ResMed S8 Elite System User Manual

1
S8 Elite™ System
ILLUSTRATIONS
ILLUSTRATIONEN / ILLUSTRATIONS
ILLUSTRAZIONI / FIGURAS
Component of 338112 /1
A
Keypad / Tastenfeld / Clavier / Tastierino / Teclado / Teclado / Toetsenpaneel
Carry handle / Tragegriff /
Poignée de transport / Maniglia
da trasporto / Asa para el
transporte / Pega de transporte /
Handvat
Air filter cover / Luftfilterabdeckung / Couvercle du logement du filtre à air / Coperchio del filtro dell'aria / Cubierta del filtro de aire / Tampa do filtro de ar / Afdekkap luchtfilter
Module / Modul / Module / Modulo /
Módulo / Módulo / Module
DC input / DC-Eingang / Entrée CC /
Ingresso CC / Entrada de CC /
Entrada CC / Gelijkstroom-ingang
AC input / AC-Eingang / Entrée CA / Ingresso CA / Entrada de CA / Entrada CA / Wisselstroom-ingang
LCD screen / LCD-Display /
Ecran LCD / Schermo a cristalli
liquidi / Visor / Mostrador
LCD / LCD-scherm
Air outlet / Luftauslassöffnung / Sortie d'air / Presa di uscita dell'aria / Salida de aire / Saída de ar / Luchtuitlaat
SYSTEM COMPONENTS / SYSTEMKOMPONENTEN / COMPOSANTS DU SYSTEME / COMPONENTI DEL SISTEMA / COMPONENTES DEL SISTEMA / COMPONENTES DO SISTEMA / SYSTEEMCOMPONENTEN
Air inlet / Lufteinlassöffnung / Entrée d'air / Presa di ingresso dell'aria / Entrada de aire / Entrada de ar / Luchtinlaat
2 m
2
34 5
OTHER ACCESSORIES / SONSTIGES ZUBEHÖR / AUTRES ACCESSOIRES / ALTRI ACCESSORI / OTROS ACCESORIOS / OUTROS ACESSÓRIOS / ANDERE ACCESSOIRES
C
1
5
D
2
3
4
B
INTEGRATED HUMIDIFIER / INTEGRIERTER ATEMLUFTBEFEUCHTER / HUMIDIFICATEUR INTEGRE / UMIDIFICATORE INTEGRATO / HUMIDIFICADOR INTEGRADO / HUMIDIFICADOR INTEGRADO / GEÏNTEGREERDE BEVOCHTIGER
32
3 m
1
4
© 2005 ResMed Ltd.
© 2005 ResMed Ltd.
© 2005 ResMed Ltd.
© 2005 ResMed Ltd.
HumidAire 3i
Note: Some products not available in all regions. Hinweis: Einige Produkte sind nicht in allen Regionen erhältlich. Remarque : certains produits ne sont pas disponibles dans tous les pays. Nota:Alcuni prodotti non sono disponibili in tutte le regioni. Nota:Algunos productos no están disponibles en todas las regiones. Nota: Alguns produtos não se encontram disponíveis em todas as regiões. Opmerking: Sommige producten zijn niet in alle regio's verkrijgbaar.
SETTING UP THE S8™/ AUFSTELLEN DES S8™ / INSTALLATION DE LA S8
/ CONFIGURAZIONE DELL'S8™/ INSTALACIÓN
DEL S8
/ MONTAGEM DO S8™ / OPSTELLEN VAN DE S8
DATA CARD MODULE / DATENKARTENMODUL / MODULE DE CARTE DE DONNEES / MODULO PER DATA CARD / MÓDULO TARJETA DE DATOS / MÓDULO DE CARTÕES DE DADOS / DATA CARD-MODULE
G
12
© 2005 ResMed Ltd.
H
USING THE DATA CARD / GEBRAUCH DER DATENKARTE / UTILISATION DE LA CARTE DE DONNEES / USO DELLA DATA CARD / USO DE LA TARJETA DE DATOS / COMO USAR O CARTÃO DE DADOS / GEBRUIK VAN DE DATA CARD
1
3
2
© 2005 ResMed Ltd.
Note: Some products not available in all regions. Hinweis: Einige Produkte sind nicht in allen Regionen erhältlich. Remarque : certains produits ne sont pas disponibles dans tous les pays. Nota: Alcuni prodotti non sono disponibili in tutte le regioni. Nota: Algunos productos no están disponibles en todas las regiones. Nota: Alguns produtos não se encontram disponíveis em todas as regiões. Opmerking: Sommige producten zijn niet in alle regio's verkrijgbaar.
HumidAire 3i
E
HUMIDAIRE 3i AND S8 / HUMIDAIRE 3i UND S8 / HUMIDAIRE 3i ET S8 / HUMIDAIRE 3i E S8 / HUMIDAIRE 3i y EL S8 / HUMIDAIRE 3i E S8 / HUMIDAIRE 3i EN S8
© 2005 ResMed Ltd.
BLANK MODULE / MODULSCHACHTABDECKUNG / MODULE VIERGE / FALSO MODULO / MÓDULO EN BLANCO / MÓDULO EM BRANCO / LEGE MODULE
F
1
2
© 2005 ResMed Ltd.
