Remote Automation Solutions Guide: TechView User's Guide Manuals & Guides

s
User Manual
Document: D5131 Part: D301430X012 November 2016
TechView User’s Guide
OpenBSI Version 5.9
Remote Automa ti on Solution
www.EmersonProcess.com/Remote
Application Safety Considerations
Protecting Operating Processes
A failure of this application – for whatever reason -- may leave an operating process without appropriate protection and could result in possible damage to property or injury to persons. To protect against this, you should review the need for additional backup equipment or provide alternate means of protection (such as alarm devices, output limiting, fail-safe valves, relief valves, emergency shutoffs, emergency switches, etc.)
CAUTION
When implementing control using this product, observe best industry practices as suggested by applicable and appropriate environmental, health, and safety organizations. While this product can be used as a safety component in a system, it is NOT intended or designed to be the ONLY safety mechanism in that system.
Contents
TechView User’s Guide
Chapter 1 – Introduction - What is TechView? 1-1
Chapter 2 – TechView Installation 2-1
Chapter 3 – Starting TechView and Establishing Communications 3-1
3.1Starting Your Session: .............................................................................................................. 3-1
3.2Establishing Communications ................................................................................................... 3-2
3.2.1Communication Setup page - BSAP ............................................................................. 3-2
3.2.2Communication Setup page - IP ................................................................................... 3-3
3.2.3Advanced Communication Parameters – BSAP ........................................................... 3-4
3.2.4Advanced Communication Parameters – IP ................................................................. 3-7
3.2.5Node Setup ................................................................................................................... 3-8
3.2.6Calibration Setup ......................................................................................................... 3-10
3.2.7Transmitter Setup (Only Visible if using Remote Access) .......................................... 3-13
3.3Signing On to the RTU ............................................................................................................ 3-15
Chapter 4 – Overview of TechView Functions 4-1
4.1Changing the Session Parameters You Entered ...................................................................... 4-2
4.2Restarting the Session .............................................................................................................. 4-3
4.3Saving Your Session File .......................................................................................................... 4-3
Copying Your Session to Another Session File Name ............................................................. 4-3
4.4Application Settings ................................................................................................................... 4-3
4.4.1Startup tab ..................................................................................................................... 4-4
4.4.2Security tab ................................................................................................................... 4-7
4.4.3General tab ................................................................................................................... 4-7
4.5Calibration and Verification Signals .......................................................................................... 4-8
4.5.1Extended Verification .................................................................................................... 4-9
4.5.2Changing the Name of a Transmitter .......................................................................... 4-11
4.6Changing the Local Address / Group Number ........................................................................ 4-11
4.6.1Starting the Flash Configuration Utility ........................................................................ 4-12
4.6.2Show Firmware Version in Node ................................................................................ 4-12
4.6.3Writing an Audit Note .................................................................................................. 4-12
4.6.4Viewing OpenBSI Workstation Communication Statistics .......................................... 4-13
4.6.5Deleting Historical Files from the RTU (Clear History) ............................................... 4-13
Chapter 5 – Calibration Overview 5-1
5.1Why is Calibration Necessary? ................................................................................................. 5-1
5.2Calibration Concepts ................................................................................................................. 5-1
5.3Before You Begin ...................................................................................................................... 5-3
5.4Equipment Required for Calibration .......................................................................................... 5-5
5.4.1Equipment Required For Pressure Calibration ............................................................. 5-5
5.4.2Equipment Required For Temperature Calibration ....................................................... 5-7
5.4.3Equipment Required for Analog Output Calibration (3808 only) ................................... 5-7
5.4.4Entering Calibration Mode / Leaving Calibration Mode ................................................ 5-7
Chapter 6 – Calibrating the 3508 series TeleTrans Transmitter 6-1
6.1Calibration of Differential/Gage Pressure ................................................................................. 6-1
Issued Nov-2016 Contents iii
TechView User’s Guide
6.2Calibration of Static Pressure ................................................................................................... 6-3
6.3Calibration of RTD Temperature ............................................................................................... 6-5
Chapter 7 – Calibrating the 3808 Transmitter 7-1
7.1Calibration of Gage or Differential Pressure (3808-10A/3808-30A only) .................................. 7-1
7.2Calibration of Static Pressure (3808-30A ONLY) ..................................................................... 7-2
7.3Calibration of RTD Temperature ............................................................................................... 7-3
7.4Calibrating the Analog Output (Analog Pressure Transducer models ONLY) ......................... 7-5
Chapter 8 – Calibrating the 3530 TeleFlow-series Flow Computers 8-1
8.1Calibration of Differential/Gage Pressure ................................................................................. 8-1
8.2Verification of Differential Pressure ........................................................................................... 8-2
8.3Calibration of Static Pressure ................................................................................................... 8-2
8.4Verification of Static Pressure ................................................................................................... 8-3
8.5Calibration of RTD Temperature ............................................................................................... 8-4
8.5.1Verification of RTD Temperature .................................................................................. 8-6
8.6Damping .................................................................................................................................... 8-6
8.6.1Configuring the Damping Time ..................................................................................... 8-7
8.7Calibrating TeleRecorder Inputs (3530-45B/55B ONLY) .......................................................... 8-7
8.7.1Calibrating the Input ...................................................................................................... 8-8
Chapter 9 – Calibrating the ControlWave EFM / GFC-CL /GFC/ XFC 9-1
9.1Calibration of Differential/Gage Pressure ................................................................................. 9-1
9.2Verification of Differential/Gage Pressure ................................................................................. 9-2
9.3Calibration of Static Pressure ................................................................................................... 9-4
9.4Verification of Static Pressure ................................................................................................... 9-5
9.5Calibration of RTD Temperature ............................................................................................... 9-6
9.6Verification of Temperature ....................................................................................................... 9-9
9.7Damping .................................................................................................................................. 9-11
9.7.1Configuring the Damping Time ................................................................................... 9-11
9.8Orifice ...................................................................................................................................... 9-12
9.8.1Specifying the Orifice Plate Size ................................................................................. 9-12
Chapter 10 – Configuring and Calibrating the 4088B 10-1
10.1Configuring the 4088B ............................................................................................................ 10-1
10.1.1DP/GP Pressure .......................................................................................................... 10-2
10.1.2Static Pressure ............................................................................................................ 10-2
10.1.3Temperature ................................................................................................................ 10-3
10.1.4Serial Port Setup ......................................................................................................... 10-4
10.1.5Transmitter Data ......................................................................................................... 10-5
10.1.6RTD Coefficients ......................................................................................................... 10-6
10.2Calibrating the 4088B .............................................................................................................. 10-6
10.2.1Calibration of Gage or Differential Pressure ............................................................... 10-7
10.2.2Calibration of Static Pressure ..................................................................................... 10-7
10.2.3Calibration of RTD Temperature ................................................................................. 10-8
Chapter 11 – Online Editing (ControlWave only) 11-1
11.1On-line Editing of Signal Lists ................................................................................................. 11-1
11.2On-line Editing of Archive Files ............................................................................................... 11-4
11.2.1Archive Fields .............................................................................................................. 11-6
11.2.2Batch Editing of Archive Files ..................................................................................... 11-7
Chapter 12 – Device and Measurement Groups 12-1
iv Contents Issued Nov-2016
TechView User’s Guide
Appendix A -Initialization Files A-1
Creating an Advanced Interface Setup File for Storage at the RTU ................................................. A-2
To create an Advanced Interface Setup Zip File for a custom application: ...................................... A-2
AISF.TVS File Format ....................................................................................................................... A-3
Configuration Initialization (*.INI) Files (Platform-dependent) ........................................................... A-5
LISTS.INI ......................................................................................................................................... A-15
TRANSLATION.INI Initialization File ..................................................................................... A-16
Notes about IP Address Formats in TVS Files ...................................................................... A-19
Index IND-1
Issued Nov-2016 Contents v
This page is intentionally left blank
TechView User’s Guide (D5131)
Chapter 1 – Introduction - What is TechView?
TechView is a standalone software package that allows a technician to: Calibrate transmitter(s) either locally (bench configuration) or via
communication with the transmitter’s master controller (RTU).
Perform certain basic configuration operations on a controller, flow
computer, or transmitter, such as changing the local address or EBSAP group number.
Collect real-time or historical data from the controller, flow
computer, or transmitter.
Start other OpenBSI programs or utilities to perform other activities,
such as the Flash Configuration utility.
Perform on-line edits to signal lists, and to the structure of archive
files in ControlWave-series controllers. (Requires 4.60 or newer ControlWave firmware.)
Typically, you install TechView software on a portable laptop computer to allow the technician to bring it to the site of the RTU/transmitter. You could also install it on a desktop computer in a lab for bench configuration of a transmitter.
You can use TechView to calibrate the RTD temperature circuitry and /or the MVT/GPT pressure transducer for the following products:
3508 TeleTrans Transmitter 3808 MVT Transmitter (Firmware Version 1.5 or newer) 3530-10B TeleFlow Electronic Gas Measurement Computer 3530-20B TeleFlow Plus 3530-45B TeleRecorder 3530-50B TeleFlow Corrector 3530-55B TeleRecorder Rosemount 4088B Transmitter ControlWave EFM ControlWave GFC ControlWave XFC
Issued Nov-2016 Introduction 1-1
Chapter 2 – TechView Installation
Note: For minimum system requirements and more detailed installation
instructions, refer to Chapter 2 of the OpenBSI Utilities Manual (D5081).
1. Log into the workstation with administrative privileges.
2. Insert the OpenBSI CD-ROM in your CD-ROM drive.
3. If your CD-ROM drive has autorun enabled, skip to the next step.
Otherwise, open a DOS prompt with administrative privileges, then set the folder to the CD root directory, and run the same “open” command that is in the autorun.inf file on the CD, for example “demo32.exe openbsi.dbd”. When the CD browser screen appears, choose the Install OpenBSI option.
A screen reminds you to close all other programs, and warns you that older OpenBSI versions will be removed. Click Next.
TechView User’s Guide (D5131)
4. A license agreement screen opens. Review the agreement, using the
scroll bar to bring it into view. Click I accept the terms of the license agreement to proceed. Then click Next.
5. On the next screen, you can specify the user files folder. The system
stores user files such as ControlWave projects, ACCOL files, recipes, and network files in sub-folders of this folder. Note: If User Account Control (UAC) is enabled in Windows, you must have write access to this folder. The default is C:\ProgramData\Bristol\Openbsi. Use the default or use Browse to specify a different user files folder. When finished click Next.
6. On the next screen, use the check-boxes to select either Network
Edition (if you purchased the full Network Edition) or BSI_Config if you are using the free BSI_Config package. These are the packages which include TechView.
7. Once you make your choices, click Next.
8. This is your last opportunity to make any changes prior to starting
the installation. If you want to make changes, you can click Back to go back to earlier pages. If you want to read the printed release notes for this version of OpenBSI, check the View the Release Notes box.
9. If you are ready to perform the installation, click Install, and the
installation process starts. Be patient, as it may take several minutes to install all of the different utilities, depending upon which you choose.
10. When the installation completes, re-boot your computer when
prompted. This must be done in order for OpenBSI to function properly. If you choose not to re-boot now, you must do so before
Issued Nov-2016 Installation 2-1
TechView User’s Guide (D5131)
After re-boot, an “OpenBSI Tools” menu selection is added to your Windows Start Programs menu through which you can access the various OpenBSI utilities. If you prefer, you can create Windows™ shortcuts to the tools to provide access through icons on the desktop. See your Windows™ documentation for information on how to do this.
running OpenBSI. Click Finish, and the installation will be complete, and re-boot will proceed, if you chose to do it now.
2-2 Installation Issued Nov-2016
TechView User’s Guide (D5131)
Chapter 3 – Starting TechView and Establishing
Communications
Note: Neither LocalView nor NetView can be running on your
computer at the same time you are running TechView. If either of those programs are running, you must shut them down, first.
Click Start > Programs > OpenBSI Tools > TechView
In This Chapter
3.1Starting Your Session: ..................................................................... 3-1
3.2Establishing Communications .......................................................... 3-2
3.2.1Communication Setup page - BSAP. ................................... 3-2
3.2.2Communication Setup page - IP .......................................... 3-3
3.2.3Advanced Communication Parameters – BSAP .................. 3-4
3.2.4Advanced Communication Parameters – IP ........................ 3-7
3.2.5Node Setup ........................................................................... 3-8
3.2.6Calibration Setup ................................................................ 3-10
3.2.7Transmitter Setup (Only Visible if using Remote Access) . 3-13
3.3Signing On to the RTU ................................................................... 3-15
3.1 Starting Your Session:
The configuration information you specify to allow TechView to communicate with controllers, flow computers and transmitters is stored in a session file. Session files have a file extension of *.TVS. .
Once you have a session file saved, you can start TechView by standard Windows™ methods such as typing ‘TechView’ on the command line, followed by the path and name of the session file (excluding the TVS extension), or you can drag a TVS file icon onto a shortcut icon for the TechView application.
Note: If you use the command line option, put quotation marks around
the path or filename if it includes spaces.
Issued Nov-2016 Starting TechView /Establishing Communications 3-1
TechView User’s Guide (D5131)
When first starting TechView, by default, the software prompts you to open an existing session file.
Figure 3-1. Opening a Session File
Note: You can change the initial prompt for the user from the
Application Settings dialog box.
Select the session file name you want to open, then click Open.

