BiPAP systems are the subject of one or more of U.S. Patents #5148802, #5239995, #531937,
#5433193, Canadian Patent #2, 024, 477, European Patent #EP0425092, German Patent
#69021681.5-08, and other pending U.S. and foreign patents. BiPAP is a registered trademark of
Respironics, Inc.
TELEPHONE COMPANY PROCEDURES .................................................................................................................................. 1-5
IF PROBLEMS ARISE .......................................................................................................................................................... 1-5
2.2.2 S MODE .............................................................................................................................................................. 2-2
3.1.2 CONTROL KEYS .................................................................................................................................................... 3-2
4.1.2 INSTALLINGTHE AIR FILTERS................................................................................................................................. 4-2
4.5.3 DATA TRANSMISSION ERRORS ................................................................................................................................ 4-9
4.5.4 USING THE RAMP FUNCTION ............................................................................................................................... 4-10
5.2.2 S MODE .............................................................................................................................................................. 5-3
5.2.3 T , PC, AND S/T MODES ....................................................................................................................................... 5-3
7.2.2 SYSTEM CODES .................................................................................................................................................... 7-3
7.2.3 PATIENT ALARM HISTORY ..................................................................................................................................... 7-4
7.2.8 CHANGINGTHE LANGUAGE ................................................................................................................................. 7-11
Index ......................................................................................................................... I-1
R
Provider Manua
iv
l
R
Provider Manua
l
Package Contents
•BiPAP Synchrony
•power cord
•flexible tubing (1.8 m x 2 mm I.D.)
•filter cap
•pollen filters
•ultra-fine filter
•optional modem cable (4.27 m)
•User Guide*
R
BiPAP Synchrony
Filter Cap
Ultra-fine Filter
Pollen Filters
R
User Guide
User Guide*
www.respironics.com
Power Cord
Flexible Tubing
Optional Modem
Cable
* Not included with all models.
R
Provider Manua
R
l
Provider Manua
l
Chapter 1: Warnings and Cautions
WARNING: Indicates the possibility of injury to the patient or the operator.
CAUTION:Indicates the possibility of damage to the device.
CAUTION!US federal law restricts this device to sale by or on the order of a physician.
NOTE:Places emphasis on an operating characteristic.
1-1
1.1 Warnings
•This manual serves as a reference. The instructions in this manual are
not intended to supersede the health care professional’s instructions
regarding the use of the Synchrony.
•The operator should read and understand this entire manual before
using the Synchrony.
•The Synchrony provides positive pressure ventilation and is indicated
for assisted ventilation. The Synchrony does not provide ventilation
with guaranteed tidal volume delivery. Patients requiring ventilation
at predetermined tidal volumes are not candidates for pressure support
ventilation.
•The Synchrony should be used only with masks and connectors
recommended by Respironics or with those recommended by the
health care professional or respiratory therapist. See Chapter 12 for
approved patient circuits. A mask should not be used unless the
Synchrony is turned on and operating properly. The exhalation port(s)
associated with the mask should never be blocked.
Explanation of the Warning: The Synchrony is intended to be used
with special masks or connectors that have exhalation ports to allow
continuous flow of air out of the mask. When the Synchrony is turned
on and functioning properly, new air from the Synchrony flushes the
exhaled air out through the mask exhalation port. However, when the
Synchrony is not operating, enough fresh air will not be provided
through the mask, and exhaled air may be rebreathed. Rebreathing of
exhaled air for longer than several minutes can in some circumstances
lead to suffocation.
•In the event of a power or device failure, audible and visual alarm
signals will activate. The Synchrony must be disconnected from the
patient immediately. As is the case with most ventilators with passive
exhalation ports, when power is lost, sufficient air will not be provided
through the circuit and exhaled air may be rebreathed.
•At low EPAP pressures, the flow through the exhalation port may be
inadequate to clear all exhaled gas from the tubing. Some rebreathing
may occur.
•If supplemental oxygen is added to the breathing circuit, the optional
oxygen valve must be installed and used as shown in Chapter 10. The
valve automatically stops the flow of oxygen when the blower is off.
•If oxygen is used with the Synchrony, the oxygen flow must be turned
off when the Synchrony is not operating.
Explanation of the Warning: When the Synchrony is not in operation,
and the oxygen flow is left on, oxygen delivered into the ventilator
tubing may accumulate within the Synchrony’s enclosure. Oxygen
accumulated in the ventilator enclosure will create a risk of fire.
