REED Instruments R7800-NIST, R7800 User manual

Coating Thickness Gauge
R7800
Instruction Manual
www.GlobalTestSupply.com
Table of Contents
Introduction ................................................................................................... 3
Product Quality .............................................................................................. 3
Safety ............................................................................................................ 4
Features ......................................................................................................... 4
Included ......................................................................................................... 4
Specications ................................................................................................ 5
Instrument Description .................................................................................. 6
Display Description ....................................................................................... 6
Operating Instructions ................................................................................... 7
Power ON/OFF ......................................................................................... 7
Measuring Procedure ............................................................................... 7
Setup Mode .............................................................................................. 8-12
Enabling/Disabling Audible Beep ............................................................. 9
Setting the LCD Brightness ...................................................................... 9
Selecting the Thickness Unit of Measure ................................................. 9
Selecting Calibration Mode ...................................................................... 9
Setting Higher Limit Value ...................................................................... 10
Setting Lower Limit Value ....................................................................... 10
Enabling/Disabling LED Indicators ......................................................... 11
Factory Reset ......................................................................................... 11
Selecting Continuous Measurement Mode ............................................ 11
Delete Recorded Data ............................................................................ 12
Minimum, Maximum, and Average Values .................................................. 12
Quick Decision Mode ............................................................................. 12-13
Single-Point Decision Configuration ...................................................... 13
Multi-Point Decision Configuration ........................................................ 13
www.GlobalTestSupply.com
continued...
Zero Calibration ......................................................................................14-16
Zero Calibration ...................................................................................... 14
Two-Point Calibration ............................................................................. 15
Calibration Verification ............................................................................ 16
Locking Screen Auto-Rotation .................................................................... 16
Data Upload ................................................................................................ 16
Software Installation .................................................................................... 16
Battery Replacement ................................................................................... 17
Applications ................................................................................................. 17
Accessories and Replacement Parts .......................................................... 17
Product Care ............................................................................................... 17
Product Warranty ........................................................................................ 18
Product Disposal and Recycling ................................................................. 18
Product Support .......................................................................................... 18
Introduction
Thank you for purchasing your REED R7800 Coating Thickness Gauge. Please read the following instructions carefully before using your instrument. By following the steps outlined in this manual your meter will provide years of reliable service. This coating thickness gauge is designed for non-invasive coating thickness measurements. The R7800 features zero or two point calibration, user adjustable high/low alarms and a USB interface to further analyze results in the software.
Product Quality
This product has been manufactured in an ISO 9001 facility and has been calibrated during the manufacturing process to meet stated product specications. If a certicate of calibration is required please contact the nearest authorized REED distributor or authorized Service Center. Please note an additional fee for this service will apply.
www.GlobalTestSupply.com
Safety
Never attempt to repair or modify your instrument. Dismantling your product, other than for the purpose of replacing batteries, may cause damage that will not be covered under the manufacturer's warranty. Servicing should only be provided by an authorized service center.
