This oxygen concentrator should be used only under
the supervision of a licensed physician. Please read and
understand this entire manual before using the device.
NRTL
C
Symbol Glossary
No Smoking
Attention, consult
accompanying
documents
Class II equipment
Drip proof
equipment
Type BF applied
part
CSA/NRTL
ComplianceMark
REF Model number
SN Serial Number
ON (power)
OFF (power)
Classification
The Millennium M10 Oxygen Concentrator is classified as:
• IEC Class II Equipment
• Type BF Applied Part
• IPX1 Drip Proof
• Not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide.
• Continuous Operation
Intended Use
The Millennium M10 Oxygen Concentrator is intended to provide supplemental
oxygen to persons requiring oxygen therapy. The device is not intended to be life
supporting or life sustaining.
1
Specifications
Input Voltage: 120 VAC +10%, -10%
Input Frequency: 60 Hz
Average Power Consumption: 600 W
Oxygen Concentration*: 92 ± 4% @ 8-10 LPM
94 ± 2% @ 3-7 LPM
92 + 4% @ 1-2 LPM
Weight: 53 ± 0.5 lbs.
Storage/transport temperature: -30 to 160°F
Operating temperature: 55 to 90°F
Storage/transport humidity: up to 95%, noncondensing
Operating humidity: up to 95%, noncondensing
Operating pressure: 10-30 psig
Altitude: up to 1,368 feet
* Device operation above or outside of the Voltage, LPM, Temperature,
Humidity and/or Altitude values specified may decrease oxygen concentration levels.
Covered by one or more of the following U.S. Patents: 5,060,506; 5,183,483;
5,916,349; 5,996,731; 5,997,617; 6,190,441; 6,348,082; 6,382,931; 6,395,065;
and 6,497,755.
2
Electromagnetic Compatibility
The Millennium M10 is designed to meet the current electromagnetic
compatibility requirements. However, if you suspect operation of this device
interferes with the normal operation of your TV, radio or other electronic
appliance try:
1. Relocating the appliance or the device until the interference stops or
2. Plugging the device into a different power outlet controlled by a
separate circuit breaker or fuse.
Important Safety Requirements
Caution! U.S. federal law restricts this device to sale by or on the
order of a physician.
Warnings: A warning represents the possibility of harm to the
operator or patient
1. For proper operation, your concentrator requires unobstructed
ventilation. The ventilation ports are located at the rear base of the
device and at the side air inlet filter. Always make sure that these
areas are not obstructed by items which may impede ventilation (like
blankets, foam, curtains, etc.). Do not place the concentrator in a
small closed space (such as a closet).
2. Do not remove the covers of this device. Servicing must be referred to
an authorized and trained Respironics home care provider.
3. In the event of an equipment alarm or if you are experiencing any
signs of discomfort consult your home care provider and/or your
health care professional immediately.
4. Oxygen generated by this concentrator is supplemental and should
not be considered life supporting or life sustaining. In certain
circumstances oxygen therapy can be hazardous; any user should
seek medical advice prior to using this device.
5. Where the prescribing health care professional has determined that
an interruption in the supply of oxygen, for any reason, may have
serious consequences to the user, an alternate source of oxygen
should be available for immediate use.
3
6. Oxygen vigorously accelerates combustion and should be kept away
from heat or open flame. Not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen or nitrous oxide.
7. Do not smoke, allow others to smoke or have open flames near the
concentrator when it is in use.
8. Do not use oil or grease on the concentrator or its components as
these substances, when combined with oxygen, can greatly increase
the potential for a fire hazard and personal injury.
9. Do not use the oxygen concentrator if either the plug or power cord is
damaged. Do not use extension cords or electrical adapters.
10. Do not attempt to clean the concentrator while it is plugged into an
electrical outlet.
11. Device operation above or outside of the Voltage, LPM, Temperature,
Humidity and/or Altitude values specified may decrease oxygen
concentration levels.
12. Your home care provider is responsible for performing appropriate
preventive maintenance at the intervals recommended by the device
manufacturer.
Cautions: A caution represents the possibility of damage to the
equipment.
1. Do not operate the concentrator without all the filters in place. The
filters must be totally dry before operating the device.
2. Do not place liquids on or near the device.
3. If liquid is spilled on the device, turn the power off and unplug from
electrical outlet before attempting to clean up spill. Call your home
care provider if device does not continue to work properly.
4
Introduction
Millennium M
1
0
Your health care professional has determined that supplemental oxygen
is of benefit to you and has prescribed an oxygen concentrator set at a
specific flow setting to meet your needs. DO NOT change the flow settings
unless your health care professional tells you to do so. The Millennium M10
oxygen concentrator is intended to provide supplemental oxygen to persons
requiring oxygen therapy. This device is not intended to be life supporting
or life sustaining. The Millennium produces concentrated oxygen from room
air for delivery to a patient requiring low flow oxygen therapy. The oxygen
from the air is concentrated using a molecular sieve and a pressure swing
absorption process. Your home care provider will show you how to operate
the concentrator and will be available to answer any questions. You should
contact him or her in the future if you have any questions or problems.
Parts of Your Concentrator
On/Off
Switch
Flow Meter
Air Filter
Filter Cap
5
Operating
Condition
Indicators
Oxygen
Outlet Port
Operating Instructions
1. Select a location that allows the concentrator to draw in room air
without being restricted. Make sure that the device is at least 3 inches
away from walls, furniture, and especially curtains that could impede
adequate airflow to the device. Do not place the device near any heat
source.
