The exclamation point within an equilateral
triangle is intended to alert the user to
the presence of important operating and
maintenance (servicing) instructions in this
Installation and Operating Instructions.
« DANGER », « AVERTISSEMENT », ou
« PRÉCAUTION »
Le point d’exclamation à l’intérieur d’un
triangle équilatéral vise à alerter l’utilisateur
à la présence d’importantes instructions
de fonctionnement et de maintenance
(service) dans ce manuel d’installation et
d’instructions de fonctionnement.
“NOTE”
Amplies a procedure, practice, or condition.
« REMARQUE »
Amplie une procédure, pratique ou condition.
!
Safe Working Load
“Attention, Consult Accompanying
Documents”
Refer to Instruction Manual/Booklet
Type B, Applied Part
Dangerous Voltage / Shock Hazard
Fuse Rating Specication
Protective Earth Ground
Burn Hazard/Hot Surface
TRANSPORTATION / STORAGE CONDITIONS
Temperature: Operating: 59 to 104º F
15 to 40º C
Storage: -22 to +140º F
-30 to +60º C
R e l a t i v e H u m i d i t y : O p e r a t i n g : 5 t o 9 5 % w i t h o u t
1.1. This manual contains information applicable only to the
Reliance® Model FX920 Chair.
1.2. Whenever you see the symbols shown below, heed their
instructions! Always follow safe operating and maintenance
practices.
“DANGER”-
IDENTIFIES SPECIAL INSTRUCTIONS OR
PROCEDURES WHICH, IF NOT CORRECTLY
FOLLOWED, COULD RESULT IN LOSS OF LIFE
OR PERSONAL INJURY.
THE DANGER SYMBOL
« DANGER » : LE SYMBOLE DANGER IDENTIFIE
DES INSTRUCTIONS OU PROCÉDURES
SPÉCIALES QUI, SI ELLES NE SONT PAS SUIVIES
CORRECTEMENT, POURRAIENT CAUSER UNE
PERTE DE VIE OU UNE BLESSURE.
“WARNING”-
IDENTIFIES SPECIAL INSTRUCTIONS OR
PROCEDURES WHICH, IF NOT CORRECTLY
FOLLOWED, COULD RESULT IN PERSONAL
INJURY.
THE WARNING SYMBOL
« AVERTISSEMENT » : LE SYMBOLE
AVERTISSEMENT IDENTIFIE DES INSTRUCTIONS
OU PROCÉDURES SPÉCIALES QUI, SI ELLES
NE SONT PAS SUIVIES CORRECTEMENT,
POURRAIENT CAUSER UNE BLESSURE.
“NOTE”-
additional information.
Note indicates points of particular or
« REMARQUE » : Remarque indique des points
d’intérêt particulier ou des renseignements
additionnels.
d’information particulière ou supplémentaire.
1.3. Should your product not perform properly, or if you
have any questions concerning the use and care of any
Reliance® product, contact the Reliance® Distributor,
where you purchased this product or contact the Technical
Service Department, Reliance® Medical Products, Inc.,
3535 Kings Mills Road, Mason, Ohio 45040-2303, or call
(800) 735-0358.
NOTE:
number available before contacting Reliance® or
your Authorized Reliance® Distributor.
la note indique des remarques
Always have the model number and serial
REMARQUE : Ayez toujours le numéro de
modèle et le numéro de série à portée de la
main avant de contacter Reliance® ou votre
distributeur Reliance® autorisé.
“CLASSIFIED BY CANADIAN STANDARDS
ASSOCIATION® CSA WITH RESPECT TO ELECTRIC
SHOCK, FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH IEC 60601-1.”
According to Clause 5 in IEC 60601-1, sec 6.8.1, this unit
is classied by the following:
“CAUTION”-
special instructions or procedures which, if
not strictly observed, could result in damage
to or destruction of equipment.
This caution symbol identifies
« PRÉCAUTION » : Ce symbole de précaution
identifie des instructions ou procédures
spéciales qui, si elles ne sont pas strictement
suivies, pourraient causer un dommage ou la
destruction de l’équipement..
IN-FX920
• The type of protection against electric shock: EQUIPMENT
energized from an external electrical power source:
CLASS I EQUIPMENT
• The degree of protection against electric shock: TYPE B
EQUIPMENT
• The degree of protection against harmful ingress of water:
ORDINARY DEGREE
• The degree of safety of application in the presence of a
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH
OXYGEN OR NITROUS OXIDE: EQUIPMENT not suitable
for use in the presence of a FLAMMABLE ANAESTHETIC
MIXTURE WITH AIR or WITH OXYGEN OR NITROUS
OXIDE
• The mode of operation: CONTINUOUS OPERATION
5
WARNING: TO PREVENT FIRE OR ELECTRICAL
SHOCK HAZARD, DO NOT EXPOSE THIS
EQUIPMENT TO RAIN OR MOISTURE.
AVERTISSEMENT : POUR ÉVITER UN
INCENDIE OU UN RISQUE DE CHOC
ÉLECTRIQUE, N’EXPOSEZ PAS CET
ÉQUIPEMENT À LA PLUIE OU À L’HUMIDITÉ.
TECHNICAL SPECIFICATIONS
Rating for 120V Low Voltage Chair
This equipment is rated Class I, Type B
Equipment Model
Low Voltage Chair FX920
Volts Hertz Amps
120V 50/60 6.3
Class I Type B
IMPORTANT USER WARNINGS
WARNING: EXPLOSION HAZARD. THIS EQUIP-
MENT MUST NOT BE USED IN THE PRESENCE
OF FLAMMABLE ANESTHETICS.
AVERTISSEMENT : RISQUE D’EXPLOSION.
CET ÉQUIPEMENT NE DOIT PAS ÊTRE UTILISÉ
EN PRÉSENCE D’ANESTHÉSIQUES INFLAMMABLES.
POTENTIAL ELECTROMAGNETIC or OTHER
INTERFERENCE
This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices
in the vicinity. However, there is no guarantee that interference
will not occur in a particular installation.
