Introduction
Thank you for purchasing your REED R5800 Voltage/Current Simulator.
Please read the following instructions carefully before using your instrument.
By following the steps outlined in this manual your meter will provide years
of reliable service.
Product Quality
This product has been manufactured to meet stated product specications.
If a certicate of calibration is required please contact the nearest authorized
REED distributor or authorized Service Center. Please note an additional fee
for this service will apply.
Safety
To prevent the user and the instrument from electric shock and other
hazards, it is necessary to follow the following regulations:
• Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries, may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center.
•
Do not operate the instrument in the presence of ammable or explosive
gas or vapor. This is extremely dangerous.
• Never apply more than 30V between any two terminals, or between
terminal and ground.
• For optimal accuracy, allow the instrument to warm up for 5 minutes
before operating.
www.GlobalTestSupply.com
3
Features
• Sources voltage and current
• Large, 5-digit LCD display
• Automatic calibration
• Low battery indicator and auto shut off
Specifications
Output Functions
Applicable range from 18 to 28°C, within one year of calibration.
OutputRange
DCV10V
Output
Range
0.000 to
11.000V
ResolutionAccuracyNote
1mV
0.05% of set
value ±2mV
current: 2mA
Max output
DCA20mA
Analog
transducer
Loop
power
0.000 to
22.000mA
-20mA
24V--±10%
0.000 to
-22.000mA
0.001mA
0.001mA
0.05% of set
value ±4μA
0.05% of set
value ±4μA
Max load:
1kΩ at 20mA
note 1
Max load:
1kΩ at 20mA
Max output
current up
to 25mA
Note 1: When the battery voltage exceeds 6.8V, the max load is 1KΩ at
20mA. When its voltage lies between 5.8V and 6.8V, the max load is 700Ω
at 20mA.
Note 2: Temperature coefcient: ±0.005% of the range per °C for the
temperature ranges 5 to 18°C and 28 to 40°C.
continued...
www.GlobalTestSupply.com
4
General Specifications
Display: 5-Digit LCD
Kick Stand: Yes
Battery Life: Approximately 20 hours
Power Supply: 2 AA Batteries
Auto Shut Off: Yes (after 15 minutes/off)
Low Battery Indicator: Yes
Replaceable Test Leads: Yes
Product Certications: CE
Operating Temperature: 32 to 122°F (0 to 50°C)
Operating Humidity Range: 0 to 85%
Storage Temperature: 14 to 122°F (-10 to 50°C)
Dimensions: 7.1 x 3.5 x 1.9" (180 x 90 x 47mm)
Weight: 8.2oz (500g)
Included
• Test Leads
• Alligator Clips
• Protective Holster
• Batteries
www.GlobalTestSupply.com
5
Instrument Description
R5800
3
4
5
6
Voltage/Current Simulator
7
8
9
10
1. LCD Screen
2. Unit V/mA Select Button
3. Power Button
4. Output ON/OFF Button
5.Output Value
Setting Button
6. Output Digit
Selection Button
www.GlobalTestSupply.com
7. Unit mA/% Select Button
8. Zero/FS Points Select Button
9. 25% Single Step Setting/Auto
Ramp Button
10. Output Terminals
6
Display Description
Indicates that the instrument is in an output state
Indicates that the instrument is in a calibration state
Indicates that the instrument is in a calibration state, denoting
that the zero point or the full scale point is now in calibration
Indicates that the battery power is low and needs to
be replaced
Indicates that the output digits need to be set
V. mA. % Indicates the current output value (unit of measure)
Indicates that the output signal is on or off
www.GlobalTestSupply.com
7
Operating Instructions
Power ON/OFF
Press the POWER button to turn the instrument on. To turn the instrument
off, press and hold the POWER button.
Automatic Power-Off
As a default the instrument will automatically turn off after 15 minutes of
inactivity. Follow the steps below to Enable/Disable the Auto-Power
Off function.
1.Turn the meter off.
2.Press the ON/OFF button (to display the full screen).
3.Quickly press the mA% button when the instrument is in the maintenance state as indicated by "1.8.8.888".
4."AP-XX" will appear on the display.
5.Press the
indicates that the automatic power-off function is disabled, while
"AP-ON" indicates the automatic power-off function is enabled.
6.Press the mA 100% button to save the required setting.
7.Press and hold the ON/OFF button to exit the maintenance state and
turn the instrument off.
Output Function
The output terminal of the instrument can produce DC voltages set by the
user or simulating resistance.
Do not apply any voltage to the output terminal during the operation. If any
improper voltage is applied to the output terminal, it will cause damage to
the internal circuits.
button to toggle between "AP-ON" & "AP-OF". "AP-OF"
www.GlobalTestSupply.com
continued...
8
Output Operation Procedure
XMT
V
30V
MAXMAX
30V
30V
MAX
+mA-
XMT
V
30V
MAX
MAX
30V
30V
MAX
+mA-
Function
Operation
DCA 10mA20mA%0.000V
Range
Operation
DisplaySet Range
0.000 to 11.000V
-00.000 to 22.000mA
DC Voltage Output
1.Insert one end of the test leads into the
V output jack and connect the other
end to the input of the meter, as shown
in the diagram.
2.Press the V/mA button to select V
function and V will appear on the display.
3.Press the buttons to select the
output digits.
4.Press the buttons to change the numerical value of the set
digits. Hold the button and the value will keep varying.
5.Press the ON/OFF button to turn the output signal on and off as
indicated by ON or OFF on the LCD display.
