ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA
ResMed (UK) Ltd (EU Authorised Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK
ResMed Offices Australia, Austria, Belgium, Brazil, China, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Malaysia,
Mexico, Netherlands, New Zealand, Norway, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com
for contact details).
Protected by patents: AU 697652, AU 699726, AU 713679, CA 2235939, EP 0661071, EP 0858352, JP 3638613, JP 4083154, US 5199424, US
5522382, US 6213119, US 6240921, US 6705315. Other patents pending.
Protected by design registrations: AU 302180, AU 302181, AU 302182, AU 302183, AU 302184, AU 302185, AU 302186, CN 200430121211.8, CN
200430121212.2, CN 200430121213.7, CN 200430121214.1, CN 200430121215.6, CN 200430121216.0, CN 200430121217.5, EU 269436, HK
0412901.9, JP 1248040, JP 1257662, JP 1257663, JP 1266956, JP 1266957, JP 1267270, JP 1267271, NZ 405614, NZ 406923, NZ 406924, NZ
406925, NZ 406926, NZ 406927, NZ 406928, US D544598, US D557406, US D557407, US D560795, US D561891. Other designs pending
SmartStart, TiCONTROL, VPAP, and Vsync are trademarks of ResMed Ltd and SmartStart and VPAP are registered in U.S. Patent
and Trademark Office.
Using the VPAP IV and VPAP IV ST . . . . . . . . . . . . . . . . . . . . . . . . . 3
Setting up your VPAP 3
Attaching a H4i Humidifier 4
How to Use the Control Panel 4
Using the Menus 5
How to Select the Mask Type 6
SmartStart™ 6
Starting Treatment 6
Stopping Treatment 7
Using the Mask-Fitting feature 7
Reminders on the VPAP LCD 8
Smart Data™ 8
Using the Data Card 9
The VPAPTM IV and the VPAPTM IV ST are intended to provide non-invasive ventilation
for patients with respiratory insufficiency or obstructive sleep apnoea (OSA), in the
hospital or home.
Contraindications
Positive airway pressure therapy may be contraindicated in some patients with the
following pre-existing conditions:
• severe bullous lung disease
• pneumothorax or pneumomediastinum
• pathologically low blood pressure, particularly if associated with intravascular
volume depletion
• dehydration
• cerebrospinal fluid leak, recent cranial surgery or trauma.
Adverse Effects
Patients should report unusual chest pain, severe headache or increased
breathlessness to their prescribing physician. An acute upper respiratory tract
infection may require temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the device:
• drying of the nose, mouth or throat
• nosebleed
• bloating
• ear or sinus discomfort
• eye irritation
• skin rashes.
WARNING
Read the entire manual before using these VPAP devices.
English
The VPAP System
Both the VPAP IV and the VPAP IV ST systems comprise the following elements:
• VPAP device
• 2 m air tubing
• Power cord
• Travel bag
• ResScan™ Data Card
• ResMed Oxygen Connector Port.
Optional components include:
• 3 m air tubing
• DC/DC Converter 24V/50W.
1Introduction
Page 5
Masks
!
Screen
Keypad
Air outlet
DC
AC
Air inlet
Data Card
module
Carry handle
The following ResMed mask systems are recommended for use with these devices:
Mask TypeName
Nasal Masks• Mirage Vista™ Nasal Mask
Nasal Pillows Systems• Mirage Swift™ Nasal Pillows System
Full Face Masks• Mirage™ Liberty Full Face Mask
For information on using masks, see your mask manual. For the latest available
masks, see www.resmed.com.
• Ultra Mirage™ Nasal Mask
• Ultra Mirage™ II Nasal Mask
• Mirage Activa™ Nasal Mask
• Mirage Micro™ Nasal Mask
• Mirage Swift™ II Nasal Pillows System
• Mirage™ Quattro Full Face Mask
• Ultra Mirage™ Full Face Mask
Humidifier
If you are experiencing dryness of the nose, throat or mouth, the H4i heated
humidifier is recommended for use with these VPAP devices.
WARNING
• Only ResMed mask systems are compatible for use with these VPAP devices.
• Only the H4i is compatible for use with these VPAP devices.
ResScan Data Card
The ResScan Data Card may be used with these VPAP devices either to help your
clinician to monitor your treatment or to provide you with updates to your device
settings.
Supplemental Oxygen
The VPAP IV and VPAP IV ST are designed to be compatible with up to 15 L/min of
supplemental oxygen.
