ENGLISH
Nasal Mask
Thank you for choosing the AirFit N20 Classic. This document provides the user instructions for the AirFit N20 Classic.

Using this guide

Please read the entire guide before use. When following instructions, refer to the images at the front of the guide.

Intended use

The AirFit N20 Classic channels airflow non-invasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel device.
The AirFit N20 Classic is:
to be used by patients weighing more than 30 kg for whom positive airway pressure has been prescribed
intended for single-patient re-use in the home environment and multi- patient re-use in the hospital/institutional environment.
GENERAL WARNINGS
• The vent holes must be kept clear.
The mask should only be used with CPAP or bilevel devices
recommended by a physician or respiratory therapist.
The mask should not be used unless the device is turned on. Once the mask is fitted, ensure the device is blowing air. Explanation: CPAP and bilevel devices are intended to be used with special masks (or connectors) which have vent holes to allow continuous flow of air out of the mask. When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask vent holes. However, when the device is not operating, insufficient fresh air will be provided through the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can, in some
English 1
Page 7
circumstances, lead to suffocation. This applies to most models of CPAP or bilevel devices.
• Follow all precautions when using supplemental oxygen.
Oxygen flow must be turned off when the CPAP or bilevel device
is not operating, so that unused oxygen does not accumulate within the device enclosure and create a risk of fire.
Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of an open flame. Only use oxygen in well ventilated rooms.
At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate. This warning applies to most types of CPAP or bilevel devices.
The technical specifications of the mask are provided for your clinician to check that they are compatible with the CPAP or bilevel device. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect the CPAP or bilevel device function.
Discontinue using this mask if you have ANY adverse reaction to the use of the mask, and consult your physician or sleep therapist.
Using a mask may cause tooth, gum or jaw soreness or aggravate an existing dental condition. If symptoms occur, consult your physician or dentist.
As with all masks, some rebreathing may occur at low CPAP pressures.
Refer to your CPAP or bilevel device manual for details on settings and operational information.
• Remove all packaging before using the mask.
The mask must be used under qualified supervision for users who
are unable to remove the mask by themselves. The mask may not be suitable for those predisposed to aspiration.
Avoid connecting flexible PVC products (eg, PVC tubing) directly to any part of the mask. Flexible PVC contains elements that can be damaging to the materials of the mask, and may cause the components to crack or break.
The N20 line of nasal CPAP masks are not intended to be used simultaneously with nebuliser medications that are in the air path of the mask/tube.
2
Page 8
The mask contains an exhaust port safety feature to enable normal breathing. The mask should not be worn if this safety feature is damaged or missing.

Using your mask

When using your mask with ResMed CPAP or bilevel devices that have mask setting options, refer to the Technical specifications section in this user guide for the correct setting.
For a full list of compatible devices for this mask, see the Mask/Device Compatibility List on www.resmed.com/downloads/masks. If you do not have internet access, please contact your ResMed representative.

Fitting

1. With both lower headgear straps released, hold the mask against your face and pull the headgear over your head.
2. Bringing the lower straps below your ears, loop the headgear into the lower hooks on the mask frame.
3. Undo the Velcro
®
tabs on the upper headgear straps. Adjust the straps evenly until the mask sits comfortably just over the nose. Reattach the Velcro tabs.
4. Undo the Velcro tabs on the lower headgear straps. Adjust the straps evenly until the mask is stable. Reattach the Velcro tabs.
5. Connect the device air tubing to the elbow. Attach the elbow to the mask by pressing the side buttons and pushing the elbow into the mask, ensuring it clicks in on both sides.
6. Your mask and headgear should be positioned as shown in the illustration.

Adjustment

If necessary, slightly adjust the position of the mask for the most comfortable fit. Ensure that the cushion is not creased and the headgear is not twisted.
1. Turn on your device so that it is blowing air.
Adjustment tips:
To resolve any leaks at the upper part of the mask, adjust the
upper headgear straps. For the lower part, adjust the lower headgear straps. Adjust only enough for a comfortable seal.
English 3
Page 9
Ensure the headgear straps are not overtightened or too loose.

Removal

1. Unloop the lower headgear straps from the lower hooks of the frame.
2. With both lower straps released, pull the headgear back over your head.

Disassembly

If the mask is connected to your device, disconnect the device air tubing from the elbow.
1. Pull the straps out of the frame.
2. Remove the elbow from the mask by pressing the side button and pulling the elbow away.
3. Pull the cushion away from the frame.

Reassembly

1. Push the cushion into the frame. Note: Ensure the cushion mask size indicator is facing downwards.
2. Attach the top headgear straps to the mask frame by looping onto each of the upper headgear hooks.
3. Attach the elbow by pressing the side buttons and pushing into the mask.

Cleaning your mask at home

It is important to follow the steps below to get the best performance out of your mask.
WARNING
As part of good hygiene, always follow cleaning instructions and use a mild liquid detergent. Some cleaning products may damage the mask, its parts and their function, or leave harmful residual vapours that could be inhaled if not rinsed thoroughly.
Regularly clean your mask and its components to maintain the quality of your mask and to prevent the growth of germs that can adversely affect your health.
4
Page 10
CAUTION
Problem/possible cause
Solution
Mask is too noisy.
If any visible deterioration of a system component is apparent (cracking, crazing, tears etc), the component should be discarded and replaced.

Daily/After each use:

Before cleaning, take the mask components apart according to the disassembly instructions.
1. Soak and agitate the cushion and elbow in warm water (approximately 30°C) using a mild liquid detergent for up to 10 minutes.
2. Clean the components with a soft bristle brush, paying particular attention to the small vent holes on the elbow.
3. Rinse the components under warm flowing drinking quality water.
4. Inspect each component to ensure that they are visually clean and free of detergent residue. If required, repeat washing.
5. Allow the components to dry out of direct sunlight before assembling.

Weekly:

1. Soak and agitate the separated headgear and frame in warm water (approximately 30°C) using a mild liquid detergent for up to 10 minutes.
2. Rinse the components under warm flowing drinking quality water until they are free of detergent residue.
3. Squeeze the headgear to remove excess water.
4. Allow the components to dry out of direct sunlight before assembling.

Reprocessing the mask between patients

Reprocess this mask when using between patients. Cleaning, disinfection and sterilisation instructions are available on www.resmed.com/downloads/masks. If you do not have internet access, please contact your ResMed representative.

Troubleshooting

Vent is dirty. Clean the vent according to the instructions.
English 5
Page 11
Problem/possible cause
Solution
Mask is incorrectly assembled. Disassemble the mask, then reassemble
Mask leaks around face.
keep the mask in the right place.
Check that the headgear is not twisted.
Mask does not fit properly.
according to the instructions.
Mask may be the wrong size.
Talk to your clinician.
according to the instructions.
Mask is leaking. Refit or reposition your mask.
Readjust the headgear straps to bring the mask closer to the face and improve your seal.
Check that your mask is assembled correctly.
Mask may have been fitted incorrectly. Pull the mask away from your face and
reposition, ensuring that the lower corners of the mask fully enclose your nose. If necessary, readjust the headgear straps to
Cushion may be dirty. Clean the cushion according to the
instructions.
Cushion membrane is creased or folded. Remove and refit your mask according to the
fitting instructions. Ensure that you position the cushion correctly over your nose before pulling the headgear over your head. Do not slide the mask down your face during fitting as this may lead to folding or curling of the cushion.
Mask is not tightened correctly. Ensure that the headgear straps are not
overtightened or too loose. You may find that this mask does not need to be tightened as much as other nasal masks.
Mask may be the wrong size. Talk to your clinician.
Mask is not positioned correctly. Check that the mask size indicator on the
bottom of the cushion is facing downward.
Mask is incorrectly assembled.
Disassemble the mask, then reassemble
6
Page 12

Technical specifications

Pressure-flow
The mask contains passive venting to protect against rebreathing.
Pressure (cm H2O)
Flow
(L/min)
4
20 7 27
11
35
14
40
17
44
20
49
24
54
27
58
30
61
Dead space information
Physical dead space is the empty volume of the mask to the end of the swivel. Using the large cushions it is 55.8 mL.
Therapy pressure
Resistance
Drop in pressure measured (nominal)
at 100 L/min: 1.5 cm H2O
Environmental
Operating temperature: 5°C to 40°C
Storage and transport humidity: up to 95% non-condensing
Sound
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance
3 dBA.
Gross dimensions
141 mm (H) x 73 mm (W) x 90 mm (D)
curve
conditions
As a result of manufacturing variations, the vent flow rate may vary.
4 to 30 cm H2O
at 50 L/min: 0.4 cm H2O
Operating humidity: 15% to 95% non-condensing Storage and transport temperature: -20°C to +60°C
with ISO 4871. The A-weighted sound power level of the mask is 26 dBA, with uncertainty of 3 dBA. The A-weighted sound pressure level of the mask at a distance of 1 m is 18 dBA, with uncertainty of
Mask fully assembled with elbow assembly (no headgear)
English 7
Page 13
Service life
The service life of the AirFit N20 Classic mask system is dependent
on the intensity of usage, maintenance, and environmental
conditions to which the mask is used or stored. As this mask system
parts.
Mask setting
For AirSense, AirCurve and S9: Select 'Pillows'.
and its components are modular in nature, it is recommended that the user maintain and inspect it on a regular basis, and replace the mask system or any components if deemed necessary or according to the ‘visual criteria for product inspection’ in the ‘Cleaning your mask at home’ section of this guide. Refer to the ‘Mask components’ section of this guide for information of how to order replacement
options
For other devices: Select 'SWIFT' (if available), otherwise select 'MIRAGE' as the mask option.
Notes:
This product is not made with PVC or phthalates such as DEHP, DBP or BBP.
This product is not made with natural rubber latex.
The manufacturer reserves the right to change these specifications
without notice.

Storage

Ensure that the mask is thoroughly clean and dry before storing it for any length of time. Store the mask in a dry place out of direct sunlight.

Disposal

This mask does not contain any hazardous substances and may be disposed of with your normal household refuse.
8
Page 14

Symbols

The following symbols may appear on your product or packaging:
Catalogue number
Humidity limitation
Fragile, handle with care
Manufacturer
Keep away from rain
Classic nasal mask
Size - small
Size - large
Batch code
Temperature limitation
Not made with natural rubber latex
European Authorised Representative
This way up
Device setting - Pillows
Size - medium
Indicates a Warning or Caution and alerts you to a possible injury or explains special measures for the safe and effective use of the device
Caution, consult accompanying documents
English 9
Page 15

Consumer Warranty

ResMed acknowledges all consumer rights granted under the EU Directive 1999/44/EC and the respective national laws within the EU for products sold within the European Union.
10
Page 16
Mask Components / Maskenkomponenten / Composants du masque / Componenti della maschera / Maskeronderdelen / عﺎﻨﻘﻟا تﺎﻧﻮﻜﻣ
AirFit N20 Classic
A
Elbow / Kniestück / Coude / Gomito / Bocht /
1 Vent / Luftauslassöffnung / Orifice de ventilation / Foro
per l'esalazione / Ventilatieopening /
2 Side buttons / Seitlich angebrachte Laschen / Boutons
latéraux / Pulsanti laterali / Zijknoppen / نﺎﯿﺒﻧﺎﺟ نارز
3 Swivel / Drehadapter / Pièce pivotante / Giunto
rotante / Draaibare kop / ةراود ﺔﻠﺻو
B Frame / Maskenrahmen / Entourage rigide /
Telaio / Frame / رﺎطإ
C Cushion / Maskenkissen / Bulle / Cuscinetto /
Kussentje /
D Headgear / Kopfband / Harnais / Copricapo /
Hoofdband /
4 Lower headgear straps / Untere Kopfbänder / Sangles
inférieures du harnais / Cinghie inferiori del copricapo / Onderste banden van de hoofdband /
ةدﺎﺳو
سأر ءﺎﻄﻏ
ﺔﯾﻮﮭﺗ بﻮﻘﺛ
ﻖﻓﺮﻣ
63765
63766
63550 (S) 63551 (M) 63552 (L)
63760 (S) 63761 (Std)
نﺎﯿﻠﻔﺴﻟا سأﺮﻟا ءﺎﻄﻏ ﺎﻄﯾﺮﺷ
5 Upper headgear straps / Obere Kopfbänder / Sangles
supérieures du harnais / Cinghie superiori del copricapo / Bovenste banden van de hoofdband /
نﺎﯾﻮﻠﻌﻟا سأﺮﻟا ءﺎﻄﻏ ﺎﻄﯾﺮﺷ
A+B+C+D
Complete system / Komplettes System / Système complet / Sistema completo / Volledig systeem /
ﻞﻣﺎﻜﻟا مﺎﻈﻨﻟا
63710 (S) 63711 (M) 63712 (L)
Page 17
S: Small / Small / Small / Piccolo / Small / ﺮﯿﻐﺻ M: Medium / Medium / Medium / Medio / Medium / ﻂﺳو L: Large / Large / Large / Grande / Large / ﺮﯿﺒﻛ Std: Standard / Standard / Standard / Standard / Standaard / ﻲﺳﺎﯿﻗ
Page 18
638123
BA C
D
1
2 3
4
ResMed Ltd
1 Elizab eth Macar thur Drive B ella Vista N SW 2153 Aust ralia
See Re sMed.com fo r other ResM ed locatio ns worldwid e. AirFit is a t rademark a nd/or regis tered trad emark of the ResM ed family of com panies. For p atent and o ther intelle ctual prop erty info rmation, s ee ResMed.c om/ip. Velcro i s a trademar k of Velcro Indus tries B.V. © 2017 Re sMed Ltd. 63 8123/2 2 017-11