If this equipment does cause harmful interference to other
devices, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference
by one or more of the following measures:
Rating for 220-240V Low Voltage Chair
This equipment is rated Class I, Type B
Equipment Model
Low Voltage Chair FX920
Volts Hertz Amps
220-240V 50/60 4.0
Class I Type B
U.S. Patent No.
6,106,065
• Reorient or relocate the receiving device.
• Increase the separation between the equipment.
• Connect the equipment into an outlet on a circuit different
from that to which the other device(s) are connected.
• Consult the manufacturer or eld service technician for
help.
WARNING - STORAGE OF THIS EQUIPMENT
BELOW 40°F (4.4ºC) WILL CAUSE THE
HYDRAULIC OIL TO BECOME THICK. THIS
MAY CAUSE UNUSUAL NOISES WHEN
THE BASE MOTOR IS OPERATED. IT IS
RECOMMENDED THAT THE EQUIPMENT
STAY AT A TEMPERATURE OF 60°F (15.6ºC)
OR ABOVE FOR AT LEAST 8 HOURS BEFORE
OPERATING.
AVERTISSEMENT : L’ENTREPOSAGE DE
CET ÉQUIPEMENT À MOINS DE 4,4°C (40°F)
FERA EN SORTE QUE L’HUILE HYDRAULIQUE
DEVIENDRA ÉPAISSE. CECI POURRAIT
CAUSER DES BRUITS INHABITUELS LORSQUE
LE MOTEUR DE LA BASE FONCTIONNE. ON
RECOMMANDE QUE L’ÉQUIPEMENT DEMEURE
À UNE TEMPÉRATURE DE 15,6°C (60°F) OU
PLUS DURANT AU MOINS 8 HEURES AVANT
DE L’UTILISER.
IN-FX9206
2. INSTALLATION
2.1. Unpacking
2.1.1. Chair
2.1.1.1. Cut the shipping bands, open the top of the
shipping carton, and remove all packing material and boxes
that can be easily reached. Remove (4) nuts and remove
both base clamping boards.
2.1.1.2. It is recommended that the skid be positioned as
close as possible to the desired chair location.
2.1.1.3. Remove at least two small shipping blocks stapled
to the top of the skid. The chair can now be slid and/or
tipped from the skid onto the oor.
2.2.2.2. To disconnect power to Chair unplug power cord
from Instrument Stand or from wall receptacle.
NOTE: If this chair is purchased in conjunction
with a Reliance® instrument stand, the chair control
cable supplied will provide proper operation. Refer
to the instrument stand manual for connection to
the stand.
REMARQUE : Si cette chaise est achetée avec
un support à instruments Reliance®, le câble
de commande de la chaise fourni permettra un
fonctionnement adéquat. Se référer au manuel du
support d’instrument pour savoir comment faire la
connexion au support.
3. OPERATION
2.1.1.4. Do not lift the chair by the upper structure.
However, the chair may be tipped or slid into position by
pushing or pulling on the upper structure.
2.1.1.5. With the chair in position, remove all remaining
paper pads, plastic, tape, strings, etc.
2.2. Assembly
2.2.1. Headrest
2.2.1.1. Remove the headrest from its carton, and mount
it into position onto the chair back. Using the proper hex
key wrench, tighten the cap screw to clamp the headrest
securely into its mounting bracket.
2.2.2. Power Receptacle
2.2.2.1. Locate the footswitch assembly, chair control
cable, and the power cord assembly. Refer to gure 1 for
receptacle location and attach all cables.
CAUTION: Caution should be taken to ensure
there are no obstructions (instruments, trays,
tables) impeding movement of the Chair.
PRÉCAUTION : On devrait faire attention
pour s’assurer qu’il n’y a pas d’obstructions
(instruments, plateaux, tables) empêchant le
mouvement de la chaise.
Each time main power is applied to the chair, the circuit
board inside the chair will respond with a one second beep.
The beep indicates the electronic are functioning properly.
3.1. Stop Switch
3.1.1. Depressing the STOP switch on either side of the chair
back will STOP ALL movement of the chair. The STOP switch
is shaped like a stop sign, and may be either RED or ORANGE
in color. It is located above the text “STOP”. (Refer to Figure
2). The STOP switch is also used to program the chair. See
Section 4 (page 12) for programming instructions.
Footswitch
Cable
Chair Control
Cable
IN-FX920
FIGURE 1
BASE
Power Cord &
Fusedrawer
7
3.2. Chair Lift
AUTO
RETURN
BACK
RECLINE
BACK
RAISE
LIFT
DOWN
UP
LIFT
3.2.1. Raising & Lowering the Chair Lift with
Membrane Switches
• The chair lift may be raised or lowered by Membrane
Switches on either side of the chair back (Figure 2).
• To raise the chair lift from these switches, depress the
triangular-shaped UP arrow located ABOVE the text ”BASE”.
This is called the RAISE BASE switch.
• To lower the chair lift, depress the triangular-shaped DOWN
arrow located BELOW the text “BASE”. This is called the
LOWER BASE switch.
REMARQUE : Les interrupteurs RAISE (LEVER)
et LOWER (BAISSER) la base sur le dossier de la
chaise sont programmables. Selon la conception
initiale, le mouvement de la chaise S’ARRÊTE
lorsque l’un ou l’autre de ces interrupteurs est
relâché. Si vous désirez que le mouvement de la
chaise CONTINUE, après que l’un ou l’autre de ces
interrupteurs est relâché, consultez la section 4
pour connaître les instructions de programmation.
3.2.2. Raising & Lowering the Chair Lift from the
Footswitch
• For operation from the footswitch (Figure 3), depress the
right side of the pedal to raise the chair lift.
• Depress the left side of the right-hand pedal to lower the
chair lift.
