REVITIVE ARTHROSE KNEE User Manual [fr]

Page 1
User’s Manual
PLEASE READ THE USER’S MANUAL CAREFULLY BEFORE USING THIS PRODUCT
Model number 2836AA
2910_IFU02_12517049.indd 1 22/07/2016 09:49
USE ONLY AS DIRECTED
If symptoms persist, consult your healthcare professional
Page 2
Table of contents
What is inside the box? 4
Parts and Controls of REVITIVE 5
Introduction to REVITIVE 6-7
Indications for use 6
How REVITIVE works 7
Important Safeguards 8-11
Who Should not use REVITIVE 8
Warnings and Cautions 9-10
Safety Precautions 11
Adverse Reactions 12
Instructions for use 13-27
How long should I use REVITIVE? 13
Step 1: Setting up REVITIVE for the first time 13
Step 2: Setting up REVITIVE for use 14
Step 3: Powering REVITIVE 15-16
Step 4: Using the foot-pads 17-18
Step 5: Using the Arthritis Pads 20-25
Step 6: Recharging the battery 26
Step 7: Cleaning and storing 27
Replacing the batteries 27
Replacing the Arthritis Pads 27
Troubleshooting 28-30
Technical Specifications
31-32
Warranty
33
2
2910_IFU02_12517049.indd 2-3 22/07/2016 09:49
3
Page 3
What is inside the box?
On opening the carton, please check that the following components are provided. If you think anything is missing or require any assistance with setting up, using or maintaining REVITIVE, please contact us using the helpline number on the back of this booklet.
Parts and Controls of REVITIVE
Carry handle
Parts of REVITIVE
G
A
AC/DC power adaptor
REVITIVE
B
Arthritis
C
(Electrode) Pads x 2 pairs
Arthritis Pad cord
D
Remote control
E
H
Power ON/OFF
AC/DC power
adaptor socket
K
L
Arthritis Pad socket
I
Control panel/display
J
Foot-pads
Advanced
M
IsoRocker
IsoRocker Disabler
N
Controls of REVITIVE
Remote control
Time Control – /+
AAA batteries x 2
F
4
2910_IFU02_12517049.indd 4-5 22/07/2016 09:49
Power ON/OFF
Intensity Control – /+
Control panel
Battery
indicator
5
Page 4
Introduction to REVITIVE
INDICATIONS FOR USE
Using REVITIVE ARTHRITIS-KNEE is intended to:
Reduce pain and discomfort in the legs – caused by swelling (oedema)
due to osteoarthritis or following surgery.
Reduce swelling and stiffness in the legs – caused by being immobile due to osteoarthritis.
Increase muscle strength to help regain mobility and stabilise knees affected by being immobile due to osteoarthritis or following surgery.
Improve circulation to reduce or prevent blood-pooling (stasis) –
caused by being immobile following surgery.
Help maintain leg vein and knee-joint health – by increasing circulation,
delivering more oxygenated-blood and reducing swelling (oedema) in the legs.
The above Indications for Use are certified under the Medical Devices Directive 93/42/EEC.
HOW REVITIVE ARTHRITIS-KNEE WORKS
With its unique DualAction REVITIVE Arthritis-Knee works in two ways:
1. Delivers Electrical Muscle Stimulation (EMS) to your feet via a
pair of cushioned foot-pads, while you are seated. The EMS causes muscles in your calf and foot to contract, which helps to increase circulation in your legs.
REVITIVE has been developed so that it is comfortable for you to use. It features a patented IsoRocker that allows for natural ankle movement when applying the EMS to your feet. Ankle movement is a key factor in increasing circulation in the lower legs. You can choose to enable or disable the IsoRocker at any time.
REVITIVE is designed to be easy to use and comes with a remote control, which means you do not have to reach down to adjust settings during use.
