renishaw PH10M-iQ PLUS User Manual

Renishaw plc
New Mills, Wotton-under-Edge, Gloucestershire GL12 8JR United Kingdom
T +44 (0)1453 524524 F +44 (0)1453 524901 E uk@renishaw.com
www.renishaw.com
PH10M-iQ PLUS
Less time calibrating, more time measuring
PH10M-iQ PLUS offers all the benefits of a traditional PH10M
PLUS but with the addition of inferred qualification. This further
demonstrates Renishaw’s commitment to continuous product
development.
to qualify each head position that is used. This allows more
time to be spent measuring.
Following an initial head localisation, a simple qualification
procedure of as few as three positions allows the user to
operate PH10M-iQ PLUS in every orientation possible without
requalifying.
Innovations
Two heads in one
PH10M-iQ PLUS provides all the functionality of Renishaw’s
industry standard PH10M PLUS but with the addition of
inferred qualification for touch-trigger probing routines.
Where an application or specific angle demands ultimate
touch-trigger accuracy then PH10M-iQ PLUS can be qualified
in the same way as a standard PH10M PLUS.
Fine tune your performance
Machine
type
Mounting
orientation
Qualification
positions
3 95 µm
8 125 µm
12 80 µm
18 65 µm
3 15 µm
PLTI - Typical
positional span*
PH10M-iQ PLUS must perform a simple qualification
procedure before inferred mode can be used, after which
any head position can be used for measurement without
having to requalify.
Depending on your metrology requirements you can
optimise the accuracy of your system by increasing the
number of qualification positions during probe qualification.
Horizontal arm machine specification
Using a PAA3 300 mm extension bar, standard force TP20 module, 10 mm x 4 mm diameter stylus on a machine with the following specification;
MPEe = ± (9 + L / 100) µm (L in mm) ISO 10360-2 (2009)
Bridge machine specification
Using a PAA1 adaptor, standard force TP20 module, 10 mm x 4 mm diameter stylus on a machine with the following specification;
MPEe = ± (1 + L / 750) µm (L in mm) ISO 10360-2 (2009)
*Positional span of sphere location (PLTI) evaluated as per ISO 10360-5 (2010) but covering all 720 head positions
Compatibility
Renishaw’s complete range of PH10M PLUS compatible probes are also compatible with PH10M-iQ PLUS. However, the
added functionality of inferred qualification can only be used with touch-trigger probes.
TP20, TP2-5W, TP200, TP6, TP6A and TP7M touch-trigger probes can be used in traditional PH10 mode or can utilise
inferred qualification; whereas SP25M and SP600M scanning probes can only be used in traditional PH10 mode and cannot
utilise inferred qualification.
TP20 TP2-5W TP200 TP6 TP6A TP7M
Specification
Head
Length
Width
Weight
Mounting
Probe mount
Controller
Repeatability
Angular movement
Total number of positions
Maximum drive output torque
Maximum
extension bar
SP25M
PH10M-iQ PLUS
117 mm (4.60 in)
62 mm (2.44 in)
645 g (22.07 oz)
Shank
Renishaw Autojoint (multiwire)
M8 threaded probes can be used with a PAA probe adaptor
PHC10-3 PLUS
0.4 µm (2σ 0.00002 in) specified at a distance of 62 mm (2.44 in) from the A-axis centre
of rotation
A-axis 0° to 105° in 7.5° steps
B-axis -180° to 180° in 7.5° steps
720 positions
0.45 Nm
300 mm (11.8 in) using PAA3 probe adaptor
450 mm (17.7 in) using PAACF special order only
extension
SP600M
300 mm (11.8 in)
using PEL4
extension
Temperature range
Operating
Storage
Head control unit
Warranty
10 °C to 40 °C (50 °F to 104 °F)
-10 °C to 70 °C (14 °F to 158 °F)
HCU1 or MCUlite-2, MCU5 or MCU W
2 years
For worldwide contact details, please visit our main website at www.renishaw.com/contact
RENISHAW HAS MADE CONSIDERABLE EFFORTS TO ENSURE THE CONTENT OF THIS DOCUMENT IS CORRECT AT THE DATE OF PUBLICATION BUT MAKES NO WARRANTIES OR REPRESENTATIONS REGARDING THE CONTENT. RENISHAW EXCLUDES LIABILITY, HOWSOEVER ARISING, FOR ANY INACCURACIES IN THIS DOCUMENT.
RENISHAW® and the probe emblem used in the RENISHAW logo are registered trademarks of Renishaw plc in the UK and other countries. apply innovation is a trademark of Renishaw plc.
© 2009 - 2013 Renishaw plc All rights reserved Issued 1013
Part no. H-1000-1272-02-A