a higher level of accuracy but will also restrict
the tolerance zone.
Setup information
It is only possible to enter geographical data when the menu access switch is open. – When the Service mode is active, geographical data can be entered in the Setup menu under
WP 1" for the first weighing platform and “WP 2" for the second. Settings are made in
the “
Calibration/adjustment: Geographical data: Input
parameters" menu.
You can enter either the “
latitude" (geographical latitude in degrees) and
altitude" (elevation in m above sea level) or the value for gravitational acceleration
gravity"
Gravitational acceleration takes precedence over the geographical latitude and elevation of the location: If it has been entered, input fields for latitude and elevation show the values “
99999.99" and “9999999" respectively. If only elevation and latitude have been
entered, “
0000000" is displayed for gravitational acceleration.
h
Then make sure that the geographical data for the adjustment location has been entered correctly. If no external adjustment is carried out, enter the data for the installation location. The data can be obtained from the relevant land registry or Ordnance Survey.
Procedure
t Remove the cap. t Slide the menu access switch to the left (= “open" position).
If the device is part of a verified weighing facility, this will only be possible if the verification seal is broken. The weighing equipment must then be verified again.
t Activate the Service mode, see “Service Mode." t Select weighing platform “WP 1" in the “Device Parameters" menu item. t If the “Internal" setting is not already activated (marked by “o"), select the setting
using the “
Q" or “q" soft key and confirm with “O."
y The menu for the “WP-1 INTERNAL" device parameters is displayed.
t Use the “Q" or “q" soft key to select and open the
Calibration/adjustment" menu using “O."
t Use the “Q" or “q" soft key to select and use “O" to open the “Geographical
data" menu.
t Use the “O" soft key to confirm “Input parameters."
Entering Geographical Latitude and Elevation
t Use the “Q" or “q" soft key to select the corresponding input field. t Enter the number via the keypad and confirm using the “l" soft key.
y The next input field is selected.
In this example, the geographical data are entered for the respective platform as a value pair “Latitude" and “Altitude." After this data was saved and the scale returned to weighing mode, this pair of values is displayed again the next time the display menu is opened. The input field for gravitational acceleration is empty.
In this example, the value for gravitational acceleration is entered for the place of installation. The fields “
Latitude" and “Altitude" are invalid. The set value is
then re-displayed after it is saved and the input menu is re-opened. If you exit the Setup menu and then open the Service mode, the set value for gravity is no longer displayed.
28
Configuring Weighing Platforms
29
Configuring Weighing Platforms
Entering Gravity
t Use the “Q" or “q" soft key to select the corresponding input field. t Enter gravity in m/s
2
via the keypad and confirm using the “l" soft key.
Permissible value range:
9.700000 d < gravity 2 d < 9.900000
In the example shown here, the value for gravity has been changed.
The new value,
9.810000 m/s2 applies to the setting “Germany (Zone D).”
t Press the “o" soft key to exit the Input menu. t Use the “Q" soft key to select the “Save parameters" menu item.
t Use the “Q" soft key to select “Yes" and use the “l" soft key to confirm.
y The message “Data stored" appears briefly.
The program then returns to the “
No" display status.
t Press M or “oo" to exit the Setup menu. t Slide the menu access switch to the right (= “closed" position) and reattach the cap.
y The display goes out and the device restarts. Then weighing mode is active.
Adjusting Scales in Operating Mode
See also “Calibration and Adjustment" in the chapter called “Operation"
t Open the Device parameters menu for the respective weighing platform,
(e.g. “
WP 1:INTERNAL").
Open the “
Calibration/adjustment" submenu.
–"
CAL key function" menu item:
Setting “
Ext. cal./adj.: factory-def. wt."
(factory setting).
–"
Cal/adj. sequence" menu item:
Setting “
Cal. then manual adj." (factory setting).
–"
Activate ext. adj." menu item
(not for a verifiable configuration): Setting “
Activated" (factory setting).
To display geographical data in the Device parameters menu, open the “
Operating parameters" submenu.
Menu item “
Display geogr. data > On".
t Press ( to unload the scale.
t Press J to start an external adjustment.
y The display “C.EXT.D" appears briefly.
In the example, the altitude and latitude of the installation location are being entered.
y The display “Altitud" appears briefly.
y The altitude at the place of installation is displayed in meters above sea level - here, the
altitude for “Germany (Zone D).”
t Press J to confirm the displayed value or press ( to cancel the adjustment.
y The display “Latitud" appears briefly.
y The geographical latitude of the place of installation is shown in degrees north or
degrees south - here the latitude setting for “Germany (Zone D).”
t Press J to confirm the displayed value or press ( to cancel the adjustment.
y You are prompted to place the required weight on the platform (e.g.: 5.0 kg).
The subsequent steps for completing the calibration/adjustment are described in the chapter entitled “Operation” under “Calibration and Adjustment.”
30
Configuring Weighing Platforms
C.EXT.D
C.EXT.DEF.
Altitud
C.EXT.DEF.
Latitud
C.EXT.DEF.
31
Configuring Weighing Platforms
If gravity is being entered instead of altitude and latitude, then “Gravity" is displayed
for a brief time after “
CAL."
The entered value appears in m/s
2
, here for the “Germany (Zone D)" setting.
t Press J to confirm the displayed value or press ( to cancel the adjustment.
y You are prompted to place the required weight on the platform (e.g.: 5.0 kg).
The subsequent steps for completing the calibration/adjustment are described in the
chapter entitled “Operation” under “Calibration and Adjustment."
t Slide the menu access switch to the right (= “closed" position) and reattach the cap.
y The display goes out and the device restarts. Then weighing mode is active.
If adjustment is carried out using a verifiable configuration data record, the lines
for display of metrological data (lines 1 and 2) show the data valid for use in legal
metrology, if the menu access switch is closed. See also the chapter “Operation",
“Configuration for Use in Legal Metrology."

Entering Adjustment and Linearization Weights

Purpose
Entering adjustment and linearization weights.
Procedure
See also “Calibration and Adjustment" in the chapter called “Operation".
t Remove the cap. t Slide the menu access switch to the left (= “open" position). t Activate the Service mode, see “Service Mode." t Select weighing platform “WP 1" in the “Device Parameters" menu item. t If the “Internal" setting is not already activated (marked by “o"), select the setting
using the “
Q" or “q" soft key and confirm with “O."
y The menu for the “WP-1 INTERNAL" device parameters is displayed.
t Use the “Q" or “q" soft key to select and open the
Calibration/adjustment" menu using “O."
t Use the “Q" or “q" soft key to select and use “O" to open the “Parameter for
external weight" menu.
t The first menu item “Cal/adj. wt.:" (for selecting the user-defined
calibration weight), is also accessible without activating the Service mode. The values for the linearization weights “
Lin.-wt. 1" to “Lin.-wt. 4" can, however,
only be changed in the Service mode.
y The current values for the user-defined calibration weight and the 4 linearization weights
are displayed. In this example, the value for the external, user-defined adjustment weight is changed to
6.000 kg.
t Press 6.000 and confirm with the “l" soft key.
y The “Lin.-wt. 1" input field is selected.
In this example, the value for linearization weight 1 is changed to 1.500 kg.
t Press 1.500 and confirm with the “l" soft key.
y The “Lin.-wt. 2" input field is selected.
t Enter or change all linearization weights in sequence as needed.
If you do not require all linearization positions, enter “
0.000" in the unused fields to
hide these lines in the display. Confirm with the “
l" soft key after each entry to move to
the next input field. In the example shown here, four linearization weights have been entered (1.5 kg, 3.0 kg,
4.5 kg and 6.0 kg). When you close the input menu by pressing the “
o" soft key, the input values are
directly applied.
t Slide the menu access switch to the right (= “closed" position) and reattach the cap.
32
Configuring Weighing Platforms
33
Configuring Weighing Platforms

Function Allocation of the allocation for the J Key for Calibration/Adjustment

Purpose
The J key is used for the calibration/adjustment function. Key settings can be changed when the Service mode is activated:
Procedure
t Remove the cap. t Slide the menu access switch to the left (= “open" position). t Activate the Service mode, see “Service Mode." t Select weighing platform “WP 1" in the “Device Parameters" menu item. t If the “Internal" setting is not already activated (marked by “o"), select the setting
using the “
Q" or “q" soft key and confirm with “O."
y The menu for the “WP-1 INTERNAL" device parameters is displayed.
t Use the “Q" or “q" soft key to select and open the
Calibration/adjustment" menu using “O."
t Use the “O" soft key to open the “Cal/adj. sequence" menu.
t Use the “Q" or “q" soft key to select the “Cal. then manual adj." menu
(factory setting) and confirm with the “
l" soft key.
t Use the “o" soft key to go to the next menu level.
t Use the “O" soft key to open the “Activate ext. adj." menu. t Use the “Q" or “q" soft key to select the “Activated" menu item (factory setting).
Not for a verifiable configuration.
t Geographical data is not displayed during calibration/adjustment (factory setting).
To display geographical data in the Device parameters menu, open the “
Operating
parameters" submenu.
Menu item “
Display geogr. data > On".
t Use the “O" soft key to open the CAL key function" menu.
y The “CAL key function" submenu is displayed.
t Use the “Q" or “q" soft key to select the corresponding menu item and confirm with
l."
y The menu item is marked by a circle “o."
Note
The functions that can be configured in the “
CAL key function" submenu
depends on the selected weighing platform and its configuration data. Functions that cannot be accessed are not displayed in the selection screen.
Important Note! The function set in the “
CAL key function" menu is carried out in the
normal weighing mode because when you exit the Setup menu the Service mode is deactivated. To perform the function on a digital weighing platform (such as an IS platform), however, it must be carried out in Service mode.
t The procedure is as follows: after selecting the desired function and exiting the Setup
menu, reactivate Service mode again and then exit the Setup menu immediately by pressing the M key or the “
oo" soft key.
y The scale is now in Service mode without this being displayed.
t Trigger the previously set function using the J key.
y The display shows “S_CAl:" indicating that the scale is in Service mode.
t The corresponding menu item can be selected using the “Select" soft key and
carried out using the J key.
t If you cancel the function using the ( key or by restarting the scale using the e key,
you will exit the Service mode.
External Calibration/Adjustment with Factory-Defined Weight (Default Weight)
Configuration
If not already selected (factory setting, marked by “
o" when active), select the “CAL
key function:Ext. cal./adj.; factory-def. wt.
"
menu item (external adjustment with a factory-defined, standard weight).
t Use the “Q" or “q" soft key to select this menu item and confirm with “l."
y The menu item is marked by a circle “o."
Note:
The menu items “
Ext. cal./adj.; factory-def. wt." (external
adjustment with a factory-defined, standard weight), “
Ext. cal./adj.;
user-defined wt." (external adjustment with a user-defined weight) and
Key blocked" can also be accessed without activating the Service mode.
t Press e to turn off the device. t Press e turn the device back on.
y The Sartorius logo is displayed briefly, after which the device is in normal weighing
mode.
t Press ( to unload and zero the scale. t Press J to start the external adjustment.
y The display “C.EXT.D" appears briefly.
Note
If the display of geographical data (elevation and latitude or gravity) is activated, this data is displayed and confirmed each with a press of the J key (you can cancel the calibration/adjustment process using the ( key). See also “Entering Geographical Data for Use in Legal Metrology" in this chapter.
34
Configuring Weighing Platforms
C.EXT.D
C.EXT.DEF.
12
C.EXT.D
C.EXT.DEF.
35
Configuring Weighing Platforms
Procedure
y The target value of the required adjustment weight (5.000 kg in the example) is displayed as
a negative value.
t Place the required adjustment weight on the platform.
Note
If the calibration/adjustment sequence is set to automatic ("
Calibration/adjustment:Cal./adj. sequence:Cal.,
then auto adjust" menu, see “Function Allocation of the allocation of the J
Key for Calibration/Adjustment") and the adjustment weight consists of several pieces, then they should be placed on the platform in series at short intervals.
When the weighing instrument has stabilized, the weight on the scale is accepted as the calibration weight after a predefined interval, and the weighing instrument is calibrated/adjusted with this weight.
The difference since the most recent span adjustment is not displayed; this value is output only on GMP-compliant printouts.
y After a brief pause, the difference since the last span adjustment is displayed (calibration).
Note
This display only appears for the “
Cal. then manual adj." setting (see
previous note). If “
Cal. then auto adj." is active, the calibration/adjustment
procedure cannot be cancelled. To stop the procedure after calibration and before adjustment takes place, press the ( key
(only if “
Cal. then manual adj." is active).
t Press J to adjust the scale (only if “Cal. then manual adj." is active).
y At the conclusion of the calibration procedure, the calibration weight is displayed as a
positive value.
When calibration/adjustment is complete, the GMP-compliant printout shown on the left is generated. If the adjustment procedure is canceled (only calibration is performed), the last two lines, “External calibration” and “Diff. + 0.000 kg” are not printed.
t Unload the scale. t Press e to turn off the device. t Press e turn the device back on.
y The Sartorius logo is displayed briefly, after which the device is in normal weighing mode.
3
If a serious operator error should occur during calibration (for example, if the menu setting “
Cal. then auto adj." is active and the wrong calibration weight is placed
on the scale), the scale might completely fail to stabilize, which means it cannot show a zero point. In this case, select the “
Adjust without weights" menu and set the mean
sensitivity of the strain-gauge weighing beam to 2.0 mV/V. Then re-adjust the scale. See also “Adjust without weights."
The zero point is only displayed for a verified scale with d=e.
--------------------
14.01.2010 13:50 Type CAW3P1-6DC-LCE Ser.no. 12345678 Vers.no. 1.02.101110 BVers. 01-63-02
-------------------­External calibration Nom. + 5.000 kg Diff. + 0.010 kg External adjustment Diff. + 0.000 kg
--------------------
14.01.2010 13:52 Name:

External Calibration/Adjustment with a User-Defined Weight

Configuration
Select the “
CAL key function:Ext. cal./adj.; user-def.
wt." menu item (external adjustment using a user-defined weight).
Note:
The menu items “
Ext. cal./adj.; factory-def. wt." (external
adjustment with a factory-defined, standard weight), “
Ext. cal./adj.;
user-defined wt." (external adjustment with a user-defined weight) and
Key blocked" can also be accessed without activating the Service mode.
t Use the “Q" or “q" soft key to select this menu item and confirm with “l."
y The menu item is marked as set by a circle “o."
t Enter the target value of the adjustment weight in the
Calibration/adjustment" menu under “Parameter for
external weight" in the “Cal/adj. wt" input field.
t Press e to turn off the device. t Press e turn the device back on.
y The Sartorius logo is displayed briefly, after which the device is in normal weighing
mode.
t Press ( to unload and zero the scale. t Press J to start the external adjustment.
y The display “C.EXT.U" appears briefly.
Note
If the display of geographical data (elevation and latitude or gravity) is activated, this data is displayed and confirmed each with a press of the J key (you can cancel the calibration/adjustment process using the ( key). See also “Entering Geographical Data for Use in Legal Metrology" in this chapter.
Procedure
t Press J to confirm the displayed value or press ( to cancel the adjustment.
y The target value of the required adjustment weight (6.000 kg in the example) is displayed
as a negative value.
t Place the required adjustment weight on the platform.
If the calibration/adjustment sequence is set to “Cal. then auto adjust”, refer to the note under “External Calibration/Adjustment with Factory-Defined Weight (Default Weight).”
y After a brief pause, the difference since the last span adjustment is displayed
(calibration).
t If you only want to perform a calibration, press the ( key to cancel the
calibration/adjustment procedure.
t Press J to adjust the scale.
y At the conclusion of the calibration procedure, the calibration weight is displayed as
a positive value.
36
Configuring Weighing Platforms
C.EXT.U
C.EXT.USE.
37
Configuring Weighing Platforms
When calibration/adjustment is complete, the GMP-compliant printout shown on the left is generated. If the adjustment procedure is canceled (only calibration is performed), the last two lines, “External calibration” and “Diff. + 0.000 kg” are not printed.
t Unload the scale. t Press e to turn off the device. t Press e turn the device back on.
y The Sartorius logo is displayed briefly, after which the device is in normal weighing mode.
3
If a serious operator error should occur during calibration (for example, if the menu setting “
Cal. then auto adj." is active and the wrong calibration weight is placed on
the scale), the scale might completely fail to stabilize, which means it cannot show a zero point. In this case, select the “
Adjust without weights" menu and set the mean
sensitivity of the strain-gauge weighing beam to 2.0 mV/V. Then re-adjust the scale. See also “Adjust without weights."

Internal Calibration/Adjustment

This function is available only if a digital weighing platform (for example, an IS platform) is connected to WP 1, either as a second weighing platform or as the only weighing platform without using the built-in A/D converter WP 1. The connection can be made both via the COM1, COM2 or UNICOM interface with a corresponding configuration as well as via the expansion PCBs for WP 1 or WP 2.
This function is also accessible without activating the Service mode.
--------------------
14.01.2010 13:50 Type CAW3P1-6DC-LCE Ser.no. 12345678 Vers.no. 1.02.101110 BVers. 01-63-02
-------------------­External calibration Nom. + 6.000 kg Diff. + 0.010 kg External adjustment Diff. + 0.000 kg
--------------------
14.01.2010 13:52 Name:

Adjustment Without Weights

Purpose
In the Service menu, adjustment without weights can be carried out by entering the characteristic data of the load cells.
3
Adjustment without weights may not be carried out on weighing equipment used in legal metrology.
Setup information
Adjustment without weights is only possible when the menu access switch is open in the
Service mode. – The “Nominal load" parameter must be entered in the “kg" unit. – The “Resolution" parameter must be entered in the “kg" unit and must correspond to the
scale interval “d" entered for the ADC configuration. – The “Sensitivity" parameter is entered in “mV/V" (see the data sheet for the value).
h
These values are converted to internal quantities.
Once the ADC configuration data have been stored (by selecting the “
Save
parameters" menu item), these parameters can no longer be read.
Procedure
t Remove the cap. t Slide the menu access switch to the left (= “open" position). t Activate the Service mode, see “Service Mode." t Select weighing platform “WP 1" in the “Device Parameters" menu item. t If the “Internal" setting is not already activated (marked by “o"), select the setting
using the “
Q" or “q" soft key and confirm with “O."
y The menu for the “WP-1 INTERNAL" device parameters is displayed.
t Use the “Q" or “q" soft key to select and open the
Calibration/adjustment" menu using “O."
t Use the “Q" or “q" soft key to select and use “O" to open the “Adjust without
weights
" menu.
t Use the “O" soft key to open the “Input parameters" menu.
38
Configuring Weighing Platforms
39
Configuring Weighing Platforms
y The Input menu is displayed.
Enter the nominal capacity and resolution of the load cells in kg and the sensitivity of the load cells in mV/V in the corresponding input fields. The maximum capacity is usually less than the value to be entered in the “
Adjust without weights" menu for the
nominal capacity of the load cell, as load cells carry additional weight (e.g., a weighing platform).
If the weighing platform has multiple load cells, multiply the nominal capacity accordingly. Example: The weighing platform has 4 load cells each at 50 kg.
The nominal capacity is 4 x 50 kg = 200 kg. In the example shown here, the weighing platform consists of one load cell with a maximum
capacity of 10 kg.
t Press 10.00 and confirm with the “l" soft key.
y The “Resolution" input field is selected.
The smallest digit “d" is entered in this field in “kg". The value must correspond to the “
D"
entry in “
ADC configuration: Standard: Ranges: Single-
range mode."
In the example, this should be d = 0.002 kg.
t Press 0.002 and confirm with the “l" soft key.
y The “Sensitivity 1" input field is selected.
If the weighing platform has multiple load cells, enter the sensitivity either
under “
Sensitivity 1"… “Sensitivity 4" as an individual value or
under “
Sensitivity 1" as an average value.
If an average value is entered for all cells, or if fewer than 4 load cells are present, enter “0" in the remaining fields. Range of permitted values: 0.01…5 mV/V.
t Press 1.944, e.g. to enter a sensitivity of 1.944 mV/V. t Confirm using the “l" soft key.
y The “Sensitivity 2" input field is selected.
t Either enter a value or set all remaining input fields to “0." t Confirm each using the “l" soft key.
t Press the “o" soft key to exit the menu.
t Use the “q" soft key to select the “Save parameters" submenu.
t To save the configuration, select “Yes" and confirm using the l soft key.
In the first line of the display, the message “
Data stored" is shown briefly.
Then the program then returns to the “No" display status.
t Use the “o" to return to the next menu level. t Press M or “oo" to exit the Setup menu. t Slide the menu access switch to the right (= “closed" position) and reattach the cap.
y The display goes out and the device restarts. Then weighing mode is active.
Function Allocation of the J Key for Linearization and
Setting/Deleting the Preload
Purpose
The J key is normally used for the calibration/adjustment function. The following
additional functions can be allocated to the key when the Service mode is activated: – External linearization with default weights – External linearization with entered linearization weights – Internal linearization (for external IS platforms only) – Set preload – Delete preload
3
Once linearization has been completed, or after a preload has been set or deleted the
function of the J key must be reallocated back to its original function in the Setup menu,
e.g. external calibration/adjustment with default weights.
Procedure
t Remove the cap. t Slide the menu access switch to the left (= “open" position). t Activate the Service mode, see “Service Mode." t Select weighing platform “WP 1" in the “Device Parameters" menu item. t If the “Internal" setting is not already activated (marked by “o"), select the setting
using the “
Q" or “q" soft key and confirm with “O."
y The menu for the “WP-1 INTERNAL" device parameters is displayed.
t Use the “Q" or “q" soft key to select and open the “Calibration/
adjustment" menu using “O."
y The “CAL key function" submenu is displayed.
t Use the “Q" or “q" soft key to select the corresponding menu item and confirm with “l."
y The menu item is marked by a circle “o."
Note
The functions that can be configured in the “
CAL key function" submenu
depends on the selected weighing platform and its configuration data.
Functions that cannot be accessed are not displayed in the selection screen.
Note
The function set in the “
CAL key function" menu is carried out in the normal
weighing mode because when you exit the Setup menu the Service mode is deactivated.
To perform the function on a digital weighing platform (such as an IS platform), however,
it must be carried out in Service mode.
t The procedure is as follows: after selecting the desired function and exiting the Setup menu,
reactivate Service mode again and then exit the Setup menu immediately by pressing the
M key or the “
oo" soft key.
y The scale is now in Service mode without this being displayed.
t Trigger the previously set function using the J key.
y The display shows “S_CAl:" indicating that the scale is in Service mode.
t The corresponding menu item can be selected using the “Select" soft key and carried
out using the J key.
t If you cancel the function using the ( key or by restarting the scale using the e key,
you will exit the Service mode.
40
Configuring Weighing Platforms
41
Configuring Weighing Platforms

External Linearization with the Factory-Defined Weight (Default Weights)