•Oxygen supports combustion. Oxygen should not be used while
smoking or in the presence of an open flame.
R
Provider Manua
1-2
l
W arnings (continued)
•When administering fixed-flow supplemental oxygen, the oxygen
concentration may not be constant. The inspired oxygen concentration
will vary, depending on the IPAP and EPAP settings, patient breathing
pattern, and the leak rate. Substantial leaks around the mask may
reduce the inspired oxygen concentration to less than the expected
concentrations shown in Chapter 10. Appropriate patient monitoring
should be used.
•Operation of the Synchrony may be adversely affected by
—electromagnetic fields exceeding the level of 10 V/m in the test
conditions of EN 60601-1-2
—the operation of high frequency (diathermy) equipment
—defibrillators, or short wave therapy equipment
—radiation (e.g., x-ray, CT)
—magnetic fields (e.g., MRI).
•The Synchrony is not suitable for use in the presence of flammable
mixtures such as gases, anesthetics and liquids.
•Do not use the Synchrony at room temperatures above 35°C. If the
Synchrony is used at room temperatures above 35°C, the temperature
of the airflow may exceed 41°C, which could cause thermal irritation
or injury to the patient’s airway.
•Do not operate the Synchrony in direct sunlight or near a heating
appliance because these conditions can increase the temperature of the
airflow delivered to the patient.
•To reduce the risk of contamination, a low resistance main flow
bacteria filter may be placed in-line between the Synchrony and
the patient.
•The Synchrony does not have an alarm to detect occlusion of the
exhalation port. Before each use, inspect the patient circuit to verify
that the port is not occluded. Occlusion or partial occlusion can reduce
airflow and result in rebreathing of exhaled air.
•Do not use antistatic or electrically conductive hoses or tubing with the
Synchrony.
•When the Synchrony is used with a humidifier, position the humidifier
such that the water level in the humidifier is lower than the patient, and
the humidifier is on the same level or lower than the Synchrony.
•If you detect any unexplained changes in the performance of the
Synchrony, if the Synchrony is dropped or mishandled, if water is
spilled into the enclosure, or if the enclosure is broken, seek the
assistance of Respironics or an authorized service center.
•Repairs and adjustments must be performed by Respironics or an
authorized service center. Service done by inexperienced or
unqualified personnel, or installation of unauthorized parts could cause
injury, invalidate the warranty, or result in costly damage.
•Electrical cords and cables should be periodically inspected for
damage or signs of wear.
•To avoid electrical shock, unplug the Synchrony before cleaning it.
•Verify the operation of the patient disconnect alarm with any changes
in the patient circuit.
R
Provider Manua
1-3
l
Warnings (continued)
1.2 Cautions
•Verify that the Patient Disconnect alarm is active if required for
medical reasons.
•The Synchrony may only be operated at temperatures between 5°C and
35°C.
•Do not immerse the Synchrony or allow any liquid to enter the
enclosure or the inlet filter.
•Condensation may damage the Synchrony. Always allow the
Synchrony to reach ambient temperature before use.
•Use the AC cable retainer to keep the AC power cord from falling out.
Additional Warnings, Cautions, and Notes are located throughout
this manual.
R
Provider Manua
1-4
l
1.3 Intended Use
WARNING! Do not connect any equipment to the Synchrony unless recommended by Respironics or the health care
professional. Verify that an exhalation port is present to exhaust CO2 from the circuit. If circuit accessories other than those recommended by Respironics are connected to the Synchrony, then pressures must
be verified. Use of these accessories may alter the pressure received, reducing the effectiveness of
treatment.
1.4 Contraindications
The Synchrony is intended to provide noninvasive ventilation in adult patients
(>30 kg) for the treatment of respiratory insufficiency (a condition in which the
patient can continue without ventilation for some period, such as overnight) or
obstructive sleep apnea. The Synchrony may be used in the hospital or home.
The Synchrony is intended for use with nasal masks and full-face masks as
recommended by Respironics.
The use of the Synchrony is contraindicated on patients without a spontaneous
respiratory drive.
The use of the Synchrony for noninvasive positive pressure therapy may be
contraindicated on patients
•unable to maintain a patent airway or adequately clear
secretions,
•at risk for aspiration of gastric contents,
•with acute sinusitis or otitis media,
•with a history of allergy or hypersensitivity to the mask materials
where the risk from allergic reaction outweighs the benefit of
ventilatory assistance,
•with epistaxis, causing pulmonary aspiration of blood, or
•with hypotension.
