ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia • ResMed Corp (US
Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA • ResMed (UK) Ltd (EU Authorized
Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK • ResMed Offices Australia, Austria, Brazil, China,
Finland, France, Germany, Hong Kong, India, Japan, Malaysia, Netherlands, New Zealand, Norway, Singapore, Spain,
Sweden, Switzerland, UK, USA (see www.resmed.com for contact details).
Protected by patents: US 5199424, US 5522382, US 6705315. Other patents pending.
Protected by design registrations: AU 302180, AU 302181, AU 302182, AU 302183, AU 302184, AU 302185, AU 302186,
CN 200430121212.2, CN 200430121213.7, CN 200430121211.8, CN 200430121214.1, CN 200430121215.6, CN
200430121216.0, CN 200430121217.5, EP 269436, HK 0412901.9, JP 1257662, JP 1257663, JP 1266956, JP 1266957, JP
1267270, JP 1267271, NZ 405614, NZ 406923, NZ 406924, NZ 406925, NZ 406926, NZ 406927, NZ 406928, US
D544598, US D553228, US D557406, US D557407. Others pending.
S8, S8 Escape, HumidAire 3i, ResScan, ResControl, ResLink and ResTraxx are trademarks of ResMed Ltd and Escape
and HumidAire are registered in U.S. Patent and Trademark Office.
The S8 Escape II continuous positive airway pressure (CPAP) system is indicated
for the treatment of obstructive sleep apnea (OSA) in patients > 66lb (30 kg).
The S8 Escape II CPAP system is intended for home and hospital use.
CAUTION
!
In the US, Federal law restricts this device to sale by or on the order of a
physician.
Contraindications
The S8 Escape II is not a life support device and may stop operating during power
failure or certain fault conditions. It should not be used by patients who are
dependent on continuous therapy.
Positive airway pressure therapy may be contraindicated in some patients with
the following pre-existing conditions:
• severe bullous lung disease
• pneumothorax
• pathologically low blood pressure
• dehydration
• cerebrospinal fluid leak, recent cranial surgery, or trauma.
Adverse Effects
Patients should report unusual chest pain, severe headache, or increased
breathlessness to their prescribing physician. An acute upper respiratory tract
infection may require temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the CPAP
device:
• drying of the nose, mouth, or throat
• nosebleed
• bloating
• ear or sinus discomfort
• eye irritation
• skin rashes.
English
1Introduction
The S8 Escape II System
Your S8 Escape II comprises:
• S8 Escape II positive airway pressure device
• 2 m air tubing
• Power cord
• Carry bag.
Carry handle
Screen
Keypad
Data
Card slot
Air inlet
Air outlet
Optional components include:
• ResScan™ Data Card
• 3 m air tubing
• DC-12 converter.
Masks
For information on using masks, see your mask manual. Recommended masks
are available on www.resmed.com on the Products page under Service & Support.
WARNING
!
Only ResMed mask systems are compatible for use with the S8 Escape II.
Please refer to “Adverse Effects” on page 1.
Humidifiers
A list of humidifiers compatible with ResMed devices can be found on
www.resmed.com, on the Products page under Service & Support.
ResScan Data Card
The ResScan Data Card may be used with the S8 Escape II either to help your
clinician to monitor your treatment or to provide you with updates to your device
settings.
DC
AC
2
Using the S8 Escape II
WARNING
!
• Make sure the power cord and plug are in good condition and the
equipment is not damaged.
• Only ResMed air tubing should be used with the device. A different type
of air tubing may alter the pressure you actually receive, reducing the
effectiveness of your treatment.
• Blocking the hose and/or air inlet of the device while in operation could
lead to overheating of the device.
CAUTION
!
• Be careful not to place the device where it can be bumped or where
someone is likely to trip over the power cord.
• If you put the device on the floor, make sure the area is free from dust
and clear of bedding, clothes or other objects that could block the air
inlet.
• Make sure the area around the device is dry and clean.
Set Up the S8 Escape II
1 Connect the power cord to the socket at the rear of the S8 Escape II.
2 Plug the other end of the power cord into the power outlet.
3 Connect one end of the air tubing firmly onto the air outlet.
English
4 Connect the assembled mask system to the free end of air tubing.
For information on assembling your mask, see your mask manual.
2
1
3
4
3Using the S8 Escape II
Start Treatment
LCD screen
Keypad
Start/Stop key
1 Make sure the power is on.
The product name is displayed briefly on the LCD screen, then the standby
(Ramp) screen appears. The key and LCD backlights also turn on.
2 Lie down and arrange the air tubing so that it is free to move if you turn in your
sleep.
CAUTION
!
Do not leave long lengths of air tubing around the top of your bed. They
could twist around your head or neck while you are sleeping.
3 Alter the ramp time if required (see “How to change the ramp time” on page 6).
4 To start treatment, press .
5 Fit your mask as described in the mask user instructions.
WARNING
!
A mask should not be used unless the S8 Escape II device is turned on and
operating properly.
When ramping begins the LCD will display the word "RAMPING", with the
current pressure indicated as dashes. Each dash is approximately equivalent to
2 cm H
pressure.
Stop Treatment
To stop treatment at any time, remove your mask and press .
4
O. For example, 6 dashes represents approximately 12 cm H2O
2
How to Use the Control Panel
The control panel of the S8 Escape II includes an LCD screen which displays the
menus and treatment screens as well as a keypad for navigating through the
menus and delivering treatment.
The S8 Escape II keypad has the following keys:
KeyFunction
Start/Stop
Starts or stops treatment.
English
Up
Down
Left
Right
Allows you to increase settings options and scroll through the
S8 Escape II menu.
Allows you to decrease settings options and scroll through
the S8 Escape II menu.
Performs the function indicated by the guiding text displayed
above it on the LCD screen. Guiding text includes menu, change, and apply.
Performs the function indicated by the guiding text displayed
above it on the LCD screen. Guiding text includes exit and
.
cancel
The keypad and LCD are equipped with backlights that come on when the device
is powered. The LCD backlight turns off after two minutes of inactivity, and
comes back on when you press a key. The keypad backlight is always on while
the S8 Escape II is on.
5Using the S8 Escape II
How to Change the Settings on your S8 Escape II
You can change settings (such as Ramp Time) on the S8 Escape II by using the
keypad and LCD. When the S8 Escape II is in standby mode you can display a
series of screens (a menu) on the LCD screen in order to view and change the
settings for a particular function.
S8 ESCAPE II
PATIENT MENU
S8 CLASSIC
S8 CLASSIC
RAMP
10min
Menu
EPR LEVEL: 1
Change Exit
USED HRS: 960
Exit
USAGE: 120/180
Exit
LANG
:
ENGLISH
Change Exit
SW: SX123456789
Exit
Use the menu screens to change the ramp time and language. You can also view
details about the usage and software version of your S8 Escape II.
Only displayed if
clinician enabled
How to change the ramp time
During ramp time, the pressure increases from a low pressure to the prescribed
treatment pressure. You can set the ramp time in 5-minute intervals, from 0
minutes (OFF) up to a maximum ramp time set by your clinician. The standby
screen displays the current ramp time in the right corner.
1. Press or until the desired ramp time appears.
2. Once a ramp time is selected, begin therapy by pressing .
When ramping begins the LCD will display the word "RAMPING", with the
current pressure indicated as dashes. Each dash is approximately equivalent
to 2 cm H
O. For example, six dashes represents approximately 12 cm H2O
2
pressure.
How to change the EPR level
EPR can be used to help ease any discomfort you may experience when
breathing out during treatment. If your clinician has allowed access you can
change the EPR level. There are four settings: OFF; 1 (lowest EPR); 2; and 3
(highest EPR).
6
1. On the standby (RAMP) screen, press .
2. Press
until you see EPR LEVEL, then press (Change).