2. After reading this entire manual, plug the power cord into an electrical outlet.
WARNING:
DO NOT USE EXTENSION
CORDS OR ELECTRICAL
ADAPTERS.
6
3. Connect the oxygen tubing to the oxygen outlet port.
7
3. Press the power switch to the On [ I ] position. Initially, all the indicator
Millennium M10
lights will turn on and the audible alarm will sound for a few seconds.
After that time, only the green indicator should remain lit.
4. After turning on the device, you should allow at least 10 minutes for
the device to warm up.
5. Adjust the flow to the prescribed setting by turning the knob on the
top of the flow meter until the ball is centered on the line marking the
specific flow rate.
6. Be sure oxygen is flowing through the cannula. If it is not, refer to the
Troubleshooting section.
7. Put the cannula on as directed by your home care provider.
8. When not using the device, press the power switch to the Off [O]
position.
Accessory Equipment
Use only the following Respironics™ accessories with the M10:
- Part Number 1014653 Cannula and Tubing
Contact your home care provider if you have questions concerning the
use of these accessories.
8
Maintenance
Warning: It is i m portant to un p lug the devi ce before any
maintenance or cleaning is performed.
Cleaning and Changing the Air Inlet Filter
Cleaning the air inlet filter is the most important maintenance activity
that you will perform and should be done at least once a week.
1. Remove the filter cap.
2. Remove the filter from the cabinet.
3. Visually inspect the filter for damage, such as holes or tears.
4. If damaged, replace with a new filter. Contact your health care
provider for a new filter if necessary.
5. Rinse and wash the filter in warm water. A mild detergent may be
used if rinsed thoroughly.
6. Squeeze out the excess water and allow the filter to air dry. The filter
should be completely dry before using again. Excess moisture may
impair the proper operation of the device.
7. Visually inspect the filter after cleaning. Make sure it is not damaged
or clogged.
8. Reinsert the filter on the cabinet.
9. Reinstall the filter cap.
Cleaning the Device
You should also clean the exterior case of the device. You may use a
damp cloth to wipe down the exterior case. If medical disinfectants are used,
be sure to follow the manufacturer's instructions.
9
Power Interrupt Alarm
service
alarm
power
normal
read m
an
u
al
low o
x
ygen
required
Millennium
oxygen
If the device is turned on but not operating, the alarm condition will be
activated. This is often the result of either the device not being plugged in or
a power failure. This feature is powered by an internal 9V battery.
To test the condition of the battery you should test this alarm state often
by unplugging the device and switching the power on. If the alarm condition
is not activated, you should contact your home care provider to replace
the battery. If the device shuts down because of a power failure, turn the
power switch off for two minutes before restarting to allow the system to
depressurize itself.
Operating Condition Indicators and Alarms
• Power: The GREEN indicator illuminates when the device is turned
on.
• Normal Oxygen (Model M10605 only): The extra GREEN indicator
illuminates when the device is delivering a normally expected
percentage of oxygen. The illumination of this indicator is delayed for
several minutes after start up to allow the system to stabilize.
• Low Oxygen: The YELLOW indicator illuminates continuously* to
alert the user when some corrective action is required. Following
the troubleshooting guide in this manual, this action can normally
be performed by the user. Call your home care provider if the
troubleshooting actions fail to end this alert condition.
• Alarm: The RED indicator illuminates continuously and the audible
alarm sounds continuously if the device detects a system malfunction.
If this happens when the device is plugged in, you should turn off the
device and call your home care professional immediately.
* Units may have intermittent RED indicator and sound an audible alarm
accompanying the YELLOW indicator under a Low Oxygen condition.
10
Minor Troubleshooting Guide
ProblemProbable CauseSolution
The device is not
working when it is
turned on; the alarm
condition (red light)
is activated and
the power indicator
(green light) is not
illuminated.
The power cord
plug is not properly
inserted into the
electrical outlet.
No power from the
electrical outlet.
Internal part failure.
Make sure the device is properly
plugged in to the electrical
outlet.
Check your household outlet
fuse or circuit breaker.
Contact your home care
provider.
Low oxygen (yellow
light) activated.*
Alarm condition
(red light) activated,
when the device is
on with the power
indicator (green
light) on.
Limited oxygen flow
to the user.
Alarm is weak or not
sounding when the
power cord is unplugged and power
switch is on.
The airflow to the
device is impeded or
blocked.
The flowmeter knob
is completely closed.
The oxygen tubing is
kinked and blocking
the delivery of
oxygen.
Internal system
failure.
Faulty oxygen
tubing, cannula.
Poor connection of
device accessory.
The 9V battery is
weak.
Check the condition of the air
inlet filter and clean or replace
as necessary. Remove any
items that appear to be blocking
the airflow into the device.
Turn flowmeter knob
counterclockwise to center ball
on prescribed LPM Flow.
Check to see that the tubing is
not kinked or blocked. Replace if
necessary.
Contact your home care provider
immediately.
Inspect and replace the items if
necessary.
Ensure that all connections are
free from leaks.
Contact your home care
provider.
* Units may have intermittent RED indicator and sound an audible alarm accompanying the YELLOW
indicator under a Low Oxygen condition.
11
12
1024574
AM 4/19/05