Remote Automation Solutions Gas Chromatograph Software Interface Manual Manuals & Guides

Remote Automation Solutions
Form A6248
Part Number D301345X012 June 2017
Gas Chromatograph Software Interface User Manual (for FloBoss 107)
Gas Chromatograph Software Interface User Manual (for FB107)
ii Revised June-2017

Revision Tracking Sheet

June 2017
This manual may be revised periodically to incorporate new or updated information. The revision date of each page appears at the bottom of the page opposite the page number. A change in revision date to any page also changes the date of the manual that appears on the front cover. Listed below is the revision date of each page (if applicable):
Page
Revision
All pages
June-2017
All pages
February-2010
All pages
February-2009
All pages
October-2008
All pages
September-2008
Page 5, 7
February-2008
Initial Release
January-2008
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Contents iii

Contents

Chapter 1 Introduction 1
1.1 Scope and Organization ..................................................................................................................... 1
1.2 Product Overview ............................................................................................................................... 1
1.2.1 Communications Wiring ......................................................................................................... 2
1.2.2 Auto-configure ........................................................................................................................ 3
1.2.3 Validating GC Data ................................................................................................................ 5
1.3 Program Requirements ...................................................................................................................... 7
1.3.1 License Key ............................................................................................................................ 7
Chapter 2 – Installation 9
2.1 Installing the License Key .................................................................................................................. 9
2.2 Downloading the Program ................................................................................................................ 10
Chapter 3 – Configuration 15
3.1 GC Interface Screen ........................................................................................................................ 16
3.2 GC Configuration Screen ................................................................................................................. 19
3.3 GC Stream Data Screen .................................................................................................................. 22
3.4 Meter Setup Screen ......................................................................................................................... 26
3.5 Saving the Configuration .................................................................................................................. 28
Chapter 4 – Reference 31
4.1 Point Type 22/25: GC User Program Configuration and Status ...................................................... 32
4.2 Point Type 23/26: GC User C Program Stream Data ...................................................................... 37
Gas Chromatograph Software Interface User Manual (for FB107)
iv Contents Revised June-2017
[This page is intentionally left blank.]
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Introduction 1

Chapter 1 – Introduction

Caution
When implementing control using this product, observe best industry practices as suggested by applicable and appropriate environmental, health, and safety organizations. While this product can be used as a safety component in a system, it is NOT intended or designed to be the ONLY safety mechanism in that system.
This chapter describes the structure of this manual and presents an overview of the Gas Chromatograph Software Interface for the FloBoss™ 107.

1.1 Scope and Organization

This document serves as the user manual for the Gas Chromatograph Software Interface, which is intended for use in a FloBoss 107 (FB107).
This manual describes how to download, install, and configure the Gas Chromatograph Software Interface user program (referred to as the “GC
Interface program” or “the program” throughout the rest of this manual).
You access and configure this program using ROCLINK™ 800
Configuration Software (version 2.40 or greater) loaded on a personal computer (PC) running Windows® 7 (32 or 64-bit).
The sections in this manual provide information in a sequence appropriate for first-time users. Once you become familiar with the procedures and the software running in FB107, the manual becomes a reference tool.
This manual has the following major sections:
Chapter 1 – Introduction Chapter 2 – Installation Chapter 3 – Configuration Chapter 4 – Reference
This manual assumes that you are familiar with the FB107 and its configuration. For more information, refer to the following manuals:
FloBoss 107 Flow Manager Instruction Manual (part
D301232X012)
ROCLINK 800 Configuration Software User Manual (for FB107)
(part D301249X012)

1.2 Product Overview

The GC Interface program enables the FB107 to communicate directly with up to two gas chromatographs (GCs) on the same EIA-232 (RS-
232), EIA-485 (RS-485), or Ethernet communications port. Gas Chromatographs supported include the Daniel (Danalyzer) and Rosemount Analytical Models 500, 570, 590, 700, 770 and
Gas Chromatograph Software Interface User Manual (for FB107)
2 Introduction Revised June-2017
1000/1000A. GC controllers supported include the Daniel and Rosemount Analytical Models 2251, 2255, 2350, 2350A and 2360. The program communicates directly with the GC using Modbus protocol (in which the FB107 has Master status). The FB107 polls data from the GC, validates that data, and updates the appropriate meter run parameters using that data. The program can poll up to eight GC streams on one GC or up to a total of ten streams on two GCs.
Note: Using MON 2000 software, you set the Daniel GC
communications port to the SIM 2251 protocol. For all other parameter configurations, refer to the MON 2000 documentation.
In order to update meter run data in the FB107, you must assign the
streams to a meter run in the FB107’s database. This allows the FB107
to log the gas component data, heating value, and specific gravity and use these values in volume, mass, and energy calculations.
Note: Two versions of the program are included. Installation and
operation are identical between the two programs, but they use different point type locations, different display numbers, and are loaded into different program slots on the FB107. GCInterface_1.bin loads into user program location 1 and uses point types 22 and 23. GCInterface_2.bin loads into user program location 2 and uses point types 25 and 26.
The manual shows installation of GCInterface_1.bin. Load
GCInterface_1.bin unless another user program is currently installed in user program location 1.

1.2.1 Communications Wiring

The GC must be connected to the communications port on the FB107 with wiring between 16 and 24 AWG. Figure 1 shows the wiring for an EIA-232 (RS-232) to a Daniel 2350A GC.
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Introduction 3
Figure 1-1. Communications Wiring