2. Delivers Electrical Muscle Stimulation to the front of the thighs/
quadriceps through the use of the Arthritis (Electrode) Pads - to strengthen the quadriceps. Strengthening your quadriceps muscles may help to:
Support and stabilise the knee
Reduce stress on the knee
Relieve knee pain
To get the most out of REVITIVE, it is recommended that you use:
REVITIVE foot-pads for 20-30 minutes each day.
REVITIVE Arthritis Pads on the thigh of your affected leg (knee)
once a day for 20 minutes, 5 days per week.
Visit www.revitive.com for more information
6
2910_IFU02_12517049.indd 7 22/07/2016 09:49
7
Page 5
IMPORTANT SAFEGUARDS
READ ALL INSTRUCTIONS BEFORE USE
Warnings
WHO SHOULD NOT USE REVITIVE
REVITIVE should not be used by some people.
Do not use if (contraindications):
You are fitted with an electronic implanted device such as a heart
pacemaker or Automatic Implantable Cardioverter Defibrillator (AICD)
You are pregnant
You are being treated for, or have the symptoms of, an existing Deep
Vein Thrombosis (“DVT”)
Long periods of inactivity can put you at greater risk of developing Deep
Vein Thrombosis (DVT). DVT is a blood clot and usually occurs in a deep leg vein. If part of the DVT breaks off it may lead to potentially life- threatening complications such as pulmonary embolism.
If you have been inactive for prolonged periods and suspect you may
have a DVT, consult your doctor immediately. To prevent dislodging the clot do not use REVITIVE.
In some cases of DVT there may be no symptoms. However it is important
to be aware of the symptoms that may include:
pain, swelling and tenderness in one of your legs (usually your calf)
Consult with your doctor before using this device if:
• You are in the care of a doctor
• You have a history of heart problems
• You have suspected or diagnosed epilepsy
• You are unsure about the suitability of REVITIVE for you
• You are unsure about the cause of your symptoms
If you have a metallic implant, you may experience pain or discomfort near
the implant when applying electrical stimulation, if this should occur discontinue use and seek advice from your doctor
The electrical stimulation may feel more intense close to a metallic implant.
It is safe to continue use provided no pain is experienced. You may need to adjust the intensity to a comfortable level
If your Arthritis symptoms worsen, or you have had knee surgery (within the
last six months), consult with your doctor before reusing REVITIVE
Do not use the Arthritis (Electrode) Pads on your head, face, neck or chest
Do not apply foot-pads or Arthritis (Electrode) Pads directly:
on open wounds or rashes, swollen, red, infected, or inflamed areas or skin eruptions (such as phlebitis, thrombophlebitis, varicose veins, cellulitis)
on or close to, malignant tumours
on areas treated with radiotherapy (within the past 6 months)
a heavy ache in the affected area
warm skin in the area of the clot
redness of your skin, particularly at the back of your leg, below the knee
Consult your doctor as soon as possible if you show any signs of the
above symptoms.
8
2910_IFU02_12517049.indd 8-9 22/07/2016 09:49
on reproductive organs
Arthritis (Electrode) Pads are for single person use only. Do not share Arthritis
Pads with others
Do not use Arthritis (Electrode) Pads in conjunction with a brace or cast
without first consulting with your doctor.
Do not use Arthritis (Electrode) Pads after their use by date
9
Page 6
IMPORTANT SAFEGUARDS
There is a chance you may experience skin irritation or burns beneath
the stimulation electrodes or hypersensitivity due to the electrical stimulation
Symptoms may worsen during the initial treatment phase before getting
better. This may occur if the body has not yet adjusted to increased muscular activity and blood circulation. If this occurs, reduce the intensity and the duration of treatment which will reduce the initial symptoms, if symptoms persist consult with your doctor
Cautions
Be careful when applying stimulation:
Over areas of skin that lack normal sensation (eg following injections around the knee) - be careful not to overstimulate the area.