FIGURE 2
MEMBRANE SWITCH
NOTE: The RAISE and LOWER BASE switches on
the Chair Back are programmable. By design, the
motion of the Chair will STOP when either of these
switches is released. If you want the motion of the
Chair to CONTINUE after either of these switches
is released, then see Section 4 for programming
instructions
.
FIGURE 3
Footswitch Operation
NOTE: The RAISE and LOWER BASE switches on
the Footswitch are programmable. By design, the
motion of the Chair will STOP when either of these
switches is released. If you want the motion of the
Chair to CONTINUE after either of these switches
is released, then see Section 4 for programming
instructions.
IN-FX9208
REMARQUE : Les interrupteurs RAISE (LEVER) et
LOWER (BAISSER) la base sur l’interrupteur au pied
sont programmables. Selon la conception initiale,
le mouvement de la chaise S’ARRÊTE lorsque l’un
ou l’autre de ces interrupteurs est relâché. Si vous
désirez que le mouvement de la chaise CONTINUE,
après que l’un ou l’autre de ces interrupteurs est
relâché, consultez la section 4 pour connaître les
instructions de programmation.
3.3. Chair Top
3.3.1. An electric motor supplies power for the movement
of the chair top.
3.3.2. Raising & Reclining the Chair Top with
Membrane Switches.
• The chair top may be raised or reclined by Membrane
Switches on either side of the chair back (Figure 2).
• To lower, depress the triangle-shaped DOWN arrow located
BELOW the “TOP”. This is called the LOWER TOP or RECLINE
switch.
• To raise the chair top from these switches, depress the
triangle-shaped UP arrow located ABOVE the next “TOP”.
This called the RAISE TOP switch.
NOTE: The RAISE and LOWER/RECLINE switches
on the Chair Back are programmable. By design,
the motion of the Chair will CONTINUE when either
of these switches is released. If you want the motion
of the Chair to STOP after either of these switches
is released, then see Section 4 for programming
instructions.
REMARQUE : Les interrupteurs RAISE (LEVER)
et LOWER/RECLINE (BAISSER/INCLINER) sur
le dossier de la chaise sont programmables.
Selon la conception initiale, le mouvement de la
chaise CONTINUE lorsque l’un ou l’autre de ces
interrupteurs est relâché. Si vous désirez que
le mouvement de la chaise S’ARRÊTE, après
que l’un ou l’autre de ces interrupteurs est
relâché, consultez la section 4 pour connaître les
instructions de programmation.
CAUTION - ANY chair movement that
CONTINUES after the switch is released may
be stopped by depressing the STOP switch
or by depressing the same switch a second
time.
PRÉCAUTION : TOUT mouvement de la
chaise qui CONTINUE après que l’interrupteur
est relâché peut être arrêté en appuyant sur
l’interrupteur STOP (ARRÊT) ou en pressant le
même interrupteur une seconde fois.
3.3.3. Raising & Lowering the Chair Top From the
Footswitch.
• For operation from the footswitch (Figure 3), depress the
right side of the left-hand pedal to raise the chair top.
• Depress the left side of the left-hand pedal to lower/recline
the chair top.
NOTE: The RAISE and LOWER/RECLINE switches
on the Footswitch are programmable. By design,
the motion of the Chair will STOP when either
of these switches is released. If you want the
motion of the Chair to CONTINUE after either of
these switches is released, then see Section 4 for
programming instructions.
REMARQUE : Les interrupteurs RAISE (LEVER)
et LOWER (BAISSER/INCLINER) sur l’interrupteur
au pied sont programmables. Selon la conception
initiale, le mouvement de la chaise S’ARRÊTE
lorsque l’un ou l’autre de ces interrupteurs est
relâché. Si vous désirez que le mouvement de la
chaise CONTINUE, après que l’un ou l’autre de ces
interrupteurs est relâché, consultez la section 4
pour connaître les instructions de programmation.
IN-FX920
9
3.4. Automatic Return
3.4.1. This feature is employed as a simple way for the
operator to lower the chair lift and raise the chair back after
completion of an examination or treatment. The automatic
return system is activated by depressing the square switch
located ABOVE the text “AUTO” on either side of the chair
back. The chair lift, if raised, will then lower, and the chair
back, if reclined, will rise to the upright position. These
movements will CONTINUE after the switch is released until
the chair lift is fully lowered and the chair back is fully upright.
These movements may also be stopped by depressing the
STOP switch or by depressing the AUTO switch a second
time. The AUTO switch is square and is located in the same
place on both Membrane Switches.
3.4.2. These instructions and operations also apply to any
AUTO switch that is installed in a footswitch, console, or
base cap.
NOTE: Memory switches on the right and left
sides of the Chair Back work independently of
each other. If the Memory Feature is activated on
the left side, the right side will continue to recline
the Chair all the way back with one touch. If you
also want to activate the Memory Feature on the
right side of the Chair, repeat Steps 1-4 on the right
side of the Chair.
REMARQUE : Les interrupteurs à mémoire sur
les côtés droit et gauche du dossier de la chaise
fonctionnent indépendamment l’un de l’autre. Si
la fonctionnalité de mémoire est activée du côté
gauche, le côté droit continuera à incliner la chaise
jusqu’au bout avec une seule touche. Si vous
désirez aussi activer la fonctionnalité de mémoire
du côté droit de la chaise, répétez les étapes 1 à
4 sur le côté droit de la chaise.
3.5. Memory Feature
3.5.1. This system allows the operator to select a preferred
examining position. The chair will automatically stop reclining
when that point is reached.
3.5.2. Programming the Memory Feature
Step 1) Press the AUTO Switch. The chair will lower the
chair base to its lowest position. It will also raise
the chair back to its most upright position. Wait
20 seconds.
Step 2) Press and hold the RAISE TOP Switch. The chair
back will NOT move since it is already upright.
While maintaining pressure on the RAISE TOP
switch, press the RECLINE switch. The chair will
enter a programming mode and begin to recline
and the switches will blink rapidly. At this point
release the switches.