1.2.2 Auto-configure

The GC Program provides an auto-configure option on the GC Interface screen (see Figure 3-2). This option enables the program to automatically configure the communication ports and Modbus parameters necessary to poll the GC(s). For this option to work, however, you must first specify a communication port in the Comm Port # field on the GC Interface screen, specify the maximum number of streams and Modbus address for each GC in use, and modify the Modbus register table location, if desired.
When you select Auto-configure on the GC Interface screen, the program sets the selected communication port parameters to the following values:
Note: The following communication port parameters are configured if
you are using an RS-232 or RS-485 module only.
Baud Rate:
9600
Data Bits:
8
Stop Bits:
1
Parity:
None
Key-On Delay:
200 mSec
Key-Off Delay:
200 mSec
Port Owner:
Modbus Master
When you select Auto-configure on the GC Interface screen, the program sets the Modbus configuration parameters for the selected communications port to the following values:
FB107
EIA-232 (RS-232) Port
Daniel 2350A
Terminals J6 (port 2), J10 (port 3), or J11 (port 4)
Rx Tx
RTS
DTR
GND
6
5
4
3
2
1
S IN_2
S OUT_2
GND
Gas Chromatograph Software Interface User Manual (for FB107)
4 Introduction Revised June-2017
Transmission Mode:
RTU
Byte Order:
MSB First
Event Log Enable
Disabled
Master Starting Request Number:
1
Master Number of Requests:
8
Master Continuous Polling Enable:
Disabled
When you select Auto-configure on the GC Interface screen, the program automatically configures the Modbus Master Table (using values in the Comm Port # and Modbus Address fields on the GC Interface screen) to poll for the following registers in the GC:
3001–3016:
Component IDs, Table 1
3017–3032:
Component IDs, Table 2
3034:
Current Stream
3035:
Mask of Streams associated with Table 1
3045:
Cycle Start Time – minutes
3046:
GC Alarm 1
3047:
GC Alarm 2
3059:
Calibration/Analysis Flag
7001–7016:
Gas Composition Values Mole % Comp 1–16
7033:
BTU (day)
7034:
BTU (saturated)
7035:
Specific Gravity
7036:
Compressibility
7037:
Wobbe Index
7038:
Total Unnormalized Mole %
7039
Total GPM CF
7040–7044
User Defined Calc Values
7070–7084
User Defined Average
9034:
Active Alarm Status
9035:
Unacknowledged Alarm Status
Note: Depending on the GC device, registers 9034 and 9035 may
require a Modbus conversion code to be manually configured.
When you select Auto-configure on the GC Interface screen, the program sets the Modbus Master Tables for each GC with a valid
address. The first GC’s master table uses the first logical point for the
communications port, and the second GC (if present) uses the second logical point for its communications port. The actual poll sequence set for each GC is:
3045–3047
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Introduction 5
3001–3032
3034–3035
3059
7001–7016
7033–7044
7070–7084
9034–9035
3045
Note: Depending on the GC device, registers 9034 and 9035 may
require a Modbus conversion code to be manually configured.
The Modbus Register to TLP Mapping assigns TLPs to registers. The program maps TLPs to the register table you specify in the Modbus Register Table Location field on the GC Interface screen. The parameters necessary for this program automatically map to the appropriate registers.
Finally, when the auto-configure process completes, the program disables the auto-configure parameter.
Note: After the auto-configure process completes, you may modify the
Modbus Master Table and/or the Modbus Register Table, but register 3045 must be the first and last register polled. The first poll must be stored in a register mapped to GC Stream parameter Sample Min Start (23/26,0,17) and the last poll must be stored in a register mapped to GC Stream parameter Sample Min End (23/26,0,16).

1.2.3 Validating GC Data

When the polls are complete, the program validates the data to ensure the polling was successful and data is correct. This validation occurs before the program copies the GC stream data to the meter run. Checks include:
The Communication Status (Point 121, Parameter 6, 12, 18, etc.)
must return valid responses (value of 8) for all registers polled. If errors are present, the meter runs are not updated.
If you disable Bypass Alarm 1 (Point 22/25, Parameter 11), the
Alarm Flag 1 (Point 23/26, Parameter 18, bits 14 & 15) from the GC is checked for errors. If errors are present, the meter runs are not updated.
If you disable Bypass Alarm 2 (Point 22/25, Parameter 12), the
Alarm Flag 2 (Point 23/26, Parameter 19, bits 0, 1, 2 & 3) from the GC is checked for errors. If errors are present, the meter runs are not updated.
Gas Chromatograph Software Interface User Manual (for FB107)
6 Introduction Revised June-2017
The program checks the Calibration Flag (Point 23/26, Parameter
20) to ensure it is in the Analysis State. If it is not in the Analysis State, the meter runs are not be updated.
The Starting Sample Minute value (Point 23/26, Parameter 17) must
be different than the previous value, or the meter runs are not updated.
The Starting Sample Minute value (Point 23/26, Parameter 17) and
Ending Sample Minute value (Point 23/26, Parameter 16) in the poll must match or the meter runs are not updated.
The current Stream Number (Point 23/26, Parameter 2) must be
assigned to a meter run.
The Total Un-Normalized Mole % value (Point 23/26, Parameter 9)
must be within plus or minus the Total Mole % Deviation value (Point 22/25, Parameter 14) of 100%. If this value is outside of this limit, the meter runs are not updated.
The Mole Sum value (Point 23/26, Parameter 21) must be within
plus or minus the Total Mole % Deviation value (Point 22/25, Parameter 14) of 100%. If this value is outside of this limit, the meter runs are not updated.
If you enable HV Limits (Point 23/26, Parameter 90) on the GC
Stream Data screen, ensure that the Stream Heating Value is between the Heating Value Low (Point 23/26, Parameter 91) and Heating Value High (Point 23/26, Parameter 92) values.
Note: If you do not enable the HV Limits (which is a per-stream
value), the module uses the Heating Value (described below) as the default.
The Heating Value (Point 23/26, Parameter 4 or 5, depending on
Wet vs. Dry) must be between Heating Value Low (Point 22/25, Parameter 9) and Heating Value High (Point 22/25, Parameter 10). If this value is outside the limits, the meter runs are not updated.
Note: The Heating Value is the default parameter the module checks
first. If you have enabled the HV Limits parameter (which is a per-stream value), that value overrides this value.
The Specific Gravity (Point 23/26, Parameter 6) must be between
0.07 and 1.52. If this value is outside of this limit, the meter runs are not updated.
Note: The program copies each GC stream component to its
corresponding component in the meter run, with the exception of neo-pentane. Neo-pentane is added to the iso-pentane component and then copied to the meter run. The heating value and specific gravity are also copied to the appropriate meter run.
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Introduction 7

1.3 Program Requirements

The GC Interface program is compatible with version 1.01 (or greater) of the FB107 firmware and with version 1.86 (or greater) of the ROCLINK 800 software.
Program specifics include:
File Name
Target Unit/
Version
User Defined
Points (UDP)
Flash Used
(in bytes)
DRAM Used
(in bytes)
ROCLINK 800
Version
Display
Number
GCInterface_1.bin
FB107 1.01
22, 23
(location 1)
29025
16,384
1.86
22, 23, 24
GCInterface_2.bin
FB107 1.01
25, 26
(location 2)
29025
16,384
1.86
24, 25, 26
For information on viewing the memory allocation of user programs, refer to the ROCLINK 800 Configuration Software User Manual (for FloBoss™ 107) (part D301249X012).

1.3.1 License Key

Some applications require that you install a license in the CPU to run the application. This license software is specific to these applications and is the property of the individual vendor (shown in the Vendor Name field on the License Key Administrator screens).
RAS (and other authorized vendors) distributes software licenses on security-enhanced universal serial bus (USB) drives.
You must install the following license keys to use the GC Interface Program.
GC Interface License Key.
Gas Chromatograph Software Interface User Manual (for FB107)
8 Introduction Revised June-2017
[This page is intentionally left blank.]
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Installation 9

Chapter 2 – Installation

This section provides instructions for installing the GC Interface program. Read Section 1.3 of this manual for program requirements.
Note: The program and license key can be installed in any order. The
manual shows the installation of the license key first.