3. Press or until you see the setting that you require.
4. Press (Apply) to save the setting.
How to change the language setting
1. On the standby (Ramp) screen, press (Menu).
2. Press
until you see LANG (Language), then press (Change).
3. Press or until you see the language that you require.
4. Press (Apply) to save the new language.
Messages on the S8 Escape II LCD
Your clinician may have set your S8 Escape II to remind you about important
events, such as when to replace your mask, when to insert your Data Card (if
your device is Data Card enabled) and so on.
The reminder message is displayed on the LCD and is visible if the device is not
delivering therapy. The backlight on the LCD flashes when a message is
displayed.
Your clinician can set any of the following reminders on your LCD:
MessageDescriptionAction
English
INSERT
CARD
REPLACE
MASK
CALL
PROVIDER
REPLACE
FILTER
SERVICE
DUE
May appear if your
device is Data Card
enabled.
May appear when
your mask is due for
replacement.
May appear as a
reminder to contact
your clinician, eg, to
discuss how your
therapy is going.
May appear as a
reminder to replace
the air filter on your
device.
May appear as a
reminder to return
your device for
service.
Insert your Data Card and follow any
instructions that your clinician has given you.
When you have done this, the message will
disappear from the LCD. Pressing (Ok)
will also remove the message.
Press (Ok) to remove the message from
your LCD, and replace your mask with a new
one.
Press (Ok) to remove the message from
your LCD, and contact your clinician.
Press (Ok) to remove the message from
your LCD, and replace the air filter.
Press (Ok) to remove the message from
your LCD, and contact your clinician.
7Using the S8 Escape II
Using the Data Card to Collect data
If your clinician needs to review your treatment, they will ask you to use the Data
Card to copy data from your S8 Escape II, and to return the card to them. The
Data Card will be provided in a protective folder.
Data that is copied to a Data Card is still stored and available on the S8 Escape II.
Copy data onto the Data Card
1. Switch on the S8 Escape II and wait until you see the standby (Ramp) screen.
2. Hold the Data Card with the arrow facing up and insert it into the Data Card
slot until it stops. Data copying starts automatically.
The “Card Inserted Please Wait” message is displayed on the S8 Escape II
LCD while data is being copied. Copying takes approximately five seconds.
The “Copy Complete Remove Card” message is displayed on the LCD when
copying has finished.
3. Remove the Data Card by gripping the end of the Data Card and pulling it out.
4. Store the Data Card in its protective folder when not in use.
5. Return the card in its protective folder to your clinician using a postal
envelope.
Using the Data Card to Update Settings
If your clinician has provided a Data Card with new device settings:
1. With the device in standby (Ramp) mode, insert the Data Card into the slot
on the Data Card module. Updating will start automatically.
The “Card Inserted Please Wait” message is displayed on the LCD while
updating is in progress. Updating takes approximately five seconds.
The “Settings Success Remove Card” message is displayed on the LCD if
the settings were updated successfully.
Note: This message only appears once. If you re-insert the Data Card after
you have updated your settings, this message is not displayed.
2. Remove the Data Card from the S8 Escape II.
3. Store the Data Card in its protective folder when not in use.
WARNING
!
If your clinician has told you to use this Data Card to update the settings on
your device and the “Settings Success” message does not appear, contact
your clinician immediately.
Cleaning and Maintenance
WARNING
!
Beware of electric shock. Do not immerse the unit or power cord in water.
Always unplug the unit before cleaning and be sure it is dry before plugging
back in.
8
CAUTION
!
• Do not hang the air tubing in direct sunlight as it may harden over time
and eventually crack.
• Do not use bleach, chlorine, alcohol, or aromatic-based solutions,
moisturising or antibacterial soaps or scented oils to clean the air tubing
or the S8 Escape II. These solutions may cause hardening and reduce
the life of the product.
Daily
Disconnect the air tubing and hang it in a clean, dry place until next use.
