renishaw PHS User Manual

Instruction leafl et
H-1000-4055-03-A
PHS temperature sensing PC card
PHS temperature sensing PC card
The PHS1 head has been equipped with a head temperature sensing facility to allow thermal compensation. Previously, the PC
card was unable to read this information but a new card has been designed to accommodate this facility.
Identifi cation of new card
The PC card level may be identifi ed by looking at the PCB part number (different to the card assembly) on the component free
NUMBER. The optional fi lter module should not be used with the new card (as the fi lter module is already incorporated onto the
card).
‘Old’ non-temperature PC card:
M-2150-0501-XX
M-2150-0501-XX
Optional fi lter module:
‘New’ non-temperature PC card:
M-2150-0501-XX
or
M-2150-1341-XX
M-2150-0501-XX
or
M-2150-1341-XX
A-2150-1097
NOTE: NOT to be used with NEW card.
Figure 1 - PHS PC card identifi cation
Renishaw plc
New Mills, Wotton-under-Edge, Gloucestershire, GL12 8JR United Kingdom
T +44 (0)1453 524524 F +44 (0)1453 524901 E uk@renishaw.com
www.renishaw.com
Switch settings
Due to the added functionality of the new card, the SYNC EDGE and CONTROLLER SELECT switches have been moved
(see fi gure 2). Switch banks 1, 2 and 3 (card address select) remain unchanged but switch 4 changes as follows. Therefore it is
important that you do not copy the DIP switch settings one to one.
PC card - OLD (0551) PC card - NEW (0881) and (1341)
SW1-1 Card address select - as before Card address select - as before
SW1-2
SW1-3
SW2-1
SW2-2
SW2-3
SW3-1
SW3-2
SW3-3 Not used Not used
SW4-1 Sync edge off (Left) = Rising edge
On (Right) = Falling edge
SW4-2 Controller select
Off (Left) = IBM-PC Bus
On (Right) = DELTA- TAU
SW4-3 Not used Controller select
Sync edge off (Left) = Rising edge;
Not used
On (Right) = Falling Edge.
Off (Left) = IBM-PC Bus
On (Right) = DELTA- TAU
SW1
SW2
SW3
SW4
1
2
3
OFF ON
Figure 2 - PC card switches (0881 / 1341 card shown)