DC Current Output
1.Insert one end of the test leads into the
+mA output jack and connect the other
end to the input of the meter, as shown
in the diagram.
2.Press the V/mA button to select mA
function and mA will appear on
the display.
3.Press the mA/% button to select mA or
% function to display the mA or % unit,
in which 0% is 4mA and 100% is 20mA.
4.Press the buttons to select the output digits.
5.Press the buttons to change the numerical value of the set
digits. Hold the button and the value will keep varying.
6.Press the ON/OFF button to turn the output signal on and off as
indicated by ON or OFF on the LCD display.
www.GlobalTestSupply.com
continued...
9
25% Step Current Output
1.Connect the test leads to the meter as stated in the Current
Output procedures.
2.Press the V/mA button to select mA function and mA will appear on
the display.
3.Press the mA 25% button and
4.Press the mA/% button to select the set output mA or % and mA or
mA% will appear on the display.
5.Press the button to change the output value in 25% increments
in which 0% indicates 4mA and 100% indicates 20mA
6.Press the ON/OFF button to turn the output signal on and off as
indicated by ON or OFF on the LCD display.
7.Press the mA 25% button to exit the 25% current output setup.
will appear on the display.
,
.
Current Output Set for Zero Point & Full Scale
1.Connect the test leads to the meter as stated in the Current
Output procedures.
2.Press the V/mA button to select the mA function and mA will appear
on the display
3.Press the mA 100% button.
4.Press the mA/% button to select the set output mA or % and mA or
mA% will appear on the LCD display.
5.Press the buttons to change the output values at 100%, in
which 0% indicates 4mA and 100% indicates 20mA.
6.Press the ON/OFF button turn the output signal on and off as indicated
by ON or OFF on the display.
7.
Press the mA 100% button to exit the full scale current output set up.
, 0 and FS will appear on the display.
www.GlobalTestSupply.com
continued...
10
Simulating Transmitter Output (XMT)
1.Insert one end of the test lead to the 'XMT' output jack of the calibrator
and connect the other end with the input terminal of the user's instrument
as shown in the following diagram:
XMT
+mA-
V
5--25VDC POWER
MAXMAX
30V
MAX
30V
30V
2.The button-operation is the same as that as indicated in the Current
Output section.
Note:
• Power supply range: 5 to 25V DC.
• During an output operation, use a external 24V DC power supply in
order to prolong battery life.
Battery Replacement
When the appears on the LCD, the batteries need to be replaced.
Follow the instructions below to properly replace the batteries.
1.Turn the power off the meter and remove any test leads that may be
attached to the meter.
2.Lift the tilt stand on the back of the unit to access the battery
compartment door, which can be removed using a Phillips
head screwdriver.
3.Replace the 2 x "AA" batteries in the lid of the battery compartment.
continued...
11
www.GlobalTestSupply.com
4.Reinstall the compartment lid by ensuring that battery terminals
touch the unit's contact points and snap into place.
5.Tighten the screw to secure the battery compartment door.
Ensure battery door is closed and secure in place before
using the meter. To ensure proper operation, wait 5 seconds
before turning meter on after changing batteries.
Fuse Replacement
Figure 1
1.Remove the test leads from the meter and turn the meter OFF.
2.Take off the protective boot (Figure 1), remove the four screws by using
a standard-blade screwdriver, and then remove the cover (Figure 2).
3.Replace the blown fuse(s).
4.Reinstall the cover.
5.Reinstall the meter's protective boot.
Figure 2
12
www.GlobalTestSupply.com
Accessories and Replacement Parts
• FC-300 Fused Test Lead Set
• R1000 Safety Test Lead Set, Double Insulated
• CA-05A Soft Carrying Case
• R9940 Hard Shell Carrying Case
Don't see your part listed here? For a complete list of all accessories and
replacement parts visit your product page on www.reedinstruments.com.
Product Care
To keep your instrument in good working order we recommend the following:
• Store your product in a clean, dry place.
• Change the battery as needed.
• If your instrument isn't being used for a period of one month or longer
please remove the battery.
• Clean your product and accessories with biodegradable cleaner. Do not
spray the cleaner directly on the instrument. Use on external parts only.
Product Warranty
REED Instruments guarantees this instrument to be free of defects in
material or workmanship for a period of one (1) year from date of shipment.
During the warranty period, REED Instruments will repair or replace, at no
charge, products or parts of a product that proves to be defective because
of improper material or workmanship, under normal use and maintenance.
REED Instruments total liability is limited to repair or replacement of the
product. REED Instruments shall not be liable for damages to goods,
property, or persons due to improper use or through attempts to utilize the
instrument under conditions which exceed the designed capabilities. In
order to begin the warranty service process, please contact us by phone at
1-877-849-2127 or by email at info@reedinstruments.com to discuss the
claim and determine the appropriate steps to process the warranty.
www.GlobalTestSupply.com
13
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling your
instrument. Your product contains electronic components and must be
disposed of separately from standard waste products.
Product Support
If you have any questions on your product, please contact your authorized
REED distributor or REED Instruments Customer Service by phone at
1-877-849-2127 or by email at info@reedinstruments.com.
Please visit www.REEDINSTRUMENTS.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice.
All rights reserved. Any unauthorized copying or reproduction of
this manual is strictly prohibited without prior written permission
from REED Instruments.
www.GlobalTestSupply.com
14
TEST & MEASURE
WITH CONFIDENCE
CHECK OUT OUR LATEST PRODUCTS!
www.GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.