3.2 Establishing Communications

You can establish communication with all RTUs/transmitters using the BSAP protocol. For some RTUs you may also use Internet (IP) protocol.
On the Communication Setup page, choose either BSAP or IP in the Select Communication Protocol field.
3.2.1 Communication Setup page - BSAP
For BSAP communication, specify the BSAP local address of the RTU or transmitter:
If you know the BSAP local address, select the Specify Local Address button, and choose the address using the list box, or enter it directly.
If you want the user to specify the local address at startup, select Ask for Local Address at startup; the user will be prompted to enter the address.
If you want TechView at startup to attempt to find the local address by sequentially polling each possible local address (1-127) at the current baud rate, select Auto Detection.
3-2 Starting TechView / Establishing Communications Issued Nov-2016
TechView User’s Guide (D5131)
Most users don’t need to adjust these advanced parameters.
Choose “BSAP” protocol.
Specify the local address of the RTU / transmitter.
Select the PC communication port.
Select the baud rate.
Click on [Next>].
Figure 3-2.Communication Setup dialog box (BSAP)
Specify the PC communication port, e.g. COM1, COM2, etc. in the What port would you like to use field.
Specify the baud rate used by the port in the What baud rate would you like to use field. If you are making an FSK connection to a 4088B, this must be 1200.
Most users don’t need to adjust the advanced parameters, but if you need to, click on Advanced Communication Parameters. (Information on this dialog box is included later in this section.)
Click on Next to proceed to the next page.
3.2.2 Communication Setup page - IP
For IP communication, you must specify the IP address of the RTU. If you want the user to specify the IP address at startup, select Ask for
IP Address at startup; the user will be prompted to enter the address. Most users don’t need to adjust the advanced parameters, but if you
need to, click Advanced Communication Parameters. (Information on this dialog box is included later in this section.)
Click Next to proceed to the next page.
Issued Nov-2016 Starting TechView /Establishing Communications 3-3
TechView User’s Guide (D5131)
Specify the IP address of the RTU.
Choose Internet Protocol (IP).
Most users don’t need to adjust these advanced parameters.
Figure 3-3. Communication Setup dialog box (IP)
3.2.3 Advanced Communication Parameters – BSAP
Most users should not edit these parameters. If however, you are having certain communication problems, you may edit these parameters to better tune your system.
Click on [Next>].
Figure 3-4. Advanced Communication Parameters dialog box (BSAP)
3-4 Starting TechView / Establishing Communications Issued Nov-2016
TechView User’s Guide (D5131)
Field Description Poll Period
Variable
Front Pad, Back Pad
This is the rate (in seconds) in which TechView polls the directly connected RTU.
This is the amount of time, in seconds, that TechView waits for a response from an RTU or transmitter before assuming that the unit is dead and will not respond.
These fields specify the number of null characters to insert at the beginning (front) or ending (back) of a message. Null characters may be useful in situations where there may be a momentary delay which could cause the start of a message to be missed, for example, while a radio link is being activated. Null characters are also necessary if you are communicating using a 2-wire RS-485 link, to ensure that DTR is not dropped prematurely. To determine the delay caused by null packing, perform the following calculation:
seconds of delay = (number of null characters x 10) / baud rate
Use RTS/CTS Signals
Disable the sending of the Time Synch
Dialup Line / Dial Parameters
On an FSK connection, use “3” back pad for model 4088B. Use “2” or “3” Back Pad for models 3508 and 3808.
If your communication line uses Ready to Send (RTS) / Clear to Send (CTS) signals (not to be confused with ACCOL signals), check this box.
Check this box to prevent TechView from sending a BSAP time synch message. (Requires OpenBSI 5.8 Service Pack 1 or newer.) The default setting for this is set in the Advanced Configuration utility. See Appendix E of the OpenBSI Utilities Manual (D5081) for details on the Advanced Configuration utility.
If using a dial line to communicate with the controller, check this box, then click the Dial Parameters button, and specify the dialing parameters.
Issued Nov-2016 Starting TechView /Establishing Communications 3-5
TechView User’s Guide (D5131)
Figure 3-5. Dial Parameters
Field
Dial-up Parameters:
Enter modem commands and phone number to be dialed
Retries
Timeout
Command Delay
Init String
Hang-up parameters:
Description
Enter any modem commands, as well as the dial string here.
This is the number of attempts TechView makes to dial a controller (RTU), before declaring that it is off-line. This must be an integer from 0 to 10.
This is the period of time (in seconds) TechView waits before declaring a dialing attempt to be a failure.
This is the period of time (in seconds) TechView waits between sending commands.
This is an initialization string for the modem. TechView immediately precedes the initialization string with the “AT” modem command. TechView sends the initialization string ahead of the dial-up string.
String1
3-6 Starting TechView / Establishing Communications Issued Nov-2016
This is the first string TechView sends to the modem when an attempt is made to hang up.
TechView User’s Guide (D5131)
String2
Retries
Timeout
No Data Timeout
DTR Support"
This is the second string TechView sends to the modem when an attempt is made to hang up.
This is the number of attempts which will be made to hang up the modem before declaring a hang-up failure. This value must range from 0 to 10.
This is the amount of time TechView waits before declaring an attempt to hang up to be a failure.
This value specifies the amount of time (in seconds) after which the modem should be hung up if there has been no communication (other than poll messages) between TechView and the controller.
If you check this box, TechView will drop DTR in order to hang up, before sending the hang-up strings, and TechView will raise DTR before dialing.
Click OK to save the dial-up and hang-up parameters.
3.2.4 Advanced Communication Parameters – IP
Most users should not edit these parameters. If however, you are having certain communication problems, you can edit these parameters to better tune your system.
Figure 3-6. Advanced Communication Parameters dialog box (IP)
Issued Nov-2016 Starting TechView /Establishing Communications 3-7
TechView User’s Guide (D5131)
Field UDP Port for the IP Driver
UDP Port for Time Synchs
Description
The UDP Port Number for IP Driver (sometimes known as the socket number) is used for communication with RTUs. It is used to split message traffic along different 'streams'. All PCs or RTUs which are to communicate with each other must have the same UDP Port Number for IP Driver. In a sense, this value is like a common password which must be known by each node in the network. If no value is entered, TechView assigns a default value for use throughout the network when the system is first created. (Note: Although the term “UDP port” is used, it has no actual relationship with the physical communication ports.) Note: The UDP Port Number for IP Driver should never be 0.
The UDP Port Number for Time Synch is used for time
synchronization of the RTUs. All PCs or RTUs must have
this value defined, or else they will be unable to receive
time synchronization messages. In a sense, this value is
like a common password which must be known by each
node in the network. If no value is entered, TechView
assigns a default value for use throughout the network
when the system is first created. (Note: Although the term
“UDP port” is used, it has no actual relationship with the
physical communication ports.) Note: The UDP Port
Number for Time Synch should never be 0.
Frequency of Time Sy nchs
Disable the sending of the Time Synch
3.2.5 Node Setup
Next, you must identify certain characteristics of the unit to which you are trying to communicate. The appearance of the Node Setup page varies somewhat, depending upon what type of RTU you are configuring:
This value specifies (in seconds) how often time
synchronization messages should be sent to IP RTUs.
Check this if you want to disable the transmission of time
synch messages.
3-8 Starting TechView / Establishing Communications Issued Nov-2016
TechView User’s Guide (D5131)
Figure 3-7. Node Setup dialog box (appearance varies based on node type)
Field Node Type
Description
The type of controller, RTU, flow computer, or transmitter. Valid choices include:
3305 RTU 3305 controller 3310 RTU 3310 controller 3330 DPC 3330 controller 3335 DPC 3335 controller 3508 Model 3508 TeleTrans transmitter 3530 3530-series TeleFlow / TeleRTU
TeleCorrector/TeleRecorder
3808 Model 3808 transmitter 4088B Rosemount 4088B transmitter ControlWave ControlWave Process Automation
Controller CW_10 ControlWave_10 controller CW_30 ControlWave_30 controller CW_35 ControlWave_35 roller CWave_EFM ControlWave Electronic Flow Meter Cwave-Exp ControlWave Express Gas Flow Computer Cwave_GFC ControlWave Gas Flow Computer ControlWave_LP ControlWave Low Power Controller CWave_Micro ControlWave Micro Controller CWave_XFC ControlWave Explosion-Proof Flow
PT Transmitter
Issued Nov-2016 Starting TechView /Establishing Communications 3-9
When checked indicates that communication will be with a gage pressure transmitter. There will be no static pressure. When not checked, indicates that the transmitter has both differential and static
Computer (3820)
TechView User’s Guide (D5131)
pressures. This applies only to 3508/3808 units.
Wet Ends
Perform Horn in operation
Control Strategy Resource
Load Type
Control Strategy file name
Certain RTUs (EFM, GFC, XFC) may include a built-in internal transmitter with sensor (wet end). Some versions of the XFC can include 2 wet ends. Specify the number of wet ends for this RTU.
If checked, the transmitter’s communication with the BBTI board will be temporarily suspended, so that a different program can communicate with the transmitter.
If this ControlWave project has multiple resources, select the resource which is currently running in the unit.
(Not shown) If communication is with a 3530 TeleFlow-series unit, specify which of the standard application ACCOL loads (or C load) is running in the unit. If needed, you can then use the Calibration Signal Configuration button to specify different signal names for the configuration signals associated with the load. Note: If the load type you specify here does not match the load type running in the controller, TechView prompts you to change the load type so that it does match the one running in the controller.
You can specify a control strategy to be used in conjunction with the RTU used in this application. For the 33XX line of RTUs, this is the name of the ACCOL load running in the RTU. For the ControlWave line of RTUs, this is the name of the .MWT project running in the RTU. Click Browse to locate the file. Note: 3508 and 3808 Transmitters don’t have a control strategy file.
Web Access Startup Page
Advanced Interface Setup
You can optionally specify a default web page for this RTU, which you may start while you are in TechView. Click Browse if necessary, to locate it.
Most users use the standard configuration INI files that come with TechView. Some advanced users, however, may want to customize the organization of pages within TechView. This button calls up a dialog box allowing you to identify special configuration files for TechView. For details on this subject, please see Appendix A of this manual.
3.2.6 Calibration Setup
This page specifies certain parameters that tell TechView how it will access the transmitters for calibration.
3-10 Starting TechView / Establishing Communications Issued Nov-2016
TechView User’s Guide (D5131)
Figure 3-8. Calibration Setup dialog box
Field Target Node
Select Transmitter Access Local / Remote
How many transmitters does the application load support (Wet ends and externals)?
Description
Displays the type of node chosen in the Node Setup dialog box.
This field only applies for BSAP communication connections.
For 3508 and 3808 series transmitters, checking “Local” means that TechView will communicate locally with a transmitter directly connected to the PC, on Level 1 of the temporary BSAP network. For Network 3000 (except for the 3530-series) checking the “Remote” box means that TechView will communicate with an RTU directly connected to the PC, on level 1 of the BSAP network. This RTU has one or more transmitters connected to it that will be on Level 2 of the temporary BSAP network. Note: The 3530-series does not support Remote access.
Note: In Remote Access, if your Network 3000 RTU is not currently part of a network, i.e. you are working in the lab, or performing bench testing, you should connect through the Slave Port. Conversely, if you are working in the field, with an RTU that is already part of a network, you should connect through the Pseudo Slave Port.
Note: If you select Remote the Transmitter Setup page will be added to your Session configuration.
If using the ControlWave Express/EFM/GFC/XFC series, with multiple transmitters, the application running in the ControlWave EFM/GFC/XFC must support multiple meter runs and TechView configures the transmitter(s) by setting parameters within the RTU’s multi-run application; and calibration success/failure will be noted as an audit record in the multi-run application. You must specify the total number of transmitters (both internal wet ends, and external transmitters) here.
Issued Nov-2016 Starting TechView /Establishing Communications 3-11
TechView User’s Guide (D5131)
Disable Damping
Do not ask user’s permission when freezing/unfreezin g live values.
Enable “As Found/As Left” Verification Functionality
Calibration Signal Configuration
Extended Verification Signal Configuration
As Found / As Left Verification Signal Configuration
Note: If you want to use TechView to communicate with
an RTU that doesn’t require any calibration, you can hide calibration options by setting this to 0. (OpenBSI
5.8 Service Pack 1 and newer.) If your ControlWave RTU has an internal transmitter that
doesn’t require damping, you can disable damping functions by selecting this box. (OpenBSI 5.8 Service Pack 1 and newer.)
When the user enters calibration mode, live values for differential pressure, static pressure, gage pressure, and temperature are frozen (i.e. the RTU will collect no new values from the transmitter, during calibration mode.) A prompt appears warning the user of this, and giving them the option of not entering calibration mode. Similarly, when the user tries to exit calibration mode, the user will be prompted to confirm that they want to exit, thereby unfreezing live values. If you check this box, however, these warning boxes will not appear, and the user will automatically enter/exit calibration mode. If you don’t check it, the warning prompts will appear, giving the user the option of aborting the starting or exiting of calibration mode.
Check this button to activate the “As Found/As Left” feature for calibration logs. (OpenBSI 5.8 and newer.) Note: You cannot use this feature if you are also using the extended verification feature; they are mutually exclusive.
If you have modified a standard application with different calibration signal or variable names, the [Calibration Signal Configuration] button allows you to specify the new signal/variable names.
If you have modified an application that uses the extended verification feature for Modbus messages, and specified different variable names for the extended verification signal configuration feature, click the Extended Verification Signal Configuration button to specify the new variable names. (OpenBSI 5.8 Service Pack 2 and newer.) Note: Extended verification only applies to external transmitters. In addition, the Extended Verification and As Found/As Left functions are mutually exclusive.
If you have modified a standard application with different verification signal or variable names, the As Found / As Left Verification Signal Configuration button allows you to specify the new signal/variable names. (OpenBSI
5.8 and newer.) Note: The As Found/As Left and Extended Verification features are mutually exclusive.
If you chose Local for Select Transmitter Access, click Finish; if you chose Remote click Next.
3-12 Starting TechView / Establishing Communications Issued Nov-2016
TechView User’s Guide (D5131)
3.2.7 Transmitter Setup (Only Visible if using Remote Access)
First, specify the number of transmitters connected below this RTU.
Specify the number of transmitters below this RTU
Figure 3-9. Transmitter Setup dialog box – Initial View
Issued Nov-2016 Starting TechView /Establishing Communications 3-13
TechView User’s Guide (D5131)
If desired, you can change the Local Address and the Type of the transmitter from the Transmitter Identification dialog box.
To modify the address or type of a transmitter in the list, either double-click on the transmitter’s name, or click on the name, and then click on Modify and make the changes in the Transmitter Identification dialog box.
Figure 3-10. Transmitter Identification
To add another transmitter to the list, click Add and fill in the fields in the Transmitter Identification dialog box.
To remove a transmitter from the list, click on the transmitter’s name, then click Remove.
To remove all of the transmitters from the list, click on Remove All. Click OK when finished. When you have finished configuration for the transmitters, click Finish.
3-14 Starting TechView / Establishing Communications Issued Nov-2016

