Product Disposal and Recycling .............................................................. 23
Product Support ....................................................................................... 24
Introduction
Thank you for purchasing your REED R5055 1000A True RMS Digital
Clamp Meter. Please read the following instructions carefully before using
your instrument. By following the steps outlined in this manual your meter
will provide years of reliable service.
www.REEDInstruments.com
www.GlobalTestSupply.com
3
Page 4
Product Quality
This product has been manufactured in an ISO9001 facility and has been
calibrated during the manufacturing process to meet stated product
specications. If a certicate of calibration is required please contact the
nearest authorized REED distributor or authorized Service Center. Please
note an additional fee for this service will apply.
Safety
• Read and understand this user manual before operating the meter.
• Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries, may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center.
• Do not exceed the maximum allowable input range of any function.
• Do not apply voltage to the meter when resistance function is selected.
• Set the rotary function switch to OFF when the meter is not in use.
• Set the rotary function switch to the appropriate position
before measuring.
• When measuring volts do not switch to current/resistance modes.
• After taking a measurement or when switching between functions
always disconnect the test leads from the circuit under test.
• Do not exceed the maximum stated input limits.
• Improper use of this meter can cause damage, shock, injury or death.
• Always remove the test leads before replacing the battery.
• Inspect the condition of the test leads and the meter itself for any
damage before operating.
• Remove the battery if the meter will be stored for an extended period
of time.
• Always discharge capacitors and remove power from the device under
test before performing diode, resistance, or continuity tests.
www.REEDInstruments.com
www.GlobalTestSupply.com
4
Page 5
Electrical Symbols
SymbolDescription
Equipment protected throughout by double insulation or
reinforced insulation
Earth (ground)
Warming or Caution
Alternating current
Direct current
Continuity buzzer
Diode
Capacitance
Alternating current or direct current
Caution, possibility of electric shock
Application around and removal from uninsulated hazardous
live conductors is permitted
Complies with European Union Standards
Conforms to UL STD 61010-1, 61010-2-032, 61010-2-033,
Certied to CSA STD C22.2 No. 61010-1, 61010-2-032,
61010-2-033
It is applicable to testing and measuring circuits connected
CAT III
CAT IV
to the distribution part of the building's low-voltage
mains installation
It is applicable to testing and measuring circuits
connected at the source of the building's low-voltage
mains installation
www.REEDInstruments.com
www.GlobalTestSupply.com
5
Page 6
Features
• Measures AC/DC current/voltage, resistance, capacitance, frequency,
duty cycle, and contact temperature
• 6,000-count backlit LCD display and analog bargraph
• Flexible current probe can expand AC current measurements up to
3000A (with R5065, sold separately)
• Ergonomic design ts in your hand and can be used while wearing
protective gloves
• Built-in LED ashlight
• Low battery and over range indicators
• Built-in non-contact voltage detector with LED indicator
• True RMS voltage and current for accurate measurements
• Low Pass Filter(LPF) and Low Impedance (LoZ) modes
Button Adjustment: Yes
Relative Mode: Yes
Diode Test: Yes
Backlit Display: Yes
Analog Bargraph: Yes (31-segment)
Continuity Check: Audible signal if resistance ≤30Ω
Duty Cycle: Yes
Non-Contact
Voltage Detector: Yes
Built-In Flashlight: Yes
Kick Stand: N/A
www.REEDInstruments.com
www.GlobalTestSupply.com
continued...
