Respironics Respironics_V60_Ventilator_User_Manual_ Model AF Series Lab Ovens Operating Manual

User Manual
For Technical Support and Customer Service, contact:
USA and Canada: 1-800-722-9377 International: www.philips.com. Select your location to access contact information.
Manufacturer
Respironics California, LLC 2271 Cosmos Court Carlsbad, CA 92011 USA
Email and web addresses
respironics.service@philips.com respironics.clinical@philips.com www.philips.com/healthcare
Authorized European representative
Respironics Deutschland GmbH&Co. KG Gewerbestrasse 17 82211 Herrsching Germany +49-8-15-29-30-60
EC REP
Australian sponsor
Philips Electronics Australia Ltd 65 Epping Road North Ryde, NSW 2113 Australia
You can find the most current version of this user manual here: http://www.philips.com/hrcmanuals
Table of contents
1. Warnings, cautions, and notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Preparing for ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Operation in high flow therapy (HFT) . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Alarms and messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Care and maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
First-time installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Communications interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Diagnostic mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
2. Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
3. General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Contraindications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
About CO
Potential side effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Physical description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
About the optional backup battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
About the graphical user interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Starting up the ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Shutting down the ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
rebreathing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
2
Patient circuits, masks/patient interfaces, and accessories . . . . . . . 3-4
Ventilator unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Navigating the graphical user interface . . . . . . . . . . . . . . . . . . . . 3-12
4. Principles of operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
System operational overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Pneumatic system operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Breath delivery characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Control variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Triggering, cycling, and leak adaptation . . . . . . . . . . . . . . . . . . . . 4-2
Baseline pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Pressure rise time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Negative pressures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Oxygen concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Auto-Trak Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Cycling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Leak adaptation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Auto-Trak+ (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
iii
High flow therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Ventilation modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
CPAP mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
PCV mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
S/T mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
AVAPS mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
PPV mode (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Oxygen mixing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
5. Setting up the ventilator for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Connecting oxygen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Installing an oxygen analyzer/monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Connecting to AC power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Installing the patient circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Connecting external devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Before placing a patient on the ventilator. . . . . . . . . . . . . . . . . . . . . . . . 5-7
Verify ventilator operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Running alarm tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
High Inspiratory Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Low Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Patient Disconnect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Patient Circuit Occluded . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Using the ventilator for intra-hospital transport . . . . . . . . . . . . . . . . . . . 5-10
Storing the ventilator between patient use . . . . . . . . . . . . . . . . . . . . . . 5-11
6. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Changing the mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Changing control settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Making batch setting changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Changing individual ventilator settings . . . . . . . . . . . . . . . . . . . . . . 6-4
Using the Ramp Time function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Using the 100% O
Using PPV. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
About Max V and Max P alarms and alarm limits . . . . . . . . . . . . . . . 6-7
Guidelines for using PPV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Changing alarm settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Selecting the mask and exhalation port . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Running the exhalation port test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Other functions: the Menu window . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Loudness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Mask/Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Vent Info (ventilator information) . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Screen Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Auto-Trak+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
Help function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
Table of modes and control settings . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
2
iv
7. High flow therapy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Circuit setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
High flow nasal cannula setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Connecting to a circuit with an FEP (filter exhalation port) installed . . 7-2
Connecting directly to a 22 mm circuit . . . . . . . . . . . . . . . . . . . . . 7-2
Changing from an NIV mode to high flow therapy . . . . . . . . . . . . . . . . . 7-3
Viewing and pausing the HFT graph. . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Changing from high flow therapy to an NIV mode . . . . . . . . . . . . . . . . . 7-4
HFT alarms and messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
8. Patient monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Display conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Table of monitored parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Scaling the waveform axes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Freezing and unfreezing waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
9. Alarms and messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Responding to alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Setting alarm loudness. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Silencing alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Resetting alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Manually resetting alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Clearing autoreset alarms from the Alarms list . . . . . . . . . . . . . . . . 9-5
Hiding/displaying alarm messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Alarms and other messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
10. Care and maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Exterior and touchscreen cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Approved cleaning agents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2
Cleaning instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2
Exterior and touchscreen disinfection . . . . . . . . . . . . . . . . . . . . . . . . 10-2
Approved disinfecting agents . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Disinfection instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Bacteria filter, patient circuit, and other accessories . . . . . . . . . . . . . . 10-3
Preventive maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Replacing the air inlet filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Cleaning or replacing the cooling fan filter . . . . . . . . . . . . . . . . . 10-6
Removing and replacing the battery . . . . . . . . . . . . . . . . . . . . . . 10-7
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
Storage between patient use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
Service and repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
Repacking and shipping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
11. Technical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
Control settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
Patient data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Menu window settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Diagnostic mode functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
v
Physical characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Environmental specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
Pneumatic specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
Electrical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
Accessory requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
Other specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
A. First-time installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Unpacking and inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Mounting the ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Installing the optional battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Installing oxygen inlet connector and AC power cord . . . . . . . . . . . . . . . . A-6
Installing the oxygen manifold kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8
Verifying ventilator operation and audible alarm . . . . . . . . . . . . . . . . . . . A-8
Configuration and screen calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8
B. Communications interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
RS-232 serial and analog I/O port . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Pinout of connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Communications protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
Commands and transmission conventions . . . . . . . . . . . . . . . . . . . B-4
Using Philips IntelliBridge or VueLink . . . . . . . . . . . . . . . . . . . . . . . . B-15
Using Philips monitors and the IntelliBridge Open Interface . . . . . B-15
Using Philips monitors and the VueLink Open Interface . . . . . . . . B-15
Data display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-16
Remote alarm port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-19
C. Parts and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
Masks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
HFT interfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Exhalation ports. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
O
analyzer/monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
2
Patient breathing circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Humidification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Bacteria filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
Operator maintenance parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
Other parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
D. Regulatory compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Electromagnetic compatibility (EMC). . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Electromagnetic compatibility declaration . . . . . . . . . . . . . . . . . . . . . . D-1
Electromagnetic emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Electromagnetic immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
WEEE recycling directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-5
Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-5
vi
E. Diagnostic mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
Entering the diagnostic mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
System settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4
Date/Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6
Pressure Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-7
Restore Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-8
Software Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-9
Baud Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-10
Alarm Volume Escalation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-11
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-12
Significant Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-12
Touchscreen calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-14
Exiting the diagnostic mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-14
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Glossary-1
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index-1
vii
(This page is intentionally blank.)
viii

Chapter 1. Warnings, cautions, and notes

Before using the Respironics V60/V60 Plus Ventilator on a patient, familiarize yourself with this user manual, particularly the safety considerations listed. Be aware, however, that this manual is a reference only. It is not intended to supersede your institution’s protocol regarding the safe use of assisted ventilation.

Definitions WARNING: Alerts the user to the possibility of injury, death, or other serious adverse

reactions associated with the use or misuse of the device.
CAUTION: Alerts the user to the possibility of a problem with the device
associated with its use or misuse, such as device malfunction, device failure, damage to the device, or damage to other property.
NOTE: Emphasizes information of particular importance.

General WARNING: An alternative means of ventilation shall be available whenever the

ventilator is in use. If a fault is detected in the ventilator, disconnect the patient from it and immediately start ventilation with such a device. The ventilator must be removed from clinical use and serviced by authorized service personnel.
WARNING: Use the Respironics V60/V60 Plus Ventilator on spontaneously breathing
patients only. It is an assist ventilator and is intended to augment the ventilation of a spontaneously breathing patient. It is not intended to provide the total ventilatory requirements of the patient.
WARNING: We do not recommend you use the Respironics V60/V60 Plus Ventilator
on patients who require ventilation at predetermined tidal volumes. The ventilator provides continuous positive airway pressure (CPAP) and positive pressure ventilation (S/T, PCV, and AVAPS, and PPV) and is indicated for assisted ventilation only. These modes do not provide ventilation with guaranteed tidal volume delivery.
1-1
Warnings, cautions, and notes
WARNING: We do not recommend you use AVAPS on patients who require rapid and
frequent IPAP adjustments to maintain a consistent tidal volume. AVAPS, a volume targeted mode, changes the IPAP setting in order to achieve the target tidal volume. During AVAPS setup, there may be a period of time before the target tidal volume is achieved. AVAPS is ideal for more stabilized patients.
WARNING: To reduce the risk of CO
rebreathing, make sure EPAP pressures and
2
exhalation times are sufficient to clear all exhaled gas through the exhalation port. In noninvasive ventilation continuous air flow through the port flushes exhaled gases from the circuit. The ability to completely exhaust exhaled gas from the circuit depends on the EPAP setting and I:E ratio. Higher tidal volumes further increase the volume of CO
rebreathed
2
by the patient.
WARNING: To reduce the risk of CO
rebreathing, monitor the patient for changes in
2
respiratory status at the start of ventilation and with each change in ventilator settings, circuit configuration, or patient condition. Pay attention to ventilator alarms that warn of increased CO
rebreathing risk.
2
WARNING: To ensure accuracy of oxygen administration and to monitor for the
presence of contamination (incorrect gas connected), use an external oxygen monitor to verify the oxygen concentration in the delivered gas.
WARNING: To reduce the risk of fire, use the ventilator in well-ventilated areas away
from flammable anesthetics. Do not use in a hyperbaric chamber or other similarly oxygen-enriched environments. Do not use near an open flame.
WARNING: To reduce the risk of electric shock from liquid entering the device, do
not put a container filled with a liquid on the ventilator.
WARNING: To reduce patient risk of oxygen toxicity, keep free-flowing oxygen away
from air inlet of ventilator.
WARNING: The nurse call/remote alarm should be considered a backup to the
ventilator’s primary alarm system.
WARNING: To ensure that the alarm will be heard, make sure the alarm loudness is
adequate and avoid blocking the alarm speakers beneath the ventilator. WARNING: Do not leave the ventilator unattended when stationed on an incline. WARNING: The V60/V60 Plus Ventilator may cause radio interference or may disrupt
the operation of nearby equipment. It may be necessary to take mitigation
measures, such as re-orienting or relocating the ventilator or shielding
the location. WARNING: Use of non-approved accessories, transducers or cables may increase
EMC emissions or decrease the EMC immunity performance of the
equipment.
CAUTION: Federal law (USA) restricts this device to sale by or on the order of a
physician. CAUTION: The Respironics V60/V60 Plus Ventilator is designed to operate in the
temperature range of 5 to 40 ºC (41 to 104 ºF). To minimize the risk
of overheating the device, do not operate adjacent to heaters or other
heat sources. NOTE: The displays shown in this manual may not exactly match what you
see on your own ventilator.
1-2
Warnings, cautions, and notes
NOTE: Pressures are indicated on the ventilator in cmH2O. Millibars and
hectopascals (hPa) are used by some institutions instead. Since 1 millibar equals 1 hPa, which equals 1.016 cmH be used interchangeably.
NOTE: The ventilator is not intended for use as an ambulance transport
ventilator or as an Automatic Transport Ventilator as described by the American Hospital Association and referenced by the FDA. It is intended to allow the patient to be transported within the hospital setting using a cart to move the ventilator.
NOTE: When attachments or other components or subassemblies are added
to the ventilator breathing system, the pressure gradient across the ventilator breathing system, measured with respect to the ventilator outlet, may increase.
NOTE: To ensure the correct performance of the ventilator and the accuracy
of patient data, use only Respironics-approved accessories with the ventilator. See Appendix C, “Parts and accessories”.
NOTE: This Respironics V60/V60 Plus Ventilator and its recommended
accessories that have patient contact are not made with natural rubber latex.
NOTE: If an alarm persists for no apparent reason, discontinue ventilator use
and contact Philips.
NOTE: If you detect any unexplained changes in the performance or visual
displays of the ventilator, discontinue ventilator use and contact Philips.
NOTE: The Respironics V60/V60 Plus Ventilator does not support automatic
record keeping.
NOTE: All ventilator mode and alarm settings, alarm messages and
significant events are retained and automatically logged, even when power is lost.
O, the units may
2