Data communications port / Datenkommunikations­anschluss / Port de communication de données / Porta dati / Puerto de comunicación de datos / Dados comunicações porta / Data communicatie poort
S8 ELITE™ SYSTEM
USERS MANUAL 1
EDIENUNGSANLEITUNG 43
B
M
ANUEL UTILISATEUR 89
M
ANUALE DELLUTENTE 135
ENGLISHDEUTSCHITALIANO FRANÇAISESPAÑOLPORTUGUÊS
MANUAL DEL USUARIO 179
ANUAL DO UTILIZADOR 223
M
ANDLEIDING VOOR DE GEBRUIKER 267
H
NEDERLANDS
Manufactured by:
ResMed Ltd 1 Elizabeth MacArthur Drive Bella Vista NSW 2153 Australia
Tel: +61 (2) 8884 1000 or 1 800 658 189 (toll free) Fax: +61 (2) 8883 3114 Email: reception@resmed.com.au
ResMed Offices:
ResMed Corp 14040 Danielson Street Poway CA 92064-6857 USA
Tel: +1 (858) 746-2400 or 1-800-424-0737 (toll free) Fax: +1 (858) 746-2900 Email: reception@resmed.com ResMed (UK) Ltd 65 Milton Park Abingdon Oxfordshire OX14 4RX UK
Tel: +44 (1235) 862 997 Fax: +44 (1235) 831 336 Email: reception@resmed.co.uk ResMed Asia Pacific Ltd 97 Waterloo Road North Ryde NSW 2113 Australia
Tel: +61 (2) 9886 5000 or 1 800 991 900 (toll free) Fax: +61 (2) 9889 1471 Email: reception@resmed.com.au ResMed BV Nieuwe Parklaan 86, 2587 BV Den Haag Nederland Tél: +31 (70) 358 6263 Fax: +31 (70) 358 4333 Email: receptiondh@resmed.nl ResMed Finland Niittykatu 6 FIN 02200 Espoo Suomi Puh: +358 9 8676820 Faksi: +358 9 86768222 Sähköposti: reception@resmed.fi ResMed Gmb H & Co. KG Fraunhoferstraße 16 82152 Martinsried Deutschland Tel: +49 (0) 89 / 99 01 - 00 or +49 (0) 800 2 777 000 (gebührenfrei) Fax: +49 (0) 89 / 99 01 - 10 55 Email: reception@resmed.de
ResMed Hong Kong Ltd Room 1714 Miramar Tower 132-134 Nathan Road Tsim Sha Tsui Hong Kong Tel: +852 2366 0707 Fax: +852 2366 4546 Email: reception@resmed.com.hk
ResMed Japan Nihonbashi Hisamatsu Bldg. 4F 2-28-1 Nihonbashi-Hamacho Chuo-Ku, Tokyo 103-0007 Japan Tel: +81 (3) 3662 5056 Fax: +81 (3) 3662 5040 ResMed Malaysia Sdn Bhd Suite E-10-20, Plaza Mon’t Kiara No. 2, Jalan 1/70C Mon’t Kiara 50480 Kuala Lumpur
Malaysia Tel: +60 3 6201 7177 Fax: +60 3 6201 2177 Email: reception@resmed.com.my ResMed NZ Ltd PO Box 51-048 Pakuranga Auckland New Zealand
Tel: +0800 737 633 (NZ toll free) or +64 274 737 633 Fax: +0800 737 634 (NZ toll free) or +64 9 239 0193 Email: reception@resmed.co.nz
ResMed SA Parc de la Bandonnière 2 rue Maurice Audibert 69800 Saint-Priest France Tel: +33 (0) 4 37 251 251 Fax: +33 (0) 4 37 251 260 Email: reception@resmed.fr
ResMed Singapore Pte Ltd 238A Thomson Road # 12-03/04 Novena Square Tower A Singapore 307684 Tel: +65 6284 7177 Fax: +65 6284 7787 Email: reception@resmed.com.sg
ResMed Spain SL C/ Arturo Soria, 245 28033 Madrid España Tel: +34 (93) 590 8154 Fax: +34 (93) 590 8153 Email: angelo@resmed.es
ResMed Sweden AB Industrigatan 2 S-461 37 Trollhättan Sverige Tel: +46 520 420 110 Fax: +46 520 397 15 Email: reception@resmed.se
Labhardt AG Thannerstrasse 57 CH-4054 Basel Schweiz Tel: +41 (061) 307 9711 Fax: +41 (061) 307 9722 Email: info@labhardt.ch
US DESIGNATED AGENT: ResMed Corp EU AUTHORISED REPRESENTATIVE: ResMed (UK) Ltd
www.resmed.com
Protected by patents: AU 697652, EP 0661071, US 4944310, US 5199424, US 5522382, US 6240921, US 6705315. Other patents pending.
Design registrations pending.
S8, S8 Elite, Activa, HumidAire, HumidAire 3i, Mirage, Mirage Swift, Mirage Vista, ResScan and Ultra Mirage are trademarks of ResMed Ltd. Papillon is a trademark of MAP Medizin-Technologie GmbH.
© ResMed Ltd 2005.