3.3 Signing On to the RTU

Once you click Finish and communications have been successfully established with the RTU, TechView prompts you for a Username and Password.
TechView User’s Guide (D5131)
Figure 3-11. SignOn to RTU
Enter the username/password combination, and click Signon. The RTU must recognize this username/password combination as valid in order for you to gain access.
Notes:
Some customers may choose to set up a default username/password,
to avoid having to enter it repeatedly. See Application Settings later in this manual, for details.
If you use TechView to call up web pages, for example, in the
Measurement group, do not sign off from the Sign-On/Off web page control unless you intend to also exit TechView, since the web pages, and TechView share the same connection to the RTU. If you only want to exit web pages, but continue with TechView, just close the web pages, from the standard windows control, don’t sign off.
If you are unable to sign on to the RTU, because it needs to be
downloaded first, you can do this without logging on. In TechView, click on Operations > Download First Level Node to call up the Downloader and download a control strategy file to the RTU. You can then proceed to download.
Issued Nov-2016 Starting TechView /Establishing Communications 3-15
TechView User’s Guide (D5131)
Chapter 4 – Overview of TechView Functions
The basic functions of TechView are available through the menu bar, and tool bar items (see Figure 4-1).
In This Chapter
4.1Changing the Session Parameters You Entered ............................. 4-2
4.2Restarting the Session ..................................................................... 4-3
4.3Saving Your Session File ................................................................. 4-3
Copying Your Session to Another Session File Name .................... 4-3
4.4Application Settings ......................................................................... 4-3
4.4.1Startup tab ............................................................................ 4-4
4.4.2Security tab ........................................................................... 4-7
4.4.3General tab ........................................................................... 4-7
4.5Calibration and Verification Signals ................................................. 4-8
4.5.1Extended Verification ........................................................... 4-9
4.5.2Changing the Name of a Transmitter ................................. 4-11
4.6Changing the Local Addres s / Group Number ............................... 4-11
4.6.1Starting the Flash Configuration Utility ............................... 4-12
4.6.2Show Firmware Version in Node ........................................ 4-12
4.6.3Writing an Audit Note ......................................................... 4-12
4.6.4Viewing OpenBSI Workstation Communication Statistics . 4-13
4.6.5Deleting Historical Files from the RTU (Clear History) ....... 4-13
Standard sets of configuration/calibration pages specific to each type of transmitter, or RTU containing a transmitter, are included when you install TechView.
Based on your choice of node type, these pages load in a series of one or more groups, accessible via icons on the right hand side of the page. When you choose a particular group, the associated pages appear on the screen, or menu items which call them appear, depending upon the number of pages. Once the group is activated, you can proceed to perform the calibration, configuration, or online editing for your RTU/transmitter.
The Calibration group is always required. Additional groups may exist for other functions. Advanced users can also change the organization of the pages in groups, by editing TechView initialization files. For more information on changing these initialization files, see Appendix A – Initialization Files.
Issued Nov-2016 Overview of TechView Functions 4-1
TechView User’s Guide (D5131)
Open new session file
Open existing session file
Save session file
Restart session with new parameters
View/Modify application parameters
Title bar
Menu bar
Tool bar
Network tree
Pop-up menu to call up other programs for use with this RTU.
Modify session parameters
View/modify calibration signal names
Change local address and group number
Start Flash Configuration Utility
Write audit note
Show/hide Node status box
Show/hide Communication Statistics Window
Show firmware version in node
Show TechView version info
Access help file
Blinks on data refresh
Sele ct the gr oup of pages you want to use right now.
Status bar
Node status box
Communication Statistics window
Most recent status message
Type of RTU/application
Firmware revision in RTU or Transmitter
Node address (IP or BSAP local)
Figure 4-1. TechView Overview

4.1 Changing the Session Parameters You Entered

If you want to change the entries you made in the Communication Setup, Node Setup, or Calibration Setup dialog boxes for this session, click on the icon shown at left, or click on Configure > Session Parameters. Enter the new parameters, as desired. After exiting the wizard, the session restarts automatically with the new parameters.
Typical calibration page
Current security level of logged on user.
4-2 Overview of TechView Functions Issued Nov-2016

4.2 Restarting the Session

If you change the local address or group number of the RTU, and want to restart the session, click on the icon shown at left. If communication is lost during calibration, you can also use this button to restart communications.

4.3 Saving Your Session File

To save your session file, click on File > Save Session, or click on the
Copying Your Session to Another Session File Name
icon shown at left. Click on File>Save Session As. The Save Copy As dialog box opens.
Enter a name for the second copy of the session file in the File name field, and click Save. The original session file remains active; the newly named session file will be a copy of the active session file, but will not be running.
TechView User’s Guide (D5131)
Enter a name for the session, then click on [Save].
Figure 4-2. Copying a Session

4.4 Application Settings

Note: Unlike session parameters, which are stored in TVS files, and
can thereby vary depending upon which session file you have running, application parameters apply to the TechView application itself. Therefore, there is only one set of application parameters that apply no matter what session file you are using
Issued Nov-2016 Overview of TechView Functions 4-3
TechView User’s Guide (D5131)
Application Settings determine the choices the user needs to make when TechView first starts. To call up the Application Settings dialog box, click Configure > Application Settings, or simply click on the icon, shown at left.
Figure 4-3. Application Settings dialog box – Startup tab
4.4.1 Startup tab
The Startup tab specifies the initial prompt a user sees, if a session file is not specified via the command line. As discussed, earlier, a *.TVS session file holds certain configuration parameters for how TechView communicates with a controller or transmitter
Field Create a new
Session
Description
If selected, when TechView starts, it prompts the user to provide a name for a new session (or to use the default name of temp.tvs.). This session file holds the communication/ configuration settings you subsequently define.
4-4 Overview of TechView Functions Issued Nov-2016
TechView User’s Guide (D5131)
Figure 4-4. Create Session File
Open an existing Session
If selected, when TechView starts, it prompts the user to choose a previously defined session file, which TechView then starts. (Default choice). Note:: Only one session can be active at any one time; if you already have a session running, when you try to open a new session file, TechView prompts you to confirm that the current session will be closed.
Figure 4-5. Open Session File
None
Issued Nov-2016 Overview of TechView Functions 4-5
If selected, no prompt appears initially. The user has two choices:
The user can choose to manually start a new session by clicking on File> New Session, or by clicking on the icon shown at left
The user can choose to open a previously defined session by clicking on File > Open Session, or by clicking on the icon shown at left, and then selecting the appropriate *.TVS file
TechView User’s Guide (D5131)
Run application maximized at startup Show Node Status and Communication Statistics at startup
If checked, when the TechView application starts, it fills the available screen space on your PC monitor
If checked, displays the Node Status box, and the Communication Statistics window.
Note: These items can be shown/hidden after startup
“In Service” = Unit is currently on-line. “Out of Service”= Unit is currently off-line.
“Alive = TechView can communicate with the unit. “Dead = TechView cannot communicate with the unit.
“Downloading”= Download in progress “Needs NRT” = IP node needs new Node Routing
“Needs Poll” = BSAP node needs to be polled. “Time Sync” = Unit needs a time-synchronization
is completed
Table.
Figure 4-6. Node Status Box
To show/hide the Node Status box, click View > Node Status or click on the icon shown at left.
Figure 4-7. Communication Statistics window
To show/hide the Communication Statistics window, click on View > Comm Statistics or click on the icon shown at left.
4-6 Overview of TechView Functions Issued Nov-2016
4.4.2 Security tab
The Security tab allows you to optionally set up a default username/password combination that TechView uses to log onto any RTU in your network. For this to work, each RTU you want to communicate with must have this same username/password combination defined locally.
Note: Beginning with OpenBSI 5.8 Service Pack 1, passwords can
The Security tab also allows you to set a timeout (in seconds) for successfully logging in or out of web pages.
TechView User’s Guide (D5131)
include up to 16 characters; previous versions were limited to six characters.
If you don’t want to enter a username and password every time you log onto an RTU, you can enable the default security feature, and then set a default username and password. You must have configured the same username / password combination for each RTU you try to access during this session.
These settings allow you to change the timeouts for logging into/out of web pages, handled via the Data Server.
Figure 4-8. Application Settings dialog box – Security tab
4.4.3 General tab
On the General tab you can specify whether or not you want the Advanced Interface Setup File detection dialog box to prompt you when AISF files already exist on the PC, and would be overwritten by files from the RTU. To disable this prompt, un-check the Show Advanced Interface Setup File Detection Dialog option.
Issued Nov-2016 Overview of TechView Functions 4-7
TechView User’s Guide (D5131)
Figure 4-9. Application Settings dialog box – General tab
You can also use the Change Font button to change the font used in properties pages from the default of MS-San Serif – 8 point.
The change of font takes effect once the application or session is re­started. You can change the point size of text immediately by clicking View > Small Fonts, View> Medium Fonts or View > Large Fonts.
Note: Bolding or larger point sizes may result in text that cannot fit on
pages and so cannot be read. Verify your changes to ensure readability.

4.5 Calibration and Verification Signals

If you modify a standard application (pre-made ACCOL load or ControlWave project that shipped from the factory with your unit), and as part of that modification, change the signal names used for calibration, verification, live values, etc., you must identify the new signal names, within TechView, otherwise calibration and verification operations will not function properly. To do this, you must call up the associated Calibration, or Verification signals dialog box.
To call up configuration signals, click Configure > Calibration Signals or click the Calibration Signal Configuration button located in the Calibration Setup page of the Session Parameters wizard. You can also click the icon shown above.
To call up verification signals, click Configure > As Found / As Left
Verification Signals or click on the As Found / As Left Signal Configuration button located in the Calibration Setup page of the
Session Parameters wizard.
4-8 Overview of TechView Functions Issued Nov-2016
TechView User’s Guide (D5131)
Figure 4-10. Modifying Calibration Signal Names
Note: The dialog box appears different depending upon which RTU
platform and application you are modifying.
Be sure, when specifying calibration signals, that you first select the correct transmitter, meter run and/or application.
4.5.1 Extended Verification
If you are using the extended verification feature (OpenBSI 5.8 Service Pack 2 and newer) which allow audit logging of Modbus messages for verification of zero and span, and you change the names of verification variables used in the application, you must specify the new names as well.
To call up the Extended Verification Signals dialog box, click
Configure > Extended Verification Signals or click on the Extended Verification Signal Configuration button located in the Calibration
Setup page of the Session Parameters wizard.
Issued Nov-2016 Overview of TechView Functions 4-9
TechView User’s Guide (D5131)
The table, below, describes the functions of each of the variables. Change the names as required, and click OK.
Figure 4-11. Extended Verification Signals dialog box
If, at any time you need to restore the default variable names, call up the dialog box and click the Restore Factory Default Signal Names button, then click OK.
Table 4-1. Default Variable Names for Extended Verification
Default Variable Name Data Type Description
@GV.MVT_Events_MB BOOL Set to TRUE to enable the extended verification
feature. TechView only enables extended verification when this variable is defined; if the default variable name or a substitute name you specify does not exist in the application, the extended verification feature is disabled.
@GV.VF_Applied_x REAL TechView stores the verification (tester) value the user
enters here. The x refers to the external transmitter number.
@GV.CV_Applied_x REAL TechView stores the actual live value read from the
process variable in the transmitter here. The x refers to the external transmitter number.
@GV.Verif_Exec_x INTEGER This field shows an execution code when the applied
variables are updated. These codes are: 1 = Verify DP Zero
2 = Verify DP Span 3 = Verify SP Zero 4 = Verify SP Span 5 = Verify T Zero 6 = Verify T Span
Note: The application programmer is responsible for resetting this value to 0 after the code is read.
4-10 Overview of TechView Functions Issued Nov-2016
4.5.2 Changing the Name of a Transmitter
To change the name displayed for a particular transmitter, click the Change Transmitter Name button and type in a new name for the transmitter. If you decide you don’t want to change the name, prior to
pressing Enter press Esc.
Figure 4-12. Changing the Name of a Transmitter
If you make a mistake, and want to change signal names back to the original names from the standard application, click Restore Factory Default Signal Names.
This feature requires OpenBSI 5.8 (or newer).

