REED Instruments R6050SD-KIT, R6050SD-NIST, R6050SD User manual

Data Logging Thermo­Hygrometer
R6050SD
Instruction Manual
www.GlobalTestSupply.com
Table of Contents
Introduction ................................................................................................ 3
Product Quality ........................................................................................... 3
Safety ......................................................................................................... 4
Features ...................................................................................................... 4
Included ...................................................................................................... 4
Specications .......................................................................................... 5-6
Instrument Display ...................................................................................... 7
Display Description .................................................................................... 8
Operating Instructions ........................................................................... 9-12
Power ON/OFF ..................................................................................... 9
Selecting Measurement Modes ............................................................ 9
Humidity/Temperature Measurement Mode ......................................... 9
Dew Point Measurement Mode .......................................................... 10
Wet Bulb Measurement Mode ............................................................ 10
Type K/J Thermocouple Measurement Mode .................................... 11
Data Hold ........................................................................................... 11
Recording Maximum and Minimum Readings ..............................11-12
Backlight ............................................................................................. 12
Setup Mode ......................................................................................... 12-15
Setting the Time and Date (dAtE) ....................................................... 13
Setting Data Decimal Format (dEC) .................................................... 13
Enabling/Disabling Auto Power Off (PoFF) ......................................... 13
Enabling/Disabling the Beeper (bEEP) ............................................... 14
Selecting Thermocouple Type (tYPE) ................................................. 14
Selecting the Temperature Unit of Measure (t-CF) ............................. 14
Setting the Data Logging Sampling Rate (SP-t) ................................. 14
Formatting the SD Card (Sd F) ........................................................... 15
Verify Set Time and Date .................................................................... 15
Verify Sampling Rate .......................................................................... 15
continued...
www.GlobalTestSupply.com
Data Logging ....................................................................................... 15-17
Data Recording Modes....................................................................... 15
Manual Data Logging Mode ............................................................... 16
Automatic Data Logging Mode .......................................................... 16
Transferring Data from the SD Memory Card to a Computer ............. 17
Humidity Calibration Procedure ............................................................... 18
Data Stream Output ............................................................................18-19
3.5mm Terminal to RS-232 Serial Connection ................................... 19
System Reset ........................................................................................... 19
Battery Replacement ................................................................................ 20
Applications .............................................................................................. 20
Accessories and Replacement Parts ..................................................20-21
FAQ's ...................................................................................................21-22
Product Care ............................................................................................ 22
Product Warranty ..................................................................................... 23
Product Disposal and Recycling .............................................................. 23
Product Support ....................................................................................... 23
Introduction
Thank you for purchasing your REED R6050SD Data Logging Thermo-Hygrometer. Please read the following instructions carefully before using your instrument. By following the steps outlined in this manual your meter will provide years of reliable service.
Product Quality
This product has been manufactured in an ISO9001 facility and has been calibrated during the manufacturing process to meet the stated product specications. If a certicate of calibration is required please contact the nearest authorized REED distributor or authorized Service Center. Please note an additional fee for this service will apply.
www.GlobalTestSupply.com
Safety
Never attempt to repair or modify your instrument. Dismantling your product, other than for the purpose of replacing batteries, may cause damage that will not be covered under the manufacturer's warranty. Servicing should only be provided by an authorized service center.
Features
• Measures relative humidity, ambient temperature and dew point/wet bulb temperatures
• Contact temperature measurement with a type J/K thermocouple option
• Temperature is measured in °C or °F
• Real time data logger with integrated SD memory card
• User selectable sampling rate from 1 to 3600 seconds
• Easy-to-read backlit LCD display
• Data hold and Min/Max functions
• Tripod mount for long-term monitoring
• Low battery indicator and auto shut off
Included
• Humidity/Temperature Probe
• Soft Carrying Case
• Batteries
www.GlobalTestSupply.com
Specifications
Temperature
Measuring Range: 32 to 122°F (0 to 50°C) Accuracy: ±1.5°F (0.8°C) Resolution: 0.1°F/°C
Contact Temperature (Probe sold separately)
Measuring Range: Type K: -148 to 2372°F (-100 to 1300°C) Type J: -148 to 2192°F (-100 to 1200°C)
Accuracy: ≥-58°F (-50°C) ±(0.4% + 1°F (0.5°C)) ≤-58°F (-50°C) ±(0.4% + 1.8°F (1°C))
Resolution: 0.1°F/°C
Dew Point
Measuring Range: -13.5 to 120.1°F (-25.5 to 48.9°C) Accuracy: Sum of temp and RH accuracy Resolution: 0.1°F/°C
Wet Bulb
Measuring Range: -6.9 to 122.0°F (-21.6 to 50.0°C) Accuracy: Sum of temp and RH accuracy Resolution: 0.1°F/°C
Humidity
Measuring Range: 5 to 95% RH Accuracy: ±(3% rdg. + 1% RH): ≥70% RH
±3% RH: < 70% RH Resolution: 0.1% RH
www.GlobalTestSupply.com
continued...