Step 3) Press the STOP Switch when the chair top arrives
at the desired recline position. This completes the
programming setup.
Step 4) To use the memory feature once it has been
programmed, press the AUTO Switch. The LED’s
will blink and the chair will return to its full upright
position. The LED’s will stop blinking. Then press
the RECLINE Switch which is programmed for
memory mode. (Specify which RECLINE Switch
will utilize this feature.) The switches will blink and
the chair will recline on its own and stop at the
programmed point.
NOTE: The Memory Feature does not control the
raise motion of the Chair Back.
REMARQUE : La fonctionnalité de mémoire
ne commande pas le mouvement d’élévation du
dossier de la chaise..
3.6. Revolving/Rotation
3.6.1. The chair upper structure may be revolved 350°,
with a built-in stop to prevent continuous rotation. The lock
for revolving is controlled by a hand lever with chrome plated
knob located below the seat at the center and on either side.
Push the lever in the direction of footrest to apply the lock
and in opposite direction to release the lock.
NOTE: The lock must be applied each time
before the Chair is used. 30 lbs. of locking force
is necessary to lock the Chair for one working
day. Vibration of the Chair during transportation or
movement on a rough surface will release the lock..
REMARQUE : Le verrouillage doit être appliqué
chaque fois avant que la chaise soit utilisée. 30
lb de force de verrouillage est nécessaire pour
verrouiller la chaise pour une journée de travail.
La vibration de la chaise durant le transport ou le
mouvement sur une surface rugueuse relâchera
le verrouillage.
IN-FX92010
3.7. Arm Movement
3.9.3. Adjustment-No.26 Headrest
3.7.1. The chair arms may be raised to provide better
access to the patient or to permit entrance to or exit from
the chair.
3.7.2. The movements of both arms are identical and
completely independent of each other.
3.7.3. To raise an arm, push rearward on the middle joint
of the arm support link while raising upward on the arm. A
stop limits upward motion of the arm to a point parallel to
the plane of the back. Lowering the arm automatically locks
it in place against vertical motion in either direction.
3.8. Footrest
3.8.1. The footrest may be retracted by lifting near its outer
edge. When fully retracted the footrest will stay in place.
3.9. Headrest
3.9.1. The No. 26 headrest is the standard headrest for
the Model FX920 Chair.
WARNING- IF THE NO. 26 HEADREST
RECEIVES A IMPACT OR STRIKES AN OBJECT
WITH FORCE CONTACT RELIANCE® TECH
SERVICE IMMEDIATELY.
AVERTISSEMENT : SI L’APPUIE-TÊTE # 26
REÇOIT UN CHOC OU FRAPPE UN OBJET
AVEC FORCE, CONTACTEZ IMMÉDIATEMENT
LE SERVICE TECHNIQUE RELIANCE®.
3.9.3.1. There are three adjustments provided on the No.
26 headrest. The rst is the tension on the pad rotation.
By tightening the two hex acorn nuts on the pad pivots with
a 3/8” open end wrench, the tension will be increased. To
adjust the drag on the pivots of the articulating arm, loosen
the set screw in the center of the pivot with (Figure 4, item
1) a 3/32” hexagon key wrench. Next, tighten the pivot
screw on the other side of the pivot, with a 1/8” hexagon
key wrench. Now, retighten the set screw to retain the
pivot screw at the new adjustment. Either pivot may be
adjusted.
ACORN NUTS
2
BLACK DOT
ACTUATOR
1
SAFETY/SHUTDOWN
3.9.2. Operation-No. 26 Headrest
3.9.2.1. The No. 26 headrest is designed so it can be
operated with just one hand. To unlock, squeeze the clear
plastic actuator into lock housing. This will allow both pivots
of the articulating arm to rotate so the headrest can be
moved to a new desired position. Once in the new position,
release the actuator. Make sure the actuator expands to its
original position fully exposing black dot on handle to insure a
positive lock. The headrest pad is free to rotate at all times.
(See Figure 4).
IN-FX920
FIGURE 4
3.9.3.2. The third adjustment is for proper clearance of the
block dot to indicate that the headrest is properly locked.
If the dot does not fully clear the lock housing when the
headrest is lock securely, turn the set screw located at the
top of the lock housing (Figure 4, item 2) counterclockwise
until the dot on the actuator is fully exposed.
11
4. PROGRAMMING THE CHAIR
The items described in Section 4. are built-in features of the
chair. Only those users who wish to utilize these features need
to be concerned with this Section. The chair will operate as
previously designed if these features are not used. The chair
can be programmed in the same fashion from either switch
located on the chair back.
4.1. Safety/Shutdown Feature
when the switch is released, this is called the MOMENTARY
MODE of operation. If the movement of the chair continues
after the switch is released, this is called the MAINTAINING
MODE of operation. The factory designed setting for the
RAISE TOP and RECLINE TOP switches located on the chair
top is MAINTAINING MODE. The factory designed setting for
the RAISE BASE and LOWER BASE switches located on the
chair top is MOMENTARY MODE. If changed, the new settings
will remain permanent as long as main power remains applied
to the chair and will apply to the similar function on the other
side.
The chair has a built-in Safety/Shutdown feature. This feature
causes the chair to shutdown or become inoperable. This
feature prevents unauthorized use of the chair. (See Figure
2).
4.1.1 To ENTER Safety/Shutdown Mode
4.1.1.1 Depress and HOLD either STOP switch.
4.1.1.2 With the STOP switch depressed, depress the
LOWER BASE switch on the same side of the chair back.
4.1.1.3 A two second beep will be heard to indicate the chair
will NOT move if any additional switches are depressed.
4.1.1.4 The LED’s on the Membrane Switches will blink every
three seconds while in Safety/Shutdown Mode.
4.1.2 To EXIT Safety/Shutdown Mode, repeat the
above sequence.
4.1.2.1 Depress and HOLD either STOP switch.
4.1.2.2 With the STOP switch depressed, depress the LOWER
BASE switch on the same side of the chair back.