2
Page 6
At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration
!
!
will vary, depending on the pressure settings, patient breathing pattern, mask
selection, and the leak rate.
WARNING
Always use the ResMed Oxygen Connector Port when adding supplemental
oxygen at the flow generator outlet.
Using the VPAP IV and VPAP IV ST
Setting up your VPAP
1 Connect the power cord to the socket at the rear of your VPAP.
2 Plug the other end of the power cord into the power outlet.
3 Connect one end of the air tubing firmly onto the air outlet.
4 Connect the assembled mask system to the free end of air tubing.
For information on assembling your mask, see your mask manual.
2
1
English
3
4
WARNING
• Make sure the power cord and plug are in good condition and the
equipment is not damaged.
• Only ResMed air tubing should be used with the device. A different type of
air tubing may alter the pressure you actually receive, reducing the
effectiveness of your treatment.
• Blocking the hose and/or air inlet of the device while in operation could lead
to overheating of the device.
3Using the VPAP IV and VPAP IV ST
Page 7
!
CAUTION
!
LCD screen
Keypad
Start/Stop Key
• Be careful not to place the device where it can be bumped or where
someone is likely to trip over the power cord.
• If you put the device on the floor, make sure the area is free from dust and
clear of bedding, clothes or other objects that could block the air inlet.
Attaching a H4i Humidifier
The H4i humidifier attaches to the front of a VPAP IVor VPAP IV ST device to provide
heated humidification. These devices automatically detect the presence of the H4i
and no other accessories are required for its use. For more information on using your
H4i, please refer to the H4i user guide.
WARNING
• Make sure that the water chamber is empty and thoroughly dried before
transporting the H4i.
• When using the travel bag, always separate the VPAP unit and the H4i and
place the H4i in its pouch.
How to Use the Control Panel
4
The control panel of your VPAP device includes an LCD screen which displays the
menus and treatment screens as well as a keypad for navigating through the menus
and delivering treatment. The keypad has the following keys:
KeyFunction
Start/Stop
Up
Down
Left
Right
Starts or stops treatment. Extended hold for at least three seconds
starts the mask-fit feature .
Allows you to increase settings options and scroll through the
menu.
Allows you to decrease settings options and scroll through the
menu.
Performs the function indicated by the guiding text displayed above
it on the LCD screen. Guiding text includes menu, change, and
apply.
Performs the function indicated by the guiding text displayed above
it on the LCD screen. Guiding text includes exit and cancel
.
Page 8
Using the Menus
RAMP: 10min
menu
WELCOME
>>>>>>
00006/00012 days
USED: 00090hrs
SW: SX459-0200
exit
MASK: ULTRA
change exit
SETTINGS
enter exit
RESULTS
enter exit
enter exit
OPTIONS
TUBE LENGTH: 2m
change exit
S
MARTSTART: OFF
exit
MASK: ULTRA
change exit
LEAK ALER
T: ON
change exit
LANGUAGE:ENGLISH
change exit
SMART DATA
enter exit
enter exit
SERVICING
SW: SX459-0200
exit
SN: 20080102356
exit
AUTO APPEAR: ON
change exit
SMART DATA
enter exit
USED: 00090hrs
00003/00122days
MASK FIT: *****
excelnt exit
PRESSURE: 8.0
exit
USAGE: 4.00hrs
exit
These menus only appear if
your clinician has enabled
Smart Data.
Standard ViewDetailed View
The VPAP IV and VPAP IV ST provide a set of functions which are arranged in menus
and submenus. Via the LCD screen, the menus and submenus allow you to view and
change the settings for a particular function. To navigate and make selections:
1 Press or to scroll through items within a level.
2 Press to enter a submenu and to apply an option choice.
3 Press to navigate out of a menu or submenu and to exit without changing
options.
Your clinician has preset the menu to either a standard view or a detailed view. The
following illustration summarises these views:
English
5Using the VPAP IV and VPAP IV ST
Page 9
How to Select the Mask Type
!
