Rhewa 940C/4, 940C/10, 940C/20 Operating Manual

Bench Scale 940C
Operating Manual
Adjustment Guide
RHEWA–WAAGENFABRIK
August Freudewald GmbH & Co. KG
2 Bench Scale 940C • Operating Manual and Adjustment Guide
Warranty
We reserve the right to make design changes or technical modifications to the product under description with­out prior notification.
We reserve the right to amend this docu me nta tio n with ou t prior notification.
RHEWA-WAAGENFABRIK accepts no liability for any te chni ca l or p ri nti ng erro rs o r de fic ien ci es i n this do cu ­mentation. In addition, RHEWA cannot be made liable for any damage directly or indirectly attributable to the use of the documentation.
Copying restriction
This documentation and the examples provided for the product under description constitute protected information that is subject to the law of copyright. All rights are protected under the law. This documentation may not be copied in whole or in part or otherwise reproduced without the p rior c onse nt of RHEW A .
Trade mark
The use of generic names, trade names and trade marks etc. in this documentation cannot be taken to mean that such names are not protected by trade mark and brand protection legislation and therefore open to use by anyone, even if this is not expressly stated.
Technical modifications
Rapidly advancing technological dev elopments and shorter product cycles prevent us from bringing this docu­mentation exactly into line with t he appli ance f unct ions and prop ertie s. Shou ld an y devi atio ns exis t, th e appli ance is to be used analogously.
Disposal information
Information on the disposal of packaging, batteries and used appliances can be found in our terms of delivery and on our website.
RHEWA-WAAGENFABRIK
August Freudewald GmbH & Co. KG
Feldstrasse 17 D-40822 Mettmann
Postfach 10 01 29 D-40801 Mettmann
Tel.: +49/(0)2104/14020 Fax: +49/(0)2104/140288
E-mail: info@rhewa.com Internet: http://www.rhewa.com
Document designation: Bench Scale 940C Written by:
Operating Manual Andreas Hensel Adjustment Guide Edited by:
Document number: 90514 Andreas Hensel Version/date: 1 Dated:29.04.2010 Reviewed by: Number of pages: 10 Friedhelm Lüling Appliance: Bench Scale 940C Translated by: Software version: Stuart Dykes
Language:
English
Chapter 1 Table of contents 3
1 Table of contents
Table of contents Chapter 1 3
Operation and adjustment Chapter 2 5
2.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Care and maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.4 Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.4.1 Inserting the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.4.2 Aligning the scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.5 Batt e ry operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.6 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6.1 Power on/off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6.2 Zero. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6.3 Tare. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6.4 Gross display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6.5 Putting mixtures together. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6.6 Discharge weighing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.7 Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.7.1 Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.7.2 Adjustment procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Specifications Chapter 3 9
3.1 Error messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Declaration of conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4 925 Tabletop Scale • Operating Manual and Adjustment Guide
Chapter 2 Operation and adjustment 5
2 Operation and adjustment
2.1 Safety information
For safe and troublefree operation of this scale, please ensure you follow the instructions laid down in this operating manual.
Before connecting a power adapter to the mains, ensure that the mains voltage indicated on the adapter corresponds to the actual mains voltage. If this is not the case, the adapter may not be used.
STOP
If the power adapter is opera ted at the wrong mains voltage, there is a danger of electric shock.
The power adapter may not be used if damaged or defective.
If a damaged power adapter is used, there is a danger of electric shock.
The scale may be connected only to a properly installed socket outlet.
The scale may be used in dry rooms only.
Never subject a powered-off scale to an excessive or long-term load.
Never attempt to open the scale enclosure.
2.2 Ambient conditions
Level, stable surface for the scale.
No moisture, water, liquids or corrosive substances.
No draughts (open windows or doors)
No significant temperature fluctuations (e.g. direct sunlight, outlets of air-conditioning or heating systems).
No vibrations.
No magnetic fields.
Do not place in th e immediate vicinity of transmitting devices ( mobile phones, radio equipment, etc.).
2.3 Care and maintenance
This scale is a precision instrument for which great importance was attached to exact meas­urement results during the manufacturing and setting process. When using the scale, please note the following:
!
Never open the scale enclosure and make modifications to its mechanical or electronic components. Noncompliance will void all warranty rights!
The scale will be damaged if subjected to a load above Max.
The scale keypad may be op erated by hand only. Never use pointed objects to press the keys.
Cleaning: Wipe the scale regularly with a damp cloth. Use mild cleaning agents only. No sol­vent-containing, aggressive and abrasive cleaning agents.
Calibration of the scale at regular intervals is recommended (annually if used normally).
If the keypad o r d i splay window are damaged, the scale may no longer be opera te d a nd mu st be protected against moisture, liquids and dust, in particular. The keypad or display window must then be replaced by your service partner.
6 Bench Scale 940C • Operating Manual and Adjustment Guide
2.4 Startup
2.4.1 Inserting the batteries
Remove the weighing platform .
Carefully turn over the scale.
Open the two battery compartments on the bottom of the scale.
Insert the four LR14 (1.5V) batteries supplied.
Note the polarity marked on the batteries.
Close the battery compartments.
T urn the scale back over and place on the levelling feet.
2.4.2 Aligning the scale
On the front of the scale there is a bubble level (spirit level) next to the display.
Using the adjustable feet, align the scale so that the air bubble in the level is in the middle of the circle.
The scale must sit firmly on all four feet.
Attach the weighing platform lengthwise or crosswise as required.
2.5 Battery operation
New batteries have a service life of approx. 500 hours.
To achieve a long battery life, power off the scale when not in use.
If the battery symbol on the display flash es on p ower up , the batter ies ar e e xhausted. Ba ckup operation (approx. 30 hours) is possible. Please replace the batteries!
Scale underside
Levelling feet
Levelling feet
Battery compartments
Bubble level
Chapter 2 Operation and adjustment 7
2.6 Keypad
2.6.1 Power on/off
To switch on, remove any load from the scale and press the key.
After a self-test, the display reads
.
To switch off, press the key again.
2.6.2 Zero
•Press the key if the display does not read
 with the scale at no load and instead
shows a slightly deviating value (to approx. ± 2% of Max).
2.6.3 Tare
Place an empty container on the weighing platform.
The scale displays the weight of the container =
TARE.
•Press the key . The display shows
 and "TARE" at the bottom left as an indication
that the tare weight has been subtracted and saved.
Fill the container. The scale shows the weight of the contents =
NET
Remove the filled container from the weighing platform.
The scale displays the weight of the empty container with a preceding minus sign.
On pressing the key again,
 appears on the display. The tare has been cleared.
2.6.4 Gross display
With a tared scale, press the key.
The display toggles between
GROSS and NET each time the key is pressed.
2.6.5 Putting mixtures together
•Press the key again each time a component is added.
2.6.6 Discharge weighing
Place full container on the weighing p latform. The scale displays the weight of the full container =
GROSS.
•Press the key . The display shows
 and "TARE" at the bottom left as an indication
that the tare weight has been subtracted and saved.
Remove something from the container. The scale shows the weight of the quantity removed with a preceding minus sign =
NET
Before removing any more material, press the key again.
Toggle gross
and net
Zero
ON/OFF
Tare
Weight display
Bubble level
(spirit level)
8 Bench Scale 940C • Operating Manual and Adjustment Guide
2.7 Adjustment
Owing to the high resolution of the scale, the diff erent gravitational force (g value) at diff erent plac­es of use must be taken into account. T o this end, the scale should be adjusted using test weights (accuracy class at least M1, preferably F2) before being taken into operation.
Adjustment should be performed only when necessary. Inaccurate values on the display or replacement of the load cell or the weighing electronics can
necessitate an adjustment.
2.7.1 Conditions
!
Switch on the scale and wait 15 minutes before performing the adjustment.
Adjustment can be performed only with fixed weight values. Weights corresponding to half of and the full maximum safe load are required.
2.7.2 Adjustment procedure
Align the scale exactly using the integral bubble level (spirit level).
Remove any load from the scale.
Power off the scale by pressing the key.
Press and hold down the key and power on the scale using the key.
Hold down the key until such time as the internal m easured value appears on the display (approx. 2s). (e.g.
)
With the weighing platform at no load, press the key.
The first adjustment point (zero point) is saved.
 appears briefly on the display.
The weight value for the second adjustment point (half the maximum safe load of the scale) appears on the display.
Subject the scale to the weight indicated and confirm by pressing the key.
 appears briefly on the display.
The weight value for the third adjustment point (maximum safe load of the scale) appears on the display.
Subject the scale to the weight indicated and confirm by pressing the key.
appears briefly on the display and then .
End adjustment by pressing the key.
Power off the scale by pressing the key.
Chapter 3 Specifications 9
3 Specifications
3.1 Error messages
Max / Scale interval 4 kg / 0.2 g 10 kg / 0.5 g 20 kg / 1 g Automatic shutdown After approx. 5 minutes of no use (scale at no load). Automatic initial zeroing To approx. ± 10% of Max. Manual zeroing To approx. ± 2% of Max. Permissible weighing
error
Approx. ± 3 scale intervals after adjustment at place of use using weight of accu­racy class F2.
Ambient temperature 0° to + +40°C Humidity Max. 85%, non-condensing, protected against splashing water. Mains operation Power adapter with coaxial power connector (DC 6 V / 0.3A, inner conductor +).
(available as an accessory)
Power supply 4 LR14 batteries (1.5 V)
Mains option using 230V/50Hz power adapter (available as an accessory)
Power consumption 0.3 W Net weight Approx. 1.5 kg without batteries.
Message Explanation Remedy
Cannot power up Batteries too weak Replace batteries.