S8 Elite™ System
USER’S MANUAL
English
ENGLISH
1
2
CONTENTS
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
OUR RESPONSIBILITY 5
Y
MEDICAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
BOUT THE S8 ELITE 5
A C
ONTRAINDICATIONS 5
W
ARNINGS 6
C
AUTIONS 7
A
DVERSE EFFECTS 7
THE S8 ELITE SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
LITE 9
S8 E M
ASKS 9
H
UMIDIFIER 10
R
ESSCAN DATA CARD MODULE 10
A
CCESSORIES 10
HOW TO USE THE S8 ELITE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
ET UP THE S8 ELITE 11
S A
SSEMBLE THE MASK 11
S
TART TREATMENT 12
S
TOPPING TREATMENT 12
U
SING THE MASK-FITTING FEATURE 13
ENGLISH
HOW TO ATTACH A HUMIDIFIER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
TTACH A HUMIDAIRE 3I HUMIDIFIER 14
A
HOW TO USE THE KEYPAD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
OW TO CHANGE THE SETTINGS 16
H
HOW TO USE THE STANDARD MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
HOW TO USE THE DETAILED MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
ETTINGS MENU 20
S R
ESULTS MENU 21
O
PTIONS MENU 21
S
ERVICING MENU 21
M
ESSAGES ON THE S8 ELITE LCD 22
S8 MODULES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
SING A BLANK MODULE 23
U U
SING A RESSCAN DATA CARD MODULE 23
CLEANING AND MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
AILY CLEANING 25
D
CONTENTS
3
EEKLY CLEANING 25
W P
ERIODIC CLEANING 25
R
EPLACING THE AIR FILTER 26
S
ERVICING 26
FREQUENTLY ASKED QUESTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
SYSTEM SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
LITE (NO ACCESSORIES) 33
S8 E A
CCESSORIES 33
A
LL DEVICES (WHERE APPLICABLE) 34
LIMITED WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4
INTRODUCTION
Thank you for choosing the S8 ELITEflow generator. The S8 ELITE is a compact, all-in-one package with an inbuilt power supply and the option of integrated data management and humidification.
This user manual contains the information you need for the correct use of your
S8 ELITE.
YOUR RESPONSIBILITY
The owner or user of this system shall have sole responsibility and liability for any injury to persons or damage to property resulting from:
operation which is not in accordance with the operating instructions supplied
maintenance or modifications carried out unless in accordance with authorised
instructions and by authorised persons.
Please read this manual carefully before use.
This manual contains special terms and icons that appear in the margins to draw your attention to specific and important information.
Warning alerts you to possible injury.
Caution explains special measures for the safe and effective use of the device.
Note is an informative or helpful note.
MEDICAL INFORMATION
ENGLISH
ABOUT THE S8 ELITE
The S8 ELITE continuous positive airway pressure (CPAP) system is indicated for the treatment of obstructive sleep apnoea (OSA) in adult patients.
The
S8 ELITE CPAP system is intended for home and hospital use.
CONTRAINDICATIONS
The S8 ELITE is not a life support device and may stop operating during power failure or certain fault conditions. It should not be used by patients who are dependent on continuous therapy.
Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions:
severe bullous lung disease
pneumothorax
pathologically low blood pressure
dehydration
cerebrospinal fluid leak, recent cranial surgery or trauma.
INTRODUCTION
5
WARNINGS
The entire manual should be read before using the S8 ELITE.
Advice contained in this manual should not supersede instructions given by the
prescribing physician.
A patient should not connect a device to the data communication port unless
instructed to do so by their health care provider or physician. Only ResMed products are designed to be connected to the data communication port. Connecting other devices could result in injury, or damage to the
The S8 ELITE should only be used with masks (and connectors)
ResMed, or by a physician or respiratory therapist. A mask should not be used unless the
S8 ELITE device is turned on and operating properly. The vent hole or
holes associated with the mask should never be blocked. Explanation: The connectors)
S8 ELITE is intended to be used with special masks (or
*
which have vent holes to allow continuous flow of air out of the
mask.
When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask vent holes. However, when the device is not operating, insufficient fresh air will be provided through the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can, in some circumstances, lead to suffocation. This applies to most models of CPAP devices.
In the event of power failure or machine malfunction, remove the mask.
The S8 ELITE can be set to deliver pressures up to 20 cm H
of certain fault conditions, pressures up to 30 cm H
If oxygen is used with this device, the oxygen flow must be turned off when the
O are possible.
2
device is not operating.
Explanation: When the CPAP device is not in operation and the oxygen flow is left on, oxygen delivered into the air delivery tubing may accumulate within the CPAP machine enclosure and create a risk of fire. This applies to most types of CPAP machines.
Oxygen supports combustion. Oxygen should not be used while you are
smoking or in the presence of an open flame.
Always ensure airflow is being generated by the device before the oxygen
supply is turned on.
Always turn the oxygen supply off before stopping the airflow from the device.
Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will vary, depending on where the oxygen is introduced, the pressure settings, patient breathing pattern, mask selection and leak rate.
Do not use the S8 ELITE if there are obvious external defects or unexplained
changes in performance.
*
recommended by
O. In the unlikely event
2
S8 ELITE device.
* Ports may be incorporated into the mask or in connectors that are near the mask.
6
Do not open the S8 ELITE case. There are no user serviceable parts inside. Repairs
and internal servicing should only be performed by an authorised service agent.
