Thank you for purchasing your REED R5340 Cable Length Tester. Please
read the following instructions carefully before using your instrument. By
following the steps outlined in this manual your meter will provide years of
reliable service.
Product Quality
This product has been manufactured in an ISO9001 facility and has been
calibrated during the manufacturing process to meet stated
product specications.
Safety
• Do not connect this unit to live voltage.
• Ensure the test leads are removed from the circuit and the unit is
powered off before opening the case.
• Do not expose this meter to rain, moisture, or high humidity.
• Do not operate the meter in extreme temperatures.
• Do not use the meter if it is wet or damaged.
• Always ensure test leads or accessories are clean and dry, and their
insulation is in good condition before use.
• Do not operate the meter with the case open.
• Using the meter near equipment that generates electromagnetic
interference may cause unstable or inaccurate readings.
• Failure to observe these safety instructions can result in serious injury
or death.
• Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center.
Kick Stand: Yes
Autoshut off: Yes (after 15 mins)
Power Supply: 6 x AA Batteries
Low Battery Indicator: Yes
Product Certications: CE
Operating Altitude: 7000' (2000m)
Operating Temperature: 32 to 104°F (0 to 40°C)
Operating Humidity: 10 to 80%
Storage Temperature: -4 to 140°F (-20 to 60°C)
Storage Humidity: 10 to 80%
Dimensions: 8.3 x 3.9 x 2.6" (212 x 100 x 67 mm)
Weight: 1.32lbs (600g)
Warning: Electric shock hazard: Do not connect the unit to live
voltage. Contact with live circuits could result in severe injury
or death.
Power ON/OFF
Rotate the function dial to any position to power on the meter. If the meter
does not turn on, check the batteries. For details on battery replacement,
refer to section ''Battery replacement'' for details.
Calibration Procedure
Every time the meter is turned ON, it must be calibrated before use.
1. Turn the function dial from the OFF position to the Ω position.
2. Connect the red Kelvin clip to the meter's red terminals.
3. Connect the black Kelvin clip to the meter’s black terminals.
Important: Ensure that both the calibration bar and the jaws of the
Kelvin clips are clean before proceeding.
I
4. Connect the two alligator clips to the supplied copper rod.
5. Adjust the clips so they are as close together as possible along
7. The display will alternate between “ YES” and “no
”, ashing
twice each time as shown below.
8. Press the MEM button. If the display briey indicates “dOnE” and
then returns to resistance measurement mode, the calibration
has succeeded.
Note: If the display indicates "FAIL" after pressing the MEM Button,
the calibration has failed. Check whether the connections and contacts
are correct and secure.
9. If you press the CAL button again, the meter will exit calibration
Warning: Electric shock hazard: Do not connect the unit to live
voltage. Contact with live circuits could result in severe injury
or death.
1. Turn the function dial from the OFF position to the Ω position.
2. Follow the steps outlined in the "Calibration Procedure" to ensure
accurate measurements.
Note: Temperature affects the accuracy of the readings. For best
results, allow the meter and the cable to reach the same ambient
temperature. Typically, this takes about 30 minutes.
3. Strip the insulation from both ends of the cable. Ensure the exposed
conductor is clean and fully visible. If necessary, use abrasive paper
to remove any oxide layer from the exposed ends.
Note: Both ends must be clean for accurate measurement.
4. Adjust the function dial to the appropriate cable gauge position
based on the cable being measured.
Example: For a cable gauge of 1 mm², set the rotary switch to
the “1” position.
5. If the cable is copper, press the Cu/AI button until "Cu" appears on
the display. If the cable is aluminum, press the Cu/AI button until "AI"
appears on the display.
6. Press the FT/M button to choose the unit of measurement (feet
or meters).
7. Attach one alligator clip to one end of the cable and the other
alligator clip to the opposite end of the cable.
8. The display will indicate the measured length of the cable in the
selected unit (feet or meters) and the ambient temperature.
Note: If the temperature falls below 23°F (-5°C), the display will
show "-OL" to indicate it is out of range. If the temperature exceeds
122°F (50°C), the display will show “OL” to indicate it is out of range.
Press and hold the Cu/AI button to toggle between Celsius (°C) and
Fahrenheit (°F) display modes.
9. Once the measurement is complete, disconnect the alligator clips
from the cable.
User-selected Programming Mode:
The User Selected Mode offers the following functionalities:
Resistance Parameter Storage: Users can save the resistance
parameters of a specic cable for length measurements of cables with the
same characteristics (i.e., same gauge and conductor material).
Accurate Length Measurement: It allows precise measurement of
standard gauge cables.
Versatility: In this mode, users can measure the length of various metal
cables, including copper and aluminum cables.
Note: Before starting, prepare a sample cable with a length between
4 m and 100 m (or 13.1 ft to 320 ft in FEET mode). This sample will
be programmed into the meter so that it can correctly measure and
store the resistance of this specic cable type. Ensure the cable being
measured is not energized or live.
1. Turn the function dial from the OFF position to the Ω position.
2. Follow the steps outlined in the “Calibration Procedure” to calibrate
the meter.
Note: Temperature affects the accuracy of the readings. For best
results, allow the meter and the cable to reach the same ambient
temperature. Typically, this takes about 30 minutes.
3. Remove the insulation from both ends of the sample cable.
Note: Ensure the two conductors are clean and fully exposed. Use
a grinding pad if needed to clean the exposed cable ends for better
conductivity and easier clamping with the test rods.
4. Use the function dial to select a
User-Dened Mode from 1 to 8.
5. The selected User-Dened mode will be
displayed in the bottom right corner of the
screen. If the selected User-Dened mode
already contains a stored cable parameter,
the screen will display "OL" as shown below.