Features
• Automatically detects ferrous and non-ferrous substrates
• Easy-to-read color LCD display
• 360° screen rotation allows users to view measured readings from any angle
• Durable gauge and probe construction
• Zero and two-point calibration ensures measurement accuracy
• Single or multi-point mode allows a user to set tolerances and quickly identify whether a coating passes or fails
• Tri-color LEDs quickly indicate when measurements are below (red), above (yellow) or within set parameters (green)
• Internal memory stores up to 500 groups of data
View stored data, charts or real-time measurement with software
• Low battery indication and auto shut off
Included
• Ferrous Substrate
• Non-Ferrous Substrate
• 5 Reference Films
• USB Cable
• Wrist Strap
• Sensor Cover
• Batteries
• Hard Carrying Case
www.GlobalTestSupply.com
Specifications
Measuring Range: 0 to 1250µm
Accuracy: 0 to 1250µm: ±3% +1µm
Resolution: 0 to 99.9µm: 0.1µm
Measurement Modes: Single and Continuous
Display: 4-digit color LCD display
Resolution: 320 x 240 pixels
Backlit Display: Yes (5 levels)
Min/Max/Average Readings: Yes
Minimum Curvature Radius: 5mm (196.85mils)
Minimum Area Diameter: 20mm (787.40mils)
Minimum Substrate Thickness: 0.5mm (19.69mils)
Auto Shut-Off: Yes (After 5 minutes)
Low Battery Indicator: Yes
Power Supply: 2 x AA Batteries
Internal Memory: Yes (Up to 500 groups)
PC Connectivity: USB Cable
Software: Yes
Software OS Compatibility: Windows 7/8/10/11
Product Certications: CE
Refresh Rate: 0.5 seconds (continuous mode)
Operating Temperature: 32 to 104°F (0 to 40°C)
Storage Temperature: -4 to 140°F (-20 to 60°C)
Operating Humidity Range: ≤80%RH
Dimensions: 5.9 x 2.6 x 1.38" (152 x 65 x 35mm)
Weight: 6.35oz (180g)
0.1 to 49.2mils
0 to 49.21mils: ±3% +0.04mils
100 to 1250µm: 1µm 0 to 4.99mils: 0.01mils
5.0 to 49.2mils: 0.1mils
www.GlobalTestSupply.com
Instrument Description
1. Wrist Strap Anchoring Point
2. LED Status Indicator
3. LCD Display
4. Power Button
5. Menu/Conrm/Calibration Button
6. Increase Value/Up/Lock Screen Button
7. USB Port
8. Cancel/Clear/Return Button
9. Decrease Value/Down/Quick Decision Mode Button
10. Battery Compartment (Back)
11. Sensor
Display Description
1. Low Limit Indicator
2. Measured Value
3. Minimum Value Indicator
4. Stored Measurements Indicator
5. Average Measurement Value
6. Maximum Value Indicator
7. Battery Level Indicator
8. Locked Screen Indicator
9. Substrate Type
10. Unit of Measure
11. Calibration Mode Indicator
12. High Limit Indicator
1
2
3
4
5
6
4
Num:
3
2
1
99.4 100
100
Lo: 90.0
Coating Thickness Gauge
OK
MENU
99.84
Avg: Max:Min:
101Hi:
R7800
CLEAR
6
NFe/Fe
um/mil
7
8
9
10
11
75
8
9
10
11
12
www.GlobalTestSupply.com
Operating Instructions
Power ON/OFF
To turn the meter ON or OFF press and hold the power button for 2 seconds. Note: If the buzzer is enabled, there will be an audible beep when the
instrument is powered ON or OFF.
Measuring Procedure
Before each test session perform a Zero Calibration. See the Zero Calibration section for details.
1. During start-up the meter begins an initialization process and enters normal operation upon completion.
2. Select the closest reference lm to the estimated coating thickness of the test area and perform a 2-point calibration on the appropriate substrate plate.
3. Proceed with the 2-point calibration. See the Two-Point Calibration section for details.
4.
Once the 2-point calibration has been completed, proceed with testing.
5. Select 3-5 measuring points on the test surface.
6. For each measuring point, it is recommended to take 5 measurements and record the average value as indicated on the LCD display.
7. After each measuring point has been tested, the overall average value should be noted as the coating thickness value of the test piece.
Note: Lightly press the sensor against the substrate being measured. Keep the sensor in close contact with the surface of the substrate without the use of excessive force as it can cause measurement errors. If the measured value is greater than 1250μm and less than 1500μm, the LCD displays "OL" indicating it is over the measurement range. If the measured value is greater than 1500μm, the gauge will not respond.
www.GlobalTestSupply.com
Setup Mode
CLEAR
OK
1. Press the
um/mil
MENU
MENU
button to enter Setup Mode.
2. Use the
and
arrows to scroll through the following parameters.
Enabling/Disabling Audible Beep
Setting the Brightness
um/mil
Selecting the thickness Unit of Measure
Selecting Calibration Mode
Setting Higher Limit Value
Setting Lower Limit Value
Turning LED Indicators ON/OFF
Factory Reset
Selecting Continuous Measurement Mode
Delete Recorded Data
Follow the instructions below to adjust each setting
3.
www.GlobalTestSupply.com
.
continued...
