25
UltraFill User Manual
emc InFormatIon
guIdance and manuFacturer’S declaratIon – electromagnetIc emISSIonS: This device is
intended for use in the electromagnetic environment specied below. The user of this device should make sure it is used in such an
environment.
emISSIonS teSt complIance electromagnetIc envIronment – guIdance
RF emissions
CISPR 11
Group 1 The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
NA No test requirements for devices that operate at 115 VAC 60 Hz.
Voltage uctuations/Flicker emissions
IEC 61000-3-3
NA
guIdance and manuFacturer’S declaratIon – electromagnetIc ImmunIty: This device is intended for use in
the electromagnetic environment specied below. The user of this device should make sure it is used in such an environment.
ImmunIty teSt Iec 60601
teSt level
complIance level electromagnetIc envIronment
– guIdance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If oors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical Fast Transient/Burst
IEC 61000-4-4
±2 kV for power supply lines ±2 kV for power supply mains Mains power quality should be that of a typical
home or hospital environment.
Surge
IEC 61000-4-5
±1 kV Line to Line ±1 kV line to line Mains power quality should be that of a typical
home or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in UT) for 0.5
cycle
40% U
T
(60% dip in UT) for 5 cycles
70% UT
(30% dip in UT) for 25 cycles
<5% UT
(>95% dip in UT) for 5 sec
<5% U
T
(>95% dip in UT) for 0.5
cycle
40% U
T
(60% dip in UT) for 5 cycles
70% UT
(30% dip in UT) for 25 cycles
<5% UT
(>95% dip in UT) for 5 sec
Mains power quality should be that of a typical
home or hospital environment. If the user of
the device requires continued operation during
power mains interruptions, it is recommended
that the device be powered from an
uninterruptible power supply.
Power frequency (60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should be at
levels characteristic of a typical location in a
typical hospital or home environment.
Note: UT is the a.c. mains voltage prior to application of the test level.