Rez?m G2200 Operator's Manual

3033-001 Revision D January 2018
©NxThera 2018
Page 1 of 90
Operator’s Manual
Rezūm® Generator
Model G2200
WARNING: Carefully read and understand all instructions, indications, warnings, and cautions
in this operator’s manual prior to using the Rezūm Generator. Failure to do so could result in compromised patient safety, patient complications and/or insufficient treatment.
3033-001 Revision D January 2018
©NxThera 2018
Page 2 of 90
This page intentionally left blank.
3033-001 Revision D January 2018
©NxThera 2018
Page 3 of 90
Table of Contents
Chapter 1 About the Rezūm® Generator ........................................................................................... 6
1.1 Trademark and Copyright Information .................................................................................... 7
1.2 Contact Information ................................................................................................................... 7
1.3 Indications for Use/Intended Use and Contraindications ..................................................... 8
1.4 Product Description .................................................................................................................... 9
1.5 Safety .......................................................................................................................................... 10
1.5.1 Dangers, Warnings and Cautions ........................................................................................ 10
1.5.1.1 DANGERS ....................................................................................................................... 11
1.5.1.2 WARNINGS .................................................................................................................... 11
1.5.1.3 CAUTIONS ..................................................................................................................... 13
1.5.2 Potential Adverse Events .................................................................................................... 16
1.5.3 Reported Adverse Events .................................................................................................... 16
1.5.4 Symbols ............................................................................................................................... 17
1.5.4.1 Generator Symbols ....................................................................................................... 17
1.5.4.2 User Interface Symbols ................................................................................................ 20
1.6 Operator Training Requirements ............................................................................................ 21
Chapter 2 Getting Started .................................................................................................................. 22
2.1 Unpacking and Inspecting ....................................................................................................... 23
2.1.1 Generator Controls and Connections ................................................................................. 24
2.1.2 Display Screen Map ............................................................................................................. 27
2.1.3 Audible Tones...................................................................................................................... 28
2.1.4 Volume Control ................................................................................................................... 29
2.1.5 Detachable Parts and Accessories ...................................................................................... 30
Chapter 3 Using the Rezūm Generator ............................................................................................ 31
3.1 Step by Step Instructions ......................................................................................................... 32
3.1.1 Connecting the Power Cord ................................................................................................ 34
3.1.2 Turning on the Rezūm Generator ....................................................................................... 35
3.1.3 Priming the Delivery Device ................................................................................................ 39
3033-001 Revision D January 2018
©NxThera 2018
Page 4 of 90
3.1.4 Performing the Therapy ...................................................................................................... 42
3.1.4.1 Turbo Flush .................................................................................................................... 45
3.1.4.2 Connecting New Delivery Device during treatment session .................................. 45
3.1.5 Turning off the Rezūm Generator ....................................................................................... 46
3.1.6 Rezūm Generator Options Menu Items .............................................................................. 47
3.1.6.1 Enter Patient Info .......................................................................................................... 48
3.1.6.1.1 Using the On-screen Keyboard ............................................................................ 50
3.1.6.2 Drain Bladder ................................................................................................................. 51
3.1.6.3 Replace Saline ............................................................................................................... 51
3.1.6.4 Remove Device.............................................................................................................. 52
3.1.6.5 View Patient Info ........................................................................................................... 53
3.1.6.5.1 Delete Record ......................................................................................................... 54
3.1.6.5.2 View Record ............................................................................................................ 55
3.1.6.5.3 Export Record .......................................................................................................... 56
3.1.6.6 More Options ................................................................................................................ 59
3.1.6.6.1 System Status .......................................................................................................... 59
3.1.6.6.2 Export Logs ............................................................................................................... 64
3.1.6.6.3 Treatment Monitor ................................................................................................. 66
3.1.6.6.4 Treatment Config.................................................................................................... 66
3.1.6.6.5 Servicing .................................................................................................................. 66
Chapter 4 Maintenance and Service ................................................................................................. 67
4.1 Recommended Maintenance and Care ................................................................................. 68
4.1.1 Visual Inspection ................................................................................................................. 69
4.1.2 Cleaning the Rezūm Generator ........................................................................................... 70
4.1.2.1 Recommended Cleaning Products ............................................................................. 70
4.1.2.2 Not Recommended Cleaning Products ..................................................................... 70
4.1.2.3 Cleaning Instructions .................................................................................................... 71
4.1.3 Maintenance Checklist ........................................................................................................ 72
4.2 Authorized Repair Service ....................................................................................................... 73
Chapter 5 Technical Specifications ................................................................................................... 74
3033-001 Revision D January 2018
©NxThera 2018
Page 5 of 90
5.1 Generator Specifications .......................................................................................................... 75
5.2 Operating, Storage, and Transportation Specifications ...................................................... 76
5.3 Electromagnetic Compatibility Requirements ...................................................................... 77
5.4 EN 60601-1-2 Compliance ...................................................................................................... 84
Chapter 6 Troubleshooting................................................................................................................ 85
6.1 Troubleshooting Steps ............................................................................................................. 86
6.2 Error Messages .......................................................................................................................... 87
6.3 Error Message Table ................................................................................................................. 88
6.4 Obtaining technical assistance ................................................................................................ 90
3033-001 Revision D January 2018
©NxThera 2018
Page 6 of 90
Chapter 1 About the Rezūm
®
Generator
Overview
This chapter contains information about the Rezūm Generator including safety. Please make sure you read this entire instruction manual prior to using the Rezūm Generator.
3033-001 Revision D January 2018
©NxThera 2018
Page 7 of 90
1.1 Trademark and Copyright Information
Emergo Europe Prinsessegracht 20 2514 AP The Hague The Netherlands
Rezūm is a registered trademark of the NxThera Corporation.
© NxThera Corporation 2018. All rights reserved.
1.2 Contact Information
NxThera, Inc.
7351 Kirkwood Lane North Suite 138 Maple Grove, MN 55369 USA
www.nxthera.com
Phone: +1-763-515-0404 Toll Free: +1-888-319-9691 Fax: +1-763-515-2085
3033-001 Revision D January 2018
©NxThera 2018
Page 8 of 90
1.3 Indications for Use/Intended Use and Contraindications
The Rezūm generator is intended to be used with the NxThera Rezūm Delivery Device Kit, Model D2201 only. Refer to the Instructions for Use for the Rezūm Delivery Device Kit for the
Indications for Use/Intended Use and Contraindications.
3033-001 Revision D January 2018
©NxThera 2018
Page 9 of 90
1.4 Product Description
The Rezūm generator is designed to heat the RF coil within the hand-held Delivery Device rapidly. The device uses radio frequency (RF) energy to heat water into vapor outside the body.
The vapor is injected into the tissue and rapidly disperses through the interstitial spaces between the tissue cells. The vapor begins to cool and condenses immediately on contact with tissue. The stored heat energy is released, gently denaturing the cell membranes and causing instantaneous cell death.
The denatured tissue is absorbed by the body over time. The vapor condensation process also causes a rapid collapse of blood vessels in the ablation treatment zone, resulting in a bloodless procedure.
The generator rapidly heats and converts sterile water into nearly pure or “dry vapor” at slightly above 100°C. The generator delivers this thermal energy in the form of dry vapor at temperatures through precise vapor emitter openings at the tip of the Vapor Emitter Needle. The rate and time over which the thermal energy in the form of vapor is delivered is monitored and regulated by the generator.
3033-001 Revision D January 2018
©NxThera 2018
Page 10 of 90
1.5 Safety
This section contains important safety information. NxThera requires that you read and understand all warnings, cautions, and the operator’s manual prior to using the Rezūm Generator.
1.5.1 Dangers, Warnings and Cautions
The symbol and signal words shown below identify potential hazard categories. The definition of each category is as follows:
DANGER
This alert identifies hazards that will cause serious personal injury or death.
WARNING
This alert identifies hazards that may cause serious personal injury or death.
CAUTION This alert identifies hazards that may cause minor personal injury, product damage, or property damage.
3033-001 Revision D January 2018
©NxThera 2018
Page 11 of 90
1.5.1.1 DANGERS
WARNING:
A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation. To avoid electrical shock, plug the power cord into a properly wired receptacle, use only the power cord supplied with the generator, and make sure the power cord is in good condition.
WARNING:
After visual inspection, if the generator is damaged or a message is indicated to not use the generator, please contact Customer Service and take the generator out of service.
WARNING:
Before conducting maintenance work, turn the power off and unplug the power cord from the outlet to prevent electric shock.
WARNING:
Do not modify this equipment without authorization of NxThera.
WARNING:
Carefully read and understand all instructions, indications, warnings, and cautions in this operator’s manual prior to using the Rezūm Generator. Failure to do so could result in compromised patient safety, patient complications and/or insufficient treatment.
WARNING:
Do not connect a grounding wire from a grounding stud to a gas pipe or water pipe.
WARNING:
Do not connect to an electrical outlet controlled by a wall switch because the device may be accidentally turned off.
WARNING:
Do not plug power cord into an outlet (or unplug it) with wet hands.
WARNING:
Do not submerge the device in liquids or pour cleaning liquids over, into or onto the generator.