Explosion hazard—do not use in the vicinity of flammable anesthetics.
CAUTIONS
At low pressures, the flow through the exhalation ports of your mask may not
clear all exhaled gas from the tubing. Some rebreathing may occur.
The airflow for breathing produced by this device can be as much as 6°C higher
than the temperature of the room. Caution should be exercised if the room temperature is warmer than 32°C.
When AC mains power (100–240V AC) is not available, always use a ResMed
DC-12 converter. (The DC-12 converter is available as an optional accessory.)
If supplied, the RESSCAN™ Data Card module is designed for use with specific
ResMed devices.
Only use the RESSCAN Data Card module (if supplied) as specified in this manual.
Note: The above are general warnings and cautions. Specific warnings, cautions and notes appear with the relevant instructions in the manual.
ADVERSE EFFECTS
Patients should report unusual chest pain, severe headache or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the S8 device:
drying of the nose, mouth or throat
nosebleed
bloating
ear or sinus discomfort
eye irritation
skin rashes
chest discomfort.
ENGLISH
MEDICAL INFORMATION
7
8
THE S8 ELITE SYSTEM
The S8 ELITE device is part of a system that consists of five elements:
S8 ELITE flow generator
Mask
Humidifier
Data management products
Accessories.
S8 ELITE
Please refer to the illustrations in section A of the illustration sheet.
FRONT VIEW (A-1)
Shows the keypad and LCD screen, and the air outlet.
REAR VIEW (A-2)
Shows where the module attaches, and the location of the air filter and power sockets.
POWER CORD (A-3)
Power cord to attach the S8 ELITE to the AC (mains) power.
CARRY BAG (A-4)
The S8 ELITE carry bag contains two pouches: one for storing the S8 device, the other for storing the integrated humidifier. When storing the humidifier, always put it in the larger, mesh pouch.
AIR TUBING (A-5)
The 2 m air tubing connects the
S8 ELITE to your mask.
ENGLISH
MASKS
You will also need a ResMed mask system (supplied separately). The ResMed mask systems that are recommended for use with the
NASAL MASKS
MODULAR NASAL MASK
MIRAGE VISTA™ NASAL MASK
ULTRA MIRAGE™ NASAL MASK
MIRAGE™ NASAL MASK
MIRAGE ACTIVA™ NASAL MASK
PAPILLON
NASAL PILLOWS SYSTEMS
MIRAGE SWIFT™ NASAL PILLOWS SYSTEM
S8 ELITE are:
THE S8 ELITE SYSTEM
9
ULL FACE MASKS
F
MIRAGE™ FULL FACE MASK
MIRAGE™ FULL FACE MASK SERIES 2
ULTRA MIRAGE™ FULL FACE MASK
Note: Not all masks are available in all regions.
WARNING
!
Only ResMed mask systems are compatible for use with the S8 Elite. Please refer to “Warnings” on page 6.
HUMIDIFIER
Please refer to the illustration in section B of the illustration sheet.
A humidifier may be required if you are experiencing dryness of the nose, throat or mouth. The
S8 ELITE is compatible for use with the HUMIDAIRE 3i integrated heated
humidifier (B-1).
RESSCAN DATA CARD MODULE
The RESSCAN Data Card module and RESSCAN Data Card (C-1) may be used with the
S8 ELITE either to help your clinician to monitor your treatment or to provide you
with updates to your device settings. See “Using a ResScan Data Card Module” on page 23 for information about how to use the Data Card module and Data Card.
ACCESSORIES
Please refer to the illustrations in section C of the illustration sheet.
Other accessories for use with the
Front cap (not available in all regions) (C-2)
Air tubing (3 m) (C-3)
DC-12 converter (C-4)
Blank module (C-5).
Note: ResMed regularly releases new products. Please check our website at <www.resmed.com>.
S8 ELITE:
10
HOW TO USE THE S8 ELITE
Please refer to the illustrations in section D of the illustration sheet.
SET UP THE S8 ELITE
1 Place the S8 ELITE on a table near the head of your bed.
CAUTION
!
Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord.
Note: You can also place your S8 Elite on the floor beside or underneath your bed. Ensure that the area is free from dust and clear of bedding, clothes or any other objects that could block the air inlet.
2 Connect the power cord to the socket at the rear of the flow generator (D-1). Plug
the other end of the power cord into the power outlet.
WARNING
!
Make sure the power cord and plug are in good condition and the equipment is not damaged.
3 Connect one end of the air tubing firmly onto the air outlet (D-2).
WARNING
!
Only ResMed air tubing should be used with your flow generator. A different
type of air tubing may alter the pressure you actually receive, reducing the effectiveness of your treatment.
Blocking the air inlet or tubing while in operation could lead to overheating of
the device.
In the clinical environment any personal computer that is used with your CPAP
system must be at least 1.5 m away from, or at least 2.5 m above, the patient. It must also comply with the relevant test standard. For personal computers the international standard is IEC 60950 or equivalent.
ENGLISH
ASSEMBLE THE MASK
1 Assemble your mask system according to the mask user instructions.
2 Connect your mask system to the free end of the air tubing (D-3).
3 Check that the mask setting on the MASK screen is the one recommended for the
mask you are using (see the table in “How to Select the Mask Type” on page 20). The
S8 ELITE is now ready for use (D-4).