4.2.2 To change these settings the chair must be placed in
Setup Mode. Once in Setup Mode, depressing one of these
switches will change its present setting from MOMENTARY
to MAINTAINING or from MAINTAINING to MOMENTARY.
Depressing a switch repeatedly will cause the setting to toggle
between the two modes of operation.
4.2.3 To ENTER Setup Mode and change current
settings
4.2.3.1 Depress and HOLD either STOP switch.
4.2.3.2 With the STOP switch depressed, depress the AUTO
switch on the same side of the chair back.
4.2.3.3 A one second beep will be heard to indicate you
have ENTERED Setup Mode. The LED’s will blink on and off
while in Setup Mode.
4.2.3.4 Depress any of the following switches (including
footswitch), one at a time, to toggle its current setting - RAISE
TOP, RECLINE TOP, RAISE BASE, OR LOWER BASE. When
the switch is depressed, a SHORT beep (0.25 seconds) will
indicate the switch is now in MOMENTARY MODE of operation.
A LONG beep (1.05 seconds) will indicate the switch is now
in MAINTAINING MODE of operation.
4.1.2.3 A two second beep will be heard to indicate the chair
is functional, and movement of the chair WILL OCCUR when
another switch is depressed.
4.1.2.4 The LED’s will come back on when the chair becomes
operational.
4.2. Individual Programming of the RAISE TOP,
RECLINE TOP, RAISE BASE, and LOWER
BASE switch located on the Chair Back or
Footswitch. (See Figure 2 & 3).
4.2.1 These individual switches may be programmed to
function in two ways. If the movement of the chair STOPS
4.2.3.5 When nished programming the chair, the chair MUST
be taken out of Setup Mode.
4.2.4.2 With the STOP switch depressed, depress the AUTO
switch on the same side of the chair back.
4.2.4.3 A one second beep will be heard to indicate the chair
has EXITED Setup Mode. The LED’s will come back on.
IN-FX92012
NOTE: Depressing a switch causes the mode
of operation to TOGGLE between MOMENTARY
and MAINTAINING operation. If a mistake is made,
depress the same switch again to return that switch
to its previous mode of operation. If, at any time,
the user becomes lost or confused during the
programming procedure, the chair can be reset
to the factory settings by disconnecting and then
reconnecting main power to the chair. Once reset,
the programming procedure can be repeated.
REMARQUE : Appuyer sur un interrupteur fait
BASCULER le mode de fonctionnement entre
MOMENTANÉ et MAINTENU. Si vous avez fait
une erreur, appuyez de nouveau sur le même
interrupteur pour le ramener à son mode de
fonctionnement préalable. Si, en tout temps,
l’utilisateur devient perdu ou confus durant la
procédure de programmation, la chaise peut être
réinitialisée aux réglages d’usine en déconnectant
et en reconnectant l’alimentation principale à la
chaise. Une fois réinitialisée, la procédure de
programmation peut être répétée..
4.3. BEEP-ON-SWITCH Feature
4.3.1 The chair can be programmed to beep whenever a
switch is depressed. This gives the user additional feedback
on the operation of the chair. The factory design is to have
NO BEEP occur when a switch is depressed. To change this
setting the chair must be in Setup Mode. See Section 4.2.4.
on how to ENTER Setup Mode, once in Setup Mode.
4.3.2 Depress either AUTO switch on the chair back
to toggle the BEEP-ON-SWITCH feature between ON and
OFF.
4.3.3 A short beep (0.25 seconds) indicates NO BEEP
will occur when a switch is depressed. A long beep
(1.05 seconds) indicates a BEEP occur when a switch is
depressed.
4.4.1 To ENTER Auto Up/Retinal Mode.
4.4.1.1 Depress and HOLD either STOP switch.
4.4.1.2 With the STOP switch depressed, depress the
LOWER TOP switch on the right side of the chair back to
program that switch for Retinal Mode.
4.4.1.3 A one second beep will be heard to indicate that
switch has ENTERED Auto Up/ Retinal Mode. When depressed,
the RECLINE TOP switch on the right-hand side of the chair will
now cause the chair lift to rise and the chair top to recline.
4.4.1.4 To program the RECLINE TOP switch on the left-
hand side of the chair, use that switch and repeat the same
steps as described above.
CAUTION: ANY Chair movement that
CONTINUES after the switch is released may
be stopped by depressing the STOP Switch or
by depressing the same switch a second time.
PRÉCAUTION : TOUT mouvement de la
chaise qui CONTINUE après que l’interrupteur
est relâché peut être arrêté en appuyant sur
l’interrupteur STOP (ARRÊT) ou en pressant le
même interrupteur une seconde fois.
4.4.2 To EXIT Auto Up/Retinal Mode, repeat the
above sequence.
4.4.2.1 Depress and HOLD either STOP switch.
4.4.2.2 With the STOP switch depressed, depress the LOWER
TOP switch on the same side of the chair back.
4.4.2.3 A one second beep will be heard to indicate that
switch has EXITED Auto Up/Retinal Mode.
4.3.4 When nished programming the chair, the MUST
be taken out of Setup Mode. See Section 4.2.4. on how to
EXIT Setup Mode.
4.4. Auto Up/Retinal Feature
The Auto Up/Retinal Feature is a convenient feature for
Retinal Specialists. When either recline switch is programmed
to operate in this mode, depressing that switch will cause
the chair lift to rise and the chair top to recline
IN-FX920
4.5. Home Feature
4.5.1 The Home Feature allows the user to program the
distance the chair lift will travel when starting from the fully
lowered position. Activating this Feature causes the chair to
move upward for a xed amount of time. The actual height
achieved by the chair will vary slightly from the programmed
height, depending on the weight of the patient.
13
4.5.2 The Home Feature can be initiated by any of the
RAISE BASE switches, located on the chair back, footswitch,
or instrument stand. Each switch can be independently
programmed for a specied height.