Preparing for ventilation

1-3
WARNING: Connect the ventilator only to an appropriate medical-grade oxygen
source.
WARNING: To reduce the risk of hypoxia, connect only oxygen to the high-pressure
connector at the rear of the ventilator.
WARNING: To reduce the risk of fire, do not use a high-pressure oxygen hose that is
worn or contaminated with combustible materials like grease or oil.
WARNING: The Respironics V60/V60 Plus Ventilator is designed to use ambient air
and high pressure 100% oxygen. No other gases should be used. WARNING: Do not use the ventilator with helium or mixtures with helium. WARNING: Do not use the ventilator with nitric oxide. WARNING: To prevent possible asphyxia and to reduce the risk of CO
take these precautions with respect to mask and exhalation port use:
- Use only an oro-nasal mask with an anti-asphyxia valve or a nasal mask for noninvasive ventilation.
rebreathing,
2
Warnings, cautions, and notes
- Do not occlude the exhalation port.
- Turn on the ventilator and verify that the port is operational before application. Pressurized gas from the ventilator should cause a continuous flow of air to exhaust from the leak port, flushing exhaled gas from the circuit.
- Never leave the mask on the patient while the ventilator is not operating. When the ventilator is not operating, the exhalation port does not allow sufficient exhaust to eliminate CO Substantial CO
rebreathing may occur.
2
from the circuit.
2
WARNING: The patient’s exhaled volume can differ from the measured exhaled
volume due to leaks around the mask during noninvasive ventilation.
WARNING: To ensure normal air circulation and exchange, do not cover or block the
ports on the ventilator. Do not block the air inlet panel on the right side of the ventilator.
WARNING: Do not cover or position the ventilator so as to adversely affect its
operation or performance. Use the V60/V60 Plus in an upright position that does not block the air inlet.
WARNING: To reduce the risk of the device overheating and possible burn injury, do
not block the fan intake at the rear of the ventilator.
WARNING: To prevent possible patient injury and possible water damage to the
ventilator, make sure the humidifier is set appropriately.
WARNING: When using a humidifier, always use either a circuit with a water trap or a
heated wire circuit to minimize patient risk from condensate in the circuit.
WARNING: To prevent the possibility of inadequate humidification, pay close
attention to the humidifier’s functioning when operating the ventilator at an ambient temperature > 30 ºC (86 ºF). The ventilator warms the air delivered to the patient above ambient temperature, which may impair the humidifier’s performance.
WARNING: To reduce the risk that the patient will aspirate condensed water from the
breathing circuit, position any humidifier lower than both the ventilator and the patient.
WARNING: To prevent possible patient injury and equipment damage, do not turn the
humidifier on until the gas flow has started and is regulated. Starting the heater or leaving it on without gas flow for prolonged periods may result in heat build-up, causing a bolus of hot air to be delivered to the patient. Circuit tubing may melt under these conditions. Turn the heater power switch off before stopping gas flow.
WARNING: To reduce the risk of fire, use only patient circuits intended for use in
oxygen-enriched environments. Do not use antistatic or electrically conductive tubing.
WARNING: To prevent patient or ventilator contamination, always use a main flow
bacteria filter on the patient gas outlet port. Filters not approved by Respironics may degrade system performance.
WARNING: During ventilation, patient exhalate is released into room air. Use of a
patient circuit with a filter on its exhalation port is recommended.
WARNING: To reduce the risk of bacterial contamination or damage, handle bacteria
filters with care.
1-4
Warnings, cautions, and notes
WARNING: Any additional accessories in the patient circuit may substantially
increase flow resistance and impair ventilation.
WARNING: Avoid adding resistive circuit components on the patient side of the
proximal pressure line. Such components may defeat the disconnect alarm.
WARNING: To reduce the risk of strangulation from patient tubing, use a tubing
support arm and secure the proximal pressure line with clips.
WARNING: To reduce the risk of electric shock, connect the ventilator to an AC
supply mains with protective earth only.
WARNING: Do not use extension cords, adapters, or power cords with the ventilator
that are not approved by Respironics.
WARNING: To prevent unintentional disconnection of the power cord, always use the
correct, Philips-supplied power cord and lock it into place with the power cord retainer before you switch the ventilator on. The retainer is designed to hold the connector end of the Philips-supplied cord securely in place.
WARNING: The V60/V60 Plus Ventilator should not be positioned in a way that makes
it difficult to disconnect from mains power if necessary. Disconnect from supply mains by removing the power cord from the wall outlet.
WARNING: To reduce the risk of electric shock, regularly inspect the AC power cord
and verify that it is not frayed or cracked.
WARNING: To reduce the risk of strangulation, route the power cord to avoid
entanglement.
WARNING: To reduce the risk of power failure to the ventilator, pay close attention to
the battery’s charge level. The battery’s operation time is approximate and is affected by ventilator settings, discharge and recharge cycles, battery age, and ambient temperature. Battery charge is reduced at low ambient temperatures or in situations where the alarm is continuously sounding.
WARNING: Always check the status of the oxygen cylinders before using the
ventilator during transport.
WARNING: Provide external oxygen monitoring to minimize patient risk in case of O
supply loss or ventilator failure.
WARNING: To ensure the ventilator’s safe operation, always verify ventilator
operation as described in “Verify ventilator operation” on page 5-7 before using the ventilator on a patient. If the ventilator fails any tests, remove it from clinical use immediately. Do not use the ventilator until necessary repairs are completed and all tests have passed.
WARNING: To prevent possible patient injury due to nonannunciating alarms, verify
the operation of any remote alarm device before use.
WARNING: To prevent possible patient injury, always return alarm settings to
hospital-standard values after verifying ventilator operation.
WARNING: Manufacturer default settings are not appropriate for all patients. Prior to
using the ventilator, verify that the current alarm settings or defaults are appropriate for each particular patient.
CAUTION: To prevent possible damage to the ventilator, ensure that the
connection to the oxygen supply is clean and unlubricated, and that there is no water in the oxygen supply gas.
2
1-5
Warnings, cautions, and notes
CAUTION: For 120 V equipment, grounding reliability can only be achieved
when it is connected to an equivalent receptacle marked “hospital only” or “hospital grade.”
CAUTION: Oxygen hose configurations using SIS connectors generate higher
resistance to flow. Therefore, a minimum supply pressure of 53 psig is recommended when adding supplemental O adapters such as the O2 transport manifold.
accessories with SIS
2

Operation WARNING: To prevent possible patient injury, avoid setting alarm limits to extreme

values, which can render the alarm system useless.
WARNING: PPV limits are not intended to be the primary ventilator alarms and should
not be substituted for the alarms found in the Alarm Settings window.
WARNING: To prevent the delivery of excessive pressure or volume, set the PPV
limits appropriately. Delivery of excessive pressure or volume can occur from a sudden increase in mask leak, inappropriate settings, or a plugged or kinked proximal pressure line. Conversely, insufficient treatment may result if limits are set too low.
WARNING: Nebulization or humidification can increase the resistance of breathing
system filters. When using a nebulizer or humidifier, monitor the breathing system filter frequently for increased resistance and blockage.
WARNING: Using a jet nebulizer can cause inadvertent alarms and affect the
accuracy of delivered FiO pneumatic nebulizers to 10 L/min or use a vibrating mesh nebulizer.
. To reduce patient risk, limit the flow of
2

Operation in high flow therapy (HFT)

WARNING: When transitioning from a high flow therapy interface to an NIV mask,
ensure that an exhalation port is placed in the circuit and is unobstructed to reduce the risk of CO
WARNING: When transitioning from ventilation to high flow therapy, remove the NIV
mask and use only a Philips-approved high flow patient interface to minimize pressure build-up and patient discomfort.
WARNING: When transitioning from high flow therapy to ventilation, remove the nasal
cannula as these are restrictive and may defeat alarms such as patient disconnect. Using a nasal cannula in an NIV mode may lead to hypercarbia due to the inability to provide pressure support.
WARNING: Patient alarms are not available during high flow therapy (HFT) as the
therapy uses an open system. A nasal cannula occupies only a portion of the nares and patients can breathe through their mouth, which prevents estimation of patient parameters such as tidal volume, respiratory rate, pressure, and minute ventilation. Provide external monitoring, including oximetry, to inform the clinician of a change in the patient's condition.
rebreathing.
2
1-6
Warnings, cautions, and notes
WARNING: During high flow therapy (HFT), verify that an occlusive patient interface
is not being used. Occlusive patient interfaces include a cannula fully sealed within the nares, an NIV mask, or a direct connection to a tracheostomy tube or endotracheal tube. Remove any occlusive interface immediately as this may expose the patient to unintended high pressures.

Alarms and messages WARNING: If AC power fails and the backup battery is not installed or is depleted, an

audible and visual alarm annunciates for at least 2 minutes. Immediately discontinue ventilator use and secure an alternative means of ventilation. As in most ventilators with passive exhalation ports, when power is lost, sufficient air is not provided through the circuit and exhaled air may be rebreathed.

Care and maintenance

WARNING: To reduce the risk of electric shock, power down the ventilator and
disconnect it from AC power before cleaning, disinfecting, or servicing it.
WARNING: To prevent patient or ventilator contamination, inspect and replace the
main flow bacteria filter between patients and at regular intervals (or as stated by the manufacturer).
WARNING: To prevent possible patient injury, inspect and verify the proper operation
of the exhalation port regularly during use.
WARNING: To reduce the risk of fire, explosion, leakage, or other hazard, take these
precautions with respect to the battery:
- Do not attempt to disassemble, open, drop, crush, bend or deform, insert foreign objects into, puncture, or shred the battery pack; modify or remanufacture it; immerse or expose it to water or other liquids; expose it to fire, excessive heat (including soldering irons); or put it in a microwave oven.
- Replace the battery only with another battery specified by the manufacturer.
- Follow all instructions for proper use of the battery.
- Do not short-circuit the battery or allow metallic or conductive objects to contact the battery connector housing.
- Use the battery with the Respironics V60/V60 Plus Ventilator only.
WARNING: Modification of the V60/V60 Plus Ventilator and associated equipment is
not permitted and may compromise ventilator operation and patient safety. Service should only be performed by qualified service personnel.
WARNING: This product consists of devices that may contain mercury, which must be
recycled or disposed of in accordance with local, state, or federal laws. (Within this system, the backlight lamps in the monitor display contain mercury.)
CAUTION: Do not attempt to sterilize or autoclave the ventilator. CAUTION: To prevent possible damage to the ventilator, use only those cleaning
and disinfecting agents listed in this manual.
1-7
Warnings, cautions, and notes
CAUTION: To prevent possible damage to the ventilator, do not drip or spray any
liquids directly onto any surface including the front panel, touchscreen, and navigation ring.
CAUTION: Never clean or disinfect the touchscreen with an abrasive brush or
device, since this will cause irreparable damage.
CAUTION: To avoid introducing foreign matter into the ventilator and to ensure
proper system performance, change the air inlet filter at regular intervals (or as stipulated by your institution).
CAUTION: To ensure proper system performance, use a Respironics-approved air
inlet filter.
CAUTION: Because some environments cause a quicker collection of lint and dust
than others, inspect the filters more often when needed. The air inlet filter should be replaced; the cooling fan filter should be cleaned.
CAUTION: To prevent possible damage to the ventilator, always ship it with the
original packing material. If the original material is not available, contact Philips to order replacements.

First-time installation WARNING: Never attempt to disconnect or connect the battery during operation.

CAUTION: To prevent possible damage to the ventilator, always secure it to its
stand or securely place it on a flat, stable surface that is free of dirt and debris. Do not use the ventilator adjacent to, or stack it with, other equipment.

Communications interface

WARNING: Connect to the ventilator only items that are specified as part of or
compatible with the ventilator system. Additional equipment connected to medical electrical equipment must comply with the respective IEC or ISO standards. Furthermore, all configurations shall comply with the requirements for medical electrical systems (see IEC 60601-1-1 or clause 16 of edition 3 of IEC 60601-1, respectively). Anybody connecting additional equipment to medical electrical equipment configures a medical system and is therefore responsible for ensuring that the system complies with the requirements for medical electrical systems. Also be aware that local laws may take priority over the above mentioned requirements. If in doubt, consult Philips.
WARNING: The USB port is not currently available for use. DO NOT connect or attempt
to power any equipment from the USB port.
WARNING: It is the responsibility of the end user to validate the compatibility and use
of information transmitted from the ventilator to the device to be connected to the ventilator.
WARNING: The data provided through the communications interface is for reference
only. Decisions for patient care should be based on the clinician’s observations of the patient.
1-8
Warnings, cautions, and notes
WARNING: To prevent possible patient injury due to nonannunciating alarms, verify
the operation of any remote alarm device before use.
WARNING: To ensure the functionality of the remote alarm, connect only Respironics-
approved cables to the remote alarm port.
CAUTION: The remote alarm port is intended to connect only to an SELV (safety
extra-low voltage and ungrounded system with basic insulation to ground), in accordance with IEC 60601-1. To prevent damage to the remote alarm, make sure the signal input does not exceed the maximum rating of 24 VAC or 36 VDC at 500 mA with a minimum current of 1 mA.

