Introduction
Thank you for purchasing your REED R6020 Temperature & Humidity USB
Data Logger. Please read the following instructions carefully before using
your instrument. By following the steps outlined in this manual your data
logger will provide years of reliable service.
www.GlobalTestSupply.com
Product Quality
This product has been manufactured in an ISO9001 facility and has been
calibrated during the manufacturing process to meet stated product
specications. If a certicate of calibration is required please contact the
nearest authorized REED distributor or authorized Service Center. Please
note an additional fee for this service will apply.
Safety
Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries, may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center.
Features
• Stores up to 16,000 temperature and 16,000 humidity readings
• Simply plug into the USB port to begin downloading - no cables or
wires required
• User selectable sampling rate from 2 secs to 24 hrs
• Visible alarm (LED) if set max/min values are exceeded
• Wall mount bracket and screws for long-term monitoring
Included
• Wall Mount Bracket and screws
• Protective USB cap
• Software
• Battery
www.GlobalTestSupply.com
Specifications
Temperature
Measuring Range: -40 to 158°F (-40 to 70°C)
Accuracy: ±1.8°F (1.0°C)
Resolution: 0.1°F/°C
Humidity
Measuring Range: 0 to 100% RH
Accuracy: 0 to 20% & 80 to 100%: ±5%
20 to 40% & 60 to 80%: ±3.5%
40 to 60%: ±3%
Resolution: 0.1% RH
General Specifications
Datalogging Capabilities: Yes
Real-Time Clock and Date Stamp: Yes
Settable Alarms: Yes (temp. and RH)
Sampling Rate: Yes (2 secs to 24 hours)
Internal Memory: 32k readings
(16,000 temp. and 16,000 humidity)
Power Supply: 3.6V battery
PC Connectivity: Yes (USB)
Software: Included
Product Certications: CE
Operating Temperature: -31 to 176°F (-35 to 80°C)
Storage Temperature: -31 to 176°F (-35 to 80°C)
Operating Humidity Range: 0 to 100%
Dimensions: 4 x 1 x 0.9" (101 x 25 x 23mm)
Weight: 6 oz (172g)
www.GlobalTestSupply.com
Instrument Description
2
3
4
5
6
7
R6020
DATA LOGGER
1. Protective Cover
2. USB Connector
3. Start Button
4. RH & Temperature Sensors
5. Mounting Clip
6. Record LED (Green)
7. Alarm LED (Red/Yellow)
Operating Instructions
Software Installation
Install the included software by placing the supplied program disc in the
PC's CD-ROM drive. If the installation program does not automatically start,
open and browse the CD-ROM drive from your computer. Double click on
the setup le and follow the on-screen prompts to install the software.
If your computer does not have a CD drive please visit
www.reedinstruments.com/software to download the latest R6020 software.
Battery Installation or Battery Replacement
A 3.6V battery is provided with the R6020. Please note that only a 3.6V
battery should be used with the instrument. In order to install the battery
follow the steps below:
1. Use a small athead screwdriver to open the casing
2. Slide the casing off to expose the battery
3. Install (or replace) the battery
4. Slide the casing back on the instrument
continued...
www.GlobalTestSupply.com
Setting up or Downloading Data
234
1. Once the software and battery have been installed, insert the R6020
in an available USB port on your PC.
2. Double click the software icon on your desktop to load the program.
3. In the bottom left corner of the software window there is a status
indicator that should show your R6020 is connected "Connect".
If you do not see "Connect" please verify your battery has been
correctly installed.
3. Set up the data logger for a new session, or download saved data
from a previous recording.
4. When setup is complete or data has been downloaded, the R6020
should be removed from the USB port in order to preserve battery life.
5. Full specications and Operating System compatibility can be found
on the product page at www.reedinstruments.com.
6. If you have specic questions related to your application and/or
questions related to software setup and functionality please contact
the nearest authorized distributor or Customer Service at
info@reedinstruments.com or 1-877-849-2127.
LED Status Guide
www.GlobalTestSupply.com
R6020
DATA LOGGER
continued...