NOTE:When assessing the relative risks and benefits, the health care professional should understand that the
Synchrony can be set to deliver pressures up to 30 cm H2O. Also, in the unlikely event of certain fault
conditions, a maximum static pressure of 40 cm H2O is possible.
1.5 Patient Precautions
•Advise the patient to immediately report any unusual chest discomfort,
shortness of breath, or severe headache.
•If skin irritation or breakdown develops from the use of the mask,
refer to the mask instructions for appropriate action.
•The following are potential side effects of noninvasive positive
pressure therapy:
—ear discomfort
—conjunctivitis
—skin abrasions due to noninvasive interfaces
—gastric distention (aerophagia).
R
Provider Manua
l
1.6 Communication Statements for Optional Modem
NOTE:An optional internal modem is available in models of the Synchrony sold in the United States and
Canada. If your model does not have this feature, you can connect the Synchrony to an external modem.
The following statements apply to use of the Synchrony with a modem through U.S. telephone lines.
1-5
TYPESOF SERVICE
TELEPHONE COMPANY
PROCEDURES
IF PROBLEMS ARISE
The Synchrony optional internal modem is designed to be used on standard
device telephone lines. It connects to the telephone line by means of a standard
jack called the USOC RJ-11C (or USOC FJ45S). Connection to telephonecompany-provided coin service (central office implemented systems) is
prohibited. Connection to party lines service is subject to state tariffs.
The goal of the telephone company is to provide you with the best service it
can. In order to do this, it may occasionally be necessary for them to make
changes in their equipment, operations or procedures. If these changes might
affect your service or the operation of your equipment, the telephone company
will give you notice, in writing, to allow you to make any changes necessary to
maintain uninterrupted service. In certain circumstances, it may be necessary
for the telephone company to request information from you concerning the
equipment that you have connected to your telephone line. Upon request of the
telephone company, provide the FCC registration number and the ringer
equivalence number (REN); both of these items are listed on the equipment
label. The sum of all the RENs on your telephone lines should be less than five
in order to assure proper service from the telephone company. In some cases, a
sum of five may not be usable on a given line.
If any of your telephone equipment is not operating properly, you should
immediately remove it from your telephone line, as it may cause harm to the
telephone network. If the telephone company notes a problem, they may
temporarily discontinue service. When practical they will notify you in advance
of the disconnection. If advance notice is not feasible, you will be notified as
soon as possible. When you are notified, you will be given the opportunity to
correct the problem and informed of your right to file a complaint with the
FCC. Contact your telephone company if you have questions about your phone
line. In the event repairs are ever needed on the Synchrony, they should be
performed by Respironics or an authorized representative of Respironics. For
information contact Respironics. (See page 2-9.)
R
Provider Manua
1-6
l
1.7 Industry Canada Notice
NOTICE: The Industry Canada Label identifies certified equipment. This certification means that the equipment meets
telecommunications network protective, operational, and safety requirements as prescribed in the appropriate Terminal
Equipment Technical Requirements documents. The Department does not guarantee the equipment will operate to the
user’s satisfaction.
Before installing this equipment, users should make sure that it is permissible to be connected to the facilities of the local
telecommunications company. The equipment must also be installed using an acceptable method of connection. The
customer should be aware that compliance with the above conditions may not prevent degradation of service in some
situations.
Repairs to certified equipment should be coordinated by a representative designated by the supplier. Any repairs or
alterations made by the user to this equipment, or equipment malfunctions, may give the telecommunications company
cause to request the user to disconnect the equipment.
Users should ensure for their own protection that the electrical ground connections of the power utility, telephone lines
and internal metallic water pipe system, if present, are connected together. The precaution may be particularly important
in rural areas.
CAUTION: Users should not attempt to make such connections themselves, but should contact the appropriate electric
inspection authority, or electrician, as appropriate.
Ringer Equivalence Number (REN): The REN assigned to each terminal device provides an indication of the maximum number of terminals allowed to be connected to a telephone interface. The termination on an interface may consist
of any combination of devices subject only to the requirement that the sum of the Ringer Equivalence Numbers of all the
devices does not exceed 5.
R
Provider Manua
l
Chapter 2: Introduction
2.1 Synchrony Overview
WARNING:The Synchrony can operate on AC or DC power. The DC power option is not intended as a battery
backup during use of AC power.
CAUTION:When DC power is obtained from a vehicle battery, the Synchrony should not be used while the
engine of the vehicle is running. Damage to the vehicle may occur.
NOTE:The Synchrony is available in two models. The BiPAP Synchrony Lab System is for use by
clinicians in a sleep lab setting. The BiPAP Synchrony Ventilatory Support System is for use in a
hospital or home setting. Throughout this manual, references to “Synchrony” will apply to both the
Lab System and the Ventilatory Support System unless otherwise noted.
NOTE:Throughout this manual, references to Bi-Flex apply only to the Synchrony Lab System.
The Synchrony, shown in Figure 2-1, is a low-pressure, electrically driven
ventilator system with electronic pressure control. The Synchrony’s pressure
controls are adjusted to deliver pressure support for patient ventilatory
R
Figure 2-1 Synchrony
assistance.
The Synchrony is intended to augment patient breathing by supplying
pressurized air through a patient circuit. It senses the patient’s breathing effort
by monitoring airflow in the patient circuit and adjusts its output to assist in
inhalation and exhalation. This assistance is provided by the administration of
two levels of positive pressure. During exhalation, pressure is variably positive
or near ambient. During inspiration, pressure is variably positive and always
higher than the expiratory level.
2-1
The Synchrony responds reliably to patient flow rates that indicate movement
to inhalation or exhalation, even in the presence of most normal leaks in the
patient circuit. Automatic adjustment of this trigger threshold in the presence of
leaks makes the system ideal for mask-applied ventilation assistance. The
patient-controllable Rise Time may enhance patient-ventilator synchrony and
patient comfort.
If the patient interface is disconnected, the Synchrony automatically reduces the
output flow to a low level. When the interface is replaced, therapy will typically
resume when the patient starts to breathe. If therapy does not resume when the
patient starts to breathe, the patient should press the RAMP/ALARM SILENCE
button to resume therapy.
A modem allows the patient to provide compliance data to the home care
provider with a few simple key presses. A communications port provides
communication to the Respironics Encore® or Encore Pro Data Management
Software.
The Synchrony operates in the following modes:
•Continuous Positive Airway Pressure (CPAP)
•Spontaneous (S)
•Spontaneous/Timed (S/T)
•Timed (T)
•Pressure Control (PC).
R
Provider Manua
2-2
l
2.2 Modes of Operation
2.2.1CPAP MODE
2.2.2S MODE
In the Continuous Positive Airway Pressure (CPAP) mode, the Synchrony
delivers a continuous pressure support ventilation at one pressure level.
The Spontaneous (S) mode of the Synchrony delivers bi-level pressure support.
The unit triggers to Inspiratory Positive Airway Pressure (IPAP) in response to
spontaneous inspiratory effort and cycles to Expiratory Positive Airway
Pressure (EPAP) during exhalation. Figure 2-2 illustrates the trigger and cycle
concepts.
Cycle
Inspiration
Trigger
Expiration
IPAP
EPAP
Rise Time
Inspiratory Time
Figure 2-2 Triggering and Cycling in S Mode
The level of pressure support (PS) delivered is determined by the difference
between the IPAP and EPAP settings (PS = IPAP - EPAP).
Bi-Flex
The Synchrony Lab System provides the Bi-Flex feature in S mode. The BiFlex attribute adjusts therapy by inserting a small amount of pressure relief
during the latter stages of inspiration and during active exhalation (the
beginning part of exhalation). In the following diagram, the bold lines represent
Bi-Flex in comparison to the dashed line representing normal BiPAP therapy.
Bi-Flex levels of 1, 2, or 3 progressively reflect increased pressure relief that
will take place at the end of inspiration and at the beginning of expiration.
IPAP
EPAP
Bi-Flex
BiPAP
1
2
3
Time
Figure 2-3 Bi-Flex in Comparison to Traditional Bi-level Therapy
R
Provider Manua
2-3
l
2.2.3S/T MODE
The Spontaneous/Timed (S/T) mode of the Synchrony is similar to the S mode,
except that it also can deliver a machine-triggered breath if the patient does not
spontaneously breath within a set time.
The S/T mode ensures that patients will receive a minimum number of breaths
per minute based on the Rate setting. If the patient fails to initiate an inspiration
within the interval determined by the Rate control, the unit triggers a timed (or
machine-triggered) breath resulting in a pressure-control (pressure-limited,
time-cycled) breath at the set IPAP level. The rate of timed breaths is
adjustable. The duration of each timed breath is controlled by an Inspiratory
Time control. See Figure 2-4 for an example of patient-triggered and machinetriggered breaths.
Example:
IPAP = 14 cm H2O
EPAP = 6 cm H2O
Rate = 10 BPM
PS = 8 cm H2O
Figure 2-4 Example of Patient-Triggered and Machine-Triggered Breaths
The Pressure Control (PC) mode is similar to the S/T mode, except that all
breaths are machine-cycled. The PC mode is a pressure-limited, machine- or
patient-triggered, time-cycled mode. Therefore, the inspiratory pressure may be
triggered by the patient or by the Synchrony, but IPAP will be pressure-limited
with a set cycle time determined by the Inspiratory Time control.
The Timed (T) mode provides mandatory pressure assist with bi-level
pressures. All breaths are machine-triggered and machine-cycled; the patient’s
breathing rate has no effect on the machine rate or pressure levels. The trigger
to IPAP is determined by the Rate control, and the cycle time is determined by
the Inspiratory Time control.
R
Provider Manua
2-4
l
2.2.6MODE SUMMARY
2.3 Digital Auto-T rak
Sensitivity™
2.3.1LEAK TOLERANCE
Table 2-1 summarizes the initiation of triggers and cycles for each bi-level
mode.
ModeTriggerLimitCycle
SPatientPressurePatient or Device
S/TPatient or MachinePressurePatient or Device
PCPatient or MachinePressureDevice
TMachinePressureDevice
Table 2-1 Initiation of Triggers and Cycles for Synchrony Modes
An important characteristic of the Synchrony is its ability to recognize and
compensate for unintentional leaks in the system and to automatically adjust its
trigger and cycle algorithms to maintain optimum performance in the presence
of leaks. This feature is known as Digital Auto-Trak Sensitivity. The
following sections examine this function in detail by describing the leak
tolerance function and sensitivity.
A microprocessor monitors the total flow of the patient circuit and calculates
patient flow values.
2.3.2SENSITIVITY
1. Leak Estimation—Average and Parabolic
The Synchrony uses two leak estimation algorithms. A conservation of mass
algorithm is used to compute the average leak for a given pressure support
relationship. This average leak is used when large leak variations are present in
the system. Average leak is a high estimate during EPAP pressure and a low
estimate during IPAP pressure.
A better leak estimate, enabled by the digital system, is the parabolic leak
algorithm. Parabolic leak is proportional to the square of the patient pressure;
therefore, the leak estimate is correlated to the changing patient pressure. Both
algorithms include unintentional circuit leak and are averaged over several
breaths.
2. Patient Flow
The total circuit flow is comprised of the circuit leak and the patient flow. The
calculated patient flow is the total flow minus the circuit leak. Patient flow is a
primary input into the triggering and cycling mechanisms.
An essential feature of the Synchrony while operating in the S, S/T, and PC
modes is its ability to effectively sense spontaneous breathing efforts, which
causes the ventilator to trigger to IPAP and cycle to EPAP. Because no preset
sensitivity threshold can assure patient and machine synchrony with changing
breathing efforts and circuit leaks, the Synchrony continuously tracks patient
breathing patterns and automatically adjusts sensitivity thresholds to ensure
optimum sensitivity as breathing patterns change or as circuit leaks change.
The algorithms used to ensure optimum sensitivity are the Volume Trigger,
Shape Signal, and the Spontaneous Expiratory Threshold (SET).
R
Provider Manua
2-5
l
Volume Trigger
(EPAP to IPAP)
Shape Trigger/Shape Cycle
(EPAP to IPAP)
(IPAP to EPAP)
The volume trigger is one method used to trigger IPAP during spontaneous
breathing in the S, S/T, and PC modes. The volume trigger threshold is 6 cc of
accumulated patient inspiratory volume. When patient effort generates
inspiratory flow causing 6 cc of volume, IPAP is triggered.
The shape trigger/cycle is another method used to trigger IPAP and/or cycle
from IPAP to EPAP during spontaneous breathing in the S, S/T, and PC modes.
This method continuously tracks patient inspiratory and expiratory flow and
adjusts the spontaneous trigger and cycle thresholds for optimum sensitivity.
The Shape Signal appears as a shadow image of the patient’s actual flow. The
shape signal functions as a sensitivity threshold at either inspiration or
expiration. When the patient’s flow rate crosses the shape signal the unit
changes pressure levels. Figure 2-5 illustrates how the shape signal is
superimposed onto the actual waveform to trigger and cycle off IPAP.
The shape signal is created by offsetting the signal from the actual patient flow
by 15 L/min and delaying it for a 300 msec period. This intentional delay
causes the shape signal to be slightly behind the patient’s flow rate. A sudden
change in patient flow will cross the shape signal, causing the pressure level to
change.
IPAP
PRESSURE
EPAP
Cycle to
EPAP
Crossover
Point
Trigger to
IPAP
Crossover
Point
FLOW
Shape
Signal
Estimated
Patient
Flow
Figure 2-5 Shape Signal
Tracking the patient’s flow pattern with the Shape Signal provides a sensitive
mechanism to trigger to IPAP or cycle to EPAP in response to changing
breathing patterns and circuit leaks.
R
Provider Manua
2-6
l
Spontaneous Expiratory
Threshold
(IPAP to EPAP)
A second method used to cycle off IPAP during spontaneous breathing in the
S, S/T, and PC modes is called Spontaneous Expiratory Threshold (SET). The
SET rises in proportion to the inspiratory flow rate on each breath. When the
Spontaneous Expiratory Threshold (SET) and actual patient flow value are
equal, the unit cycles to EPAP.
IPAP
PRESSURE
EPAP
Spontaneous
Expiratory
Threshold
FLOW
Maximum IPAP Time
(IPAP to EPAP)
Flow Reversal
(IPAP to EPAP)
Summary
Figure 2-5 Spontaneous Expiratory Threshold
A maximum IPAP time of 3.0 seconds acts as a safety mechanism to limit the
time spent at the IPAP level during spontaneous breathing in the S, S/T, and PC
modes. Once the time limit is reached, the unit automatically cycles off IPAP
to the EPAP level.
As flow begins to decrease during IPAP, a flow reversal can occur due to a
large leak around the mask or because the patient’s mouth is open. When the
Synchrony unit senses this flow reversal, the unit automatically cycles to the
EPAP level.
The sensitivity criteria for spontaneous breathing in the S, S/T, and PC modes
can be summarized as follows:
Spontaneous Trigger to IPAP
A transition from EPAP to IPAP will occur when one of the following
conditions is met:
• Patient flow exceeds the shape signal
• 6 cc inspired patient volume
Cycle to EPAP
The transition from IPAP to EPAP will occur when one of the following
conditions is met:
• Patient flow is less than the shape signal
• Spontaneous Expiratory Threshold (SET) is achieved
• A 3.0 second maximum IPAP time has occurred (safety feature)
• Flow reversal occurs during IPAP (safety feature)
R
Provider Manua
2-7
l
2.4 Access Levels
The Synchrony can be set to a Lockout mode so that certain settings and
displays cannot be accessed. In the Lockout mode, the IPAP Rise Time can be
adjusted if Bi-Flex is not enabled. If Bi-Flex is enabled, then it is adjustable in
Lockout Mode. Rise Time cannot be adjusted in Lockout mode when Bi-Flex is
enabled. Only the IPAP, EPAP, and Respiratory Rate are displayed in Lockout
Mode.
In the Full Access level, all ventilator functions and displays are accessible.
Section 7.2.6 describes the procedure to set the access level.
R
Provider Manua
2-8
l
2.5 Symbol Key
SymbolMeaning
The following symbols are used on the Synchrony:
AC Power Indicator
DC Power Indicator
Alarm Indicator
Type BF Applied Part
Class II for Protection Against Electric Shock
Attention, consult accompanying documents
Standby Key (Start/Stop)
Up Key
Down Key
Enter Key
Ramp Start/Alarm Silence Key
R
Provider Manua
l
2.6 Definitions, Acronyms, and Abbreviations
The following terms are used in this manual:
Bi-Flex—A therapy feature that establishes a level of pressure relief taking
place at the end of inhalation and at the start of exhalation.