2.1 Installing the License Key

A license key is required to use the GC Interface program. To install a USB key-based license.
Caution
Failure to exercise proper electrostatic discharge precautions, such as wearing a grounded wrist strap may reset the processor or damage electronic components, resulting in interrupted operations.
When working on units located in a hazardous area (where explosive gases may be present), make sure the area is in a non-hazardous state before performing these procedures. Performing these procedures in a hazardous area could result in personal injury or property damage.
To install a USB key-based license on the FB107:
1. Insert the USB license key in a USB port on your PC.
2. Select Utilities > License Key Administrator > Transfer Between
Device and Key from the ROCLINK 800 menu bar. The Transfer Licenses Between a Device and a Key screen displays.
Figure 2-1. Transfer Licenses Between a Device and a Key
Gas Chromatograph Software Interface User Manual (for FB107)
10 Installation Revised June-2017
Note: This screen has three sections. The upper portion (Licenses
on Device) shows any software licenses installed on the FB107. The middle portion (Licenses on Key) shows software licenses on the license key. The lower portion of the screen (License Key Event Log) provides a rolling log of the last eight events related to this license key.
3. Select the key-based license you want to transfer to the FB107
(GC Interface, as shown in Figure 2-1).
4. Click Move to Device. ROCLINK moves the license from the key
to the FB107 and updates the screen.
Figure 2-2. License Installed
Note: An FB107 can hold up to six different licenses, although you
can install only one instance of each license on the FB107. When you click Move to Device, ROCLINK 800 moves only one instance of the license onto the FB107 and automatically decreases the license quantity on the KEY.
5. Verify the license name displays in the Licenses on Device section
of the screen. Proceed to Section 2.2 to download the user program.

2.2 Downloading the Program

This section provides instructions for installing the program into the Flash memory on the FB107.
Note: Connect a PC to the FloBoss’s LOI port before starting the
download.
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Installation 11
Note: Two versions of the program are included. Installation and
operation are identical between the two programs, but they use different point type locations, different display numbers, and are loaded into different program slots on the FB107. GCInterface_1.bin loads into user program location 1 and uses point types 22 and 23. GCInterface_2.bin loads into user program location 2 and uses point types 25 and 26.
The manual shows installation of GCInterface_1.bin. Load
GCInterface_1.bin unless another user program is currently installed in user program location 1.
To download the user program using ROCLINK 800 software:
1. Start and logon to the ROCLINK 800.
2. Select ROC > Direct Connect to connect to the ROC800.
3. Select Utilities > User Program Administrator from the
ROCLINK menu bar. The User Program Administrator screen displays (see Figure 2-3):
Figure 2-3. User Program Administrator
4. Click Browse in the Download User Program File frame. The Select
User Program File screen displays (see Figure 2-4).
5. Select the path and user program file to download from the CD-
ROM (Program files are typically located in the Program Files
Gas Chromatograph Software Interface User Manual (for FB107)
12 Installation Revised June-2017
folder on the CD-ROM). As Figure 2-4 shows, the screen lists all valid user program files with the .BIN extension:
Note: Load only one of the included program files. The manual
shows installation of GCInterface_1.bin. Load GCInterface_1.bin unless another user program is currently installed in user program location 1.
Figure 2-4. Select User Program File
6. Click Open to select the program file. The User Program
Administrator screen displays. As shown in Figure 2-5, note that the Download User Program File frame identifies the selected program and that the Download & Start button is active:
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Installation 13
Figure 2-5. User Program Administrator
7. Click Download & Start to begin loading the selected program.
The following message displays:
Figure 2-6. Confirm Download
8. Click Yes to begin the download. When the download completes the
following message displays:
Figure 2-7. ROCLINK 800 Download Confirmation
Gas Chromatograph Software Interface User Manual (for FB107)
14 Installation Revised June-2017
9. Click OK. The User Program Administrator screen displays (see
Figure 2-8). Note that:
The User Programs Installed in Device frame identifies the
installed program(s).
The Status field indicates that the program is running.
Figure 2-8. User Program Administrator
Note: If you install the program before you install the license key,
the Status field reads “License Key Not Found.”
10. Click Close. Proceed to Chapter 3 Configuration to configure the
program.
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Configuration 15

Chapter 3 – Configuration

This section provides information on how to configure the Gas Chromatograph user program.
After you have loaded the GC Interface program on the FB107, you configure the program using three program-specific screens (GC Interface, GC Configuration, and GC Stream Data) and one ROCLINK 800 screen (Meter Setup):
Use the GC Interface screen to configure one or more GC addresses,
select a Comm port, modify the Modbus Register Table location (if necessary), assign the GC streams to meter runs, enable GC polling, and enable auto-configuration.
Use the GC Configuration screen to set GC-specific parameters,
including component IDs, GC alarm options, hexane plus options, data limits, and heating value adjustment parameters.
Use the GC Stream Data screen to verify communications between
the GC and the FB107, set SRBX alarms, and configure stream­specific heating value checks.
Use the Fluid Properties tab on the ROCLINK 800 Meter Setup
screen to indicate the type of heating value read from the GC and to select the “live” gas quality option.
You must configure the software before you can establish communications with the GC. To configure the program (after logging onto ROCLINK 800 and successfully installing the program and license key), proceed through the program screens as shown in this section.
Note: Using MON 2000 software, you set the Daniel GC
communications port to the SIM 2251 protocol. For all other parameter configurations, refer to the MON 2000 documentation.
You can access all the program-specific screens from the main ROCLINK 800 screen:
Gas Chromatograph Software Interface User Manual (for FB107)
16 Configuration Revised June-2017
Figure 3-1. Main ROCLINK 800 screen