Weekly
1. Remove the air tubing from the S8 Escape II and the mask.
2. Wash the air tubing in warm water using mild detergent.
3. Rinse thoroughly, hang, and allow to dry.
4. Before next use, reconnect the air tubing to the air outlet and mask.
Monthly
1. Clean the exterior of the S8 Escape IIwith a damp cloth and mild liquid soap.
2. Check the air filter for holes and blockage by dirt or dust. Replace it every six
months, or more often in a dusty environment (see “Replacing the Air Filter”
on page 9).
English
Replacing the Air Filter
WARNING
!
• Do not wash the air filter. The air filter is not washable or reusable.
• The air filter cover protects the device in the event of accidental liquid
spillage onto the device. Ensure the air filter and air filter cover are fitted
at all times.
1. Remove the air filter cover at the back of the S8 Escape II.
Air filter cover
2. Remove and discard the old air filter.
3. Insert a new filter with the blue tinted side facing out from the device.
4. Replace the air filter cover.
9Cleaning and Maintenance
Servicing
CAUTION
!
• Do not attempt to open the S8 Escape II case. There are no user
serviceable parts inside.
• Inspection and repair should only be performed by an authorized agent.
Under no circumstances should you attempt to service or repair the S8
Escape II yourself.
This product (S8 Escape II) should be inspected by an authorized ResMed
Service Center five years from the date of manufacture. Prior to this, the device
is intended to provide safe and reliable operation provided that it is operated and
maintained in accordance with the instructions provided by ResMed. Applicable
ResMed warranty details are provided with the device at the time of original
supply. Of course, as with all electrical devices, if any irregularity becomes
apparent, you should exercise caution and have the device inspected by an
Authorized ResMed Service Center.
If you feel that your device is not performing properly, see “Troubleshooting” on
page 11.
Frequently Asked Questions
How will it feel when I am breathing with the S8 Escape II?
When you first use your S8 Escape II, you may feel uncomfortable breathing
against the airflow. This is normal. You will continue to breathe normally while
you sleep. The EPR setting may also improve comfort. Take time to adjust to this
new sensation.
What do I do when I want to get up in the night?
If you need to get up during the night, remove your mask and stop therapy.
Always remember to restart therapy and put your mask on when you return to
bed.
What do I do if I feel air leaking out of the mask?
The S8 Escape II delivers the most effective treatment when the mask is well
fitted and comfortable. Treatment can be affected by leaks, so it is important to
eliminate any leaks that may arise. For example, try repositioning the mask and
adjusting the straps.
If you are using a nasal mask, try to keep your mouth closed during treatment.
Air leaks from your mouth can decrease the effectiveness of your treatment. If
mouth leaks are a problem, a full face mask or chin strap may help. Contact your
clinician or equipment supplier for further details.
If you are using a full face mask and you have problems trying to get a
comfortable mask fit, contact your clinician or equipment supplier. You may
benefit from a different size or style of mask.
10
Anything special I should do before putting the mask on?
Before wearing your mask, wash your face to remove excess facial oils. This will
allow a better fit and prolong the life of the mask cushion.
Do not use moisturizer when wearing a mask.
My mouth feels dry during treatment—what can I do?
You may experience dryness of the nose, mouth, and/or throat during the course
of treatment, especially during winter. In many cases, a humidifier may resolve
this discomfort. Contact your clinician for advice.
My nose is irritated during treatment—what can I do?
You may experience sneezing and/or a runny or blocked nose during the first few
weeks of treatment. In many cases, nasal irritation can be resolved with a
humidifier. Consult your clinician or equipment supplier for advice.
Can I travel with the S8 Escape II?
Your S8 Escape II has an internal power supply that enables it to operate in other
countries. It will operate on power supplies of 100–240V and 50/60Hz. No special
adjustment is necessary, but you will require an approved power cord for the
region in which you are using the device.
Please consult the medical services department of your carrier if you intend to
use your S8 Escape II on an aircraft. The S8 Escape II complies with US FCC Part
15, Class B requirements.
Note: Do not use your S8 Escape II while the aircraft is taking off or landing.