Setup information
– This function is accessible only if the software and the functionality of the connected
weighing platform permit this operation.
External linearization when weighing in legal metrology is only possible when the menu
access switch is open.
– The J key must be assigned the “External linearization" function, see “Function Allocation
of the J Key for Linearization and Setting/Deleting the Preload."
Activating the display of geographical data has no effect on this function.
3
Once linearization has been completed, the J key must be reallocated back to its original function in the Setup menu, e.g. external calibration/adjustment with default weights.
Procedure
t For scales used in legal metrology, slide the menu access switch to the left
(= “open" position).
y The display goes out and the device restarts. Then weighing mode is active.
t Press ( to unload and zero the scale. t Press J to start the external linearization.
y The display “L.EXT.D" appears briefly.
Additional steps are explained in “External Linearization with User-Defined Weights."
External Linearization with User-Defined Weights
Setup information
– This function is accessible only if the software and the functionality of the connected
weighing platform permit this operation. – External linearization when weighing in legal metrology is only possible when the menu
access switch is open.
– The J key must be assigned the “External linearization" function, see “Function Allocation
of the J Key for Linearization and Setting/Deleting the Preload." –
Activating the display of geographical data has no effect on this function.
3
Once linearization has been completed, the J key must be reallocated back to its original
function in the Setup menu, e.g. external calibration/adjustment with default weights.
Configuration
t Set the linearization weights, see “Entering Adjustment and Linearization Weights."
Procedure
t For scales used in legal metrology, slide the menu access switch to the left
(= “open" position).
y The display goes out and the device restarts. Then weighing mode is active.
t Press ( to unload and zero the scale. t Press J to start the external linearization.
y The display “L.EXT.U" appears briefly.
y After approx. 2 seconds, the target value for linearization weight 1 is shown as a negative
value on the display (in example shown here, 1.500 kg) .
t Place the required linearization weight 1 on the platform. t Press J to apply linearization weight 1 or press ( to cancel the linearization function.
y After a short time the difference between the measured value and the true weight of the
sample will be displayed.
42
Configuring Weighing Platforms
43
Configuring Weighing Platforms
t Press J to apply linearization weight 1 or press ( to cancel the linearization function.
y After the linearization weight 1 has been saved you will be prompted to place the second
linearization weight on the weighing pan.
t Repeat the procedure for all required linearization weights.
y After the last linearization weight has been saved you will be prompted to remove any load
from the weighing pan.
t Remove all linearization weights from the weighing platform.
y The zero point is applied automatically after a brief time. The indicator automatically
switches to weighing operation. When linearization is complete, the GMP-compliant printout shown on the left is generated.
t Slide the menu access switch to the right (= “closed" position).
--------------------
14.01.2010 13:00 Type CAW3P1-6DC-LCE Ser.no. 12345678 Vers.no. 1.02.101110 BVers. 01-63-02
-------------------­Linearization Wt.1 + 1.500 kg Wt.2 + 3.000 kg Wt.3 + 4.000 kg Wt.4 + 6.000 kg completed
--------------------
14.01.2010 13:02 Name:

Set preload

Setup information
Setting the preload when weighing in legal metrology is only possible when the menu access
switch is open.
– The J key must be assigned the “Set preload" function, see “Function Allocation of the
J Key for Linearization and Setting/Deleting the Preload." –
Activating the display of geographical data has no effect on this function.
3
Once the preload has been set, the J key must be reallocated back to its original function
in the Setup menu, e.g. external calibration/adjustment with default weights.
Procedure
t For scales used in legal metrology, slide the menu access switch to the left (= “open"
position).
t Press e turn the device back on.
y The Sartorius logo is displayed briefly, after which the device is in normal weighing mode.
t Press ( to unload and zero the scale.
y Display after the scale has been zeroed.
t Place the preload weight on the weighing platform. t Press J to start “Set preload."
y The display “SET PREL" appears briefly.
After a short period of time the preload will be applied and the indicator will automatically
switch back to weighing mode.
After the “Set Preload” operation has been completed, the scale is zeroed.
When the “Set preload" function is complete, the GMP-compliant printout shown on the left
is generated.
t Slide the menu access switch to the right (= “closed" position).
44
Configuring Weighing Platforms
--------------------
14.01.2010 13:50 Type CAW3P1-6ED-LCE Ser.no. 12345678 Vers.no. 1.02.101110 BVers. 01-63-02
-------------------­Set preload completed
--------------------
14.01.2010 13:52 Name:
45
Configuring Weighing Platforms

Delete preload

Setup information
Deleting the preload when weighing in legal metrology is only possible when the menu
access switch is open.
– The J key must be assigned the “Delete preload" function, see “Function Allocation of
the J Key for Linearization and Setting/Deleting the Preload."
Activating the display of geographical data has no effect on this function.
3
Once the preload has been deleted, the J key must be reallocated back to its original function in the Setup menu, e.g. external calibration/adjustment with default weights.
Procedure
t For scales used in legal metrology, slide the menu access switch to the left
(= “open" position).
t Press e turn the device back on.
y The Sartorius logo is displayed briefly, after which the device is in normal weighing mode.
t Remove the preload weight from the weighing platform.
y The display shows the removed preload weight as a negative value.
t Press J to start “Delete preload."
y The display “CLR PREL" appears briefly.
After a short period of time the preload will be deleted and the indicator will automatically switch back to weighing mode.
After the “Delete Preload” operation has been completed, the scale is zeroed.
When deleting the preload is complete, the GMP-compliant printout shown on the left is generated.
t Slide the menu access switch to the right (= “closed" position).
--------------------
14.01.2010 13:50 Type CAW3P1-6DC-LCE Ser.no. 12345678 Vers.no. 1.02.101110 BVers. 01-63-02
-------------------­Delete preload completed
--------------------
14.01.2010 13:52 Name:
You can use the Combics 3 to record weight values from 1 to 3 weighing platforms, calculate and display weight values through application programs, and assign IDs to the samples weighed.
Configure the indicator first, using the Setup menu for the desired application program (printer settings, etc.). Then you can begin weighing.
When a key is pressed that does not have an active operating mode function, an acoustical signal (double beep) sounds and the message “——-—" is displayed for 2 seconds. The display then returns to the previous screen content.

Operating Design

Weighing Mode Operating Design

Labeled Keys
Some keys have a second function, activated by pressing and holding the key for over 2 seconds. Whether a function is available depends on the indicator operating state and operating menu settings.
e On/Off key
y standby is displayed in Standby mode.
n If a second weighing platforms is connected, this key toggles the display between the
two readouts.
( – Press briefly: Zero
– Press briefly: cancels calibration/adjustment – Press longer than 2 seconds: displays the adjustment/configuration counter
) Tare the scale: press briefly.
k Displays the second weight unit or SQmin (depending on the settings, see “Operation,"
“SQmin Function").
J Starts calibration or adjustment.
p – For printing:
Press briefly.
– Prints GMP footer:
Press longer than 2 seconds.
K Toggles unit between normal and 10-fold higher resolution.
L Toggles the display between
gross value (net value plus tare) and net value (gross value minus tare).
c – Zum Beenden der Programme oder Löschen einzelner Zeichen:
Taste kürzer als 2 Sek. drücken.
– Zum Löschen der Eingaben:
Taste länger als 2 Sek. drücken.
D Toggles to the Info mode:
Press longer than 2 seconds.
R Product data memory:
Saves initialization and user data (product and tare values). The product data memory can store over 400 product and tare values.
M Opens/Exits the Setup program
0, 1, 2 … 9, .
Enters numbers, letters and other characters.
a Toggles between numeric and alphabetic input.
46
47
Soft keys
The functions of active soft keys are indicated by symbols and abbreviations in the last line on the display.
Abbreviation examples:
ID: ID list DELETE: Delete entry
Symbols used for soft key functions:
oo: Return to initial state o: Next higher menu level O: Show items under selected entry Q: Scroll up in the input/output window q: Scroll down in the input/output window l: Set the selected menu parameter
Numeric Input via the Keypad
t To enter numbers (one digit at a time):
Press 0, 1, 2 … 9
t To save input:
press the corresponding key. For example, press ) to save manual tare input.
Text Input via the Keypad
t Press the a key.
y “ABC" is displayed.
t For example, press the 4 key.
y The corresponding letter is displayed. The flashing cursor marks the first letter.
t Press the key as many times needed to select the desired letter. t Press the F1 soft key (l) or wait 2 seconds.
y The selected letter appears on the display.
Entering Spaces via the Keypad
t Press the a key.
y “ABC" is displayed.
t Press the 0 key.
y The corresponding selection is displayed. The flashing cursor marks the space.
t Press the F1 soft key (l) or wait 2 seconds.
y The space appears on the display.
Operating Design
ID
F5 F4 F3 F2 F1
Operating Design
Special Character Input via the Keypad
t Press the a key.
y “ABC" is displayed.
t Press the 1 key.
y The corresponding character selection is displayed. The flashing cursor marks the first
character.
t Press the F1 soft key (l) or wait 2 seconds.
y The special character appears on the display.
Deleting characters
t Press c.
Deleting entire input string
t Press the F4 soft key (ESC).
Saving Input
t Enter a value.
t Press the corresponding soft key (e.g. 1.ID to define an identifier).

Saving Settings in Weighing Mode

You can specify the save type used in the Application Parameters Setup menu.
By default, all application parameters saved (e. g., reference values) remain in memory and are available when
the device has been switched off and then on again
you return to the originally selected application from a second one (e. g., when you
switch from Averaging back to Counting all parameters saved for Counting are available).

Applying the Tare Weight

To save the weight on the weighing platform as a tare weight:
t Place the tare object on the weighing platform.
t Press ).
y The value is applied as the tare value.

Input Through the Digital Control Port

You can connect a remote hand switch or foot switch to the input control line, for use with all application programs. The following functions can be assigned in the Setup menu
Device Parameters:Control I/O Ports:Universal
Switch Key:
Off – Zeroing key
Print key On key
Print key 2 sec CF key
Tare key F1 function key
ISO test key Appl. toggle key
Fn key function 10x higher resolution key
WP toggle key Net/gross value key
Combined zero/
tare function
48
49

Display in Weighing Mode

Weighing Mode: Display of Measured and Calculated Values
The display is divided into several sections.
Lines for metrological data
The following parameters are shown here:
Max Maximum capacity (upper weighing range limit) of the active weighing
platform
Min Minimum capacity (lower weighing range limit) of the active weighing
(verified models only)
e Verification scale interval of the active weighing platform (verified models only) d Smallest display digit of the active weighing platform R1 R2 Display of the current weighing range of the active weighing platform
(with multiple range scale connected)
Symbols and Their Meaning
F The Busy symbol appears when the scale is processing a function activated
by pressing a key.
+ - The plus or minus sign of the weight or other measured value A Verified models only:
Identifies “zero” as a weight value (after the scale or the active weighing platform has been zeroed)
Measured Value Line/Calculated Values
5.23r The current weight value (on verified scales or platforms with e = d, the last
digit is bordered for identification)
20 A calculated value when using an application program, such as Counting or
Weighing in Percent
Unit and Stability
g The current weight unit (e.g. “g") pcs Identifies additional characteristics (e.g. “pieces" for the “Counting"
application)
When the weighing system reaches stability, the weight unit or calculation unit is displayed here.
Data in Tare Memory, Calculated Value, Identification of the Active Weighing Platform when More Than One Platform is Used
a Identification of calculated values (values not used in legal metrology)
B/G NET Indicates gross value or net value (data in tare memory) PT Indicates manual tare input (using a barcode scanner) when viewing tare
information.
WP1 Display of the active weighing platform when 2 platforms are connected.
The display flashes for a ISO Cal adjustment prompt if WP 1 is an IS-weighing platform.
WP When the timer is active (Setup:Device Parameters:
Operating Parameters:Timer), the symbol flashes to indicate
that the set time has halfway expired.
Operating Design
Lines for metrological data
Bar graph
Info/Status line
Line for measured values
Text lines Soft key labels
Plus or minus sign Stability
Application symbol
Tare memory
Calculated values
Battery symbol
Unit
Application, printing and battery symbols
S Printing in progress T GMP-compliant printout active
, Battery status: ‘Battery fully charged’ or ‘Battery empty’
Bar graph
The bar graph shows the percentage of the weighing platform’s capacity that is “used up” by the load on the scale (gross value).
0% Lower load limit 100% Upper load limit
The following symbols indicate tolerance levels for checkweighing:
Bar graph showing 10% intervals
Minimum for “checkweighing"
Target value for “checkweighing"
Maximum for “checkweighing"
Application symbols
A Symbol for the Counting application Y Symbol for “checkweigher" counting down to “zero"
L H W M B A D Symbols for the Totalizing, Checkweighing, Classification,
Net-total Formulation, Weighing in Percent, Counting (with or without reference sample updating) and Neutral Measurement application programs. For details on the symbols, please see the chapters for the respective application programs in the manual.
Text Lines
The text lines show operator support information, such as IDs and user guidance prompts.
Soft key line
This line shows the abbreviations or symbols that indicate soft key functions.