8
Page 9
Autoshut Off: Yes (after 15 mins)
Flexible Current
Probe (optional): Yes (R5065, Extends to 3000A)
Power Supply: 3 x AAA batteries
Low Battery Indicator: Yes
Jaw Opening: 1.7" (42mm), up to 1500 MCM
Overvoltage Category: CAT. III 1000V, CAT. IV 600V
Product Certications: CE, ETL
Operating Temperature: 32 to 122°F (0 to 50°C)
Storage Temperature: 14 to 140°F (-10 to 60°C)
Dimensions: 10.7x 3.2 x 1.7" (272 x 81 x 43.5mm)
Weight: 15.8oz (447g)
www.REEDInstruments.com
www.GlobalTestSupply.com
9
Page 10
Instrument Description
1
5
R5055
2
1000A AC/DC Clamp Meter
3
4
11
1. Current Clamp
2. Clamp Trigger
3. LCD Display
4. Function Buttons
5. Non-Contact Voltage Sensor
6. Non-Contact AC Voltage
Indicator Light
6
7
8
7. Flashlight Button
8. Rotary Function Switch
9. Flashlight LED Light
10. Battery Cover
11. Input Jacks
9
10
www.REEDInstruments.com
www.GlobalTestSupply.com
10
Page 11
Display Description
1
2
3
4
5
6
7
8
9
10
11 12
13
15 16
14
1819
17
1. Low Battery Indicator
2. Inrush Current Measurement
Indicator
3. High Voltage Indicator
4. Auto Power Off Indicator
5. Relative Mode Indicator
6. AC Signal Indicator
7. DC Signal Indicator
8. Flexible Current Probe
Indicator (R5065 required)
9. Data Hold Indicator
10. Zero Mode Indicator
11. Diode Test Mode Indicator
12. Audible Continuity Indicator
www.REEDInstruments.com
www.GlobalTestSupply.com
20
21 22
13. Non-Contact Voltage
Indicator
14. Low Impedance
Measurement Indicator
15. Units of Measurement
16. Flexible Current Probe Range
Indicator (R5065 required)
17. Analog Bar
18. Auto Range Mode Indicator
19. Max Measurement Indicator
20. Min Measurement Indicator
21. Temperature Measurement
Indicator
22. Low Pass Filter Measurement
Indicator
11
Page 12
Operating Instructions
AC/DC Current Measurements
1. Set the rotary function switch to the
position and select either AC or DC by
pressing the SELECT button.
Prior to taking DC current measurements
wait for the display to stabilize. When stable,
push the Zero button to ensure accurate
readings. Zeroing the meter removes DC
offset from the measurement reading.
Note: The Zero function works only when the
dial is set to the DC current measurement
position. Before zeroing the meter, make
sure the jaws are closed and no conductor is
inside the jaw.
2. Press the trigger to the open jaw and fully
enclose one conductor.
NOTE: Only one conductor can be measured
at a time.
3. The LCD will display the reading.
Current Frequency Measurement
1. While in AC current measurement mode and the meter is under test,
press Hz/INRUSH button to enter the Frequency measurement mode.
2. The LCD will display the reading.
3. Press the Hz/INRUSH button again to exit the Frequency
measurement mode.
R5055
1000A AC/DC Clamp Meter
Inrush Current Measurement
1. While the device to be tested is turned off, set the rotary function
switch to the
position.
2. Open the jaw and fully enclose one conductor.
continued...
www.REEDInstruments.com
www.GlobalTestSupply.com
12
Page 13
3. Press and hold the Hz/INRUSH button for 2 seconds to enter the
Inrush current measurement mode.
4. Turn on the device under test.
5. The LCD will display the reading.
NOTE: Inrush current is the highest AC current (True RMS) within 100ms of
start time, as shown below.
Measured Inrush
CURRENT
Current
Detection
Threshold
*
0A
StartStop
Current
TIME
100ms Sample Window (60Hz)
AC/DC Voltage Measurements
1. Insert the black test lead into the negative COM terminal and the red
test lead into the positive V terminal.
2. Set the rotary function switch to the position and select either AC or
DC by pressing the SELECT button.
3. Connect the test leads in parallel to the circuit under test.
4. The LCD will display the reading.
Voltage Frequency Measurement
1. While in AC voltage measurement mode and the meter is under test,
press Hz/INRUSH button to enter the Frequency measurement mode.
2. The LCD will display the reading.
3. Press the Hz/INRUSH button again to exit the Frequency
measurement mode.
www.REEDInstruments.com
www.GlobalTestSupply.com
continued...
13
Page 14
Low Pass Filter (LPF) ACV Measurement
1. While in AC voltage measurement mode, press and hold the SELECT
button to enter the LPF ACV measurement mode.
NOTE: Low Pass Filter can measure combined sine wave signals
produced by inverters and variable frequency drives, as shown below.
2. 5KHz
LPF
V
2. Once the LPF ACV function is activated, press the Hz/INRUSH to
enter the Frequency measurement mode.
3. The LCD will display the reading.
4. Press the Hz/INRUSH again to exit the Frequency measurement mode.
Low Impedance (LoZ) ACV Measurement
1. Insert the black test lead into the negative COM terminal and the red
test lead into the positive V terminal.
2. Set the rotary function switch to the position.
3. Connect the test leads in parallel to the circuit under test.
4. The LCD will display the reading.
LoZ ACV Frequency Measurement
1. While the rotary function switch is set to the position and
the meter is under test, press the Hz/INRUSH button to enter the
Frequency measurement mode.
2. The LCD will display the reading.
3. Press the Hz/INRUSH again to exit the Frequency measurement mode.
www.REEDInstruments.com
www.GlobalTestSupply.com
continued...
14
Page 15
Resistance Measurement
Ω
Ω
NOTE: Remove power before making resistance measurements.
1. Insert the black test lead into the negative COM terminal and the red
test lead into the positive V terminal.
2. Set the rotary function to the
position.
3. Press the SELECT button to switch to resistance measurement
mode (Ω).
4. Connect the test leads in parallel to the circuit under test. It is best to
disconnect one side of the device under test so the rest of the circuit
will not interfere with the reading.
5. The LCD will display the reading.
Continuity Measurement
1. Insert the black test lead into the negative COM terminal and the red
test lead into the positive V terminal.
2. Set the rotary function to the Ω position.
3. Press the SELECT button to switch to continuity
measurement mode ( ).
4. Connect the test leads in parallel to the circuit under test.
5. If the resistance is ≤30Ω, the circuit is in good condition and the
device will emit a continuous buzzer sound.
NOTE: If the measured resistance is ≥70Ω, the device will not emit a
buzzer sound.
Diode Test
1. Insert the black test lead into the negative COM terminal and the red
test lead into the positive V terminal.
2. Set the rotary function to the
3. Press the SELECT button to switch to diode test mode ( ).
4. Connect the red probe to the diode anode, and the black probe to
the diode cathode.
position.
continued...
www.REEDInstruments.com
www.GlobalTestSupply.com
15
Page 16
5. The LCD will display the reading. Forward voltage will indicate 0.5V
Forward Test Reverse Test
to 0.8V. Reverse voltage will be indicated by "OL". Shorted devices
will indicate near 0mV and an open device will be indicated by "OL"
in both polarities.
Red
Probe
Black
Probe
Black
Probe
Red
Probe
Capacitance Measurement
1. To avoid electric shock, before taking any capacitance
measurements disconnect power to the unit under test and discharge
all capacitors. Remove the batteries and unplug the line cords.
2. Insert the black test lead into the negative COM terminal and the red
test lead into the positive V terminal.
3. Set the rotary function to the position.
4. Touch the test leads to the capacitor under test.
5. The LCD will display the reading.
NOTE: If the measured capacitor is short-circuited or the capacitance
exceeds the maximum range, the LCD will display "OL". The analog bar
graph is disabled in capacitance measurement mode. When measuring
capacitance >600μF, please allow the meter time to stabilize.
Frequency/Duty Ratio Measurement
1. Insert the black test lead into the negative COM terminal and the red
test lead into the positive V terminal.
2. Set the rotary function to the Hz% position.
3. Press the SELECT button to switch to Frequency measurement
mode (Hz).
4. Connect the test leads in parallel to the circuit under test.
5. The LCD will display the Frequency reading.
6. Press the SELECT button again to switch to Duty Ratio
measurement mode (%).
7. The LCD will display the duty ratio reading.
continued...
www.REEDInstruments.com
www.GlobalTestSupply.com
16
Page 17
Contact Temperature Measurement
1. To avoid electric shock, before taking a contact temperature
measurement disconnect both test probes from any source of voltage.
2. Set the rotary function to the °C/°F position.
3. Insert the included Temperature Probe into the negative COM and
the positive V terminals, making sure to observe the correct polarity.
4. Press the SELECT button to change between °C and °F.
5. Place the Temperature probe onto the area being measured. Keep
the probe in place until the reading stabilizes (approx. 30 seconds).
6. The LCD will display the reading.
NOTE: The meter will display "OL" if there is an open input or
a temperature over range. To avoid electric shock, be sure the
thermocouple has been removed before changing to another function.