WARNING:
Do not use the generator if it is damaged, is not functioning properly, or fails to meet an electrical safety check. Notify the appropriate personnel to ensure the generator is removed from service and properly repaired.
DANGER: Do not take or use the device in locations where combustible anesthetics or
flammable gases are used or in high-pressure oxygen rooms or inside oxygen tents.
1.5.1.2 WARNINGS
3033-001 Revision D January 2018
©NxThera 2018
Page 12 of 90
WARNINGS CONTINUED
WARNING:
Failure on the part of all responsible individuals, hospitals, or institutions, employing the use of Rezūm generator, to implement the recommended maintenance schedule may cause equipment failure and possible health hazards. The manufacturer does not, in any manner, assume the responsibility for performing the recommended maintenance schedule. The sole responsibility rests with the individuals, hospitals, or institutions utilizing the Rezūm Generator.
WARNING:
If a critical error message is displayed, take the generator out of service and call NxThera Customer Service. Do not attempt to service or maintain the generator while in use with a patient.
WARNING:
If the generator measurement readings or messages seem dubious or abnormal, check the condition of the patient first and stop using the generator.
WARNING:
In the event of power failure, the generator will automatically shut off. Turn the power button off. Please remove the Delivery Device from the patient immediately per instructions in the IFU, 3032-0XX Section 9. Turn on again to restart the generator to begin a new therapy session.
WARNING:
No modification of this equipment is allowed. Do not attempt to service or maintain the generator while in use with a patient.
WARNING:
RF Interference -Known RF sources, such as cell phones, radio or TV stations, and two-way radios, may cause unexpected or adverse operation of this generator. Consult qualified personnel regarding system configuration.
WARNING:
Shock Hazard - Do not open, disassemble, or alter the Rezūm Generator! Failure to observe this warning can result in personal injury or death. Refer maintenance issues to authorized service personnel.
WARNING:
The generator contains magnets in the LCD lid. Avoid close or prolonged contact with electrical devices or devices that have strong magnetic fields.
WARNING:
The generator is not intended to be deployed in settings or situations that promote use by untrained personnel. Operation by untrained personnel can result in injury or death.
WARNING:
The generator needs special precautions regarding Electromagnetic Compatibility (EMC) and needs to be installed and put into service according to the EMC information in Chapter 5 provided in this operator’s manual.
WARNING:
The generator should not be used adjacent to, or stacked with other equipment. If adjacent or stacked use is necessary, test the generator to verify normal operation. Refer to the Electromagnetic Immunity information in Chapter 5.
WARNING:
The Rezūm Generator is equipped with a USB port that is sensitive to ESD that may
potentially result in injury or device failures.
3033-001 Revision D January 2018
©NxThera 2018
Page 13 of 90
WARNINGS CONTINUED
WARNING:
The Rezūm Generator is reusable, but is restricted to a single patient at a time for a therapy session.
WARNING:
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
WARNING:
Use a grounded AC outlet for the power supply and ground this generator.
WARNING:
Use of accessories other than those specified in this document may result in increased emission or decreased immunity of the Rezūm Generator.
WARNING:
Use only NxThera approved and specified parts, accessories, optional parts, consumables, and components.
WARNING:
Use only NxThera specified power cord.
WARNING:
Use with the specified AC voltage and frequency.
WARNING:
When transporting the generator, it is important to position it with the display facing away from the body.
1.5.1.3 CAUTIONS
CAUTION:
After cleaning, allow complete drying before plugging into an outlet.
CAUTION:
Before conducting maintenance work, turn the power OFF and unplug the power cord from the outlet to prevent electric shock.
CAUTION:
Before moving this generator, turn the power OFF, remove all accessories from the patient, and unplug the power cord from the outlet.
CAUTION:
Do not install this generator in the following locations:
Locations where gases and flames are used
Locations where the air includes dust, salt, or sulfur
Locations exposed to prolonged direct sunlight
Locations that vibrate or are subject to sharp impacts Locations near heating equipment Locations where chemicals are stored This generator cannot be used in any room in which noise-generating
apparatuses are used (such as an MRI room, CT room, X-ray room, etc.)
CAUTION:
Do not place anything on this generator.
CAUTION:
Do not soak the generator or accessories in any medical liquid. Also, keep liquids out of the generator and accessories.
3033-001 Revision D January 2018
©NxThera 2018
Page 14 of 90
CAUTIONS CONTINUED
CAUTION:
Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference (RFI), which could affect the performance of this device. Avoid operating the Rezūm Generator near cauterizers, diathermy equipment, FM 2-way radios, or cellular phones. Turn power off to radio, cellular and other like equipment near the Rezūm Generator. Refer to the EMI tables in Chapter 5.
CAUTION:
Exposing the Rezūm Generator to extreme environmental conditions outside of its specified parameters may compromise the ability of the Rezūm Generator to function properly and/or cause the plastic to warp and/or crack.
CAUTION:
Follow your facility’s procedures and applicable regulations when disposing of anything that has been used on patients.
CAUTION:
If the Rezūm Generator is stored in an environment with a temperature below the operating temperature, the unit should be allowed to warm up to the needed operating temperature before using.
CAUTION:
If there is condensation on the generator, dry it thoroughly before turning the power on.
CAUTION:
Input voltage range is 100 to 240V at 50 to 60 Hertz. Verify this voltage matches that of the power outlet.
CAUTION:
Observe the following cautions when connecting this generator with other equipment:
Ensure that the connected equipment is in accordance with the IEC60601-1
or IEC safety standards.
Employ additional protective measures (e.g., additional protective earthing)
as necessary.
CAUTION:
Only NxThera approved equipment and accessories shall be connected to the generator.
CAUTION:
The Rezūm Generator USB port is only intended for use during maintenance by authorized service personnel or to export therapy data.
CAUTION:
The generator conforms to the requirements of the EMC standards (IEC 60601-1­2:2007 and IEC 60601-1-2:2014), so it can be used at the same time as other electrical simulators. However, it may be affected by electrical scalpels and microwave treatment devices and there may be an impact on measurement precision for patients using cardiac pacemakers and the like. Check the operation of this generator during and after use of such equipment and with such patients.
CAUTION:
The Rezūm Generator is intended to be used indoors at a medical facility or
physician office environment only.
3033-001 Revision D January 2018
©NxThera 2018
Page 15 of 90
CAUTIONS CONTINUED
CAUTION:
The Rezūm Generator cables may cause a trip hazard while cables are attached to the generator.
CAUTION:
The Rezūm Generator needs special precautions regarding Electromagnetic compatibility (EMC) and care should be taken in accordance to the EMC information provided in Chapter 5 of this document.
CAUTION
To prevent damage to equipment, do not clean any part of the generator with phenolic compounds. Do not use abrasive or flammable cleaning agents. Do not steam, autoclave, or gas-sterilize the generator.
CAUTION:
Use of portable and mobile RF communications equipment near the Rezūm Generator may affect its operation.
CAUTION:
Using this generator with the air vent blocked could cause a breakdown. Clean this generator with care.
CAUTION:
When any of the following occur, turn the power off, remove all accessories from the patient, and unplug the power cord from the outlet.
There is smoke or a strange odor leaking out of the generator. The generator has been dropped or impacted by an object. Liquid or foreign matter gets inside the generator. If you think the generator may have been damaged.
CAUTION:
When using disinfectant solutions, follow the manufacturer’s directions.
CAUTION:
Input voltage is pre-selected as labeled on the generator
CAUTION:
Please use caution in opening the shipping box and try not to use sharp utility knives and such, as you risk cutting into yourself and/or product.
3033-001 Revision D January 2018
©NxThera 2018
Page 16 of 90
1.5.2 Potential Adverse Events
The following adverse events have not been reported in the clinical trials: erectile dysfunction, pelvic abscess, rectal wall injury, and fistula. Delivering a form of thermal therapy or misuse of the device has potential for producing these adverse effects.
1.5.3 Reported Adverse Events
The types of device related, or procedure related adverse events reported are typical of thermal BPH ablation procedures. There were no clinically significant complications resulting from the treatment.
A summary of the adverse events observed in NxThera clinical studies is presented in the Rezūm Delivery Device Kit Instructions for Use for the Rezūm Delivery Device Kit, 3032-0XX.
3033-001 Revision D January 2018
©NxThera 2018
Page 17 of 90
1.5.4 Symbols
Symbol
Description of the Symbol
Standard
Title of symbol and (reference number)
Indicates on the rating plate that the equipment is suitable for alternating current only; to identify relevant terminals
IEC 60417
Alternating current (5032)
Manufacturer’s catalog number, so that the medical device can be identified
ISO 15223-1:2016
Catalog number (5.1.6)
Electrostatic sensitive device
Caution: Contains parts and assemblies susceptible to damage by electrostatic discharge (ESD)
Contact with generator package should be avoided at low levels of relative humidity, especially if insulating footwear is being worn or the ground/floor is nonconductive. Low levels of relative humidity must in particular be expected on hot, dry summer days and very cold winter days.