LEDs

LEDs indicate – whether the weight exceeds tolerance limits for checkweighing – the weight value classification for the Classification application

Error Codes

– If a key is inactive, “-------” and/or “No function” is displayed (2 sec.) and
an acoustic signal (double- beep) is emitted – Temporary errors are displayed for 2 seconds in the measured value/result line via an
error code (e .g.,
INF 09); fatal errors (e .g., ERR 101) are displayed continuously
until cleared via a restart.
Error codes are described in detail in “Data Interfaces" in “Error Codes."
Operating Design
50
51

Menu Operating Design

Operating example: Setup:Device parameters:
WP 1:Internal:Adapt filter
o Marks the current menu setting
Configuring parameters:
t Soft key “Q" or “q": Parameter settings t Soft key “l": Confirm parameter t M or “oo": Exit Setup menu

Menu display

Display Mode for Configuration and Information (Setup)
The display is divided into three sections.
Status Line
Indicates the function of the current screen page. In the Setup program, this line shows the “path” to the data displayed.
Input and Output Window
For input and display of detailed information; e.g., for the selected application. Selected items are displayed inversely (white letters on a black background).
Soft key labels
See the description in this section.

Saving Menu Settings

The parameters selected in the menu remain saved when you switch to weighing mode or turn off the device.
You can block access to the Device parameters menu by requiring a password to prevent unauthorized or accidental setting changes (see also “Setting up Password Protection").
Operating Design
Status line
Input/output window
Soft key labels

Configuration

You can configure the indicator by selecting parameters in the Setup menu. These are
divided into the following groups (menu level 1), menu structure see section the “Setup
Overview (Parameters)" section –
Application parameters
Fn key functionDevice parametersInfo (device-specific information) –
Language
When used in legal metrology, not all parameters can be accessed. The indicator only
displays parameters that can be selected.
Factory-set parameters are identified by an “o” in the list starting on the next page.
Setting the Language
You can choose from the following languages for displaying information: –
Deutsch
English (factory setting) –
U.S. mode
FrançaisItalianoEspañol
Example: Selecting “U.S. Mode" for the language
t Press e to turn on the device. t Press M.
y The menu appears on the display.
t Press the “q" soft key several times to select the “Language" line.
t Press the “O" soft key.
y The language selection appears on the display.
t Press the “Q" soft key to select the “U.S. mode" line.
t Press the l soft key to save the setting.
y The setting selection “o" moves to “U.S. mode."
t Press M or oo to exit the Setup menu.
Operating Design
52
53
Operating Design
Setting up Password Protection
t Press e to turn on the device. t Press M.
y The menu appears on the display.
t Press the “q" soft key several times to select the “Device parameters" line.
t Press the “O" soft key.
t Press the “q" soft key several times to select the “Password" line.
y The input line appears on the display.
t Enter numbers and/or letters via the keypad, see “Numeric Input via the Keypad" and
“Letter Input via the Keypad."
t Press the “l" soft key to save the setting. t Press M or “oo" to exit the Setup menu.
Operating Design
54
Setting up Password Protection
t Press e to turn on the device. t Press M.
y The menu appears on the display.
t Press the “q" soft key several times to select the “Device parameters" line.
t Press the “O" soft key.
y The Access window appears on the display.
t Enter the password via the keypad, see “Numeric Input via the Keypad" and “Letter Input
via the Keypad."
t Press the “l" soft key.
y The device selection appears on the display. t Press the “q" soft key several times to select the “Password" line.
y The input line appears on the display.
t Press c several times to delete the password. t Press the “l" soft key to save the delete.
If not yet saved, the process can be canceled using the “
ESC" soft key.
t Press M or “oo" to exit the Setup menu.
55
Operating Design
Printing Parameter Settings
Example: Maximum 20 characters per line.
--------------------
12.01.2010 09:46 Type CAIS3 Ser.no 12345678 Vers. 1.02.101110 BVers. 01-63-02
-------------------­SETUP
DEVICE
-------------------­WP 1
Internal WP 2 off COM1
Data communcations
SBI
Baud rate
1200 Baud
Parity
Odd
Number of Stop bits
1 Stop bit
Handshake mode
Hardware 1-char
Number of data bits
7 Bit
Data output
Printout printer 1
etc.
Operating Design

Setup Overview (Parameters)

o = Factory settings W = User-defined setting
Setup
Application parameters: Please refer to the “Basic Application Programs” manual
Fn key function
o Off
Unit conversion SQMIN (only possible when the display has been turned on in the “Device parameters:SQMin" menu item)
Device parameters
WP 1
RS-232 1)
SBI standard SBI verifiable
o IS-232
ADC-232
RS-485 1)
o IS-485
ADC-485
Internal
Calibration/Adjustment
CAL key function
o Ext. cal./adj.; factory-def. wt.
Ext. cal./adj.; user-defined wt. Key blocked
Cal./adj. sequence
Cal. then auto adjust
o Cal. then manual adjust
isoCAL function 3)
o Off
Adjustment prompt
Activate external adjustment 2)
o Activated
Deactivated
Parameter for external weight
Cal./adj. weight:
Calibration/adjustment unit
Grams /g Kilograms /kg Tons /t Pounds /lb
Adapt filter
Min. vibration
o Normal vibration
Strong vibration Extreme vibration
Application filter
o Final readout
Filling mode Low filtering Without filtering
Stability range
1/4 digit 1/2 digit 1 digit 2)
o 2 digits 2)
4 digits 2) 8 digits 2)
1
) Equipment version: – then blocked internally
2
) Not available on devices verified for use in legal metrology
3
) Only when operated with Sartorius IS weighing platforms or an external ADC
56
57
Device Parameters
WP 1
Internal
Stability delay
Without delay
o Short delay
Average delay Long delay
Taring 1)
Without stability
o After stability
Autozero
o On
Off
Weight unit 1 2)
User-definable / o (factory setting: grams) 1) Grams /g
o Kilograms / kg
Carats / ct 1) Pounds / lb 1) Ounces / oz v) Troy ounces / ozt 1) Hong Kong taels / tlh 1) Singapore taels / tls 1) Taiwan taels / tlt 1) Grains / GN 1) Pennyweights / dwt 1) Milligrams / mg 1) Parts per pound / lb 1) Chinese taels / tlc 1) Mommes / mom 1) Austrian carats / K 1) Tola / tol 1) Baht / bat 1) Mesghal / MS 1) Tons / t lb / oz
Display accuracy 1
o All digits
One less for wt. change Index +1 1) Index +2 1) One less 1)
Zero range
1 percent/max. cap.
o 2 percent/max. cap.
Zero at power on
2 percent/max. cap. 5 percent/max. cap.
o 10 percent/max. cap.
Tare/zero at power on:
o On
Off Only zero at power on
1
) Not available on devices verified for use in legal metrology
2
) Depends on weighing platform type
Operating Design
Operating Design
Device Parameters
WP 1
Internal
Weight unit 2
2
)
User-definable / o (factory setting: grams)
1
)
o Grams / g
Kilograms / kg Carats / ct
1
)
Pounds / lb
1
)
Ounces / oz
1
)
Troy ounces / ozt
1
)
Hong Kong taels / tlh
1
)
Singapore taels / tls
1
)
Taiwan taels / tlt
1
) Grains / GN 1) Pennyweights / dwt
1
) Milligrams / mg 1) Parts per pound / / lb
1
)
Chinese taels / tlc
1
)
Mommes / mom
1
)
Austrian carats / K
1
)
Tola / tol
1
)
Baht / bat
1
)
Mesghal / MS
1
) Tons / t lb / oz
Display accuracy 2
o All digits
One less for wt. change Index +1
1
)
Index +2
1
)
One less
1
)
Factory settings: weighing parameters only
Yes
o No
1
) Not available on devices verified for use in legal metrology
2
) Depends on weighing platform type
58
59
Device Parameters
WP 1
Internal
Weight unit 3
2
)
User-definable / o (factory setting: grams)
1
)
o Grams / g
Kilograms / kg Carats / ct
1
)
Pounds / lb
1
)
Ounces / oz
1
)
Troy ounces / ozt
1
)
Hong Kong taels / tlh
1
)
Singapore taels / tls
1
)
Taiwan taels / tlt
1
) Grains / GN 1) Pennyweights / dwt
1
) Milligrams / mg 1) Parts per pound / lb
1
)
Chinese taels / tlc
1
)
Mommes / mom
1
)
Austrian carats / K
1
)
Tola / tol
1
)
Baht / bat
1
)
Mesghal / MS
1
) Tons / t lb / oz
Display accuracy 3
o All digits
One less for wt. change Index +1
1
)
Index +2
1
)
One less
1
)
Factory settings: weighing parameters only
Yes
o No
Off COM 1 (when the WP is assigned to this interface) COM 2 (when the WP is assigned to this interface) UNICOM (when the WP is assigned to this interface)
Device parameters
WP 2
RS-232
3
) o ADC-232 similar to “Internal" menu for WP 1
RS-485
3
) o ADC-485 similar to “Internal" menu for WP 1
o Off
COM 1 similar to WP 1 COM 2 similar to WP 1 UNICOM similar to WP 1
1
) Not available on devices verified for use in legal metrology
2
) Depends on weighing platform type
3
) Equipment version
Operating Design
Operating Design
Device Parameters
COM 1
o Off
WP 3
o RS-232
SBI standard (9600 baud) SBI verifiable (9600 baud)
o IS-232
1
)
ADC-232
1
)
Data communications
o SBI
Baud rate
150 baud 300 baud 600 baud
o 1200 baud
2400 baud 4800 baud 9600 baud
19200 baud
Parity
Space
1
)
o Odd
Even None
2
)
Number of stop bits
o 1 stop bit
2 stop bits
Handshake mode
Software handshake
o Hardware 1-char
Number of data bits
o 7 bits
8 bits
Data output
On request, without stability
o On request, after stability
Automatic, without stability
o 1 display update
2 display updates
10 display updates
100 display updates
On request, without stability
o 1 display update
2 display updates
10 display updates
100 display updates
Printout printer 1 Printout printer 2
1
) Not with 8 data bits
2
) Not with 7 data bits
60
61
Device Parameters
COM 1
Data communications
o SBI
Line format
For raw data (16 characters)
o For other apps. (22 characters)
Sign format
Do not output + sign
o Output + sign
Factory setting
Yes
o No
XBPI RS-232
SMA
Baud rate
150 baud 300 baud 600 baud
o 1200 baud
2400 baud 4800 baud 9600 baud
19200 baud
Parity
Space
1
) Odd Even
o None
2
)
Number of stop bits
o 1 stop bit
2 stop bits
Handshake mode
Software handshake
o Hardware 1-char
Number of data bits
o 8 bits
1
) Not with 8 data bits
2
) Not with 7 data bits
Operating Design
Operating Design
Device Parameters
COM 1
Printer 1
1
)
YDP20
Baud rate
o 1200 baud
2400 baud 4800 baud 9600 baud
19200 baud
Parity
Space
o Odd
Even
Number of stop bits
o 1 stop bit
2 stop bits
Handshake mode
Software handshake
o Hardware 1-char
YDP14IS
o Strip
Label
Universal
Baud rate
150 baud 300 baud
600 baud 1200 baud 2400 baud 4800 baud
o 9600 baud
19200 baud
Parity
Odd Even
o None
Number of stop bits
o 1 stop bit
2 stop bits
Handshake mode
o Software handshake
Hardware, 1-char
Number of data bits
o 8 bits
o YDP14IS
o Strip
Label Label, manual form feed
Printer 2
1
) Similar to printer 1
1
) Max. 2 printers can be configured
62
63
Device Parameters
COM 2 similar to COM 1
UNICOM (optional interface)
o Off
WP 3
o RS-232
SBI standard (9600 baud) SBI verifiable (9600 baud)
o IS-232
1
)
ADC-232
1
)
RS-485
o IS-485
1
)
ADC-485
1
)
Data communications
o SBI similar to COM 1
xBPI-232 similar to COM 1 xBPI-485 address 0 to 31 can be selected SMA similar to COM 1 Profibus address: 0 to 126 can be selected, factory setting: 126
Ethernet
Printer 1
1
) Similar to COM 1
Printer 2 1) Similar to COM 1
Analog output
Analog output: value
o Net value
Gross value
Analog Out: Error Signal
o High level (20 mA)
Low level (0/4 mA). When menu is open or during calibration (0/4 mA) on this interface.
Analog output: mode
o Zero to maximum capacity
Minimum/Maximum values
Analog Out: Data Output Min./Max.
o Min. (0/4 mA) input in kg
Max. (20 mA) input in kg
1
) Not with 8 data bits
Operating Design
Operating Design
Device Parameters
I/O control
Input ports
Universal switch key
Off
p Print key p Print key - long ) Tare key
o J ISO-Test key
k Fn key n WP toggle key
Comb. tare/zero function (zero if possible, otherwise tare.)
( Zero key e On key c CF key v F1 function key D Toggle between applications K 10x higher resolution key L Net/gross value key
Level internal output ports
o All active low
All active high
Bar code
o Reference
Tare ID1 Input Input without activating function External keyboard
Config. printout
Headers
Line 1: Line 2:
ID codes
ID1: ID2: ID3: ID4: ID5: ID6:
ISO/GLP/GMP printout
o Off
For several application results
Date/time
o Date with time
Date only
Once at stability
o Off
On
FlexPrint
o Off
On
64
65
Device Parameters
Config. printout
Printer 1
Number of printouts
o 1 printout
2 printouts
Indiv.: Printout f. app./weighing
List <-> Selection Gross (G#) Space line Tare -------­Net (N) Form feed Space line Date/time
Time GMP header GLP footer Transaction no. ID1…ID6 Tare 2 (T2/PT2) Header line 1 Header line 2 Scale ser. no. Product no. Product name Product ID1…ID4
(application-dependent values)
Comp.: Printout after saving val.
2
) Similar to individual
Total: Printout after pressing CF
2
) Similar to individual
Printer 2
1
) Similar to printer 1
Decimal separator
o Period
Comma
Data records (e.g. alibi memory, product data memory)
All data records
o Specified quantity
Printout factory setting
Yes
o No
1
) Max. 2 printers can be configured
2
) Only when “Totalizing" and/or “Net total" has been selected under “Application parameters:Application 3"
Operating Design
Operating Design
Device Parameters
Operating parameters
Acoustic signal
oOn
Off Linked to the green LED
Keypad
Block key functions
All keys unblocked All blocked except e Alphanumeric keys blocked Weighing platform switch disabled Zero key blocked Tare key blocked Fn key blocked ISO-Test key blocked Print key blocked 10X higher resolution key blocked Net/gross value key blocked CF key blocked Softkey 5 key blocked Softkey 4 key blocked Softkey 3 key blocked Softkey 2 key blocked Softkey 1 key blocked Application switch disabled Setup key blocked Mem key blocked
Display
Contrast
0 1 2 3 4
o5
6 7
Backlighting
oOn
Auto shutoff acc. to. timer
Model
o2
1
Automatic shutdown
Auto-off acc. to. timer
o No automatic shutoff
Timer
o 1 + 1 minute
2 + 2 minutes 5 + 5 minutes
Main scale
oWP 1
WP 2 WP 3
Display geogr. data
Off
oOn
Factory settings: operat. param.
Yes
oNo
66
67
Device Parameters
Clock
Time: Date:
Test I/O ports
Set internal outputs
Int. output 1 (Lighter) 0 Int. output 2 (Equal) 0 Int. output 3 (Heavier) 0 Int. output 4 (Set) 0
Read internal inputs
Int. input 1: 0
Passwords
Password:
SQmin
Display
oNo
Yes
GMP print
oNo
Yes
Alibi memory
Alibi memory period
In days 90
Operating Design
Operating Design
Setup
Info
Service
Service date:
Terminal
Model: Serial no. Basic ID: Version no.: (application software version)
1. WP 1 Model 2): Version no.: (software version) Serial no. 2): Latitude: 1) Altitude: 1) Grav. acc.: 1) Access switch
2. WP 2, see WP 1
3. WP 3, see WP 1
FlexInfo
ID--­V.---
Alibi memory
Date Time Transaction no.
Language
Deutsch
o English
U.S. mode Français Italiano Español
1
) Depending on configuration prior to verification: either latitude/altitude or gravitational acceleration
2
) Not displayed for internal ADC
68
69

