Remote Automation Solutions FB1200 Flow Computer 6-Channel Expansion I/O Board Field Replacement Guide Manuals & Guides

FB1200 Flow Computer 6-Channel I/O Expansion Board Field Replacement Guide
FB1200 Flow Computer 6-Channel I/O Expansion Board Field Replacement Guide
D301819X012
November 2020
400215010-KIT
Remote Automation Solutions
FB1200 Flow Computer 6-Channel I/O Expansion Board Field Replacement Guide
D301819X012 November 2020
Device Safety Considerations
Reading these Instructions
Before operating the device, read these instructions carefully and understand their safety implications. In some situations, improperly using this device may result in damage or injury. Keep this manual in a convenient location for future reference. Note that these instructions may not cover all details or variations in equipment or cover every possible situation regarding installation, operation, or maintenance. Should problems arise that are not covered sufficiently in the text, immediately contact Customer Support for further information.
Protecting Operating Processes
A failure of this device – for whatever reason -- may leave an operating process without appropriate protection and could result in possible damage to property or injury to persons. To protect against this, you should review the need for additional backup equipment or provide alternate means of protection (such as alarm devices, output limiting, fail-safe valves, relief valves, emergency shutoffs, emergency switches, etc.). Contact Remote Automation Solutions for additional information.
Returning Equipment
If you need to return any equipment to Remote Automation Solutions, it is your responsibility to ensure that the equipment has been cleaned to safe levels, as defined and/or determined by applicable federal, state and/or local law regulations or codes. You also agree to indemnify Remote Automation Solutions and hold Remote Automation Solutions harmless from any liability or damage which Remote Automation Solutions may incur or suffer due to your failure to ensure device cleanliness.
Grounding Equipment
Ground metal enclosures and exposed metal parts of electrical instruments in accordance with OSHA rules and regulations as specified in Design Safety Standards for Electrical Systems, 29 CFR, Part 1910, Subpart S, dated: April 16, 1981 (OSHA rulings are in agreement with the National Electrical Code). You must also ground mechanical or pneumatic instruments that include electrically operated devices such as lights, switches, relays, alarms, or chart drives. Important: Complying with the codes and regulations of authorities having jurisdiction is essential to ensuring personnel safety. The guidelines and recommendations in this manual are intended to meet or exceed applicable codes and regulations. If differences occur between this manual and the codes and regulations of authorities having jurisdiction, those codes and regulations must take precedence.
Protecting from Electrostatic Discharge (ESD)
This device contains sensitive electronic components which be damaged by exposure to an ESD voltage. Depending on the magnitude and duration of the ESD, it can result in erratic operation or complete failure of the equipment. Ensure that you correctly care for and handle ESD-sensitive components.
System Training
A well-trained workforce is critical to the success of your operation. Knowing how to correctly install, configure, program, calibrate, and trouble-shoot your Emerson equipment provides your engineers and technicians with the skills and confidence to optimize your investment. Remote Automation Solutions offers a variety of ways for your personnel to acquire essential system expertise. Our full-time professional instructors can conduct classroom training at several of our corporate offices, at your site, or even at your regional Emerson office. You can also receive the same quality training via our live, interactive Emerson Virtual Classroom and save on travel costs. For our complete schedule and further information, contact the Remote Automation Solutions Training Department at 800-338-8158 or email us at education@emerson.com.
Ethernet Connectivity
This automation device is intended to be used in an Ethernet network which does not have public access. The inclusion of this device in a publicly accessible Ethernet-based network is
not recommended.
FB1200 Flow Computer 6-Channel Expansion I/O Board Field Replacement Guide
Item
Field Replacement Kit Part Number
WARNING
WARNING
D301819X012
November 2020

Replacing the 6-Channel Expansion I/O Board

You can replace the optional 6-channel expansion I/O board provided you replace it with the exact same part.
Restriction
Hazardous area approvals require that any part replaced in the field be the exact same part (“like­for-like”). Upgrading or substituting different parts violates hazardous area certification.
Refer to the table below for the correct field replacement part number.
Board Type UL Field Replacement Kit
Part Number
6-Channel Expansion I/O Board 400215010-KIT
Ambient Temperature Range
May be used up to a maximum ambient temperature of 80°C and a minimum ambient temperature of –40°C; refer to the data plate attached to the device for ambient temperature.
Required Tools
#1 Phillips-head screwdriver
#2 Phillips-head screwdriver
3/32-inch flat head screwdriver (for 3.81 mm pitch terminal block connections and disconnecting the internal connector to the sensor)
Hexagonal torque wrenches with 3mm, #1, and #2 Phillips-head bits. Ranges must include 1 to 2 in-lbs (0.1 to 0.2 N-m), 4 to 6 in-lbs (0.5 to 0.7 N-m), and 10 to 12 in-lbs (1.1 to 1.4 N-m).
Electrical Ratings
Input Voltage: 5.7-30Vdc, 10 Watts Max
I/O Ratings (FB1200 Flow Computer):
o
4 Digital Outputs: 0.5A, 24Vdc
o
2 Analog Outputs: 4-20 mA
EXPLOSION HAZARD – Substitution of any components may impair suitability for Class I, Division 1 or Class I, Division 2.
EXPLOSION HAZARD - Do not disconnect equipment unless power has been removed or the area
is known to be non-hazardous.
1
FB1200 Flow Computer 6-Channel Expansion I/O Board Field Replacement Guide
DANGER
1
Screw
2
Retaining Clamp
3
Washer
D301819X012 November 2020
DANGER
EXPLOSION HAZARD: Ensure the area in which you perform this operation is non-hazardous. Performing this operation in a hazardous area could result in an explosion.
EXPLOSION HAZARD: Never remove end cap(s) in a hazardous location. Removing cover(s) in a hazardous location could result in an explosion.

Removing/Replacing the 6-Channel Expansion I/O Board

UL Listed 6-Channel I/O Expansion Board Field Installed Accessory Kit for Use in Class I, Division 1, Groups C, and D; and Class I, Division 2, Groups A, B, C, and D, Part No. 400215010-KIT for use with UL Listed Model Series FB1200.
1. Remove the retaining clamps on the end caps (if present) using a 3 mm hexagonal wrench.
Components of Retaining Clamp Assembly
2
2. Grasp the end cap.
Front End Cap Rear End Cap
3. Unscrew the end cap by turning it counter-clockwise until it comes off. Set the end cap aside in a safe location.