General Specifications
Display: Dual Display, LCD Backlit Display: Yes Data Hold: Yes Min: Yes Max: Yes Datalogging Capabilities: Yes Real-Time Clock and Date Stamp: Yes Sampling Rate: Yes (1, 2, 5, 10, 30, 60, 120, 300,
600, 1800, 3600 seconds) External Memory: Yes, expandable up to 16GB
with SD card (optional) Auto Shut-off: Yes (after 10 minutes/off) Kick Stand: Yes Tripod Mountable: Yes Low Battery Indicator: Yes Power Supply: 6 x AA batteries or AC Adapter
(optional) Data Output: Yes (RS-232) Battery Life: Sampling Time Dependent Product Certications: CE Operating Temperature: 32 to 122°F (0 to 50°C) Storage Temperature: 14 to 140°F (-10 to 60°C) Operating Humidity Range: 10 to 80% Dimensions: 7 x 2.7 x 1.9" (177 x 68 x 45mm) Weight: 1.13lbs (515g)
www.GlobalTestSupply.com
Instrument Description
17
Temperature and
Humidity Meter
FUNC.
POWER
2
3
4
HOLD
ESC
NEXT
SET
TIME
CHECK
SD Series Datalogger
LOG
SAMPLING
CHECK
R6050SD
5
REC
ENTER
6
7
8
1. LCD Display
2. POWER/Backlight ESC Button
3. SET/Down/TIME CHECK Button
4. Type K/J Thermocouple Input Jack
5. FUNCTION/HOLD/ NEXT Button
6. REC/ENTER Button
7. LOG/Up/SAMPLING CHECK Button
8. Humidity/Temperature Probe Input Socket
12
13
14
9
10 11
15
16
9. RS-232 Output Jack
10. Reset Pin
11. Power Adapter Input
12. Tripod Mounting Screw
13. Battery Cover
14. Kickstand
15. SD Card Slot
16. Humidity/Temperature Probe Plug
17. Humidity/Temperature Probe Sensing Head
18. Humidity/Temperature Probe Handle
18
www.GlobalTestSupply.com
Display Description
6
F
%RH
3
1. Relative Humidity/Dew Point/Wet Bulb/Type K/J Thermocouple Measurement Reading
2. Date & Time Values/Temperature Measurement Reading
3. Date/Time Stamp
4. Temperature Unit of Measure
5. Relative Humidity Unit of Measure
6. Maximum and Minimum Indicator
7. Record Mode Indicator
8. Data Hold Indicator
9. Low Battery Indicator
www.GlobalTestSupply.com
Operating Instructions
Power ON/OFF
1. Turn the meter on by pressing the POWER button. To turn the meter off, press and hold the POWER button for 2 seconds.
Note: This meter can be powered by either 6 x "AA" batteries or an AC adapter (sold separately).
Selecting Measurement Modes
The R6050SD offers 4 types of measurement modes:
• Humidity/Temperature
• Dew Point
• Wet Bulb
• Type K/J Thermometer
1. When the meter is on, press and hold the FUNCTION button to scroll through the measurement modes as indicated by "rH" (for Humidity/ Temperature measurement), "dp" (for Dew Point measurement),"_Et" (for Wet Bulb measurement) and "tP" (for Type K/J Thermometer measurement.)