Weighing

This application is always available during operation.
Features
– Zeroing by pressing ( – Storing the weight on the platform as a tare by pressing ) – Tare container weight automatically – Use a barcode scanner to enter tare weight – Use the numeric keys 0 to enter a tare weight and press ) to save – Delete tare values using the c key and save using the ) key – Toggle the display using the k key between: – 1st and 2nd weight unit – SQmin – You can configure the k key function in the “
Fn key" Setup menu
– 10-fold increased resolution using the K key – Weighing using up to three weighing platforms – Individual numeric ID codes for weight values – Print weight values: – Manually, by pressing p – GMP printout (see “Data Interfaces") – Automatic printout – Automatic data output (see “Data Interfaces") – Restore factory settings in the Setup menu:
"
Application parameters:Factory settings: only
application:Yes"
Soft Key Functions
ID Enter up to six ID codes for identifying results on the printout 1st ID Save the value entered as the first ID code.
Preparation
t Select Setup: Press M. t Select Application parameters: Press the “O" soft key. t Select “Application 1 (basic settings)":
Press the “
O" soft key.
t Confirm the “Weighing" application: Press the “l" and “o" soft keys. t Select the “Min. load f. auto. taring/printout" function:
Press the “
O" soft key.
t Confirm the “o 10 digits" setting:
Press the “
l" and “o" soft keys.
t Select the “Factory setting" menu item: Press the “O" soft key. t Confirm the “o No" setting: Press the “l" and “o" soft keys. t To save settings and exit the Setup menu: Press the M key or the “oo" soft key.
Automatic Taring
The first weight on the scale that exceeds the preset minimum capacity is stored in the tare memory at stability. The values for subsequent loads are stored as weight values.
The scale returns to the initial state when the load on the scale is less than 50% of the minimum load. Operating menu setting: “
Application parameters:
Autotare 1st weight:Yes"

Operation

Operation
Minimum load for automatic taring and automatic printing
To tare container weights automatically, you need to set a minimum load in the Setup menu, under:
Application parameters:Min. load f. auto.
taring/printout:o 10 digits"
You can set the following for the minimum load: 1 digit (no minimum load)
2 digits 5 digits 10 digits 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits
The “digits" here refer to the scale intervals for the connected weighing platform. If the interval is 1 g and 1000 digits are required, the minimum load is 1000 g (1000 intervals).
If the weighing platform interval is 5 g and the same number of digits as above are required, the minimum load is 5000 g.
Once the minimum load has been exceeded, the weighing platform is tared automatically and/or an automatic printout is generated.
Automatic Printing
The first weight value that exceeds the minimum load is printed. If the menu item is also activated for automatic taring, it is only tared when the minimum load is exceeded. In this case, an automatic printout would only be generated when the second weight value exceeds the minimum load. Setting in the Setup menu:
Device parameters:Config. printout:Once at
stability:On"
Weighing using up to three weighing platforms
Press the n key to toggle between three weighing platforms.
Main scale: first platform displayed on start-up
You can select the weighing platform to be displayed first when the device is turned on in the Setup menu:
Device parameters:Operating parameters:Main
scale:oWP 1"
Press n to toggle the readout between platforms.
Using a barcode scanner to enter tare weight
The tare weight of the container can be entered via a barcode scanner. Operating menu setting: “
Device parameters:Bar
code:oTare-value"
The value is applied and saved automatically. The content of the tare memory can be displayed in Info mode by pressing the D key.
Using a barcode scanner to enter application parameters
Application parameters (reference value) can be entered via a barcode scanner. Operating menu setting: “
Device parameters:Bar code:
oReference-value
"
The value is applied and saved automatically.
70
71
Using a barcode scanner to enter an identifier
Identifiers can be entered via a barcode scanner. Operating menu setting: “
Device parameters:Bar code:oID"
The value is applied and saved automatically.
Scanning barcodes directly
You can directly scan a barcode using the barcode scanner. Operating menu setting: “
Device parameters:Bar code:oInput"
The barcode can contain the following codes: – “I" for write identifier – “T" for save tare memory – “R" for write reference weight – “A" for activate product data memory
Examples:
“I4Anton" = write the character string to ID 4: Anton
“TC1" = write 1 Kg to the preset tare memory.
“C" = unit: Kilograms
“B" = grams
“D" = carat,
etc.
“RC0.0023" = write 0.0023 kg as the reference weight
“A1" = load product data memory 1
Operation
Operation

Adjustment/Configuration counter for standard scales

Purpose
Automatic recording of changes to adjustment and weighing parameters using two independent counters. The values remain saved for the life of the component.
t To display both counters, press and hold the ( key for longer than 2 seconds. y The “Configuration counter" is then shown in the weight display for 3 seconds
(identified by a “P"). Then the “Adjustment counter" is displayed for another 3 seconds (identified by a “C"). After 6 seconds, the information display turns off automatically.
Adjustment counter features:
– Counter limited to 9999 – Counter at “C 0000" for hardware commissioning – Counter cannot be reset – Counter is updated automatically when:
– linearization, calibration/adjustment/preload change is successful – user calibration, adjustment or linearization weight is changed – When the following parameters are changed:
– Function J key – Zeroing range – Initial zero point range – The above parameters are reset to factory settings
Configuration counter features:
– Counter limited to 9999 – Counter at “P 0000" for hardware commissioning – Counter cannot be reset – Counter is updated automatically when:
– When the following parameters are changed:
– Installation location – Application filter – Stability range – Taring – Auto zero – Weight unit 1 – Weight unit 2 – Weight unit 3 – The above parameters are reset to factory settings
– Turning the application automatic taring on/off – Restore application parameters to factory settings
72
73