Indication is unstable. A stable indication is not obtained.
Wait for a stable indication. Keep away from draughts and vibrations. Renew batteries.

AD converter is defective. Contact your service partner.

Incorrect adjustment weight. Adjust scale using correct adjustment
weight.

Weighing platform is over­loaded.
Load the scale to Max only.

- Weighing error
- Defective load cell.
Remove weight from the weighing platform. Switch appliance off and back on again.

or

- Defective main board.
- Defective load cell.
Contact your service partner.

Switch appliance off and back on again.
10 Bench Scale 940C • Operating Manual and Adjustment Guide
3.2 Declaration of conformity
KONFORMITÄTSERKLÄRUNG
Declaration of conformity
Déclaration de conformité
Die nichtselbsttätige Waage The non-automatic weighing instrument L’instrument de pesage à fonctionnement non automatique
entspricht den Anforderungen der EG-Richtlinien in den jeweils geltenden Fassungen: corresponds to the requirements of the following EC directives: correspond aux exigences des directives CE suivantes:
2004 / 108 / EC
Applied standards: EN 61000 - 6-1: 2007
EN 61000 - 6-3: 2007 EN 61000 - 3-2: 2006 EN 61000 - 3-3: 1995 + A1:2001 + A2:2005
RHEWA–WAAGENFABRIK A.Freudewald GmbH & Co
Hersteller:
Manufacturer: Fabricant:
RHEWA–WAAGENFABRIK August Freudewald GmbH & Co. KG Postfach 10 01 29 D-40801 Mettmann
Typ:
Type:
Type:
940C/ 4 940C/ 10 940C/ 20
Fabriknummer:
Serial number: (to be entered by the customer) Numero de série:
See bottom of indicator
Signature Signature: Signature:
Date: 29.04.2010 Date: Date: