ResMed 61850-2 User Manual

Page 1
Mirage Liberty™
61850
FULL FACE MASK
User Guide
English
61850/2
07 04
MIRAGE LIBERTY
APAC ENG
Global leaders in sleep and respiratory medicine www.resmed.com
Page 2
ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA ResMed (UK) Ltd (EU Authorized Representative) 65 Milton Park Abingdon Oxfordshire OX14 4RX UK ResMed Offices Australia, Austria, Brazil, Finland, France, Germany, Hong Kong, Japan, Malaysia, Netherlands,
New Zealand, Norway, Singapore, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com for contact details).
Mirage Liberty Full Face Mask Protected by patents: EP 1314445, HK 1057714, US 6412487, US 6581594, US 6796308, US 6823865, US 7021311.
Other patents pending.
Protected by design registrations: AU 312036, AU 312037, AU 312038, AU 312039, AU 312040, AU 312971, AU 312972, AU 312973, AU 312974, AU 312975, AU 312976, AU 312977, AU 312978, AU 312979, AU 312980, AU 312981, AU 312982, AU 312983, AU 312984, AU 313118, AU 313120, AU 313121, AU 313122, AU 313123, AU 313124, AU 313125. Others pending.
Mirage and Mirage Liberty are trademarks of ResMed Ltd and Mirage is registered in U.S. Patent and Trademark Office.
Velcro is a registered trademark of Velcro Industries B.V. STERRAD and CIDEX are trademarks of Johnson & Johnson. Alconox is a trademark of Alconox Ltd. Neodisher Medizyme is a trademark of Chemische Fabrik Dr Weigert.
© 2007 ResMed Ltd.
Page 3
A
M
i
r
a g
e
L
i
b
e
r
t
y
1
2
Mask parts
Mouth cushion
B
Fitting the mask
Valve hinge
Valve flap
Valve
Cushion alignment guide
Valve clip
Elbow
Inlet tube
Swivel
Mask frame
Ports cap
Nasal pillows
Exhaust vents
1 2
C
1 8
Disassembling for cleaning
2
3 4 5 6 7
3 4 5
© 2007 ResMed Ltd.
© 2007 ResMed Ltd.
Headgear parts
Right upper clip
Lower straps
Lower clip
Upper straps
Left upper clip
Lower clip
© 2007 ResMed Ltd.
Page 4
D
Reassembling the mask
E
Reassembling the headgear
Mirage Liberty
Full Face Mask
Illustrations Sheet
Note: For complete instructions, please read the text section
in conjunction with the illustrations on this sheet.
1 2 3 4
7 8
5
1
96
© 2007 ResMed Ltd.
2
© 2007 ResMed Ltd.
Component of: 61850/2
Page 5
Mirage Liberty™
FULL FACE MASK
Thank you for choosing ResMed’s Mirage Liberty. The Mirage Liberty is a full face mask that combines nasal pillows and a mouth cushion, creating a mask that is both unobtrusive and easy to use without compromising seal, stability and comfort.
The mask incorporates vent holes and a built-in valve so that you can continue to breathe fresh air if the airflow to your mask is impeded for any reason.
Intended Use
The Mirage Liberty channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.
The Mirage Liberty is to be used by adult patients (>30 kg) for whom positive airway pressure has been prescribed.
The Mirage Liberty is intended for single-patient re-use in the home
environment and multipatient re-use in the hospital/institutional environment.
English
Before Using the Mask
WARNING
The Mirage Liberty must be used under qualified supervision with
patients who are unable to remove the mask by themselves. The mask may not be suitable for those predisposed to aspiration.
The elbow and valve assembly have specific safety functions. They
allow you to breathe fresh air if the airflow from your flow generator to your mask is impeded. The mask should not be worn if the valve is damaged as it will not be able to perform its safety function. The valve should be replaced if it is damaged, distorted or torn.
The vent holes must be kept clear.
Follow all precautions when using supplemental oxygen.
1
Page 6
Any supplemental oxygen flow must be turned off when the flow generator is not operating, so that unused oxygen does not accumulate within the flow generator enclosure and create a risk of fire.
At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen
concentration will vary, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate.
The technical specifications of the mask and flow generator are
provided for your clinician to check that they are compatible. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect the flow generator function.
Discontinue using the Mirage Liberty if you have ANY adverse
reaction to the mask system, and consult your physician or sleep therapist.
Discontinue using the Mirage Liberty if you experience nasal irritation,
and consult your physician or sleep therapist.
Refer to your flow generator manual for details on settings and
operational information.
Using the Mask
Select ‘Mir Full’ or ‘Full Face’ as the mask option when using the Mirage Liberty with ResMed flow generators that have mask setting options.
Fitting the mask
The diagrams in section A on the illustrations sheet show how the parts of the mask fit together.
The diagrams in section B show the suggested sequence for obtaining the best fit and comfort in using your mask. Please follow these instructions carefully.
1 Wash your face thoroughly before fitting the mask. It is recommended that
you do not apply moisturiser where the mask contacts your face as it can affect the mask seal.
2 Place the headgear on your head, like you would a cap, and pull the mask
forward onto your face (B-1). Make sure that the headgear fits evenly on the top of your head.
3 Fit the mask over your mouth. Insert the pillows so that they sit gently just
inside the entrance to your nostrils.
4 Push the lower headgear clips into the slots on the mask frame until they
click into place (B-2).
2
Page 7
5 To tighten or loosen the headgear straps, peel the Velcro® tab off the
headgear and reposition it. Gently and evenly adjust the upper headgear straps (B-3), then the lower straps (B-4). Alternate the adjustment of each set of straps until they are comfortable and firm and so that the upper headgear clips rest between your eye and your ear. The mask should gently ‘float’ on your face. Do not overtighten the headgear straps.
6 Connect one end of the main air tubing to your flow generator and the other
end to the swivel.
7 Connect the elbow to the mask frame, holding the sides of the elbow then
firmly pushing it onto the frame until it clicks into place (B-5). Rotate the elbow to check that it rotates freely and does not separate from the frame.
8 Turn your flow generator on. 9 If you feel air leaks, you may need to adjust your headgear. Return to step 5
or check through the suggestions in “Optimising mask seal and comfort” below.
Optimising mask seal and comfort
To receive effective therapy it is important to have the best seal possible between your mask and your face. If you are experiencing leaks, please ensure that you have followed the fitting instructions carefully. Carry out a quick check after fitting your mask:
Is the mouth cushion creased? Creases can allow air to leak out of the
mask.
Have you cleaned your mask recently? Oils on the cushion and pillows can
affect the seal.
Are the pillows correctly seated in your nostrils? Re-seat or adjust the
pillows to obtain a proper seal.
Are your headgear straps overtightened? Try not to alter the upper headgear
straps once they have been tightened by your clinician.
If you cannot establish a comfortable seal, do not overtighten the headgear.
Consult your clinician.
Try another nasal pillow size.
If you cannot resolve any problems, look at the Troubleshooting section, or talk to your clinician.
English
Removing the Mask
To remove the Mirage Liberty:
1 Detach the elbow by squeezing the elbow on both sides (C-1). 2 Unclip both of the lower headgear clips (C-2) and remove the mask.
3
Page 8
Note: You can detach the elbow from the mask frame and leave the mask on your face. You may prefer to do this when briefly stopping therapy, for example to go to the bathroom.
Disassembling the Mask
See section C on the illustrations sheet. 1 Unclip the upper headgear clips from the frame (C-3). 2 Carefully pull the mouth cushion away from the frame (C-4). 3 Push the two nasal pillows down through the mouth cushion (C-5). 4 Remove the ports cap (C-6). 5 Detach the inlet tube assembly from the elbow.
Note: The swivel does not need to be disconnected from the inlet tube.
6 Detach the valve clip from the elbow (C-7). 7 Peel the valve from the valve clip (C-8).
Note: Do not pull on the valve flap, as this may cause it to tear.
Cleaning the Mask in the Home
Notes:
The mask and headgear can only be cleaned by handwashing.
The headgear may be washed without removing the headgear clips.
Do not soak any components of the mask.
WARNING
Do not use aromatic-based solutions or scented oils (eg, eucalyptus or
essential oils), bleach, alcohol or products that smell strongly (eg, citrus) to clean any of the mask components. Residual vapours from these solutions can be inhaled if not rinsed thoroughly. They may also damage the mask, causing cracks.