4.5.3 To PROGRAM the Home Mode Feature.
REMARQUE : La fonctionnalité Accueil et la
fonctionnalité Mémoire peuvent être utilisées
seulement lorsque la base est en bas et que
l’arrière est entièrement relevé. Consultez la
Section 3.5.2 Étape 4..
4.5.3.1 The chair lift must be fully lowered. To make sure the
lift is all the way down, DEPRESS the AUTO switch and wait
at least fteen seconds.
4.5.3.2 The RAISE BASE switches that are to be programmed
for Home Mode, must be in Momentary Mode.
4.5.3.3 DEPRESS and HOLD either STOP switch located on
the chair back.
4.5.3.4 With the stop switch depressed, DEPRESS the RAISE
BASE switch to be programmed. The chair lift will begin to
rise and the LED’s will blink. Release both switches.
4.5.3.5 When the chair rises to the desired height, DEPRESS
either STOP switch to program that height into the chair.
4.5.3.6 Repeat steps 1 thru 5 to program any other RAISE
BASE switch for Home Mode.
4.5.4 To UTILIZE the Home Mode Feature.
4.5.4.1 When the chair is all the way down, depressing a Raise
Base Switch that has been programmed for Home Mode will
cause the chair to rise to the programmed height.
4.5.4.2 If a RAISE BASE switch is programmed for Home
Mode, but the chair lift is not all the way down, the Raise Base
switch will function like a momentary switch.
5.0 LOCATING AND REPLACING FUSES
WARNING: DISCONNECT EQUIPMENT
FROM MAIN INPUT POWER BEFORE
PROCEEDING WITH ELECTRICAL
INSPECTIONS OR MAINTENANCE.
AVERTISSEMENT : DÉCONNECTEZ
L’ÉQUIPEMENT DE L’ALIMENTATION
ÉLECTRIQUE PRINCIPALE AVANT
DE PROCÉDER À DES INSPECTIONS
ÉLECTRIQUES OU À DE L’ENTRETIEN.
CAUTION: Replace fuse(s) as marked. All
fuses must be replaced with a fuse of the same
size and rating. Refer to the Wire Diagrams
at the end of this Manual.
4.5.4.3 If any RAISE BASE switch is programmed for Home
Mode, the LED’s will blink for 15 seconds after the AUTO
DOWN switch is depressed. When the LED’s stop blinking,
the chair lift is all the way down.
4.5.5 To EXIT the Home Mode Feature.
4.5.5.1 If the Home Feature is no longer needed for any of
the RAISE BASE switches, depress a STOP switch and the
desired RAISE BASE switch at any time when the chair lift is
not all the way down.
NOTE: The Home Feature and Memory Feature can
only be operated with the Base down and the Back
fully upright See Section 3.5.2 Step 4
PRÉCAUTION : Remplacez le(s) fusible(s)
comme indiqué. Tous les fusibles doivent
être remplacés par un fusible de la même
dimension et de la même valeur. Se référer
aux schémas de câblage à la fin de ce manuel.
The chair contains two fuses located inside the AC Input
Module in the Base Assembly.
IN-FX92014
5.1. Fuse Replacement (AC Input Module)
Unplug the power cord from the chair base. With a tool
similar to a small common screwdriver, unsnap the locking
tab on the bottom of the fuse drawer. Remove and examine
the fuses. Replace as necessary. See Figure 5.
FIGURE 5
AC INPUT MODULE
Method for removing fuse drawer:
1. The fuse drawer is located above the power cord.
2. Remove the power cord.
3. Place a small bladed screwdriver into appropriate slot
(shown above) and push the at spring towards the center
of the power module until a click is heard. (Click may not
be heard in all cases.) The drawer will come forward
slightly.
4. Once the ap has been pushed towards the center, the
drawer will extend slightly beyond the power module.
5. Remove the drawer by pulling it out by your ngers.
CAUTION: To avoid a bind of the Fuse Drawer
within the AC Input Module, never install the
Fuse Drawer into the AC Input Module without
fuses.
PRÉCAUTION : Pour éviter que le tiroir à
fusible reste bloqué dans le module d’entrée
CA, n’installez jamais le tiroir à fusible
à l’intérieur du module d’entrée CA sans
fusibles.
IN-FX920
15
TROUBLESHOOTING GUIDE
PROB L E MPROB L E M C A U SESOLU T I ON
CHAIR WILL NOT OPERATE.1) POWER CORD NOT PLUGGED IN.
2) NO AC POWER AT WALL
RECEPTACLE.
3) BLOWN LINE FUSE.
OUTLET POWER OK, BUT NO
BEEP IS HEARD WHEN
POWER IS APPLIED.
FOOTSWITCH FUNCTIONS
NOT OPERATING.
BACKSWITCH FUNCTIONS
NOT OPERATING.
CHAIR BASE WILL NOT STAYUP1) SOLENOID VALVE NOT CLOSED.
CHAIR BASE WILL NOT
LOWER.
CHAIR BASE WILL NOT RAISE.
PUMP MOTOR DOES NOT RUN.
1) BLOWN LINE FUSE.
2) LOOSE WIRING IN BASE.
3) OTHER.
1) FOOTSW
CONNECTED.
1) DAMAGED OR DISCONNECTED WIRES
IN CHAIR BACK, UNDER CHAIR SEAT,
OR IN CHAIR BASE
( DIRT IN VALVE.)
2) CONSTANT POWER APPLIED TO
SOLENOID.
3) OTHER
1) NO POWER TO PUMP MOTOR ASS'Y.
2) SOLENOID COIL BURNED-OUT
3) OTHER
1) NO POWER TO PUMP MOTOR ASS'Y
2) UP-LIMIT SWITCH STUCK "OPEN".
3) PUMP MOTOR BURNED-OUT.
4) OTHER
ITCH CORD NOT
1) CHECK ALL WIRING
CONNECTIONS.
2) RESTORE OUTLET POWER.
3) REPLACE FUSE.
1) REPLACE FUSE.
2) CHECK ALL WIRING
CONNECTIONS.