3.1 GC Interface Screen

Use this screen to configure one or more GC addresses, select a Comm port, modify the Modbus Register Table location (if necessary), assign the GC streams to meter runs, enable GC polling, and enable automatic configuration of the Modbus parameters and communications ports. To access this screen:
1. From the Directory Tree, select User Program > GC Interface.
2. Double-click Display #22 GC Interface. The GC Interface screen
displays:
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Configuration 17
Figure 3-2. GC Interface Screen
3. Review the values in the following fields:
Field
Description
Max Streams
Sets the maximum number of streams from the gas chromatograph. Valid values are 1–8. You cannot define more than 10 streams for both GCs.
Modbus Address
Sets the address the FB107 uses to communicate with the gas chromatograph. You must set this value before the program can poll data and write it to the meter run. If you set this field to 0 (the default), GC polling is disabled.
Note: The Auto-configure option uses this value
(see Section 1.2.2).
Gas Chromatograph Software Interface User Manual (for FB107)
18 Configuration Revised June-2017
Field
Description
Modbus Register
Table Location
Identifies the location of the Modbus Register Table. Valid values are 1–12. The default value is 12.
Note: The program uses this value for
automatic configuration (see Section
1.2.2).
Comm Port #
Indicates which FB107 communications port the program uses for the connection to the GC(s). Valid values are Comm 1 (RS-485), Comm 2 (RS-232), and Comm 3 [RS-485, RS-232, or Enhanced Communication Module (Ethernet)]. Click to display all valid values.
Notes: The program uses this value for automatic
configuration (see Section 1.2.2).
Unlike other programs, the Comm Port
owner is Modbus Master for all values except for an ECM module. For that reason, the Comm Port owner should be something other than this program.
Polling Interval
Sets the delay, in seconds, the program waits before asking the GC for the next set of results.
Note: The GC typically takes 3 to 6 minutes to
update results.
Next Poll Req
This read-only field shows the number of seconds remaining until the next polling cycle. After a poll cycle completes, the program resets this field to the value stored in the Polling Interval field.
Comm Timeout
Sets the period, in seconds, the system waits for the GC to respond to a Modbus poll.
IP Address
Identifies the IP address of the GC to be polled. Note: This field only appears if an Enhanced
Communication Module (Ethernet) is installed in Slot 1 of the FB107 and Comm 3 is selected in the Comm Port # drop-down box, or an Enhanced Communication Module module is installed in Slot 2 of the FB107 and Comm 2 is selected in the Comm Port # drop-down box.
Meter Run Map
Associates GC streams with configured FB107 meter runs. Valid values are 18. Enter 0 to disable meter run updating.
You must configure this parameter before the program can poll data and write it to the meter run. The program stores the gas composition data for the specified stream in the gas quality parameters of the specified meter run (see Section 3.4).
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Configuration 19
Field
Description
Poll Mode
Indicates whether the program attempts to poll the GC(s).
Note: A valid license key and at least one valid
GC address must be present before polling can be enabled.
Autoconfigure
Indicates whether the program automatically determines the configuration values for communications ports and Modbus settings for a Daniel gas chromatograph.
Polling must be disabled and at least one valid GC address configured before you can initiate Autoconfigure.
Notes: See Section 1.2.2 for further information on
auto-configuration.
If you do not select this option, you must
configure the comm port, Modbus comm configuration, Modbus register mapping, and Modbus master table. Alternately, you can select Auto-configure and then use the ROCLINK 800 Modbus Configuration screen (Configure > MODBUS > Configuration) to change the values to suit the application.
4. Click Apply to save any changes you have made to this screen.
5. Click Close to return to the ROCLINK 800 screen. Proceed to
Section 3.2 to define GC configuration settings.

3.2 GC Configuration Screen

Use this screen to configure GC settings. It also displays communication status for each polling, configurable AGA update parameters, and some returned data values. To access this screen:
1. From the Directory Tree, select User Program > GC Interface >
Display #23, GC Configuration.
2. Double-click #1, GC Config. The GC Configuration screen
displays:
Gas Chromatograph Software Interface User Manual (for FB107)
20 Configuration Revised June-2017
Figure 3-3. GC Configuration Screen
3. Complete the screen fields based on your organization’s
requirements.
Field
Description
Point Number
Identifies the GC number for this screen. The program provides one instance for each GC. Click to display all defined instances.
Component IDs
Assigns each gas component a value in component data tables 1 and 2. Select Disable to identify any gases the FB107 supports but for which the gas chromatograph does not provide data.
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Configuration 21
Field
Description
Hexane Split
Enable
Enables the hexane(+) composition to be split between hexane, heptane, octane, nonane, and decane. The program uses a pre-defined split if the component ID for hexane(+) is set to 8, 9, 10, or 11 (see table below).
ID
Hexane
%
Heptane
%
Octane
%
Nonane
%
Decane
%
8
47.466
35.34
17.194
0
0
9
50
50 0 0 0 10
50
25
25 0 0
11
57.143
28.572
14.285
0
0
Alarm Bypass 1 and Alarm Bypass 2
Select this check box to allow the program to update the meter run values even if the Alarm 1 or Alarm 2 field in the GC displays an alarm.
Heating Value Low Limit
Sets the minimum heating value the FB107 accepts for a meter run update. The FB107 considers any heating values sent by the GC that are lower than this value to be invalid and does not forward them to the meter run.
Heating Value High Limit
Sets the maximum heating value the FB107 accepts for a meter run update. The FB107 considers any heating values sent by the GC that are greater than this value to be invalid and does not forward them to the meter run.
Total Mole Deviation
Sets, as a percentage, either the difference plus or minus from 100% that the program allows for Total Unnormalized Mole % or the sum of the component mole percentages. The program subtracts or adds this value to 100% to determine the range. The FB107 considers any stream gas compositions that exceed this value to be invalid and does not forward them to the meter run. Valid values are 0-100%.
Specific Gravity Low Limit
Sets the minimum specific gravity value the FB107 accepts for a meter run update. The FB107 considers any specific gravity values sent by the GC that are lower than this value to be invalid and does not forward them to the meter run.
Specific Gravity High Limit
Sets the maximum specific gravity value the FB107 accepts for a meter run update. The FB107 considers any specific gravity values sent by the GC that are greater than this value to be invalid and does not forward them to the meter run.
Heating Value Adjustment
Sets whether the program (in case of differing base pressures between the GC and the meter
run point) adjusts the GC’s heating value before
storing the value in the meter run parameter. If you click Enable, the heating value stored in the meter run equals the GC heating value multiplied by the ratio of the meter run base pressure to the GC’s base pressure.
Gas Chromatograph Software Interface User Manual (for FB107)
22 Configuration Revised June-2017
Field
Description
GC Base Pressure
Sets the base pressure, in PSIA or kPa, the GC uses to determine the heating value that is returned on a Modbus request.
Note: This field is active only if you enable the
Heating Value Adjustment.
4. Click Apply to save your changes.
5. Click Close to return to the ROCLINK 800 screen. Proceed to
Section 3.3 to review stream data.