IEC 60417
Electrostatic sensitive devices (5134)
CE Marked per the Medical Device Directive 93/42/EEC of the European Union. The Notified Body is SEMKO (ID#0413)
MDD 93/42/EEC
Compliance with MDD 93/42/EEC
Consult accompanying documents
(Printed in blue on generator label)
ISO 7010:2011
Refer to instruction manual/booklet (M002)
Indicates the date when the medical device was manufactured
ISO 15223-1:2016
Date of Manufacture (5.1.3)
The following symbols may appear in this manual, on the generator labeling and/or packaging. Some of the symbols represent standards and compliances associated with the generator and its use.
1.5.4.1 Generator Symbols
3033-001 Revision D January 2018
©NxThera 2018
Page 18 of 90
Symbol
Description of the Symbol
Standard
Title of symbol and (reference number)
Authorized Representative in the European Union
ISO 15223-1:2016
Indicates the Authorized representative in the European Community (5.1.2)
Indicates a medical device that can be broken or damaged if not handled carefully.
The generator package should be handled carefully and should never be tipped over or slung
ISO 15223-1:2016
Fragile, handle with care (5.3.1)
Indicates a medical device that needs to be protected from moisture.
The generator package must be protected from excessive humidity and must accordingly be stored under cover.
ISO 15223-1:2016
Keep dry (5.3.4)
Indicates the medical device manufacturer
ISO 15223-1:2016
Manufacturer (5.1.1)
Manufacturing internal part number reference
Not Applicable
Not Applicable
Model Number
Not Applicable
Not Applicable
Indicates generally elevated, potentially hazardous, levels of non­ionizing radiation, or to indicate equipment or systems e.g. in the medical electrical area that include RF transmitters or that intentionally apply RF electromagnetic energy for diagnosis or treatment.
IEC 60417
Non-Ionizing electromagnetic radiation (RF) (5140)
Indicates disconnection from the mains, at least for mains switches or their positions, and all those cases where safety is involved.
IEC 60417
“OFF” (power)
(5008)
Indicates connection to the mains, at least for mains switches or their positions, and all those cases where safety is involved.
IEC 60417
“ON” (power)
(5007)
3033-001 Revision D January 2018
©NxThera 2018
Page 19 of 90
Symbol
Description of the Symbol
Standard
Title of symbol and (reference number)
US Federal law restricts this device to sale by or on the order of a physician.
FDA Register Vol. 81, No. 115
Prescription device (IIG, page 38919)
IPx0
Protection against Ingress of Solids and Liquids
IEC 60529
Non-protected
Identifies any terminal which is intended for connection to an external conductor for protection against electric shock in case of a fault, or the terminal of a protective earth (ground) electrode.
IEC 60417
Protective earth; protective ground (5019)
Indicates the range of humidity to which the medical device can be safely exposed.
Relative Humidity limits (non­condensing)
ISO 15223-1:2016
Humidity limitation (5.3.8)
Indicates the manufacturer's serial number so that a specific medical device can be identified.
ISO 15223-1:2016
Serial number (5.1.7)
Temperature limitations
This symbol will specify an upper ("...°C max.") and lower ("...°C min.") temperature limit.
ISO 15223-1:2016
Temperature limit (5.3.7)
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.
BS EN 50419:2006
WEEE wheeled bin Identifies a type BF applied part complying with IEC 60601-1.
IEC 60417:2002
ANSI/AAMI/IEC TIR60878:2003
IEC 60601-1
Type BF applied part (5333)
Universal Serial Bus (USB) port connector
Not Applicable
Not Applicable
3033-001 Revision D January 2018
©NxThera 2018
Page 20 of 90
Symbol
Description of the Symbol
Standard
Title of symbol and (reference number)
Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot, for a variety of reasons, be presented on the medical device itself.
ISO 15223-1:2016
Caution (5.4.4)
Medical Equipment as to electrical shock, fire and mechanical hazards only in accordance with ANSI/AAMI ES 60601-1 AMD 1 (2012), CAN/CSA C22.2 No 60601-1 (2014)
ANSI/AAMI ES 60601­1 AMD 1 (2012), CAN/CSA C22.2 No 60601-1 (2014)
Medical Equipment recognized by Underwriters Laboratory Inc.
1.5.4.2 User Interface Symbols
Symbol
Description
Priming the Delivery Device
Perform a pre-treatment vapor cycle
Back to previous screen
Complete
Confirmation message
Continue
Critical error message
Displayed for errors that are not of critical severity.
Export
Table 1: Generator Symbols
3033-001 Revision D January 2018
©NxThera 2018
Page 21 of 90
Symbol
Description
Generator question that requires user response
Home screen menu Navigation buttons for back, forward, up and down
Options Menu
Volume Control
Table 2: User Interface Symbols
1.6 Operator Training Requirements
WARNING: The generator is not intended to be deployed in settings or situations that promote use by untrained personnel. Operation by untrained personnel can result in injury or death.
Persons authorized to operate the generator must have all the following minimum training.
Training as required by state, province, or country regulations. Training on operation and use of the generator. Additional training as required by a physician or Medical Director. A thorough understanding of the procedures in this manual.
3033-001 Revision D January 2018
©NxThera 2018
Page 22 of 90
Chapter 2 Getting Started
Overview
This chapter contains information on how to get started with your Rezūm Generator.
3033-001 Revision D January 2018
©NxThera 2018
Page 23 of 90
2.1 Unpacking and Inspecting
CAUTION: Please use caution in opening the shipping box and try not to use sharp utility knives and such, as you risk cutting into yourself and/or product.
Every attempt is made to ensure your order is accurate and complete. However, to be sure that your order is correct, verify the contents of the box against your packing slip.
The Rezūm Generator is designed for simplicity of operation and set-up and requires minimal assembly. The following items are included in the Rezūm Generator box:
One (1) Rezūm Generator
One (1) Operator’s Manual
One (1) Power cord
1. Carefully inspect each item as it is unpacked for any signs of damage that may have occurred during
shipment.
2. Check the components according to the packing list.
3. Check for any damage or defects. Do not attempt to set up the Rezūm Generator if anything is
damaged or defective. Contact Customer Service immediately if anything is damaged or defective.
3033-001 Revision D January 2018
©NxThera 2018
Page 24 of 90
2.1.1 Generator Controls and Connections
WARNING: The generator contains magnets in the LCD lid. Avoid close or prolonged contact with electrical devices or devices that have strong magnetic fields.
WARNING: The Rezūm Generator is equipped with a USB port that is sensitive to ESD that may
potentially result in injury or device failures.
WARNING: Do not connect a grounding wire from a grounding stud to a gas pipe or water pipe.
CAUTION: Using this generator with the air vent blocked could cause a breakdown. Clean this generator with care.
CAUTION: Only NxThera approved equipment and accessories shall be connected to the generator.
CAUTION: The Rezūm Generator USB port is only intended for use during maintenance by authorized service personnel or to download therapy data.
The following figures and table explain the controls, connections, and their function.
3033-001 Revision D January 2018
©NxThera 2018
Page 25 of 90
ID
Item
Description
A
Display Screen
Touch screen display to provide system feedback to the user B
Syringe / Syringe Cradle
Holds water for vapor treatment
C
Electrical cable port
The electrical cable is the RF energy line and the connections for the switches and thermocouples
D
Roller pump
Delivers saline during procedure
E
Power Indicator
Displays system status
F
Power switch
Turns system on / off
G
USB port
Allows data from system to be exported onto USB device
Figure 1: Front and Top of Generator
Table 3: Front and Top of Generator Description Table
3033-001 Revision D January 2018
©NxThera 2018
Page 26 of 90
ID
Item
Description
I
Grounding stud
Grounding stud used for grounding product (Required in Europe)
J
Air vent
Outgoing air vent
K
Product Label
Provides information about the generator
L
Fuse box
Holds generator fuses
M
Power cord plug
Connection plug for electrical power cord
N
Lid
Cover to protect display screen, syringe and pressure sensor ports
O
Handle
Use to transport the device by hand
P
Air vent and speaker
Incoming air vent (both sides) and speaker (left side only)
Q
Rubber feet (on bottom and side)
Allows product to be stored on base or bottom end
J I K M L N P
Q
O
Figure 2: Back of Generator
Figure 3: Side of Generator
Table 4: Back and Side of Generator Description Table
3033-001 Revision D January 2018
©NxThera 2018
Page 27 of 90
2.1.2 Display Screen Map
A-Title Bar
B-Information Screen
C-Lower Toolbar
Item
Description
A-Title Bar
Brief title of the screen displayed
B-Information
Screen
Main area for screen information and error messages will be shown to the user. Replace saline button is also available on the saline bag.
C-Lower
Toolbar
The bottom of all screens, except the start-up screens, will include the same lower bar. This will have three features on it: volume adjustment, product logo, and the Options menu button (if applicable).
The Rezūm Generator is equipped with a color touchscreen that can be viewed up to 8 feet away from the generator. The display allows interaction with the generator using screen buttons icons and menus with the touch of a finger with or without latex gloves.
Figure 4: Display Screen Map
Table 5: Display Screen Map
3033-001 Revision D January 2018
©NxThera 2018
Page 28 of 90
2.1.3 Audible Tones
Tone Name
Tone Description
1. Start-up
A musical tone is played during generator power on sequence.
2. Treatment Warning
A single quick tone played when there is a warning after treatingrunning low on treatments, running low on treatment time, and other alerts that may require user intervention.
3. Partial Treatment Tone
Tone emitted when the delivery device vapor activation button has been deactivated prior to the completion of a treatment.