6.Press the FT/M button to toggle between feet (FT) and meters (M) as
MEM
T-CAL
MEM
the unit of measurement.
7.Press the CAL button. to enter the sample cable length setting
mode. In this mode, the display alternates between " YES" and
"no
" with “MEM” and “TCAL” ashing twice each cycle.
8.Use the Up/Down buttons to adjust the displayed value.
9.Adjust the display until it matches the actual length of the sample
cable to be measured.
10. Attach one alligator clip to each end of the sample cable, ensuring
secure connections at both ends.
11. Press the MEM button. If the display briey indicates “dOnE” and
then returns to measurement mode, the setting has succeeded.
Note: If the display indicates "FAIL" and "ERROR", the setting has
failed. In this case, verify that all connections and contacts are secure
and correct.
12. Press the CAL button to resume normal operation.
13. Remove the two alligator clips from the sample cable.
Measuring Cable Length in User-Defined Mode
1.Turn the function dial from the OFF position to the Ω position.
2.Follow the steps outlined in the “Calibration Procedure” to calibrate
the meter.
Note: Temperature affects the accuracy of the readings. For best
results, allow the meter and the cable to reach the same ambient
temperature. Typically, this takes about 30 minutes.
3.Remove the insulation from both ends of the sample cable.
Note: Ensure the two conductors are clean and fully exposed. Use
a grinding pad if needed to clean the exposed cable ends for better
conductivity and easier clamping with the test rods.
4. Use the function dial to select a
User-Dened Mode from 1 to 8.
5. The selected User-Dened mode will be
displayed in the bottom right corner of the
screen. If the selected User-Dened mode
already contains a stored cable parameter,
the screen will display "OL" as shown below.
6. Press the FT/M button to toggle between feet (FT) and meters (M) as
the unit of measurement.
7. Attach one alligator clip to each end of the cable under test, ensuring
secure connections at both ends.
8. The display will indicate the measured length of the cable in the
selected unit (feet or meters) and ambient temperature.
Note: If the temperature falls below 23°F (-5°C), the display will
show "-OL" to indicate it is out of range. If the temperature exceeds
122°F (50°C), the display will show “OL” to indicate it is out of range.
Press and hold the Cu/AI button to toggle between Celsius (°C) and
Fahrenheit (°F) display modes.
9. Once the measurement is complete, disconnect the alligator clips
Use this table to log the cable sizes assigned to each programmable
switch position. For exibility, make entries using an erasable
medium (e.g., pencil or erasable marker), as switch positions may be
reprogrammed in the future.
Switch
Position
Size Type
Gauge
Material
Type
1mm² AWG kcmil aughtCu or AI
2mm² AWG kcmil aughtCu or AI
3mm² AWG kcmil aughtCu or AI
4mm² AWG kcmil aughtCu or AI
5mm² AWG kcmil aughtCu or AI
6mm² AWG kcmil aughtCu or AI
7mm² AWG kcmil aughtCu or AI
8mm² AWG kcmil aughtCu or AI
Clearing Memory Locations in User-Selected Mode
1.
Ensure the test rod is disconnected from the meter before proceeding.
2.Turn the function dial to select the user-dened mode whose
memory you want to clear.
3.Press the CAL button. The display will indicate " YES".
4.To clear the memory, press the MEM button. The display will indicate
" CLr" conrming that the stored data in the selected memory
location will be erased.
1. Turn the function dial from the OFF position to the “Ω‘’ position.
2. Follow the steps outlined in the “Calibration Procedure” to calibrate
the meter.
Note: Temperature affects the accuracy of the readings. For best
results, allow the meter and the cable to reach the same ambient
temperature. Typically, this takes about 30 minutes.
3. Remove the insulation from both ends of the cable.
Note: Ensure the two conductors are clean and fully exposed. Use
a grinding pad if needed to clean the exposed cable ends for better
conductivity and easier clamping with the test rods.
4. Attach one alligator clip to each end of the cable under test, ensuring
secure connections at both ends.
5.
The display will indicate the impedance value and ambient temperature.
Note: If the temperature falls below 23°F (-5°C), the display will
show “-OL” to indicate it is out of range. If the temperature exceeds
122°F (50°C), the display will show “OL” to indicate it is out of range.
Press and hold the Cu/AI button to toggle between Celsius (°C) and
Fahrenheit (°F) display modes.
6. Once the measurement is complete, disconnect the alligator clips
from the cable.
Data Hold
1. While taking a measurement, press the HOLD button to freeze the
current reading on the LCD display.
2. While in this mode an "H" symbol will appear.
3. Press the HOLD button again to resume normal operation.
REED Instruments guarantees this instrument to be free of defects in
material or workmanship for a period of one (1) year from date of shipment.
During the warranty period, REED Instruments will repair or replace, at no
charge, products or parts of a product that proves to be defective because
of improper material or workmanship, under normal use and maintenance.
REED Instruments total liability is limited to repair or replacement of the
product. REED Instruments shall not be liable for damages to goods,
property, or persons due to improper use or through attempts to utilize the
instrument under conditions which exceed the designed capabilities. In
order to begin the warranty service process, please contact us by phone
at 1-877-849-2127 or by email at [email protected] to discuss
the claim and determine the appropriate steps to process the warranty.
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling your
instrument. Your product contains electronic components and must be
disposed of separately from standard waste products.
Product Support
If you have any questions on your product, please contact your authorized
REED distributor or REED Instruments Customer Service by phone at
1-877-849-2127 or by email at [email protected].
Please visit www.REEDInstruments.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice.
All rights reserved. Any unauthorized copying or reproduction of
this manual is strictly prohibited without prior written permission
from REED Instruments.