Non-Contact Voltage (NCV) Detection
1. Set the rotary function to the NCV position.
NOTE: Remove test leads from the meter prior to attempting NCV tests.
Always verify the NCV function on a known live circuit before performing
your tests. Do not use the NCV function if the display does not show EF
when the function switch is turned to the NCV position. If the meter does
not switch on when the NCV mode is initiated, check the batteries.
2. Place the probe near an AC voltage location. If AC voltage is
present, the meter will emit an audible beep, the LCD will display
dashes "---" and the NCV indicator will light up.
NOTE: As the AC voltage signal increases, the frequency of the beeps
and the number of dashes on the LCD screen will increase. Receptacles
in recessed sockets or differences in socket may result in false readings.
www.REEDInstruments.com
www.GlobalTestSupply.com
continued...
17
Page 18
Auto and Manual Range
When the meter is rst turned on, it automatically goes into Auto Ranging
mode. This automatically selects the best range for the measurement
being made and is generally the best mode for most applications. For
applications that require a manual range to be set, perform the following:
1. Press the RANGE button. The "Auto" range display indicator will
turn off.
2. Continuously press the RANGE to skip through the available ranges
until the required range is selected.
3. Press and hold the RANGE button for 2 seconds to exit "Manual
Ranging" mode and return to "Auto Ranging".
Relative Mode
Relative mode stores the current measurement reading displayed on the
LCD, resets the meter to zero and maintains the stored measurement
reading as a reference point for all subsequent measurements.
1. Press the REL button to enter relative mode. The indicator will
appear on the LCD along with the relative reading.
2. Press the REL button again to exit relative mode and resume
normal operation.
Note: The meter does not auto range while relative mode is enabled
and does not work under Continuity, Resistance or Diode Test
measurement modes.
Auto Power Off
1. To preserve battery life, the meter is programmed to turn off after
approx. 15 minutes of inactivity.
2. To turn this function off, press and hold the SELECT button while
powering on the meter.
3. If the meter is turned off then back on, the "Auto Power Off" feature
will be enabled again.
www.REEDInstruments.com
www.GlobalTestSupply.com
continued...
18
Page 19
MAX/MIN Measurement
1. Press the MAX/MIN button to select the maximum reading as
indicated by the "MAX" symbol. The Max value is updated when a
new maximum data value has been attained.
2. Press the MAX/MIN button again to select the minimum reading as
indicated by the "MIN" symbol. The Min value is updated when a
new minimum data value has been attained.
3. Press and hold the MAX/MIN button to exit and resume
normal operation.
Data Hold
1. While taking a measurement, press the HOLD button to freeze the
current reading on the display.
2. While in this mode, an "H" symbol will appear.
3. Press the HOLD button again to exit and resume normal operation.
Backlight
Press and hold the button to turn the LCD Backlight on and off.
Battery Replacement
1. Remove the Phillips screw on
the back of the meter.
2. Open the battery compartment.
3. Replace the 3 X AAA batteries.
4. Properly secure the cover and
tighten the screw.
1
2
Applications
• Industrial maintenance teams performing scheduled and preventative
maintenance on electro-mechanical equipment and systems.
• Facilities, building maintenance and electricians looking to troubleshoot
electrical equipment installation problems.
www.REEDInstruments.com
www.GlobalTestSupply.com
19
3
Page 20
Accessories and Replacement Parts
• R5400 Line Splitter
• R2990 Thermocouple Adapter
• R2920 Surface Thermocouple Probe
• R2930 Right Angle Thermocouple Surface Probe
• R2940 Air/Gas Thermocouple Probe
• R2950 Immersion Thermocouple Probe
• R2960 Needle Tip Thermocouple Probe
• R1020 Fused Test Lead Set
• R1000 Safety Test Lead Set
• CA-05A Medium Soft Carrying Case
Don't see your part listed here? For a complete list of all accessories and
replacement parts visit your product page on www.reedinstruments.com.
Appendix A
Flexible AC Current Probe Measurement (R5065 sold separately)
AC Current Measurement with Flexible Probe
1. Insert the exible current probe into the negative COM terminal and
the red test lead into the positive V terminal.
