Specifications
Measuring Range 0 to 999 ppm Accuracy ±5% of reading or ±10 ppm Resolution 1 ppm Sensor Type Stabilized electrochemical sensor Measuring Range 32 to 122°F (0 to 50°C) Accuracy ±1.2°F (0.6°C) Resolution 0.1°F (0.1°C) Measuring Range 10-90% RH Accuracy ±3% RH (at 25°C, otherwise ±5%) Resolution 0.1% RH Display LCD F/C Switchable Yes Alarm Audible (buzzer), Visible (LED) Adjustable CO Alarm Limit Yes Max Hold Yes Date & Time Yes Warm-up Time 60 seconds Desktop Stand Yes Wall Mountable Yes Power Supply 4 “AAA” batteries, AC Adaptor (optional) Product Certications CE Operating Temperature 32 to 122°F (0 to 50°C) Storage Temperature -4 to 122°F (-20 to 50°C) Operating Humidity Range 0-90% (non-condensation) Dimensions 4.4 x 4.3 x 1.1” (113 x 108 x 28mm) Weight 5.6 oz (159g)
Operating Instructions
Taking Measurements
This meter starts taking measurements when powered on and updates readings every 2 seconds. When turning on the meter, the CO sensor needs 1 minute warm up. The CO alarm default is 30 ppm and may be adjusted by the user.
0 to 1 ppm Normal background levels
9 ppm ASHRAE Standard 62-1989 for living areas
Maximum Function
Press MODE button, LCD displays MAX CO, Temperature and Humidity reading. To reset the max, hold MAX button while in MAX mode and the display will show Clr, confirming memory has been cleared.
35 ppm The average exposure level per U.S. OSHA workplace standards
50 ppm OSHA enclosed space 8-hour average level*
100 ppm OSHA exposure limit*
200 ppm Mild headache, fatigue, nausea and dizziness
400 ppm Headache occurs in 2.5 to 3.5 hours
800 ppm Dizziness, nausea and convulsions Death within 2 to 3 hours
* U.S. Department of Labor, Occupational Safety & Health Adminis­tration (OSHA) Regulation 1917.24: The CO content in any enclosed space shall be maintained at not more than 50 ppm (0.005%). Remove employees from enclosed space if the CO concentration exceeds 100 ppm (0.01%).
Features
• Measures carbon monoxide (CO), air temperature and
Setting the Date & Time
The meter displays current date and time alternately, each cycle lasts 6 seconds. To enter the date/time setting press MODE and POWER at the same time. Press SEL/R to increase the digit and TRH/M to decrease the digit. To edit the Year, Month/Date and Hour/Minute press MODE. Press UNIT to select the Month or Date, Hour or Minute. Once complete, press MODE and POWER to save and exit. Please note time is reset when meter is powered off.
Toggle Temperature
Press the UNIT button to toggle unit of measure.
Power ON/OFF
Press the POWER button and the LCD will show current CO reading, Temperature, Humidity, Date and Time.
Alarm Setting
In normal measurement mode, press the POWER button to toggle alarm ON or OFF. While alarm is on, the red LED will flash if the data is exceeding the set alarm limit. Press SEL/R and POWER to set CO alarm threshold. Press UNIT to change digit. Press SEL/R to increase the digit and TRH/M to decrease the digit. Once complete, press SEL/R and POWER to save and exit.
relative humidity
• User adjustable alarm limit
• Built-in clock and calendar
• Max hold function
• Can be mounted on a wall or placed on desktop
• Manual CO calibration to zero
• Low battery indicator
• Includes batteries
Manual Calibration to Zero
Before starting make sure the area is well ventilated and there is no CO in the environment. When unit is turned on, hold UNIT for 6 seconds. The CO will display “0000 and flash for 60 seconds.
Error Codes Er 1 Sensor Failed Er 2 Out of measured range
Display Description
1) Carbon Monoxide (CO)
2) CO alarm indicator
3) Relative Humidity Temperature
4) Air Temperature
Find Quality Products Online at: sales@GlobalTestSupply.com
5) Year / Month / Day / Time (Rotates)