Enabling/Disabling Audible Beep
1. Press the OK button when the audible beep icon Setup Mode.
and
2. Use the
arrows to select between ON
3. Press the OK button to conrm selection.
4. Press the
CLEAR
button to exit Setup Mode and resume normal operation.
is selected in
or OFF
X.
Setting the LCD Brightness
1. Press the OK button when the brightness icon Setup Mode.
2.
Use the
and
arrows to increase or decrease the brightness level.
3. Press the OK button to conrm selection.
4. Press the
CLEAR
button to exit Setup Mode and resume normal operation.
is selected in
Selecting the Thickness Unit of Measure
1. Press the OK button when the unit of measure icon um/mil is selected in Setup Mode.
2. Use the
and
arrows to select between μm and mil.
3. Press the OK button to conrm selection.
4. Press the
CLEAR
button to exit Setup Mode and resume normal operation.
Selecting Calibration Mode
1. Press the OK button when the calibration mode icon in Setup Mode.
2. Use the
and
arrows to select between zero and
2-point calibration.
3. Press the OK button to conrm selection.
4. Press the
CLEAR
button to exit Setup mode and resume normal operation.
is selected
www.GlobalTestSupply.com
continued...
Setting Higher Limit Value
1. Press the OK button when the higher limit value icon in Setup Mode.
2. Use the
and
arrows to increase or decrease the higher
limit value.
3. Press the OK button to conrm selection.
4. Press the
CLEAR
button to exit Setup Mode and resume normal
operation.
Note: If the higher limit value is exceeded while in normal measuring mode and LED indicators are enabled (See Enabling/Disabling LED Indicators for details), the LED warning light will ash yellow. When measurements are within set parameters, the LED warning light ashes green.
is selected
Setting Lower Limit Value
1. Press the OK button when the lower limit value icon Setup Mode.
2. Use the
and
arrows to increase or decrease the lower
limit value.
3. Press the OK button to conrm selection.
4.
Press the
CLEAR
button to exit Setup Mode and resume normal operation.
Note: If the lower limit value is exceeded while in normal measuring mode and LED indicators are enabled (See Enabling/Disabling LED Indicators for details), the LED will ash red. When measurements are within set parameters, the LED warning light ashes green.
is selected in
www.GlobalTestSupply.com
continued...
10
Enabling/Disabling LED Indicators
1. Press the OK button when the LED alarm icon Setup Mode.
2. Use the
and
arrows to turn the LED alarm ON or OFF.
3. Press the OK button to conrm selection.
4.
Press the
CLEAR
button to exit Setup Mode and resume normal operation.
is selected in
Factory Reset
1. Press the OK button when the factory reset icon Setup Mode.
2. Use the
and
arrows to turn the factory reset ON or OFF.
3. Press the OK button to conrm selection.
4.
Press the
CLEAR
button to exit Setup Mode and resume normal operation
is selected in
Selecting Continuous Measurement Mode
1. Press the OK button when the continuous measurement mode icon
is selected in Setup Mode.
2. Use the mode ON or OFF.
3. Press the OK button to conrm selection.
4.
Press the
Note: When Continuous Measurement is turned on, the gauge will continuously measure until it is powered off.
and
arrows to turn the continuous measurement
CLEAR
button to exit Setup Mode and resume normal operation.
.
www.GlobalTestSupply.com
continued...
11
Delete Recorded Data
CLEAR
1. Press the OK button when the delete recorded data icon selected in Setup Mode.
2. Use the
and
arrows to turn the delete recorded data mode
ON or OFF.
3. Press the OK button to conrm selection.
4. Press the
CLEAR
button to exit Setup Mode and resume normal operation.
is
Maximum, Minimum, and Average Values
Minimum, maximum and average values are displayed during the measuring process. To clear and reset values to zero, press and hold the
CLEAR
button for 2 seconds.
Quick Decision Mode
The Quick Decision Mode is intended for quick measurement taking and analyzing. While in normal operation, press and hold the enter Quick Decision Mode. Use the
single-point/multi point setup. Press the
and
arrows to select between
OK
MENU
button to enter the selected
mode and follow the steps below to congure.
arrow to
continued...