Can I use my S8 Escape II if mains (AC) power is not available?
A DC-12 converter is available to connect the S8 Escape IIto a 12V or 24V DC
power source. Contact your clinician for details.
English
CAUTION
!
The device should not be connected to both AC and DC power sources
simultaneously.
Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be
solved, contact your equipment supplier or ResMed. Do not attempt to open the
device.
Problem/Possible CauseSolution
No display
Power is not connected.Ensure the power cable is connected and the power
switch (if available) is on.
Displays error message:
MOTOR FAULT Call Service or UPLOAD LANGUAGES Call Service or
DATA LOST Call Service
Component failure.Call service.
Exxxx Call Service (where xxxx defines an error) or
11Troubleshooting
Insufficient air delivered from S8 Escape II
Ramp Time is in use.Wait for air pressure to build up or change ramp time.
Air filter is dirty.Replace air filter.
Air tubing is kinked or
punctured.
Air tubing is not connected
properly.
Mask and headgear are not
positioned correctly.
Cap(s) are missing from
access port on mask.
Pressure required for
treatment may have changed.
Straighten or replace tubing.
Check air tubing.
Adjust position of mask and headgear.
Replace cap(s).
See your clinician to adjust the pressure.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card: Card Error Remove Card
Data Card is not inserted
correctly.
The Data Card is only partly
inserted.
You may have removed the
Data Card before settings
were copied to the
S8 Escape II.
Check that the patient is inserting the Data Card into the
Data Card module with
Escape II is switched on and not delivering therapy
(standby mode).
Ensure that the Data Card is inserted into the Data Card
slot as far as it will go.
Reinsert the Data Card and wait for the Settings Success Remove Card message to appear on the
LCD.
the arrow facing up, while the S8
The card is not a ResScan Data
Card.
Only use a ResScan Data Card.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:
The identification details on
the Data Card do not match
the identification details on
your device.
Settings Invalid Remove Card
Contact your clinician immediately.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:
There is a data error on the
Data Card.
Settings Error Remove Card
Contact your clinician immediately.
The following message is NOT displayed on the LCD after you try to update the
settings using the Data Card: Settings Success Remove Card
The settings were not
updated.
Contact your clinician immediately.
12
Technical Specifications
.
Operating pressure
range
Dimensions
(H x W x D)
Weight2.9lb (1.3 kg)
Power Supply• Input range for S8 Escape II with HumidAire 3i:100–240V, 50/60Hz;
Air FilterTwo-layered, powder-bonded, polyester non-woven fiber
Air TubingFlexible plastic, 6’6” (2 m)
Air OutletThe 22 mm conical outlet complies with EN 1281-1
IEC 60601-1
Classifications
Noise LevelRadiated sound pressure is measured at <30dBA at 10 cm H2O and
Te m p e r a t u r e
Operating Humidity10%–95% non-condensing
Storage and Transport
Te m p e r a t u r e
Storage and Transport
Humidity
Operating AltitudeSea level to 8500’ (2591 m)
Electromagnetic
Compatibility
4 to 20 cm H2O
4.6” x 6.5” x 5.7”
(112 mm x 164 mm x 145 mm)
110V, 400Hz; 2.5A < 140VA (110W) (maximum power consumption).
Instantaneous peak power consumption <340VA.
• Refer to the DC-12 converter instructions for DC ratings. Only use
the ResMed DC-12 converter for DC input.
• Typical power consumption at 20 cm H
Class II (double insulation), Type CF
3’3.36” (1 m) according to ISO17510-1.
+41ºF to +95ºF (+5ºC to +35ºC)
-4ºF to +140ºF (-20ºC to +60ºC)
10%–95% non-condensing
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments.
The electromagnetic compatibility tables for these ResMed devices can
be found on www.resmed.com, on the Products page under Service & Support. Click on the PDF file for your device.
0 is 34VA or 20W.
2
English
Note: The manufacturer reserves the right to change these specifications without notice.
