Working with other optional accessories . . . . . . . . . . . . . . . . . . . . . . . 7
Attaching the pulse oximeter 7
Adding supplemental oxygen 7
Starting therapy using oxygen 8
Stopping therapy using oxygen 8
Using the FiO2 monitoring sensor 8
Attaching an antibacterial filter 9
Performing a functional test 11
Starting therapy 12
Stopping therapy 12
Turning off the power 13
Working with alarms 13
Tailoring treatment setup options 14
Using mask-fit 14
Read the entire manual before using the device. This User Guide is for a non-clinician user,
and does not contain all the information provided in the Clinical Guide.
Indications for use
The Stellar 100/150 is intended to provide ventilation for non-dependent, spontaneously
breathing adult and paediatric patients (13 kg and above) with respiratory insufficiency, or
respiratory failure, with or without obstructive sleep apnoea. The device is for noninvasive
use, or invasive use (with the use of the ResMed Leak Valve). Operation of the device
includes both stationary, such as in hospital or home, or mobile, such as wheelchair usage.
Contraindications
The Stellar is contraindicated in patients who are unable to endure more than brief
interruptions in ventilation. The Stellar is not a life support ventilator.
If you have any of the following conditio
• pneumothorax or pneumomediastinum
• pathologically low blood pressure, particularly if associated w
depletion
• cerebrospinal fluid leak, recen
• severe bullous lung disease
• dehydration.
The use of the Stellar or pulse
environment.
oximetry (including XPOD) is contraindicated in an MRI
Adverse effects
You should report unusual chest pain, severe headache or increased breathlessness to
your prescribing physician.
The following side effects may arise during the cou
device:
• drying of the nose, mouth or throat
• nosebleed
• bloating
• ear or sinus discomfort
• eye irritation
• skin rashes.
ns, tell your doctor before using this device:
ith intravascular volume
t cranial surgery or trauma
rse of noninvasive ventilation with the
English
1Introduction
Stellar at a glance
Handle
Infrared connection
for humidifier
Air outlet
H4i connector
plug and AC
connection
Oxygen inlet
Air filter cover
XPOD pulse oximeter connection
FiO2 sensor connection
Data port for USB stick
Data port for direct PC connection
DC power socket
AC power socket
Power on/standby switch
Remote alarm connection
The Stellar comprises:
• Stellar device • Hypoallergenic air filter • AC power cord • Carry bag
ResMed USB stick • Low pressure oxygen connector.
The Stellar should only be used with air tubing and accessories recommended by
ResMed. Connection of other air tubing or accessories could result in injury or
damage to the device.
ResMed regularly releases new products. Please check the catalogu
accessories on our website at www.resmed.com.
Both masks and tracheostomy tubes can be used with Stellar. To set the patient interface
type go to the Setup menu, select Clinical Settings, then Advanced Settings.
For information on using masks, see
for this device, see the Mask/Device Compatibility List on www.resmed.com on the
Products page under Service & Support. If you do not have internet access, please
contact your ResMed representative.
the mask manual. For a full list of compatible masks
Humidification
A humidifier is recommended especially for patients experiencing dryness of the nose,
throat or mouth. For information on using a humidifier as part of:
• noninvasive ventilation, see “Setting up for noninvasive use” on page 4.
• invasive ventilation, see “Setting up for invasive use” on page 5.
Internal battery
CAUTION
Over time, the capacity of the internal battery will decrease. This depends on
individual use and ambient conditions. As the battery degrades, it could lead to
battery related alarms and or system errors. ResMed recommends the battery be
tested after two years to assess the remaining battery life. Testing and replacement
of the internal battery should only be performed by an authorised service agent.
Note: T
he battery duration depends on the state of charge, the environmental conditions,
the condition and age of battery, the device settings and the patient circuit configuration.
In case of a mains power disruption, the device will op
there is no external battery connected to the device. The internal battery will operate for
approximately two hours under normal conditions (see “Technical specificatio
page 27). The power status of the battery is displayed on to
battery status regularly while ope
device in time to mains power or alternatively to the external battery.
Additionally the Internal battery use alarm
to clear the alarm.
To recharge the internal battery, connect the device to main
hours to fully recharge the internal battery, however this can vary depending on
environmental conditions and if the device is in use.
rating the device with the internal battery and connect the
will be displayed. Press the Alarm mute button
erate using the internal battery if
ns” on
p of the LCD screen. Check the
s power. It can take up to three
Maintenance
The internal battery has to be discharged and recharged every six months.
1Remove the power cord while the Stellar is turned on (standy-by or in operation) and let
the device operate with the internal battery to a charge level of 50%.
2Reconnect the power cord to the mains power. The internal battery will be fully recharged.
Storing
If the device is stored for a longer period the internal battery should be at the charge level
of approximately 50% to increase the durability.
Note: Chec
to a charge level of 50%.
k the charge level every six months, if necessary recharge the internal battery
ResMed USB stick
A ResMed USB stick may be used with the device either to help the clinician to monitor
your treatment or to provide you with updated device settings. For more information, see
“Data management” on page 19.
Use on an aircraft
ResMed confirms that the Stellar can be used during all phases of air travel without further
testing or approval by the airline operator. See “Technical specifications” on page 27.
English
3Stellar at a glance
Mobile use
1
2
3
4
AC locking clip
AC power cord
5
The Stellar Mobility bag allows the Stellar to be used in mobile situations, eg, in a
wheelchair. For setup and correct use, see the Stellar Mobility Bag User Guide. For
extended mobile use, the ResMed Power Station II external power supply unit can be used
as an additional power source. Limitations apply to the use of oxygen with the Stellar
Mobility Bag. For more information, contact your local ResMed representative.
Remote Alarm
The Remote Alarm may be used to transfer audible and visual alarms by a direct cable
connection. For more information on using the Remote Alarm, see the Remote Alarm user
guide.
Setting up for noninvasive use
WARNING
• The air filter cover protects the device in the event of accidental liquid spillage
onto the device. Ensure that the air filter and air filter cover are fitted at all times.
• Make sure that all the air inlets at the rear of th
vents at the mask are unobstructed. If you put the device on the floor, make sure
the area is free from dust and clear of bedding, clothes or other objects that could
block the air inlets.
• Ho
ses or tubes must be non-conductive and antistatic.
not leave long lengths of the air tubing or the cable for the finger pulse sensor
• Do
around the top of your bed. It could twist around your head or neck while you are
sleeping.
CAUTION
• Be careful not to place the device where it can be bumped or where someone is
likely to trip over the power cord.
• Make sure the area around the device is dry and clean.
Notes:
• ResMed recommends using the AC power cord supplied with
power cord is required, contact your ResMed Service Centre.
• Place the device on a flat surface near th
e head of the bed.
e device and under the device and
the unit. If a replacement
1Connect the power cord.
2 Plug the free end of the power cord into a power outlet.
3 Attach the H4i heated humidifier to the front of the Stellar.
If the H4i is not in use, and if applicable,
outlet of the device (see “Attaching an antibacterial filter” on page 9).
4
connect the antibacterial filter firmly onto the air
Note: Other external humidifiers can be used for noninvasive setup. ResMed
recommends using a breathing system (including antibacterial filter, air tubings, external
humidifier) with an impedance of maximum 2 cm H
and 16 cm H
O at 120 L/min.
2
O at 30 L/min, 5 cm H2O at 60 L/min
2
4Connect one end of the air tubing firmly onto the air outlet of the H4i.
5Connect the mask system to the free end of the air tubing.
6Select the mask type (select Setup menu, then Clinical Settings, then Advanced Settings).
7Perform Learn Circuit (see “Setup menu: Options” on page 16).
Attaching the H4i heated humidifier for noninvasive use
Use of the H4i heated humidifier may be recommended by your clinician. For information
on using the H4i, see the H4i user guide.
WARNING
• Always place the H4i on a level surface below the level of the patient to prevent
the mask and tubing from filling with water.
• M
ake sure that the water chamber is empty and thoroughly dried before
transporting the humidifier.
e H4i heated humidifier is not intended for mobile use.
• Th
• Do
not overfill the water chamber, as during use this will cause water to spill into
the air circuit.
• For optimal accuracy and synchrony, perform Learn Circuit with a change of the
ircuit configuration, in particular when adding or removing high impedance
c
components (eg, antibacterial filter, external humidifier, water trap, nasal pillow
type mask or air tubing). See “Setup menu: Options” on page 16.
CAUTION
Check the air circuit for water condensation. Use a water trap or a tubing wrap if
humidification is causing water condensation within the tube.
Notes:
A humidifier increases resistance in the air circuit and may affect triggering and cycling, and
•
accuracy of display and delivered pressures. Therefore perform the Learn Circuit function
(see “Setup menu: Options” on page 16). The device adjusts the airflow resistance.
• The heating feature of the H4i is disabled when the device is not mains powered.
English
Setting up for invasive use
The Stellar can be used invasively only with the ResMed Leak Valve.
WARNING
• Make sure that all the air inlets at the rear of the device and under the device and
vents at the mask or at the ResMed Leak Valve are unobstructed. If you put the
device on the floor, make sure the area is free from dust and clear of bedding,
clothes or other objects that could block the air inlets.
• When using a Heated Moisture Exchange Filter (HMEF), replace the HMEF
regu
larly as specified in the instructions provided with the HMEF.
• The H4i is contraindicated for invasive use. An external humidifier approved for
in
vasive use is recommended according to EN ISO 8185 with an absolute
humidity of > 33 mg/L.
• For optimal accuracy and synchrony, perform Learn Circuit with a change of the
c
ircuit configuration, in particular when adding or removing high impedance
components (eg, antibacterial filter, external humidifier, water trap, nasal pillow
type mask or air tubing). See “Setup menu: Options” on page 16.
5Setting up for invasive use
CAUTION
Antibacterial filter
External humidifier*
Air tubing
ResMed Leak Valve
Catheter mount**
1
2
3
4
5
HMEF*
6
10
AC locking clip
AC power cord
Learn Circuit
configuration
When using a humidifier check the breathing system regularly for accumulated water.
Note: When the mask type is set to Tr a c h , the Non-Vented Mask alarm will be
automatically enabled to alert you when the ResMed Leak Valve has no vent holes or when
the vent holes are blocked.
1Connect the power cord.
2Plug the free end of the power cord into a power outlet.
3 Connect the antibacterial filter firmly onto the air outlet of the device.
4
Connect the external humidifier to the other side of the antibacterial filter.
*If not using an external humidifier, the HMEF can be connected to the ResMed Leak Valve
(at step 9).
5Connect the air tubing to the external humidifier.
6Connect the ResMed Leak Valve to the air tubing. Position the ResMed Leak Valve so that
air from the vent holes does not blow directly onto the patient's chest.
WARNING
Always set up the ResMed Leak Valve in the breathing circuit with the arrows and the
symbol p
7Select the mask type Tr a c h (select Setup menu, then Clinical Settings, then Advanced
Settings).
ointing in the direction of air flow from the Stellar to the patient.
8Perform Learn Circuit (see “Setup menu: Options” on page 16).
9If an external humidifier is not being used, you can connect the HMEF to the patient side
of the ResMed Leak Valve.
10 Connect the catheter mount.
** The ResMed Leak Valve or the HMEF can be con
interfaces including connector pieces like catheter mounts.
The catheter mount and the external humidifier are not part o
nected to standardised tracheostomy
f the ResMed component.
6
Working with other optional accessories
1
2
Attaching the pulse oximeter
Use of a pulse oximeter may be recommended by your clinician.
Contraindication
The pulse oximeter does not meet defibrillation-proof requirement per IEC 60601-1: 1990,
clause 17.h.
1Connect the plug of the finger pulse sensor to the plug of the pulse oximeter.
2Connect the plug of the pulse oximeter at the rear of the device.
To view the oximetry values, from the Monitoring menu, select Monitoring.
Adding supplemental oxygen
Oxygen may be prescribed by your clinician.
Note: Up to 30 L/min at maximum oxygen pressure of 50 mbar (0.73 psi) can be added.
WARNING
• Oxygen flow must be turned off when the device is not operating, so that unused
oxygen does not accumulate within the device and create a risk of fire.
• Use o
• R
• O
• Th
•
nly certified, clean oxygen sources.
esMed strongly recommends adding oxygen into Stellar’s oxygen inlet at the
rear of the device. Entraining oxygen elsewhere, ie into the breathing system via
a side port or at the mask, has potential to impair triggering and accuracy of
therapy/monitoring and alarms (eg, High Leak alarm, Non-vented mask alarm). If
used in this way, therapy and alarm operation must be verified each time oxygen
flow is adjusted.
xygen supports combustion. Oxygen must not be used while smoking or in the
presence of an open flame. Only use oxygen supply in well-ventilated rooms.
e breathing system and the oxygen source must be kept at a minimum
distance of 2 m away from any sources of ignition (eg, electrical devices).
Oxygen must not be used while the device is being operated within the mobility bag.
English
7Working with other optional accessories
Starting therapy using oxygen
1
2
Oxygen
supply
Press spring to release
connection after use
1
2
1
2
3
4
5
1Fit the oxygen connector to the oxygen inlet of the device.
2Attach the other end of the oxygen supply tubing to the oxygen supply.
3 Press to start treatment.
4Turn on oxygen.
Stopping therapy using oxygen
1Turn off oxygen.
2Press to stop treatment.
Using the FiO2 monitoring sensor
Use of the FiO2 monitoring sensor may be recommended by your clinician.
CAUTION
Do not use the FiO2 monitoring sensor with the H4i humidifier.
Preparing to use a new sensor
1Before use, leave the FiO
2Attach a new FiO
2
3Perform the sensor calibration (see “Setup menu” on page 16).
he FiO2 monitoring sensor must be replaced every 12 months.
Note: T
monitoring sensor open to the air for 15 minutes.
2
monitoring sensor (as shown below).
8
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.