4. Treatment Ready
Played when the system is ready to perform a treatment
5. Treatment and Priming
Beep tone is repeated once per second while performing Priming and Treatment operations.
6. Success
Tone emitted when a full treatment has been administered, after successful priming, and as feedback while adjusting the volume.
7. Error Message
Two quick tones in succession are played whenever the generator displays an error message on the generator screen.
8. Critical Error Message
A single high-pitched tone is played when the generator encounters a critical error. In addition, three quick tones in succession are played whenever the generator displays a critical error message on the generator screen.
9. Delivery Device Disable Tone
A high-pitched tone followed by a low-pitched tone played when the delivery device is disabled.
The generator emits different audible tones to indicate to the user different events. These tones vary with the type of message and its content. The description of each tone is explained in Table 6.
Table 6: Tone Description
3033-001 Revision D January 2018
©NxThera 2018
Page 29 of 90
2.1.4 Volume Control
The generator has an on-screen volume control as depicted in Figure 5.
Figure 5: Volume Control
Touch the minus symbol to decrease or silence the volume and touch the plus symbol to
increase the volume. A tone will be emitted upon each button press.
The Critical Error and Delivery Device Disabled alarm tones cannot be silenced by the
volume control.
The volume resets to the default when the generator is turned off and on again.
Set the volume loud enough to be heard adequately in the actual use environment.
3033-001 Revision D January 2018
©NxThera 2018
Page 30 of 90
2.1.5 Detachable Parts and Accessories
Model Number
Description, Function
Type
1519-001
Power cord
3.05 meter hospital grade power cord North American, 15 amp
Reusable
1519-002
Power cord
3.05 meter hospital grade power cord Continental Europe, 10 amp
Reusable
1519-003
Power cord
2.50 meter hospital grade power cord Swiss, 10 amp
Reusable
Model Number
Description, Function
Type
D2201
Delivery Device,
Delivers the vapor into the tissue
Disposable, single use
WARNING: Use only NxThera approved and specified parts, accessories, optional parts, consumables, and components.
CAUTION: The Rezūm Generator cables may cause a trip hazard while cables are attached to the generator.
The Rezūm Generator has the following detachable parts and accessories:
Detachable Parts – Supplied with the Rezūm Generator
Accessories – Supplied Separately
Table 7: Detachable Parts
Table 8: Accessories
3033-001 Revision D January 2018
©NxThera 2018
Page 31 of 90
Chapter 3 Using the Rezūm Generator
Overview
This chapter provides step by step instructions and sequence of operation for the Rezūm Generator.
3033-001 Revision D January 2018
©NxThera 2018
Page 32 of 90
3.1 Step by Step Instructions
This section contains step by step instruction on how to connect the power cord, turn on and off the generator, prime the delivery device, perform treatment, and use the option menu.
WARNING: When transporting the generator, it is important to position it with the display facing away from the body.
CAUTION: Follow your facility’s procedures and applicable regulations when disposing of
anything that has been used on patients.
CAUTION: Do not install this generator in the following locations:
Locations where gases and flames are used
Locations where the air includes dust, salt, or sulfur
Locations exposed to prolonged direct sunlight
Locations that vibrate or are subject to sharp impacts Locations near heating equipment Locations where chemicals are stored
This generator cannot be used in any room in which noise-generating apparatuses are used (such as an MRI room, CT room, X-ray room, etc.)
CAUTION: Do not place anything on this generator.
CAUTION: Observe the following cautions when connecting this generator with other
equipment:
Ensure that the connected equipment is in accordance with the IEC60601-1 or IEC safety
standards.
Employ additional protective measures (e.g., additional protective earthing) as necessary.
CAUTION: Exposing the Rezūm Generator to extreme environmental conditions outside of its specified parameters may compromise the ability of the Rezūm Generator to function properly and/or cause the plastic to warp and/or crack.
CAUTION: The Rezūm Generator cables may cause a trip hazard while cables are attached to the generator.
CAUTION: If the Rezūm Generator is stored in an environment with a temperature below the operating temperature, the unit should be allowed to warm up to the needed operating temperature before using.
3033-001 Revision D January 2018
©NxThera 2018
Page 33 of 90
CAUTION: Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference (RFI), which could affect the performance of this device. Avoid operating the Rezūm Generator near cauterizers, diathermy equipment, FM 2-way radios, or cellular phones. Turn power off to radio, cellular and other like equipment near the Rezūm Generator. Refer to the EMI tables in Chapter 5.
CAUTION: The Rezūm Generator is intended to be used indoors at a medical facility or physician office environment only.
3033-001 Revision D January 2018
©NxThera 2018
Page 34 of 90
3.1.1 Connecting the Power Cord
WARNING: Do not connect to an electrical outlet controlled by a wall switch because the generator may be accidentally turned off.
WARNING: Do not plug power cord into an outlet (or unplug it) with wet hands.
WARNING: Use only NxThera specified power cord.
WARNING: To avoid the risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
WARNING: A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation. To avoid electrical shock, plug the power cord into a properly wired receptacle, use only the power cord supplied with the generator, and make sure the power cord is in good condition.
WARNING: Use a grounded AC outlet for the power supply and ground this generator.
WARNING: Use only NxThera approved and specified parts, accessories, optional parts,
consumables, and components.
WARNING: Use with the specified AC voltage and frequency.
1. Make sure that the AC outlet is properly grounded and supplies the specified voltage
and frequency.
2. Connect the female connector end of the power cord to the AC power connector on the
back of the generator.
3. Plug the male connector end of the power cord into a properly grounded AC power
outlet.
3033-001 Revision D January 2018
©NxThera 2018
Page 35 of 90
3.1.2 Turning on the Rezūm Generator
Power button
I
O
DANGER: Do not take or use the device in locations where combustible anesthetics or flammable gases are used or in high-pressure oxygen rooms or inside oxygen tents.
WARNING: The Rezūm Generator is reusable, but is restricted to a single patient at a time for a therapy session.
CAUTION: If there is condensation on the generator, dry it thoroughly before turning the power on.
CAUTION: When any of the following occur, turn the power off, remove all accessories from the patient, and unplug the power cord from the outlet.
There is smoke or a strange odor leaking out of the generator. The generator has been dropped or impacted by an object. Liquid or foreign matter gets inside the generator. If you think the generator may have been damaged.
To turn ON:
1. Open the lid of the generator to show the display screen and make sure it is fully open.
2. Turn on the generator by pushing the top of the Power button located on the front of the
generator as depicted in Figure 6.
Figure 6: Power button
3033-001 Revision D January 2018
©NxThera 2018
Page 36 of 90
3. While the generator is powering up, it initially displays two start-up screens and a test
screen.
WARNING: In the event of power failure, the generator will automatically shut off. Turn the power button off. Please remove the Delivery Device from the patient immediately per instructions in the IFU, 3032-0XX Section 9. Turn on again to restart the generator to begin a new therapy session.
Figure 7: NxThera Start-up Screens
3033-001 Revision D January 2018
©NxThera 2018
Page 37 of 90
Figure 8: NxThera Test Screen
4. It will automatically run the Start-up Diagnostics. During this time the test screen is
displayed, and a bar and text will indicate status of the tests on the screen.
3033-001 Revision D January 2018
©NxThera 2018
Page 38 of 90
5. After the Start-up Diagnostics are completed, the generator will display the Delivery Device
Setup screen.
Figure 9: Deliver Device Setup
After all connections are properly made, a message will be displayed for 5 seconds letting the user know that items have been connected and then the Priming screen will be displayed.
Figure 10: Connection Message
3033-001 Revision D January 2018
©NxThera 2018
Page 39 of 90
3.1.3 Priming the Delivery Device
CAUTION: When any of the following occur, turn the power off, remove all accessories from the patient, and unplug the power cord from the outlet.
There is smoke or a strange odor leaking out of the generator. The generator has been dropped or impacted by an object. Liquid or foreign matter gets inside the generator. If you think the generator may have been damaged.
Figure 11: Ready for Priming
When the operator starts priming the Delivery Device, a progress bar will be initiated and displayed. The text will change, and a tone will be generated to indicate that the operation is in progress. Priming is initiated by deploying the needle and holding the Delivery Device vapor activation button for the priming duration, which lasts approximately 30 seconds. A message will be displayed indicating when Priming is complete.
If the vapor activation trigger is activated while the needle is being deployed, priming, pre­treatment, or treatment operation will not be initiated until the vapor trigger is released and reengaged.
3033-001 Revision D January 2018
©NxThera 2018
Page 40 of 90
Figure 12: Priming Screen
Figure 13: Priming Complete
The Setup Complete will be displayed when the Delivery Device has been successfully primed. There will also be an audible tone that indicates success. Pre-treatment vapor cycle shall be performed prior to treatment.
3033-001 Revision D January 2018
©NxThera 2018
Page 41 of 90
Figure 14: Successful Primed Screen
If the device is not properly primed, an error message will be displayed to the user. Upon resolution of the issue and closure of the error message, the screen will display the Priming Screen and the steps should be repeated.
3033-001 Revision D January 2018
©NxThera 2018
Page 42 of 90
3.1.4 Performing the Therapy
WARNING: If the generator measurement readings or messages seem dubious or abnormal, check the condition of the patient first and stop using the generator.
