Renishaw MCR20 User Instruction

Renishaw plc
New Mills, Wotton-under-Edge, Gloucestersh ire, GL12 8JR, United Kingdom. Telephone: +44 (0) 1453 5245 24 Fax: +44 (0) 1453 524901 email uk@renishaw.com Internet http://www.renishaw.com
Renishaw Inc, USA Renishaw (Hong Kong) Ltd
Tel +1 847 843 3666 Hong Kong Fax +1 847 843 1744 Tel +852 2753 0638 email usa@renishaw.com Fax +852 2756 8786
email hongkon g@renishaw. com
Renishaw KK, Japan
Tel +81 3 5332 6021 Renishaw Metrology Systems Fax +81 3 5332 602 5 Private Limited, India email japan@renishaw.com Tel +91 80 5320 144
Fax +91 80 5320 140 Renishaw GmbH, Germany email india@renishaw.com Tel +49 7127 9810 Fax +19 7127 88237 Renishaw’s Representative Offices email germany@renishaw.com
Austra lia Renishaw SA, France Tel +61 3 9553 8267
Tel +33 1 64 61 84 84 Fax +61 3 9552 6738 Fax +33 1 64 61 65 26 email aust ralia@renisha w.com email france@renishaw.com
Indonesia Renishaw SpA, Italy Tel +65 438 2779
Tel +39 011 9 66 10 52 Fax +65 438 2780 Fax +39 011 9 66 40 83 email indon esia@renisha w.com email italy@renishaw.com
Singapore Renishaw Ibe ricia S. A, Spain Tel +65 438 2779
Tel +34 93 478 21 31 Fax +65 438 2780 Fax +34 93 478 16 08 email singapore@renishaw.com email spain@renishaw.com
Taiwan Renishaw A.G., Switzerland Tel +886 4 2513665
Tel +41 55 410 66 66 Fax +886 4 2513621 Fax +41 55 410 66 69 email taiwan@renishaw.com email switzerland@renishaw.com
The People’s Republic of China Renishaw Latino Americia Ltda, Tel +86 10 641 07993 Brazil Fax +86 10 641 07992
Tel +55 11 4195 2866 email china@renishaw.com Fax +55 11 4195 1641 email brazil@renishaw.com Renishaw’s Liaison Office
South Korea
Tel +82 2 565 68 78
Fax +82 2 565 6879
email southkorea@renishaw.com
=>?@D
www.renishaw.com
MCR20 ENDCAP
REPLACEMENT KIT
)>?@D
User’s Instruction
INTRODUCTION
This User Instruction leaflet details the procedure to change the MCR20 end caps. There has been a design modification to the rack to ensure that ports 1 and 6 reliably inhibit the TP20 during a module change. Some early design racks may need these replacement end caps to improve probe inhibit in the change cycle.
Caution
Renishaw recommends that the CMM is
stationary during this procedure
Procedure
(Please refer to figure 1)
1. Remove screws located in A and B using the 2mm A/F allen key
provided.
2. Remove the end cap from the MCR20 rack by sliding the cap away
from the ports.
3. Re-fit the modified MCR20 end cap supplied.
4. Fit the M3 x 16mm screws into location A and B using the 2mm A/F
allen key.
Where possible, this operation should be completed without removing the MCR20 from its location on the CMM’s bed. This prevents the need to re­datum the rack for automatic stylus change. If it is necessary to remove the MCR20 to complete this operation, please refer to your CMM supplier for instructions to re-datum the rack.
Please refer to the TP20 and MCR20 Installation and User’s Guide (part no. H-1000-5008) for more information about using the TP20 probe system.
Renishaw reserves the right to change specifications without notice
Remove cap
End cap
A
B
Figure 1 – MCR20 viewed from underneath