ResMed Positive AirwAy Pressure Device S8 Series II User Manual

S8 Series II
S8 AutoSet Spirit™ II S8 AutoScore™ II
POSITIVE AIRWAY PRESSURE DEVICES
User Guide
English
ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA ResMed (UK) Ltd (EU Authorized Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK ResMed Offices Australia, Austria, Belgium, Brazil, China, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Malaysia, Mexico,
Netherlands, New Zealand, Norway, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com for contact details).
Protected by patents: AU 691200, AU 697652, AU 709279, AU 724589, AU 730844, AU 736723, AU 750095, AU 750761, AU 764761, AU 779327, AU 2002231456, CN 02804930.6, EP 0651971, EP 0661071, EP 0920845, EP 0934723, EP 1028769, EP 1126893, EP 1502618, EP 1144036, JP 3580776, JP 3778797, NZ 504595, US 5199424, US 5245995, US 5522382, US 5704345, US 6029665, US 6138675, US 6240921, US 6363933, US 6367474, US 6425395, US 6502572, US 6591834, US 6705315, US 6745768, US 6817361, US 6988498, US 7040317, US 7100608. Other patents pending. Protected by design registrations: AU 302180, AU 302181, AU 302182, AU 302183, AU 302184, AU 302185, AU 302186, CN 200430121212.2, CN
200430121213.7, CN 200430121211.8, CN 200430121214.1, CN 200430121215.6, CN 200430121216.0, CN 200430121217.5, EU 269436, EU 815071, HK
0412901.9, JP 1257662, JP 1257663, JP 1266956, JP 1266957, JP 1267270, JP 1267271, NZ 405614, NZ 406923, NZ 406924, NZ 406925, NZ 406926, NZ 406927, NZ 406928, US D544598, US D553228, US D557406, US D557407, US D560795, US D561891, US D565160. Others pending. S8, S8 AutoSet, S8 Elite and SmartStart are trademarks of ResMed Ltd and AutoSet and HumidAire are registered in U.S. Patent and Trademark Office.
© 2008 ResMed Ltd. 338575/2 08 10

Contents

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Contraindications 1 Adverse Effects 1
Parts of the S8 Series II system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Masks 2 Humidifiers 2 ResScan Data Card 2
Using the S8 Series II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Setting Up 3 How to Use the Control Panel 4 Selecting the Mask Type 4 Using the Mask-fitting Feature 5 Starting Treatment 5 Stopping Treatment 6
Changing the Settings on your S8 II Device . . . . . . . . . . . . . . . . . 6
The Standard Menu 6 The Detailed Menu 7
Messages on the S8 Series II LCD . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Using the Data Card to Collect Data . . . . . . . . . . . . . . . . . . . . . . . . 8
Copying Data onto the Data Card 8
Cleaning and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Daily 9 Weekly 9 Monthly 9 Replacing the Air Filter 10 Servicing 10
Frequently Asked Questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
General Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Introduction

S8 AutoSet Spirit II

The S8 AutoSet Spirit II self-adjusting sleep apnea system is indicated for the treatment of obstructive sleep apnoea (OSA) in adult patients. The S8 AutoSet Spirit II self­adjusting sleep apnoea system has two treatment modes: AutoSet and fixed-pressure CPAP. The S8 AutoSet Spirit II

S8 AutoScore II

The S8 AutoScore II continuous positive airway pressure (CPAP) system is indicated for the treatment of obstructive sleep apnoea (OSA) in adult patients. The S8 AutoScore II CPAP system is intended for home and hospital use.

Contraindications

Your S8 Series II is not a life-support device and may stop operating during power failure or certain fault conditions. It should not be used by patients who are dependent on continuous therapy. Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions:
severe bullous lung disease
pneumothorax
pathologically low blood pressure
dehydration
cerebrospinal fluid leak, recent cranial surgery, or trauma.

Adverse Effects

Patients should report unusual chest pain, severe headache, or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment. The following side effects may arise during the course of therapy with the CPAP device:
drying of the nose, mouth, or throat
nosebleed
bloating
ear or sinus discomfort
eye irritation
skin rashes
chest discomfort.
system is intended for home and hospital use.
1Introduction

Parts of the S8 Series II system

Carry handle
Screen
Keypad
Air outlet
DC
AC
Data Card slot
Your S8 Series II device comprises:
S8 Series II device
Power cord
Carry bag
2 m air tubing.
Optional components include:
Integrated humidifier
Front cap
3 m air tubing
DC-12 converter
ResMed Tubing Wrap.