CAUTION: When any of the following occur, turn the power off, remove all accessories from the patient, and unplug the power cord from the outlet.
There is smoke or a strange odor leaking out of the generator. The generator has been dropped or impacted by an object. Liquid or foreign matter gets inside the generator. If you think the generator may have been damaged.
After setup is complete, perform a Pre-Treatment Vapor Cycle immediately prior to procedure to initiate treatment stage.
3033-001 Revision D January 2018
©NxThera 2018
Page 43 of 90
After performing a pre-treatment vapor cycle, a screen will be displayed indicating the generator is ready for treatment.
Figure 15: Ready for Treatment Screen
A treatment is initiated by activating flush, deploying the needle, and pulling in and holding the vapor activation button. When a treatment is in progress, the time clock will count in whole seconds to the maximum preset treatment time. After treatment time has elapsed, the generator will automatically end the treatment and will once again be ready for treatment. Releasing and reactivating the vapor activation button will initiate another treatment after the required rest period has elapsed.
Figure 16: Treatment in Progress Screen
3033-001 Revision D January 2018
©NxThera 2018
Page 44 of 90
If at any time the generator is not ready to perform a treatment (e.g. rest period between treatments is in effect, etc.) the screen specified Waiting as depicted below and is grayed out. The generator will automatically be ready for treatment (Figure 15) when the pending condition is resolved. Monitor the ALERTS message area during treatment and take action when specified to do so.
Figure 17: Pending Ready Indicator on screen
3033-001 Revision D January 2018
©NxThera 2018
Page 45 of 90
3.1.4.1 Turbo Flush
If visualization becomes cloudy during the procedure, Turbo Flush can be activated to improve the visualization by increasing the saline flow rate. To activate Turbo Flush, press twice and hold vapor release button. Treatments will not be performed during this mode.
When visualization has been cleared, turn off Turbo Flush by releasing the button.
3.1.4.2 Connecting New Delivery Device during treatment session
The generator can detect if a new delivery device has been connected to the generator during a therapy session.
If a new Delivery Device is connected during the therapy session, select New to create a new patient record or Continue to continue with current record.
Figure 18: New Delivery Device
3033-001 Revision D January 2018
©NxThera 2018
Page 46 of 90
3.1.5 Turning off the Rezūm Generator
Power button
I
O
In the event of power failure, the generator will automatically shut off. Turn the power button off. Please remove the Delivery Device from the patient immediately per instructions in the IFU, 3032-0XX Section 9. Turn on again to restart the generator to begin a new therapy session.
CAUTION: Before moving this generator, turn the power OFF, remove all accessories from the patient, and unplug the power cord from the outlet.
1. Turn off the generator by pushing the bottom Power button located on the front of the
generator.
Figure 19: Power button
Note: In the event of loss of power, a new patient record is created. Previously completed treatments will be saved in the prior patient record.
3033-001 Revision D January 2018
©NxThera 2018
Page 47 of 90
3.1.6 Rezūm Generator Options Menu Items
To configure the generator and setup before the therapy session begins, select the Options button on the lower toolbar.
Figure 20: Options Menu button
The following options (if applicable) are available and described in further detail in the sections below:
Figure 21: Options Menu choices
3033-001 Revision D January 2018
©NxThera 2018
Page 48 of 90
3.1.6.1 Enter Patient Info
Patient information may be entered for reference or future retrieval at any time during or after the therapy session. When a new Delivery Device is connected, a blank patient record is automatically created. The Patient Info screen allows data entry. The data is saved whenever the cursor enters another field or when you leave the screen.
The Doctor field will allow for selection from a drop-down list of previous entries.
Figure 22: Patient Information Screen
3033-001 Revision D January 2018
©NxThera 2018
Page 49 of 90
Field
Description
First Name
Field can store up to 50 characters.
Field can store alphanumeric and symbols.
Field can store upper or lower-case characters.
Last Name
Field can store up to 50 characters.
Field can store alphanumeric and symbols.
Field can store upper or lower-case characters.
Birth Date
Note: The device is for use with adult patients only.
Consists of pull down menus for day, month, and year. There is an option
in each pull down for an unknown value – ‘?’
Months will be displayed using their equivalent three letter abbreviations.
ID #
Field can store up to 50 characters.
Field can store alphanumeric and symbols.
Field can store upper or lower-case characters.
Study ID
Field can store up to 50 characters.
Field can store alphanumeric and symbols.
Field can store upper or lower-case characters.
Doctor
Field can store up to 50 characters
Field can store alphanumeric and symbols.
Field can store upper or lower-case characters.
Notes
Field can store up to 2000 characters
Field can store alphanumeric and symbols.
Field can store upper or lower-case characters.
Table 9: Patient Information Fields and Descriptions
3033-001 Revision D January 2018
©NxThera 2018
Page 50 of 90
3.1.6.1.1 Using the On-screen Keyboard
Button
Description
Tab
To move from field to field, select the Tab button.
Shift
To switch between upper case and lower case, toggle using the Shift button.
Back Arrow
To delete text, touch the space after the character you’d like to delete to move
the cursor to that position. Press the back arrow.
Enter
To enter a selection, select Enter.
123
The 123 button allows access to the full Qwerty keyboard. Select this button to show all available keyboard alpha numeric characters.
Space
To enter a space, select space.
âéÖ
To enter additional alphabet characters for different languages.
To enter textual information from single words to paragraphs using the touchscreen, an On­screen QWERTY keyboard is automatically provided on screens that allow data entry. Touch the screen to place the cursor in a field.
Table 10: QWERTY keyboard key descriptions
Figure 23: On-Screen Keyboard Layout
3033-001 Revision D January 2018
©NxThera 2018
Page 51 of 90
3.1.6.2 Drain Bladder
When 750ml of saline has been used, a message will be displayed to the physician to drain the bladder. When this has occurred, select from the Options Menu, Drain Bladder. A “Confirm
Bladder Drain” dialog box shall appear when Drain Bladder is selected. Select Confirm to
confirm the physician has done so.
Figure 24: Drain Bladder Confirmation
3.1.6.3 Replace Saline
When saline is replaced, select the Replace Saline from the Options Menu and select the appropriate size saline from the available options. The saline source volume will be set to the previously selected value upon power up of the generator.
Figure 25: Replace Saline
3033-001 Revision D January 2018
©NxThera 2018
Page 52 of 90
3.1.6.4 Remove Device
From the Options Menu, select Remove Device. A “Confirm Syringe Release” dialog box shall appear when Remove Device item is selected. If the user selects “Release”, then the syringe shall be released. If the user selects “Cancel”, no action shall be taken.
Figure 26: Remove Device
If the Delivery Device is primed and a release syringe operation occurs, then the Delivery Device shall require re-priming before treatments can be resumed.
After the device is removed, patient information will be displayed. From this screen patient information, procedure summary and options to continue, complete or export are available.
Figure 27: Patient Information
3033-001 Revision D January 2018
©NxThera 2018
Page 53 of 90
3.1.6.5 View Patient Info
This option is used to access all stored patient records. Password protection is available via the Set Records Password button in the System Status screen to ensure patient privacy is maintained and helps comply with HIPAA. In this screen you may select a patient record stored within the system for deleting, viewing and editing.
Figure 28: List of Patient Records Screen
Patient records can be selected by pressing the desired record row or Select All button. They can be deselected by pressing a selected row or the Deselect All button. The box will be highlighted blue when selected and gray when deselected. Refer to the following sections on how to delete, view, or export records.
Figure 29: Select Patient Records
3033-001 Revision D January 2018
©NxThera 2018
Page 54 of 90
3.1.6.5.1 Delete Record
The device will store up to a maximum of 1000 records. Once the generator reaches it maximum record capacity, it is necessary to export the records if desired and delete the record(s) in order to perform another therapy session. A password, if one has been established, shall be required to delete patient therapy records. The generator does not automatically delete records. To delete one or more records, select the desired patient records to be deleted by pressing the desired record row or Select All button. Next, select the Delete button at the bottom of the screen. You will be prompted with a “Delete selected record(s)?” dialog box to confirm deletion.
Figure 30: Delete Patient Records
If sure, select the “Yes” button to permanently delete the selected records. You must then cycle power on the generator before a new record can be created and therapy session started.
3033-001 Revision D January 2018
©NxThera 2018
Page 55 of 90
3.1.6.5.2 View Record
The View Record screen will display the selected patient record. This screen is not editable. You may select the Edit Patient Info button at the bottom of the screen to edit the patient data via the Patient Info screen. You can also export the displayed record by selecting the Export Record button.
Figure 31: View Record Screen
3033-001 Revision D January 2018
©NxThera 2018
Page 56 of 90
3.1.6.5.3 Export Record
CAUTION: Only NxThera approved equipment and accessories shall be connected to the generator.
CAUTION: The Rezūm Generator USB port is only intended for use during maintenance by authorized service personnel or to export treatment data.
This option is used to export selected records. Password protection is available via the Set Records Password button in the System Status screen to ensure patient privacy is maintained. Therapy and patient information can be exported to a USB memory stick.
Touch the desired folder to select the location on the USB drive to export the records. Touch Save to export the records.