3) CONTACT FACTORY
1) CHECK ALL INTERNAL AND
EXTERNAL FOOTSWITCH WIRING
1) CHECK ALL WIRING
CONNECTIONS.
1) REPAIR OR REPLACE SOLENOID
VALV
2) CHECK FOR SWITCHES STUCK
"CLOSED" .
3) CONTACT FACTORY.
1) CHECK FOR VOLTAGE.
2) REPLACE SOLENOID COIL.
3) CONTACT FACTORY.
1) CHECK FOR VOLTAGE.
2) RESWITCH.
3) REPLACE PUMP MOTOR ASS'Y.
4) CONTACT FACTORY.
E.
ADJUST OR REPLACE
CHAIR BASE WILL NOT RAISE.
PUMP MOTOR RUNS.
BASE "STUTTERS" DURING
ASCENT.
CHAIR BASE WILL NOT RAISE
OR LOWER FROM FLOOR
UNIT.
CHAIR TOP "BINDS-UP".1) CHAIR TOP MOTOR IS OUT OF
CHAIR TOP "WOBBLES" ON
CHAIR BASE
MEMORY SYSTEM
INOPERATIVE.
1) SOLENOID VALVE NOT CLOSED.
( DIRT IN VALVE.)
2) CONSTANT POWER APPLIED TO
SOLENOID.
3) OTHER.
1) HYDRAULIC FLUID LEVEL IS LOW.1) CONTACT DEALER OR F
1) CHAIR CONTROL CABLE NOT
PLUGGED
IN.
2) BLOWN FUSE IN CHAIR BASE.
ADJUSTMENT.
1) LOCK NUT IS LOOSE.
( UNDER SEAT BOARD.)
1) NOT PROGRAMMED PROPERLY.
2) LOOSE CONNECTIONS UNDER
CHAIR
SEAT.
1) REPAIR OR REPLACE SOLENOID
VALVE.
2) CHECK FOR SWITCHES STUCK
"CLOSED".
3) CONTACT FACTORY.
ACTORY
FOR INFORMATION ON ADDING
OIL.
1) CHECK CABLE CONNECTIONS AT
STAND AND CHAIR
2) REPLACE FUSE.
1) CONTACT DEALER OR FACTORY
FOR ADJUSTMENT PROCEDURE.
1) TIGHTEN LOCK NUT.
1) SEE SECTION 3.5.1 OR SECTION
4.
2) CHECK ALL WIRING
CONNECTIONS.
IN-FX92016
CLEANING AND MAINTENANCE
Cleaning the Upholstery
When cleaning upholstery, it is suggested that you use all
recommended commercial cleaning agents in an inconspicuous spot to check for potential damage to the upholstery
material before applying to the entire area to be cleaned.
Day to Day soil:
Remove ordinary dirt and smudges with a mild soap and
warm water solution. Dry with a soft, lint-free cloth or towel.
For more difcult stains, use of a stronger detergent is
recommended; however, follow the detergent manufacturer’s
instructions closely.
PRÉCAUTION: Ces solvants sont extrêmement
inflammables. Faites très attention lors du
nettoyage et avisez le personnel dans la
zone de tout danger. Portez des gants de
caoutchouc durant toute activité de nettoyage.
Faites attention en nettoyant autour des
boutons, des coutures et autres finitions de
bois ou décoratives, puisque ces solvants
pourraient endommager sérieusement ces
endroits.
Cleaning Painted Surfaces
Disinfecting Material:
In order to disinfect, use a 1:9 solution of bleach and water (1
part bleach to 9 parts water). Wipe surface with damp cloth
then rinse with water immediately, using a damp cloth.
Special cleaning problems:
The following methods are recommended to clean stains
on PreFixx-protected vinyl upholstery. Most stains will be
removed when these cleaning agents are used in the following order.
Type 1 Cleaners: Nonabrasive household cleaners to be
used with water and a soft cloth.
• Formula 409~All-Purpose Spray Cleaner
• Fantastik® Spray Cleaner
• Other similar household cleaners and bleaches
Type 2 Cleaners: Solvent-type cleaner to be liberally applied with a soft cloth. Dry area with another cloth, rinse
with clean water and dry.
• Rubbing alcohol (isopropyl alcohol)
• Lighter uid (naphtha)
Type 3 Cleaners: Strong, active solvent cleaners such as
nail polish remover (acetone/water) to be applied with a soft
cloth. Stain should be removed with less than six (6) rubs; if
stain persists after six rubs, contact manufacturer. Dry area
with another cloth, rinse with clean water and dry.
The painted metal surfaces are covered in durable, powdercoated paint which is resistant to scratching and scufng.
It may be cleaned with a clean cloth dampened with mild,
soapy water or equivalent household product that uses spray
application.
Recommended Infection Control Products
The disinfectant/cleaner products that we have tested and
recommend for our upholstery, painted items and plastic
covers are as follows.
• Cavicide® Hospital Disinfectant
• Precise® Hospital Foam Cleaner Disinfectant
CAUTION - Follow manufacturer’s directions
for concentration and application of disinfecting
cleaner. Avoid prolonged application of any
disinfectant/cleaner products, because
they may cause staining or discoloration of
material.
PRÉCAUTION: Suivez les instructions
du fabricant pour la concentration et
l’application de nettoyeurs désinfectants.
Évitez une application prolongée de tout
produit désinfectant/nettoyant, parce que
ceux-ci pourraient causer des taches ou une
décoloration du matériel.
CAUTION- These solvents are highly
flammable. Exercise proper care in cleaning
and notify personnel in area of danger. Wear
rubber gloves during all cleaning activity. Use
caution in cleaning around buttons, stitching
and wooden or other decorative trim, since
these solvents could seriously damage such
areas.
IN-FX920
WARNING: REFER TO LOCAL OR
HEALTHCARE PROFESSION GUIDELINES
FOR SELECTING EFFECTIVE INFECTION
CONTROL PRODUCTS.