Over the abdomen during menstruation
After recent surgical procedures (within the last six months) as stimulation may disrupt your healing process
If you are likely to bleed internally, such as following an injury or fracture – use a low intensity and/or shorter time to avoid over- stimulation
Safety Precautions
Do not stand on the machine. Use only when sitting down
Use the device only for its intended purpose
Do not expose the device to extreme heat
Do not spill liquid on the device or its accessories
Do not overload the electrical outlet
Keep this device out of the reach of children
REVITIVE may be used by multiple persons, ensure the device is
cleaned after each use
Use REVITIVE only with the accessories supplied by, or purchased from,
the manufacturer
Check body pads, cords and cables periodically for damage
Do not open REVITIVE or repair it yourself. This will invalidate your
warranty and may cause serious harm
In the unlikely event your REVITIVE malfunctions, disconnect it from the
power source and contact your nearest authorised agent
REVITIVE has passed the required tests for EMI, it may still be affected
by excessive emissions and/or may interfere with more sensitive equipment
After a long period of immobility or inactivity – use a low intensity and shorter time to avoid overstimulation or muscle fatigue
This product is not intended for use by persons with reduced physical,
sensory or mental capabilities, unless they are supervised by a person responsible for their safety
It has been reported that some universal remote control devices
(eg for TV etc) can change the settings on REVITIVE if used during a treatment. If this should occur, simply adjust the time or intensity settings on REVITIVE back to where you want them to be using the REVITIVE Remote Control, or using the device Control Panel
After any exposure to hot or cold temperatures outside the specified
operating range of 10 – 40°C allow the product to re-adjust to the recommended operating temperatures to ensure continued product performance
10
2910_IFU02_12517049.indd 10-11 22/07/2016 09:49
11
Page 7
AC/DC power adaptor
IMPORTANT SAFEGUARDS
Adverse Reactions
If you experience adverse reactions, stop using REVITIVE and
consult with your doctor
SAVE THESE INSTRUCTIONS
Instructions for use
How long should I use REVITIVE?
It is recommended to use REVITIVE foot-pads for 20-30 minutes each day, seven days a week.
It is recommended to use REVITIVE Arthritis Pads on your affected leg (knee) once a day for 20 minutes, 5 days per week.
Do not use REVITIVE more than 6 sessions of 30 minutes (or the equivalent) per day. This may result in muscle fatigue.
STEP 1: SETTING UP REVITIVE FOR THE FIRST TIME
a Remove all parts from the box:
REVITIVE device
Arthritis Pads x 2 pairs
Arthritis Pad cord
Remote control
• AAA batteries x 2
See What is inside the Box? on pages 4-5 to help identify the parts.
b Locate the remote control and insert the batteries.
Before you use REVITIVE for the first time, read the Warnings and Cautions on pages 8-12. If in doubt, consult your doctor before using the product.
12
2910_IFU02_12517049.indd 12-13 22/07/2016 09:49
13
Page 8
Instructions for use
STEP 2: SETTING UP REVITIVE FOR USE
a Turn the device over so that you can see the
IsoRocker and the IsoRocker Disabler.
• The IsoRocker allows REVITIVE to rock while stimulating your legs and feet.
• REVITIVE will begin rocking on the IsoRocker once muscle contractions are strong enough.
• The Isorocker feature can be disabled, by opening the IsoRocker Disabler. REVITIVE will then be static and it will not rock during use.
b If you want to use the IsoRocker, leave the IsoRocker
Disabler in place (ie flat) on the bottom of the device, as shown.
c If you do not want to use the IsoRocker
• Open the IsoRocker Disabler on the bottom of REVITIVE. You may have to pull hard.
STEP 3: POWERING REVITIVE
REVITIVE can be powered from either the rechargeable battery or by connecting it to an electrical socket.
BATTERY POWER OPTION
If you want to use the device without connecting it to an electrical socket, it can run from the battery.
When the device is running from the battery, an indicator will show the remaining battery level.