Device Parameters

Password protection
Access to device and application parameters can be password-protected against
unauthorized changes in the Setup menu under “
Device parameters:
Password," see the “Setting up Password Protection" section in the “Operating
Design" chapter.
Acoustic Signal
An acoustic signal (single beep for active, double beep for inactive keys) is emitted when
you press a key.
The acoustic signal can be turned on/off and linked to the green LED in the Setup menu
under “
Device parameters:Operating
parameters:Acoustic signal."
Keys
Keys can be blocked/unblocked in the Setup menu under “
Device
parameters:Operating parameters:Keys." The following
settings are available: – All keys unblocked (default setting) – All blocked except e and H – Alphanumeric keys blocked – n blocked – ( blocked – ) blocked – k blocked – J blocked – p blocked – K blocked – L blocked – c blocked – Soft key 1…5 blocked – D blocked – H blocked – c blocked
Automatic shutoff of Combics
You can specify that the indicator will shut off after a set time has elapsed in the Setup
menu under “
Device parameters:Operating parameters:
Automatic shutoff
,"
Display lighting
The following settings can be made for display lighting in the Setup menu under
Device parameters:Operating parameters:Display:"
– Contrast – Backlighting – Model
Timer
The timer used to turn the device and/or display lighting on/off can be set to 2, 4, or
10 minutes in the Setup menu under “
Device parameters:Operating
parameters:Timer
."
Operation
Operation
Tare the scale by placing a container on the weighing platform
t Press the e key to turn on the indicator. y The automatic self-test runs.
When the weight readout is shown, the scale is ready to operate and automatically set to zero. With no load on the platform, you can zero the weighing platform at any time by pressing (.
t Place empty container on the scale. y The container weight is displayed. t Press the ) key to tare the scale.
Note: If the automatic tare function is enabled, you do not need to press the ) key. The tare weight is saved automatically when you place the container on the platform.
t Wait until a zero value is displayed together with the “NET" symbol (net weight).
t Place sample on the platform.
t Wait until the weight unit symbol is displayed (indicating stability). t Read the weight value.
74
75
Weighing with numeric entry of tare value and printing the results.
t Press the e key to turn on the indicator.
y The automatic self-test runs.
When the weight readout is shown, the scale is ready to operate and automatically set to
zero. With no load on the platform, you can zero the weighing platform at any time by
pressing (.
t Press the .25 key to enter a known tare weight via the keypad (in this example,
0.25 kg).
t Press the ) key to apply the tare weight.
t Place the container with sample on the platform.
t Read the results.
t Press the L key to toggle the display to the net weight value.
y The display shows the gross weight (in this example, 0.250 kg for the container plus
2.000 kg for the sample).
Operation
Operation
t Press the L key to toggle back to the previous display.
t Press the p key to print the results.
y Start of GMP header (only if GMP-compliant printout is configured).
End of GMP header Headers
Identifiers
Start of GMP footer (only if GMP-compliant printout is configured)
End of GMP footer
76
--------------------
24.10.2010 10:09 Type CAW3P1-12ED-L Vers.no. 1.02.101110 BVers. 01-63-02
-------------------­EISENSCHMIDT GOETTINGEN Lot no. 123456 Customer Schulze
24.10.2010 10:09
--------------------
G# + 2.250 kg T + 0.000 kg PT2 + 0.250 kg N + 2.000 kg
--------------------
--------------------
24.10.2010 10:10 Name:
--------------------
77
Weighing with variable tare values, printing results, deleting tare values
t Press the e key to turn on the indicator.
y The automatic self-test runs.
When the weight readout is shown, the scale is ready to operate and automatically set to
zero. With no load on the platform, you can zero the weighing platform at any time by
pressing (.
t Place empty container on the scale. t Press the ) key to tare the scale.
Note: If the automatic tare function is enabled, you do not need to press the ) key.
The tare weight is saved automatically when you place the container on the platform.
t Wait until a zero value is displayed together with the “NET" symbol (net weight).
t Place a packaged sample in the container. t Press the .25 key to enter the known tare weight of the package via the keypad
(in this example, 0.25 kg).
t Press the ) key to apply the package weight that was entered.
Both tare weights are added together.
Operation
Operation
t Read the Net weight.
t Press the p key to print the results.
t Press the 0 key. t Press the ) key to apply the entered value.
y The tare values are deleted. The gross value appears on the display.
t Press the p key to print the results.
78
G# + 6.433 kg T + 0.000 kg N + 6.433 kg
G# + 6.433 kg T + 4.183 kg PT2 + 0.250 kg N + 2.000 kg
--------------------
79

Calibration and Adjustment

Purpose
Perform calibration to determine the difference between the value displayed and the actual weight on the platform. Calibration does not entail making any changes within the weighing equipment.
During adjustment, the difference between the measured value displayed and the true weight of a sample is corrected, or is reduced to an allowable level within maximum permissible error limits.

Configuration for Use in Legal Metrology

Configuration of the weighing instrument for use in legal metrology is set by a switch. The switch is located on the back of the weighing platform and covered by a protective cap.
Using a verified scale in legal metrology in the EU: The Type-Approval Certificate for verified scales is only valid for non-automatic weighing
instruments. For automatic operation with or without additional, integrated equipment, please follow the applicable national regulations for the installation location.
Externally connected IS scales
Before use in legal metrology, the scale should be calibrated via the internal calibration equipment at the installation location: see chapter “Configuring Weighing Platforms, section “Internal Calibration/Adjustment."
3
The temperature range (°C) listed on the ID label should not be exceeded during operation.
For Servicing
External calibration for verified scales of accuracy class l:
– External calibration is blocked in legal metrology (switch cover is sealed) – External calibration is only possible by removing the seal.
If the seal is broken, the validity of verification will become void and you must have your scale re-verified.
Using a verified scale in legal metrology with internal calibration equipment:
t Before use in legal metrology, the “internal calibration" function should be carried out at the
installation location:
Opening the menu access switch
The menu access switch is located on the back of the indicator right next to the weighing platform connection.
t Remove the cap. t Slide the switch to the left (= “open" position, for use in legal metrology).
h
You can determine the position of the switch in the Setup menu under “Info:WP."
Operation
P
S
2
Operation
Features
Which of the following features are available depends on the connected weighing platform. These features are configured in the Setup menu:
– External calibration/adjustment blocked in verified weighing instruments
- External calibration/adjustment with the default weight value or user-defined weight (not available on verified instruments) under “…
Calibration/adjustment:CAL key function"
- Specify the weight for external calibration/adjustment under “…
Calibration/adjustment:Parameter for external
weight"
- Internal calibration/adjustment of IS weighing platforms (configuration in “
COM 1:, COM 2: or UNICOM:WP 3 )
- Block the J key to prevent the use of the two functions described above under “…
Calibration/adjustment:CAL key function"
- Calibrate first; then adjust automatically or manually (not for verified weighing instruments) under “…
Calibration/adjustment:Cal/adj.
sequence"
- For Sartorius IS weighing platforms only: Flashing WP symbol as adjustment prompt (If more than one weighing platform is connected, the platform number is also displayed) under “…
Calibration/adjustment:CAL key function"
- Block/Unblock external calibration/adjustment under “…
Calibration/adjustment:Activate ext. adj."
- Altitude and latitude or gravitational acceleration displayed after “
Cal" is displayed at
the start of the calibration/adjustment process if these values are supported by the weighing platform under “
Device parameters:Operating
parameters:Display geogr. data"
For each of these parameters, the term is displayed first (Altitude, Latitude or Gravity) for 1 second, and then the corresponding value is displayed continuously until you press J.
80
81

External Calibration and Manual Adjustment with Default Weights (weighing parameters: factory settings)

Preparation
t Select Setup: Press the M key. t Select device parameters: Press the “O" soft key. t Select weighing platform “WP 1," “WP 2:" Press the “O" soft key or t Select interface “COM 1," “COM 2"or “UNICOM" (depending on the interface):
Press the “
O" soft key.
t Select weighing platform WP 3: Press the “O" soft key.
Calibration/Adjustment
CAL key function
o Ext. cal./adj.; factory-def. wt.
Ext. cal./adj.; user-defined wt. Key blocked
Cal./adj. sequence
Cal. then auto adj.
o Cal. then manual adj.
isoCAL function (for Sartorius IS weighing platforms only)
oOff
Adjustment prompt
Activate ext. adj.
o Activated
Deactivated
Parameter for external weight
o = Factory setting
t To save settings and exit the Setup menu: Press the M key or the “oo" soft key.
Procedure
t Zeroing the scale by pressing the ( key.
t Start calibration/adjustment using the J key.
y The display “C.EXT.D" appears for two seconds.
y You are prompted to place the required weight on the platform (e.g., 10,000 g).
Operation
12
C.EXT.D
C.EXT.DEF.
Operation
t Place the calibration/adjustment weight on the weighing platform.
y The difference between the measured value and the true weight of the sample will be
displayed with plus/minus signs.
y A printout will be generated if the adjustment is not carried out and the procedure is
stopped by pressing the ( key.
t Start calibration/adjustment via the J key (cancel via ().
y The adjustment weight is displayed once adjustment is finished.
y A GMP-compliant printout is generated
82
External calibration Nom. + 10000 g Diff. + 1 g
--------------------
24.10.2010 10:15 Type CAIS3 Vers.no. 1.02.101110 BVers. 01-63-02
-------------------­External calibration Nom. + 10000 g Diff. + 1 g External adjustment Diff. + 0 g
--------------------
24.10.2010 10:15 Name:
--------------------
83

SQmin Function

Purpose
To display the allowable minimum sample quantity “SQmin” (sample quantity minimum) in accordance with the United States Pharmacopoeia (USP). According to USP guidelines, the uncertainty of measurement may not exceed 0.1% of the sample quantity when substances are weighed with the highest degree of accuracy for volume determination. This additional function ensures that weighing results lie within defined tolerance limits corresponding to the requirements of your quality assurance system.
System Requirements
The scale must be set up by a service technician to be able to use the SQmin function. The technician will determine the permitted minimum sample quantity and set this on your balance using your QA system's guidelines. He will document this setting via a “Weighing module test as per USP" certificate in which the measurements and min. sample quantity are logged. The SQmin function ensures that the weighing results correspond to USP guidelines. These SQmin settings cannot be changed by the user.
Features
– Displaying the minimum sample quality: The value is displayed in the text line for
4 seconds after pressing the k key.
– If the minimum sample quantity has not been reached: The
ksymbol
is displayed and weight values are marked with a “!" in the printout.
– GLP header: The minimum sample quantity entered for SQmin can be included on the
printout.
Operation
Operation

SQmin Operation

Configuration in Service Mode
The SQmin value can only be entered in Service mode.
t Remove the cap. t Slide the menu access switch to the left (= “open" position).
If the device is part of a verified weighing facility, this will only be possible if the verification seal is broken. The weighing equipment must then be verified again.
t Activate the Service mode, see “Service Mode." t In the “Device parameters" menu, select “SQmin" and open using the
O" soft key.
t Select “SQmin input" using the “O" soft key.
t Select “SQmin WP1" using the “O" soft key, enter the value via the keypad (in this
example 0.100 kg) and save using the “
l" soft key or cancel using “ESC."
t Then enter the values for WP 2 and WP 3. t Press M or “oo" to exit the Setup menu. t Slide the menu access switch to the right (= “closed" position) and reattach the cap.
y The device is now in normal weighing mode.
Configuration in Weighing Mode
t In the “Device parameters" menu, select “SQmin" and open using the
O" soft key.
t Select “Display.". t Open using the “O" soft key. t Confirm the factory setting “oYes" using the “l" soft key. t Press the “o" soft key several times to switch to the next highest menu.
The SQmin display must be turned on to use the SQmin function.
t In the “Fn key:SQmin display:" menu, select using the “O" soft key and
confirm using the “
l" soft key.
t Press M or “oo" to exit the Setup menu.
84
85
Procedure
t Place the container for the sample on the scale and press the ) key to tare.
t Place the sample on the scale.
y The minimum sample quantity is not reached (symbol k).
t Press the p key to generate the printout.
t Place another sample on the scale.
y The minimum sample quantity is exceeded (no symbol k).
t Press the p key to generate the printout.
t Briefly press the k key to toggle between the measured value and SQmin value.
y The value for the minimum sample quantity is displayed for four seconds.
Operation
Operation