LEDsStatusIndicationAction
REC ALM
Both LED
lights OFF
Single Green
ash every
Logging not activeStart logging
Low battery or no
battery installed
Logging, no set
Install or replace
battery
No action required
alarm triggered**
10 seconds*
REC ALM
Double Green
ash every
10 seconds*
Manual start was
chosen from setup
Press and hold the
start button until the
Green and Yellow
LEDs ash indicating
recording has started
REC ALM
REC ALM
Single Red
ash every
10 seconds*
Double Red
ash every
10 seconds*
Single Red
ash every 60
seconds
Single Yellow
ash every 10
seconds*
Double Yellow
ash every
10 seconds*
Single Yellow
Logging, Low
No action required
alarm for relative
humidity triggered
Logging, High
No action required
alarm for relative
humidity triggered
Low BatteryReplace Battery,
download saved
data
Logging, Low alarm
No action required
for temperature
triggered
Logging, High alarm
No action required
for temperature
triggered
Memory fullDownload data
ash every
60 seconds*
*To conserve battery power, the LED ashing frequency can be adjusted between 15 and 30
seconds in the software.
**To conserve battery power, the alarm LEDs for temperature and humidity can be disabled
via software.
www.GlobalTestSupply.com
LED Status Notes
• When the battery is low, all recording operations will be disabled
automatically. When the recording is disabled the R6020 will still retain
logged data. The supplied software is required to restart recording and to
download data.
•
When both temperature and humidity alarms have been triggered, the
LED status will alternate every cycle. For example, if only one alarm has
been triggered, the REC LED blinks Green for one cycle (indicating the
instrument is logging) and the ALM indicator will blink RED during the next
cycle (indicating an alarm has been triggered). If two alarms have been
triggered, the REC LED will not blink green at all. The rst alarm will blink
during the rst cycle and the second alarm will blink during the next cycle.
FAQ's
How do I start recording? What's the difference between "Instant" and
"Manual" recording options during the software setup process?
By selecting "Manual" for the start mode, you will have to press the start
button on the instrument to begin recording. When the "Instant" mode is
selected, data logging will automatically start as soon as the instrument is
pulled out of the USB port.
How do I stop recording?
Unfortunately there is no way to stop recording in the eld. In order to stop
recording the R6020 needs to be connected to your PC.
Is the R6020 compatible with my Apple (MAC) computer or tablet?
At this time the R6020 is only compatible with PC computers.
I cannot install my software on my work computer, do you know why?
Please note that you will require administrative rights to install and run the
R6020 software. In order to install and run the software you will have to
contact your IT department/personnel.
Can I use my R6020 without the software?
In order to setup the instrument and retrieve data the R6020 software
is required.
www.GlobalTestSupply.com
Product Care
To keep your instrument in good working order we recommend
the following:
• Store your product in a clean, dry place.
• Change the battery as needed.
• If your instrument isn't being used for a period of one month or longer
please remove the battery.
Product Warranty
REED Instruments guarantees this instrument to be free of defects in
material or workmanship for a period of one (1) year from date of shipment.
During the warranty period, REED Instruments will repair or replace, at no
charge, products or parts of a product that proves to be defective because
of improper material or workmanship, under normal use and maintenance.
REED Instruments total liability is limited to repair or replacement of the
product. REED Instruments shall not be liable for damages to goods,
property, or persons due to improper use or through attempts to utilize the
instrument under conditions which exceed the designed capabilities. In
order to begin the warranty service process, please contact us by phone at
1-877-849-2127 or by email at info@reedinstruments.com to discuss the
claim and determine the appropriate steps to process the warranty.
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling your
instrument. Your product contains electronic components and must be
disposed of separately from standard waste products.
www.GlobalTestSupply.com
Product Support
If you have any questions on your product, please contact your authorized
REED distributor or REED Instruments Customer Service by phone at
1-877-849-2127 or by email at info@reedinstruments.com.
Please visit www.REEDINSTRUMENTS.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice.
All rights reserved. Any unauthorized copying or reproduction of
this manual is strictly prohibited without prior written permission
from REED Instruments.
www.GlobalTestSupply.com
TEST & MEASURE
WITH CONFIDENCE
CHECK OUT OUR LATEST PRODUCTS!
www.REEDINSTRUMENTS.com
www.GlobalTestSupply.com
TEMPERATURE
& HUMIDITY
SOUND
MOISTURE
AIR VELOCITY
ELECTRICAL
www.REEDINSTRUMENTS.com
www.GlobalTestSupply.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.