3.3 GC Stream Data Screen

Use this screen to review stream data the FB107 has received from the GC(s). The program provides one iteration of this screen for each active stream in each GC. You can move between stream data displays using either the Point Number drop-down box on this screen or from the list on the Directory Tree. With the exception of the Alarm Logging Mode and HV Limits, the fields on this screen are read-only. To access this screen:
1. From the Directory Tree, select User Program > GC Interface >
Display #24, GC Stream Data.
2. Double-click #1, Cur Stream. The GC Stream Data screen displays:
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Configuration 23
Figure 3-4. GC Stream Data Screen
3. Reviewand change as necessarythe values in the following
fields:
Field
Description
Point Number
Identifies the stream number for this screen. Click to display all defined streams.
Calibration/
Analysis Flag
This read-only field indicates the
chromatograph’s status (self-calibration or analysis).
HV Limits
Select the Enable checkbox to allow user­defined Heating Value (HV) limits to be set for the selected run. If the HV Limits are enabled, the BTU Low/High values override the Heating Value High/Low Limits set on the GC Configuration screen.
Gas Chromatograph Software Interface User Manual (for FB107)
24 Configuration Revised June-2017
Field
Description
BTU Low Limit
If HV Limits are enabled, sets the minimum heating value for the selected run that the FB107 accepts for a meter run update. The FB107 considers any heating values sent by the GC that are greater than this value to be invalid and does not forward them to the meter run.
Note: Values entered in this field override the
Heating Value Low Limits set on the GC Configuration screen for the selected
run only.
BTU High Limit
If HV Limits are enabled, sets the maximum heating value for the selected run that the FB107 accepts for a meter run update. The FB107 considers any heating values sent by the GC that are greater than this value to be invalid and does not forward them to the meter run.
Note: Values entered in this field override the
Heating Value High Limits set on the GC Configuration screen for the selected
run only.
Stream
This read-only field shows the selected stream for the GC.
User Calcs
This read-only field shows data returned by the GC, but is not used by the program. For more information, refer to the GC’s user manual.
Diagnostics
This read-only field shows any diagnostic codes for the selected stream.
Stream Data
This read-only field shows values for the selected stream.
Component Mask
This read-only field shows which of the two Component ID tables each stream uses. Bit 0 of the Component Mask represents stream 1, bit 1 represents stream 2, and so on. If the bit is set, the program uses Component ID table 1. If the bit is not set, the program uses Component ID table 2.
Note: This value reflects settings from the GC.
Error Code
This read-only field provides a color-coded error display. Red indicates an alarm condition.
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Configuration 25
Field
Description
Alarm Logging
Mode
Sets how the program logs alarms with Spontaneous Report by Exception (SRBX) notification. Click to select a mode (described below).
Note: The system generates one alarm,
regardless of the number of different errors that may occur in the time before the alarm clears. SRBX notification occurs based on the Alarm Logging Mode. For the system to generate an alarm (such as for a Poll Sequence Failure error), you must first enable the Alarm Logging Mode parameter on the current logical stream.
Disable
Logging
No logging occurs.
Enable
Logging,
No SRBX
Logging occurs, but without generating SRBX notifications.
Enable
Logging,
SBRX in Set
Logging occurs, and SRBX notifications occur on alarm set.
Enable Logging, SRBX on
Clear
Logging occurs, and SRBX notifications occur on alarm clear.
Enable Logging, SRBX on
Both
Logging occurs, and SRBX notifications occur on both alarm set and alarm clear.
Component Data
This read-only field shows component values for the selected stream.
GC Defined
Averages
This read-only field shows data returned by the GC, but is not used by the program. For more information, refer to the GC’s user manual.
Specific Gravity
Limits
Select the Enable checkbox to allow user­defined specific gravity limits to be set for the selected run. If enabled, the Specific Gravity Low/High values for the selected run will override the Specific Gravity High/Low Limits set on the GC Configuration screen.
Gas Chromatograph Software Interface User Manual (for FB107)
26 Configuration Revised June-2017
Field
Description
Specific Gravity
Low Limit
If Specific Gravity Limits are enabled, sets the minimum specific gravity value for the selected run that the FB107 accepts for a meter run update. The FB107 considers any specific gravity values sent by the GC that are greater than this value to be invalid and does not forward them to the meter run.
Note: Values entered in this field override the
Specific Gravity Low Limits set on the GC Configuration screen for the
selected run only.
Specific Gravity
High Limit
If Specific Gravity Limits are enabled, sets the maximum specific gravity value for the selected run that the FB107 accepts for a meter run update. The FB107 considers any specific gravity values sent by the GC that are greater than this value to be invalid and does not forward them to the meter run.
Note: Values entered in this field override the
Specific Gravity High Limits set on the GC Configuration screen for the
selected run only.
Disable Event
Logging
Select the Enable checkbox to disable error
events from being written to the FB107’s event
log. Note: This option disables logging for ALL error
events except parameter change events. For example, an event is not logged if the GC reports an out-of-range specific gravity value and event logging is disabled.
4. Click Apply to save your changes.
5. Click Close to return to the ROCLINK 800 screen. Proceed to
Section 3.4 to set up meter run parameters.

3.4 Meter Setup Screen

Use this screen to set gas quality and heating values. To access this screen:
1. Select Meter > Setup from the ROCLINK 800 menu bar.
2. Select the Fluid Properties tab. The Fluid Properties tab displays:
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Configuration 27
Figure 3-5. Meter Setup, Fluid Properties tab
3. Reviewand change as necessarythe values in the following
fields:
Field
Description
Meter Number
Select the meter number to which the GC stream has been assigned.
Component Mole %
Indicates the mole percentage of each gas component. The program retrieves these values from the GC.
Heating Value
Indicates the heating value of a specified quantity of gas. The program sets this option to
Enter and retrieves this value from the GC.
Heating Value
Basis
Determines the heating value (dry or saturated) the program copies to the meter run. Valid values for this program are Dry or Wet. If you select Wet, the program copies the saturated heating value to the meter run. If you select As Delivered or Dry, the program resets the value to Dry and copies the dry heating value to the meter run.
Gas Chromatograph Software Interface User Manual (for FB107)
28 Configuration Revised June-2017
Field
Description
Specific Gravity
Indicates the specific gravity ratio of the molar mass gas to the molar mass of air. The program sets this option to Enter and retrieves this value from the GC.
Gas Quality
Indicates the source of the gas quality. The GC Interface program sets this field to Live when it copies data to the meter run.
4. Click Apply to save your changes. Proceed to Section 3.5 to save
the configuration.

3.5 Saving the Configuration

Whenever you modify or change the configuration, it is a good practice to save the final configuration to memory. To save the configuration:
1. Select ROC > Flags. The Flags screen displays:
Figure 3-6. Flags
2. Click Save Configuration. A verification message displays:
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Configuration 29
Figure 3-7. Save Verification
3. Click Yes to begin the save process. The Flash Write Status field on the
Flags screen displays In Progress. The following message displays:
Figure 3-8. Save Confirmation
4. Click OK. The Flash Write Status field on the Flags screen displays
Completed.
5. Click Update on the Flags screen. This completes the process of saving
your new configuration.
Note: For archive purposes, you should also save this configuration to
your PC’s hard drive or a removable media (such as a flash drive) using the File > Save Configuration option on the ROCLINK 800 menu bar.
Gas Chromatograph Software Interface User Manual (for FB107)
30 Configuration Revised June-2017
[This page is intentionally left blank.]
Gas Chromatograph Software Interface User Manual (for FB107)
Revised June-2017 Reference 31

Chapter 4 – Reference

This section provides tables of information on the user-defined point types used by the GC Interface program.
Point Type 22/25 (GC User Program Configuration and Status) Point Type 23/26 (GC User C Program Stream Data)
Gas Chromatograph Software Interface User Manual (for FB107)
32 Reference Revised June-2017