Do not iron the headgear as the material is heat sensitive and will be
damaged.
Daily/After Each Use
To optimise the mask seal, facial oils should be removed from the cushion and pillows before use.
Handwash the mask by gently rubbing in warm (30°C) water using mild soap or diluted dishwashing detergent.
Clean the walls of the nasal pillows, including the space between the walls, of any debris using a small brush.
Do not twist the valve during cleaning as this may damage or tear the valve membrane or hinge. Take care when handling the valve.
4
Page 9
If the vent requires cleaning, use a soft bristle brush. Rinse all components well with drinking quality water and allow them to air dry
out of direct sunlight.
Weekly
Handwash the headgear by gently rubbing in warm (30°C) water using mild soap or diluted dishwashing detergent. Rinse the headgear well and allow it to air dry out of direct sunlight.
Cleaning the Mask between Patients
The Mirage Liberty should be disinfected or sterilised when used between patients.
Detailed disinfection and sterilisation instructions are available from the ResMed website, www.resmed.com/masks/sterilization, and must be consulted before multipatient use.
Reassembling the Mask
See section D on the illustrations sheet. Before reassembling the mask, check that all components are clean and dry.
Check that the valve is not damaged, distorted or torn, and that none of the other components are broken or cracked. Replace any parts that are damaged. Flexible PVC contains elements that can be detrimental to the materials of the mask. If flexible PVC tubing is used continuously, the attachment port should be inspected for cracking. Replace any parts that are damaged.
Refer to “Ordering Information” for re-order numbers.
1 Attach the nasal pillows to the mouth cushion:
a Push the nasal pillows through the holes in the mouth cushion (D-1). b Pull the nasal pillows up until the grooves at their base fully engage
around the mouth cushion holes (D-2).
c Check that the tabs on each pillow fit into the corresponding notches on
the cushion (D-3). Check that the seal between the pillows and cushion is complete.
2 Attach the mouth cushion to the frame:
a Position the cushion alignment guide between the two nasal pillows and
line up the triangle symbols on the bottom of the mask frame with those on the mouth cushion.
b Press the mouth cushion into the groove in the frame (D-4).
c Check that the mouth cushion is fully seated in the groove. 3 Insert the ports cap (D-5). 4 Clip the upper headgear clips onto the mask (D-6).
English
5
Page 10
5 Assemble the elbow assembly:
a Push the tab of the valve up through the slot in the valve clip (D-7).
Ensure that the valve locks into the slot.
b Feed the flap of the valve through the slot in the elbow (D-8). c Press the valve clip onto the elbow until you hear both sides click into
place (D-9).
6 Push the inlet tube assembly onto the elbow assembly.
Reassembling the Headgear
If taken apart, the headgear has five parts: headgear straps, left and right upper headgear clips, and two lower clips.
1 Attach the upper headgear clips to the mask frame (D-6). The mask
incorporates lock-out features to ensure that each upper clip attaches to the correct side of the mask frame.
2 Clip the lower headgear clips into the mask frame. 3 Lay out the headgear straps, label side down. The label is on the outer side
of the headgear. The internal side is smooth, to be comfortable against your face.
4 Thread the top (short) straps through the upper headgear clips (E-1, E-2). 5 Thread the bottom (long) straps through the lower clips (E-1, E-2). 6 Check that the headgear is not twisted.
Storage
Ensure that the mask is thoroughly clean and dry before storing it for any length of time. Store the mask in a cool dry place out of direct sunlight.
Disposal
The Mirage Liberty does not contain any hazardous substances and may be disposed of with your normal household refuse.
6
Page 11
Troubleshooting
Problem Possible reason Possible solution
Mask won’t seal properly or is uncomfortable
Mask is too noisy Vents are blocked or partially
Mouth cushion or pillows may be dirty.
Mask may have been fitted incorrectly.
Mask is incorrectly adjusted. Check headgear is not over-
Mouth cushion is wrong size. Talk to your clinician. Mouth cushion is skewed in
mask frame. Mouth cushion is misaligned or
creased on face.
Mask pillows are the wrong size. Try another size or talk to your
Mask pillows are not correctly seated in your nostrils.