2. Connect the probe's exible tubing around the conductor under test.
If you are opening the end of the exible probe
to make the connection, be sure to close and lock it back into place.
3. Keep the probe coupling more than 1" (2.5cm) away from
the conductor.
www.REEDInstruments.com
www.GlobalTestSupply.com
continued...
20
Page 21
4. Turn on the exible probe by
setting the required current range.
NOTE: When measuring current,
center the conductor in the exible
current Probe. If possible, avoid taking
measurements close to other
current-carrying conductors.
R5055
1000A AC/DC Clamp Meter
Current Probe
R5065
Flex Clamp
5. Set the rotary function switch to
the
position.
6. The meter will automatically switch
to the exible current probe with
the extended current range.
7. The LCD will display the reading.
AC Current Frequency Measurement with Flexible Probe
1. Insert the exible current probe into the negative COM terminal and
the red test lead into the positive V terminal.
2. Connect the probe's exible tubing around the conductor under
test. If you are opening the end of the exible probe to make the
connection, be sure to close and lock it back into place.
3. Turn on the exible probe by setting the required current range.
4. Set the rotary function switch to the position.
5. The meter will automatically switch to the exible current probe with
the extended current range.
6. Press Hz/INRUSH button to enter the Frequency measurement mode.
7. The LCD will display the reading.
8. Press the Hz/INRUSH button again to exit the Frequency
measurement mode.
www.REEDInstruments.com
www.GlobalTestSupply.com
continued...
21
Page 22
Inrush AC Current Measurement
1. Insert the exible current probe into the negative COM terminal and
the red test lead into the positive V terminal.
2. Connect the probe's exible tubing around the conductor under test.
If you are opening the end of the exible probe to make the
connection, be sure to close and lock it back into place.
3. Turn on the exible probe by setting the required current range.
4. While the device to be tested is turned off, set the rotary function
switch to the position.
5. The meter will automatically switch to the exible current probe with
the extended current range.
6. Press and hold the Hz/INRUSH button for 2 seconds to enter the
Inrush current measurement mode.
7. Turn on the device under test.
8. The LCD will display the reading.
NOTE: Inrush current is the highest AC current (True RMS) within 100ms
of start time, as shown below.
Measured Inrush
Current
TIME
Current
Detection
Threshold
CURRENT
*
0A
100ms Sample Window (60Hz)
StartStop
www.REEDInstruments.com
www.GlobalTestSupply.com
22
Page 23
Product Care
To keep your instrument in good working order we recommend the following:
• Store your product in a clean, dry place.
• Change the battery as needed.
• If your instrument isn't being used for a period of one month or longer
please remove the battery.
• Clean your product and accessories with biodegradable cleaner. Do not
spray the cleaner directly on the instrument. Use on external parts only
Product Warranty
REED Instruments guarantees this instrument to be free of defects in
material or workmanship for a period of one (1) year from date of shipment.
During the warranty period, REED Instruments will repair or replace, at no
charge, products or parts of a product that proves to be defective because
of improper material or workmanship, under normal use and maintenance.
REED Instruments total liability is limited to repair or replacement of the
product. REED Instruments shall not be liable for damages to goods,
property, or persons due to improper use or through attempts to utilize
the instrument under conditions which exceed the designed capabilities.
In order to begin the warranty service process, please contact us by email
at 1-877-849-2127 or info@reedinstruments.com to discuss the claim and
determine the appropriate steps to process the warranty.
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling your
instrument. Your product contains electronic components and must be
disposed of separately from standard waste products.
www.REEDInstruments.com
www.GlobalTestSupply.com
23
Page 24
Product Support
If you have any questions on your product, please contact your authorized
REED distributor or REED Instruments Customer Service by phone at
1-877-849-2127 or by email at info@reedinstruments.com.
Please visit www.REEDInstruments.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice.
All rights reserved. Any unauthorized copying or reproduction of
this manual is strictly prohibited without prior written permission
from REED Instruments.
www.REEDInstruments.com
www.GlobalTestSupply.com
24
Page 25
TEST & MEASURE
WITH CONFIDENCE
CHECK OUT OUR LATEST PRODUCTS!
www.REEDInstruments.com
www.GlobalTestSupply.com
Page 26
www.REEDInstruments.com
www.GlobalTestSupply.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.