4.1 Point Type 22/25: GC User Program Configuration and Status

Point type 22 and 25 contains the parameters for configuring the GC Interface program and houses the status information from the gas chromatograph. The program maintains two logical points of this point type.
Point Type 22/25: GC User Program Configuration and Status
Param
#
Name
Access
System or
User
Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
0
Point Tag ID
R/W
User
AC
10
0x20
0x7E for
each ASCII
character
“GC
Config“
1.00
Point Type Description
1
Not used
2
User Program Status
R/O
System
UINT8
1
0  3
0
1.00
User Program Status
0 = No Error 1 = License Unavailable 2 = Comm Configuration Failed
3
Autoconfiguration
R/W
Both
UINT8
1
0  1
1
1.00
Auto-configuration
0 = Disabled 1 = Enabled
4
Comm Port Number
R/W
User
UINT8
1
2 2
2
1.00
Comm Port used to communicate with the GC
5
GC Address
R/W
User
UINT8
1
0 255
0
1.00
GC Address (Modbus address of GC)
6
MODBUS Location
R/W
User
UINT8
1
1 24
24
1.00
Modbus Register Mapping Table used by GC
7
Polling Interval
R/W
User
FL
4
Any Positive
Floating
Point
Number
60
1.00
Interval at which this program polls the GC for new data (In Seconds).
8
Next Poll Request
R/O
System
FL
4
Any Positive
Floating
Point
Number
0
1.00
Amount of time (In Seconds) until the next time the GC will be polled for new data.
9
Heating Value Low Limit
R/W
User
FL
4
Any Floating
Point
Number
900.0
1.00
Heating Value Low Limit. If the Heating Value returned from the GC is less than this value, an alarm will be set.
10
Heating Value High Limit
R/W
User
FL
4
Any Floating
Point
Number
1300.0
1.00
Heating Value High Limit. If the Heating Value returned from the GC is greater than this value, an alarm will be set.
Liquid Calculations Application User Manual (for DL8000)
Revised June-2017 Reference 33
Point Type 22/25: GC User Program Configuration and Status
Param
#
Name
Access
System or
User
Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
11
Bypass Alarm 1
R/W
User
UINT8
1
0 1
0
1.00
Bypass Alarm 1 from GC
0 = Alarm Allowed 1 = Alarm Bypassed
12
Bypass Alarm 2
R/W
User
UINT8
1
0 1
0
1.00
Bypass Alarm 2 from GC
0 = Alarm Allowed 1 = Alarm Bypassed
13
Max Streams
R/W
User
UINT8
1
0 255
6
1.00
Maximum number of streams available from GC
14
Total Mole % Deviation
R/W
User
FL
4
0 5.0
2
1.00
Total Mole % Deviation
15
Hexane Split Enable
R/W
User
UINT8
1
0 1
1
1.00
Enable the Hexane Split functionality of the GC Program
16
Communication Timeout
R/W
User
FL
4
0 60.0
45.0
1.00
Amount of time to wait for a response from a GC (In Seconds).
17
Meter Run 1 Stream
R/W
User
UINT8
1
0 8
0
1.00
Meter Run 1 Stream
0 = Disable Meter Run Updating 1-8 = Stream number to use to update meter
run’s gas composition
18
Meter Run 2 Stream
R/W
User
UINT8
1
0 8
0
1.00
Meter Run 2 Stream
0 = Disable Meter Run Updating 1-8 = Stream number to use to update meter
run’s gas composition
19
Meter Run 3 Stream
R/W
User
UINT8
1
0 8
0
1.00
Meter Run 3 Stream
0 = Disable Meter Run Updating 1-8 = Stream number to use to update meter
run’s gas composition
20
Meter Run 4 Stream
R/W
User
UINT8
1
0 8
0
1.00
Meter Run 4 Stream
0 = Disable Meter Run Updating 1-8 = Stream number to use to update meter
run’s gas composition
21
Not used
22
Not used
23
Not used
24
Not used
25
Not used
26
Not used
27
Not used
28
Not used
Gas Chromatograph Software Interface User Manual (for FB107)
34 Reference Revised June-2017
Point Type 22/25: GC User Program Configuration and Status
Param
#
Name
Access
System or
User
Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
29
Methane Component ID
R/W
User
UINT8
1
0 255
0
1.00
Methane Component ID. This is the Component ID number that is associated with Methane in the GC.
30
Ethane Component ID
R/W
User
UINT8
1
0 255
1
1.00
Ethane Component ID. This is the Component ID number that is associated with Ethane in the GC.
31
Propane Component ID
R/W
User
UINT8
1
0 255
2
1.00
Propane Component ID. This is the Component ID number that is associated with Propane in the GC.
32
i-Butane Component ID
R/W
User
UINT8
1
0 255
3
1.00
i-Butane Component ID. This is the Component ID number that is associated with i-Butane in the GC.
33
n-Butane Component ID
R/W
User
UINT8
1
0 255
4
1.00
n-Butane Component ID. This is the Component ID number that is associated with n-Butane in the GC.
34
Neo-Pentane Component ID
R/W
User
UINT8
1
0 255
7
1.00
Neo-Pentane Component ID. This is the Component ID number that is associated with Neo-Pentane in the GC.
35
i-Pentane Component ID
R/W
User
UINT8
1
0 255
5
1.00
i-Pentane Component ID. This is the Component ID number that is associated with i-Pentane in the GC.
36
n-Pentane Component ID
R/W
User
UINT8
1
0 255
6
1.00
n-Pentane Component ID. This is the Component ID number that is associated with n-Pentane in the GC.
37
Hexane Component
R/W
User
UINT8
1
0 255
255
1.00