Full Battery
Low Battery
• Pull the IsoRocker Disabler back until it clicks open, as shown.
• To re-enable the IsoRocker, pull the IsoRocker Disabler and press into place on the bottom of the device (ie flat).
Follow Step 4 - Using the foot-pads or Step 5 - Using the Arthritis Pads to continue setting up REVITIVE.
When the battery drops to the low level, it will require recharging. See Step 6, Recharging Battery on page 26.
If you turn the device off and the low battery level indictor blinks for 10 seconds, the battery requires recharging.
Please note that if the battery runs out, the device can still be used, by connecting to an electrical socket.
14
2910_IFU02_12517049.indd 14-15 22/07/2016 09:49
15
Page 9
Instructions for use
MAINS POWER OPTION
If you want to use REVITIVE when the battery is low or to preserve the battery level, REVITIVE can be used by connecting it to an electrical socket.
a Locate the AC/DC power adaptor.
b Plug the small end of the power cord into the
opening on the side of REVITIVE.
Sit with both bare feet on the foot-pads
STEP 4: USING THE FOOT-PADS
a Get into a comfortable seated position.
For best results sit with your knees at a 90 degree angle (right angle).
b Place REVITIVE on the floor in front of you.
REVITIVE is designed to be used while seated. Never stand on the REVITIVE unit.
c Plug the power adaptor into the nearest electrical
socket. This will start charging the (integrated)
rechargeable battery.
Once you have connected REVITIVE to the electrical socket follow Step 4 Using the foot-pads or Step 5 Using the Arthritis Pads to continue setting up REVITIVE.
c Place both bare feet on the foot-pads.
Make sure you remove all footwear, including socks/ stockings.
d If desired, apply a moisturising lotion to the soles of
your feet.
Hint: Applying a moisturising lotion to the soles of the feet before using REVITIVE can help improve the delivery of electrical stimulation.
e Press the power button on the device or on the remote
control to turn REVITIVE on.
You will hear a beep and the display lights will light up. The time will read 30 minutes and Intensity will read 00.
To turn off REVITIVE at any time during operation, press the power button on the device or on the remote control.
16
2910_IFU02_12517049.indd 16-17 22/07/2016 09:49
17
Page 10
Instructions for use
f Set the time for REVITIVE. The default time is
30 minutes. You can set the timer for up to 60 minutes. If you do not want to use the default time of 30 minutes, set the time for the number of minutes you want.
You can use the buttons on the device or on the remote control.
Press the (+) Time Control to increase the time. Press the (–) Time Control to decrease the time.
NB It is recommended to use REVITIVE foot-pads for 20-30 minutes per day.
Do not use REVITIVE more than 6 sessions of 30 minutes (or the equivalent) per day. This may result in muscle fatigue.
g Press the (+) Intensity Control once to start the
stimulation.
h Use an intensity that provides you with strong but
comfortable muscle contractions and note that the intensity required may vary from day to day - eg using it on 25 one day, does not necessarily mean you will need to use it at 25 the next day. The required intensity could be higher or lower than previous usage.
It is normal to feel tingling or varying sensations in your calf muscles.
REVITIVE does not vibrate.
If you feel pain or discomfort:
• Remove one or both feet from the foot-pads to stop the stimulation.
• Lower the intensity of the foot-pads using the (–) Intensity Control on the control panel, or remote control, before replacing your feet on
REVITIVE.
18
Continue to increase the intensity until you can feel or see the muscles in your calves contract and relax. You can set the intensity between 1-99.
You can use the buttons on the device or remote control.
Press the (+) button to increase the level. Press the (–) button to decrease the level.
i The Time Display will start to count down in
minutes as REVITIVE cycles through its program.
j When REVITIVE times out, the Time Display
reads 00 and you will hear three beeps.
k REVITIVE switches itself off automatically.
19
2910_IFU02_12517049.indd 18-19 22/07/2016 09:49
Page 11
Instructions for use
STEP 5: USING THE ARTHRITIS PADS
Arthritis Pads are for single person use only.