BiPAP—Bi-level Positive Airway Pressure
CPAP—Continuous Positive Airway Pressure
Cycle—The transition from inspiration to expiration
EPAP—Expiratory Positive Airway Pressure
High Priority Alarm—Alarm signal indicating a condition that requires
immediate attention
IPAP—Inspiratory Positive Airway Pressure
Low Priority Alarm—Signal indicating an information message
2-9
2.7 Service
Medium Priority Alarm—Alarm signal indicating a condition that requires
operator awareness
OSA—Obstructive Sleep Apnea
Ramp—A feature that may increase patient comfort when therapy is started.
The Synchrony IPAP starts at the EPAP level and is increased gradually (breath
by breath over several breaths) until the IPAP prescription pressure is reached.
RR—Respiratory Rate
Trigger—The transition from expiration to inspiration
If you need product support, call Respironics’ Customer Service Department:
1-800-345-6443 (USA or Canada only) or 1-724-387-4000.
R
Provider Manua
2-10
l
R
Provider Manua
l
Chapter 3: Controls and Displays
This chapter describes the control panel and displays, the patient circuit
connections, and the rear panel connections.
3.1 Control Panel
Display
Screen
AC Power
Indicator
DC Power
Indicator
3-1
R
Red
Alarm Indicator
Yellow
Alarm Indicator
3.1.1DISPLAY SCREEN
Down
Key
Standby
Key
Up
Key
Figure 3-1 The Synchrony Control Panel
Figure 3-1 illustrates the Synchrony control panel. The control panel includes:
•display screen
•control keys
•alarm indicators
•power indicators.
The display screen is used to display operating parameters, instructions, and
messages. A backlight is activated when the , , or key is pressed, and
remains on until there are no keystrokes for one minute. If the backlight is off,
the first press of any of these keys will only turn the backlight on; the normal
key action is suppressed until the key is pressed a second time.
Ramp/
Alarm Silence
Key
Enter
Key
R
Provider Manua
3-2
l
3.1.2CONTROL KEYS
Standby Key
Down Key
Enter Key
Ramp/Silence Key
3.1.3INDICATORS
Up Key
The Synchrony control panel has five membrane-type switches that perform
different functions depending on the screen selected, alarm status, and
access level:
Starts and stops the Synchrony. The key must be pressed and released.
Navigates between screens or increases the value of a selected parameter by one
increment.
Navigates between screens or decreases the value of a selected parameter by
one increment.
Switches the function of the Up and Down keys between screen navigation and
parameter adjustment, and implements changes.
Activates ventilation ramp and silences the audible alarm signal.
The Synchrony uses four LED indicators on the control panel for alarm
indications and power indications. It also uses five indicators on the display
screen.
AC Power Indicator
DC Power Indicator
Green LED that illuminates when the Synchrony is connected to AC power.
See Figure 3-2.
Green
Figure 3-2 AC Power Indicator
Green LED that illuminates when the Synchrony is connected to DC power.
See Figure 3-3.
Green
Figure 3-3 DC Power Indicator
R
Provider Manua
3-3
l
Red Alarm Indicator
Yellow Alarm Indicator
Attention or Activity Indicator
( )
• When flashing indicates a new high priority alarm has occurred.
• When on continuously indicates a loss of power or silenced high priority
alarm.
• When flashing indicates a new medium priority alarm.
• When on continuously indicates a low priority alarm or silenced medium
priority alarm.
Red
Yellow
Figure 3-4 Alarm Indicators
Used in several screens to indicate pressure level, machine-triggered breaths, or
alarm messages:
• When appears next to EPAP or IPAP on the display, it
indicates the current pressure phase. See Figure 3-5.
• When appears next to an alarm message, it indicates that the
alarm is active. See Figure 3-6.
• When appears next to RR on the display, it indicates that the
current pressure phase was device-triggered. See Figure 3-7.
MODE: S 1/2
IPAP 15 cm H2O
EPAP 9 cm H2O
RR 20 BPM
Figure 3-5 EPAP Indicator
ALARM: 1/1
Patient Disconnect
Figure 3-6 Alarm Indicator
MODE: S/T 1/4
IPAP 15 cm H2O
EPAP 9 cm H2O
RR 20 BPM
Figure 3-7 Device-Triggered Breath
Indicator
R
Provider Manua
3-4
l
Selection Indicator ( )
Navigation Indicator ( )
appears next to the item that can be selected with the ENTER key, as shown
in Figure 3-8. The indicator is moved by pressing the UP or DOWN key.