Symbols which appear on the device
Attention, consult accompanying documents; Drip proof; Type CF equipment;
Dangerous voltage; Class II equipment; Start/Stop.
13Technical Specifications
General Warnings and Cautions
Warnings
• Read the entire guide before using the S8 Escape II.
• Advice contained in this guide should not supersede instructions given by the
prescribing physician.
• A patient should not connect a device to the data communication port unless
instructed to do so by their health care provider or physician. Only ResMed
products are designed to be connected to the data communication port.
Connecting other devices could result in injury, or damage to the S8 Escape II
device.
• The S8 Escape II should be used with masks (and connectors)*
recommended by ResMed, or by a physician or respiratory therapist. A mask
should not be used unless the S8 Escape II device is turned on and operating
properly. The vent hole or holes associated with the mask should never be
blocked.
Explanation: The S8 Escape II is intended to be used with special masks (or
connectors)* which have vent holes to allow continuous flow of air out of the
mask. When the device is turned on and functioning properly, new air from
the device flushes the exhaled air out through the mask vent holes. However,
when the device is not operating, insufficient fresh air will be provided
through the mask, and the exhaled air may be rebreathed. Rebreathing of
exhaled air for longer than several minutes can, in some circumstances, lead
to suffocation. This applies to most models of CPAP devices.
• In the event of power failure or machine malfunction, remove the mask.
• The S8 Escape II can be set to deliver pressures up to 20 cm H
unlikely event of certain fault conditions, pressures up to 30 cm H
possible.
• If oxygen is used with this device, the oxygen flow must be turned off when
the device is not operating.
If the oxygen has been left on, turn off the device, then wait 30 minutes
before turning on the device again.
Explanation: When the CPAP device is not in operation and the oxygen flow
is left on, oxygen delivered into the air delivery tubing may accumulate within
the CPAP machine enclosure and create a risk of fire. This applies to most
types of CPAP machines.
• Oxygen supports combustion. Oxygen should not be used while you are
smoking or in the presence of an open flame.
• Always ensure airflow is being generated by the device before the oxygen
supply is turned on.
• Always turn the oxygen supply off before stopping the airflow from the
device.
O. In the
2
O are
2
14
* Ports may be incorporated into the mask or in connectors that are near the mask.
Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen
concentration will vary, depending on where the oxygen is introduced, the
pressure settings, patient breathing pattern, mask selection, and leak rate.
• Do not use the S8 Escape II if there are obvious external defects or
unexplained changes in performance.
• Do not open the S8 Escape II case. There are no user serviceable parts
inside. Repairs and internal servicing should only be performed by an
authorized service agent.
• Explosion hazard—do not use in the vicinity of flammable anesthetics.
Cautions
• At low pressures, the flow through the exhalation ports of your mask may not
clear all exhaled gas from the tubing. Some rebreathing may occur.
• The airflow for breathing produced by this device can be as much as 11ºF
(6ºC) higher than the temperature of the room. Caution should be exercised if
the room temperature is warmer than 90ºF (32ºC).
• When AC mains power (100–240V AC) is not available, always use a ResMed
DC-12 converter. (The DC-12 converter is available as an optional accessory. It
is not supplied with all models.)
Note: The above are general warnings and cautions. Specific warnings,
cautions, and notes appear with the relevant instructions in the manual.
English
15General Warnings and Cautions
Limited Warranty
ResMed warrants that your ResMed product shall be free from defects in
material and workmanship for the period specified below from the date of
purchase by the initial consumer. This warranty is not transferable.
ProductWarranty Period
ResMed humidifiers, ResControl™, ResLink™, ResTraxx™1 Year
ResMed flow generators2 Years
Accessories, mask systems (including mask frame, cushion,
headgear and tubing). Excludes single-use devices.
Note: Some models are not available in all regions.