Data ID Codes

You can assign codes (such as product name, batch number, etc.) for identification of measured values in all application programs.
Features
– Assign up to six ID codes. – Assign both a name and a value to each ID code. – Displaying individual IDs: press the “
ID" soft key.
– The name is left-justified and the value is right-justified on the printout. If the entire
code is too long for one line, additional lines are printed.
– Enter ID code names in Setup under: “
Device parameters:Config.
printout:ID codes."
– The name can have a max. of 20 characters. When entering the
ID value, no more than
11 characters are displayed during input; however, all 20 characters are printed.
– Enter up to 21 characters for the value of the ID code. Press the “
ID" soft key to
activate the input mode.
– Enter the first ID value directly through the numeric keypad. Press the “
1st ID" soft
key to save the value.
– Individual characters of the ID can be deleted using the ( key. Complete ID code
values can be deleted using the “
Delete" soft key.
– If both the name and value fields are empty, no ID code is printed. – In the Setup menu, you can configure when and whether ID codes are printed (see
included “Application Programs" instructions, “Configuring Printouts").
Settings for individual ID codes
Setup menu: “
Device parameters:Config. printout:
ID codes"
Factory settings for ID code names: ID1: “
ID1"
ID2: “
ID2"
ID3: “
ID3"
ID4: “
ID4"
ID5: “
ID5"
ID6: “
ID6"
There are no factory settings for ID code values.
86
87

Using Individual ID Code

Example
Enter ID code names. “Batch no." and “Customer" should be entered for ID 1 and ID 2.
t Press the M key and select “Device parameters:Config.
printout:ID codes."
y The first line is selected.
t Press the a key to enter “Batch no." via the keypad, see also chapter “Operating
Design," sections “Text Input via the Keypad" and “Special Character Input via the Keypad."
t Press the “l" soft key to save the entry. t Press the “q" soft key to select the second line. t Press the a key to enter “Customer" via the keypad, see also “Operating Design,"
“Text Input via the Keypad."
t Press the “l" soft key to save the entry. t Press M or “oo" to exit the Setup menu.
Example
Enter ID code values. The values “123456" and “Smith" should be entered for ID 1 and ID 2.
t Press the “ID" soft key.
y The first line is selected.
t Press the key to enter “123456" via the keypad, see also “Operating Design,"
“Number Input via the Keypad."
t Press the “l" soft key to save the entry.
t Press the “q" soft key to select the second line. t Press the a key to enter “Smith" via the keypad, see also “Operating Design,"
“Text Input via the Keypad."
t Press the “l" soft key to save the entry. t Press the “oo" soft key to exit the menu.
Operation

Data Interfaces

The indicator is equipped with the following data interfaces: – COM1: Standard data interface (RS-232) – COM2: Standard data interface (RS-232) – UniCOM: Universal data interface (optional)
Interfaces can be configured in the Setup menu for different input and output functions
(e.g. printer, 3rd weighing platform, PC, checkweighing).
The optional UniCOM interface can be used as an RS-232, RS-485/RS-422 or analog output
(voltage/current interface, galvanically separated digital I/Os, profibus, Ethernet), see also
“Accessories."
A barcode scanner or a keyboard can be connected via the PS/2 -socket or the corresponding
screw terminals (IP69K).
Features
– IP44 indicator:
Connect via a 25-pin D-Sub female connector – IP69K indicator:
Route the connecting cable from the peripheral device to the indicator via a cable gland.
The free cable ends are connected via the screw terminals.
3
Warning when using third-party RS-232 connecting cables:
The pin assignments may not be compatible with Sartorius equipment.
Check the pin assignment against the cabling diagrams and disconnect any lines that are
not assigned. Failure to do so may cause malfunction, damage or even completely ruin your
indicator and/or peripheral device(s).
88
89

Specifications

Serial interface: Interface operating mode: Full duplex
Level: COM1: RS-232,
COM2: RS-232, UniCOM (optional): RS-232 or
RS-422/485 half duplex
Connection: IP44 devices:
25-pin D-Sub socket IP69K devices: Connection via screw terminals in the housing, cable routed into the housing via a cable gland.
Transmission rate: 150, 300, 600, 1200, 2400, 4800, 9600, 19200 baud
(depending on the operating mode) Number of data bits: 7, 8 bits Parity: Space, odd, even, none (depending on the
operating mode) Number of stop bits: 1 or 2 Handshake mode: Software (XON/XOFF), hardware (1 character after CTS) Protocols: SBI, XBPI-232
2)
, XBPI-485
1)2)
, SMA, Profibus (UniCOM only) various printers: – YDP01IS – YDP02IS label – YDP01IS label – Universal – YDP02 – YDP04IS – YDP03 – YDP04IS label – YDP02IS – YAM01IS Alibi memory
Network address3): 0, 1, 2, ..., 31 SBI: Manual data output: Without stability, after stability, configurable printout SBI: Auto. data output: Without stability, at stability, at user-defined intervals SBI: Output format: 16 or 22 characters Printout of application data: Output of a configurable printout
Factory Settings Depends on the device configured; for example: “SBI," “Data communication" setting
Transmission rate: 1200 baud Number of data bits: 7 bits Parity: Odd Number of stop bits: 1 stop bit Handshake: Hardware handshake, 1 character after CTS Activation of data output: Individual print out after stability Time-dependent autoprint: 1 display update Output format: 22 characters
Analog UniCOM interface (optional) Level: 4…20 mA, 0…20 mA, 0…5 V
Power supply: Internal or external Factory setting: 4...20 mA, internal power supply
Connection: CAISL3 devices (IP44):
25-pin D-Sub socket CAIS3 devices (IP69K): Connection via screw terminals in the housing, cable routed into the housing via a cable gland.
1)
Optional UniCOM universal data interface
2)
xBPI operating mode: 9600 baud, 8 data bits, parity: odd, 1 stop bit
3)
Network address is valid only in xBPI-RS-485 operating mode
Data Interfaces
Data Interfaces

Connection Options

3
You may need to use an external power supply to operate peripheral devices.
Preparation
See chapter “Getting Started," sections “COM1 Pin Assignment Chart, “COM2 Pin Assignment Chart," “PS2 Pin Assignment Chart" for pin assignments and cabling diagrams.
Printer connection options
The following printers can be connected to the standard COM1 and COM2 interfaces or
the optional UniCOM universal interface: – YDP20 (user-defined interface parameters) – YDP14IS (strip or label printer) – YDP04IS (strip or label printer) – Universal printer (user-definable transmission parameters)
Device connection options
In addition, the following devices can be connected to the standard COM1 and COM2
interfaces: – Foot / Hand switch, COM1 only – Second printer – Remote display – PC (RS-232 interface) – Third weighing platform (RS-232 interface) – External checkweighing display (stop light) via a digital I/O (Sartorius standard), COM1 only – To PS2: Barcode scanner/External keyboard
The following devices can be connected to the optional UniCOM universal interface: – WP3 RS-232/RS-485 – PC (RS-232 interface) – Second printer (external power source required) – Remote display – Digital I/O – Current interface (0/4...20 mA), voltage (0...10 V) – PLC with Profibus DP – Ethernet
Connection options for other weighing platforms
The Combics 3 model enables you to connect a 2nd or 3rd weighing platform. You can use
both the COM1, COM2 or the UniCOM port.
The standard COM1 and COM2 port is operated in RS-232 mode. The following modes are
available for a 3rd weighing platform: –SBI – IS-232 (factory setting) – ADC-232
Operating modes
UniCOM interfaces can be operated in either the RS-232 or RS-485 mode.
The third weighing platform can be operated in any of the following modes: – SBI (RS-232 mode) – IS-232 (RS-232 mode) – ADC-232 (RS-232 mode) – IS-485 (RS-485 mode, xBPI mode, factory setting) – ADC-485 (RS-485 mode)
The standard COM1 and COM2 ports or the optional universal UniCOM interface can be used
as a printer interface.
90
91

Configuring the Data Interface as a COM Port

For operation as a COM port, you can adapt data records to the following operating
modes: – SBI (factory setting) – XBPI-232 – XBPI-485 – SMA
In the SBI communication mode, you can control a display unit and a connected
weighing platform by sending ESC commands from a PC to the communications port
(COM1, COM2 or UniCOM). See also “Data Input Format."
COM1 o Off
WP 3 RS-232 SBI standard version
SBI trade version (for legal metrology)
o IS-232
ADC-232
Data o SBI communication XBPI-232
SMA
Printer 1
1
) YDP20 or YDPO4IS o Strip Printer 2
1
) Label
Universal (printer) YDP04IS o Strip
Label Label, man. form feed
COM2 o Off
WP 3 RS-232 SBI standard version
(9600 baud) SBI trade version
(for legal metrology)
o IS-232
ADC-232
Data o SBI communication XBPI-232
SMA
Printer 1
1
) YDP20 or YDPO4IS o Strip Printer 2
1
) Label
Universal (printer) YDP04IS o Strip
Label Label, man. form feed
UNICOM o Off (option) WP 3 RS232
RS485
Data o SBI communication XBPI-232
XBPI-485 SMA Profibus Ethernet
Printer 1
1
) see COM2 or Printer 2
1
) Analog output External Multi-I/O converter
o Factory setting
1
) Max. 2 printers can be configured
Data Interfaces
Data Interfaces

Setting the SBI Data Output

Data output settings can be made in the Setup menu under “Data
communications:SBI:Data output."
The following options are available: – The displayed value, with or without stability check – Automatic output of the displayed value, either with or without stability check,
or automatically at defined intervals – Output of a printout as configured in the “
Device parameters:
Config. printout:Printer 1" or “Printer 2" (see next page).
You can define the printout content by specifying which blocks of information are to be
included (see “Configuring Printouts”).
In general, the current display value is output (weight with unit, calculated value,
alphanumeric display) unless you configure a user-definable data record.
Generally, data is output only after the weighing instrument has stabilized (factory
setting). Here you can define whether data is output on request or automatically, at
stability or without stability check; and configure a user-defined printout. The data
output for “without stability" must be set accordingly (SBI: Data output).
If you select time-dependent automatic printout, you need to define the print interval
(in display updates) as well.
Each line of a printout can contain up to 20 characters. The first 6 characters, called the
“data header”, identify the subsequent value. Only 14 characters are available if you
disable the header. This is carried out in the “Line format" menu item (see also “Setup
Overview" in “Configuration").
Automatic Data Output (SBI)
You can have the weight readout printed automatically
1
). This printout can be generated
after a certain number of display updates
2
). You can also configure whether or not the
auto-print function is dependent on the stability of the scale
3
). The display update frequency depends on both the scale model and the operating status. Examples:
N + 153.00 g Net weight Stat Display blank Stat L Display underload Stat H Display overload
"Data output" setting:
1) 3
)Automatic, without stability or automatic with stability.
Factory setting: Manual after stability; i.e., automatic data output function off.
2
) Time-dependent automatic data output:
Interval: 1, 2, 10 or 100 display updates Factory setting: 1 display update

Data Input Format

You can connect a computer to your scale to send commands controlling weighing instrument functions and applications via the interface port.
All commands use the same data input format. They begin with the ESC (ASCII: 27) character and end with a carriage return CR (ASCII: 13) and LF (ASCII: 10). The total length of a command is anywhere from 4 characters (1 command character between the start and end described above) to a max. of 7 characters (4 command characters). This number can also be higher when sending texts.
The commands listed in the following table must each be supplemented with ESC ... CR LF.
92
93
Example
The command character for output is “P" (“output to Port"). To trigger this command, send the string: “ESC P CR LF”.
Command Meaning
K Weighing mode 1 L Weighing mode 2 M Weighing mode 3 N Weighing mode 4 O Block keys P Send display value to data interface Q Output acoustic signal R Unblock keys D Tare and zero
(combination tare function) f3_ Zero (see also the “kZE_” command) f4_ Tare without zeroing (see also the “kT_” command) i_ Information about the indicator, example of output: “CAI/016302/1"
Meaning: Indicator: Combics 3, software version: 016302,
Active weighing platform: 1 kF1_ Trigger soft key F1 function kF2_ Trigger soft key F2 function kF3_ Trigger soft key F3 function kF4_ Trigger soft key F4 function kF5_ Trigger soft key F5 function kF6_ Trigger R key function kF7_ Trigger H key function kF8_ Trigger D key function kF9_ Trigger k key function kF10_ Trigger J key function kF11_ Trigger K key function kF12_ Trigger L key function kP_ Trigger p key function (print at printer interface) kT_ Trigger T key (tare) kNW_ Trigger n key function (toggle the weighing platform) kZE_ Trigger ( key function (zero the instrument) kCF_ Trigger c key function x1_ Output model designation of active weighing platform, example:
“LP6200S-0C" x2_ Output serial number of active weighing platform, example: “0012345678" x3_ Output software version of active weighing platform, example: “00-43-04" x4_ Output software version of indicator, example: “01-63-02" x9_ Output serial number of indicator, example: “0012345678" x10_ Output model of indicator, example: “CAIS3" z1_ Input: printout header 1 z2_ Input: printout header 2
The ASCII code for the “underline” character (“_”) is 95. Format for entering printout header lines: “ESC z x a ... a _ CR LF" with x=1 or 2 and a ... a: 1 to 20 characters for header x, followed by the underline, CR and LF characters. txx…x_ Write text in display. xx…x is the text to be displayed.
Data Interfaces
Data Interfaces