S PARAMETERS: 1/2
IPAP 15 cm H2O
EPAP 9 cm H2O
Rise Time 3
Figure 3-8 Selection Indicator; Pressing
the ENTER Key will Select EPAP for
Adjustment
appears next to a page number display to indicate that the UP and DOWN
keys can be used to change pages, as shown in Figure 3-9. If it appears next to a
value, it flashes and indicates that the UP and DOWN keys can be used to
adjust the value, as shown in Figure 3-10.
MODE: S 1/2
IPAP 15 cm H2O
EPAP 9 cm H2O
RR 20 BPM
Ventilation Ramp Indicator ( )
Figure 3-9 Page Indicator
S PARAMETERS: 1/2
IPAP 15 cm H2O
EPAP 9 cm H2O
Rise Time 3
Figure 3-10 Value Adjust Indicator; IPAP
can be adjusted with the UP or DOWN key
appears at the top of the display when ventilation ramp has been initiated.
See Figure 3-11.
MODE: S 1/2
IPAP 15 cm H2O
EPAP 9 cm H2O
RR 20 BPM
Figure 3-11 Ventilation Ramp Indicator
R
Provider Manua
3-5
l
Progress Indicator ( )
Alarm Enabled Indicator (A)
The consists of seven segments that empty from top to bottom to indicate the
progress of certain operations. See Figure 3-12. The four indicators work
simultaneously.
SYSTEM SELF TEST
VER:X.XX XXXXX.X HRS
XXX XXX XXX XXX XXX
XXX XXX XXX XXX XXX
Figure 3-12 Progress Indicators
The A appears in the Monitoring screen if any of the patient alarms (Patient
Disconnect, Apnea, or Low Minute Ventilation) is enabled (see Chapter 8).
Indicates a Patient
Alarm is Enabled
MODE: S/T 1/4
IPAP 15 cm H2O
EPAP 9 cm H2O
RR 20 BPM
A
3.2 Navigating the
Screens
Figure 3-13 Monitoring Screen with
Alarm Enabled Indicator
When the screen indicates that there are multiple pages (1/3, 2/4, etc. appear in
the upper right corner), you can scroll through the pages by pressing either the
DOWN arrow key when the last item on the screen is selected, or by pressing
the UP arrow key when the first item is selected.
Total Number
Page
Indicator
NEW MODE: S/T 2/3
Timed Insp 1.0 sec
Rise Time 3
Figure 3-14 Example of Page Indicator on Screen
of Pages
In the example of Figure 3-14, if you press the UP key, the Synchrony displays
page 1 of 3; if you press the DOWN key twice, the Synchrony displays page 3
of 3.
R
Provider Manua
3-6
l
3.3 Display Data
NOTE: The Estimated Exhaled Tidal
Volume and the Estimated Exhaled
Minute Ventilation displays represent
estimates. There are some circumstances, such as low tidal volumes,
erratic breathing, or rapidly changing
leak where the accuracy of the
estimates will vary greatly. The
Estimated Exhaled Tidal Volume and
the Estimated Exhaled Minute
Ventilation displays will flash during
some transient conditions.
The display screen is used to display measured parameters in the Monitoring
screens. The data changes depending on the access level and the mode of
operation.
The Synchrony is capable of displaying:
•CPAP, IPAP, and EPAP - These displays are estimates of measure-
ments at the patient connection. Displays are updated as follows:
CPAP is updated at the end of inspiration; IPAP is updated at the end
of IPAP; and EPAP is updated at the end of EPAP.
•RR (Respiratory Rate) - The breathing rate is a running average from
the six previous breaths. If the Synchrony is operating in a mode that
supports machine-triggered breaths, this display will be the total
breathing rate (spontaneous + machine). The display value is updated
at the end of each expiration.
•VTE (Estimated Exhaled Tidal Volume) - Displays an estimated
exhaled tidal volume measurement for the previous breath. The display
is updated at the end of each breath. The estimated exhaled tidal
volume is obtained by integration of patient flow.
•MinVent (Estimated Exhaled Minute Ventilation) - Displays the
estimated exhaled minute ventilation based on an average of the
previous six breaths. The display is updated at the end of each breath.
•Leak - Displays the average leak value calculated during the previous
breath cycle. The display value is updated after each breath.
Table 3-1 summarizes the displayed parameters.
ParameterModeUnits
CPAP
IPAP
EPAP
RR
VT
E
MinVent
Leak
Different display screens are available based on the access level: