Introduction
Thank you for purchasing your REED R6050SD Data Logging
Thermo-Hygrometer. Please read the following instructions carefully
before using your instrument. By following the steps outlined in this
manual your meter will provide years of reliable service.
Product Quality
This product has been manufactured in an ISO9001 facility and has been
calibrated during the manufacturing process to meet the stated product
specications. If a certicate of calibration is required please contact the
nearest authorized REED distributor or authorized Service Center. Please
note an additional fee for this service will apply.
www.GlobalTestSupply.com
3
Safety
Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries, may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center.
Features
• Measures relative humidity, ambient temperature and dew point/wet
bulb temperatures
• Contact temperature measurement with a type J/K thermocouple option
• Temperature is measured in °C or °F
• Real time data logger with integrated SD memory card
• User selectable sampling rate from 1 to 3600 seconds
• Easy-to-read backlit LCD display
• Data hold and Min/Max functions
• Tripod mount for long-term monitoring
• Low battery indicator and auto shut off
Included
• Humidity/Temperature Probe
• Soft Carrying Case
• Batteries
www.GlobalTestSupply.com
4
Specifications
Temperature
Measuring Range: 32 to 122°F (0 to 50°C)
Accuracy: ±1.5°F (0.8°C)
Resolution: 0.1°F/°C
Contact Temperature (Probe sold separately)
Measuring Range: Type K: -148 to 2372°F
(-100 to 1300°C)
Type J: -148 to 2192°F
(-100 to 1200°C)
600, 1800, 3600 seconds)
External Memory: Yes, expandable up to 16GB
with SD card (optional)
Auto Shut-off: Yes (after 10 minutes/off)
Kick Stand: Yes
Tripod Mountable: Yes
Low Battery Indicator: Yes
Power Supply: 6 x AA batteries or AC Adapter
(optional)
Data Output: Yes (RS-232)
Battery Life: Sampling Time Dependent
Product Certications: CE
Operating Temperature: 32 to 122°F (0 to 50°C)
Storage Temperature: 14 to 140°F (-10 to 60°C)
Operating Humidity Range: 10 to 80%
Dimensions: 7 x 2.7 x 1.9" (177 x 68 x 45mm)
Weight: 1.13lbs (515g)
2. Date & Time Values/Temperature Measurement Reading
3. Date/Time Stamp
4. Temperature Unit of Measure
5. Relative Humidity Unit of Measure
6. Maximum and Minimum Indicator
7. Record Mode Indicator
8. Data Hold Indicator
9. Low Battery Indicator
www.GlobalTestSupply.com
8
Operating Instructions
Power ON/OFF
1. Turn the meter on by pressing the POWER button. To turn the meter
off, press and hold the POWER button for 2 seconds.
Note: This meter can be powered by either 6 x "AA" batteries or an AC
adapter (sold separately).
Selecting Measurement Modes
The R6050SD offers 4 types of measurement modes:
• Humidity/Temperature
• Dew Point
• Wet Bulb
• Type K/J Thermometer
1. When the meter is on, press and hold the FUNCTION button to scroll
through the measurement modes as indicated by "rH" (for Humidity/
Temperature measurement), "dp" (for Dew Point measurement),"_Et"
(for Wet Bulb measurement) and "tP" (for Type K/J Thermometer
measurement.)
2. Release the FUNCTION button to conrm selection.
Humidity/Temperature Measurement Mode
1. While the meter is off, plug the Probe into the Probe Input Socket,
then power on the meter by pressing the POWER button.
2.
Select the "rH" function (see Selecting Measurement Modes for details).
3. Hold the Probe Handle and face the Probe Head against the
measuring source. The top part of the display will indicate the
humidity value, and the lower part of the display will indicate the
temperature value as shown below.
%RH
www.GlobalTestSupply.com
continued...
9
Dew Point Measurement Mode
1. Select the "dP" function (see Selecting Measurement Modes for details).
2. Hold the Probe Handle and face the Probe Head against the
measuring source. The top part of the display will indicate the dew
point value, and the lower part of the display will indicate "dP" as
shown below.
Wet Bulb Measurement Mode
1.
Select the "_Et" function (see Selecting Measurement Modes for details).
2. Hold the Probe Handle and face the Probe Head against the measuring
source. The top part of the display will indicate the wet bulb value,
and the lower part of the display will indicate "_Et" as shown below.
www.GlobalTestSupply.com
continued...
10
Type K/J Thermocouple Measurement Mode
1. Select the "tP" function (see Selecting Measurement Modes for details).
2. Plug the optional Type K or J Thermocouple Probe into the Type K/J
thermocouple input jack.
3. The screen will display a "K", conrming that is congured by default
to a Type K thermocouple probe. The measured temperature will now
be displayed on the display screen as shown below.
Note: If using a type J thermocouple probe, see Selecting Thermocouple
Type for details.
K
Data Hold
1. While taking a measurement, press the HOLD button to freeze the
current measurement(s) on the display.
2. While in this mode a "HOLD" symbol will appear on the display.
3. Press the HOLD button again to resume normal operation.
Note: When the Data Hold feature is active all buttons except the POWER
button are disabled.
Recording Maximum and Minimum Readings
1. Press the REC button to enter recording mode as indicated by
"REC" on the LCD. The meter will now begin recording maximum
and minimum readings.
2. While in recording mode;
A) Press the REC button once and the maximum value will appear
on the display as indicated by "REC MAX".
www.GlobalTestSupply.com
continued...
11
B) Press the REC button again and the minimum value will appear
on the display as indicated by "REC MIN".
C) To exit recording mode and resume normal operation, press and
hold the REC button for two seconds.
When in recording mode the POWER button is disabled
and the meter cannot be turned off.
Backlight
After powering the meter ON, the LCD Backlight will turn on automatically.
Press the BACKLIGHT button to turn the LCD Backlight on or off.
Setup Mode
1. Press and hold the SET button for 2 seconds to enter Setup Mode.
2. Press the NEXT button continuously to scroll through the following
parameters.
ParameterDescription
dAtESet the time and date
dECSet the decimal format (USA (20.00) or European (20,00))
PoFFEnable or disable the auto-power off function
bEEPTurn the beeper on or off
tYPESelect thermocouple type (K or J)
t-CFSelect the temperature unit of measure (°C or °F)
SP-tSet the data logging sampling rate
Sd FFormat the SD memory card
3. Once the appropriate parameter has been selected, follow the
instructions below.
Note: The meter will automatically exit out of the Setup mode if no key is
pressed within 7 seconds.
www.GlobalTestSupply.com
continued...
12
Setting the Time and Date (dAtE)
1. Press the ENTER button when "dAtE" appears on the LCD.
2. Press the and buttons to adjust the year as indicated by "YY".
3. Press the ENTER button to conrm selection.
4. Repeat steps 2 and 3 for month, day, hour, minute and seconds as
indicated by (mm/dd/h/m/s).
5. After each value has been selected and conrmed, the meter will
automatically skip to the next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation. The internal clock will keep accurate time
when the meter is powered off. When new batteries are installed the clock
will have to be reset.
Setting Data Decimal Format (dEC)
Numeric formats vary in different countries. By default the meter is set to
bASIC mode where a decimal point is used to separate units, (i.e. 20.00).
The European format uses a comma (i.e. 20,00) to separate units. To change
this setting, follow steps 1 and 2 when "dEC" appears on the LCD.
1. Press the
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation
and buttons to select between bASIC and Euro.
Enabling/Disabling Auto Power Off (PoFF)
Follow steps 1 and 2 when "PoFF" appears on the LCD.
1. Press the
(disabled). With the Auto Power OFF feature enabled, the meter will
automatically switch OFF after 10 minutes of inactivity to preserve
battery life.
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
and buttons to select between YES (enabled) or NO
continued...
13
www.GlobalTestSupply.com
Enabling/Disabling the Beeper (bEEP)
Follow steps 1 and 2 when "bEEP" appears on the LCD.
1. Press the
OFF (disabled).
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
and buttons to select between ON (enabled) or
Selecting Thermocouple Type (tYPE)
Follow steps 1 and 2 when "tYPE" appears on the LCD.
1. Press the
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
and buttons to select between type K and J.
Selecting the Temperature Unit of Measure (t-CF)
Follow steps 1 and 2 when "t-CF" appears on the LCD.
1. Press the
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
and buttons to select between °C and °F.
Setting the Data Logging Sampling Rate (SP-t)
Follow steps 1 and 2 when "SP-t" appears on the LCD.
1. Press the
2, 5, 10, 30, 60, 120, 300, 600, 1800, and 3600 seconds.
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
and buttons to adjust the sampling rate between 0, 1,
www.GlobalTestSupply.com
continued...
14
Formatting the SD Card (Sd F)
Follow steps 1 through 5 when "Sd F" appears on the LCD.
1. Press the
"NO" to abort.
2. Press the ENTER button to conrm selection.
3. Press the ENTER button again to re-conrm.
4. The meter will format the SD card and automatically return to the
Setup menu when formatting is complete.
5. Press the ESC button to exit the Setup mode and resume
normal operation.
Note: It is recommended that new SD cards be formatted prior to rst
use. If the formatting process fails, the meter will display "-E-".
and buttons to select "YES" to format the card. Select
Verify Set Time and Date
During normal operation press the TIME CHECK button to display the
year, month, day, hour and minutes.
Verify Sampling Rate
During normal operation, press the SAMPLING CHECK button to view the
selected sampling rate.
Data Logging
Data Recording Modes
Manual Data Logging: Press the LOG button to manually log up to 99
readings on a SD memory card (see Manual Data Logging Mode for details).
Automatic Data Logging: Setup the meter parameters in order to
automatically log data on a SD memory card. The number of data points
is limited by the size of the memory card.
Note: It is recommended to use a class 4 SDHC memory card between 1
and 16Gb. Insert the SD memory card in the slot at the bottom of the meter.
The card must be inserted with the label side facing the rear of the meter.
www.GlobalTestSupply.com
continued...
15
Manual Data Logging Mode
In manual mode, the LOG button is pressed to manually log a reading on
the inserted SD card at any time. In order to setup the meter for manual
data logging, follow steps 1 through 8 below.
1. Set the data logging sampling rate to "0" seconds (see Setting the Data Logging Sampling Rate section for details).
2. Press the REC button and "REC" will appear on the LCD.
3. The meter will also display P-X (X = memory position number
between 1 and 99).
4. Press the SET button to enter the data memory positions selection.
5. Press the and buttons to select one of the 99 data memory
positions in which to record.
6. Press the ENTER button to conrm selection.
7. Press the LOG button again to save a reading to memory. "REC" will
ash each time a data point is stored.
8. To exit manual data logging mode, press and hold the REC button
2 seconds to resume normal operation.
Automatic Data Logging Mode
In automatic mode the desired data logging sampling rate is set to 1, 2,
5, 10, 30, 60, 120, 300, 600, 1800 or 3600 seconds prior to recording (see
Setting the Data Logging Sampling Rate section for details). In order to
setup the meter for automatic data logging, follow steps 1 through 5 below.
1. To begin a data logging session Press the REC button and "REC"
will appear on the LCD.
2. Press the LOG button. The meter will scan for a SD memory card.
If a card is not inserted or the card is defective, the meter will ash
"CArD -E-" indenitely. In this case, power the meter OFF and try
again with another SD memory card.
3. The "REC" icon will appear on the LCD and will continuously ash
while in a data logging session.
4. To pause the data logging session press the LOG button and "REC"
will stop ashing. To resume press the LOG button again.
5. To end the current data logging session pause the data logging
session and hold the REC button until "REC" disappears.
continued...
16
www.GlobalTestSupply.com
Transferring Data from the SD Memory Card to a Computer
70
1 2 3 4 5 6 7 8 9 10 11 12
1. While the meter is powered Off remove the SD memory card from the
SD card slot.
2. Insert the SD memory card into the computer.
3. Open the le(s) with Excel. See sample below:
Place DateTimeValue UnitValue Unit
119-07-10 1:13:10 18.4%RH22.6Degree C
219-07-10 1:13:12 30%RH25Degree C
319-07-10 1:13:14 25%RH35Degree C
419-07-10 1:13:16 45%RH30Degree C
519-07-10 1:13:18 30%RH35Degree C
619-07-10 1:13:20 45%RH25Degree C
719-07-10 1:13:22 50%RH35Degree C
819-07-10 1:13:24 60%RH30Degree C
919-07-10 1:13:26 45%RH28Degree C
1019-07-10 1:13:28 30%RH22.6Degree C
1119-07-10 1:13:30 17.5%RH22.6Degree C
1219-07-10 1:13:32 17.5%RH22.6Degree C
The raw data can be used to create a graph in Excel. See sample below:
In order to calibrate the meter humidity calibration standards are required
(REED R9933, R9975 or R9980). If the meter has been exposed to
extreme conditions or it has not been used in a long period of time the
meter should be calibrated by following steps 1 through 5 below.
1. While in normal %RH measurement mode, hold the SET and LOG
buttons simultaneously for approx. 2 seconds to enter the calibration
mode of %RH.
2. The LCD display will show the same relative humidity measuring
value along with the %RH symbol.
3. Press the and buttons to adjust the upper display until it
matches the %RH Humidity Calibration Standard.
4. Once the calibration is complete, press "REC" button to save the
calibration value into the memory which will ash conrming it has
been saved.
5. Press the NEXT button twice to exit the calibration mode and
resume normal operation.
Note: At any time, you can press the ESC button to exit the calibration
mode and resume normal operation.
Data Stream Output
Using Terminal application, a data stream can be viewed from the RS-232
output, connect RS-232/USB cable between the product and terminal/PC
and use the following setting to view the data stream.
Terminal Settings:
• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
www.GlobalTestSupply.com
continued...
18
41010100000501 = Upper Value, +50.1% RH
Reading
4
RS-232Mono Jack
Pin 2 (RX)
Pin 4 (DTX)
Sleeve
Tip
Decimal Place 0-3
Positive (+) = 0,
Negative (-) = 1
°C = 01, °F = 02, %RH = 04
1 = Upper display value,
2 = Lower display value
3.5mm Terminal to RS-232 Serial Connection
System Reset
If the meter becomes unresponsive or if the display freezes the Reset pin
can be used to reset the instrument.
1. Use a paper clip or any similar small object to press the Reset pin.
2. After pressing the Reset pin, power the meter back up by holding the
POWER button for 2 seconds.
3. If resetting the meter does not resolve the issue, please return the
meter to the nearest authorized REED distributor or authorized
Service Center for repair.
www.GlobalTestSupply.com
19
Battery Replacement
When the low battery icon appears on the LCD, the batteries must
be replaced.
1. Remove the two (2) Phillips screws on the back of the meter directly
located above the top of the kickstand.
2. Remove the battery cover.
3. Replace 6 x "AA" batteries.
4. Secure the battery cover with the two (2) Phillips screws.
R1500 Tripod
RSD-ADP-NA Power Supply, 110V
RSD-ADP-EU Power Supply, 220V
CA-05A Soft Carrying Case
R8888 Deluxe Hard Carrying Case
SD-4GB 4GB Class 4 SDHC Memory Card
RSD-16GB 16GB Micro SD Memory Card w/ Adapter
Don't see your part listed here? For a complete list of all accessories and
replacement parts visit your product page on www.REEDInstruments.com.
Frequently Asked Questions (FAQ's)
I cannot turn off my meter, do you know why?
Often times, the meter cannot be turned off because it is in recording
mode ("REC" mode). Make sure to exit that function by holding down the
REC button before attempting to turn the meter off.
How long can I record for?
Battery life will depend on a number of factors:
• Quality of Battery
• Whether the beeper is ON or OFF
• Whether the backlight is ON or OFF
• Environmental Conditions during data logging
As a reference, using alkaline batteries with both the backlight and beeper
on while automatic data logging at a 2 second sampling rate; the batteries
will last approximately 14 hours.
Is there any way to make the R6050SD record temp, humidity, dew
point, wet bulb, and Type K readings all at the same time?
The R6050SD is unable to measure all the parameters simultaneously.
www.GlobalTestSupply.com
continued...
21
If the screen displays the "_ET" function, will it still record humidity/
temperature? How do I select the "rH" function to display on the
screen. I want to record the humidity/temperature for an extended
period of time.
The R6050SD offers 4 types of measurement modes:
1. Humidity/ Temp
2. Dew Point
3. Wet Bulb
4. Type K/J Thermometer
When the meter is on, press and hold the FUNCTION button to scroll
through the measuring modes. The display will ash "rH", for Humidity/
Temperature measurement, "dP" for Dew Point measurement, "_Et" for
Wet Bulb measurement, and "tP" for Type K/J Thermometer measurement.
Release the FUNCTION button to conrm selection. Select the "rH''
option for extended recording of temperature and humidity.
Product Care
To keep your instrument in good working order we recommend the following:
• Store your product in a clean, dry place.
• Change the battery as needed.
• If your instrument isn't being used for a period of one month or longer
please remove the battery.
• Clean your product and accessories with biodegradable cleaner. Do not
spray the cleaner directly on the instrument. Use on external parts only.
www.GlobalTestSupply.com
22
Product Warranty
REED Instruments guarantees this instrument to be free of defects in
material or workmanship for a period of one (1) year from date of shipment.
During the warranty period, REED Instruments will repair or replace, at no
charge, products or parts of a product that proves to be defective because
of improper material or workmanship, under normal use and maintenance.
REED Instruments total liability is limited to repair or replacement of the
product. REED Instruments shall not be liable for damages to goods,
property, or persons due to improper use or through attempts to utilize the
instrument under conditions which exceed the designed capabilities. In
order to begin the warranty service process, please contact us by phone
at 1-877-849-2127 or by email at info@REEDInstruments.com to discuss
the claim and determine the appropriate steps to process the warranty.
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling
your instrument. Your product contains electronic components and
must be disposed of separately from standard waste products.
Product Support
If you have any questions on your product, please contact your authorized
REED distributor or REED Instruments Customer Service by phone at
1-877-849-2127 or by email at info@REEDInstruments.com.
Please visit www.REEDInstruments.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice.
All rights reserved. Any unauthorized copying or reproduction of this
manual is strictly prohibited without prior written permission from
REED Instruments.
23
www.GlobalTestSupply.com
TEST & MEASURE
WITH CONFIDENCE
Access our
Product Catalog
Over 200 portable test and
measurement instruments
www.GlobalTestSupply.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.