Diagnostic mode WARNING: To prevent possible patient injury, do not enter the diagnostic mode while

a patient is connected to the ventilator. Verify that the patient is disconnected before proceeding.
1-9
Warnings, cautions, and notes
(This page is intentionally blank.)
1-10

Chapter 2. Symbols

Refer to these tables to interpret symbols used on the ventilator labels and packaging and on the ventilator screen. To interpret symbols pertaining to accessories, refer to their instructions for use.
?
Table 2-1: Symbols used on ventilator labels and packaging
Symbol Description
Warning: Risk of explosion. Do not use in the presence of flammable an­esthetics.
Attention, consult the accompanying documents.
Read the user manual before using the ventilator.
(Blue) It is mandatory for the operator to consult the accompanying docu­ments.
Protective earth (ground)
Type B applied part, which is equipment that provides a particular degree of protection against electric shock, particularly in regard to allowable leakage current and of the protective earth connection
Requires alternating current (AC)
Degree of fluid ingress protection provided by the enclosure (drip-proof)
Caution: Federal law restricts this device to sale by or on the order of a
physician
Alarm and remote alarm
Two states of control: ON and Shutdown
Battery
European Conformity. Symbol is on rear panel of ventilator.
2-1
Symbols
EC REP
Table 2-1: Symbols used on ventilator labels and packaging (continued)
Symbol Description
Brazilian Conformity. Certification by INMETRO (National Institute of Me­trology, Standardization and Industrial Quality)/SGS (Societe Generale de Surveillance).
EurAsian Conformity mark - EAC
Date of manufacture
Manufacturer
EC representative
Serial number
Order number
Lot or batch number
Model number
Use by date
RS-232 serial input/output
USB port
Oxygen
(Yellow) Warning
Ethernet connection
2-2
Table 2-1: Symbols used on ventilator labels and packaging (continued)
(On power cord)
Symbol Description
Accept button on the navigation ring
Adjustment direction on the navigation ring
Canadian Standards Association approval
Do not disassemble. Refer to authorized service personnel.
Product must be disposed of in accordance with the WEEE directive.
Symbols
Noninvasive ventilation (patient with mask)
Invasive ventilation (intubated patient)
Do not block the cooling fan Inlet (at the rear of the ventilator).
No pushing. Do not push on the ventilator screen. Tipping hazard.
Total mass (weight) of the ventilator, ventilator stand, and standard setup. See page 11-5 for more information.
Hospital-grade
2-3
Symbols
廢電池請回收
Table 2-1: Symbols used on ventilator labels and packaging (continued)
Symbol Description
Recycle
Recycle (Taiwan)
RoHS (China). Administrative Measure on the Control of Pollution Caused by Electronic Information Products. Contains RoHS substances with 50 years environmentally friendly use period (EFUP).
uR UL recognition symbol
Direct current (DC). Symbol is on backup battery.
Rechargeable battery. Symbol is on backup battery.
Lithium-ion battery. Battery must be recycled or disposed of properly. Symbol is on backup battery.
Atmospheric pressure limitation. Indicates the acceptable upper and low­er limits of atmospheric pressure for transport and storage.
Humidity limitation. Indicates the acceptable upper and lower limits of relative humidity for transport and storage.
Temperature limit. Indicates the maximum and minimum temperature limits at which the item shall be stored or transported.
Battery option
C-Flex feature
AVAPS mode (included)
2-4
Table 2-1: Symbols used on ventilator labels and packaging (continued)
Symbol Description
PPV software option
Auto-Trak+ software option
High flow therapy
Note: 3.00 software and above. HFT is optional for model V60 and included with model V60 Plus.
Table 2-2: Symbols used on graphical user interface
Symbol Description
Alarm (audible)
Symbols
Alarm is silenced
High priority alarm
Low priority alarm
Alarm reset
Informational message
Alarm message is displayed. Touch to hide alarm messages.
Alarm message is hidden. Touch to display alarm messages.
Do not use an NIV mask during high flow therapy (3.00 software and above, and V60 Plus).
2-5
Symbols
Table 2-2: Symbols used on graphical user interface (continued)
Symbol Description
Increase and decrease (adjustment arrow) buttons. Adjusts a setting or selects a value.
Accept button. Accepts set values.
Cancel button. Cancels set values.
+2:00 minutes button. Adds two minutes to 100% O
Ventilator is powered by AC power and the optional battery is installed.
Ventilator is powered by AC power and the optional battery is not in­stalled.
Ventilator is powered by the battery. This symbol shows the approximate battery time remaining in hours and minutes, and it shows the capacity graphically.
Help button. Touch to display onscreen help information.
Vertical autoscale button. Autoscales the Y axis of the graphs to fit the data currently displayed.
delivery.
2
Pause button. Freezes waveforms in the Waveform window.
Pause in progress
Resume button. Resumes all waveform graphs from a paused state.
2-6
Table 2-2: Symbols used on graphical user interface (continued)
Symbol Description
Time base adjust button. Rescales the X axis of the graph display data at 3, 6, 12, and 24 second increments.
Symbols
V
E
V
T
T
I/TTOT
Estimated minute ventilation
Estimated exhaled tidal volume Duty cycle. Inspiratory time divided by total cycle time. No valid data to display
Data is under range
Data is over range
Pressure, centimeters of water
Flow, liters per minute. BTPS compensated.
Volume, milliliters
User-set Ramp Time. Ramp graphic fills in as Ramp Time progresses.
Ramp Time is OFF (no ramp time set).
Intentional leak. The number corresponds to the leak symbol printed on Philips Respironics masks.
2-7
Symbols
(This page is intentionally blank.)
2-8

Chapter 3. General information

This manual covers the Respironics V60 and V60 Plus Ventilator configurations. Both share the same platform. The V60 Plus Ventilator comes standard with High Flow Therapy (HFT). The V60 Ventilator can be field­upgraded with HFT, subject to local regulations. For a full list of features, modes, and options, see "General description" on page 3-2.
NOTE: The 3.00 software upgrade, which permits the activation of HFT, and
the V60 Plus Ventilator are not available in all countries
NOTE: .

Intended use The Respironics V60/V60 Plus Ventilator is an assist ventilator and is intended

to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional settings under the direction of a physician.
The ventilator is intended to support pediatric patients weighing 20 kg (44 lb) or greater to adult patients. It is also intended for intubated patients meeting the same selection criteria as the noninvasive applications. The ventilator is intended to be used by qualified medical professionals, such as physicians, nurses, and respiratory therapists. The ventilator is intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.

Contraindications The Respironics V60/V60 Plus Ventilator is contraindicated for patients with

any of the following conditions:
Lack of spontaneous respiratory drive
Inability to maintain a patent airway or adequately clear secretions
At risk for aspiration of gastric contents
Acute sinusitis or otitis media
Hypotension
Untreated pertussis
Epistaxis (nosebleed)
3-1
General information
Model number

About CO2 rebreathing

As with mask ventilation in general, patient CO2 rebreathing may occur under some circumstances. Follow these guidelines to minimize the potential for CO rebreathing. If rebreathing is a significant concern for a particular patient and these guidelines are not sufficient to acceptably reduce the potential for CO rebreathing, consider an alternative means of ventilation.
Increase EPAP to decrease the potential for CO pressures produce more flow through the exhalation port, which helps to purge all CO
Be aware that the potential for CO inspiratory time increases. A longer inspiratory time decreases exhalation time, allowing less CO the next cycle. In such circumstances, higher tidal volumes further increase the volume of CO
from the circuit to prevent rebreathing.
2
rebreathing increases as
2
to be purged from the circuit before
2
rebreathed by the patient.
2
rebreathing. Higher
2

Potential side effects Advise the patient to immediately report any unusual chest discomfort,

shortness of breath, or severe headache. Other potential side effects of noninvasive positive pressure ventilation include: ear discomfort, conjunctivitis, skin abrasions due to mask/patient interface, and gastric distention (aerophagia). If skin irritation or breakdown develops from the use of the mask, refer to the accompanying mask instructions for appropriate action.
2
2

General description The Respironics V60/V60 Plus Ventilator (Figure 3-1) is a microprocessor-

controlled, bilevel positive airway pressure (BiPAP) ventilatory assist system that provides noninvasive positive pressure ventilation (NPPV) and invasive ventilatory support for spontaneously breathing adult and pediatric patients.
Figure 3-1: Respironics V60 Ventilator shown
Ventilation modes. The ventilator offers a range of conventional pressure modes, CPAP (continuous positive airway pressure), PCV (pressure-controlled ventilation), and S/T (spontaneous/timed). The volume-targeted AVAPS (average volume-assured pressure support) mode combines the attributes of pressure-controlled and volume-targeted ventilation. The optional PPV mode provides pressure ventilation in proportion to the patient’s efforts.
3-2
General information
Modes, therapies and features. Table 3-1 shows which modes, therapies and
features are included or optional for the V60 and V60 Plus models.
Table 3-1: V60 and V60 Plus comparison
Ventilator Model
V60 Included Optional Optional Optional Included
V60 Plus Included Optional Included Optional Included
Modes Therapy Features
AVAPS PPV HFT Auto-Trak+ C-Flex
High flow therapy (HFT). High flow therapy provides a set flow of mixed air and
oxygen. Flow and O
percentage settings are selected by the clinician. HFT is
2
available for 3.00 software and above, as well as for the V60 Plus.
Auto-Trak Sensitivity allows the ventilator to automatically compensate for intentional and unintentional leaks by maintaining a stable baseline and adjusting trigger and cycle thresholds for optimum patient-to-ventilator synchrony. The optional Auto-Trak+ feature lets you further adjust the level of Auto-Trak Sensitivity.
User interface. The ventilator’s ergonomic design, including a 12.1-inch (31­cm) color touchscreen, a navigation ring, and key panel, lets you easily access ventilator settings and monitored parameters.
Monitoring. The ventilator displays monitored parameters as numbers and as real-time waveforms (curves or scalars).
Alarms. The ventilator’s operator-adjustable and nonadjustable alarms help ensure the patient’s safety.
Power and gas supplies. The ventilator uses as its primary power source AC mains. An optional internal backup battery powers the ventilator typically for 6 hours.
The ventilator uses high-pressure oxygen. An integral blower pressurizes gas for delivery to the patient.
NOTE: Oxygen delivered through the compressed gas hose and blower is
used as fresh gas.
Mounting. The ventilator can be mounted to a stand. When equipped with the optional cylinder holder, the stand can accommodate two E-size oxygen cylinders. An oxygen manifold kit is available, which allows two oxygen cylinders and one wall oxygen supply line to be used as inputs to the ventilator.
Communications interface. The ventilator can output data through the RS-232 serial port upon receiving a command from a host computer or bedside monitoring system. The ventilator is equipped with a remote alarm/nurse call connection to activate alarms remotely.
Upgradability via Respi-Link
remote diagnostic system. The Respi-Link interface
permits software upgrade and remote troubleshooting of the ventilator through the RS-232 port.
3-3
General information
Oxygen monitor
Humidifier
Oxygen
cylinder
Patient circuit
Mask
Bacteria filter

Physical description Patient circuits, masks/patient interfaces, and accessories

Figure 3-2 shows the Respironics V60/V60 Plus Ventilator with its patient circuit and accessories. Table 3-2 on page 3-5 lists recommended patient circuits, masks/patient interfaces, and other accessories for use with the ventilator. Appendix C provides ordering information for parts and accessories.
Figure 3-2: Respironics V60/V60 Plus Ventilator with accessories
3-4
General information
Table 3-2: Recommended parts and accessories
Part Use...
Patient circuit Single-limb patient circuit intended for
noninvasive ventilation, invasive ventilation, or high flow therapy (if applicable). Use a circuit listed in Appendix C.
WARNING: During ventilation, patient exhalate is
released into room air. Use of a patient circuit with a filter on its exhalation port is recommended.
Patient interface (noninvasive or invasive) • Respironics masks listed in Appendix C
• Invasive interface (tracheostomy or ET tube)
Patient interface (HFT) Nasal high flow cannula and tracheostomy
interface listed in Appendix C. (For use with 3.00 software and above, as well as V60 Plus.)
Exhalation port Philips Respironics exhalation port listed in
Appendix C. Contact your Philips representative.
Inspiratory filter Main flow (inspiratory) bacteria filter listed
in Appendix C
Humidifier • Fisher & Paykel
• Fisher & Paykel
Oxygen analyzer/monitor • Analytical Industries AII-2000M oxygen
analyzer/monitor*
*NOTE: Contact Philips representative for availability outside the U.S.
Nebulizer • Vibrating mesh nebulizer, such as the
Aerogen
• Jet nebulizer with low drive flow requirements (< 10 L/min)
MR850 MR290 chamber
3-5
General information
1
6
5
4
2
3
Key panel
789
Ventilator unit
Figure 3-3 through Figure 3-5 show the controls, indicators, and other important parts of the ventilator unit.
Figure 3-3: Front view
Number Description
1 Graphical user interface. Color LCD (liquid crystal display) with touchscreen.
2 Navigation ring. Lets you adjust values and navigate the graphical user interface
by rotating the finger on its touchpad.
3 Accept button. Activates selections.
4 Proximal pressure port. Connection for tubing that monitors patient pressure in
5 Ventilator outlet (To patient) port. Main connection for the patient circuit.
6 Alarm speakers (beneath ventilator)
7 Alarm LED. Flashes during a high-priority alarm. On continuously during a
8 Battery (charged) LED. Flashes when battery is charging. On continuously when
9 ON/Shutdown key with LED. Turns on AC power and initiates ventilator shutdown.
the patient circuit.
Delivers air and oxygen in prescribed pressures to the patient.
ventilator inoperative condition.
battery is charged. Off when ventilator is running on battery or when the ventilator is off and AC power is not connected.
LED is continuously on when AC power is connected.
3-6
Number Description
1 2
General information
Figure 3-4: Side view
1 Ventilation vents. Allow intake of air for delivery to the patient.
2 Air inlet filter (under side panel). Filters the air for delivery to the patient.
3-7
General information
83
1
7 3 6 5 4 3 2
9
Figure 3-5: Rear view
Number Description
1 Backup battery (compartment under side panel). Optional, 6-hour backup
battery.
2 Remote alarm/nurse call connector
3 Reserved for future use
4 Power cord retainer
5 Power cord
6 RS-232 serial and analog I/O connector (female DB-25). Connects to hospital
information systems and other serial devices, and functions as an interface for analog signals. Connects Respi-Link for software updates.
7 Cooling fan filter
8 High-pressure oxygen inlet connector
9 Option labels
remote diagnostic system gateway
3-8
General information