4.6 Changing the Local Address / Group Number

For those RTUs in which the local address is stored in FLASH (ControlWave-series, TeleFlow-series) TechView can change the local address of the unit, and simultaneously change the local address for this session so communications will continue. The session is temporarily suspended while this change is occurring.
TechView User’s Guide (D5131)
By default, the Group Number is always set to 0, however, if you are using Expanded Node Addressing (EBSAP) you may assign different group numbers. Note, however, that once you have changed the group number, you must drag the icon for the node in the NetView tree so that it is underneath the correct virtual node.
Figure 4-13. Changing the Local Address / Group Number
Issued Nov-2016 Overview of TechView Functions 4-11
TechView User’s Guide (D5131)
p
p
To change the local address or group number, use the list boxes provided, then click on the associated Change button.
Note: The change local address and change group number functions for
4.6.1 Starting the Flash Configuration Utility
The Flash Configuration utility allows configuration of soft switches,
orts, IP parameters, and for ControlWave-series units, additional arameters, as well. To start the Flash Configuration utility, click
Operations > Access Flash Parameters, or click on the icon, shown at left. For information on the Flash Configuration utility, see the Chapter 5 of the OpenBSI Utilities Manual (document# D5081).
4.6.2 Show Firmware Version in Node
To find out the version of system firmware within the RTU, click on View > Version Information, or click on the icon shown at left
ControlWave require 04.60 or newer system firmware
Figure 4-14. Version Information dialog box
4.6.3 Writing an Audit Note
The Audit system configured in a ControlWave RTU’s AUDIT function block maintains a history of certain system events and alarms. In addition, beginning with TechView, it is possible to have an operator or technician send their own audit note to be stored in the RTU.
To send an audit note to the RTU, type the text in the box (up to 37 characters), then click on the Send button. TechView adds the note to the audit file, with a timestamp.
4-12 Overview of TechView Functions Issued Nov-2016
TechView User’s Guide (D5131)
Figure 4-15. Notes dialog box
To close the Notes dialog box, click Exit.
4.6.4 Viewing OpenBSI Workstation Communication Statistics
In addition to RTU communication statistics which may be displayed at the bottom of TechView screens, you can optionally view OpenBSI workstation communication statistics. To see these, click View > OpenBSI Processes from the menu bar.
For information on what the various statistics mean, see Chapter 6 of the OpenBSI Utilities Manual (D5081).
Figure 4-16. OpenBSI System Processes
4.6.5 Deleting Historical Files from the RTU (Clear History)
A ControlWave-series controller stores archive files of historical data, and audit records of significant system events. Normally, the Harvester, or some other program periodically collects this data for export to
Issued Nov-2016 Overview of TechView Functions 4-13
TechView User’s Guide (D5131)
OpenEnterprise or a third-party package, and eventually new data overwrites the existing data.
If, however, you want to permanently delete audit or archive data residing in the ControlWave, you can use the Clear History function.
Note: The ControlWave-series controller must have 04.80 or newer
If there should be a power failure to the ControlWave during the “Clear
Caution
History” operation, files will not delete properly.
If you want to save any Audit or Archive data, you must do this before
Caution
you use the Clear History function. We also recommend you save your current historical configuration to
an FCP file prior to deleting audit or archive files, so that you can restore the structure (though not the data), if a failure occurs.
1. To delete historical data, click Operations > Clear History.
firmware to use this function. This operation requires OpenBSI
5.7 or newer.
2. The Clear RTU History dialog box opens.
Choose the type of historical data you want to delete, then click “Start.”
Figure 4-17. Deleting Historical Data
3. You have four possible choices on what to delete. See explanations
for the fields, below.
4-14 Overview of TechView Functions Issued Nov-2016
TechView User’s Guide (D5131)
Field Description Clear Audit Records
Clear All Archive Files
Clear Single Archive File
Clear All History (Audit and All Archives)
This choice deletes all Audit records residing in the ControlWave.
This choice deletes all Archive Files residing in the ControlWave.
This choice deletes a single Archive File. When choosing this, use the list box to specify which Archive File you want to delete.
This choice deletes all Audit Records and all Archive Files.
4. After you make your choice, click OK to proceed with the deletions,
or Cancel to abort the operation. After you click OK TechView gives you a warning prompt and asks you to confirm that you want to perform the deletions.
Once you click OK to the deletion confirmation prompt, there is no way
Caution
to reverse the deletion operation; the delete command removes the chosen files from the unit and the unit re-boots twice to accomplish the deletion operation.
Note: If you see the status message Not enough memory to complete
parameter setting, it means the flash memory area is full and the deletion could not be completed because files are shifted during the deletion. If this occurs during the Clear All History operation, try clearing audit files first, then clear archives; if this doesn’t resolve the issue, use the Flash File Access tool to delete individual files and free up space. See the ControlWave Designer Programmer’s Handbook (D5125) for information on the Flash File Access tool.
Issued Nov-2016 Overview of TechView Functions 4-15
Chapter 5 – Calibration Overview
This chapter provides general information on calibration and then provides procedures for calibrating different devices.
In This Chapter
5.1Why is Calibration Necessary? ........................................................ 5-1
5.2Calibration Concepts ....................................................................... 5-1
5.3Before You Begin ............................................................................. 5-3
5.4Equipment Required for Calibration................................................. 5-5
5.4.1Equipment Required For Pressure Calibration .................... 5-5
5.4.2Equipment Required For Temperature Calibration .............. 5-7
5.4.3Equipment Required for Analog Output Calibration (3808 only)
............................................................................................ 5-7
5.4.4Entering Calibration Mode / Leaving Calibration Mode ........ 5-7

5.1 Why is Calibration Necessary?

TechView User’s Guide (D5131)
When a transmitter ships from the factory, it is already factor y­compensated for a specific pressure range. When you install the transmitter on site, however, the accuracy of the transmitter may have been affected by environmental conditions (temperatures, atmospheric pressures, etc.) at that particular site. For this reason you must calibrate the transmitter. In addition, you must periodically re-calibrate the transmitter to ensure that it maintains its accuracy.

5.2 Calibration Concepts

You must understand certain concepts and terminology before performing the calibration procedures:
Measured Variable
Applied Value
The measured variable is the actual “live” process variable reading, coming from the transmitter. This could be a differential/gage pressure reading, a static pressure reading, or a resistance temperature detector (RTD) temperature reading.
Applied values are used during calibration. They represent known values, which are used as a standard or benchmark from which other values can be calculated by the transmitter.
For pressure, the applied value is a known pressure, applied from an external pressure source, to the transmitter.
For temperature, the applied value is a known resistance applied by the user, by connecting a resistor, in place of the RTD.
Upper Range Limit (URL) and Lower Range Limit (LRL)
Issued Nov-2016 Calibration Overview 5-1
Transmitters are ordered with specific pressure ranges. An upper range limit (URL) and a lower range limit (LRL) are preset at the factory to define the range of pressures that the transmitter can
TechView User’s Guide (D5131)
Upper Range Value (URV) and Lower Range Value (LRV)
(applies to analog models only)
read. The URL is the maximum value for the pressure, which can be read by the transmitter. The LRL is the minimum value for the pressure, which can be read by the transmitter. Any pressure value which is higher than 105% of the value of the URL or lower than the value of (LRL -(5% of the URL)) is considered out-of-range and cannot be considered accurate.
In contrast to the URL and LRL which define the fixed, pre­defined factory range for the transmitter, the Upper Range Value (URV) and Lower Range Value (LRV) may be used to establish the range being measured for the actual process. For example, suppose the transmitter has an LRL of 0 psi, which corresponds to 4 mA and a URL of 1000 psi, which corresponds to 20 mA, but the range of values you want to measure is 0 to 500 psi, then the LRV would be 0 (4 mA) and the URV would be 500 (20 mA).In this example, the LRV and URV settings represent a 2 to 1 turndown.
There are certain rules to be observed when setting LRV and URV.
The LRV cannot be set below the LRL. The URV cannot be set above the URL. URV minus LRV must be greater than 1/20 of the URL.
For more information on LRV/URV, please consult the hardware documentation for the transmitter.
SPAN
ZERO
MVT
RTD
GPT
The SPAN value is a positive value which (when added to the ZERO value) defines the highest expected operational value for a
particular measured variable. This could be the same as the URL, but in some cases, the SPAN value might not equal the URL. For example, a transmitter having a 2000 psi URL might only be used over a 0 to 1000 psi range, and so a span calibration at the 1000 psi point is performed. The minimum value for the span is URL divided by 20. If, when calibrating the SPAN value for a pressure transducer, you apply a pressure value significantly less than the URL, accuracy may be reduced for readings outside of the span.
The ZERO value is the measured variable value that corresponds to 0% of the transmitter’s SPAN, i.e. 4 mA for a 4 to 20 mA range, or 1V for a 1 to 5V range.
Multi-variable transducer.
Resistance Temperature Detector
Gage Pressure Transducer
Verification
This is a periodic check in which you compare the live reading of a transmitter against a known measurement standard in order to
5-2 Calibration Overview Issued Nov-2016
TechView User’s Guide (D5131)
p
determine if the transmitter measures accurately and remains properly calibrated. You apply an external source, such as a dead­weight tester, to the transmitter and check points along the measurement curve.
Damping
Damping Time
As Found / As Left (OpenBSI 5.8 and
newer)
The output damping feature is applicable to ControlWave EFM/GFC/XFC flow/gas computers, 3508-series Transmitters and 3808-series Transmitters and controls the rate at which the output responds to a given change of input. It is used to slow down the output response to a rapid or oscillatory change of the measured variable.
The period of time during which the indicated value changes only 63% of the difference between the “present measured variable” and the “present indicated pressure” in one damping time period. It would take 5 times the DP Floating Point Damping Time for the “present indicated pressure” to equal the “present measured pressure” (if a change in the “present measured pressure” didn’t occur). An increase in the DP Floating Point Damping Time results in an increased smoothing of the indicated value. Note: In the case of a model 3530-series unit, if the External Transmitter is a model 3808, Damping Mode is always enabled and can’t be changed; Damping Time can’t be changed either.
“As Found/As Left” verification is a systematic method to capture the state of a transmitter, both before and after linearization or calibration of the transmitter.
At the start of the process, you perform the “As Found” verification. This determines the current state of the transmitter.
You evaluate the results against a set of limits to determine whether you need to:
a) leave the transmitter as is b) perform a physical calibration of the transmitter or c) take the transmitter out of service and replace it. You can compare the results to values from the previous “As Left”
verification.
Note: As Found/As Left operations are logged in the audit buffer
at the RTU.

5.3 Before You Begin

Before you start the calibration procedure, there are several things you need to be aware of.
Calibration in the lab vs. calibration on site
Issued Nov-2016 Calibration Overview 5-3
Although it is more convenient and recommended to perform this
rocedure using a laboratory setup, calibration can also be performed
TechView User’s Guide (D5131)
on site providing that the connecting line or flange is equipped with a calibration tap and appropriate shutoff and bypass valves. This added equipment allows you to feed in an external test pressure source or use the process pressure as a reference signal. In the latter setup, the valves are closed to seal a fixed pressure in the connecting line.
Note: Before starting any calibration procedures, make sure that the
unit in question is firmly anchored in its intended operating position. A different mounting position can affect zero calibration for some ranges and necessitate re-calibration.
WARNING
Communication Port Considerations
Only perform calibration activities if the area is non-hazardous. Before attempting on-site calibration, carefully check the application. If
the unit in question is operated in a closed control loop configuration, either the MVT/GPT Pressure Transducer must be isolated from the process, or the process must be turned off. If this is not done, a critical process could accidentally be driven into a dangerous region causing damage to equipment and property, and injury to persons
Note: If you modified your application to use different calibration /
verification signal names, you must identify these within TechView. See Section 4.5.
Notes:
When performing calibration in the field for a Network 3000
device, use the Pseudo-Slave Port of the RTU, not the Slave Port. Connecting through the Slave port would force you to break communications with upper levels of the network. In addition, if you use the Slave Port, it would receive a node routing table (NRT) from TechView which would only encompass the locally connected node, and its transmitters. When the connection is subsequently restored with the network, this would be an incorrect node routing table, and communications would still not be possible with higher level nodes. To avoid these problems, perform calibration through a Pseudo-Slave Port.
When performing calibration through a ControlWave unit’s port,
you must set its _Px_NRT_DIS system variable to TRUE to avoid the same problem described above.
When performing calibration through a ControlWave unit’s port,
you must set its _Px_NRT_DIS system variable to TRUE to avoid the same problem described above.
If you are using Expanded Node Addressing (EBSAP) with
ControlWave, and the ControlWave unit is in a group other than Group 0, the Local Port MUST be marked as LOCAL via the System Variable Wizard in ControlWave Designer. Otherwise, communications will not be possible.
5-4 Calibration Overview Issued Nov-2016
TechView User’s Guide (D5131)
The _Px_LOCAL_PORT must be set to TRUE
Figure 5-1. Identifying the Local Port

5.4 Equipment Required for Calibration

The equipment required depends on what type of calibration you need to perform.
5.4.1 Equipment Required For Pressure Calibration
Pressure calibration (static, differential or gage) requires the following items:
Instrument Mounting Fixture: This fixture must secure the
instrument such that the MVT is positioned identically to its installation site. GPT equipped units must be positioned vertically.
A Pressure Gauge (must cover the range of the MVT/GPT
Transducer .025% accuracy)
An adequate Pressure Regulator A Pressure Source (and Interconnect Lines) capable of generating
fixed pressure values equivalent to 0%, and 100% values of transmitter’s range (URL).
The pressure calibration procedures in this manual require that a text fixture incorporating these items be used.
Issued Nov-2016 Calibration Overview 5-5
TechView User’s Guide (D5131)
A typical MVT pressure calibration interconnection diagram is shown, below:
: For static pressure calibration setup, connect an identical supply
Note
pressure to the low side port of the MVT.
Figure 5-2. Typical Setup for MVT Calibration
A typical gage pressure calibration interconnection diagram is shown, below:
Figure 5-3. Typical Calibration Setup of GPT Using Test Pressure
5-6 Calibration Overview Issued Nov-2016
TechView User’s Guide (D5131)
5.4.2 Equipment Required For Temperature Calibration
In order to calibrate the RTD Temperature properly, you must have a
Caution
resistor with accuracy of 0.01% accuracy or better. Other wise: you could actually make the RTD calibration less accurate than it was prior to your calibration/verification. For example, with a 0.1% resistor, a temperature reading could be off 0.25 degrees C at 100 ohms, and 0.9 degrees C at 300 ohms. Factory calibration accuracy is 0.05°C at 100°C (212°F) and 0.30°C at 300°C (572°F).
RTD Calibration Resistors
R
zero
R
20%Scale
R
40%Scale
R
50%Scale
R
60%Scale
R
80%Scale
R
span
= 100.00, ±0.01%, 1/4W (0°C = +32°F) = 140.23, ±0.01%, 1/4W (104.55°C = +220°F) = 180.57, ±0.01%, 1/4W (212.85°C = +415°F) = 200.01, ±0.01%, 1/4W (266.38°C = +511°F) = 220.12, ±0.01%, 1/4W (322.74°C = +613°F) = 260.22, ±0.01%, 1/4W (438.34°C = +821°F) = 300.00, ±0.01%, 1/4W (+557.69°C = +1035.84°F)
5.4.3 Equipment Required for Analog Output Calibration (3808 only)
Analog output calibration requires: electrical supply source: +5Vdc (Min.), +42Vdc (Max.) – powers
the transmitter
digital multimeter (DMM) or ammeter with a 5-1/2 digit scale
(.005% accuracy)
Transmitter Interface Unit (Part No. 389959-01-4) 250-ohm resistor (.01%, 1/4 watt)
5.4.4 Entering Calibration Mode / Leaving Calibration Mode
You enter calibration mode by clicking on the Calibration icon on the right hand side of the screen.
.
Issued Nov-2016 Calibration Overview 5-7
TechView User’s Guide (D5131)
When you enter calibration mode, live values for differential pressure, static pressure, gage pressure, and temperature are frozen (i.e. the RTU will collect no new values from the transmitter, during calibration mode.) By default, you’ll see a prompt warning you of this and giving you the option of not entering calibration mode. Click Yes to enter calibration mode.
Figure 5-4. Entering Calibration Mode
Figure 5-5. Exiting Calibration Mode
Similarly, by default, when you click on a different group icon to exit calibration mode, you’ll see a prompt asking you to confirm that you want to exit calibration mode, thereby unfreezing live values. Click Yes to exit calibration mode.
Note: You can disable these confirmation prompts in the Calibration
Setup session parameters dialog box.
5-8 Calibration Overview Issued Nov-2016
TechView User’s Guide (D5131)
p
Chapter 6 – Calibrating the 3508 series TeleTrans
Transmitter
This chapter covers calibration activities for the Bristol 3508-series TeleTrans transmitter.
For a general overview of calibration concepts, see Chapter 5. For instructions on starting calibration mode, see Section 5.4.4.
In This Chapter
6.1Calibration of Differential/Gage Pressure ........................................ 6-1
6.2Calibration of Static Pressure .......................................................... 6-3
6.3Calibration of RTD Temper ature ...................................................... 6-5

6.1 Calibration of Differential/Gage Pressure

Figure 6-1. DP Tab – Calibrating Differential/Gage Pressure
Calibrating Zero and Span
1. To calibrate the zero, vent the transmitter to atmosphere, so no
differential pressure is applied. Click Calibrate Zero and the Live Value will be trimmed to be as close to the zero as possible.
2. To calibrate the span, enter the desired span in the Operating Span
field, then apply a pressure equivalent to the desired span. This value must be less than or equal to the Data Plate DP URL value. Then click Calibrate Operating Span. The Live Value will be trimmed to be as close to the desired span, as possible.
Shifting the DP/GP Floating Point Number
If needed, the span and zero of the DP/GP floating point number may be shifted a minor amount (+3% of URL) so that it correlates with the
ressure standard. A correction is achieved by calculating a shift
Issued Nov-2016 3508 Calibration 6-1
TechView User’s Guide (D5131)
correction factor and applying it to the floating point number by direct entry.
1. Compare the floating point values in the Live Value field with the
2. Calculate the amount of ZERO shift necessary to correlate the
Example: Assume that a DP Transmitter with a 300 inH2O URL
3. Click Shift Floating Point Zero. The adjusted value will appear in
pressure standard. Choose values at, or close to, zero and the URV. If the zero pressure reading is offset from zero, proceed to step 2. However, if the zero readings agree but an offset occurs at the upper end of the range, proceed to step 4.
floating point number with the external reading as described in the following example, and enter it in the Zero Shift field.
has been properly calibrated for a 0 to 100 inH2O range. If the external device reads 1.400, the difference is: 0 - 1.400 = -1.400 = -
1.400. A Zero Shift entry of -1.4 would provide correlation between the floating point number and the external device. Note: Always set the zero before attempting span calibration. Any zero errors result in an equivalent span offset that complicates calibration.
the Live Value field. If an error occurs, proceed to step 6.
4. Calculate the amount of SPAN shift necessary to correlate the
floating point number with the external reading at the upper end of the range (this assumes that zero was previously calibrated as noted above). Use the following example as a guide then enter the value in the Span Shift field.
Example: Assume that the URL of the transmitter is 300 inH2O and
that the calibrated range is 0 to 100 inH2O. If the external device reads 0.000 and 99.125 inH2O for equivalent DP/GP floating point readings of 0.000 and 100.000 inH2O, the difference would be
100.000 - 99.125 = 0.875. The required span shift entry would be calculated as follows:
Span Shift = (+inH2O error) URL/URV = (+0.875) 300/100 =
+2.625 inH2O
5. Click Shift Floating Point Span. The adjusted value will appear in
the Live Value field. If an error occurs, proceed to step 6.
6. If the error message "Correction not allowed, change exceeds +3%
of URL" appears, the problem may be caused by: a) the pressure source is not accurate; b) an incorrect value was entered; or c) the zero pressure point was not set accurately. Check out each item and make corrections as required.
Changing the Pressure Units
Select the pressure units for the transmitter, from the Select Pressure Units list box, then click the Change Pressure Units button.
Note: To force a refresh after changing pressure units, click the Restart
button.
6-2 3508 Calibration Issued Nov-2016

6.2 Calibration of Static Pressure

p
For static pressure, the same pressure must be applied to both the HI and LO inputs of the transmitter, thus producing a zero differential across the HI and LO ports, and a SP value at the HI port.
TechView User’s Guide (D5131)
Figure 6-2. SP Tab – Calibrating Static Pressure
Calibrating Zero and Span
Shifting the SP Floating Point Number
1. To calibrate the zero, vent the transmitter to atmosphere, so only
atmospheric pressure is applied. Click Calibrate Zero, and the Live Value will be trimmed to be as close to the zero as possible.
2. To Calibrate the span, enter the desired span in the Operating Span
field, then apply a pressure equivalent to the desired span. This value must be less than or equal to the Data Plate SP URL value. Then click Calibrate Operating Span. The Live Value will be trimmed to be as close to the desired span, as possible.
If needed, the span and zero of the SP floating point number may be shifted a minor amount (+3% of URL) so that it correlates with the
ressure standard used by the external process measuring device. A correction is achieved by calculating a shift correction factor and applying it to the floating point number by direct entry.
If it is determined that the difference at zero and upper end of the SP range is the same, then only a zero shift will be required.
Example: Assume that the URL of the transmitter is 2000 psi
and that it has been calibrated for a 0 to 500 psi range. If the 0 check point was read as 1.200, the difference would be: 0 - 1.200 = -1.200. A zero entry of -1.2 psi would provide the correlation between the floating point number and the standard of the
Issued Nov-2016 3508 Calibration 6-3
TechView User’s Guide (D5131)
To implement a zero correction, enter the correction value in the Zero Shift field, and click the Shift Floating Point Zero button. After a
moment the reading in the Live Value field will reflect the change. The attempted entry of values that exceed +3% will result in an error
message. Should a correlation difference exist at the upper end of the range but
none occurs at the zero end, only a span shift will be required. Example: Assume that the URL of the static pressure sensor is
external device.
Note: Always set the zero point before attempting span
calibration. Any zero errors will result in an equivalent span offset that complicates calibration.
2000 psi and that the calibrated range is 0 to 500 psi. If the Live Value field reads 0.000 and 500 psi and the external device reads 0.000 and 497.445 psi for these same points, the difference is 0 psi at the low end and 2.555 psi at the upper end. The span shift is calculated as follows: Span Shift = (+ psi error) URL/Operating Point = (+2.555) 2000/500 = +10.22 psi
Changing the Pressure Units
To implement a span correction, enter the correction value in the Span Shift field, and click the Shift Floating Point Span button. After a
moment the reading in the Live Value field will reflect the change. Values that exceed +3% will elicit an error message and the previous
values will be retained. Select the pressure units for the transmitter, from the Select Pressure
Units list box, then click the Change Pressure Units button.
Note: To force a refresh after changing pressure units, click Restart.
6-4 3508 Calibration Issued Nov-2016

6.3 Calibration of RTD Temperature

Note:: The maximum rated RTD cable length for the TeleTrans
Transmitter (Model 3508) is 100 feet. This limitation is imposed to reduce noise pickup at electrically noisy sites and also to limit error due to line resistance. Users may employ longer cables provided they verify proper operation in their specific application.
TechView User’s Guide (D5131)
Changing the Temperature Units
Calibrating the Zero and Span
Figure 6-3. T tab – Calibrating RTD Temperature
The factory calibrates the internal temperature sensor to provide optimum transmitter accuracy and performance over a wide range of temperature conditions and cannot be changed in the field. However, the temperature output reading can be calibrated for the user's application.
Readings can be displayed in either degrees Celsius or degrees Fahrenheit. To toggle the units from one to the other, click on the Change Units button. Output values will be scaled to the selected units.
Note: To force a refresh after changing units, click Restart.
1. Disconnect the regular RTD and connect a 100 Ohm (±0.01%)
precision resistor across the RTD terminals (this is equivalent to 32°F).
2. Observe the Live Reading and wait for it to stabilize.
3. Click on the Calibrate Zero (at 100 OHMs) button to calibrate the
RTD zero.
Issued Nov-2016 3508 Calibration 6-5
TechView User’s Guide (D5131)
p
4. Disconnect the resistor of step 1 and connect a 140.23 Ohm (±0.01%)
5. Observe the Live Reading and wait for it to stabilize.
6. Enter the desired operating span in the Operating Span field.
7. Click on the Calibrate Operating Span button.
8. Disconnect the resistor, and reconnect the RTD.
precision resistor across the RTD terminals (this is equivalent to 220°F).
Figure 6-4. Model 3508 TeleTrans - Connections for RTD Calibration (See page 5-7 for resistor
values)
Adjusting the Live Reading
If after re-connecting the unit to the process it is determined th at the number in the Live Reading field does not correlate with that measured by an external process monitoring device, offset compensation may be
erformed. To do this, enter the desired reading in the Live Reading
field, then click the Adjust Live Reading button.
6-6 3508 Calibration Issued Nov-2016
TechView User’s Guide (D5131)
Chapter 7 – Calibrating the 3808 Transmitter
This chapter covers calibration activities for the Bristol 3808-series transmitter.
For a general overview of calibration concepts, see Chapter 5. For instructions on starting calibration mode, see Section 5.4.4.
In This Chapter
7.1Calibration of Gage or Differential Pressure (3808-10A /3808-30A only)
......................................................................................................... 7-1
7.2Calibration of Static Pressure (3808-30A ONLY) ............................ 7-2
7.3Calibration of RTD Temper ature ...................................................... 7-3
7.4Calibrating the Analog Output (Analog Pressure Transducer models
ONLY) .............................................................................................. 7-5

7.1 Calibration of Gage or Differential Pressure (3808-10A/3808-30A only)

The Sensor LRL and Sensor URL display the lower-range limit and upper range limit, respectively, of the gage pressure / differential pressure sensor.
Figure 7-1. DP tab – Calibrating Gage or Differential Pressure
1. In the “Select Operation for DP Sensor” box, select Zero
Calibration.
Issued Nov-2016 3808 Calibration 7-1
TechView User’s Guide (D5131)
2. To perform the zero calibration, vent the transmitter to atmosphere,
3. To calibrate the span, select Span Calibration in the “Select
Note: If, for some reason, you want to return to the calibration that was
so no pressure is applied. Click on the Calibrate Zero button, and the Measured Value will be trimmed to be as close to the zero as possible.
Operation for DP Sensor” box, then apply a pressure equivalent to the desired span. Now enter the pressure you applied in the New
Applied Value field. If the transmitter accepts the New Applied Value, the value you entered will be displayed in the Target Trim Value field. This value must be less than or equal to the Sensor URL value. Click on Calibrate Span. The Measured Value will be
trimmed to be as close to the desired span, as possible. Note: Span calibration may only be performed successfully if the applied pressure and span are within +3% of one another.
performed by the factory, click on Restore Factory Defaults.

7.2 Calibration of Static Pressure (3808-30A ONLY)

Note: For static pressure calibration setup, connect identical supply
pressures to both the low and high side ports of the MVT.
Figure 7-2. SP tab - Calibration of Station Pressure
The Sensor LRL and Sensor URL display the lower-range limit and upper range limit, respectively, of the static pressure sensor.
1. In the “Select Operation for SP Sensor” box, select Zero
Calibration.
7-2 3808 Calibration Issued Nov-2016
TechView User’s Guide (D5131)
2. To perform the zero calibration, vent the transmitter to atmosphere,
so only atmospheric pressure is applied. Click on the Calibrate Zero button, and the Measured Value will be trimmed to be as
close to the zero as possible.
3. To calibrate the span, select Span Calibration in the “Select
Operation for SP Sensor” box, then apply a pressure equivalent to the desired span. Now enter the pressure you applied in the New
Applied Value field. If the transmitter accepts the New Applied Value, the value you entered will be displayed in the Target Trim Value field. This value must be less than or equal to the “Sensor URL” value. Click on Calibrate Span. The Measured Value will
be trimmed to be as close to the desired span, as possible. Note: Span calibration may only be performed successfully if the applied pressure and span are within +3% of one another.
Note: If, for some reason, you want to return to the calibration that was
performed by the factory, click on Restore Factory Defaults.

7.3 Calibration of RTD Temperature

Figure 7-3. T tab – Calibration of RTD Temperature
Issued Nov-2016 3808 Calibration 7-3
TechView User’s Guide (D5131)
Lower TB
Not Shown
+
3808 RTD Co nnect ion s
Figure 7-4. Model 3808 Transmitter RTD Calibration Connection Diagram (See Section 5.4.2 for
resistor values)
1. Disconnect the regular RTD and connect a 100 Ohm (0.01%)
precision resistor across the RTD terminals (this is equivalent to 32°F).
2. Observe the Measured Value and wait for it to stabilize.
3. Click the Calibrate Zero button to calibrate the RTD zero.
4. Disconnect the resistor of step 1 and connect a precision resistor
across the RTD terminals to represent your desired span. To calibrate to the default temperature span value of 558 °C (1036
°F) requires a resistance value of 300 ohms +/- 0.01%. To calibrate the temperature span to any other desired value, you
must know the proper resistance value for that span. You must calculate this yourself or refer to a 100 ohm platinum RTD chart. For example, to calibrate the span to the upper range limit (URL) of the transmitter (660 °C, 1220 °F) requires a 332 ohm resistance +/-
0.01%. After you connect the required resistance, enter the desired temperature span value in the Custom Span field.
5. Observe the Live Value and wait for it to stabilize.
6. If you used a 300-ohm resistance in step 4, click the Calibrate Span
(at 300 ohm) button; if you connected any other resistance value click the Calibrate Custom Span button.
7. Disconnect the resistor, and reconnect the RTD.
7-4 3808 Calibration Issued Nov-2016
TechView User’s Guide (D5131)
Notes:
If, for some reason, you want to return to the calibration that was
performed by the factory, click on Restore Factory Defaults.
To change the reading from the RTD for the 3808, enter the desired
reading (which should be similar to the actual reading, but presumably off slightly), and click on Adjust Live Reading.

7.4 Calibrating the Analog Output (Analog Pressure Transducer models ONLY)

3808 analog output calibration requires the following items: electrical supply source: +6Vdc (Min.), +42Vdc (Max.) – powers
the transmitter
digital multimeter (DMM) or ammeter with a 5-1/2 digit scale
(0.005% accuracy)
Transmitter Interface Unit (Part No. 389959-01-4) 250-ohm resistor (0.01%, ¼ watt)
Lower TB
Not Shown
Figure 7-5. 3808 Analog Output Calibration - Voltage Measurement Diagram
Issued Nov-2016 3808 Calibration 7-5
TechView User’s Guide (D5131)
Figure 7-6. 3808 Analog Output Calibration - Current Measurement Diagram
Figure 7-7. AO tab – Calibrating the Analog Output
Any of the three process variables (differential/gauge pressure, static pressure, or RTD temperature) can control the analog output of the transmitter. The analog output ranges from 4 mA to 20 mA, but the 4 mA and 20 mA points may change over time and require calibration. The measured variable in this case is the value read by an accurate milli­ammeter or voltmeter connected to the analog output.
To calibrate the analog output:
7-6 3808 Calibration Issued Nov-2016
TechView User’s Guide (D5131)
1. For Select the output to be calibrated choose “Cal 4 mA”.
Measure the current or voltage externally, and enter the measured value in the Enter new measured value for the selected Output field, in milliamperes.
2. To calibrate, select the Calibrate Analog Output button. The 3808
will adjust and save the new settings.
3. Repeat steps 1 and 2, but this time select “Cal 20 mA” for the Select
the output to be calibrated choice.
Note: When calibrating the 4mA point, only measured value entries
between 3mA and 5mA are accepted. When calibrating the 20mA point, only measured value entries between 19mA and 21mA are accepted.
Issued Nov-2016 3808 Calibration 7-7
TechView User’s Guide (D5131)
Chapter 8 – Calibrating the 3530 TeleFlow-series Flow
Computers
This chapter covers calibration activities for the Bristol 3530-series flow computers (TeleFlow, TeleRecorder, TeleCorrector).
For a general overview of calibration concepts, see Chapter 5. For instructions on starting calibration mode, see Section 5.4.4.
In This Chapter
8.1Calibration of Differential/Gage Pressure ........................................ 8-1
8.2Verification of Differential Pressure ................................................. 8-2
8.3Calibration of Static Pressure .......................................................... 8-2
8.4Verification of Static Pressure .......................................................... 8-3
8.5Calibration of RTD Temper ature ...................................................... 8-4
8.5.1Verification of RTD Temperature.......................................... 8-6
8.6Damping ........................................................................................... 8-6
8.6.1Configuring the Damping Time ............................................. 8-7
8.7Calibrating TeleRecorder Inputs (3530-45B/55B ONLY) ................ 8-7
8.7.1Calibrating the Input ............................................................. 8-8
Important
Whenever performing calibration operations on a 3530-series unit, you must be connected to the local port of the 3530. You cannot perform calibration via the network port of the 3530

8.1 Calibration of Differential/Gage Pressure

Figure 8-1. DP tab – Calibration of Differential / Gage Pressure
Issued Nov-2016 3530 TeleFlow Calibration 8-1
TechView User’s Guide (D5131)
1. If this flow computer is handling more than one meter run, first
select the run you want to configure from the Run list box.
2. In the “Select Operation” box, select Calibrate.
3. To calibrate the zero, vent the transmitter to atmosphere, so
atmospheric pressure is applied. Click on the Calibrate Zero button, and the Live Value will be trimmed to be as close to the zero as possible.
4. To calibrate the span, enter the desired span in the Operating Span
field, then apply a pressure equivalent to the desired span. This value must be less than or equal to the Data Plate URL value. Then click on Calibrate Operating Span. The Live Value will be trimmed to be as close to the desired span, as possible.

8.2 Verification of Differential Pressure

Note: This must be performed using an external tes t fixture that can
apply various pressures.
1. If this flow computer is handling more than one meter run, first
select the run you want to configure from the Run list box.
2. In the “Select Operation” box, select Verify.
3. Using an external test fixture, apply a known pressure. When the
Reported Value is steady, enter the pressure you applied in the Applied Value field, then click on Verify Differential Pressure Calibration. Repeat this process using the external test fixture to
apply known pressures of 0, 50, 100, 80, 20, and 0 percent of span.
Note: If, during the verification procedure, the transmitter is found to
be out of calibration, perform the calibration procedure to correct the discrepancy.

8.3 Calibration of Static Pressure

For static pressure, the same pressure must be applied to both the HI and LO inputs of the transmitter, thus producing a zero differential across the HI and LO ports, and a SP value at the HI port.
8-2 3530 TeleFlow Calibration Issued Nov-2016
TechView User’s Guide (D5131)
Figure 8-2. SP tab – Calibration of Static Pressure
1. If this flow computer is handling more than one meter run, first
select the run you want to configure from the Run list box.
2. In the “Select Operation” box, select Calibrate.
3. To calibrate the zero, vent the transmitter to atmosphere, so no
pressure is applied. Click on the Calibrate Zero button, and the Live Value will be trimmed to be as close to the zero as possible.
4. To calibrate the span, enter the desired span in the Operating Span
field, then apply a pressure equivalent to the desired span. This value must be less than or equal to the Data Plate URL value. Then click on Calibrate Operating Span. The Live Value will be trimmed to be as close to the desired span, as possible.

8.4 Verification of Static Pressure

Note: This must be performed using an external test fixture that can
apply various pressures.
1. If this flow computer is handling more than one meter run, first
select the run you want to configure from the Run list box.
2. In the “Select Operation” box, select Verify.
3. Using an external test fixture, apply a known pressure. When the
Reported Value is steady, enter the pressure you applied in the Applied Value field, then click on Verify Static Pressure Calibration. Repeat this process using the external test fixture to
apply known pressures of 0, 50, 100, 80, 20, and 0 percent of span.
Issued Nov-2016 3530 TeleFlow Calibration 8-3
TechView User’s Guide (D5131)
Note: If, during the verification procedure, the transmitter is found to
be out of calibration, perform the calibration procedure to correct the discrepancy.

8.5 Calibration of RTD Temperature

Figure 8-3. T tab – Calibration of RTD Temperature
1. If this flow computer is handling more than one meter run, first
select the run you want to configure from the Run list box.
2. In the Select Operation box, select Calibrate.
3. You have the option of displaying the temperature in either
Fahrenheit or Celsius; to switch the units displayed, click on the Change Units button.
4. Disconnect the regular RTD and connect a 100 Ohm (0.01%)
precision resistor across the RTD terminals (this is equivalent to 32°F).
5. Observe the Live Reading and wait for it to stabilize.
6. Click on the Calibrate Zero (at 100 OHMs) button to calibrate the
RTD zero.
8-4 3530 TeleFlow Calibration Issued Nov-2016
TechView User’s Guide (D5131)
7. Disconnect the resistor of step 4 and connect a 140.23 Ohm
(0.01%) precision resistor across the RTD terminals (this is equivalent to 220°F).
8. Observe the Live Reading and wait for it to stabilize.
9. Enter the desired operating span in the Operating Span field.
10. Click on the Calibrate Operating Span button.
11. Disconnect the resistor, and reconnect the RTD.
12. If after re-connecting the unit to the process it is determined that the
number in the Live Reading field does not correlate with that measured by an external process-monitoring device, offset compensation may be performed. To do this, enter the desired reading, and click on the Adjust Live Reading button.
Figure 8-4. TeleFlow CPU Board RTD Calibration Connection Diagram (See page 5-7 for resistor
values)
Issued Nov-2016 3530 TeleFlow Calibration 8-5
TechView User’s Guide (D5131)
8.5.1 Verification of RTD Temperature
1. If this flow computer is handling more than one meter run, first
2. In the “Select Operation” box, select Verify.
3. Disconnect the regular RTD and connect a 100 Ohm (0.01%)
4. The regular RTD must be disconnected, and replaced with a resistor
5. When the Reported Value is steady, enter the value you applied in
6. Repeat this process with other test values (typically 0, 50, 100, 80,
select the run you want to configure from the Run list box.
precision resistor across the RTD terminals (this is equivalent to 32°F).
which simulates the desired temperature.
the Applied Value field, then click on the Verify Temperature Calibration button.
20, 0 percent of scale values would be applied.)

8.6 Damping

The output damping feature controls the rate at which the output responds to a given change of input. It is used to slow down the output response to a rapid or oscillatory change of the measured variable.
Figure 8-5. Damping tab
The Damping Time is a period of time during which the indicated value changes only 63% of the difference between the “present measured variable” and the “present indicated pressure” in one damping time period. It would take 5 times the DP floating point damping time
8-6 3530 TeleFlow Calibration Issued Nov-2016
for the “present indicated pressure” to equal the “present measured pressure” (if a change in the “present measured pressure” didn’t occur).
An increase in the DP floating point damping time results in an increased smoothing of the indicated value.
Note: In the case of a model 3530-series unit, if the external transmitter
is a model 3808, damping mode is always enabled and can’t be changed; damping time can’t be changed either.
8.6.1 Configuring the Damping Time
1. If this flow computer is handling more than one meter run, first select
the run you want to configure from the Run list box.
2. Enable the damping feature (if it isn’t already active) by clicking on
the Change DP Floating Point Damping Mode button. (The currently active mode is displayed in the Damping Mode field).
3. To change the Damping Time enter a new damping time (in
seconds) in the Damping Time field, then click on the Change DP Floating Point Damping Time button.
TechView User’s Guide (D5131)
Note: If you later decide to disable the damping feature, click again on
Change DP Floating Point Damping Mode.

8.7 Calibrating TeleRecorder Inputs (3530-45B/55B ONLY)

Important
The “Inputs” page provides for gage pressure calibration of inputs to remove the effect of sensor zero offset and slope variation. You must calibrate the zero first.
Gage pressure calibration requires the following items: Instrument mounting fixture: This fixture must secure the instrument
such that the presure transducer is mounted vertically.
a pressure gauge (must cover the range of the pressure transducer) an adequate pressure regulator a pressure source and interconnect lines
A test fixture that incorporates the items listed above may be used. A typical 3530-series pressure transducer calibration interconnection diagram is provided, below:
Issued Nov-2016 3530 TeleFlow Calibration 8-7
TechView User’s Guide (D5131)
Figure 8-6. 3530-45B/-55B TeleRecorder Pressure Transducer Connections
Figure 8-7. Main tab
8.7.1 Calibrating the Input
Note: Calibration mode only remains active for the length of time
indicated in the Calibrate Time field. Otherwise, Zero Offset and Span Factor will be set to 0.000 and 1.000, respectively.
1. First, calibrate the ZERO. You should apply atmospheric pressure
when calibrating the ZERO. Click on Calibrate Zero. The load calculates a Zero Offset that will be added to the sensor reading to
8-8 3530 TeleFlow Calibration Issued Nov-2016
TechView User’s Guide (D5131)
remove any existing offset at zero pressure. For example, a 1000 psig sensor may be reading 1002 psig with no pressure applied. After calibration a -2.0 zero offset will be used on the Input Value to remove the +2 psig offset.
2. To calibrate the span, apply 25% to 100% of sensor URL pressure,
and enter the applied value in the Applied Pressure field, then click the Calibrate Span button to do a span factor calculation. The load calculates a Span Factor that will be applied to the Input Value reading to compensate for slope variation. For example, with zero offset already calculated and 50 psig applied a 100 psig sensor may be reading 49 psig because the sensor slope is slightly off. The load will calculate and apply a span correction factor of 50/49 = 1.0205 to the Input Value to compensate for the slope variation. THIS CALIBRATION DOES NOT LINEARIZE THE SENSOR INPUT READING - IT ONLY COMPENSATES FOR ZERO OFFSET AND SLOPE VARIATION. The result of applying zero offset and span factor to the Input Value is the Value in Use. The ACCOL load executes every 15 seconds; so allow time for results to appear.
Note: If desired, you can clear the Zero Offset and Span Factor values
by clicking on Restore Defaults; the offset and span factors are set to 0.0 and 1.0 respectively.
Issued Nov-2016 3530 TeleFlow Calibration 8-9
TechView User’s Guide (D5131)
Chapter 9 – Calibrating the ControlWave EFM / GFC-CL
/GFC/ XFC
This chapter covers calibration activities for the ControlWave EFM, GFC, XFC, and GFC-CL.
For a general overview of calibration concepts, see Chapter 5. For instructions on starting calibration mode, see Section 5.4.4.
In This Chapter
9.1Calibration of Differential/Gage Pressure ........................................ 9-1
9.2Verification of Differential/Gage Pressure ....................................... 9-2
9.3Calibration of Static Pressure .......................................................... 9-4
9.4Verification of Static Pressure .......................................................... 9-5
9.5Calibration of RTD Temper ature ...................................................... 9-6
9.6Verification of Temperature ............................................................. 9-9
9.7Damping ......................................................................................... 9-11
9.7.1Configuring the Damping Time ........................................... 9-11
9.8Orifice ............................................................................................. 9-12
9.8.1Specifying the Orifice Plate Size ........................................ 9-12

9.1 Calibration of Differential/Gage Pressure

Figure 9-1. DP tab – Calibration of Differential / Gage Pressure
Issued Nov-2016 ControlWave EFM/GFC/XFC Calibration 9-1
TechView User’s Guide (D5131)
1. Select the transmitter you want to calibrate using the Transmitter
2. In the “Select Operation” box, select Calibrate.
3. To perform the zero calibration, vent the transmitter to atmosphere,
4. Wait for the Live Value to become steady, and enter a Tester Value
5. Click on Calibrate Zero and the Live Value will be trimmed to be
6. To calibrate the span, apply a pressure equivalent to the desired
list box. “Wet ends” refer to internal transmitters; all other transmitters are external.
so no pressure is applied.
equivalent to the zero value.
as close to the zero as possible. If you are calibrating the internal transmitter, the Calibration Status will be set to SUCCESS or FAIL based on whether the trim operation was successful. Calibration status is not available for external transmitters. The page displays the Live Value and Tester Value in the Actual and Tester value fields, respectively, for the Zero test point.
span, and wait for the Live Value to settle.
7. Now enter the pressure you applied in the Tester Value field and
click on Calibrate Span. The Live Value will be trimmed to be as close to the desired span, as possible. If you are calibrating the internal transmitter, the Calibration Status will be set to SUCCESS or FAIL based on whether the trim operation was successful. Calibration status is not available for external transmitters. The page displays the Live Value and Tester Value in the Actual and Tester value fields, respectively, for the span test point.
Note: If, for some reason, you want to return to the calibration that was
performed by the factory, click on Restore Factory Defaults.

9.2 Verification of Differential/Gage Pressure

You perform verification as a check to see whether calibration is required, or after calibration is completed to verify proper calibration. You choose As Found to record that verification readings are before calibration, or As Left to record that verification readings are after calibration.
For a DP, you must perform verification at a minimum of two points, typically at zero (the high and low side of the transmitter equalized) and with a differential pressure applied across the high and low sides of the transmitter (typically, the upper range limit). Typically you would use three, five, or seven verification points to verify differential pressure.
9-2 ControlWave EFM/GFC/XFC Calibration Issued Nov-2016
TechView User’s Guide (D5131)
Figure 9-2. DP tab – Verification of Differential / Gage Pressure
1. Select the transmitter you want to verify using the Transmitter list
box. “Wet ends” refer to internal transmitters; all other transmitters are external.
2. In the “Select Operation” box, select Verify.
3. If you are performing the verification prior to calibration, choose As
Found. If you are performing the verification after calibration, choose As Left. This choice determines where readings are stored for the calibration report.
4. Depending upon your application, you may have multiple test points
to check in sequence. For the first test point, apply a known pressure using a deadweight tester, or similar device and let the reading stabilize. Enter the Tester Value which represents the known pressure value you applied, then click on the button labeled with that test point. The grid in the right side of the screen displays the Actual Value read from the transmitter, the Tester Value you entered, and then calculates the difference between these values and displays it in the Deviation field. The cursor advances to the next test point.
5. Repeat step 4 for each additional test point. If you make a mistake
and want to re-run the verification for a particular test point, click the Back button. If you want to skip over a test point, click the Skip Test Point button.
Issued Nov-2016 ControlWave EFM/GFC/XFC Calibration 9-3
TechView User’s Guide (D5131)

9.3 Calibration of Static Pressure

Note: For static pressure calibration setup, connect identical supply
pressures to both the low and high side ports of the MVT.
Figure 9-3. SP tab – Calibration of Static Pressure
1. Select the transmitter you want to calibrate using the Transmitter
list box. “Wet ends” refer to internal transmitters; all other transmitters are external.
2. To perform the zero calibration, vent the transmitter to atmosphere,
so atmospheric pressure is applied. Wait for the Live Value to become steady, and enter a Tester Value equivalent to the zero value.
3. Click on Calibrate Zero and the Live Value will be trimmed to be
as close to the zero as possible. If you are calibrating the internal transmitter, the Calibration Status will be set to SUCCESS or FAIL based on whether the trim operation was successful. Calibration status is not available for external transmitters. The page displays the Live Value and Tester Value in the Actual and Tester value fields, respectively, for the Zero test point.
4. To calibrate the span, apply a pressure equivalent to the desired
span, and wait for the Live Value to settle. Now enter the pressure you applied in the Tester Value field and click on Calibrate Span. The Live Value will be trimmed to be as close to the desired span, as possible. If you are calibrating the internal transmitter, the Calibration Status will be set to SUCCESS or FAIL based on whether the trim operation was successful. Calibration status is not available for external transmitters. The page displays the Live Value and Tester Value in the Actual and Tester value fields, respectively, for the Span test point.
9-4 ControlWave EFM/GFC/XFC Calibration Issued Nov-2016
Note: If, for some reason, you want to return to the calibration that was
performed by the factory, click on Restore Factory Defaults.

9.4 Verification of Static Pressure

You perform verification as a check to see whether calibration is required, or after calibration is completed to verify proper calibration. You choose As Found to record that verification readings are before calibration, or As Left to record that verification readings are after calibration.
For static pressure transmitters, you must perform verification at a minimum of two points, typically at zero (the transmitter vented to atmosphere) and under pressure (often, the upper range limit, or if that is not practical, near the typical operating range of the transmitter).
TechView User’s Guide (D5131)
Figure 9-4. SP tab – Verification of Static Pressure
1. Select the transmitter you want to verify using the Transmitter list
box. “Wet ends” refer to internal transmitters; all other transmitters are external.
2. In the “Select Operation” box, select Verify.
3. If you are performing the verification prior to calibration, choose As
Found. If you are performing the verification after calibration, choose As Left. This choice determines where reading are stored for the calibration report.
Issued Nov-2016 ControlWave EFM/GFC/XFC Calibration 9-5
TechView User’s Guide (D5131)
4. Depending upon your application, you may have multiple test points
to check in sequence. For the first test point, apply a known pressure using a deadweight tester, or similar device and let the reading stabilize. Enter the Tester Value which represents the known pressure value you applied, then click on the button labeled with that test point. The grid in the right side of the screen displays the Actual Value read from the transmitter, the Tester Value you entered, and then calculates the difference between these values and displays it in the Deviation field. The cursor advances to the next test point.
5. Repeat step 4 for each additional test point. If you make a mistake
and want to re-run the verification for a particular test point, click the Back button. If you want to skip over a test point, click the Skip Test Point button.

9.5 Calibration of RTD Temperature

Figure 9-5. T tab - Calibration of RTD Temperature
1. Select the transmitter you want to calibrate using the Transmitter
list box. “Wet ends” refer to internal transmitters; all other transmitters are external.
2. Disconnect the regular RTD and connect a 100 Ohm (0.01%)
precision resistor across the RTD terminals (which is equivalent to 32°F)
3. Observe the Live Value and wait for it to stabilize.
4. Click on the Calibrate Zero (at 100 ohm) button to calibrate the
RTD zero. If you are calibrating the internal transmitter, the
9-6 ControlWave EFM/GFC/XFC Calibration Issued Nov-2016
TechView User’s Guide (D5131)
Calibration Status will be set to SUCCESS or FAIL based on whether the trim operation was successful. Calibration status is not available for external transmitters.
5. Disconnect the resistor of step 2 and connect a precision resistor
across the RTD terminals to represent your desired span. To calibrate to the default temperature span value of 558 °C (1036
°F) requires a resistance value of 300 ohms +/- 0.01%. To calibrate the temperature span to any other desired value, you
must know the proper resistance value for that span. You must calculate this yourself or refer to a 100 ohm platinum RTD chart. For example, to calibrate the span to the upper range limit (URL) of the transmitter (660 °C, 1220 °F) requires a 332 ohm resistance +/-
0.01%. After you connect the required resistance, enter the desired temperature span value in the Custom Span field.
6. Observe the Live Value and wait for it to stabilize.
7. If you used a 300-ohm resistance in step 5, click the Calibrate Span
(at 300 ohm) button; if you connected any other resistance value click the Calibrate Custom Span button. In either case, if you are calibrating the internal transmitter, the Calibration Status will be set to SUCCESS or FAIL based on whether the trim operation was successful. Calibration status is not available for external transmitters. Disconnect the resistor, and reconnect the RTD.
8. To change the reading from the RTD, enter the desired reading in
the Live Reading field (which should be similar to the actual reading, but presumably off slightly), and click on Adjust Live Reading. The page displays the actual reading, and the live reading you entered in the Actual and Tester fields, respectively.
Note: If, for some reason, you want to return to the calibration that was
performed by the factory, click on Restore Factory Defaults.
Issued Nov-2016 ControlWave EFM/GFC/XFC Calibration 9-7
TechView User’s Guide (D5131)
.
TB6
RTD
}
Input
TB2 RTD Input
Control
GFC-CL
Battery Charger
& I/O Bd.
Wave
TB10
RTD
Input
}
Control
Wave
GFC
CPU/System
Controller Bd
Figure 9-6. Connections for RTD Temperature Calibration – ControlWave EFM, ControlWave GFC-
CL, ControlWave GFC (See page 5-7 for resistor values)
9-8 ControlWave EFM/GFC/XFC Calibration Issued Nov-2016
RTD+ ( E xci tati on) RTD+ ( S ens e)
_
RTD (Return)
TechView User’s Guide (D5131)
Figure 9-7. Connections for RTD Temperature Calibration – ControlWave XFC (see page 5-7 for
resistor values)

9.6 Verification of Temperature

You perform verification as a check to see whether calibration is required, or after calibration is completed to verify proper calibration. You choose As Found to record that verification readings are before calibration, or As Left to record that verification readings are after calibration.
For temperature transmitters, you can perform verification at two points, but more often, you would use single point verification. You perform the single point verification by measuring the process temperature with a calibrated thermometer, and comparing the results to the temperature indicated by the temperature transmitter.
Issued Nov-2016 ControlWave EFM/GFC/XFC Calibration 9-9
TechView User’s Guide (D5131)
Figure 9-8. T tab – Verification of Temperature
1. Select the transmitter you want to verify using the Transmitter list
box. “Wet ends” refer to internal transmitters; all other transmitters are external.
2. In the “Select Operation” box, select Verify.
3. If you are performing the verification prior to calibration, choose As
Found. If you are performing the verification after calibration, choose As Left. This choice determines where reading are stored for the calibration report.
4. Depending upon your application, you may have multiple test points
to check in sequence. For the first test point, apply a known pressure using a deadweight tester, or similar device and let the reading stabilize. Enter the Tester Value which represents the known pressure value you applied, then click on the button labeled with that test point. The grid in the right side of the screen displays the Actual Value read from the transmitter, the Tester Value you entered, and then calculates the difference between these values and displays it in the Deviation field. The cursor advances to the next test point.
5. Repeat step 4 for each additional test point. If you make a mistake
and want to re-run the verification for a particular test point, click the Back button. If you want to skip over a test point, click the Skip Test Point button.
9-10 ControlWave EFM/GFC/XFC Calibration Issued Nov-2016

9.7 Damping

TechView User’s Guide (D5131)
The output damping feature controls the rate at which the output responds to a given change of input. It is used to slow down the output response to a rapid or oscillatory change of the measured variable.
Figure 9-9. Damping tab
The Damping Time is a period of time during which the indicated value changes only 63% of the difference between the “present measured variable” and the “present indicated pressure” in one damping time period. It would take 5 times the DP Floating Point Damping Time for the “present indicated pressure” to equal the “present measured pressure” (if a change in the “present measured pressure” didn’t occur).
An increase in the DP Floating Point Damping Time results in an increased smoothing of the indicated value.
Note: If the External Transmitter is a model 3808, damping mode is
always enabled and can’t be changed; damping time can’t be changed either.
9.7.1 Configuring the Damping Time
1. Select the transmitter you want to change the damping time for
using the Transmitter list box. “Wet ends” refer to internal transmitters; all other transmitters are external.
2. To change the Damping Time enter a new damping time (in
seconds) in the Damping Time field, then click on the Change DP Floating Point Damping Time button. This writes the new
damping time to your application in the ControlWave.
Issued Nov-2016 ControlWave EFM/GFC/XFC Calibration 9-11
TechView User’s Guide (D5131)

9.8 Orifice

If you change the orifice plate on your pipe, you can use the Orifice tab to update your application with the new orifice plate diameter..
Figure 9-10. Orifice tab
9.8.1 Specifying the Orifice Plate Size
1. Select the transmitter using the Transmitter list box. “Wet ends”
refer to internal transmitters.
2. Enter the New Plate Diameter and click the Apply new Orifice
Plate Diameter button. This writes the new orifice diameter to your
application in the ControlWave. The Current Plate Diameter field updates with the new value.
9-12 ControlWave EFM/GFC/XFC Calibration Issued Nov-2016
TechView User’s Guide (D5131)
Chapter 10 – Configuring and Calibrating the 4088B
This chapter covers configuration and calibration activities for the Rosemount 4088B transmitter.
For a general overview of calibration concepts, see Chapter 5. For instructions on starting calibration mode, see Section 5.4.4.
In This Chapter
10.1Configuring the 4088B ................................................................... 10-1
10.1.1DP/GP Pressure ................................................................. 10-2
10.1.2Static Pressure ................................................................... 10-2
10.1.3Temperature ....................................................................... 10-3
10.1.4Serial Port Setup ................................................................ 10-4
10.1.5Transmitter Data ................................................................. 10-5
10.1.6RTD Coefficients ................................................................ 10-6
10.2Calibrating the 4088B .................................................................... 10-6
10.2.1Calibration of Gage or Differential Pressure ...................... 10-7
10.2.2Calibration of Static Pressure ............................................. 10-7
10.2.3Calibration of RTD Temperature ........................................ 10-9

10.1 Configuring the 4088B

Click the Configuration icon to bring up the Config tab. The Config tab includes buttons for calling up menus to configure the
sensor.
Figure 10-1. Config tab
Note: The menus may be re-sized so they may appear different from the
illustrations shown here.
Issued Nov-2016 4088B Configuration and Calibration 10-1
TechView User’s Guide (D5131)
10.1.1 DP/GP Pressure
Configuration options for the differential pressure variable consist of setting the damping, selecting the engineering units, and setting lower and upper range values.
Figure 10-2. Configuring the Differential/Gage (DP/GP) Pressure Variable
1. Click DP/GP Pressure to open the DP/GP Pressure menu.
2. Specify a Damping value from 0.0 (none) to 30.0 seconds. For a
step change in input it takes five times the damping value for the reading to reach the new input. For example, an input change from 60 inH2O to 70 inH2O will take five seconds when damping is set to 1. Similarly, an input change from 60 inH2O to 70 inH2O will take 20 seconds when damping is set to 4.
3. Select the engineering Units for the differential pressure variable.
4. Specify the Lower Range Value (LRV). The LRV should be the
lowest value for the DP/GP variable you want to measure for your process. This cannot be less than the Lower Range Limit (LRL) which represents the factory set lowest value for the range of values the sensor can read.
5. Specify the Upper Range Value (URV). The URV should be the
highest value for the DP/GP variable you want to measure for your process. This cannot be more than the Upper Range Limit (URL) which represents the factory set highest value for the range of values the sensor can read.
10.1.2 Static Pressure
Configuration options for the static pressure variable consist of setting the damping, selecting the engineering units, and setting lower and upper range values.
10-2 4088B Configuration and Calibration Issued Nov-2016
TechView User’s Guide (D5131)
Figure 10-3. Configuring the Static (SP) Pressure Variable
1. Click Static (SP) Pressure to open the Static (SP) Pressure menu.
2. Click Set On to enable SP conversion; click Set Off to disable SP
conversion.
3. Specify a Damping value from 0.0 (none) to 30.0 seconds. For a
step change in input it takes five times the damping value for the reading to reach the new input. For example, an input change from 60 psi to 70 psi will take five seconds when damping is set to 1. Similarly, an input change from 60 psi to 70 psi will take 20 seconds when damping is set to 4.
4. Select the engineering Units for the static pressure variable.
5. Specify the Lower Range Value (LRV). The LRV should be the
lowest value for the SP variable you want to measure for your process. This cannot be less than the Lower Range Limit (LRL) which represents the factory set lowest value for the range of values the sensor can read.
6. Specify the Upper Range Value (URV). The URV should be the
highest value for the SP variable you want to measure for your process. This cannot be more than the Upper Range Limit (URL) which represents the factory set highest value for the range of values the sensor can read.
10.1.3 Temperature
Configuration options for the temperature variable consist of setting the damping, selecting the engineering units, and setting lower and upper range values.
Issued Nov-2016 4088B Configuration and Calibration 10-3
TechView User’s Guide (D5131)
1. Click Temperature to open the Temperature menu.
2. Click Set On to enable RTD conversion; click Set Off to disable
Figure 10-4. Configuring the Temperature Variable
RTD conversion.
3. Choose either 3-Wire or 4-Wire based on the type of RTD you are
using.
4. Specify a Damping value from 0.0 (none) to 30.0 seconds. For a
step change in input it takes five times the damping value for the reading to reach the new input. For example, an input change from 60 degrees to 70 degrees will take five seconds when damping is set to 1. Similarly, an input change from 60 degrees to 70 degrees will take 20 seconds when damping is set to 4.
5. Select the engineering Units for the temperature variable.
6. Specify the Lower Range Value (LRV). The LRV should be the
lowest value for the temperature variable you want to measure for your process. This cannot be less than the Lower Range Limit (LRL) which represents the factory set lowest value for the range of values the sensor can read.
7. Specify the Upper Range Value (URV). The URV should be the
highest value for the temperature variable you want to measure for your process. This cannot be more than the Upper Range Limit (URL) which represents the factory set highest value for the range of values the sensor can read.
8. If you want the sensor to report a preset temperature value if the
RTD connection fails due an open or short, set the Fail-To option to ON, and enter the preset temperature value you want to use during a failure in the Fail-To value field.
10.1.4 Serial Port Setup
The Serial Port Setup menu lets you configure the RS-485 serial port for BSAP communication.
10-4 4088B Configuration and Calibration Issued Nov-2016
TechView User’s Guide (D5131)
Figure 10-5. Configuring the Serial Port
1. Specify the BSAP local address for the sensor in the Node Address
field.
2. If your network uses expanded BSAP addressing (EBSAP), enter
the EBSAP group number in the Group Number field; otherwise set to 0.
3. Choose the Local Port Address Mode.
4. Specify the Baud Rate the serial port uses to communicate.
5. Specify the Transmit Delay in milliseconds. This is the amount of
time the sensor will wait before it responds to a request.
6. Ignore the MODBUS Address; this is not useful for the 4088B.
10.1.5 Transmitter Data
The Transmitter Data menu shows certain identifying information about the sensor.
Figure 10-6. Viewing Transmitter Identification Information
Field Description Transmitter
Tagname
Issued Nov-2016 4088B Configuration and Calibration 10-5
Shows the user-configurable tag name for the transmitter.
TechView User’s Guide (D5131)
Transmitter Serial Number
Sensor Block Number
Firmware Revision
10.1.6 RTD Coefficients
The sensor computes RTD resistance using the R0 value and the A and B coefficients in the modified DIN equation RT = R0 * (1+A*T+B*T^2). The sensor then calculates RTD temperature from the resistance.
Shows the transmitter serial number.
Shows the transmitter block number.
Shows the revision of firmware running in the transmitter.
Field Description RTD Temperature
A Coefficient
B Coefficient
R0 Value
Restore Factory Defaults

10.2 Calibrating the 4088B

Figure 10-7. Configuring the RTD
The RTD temperature computed from the RTD resistance.
Shows the A coefficient currently in use. If needed, you can specify a different A coefficient to use in the (A*T) term.
Shows the B coefficient currently in use. If needed, you can specify a different B coefficient to use in the (B*T^2) term.
Shows the R0 value currently in use. If needed, you can specify a different R0 value to use in the equation.
Click here to restore the factory default values for the A, B, and R0 coefficients. These are the coefficients for a platinum RTD with an “alpha” of 0.00385.
Click the Calibration icon to bring up the Calibration tabs.
10-6 4088B Configuration and Calibration Issued Nov-2016
TechView User’s Guide (D5131)
10.2.1 Calibration of Gage or Differential Pressure
The Sensor LRL and Sensor URL display the lower-range limit and upper range limit, respectively, of the gage pressure / differential pressure sensor.
Figure 10-8. Calibrating Gage or Differential Pressure
1. In the “Select Operation for DP Sensor” box, select Zero
Calibration.
2. To perform the zero calibration, vent the transmitter to atmosphere,
so no pressure is applied. Click on the Calibrate Zero button, and the Measured Value will be trimmed to be as close to the zero as possible.
3. To calibrate the span, select Span Calibration in the “Select
Operation for DP Sensor” box, then apply a pressure equivalent to the desired span. Now enter the pressure you applied in the New
Applied Value field. If the transmitter accepts the New Applied Value, the value you entered will be displayed in the Target Trim Value field. This value must be less than or equal to the Sensor URL value. Click on Calibrate Span. The Measured Value will be
trimmed to be as close to the desired span, as possible. Note: Span calibration may only be performed successfully if the applied pressure and span are within +3% of one another.
Note: If, for some reason, you want to return to the calibration that was
performed by the factory, click on Restore Factory Defaults.
10.2.2 Calibration of Static Pressure
Note: For static pressure calibration setup, connect identical supply
pressures to both the low and high side ports of the MVT.
Issued Nov-2016 4088B Configuration and Calibration 10-7
TechView User’s Guide (D5131)
The Sensor LRL and Sensor URL display the lower-range limit and upper range limit, respectively, of the static pressure sensor.
1. In the “Select Operation for SP Sensor” box, select Zero
Figure 10-9. SP tab - Calibration of Station Pressure
Calibration.
2. To perform the zero calibration, vent the transmitter to atmosphere,
so only atmospheric pressure is applied. Click on the Calibrate Zero button, and the Measured Value will be trimmed to be as
close to the zero as possible.
3. To calibrate the span, select Span Calibration in the “Select
Operation for SP Sensor” box, then apply a pressure equivalent to the desired span. Now enter the pressure you applied in the New
Applied Value field. If the transmitter accepts the New Applied Value, the value you entered will be displayed in the Target Trim Value field. This value must be less than or equal to the “Sensor URL” value. Click on Calibrate Span. The Measured Value will
be trimmed to be as close to the desired span, as possible. Note: Span calibration may only be performed successfully if the applied pressure and span are within +3% of one another.
Note: If, for some reason, you want to return to the calibration that was
performed by the factory, click on Restore Factory Defaults.
10-8 4088B Configuration and Calibration Issued Nov-2016
10.2.3 Calibration of RTD Temperature
Figure 10-10. T tab – Calibration of RTD Temperature
1. Disconnect the regular RTD and connect a 100 Ohm (0.01%)
precision resistor across the RTD terminals (this is equivalent to 32°F).
TechView User’s Guide (D5131)
2. Observe the Measured Value and wait for it to stabilize.
3. Click the Calibrate Zero button to calibrate the RTD zero.
4. Disconnect the resistor of step 1 and connect a precision resistor
across the RTD terminals to represent your desired span. To calibrate to the default temperature span value of 558 °C (1036
°F) requires a resistance value of 300 ohms +/- 0.01%. To calibrate the temperature span to any other desired value, you
must know the proper resistance value for that span. You must calculate this yourself or refer to a 100 ohm platinum RTD chart. For example, to calibrate the span to the upper range limit (URL) of the transmitter (660 °C, 1220 °F) requires a 332 ohm resistance +/-
0.01%. After you connect the required resistance, enter the desired temperature span value in the Custom Span field.
5. Observe the Live Value and wait for it to stabilize.
6. If you connected a 300-ohm resistance in step 4, click the Calibrate
Span (at 300 ohm) button; if you connected any other resistance value click the Calibrate Custom Span button.
7. Disconnect the resistor, and reconnect the RTD. Notes:
If, for some reason, you want to return to the calibration that was
performed by the factory, click on Restore Factory Defaults.
To change the reading from the RTD for the 4088B, enter the
desired reading (which should be similar to the actual reading, but presumably off slightly), and click on Adjust Live Reading.
Issued Nov-2016 4088B Configuration and Calibration 10-9
TechView User’s Guide (D5131)
Chapter 11 – Online Editing (ControlWave only)
TechView supports online editing of signal lists, and archive file structures in ControlWave-series controllers with version 04.60 (or
In This Chapter

11.1 On-line Editing of Signal Lists

newer) firmware.
10.1On-line Editing of Signal Lists ........................................................ 11-1
10.2On-line Editing of Archive Files ...................................................... 11-4
10.2.1Archive Fields ..................................................................... 11-6
10.2.2Batch Editing of Archive Files ............................................. 11-7
Signal lists are a convenient way to organize and view signals used in your control strategy. For example, you might have a list of configuration parameters that apply to a particular portion of the control strategy; or you might have a list containing the current values of various process variables.
Note: Beginning with OpenBSI 5.8 Service Pack 1, lists can contain
up to 10,000 signals; earlier versions limited this to 5000 signals.
If your controller is running a standard application, created by Emerson, certain signals will already be stored in signal lists for you to look at. You may decide, however, that there are other signals in the application that you want to add to an existing list, or maybe there are signals you want to remove from a list, because you don’t use them much. These changes can be made via on-line signal list editing.
Issued Nov-2016 Online Editing 11-1
TechView User’s Guide (D5131)
By default, online signal list editing is accessed by clicking on the On Line Edits group icon.
These are all the signals in the currently selected list. Signals appear in the order they are defined in the list; you cannot sort them.
This is a list of other signals in the application that are available for you to add to the currently selected list.
First, select the list you want to modify here. If “Description”
You can search for particular signals based on their signal name or signal descriptor. Choose either “Search for Signal” or “Search for Descriptor” and type a partial signal name or descriptor;
TechView automatically scrolls the list to a matching signal or descriptor and displays a “hand icon”
signal.
is blank, no description
Use the scroll bars to bring additional signals into view.
You can sort the available signals alphabetically by signal name or descriptor by clicking on
next to a matching
the label.
= Sort alphabetically from top
= Sort
alphabetically from bottom
Figure 11-1. Lists tab – Online Editing of Signal Lists
Field Description Select a list to
modify
Search for Signal / Search for Descriptor
List number n
Available Signals
Delete
This box displays all the lists marked for on-line modification in the RTU. Lists are shown either by their list number, or a textual description. Choose the list you want to modify. This will be referred to as the “currently selected list”.
To locate a particular signal (variable), you can search based on the signal name or the signal descriptor. Choose either Search for Signal or Search for Descriptor then enter partial search text in the sea rch fiel d to the right. TechView searches through the list and scrol ls the window to the first mat ch.
It displays a hand descriptor. (OpenBSI 5.8 and newer.)
This list shows all the signals included in the currently selected list you chose previously under Select a list to modify. Signals are displayed in the order they appear in the list, along with their associated signal descriptor, if configured. (If no signal descriptor was configured, the descriptor field is left blank.) If necessary, use the scroll bar to bring more signals from the list into view.
This is a list of all signals in the RTU that have been marked as “PDD”, along with their associated signal description. (If no signal descriptor was configured, the descriptor field is left blank.) Any of these available signals may be added to the currently selected list, or may be used to replace a particular signal already in the list. You may sort the signals in this list either by signal name, or signal description by clicking on the “Signal” or “Description” titles at the top of the list box. To remove one or more signals from the currently selected list, click on
icon next to the matching signal name or signal
11-2 Online Editing Issu ed Nov-2016
TechView User’s Guide (D5131)
the signals to be removed, so they are highlighted, and then click on the Delete button. As in standard Windows, you can hold down the Ctrl key to select multiple signals, individually, or select the first signal in a range, and then select the final signal in the range while clicking on Shift. Once you click on Delete the highlighted signals will be removed from the currently selected list. Note: The signal is still in the control st rategy file, but once you send the modified list to the RTU, it will no longer be part of this list.
Move Up
Move Down
Insert
Replace
This button allows you to change the position of the currently selected signal so that is moved one position closer to the top of the list. Click on the signal you want to move, then click on Move Up and the signal will be moved one position up in the list.
This button allows you to change the position of the currently selected signal so that is moved one position closer to the end of the list. Click on the signal you want to move, then click on Move Down and the signal will be moved one position down in the list.
This button allows you to add any signals selected from the Available
Signals into the currently selected list. To do this, click on signals in the Available Signals box, so they are highlighted, then click on the Insert
button. As in standard Windows, you can hold down the Ctrl key to select multiple signals, individually, or select the first signal in a range, and then select the final signal in the range while clicking on Shift. Once you click on Insert the signals will be added to the currently selected list. If a signal is already highlighted the list, TechView inserts the signals immediately below it. If no signal is highlighted in the list, or multiple signals are already highlighted in the list, TechView inserts the signals at the top of the list. If desired, you can move them, one at a time, to a different position in the list by clicking on the signal, then using the Move Up or Move Down buttons.
This button allows you to replace any signal in the currently selected list, with a signal from the Available Signals box. To do this, click on the signal you want to be replaced, in the currently selected list, so it is highlighted, then click on the signal you want to replace it with in the Available Signals box, then finally click on the Replace button. The highlighted signal in the currently selected list will be replaced with the highlighted signal from the Available Signals box.
Send list to RTU
When you have finished modifying the signal list, click on Send list to RTU and the modified list will be sent to the controller, thereby completing
you online edits.
What if I’m not seeing any lists or signals at all?
You can only perform online editing for lists created with initialization files; lists generated with the LIST function block in ControlWave Designer cannot be edited online through TechView.
If, when you bring up the “Lists” on-line editing page, there are no signal lists in the Select a list to modify selection box, and consequently, no signals in the Available Signals list box, it’s probably because your translation initialization file has not been properly specified for TechView.
To specify the translation initialization file, click on the Session Parameters icon, and click on Next in the Communications Setup dialog box. In the Node Setup dialog box, click on the Advanced Interface Setup button and specify the translation initialization file.
For details on the format of INI files, please see Appendix A.
Issued Nov-2016 Online Editing 11-3
.
TechView User’s Guide (D5131)

11.2 On-line Editing of Archive Files

Through on-line editing of archive files, you can: Change the number of records (rows) that will be saved in an
existing archive file.
Add, delete, or modify the column definitions for an existing archive
file.
You cannot, however, create new archive files, nor can you edit saved archive file data.
Important
Archive files larger than 1 sector cannot be edited online. To determine whether or not your archive file exceeds 1 sector, please review the Historical System overview section of the ACCOL3 online help in ControlWave Designer.
The lists identified by the iiOutList and iiArchiveList parameters of the Archive function block must be modifiable, in order for your archive structure to be modified.
ASCII archive files (introduced in ControlWave firmware version 05.10) cannot be edited online.
Figure 11-2. Archives tab – Online editing of Archives
When you call up the Archives tab, information on archive files is loaded.
Note: If you see archive information colored red, there is a
configuration error.
11-4 Online Editing Issu ed Nov-2016
Loading...