HOW TO USE THE S8 ELITE
11
START TREATMENT
1 Make sure the power is on.
A welcome message is displayed briefly on the LCD screen, then the standby (Ramp) screen appears. The key and LCD backlights also turn on.
2 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep.
CAUTION
!
Do not leave long lengths of air tubing around the top of your bed. They could
twist around your head or neck while you are sleeping.
Make sure the area around the flow generator is dry and clean. It should also be
clear of bedding, clothes and other potential blockages.
3 Alter the ramp time if required (see “How to Change the Ramp Time” on page 16).
To start treatment, press the Start/Stop key, unless your clinician has set up your device to start delivering therapy automatically. If so, your device will start when you breathe into your mask (see “SmartStart” for details).
4 Fit your mask as described in the mask user instructions.
WARNING
!
A mask should not be used unless the S8 Elite device is powered and operating properly.
5 When ramping begins, the LCD will display the word "RAMPING", with the current
pressure indicated as dashes. Each dash is approximately equivalent to 2 cm H example, 6 dashes represents approximately 12 cm H
O pressure.
2
SMARTSTART
The S8 ELITE has a function called SmartStart that your clinician can enable. If SmartStart is enabled, your device will start automatically when you breathe into your mask, and will stop automatically when you take your mask off. This means you do not have to press the Start/Stop key to begin or end treatment.
Note: Do not use SmartStart when using a mask with an anti-asphyxia valve, such as the Mirage Full Face Mask.
O. For
2
12
STOPPING TREATMENT
To stop treatment at any time, remove your mask and press the Start/Stop key
or
if your clinician has enabled SmartStart, simply remove your mask and treatment will end.
USING THE MASK-FITTING FEATURE
You can use the S8 ELITE mask-fitting feature to help you fit your mask properly. This feature delivers air pressure for a three-minute period, prior to starting treatment, during which you can check and adjust your mask fit to minimise leaks.
Note: Make sure that the mask setting on the MASK screen is the same as the mask you are using.
To use the mask-fitting feature:
1 Fit your mask as described in the mask user instructions.
2 Hold down the Start/Stop key for at least three seconds until air pressure delivery
starts.
The following display will appear on the LCD screen indicating that the mask-fitting feature is operating. Your S8 device will ramp to the mask-fit pressure and will remain at this pressure for three minutes.
The LCD also displays a mask-fit star rating. See Table 1, “Definitions of Mask-Fit Star Rating,” on page 14.
ENGLISH
MASK FIT:
*****
mask-fit star rating
excelnt
Note: The mask-fit star rating display disappears after three minutes.
The mask-fitting feature can only be started from the RAMP screen.
The mask-fit pressure is the set treatment pressure or 10 cm H
3 Adjust your mask, mask cushion, and headgear until you have a secure and comfortable
fit.
Once you have a secure and comfortable fit, check your mask-fit star rating on the LCD screen. Table 1 explains the mask-fit star ratings.
Note: If there is another person nearby to check your mask-fit star rating, you can adjust your mask, mask cushion and headgear while lying down.
4 After three minutes, the pressure reverts to the set pressure and treatment will begin.
If you do not wish to wait three minutes, hold down the Start/Stop key for at
least three seconds to start treatment immediately.
If you press the Start/Stop key for less than three seconds, the unit will return
to standby mode (the RAMP screen is displayed).
0, whichever is greater.
2
HOW TO USE THE S8 ELITE
13
Table 1: Definitions of Mask-Fit Star Rating
Star rating Definition
***** Excellent
****- Very good
***-- Good
**--- Adjust mask
*---- Adjust mask
----- Adjust mask
HOW TO ATTACH A HUMIDIFIER
ATTACH A HUMIDAIRE 3i HUMIDIFIER
Please refer to the illustration in section E of the illustration sheet, and your HumidAire 3i manual.
The
HUMIDAIRE 3
humidification. No other accessories are required for its use
Notes:
a. Make sure the water chamber of the HumidAire 3i is empty before you attach or
detach the humidifier.
b. Remove the connector plug on the S8 Elite before you attach the HumidAire 3i.
Reattach the connector plug firmly after you detach the humidifier.
STORING THE HUMIDAIRE 3
Store the HUMIDAIRE 3 not in use.
i
attaches to the front of the S8 ELITE to provide heated
.
i
i
in the larger, mesh-top pouch of the S8 ELITE carry bag when
14
HOW TO USE THE KEYPAD
The control panel of the S8 ELITE includes an LCD screen and keypad for navigating through the menus and delivering treatment.
LCD screen
Up key
Left key
Down key
LCD SCREEN
The LCD screen displays the menus and treatment screens.
The LCD backlight turns off after two minutes of inactivity. It turns back on when you press a key.
KEYPAD KEYS
The S8 ELITE keypad has the following keys:
Right key
Start/Stop key
ENGLISH
Key Function
Start/Stop • Starts or stops treatment.
• Extended hold (for at least three seconds) starts the mask-fitting feature.