AVERTISSEMENT : SE RÉFÉRER
AUX DIRECTIVES LOCALES OU DES
PROFESSIONS DE LA SANTÉ POUR
SÉLECTIONNER DES PRODUITS DE
CONTRÔLE D’INFECTION EFFICACES.
17
PARTS LIST
RELIANCE® MODEL FX920
LOW VOLTAGE CONTROL CHAIR
Note: When ordering parts, please:
1. Advise dealer or factory of model and serial number of unit. These numbers are on the label that
is located at rear, near bottom, of chair back.
2. Specify color of painted parts. Painted parts have an asterisk(*) behind the part description.
Note : Quand pièces de commande, s’il vous plaît :
1. Informez le distributeur ou l’usine du modèle et du numéro de série de l’ensemble. Ces numéros
sont de la plaque qui est située à l’arrière, près de la base, de l’unité centrale.
2. Indiquez la couleur des pièces peintes. Les pièces peintes ont un astérisque (*) derrière la
description de pièce.
Should your Exam Chair not perform properly, or if replacement parts
are needed, contact the Reliance® Distributor, where you purchased this
product or contact the Technical Service Department, Reliance® Medical
Products, Inc., 3535 Kings Mills Road, Mason, Ohio 45040-2303, or call
(800) 735-0358.
IN-FX92018
FIGURE 6
FX920 CHAIR ASSEMBLY
ADJUSTABLE
13.3"
3
1
5, 6
4
13
7, 8
11, 12
14
SEAT HEIGHT (MIN)
MODEL AND SERIAL
NUMBER PLATE
2
9, 10
HIGH BASE 24 3/4"
(10 5/8" TRAVEL)
LOW BASE 22 1/4"
(8 1/2" TRAVEL)
5.5"
47.25"±.25 HIGH BASE
44.75"±.25 LOW BASE
AT LOWEST POSITION
BASE
24.58"
25.8"
32.5"
21.3"
32.5"
36.5"
ITEMPART NO.DESCRIPTIONITEMPART NO.DESCRIPTION
1
2
3
1760488Footstep
4
1764073Back Board Assembly**
5
1761273Upholstered Back Assembly**
6
1779273Seat Board Assembly**
7
Chair Top Assembly* (See Figure 7 )
Base Assembly* (See Figure 8)
#26 Headrest Assembly* (See Figure10)
10
12
13
14
1761073Upholstered Seat Assembly**
8
1764273Apron Board Assembly**
9
1760973Upholstered Apron Assembly**
1767873Center Board Assembly**
11
1761173Upholstered Center Assembly**
1761373Upholstered Arm Assembly
1765073Scuff Pad**
** Please specify upholstery color when ordering.
To order complete Upholstery Set use P/N.
**
1782073 Upholstery Set - FX920
IN-FX920
19
38
FIGURE 7
CHAIR TOP ASSEMBLY
39
5
40
45
30
37
41
42
11
52
(2)
22
43
35
29
1
44
9
14
47
48
49
36
6
51
13
50
18
48
17
34
20
26
46
31
24
23
44
4
33
19
16
8
25
2
27
10
28
12
3
43
21
52
(2)
15
16
IN-FX92020
7
ITEM
PART NO.DESCRIPTIONITEMPART NO.DESCR IPTION
1
1766585Cradle270643582Lock Plug
2
1564995Seat Support281766485Apron Platform Support
3
1518685Apron Bracket - R.H.*290722699Set Screw 1/4 x 1/4
*Please specify paint color when ordering. 2721792 Bottom Joint Screw Assembly
** Please specify upholstery color when ordering.
Includes: Item’s 5, 6, 8 & 18
IN-FX92027
FIGURE 10
WIRING DIAGRAM - FX920
IN-FX92028
IN-FX920
29
IN-FX92030
LIMITED WARRANTY
The Reliance® product must be used only for the purposes and in the manner described in the literature
distributed with the product. The products are warranted against defective materials and workmanship for a
period of one (1) year from date of installation. Products or parts thereof will be repaired or replaced as required
at Reliance® Medical Products, Inc. Such repair or replacement shall be the sole remedy under this warranty.
This warranty extends only to the original purchaser from an authorized Reliance® dealer and is subject to the
following conditions:
1. The warranty card must be completed and returned to Reliance® Medical Products, Inc. within two (2) weeks
from the date of installation.
2. Installation and servicing of the products must be performed by trained Reliance® equipment dealer service
personnel in accordance with the appropriate instructions manual for the products. Warranty will be voided if
installation is performed by a non-authorized individual
3. The products have not sustained breakage or other types of damage due to accident or misuse.
4. This warranty will not apply to Reliance® products which have had the serial number removed, altered or
effaced.
EXCEPT FOR THE EXPRESS WARRANTY SET FORTH ABOVE, RELIANCE® DOES NOT GRANT ANY
WARRANTIES, EITHER EXPRESS OR IMPLIED, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILTY
OR FITNESS FOR A PARTICULAR USE.
Reliance® Medical Products neither assumes nor authorizes any person to assume for it, any other liability
in connection with the sale and use of its products.
REMEDIES ARE LIMITED EXCLUSIVELY TO REPAIR OR REPLACEMENT OF PARTS. RELIANCE®
MEDICAL PRODUCTS, INC. EXPRESSLY DISCLAIMS LIABILITY FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGE RESULTING FROM THE USE OF THE EQUIPMENT.
Claims covered by this warranty will be honored when presented within one (1) month from discovery of a defect.
IN-FX92031
EC
Authorized European Community Representative
REP
HAAG-STREIT UK
Edinburgh Way
Harlow, Essex, England
CM20 2TT
United Kingdom
MADE IN AMERICA
Reliance MEdical Products, Inc.
3535 Kings Mills Road
Mason, Ohio 45040-2303 USA
1-800-735-0357(Customer Service)
1-800-735-0358 (Technical Service)
(513) 398-3937
Http://www.reliance-medical.com