About the optional backup battery

WARNING: To reduce the risk of power failure to the ventilator, pay close attention to
the battery’s charge level. The battery’s operation time is approximate and is affected by ventilator settings, discharge and recharge cycles, battery age, and ambient temperature. Battery charge is reduced at low ambient temperatures or in situations where the alarm is continuously sounding.
NOTE: The backup batteries are intended for short-term use only. They are
not intended to be a primary power source.
NOTE: We recommend that the ventilator’s batteries be fully charged before
you ventilate a patient. If the batteries are not fully charged and AC power fails, always pay close attention to the level of battery charge.
NOTE: A new backup battery should be installed and charged within one year
of the date of manufacture identified on the battery and on the shipping box.
The optional internal backup battery protects the ventilator from low, or failure of, AC (mains) power. If AC power fails, the ventilator automatically switches to operation on backup battery with no interruption in ventilation. The battery powers the ventilator until AC power is again adequate or until the battery is depleted. The battery powers the ventilator typically for 6 hours.
As a safeguard, the ventilator provides a low battery alarm. It also has a capacitor-driven backup alarm that sounds for at least 2 minutes when battery power is completely lost.
The ventilator charges the battery whenever the ventilator is connected to AC, with or without the ventilator switched on. The Battery (charged) LED flashes to show that the battery is being charged.
Check the battery charge level before putting a patient on the ventilator and before unplugging the ventilator for transport or other purposes. The power source symbol at the bottom right-hand corner of the screen shows the power source in use and, if the ventilator is running on battery, the level of battery charge (Figure 3-6). If the battery is not fully charged, recharge it by connecting the ventilator to AC power for a minimum of 5 hours. Pressing the Help button shows you the time remaining until the battery is full. If the battery is not fully charged after this time, have the ventilator serviced.
3-9
General information
ON/Shutdown LED
On continuously: AC power is connected
Battery (charged) LED
Flashes: Battery is charging On continuously: Battery is charged
(90 to 100%) Off: Ventilator is running on battery,
or the ventilator is off and AC power is not connected
Power source symbol
The ventilator is powered by AC and the battery is installed.
The ventilator is powered by AC and the battery is not installed.
The ventilator is powered by the battery. This symbol shows the approximate battery time remaining in hours and minutes, and it shows the capacity graphically.
Figure 3-6: Power indicators
3-10
General information
15
Window/window tabs
(see page 6-1)
Help button (see page 6-21)
Waveforms window
(see page 8-1)
Patient data window
(see page 8-1)
Power symbols
(see page 5-2)
Alarm status bar
(see page 9-2)
Compressed waveforms window with Alarms/ Messages list (see page 9-2)
100% O
2
button
(see page 6-5)

About the graphical user interface

Through the graphical user interface (Figure 3-7) you make ventilator settings and view ventilator and patient data. During ventilation, the upper screen displays alarms and patient data. The middle screen displays real-time waveforms and alarm and informational messages. The lower screen lets you access modes and other ventilator settings, display help information, and see the power status.
Figure 3-7: Parts of graphical user interface
3-11
General information
Adjustment arrow button
Slider flag
Setting range scale
Proposed value
Navigating the graphical user interface
Select a function by touching the desired tab or button on the touchscreen. Use this as the primary method to control the ventilator.
You can use the navigation ring as an alternative to the following touchscreen functions:
Touchscreen equivalent Navigation ring equivalent
Touch increase button (adjustment arrow). Press and hold for faster adjustments.
Touch decrease button (adjustment arrow). Press and hold for faster adjustments.
* Available in Revision 2.30 software and above.
*
*
Touch Accept button (applies selection)
Touch and rotate finger clockwise to increase value or move cursor forward
Touch and rotate finger counterclockwise to decrease value or move cursor backward
Press Accept (checkmark) button (applies selection)
After making selections and adjusting values, accept selections by pressing the circular Accept button (the checkmark) in the middle of the navigation ring to accept and apply the change.
To open a window, touch the window tab.
To cancel a function and close the window, either select Cancel or touch another
window tab.
To adjust a parameter, select the value with the navigation ring or touch the arrow button. Each touch changes the value in single increments or, for parameters with wide ranges, press and hold the arrow key to make faster changes. The slider flag moves along the setting range scale. Select Accept to apply.
The navigation ring also lets you adjust the position of the cursor in the waveforms window while the screen is frozen. See Freezing and unfreezing waveforms on page 8-3 for more information.
3-12
General information

Starting up the ventilator

Shutting down the ventilator

NOTE: Upon power-on the ventilator automatically runs a test of the backup
audible alarm followed by the primary audible alarm. You should hear a high-pitched tone, followed by a beep. If you do not hear all of these sounds, discontinue use of the ventilator and have it serviced.
1. Power on the ventilator with the ON/Shutdown key.
2. Verify the ventilator operation, as described on page 5-7.
Shut down the ventilator as follows:
1. Press and release the ON/Shutdown key. The Shutdown window opens.
2. Select Ventilator Shutdown. The ventilator shuts down.
NOTE: Improper shutdown may cause a Power has been restored message the
next time the ventilator is turned on.
NOTE: If the screen is blank and the dialogue box cannot be displayed, shut
down the ventilator by pressing the ON/Shutdown key, then the Accept button on the navigation ring.

Training Product training is available. Contact your local Philips sales representative or

Philips Customer Support for assistance. Call 1-800-225-0230 for ordering and 1-800-722-9377 for service.
3-13
General information
(This page is intentionally blank.)
3-14

Chapter 4. Principles of operation

System operational overview

The Respironics V60/V60 Plus Ventilator is a microprocessor-controlled pneumatic system that delivers a mixture of air and oxygen. It is powered by AC with optional battery backup to protect against power failure or unstable power and to facilitate intrahospital transport. The ventilator’s pneumatics deliver gas and its electrical systems control pneumatics, monitor the patient, and distribute power.
The user provides inputs to the ventilator through a touchscreen, keys, and a navigation ring. These inputs become instructions for the pneumatics to deliver a precisely controlled gas mixture to the patient. Pressure and flow sensors provide feedback, which is used to adjust gas delivery to the patient. Monitored data based on sensor inputs is also displayed by the graphical user interface.
The ventilator’s gas delivery and monitoring functions are cross-checked. This cross-checking helps prevent simultaneous failure of these two main functions and minimizes the possible hazards of system failure.
A comprehensive system of visual and audible alarms helps ensure the patient’s safety. Clinical alarms can indicate an abnormal physiological condition. Technical alarms, triggered by the ventilator’s self-tests, can indicate a hardware or software failure. In the case of some technical alarms, limited ventilation is provided to give the user time for corrective actions. When a condition is critical enough to possibly compromise safe ventilation, the ventilator is placed into the ventilator inoperative state, in which oxygen flow and blower operation are disabled.
The ventilator has several means to ensure that safe patient or respiratory pressures are maintained. The maximum working pressure is ensured by the high inspiratory pressure (HIP) alarm limit. If the set high pressure limit is reached, the ventilator cycles into exhalation.
4-1
Principles of operation
Respironics V60/V60 Plus Ventilator
Blower
High-
pressure
oxygen
Ambient
air
Proximal (patient)
pressure
Ventilator outlet
(machine)
pressure
Air flow
sensor
Mixer
Propor-
tioning
valve
O2 flow
sensor
Air inlet
filter
Main flow
filter
Patient
Exhalation
port
O2 pressure

Pneumatic system operation

Figure 4-1: Respironics V60/V60 Plus Ventilator gas delivery system
The ventilator uses ambient air and high-pressure oxygen (Figure 4-1). Air enters through an inlet filter. Oxygen enters though a high-pressure inlet, and a proportioning valve provides the operator-set concentration. The system mixes the air and oxygen, pressurizes it in the blower, and then regulates it to the user-set pressure. To do this, the ventilator compares the proximal (patient) pressure measurement with the ventilator outlet (machine) pressure, and adjusts the machine pressure to compensate for the pressure drop across the inspiratory filter, patient circuit, and humidifier. This helps ensure accurate and responsive pressure delivery and leak compensation.
The ventilator delivers gas to the patient through a main flow (inspiratory) bacteria filter, a single-limb patient breathing circuit, a humidification device (optional) and a patient interface such as a mask or ET tube. A pressure tap proximal to the patient is used to monitor patient pressure. The exhalation port continually exhausts gas from the circuit during inspiration and exhalation to minimize rebreathing and ensure CO
removal.
2

Breath delivery characteristics

Control variable
Breaths delivered by the Respironics V60/V60 Plus Ventilator are pressure controlled. In the AVAPS mode, the ventilator’s applied pressure is automatically adjusted over a period of time to maintain a target tidal volume.
Triggering, cycling, and leak adaptation
Unlike other ventilators, the Respironics V60/V60 Plus Ventilator does not require you to set triggering and cycling sensitivity or to adjust baseline flow.The ventilator’s unique Auto-Trak Sensitivity algorithm adjusts these automatically; see “Auto-Trak Sensitivity” on page 4-3.
Baseline pressure
4-2
A positive baseline pressure (EPAP or CPAP) may be set for all breaths in all modes.
Principles of operation
Pressure rise time
The operator-set Rise Time defines the time required for inspiratory pressure to rise to the set (target) pressure.
Negative pressures
There are no negative pressures generated during exhalation.
Oxygen concentration
The Respironics V60/V60 Plus Ventilator incorporates an oxygen mixer. Oxygen concentration can be set in all modes.

Auto-Trak Sensitivity An important characteristic of the Respironics V60/V60 Plus Ventilator is its

ability to recognize and compensate for intentional and unintentional leaks in the system, and to automatically adjust its triggering and cycling algorithms to maintain optimum performance in the presence of leaks. This is called Auto­Trak Sensitivity. The following subsections describe this function in detail.
Triggering
Breaths are patient (flow) triggered in all modes, typically when patient effort causes a certain volume of gas to accumulate above method). An inspiration is also triggered when the patient inspiratory distorts the expiratory flow waveform sufficiently (shape signal method; see page 4-4).
baseline flow (volume
effort
Cycling
Cycling to exhalation occurs in these cases:
Patient expiratory sufficiently (shape signal method). See “Shape signal method of cycling and triggering.” on page 4-4.
Patient flow reaches the spontaneous exhalation threshold (SET). See “SET method of cycling.” on page 4-4.
After 3 seconds at the IPAP level (timed backup safety mechanism)
When a flow reversal occurs, typically due to a mask or mouth leak
effort distorts the inspiratory flow waveform
4-3
Principles of operation
Estimated
patient flow
Shape
signal
Cycle to exhalation crossover point
Trigger to inspiration crossover point
Spontaneous exhalation threshold
Shape signal method of cycling and triggering. The shape signal or “shadow trigger” method uses a mathematical model derived from the flow signal. A new flow signal (shape signal) is generated by offsetting the signal from the actual flow and delaying it (Figure 4-2). This intentional delay causes the flow shape signal to be slightly behind the patient’s flow signal. If there is a sudden change in patient flow, the patient’s flow signal crosses the shape signal; this results in a trigger or a cycle. As a result, a sudden decrease in expiratory flow from an inspiratory effort will cross the shape signal and create a signal for ventilator triggering.
Figure 4-2: Shape signal
SET method of cycling. Patient flow reaches the spontaneous exhalation threshold (SET); see Figure 4-3. The SET represents the intersection of the flow waveform and a line of a given slope. SET is updated each breath.
Figure 4-3: Spontaneous exhalation threshold (SET)
4-4
Principles of operation
Flow
TIME
Original baseline
Inspiration
Spontaneous trigger
Cycle to exhalation
End exhalation
Adjustment of baseline
New
baseline
Patient
flow
Additional leak
Leak adaptation
Noninvasive ventilation in particular may involve considerable leakage around the mask or through the mouth. Some leakage is known or intentional: it is a characteristic of the mask/patient interface design. So that it can accurately adjust its baseline flow, the ventilator has you enter the intentional leakage value specific to the mask/patient interface (“Selecting the mask and exhalation port” on page 6-12). Other leakage is unpredictable or unintentional, and it changes as the patient’s breathing pattern changes.
To maintain prescribed pressures in the presence of leakage, the ventilator adjusts its baseline flow. Because the unintentional part of the leakage may constantly change, the ventilator recalculates the baseline flow each breath at the end of exhalation. The ventilator uses two main mechanisms to update its baseline flow: expiratory flow adjustment and tidal volume adjustment.
Expiratory flow adjustment. Every breath, at end-exhalation, the ventilator updates its flow baseline. At end-exhalation patient flow is assumed to be zero, so any difference between actual patient flow and the original baseline flow indicates a change in leakage. Figure 4-4 shows how the ventilator adjusts the baseline.
Figure 4-4: Expiratory flow adjustment
4-5
Principles of operation
Additional leak introduced
New baseline
Volume adjustment
Flow
Volume
Tidal volume adjustment. Every breath, the ventilator compares the inspiratory and expiratory tidal volumes. Any difference is assumed to be due to an unintentional circuit leak. The ventilator adjusts the baseline to reduce this tidal volume difference for the next breath. Figure 4-5 shows how the ventilator adjusts the baseline.
Figure 4-5: Tidal volume adjustment

Auto-Trak+ (optional) The Auto-Trak+ option for the Respironics V60/V60 Plus Ventilator lets you

further adjust the level of Auto-Trak Sensitivity, a feature that recognizes and compensates for intentional and unintentional leaks. This algorithm has multiple breath trigger and cycle thresholds. When you adjust Auto-Trak+ settings, you adjust these multiple trigger and/or cycle thresholds simultaneously, retaining all the auto-adaptive features of Auto-Trak Sensitivity.
The Normal Auto-Trak settings work well for most patients. Pediatric patients, however, may benefit from more sensitive trigger settings, while some adult patients may benefit from more or less sensitive cycle settings.

High flow therapy High flow therapy (HFT) enables delivery of a humidified gas mixture at an

operator-set flow rate via a nasal cannula interface or tracheal adapter. The principle mechanism of action for high flow therapy is delivering a known FiO at a flow rate equal to or greater than the patient's peak flow, thus minimizing dilution of the gas.
HFT provides blended gas to the patient at a targeted flow. Both O
2
concentration and flow are set by the clinician. Heated humidification is recommended during high flow therapy.
2
4-6
Principles of operation
High flow therapy (HFT) controls flow instead of pressure and is accessed only while in Standby mode. Patient alarms are not available during high flow therapy. This therapy is not considered a breath delivery mode.
Available for 3.00 software and above, as well as the V60 Plus.