Hexane Component ID. This is the Component ID number that is associated with Hexane in the GC.
38
Hexane (+) Component ID
R/W
User
UINT8
1
0 255
8
1.00
Hexane (+) Component ID. This is the Component ID number that is associated with Hexane (+) in the GC.
39
Nitrogen Component ID
R/W
User
UINT8
1
0 255
14
1.00
Nitrogen Component ID. This is the Component ID number that is associated with Nitrogen in the GC.
40
Carbon Dioxide Component ID
R/W
User
UINT8
1
0 255
17
1.00
Carbon Dioxide Component ID. This is the Component ID number that is associated with Carbon Dioxide in the GC.
41
H2S Component ID
R/W
User
UINT8
1
0 255
255
1.00
H2S Component ID. This is the Component ID number that is associated with H2S in the GC.
Liquid Calculations Application User Manual (for DL8000)
Revised June-2017 Reference 35
Point Type 22/25: GC User Program Configuration and Status
Param
#
Name
Access
System or
User
Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
42
Water Component ID
R/W
User
UINT8
1
0 255
255
1.00
Water Component ID. This is the Component ID number that is associated with Water in the GC.
43
Helium Component ID
R/W
User
UINT8
1
0 255
255
1.00
Helium Component ID. This is the Component ID number that is associated with Helium in the GC.
44
Oxygen Component ID
R/W
User
UINT8
1
0 255
255
1.00
Oxygen Component ID. This is the Component ID number that is associated with Oxygen in the GC.
45
Carbon Monoxide Component ID
R/W
User
UINT8
1
0 255
255
1.00
Carbon Monoxide Component ID. This is the Component ID number that is associated with Carbon Monoxide in the GC.
46
Hydrogen Component ID
R/W
User
UINT8
1
0 255
255
1.00
Hydrogen Component ID. This is the Component ID number that is associated with Hydrogen in the GC.
47
Heptane Component ID
R/W
User
UINT8
1
0 255
255
1.00
Heptane Component ID. This is the Component ID number that is associated with Heptane in the GC.
48
Octane Component ID
R/W
User
UINT8
1
0 255
255
1.00
Octane Component ID. This is the Component ID number that is associated with Octane in the GC.
49
Nonane Component ID
R/W
User
UINT8
1
0 255
255
1.00
Nonane Component ID. This is the Component ID number that is associated with Nonane in the GC.
50
Decane Component ID
R/W
User
UINT8
1
0 255
255
1.00
Decane Component ID. This is the Component ID number that is associated with Decane in the GC.
51
Argon Component ID
R/W
User
UINT8
1
0 255
255
1.00
Argon Component ID. This is the Component ID number that is associated with Argon in the GC.
52
Heating Value Adjust Option
R/W
User
UINT8
1
0 1
0
1.00
Heating Value Adjust Option
0 = Disabled 1 = Enabled
53
GC Base Pressure
R/W
User
FL 4 0Valid Positive Floating
Point
Number
14.73
1.00
Base Pressure Configured in GC
54
Poll Mode
R/W
User
UINT8
1
0 1
0
1.00
GC Polling Mode
0 = Disabled 1 = Enabled
Gas Chromatograph Software Interface User Manual (for FB107)
36 Reference Revised June-2017
Point Type 22/25: GC User Program Configuration and Status
Param
#
Name
Access
System or
User
Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
55
Hexane Percentage
R/W
Both
FL 4 0Valid Positive Floating
Point
Number
47.466
1.00
Hexane Split C6 (Hexane) Component Percentage. This is the percentage of the Hexane (+) value returned from the GC that will be attributed to Hexane.
56
Heptane Percentage
R/W
Both
FL 4 0Valid Positive Floating
Point
Number
35.34
1.00
Hexane Split C7 (Heptane) Component Percentage. This is the percentage of the Hexane (+) value returned from the GC that will be attributed to Heptane.
57
Octane Percentage
R/W
Both
FL 4 0Valid Positive Floating
Point
Number
17.194
1.00
Hexane Split C8 (Octane) Component Percentage. This is the percentage of the Hexane (+) value returned from the GC that will be attributed to Octane.
58
Nonane Percentage
R/W
Both
FL 4 0Valid Positive Floating
Point
Number
0
1.00
Hexane Split C9 (Nonane) Component Percentage. This is the percentage of the Hexane (+) value returned from the GC that will be attributed to Nonane.
59
Decane Percentage
R/W
Both
FL 4 0Valid Positive Floating
Point
Number
0
1.00
Hexane Split C10 (Decane) Component Percentage. This is the percentage of the Hexane (+) value returned from the GC that will be attributed to Decane.
60
Spec. Grav Low Limit
R/W
User
FL 4 0Valid Positive Floating
Point
Number
0.07
1.01
Specific Gravity Low Limit. If the specific gravity value returned from the GC is less than this value, an alarm will be set.
61
Spec. Grav High Limit
R/W
User
FL 4 0Valid Positive Floating
Point
Number
1.52
1.01
Specific Gravity High Limit. If the specific gravity value returned from the GC is more than this value, an alarm will be set.
62
IP Address of GC
R/W
User
AC
20
20
Characters
“ “
IP address of the GC.
63
IP Port of GC
R/W
User
UINT16
2
0-65535 0
IP port of the GC.
Liquid Calculations Application User Manual (for DL8000)
Revised June-2017 Reference 37

