Patient Manual
This oxygen concentrator should be used only under
the supervision of a licensed physician. Please read and
understand this entire manual before using the device.

Symbol Glossary
No Smoking
Attention, consult
accompanying
documents
Class II equipment
Drip proof
equipment
Type BF applied
part
CSA/NRTL
ComplianceMark
REF Model number
SN Serial Number
ON (power)
OFF (power)
Classification
The Millennium M10 Oxygen Concentrator is classified as:
• IEC Class II Equipment
• Type BF Applied Part
• IPX1 Drip Proof
• Not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide.
• Continuous Operation
Intended Use
The Millennium M10 Oxygen Concentrator is intended to provide supplemental
oxygen to persons requiring oxygen therapy. The device is not intended to be life
supporting or life sustaining.
1

Specifications
Input Voltage: 120 VAC +10%, -10%
Input Frequency: 60 Hz
Average Power Consumption: 600 W
Oxygen Concentration*: 92 ± 4% @ 8-10 LPM
94 ± 2% @ 3-7 LPM
92 + 4% @ 1-2 LPM
Weight: 53 ± 0.5 lbs.
Storage/transport temperature: -30 to 160°F
Operating temperature: 55 to 90°F
Storage/transport humidity: up to 95%, noncondensing
Operating humidity: up to 95%, noncondensing
Operating pressure: 10-30 psig
Altitude: up to 1,368 feet
* Device operation above or outside of the Voltage, LPM, Temperature,
Humidity and/or Altitude values specified may decrease oxygen concentration levels.
Covered by one or more of the following U.S. Patents: 5,060,506; 5,183,483;
5,916,349; 5,996,731; 5,997,617; 6,190,441; 6,348,082; 6,382,931; 6,395,065;
and 6,497,755.
2

Electromagnetic Compatibility
The Millennium M10 is designed to meet the current electromagnetic
compatibility requirements. However, if you suspect operation of this device
interferes with the normal operation of your TV, radio or other electronic
appliance try:
1. Relocating the appliance or the device until the interference stops or
2. Plugging the device into a different power outlet controlled by a
separate circuit breaker or fuse.
Important Safety Requirements
Caution! U.S. federal law restricts this device to sale by or on the
order of a physician.
Warnings: A warning represents the possibility of harm to the
operator or patient
1. For proper operation, your concentrator requires unobstructed
ventilation. The ventilation ports are located at the rear base of the
device and at the side air inlet filter. Always make sure that these
areas are not obstructed by items which may impede ventilation (like
blankets, foam, curtains, etc.). Do not place the concentrator in a
small closed space (such as a closet).
2. Do not remove the covers of this device. Servicing must be referred to
an authorized and trained Respironics home care provider.
3. In the event of an equipment alarm or if you are experiencing any
signs of discomfort consult your home care provider and/or your
health care professional immediately.
4. Oxygen generated by this concentrator is supplemental and should
not be considered life supporting or life sustaining. In certain
circumstances oxygen therapy can be hazardous; any user should
seek medical advice prior to using this device.
5. Where the prescribing health care professional has determined that
an interruption in the supply of oxygen, for any reason, may have
serious consequences to the user, an alternate source of oxygen
should be available for immediate use.
3

6. Oxygen vigorously accelerates combustion and should be kept away
from heat or open flame. Not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen or nitrous oxide.
7. Do not smoke, allow others to smoke or have open flames near the
concentrator when it is in use.
8. Do not use oil or grease on the concentrator or its components as
these substances, when combined with oxygen, can greatly increase
the potential for a fire hazard and personal injury.
9. Do not use the oxygen concentrator if either the plug or power cord is
damaged. Do not use extension cords or electrical adapters.
10. Do not attempt to clean the concentrator while it is plugged into an
electrical outlet.
11. Device operation above or outside of the Voltage, LPM, Temperature,
Humidity and/or Altitude values specified may decrease oxygen
concentration levels.
12. Your home care provider is responsible for performing appropriate
preventive maintenance at the intervals recommended by the device
manufacturer.
Cautions: A caution represents the possibility of damage to the
equipment.
1. Do not operate the concentrator without all the filters in place. The
filters must be totally dry before operating the device.
2. Do not place liquids on or near the device.
3. If liquid is spilled on the device, turn the power off and unplug from
electrical outlet before attempting to clean up spill. Call your home
care provider if device does not continue to work properly.