Up • Allows you to scroll through the S8 Elite menus and
Down • Allows you to scroll through the S8 Elite menus and
Left • Performs the function indicated by the guiding text displayed
Right • Performs the function indicated by the guiding text displayed
When the
S8 device is in standby mode all keys remain lit. During therapy, the keypad
submenus, and increase the setting options.
submenus, and decrease the setting options.
above it on the LCD screen. Guiding text includes menu, change and apply.
above it on the LCD screen. Guiding text includes exit and
.
cancel
backlight turns off after two minutes of inactivity and the Start/Stop key dims.
HOW TO USE THE KEYPAD
15
HOW TO CHANGE THE SETTINGS
You can change settings on the S8 ELITE by using the keypad and LCD. When the
S8 ELITE is in standby mode you can display a series of screens (a menu) on the LCD
screen in order to view and change the settings for a particular function. Your clinician will preset the menu to either a standard view or a detailed view. See “How to Use the Standard Menu” on page 17 and “How to Use the Detailed
Menu” on page 19 for more information.
HOW TO ENTER THE MENU SCREENS
When the S8 ELITE standby (RAMP) screen is displayed you will see the word ‘menu’ in the left-hand corner. To select an
HOW TO SCROLL THROUGH THE MENU
When you are in the menu, scroll through items by pressing the Up or Down keys.
HOW TO CHANGE A SETTING OPTION
When you have scrolled to the item you require:
1. Press the Left key (Change).
2. Press the Up or Down keys until you see the setting option that you require.
3. Press the Left key (Apply) to select the setting option.
If you wish to exit without changing the option, press the Right key (Cancel).
HOW TO EXIT OUT OF A MENU
Press the Right key (Exit). This will take you back to the standby (RAMP) screen.
HOW TO EXIT OUT OF AN ITEM WITHIN THE MENU
Press the Right key (Exit).
S8 ELITE menu, press the Left key.
16
HOW TO CHANGE THE RAMP TIME
During ramp time, the pressure increases from a low pressure to the prescribed treatment pressure. You can set the ramp time in 5-minute intervals, from 0 minutes (OFF) up to a maximum ramp time set by your clinician.
The standby screen displays the current ramp time in the right-hand corner. To change the ramp time, press the Up or Down key until the desired ramp time appears.
Once a ramp time is selected, begin therapy by pressing the Start/Stop key.
HOW TO USE THE STANDARD MENU
Figure 1 shows the standard menu.
WELCOME >>>>>>
RAMP: 10min menu
EPR SETTING: 1
change exit
USED HRS: 08000
exit
USAGE: 1000/1068
exit
MASK: ULTRA change exit
SW: SX3960200 exit
Figure 1 : Standard menu screen
You enter the standard menu from the standby (RAMP) screen, by pressing the Left (menu) key.
HOW TO CHANGE THE EXPIRATORY PRESSURE RELIEF SETTING
Note: EPR may not be available in all regions.
Your clinician may have set up your device to deliver expiratory pressure relief (EPR). This can be used to help ease any discomfort you may experience when breathing out during treatment. There are four settings:
OFF (no EPR)
1 (lowest EPR)
2
3 (highest EPR).
The EPR SETTING screen only appears if your clinician has enabled access to EPR settings. To change the EPR setting, select Change (Left key). Press the Up or Down
keys until you see the setting that you require.
HOW TO SELECT THE MASK TYPE
Scroll to MASK and select Change (Left key). Press the Up or Down keys until you see the mask type that you require.
ENGLISH
HOW TO USE THE STANDARD MENU
17
The following table shows the setting that should be selected for each mask type.
Table 2: Settings for mask types
Settings Mask
ULTRA Ultra Mirage Nasal Mask
MIR FULL Mirage Full Face Mask
ACTIVA Mirage Activa Nasal Mask
SWIFT Mirage Swift Nasal Pillows System
MIRAGE Mirage Nasal Mask
STANDARD Mirage Vista Nasal Mask
Mirage Full Face Mask Series 2 Ultra Mirage Full Face Mask
Modular Nasal Mask Papillon
OTHER VIEWABLE INFORMATION
You can also view the total number of hours of treatment (USED HRS); the number of sessions for which the
S8 ELITE was used in a usage period (USAGE); and the current
software version (SW).
18
HOW TO USE THE DETAILED MENU
Figure 2 shows the detailed menu.
WELCOME >>>>>>
RAMP: 10 min
menu
SETTINGS enter
RESULTS
enter exit
OPTIONS
enter exit
exit
EPR SETTING: 1 change
exit
MASK: ULTRA change
exit
TUBE LENGTH: 2m change
exit
HUMID: NONE change
exit
SMARTSTART: ON change
exit
LEAK ALERT: OFF change
exit
USED HRS: 08000
exit
USAGE: 1000/1068
exit
LANG: ENGLISH
change exit
ENGLISH
SERVICING
enter exit
SN: 12345678912
exit
PCB: 123456789123
4567 exit
SW: SX3960200
exit
Figure 2 : Detailed menu screens
You enter the detailed menu from the standby (RAMP) screen, by pressing the Left (menu) key. The SETTINGS menu screen is displayed.
HOW TO USE THE DETAILED MENU
19
SETTINGS MENU
Select Enter (Left key) on the Settings menu to access the options in this menu.
HOW TO CHANGE THE EXPIRATORY PRESSURE RELIEF SETTING
Note: EPR may not be available in all regions.