Mask pillows are skewed in cushion.
Mask is assembled incorrectly. Reassemble mask.
blocked.
Valve is misassembled. Reassemble valve. Cushion is not properly inserted. Check insertion of cushion. Mask is assembled incorrectly. Reassemble mask. Valve is damaged, torn or
distorted. Pillows are not inserted correctly
into cushion. Mask is not sealing. Check and adjust mask as
Clean mask according to instructions in “Cleaning the Mask in the Home”.
Carefully follow instructions in “Fitting the mask”.
tightened.
Check insertion of cushion.
Ensure cushion is properly aligned and that there are no creases in it: undo headge ar clip, check cushion and reposition the mask on your face.
clinician. Re-seat pillows into nostrils.
Check insertion of pillows.
Clean vents according to instructions in “Cleaning the Mask in the Home”.
Replace valve.
Check insertion of pillows.
necessary.
English
7
Page 12
Technical Specifications
Note: The mask system does not contain latex, PVC or DEHP materials.
Pressure–Flow Curve
70.0
60.0
50.0
40.0
30.0
20.0
10.0
Vent Flow Rate (L/min)
0.0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Mask Pressure (cm H2O)
Note: As a result of manufacturing variations, the vent flow may vary from that indicated in the graph.
Dead Space Information
Dead space is the empty volume of the mask up to the elbow. The maximum dead space of the Mirage Liberty for all mask size combinations
is 151 mL.
Standards
CE designation in accordance with EC directive 93/42/EEC, class IIa. ISO 14971, ISO 10993-1, ISO 17510-2, ISO 13485.
Therapy Pressure
4 to 20 cm H2O
Resistance
Drop in pressure measured (nominal) at 50 L/min: 0.5 cm H
O
2
at 100 L/min: 1.6 cm H2O
Sound
The A-weighted sound power of the mask is 28 dBA.
Environmental Conditions
Operating temperature: +5°C to +40°C Operating humidity: 10%–95% relative humidity non-condensing Storage and transport temperature: -20°C to +60°C Storage and transport humidity: 10%–95% relative humidity non-condensing
Pressure (cm H2O)
422 832 12 41 16 48 20 54
Flow (± 6.0 L/min)
8
Page 13
Mask sizes
(excludes inlet tube and headgear)
size width (mm) height (mm) depth (mm)
small 110 92 95 la rge 110 100 100
Symbols
System and Packaging
Caution, consult accompanying documents
Lot number
LOT
Part number/Product code
REF
No latex
Humidity limitation
Temperature limitation
Manufacturer
English
User Guide
Indicates a warning or caution and alerts you to a possible injury or explains special measures for the safe and effective use of the device.
9
Page 14
Ordering Information
Ref. Item
System
61310 Mirage Liberty Full Face Mask, Small – ANZ 61311 Mirage Liberty Full Face Mask, Large – ANZ 61312 Mirage Liberty Full Face Mask, Small – Asia-Pacific 61313 Mirage Liberty Full Face Mask, Large – Asia-Pacific
Replacement parts
61330 Mouth cushion, Small (1 pk) 61331 Mouth cushion, Large (1 pk) 61333 Nasal pillows, Small (1 pk) 61334 Nasal pillows, Medium (1 pk) 61335 Nasal pillows, Large (1 pk) 61337 Mask frame, Small (1 pk) 61338 Mask frame, Large (1 pk) 61340 Elbow assembly (elbow, valve and clip) (1 pk) 61342 Valve and clip (1 pk) 61343 Valve and clip (10 pk) 60926 Inlet tube and swivel (1 pk) 60927 Inlet tube and swivel (10 pk) 61348 Headgear assembly, Small (1 pk) 61349 Headgear assembly, Standard (1 pk) 61352 Upper clips (2 pk) 61353 Lower clips (2 pk) 16570 Ports cap (2 pk) 16571 Ports cap (10 pk) 60014 Swivel (10 pk)
Limited Warranty
ResMed warrants that your ResMed mask system, including its components: mask frame, cushion, headgear, tubing and other accessories (“Product”) shall be free from defects in material and workmanship for a period of ninety (90) days from the date of purchase by the initial consumer.
For further details refer to the Warranty.