Revox B77 Service Manual

Page 1
REVOX service
6.4.3. Gain of input amplifier
Adjust operating controls as per fig.6.4.- 1.
Connect audio voltmeter to OUTPUT 51 (pot-meters LEVEL [50] fully open). Connect audio generator to feed the sockets AUX INPUT [48] in parallel. Select
frequency of 1000 Hz and adjust generator level until a reading of 0.775 V is obtained on the audio vol tmeter .
The required input level should read 20 mV approx. Note:
At all positions of the mode selector (MONITOR 17) the difference in output level must not exceed 0.5 dB.
1 NSL/2016
Page 2
REVOX service
6.4.4. Testing the mixing facility with a single
Adjust operating controls as per fig. 6.4.-2.
Audio generator adjusted as for previous test ( 1000 Hz. 0.775 V output level) and connected to AUX INPUT [48]
Both VU-meters indicate 0 VU.
Switch record pre-selector REC R [22] to ON. Only the left-hand VU-meter indicates 0 VU.
Turn potentiometer INPUT LEVEL LEFT fully counterclockwise to position "0". Turn potentiometer INPUT LEVEL RIGHT fully clockwise to position "10" Only the right-hand VU-meter will read 0 VU.
Switch record pre-selector REC L to position OFF. Both VU-meters indicate 0 VU
2 NSL/2016
channel feed
Page 3
REVOX service
6.4.5. VU-meter calibration
Set operating controls as per fig 6.4 -3.
Audio generator still adjusted to same level and frequency (1000 Hz. 0.775 V on output [51] )
Adjust potentiometers METER CAL L and R (fig. 6.4.-4) to obtain a 0 VU deflection on the respective VU-meter.
3 NSL/2016
Page 4
REVOX service
6.4.6. Checking the overl oa d indicators
When raising the signal level by more than 6 dB, the overload indicators in the VU­meters must become illuminated.
The difference in triggering level between the two indicators must not exceed
0.5dB.
When exceeding this tolerance. alter the values of resistors R37 or R34 respectively in the monitor amplifier 1.177 260.
On newer versions, resistors R34/R37 are replaced by additional trim-pots R30. R38
6.4.7. Frequency response "be f ore tape"
Set operating controls as per fig 6.4.-3.
Connect audio generator to both AUX INPUTS [51] in parall el . Connect audio millivolt meter to OUTPUT [48]
Sweep generator from 30 Hz to 20 000 Hz and check frequency response by observing the level variation on the audio mill i volt meter.
Permissible tolerance+ 0/-1 dB.
Check frequency respons e on OU TPU T L and R.
4 NSL/2016
Page 5
REVOX service
6.4.8. Reproducing head azimut h, coarse adjustment (Level maximum)
Double check: Check to make sure that the head gap is centered inside the wear pattern on the
head.
With grease pencil make horizontal mark on the head face, run tape for several seconds.
Then check position of the head gap inside the polished area. lf necessary adjust head position to bring gap into center of wear area.
Set operating controls as per fig. 6.4.-5. Connect audio millivolt meter to OUTPUT L.
Load recorder with calibration tape 19 cm/s and run azimuth adjusting secti on (10 kHz) by pressing button PLAY.
Rotate screw “P” to obtain maximum output voltage.
5 NSL/2016
Page 6
REVOX service
6.4.9. Reproducing head azimut h, f ine a djustment
(Phase check method)
When using the phase check method for align ment of the head azimuth, it is essential that the azimuth is first adjusted for level maximum (section 6.4.8.) to avoid the possibility of a phase error in excess of 90°
Switch mode selector MONITOR to position MONO. Connect audio millivolt meter to both outputs OUTPUT L + R @ in parallel.
Load recorder with calibration tape 19 cm/s and run azimuth section (10 kHz) while rotating screw “P” until a pronounced maximum in output voltage is
obtained.
6.4.10. Reference level from calibration tape
(REVOX test tape no. 46001)
Set operating controls as per fig. 6.4.-6.
Connect audio millivolt meter (1 V range) to OUTPUT L . Search for reference level 257 nWb/m on the calibration tape and run that
section by pressing the button PLAY.
Adjust trim-pot REPR LEVEL L (fig.6.4.-7) to obtain an output level of 0.775 V.