Masks

A list of masks (and their settings) compatible with these devices can be found on www.resmed.com, on the Products page under Service and Support.

Humidifiers

A list of humidifiers compatible with ResMed devices can be found on www.resmed.com, on the Products page under Service and Support.

ResScan Data Card

The ResScan Data Card may be used with the S8 Series II either to help your clinician monitor your treatment or provide you with updates to your device settings.
2

Using the S8 Series II

!
!

Setting Up

1 Connect the power cord to the socket at the rear of the S8 Series II device.
2 Plug the other end of the power cord into the power outlet.
3 Connect one end of the air tubing firmly onto the air outlet.
4 Connect the assembled mask system to the free end of air tubing.
For information on assembling your mask, see your mask manual.
2
1
3
4
WARNING
Make sure the power cord and plug are in good condition and the
equipment is not damaged.
Only ResMed air tubing should be used with the device. A different type
of air tubing may alter the pressure you actually receive, reducing the effectiveness of your treatment.
Blocking the hose and/or air inlet of the device while in operation could
lead to overheating of the device.
CAUTION
Be careful not to place the device where it can be bumped or where
someone is likely to trip over the power cord.
If you put the device on the floor, make sure the area is free from dust
and clear of bedding, clothes or other objects that could block the air inlet.
Make sure the area around the device is dry and clean.
3Using the S8 Series II

How to Use the Control Panel

LCD screen
Start/Stop key
Keypad
The control panel of the S8 Series II includes an LCD screen which displays the menus and treatment screens as well as a keypad for navigating through the menus and delivering treatment.
The S8 Series II keypad has the following keys:
Key Function
Start/Stop
Starts or stops treatment.
Up
Down
Left
Right
The keypad and LCD are equipped with backlights that come on when the device is powered. The LCD backlight turns off after two minutes of inactivity, and comes back on when you press a key. The keypad backlight is always on while the S8 Series II is on.
Allows you to increase settings options and scroll through the S8 Series II menu.
Allows you to decrease settings options and scroll through the S8 Series II menu.
Performs the function indicated by the guiding text displayed above it on the LCD screen (eg, change).
Performs the function indicated by the guiding text displayed above it on the LCD screen (eg, exit)
.

Selecting the Mask Type

To ensure your device functions as intended with your selected mask, you need to select the mask type that you are using.
1 With the RAMP or SETTLING screen displayed, press . (In CPAP mode, the
standby screen displays RAMP. In AutoSet mode (S8 AutoSet Spirit II only) the standby screen displays SETTLING.)
Press until MASK is displayed.
2
Press to change the setting.
3
4
Press or until you see the setting that you require.
!
MASK FIT:
*****
excelnt
4
Press to apply.
5

Using the Mask-fitting Feature

You can use the S8 Series II mask-fitting feature to help you fit your mask properly. This feature delivers air pressure for a three-minute period, prior to starting treatment, during which you can check and adjust your mask fit to minimise leaks.
Note: Make sure that the mask setting on the MASK screen is the same as the mask you are using.
1 Fit your mask as described in the mask user instructions.
2 With the RAMP or SETTLING screens displayed, hold down for at least
three seconds until air pressure delivery starts. Your device will ramp to the mask-fit pressure and will remain at this pressure for three minutes. The following displays on the LCD screen indicating that the mask-fitting feature is operating, where: ***** is an excellent fit;****- is a very good fit; ***-- is a good fit; while **---, *---- and ----- each require you to adjust your mask.
3 Adjust your mask, mask cushion and headgear until you have a secure and
comfortable fit. After three minutes, the pressure reverts to the set pressure and treatment will begin. If you do not wish to wait three minutes, hold down for at least three seconds to start treatment immediately. If you press for less than three seconds, the unit will return to standby mode (the RAMP or SETTLING screen).

Starting Treatment

1 Make sure the power is on.
The product name is displayed briefly on the screen, then the standby screen displays RAMP (or SETTLING). The key and LCD backlights also turn on.
Lie down and arrange the air tubing so that it is free to move if you turn in your
2
sleep.
CAUTION
Do not leave long lengths of air tubing around the top of your bed. They could twist around your head or neck while you are sleeping.
Alter the ramp or settling time if required.
3
5Using the S8 Series II