2. Release the FUNCTION button to conrm selection.
Humidity/Temperature Measurement Mode
1. While the meter is off, plug the Probe into the Probe Input Socket, then power on the meter by pressing the POWER button.
2.
Select the "rH" function (see Selecting Measurement Modes for details).
3. Hold the Probe Handle and face the Probe Head against the measuring source. The top part of the display will indicate the humidity value, and the lower part of the display will indicate the temperature value as shown below.
%RH
www.GlobalTestSupply.com
continued...
Dew Point Measurement Mode
1. Select the "dP" function (see Selecting Measurement Modes for details).
2. Hold the Probe Handle and face the Probe Head against the measuring source. The top part of the display will indicate the dew point value, and the lower part of the display will indicate "dP" as shown below.
Wet Bulb Measurement Mode
1.
Select the "_Et" function (see Selecting Measurement Modes for details).
2. Hold the Probe Handle and face the Probe Head against the measuring source. The top part of the display will indicate the wet bulb value, and the lower part of the display will indicate "_Et" as shown below.
www.GlobalTestSupply.com
continued...
10
Type K/J Thermocouple Measurement Mode
1. Select the "tP" function (see Selecting Measurement Modes for details).
2. Plug the optional Type K or J Thermocouple Probe into the Type K/J thermocouple input jack.
3. The screen will display a "K", conrming that is congured by default to a Type K thermocouple probe. The measured temperature will now be displayed on the display screen as shown below.
Note: If using a type J thermocouple probe, see Selecting Thermocouple Type for details.
K
Data Hold
1. While taking a measurement, press the HOLD button to freeze the current measurement(s) on the display.
2. While in this mode a "HOLD" symbol will appear on the display.
3. Press the HOLD button again to resume normal operation.
Note: When the Data Hold feature is active all buttons except the POWER button are disabled.
Recording Maximum and Minimum Readings
1. Press the REC button to enter recording mode as indicated by "REC" on the LCD. The meter will now begin recording maximum and minimum readings.
2. While in recording mode; A) Press the REC button once and the maximum value will appear
on the display as indicated by "REC MAX".
www.GlobalTestSupply.com
continued...
11
B) Press the REC button again and the minimum value will appear
on the display as indicated by "REC MIN".
C) To exit recording mode and resume normal operation, press and
hold the REC button for two seconds.
When in recording mode the POWER button is disabled
and the meter cannot be turned off.
Backlight
After powering the meter ON, the LCD Backlight will turn on automatically. Press the BACKLIGHT button to turn the LCD Backlight on or off.
Setup Mode
1. Press and hold the SET button for 2 seconds to enter Setup Mode.
2. Press the NEXT button continuously to scroll through the following parameters.
Parameter Description
dAtE Set the time and date dEC Set the decimal format (USA (20.00) or European (20,00)) PoFF Enable or disable the auto-power off function bEEP Turn the beeper on or off tYPE Select thermocouple type (K or J) t-CF Select the temperature unit of measure (°C or °F) SP-t Set the data logging sampling rate Sd F Format the SD memory card
3. Once the appropriate parameter has been selected, follow the instructions below.
Note: The meter will automatically exit out of the Setup mode if no key is pressed within 7 seconds.
www.GlobalTestSupply.com
continued...
12
Setting the Time and Date (dAtE)
1. Press the ENTER button when "dAtE" appears on the LCD.
2. Press the and buttons to adjust the year as indicated by "YY".
3. Press the ENTER button to conrm selection.
4. Repeat steps 2 and 3 for month, day, hour, minute and seconds as indicated by (mm/dd/h/m/s).
5. After each value has been selected and conrmed, the meter will automatically skip to the next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode and resume normal operation. The internal clock will keep accurate time when the meter is powered off. When new batteries are installed the clock will have to be reset.
Setting Data Decimal Format (dEC)
Numeric formats vary in different countries. By default the meter is set to bASIC mode where a decimal point is used to separate units, (i.e. 20.00). The European format uses a comma (i.e. 20,00) to separate units. To change this setting, follow steps 1 and 2 when "dEC" appears on the LCD.
1. Press the
2. Press the ENTER button to conrm selection and skip to next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode and resume normal operation
and buttons to select between bASIC and Euro.
Enabling/Disabling Auto Power Off (PoFF)
Follow steps 1 and 2 when "PoFF" appears on the LCD.
1. Press the (disabled). With the Auto Power OFF feature enabled, the meter will automatically switch OFF after 10 minutes of inactivity to preserve battery life.
2. Press the ENTER button to conrm selection and skip to next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode and resume normal operation.
and buttons to select between YES (enabled) or NO
continued...
13
www.GlobalTestSupply.com
Enabling/Disabling the Beeper (bEEP)
Follow steps 1 and 2 when "bEEP" appears on the LCD.
1. Press the OFF (disabled).
2. Press the ENTER button to conrm selection and skip to next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode and resume normal operation.
and buttons to select between ON (enabled) or
Selecting Thermocouple Type (tYPE)
Follow steps 1 and 2 when "tYPE" appears on the LCD.
1. Press the
2. Press the ENTER button to conrm selection and skip to next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode and resume normal operation.
and buttons to select between type K and J.
Selecting the Temperature Unit of Measure (t-CF)
Follow steps 1 and 2 when "t-CF" appears on the LCD.
1. Press the
2. Press the ENTER button to conrm selection and skip to next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode and resume normal operation.
and buttons to select between °C and °F.
Setting the Data Logging Sampling Rate (SP-t)
Follow steps 1 and 2 when "SP-t" appears on the LCD.
1. Press the 2, 5, 10, 30, 60, 120, 300, 600, 1800, and 3600 seconds.
2. Press the ENTER button to conrm selection and skip to next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode and resume normal operation.
and buttons to adjust the sampling rate between 0, 1,
www.GlobalTestSupply.com
continued...
14
Formatting the SD Card (Sd F)
Follow steps 1 through 5 when "Sd F" appears on the LCD.
1. Press the "NO" to abort.
2. Press the ENTER button to conrm selection.
3. Press the ENTER button again to re-conrm.
4. The meter will format the SD card and automatically return to the Setup menu when formatting is complete.
5. Press the ESC button to exit the Setup mode and resume normal operation.
Note: It is recommended that new SD cards be formatted prior to rst use. If the formatting process fails, the meter will display "-E-".
and buttons to select "YES" to format the card. Select
Verify Set Time and Date
During normal operation press the TIME CHECK button to display the year, month, day, hour and minutes.
Verify Sampling Rate
During normal operation, press the SAMPLING CHECK button to view the selected sampling rate.
Data Logging
Data Recording Modes
Manual Data Logging: Press the LOG button to manually log up to 99 readings on a SD memory card (see Manual Data Logging Mode for details).
Automatic Data Logging: Setup the meter parameters in order to automatically log data on a SD memory card. The number of data points is limited by the size of the memory card.
Note: It is recommended to use a class 4 SDHC memory card between 1 and 16Gb. Insert the SD memory card in the slot at the bottom of the meter. The card must be inserted with the label side facing the rear of the meter.
www.GlobalTestSupply.com
continued...
15
Manual Data Logging Mode
In manual mode, the LOG button is pressed to manually log a reading on the inserted SD card at any time. In order to setup the meter for manual data logging, follow steps 1 through 8 below.
1. Set the data logging sampling rate to "0" seconds (see Setting the Data Logging Sampling Rate section for details).
2. Press the REC button and "REC" will appear on the LCD.
3. The meter will also display P-X (X = memory position number between 1 and 99).
4. Press the SET button to enter the data memory positions selection.
5. Press the and buttons to select one of the 99 data memory positions in which to record.
6. Press the ENTER button to conrm selection.
7. Press the LOG button again to save a reading to memory. "REC" will ash each time a data point is stored.
8. To exit manual data logging mode, press and hold the REC button 2 seconds to resume normal operation.
Automatic Data Logging Mode
In automatic mode the desired data logging sampling rate is set to 1, 2, 5, 10, 30, 60, 120, 300, 600, 1800 or 3600 seconds prior to recording (see Setting the Data Logging Sampling Rate section for details). In order to setup the meter for automatic data logging, follow steps 1 through 5 below.
1. To begin a data logging session Press the REC button and "REC" will appear on the LCD.
2. Press the LOG button. The meter will scan for a SD memory card. If a card is not inserted or the card is defective, the meter will ash "CArD -E-" indenitely. In this case, power the meter OFF and try again with another SD memory card.
3. The "REC" icon will appear on the LCD and will continuously ash while in a data logging session.
4. To pause the data logging session press the LOG button and "REC" will stop ashing. To resume press the LOG button again.
5. To end the current data logging session pause the data logging session and hold the REC button until "REC" disappears.
continued...
16
www.GlobalTestSupply.com
Transferring Data from the SD Memory Card to a Computer
70
1 2 3 4 5 6 7 8 9 10 11 12
1. While the meter is powered Off remove the SD memory card from the SD card slot.
2. Insert the SD memory card into the computer.
3. Open the le(s) with Excel. See sample below:
Place Date Time Value Unit Value Unit
1 19-07-10 1:13:10 18.4 %RH 22.6 Degree C 2 19-07-10 1:13:12 30 %RH 25 Degree C 3 19-07-10 1:13:14 25 %RH 35 Degree C 4 19-07-10 1:13:16 45 %RH 30 Degree C 5 19-07-10 1:13:18 30 %RH 35 Degree C 6 19-07-10 1:13:20 45 %RH 25 Degree C 7 19-07-10 1:13:22 50 %RH 35 Degree C 8 19-07-10 1:13:24 60 %RH 30 Degree C 9 19-07-10 1:13:26 45 %RH 28 Degree C 10 19-07-10 1:13:28 30 %RH 22.6 Degree C 11 19-07-10 1:13:30 17.5 %RH 22.6 Degree C 12 19-07-10 1:13:32 17.5 %RH 22.6 Degree C
The raw data can be used to create a graph in Excel. See sample below:
60
50
40
30
20
10
0
1:13:10 1:13:12 1:13:14 1:13:16 1:13:18 1:13:20 1:13:22 1:13:24 1:13:26 1:13:28 1:13:30 1:13:32
19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10 19-07-10
www.GlobalTestSupply.com
Value
Unit
Value
Unit
17
Humidity Calibration Procedure
In order to calibrate the meter humidity calibration standards are required (REED R9933, R9975 or R9980). If the meter has been exposed to extreme conditions or it has not been used in a long period of time the meter should be calibrated by following steps 1 through 5 below.
1. While in normal %RH measurement mode, hold the SET and LOG buttons simultaneously for approx. 2 seconds to enter the calibration mode of %RH.
2. The LCD display will show the same relative humidity measuring value along with the %RH symbol.
3. Press the and buttons to adjust the upper display until it matches the %RH Humidity Calibration Standard.
4. Once the calibration is complete, press "REC" button to save the calibration value into the memory which will ash conrming it has been saved.
5. Press the NEXT button twice to exit the calibration mode and resume normal operation.
Note: At any time, you can press the ESC button to exit the calibration mode and resume normal operation.
Data Stream Output
Using Terminal application, a data stream can be viewed from the RS-232 output, connect RS-232/USB cable between the product and terminal/PC and use the following setting to view the data stream.
Terminal Settings:
• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
www.GlobalTestSupply.com
continued...
18
41010100000501 = Upper Value, +50.1% RH
Reading
4
RS-232 Mono Jack
Pin 2 (RX) Pin 4 (DTX)
Sleeve Tip
Decimal Place 0-3
Positive (+) = 0, Negative (-) = 1
°C = 01, °F = 02, %RH = 04
1 = Upper display value, 2 = Lower display value
3.5mm Terminal to RS-232 Serial Connection
System Reset
If the meter becomes unresponsive or if the display freezes the Reset pin can be used to reset the instrument.
1. Use a paper clip or any similar small object to press the Reset pin.
2. After pressing the Reset pin, power the meter back up by holding the POWER button for 2 seconds.
3. If resetting the meter does not resolve the issue, please return the meter to the nearest authorized REED distributor or authorized Service Center for repair.
www.GlobalTestSupply.com
19
Battery Replacement
When the low battery icon appears on the LCD, the batteries must be replaced.
1. Remove the two (2) Phillips screws on the back of the meter directly located above the top of the kickstand.
2. Remove the battery cover.
3. Replace 6 x "AA" batteries.
4. Secure the battery cover with the two (2) Phillips screws.
Applications
• Environmental studies
• Food distribution
• HVAC/R system installation
• Greenhouses and industrial spaces
• Saunas
• Humidors
• Museums
• Lab Monitoring
Accessories and Replacement Parts
R9933 Humidity Standard, 33% R9975 Humidity Standard, 75% R9980 Humidity Standard, 33% and 75% R6050SD-PROBE Replacement Probe for R6050SD R2920 Surface Thermocouple Probe R2930 Right Angle Thermocouple Probe R2940 Air/Gas Thermocouple Probe R2950 Immersion Thermocouple Probe R2960 Needle Tip Thermocouple Probe TP-01 Beaded Thermocouple
continued...
20
www.GlobalTestSupply.com
R1500 Tripod RSD-ADP-NA Power Supply, 110V RSD-ADP-EU Power Supply, 220V CA-05A Soft Carrying Case R8888 Deluxe Hard Carrying Case SD-4GB 4GB Class 4 SDHC Memory Card RSD-16GB 16GB Micro SD Memory Card w/ Adapter
Don't see your part listed here? For a complete list of all accessories and replacement parts visit your product page on www.REEDInstruments.com.
Frequently Asked Questions (FAQ's)
I cannot turn off my meter, do you know why?
Often times, the meter cannot be turned off because it is in recording mode ("REC" mode). Make sure to exit that function by holding down the
REC button before attempting to turn the meter off.
How long can I record for?
Battery life will depend on a number of factors:
• Quality of Battery
• Whether the beeper is ON or OFF
• Whether the backlight is ON or OFF
• Environmental Conditions during data logging
As a reference, using alkaline batteries with both the backlight and beeper on while automatic data logging at a 2 second sampling rate; the batteries will last approximately 14 hours.
Is there any way to make the R6050SD record temp, humidity, dew point, wet bulb, and Type K readings all at the same time?
The R6050SD is unable to measure all the parameters simultaneously.
www.GlobalTestSupply.com
continued...
21
If the screen displays the "_ET" function, will it still record humidity/ temperature? How do I select the "rH" function to display on the screen. I want to record the humidity/temperature for an extended period of time.
The R6050SD offers 4 types of measurement modes:
1. Humidity/ Temp
2. Dew Point
3. Wet Bulb
4. Type K/J Thermometer
When the meter is on, press and hold the FUNCTION button to scroll through the measuring modes. The display will ash "rH", for Humidity/ Temperature measurement, "dP" for Dew Point measurement, "_Et" for Wet Bulb measurement, and "tP" for Type K/J Thermometer measurement. Release the FUNCTION button to conrm selection. Select the "rH'' option for extended recording of temperature and humidity.
Product Care
To keep your instrument in good working order we recommend the following:
• Store your product in a clean, dry place.
• Change the battery as needed.
• If your instrument isn't being used for a period of one month or longer
please remove the battery.
• Clean your product and accessories with biodegradable cleaner. Do not
spray the cleaner directly on the instrument. Use on external parts only.
www.GlobalTestSupply.com
22
Product Warranty
REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one (1) year from date of shipment. During the warranty period, REED Instruments will repair or replace, at no charge, products or parts of a product that proves to be defective because of improper material or workmanship, under normal use and maintenance. REED Instruments total liability is limited to repair or replacement of the product. REED Instruments shall not be liable for damages to goods, property, or persons due to improper use or through attempts to utilize the instrument under conditions which exceed the designed capabilities. In order to begin the warranty service process, please contact us by phone at 1-877-849-2127 or by email at info@REEDInstruments.com to discuss the claim and determine the appropriate steps to process the warranty.
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling your instrument. Your product contains electronic components and must be disposed of separately from standard waste products.
Product Support
If you have any questions on your product, please contact your authorized REED distributor or REED Instruments Customer Service by phone at 1-877-849-2127 or by email at info@REEDInstruments.com.
Please visit www.REEDInstruments.com for the most
up-to-date manuals, datasheets, product guides and software.