12
www.GlobalTestSupply.com
Single-Point Decision Configuration
CLEAR
CLEAR
CLEAR
CLEAR
1. Press the press the
2. Press the
3.
OK
MENU
the Use the gauge to measure the coating thickness of the measured object.
4. The screen displays the measured value and whether or not the thickness passes or fails ("PASS" or "FAIL").
5. Press the screen or hold the
and
OK
MENU
button to conrm selection.
arrows to set the target thickness value, and
and
arrows to set the tolerance value, and press
button to conrm.
CLEAR
button to return to the single-point analyzing setup
CLEAR
button to exit and resume normal operation.
Multi-Point Decision Configuration
1. Press the press the
2. Press the the
OK
MENU
3. Use the gauge to measure the coating thickness of the measured object. After measuring the same position 3 times, the gauge will automatically display the average value of all 3 measurements under point A.
4. Change measurement position. After measuring the same position 3 times, the gauge will automatically display the average value of all 3 measurements under point B.
5. Proceed by measuring positions C, D, and E the same way as positions A & B.
6. After the measurements are complete, the screen displays the average value of all 5 points and whether or not the thickness passes or fails ("PASS" or "FAIL").
7. Press the screen or press and hold the normal operation.
and
OK
MENU
button to conrm selection.
arrows to set the target thickness value, and
and
arrows to set the tolerance value, and press
button to conrm.
CLEAR
button to return to the multi-point-point analyzing setup
CLEAR
button to exit and resume
13
www.GlobalTestSupply.com
Zero Calibration
CLEAR
um
Fe
Calibration completed!
0.0
um/mil
MENU
um/mil
OK
While in normal operation press and hold the selected calibration mode from the Setup Mode. See Setting Calibration Mode for details.
Calibration Mode Icon Description
MENU
button to enter the
Zero Calibration
Simply place the sensor on an uncoated metal substrate for zero calibration.
Place a standard coating thickness lm
Two-Point Calibration
with a known thickness on an uncoated substrate for calibration to obtain a more accurate measurement result for testing.
Zero Calibration
1. When Figure 1 appears on screen, place the instrument vertically on the uncoated substrate.
2. Lift up after approximately 2 seconds. Zero will be displayed (as shown in Figure 2).
3. When completed, the gauge will automatically exit the zero calibration mode and resume normal operation.
Note: At any time, you can press the and resume normal operation.
R7800
Coating Thickness Gauge
ZERO
OK
MENU
Figure 1 Figure 2
www.GlobalTestSupply.com
CLEAR
button to cancel the calibration
0.0
Calibration completed!
Fe um
continued...
14
Two-Point Calibration
CLEAR
OK
MENU
ZERO
R7800
Coating Thickness Gauge
1/2
um
Fe
509
1/2
um
Fe
500
um
Fe
Calibration completed!
0.0
MENU
um/mil
MENU
um/mil
1/2
um
Fe
500
um
Fe
Calibration completed!
0.0
MENU
um/mil
OK
MENU
ZERO
R7800
Coating Thickness Gauge
1/2
OK
MENU
ZERO
R7800
Coating Thickness Gauge
1/2
um
Fe
509
1/2
um
Fe
500
um
Fe
Calibration completed!
0.0
2/2
um
Fe
Calibration completed!
0.0
MENU
um/mil
1/2
um
Fe
509
1/2
um
Fe
500
um
Fe
Calibration completed!
0.0
MENU
um/mil
1. When Figure 3 appears on the screen, place a standard coating thickness lm with known thickness (e.g.: 500μm) on top of an uncoated substrate.
2. Place the gauge vertically on both the thickness lm and uncoated substrate.
3. Lift the instrument after approximately 2 seconds and the measured value will be displayed (as shown in Figure 4).
4. Press the of the standard coating thickness lm (as shown in Figure 5).
5. Press the
6. Figure 6 will now appear on the screen, prompting users to place the instrument vertically on the uncoated substrate again.
7. Lift the instrument after 2 seconds. Zero will be displayed (as shown in Figure 7).
8. When completed, the gauge will automatically exit the two-point calibration mode and resume normal operation.
Note: At any time, you can press the and resume normal operation.
and
arrows to adjust the value to the actual thickness
OK
MENU
button to conrm adjustment.
1/2
R7800
Coating Thickness Gauge
ZERO
OK
MENU
509
CLEAR
button to cancel the calibration
1/2
Fe um
1/2
500
Fe um
Figure 3
2/2
Coating Thickness Gauge
OK
MENU
Figure 4
R7800
ZERO
Figure 6 Figure 7
www.GlobalTestSupply.com
2/2
0.0
Calibration completed!
Figure 5
Fe um
continued...
15
Calibration Verification
While in normal operation, measure the standard coating thickness lm. The measured value of the instrument should be within the indicated accuracy on the coating thickness lm. If it is out of tolerance, recalibration is required.
Note: If the calibration result is inaccurate try restoring the factory settings and recalibrating.
Locking Screen Auto-Rotation
The instrument has a built-in gravity sensor, which automatically rotates the display during testing. This feature allows users to read the measure­ment values in any direction. To lock the screen auto-rotation, hold the
arrow while in normal operation to turn auto-rotation function ON/OFF.
When disabled, the lock screen symbol
appears on the screen.
Data Upload
Connect the R7800 via the included cable to a USB port on your PC, to download the recorded data and generate reports or view the data in real-time.
Note: The USB interface cannot power the gauge or charge the batteries.
Software Installation
Visit www.reedinstruments.com/software to
download the R7800 software.
Full specications and Operating System compatibility can be found on the product page at www.reedinstruments.com.
If you have specic questions related to your application and/or questions related to software setup and functionality please contact the nearest authorized distributor or Customer Service at info@reedinstruments.com or 1-877-849-2127.
www.GlobalTestSupply.com
16
Battery Replacement
When the " " icon appears on the LCD, the batteries should be replaced.
1.
Remove the screw with a Philips screwdriver and open the battery cover.
2. Replace the 2 x "AA" batteries.
3. Secure the battery cover back and tighten the screw.
Applications
• Verication of paint thickness
• Automotive industry
• Plating thickness
• Industrial manufacturing
• Metal processing
• Aerospace
• Scientic research
Accessories and Replacement Parts
CA-52A Soft Carrying Case
R8888 Deluxe Hard Carrying Case
R9050 Coating Thickness Calibration Kit
Don't see your part listed here? For a complete list of all accessories and replacement parts visit your product page on www.reedinstruments.com.
Product Care
To keep your instrument in good working order we recommend the following:
• Store your product in a clean, dry place.
• Change the battery as needed.
• If your instrument isn't being used for a period of one month or longer
please remove the battery.
• Clean your product and accessories with biodegradable cleaner. Do not
spray the cleaner directly on the instrument. Use on external parts only.
www.GlobalTestSupply.com
17
Product Warranty
REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one (1) year from date of shipment. During the warranty period, REED Instruments will repair or replace, at no charge, products or parts of a product that proves to be defective because of improper material or workmanship, under normal use and maintenance. REED Instruments total liability is limited to repair or replacement of the product. REED Instruments shall not be liable for damages to goods, property, or persons due to improper use or through attempts to utilize the instrument under conditions which exceed the designed capabilities. In order to begin the warranty service process, please contact us by phone at 1-877-849-2127 or by email at info@reedinstruments.com to discuss the claim and determine the appropriate steps to process the warranty.
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling your instrument. Your product contains electronic components and must be disposed of separately from standard waste products.
Product Support
If you have any questions on your product, please contact your authorized REED distributor or REED Instruments Customer Service by phone at 1-877-849-2127 or by email at info@reedinstruments.com.
Please visit www.REEDINSTRUMENTS.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice. All rights reserved. Any unauthorized copying or reproduction of this manual is strictly prohibited without prior written permission from REED Instruments.
www.GlobalTestSupply.com
18
TEST & MEASURE
WITH CONFIDENCE
CHECK OUT OUR LATEST PRODUCTS!
www.GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...