If the product fails under conditions of normal use, ResMed will repair or replace,
at its option, the defective product or any of its components. This Limited
Warranty does not cover:
a) any damage caused as a result of improper use, abuse, modification or
alteration of the product;
b) repairs carried out by any service organization that has not been expressly
authorized by ResMed to perform such repairs;
c) any damage or contamination due to cigarette, pipe, cigar or other smoke;
d) any damage caused by water being spilled on or into a flow generator.
Warranty is void on product sold, or resold, outside the region of original
purchase.
Warranty claims on defective product must be made by the initial consumer at
the point of purchase.
This warranty is in lieu of all other express or implied warranties, including any
implied warranty of merchantability or fitness for a particular purpose. Some
regions or states do not allow limitations on how long an implied warranty lasts,
so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages
claimed to have occurred as a result of the sale, installation or use of any ResMed
product. Some regions or states do not allow the exclusion or limitation of
incidental or consequential damages, so the above limitation may not apply to
you. This warranty gives you specific legal rights, and you may also have other
rights which vary from region to region.
For further information on your warranty rights, contact your local ResMed dealer
or ResMed office.
Installation de la S8 Escape II 3
Démarrage du traitement 4
Arrêt du traitement 4
Utilisation du panneau de contrôle 5
Modification des réglages de votre S8 Escape II 6
Utilisation de la carte de données pour
collecter des données
L'appareil de pression positive continue (PPC) S8 Escape II est prévu pour le
traitement du syndrome d'apnées obstructives du sommeil (SAOS) chez les
patients de plus de 66 lb (30 kg).
La S8 Escape II est prévue pour une utilisation à domicile ou en milieu médical.
ATTENTION
!
Selon la loi fédérale américaine, ce dispositif ne peut être vendu aux U.S.A.
que par un médecin ou sur prescription médicale.
Contre-indications
La S8 Escape II n'est pas un appareil de suppléance fonctionnelle et son
fonctionnement peut être interrompu en cas de coupure de l'alimentation ou
dans le cas d'une panne. Elle ne doit pas être utilisée chez les patients dont le
traitement ne peut pas être interrompu.
Le traitement par PPC est contre-indiqué chez certains patients présentant l'une
des pathologies préexistantes suivantes :
• affection pulmonaire bulleuse grave
• pneumothorax
• hypotension
• déshydratation
• fuite de liquide céphalo-rachidien, récent traumatisme ou intervention
chirurgicale crânienne.
Français
Effets secondaires
Les patients doivent avertir leur médecin traitant en cas de douleurs thoraciques
inhabituelles, de maux de tête sévères ou d'une dyspnée accrue. Une infection
aiguë des voies aériennes supérieures peut nécessiter l'arrêt temporaire du
traitement.
Les effets secondaires suivants peuvent apparaître au cours du traitement par
PPC :
• sécheresse nasale, buccale ou de la gorge
• saignements de nez
• ballonnements
• gêne au niveau de l'oreille ou des sinus
• irritation des yeux
• érythèmes cutanés.
1Introduction
Le système S8 Escape II
Votre S8 Escape II se compose des éléments suivants :
• Appareil de pression positive S8 Escape II
• Circuit respiratoire 2 m
• Câble d'alimentation
• Sac de transport.
Poignée de transport
Écran
Clavier
Fente de la carte de données
Entrée
d'air
Sortie d'air
Les composants en option sont les suivants :
• Carte de données ResScan™
• Circuit respiratoire 3 m
• Convertisseur DC-12.
Masques
Pour les informations concernant l'utilisation des masques, veuillez vous référer
au manuel de votre masque. Vous trouverez la liste des masques recommandés
sur le site www.resmed.com, à la page Produits de la rubrique SAV et assistance.
AVERTISSEMENT
!
Seuls les masques ResMed peuvent être utilisés avec la S8 Escape II.
Veuillez vous référer à la section « Effets secondaires » page 1.
Humidificateurs
Vous trouverez la liste des humidificateurs compatibles avec les appareils
ResMed sur le site www.resmed.com, page Produits, rubrique SAV et
assistance.
Carte de données ResScan
La carte de données ResScan peut être utilisée avec la S8 Escape II pour aider