Scroll to MASK and select . Press or until you see the setting
you require. The following table shows the setting that should be selected for each
mask:
SettingsMask
ULTRAUltra Mirage Nasal Mask
MIR FULLMirage Liberty Full Face Mask
ACTIVAMirage Activa Nasal Mask
SWIFTMirage Swift Nasal Pillows System
STANDARDMirage Vista Nasal Mask
MIRAGEMirage Nasal Mask
Ultra Mirage II Nasal Mask
Mirage Quattro Full Face Mask
Ultra Mirage Full Face Mask
Mirage Swift II Nasal Pillows System
Mirage Micro Nasal Mask
SmartStart™
If your clinician has enabled SmartStart/Stop, your device will start automatically
when you breathe into your mask and stop automatically when you take your mask
off.
Starting Treatment
1 Make sure the power is on.
The product name is displayed briefly on the LCD screen, then the standby (Ramp)
screen appears. The key and LCD backlights also turn on.
2 Fit your mask as described in the mask user instructions.
WARNING
A mask should not be used unless your VPAP device is turned on and operating
properly.
3 Alter the ramp time if required.
4 To start therapy, simply breathe into the mask or press .
5 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep.
6
Page 10
6 After starting therapy, an introductory treatment screen will display:
This screen only dispays if
ResLink and an oximeter are
connected
Inspiration information
Leak and breath information
Mode and pressure display
S mode
example
Mask-fit star rating
S:RAMP 4.0-10.0
>>>> PS 6.0
LK: 10L/min RR: 15
MV: 10.2 VT: 680
S**TiMn TiMx 2.0
Ti 1.5s 1: 1.6
SpO2: 98%
HR : 60 exit
Stopping Treatment
To stop treatment at any time, remove your mask and press . If your clinician
has enabled SmartStart/Stop, simply remove your mask and treatment will end.
Note: SmartStart/Stop does not operate if:
• you have a Full Face Mask; or
• Leak Alert is enabled.
Using the Mask-Fitting feature
The VPAP IV and VPAP IV ST include a mask-fitting feature to help you fit your mask
properly. The mask-fitting feature delivers air pressure for a three-minute period, prior
to starting treatment, for checking and adjusting your mask fit to minimise leaks.
To use the mask-fitting feature:
1 Fit your mask as described in the mask user instructions.
2 Hold down for at least three seconds until air pressure delivery starts and
the following screen appears:
MASK FIT:
excelnt
*****
English
The LCD displays a mask-fit star rating from zero to five stars. Three to five stars
indicates a good fit or better. Zero to two stars indicates that the mask needs to be
adjusted.
7Using the VPAP IV and VPAP IV ST
Page 11
Reminders on the VPAP LCD
Your clinician may have set your VPAP device to remind you about important events,
such as when to replace your mask, when to insert your Data Card and so on.
The reminder message is displayed on the LCD and is visible if the device is not
delivering therapy. The backlight on the LCD flashes when a message is displayed.
Your clinician can set any of the following reminders on your LCD:
MessageDescriptionAction
INSERT
CARD
REPLACE
MASK
CALL
PROVIDER
REPLACE
FILTER
SERVICE
DUE
Customised
messages
May appear if your device is
Data Card enabled.
Reminds you that your mask
is due for replacement.
Reminds you to contact your
clinician; for example to
discuss your therapy.
Reminds you to replace the
air filter on your device.
Reminds you to return your
device for service.
Your clinician may also set
reminders for other reasons;
for example to call a
particular person or number.
Insert your Data Card and follow any
instructions that your clinician has given you.
When you have done this, the message will
disappear from the LCD. Pressing (Ok)
will also remove the message.
Press (Ok) to remove the message
from your LCD and replace your mask with a
new one.
Press (Ok) to remove the message
from your LCD and contact your clinician/
service provider.
Press (Ok) to remove the message
from your LCD and replace the air filter.
Press (Ok) to remove the message
from your LCD and contact your clinician/
service provider.
Press (Ok) to remove the message
from your LCD and contact your clinician/
service provider.
Smart Data™
Smart Data menus only appear if enabled by the clinician. Your clinician can set any
of the following Smart Data options:
MessageDescription
PRESSUREDisplays the therapy pressure from the previous session.
MASK FITDisplays the mask-fit star rating from the previous session.
USAGEDisplays the usage hours from the previous session.
AUTO APPEARIf ON is selected, Smart Data screens are automatically displayed
when you turn on your device.
8
Page 12
Using the Data Card
!
If your clinician needs to review your treatment, they will ask you to use the Data Card
to copy data from your VPAP device and return the card to them.
Copying Data onto a Data Card
1 Switch on your VPAP and wait until you see the standby (Ramp) screen.
2 Hold the Data Card with the arrow facing up and insert it into the Data Card slot until
it stops. Data copying starts automatically.
