Replacement Cone Tip (CONE)
Reective Tape (RT100)
Soft Carrying Case (CA-05A)
Soft Carrying Case (C-820)
3
www.GlobalTestSupply.com
INSTRUMENTS
Instrument Description
1.Laser Beam
2.RPM Adapter
3.Display
4.Measure Button
5.Memory Call Button
6.Function Switch
1
43
5
6
2
Operating Procedures
Photo laser tachometer measuring procedure
1.Move the Function Switch to the Photo RPM position.
2.Apply reective tape to the object being measured. Press the
Measure Button and align the Laser Beam with the applied
target. Verify that the Monitor Indicator provides a reading when
the target passes through the eld of view. Release the MeasureButton when the reading stabilizes.
Note: If the measured RPM value is very low ( ie: 50 RPM ), we recommend
attaching more Reflective tape.
Contact tachometer measuring procedure
RPM measurement
1.Move the Function Switch to the Contact RPM position.
2.Press the Measure Button and lightly pressing the RPM Adapter
against the rotating axis. Release the Measure Button when the
reading stabilizes.
4
www.GlobalTestSupply.com
INSTRUMENTS
Surface Speed Measurement
1.Move the Function Switch to the m/min. or ft/min. position.
2.Press the Measure Button and attach the surface speed test
wheel to the detector. Release the Measure Button when the
reading stabilizes.
Operation procedure for memory recall
1.Last value, Max. value and min. value are automatically stored
in memory.
2.To recall measurements saved in memory:
a. Push the Memory Call Button - To display the last
value ( LA and the last value will be displayed alternately).
b. Push the Memory Call Button again - To display
the maximum value (UP and the max. value will be
displayed alternately).
c. Push the Memory Call Button again - To display the
minimum value ( dn and the min. value will be displayed
alternately).
Max Value
Display
reading
Min Value
Measuring Time
www.GlobalTestSupply.com
Last Value
5
INSTRUMENTS
Battery Replacement
• When the LCD displays LO, it is necessary to replace the batteries,
however in-spec measurements may still be made for several hours
after the low battery indicator appears.
• Open the Battery Cover, using a at screwdriver or small coin.
• Replace batteries correctly and place the cover back on.
For service on this or any other REED product or information on other
REED products, contact REED Instruments at info@reedinstruments.com
Notes _________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
www.GlobalTestSupply.com
6
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.