Your clinician may have set up your device to deliver expiratory pressure relief (EPR). This can be used to help ease any discomfort you may experience when breathing out during treatment. There are four settings: 1(lowest EPR), 2, 3 (highest EPR) and OFF (no EPR).
If you have access and want to change the EPR setting, scroll to EPR SETTING and select Change (Left key). Press the Up or Down keys until you see the setting that
you require.
HOW TO SELECT THE MASK TYPE
Scroll to MASK and select Change (Left key). Press the Up or Down keys until you see the mask type that you require.
The following table shows the setting that should be selected for each mask type.
Table 3: Settings for mask types
Settings Mask
ULTRA Ultra Mirage Nasal Mask
MIR FULL Mirage Full Face Mask
Mirage Full Face Mask Series 2 Ultra Mirage Full Face Mask
20
ACTIVA Mirage Activa Nasal Mask
SWIFT Mirage Swift Nasal Pillows System
MIRAGE Mirage Nasal Mask
STANDARD Mirage Vista Nasal Mask
Modular Nasal Mask Papillon
HOW TO CHANGE THE TUBE LENGTH
Scroll to TUBE LENGTH and select Change (Left key). Press the Up or Down keys until you see the tube length that you require.
HOW TO ADD A HUMIDIFIER OR FRONT CAP
Scroll to HUMID and select Change (Left key). Press the Up or Down keys until you see the setting that you require. You can use the device with or without a front cap.
HOW TO CHANGE THE SMARTSTART SETTING
SmartStart enables your device to start automatically when you breathe into the mask. Scroll to SMARTSTART and select Change (Left key). Press the Up or Down keys
to change the option to ON or OFF.
HOW TO CHANGE THE LEAK ALERT SETTING
Leak alert indicates when a high mask leak is present. Scroll to LEAK ALERT and select Change (Left key). Press the Up or Down keys to
change the option to ON or OFF.
RESULTS MENU
Select Enter (Left key) on the Results menu to access the options in this menu.
Note: Your clinician may have customised your device to display Smart Data. Contact your clinician for further details about Smart Data.
VIEWING USAGE INFORMATION
You can view the total number of hours of treatment (USED HRS) and the number of sessions for which the
S8 ELITE was used in a usage period (USAGE).
OPTIONS MENU
Select Enter (Left key) on the Options menu to access the options in this menu.
HOW TO CHANGE THE LANGUAGE SETTING
Scroll to LANG and select Change (Left key). Press the Up or Down keys until you see the language that you require.
ENGLISH
SERVICING MENU
Select Enter (Left key) on the Servicing menu to access the options in this menu. You can view the current software version installed on your unit (SW); the serial
number of the printed circuit board in your device (PCB); and the serial number of the device (SN).
HOW TO USE THE DETAILED MENU
21
MESSAGES ON THE S8 ELITE LCD
Your clinician may have set your S8 ELITE to remind you about important events, such as when to replace your mask, when to insert your Data Card (if your device is Data Card enabled), and so on.
The reminder message is displayed on the LCD, and is visible if the device is not delivering therapy. The backlight on the LCD flashes when a message is displayed.
Your clinician can set any of the following reminders on your LCD:
Message Description Action
INSERT CARD May appear if your
device is Data Card enabled.
REPLACE MASK
CALL PROVIDER
REPLACE FILTER
SERVICE DUE May appear as a
Customised messages
May appear when your mask is due for replacement.
May appear as a reminder to contact your clinician; for example to discuss how your therapy is going.
May appear as a reminder to replace the air filter on your device.
reminder to return your device for service.
Your clinician may also set reminders for other reasons; for example to return equipment or to call a particular person or number.
Insert your Data Card (see Section H of the illustrations sheet), and follow any instructions that your clinician has given you. When you have done this, the message will disappear from the LCD. Pressing the Left (Ok) key will also remove the message.
Press the Left (Ok) key to remove the message from your LCD, and replace your mask with a new one.
Press the Left (Ok) key to remove the message from your LCD, and contact your clinician.
Press the Left (Ok) key to remove the message from your LCD, and replace the air filter.
Press the Left (Ok) key to remove the message from your LCD.
Press the Left (Ok) key to remove the message from your LCD.
22
S8 MODULES
Note: Do not remove modules while power is connected to the S8 Elite.
Your
S8 ELITE will have one of two modules fitted to the back: a blank module or a
RESSCAN Data Card module.
USING A BLANK MODULE
Please refer to the illustrations in section F of the illustration sheet.
The blank module is a protective cover on your remove or attach the module, for example to use another module.
To remove the blank module, pull it off the back of the communications port (F-1).
To attach the blank module, push the blank module onto the data communications port, until the module clicks into place (F-2).
Note: To protect the data communications port, always ensure there is a module attached to the back of the device.
USING A RESSCAN DATA CARD MODULE
Please refer to the illustrations in section G of the illustration sheet.
The
RESSCAN Data Card module is an accessory to the S8 ELITE. Use a RESSCAN
Card with the Data Card module to collect data and update settings in the The module and card give your clinician convenient access to data about your treatment, and can provide you with new device settings from your clinician without your having to leave home.
S8 ELITE. At times, you may wish to
S8 ELITE to show the data
Data
S8 ELITE.
ENGLISH
ATTACHING THE MODULE
To attach the Data Card module, push it onto the data communications port at the back of your
S8 ELITE until it clicks into place (G-1).