REVITIVE comes with two pairs of reusable (up to 20-30 applications) Arthritis (Electrode) Pads that can be used to deliver Electrical Muscle Stimulation (EMS) to strengthen the muscles at the front of your thigh (quadriceps).
Strengthening your quadriceps muscles may help to support and stabilise the knee, reduce stress on the knee and relieve knee pain.
The Arthritis Pads cannot be used at the same time as the foot-pads. Using
the Arthritis Pads will automatically turn the foot-pads off.
It is recommended that the Arthritis Pads be used on the affected leg (knee) for 20 minutes, once a day, 5 days per week. Find a routine that works for you, perhaps using the Arthritis Pads Monday - Friday and not at the weekend.
Do not use REVITIVE more than 6 sessions of 30 minutes (or the equivalent) per day. This may result in muscle fatigue.
d Plug the other end of the Arthritis Pad cord into
the opening on the side of REVITIVE.
e Remove the plastic liner from the Arthritis Pads, by
peeling it off carefully. Replace the Arthritis Pads onto the plastic liner after use.
f Gently press one Arthritis Pad across your thigh (1), a
few centimetres away from your groin area, with the adhesive side against the skin.
1
a Clean and dry the thigh where you will place the Arthritis Pads (see next
page for positioning instructions).
b Make sure REVITIVE is switched off. If the control
panel is still illuminated, press the power button and turn REVITIVE off.
c Attach the ends of the Arthritis Pads to the Arthritis Pad cord.
Press connectors until there is no gap
between them. Leaving a gap may cause discomfort.
20
2910_IFU02_12517049.indd 21 22/07/2016 09:49
2
g Gently press second Arthritis Pad across your thigh
(2), a few centimetres from the top of your kneecap, with the adhesive side against the skin.
The Arthritis Pads do not need to be lined up exactly with each other. They need to be in a position that is comfortable and that allows you to see your thigh muscles contract & relax.
REVITIVE Arthritis Pads are designed to be used whilst seated.
21
Page 12
Instructions for use
NB The Arthritis Pads are only suitable for use with the REVITIVE Arthritis­Knee device and will not be effective for any other areas of the body.
Do NOT use the Arthritis Pads on your head, face, neck or chest.
h Press the power button on the device or on the
remote control to turn REVITIVE on.
You will hear a beep and the display lights will light up. The time will read 20 minutes and Intensity will read 00.
To turn off REVITIVE at any time during operation, press the power button on the device or the remote control.
Adjusting the intensity level - At the start of therapy there will be 30 seconds of stimulation to help you set the right intensity level. This is followed by a rest period of 30 seconds, when you will feel no stimulation.
For the remainder of the therapy session you will feel 10 seconds of stimulation, followed by a 50 second rest period which repeats until the end of therapy.
Adjust the intensity when you can feel the stimulation being applied to your thigh.
It is not possible to adjust the intensity during the rest period.
j Press the (+) Intensity Control once to start the
stimulation.
Continue to increase the intensity until you can feel or see the muscles in your thigh contract and relax. You can set the intensity between 1-99.
You can use the buttons on the device or remote control.
i Set the time for the number of minutes you want to
use REVITIVE. The default time for the Arthritis Pads
Press the (+) button to increase the level. Press the (–) button to decrease the level.
is 20 minutes. You can use the buttons on the device or
remote control. Press the (+) Time Contol to increase the time. Press the (-) Time Contol to decrease the time.
Use an intensity that provides you with strong but comfortable muscle contractions and note that the intensity required may vary from day to day - eg using it on 25 one day, does not necessarily mean you will need to use it at 25 the next day. The required intensity could be higher or lower than previous usage.
It is normal to feel tingling or varying sensations in your thigh muscles.