Figure 32: Export Patient Records
Figure 33: Export Patient Records Success
3033-001 Revision D January 2018
©NxThera 2018
Page 57 of 90
When the records are properly transferred to the USB memory stick, a confirmation message will be displayed on the screen.
Figure 34: Export Patient Records Success
3033-001 Revision D January 2018
©NxThera 2018
Page 58 of 90
The records are saved in both a .csv and .txt format. The date/time created and patient ID plus
Patient Id
Serial Number
Date/Time
Treatment Time
123456
D2200-1001
16-09-17 12:29
9
123456
D2200-1001
16-09-17 12:29
9
123456
D2200-1001
16-09-17 12:30
9
Patient Information Area
Note Area
Treatment Information Area
the patient first, last name, if available, are saved as the file name. The .txt file contains all the user-viewable information stored in the record (Figure 35). The .csv file contains details of individual treatments stored in that record (Figure 36). Specifically, each comma-separated line of data contains: patient ID, serial number of the Delivery Device, treatment start date/time stamp, treatment duration.
Figure 35: Example of Exported Patient .txt File
Figure 36: Example of Exported Patient .csv File
3033-001 Revision D January 2018
©NxThera 2018
Page 59 of 90
3.1.6.6 More Options
From the Options menu, select More Options. The More Options screen allows the 5 options that can be selected.
Figure 37: More Options Screen
3.1.6.6.1 System Status
The System Status screen contains information.
Information on the generator and Delivery Device internal device identifiers. Software versions Ability to set date and time Ability to set language Ability to set records password
Figure 38: System Status
3033-001 Revision D January 2018
©NxThera 2018
Page 60 of 90
3.1.6.6.1.1 Set Date and Time
The Select New Date and Time screen allows the date and time of generator to be updated from factory defaults for time zone or daylight-saving time changes. The generator does not automatically adjust for daylight savings time changes. Valid dates are between 1900 and the current year.
Figure 39: Set Date and Time
3.1.6.6.1.2 Set Language
The Set Language screen allows the language to be updated from English to preloaded language options and desired number format. These settings do not change when powering on and off the generator.
Figure 40: Language Settings
3033-001 Revision D January 2018
©NxThera 2018
Page 61 of 90
Select the desired language from the dropdown language list and click OK to change the language from English to desired language.
Figure 41: Language Selections
3.1.6.6.1.3 Set Records Password
The Set Records Password screen allows password protection to ensure patient privacy is maintained. Passwords can be changed via the Set Records Password screen. If a password is forgotten, contact NxThera to reset.
After the password has been successfully changed or cleared, a Success confirmation is displayed. Select OK to close.
Figure 42: Set Records Password
3033-001 Revision D January 2018
©NxThera 2018
Page 62 of 90
Figure 43: Set Records Password Success
Figure 44: Passwords cleared successfully
3033-001 Revision D January 2018
©NxThera 2018
Page 63 of 90
If passwords do not match, an error will be displayed indicating the passwords do not match.
Figure 45: Passwords do not match
3033-001 Revision D January 2018
©NxThera 2018
Page 64 of 90
3.1.6.6.2 Export Logs
CAUTION: Only NxThera approved equipment and accessories shall be connected to the generator.
CAUTION: The Rezūm Generator USB port is only intended for use during maintenance by authorized service personnel or to export treatment data.
Encrypted log files can be exported to a USB drive for use by NxThera service personal only.
Select files to export from the list of options and Click OK.
Figure 46: Select Files to Export
Figure 47: Reading Files
3033-001 Revision D January 2018
©NxThera 2018
Page 65 of 90
Select export path and select Save.
Figure 48: Reading Files
After saving the files, the generator will build the archive, export the files, and prompt when it is successful.
Figure 49: Building archive
3033-001 Revision D January 2018
©NxThera 2018
Page 66 of 90
Figure 50: Exporting file
Figure 51: Export Successful
3.1.6.6.3 Treatment Monitor
This screen is password protected and accessed by NxThera service personal only.
3.1.6.6.4 Treatment Config
This screen is password protected and accessed by NxThera service personal only.
3.1.6.6.5 Servicing
This screen is password protected and accessed by NxThera service personal only.
3033-001 Revision D January 2018
©NxThera 2018
Page 67 of 90
Chapter 4 Maintenance and Service
Overview
Proper maintenance of the Rezūm Generator is very simple, yet it is an important factor in its reliability. This section describes the maintenance and service required for the generator.
WARNING: No modification of this equipment is allowed. Do not attempt to service or maintain the generator while in use with a patient.
WARNING: Do not modify this equipment without authorization of NxThera.
WARNING: If this equipment is modified with authorization from NxThera, appropriate
inspection and testing must be conducted to ensure continued safe use of the equipment.
CAUTION: When any of the following occur, turn the power off, remove all accessories from the patient, and unplug the power cord from the outlet.
There is smoke or a strange odor leaking out of the generator. The generator has been dropped or impacted by an object. Liquid or foreign matter gets inside the generator. If you think the generator may have been damaged.
3033-001 Revision D January 2018
©NxThera 2018
Page 68 of 90
4.1 Recommended Maintenance and Care
WARNING: Failure on the part of all responsible individuals, hospitals, or institutions, employing the use of Rezūm Generator, to implement the recommended maintenance schedule may cause equipment failure and possible health hazards. The manufacturer does not, in any manner, assume the responsibility for performing the recommended maintenance schedule. The sole responsibility rests with the individuals, hospitals, or institutions utilizing the Rezūm Generator.
WARNING: Do not submerge the device in liquids or pour cleaning liquids over, into or onto the generator.
WARNING: Before conducting maintenance work, turn the power OFF and unplug the power cord from the outlet to prevent electric shock.
CAUTION: To prevent damage to generator, do not clean any part of the device with phenolic compounds. Do not use abrasive or flammable cleaning agents. Do not steam, autoclave, or gas-sterilize the generator.
CAUTION: Using this generator with the ventilation ports blocked could cause a breakdown. Clean this generator with care.
CAUTION: Only NxThera approved equipment and accessories shall be connected to the generator.
CAUTION: The Rezūm Generator USB port is only intended for use during maintenance by authorized service personnel or to export treatment data.
To ensure the Rezūm Generator is always functional when required, NxThera recommends performing the following maintenance activities:
Performing a Visual Inspection Cleaning the Rezūm Generator Maintenance per checklist in this section
It is important that the generator is stored at the operating temperature range if it is expected to be used. See Technical Specifications for temperature specifications.
The Rezūm Generator requires no calibration.
3033-001 Revision D January 2018
©NxThera 2018
Page 69 of 90
4.1.1 Visual Inspection
WARNING: After the visual inspection, if the device is damaged or a message indicating the generator is not to be used, take the generator out of service and call NxThera Customer service.
The generator should be carefully inspected prior to installation, use, and each time the equipment is serviced.
Carefully inspect the generator case for stress or physical damage.
Inspect all external connections for loose connectors.
Inspect all external cables for damage or cracking.
Inspect the display for marks, scratches, or other damage.
Verify that the Product label on the device is clearly legible and present.
3033-001 Revision D January 2018
©NxThera 2018
Page 70 of 90
4.1.2 Cleaning the Rezūm Generator
It is recommended that the generator be inspected after each use according to the Checklist in this manual and cleaned when appropriate. Listed below are recommendations for cleaning the generator. The generator does not need to be sterilized before or after use.
WARNING: Do not submerge the device in liquids or pour cleaning liquids over, into or onto the generator.
CAUTION: After cleaning, allow complete drying before plugging into an outlet by wiping with a dry, soft cloth.
CAUTION: Do not soak the generator or accessories in any medical liquid. Also, keep liquids out of the generator and accessories.
CAUTION: When using disinfectant solutions, follow the manufacturer’s directions.
CAUTION: Using this generator with the air vent blocked could cause a breakdown. Clean this
generator with care.
CAUTION: To prevent damage to equipment, do not clean any part of the generator with phenolic compounds. Do not use abrasive or flammable cleaning agents. Do not steam, autoclave, or gas-sterilize the generator.
4.1.2.1 Recommended Cleaning Products
The following cleaning products may be used to clean the exterior surfaces of the generator:
Water 70% Isopropyl Alcohol Super Sani-Cloth® Germicidal Disposable Wipes by PDI only Cidex®
4.1.2.2 Not Recommended Cleaning Products
Do not use abrasive cleaners or strong solvents such as acetone or acetone-based cleaners. Do not mix disinfecting solutions (such as bleach and ammonia) as hazardous gases may
occur
Do not clean electrical contacts or connectors with bleach.
3033-001 Revision D January 2018
©NxThera 2018
Page 71 of 90
4.1.2.3 Cleaning Instructions
1. Before cleaning the generator, turn the generator off and disconnect the power cord.
2. To remove any foreign material and fluid (e.g. dust, paper, etc.), wipe thoroughly with a soft
cloth lightly dampened with water or 70% isopropyl alcohol. Super Sani-Cloth® wipes and Cidex® may be used per manufacturer instructions.
To prevent scratching the display, the use of a soft cloth is recommended.
3. When cleaning, do not immerse.
4. Wring any excess moisture from the cloth before and during cleaning.
5. Avoid pouring fluids on the generator, and do not allow fluids to penetrate the exterior
surfaces of the generator.
6. To dry the generator after cleaning, wipe with a dry, soft cloth.
3033-001 Revision D January 2018
©NxThera 2018
Page 72 of 90
4.1.3 Maintenance Checklist
Visual Inspection
WARNING: After the visual inspection, if the generator is damaged or a message indicates to not use, take the generator out of service and call NxThera Customer Service.
Carefully inspect the generator case for stress or physical damage.
Inspect all external connections for loose connectors.
Inspect all external cables for damage or cracking.
Inspect the display for marks, scratches, or other damage.
Verify that the Safety label on the device is clearly legible and present.
Operating Test
WARNING: If a Critical Error message is displayed, take the generator out of service and call NxThera Customer Service.
Set-up generator and turn on power to check the start-up diagnostics.
Maintenance activities involve verifying operation and safety.
Maintenance should always be performed by the Customer at least once every 12 months. The following checklist is recommended to be utilized when checking the generator:
Table 11: Maintenance Checklist
3033-001 Revision D January 2018
©NxThera 2018
Page 73 of 90
4.2 Authorized Repair Service
WARNING: Shock Hazard - Do not open, disassemble, or alter the Rezūm Generator! Failure to observe this warning can result in personal injury or death. Refer maintenance issues to authorized service personnel.
WARNING: Do not use the generator if it is damaged, is not functioning properly, or fails to meet an electrical safety check. Notify the appropriate personnel to ensure the generator is removed from service and properly repaired.
The generator has no user-serviceable internal components. Try to resolve any maintenance issues with the generator by using the Troubleshooting Table presented in Section 6, Troubleshooting. If you are unable to resolve the problem, contact Customer Service.
The warranty will be void upon unauthorized disassembly or service of the Rezūm Generator.
3033-001 Revision D January 2018
©NxThera 2018
Page 74 of 90
Chapter 5 Technical Specifications
Overview
This chapter contains specifications for the Rezūm Generator and EMC information.
3033-001 Revision D January 2018
©NxThera 2018
Page 75 of 90
5.1 Generator Specifications
Description
Specification
Protection against Electric Shock
Class I Equipment (generator)
Model Number
G2200
100-240 VAC, 50-60Hz
Power Input
10 Amps maximum
External Fuses
Two, 10AH-250V, 5x20mm
Mode of Operation
Continuous Operation
System Control
Provides controlled flow of water vapor at ambient temperatures below 25˚C
Case dimensions
23L x16W x9H inches
Weight
50 pounds or less (generator only)
Power cable length
9 feet
Applied Parts protection
Type BF
Protection against ingress of fluids and particulate matter
IPx0
CAUTION: Exposing the Rezūm Generator to extreme environmental conditions outside of its specified parameters may compromise the ability of the Rezūm Generator to function properly and/or cause the plastic to warp and/or crack.
CAUTION: If the Rezūm Generator is stored in an environment with a temperature below the operating temperature, the unit should be allowed to warm up to the needed operating temperature before using.
Table 12: Generator Specification
3033-001 Revision D January 2018
©NxThera 2018
Page 76 of 90
5.2 Operating, Storage, and Transportation Specifications
Operating Conditions
Temperature
Humidity
Pressure
20˚C – 25˚C
10% – 85% RH (non-condensing)
70kPa to 106kPa
Storage Conditions
Temperature
Humidity
Pressure
(Prevent from freezing.)
-10˚C – 40˚C (See USP Controlled Room Temperature.)
10% – 85% RH (non-condensing)
70kPa to 106kPa
Transportation
Temperature
Humidity
Pressure
(for up to 72 hours)
-29˚C – 60˚C
10% – 85% RH (non-condensing)
70kPa to 106kPa
Table 13: Operating, Storage, and Transportation Specifications
3033-001 Revision D January 2018
©NxThera 2018
Page 77 of 90
5.3 Electromagnetic Compatibility Requirements
WARNING: Use of accessories other than those specified in this document may result in increased emission or decreased immunity of the Rezūm Generator.
WARNING: The Rezūm Generator should not be used adjacent or stacked with other equipment and, if necessary, observe its operation to verify its normal operation during use. Refer to the Electromagnetic Immunity information in this chapter.
CAUTION: The Rezūm Generator needs special precautions regarding Electromagnetic compatibility (EMC) and care should be taken in accordance to the EMC information provided in Chapter 5 of this document.
CAUTION: Use of portable and mobile RF communications equipment near the Rezūm Generator may affect its operation.
CAUTION: Observe the following cautions when connecting this generator with other equipment:
Ensure that the connected equipment is in accordance with the IEC60601-1 or IEC safety
standards.
Employ additional protective measures (e.g., additional protective earthing) as necessary.
CAUTION: Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference (RFI), which could affect the performance of this device. Avoid operating the Rezūm Generator near cauterizers, diathermy equipment, FM 2-way radios, or cellular phones. Turn power off to radio, cellular and other like equipment near the Rezūm Generator. Refer to the EMI tables in Chapter 5.
3033-001 Revision D January 2018
©NxThera 2018
Page 78 of 90
Table 14: Electromagnetic Emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The Rezūm Generator is intended for use in the electromagnetic environment specified below. The customer or the user of the Rezūm Generator should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment –
guidance
Radiated emissions
CISPR 11:2015
Group 2
The Rezūm Generator uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Conducted emissions
CISPR 11:2015
Class A
The Rezūm Generator is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purpose.
Harmonic current emissions
IEC 61000-3-2:2014
Class A
Voltage fluctuations and flicker
IEC 61000-3-3:2013
Complies
3033-001 Revision D January 2018
©NxThera 2018
Page 79 of 90
Table 15: Electromagnetic Immunity
Immunity
test
Compliance
level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4­2:2008
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%
Electrical fast transient/burst immunity
IEC 61000-4­4:2012
±1 kV common mode; 5kHz and 100kHz
Mains power quality should be that of a typical commercial or hospital environment.
Surge immunity
IEC 61000-4­5:2014
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Power frequency magnetic field
IEC 61000-4­8:2009
30 A/m,
50/60 Hz
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
3033-001 Revision D January 2018
©NxThera 2018
Page 80 of 90
Immunity
test
Compliance
level
Electromagnetic environment - guidance
Voltage dips and interruptions immunity
61000-4­11:2004
Six dips each at 100%, 60%, 30% voltage reduction; one interrupt
Three 100% dips each at phase angles of 0, 45, 90, 135, 180, 225, 270, and 315; one interrupt
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Rezūm Generator requires continued operation during power
mains interruptions, it is recommended that the Rezūm
Generator be powered from an uninterruptible power supply.
3033-001 Revision D January 2018
©NxThera 2018
Page 81 of 90
Table 16: Electromagnetic Immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The Rezūm Generator is intended for use in the electromagnetic environment specified below. The customer or the user of the Rezūm Generator should assure that it is used in such an environment.
Immunity
test
Compliance
level
Electromagnetic environment - guidance
Conducted RF IEC 61000-4­6:2013
Radiated RF IEC 61000-4­3:2010
3 Vrms 150 kHz to 80 MHz
6 Vrms in ISM bands
3 V/m 80 MHz to 2.7 GHz
Portable and mobile RF communications equipment should be
used no closer to any part of the Rezūm Generator, including
cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √P d = 1.2 √P 80 MHz to 800 MHz d = 2.3 √P 800 MHz to 2.57GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and is the recommended separation distance in meters (m)b.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,c should be less than the compliance level in each frequency range.d
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
3033-001 Revision D January 2018
©NxThera 2018
Page 82 of 90
a
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765
MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 to
40.70 MHz.
b
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the
frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in these frequency ranges.
c
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Rezūm Generator is used exceeds the applicable RF compliance level above, the Rezūm Generator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Rezūm Generator.
d
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the Rezūm Generator
The Rezūm Generator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Rezūm Generator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and Rezūm Generator as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter (m)
Table 17: Separation Distances
3033-001 Revision D January 2018
©NxThera 2018
Page 83 of 90
Rated maximum output
power of transmitter
(W)
150 kHz to 80
MHz
d = 1.2√P
80 MHz to 800
MHz
d = 1.2√P
800 MHz to 2.7 GHz
d = 2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are
6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and
40.66 to 40.70 MHz.
NOTE 3 An additional factor of 10/3 is used in calculating the recommended separation distance
for transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas.
NOTE 4 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
3033-001 Revision D January 2018
©NxThera 2018
Page 84 of 90
5.4 EN 60601-1-2 Compliance
WARNING: RF Interference - Known RF sources, such as cell phones, radio or TV stations, and two-way radios, may cause unexpected or adverse operation of this generator. Consult qualified personnel regarding system configuration.
WARNING: The generator should not be used adjacent to, or stacked with other equipment. If adjacent or stacked use is necessary, test the generator to verify normal operation. Refer to the Electromagnetic Immunity information in Chapter 5.
WARNING: The generator needs special precautions regarding Electromagnetic Compatibility (EMC) and needs to be put into service according to the EMC information in Chapter 5 provided in this operator’s manual.
CAUTION: The generator conforms to the requirements of the EMC standard (IEC 60601-1­2:2009). However, it may be affected by electrical scalpels and microwave treatment devices and there may be an impact on measurement precision for patients using cardiac pacemakers and other similar devices. Check the operation of this generator during and after use of the above mentioned equipment and with patients potentially affected.
3033-001 Revision D January 2018
©NxThera 2018
Page 85 of 90
Chapter 6 Troubleshooting
Overview
This chapter contains troubleshooting steps, error message description, error message table, and how to obtain technical assistance.
3033-001 Revision D January 2018
©NxThera 2018
Page 86 of 90
6.1 Troubleshooting Steps
If you experience a problem while using the generator, please use the Error Message table to troubleshoot the issues. If you are unable to correct it, write down the error message and error code, if applicable, and contact qualified service personnel in your institution or contact Customer Service.
3033-001 Revision D January 2018
©NxThera 2018
Page 87 of 90
6.2 Error Messages
Error message
Critical Error message
Error messages will be displayed on the screen. There are 2 types of error messages – Error and Critical Error.
Error Message
Figure 52: Example Error Message
Figure 53: Example Critical Error Message
3033-001 Revision D January 2018
©NxThera 2018
Page 88 of 90
6.3 Error Message Table
Code
Error Title
Error Cause Text
Error Message
400
Generator Error
RF Power Supply Error
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact Technical Support.
405
Generator Error
Power Supply Initialization Error
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact Technical Support.
425
Generator Error
SBC Communications Critical Timeout
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact Technical Support.
430
Generator Error
RF Power Supply Communication Error
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact Technical Support.
435
Generator Error
MCU Processing Error
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact Technical Support.
440
Generator Error
CPLD Self-Test Error
Power off/on the Generator. If problem persists, contact Technical Support.
450
Generator Error
Delivery Device Interface Self-Test Error
Power off/on the Generator. If problem persists, contact Technical Support.
455
Generator Error
Saline Pump Self-Test Error
Power off/on the Generator. If problem persists, contact Technical Support.
460
Generator Error
Syringe Pump Self-Test Error
Power off/on the Generator. If problem persists, contact Technical Support.
465
Generator Error
Water Pressure Self-Test Error
Power off/on the Generator. If problem persists, contact Technical Support.
470
Generator Error
Delivery Device Temperature Excessive
Retract the needle and remove the Delivery Device from patient. Power off/on the Generator. Replace Delivery Device.
475
Generator Error
Software Compatibility Self­Test Error
Power off/on the Generator. If problem persists, contact Technical Support.
480
Generator Error
Sensor Interface Error
Retract the needle and remove the Delivery Device from patient. If problem persists, contact Technical Support.
485
Generator Error
Internal Generator Temperature Error
Retract the needle and remove the Delivery Device from patient. Power off the Generator and allow it to cool down before using again.
490
Generator Error
RF Power Supply Operational Error
Retract the needle and remove the Delivery Device from patient. If problem persists, contact Technical Support.
495
Generator Error
RF Power Supply Self-Test Error
Power off/on the Generator. If problem persists, contact Technical Support.
35000
Generator Error
GUI Program Files Corrupted
Power off/on the Generator. If problem persists, contact Technical Support.
35001
Generator Error
Unexpected GUI Program Exit
Power off/on the Generator. If problem persists, contact Technical Support.
35002
Generator Error
Unable to Start GUI Program
Power off/on the Generator. If problem persists, contact Technical Support.
40000
Generator Error
GUI Unable to Communicate with MCU
Power off/on the Generator. If problem persists, contact Technical Support.
41010
Generator Error
Patient Database Full
Export Patient Records to USB flash drive as needed and then delete old Patient Records to continue.
41020
Generator Error
MCU Reboot Detected
Retract the needle and remove the Delivery Device from patient. Remove the Generator from service and contact Technical Support.
The following tables list all error messages that are displayed by the generator. Follow error message instructions to resolve the error.
Table 18: Critical Error Message Table
3033-001 Revision D January 2018
©NxThera 2018
Page 89 of 90
Code
Error Title
Error Cause Text
Error Message
200
Faulty Delivery Device
Unable to Read from Delivery Device Memory
Remove and reconnect Delivery Device electrical cable. If problem persists, replace Delivery Device.
205
Faulty Delivery Device
Unable to Write to Delivery Device Memory
Remove and reconnect Delivery Device electrical cable. If problem persists, replace Delivery Device.
210
Faulty Delivery Device
Faulty Delivery Device Thermocouple
Replace Delivery Device.
211
Faulty Delivery Device
Faulty Delivery Device Trigger Signals
Replace Delivery Device.
215
Faulty Delivery Device
Invalid Therapy Code
Remove and reconnect Delivery Device electrical cable. If problem persists, replace Delivery Device.
218
Faulty Delivery Device
Delivery Device Impedance Error
Replace Delivery Device.
219
Faulty Delivery Device
Delivery Device Frequency Error
Replace Delivery Device.
220
Expired Delivery Device
Maximum Full Treatments Exceeded
Replace Delivery Device.
225
Faulty Delivery Device
Delivery Device is Permanently Disabled
Replace Delivery Device.
230
Expired Delivery Device
Maximum Vapor Time Exceeded
Replace Delivery Device.
235
Prime Failed
Low Temperature (Prime)
Replace Delivery Device.
236
Pre-Treatment Failed
Low Temperature (Pre­Treatment)
Replace Delivery Device.
240
Prime Failed
Low Water Pressure (Prime)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace Delivery Device.
241
Prime Failed
High Water Pressure (Prime)
Check water line for kinks. Resume priming. If problem persists, replace Delivery Device.
242
Pre-Treatment Failed
Low Water Pressure (Pre­Treatment)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace Delivery Device.
243
Pre-Treatment Failed
High Water Pressure (Pre­Treatment)
Check water line for kinks. Resume pre-treatment vapor cycle. If problem persists, replace Delivery Device.
245
Prime Failed
High Temperature (Prime)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace Delivery Device.
246
Pre-Treatment Failed
High Temperature (Pre­Treatment)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace Delivery Device.
250
Prime Paused
Trigger Released
Delivery Device trigger was released before priming completed. Press and hold Vapor Activation Button (blue) to continue priming.
251
Pre-Treatment Paused
Trigger Released
Delivery Device trigger was released before pre-treatment vapor cycle completed. Press and hold Vapor Activation Button (blue) to restart pre­treatment vapor cycle.
255
Treatment Halted
Low Temperature (Treatment)
Retract the needle and remove the Delivery Device from patient. Replace Delivery Device.
260
Treatment Halted
High Water Pressure (Treatment)
Check water line for kinks. Resume treatment. If problem persists, replace Delivery Device.
265
Treatment Halted
Low Water Pressure (Treatment)
Check syringe and water line for bubbles or leaks. If bubbles are found, replace syringe and reprime Delivery Device. If leaks are found, replace Delivery Device. If no bubbles or leaks are observed, replace Delivery Device.
270
Syringe is Empty
Syringe Empty
Retract the needle and remove the Delivery Device from patient. Replace syringe and reprime Delivery Device.
3033-001 Revision D January 2018
©NxThera 2018
Page 90 of 90
Code
Error Title
Error Cause Text
Error Message
275
Prime Failed
Syringe Water Fill Error
Refill syringe and reprime Delivery Device.
280
Treatment Halted
Elevated Coil Temperature
Partial treatment delivered. Check syringe and water line for bubbles or leaks. If no bubbles or leaks are found, resume treatment. If problem persists, replace Delivery Device. If bubbles are found, replace syringe and reprime Delivery Device. If leaks are found, replace Delivery Device.
290
Faulty Delivery Device
High Temperature (Idling)
Replace Delivery Device.
291
Faulty Delivery Device
High Water Pressure (Idling)
Replace Delivery Device.
295
Faulty Delivery Device
Needle Deployment Error
Ensure needle is retracted. Replace Delivery Device.
296
Faulty Delivery Device
Needle Retraction Error
Reattempt needle retraction. If problem persists, retract needle manually and replace Delivery Device.
300
Saline Pump Error
Saline Pump Encoder Error
Ensure Delivery Device saline flush line is correctly inserted into saline pump and pump door is closed. If problem persists, contact Technical Support.
325
Confirm Bladder Drain
Saline Instilled Limit Exceeded
Saline instilled limit exceeded. Please confirm the physician has drained the bladder.
341
Prime Paused
RF Power Tolerance Limit Exceeded
Wait for RF Power Supply reset to complete.
342
Pre-Treatment Paused
RF Power Tolerance Limit Exceeded
Wait for RF Power Supply reset to complete.
343
Treatment Halted
RF Power Tolerance Limit Exceeded
Wait for RF Power Supply reset to complete.
41000
Export Error
USB Drive Not Present or Invalid
Re-insert USB flash drive and try again. If problem persists, replace USB flash drive.
41002
Export Error
USB Drive Export Error
Insert a valid USB flash drive with sufficient available memory.
Table 19: Non-Critical Error Message Table
6.4 Obtaining technical assistance
For technical information and assistance, contact:
NxThera, Inc. 7351 Kirkwood Lane North Suite 138 Maple Grove, MN 55369 USA
www.nxthera.com
Customer Service: Phone: +1-763-515-0404 Toll Free: +1-888-319-9691 Fax: +1-763-515-2085
Loading...