REMOVING THE MODULE
To remove the Data Card module, pull it off the back of the S8 ELITE (G-2).
Note: To protect the data communications port, always ensure there is a module attached to the device.
USING THE DATA CARD TO COLLECT DATA
Please refer to the illustrations in section H of the illustration sheet.
If your clinician needs to review your treatment, they will ask you to use the Data Card to copy data from your be provided in a convenient mailback envelope.
Data that is copied to a Data Card is still stored and available on the It takes up to 30 seconds to copy data onto the Data Card.
S8 ELITE, and to return the card to them. The Data Card will
S8 ELITE.
1 INSERT THE DATA CARD
Switch on the S8 ELITE and wait until you see the standby (RAMP) screen.
S8 MODULES
23
Hold the Data Card with the arrow facing up, and insert it into the slot in the
Data Card module (H-1).
Push the card in smoothly until it stops moving (H-2) and messages start
appearing on the LCD.
2 COPY DATA ONTO THE DATA CARD
Data copying starts automatically when the Data Card is inserted into the Data Card module.
The “Card Inserted Please Wait” message is displayed on the S8 ELITE LCD
while data is being copied. Copying takes up to 30 seconds.
The “Copy Complete Remove Card” message is displayed on the LCD when
copying has finished.
Remove the Data Card from the Data Card module.
3 REMOVE THE DATA CARD
Grip the end of the Data Card and pull it out of the Data Card module (H-3).
4 STORE THE DATA CARD
Your Data Card should be stored in its mailback envelope when not in use.
5 MAILING THE DATA CARD TO YOUR CLINICIAN
A bubble mailer envelope is provided to ensure safe transportation of the Data Card. When mailing the Data Card to your clinician:
1. Insert the Data Card into the pocket provided on the mailback envelope. Fold up the mailback envelope.
2. Place the mailback envelope containing the Data Card in the bubble mailer envelope and seal it.
3. Make sure that your clinician’s name and address details are on the bubble mailer envelope, and mail it back immediately.
24
USING THE DATA CARD TO UPDATE SETTINGS
If your clinician has provided a Data Card with new device settings:
With the device in standby (Ramp) mode, insert the Data Card into the slot on
the Data Card module. Updating will start automatically.
The “Card Inserted Please Wait” message is displayed on the LCD while updating
is in progress. Updating takes up to 30 seconds.
The “Settings Success Remove Card” message is displayed on the LCD if the
settings were updated successfully.
Note: This message only appears once. If you re-insert the Data Card after you have updated your settings, this message is not displayed.
Remove the Data Card from the Data Card module.
Store the Data Card in its mailback envelope when not in use.
WARNING
!
If your clinician has told you to use this Data Card to update the settings on your device and the “Settings Success” message does not appear, contact your clinician immediately.
CLEANING AND MAINTENANCE
You should regularly carry out the cleaning and maintenance described in this section. Refer to your mask manual for detailed instructions.
DAILY CLEANING
MASK
Clean the mask according to the instructions supplied with the mask.
AIR TUBING
Disconnect the air tubing from the S8 ELITE (and humidifier, if used) and hang the tubing and mask in a clean, dry place until next use.
HUMIDIFIER
If you are using a humidifier, remove it from the S8 ELITE and clean it according to the instructions in the humidifier user’s manual.
WEEKLY CLEANING
1. Remove the air tubing from the S8 ELITE and the mask.
2. Wash the air tubing in warm water using mild detergent. Rinse thoroughly, hang, and allow to dry.
3. Before next use, assemble the mask and headgear according to the user instructions.
4. Reconnect the air tubing to the air outlet and mask.
PERIODIC CLEANING
1. Clean the exterior of the S8 ELITE with a damp cloth and mild liquid soap.
2. Check the air filter for blockage or holes (see “Replacing the Air Filter” on page 26).
ENGLISH
WARNING
!
Beware of electric shock. Do not immerse the unit or power cord in water.
Always unplug the unit before cleaning and be sure it is dry before plugging back in.
The mask system and air tubing are subject to normal wear and tear. Inspect
them regularly for damage.
CLEANING AND MAINTENANCE
25
REPLACING THE AIR FILTER
To replace the air filter:
1. Remove the air filter cover at the back of the
2. Remove and discard the old air filter.
3. Insert a new filter with the blue tinted side facing out from the device.
4. Replace the air filter cover.
Inspect the air filter every month to check if it is blocked by dirt or contains holes. With normal use of an often if your device is in a dusty environment).
S8 ELITE, the air filter needs to be replaced every six months (or more
S8 ELITE.
Air filter cover
WARNING
!
Do not wash the air filter. The air filter is not washable or reusable.
The air filter cover protects the device in the event of accidental liquid spillage
onto the device. Ensure the air filter and air filter cover are fitted at all times.
SERVICING
CAUTION
!
Do not attempt to open the S8 Elite case. There are no user serviceable parts
inside.
Inspection and repair should only be performed by an authorised agent. Under
no circumstances should you attempt to service or repair the flow generator yourself.
This product ( 5 years from the date of manufacture. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorised ResMed service centre.
If you feel that your flow generator is not performing properly, see “Troubleshooting” on page 29.
S8 ELITE) should be inspected by an authorised ResMed service centre