22
23
2910_IFU02_12517049.indd 23 22/07/2016 09:49
Page 13
Instructions for use
If you feel pain or discomfort - lower the intensity of the Arthritis Pads using either the (-) Intensity Control on the control panel of the device or on the remote control.
o Replace the Arthritis Pads onto their plastic liner
after use.
You can also choose to stop therapy before the end of the 20-minute therapy session to avoid muscle fatigue. You can do this by pressing the power button on the control panel of the device or on the remote control.
k The time control/display will start to count down in
minutes as REVITIVE cycles through its program.
l When REVITIVE times out, the time display
reads 00 and you will hear three beeps.
m REVITIVE switches itself off automatically.
n Ensuring that the device has switched itself off, then
gently peel the Arthritis Pads from the skin.
p Unplug the Arthritis Pad cord from the opening
on REVITIVE. You do not need to detach the Arthritis Pads from
the Arthritis Pads cords.
Do not pull the Arthritis Pads off the skin using
the Arthritis Pad cords as this may damage the cords.
24
2910_IFU02_12517049.indd 24-25 22/07/2016 09:49
25
Page 14
Do not clean with chemicals.
Store out of direct sunlight.
Instructions for use
STEP 6: RECHARGING THE BATTERY
REVITIVE can be powered from either the rechargeable battery or by connecting it to an electrical socket.
a Locate the AC/DC power adaptor.
b Plug the small end of the power cord into the
opening on the side of the REVITIVE.
STEP 7: CLEANING AND STORING REVITIVE
a Ensure REVITIVE is switched off.
b Wipe down the foot-pads with a soft damp cloth.
Do not immerse REVITIVE in water.
c Store REVITIVE in a cool, dry and dust-free location.
Store out of direct sunlight.
d Place Arthritis Pads onto the plastic liner.
e Store the Arthritis Pads in a dry and ventilated location.
You do not have to detach the Arthritis Pads from the Arthritis Pad cords to store them.
c Plug the power adaptor into the nearest electrical socket. This will start charging
the (integrated) rechargeable battery.
The battery indicator will illuminate constantly to indicate the device is charging
and to show the battery level, as below.
Low Battery
A full charge will take approximately 3 hours.
d Once fully charged, the full battery indicator will stay illuminated until the power
adaptor is unplugged from the device or the electrical socket.
A full charge enables approximately 3½ hours of use (the equivalent of one 30
minute session each day for one week).
Please note you can still use the device while it is charging but the battery
indicator will not show.
26
Full Battery
Replacing the Batteries
The remote control requires two AAA 1.5V batteries to operate.
Replacing the Arthritis Pads
The Arthritis Pads are reusable up to 20-30 applications. If the Arthritis Pads start to lose their adhesiveness, rub a few drops of water into the surface.
Replacement Arthritis Pads for use with REVITIVE Arthritis-Knee can be obtained from the REVITIVE website, www.revitive.com.
27
2910_IFU02_12517049.indd 26-27 22/07/2016 09:49
Page 15
Troubleshooting
Problem Possible Cause Solution
REVITIVE is on (lights illuminated on the LED display) but I cannot feel the electrical stimulation through the foot-pads.
Not placing both bare feet on the foot-pads at the same time.
Your feet may be dry. Moisturise the soles of your feet
You may be dehydrated. Drink plenty of water before
The intensity level may be on too low a setting.
Arthritis Pad cord is connected to device.
If, having tried the solutions above, you still cannot feel the stimulation:
Ensure that your feet are bare and each foot is placed on each of the foot-pads at the same time. Keep increasing the intensity up to a maximum of 99 until you feel the stimulation.
to improve conductivity and stimulation and try the procedure again. You may also have to increase the intensity level.
and after using the device. The device uses your body to create the electrical circuit. Water is an excellent conductor of electricity and if your body is less hydrated (below 60%) then the stimulation may be less, therefore it is important to always remain well hydrated.
This is a very safe device. Keep increasing the intensity level towards 99 until you feel the stimulation. You may find that you have to increase the intensity level as you get used to the therapy. The aim is not to get to 99 but to find a setting that is comfortable for you.
Disconnect the Arthritis Pad cord - the foot-pads will not work while the Arthritis Pad cord is plugged in.
Test the device by placing one hand across both foot-pads at the same time (The heel end of the foot-pads is easiest). With your other hand, and starting from zero, increase the intensity level until you can feel the stimulation. If you can feel the stimulation through your hand then the device is working. If on 99 you still cannot feel the stimulation then please contact your authorised dealer.
Problem Possible Cause Solution
No power or lights to REVITIVE when it is switched on.
REVITIVE is not vibrating.
The IsoRocker is not "rocking".
My legs are aching after treatment.
When using the IsoRocker on a hard floor it makes a tapping noise.
AC Adaptor not switched on at the electrical socket or AC Adaptor not plugged into device properly.
Flat battery Plug device into an electrical
REVITIVE IS NOT DESIGNED TO VIBRATE.
It is the muscles in your legs that cause the "rocking", the ISOROCKER IS NOT MECHANICAL.
The device is too far in front of you.
The IsoRocker system is not enabled.
You may have the intensity on too high a setting and your muscles are being overworked.
Incorrect positioning of the device or too high an intensity level.
Check electrical socket power is switched on and the AC Adaptor is plugged into the device correctly. If still not working – contact your authorised dealer.
socket and charge battery or run using mains power option.
The ISOROCKER will only rock when increased to an intensity which causes sufficient calf muscle contraction. It may be that you cannot comfortably increase the intensity high enough until you get used to the sensation. It is important that intensity is adjusted to a level that is comfortable.
Sit with your knees at a 90 degree angle.
Ensure IsoRocker Disabler is “flat”, in centre position (see page 14).
Leave adequate time after each treatment to allow the muscles to recover (just like after vigorous exercise!). On your next session start on a lower setting (where you can feel the mild electrical stimulation and it is comfortable) and reduce the duration until your muscles have acclimatized to the stimulation.
Adjust the positioning of the device or lower the intensity level to reduce the device tapping. Alternatively use a floor mat under the device to cushion the sound.
28
2910_IFU02_12517049.indd 28-29 22/07/2016 09:49
29
Page 16
Troubleshooting
Technical Specifications
Problem Possible Cause Solution
When using the Arthritis Pads, my thigh muscles are aching before the end of the treatment.
Your muscles may not yet be strong enough to carry out a full 20 minute session.
Turn the device off and end the session before the end of the 20 minutes. On your next session start on a lower setting (where you can feel the mild electrical stimulation and it is comfortable) and reduce the duration until your muscles have acclimatized to the stimulation.
When using the Arthritis Pads, I can’t feel the stimulation in
Therapy has started and has reached the rest period.
You will need to wait for up to 50 seconds before the stimulation will start again.
my thigh.
For more information including Frequently Asked Questions,
please visit www.revitive.com
Name of product REVITIVE Arthritis-Knee
Model 2836AA
Frequency 1Hz – 50Hz
Output current Max 15mA
Weight 1.5kg
Dimensions (with IsoRocker enabled and disabled)
Power consumption 5W
AC adaptor CE Approved
Power source Input (adaptor used) Output
Battery & Battery Life Li-Ion / 500 recharge cycles
Contraindications
This describes situations where you should not use the REVITIVE
enabled: 355(W)x355(H)x75(D) mm disabled: 355(W)x355(H)x90(D) mm
100-240V AC ( ), 50/60Hz, 0.18A 5V ( ) DC ,1.0A
30
Warnings and Cautions Make sure you understand these before using the REVITIVE
Power
Time Remaining
Intensity Level
Battery Indicator
IsoRocker Disabler
Center Positive Polarity
Type BF medical electrical equipment
Class II medical electrical equipment double insulated
31
2910_IFU02_12517049.indd 30-31 22/07/2016 09:49
Page 17
38 KH
z
Technical Specifications
Your international two year warranty
Complies with European Medical Devices Directive (93/42/EEC)
The Waste Electrical and Electronic Equipment Directive (WEEE Directive). At the end of the product lifecycle, do not throw this product into normal household garbage, but take it to a collection point for the recycling of electronic equipment
Legal manufacturer of the device
Keep dry
Consult instructions for use
Device LOT and serial number including year (YYYY) and month (MM) of manufacture can be found on the box and back of unit
Humidity and temperature limit for storage
Indoor Use Only
Do not disassemble
Item number
Date of manufacture
Use-by date
LOT
30%
The Remote Control replicates the controls found on the device
Operating Frequencies
#YYYYMMXXXXX
90%
10°C
40°C
REF
It is important to retain the retailer’s receipt as proof of purchase. Staple your receipt to this back cover for future reference.
Please quote the following information if the product develops a fault. These numbers can be found on the base of the product:
Model no: ............................................................
Lot no: .................................................................
All REVITIVE devices are individually tested before leaving the factory. In the unlikely event of any device proving to be faulty within 30 days of purchase, it should be returned to the place of purchase for it to be replaced.
If the fault develops after 30 days and within 24 months of original purchase, you should contact your local distributor quoting model number and LOT number on the product, or write to your local distributor at the address shown.
You will be asked to return the product (in secure, adequate packaging) to the address shown with a copy of proof of purchase.
Subject to the exclusions set out below (see Exclusions) the faulty device will then be replaced and dispatched usually within 14 working days of receipt.
If, for any reason, this item is replaced during the 2 year guarantee period, the guarantee on the new item will be calculated from the original purchase date. Therefore, it is vital to retain your original till receipt or invoice to indicate the date of initial purchase.
To qualify for the 2 year guarantee, the device must have been used according to the manufacturer’s instructions supplied.
EXCLUSIONS:
1 Actegy Ltd, manufacturer of REVITIVE
devices, shall not be liable to replace the goods under the terms of the guarantee where:
The fault has been caused or is
attributable to accidental use, misuse, negligent use or used contrary to the manufacturer’s recommendations or where the fault has been caused by power surges or damage caused in transit.
The device has been used on a voltage
supply other than that stated on the product or used with a power adaptor other than the one supplied with the product.
Repairs have been attempted by persons
other than our service staff (or authorised dealer).
The device has been used for hire
purposes or non-domestic use.
The device is second hand.
2 Actegy Ltd are not liable to carry out any
type of servicing work, under the guarantee.
3 Accessories such as Arthritis Pads are not
covered by the guarantee.
4 Batteries and damage from leakage are not
covered by the guarantee.
5 This guarantee does not confer any
rights other than those expressly set out above and does not cover any claims for consequential loss or damage. This guarantee is offered as an additional benefit and does not affect your statutory rights as a consumer.
0 to 25m at horizontal
Operating Range Distance
32
2910_IFU02_12517049.indd 32-33 22/07/2016 09:49
0 to 18m at +/- 30 degree angle from the horizontal.
To activate your free 2 year warranty please register your device at the
REVITIVE website (see warranty/guarantee card).
If you prefer you can also post your registration.
33
Page 18
Notes
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
34
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
.......................................................................................................................................
35
2910_IFU02_12517049.indd 34-35 22/07/2016 09:49
Page 19
®
Actegy Ltd REFLEX, Cain Road Bracknell, RG12 1HL United Kingdom
Helpline: +44 (0)845 871 5989
Copyright © 2016 Actegy Ltd. All rights reserved. ACTEGY® and REVITIVE™ and Circulation Booster™ are registered trademarks of Actegy Ltd. The REVITIVE product is a proprietary design and is protected by applicable design laws.
AW 2910_IFU02_12517049 P.D 08.2016
36
2910_IFU02_12517049.indd 36 22/07/2016 09:49