Ventilation modes The Respironics V60/V60 Plus Ventilator operates in the following ventilation

modes:
CPAP (continuous positive airway pressure) mode
S/T (spontaneous/timed) mode
PCV (pressure-controlled ventilation) mode
AVAPS (average volume-assured pressure support) mode
PPV (proportional pressure ventilation) mode (optional)
Table 4-1 summarizes the characteristics of these modes. Note that on the ventilator, the Timed breath indicator means the breath is ventilator triggered, while the Spont breath indicator means the breath is patient triggered.
Table 4-1: Characteristics of Respironics V60/V60 Plus ventilation modes
Timed breaths Spont breaths
Mode
CPAP N/A N/A N/A Auto-Trak Pressure Auto-Trak
PCV Time Pressure Time Auto-Trak Pressure Time
S/T Time Pressure Time Auto-Trak Pressure Auto-Trak
AVAPS Time Pressure Time Auto-Trak Pressure Auto-Trak
PPV Time Pressure Time Auto-Trak Pressure Auto-Trak
* A trigger variable starts inspiration. † A limit variable can reach and maintain a preset level before inspiration ends but it does not
end inspiration.
‡ A cycle variable is a measured parameter used to end inspiration.
Trigger
*
Limit
Cycle
Trigger Limit Cycle
4-7
Principles of operation
Volume Flow
Pressure
CPAP
I E
Time
CPAP mode
In the CPAP (continuous positive airway pressure) mode, the ventilator functions as a demand flow system, with the patient triggering all breaths and determining their timing, pressure, and size. You set no triggering or cycling sensitivities: the patient triggers and cycles based on the ventilator’s Auto-Trak Sensitivity algorithms. The control settings active in the CPAP mode are shown in Figure 4-6. Figure 4-7 shows CPAP mode waveforms.
The C-Flex feature setting enhances traditional CPAP by reducing the pressure at the beginning of exhalation – a time when patients may be uncomfortable with CPAP – and returning it to the set CPAP level before the end of exhalation.
Figure 4-6: CPAP controls
Figure 4-7: CPAP waveforms
4-8
Principles of operation
Pressure
EPAP
IPAP
I-Time
1/Rate
Machine-triggered (Timed) breath
Rise
Time
Patient-triggered (Spont) breath
PCV mode
The PCV (pressure-controlled ventilation) mode delivers pressure-controlled breaths, either triggered by the ventilator (Timed) or the patient (Spont). The control settings active in the PCV mode are shown in Figure 4-8. The IPAP setting defines the applied inspiratory pressure for all breaths. If the patient fails to trigger a breath through Auto-Trak within the interval determined by the rate setting, the ventilator triggers a mandatory breath. The I-Time setting is the cycle criterion for all breaths. Figure 4-9 shows a PCV mode pressure waveform.
Figure 4-8: PCV controls
Figure 4-9: PCV pressure waveform
4-9
Principles of operation
Pressure
EPAP
IPAP
I-Time
1/Rate
Patient-triggered (Spont)
spontaneous breath with
pressure support
Mandatory (Timed) breath
Time
Rise
S/T mode
The S/T (spontaneous/timed) mode guarantees breath delivery at the user-set rate. It delivers pressure-controlled, time-cycled mandatory and pressure­supported spontaneous breaths, all at the IPAP pressure level. If the patient fails to trigger a breath within the interval determined by the Rate setting, the ventilator triggers a mandatory breath with the set I-Time. You set no patient triggering or cycling sensitivities: the patient triggers and cycles based on the ventilator’s Auto-Trak Sensitivity algorithms. The control settings active in the S/T mode are shown in Figure 4-10. Figure 4-11 shows an S/T mode pressure waveform.
Figure 4-10: S/T controls
Figure 4-11: S/T pressure waveform
4-10
Principles of operation
AVAPS mode
NOTE: When you adjust AVAPS minimum and maximum pressures,
remember that IPAP is adjusted to meet the target value. If the calculated target pressure is outside of the minimum and maximum pressure range, the target volume will not be achieved.
Unlike most pressure modes, the AVAPS (average volume-assured pressure support) mode delivers a target tidal volume. It achieves the target volume by regulating the pressure applied following an initial pressure ramp-up. The AVAPS mode delivers time-cycled mandatory breaths and pressure-supported spontaneous breaths.
If the patient fails to trigger a breath within the interval determined by the Rate control, the ventilator triggers a mandatory breath with the set I-Time. Mandatory and spontaneous breaths are delivered at a pressure that is continually adjusted over a period of time to achieve the volume target, V Min P and Max P define the minimum and maximum pressures that can be applied. You set no patient triggering or cycling sensitivities: the patient triggers and cycles based on the ventilator’s Auto-Trak Sensitivity algorithms.
.
T
At start-up, AVAPS applies an inspiratory pressure equal to one of the following, whichever is greater:
EPAP + (target volume / 60 ml/cmH
EPAP + 8 cmH
O
2
O)
2
•Pmin
The control settings active in the AVAPS mode are shown in Figure 4-12. Figure 4-13 shows AVAPS mode waveforms.
Figure 4-12: AVAPS controls
4-11
Principles of operation
Volume Flow
Pressure
Max P
Min P
V
T
EPAP
1/Rate
I-Time
Mandatory (Timed)
breath
Patient-triggered (Spont) spontaneous breath
Rise
Time
Figure 4-13: AVAPS waveforms
4-12
Principles of operation
PPV mode (optional)
The PPV (proportional pressure ventilation) mode provides patient-triggered breaths that deliver pressure in proportion to patient effort. Additionally a user­settable backup rate activates machine-triggered, pressure-limited, and time­cycled breaths in the case of apnea. In the PPV mode, patient effort determines the pressure, flow, and tidal volume delivered by the ventilator. The ventilator responds to patient effort, allowing the patient to determine when to start and end a breath.Additionally flow and pressure change based on the patient’s efforts throughout inspiration.
The physics behind PPV. Two forces oppose ventilation, resistance and elastance.
Resistance is the impedance to air movement in the airways:
Pressure/Flow = Resistance
Airway resistance in healthy adults ranges from approximately 0.5 to
2.5 cmH
Elastance is the elastic opposition to ventilation or the tendency of the lungs to resist inflation (elastance is the reciprocal of compliance):
O/L/s.
2
Pressure/Volume = 1/Compliance = Elastance
The compliance of lungs and chest wall for a healthy adult is approximately
0.1 L/cmH
O, resulting in an elastance value of 10 cmH2O/L.
2
The inspiratory muscles, therefore, must generate force to overcome the resistance and elastance of the respiratory system. The proximal airway pressure is the net result of this contraction of these muscles: it is the force of the inspiratory muscle contraction minus both the pressure needed to generate air flow (overcome respiratory system resistance) and the pressure generated to inflate the lungs (overcome respiratory system elastance).
PPV is based on the equation of motion:
Pressure = Volume x Elastance + Flow x Resistance
where Pressure is the sum of patient effort (P
) and the ventilator-
muscle
generated pressure.
How PPV works. The delivery of a PPV breath is controlled by the maximum elastance (volume) assist (Max E), maximum resistance (flow) assist (Max R), and PPV % settings. The actual delivered assistance to overcome elastance is the product of PPV % and Max E. The actual delivered assistance to overcome resistance is the product of PPV % and Max R. In general, Max E should be set relative to the respiratory elastance and Max R should be set relative to the respiratory resistance, although you do not need to know the actual value of either to apply PPV. You adjust assist levels to optimize patient comfort. The resultant pressure support delivered in the PPV mode is the resistance assist times patient flow plus the elastance assist times the patient volume. The end
4-13
Principles of operation
result is that the level of pressure support is controlled by the inspiratory effort of the patient. Because the patient completely controls ventilatory output,
1
PPV may significantly improve patient-ventilator synchrony and ultimately, patient comfort.
The PPV backup rate ensures that the patient receives a minimum number of breaths per minute if the spontaneous breathing rate falls below the Rate setting. If the patient fails to trigger a breath within the interval determined by the Rate control, the ventilator triggers a Timed (backup) breath with the set I-Time, Rise, and IPAP settings.
The control settings active in the PPV mode are shown in Figure 4-14.
Figure 4-14: PPV controls
Figure 4-15 shows PPV mode waveforms. Note how volume and pressure increase as does the ventilatory demand of the patient. Max V (PPV maximum volume limit) and Max P (PPV maximum pressure limit) are used to prevent the delivery of excessive pressure or volume. More information about these limits is provided in “About Max V and Max P alarms and alarm limits” on page 6-7.
1 Marantz, S., Patrick, W., Webster, K., et al. “Response of ventilator-dependent
patients to different levels of proportional assist.” Journal of Applied Physiology, Vol. 80: 397-403, 1996.
4-14
Principles of operation
VolumeFlowPressure
Max V
Increase in patient demand
Time
Max P
EPAP
Figure 4-15: PPV waveforms
4-15
Principles of operation
40
45
50
55
60
65
70
75
80
85
90
95
100
130 140 150 160 170 180 190 200 210 220 230 240
Total Ventilator Flow L/min
O2% Concentration
Assumptions: At an O2 setting of 100% and an oxygen supply with a 50 psig inlet pressure capable of delivering up to160 L/min.
Total ventilator flow (L/min)
Oxygen concentration (%)

Oxygen mixing The ventilator’s oxygen mixer regulates and proportions oxygen into the air

from the blower according to the O ±5% of the set value up to the maximum oxygen flow available. The ventilator can deliver up to 240 L/min of air/oxygen mix to assist in managing uncontrolled leaks during noninvasive ventilation.
Many hospital oxygen supply systems, however, cannot meet such high flow demands. Under extraordinary conditions (high O or high patient demand) where demand exceeds available oxygen system flow, the ventilator provides additional air flow from the blower to ensure the target pressure is met. Under such conditions, the accuracy of delivered oxygen may be affected. Figure 4-16 shows the effect on the delivered oxygen concentration as the maximum oxygen system flow is exceeded. This graph assumes a continuous flow demand. Normally the higher “peak” flow is only needed during inspiration, so this is a worst case scenario.
setting. The delivered oxygen accuracy is
2
setting plus high leak, and/
2
4-16
Figure 4-16: O
2
concentration as a function of total ventilator flow

Chapter 5. Setting up the ventilator for use

Set up the ventilator for each patient use as described in this chapter. For first­time installation, refer to Appendix A. For use with high flow therapy (HFT), set up the ventilator as described in this chapter, then refer to Chapter 7, High flow therapy.

Connecting oxygen WARNING: Connect the ventilator only to an appropriate medical-grade oxygen

source.
WARNING: To ensure accuracy of oxygen administration and to monitor for the
presence of contamination (incorrect gas connected), use an external oxygen monitor to verify the oxygen concentration in the delivered gas.
WARNING: To reduce the risk of fire, do not use a high-pressure oxygen hose that is
worn or contaminated with combustible materials like grease or oil.
WARNING: To reduce the risk of hypoxia, connect only oxygen to the high-pressure
connector at the rear of the ventilator.
WARNING: To reduce patient risk of oxygen toxicity, keep free-flowing oxygen away
from air inlet of ventilator.
CAUTION: To prevent possible damage to the ventilator, ensure that the
connection to the oxygen supply is clean and unlubricated, and that there is no water in the oxygen supply gas.