Data Output Format

You can output the value displayed in the measured value line and the weight unit, with or without a data ID code. Whether the data ID code is included in the output depends on your settings under “Line Format."
Examples
+ 235 pcs without ID code
Qnt + 235 pcs with ID code
“Line Format" settings: For raw data (16 characters): without header, for other apps. (22 characters): with “header" (factory setting).
Data Output Format with 16 Characters
Display segments that are not activated are output as spaces. Values with no decimal point are output without a decimal point.
The type of character that can be output depends on the character’s position: Normal Operation
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * A A A A A A A A * E E E CR LF or – * A A A A A A A A * E E E CR LF or * * * * * * * * * * * * * CR LF
+-: Signs; the “+" sign can be hidden under “
Device parameters:
COMx:Data communications:SBI:Sign format."
*: Space A: Digit or letter (max. 7 characters plus decimal point) E: Unit symbol
1)
(1 to 3 letters followed by 2-0 spaces) CR: Carriage return LF: Line feed
Special Codes Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * * * * * * * * * * CR LF or * * * * * * H * * * * * * * CR LF or * * * * * * H H * * * * * * CR LF or * * * * * * L * * * * * * * CR LF or * * * * * * L L * * * * * * CR LF or * * * * * * L * * * * * * * CR LF
*: Space – –: Final readout H: Overload HH: Overload in checkweighing L: Underweight L L: Underweight in checkweighing C: Calibration/Adjustment
Error Messages Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
***E r r * * # # * * * * CRLF
* * * E r r * * # # # * * * CR LF *: Space
#: Number (2 or 3 digit error number)
1
) based on the model type, e.g. not all units are available for use in legal metrology
94
95
Output of the weight value +1255.7 g Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * * * 1 2 5 5 . 7 * g * * CR LF
Position 1: Plus +, or minus - or space Position 2: Space Positions 3-10: Weight value with decimal point; leading zeros are output as spaces. Position 11: Space Positions 12-14: Characters for unit of measure, space or ! sign as a symbol Position 15: Carriage return Position 16: Line feed
Data Output Format with 22 Characters
When data is output with an ID code, the 6-character code precedes the 16-character string described above. These six characters identify the subsequent value.
Normal Operation 1 2 3 4 5 6 7 8 9 1011121314 1516 17 18 19 20 21 22 K K K K K K + A A A A A A A A A * E E E CR LF K K K K K K - A A A A A A A A A * E E E CR LF * * * * * * * * * * * * * * * * * * * * CR LF
K: ID code character, right-justified with spaces +-: Plus or minus sign *: Space A: Digit or letter (max. 7 characters plus decimal point) E: Measurement unit symbol
1)
(1 to 3 letters followed by 2-0 spaces) CR: Carriage return LF: Line feed
Special Codes 1 2 3 4 5 6 7 8 9 1011121314 1516 17 18 19 20 21 22 S t a t * * * * * * * * - - * * * * * * CR LF S t a t * * * * * * * * H * * * * * * * CR LF S t a t * * * * * * * * H H * * * * * * CR LF S t a t * * * * * * * * L * * * * * * * CR LF S t a t * * * * * * * * L L * * * * * * CR LF S t a t * * * * * * * * C * * * * * * * CR LF
*: Space – –: Final readout H: Overload HH: Overload in checkweighing L: Underweight L L: Underweight in checkweighing C: Calibration/Adjustment
Error Message 1 2 3 4 5 6 7 8 9 1011121314 1516 17 18 19 20 21 22 S t a t * * * * * S r r * * # # * * * * CR LF S t a t * * * * * S r r * # # # * * * * CR LF
*: Space #: Number (2 or 3 digit error number)
1
) based on the model type, e.g. not all units are available for use in legal metrology
Data Interfaces
96
G# Gross value N Net value T Application tare memory 1 T2 Application tare memory 2 Diff Difference from calibration value Targ. Exact adjustment weight value Nom. Exact calibration weight for SBI
protocol output
nRef Reference sample quantity pRef Percentage of reference wRef Reference piece weight Qnt Result from Counting application
Result from Counting (piece count) and Neutral Measurement applications
mDef Target value for animal weighing x-Net Animal weighing results Setp Target value for checkweighing Diff.W Absolute difference (e. g., in kg) in
Checkweighing
Lim Deviation in % in Checkweighing Max Upper tolerance for checkw. Min Min. tolerance for checkw.
Stat Status Classx Classification, class x Limx Class limit D Percentage (as loss) Prc Percentage (as residue) Wxx% Reference percentage weight Cmpxxx Component xxx Cont.T Contents of the tare memory
in Net-total Formulation
S-Comp Total of initial weighings for
Net-total Formulation
PT2 Preset tare n Transaction counter *G Sum of gross weights in Totalizing *N Sum of net weights in Totalizing Ser.no Serial number of the platform or
indicator
Example: Output of the weight value +1255.7 g 1 2 3 4 5 6 7 8 9 10111213141516171819 2021 22
G # * * * * + * * * 1 2 5 5 . 7 * g * * CR LF
Position 1-6: ID code, right-justified with spaces Position 7: Plus +, or minus - or space Position 8: Space Positions 9-16: Weight value with decimal point; leading zeros are output as spaces
(a comma can also be set instead of a decimal point).
Position 17: Space Positions 18-20: Characters for unit of measure, space or ! sign as a symbol Position 21: Carriage return Position 22: Line feed
h
If the weight value is output with 10-fold increased resolution, this value is not permitted to be printed or saved in a weighing instrument operated in legal metrology in the SBI mode. In this case, the unit symbol is not included with output.
Data Interfaces
97
Data Interfaces

External Keyboard Functions (PC Keyboard)

For settings, go to the Setup menu under “Device parameters:
Bar code:External keyboard."
The alphanumeric key codes implemented here are specific to the German keyboard layout. The following alphanumeric characters are used (some require the “Shift” key):
a - z, A - Z, 0 - 9, <space>, and these characters: ,.\+’<>/»$@%/();=:_?* Function keys:
PC keyboard Combics 3
F1 ) key F2 ( key F3 n key F4 F5 soft key (far left) F5 F4 soft key (second from left) F6 F3 soft key (middle) F7 F2 soft key (second from right) F8 F1 soft key (far right) F9 D key F10 D key - long (info function) F11 H key F12 k key Print p key Return F1 soft key (far right) Cursor up F3 soft key (middle) Cursor left F4 soft key (second from left) Cursor down F2 soft key (second from right) Cursor right F1 soft key (far right) Pos 1 c key Backspace c key ESC c key
98

Configuring the Data Interface as a Printer Port

Device Parameters
Config. printout
Headers Line 1:
Line 2:
Identifiers ID1:
ID2: ID3: ID4:
ISO/GMP o Off Protocol For several application results
Date/ o Date with time Time Date only
Once at Stability
2
)oOff
On
FlexPrint o Off
On
Printer 1 Number of
printouts o 1 printout
2 printouts
Single o Max. 30 print items printout
1
) can be selected
o ID1, ... ID4
Component/ o Max. 30 print items Printout
1
) can be selected
o
Total data Max. 30 print items printout can be selected
Printer 2 as for Printer 1
Factory setting Printout Reset
Do not reset
o Factory setting
1
) Multiple selections possible
2
) When the minimum load is exceeded (can be set under menu item:
“Application parameters: ... : Min. load for autom. taring")
Data Interfaces
99
Data Interfaces
There are several actions that generate the command for outputting data to the printer port:
– Pressing the p key.
If the operating menu is active, all menu settings under the active menu level are printed.
– In some applications, pressing a given key (e. g., to save a value or start a routine) also
generates a print command, or it is generated automatically depending on the application configuration. In this case, a configurable printout is generated with application-specific data.
The P and l symbols are displayed when data is being output to the printer port.

Configuring a Printout

A printout can be configured in the Setup menu under “Device parameters:
Config. printout." This should be carried out after configuring the
application since some data in the printout is application-dependent. You can configure a separate printout for each interface. Each printout is comprised of
different information blocks that can be activated or deactivated via multiple selection in the menu.
For the “Totalizing" and “Net-total Formulation" applications, the totalizing/results printout can be configured independent of the individual/component printout.
The individual information blocks are shown below with detailed explanations. Samples of complete printouts are provided following the end of this section.
Headers
2 headers each with a max. of 20 characters are available (e. g. for printing the company name). Print image example:
ACE HARDWARE
GOETTINGEN
Date/Time
Print image example:
21.01.2010 16:02
To maintain a uniform time (e. g. for documentation within a complete automatic system), the printing of the time can be suppressed in “
Device parameters:
Config. printout:Date/time." When you set “Date only,"
the time can be added, e. g. from a superordinate control in order to always have the same time throughout the system. This setting is mainly designed for communication with a PC.
Identifiers
Use the “
ID" soft key to enter text in the named IDs via the keypad.
100
Application Initialization Data
Which data is included in this block depends on the active application. In the “Counting" application, for example, the reference sample quantity and reference weight are printed (plus a blank line). "Counting" print block example:
nRef 10 pcs wRef + 0.035 kg
Scale Identification
Print image example for weighing platform serial number:
Ser.no. 1234567890
Scale Identification
This content is application-dependent. If provided in the application, the gross, net and tare weights are usually printed, followed by a blank line. This block is terminated by a dotted line. “Counting" application print image example:
G# + 1.402 kg T + 0.200 kg N + 1.202 kg
Qnt 34 pcs
--------------------

GMP-compliant printouts

When this function, the printout is supplemented with a GMP header and a GMP footer (GMP: “Good Manufacturing Practice”).
If the GMP-compliant printout is activated, the A symbol remains displayed until the GMP footer is printed.
Setting: Setup menu under “
Device parameters:Config.
printout:ISO/GLP/GMP- printout."
You can choose from the following settings: – GMP-compliant printout off (factory setting) – GMP-compliant printout for multiple application results
The GMP header is included from the first printout generated subsequent to the
activation of the GMP printout.
The GMP footer is printed after several measurement results by pressing and holding
the p key, e.g. for component printing (menu item “For multiple application results").
In this case, the
T symbol remains visible until the GMP footer is printed.
If you toggle to a different platform while a GMP printout of several measured results
is being generated, the GMP footer for the platform used up to that point is generated
when you press the n key. The GMP header for the other platform is included on the
next printout generated.
A GMP-compliant printout is generated automatically at the conclusion of calibration/
adjustment and linearization routines, as well as when you set or delete a preload.
If you use a label printer for GMP-compliant printouts, you may find that a single label is
not long enough for the data printed. If this is the case, you can activate the automatic
form feed after each printout of a GMP header and measurement results. The following
provides sample GMP headers and footers (see “Sample Printouts").
Data Interfaces
Loading...