ResMed 368870-1 User Manual

S9 Escape
POSITIVE AIRWAY PRESSURE DEVICES
User Guide
English
A
3
2
5
4
1
English
S9 Escape indications for use
The S9 Escape CPAP system is indicated for the treatment of obstructive sleep apnoea (OSA) in patients weighing more than 30 kg. The S9 Escape CPAP system is intended for home and hospital use.
Contraindications
Positive airway pressure therapy may be contraindicated in some patients with the following pre­existing conditions:
severe bullous lung disease
pneumothorax
pathologically low blood pressure
dehydration
cerebrospinal fluid leak, recent cranial surgery, or trauma.
Adverse effects
Patients should report unusual chest pain, severe headache, or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment. The following side effects may arise during the course of therapy with the device:
drying of the nose, mouth, or throat
nosebleed
bloating
ear or sinus discomfort
eye irritation
skin rashes.
Masks and humidifiers
Recommended masks and humidifiers are available on www.resmed.com on the Products page under Service & Support. For information on using your mask or humidifier, refer to the manual supplied with your mask or humidifier.
Setup
Refer to illustration A.
1. Connect the DC plug of the power supply unit to the rear of the S9.
2. Connect the power cord to the power supply unit.
3. Plug the other end of the power cord into the power outlet.
4. Connect one end of the air tubing firmly onto the air outlet.
5. Connect the assembled mask system to the free end of the air tubing.
English
1
Control panel
Refer to illustration B. The control panel of your S9 device includes the following items:
1. Start/Stop button: Starts or stops treatment
2. Home menu
3. LCD screen
4. Info menu*: Allows you to view your sleep statistics
5. Setup menu*: Allows you to make changes to settings
6. Push dial: Turning the dial allows you to scroll through the menu and change settings. Pushing the dial allows you to enter into a menu and confirm your choice.
*The Info and Setup menus are disabled if S9 Essentials has been enabled by your clinician.
Navigating the menus
Refer to illustration C. In general, to navigate the menus:
1. Turn until the parameter you require is displayed in blue.
2. Press . The selection is highlighted in orange.
3. Turn until you see the setting that you require.
4. Press to confirm your choice. The screen returns to blue.
Getting started
1. Make sure the power is connected.
2. Adjust the ramp time if required.
3. Fit your mask as described in your mask user guide.
4. To start therapy, simply press .
5. Lie down and arrange the air tubing so that it is free to move if you turn in your sleep.
6. To stop treatment at any time, press .
Cleaning and maintenance
Daily: Remove the air tubing by pulling on the finger grips on the cuff. Hang it in a clean, dry place until next use.
Notes:
Do not hang the air tubing in direct sunlight as it may harden over time and eventually crack.
Do not wash the air tubing in a washing machine or dishwasher.
2 English
Weekly:
1. Remove the air tubing from the S9 device and the mask.
2. Wash the air tubing in warm water using mild detergent.
3. Rinse thoroughly, hang, and allow to dry.
4. Before next use, reconnect the air tubing to the air outlet and mask.
Monthly:
1. Wipe the exterior of the S9 with a damp cloth and mild detergent.
2. Check the air filter for holes and blockage by dirt or dust. Replace the air filter if necessary.
Replacing the air filter:
Replace the air filter every six months (or more often if necessary).
1. Remove the air filter cover from the back of the S9 device.
2. Remove and discard the old air filter.
3. Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover.
4. Replace the air filter cover.
Notes:
Ensure the air filter and air filter cover are fitted at all times.
Do not wash the air filter. The air filter is not washable or reusable.
Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.
Problem/Possible cause Solution
No display
Power is not connected. Ensure the power cable is connected and the power outlet
The DC plug is partially inserted into the back of the device.
Insufficient air delivered from the device
Ramp time is in use. Wait for air pressure to build up or change ramp time. Air filter is dirty. Replace air filter. Air tubing is not connected properly. Check air tubing. Air tubing is blocked, pinched or
punctured.
(if available) is on. Fully insert the DC plug.
Unblock or free the air tubing. Check the air tubing for punctures.
English
3
Loading...
+ 9 hidden pages