Installing an oxygen analyzer/monitor

Connect the oxygen hose to an appropriate high-pressure oxygen source.
Use of SIS connectors and supplemental oxygen accessories such as the O manifold requires higher oxygen supply pressures. Consult Table 11-9 on page 11-6 for appropriate oxygen pressure ranges.
Install an Analytical 2000M oxygen analyzer/monitor, or the equivalent, and follow the manufacturer’s instructions for setup and calibration.
5-1
2
Setting up the ventilator for use

Connecting to AC power

WARNING: To reduce the risk of electric shock, connect the ventilator to an AC
supply mains with protective earth only.
WARNING: Do not use extension cords, adapters, or power cords with the ventilator
that are not approved by Respironics.
WARNING: To prevent unintentional disconnection of the power cord, always use the
correct, Philips-supplied power cord and lock it into place with the power cord retainer before you switch the ventilator on. The retainer is designed to hold the connector end of the Philips-supplied cord securely in place.
WARNING: To reduce the risk of electric shock, regularly inspect the AC power cord
and verify that it is not frayed or cracked.
WARNING: To reduce the risk of strangulation, route the power cord to avoid
entanglement.
CAUTION: For 120 V equipment, grounding reliability can only be achieved
when it is connected to an equivalent receptacle marked “hospital only” or “hospital grade.”
Plug the power cord into a grounded outlet that supplies AC power between 100 and 240 V, 50/60 Hz.
Always check the reliability of the AC outlet. If you are using a 120 V outlet, make sure that it is hospital grade.
5-2
Setting up the ventilator for use

Installing the patient circuit

WARNING: To reduce the risk of strangulation from patient tubing, use a tubing
support arm and secure the proximal pressure line with clips.
WARNING: To prevent possible patient injury and possible water damage to the
ventilator, make sure the humidifier is set appropriately.
WARNING: To prevent possible patient injury and equipment damage, do not turn the
humidifier on until the gas flow has started and is regulated. Starting the heater or leaving it on without gas flow for prolonged periods may result in heat build-up, causing a bolus of hot air to be delivered to the patient. Circuit tubing may melt under these conditions. Turn the heater power switch off before stopping gas flow.
WARNING: To reduce the risk that the patient will aspirate condensed water from the
breathing circuit, position any humidifier lower than both the ventilator and the patient.
WARNING: To reduce the risk of fire, use only patient circuits intended for use in
oxygen-enriched environments. Do not use antistatic or electrically conductive tubing.
WARNING: To prevent patient or ventilator contamination, always use a main flow
bacteria filter on the patient gas outlet port. Filters not approved by Respironics may degrade system performance.
WARNING: During ventilation, patient exhalate is released into room air. Use of a
patient circuit with a filter on its exhalation port is recommended.
WARNING: To reduce the risk of bacterial contamination or damage, handle bacteria
filters with care.
WARNING: Any additional accessories in the patient circuit may substantially
increase flow resistance and impair ventilation.
WARNING: Avoid adding resistive circuit components on the patient side of the
proximal pressure line. Such components may defeat the disconnect alarm.
NOTE: Bacteria filter must be installed onto gas outlet. NOTE: Under extreme conditions and a missing, ruptured, or defective
bacteria filter, the entire gas pathway can become contaminated with bodily fluids or exhaled gas.
Install the patient circuit as shown in this section. For a list of compatible parts and accessories offered by Philips, see “Parts and accessories” on page C-1.
Assemble the patient circuit, including the main flow (inspiratory) bacteria filter, proximal pressure line, oxygen sensor tee, and if desired, humidifier and nebulizer. Figure 5-2 and Figure 5-3 show circuit configurations for noninvasive and invasive ventilation. Follow the manufacturers’ instructions for use for the individual parts.
NOTE: If you are using a jet nebulizer, use the lowest possible flow rate
recommended by the manufacturer. The flow rate must not exceed 10 L/min.
5-3
Setting up the ventilator for use
Proximal pressure line
Ventilator outlet
Bacteria filter
Proximal pressure port
Humidifier
Oxygen sensor tee
NOTE: This circuit setup is recommended for noninvasive ventilation. It is
also recommended for high flow therapy when using the AC611 FEP Connect to block the exhalation port.
Figure 5-1: Noninvasive patient circuit, with heated-wire and humidification
5-4
Setting up the ventilator for use
Proximal pressure line
Ventilator outlet
Bacteria filter
Proximal pressure port
Oxygen sensor tee
NOTE: This circuit setup is recommended for noninvasive ventilation.
Figure 5-2: Noninvasive patient circuit, without humidification
5-5
Setting up the ventilator for use
Ventilator outlet
Proximal pressure line
Water trap
Humidifier
Proximal filter
Bacteria filter
Proximal pressure port
Oxygen sensor tee
NOTE: This circuit setup is recommended for both noninvasive and invasive
ventilation.

Connecting external devices

Figure 5-3: Invasive patient circuit, with humidification
Connect the ventilator to a remote alarm (nurse call) device and a patient monitor or other external device, if applicable.
The Respironics V60/V60 Plus Ventilator can communicate with a Philips patient monitor using the IntelliBridge Open Interface. See “Using Philips monitors and the IntelliBridge Open Interface” on page B-15. The ventilator also supports the VueLink Open Interface. VueLink has been replaced by IntelliBridge, but information is included in this manual for backwards compatibility. See “Using Philips monitors and the VueLink Open Interface*” on page B-15.
For more information about connecting with non-Philips systems, contact your Philips representative.
5-6
Setting up the ventilator for use

Before placing a patient on the ventilator

WARNING: Always verify ventilator operation before placing the patient on a
ventilator. If the ventilator fails any verification steps, remove it from clinical use immediately. Do not use the ventilator until necessary repairs are completed and the ventilator passes verification.
WARNING: To reduce the risk of power failure to the ventilator, pay close attention to
the battery’s charge level. The battery’s operation time is approximate and is affected by ventilator settings, discharge and recharge cycles, battery age, and ambient temperature. Battery charge is reduced at low ambient temperatures or in situations where the alarm is continuously sounding.
NOTE: If the ventilator has a backup battery, the battery must be adequately
charged to verify operation. Recharge as necessary before verifying operation.
NOTE: The backup batteries are intended for short-term use only. They are
not intended to be a primary power source.
NOTE: We recommend that the ventilator’s batteries be fully charged before
you ventilate a patient. If the batteries are not fully charged and AC power fails, always pay close attention to the level of battery charge.
Verify ventilator operation
1. Power on the ventilator. The ventilator automatically runs a test of the
backup audible alarm followed by the primary audible alarm. Verify that you hear a high-pitched tone, followed by a beep.
2. Create a patient alarm, such as a disconnect alarm.
a. VERIFY that the proper alarm is annunciated (audio, visual,
flashing, and alarm LED).
b. VERIFY that the volume setting is adequate for the environment in
which it will be used.
c. VERIFY remote alarm setup, if applicable.
3. Resolve the alarm condition and manually reset the alarm.
4. If the backup battery is installed, disconnect the ventilator from AC power while the ventilator is running. If the backup battery is not installed, go to step 5.
a. VERIFY that the ventilator switches over to battery power (battery
symbol in right-hand corner of screen is displayed).
b. VERIFY that the audible alarm sounds intermittently.
5. Reconnect the ventilator to AC power.
5-7
Setting up the ventilator for use
V

Running alarm tests The ventilator performs a self-check during start-up and continuously during

operation. Alarm functionality is verified by this self-check. You may also want to run alarm tests, which demonstrate the alarms’ operation.
WARNING: To prevent possible patient injury, always return alarm settings to
hospital-standard values after verifying ventilator operation.
Preparation
1. Set the ventilator up as for normal ventilation, complete with breathing circuit (PN 582073 or the equivalent) and a 1-liter test lung assembly (PN 1021671).
2. Set the mode to S/T and make the following control settings: Rate: 4 BPM, IPAP: 10 cmH Ramp: Off, O
: 21%.
2
3. Make the following alarm settings: Hi Rate: 90 BPM, Lo Rate: 1 BPM, Hi V
: 2000 mL, Lo VT: OFF, HIP: 50 cmH2O, LIP: OFF, Lo E: OFF,
T
LIP T: 5 secs.
O, EPAP: 6 cmH2O, I-Time: 1 sec, Rise: 1,
2
High Inspiratory Pressure
1. Lower the HIP alarm limit to 8 cmH2O.
2. VERIFY that the High Inspiratory Pressure alarm is activated, the ventilator cycles into exhalation, and pressure falls to 6 cmH EPAP level).
3. Raise the HIP alarm limit to 15 cmH
O.
2
Low Tidal Volume
1. Raise the Lo VT alarm setting above the displayed, measured VT.
2. VERIFY that the Low Tidal Volume alarm is activated.
3. Turn the Lo V
4. VERIFY that the alarm resets.
alarm setting OFF.
T
O (the
2
5-8
Setting up the ventilator for use
Patient Disconnect
1. Disconnect the test lung.
2. VERIFY that the Patient Disconnect alarm is activated.
3. Reconnect the test lung.
4. VERIFY that the alarm resets and that the ventilator automatically
resumes ventilation.
Patient Circuit Occluded
1. Disconnect the patient circuit (including bacteria filter) from the
ventilator outlet, and block the ventilator outlet.
2. VERIFY that the Patient Circuit Occluded alarm is activated.
3. Unblock the outlet, and reconnect the circuit.
4. VERIFY that the alarm resets.
5-9
Setting up the ventilator for use

Using the ventilator for intra-hospital transport

WARNING: Always check the status of the oxygen cylinders before using the
ventilator during transport.
WARNING: To reduce the risk of power failure to the ventilator, pay close attention to
the battery’s charge level. The battery’s operation time is approximate and is affected by ventilator settings, discharge and recharge cycles, battery age, and ambient temperature. Battery charge is reduced at low ambient temperatures or in situations where the alarm is continuously sounding.
WARNING: The V60/V60 Plus Ventilator requires a pressurized oxygen supply that
provides a minimum flow of 175 SLPM. Do not use any devices such as valves, hoses, Grab n' Go regulators or other brands of combined cylinder/ regulators that limit supply of oxygen flow below 175 SLPM.
WARNING: Do not leave the ventilator unattended when stationed on an incline.
Do the following to conserve oxygen during transport with the ventilator.
Make sure all cylinders are full (13,790 kPa/2000 psig or more).
Make sure the cylinder regulators are turned off while the ventilator is connected to wall oxygen.
Never turn the cylinder regulator on until you are ready to begin transport.
Only turn one cylinder regulator on at a time. If you turn on both cylinders, they may become depleted simultaneously, leaving you with no backup oxygen.
Whenever possible, reduce the O
Minimize all inadvertent leaks. Tighten masks prior to transport, and loosen up when patient is back on wall oxygen.
setting before transport.
2
Avoid using masks that have an exhalation port built into the mask when there is already an exhalation port in the circuit.
Be aware that oxygen is more rapidly depleted at higher leak rates (see Figure 5-4).
5-10
Setting up the ventilator for use
Leak of 50 L/min
a. VT = 500 mL, Rate = 40 BPM, EPAP = 6 cmH2O, IPAP = 18 cmH2O
Leak of 10 L/min Leak of 20 L/min Leak of 30 L/min Leak of 40 L/min
FIO
2
(%)
0
10
20
30
40
50
60
70
80
90
Duration (min)
30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
b. VT = 500 mL, Rate = 20 BPM, EPAP = 6 cmH2O, IPAP = 18 cmH2O
20
40
60
80
100
120
140
0
Duration (min)
Leak of 10 L/min
Leak of 20 L/min
Leak of 30 L/min Leak of 40 L/min Leak of 50 L/min
30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
FIO2 (%)

Storing the ventilator between patient use

Figure 5-4: Duration of cylinder oxygen (13,790 kPa/2000 psig) at various
leak rates
See “Storage between patient use” on page 10-7 for information about storing the ventilator.
5-11
Setting up the ventilator for use
(This page is intentionally blank.)
5-12

Chapter 6. Operation

page 6-3 page 6-12 page 6-2 page 6-17 page 6-19 page 6-21
WARNING: To ensure the ventilator’s safe operation, always verify ventilator
operation as described in “Verify ventilator operation” on page 5-7 before using the ventilator on a patient. If the ventilator fails any tests, remove it from clinical use immediately. Do not use the ventilator until necessary repairs are completed and all tests have passed.
NOTE: Before operation, prepare the ventilator as instructed in Chapter 5.
After power-on, the ventilator starts up in the mode and with the settings that were active before last power down. Check these settings and adjust as required. You must be familiar with using the touchscreen and navigation ring to select, adjust, activate, and confirm parameters. For details, see “Before placing a patient on the ventilator” on page 5-7.
Access the ventilator setting windows from the tabs at the bottom of the screen.
6-1
Operation
Active mode

Changing the mode The active ventilation mode is displayed in the bottom, left-hand corner of the

screen. Change the mode as follows. For details on modes, see “Ventilation modes” on page 4-7.
1. Open the Modes window.
2. Select the desired mode.
3. Adjust settings as desired (see “Changing individual ventilator settings” on page 6-4). Newly adjusted setting values are shown in yellow.
4. Select Activate Mode to apply.
6-2
Operation
Active mode

Changing control settings

Table 6-3 on page 6-22 is an alphabetical list of the control settings with their ranges. Table 11-2 on page 11-2 shows the control settings applicable to the different modes. For more information on control settings as they apply in the different ventilation modes, see “Ventilation modes” on page 4-7.
Making batch setting changes
NOTE: During a batch setting change, you cannot change the Ramp Time
setting when a ramp is active.
This process applies to ventilation settings only, not to alarm settings.
1. Open the Modes window.
2. Select the active mode.
3. Adjust settings as desired (see “Changing individual ventilator
settings” on page 6-4). Newly adjusted setting values are shown in yellow.
4. Select Activate Batch Change to apply.
6-3
Operation
Adjustment arrow button
Current
settings area
Setting range scale
Additional
information
Proposed value
Changing individual ventilator settings
You can make ventilator settings from the Settings window.
1. Open the Settings window.
2. Select the desired setting. As an example we will show the IPAP adjustment.
3. The setting window opens. Adjust the setting. Select Accept to apply.
6-4
Operation
Ramp status Bar

Using the Ramp Time function

The Ramp Time function helps your patient adapt to ventilation by gradually increasing inspiratory and expiratory pressure (IPAP and EPAP/CPAP) from subtherapeutic to user-set pressures over a user-set interval. Table 6-3 on page 6-22 describes this function’s principles of operation.
Follow these instructions to use the Ramp Time function:
1. Select the Ramp Time button in the Settings window.
The ramp starts. As the ramp progresses, the Ramp Time button graphic fills in.
2. To change the ramp interval or to end the ramp, select the Ramp Time
button again. The Ramp in Progress window opens.

Using the 100% O2 function

3. To end the ramp and apply the full IPAP and EPAP/CPAP immediately,
select End Ramp.
4. To end the ramp and start a new one, select Start New Ramp. The Ramp
Time setting window opens again so that you can set up a new ramp.
NOTE: The 100% O2 feature is available in Revision 2.30 software and
above.
The 100% O2 function delivers 100% oxygen to the patient. It is available during Screen Lock status.
Follow these instructions to use the 100% O
1. Select the 100% O2 button in the main GUI window.
6-5
function:
2
Operation
2. The ventilator delivers 100% oxygen for two minutes. A countdown timer displays.
While 100% oxygen delivery is active, you can press the +2:00 button to add two minutes more. Press Cancel to stop.
6-6
Operation

Using PPV

Follow these instructions to set up the ventilator in the PPV mode, referring to Figure 6-3. For principles of operation, see “PPV mode (optional)” on page 4-13.
1. Open the PPV Settings window.
2. Set EPAP, O
, alarm limits, and backup settings to appropriate values.
2
The HIP alarm limit should be greater than Max P. See “Principles of operation” on page 4-1 for a detailed explanation of these settings.
3. Set the Max V and Max P limits.
4. Set alarm limits to appropriate values. The HIP alarm limit should be
greater than the Max P.
About Max V and Max P alarms and alarm limits
Max V (PPV maximum volume limit) and Max P (PPV maximum pressure limit)
are used to prevent the delivery of excessive pressure or volume.
WARNING: PPV limits are not intended to be the primary ventilator alarms and should
not be substituted for the alarms found in the Alarm Settings window.
WARNING: To prevent the delivery of excessive pressure or volume, set the PPV
limits appropriately. Delivery of excessive pressure or volume can occur from a sudden increase in mask leak, inappropriate settings, or a plugged or kinked proximal pressure line. Conversely, insufficient treatment may result if limits are set too low.
When the Max V (PPV maximum volume limit) is reached, the breath is terminated and a message is displayed. After the limit is reached in three consecutive breaths, the audible alarm sounds. A PPV waveform with Max V is shown in Figure 6-1.
6-7
Operation
VolumeFlow
Max V
VT ≥ Max V
Inspiration terminated
Sudden increase
in flow
Time
Pressure
EPAP
Max P
Sudden increase
in target pressure
Pressure limited Exhalation cycled by patient
Figure 6-1: PPV waveform – Max V limit
When the Max P (PPV maximum pressure limit) is reached, pressure is limited but the breath is not terminated, and a message is displayed. After the limit is reached in three consecutive breaths, the audible alarm sounds. A PPV waveform with Max P is shown in Figure 6-2.
Figure 6-2: PPV waveform – Max P limit
Frequent annunciation of one or both alarms typically indicates improved patient status. It may, however, indicate that the patient is more actively breathing, possibly due to agitation or a change in the patient’s level of sedation. It may also indicate an increase in leakage.
The V
(estimated exhaled tidal volume) measurement may remain below the
T
set Max V limit even though the inspired volume exceeds Max V. This results from variable leakage, which reduces the exhaled volume in relation to the inspired volume.
6-8
Operation
Guidelines for using PPV
NOTE: The guidelines below are based on recommendations by
clinicians. They do not replace the clinical judgment of a physician and should not, on their own, be used for clinical decision making.
Determining Max R and Max E settings
It is recommended you set Max R (flow assist) and Max E (volume assist) to initial values and then titrate them based on the patient’s disease process:
Obstructive disease (COPD, asthma): Focus on Max R. Overcoming
increased resistance is typically the emphasis, not volume delivery.
Restrictive disease (neuromuscular, chest-wall deformities, obesity hypoventilation): Focus on Max E. Maintaining sufficient volume is typically the emphasis, not overcoming increased resistance.
Mixed disease processes affecting both resistance and elastance: Titrate both Max R and Max E settings.
6-9
Operation
Suggested titration procedure Follow this procedure to titrate settings to
optimize patient comfort while avoiding overassisting. See also the flow chart in Figure 6-3.
NOTE: You may also need to adjust
you do for the other PPV settings described below. Mask leakage, especially a sudden increase, is interpreted as patient effort by the ventilator and assisted accordingly; this may necessitate lowering the
PPV %
setting. However, the best solution is to maintain a minimal leak.
PPV %
according to patient response, as
1. Set EPAP, O2, alarm limits, and backup settings to appropriate values. The HIP alarm limit should be greater than Max P.
Suggested starting settings: EPAP 4 cmH
O
2
Current setting or per prescription
Max P 25 cmH
*
O
2
O
2
Max V 1000 to 1500 mL PPV % 80 to 100% Max E 5 cmH Max R 2 cmH All other backup settings
Per usual protocol
O/L
2
O/L/s
2
and alarms
* Consider higher EPAP settings for COPD patients to treat autoPEEP as evidenced by
missed triggers
2. Adjust Max E: a. Evaluate the patient. Check whether any of these conditions is
true:
The patient says they are getting too much air, pressure, or volume
The patient is using accessory muscles to actively stop inspiration
•The Max V or Max P limit is reached
The mask leak has suddenly increased
b. If none is true, increase Max E in increments of 2 cmH
O/L while
2
continuing to evaluate the patient’s response.
c. If any is true, decrease Max E by 2 cmH
O/L, and re-evaluate.
2
Repeat to optimize patient comfort.
3. Repeat the process above adjusting Max R, increasing and decreasing
in increments of 1 cmH
O/L/s to optimize patient comfort.
2
4. Repeat adjustment for Max E as needed.
5. Adjust PPV % downward as tolerated.
6-10
Start
Make initial settings:
EPAP: 4 cmH
2
O
O
2
: Current setting or per prescription
Max P: 25 cmH
2
O
Max V: 1000 to 1500 mL PPV %: 80 to 100% Max E: 5 cmH
2
O/L
Max R: 2 cmH
2
O/L/s
HIP: > Max P Other settings: Per usual protocol
Yes
No
Is the patient using
accessory muscles to ac-
tively stop inspiration
?
Was the Max V or Max P
limit reached
?
Has the mask leak suddenly
increased?
No
No
No
Increase Max E by
2 cmH
2
O/L
Yes
Yes
Yes
No
Does the patient say the air
is coming too fast
?
Is the patient using
accessory muscles to ac-
tively stop inspiration
?
Was the Max V or Max P
limit reached
?
Has the mask leak suddenly
increased?
No
No
Increase Max R by
1 cmH
2
O/L/s
Reduce Max R by 1 cmH2O/L/s for
patient comfort
Yes
Yes
Yes
Yes
No
End
Titrate PPV % downward as tolerated
Does the patient say they
are getting too much air,
pressure, or volume
?
Reduce Max E
by 2 cmH
2
O/L for patient comfort
Max E
titration
Repeat Max E titration as needed
Max R
titration
Operation
Figure 6-3: PPV initial setup
6-11
Operation

Changing alarm settings

WARNING: To prevent possible patient injury, avoid setting alarm limits to extreme
values, which can render the alarm system useless.
Some ventilator alarm settings are operator adjustable. You can adjust these at any time. Table 6-4 on page 6-25 lists the alarm settings and their ranges.
Review and adjust the alarm settings as follows:
1. Open the Alarm Settings window.
2. Select the desired setting, adjust it, and select Accept to apply.
The ventilator annunciates an alarm when a monitored value goes out of the range bounded by the alarm limits.

Selecting the mask and exhalation port

To be able to display full leakage data plus accurate tidal and minute volumes, the ventilator must know the intentional leak characteristics of the specific mask/patient interface and exhalation port.
After power-on, the Messages list displays the current mask and port settings for 5 minutes.
Change these settings as follows:
1. Open the Menu window.
2. Select Mask/Port.
6-12
3. Select the desired mask/patient interface type (Table 6-1). Select
Accept to apply.
For information concerning mask/port leak characteristics, see the instructions provided with each mask/port. See Appendix C for a list of masks, circuits, and related components used with the ventilator.
Operation
6-13
Operation
Table 6-1: Mask/patient interface selections
Mask/patient interface type
ET/Trach ET or tracheostomy tube
Leak 1 Mask with minimal intentional leak characteristics. Enter
Leak 2 Mask with medium intentional leak characteristics. Enter
Leak 3 AP111
Leak 4 Reserved for future use
Other Mask not manufactured by Philips Respironics
*
Description
Leak 1 for any of these Philips Respironics masks:
• Contour Deluxe nasal mask
• PerformaTrak mask
• AF421, AF531, AF541, AF811 (EE)
Leak 2 for this mask:
• Philips Respironics PerforMax oro-nasal mask [EE]
• AF421, AF531, AF541 (EE)
NOTE: If you select Other, the ventilator displays
Tot.Leak rather than Pt. Leak.
* A leak symbol is printed on Respironics masks.
4. Select the desired exhalation port type (Table 6-2). Select Accept to
apply.
If you select an exhalation port that is not compatible with the selected mask, Not allowed with current mask is displayed.
NOTE: In ventilation modes, ET/tracheostomy tubes and most Philips
Respironics masks require the use of an exhalation port. If you selected ET/Trach or Leak 1 as a mask/patient interface, you may not select None as an exhalation port.
6-14
Port type
Operation
Table 6-2: Exhalation port selections
Exhalation port test recommended?
DEP
Philips Respironics Disposable Exhalation Port
Whisper Swivel
Philips Respironics Whisper Swivel
PEV
Philips Respironics Plateau Exhalation Valve
Other
Exhalation port not supplied by Philips Respironics.
None
No inline circuit exhalation port
NOTE: If you select None, refer to the manufacturer’s
instructions to make sure the mask selected contains an exhalation port.
No
No
Yes
Yes
No
5. Run the exhalation port test if indicated in the table (see “Running the exhalation port test” on page 6-16 for instructions).
CAUTION: If you selected PEV or Other as an exhalation port, you must run an
exhalation port test.
NOTE: If the exhalation port test is not run or if it fails, the intentional leak is
unknown. Tot.Leak rather than Pt. Leak is displayed in the patient data window.
6-15
Operation
Progress
bar

Running the exhalation port test

The exhalation port test is required and its window is automatically displayed when PEV or Other is selected.
Procedure
Run the test as follows:
1. Disconnect the patient circuit from the mask/patient interface.
2. Occlude the circuit outlet. Select Start Test.
3. Wait while the test runs.
4. Verify that Test Passed is displayed.
6-16
Operation
5. Reconnect the patient circuit to the mask/interface.
6. Select Start Ventilation to initiate ventilation.
Troubleshooting
If Test Failed is displayed, check for leaks in the patient circuit, and install an exhalation device with lower leak characteristics. Repeat test. If the exhalation port test fails again, the intentional leak is unknown and Tot.Leak rather than
Pt. Leak is displayed in the patient data window.

Other functions: the Menu window

From the Menu window you can adjust user preferences.
Brightness
Use Brightness to adjust the screen for optimum daytime or nighttime viewing.
Loudness
Use Loudness to adjust the volume of the alarm and touchscreen audible feedback. You will hear audible feedback as you go through the selections.
The Alarm Volume Escalation status is also displayed on this screen. See “Alarm Volume Escalation” on page E-11 for more information.
Mask/Port
See “Selecting the mask and exhalation port” on page 6-12.
6-17
Operation
Vent Info (ventilator information)
The Ventilator Information window displays software version and other information specific to your ventilator.
Screen Lock
Screen Lock deactivates all buttons and tabs on the touchscreen except Alarm Silence, Alarm Reset, the Alarm/Message button, and Help. Tabs are grayed out
as in this example.
This message bar is displayed at the top of the screen:
To unlock the screen, press the Accept button in the center of the navigation ring.
NOTE: If Screen Lock is active, the touchscreen remains locked even if an
alarm becomes active.
Auto-Trak+
The Normal Auto-Trak settings work well for most patients. Pediatric patients, however, may benefit from more sensitive trigger settings, while some adult patients may benefit from more or less sensitive cycle settings.
Changing Auto-Trak+ settings
1. Select Auto-Trak+ from the Menu window.
6-18
Operation
Proposed value
Auto-Trak+ active
2. Select the desired adjustment. As an example, the E-Cycle adjustment is shown below.
3. The setting window opens. Adjust the setting, referring to the pressure-time graphic which represents the effect on I-Time. Select Accept to apply.
When Auto-Trak+ is active (when either Trigger or E-Cycle is set to a value other than Normal), the ventilator setting window displays Auto-Trak+.
Additionally, after power-on the Messages list displays the Auto-Trak+ settings for 5 minutes.

Standby Standby lets you safely suspend ventilation to temporarily disconnect the

patient from the ventilator or to set up the ventilator before connecting the patient. Alarms are disabled during standby.
You can also change ventilator settings and most menu functions during standby. The settings changes are effective when you exit standby. Enter standby as follows:
6-19
Operation
1. Select Standby. The Entering Standby window opens.
NOTE: Remove the mask/patient interface in order to enter standby. The
ventilator will not enter standby with a patient connected. If the patient is not disconnected, the ventilator continues breath delivery while waiting for the patient to be disconnected. The standby mode request cancels in 60 seconds if the patient remains connected.
NOTE: Standby mode disables alarms and should be used when the patient
is disconnected.
2. Disconnect the patient from the ventilator now. The ventilator enters
standby and displays the Standby screen.
6-20
3. To resume ventilation, reconnect the patient. When the ventilator
Help
message
senses a patient breathing effort, ventilation automatically resumes in the previous mode.
NOTE: You can also manually resume ventilation with the Restart Mode
button.

Help function Select the help button to display additional information.