4.2 Point Type 23/26: GC User C Program Stream Data

Point type 23 and 26 contains the parameters for configuring the GC Interface program and houses the status information from the gas chromatograph. The program maintains 11 logical points of this point type. Logical 0 is the current steam, and logical 1 through 10 are mapped to streams on the gas chromatographs, as assigned by the maximum streams per GC (point type 22/25, parameter 13).
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
0
Point Tag Id
R/W
User
AC
10
0x20
0x7E for
each ASCII
character
“Cur
Stream”
or
“Stream #“
1.00
Point Type Description
1
Not used
2
Stream Number
R/O
System
UINT16
2
0 1
1-8
depending
on logical
1.00
Stream Number
3
Component Table Mask
R/W
System
UINT16
2
0 0xFFFF
0
1.00
Component Table Mask. Each bit corresponds to a stream. A bit value of 1 means that the GC will return data from Table 1 will be used. A bit value of 0 means that the GC will return data from Table 2.
4
Dry Heating Value
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Dry Heating Value
5
Saturated Heating Value
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Saturated (Wet) Heating Value
6
Specific Gravity
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Specific Gravity
Gas Chromatograph Software Interface User Manual (for FB107)
38 Reference Revised June-2017
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
7
Compressibility
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Compressibility
8
Wobbe Index
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Wobbe Index
9
Total Un­Normalized Mole %
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Total Un-Normalized Mole %
10
Total GPM
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Total GPM
11
User Defined Calc 1
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Calc 1
12
User Defined Calc 2
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Calc 2
13
User Defined Calc 3
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Calc 3
14
User Defined Calc 4
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Calc 4
Liquid Calculations Application User Manual (for DL8000)
Revised June-2017 Reference 39
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
15
User Defined Calc 5
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Calc 5
16
Sample Minute End
R/W
System
UINT16
2
0 0xFFFF
0xFFFF
1.00
The minute value read from the GC at the end
of the GC Interface Program’s polling
sequence.
17
Sample Minute Start
R/W
System
UINT16
2
0 60
0
1.00
The minute value read from the GC at the start of the GC Interface Program’s polling sequence.
18
Alarm 1
R/W
System
UINT16
2
0 255
0
1.00
GC Alarm 1
19
Alarm 2
R/W
System
UINT16
2
0 255
0
1.00
GC Alarm 2
20
Calibration Flag
R/W
System
UINT16
2
0 255
1
1.00
Calibration Flag
0 = Calculation data 1 = Analysis data
21
Mole Sum
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole Sum
22
Error Code
R/W
System
UINT8
1
0 8
0
1.00
Error Code
0 = All Checks Pass 1 = Poll Sequence Failed 2 = Alarm Check Failed 3 = Calibration Check Failed 4 = Sample Minute Changed 5 = Total Un-normalized Mole Percentage
Failed
6 = Mole Sum Check Failed 7 = Heating Value Range Check Failed 8 = Specific Gravity Range Check Failed
Gas Chromatograph Software Interface User Manual (for FB107)
40 Reference Revised June-2017
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
23
Alarm Logging Mode
R/W
User
UINT8 1 0 1.00
Alarm Logging Mode
0 = Disable Logging 1 = Enable Logging, No SRBX 2 = Enable Logging, SRBX on Set only 3 = Enable Logging, SRBX on Clear only 4 = Enable Logging, SRBX, on both Set and
Clear
24
Component Index #1 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #1 Table 1
25
Component Index #2 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #2 Table 1
26
Component Index #3 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #3 Table 1
27
Component Index #4 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #4 Table 1
28
Component Index #5 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #5 Table 1
29
Component Index #6 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #6 Table 1
30
Component Index #7 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #7 Table 1
31
Component Index #8 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #8 Table 1
32
Component Index #9 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #9 Table 1
33
Component Index #10 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #10 Table 1
34
Component Index #11 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #11 Table 1
35
Component Index #12 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #12 Table 1
36
Component Index #13 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #13 Table 1
Liquid Calculations Application User Manual (for DL8000)
Revised June-2017 Reference 41
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
37
Component Index #14 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #14 Table 1
38
Component Index #15 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #15 Table 1
39
Component Index #16 Table 1
R/W
System
UINT8
1
0 255
0
1.00
Component Index #16 Table 1
40
Component Index #1 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #1 Table 2
41
Component Index #2 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #2 Table 2
42
Component Index #3 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #3 Table 2
43
Component Index #4 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #4 Table 2
44
Component Index #5 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #5 Table 2
45
Component Index #6 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #6 Table 2
46
Component Index #7 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #7 Table 2
47
Component Index #8 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #8 Table 2
48
Component Index #9 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #9 Table 2
49
Component Index #10 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #10 Table 2
50
Component Index #11 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #11 Table 2
51
Component Index #12 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #12 Table 2
Gas Chromatograph Software Interface User Manual (for FB107)
42 Reference Revised June-2017
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
52
Component Index #13 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #13 Table 2
53
Component Index #14 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #14 Table 2
54
Component Index #15 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #15 Table 2
55
Component Index #16 Table 2
R/W
System
UINT8
1
0 255
0
1.00
Component Index #16 Table 2
56
Mole % Component #1
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #1
57
Mole % Component #2
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #2
58
Mole % Component #3
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #3
59
Mole % Component #4
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #4
60
Mole % Component #5
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #5
Liquid Calculations Application User Manual (for DL8000)
Revised June-2017 Reference 43
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
61
Mole % Component #6
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #6
62
Mole % Component #7
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #7
63
Mole % Component #8
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #8
64
Mole % Component #9
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #9
65
Mole % Component #10
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #10
66
Mole % Component #11
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #11
67
Mole % Component #12
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #12
68
Mole % Component #13
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #13
Gas Chromatograph Software Interface User Manual (for FB107)
44 Reference Revised June-2017
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
69
Mole % Component #14
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #14
70
Mole % Component #15
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #15
71
Mole % Component #16
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Mole % Component #16
72
User Defined Avg 1
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 1
73
User Defined Avg 2
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 2
74
User Defined Avg 3
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 3
75
User Defined Avg 4
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 4
76
User Defined Avg 5
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 5
Liquid Calculations Application User Manual (for DL8000)
Revised June-2017 Reference 45
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
77
User Defined Avg 6
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 6
78
User Defined Avg 7
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 7
79
User Defined Avg 8
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 8
80
User Defined Avg 9
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 9
81
User Defined Avg 10
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 10
82
User Defined Avg 11
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 11
83
User Defined Avg 12
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 12
84
User Defined Avg 13
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 13
Gas Chromatograph Software Interface User Manual (for FB107)
46 Reference Revised June-2017
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
85
User Defined Avg 14
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 14
86
User Defined Avg 15
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
User Defined Average 15
87
Active Alarm Status
R/W
System
UINT8
1
0 1
0
1.00
Active Alarm Status (Red light on GC Controller)
0 = Disabled 1 = Enabled
88
Unacknowledged Alarm Status
R/W
System
UINT8
1
0 1
0
1.00
Unacknowledged Alarm Status (Yellow Light on GC Controller)
0 = Disabled 1 = Enabled
89
GC Number
R/W
System
UINT8
1
1 2
0
1.00
GC Number. The data in this stream was received from this GC.
90
Stream Heating Value
R/W
System
UINT8
1
0 1
0
1.00
Stream Heating Value Limits
0 = Disabled 1 = Enabled
91
Heating Value Low Limit
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Stream Heating Value Low Limit. If the Stream Heating Value Limits parameter is enabled and the Heating Value in this stream is less than this value, then an alarm will be set.
92
Heating Value High Limit
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.00
Stream Heating Value High Limit. If the Stream Heating Value Limits parameter is enabled and the Heating Value in this stream is greater than this value, then an alarm will be set.
93
Spec Grav Limit
R/W
System
UINT8
1
0 1
0
1.01
Specific Gravity Limit
0 = Disabled 1 = Enabled
Liquid Calculations Application User Manual (for DL8000)
Revised June-2017 Reference 47
Point Type 23/26: GC User C Program Stream Data
Param
#
Name
Access
System
or User Update
Data
Type
Length
Range
Default
Version
Description of Functionality and Meaning
of Values
94
Spec Grav Low Limit
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.01
Specific Gravity Low Limit. If the specific gravity value returned from the GC is less than this value, an alarm will be set.
95
Spec Grav Hi Limit
R/W
System
FL 4 0Valid Positive
Floating
Point
Number
0
1.01
Specific Gravity High Limit. If the specific gravity value returned from the GC is more than this value, an alarm will be set.
96
Event Disable
R/W
System
UINT8
1
0 1
0
1.01
Error Event Disable
0 = Error Event Logging Enabled 1 = Error Event Logging Disabled
Gas Chromatograpgh Software Interface User Manual (for FB107)
For customer service and technical support, visit www.EmersonProcess.com/Remote/Support.
Global Headquarters, North America, and Latin America:
Emerson Automation Solutions Remote Automation Solutions 6005 Rogerdale Road Houston, TX 77072 U.S.A. T +1 281 879 2699 | F +1 281 988 4445
www.EmersonProcess.com/Remote
© 2008-2017 Remote Automation Solutions, a business unit of Emerson Automation Solutions. All rights reserved.
This publication is for informational purposes only. While every effort has been made to ensure accuracy, this publication shall not be read to include any warranty or guarantee, express or implied, including as regards the products or services described or their use or applicability. Remote Automation Solutions (RAS) reserves the right to modify or improve the designs or specifications of its products at any time without notice. All sales are governed by RAS terms and conditions which are available upon request. RAS accepts no responsibility for proper selection, use or maintenance of any product, which remains solely with the purchaser and/or end-user.
Europe:
Emerson Automation Solutions Remote Automation Solutions Unit 8, Waterfront Business Park Dudley Road, Brierley Hill Dudley UK DY5 1LX T +44 1384 487200 | F +44 1384 487258
Middle East/Africa:
Emerson Automation Solutions Remote Automation Solutions Emerson FZE P.O. Box 17033 Jebel Ali Free Zone – South 2 Dubai U.A.E. T +971 4 8118100 | F +971 4 8865465
Asia-Pacific:
Emerson Automation Solutions Remote Automation Solutions 1 Pandan Crescent Singapore 128461 T +65 6777 8211| F +65 6777 0947
Remote Automation Solutions
Loading...