Connect audio millivolt meter to OUTPUT R. Adjust trim-pot REPR LEVEL R to obtain an output level of 0.775 V
6 NSL/2016
Page 7
REVOX service
6.4.11. Reproducing system frequency response
(measured with calibration tape)
Set operating controls as per fig. 6.4.-6.
Search for the beginning of the frequency response section on the calibration tape.
Connect audio millivolt meter to OUTPUT L + R in parallel. Start recorder in PLAY and measure frequency response relative to 1000 Hz.
The frequency response of both channels may be checked in one run by alternating the setting of the MONITOR mode selector between the positions LEFT and RIGHT
7 NSL/2016
Page 8
REVOX service
6.4.12. Checking oscillator frequency and RF voltages
Set operating controls as per fig.6.4.-8.
Load recorder with blank tape (REVOX 631) and start in the recording mode.
Erase voltage: Use electronic 'Voltmeter to measure the RF-voltage on the erase head:
approx. value: 30 V ... 32 V
Oscillator frequency: Measure the oscillator frequency on the erase head with a digital counter.
The frequency must read 150 kHz ± 5 kHz.
When measuring with an oscilloscope, the display must consist of 15 cycles on the 100 µsec time base.
If the frequency deviates beyond the above limits. adjust the slug in Tl on the oscillator board .
For this adjustment, the extension board 1.177.241 is required in order to gain access to the coil.
8 NSL/2016
Page 9
REVOX service
6.4.13. Adjustment of bias trap in recording amplifier
Set operating controls as per fig. 6.4.-8.
Connect millivolt meter to the test point “X” of the respective recording channel fig.6.4.-9 Refer to circuit diagram 1.177.230.
Load recorder with blank tape and start in the recording mode. Adjust the controls BIAS TRAP L + R to obtain a minimum voltage indication
(<350 mV). fig 6.4.-10.
9 NSL/2016
Page 10
REVOX service
6.4.14. Recording head azim ut h c oar s e adj us t m e nt
(Level maximum)
Set operating controls as per fig. 6.4.-11.
Connect audio millivolt meter to OUTPUT [51] and select 100 mV range. Connect audio generator in parallel to AUX INPUTs . Select frequency of 10 kH z.
Load recorder with blank tape (REVOX 631 ) and start in the recording mode.
Adjust generator level to obtain an out put r ea di ng of 0.775 V.
Rotate screw R (fig. 6.4.-12) to obtain maximum output level. Because there is a time lag before the effects of any alteration in record head azimuth will be
shown on the meter, screw “R” should be rotated very slowly.
10 NSL/2016
Page 11
REVOX service
6.4.15.Recording head azimuth fine adjustment
(Phase check method)
In order to avoid the possibility of phase errors in excess of 90° , it is essential that the azimuth adjustment as described under section 6.4.14. is first completed.
Turn mode selector MONITOR to position MONO. Connect audio millivolt meter to both sockets OUTPUT in parallel.
Connect audio generator to AUX INPUT in parallel. Select frequency of 10 kHz and adjust to a level of 20 dB below 0VU.
Load recorder with blank tape (REVOX 631 ) and start in the recording mode. Adjust screw R (fig. 6.4.-12) until a sharply rising. well defined maximum in output
voltage is obtained.
11 NSL/2016
Page 12
REVOX service
6.4.16. Bias adjustment
Since tape magnetization varies with bias current, a sharply defined maximum output is exhibited at high frequencies when gradually increasing the bias current from zero upwards.
This relationship may be utilized to determine the optimum operating point (6 4 -13)
Set operating controls as per fig 6.4.-14.
Connect audio millivolt meter to OUTPUT L or R respectively . Connect audio generator to AUX IN- PUTs in parallel. Select frequency of 10 kHz and adjust to a level of 20 dB below 0 VU
Load recorder with blank tape (REVOX 631 ) and stat in the recording mode
12 NSL/2016
Page 13
REVOX service
The potentiometers:
BIAS ADJ LEFT (SLOW+ FAST) BIAS ADJ RIGHT (SLOW+ FAST)
Are to be adjusted at their respective tape speeds by turning them from their fully counter clock wise position in clockwise direction until the signal level, as measured on the recorder's output, reaches its maximum (mode selector switched to either LEFT or RIGHT as required)
After having reached the maximum, slowly keep on rotating the respective trim-pot in clockwise direction until the 10 kHz output signal has dropped by the amount shown in table 6.4 -15
13 NSL/2016
Page 14
REVOX service
6.4.17. Recording level
Set operating controls as per fig. 6.4.-16.
The levels of the reproducing system must already be adjusted as per section
6.4.10.
Connect audio generator in parallel to AUX INPUTS Connect audio millivolt meter to OUTPUT Load recorder with blank tape and start in the recording mode.
Select frequency of 1000 Hz on audio generator and adjust its level until a reading of 0.775 V is obtained on the audio millivolt meter .
Switch MONITOR selector to position TAPE. Adjust the trim-pots REC LEVEL, LEFT or RIGHT respectively to obtain an output
level of 0.775 volts. Turn mode selector to either LEFT or RIGHT as required.
Double check: There must be no level difference when switching the monitor selector between the
positions TAPE and INPUT.
At the tape speed of 9,5 cm/s, a level difference of 1 dB is permissible
14 NSL/2016
Page 15
REVOX service
6.4.18. Adjusting the recording equal ization
Set operating controls as per fig. 6.4.-17.
Connect audio millivolt meter to OUTPUT Connect audio generator in parallel to both AUX INPUTS
Signal condition: 12 kHz, level 20 dB below 0 VU
Load recorder with blank tape and start in the recording mode.
While recording the 12 kHz signal, adjust the trim-pots EQ SLOW. FAST at the respective tape speeds until the signal level at the recorder's output falls within the range from 0 dB to + 1 dB relative to 1 kHz (77,5 mV / -20dB).
Turn mode selector to position RIGHT and adjust equalization for the right channel as well.
15 NSL/2016
Page 16
REVOX service
6.4.19. Frequenc y respons e "overall"
Set operating controls as per fig. 6.4.-18.
Output level of audio generator remains set as for previous test (20 dB below 0 VU). Load recorder with blank tape and start in the recording mode.
Calibrate audio millivolt meter for frequency response measurement by adjusting the audio generator level at 1 kHz to obtain a convenient mid-scale deflection on the meter (e.g. -5dB on the 100 mV range).
Check frequency response at the two tape speeds. Both channels may be checked in one run by alternating the mode selector between the positions LEFT and RIGHT at each individual frequency .
For permissible response deviations check section 2. Techni cal Specifications of this manual.
Frequency Response:
(measured via tape, at -20 VU)
At 9.5 cm/s 30 Hz ... 16 kHz + 2/-3 dB
50Hz ... 10kHz ±1,5dB
At 19 cm/s 30 Hz ... 20 kHz + 2/-3 dB
50 Hz ... 15 kHz ± 1,5 dB
16 NSL/2016
Page 17
REVOX service
6.5.1. Distorsion via tape
Set operating controls as per fig. 6.5.-1. Connect audio generator to AUX INPUT Frequency: 500 Hz Level: 0VU + 6 dB
Connect distortion factor meter to OUTPUT Load recorder with blank tape and start in the recording mode.
Measure distortion on both channels. Switch MONITOR mode selector between positions LEFT and RIGHT as required.
The permissible amount of distortion (pure third harmonic) at the respective tape speeds may be seen from table 6.5.-2.
17 NSL/2016
Page 18
REVOX service
6.5.2. Weighted and unweighted noise m easurements via tape
Noise voltage measurements are taken with reference to the peak level modulation of 514 nWb/m (output level 1.55 V).
In order to avoid erroneous test results due t o bias frequency pick-up, al l measurements are taken by running tape. which has been erased on the recorder in a previous run.
Set operating elements as per fig. 6.5.-3.
Connect audio millivolt meter to OUTPUT L . With previously erased tape on the recorder, start the machine in PLAY. Read noise voltage of the left channel through ASA A weighting network.
Switch mode selector to position RIGHT. Read noise voltage of the right channel through ASA A weighting network
Measure noise voltage at the second tape speed as described above The RMS signal to noise ratios (weighted as per ASA A) can be seen from table 6.5.-4.
If the signal to noise ratios are less than those shown in the table, carefully demagnetize the heads and tape guides and repeat the test
18 NSL/2016
Page 19
REVOX service
For the following measurements
- Erase depth
- Crosstalk MONO
- Crosstalk STE EO
An audio millivolt meter with a narrow band-pass filter of less than 100 Hz bandwidth (wave analyzer) will be required.
19 NSL/2016
Page 20
REVOX service
6.5.3. Erase efficiency
To measure the erase depth, a sig nal of appr oximately 1000 Hz has to be recorded at peak level and that same recording has to be erased again in a subsequent run. The residual level of that signal will then be measured.
Set operating controls as per fig 6.5 -5.
Connect wave analyzer to OUTPUT Connect audio generator in parallel to both AUX INPUTS
Select 1000 Hz and adjust level to obtain peak level indication on the VU-meter ( 1000 Hz. 0 VU+ 6 dB) .
Load recorder with blank tape and start in the recording mode.
Establish reference reading on wave analyzer.
STOP and rewind to the beginning of the recording. Close fader INPUT LEVEL LEFT to position "0".
Start the machine again in the recording mode and measure the residual signal level on tape.
The magnetic recording remaining on tape must be attenuated by at least 75 dB below the above established reference (typical value 80 dB).
Switch mode selector MONITOR to position RIGHT and repeat the above measurements.
20 NSL/2016
Page 21
REVOX service
6.5.4. Crosstalk MONO
It is advisable le to use virgin (or bulk erased) tape for this test. Record a signal of 1000 Hz at peak level for approximately 1 minute on the left
channel only. then record the same signal again for 1 minute on the right channel.
In making that recording the preselector switch of the unused channel is switched to the position OFF
Set operating controls as per fig. 6.5.-6.
Select 1000 Hz on audio generator and adjust for peak level recording.
Load recorder with tape and start in the recording mode as described above. After approximately 1 minute switch recording preselector REC-L to position
OFF. Turn potentiometer INPUT LEVEL LEFT to position 0.
Switch recording preselector REC-A to position ON.
Turn potentiometer INPUT LEVEL RIGHT to position 10. Continue to record for approximately 1 minute. then press button STOP and
rewind to the beginning of the recording.
Connect wave analyzer to OUTPUT R. Start recorder in PLAY . Measure crosstalk L – R (>60dB). Wind forward to the recording on the right channel. Connect wave analyzer to OUTPUT L and switch mode selector MONITOR to LEFT. Start recorder in PLAY. Measure cross- talk R - L (> 60 dB).
21 NSL/2016
Page 22
REVOX service
6.5.5. Crosstalk STEREO
To measure the stereo crosstalk performance connect audio generator to AUX INPUT L.
Set operating controls as per fig. 6.5.-7.
Crosstalk L .- R.
Connect wave analyzer to OUTPUT R
Select 1000 Hz on the audio generator and adjust for peak level recording. Start the recorder in the recording mode. The signal level appearing on OUTPUT R must be 45 dB below that of a peak level
recording.
Crosstalk R .- L.
Connect audio generator to AUX INPUT R. Connect wave analyzer to OUTPUT L. Close potentiometer INPUT LEVEL LEFT to position "0".
Open potentiometer INPUT LEVEL RIGHT to position "10". Continue to rerord 1000 Hz at peak level.
Measure crosstalk. The signal level appearing on OUTPUT L must be 45 dB below a peak level recording.
22 NSL/2016
Page 23
REVOX service
RADIO
2.8 mV
open
60dB
MIC Hi
2.8 mV
200Ω
70dB
MIC LO
0.17 mV
200Ω
50dB
LINE AMP
-
Input level pot
80dB
Output voltages from peak level recording
OUTPUT
LEVEL min.
70mV
1.55 v
RADIO
LEVEL min.
70mV
PHONES
max. 5.6 V AL=
PHONES
max. 2.8 V AL = 220
6.5.6. Audio Measurements "Before tape"
Input
AUX
*B = 20 kHz linear
OUTPUT
Input voltages
for peak
level
recording (0
VU + 6 dB)
38mY
Terminating
resistance for
noise
measurements
10 kΩ
pos. "0"
Unwe ighted * S/N ratios
relative to peak level
73dB
OUTPUT
RADIO
0 Tolerance includes differences between channels and differences in amplification factors.
LEVEL max.
LEVEL max.
23 NSL/2016
1 55V
±0.75dB
± 0.75 dB
Page 24
REVOX service
6.5.7. Wow and flutter
The wow and flutter values as specified in the technical data section are measured with test equipment which conforms to the German standard DIN 45507 (consistent with IEEE standard 193-1971).
24 NSL/2016
Page 25
REVOX service
25 NSL/2016
Page 26
REVOX service
26 NSL/2016
Loading...