Help messages are displayed:
Operation
6-21
Operation
Pressure relief
Must be less than or equal to IPAP
EPAP IPAP
Must be greater than or equal to EPAP
EPAP IPAP

Table of modes and control settings

Table 6-3: Modes and control settings with ranges
Setting Description Range
Modes
Modes Ventilation mode AVAPS, CPAP,
S/T, PCV Optional: PPV
Control settings
C-Flex Enhances traditional CPAP by reducing the
pressure at the beginning of exhalation––a time when patients may be uncomfortable with CPAP––and returning it to the set CPAP pressure before the end of exhalation. The amount of pressure relief is determined by the C-Flex setting and the expiratory flow. The higher the setting number (1, 2 or 3) and the greater the expiratory flow, the greater the pressure relief (during the active part of exhalation only).
Applies in CPAP mode only.
CPAP Continuous positive airway pressure. The baseline pressure applied during the expiratory
phase. Applies in CPAP mode only.
E-Cycle (optional)
EPAP Expiratory positive airway pressure. The
Expiratory Cycle Sensitivity. Auto-Trak+ employs several algorithms to determine the point at which the ventilator cycles into exhalation. This setting adjusts all algorithms simultaneously. At the lowest setting (-2), inspiration terminates later, resulting in the longest inspiratory time. At the highest setting (+6), inspiration terminates earlier, resulting in the shortest inspiratory time. Normal is the Auto-Trak setting used when Auto­Trak+ is not enabled.
Applies only when the optional Auto-Trak+ feature is installed.
application and maintenance of pressure above atmospheric at the airway throughout the expiratory phase of positive-pressure mechanical ventilation.
OFF, 1 to 3
4 to 25 cmH
-2, -1, Normal, +1 to +6
4 to 25 cmH
2
2
O
O
IPAP Inspiratory positive airway pressure. The
application and maintenance of pressure above atmospheric at the airway throughout the inspiration phase of positive-pressure mechanical ventilation.
6-22
4 to 40 cmH
O
2
Table 6-3: Modes and control settings with ranges (continued)
Shows where I:E ratio becomes inverse
Resulting I:E ratio
Setting Description Range
Operation
I-Time (Inspiratory
Time to deliver the required gas. Inverse ratio ventilation is not allowed.
Time)
Max E The maximum elastance (volume assist) value used by the PPV mode to overcome the
elastance of the patient's lungs. See also PPV % setting. Applies in PPV mode only.
Max P (AVAPS
The maximum pressure to be applied. Maximum IPAP Pressure)
NOTE: When you adjust the AVAPS minimum and maximum pressures, remember that
IPAP is adjusted to meet the target value. If the calculated target pressure is outside of the minimum and maximum pressure range, the target volume will not be achieved.
Applies in AVAPS mode only.
Max P (PPV Maximum Pressure Limit)
The maximum pressure to be applied. When the limit is reached, the ventilator limits the
pressure and displays a PPV Max P alarm message. If the condition persists for three
consecutive PPV inspirations, an audible alarm also sounds.
Applies in PPV mode only.
0.30 to
3.00 secs
0 to 100 cmH
6 to 40 cmH
2
O/L
2
5 to 40 cmH2O
O
WARNING: PPV limits are not intended to be the primary ventilator alarms and should not be
WARNING: To prevent the delivery of excessive pressure or volume, set the PPV limits appropriately.
substituted for the alarms found in the Alarm Settings window.
Delivery of excessive pressure or volume can occur from a sudden increase in mask leak, inappropriate settings, or a plugged or kinked proximal pressure line. Conversely, insufficient treatment may result if limits are set too low.
Max R The maximum resistance (flow assist) value used by the PPV mode to overcome
pulmonary resistance. See also PPV % setting.
Applies in PPV mode only.
Max V (PPV Maximum Volume Limit)
The maximum volume to be delivered. When the limit is reached, the ventilator terminates
the breath and displays a PPV Max V alarm message. If the condition persists for three
consecutive PPV inspirations, an audible alarm also sounds.
Applies in PPV mode only.
WARNING: PPV limits are not intended to be the primary ventilator alarms and should not be
substituted for the alarms found in the Alarm Settings window.
WARNING: To prevent the delivery of excessive pressure or volume, set the PPV limits appropriately.
Delivery of excessive pressure or volume can occur from a sudden increase in mask leak, inappropriate settings, or a plugged or kinked proximal pressure line. Conversely, insufficient treatment may result if limits are set too low.
0 to 50 cmH L/s
200 to 3500 mL
O/
2
6-23
Operation
Max E and Max R are multiplied by PPV % to obtain the applied Elastance assist and Resistance assist values. Here a Max R
setting of 4 cmH
2
O/L/s and a PPV %
setting of 30% yield a Resistance assist value of 1.2 cmH
2
O/L/s.
PPV %
Initial CPAP/EPAP
CPAP/EPAP 4 cmH2O+
2
-------------------------------------------------------------=
Initial IPAP Initial EPAP
IPAP EPAP()
2
-------------------------------------------
+=
Ramp duration
Ramp
start
pressures
Ramp end pressures
Shows where I:E ratio becomes inverse
Resulting I:E ratio
Proposed rise slope in relation to EPAP and IPAP
Table 6-3: Modes and control settings with ranges (continued)
Setting Description Range
Min P (AVAPS
The minimum pressure to be applied. Minimum IPAP Pressure)
NOTE: When you adjust the AVAPS minimum and maximum pressures, remember that
IPAP is adjusted to meet the target value. If the calculated target pressure is outside of the minimum and maximum pressure range, the target volume will not be achieved.
Applies in AVAPS mode only.
O
2
Oxygen concentration to be delivered. 21 to 100%
PPV % Percentage of PPV assist or gain. This gain
is applied to the Max E and Max R settings,
yielding the applied Elastance and
Resistance assist values.
Applies in PPV mode only.
Ramp Time An interval during which time the ventilator
linearly increases pressure, helping to
reduce patient
anxiety.
5 to 30 cmH
0 to 100%
OFF, 5 to 45 min
O
2
Rate (Respiratory Rate)
Respiratory frequency or number of breaths
per minute.
Inverse ratio ventilation is not allowed.
Rise (Rise Time) Speed with which inspiratory pressure rises
to the set (target) pressure.
If the Rise Time is insufficient to reach the
target IPAP pressure, adjust the Rise Time
or I-Time setting.
6-24
4 to 60 BPM
1 to 5 (1 is fastest)
Table 6-3: Modes and control settings with ranges (continued)
HIP alarm
LIP alarm
IPAP
EPAP
Setting Description Range
Operation
Trigger (optional)
V
(AVAPS
T
Target Tidal Volume)
Trigger Sensitivity. Auto-Trak+ employs several algorithms to determine the point at which the inspiration begins. The larger the value, the more sensitive the trigger (that is, the patient can trigger inspiration with less effort). Normal is the Auto-Trak setting used when Auto-Trak+ is not enabled.
Applies only when the optional Auto-Trak+ feature is installed.
Target tidal volume to be delivered during inspiration. The ventilator meets this target by adjusting the inspiratory pressure with each breath.
Applies in AVAPS mode only.
Table 6-4: Alarm settings
Setting Description Range
Hi Rate (High Rate Alarm) High total breath rate. 5 to 90 BPM
Lo Rate (Low Rate Alarm) Low total breath rate. 1 to 89 BPM
NOTE: In non-CPAP modes, the Low Rate Alarm is
essentially off if set below the Respiratory Rate setting.
Hi V
(High Tidal Volume
T
Alarm)
High exhaled tidal volume. 200 to 3500 mL
Normal, +1 to +7
200 to 2000 mL
Lo V
(Low Tidal Volume
T
Alarm)
HIP (High Inspiratory Pressure Alarm)
LIP (Low Inspiratory Pressure Alarm)
Low exhaled tidal volume. OFF to 1500 mL
High pressure at the patient
5 to 50 cmH
O
2
airway.
Low pressure at the patient
OFF to 40 cmH
O
2
airway.
NOTE: In the S/T and PCV modes, the LIP alarm should
be set 3-5 cmH
O below the IPAP level. When
2
set in this manner, the alarm works in conjunction with the LIP T alarm to indicate if there is a failure to trigger between the two pressure levels. It will also alert the clinician to pressure degradation due to excessive leaks. See figure below.
6-25
Operation
V
Table 6-4: Alarm settings (continued)
Setting Description Range
LIP T (Low Inspiratory Pressure Delay Time)
Lo
(Low Minute
E
Ventilation Alarm)
The interval from the detection of low inspiratory pressure until the alarm becomes active.
Low expiratory minute volume.
5 to 60 secs
OFF to 99.0 L/min
6-26

Chapter 7. High flow therapy

The high flow therapy (HFT) feature is available for 3.00 software and above, as well as V60 Plus. HFT is accessed from the Standby mode. For more information, see “Standby” on page 6-19.
For principles of operation, see “High flow therapy” on page 4-6.
WARNING: When transitioning from a high flow therapy interface to an NIV mask,
ensure that an exhalation port is placed in the circuit and is unobstructed to reduce the risk of CO2 rebreathing.
WARNING: When transitioning from ventilation to high flow therapy, remove the NIV
mask and use only a Philips-approved high flow patient interface to minimize pressure build-up and patient discomfort.
WARNING: When transitioning from high flow therapy to ventilation, remove the nasal
cannula as these are restrictive and may defeat alarms such as patient disconnect. Using a nasal cannula in an NIV mode may lead to hypercarbia due to the inability to provide pressure support.
WARNING: Patient alarms are not available during high flow therapy (HFT) as the
therapy uses an open system. A nasal cannula occupies only a portion of the nares and patients can breathe through their mouth, which prevents estimation of patient parameters such as tidal volume, respiratory rate, pressure, and minute ventilation. Provide external monitoring, including oximetry, to inform the clinician of a change in the patient's condition.
WARNING: During high flow therapy (HFT), verify that an occlusive patient interface
is not being used. Occlusive patient interfaces include a cannula fully sealed within the nares, an NIV mask, or a direct connection to a tracheostomy tube or endotracheal tube. Remove any occlusive interface immediately as this may expose the patient to unintended high pressures.
NOTE: High flow therapy (HFT) is accessed only from the Standby window.
Standby mode cannot be entered if a nasal cannula is connected to the circuit.

Circuit setup See “Installing the patient circuit” on page 5-3 for circuit configuration.

High flow nasal cannula setup

Use either the AC611 with FEP Connector (Figure 7-1) or the AC611 22 mm (Figure 7-2), which connects directly to the patient circuit.
7-1
High flow therapy
FEP
FEP Connect
DEP/FEP
22 mm connector
Connecting to a circuit with an FEP (filter exhalation port) installed
Insert the AC611 FEP Connect into the FEP, making sure the perforations in the port are completely blocked.
Figure 7-1: High flow nasal cannula using the FEP Connect
Connecting directly to a 22 mm circuit
Remove the DEP/FEP. Connect the high flow nasal cannula with 22 mm connector directly to the circuit.
Figure 7-2: High flow nasal cannula, 22 mm connection
7-2
High flow therapy

Changing from an NIV mode to high flow therapy

Follow these instructions to use the V60/V60 Plus Ventilator for high flow therapy (HFT).
1. Select Standby. The Entering Standby window opens.
2. Remove the patient mask or ET interface to enter Standby.
3. Install a Philips-approved nasal cannula (Figure 7-1 and Figure 7-2 above) or a high flow tracheostomy interface on the patient circuit.
4. Select HFT.
5. From the Active Mode window, you can adjust Flow and O
%.
2
6. Press Start HFT.
7. The High Flow Therapy Active message is displayed during HFT.
8. Apply the HFT interface to the patient.
7-3
High flow therapy
9. Note the low priority alarm stating that patient alarms are disabled during HFT. Press alarm reset to confirm this message.

Viewing and pausing the HFT graph

Changing from high flow therapy to an NIV mode

A flow graph is displayed during high flow therapy. Press the Pause button to view an event.
1. Verify that the nasal cannula is removed from the patient and disconnected from patient circuit.
2. Select Standby to open the Standby window.
3. Press the Enter Standby button.
4. In the Select Therapy window, press Ventilation.
5. Replace the high flow patient interface with a Philips-approved NIV mask.
7-4
High flow therapy
6. Review patient settings and alarms.
7. Install the appropriate interface on the patient.
8. Verify that the ventilator detects the patient's breath to activate
ventilation, or press the Start Mode button.

HFT alarms and messages

Table 7-1 is a list of alarms and other messages displayed by the ventilator, along with descriptions, suggested corrective actions, and other information. The ID (identifier) listed with the priority type is the priority number of the alarm. This priority number determines the order of alarm message display. Unless otherwise indicated, alarms listed as autoresettable are reset when the alarm condition is removed.
Table 7-1: HFT Alarm and other messages: summary and troubleshooting
Message Description Corrective Action
Cannot Reach Target Flow
Patient alarms are disabled during HFT
Displays when HFT (high flow therapy) is active. Indicates that flow target is not achieved.
Displays when HFT (high flow therapy) is active. Patient alarms are not available in this therapy.
Check the patient. Check that the nasal cannula size is appropriate for the flow setting. Check that an occlusive interface is NOT in use (a cannula fully sealed within the nares, an NIV mask or direct connection to an ETT/trach). Check for an occlusion, kink or liquid in the patient circuit.
Manually reset to confirm and clear the audible alarm.
Priority type (ID)
Low (66)
Low/ Infor­mation (68)
Manually resettable
No Yes Yes
Yes No Yes
Autore­settable
Silence­able
Patient Circuit Occluded
Displays when HFT (high flow therapy) is active. Gas flow to the patient is obstructed.
Check the patient. Check that an occlusive interface is NOT in use (a cannula fully sealed within the nares, an NIV mask or direct connect to an ETT/trach). Check for an occlusion, kink or liquid in the patient circuit. If problem persists, provide alternative ventilation.
7-5
High (67)
No Yes Yes
High flow therapy
(This page is intentionally blank.)
7-6
Loading...