The “Card Inserted Please Wait” message is displayed on the LCD while data is being
copied. Copying takes up to 30 seconds.
The “Copy Complete Remove Card” message is displayed on the LCD when copying
has finished.
3 Remove the Data Card by gripping the end of the Data Card and pulling it out.
4 Store the Data Card in its protective folder when not in use.
5 Return the card in its protective folder to your clinician using a postal envelope.
Updating Settings on your VPAP
If your clinician has provided a Data Card with new device settings:
1 With the device in standby (Ramp) mode, insert the Data Card into the slot on the
Data Card module. Updating will start automatically.
The “Card Inserted Please Wait” message is displayed on the LCD while updating is
in progress. Updating takes approximately five seconds.
The “Settings Success Remove Card” message is displayed on the LCD if the
settings were updated successfully.
Note: This message only appears once. If you re-insert the Data Card after you have
updated your settings, this message is not displayed.
2 Remove the Data Card from the VPAP device.
3 Store the Data Card in its protective folder when not in use.
WARNING
If your clinician has told you to use the Data Card to update the settings on your
device and the “Settings Success” message does not appear, contact your
clinician immediately.
English
9Using the VPAP IV and VPAP IV ST
Page 13
Traveling with the VPAP IV and VPAP IV ST
!
International Use
Your VPAP flow generator has an internal power adapter that enables it to operate in
other countries. It will operate on power supplies of 100–240V and 50–60Hz. No
special adjustment is necessary, but you will require an approved power cord for that
country.
Use on an Aircraft
Please consult the medical services department of your carrier if you intend to use
your VPAP device on an aircraft.
Note: You should not use your VPAP device while the aircraft is taking off or landing.
Use with DC Power
You must use a ResMed DC/DC Converter 24V/50W to connect your VPAP to a 12V
or 24V DC power source. Contact your equipment supplier or ResMed for details.
WARNING
The device should not be connected to both AC and DC power sources
simultaneously.
10
Page 14
Cleaning and Maintenance
!
!
Air filter cover
You should regularly carry out the cleaning and maintenance described in this section.
Refer to your mask and humidifier manuals for detailed instructions regarding the care
of those devices.
Daily
Disconnect the air tubing and hang it in a clean, dry place until next use.
Weekly
1 Remove the air tubing from the VPAP device and the mask.
2 Wash the air tubing in warm water using mild detergent.
3 Rinse thoroughly, hang, and allow to dry.
4 Before the next use, reconnect the air tubing to the air outlet and mask.
Monthly
1 Clean the exterior of the VPAP with a damp cloth and mild liquid soap.
2 Check the air filter for holes and blockage by dirt or dust.
WARNING
• Beware of electric shock. Do not immerse the unit or power cord in water.
Always unplug the unit before cleaning and be sure it is dry before plugging
back in.
• The mask system and air tubing are subject to normal wear and tear. Inspect
them regularly for damage.
English
Replacing the Air Filter
Replace the air filter every six months (or more often if necessary).
1 Remove the air filter cover at the back of the VPAP device.
2 Remove and discard the old air filter.
3 Insert a new filter with the blue-tinted side facing out from the device.
4 Replace the air filter cover.
WARNING
• Do not wash the air filter. The air filter is not washable or reusable.
• The air filter cover protects the device in the event of accidental liquid
spillage onto the device. Ensure the air filter and air filter cover are fitted at
all times.
11Cleaning and Maintenance
Page 15
Servicing
!
This device is intended to provide safe and reliable operation when operated and
maintained in accordance with the instructions provided by ResMed. To ensure that
your device continues to provide optimum performance it is recommended that this
product be inspected by an authorised ResMed Service Centre five years from the
date of purchase. Applicable ResMed warranty details are provided with the device
at the time of original supply. Of course, as with all electrical devices, if any irregularity
becomes apparent, you should exercise caution and have the device inspected by an
authorised ResMed service centre.
CAUTION
• Do not attempt to open the VPAP case. There are no user serviceable parts
inside.
• Inspection and repair should only be performed by an authorised agent.
Under no circumstances should you attempt to service or repair the flow
generator yourself.
12
Page 16
Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be solved,
contact your equipment supplier or ResMed. Do not attempt to open the device.
Problem/Possible CauseSolution
No display
Power is not connected.Ensure the power cable is connected and the power switch (if
Insufficient air delivered from the VPAP device
Ramp time is in use.Wait for air pressure to build up or change ramp time.
Humidifier control dial set too
high, resulting in accumulation of
water in the air tubing.
Device does not start when you breathe into the mask
Breath is not deep enough to
trigger SmartStart/Stop.
There is excessive leak.Adjust position of mask and headgear.
SmartStart/Stop is disabled.
Note: SmartStart/Stop is not
available if you are using a Full
Face Mask or if Leak Alert is
enabled.
Device does not stop when you remove your mask
SmartStart/Stop is not enabled.
Note: SmartStart/Stop is not
available if you are using a Full
Face Mask or if Leak Alert is
enabled.
SmartStart/Stop is enabled but the flow generator does not stop automatically when you
remove your mask
Incompatible mask system being
used.
Displays error message: Check tube!! Key if done
The air tubing is loose or blocked.Check that the air tubing is connected securely to your mask
available) is on.
Check air tubing.
Adjust position of mask and headgear.
Adjust headgear or re-fit cushion.
Turn humidifier control down and empty the water from the air
tubing.
Take a deep breath in and out through the mask.
Air tubing not connected properly. Connect firmly at both ends.
Enable SmartStart/Stop.
Enable SmartStart/Stop.
Use only equipment recommended by ResMed.
and the air outlet. Press the Start/Stop key to restart the
device. If this does not clear the message, disconnect the
power cord and then reconnect it to restart the device.
English
13Troubleshooting
Page 17
Problem/Possible CauseSolution
Displays error message: Exxxx Call Service (where xxxx defines an error)
Component failure.Record error number and contact your ResMed service centre.
Displays error message: HIGH LEAK!!! Adjust Mask
You have experienced excessively
high leak levels for more than 20
seconds.
Check that your air tubing is connected properly.
Adjust headgear.
The following message is displayed on the LCD after you try to update settings or copy data to
the Data Card: Card Error Remove Card
Data Card is not inserted
correctly.
You may have removed the Data
Card before settings were copied
to the VPAP device.
Ensure that the Data Card is inserted with the arrow facing up
as far as it can go.
Reinsert the Data Card and wait for the
Settings Success
Remove Cardor Copy Complete Remove Card
message to appear on the LCD.
The following message is displayed on the LCD after you try to update the settings using the
Data Card: Settings Invalid Remove Card
The identification details on the
Data Card do not match the
details on your device.
Contact your clinician/service provider immediately.
The following message is displayed on the LCD after you try to update the settings using the
Data Card: Settings Error Remove Card
There is a data error on the Data
Card.
Contact your clinician/service provider immediately.
The following message is NOT displayed on the LCD after you try to update the settings using
the Data Card: Settings Success Remove Card
The settings were not updated.Contact your clinician/service provider immediately.
14
Page 18
Technical Specifications
Operating pressure range2 to 25 cm H2O
Maximum single fault pressure40 cm H
Pressure measurement
tolerance
Flow measurement tolerance±0.1 or 20% of reading, whichever is greater
S, ST and T modesIPAP: 4 to 25 cm H2O (measured at the mask); EPAP: 2 to 25 cm H2O
CPAP mode4 to 20 cm H
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:
Sound pressure level26 dBA with uncertainty of 2 dBA as measured according to
Sound power level36 dBA with uncertainty of 2 dBA as measured according to
Dimensions (L x W x H)
Weight1.3 kg
Power supplyInput range 100–240V, 50–60Hz, 40VA (typical power consumption),
Storage and transport humidity10–95% non-condensing
Operating altitudeSea level to 2,600 m
Electromagnetic compatibilityProduct complies with all applicable electromagnetic compatibility
Air filterTwo-layered, powder-bonded, polyester non-woven fiber
Air tubingFlexible plastic, 1 x 2 m
Air outletThe 22 mm conical air outlet complies with ISO 5356-1
IEC 60601-1 classificationClass II (double insulation), Type CF
112 mm x 164 mm x 145 mm
±0.5 cm H
(measured at the mask); Pressure Support: 0 to 23 cm H
ISO 17510-1: 2002
28 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1: 2007
ISO 17510-1: 2007
< 100VA (maximum power consumption)
-20
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments
Notes:
• The manufacturer reserves the right to change these specifications without notice.
• Pressure may be displayed in cm H
O
2
O ± 4% of the measured reading
2
O (measured at the mask)
2
ºC to +35ºC
ºC to +60ºC
0 or hPa.
2
O
2
English
15Technical Specifications
Page 19
Symbols Which Appear On The Device
Caution; Follow instructions for use; Drip proof; Type CF equipment;
Dangerous voltage; Class II equipment; Start/Stop and mask-fit;
Manufacturer; Environmental information
WEEE 2002/96/EC is a European Directive that requires the proper disposal of electrical and
electronic equipment. This device should be disposed of separately, not as unsorted municipal
waste. To dispose of your device, you should use appropriate collection, reuse and recycling
systems available in your region. The use of these collection, reuse and recycling systems is
designed to reduce pressure on natural resources and prevent hazardous substances from
damaging the environment.
If you need information on these disposal systems, please contact your local waste
administration. The crossed-bin symbol invites you to use these disposal systems. If you require
information on collection and disposal of your ResMed device please contact your ResMed
office, local distributor or go to www.resmed.com/environment.
16
Page 20
General Warnings and Cautions
!
!
Warnings
• Advice contained in this manual should not supersede instructions given by the
prescribing physician.
• A patient should not connect a device to the data communication port unless instructed
to do so by their health care provider or physician. Only ResMed products are designed
to be connected to the data communication port. Connecting other devices could result
in injury, or damage to the VPAP device.
• These VPAP devices should only be used with masks (and connectors
by ResMed, or by a physician or respiratory therapist. A mask should not be used unless
the VPAP device is turned on and operating properly. The vent hole or holes associated
with the mask should never be blocked.
Explanation: These VPAP devices are intended to be used with special masks (or
connectors
When the device is turned on and functioning properly, new air from the device flushes
the exhaled air out through the mask vent holes. However, when the device is not
operating, insufficient fresh air will be provided through the mask, and the exhaled air
may be rebreathed. Rebreathing of exhaled air for longer than several minutes can, in
some circumstances, lead to suffocation. This applies to most Positive Airway Pressure
devices.
• In the event of power failure
• These VPAP devices can be set to deliver pressures up to 25 cm H
event of certain fault conditions, pressures up to 40 cm H
• Follow all precautions when using supplemental oxygen.
• Oxygen flow must be turned off when the flow generator is not operating, so that
unused oxygen does not accumulate within the flow generator enclosure and create a
risk of fire.
• If the oxygen has been left on, turn off the device, then wait 30 minutes before turning on
the device again.
• At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration varies,
depending on the pressure settings, patient breathing pattern, mask, point of application
and leak rate.
• Do not use the VPAP IV or VPAP IV ST if there are obvious external defects or
unexplained changes in performance.
• Do not open the VPAP case. There are no user serviceable parts inside. Repairs and
internal servicing should only be performed by an authorised service agent.
• Explosion hazard—do not use in the vicinity of flammable anesthetics.
*
) which have vent holes to allow continuous flow of air out of the mask.
†
or machine malfunction, remove the mask.
O are possible.
2
Cautions
• At low pressures, the flow through the exhalation ports of your mask may not clear all
exhaled gas from the tubing. Some rebreathing may occur.
• The airflow for breathing produced by this device can be as much as 11ºF (6ºC) higher
than the temperature of the room. Caution should be exercised if the room temperature
is warmer than 90ºF (32ºC).
• When AC mains power (100–240V AC) is not available, always use a ResMed DC/DC
Converter 24V/50W. (The DC/DC Converter 24V/50W is available as an optional
accessory. It is not supplied with all models.)
Note: The above are general warnings and cautions. Specific warnings, cautions and
notes appear with the relevant instructions in the manual.
*
) recommended
O. In the unlikely
2
English
* Ports may be incorporated into the mask or in connectors that are near the mask.
† During partial (below rated minimum voltage) or total power failure, IPAP and EPAP pressures will not
be delivered. When power is restored, operation will recommence with no change to settings.
17General Warnings and Cautions
Page 21
Guidance and Manufacturer’s Declaration—Electromagnetic Emissions and Immunity
Guidance and manufacturer’s declaration—electromagnetic emissions
The VPAP IV and VPAP IV ST devices are intended for use in the electromagnetic environment specified below. The customer or
the user of the VPAP device should assure that the device is used in such an environment.
RF emissions CISPR11 Group 1The VPAP device uses RF energy only for its
RF emissions CISPR 11 with serial adapterClass BThe VPAP device is suitable for use in all
RF emissions CISPR 11 with USB adapterClass B
Harmonic Emissions IEC 610 00-3-2Class A
Voltage Fluctuations/Flicker Emissions IEC
61000-3-3
Warnings: The VPAP device should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the VPAP device should be observed to verify normal operation in the configuration in
which it will be used. The use of accessories (eg, humidifiers) other than those specified in this manual is not recommended.
They may result in increased emissions or decreased immunity of the VPAP device.
Recommended separation distances between portable and mobile RF communications equipment and the VPAP
series of devices
The VPAP device is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user
of the VPAP device can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the VPAP device as recommended below, according to the maximum
output power of the communications equipment.
Complies
Separation distance according to frequency of transmitter (m)
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
establishments, including domestic
establishments and those directly connected to
the public low-volt age network that supplies
buildings used for domestic purposes.
18
Rated maximum output
power of transmitter (W)
0.010.170.040.07
0.10.370.110.22
11.170.350.7
103.691.112.21
10011.703.507.0
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be
determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
150 kHz to 80 MHz
d = 1.17 √P
80 MHz to 800 MHz
d = 0.35 √P
800 MHz to 2.5 GHz
d = 0.7 √P
Page 22
Guidance and manufacturer’s declaration—electromagnetic immunity
The VPAP device is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP
device should assure that the device is used in such an environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
Power frequency
(50/60Hz) magnetic
field IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95%dip in
Ut) for 5 sec
3 A/m 3 A/m Power frequency magnetic fields should be at levels
3 Vrms
150 kHz to 80 MHz
10 V /m
80 MHz to 2.5 GHz
±6 kV contact
±8 kV air
±2 kV
Not Applicable
±1 kV differential
mode
±2 kV common
mode
< 12V (>95% dip in
240V) for 0.5 cycle
96V (60% dip in
240V) for 5 cycles
168V (30% dip in
240V) for 25 cycles
<12V (>95%dip in
240V) for 5 sec
3 Vrms
10 V /m
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typic al
commercial or hospital environment.
Mains power quality should be that of a typic al
commercial or hospital environment.
Mains power quality should be that of a typic al
commercial or hospital environment.
If the user of the VPAP device requires continued
operation during power mains interruptions, it is
recommended that the VPAP device be powered from
an uninterruptible power source.
characteristic of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF communications equipment
should be used no closer to any part of the VPAP
device, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.17 √P
d = 0.35 √P 80 MHz to 800 MHz
d = 0.70 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with this symbol .
b
a
NOTE 1: Ut is the AC mains voltage prior to application of the test level.
NOTE 2: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the VPAP device is used exceeds the applicable RF compliance level above, the VPAP device
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the VPAP device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
19General Warnings and Cautions
Page 23
Limited Warranty
R001-307/2 05 06
ResMed warrants that your ResMed product shall be free from defects in material
and workmanship for the period specified below from the date of purchase by the
initial consumer. This warranty is not transferable.
ProductWarranty Period
ResMed humidifiers, ResControl™, ResLink™, ResTraxx™1 Year
ResMed flow generators2 Years
Accessories, mask systems (including mask frame, cushion,
headgear and tubing). Excludes single-use devices.
Note: Some models are not available in all regions.
If the product fails under conditions of normal use, ResMed will repair or replace, at
its option, the defective product or any of its components. This Limited Warranty
does not cover:
a) any damage caused as a result of improper use, abuse, modification or alteration
of the product;
b) repairs carried out by any service organisation that has not been expressly
authorised by ResMed to perform such repairs;
c) any damage or contamination due to cigarette, pipe, cigar or other smoke;
d) any damage caused by water being spilled on or into a flow generator.
Warranty is void on product sold, or resold, outside the region of original purchase.
Warranty claims on defective product must be made by the initial consumer at the
point of purchase.
This warranty is in lieu of all other express or implied warranties, including any implied
warranty of merchantability or fitness for a particular purpose. Some regions or states
do not allow limitations on how long an implied warranty lasts, so the above limitation
may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed
to have occurred as a result of the sale, installation or use of any ResMed product.
Some regions or states do not allow the exclusion or limitation of incidental or
consequential damages, so the above limitation may not apply to you. This warranty
gives you specific legal rights, and you may also have other rights which vary from
region to region.
For further information on your warranty rights, contact your local ResMed dealer or
ResMed office.
90 Days
20
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.