REED Instruments R5015, R5015-NIST User manual

Model
INSTRUMENTS
Instruction Manual
True RMS Mini Clamp Meter
www.GlobalTestSupply.com
INSTRUMENTS
Table of Contents
Safety ..........................................................................................................2
Features .......................................................................................................3
Specications .......................................................................................... 3-5
Electrical Symbols .......................................................................................6
Product Description ....................................................................................6
LCD Description ..........................................................................................7
Operating Instructions .......................................................................... 8-11
Battery Replacement .................................................................................11
Safety
The unit complies with CE/ETL certication (EN61010-1, EN61010-2-30, and EN61010-2-32), CAT II 600V & CAT III 300V overload protection.
Please read the manual carefully before use.
• Wear protective equipment to avoid damage caused by arc discharge when the live conductor is exposed.
• Before each use check the product carefully to ensure no damage.
• Hold the instrument by the handle when in use.
• Before changing the battery ensure unit is not connected to live wires.
• Do not use clamp meter in environments above 600V or 400Hz.
• Pay close attention to 60V DC, 30V AC or 42V AC (peak value) or above voltage as they may cause electric shock.
• Clean the instrument with a damp cloth and avoiding abrasive compounds or solvents.
• In case of any abnormality, stop using this instrument immediately.
2
www.GlobalTestSupply.com
INSTRUMENTS
Features
• Measures AC/DC current/voltage, resistance and capacitance
• High resolution 0.1mA (AC) and 1mA (DC)
• True RMS for AC current/voltage
• 6000-count backlit LCD display
• Built-in non-contact voltage detector with LED indicator
• V.F.C (variable frequency control) mode for air-conditioner maintenance and diagnostics
• Low battery and over range indicators
• Diode test and continuity check functions
• Durable double molded plastic housing
• CAT. II 600V, CAT. III 300V safety rating
• Auto shut-off
• Includes test leads, batteries and carrying case
Specifications
AC/DC Current
Range: AC: 600mA, 6000mA, 60A / V.F.C 600mA-60A DC: 6000mA, 60A
Accuracy: AC (@ 50/60Hz): 600mA ±(1.5% rdg. +10 dgt.) 6000mA ±(2.5% rdg. +5 dgt.) 60A ±(2% rdg. +5 dgt.) V.F.C ±(4% rdg. +10 dgt.) DC: 6000mA ±(2% rdg. +5 dgt.) 60A ±(2% rdg. +3 dgt.)
Resolution: AC: 0.1mA, 1mA, 0.01A / 0.1mA, 0.01A DC: 1mA, 0.01A
3
www.GlobalTestSupply.com
INSTRUMENTS
AC/DC Voltage
Range: AC: 6, 60, 600V / V.F.C 200-600V DC: 600mV, 6, 60, 600V
Accuracy: AC: ±(0.8% rdg. +3 dgt.) / ±(4.0% rdg. +3 dgt.) DC: 600mV ±(0.7% rdg. +5 dgt.) ±(0.7% rdg. +3 dgt.)
Resolution: AC: 1mV, 10mV, 0.1V, 1V / 0.1V DC: 10µV, 1mV, 10mV, 0.1V
Resistance
Range: 600Ω, 6kΩ, 60kΩ, 600kΩ, 6MΩ, 60MΩ Accuracy: 600Ω: ±(1% rdg. +2 dgt.)
6/60/600kΩ: ±(0.8% rdg. +2 dgt.) 6MΩ: ±(1.2% rdg. +3 dgt.) 60MΩ: ±(1.5% rdg. +5 dgt.)
Resolution: 1pF, 10pF, 100pF, 1nF, 10nF, 100nF, 1μF, 10μF
Capacitance
Range: 6.2nF, 62nF, 620nF, 6.2μF, 62μF, 620μF, 6.2mF, 62mF
Accuracy: 6.200nF ±(4%+10)
62.00nF~ 620.0μF ±(4%+5)
6.200mF~ 62.00mF ±10%
Resolution: 1pF, 10pF, 0.1µF, 1µF, 10µF
Frequency
Range: 10Hz ~ 60kHz Accuracy: ±(0.1%+4) Resolution: 0.001Hz ~ 0.01kHz
General Specifications
Range Selection: Autoranging True RMS: Yes Display: 6,000 count LCD display Display Hold: Yes Zero Push
Button Adjustment: Yes
4
www.GlobalTestSupply.com
INSTRUMENTS
Diode Test: Yes Backlit Display: Yes Overrange Indicator: Yes Continuity Check Audible signal if resistance <10Ω Non-Contact
Voltage Detector: Yes Autoshut off: Yes (after 15 minutes) Power Supply: 2 AAA Batteries Low Battery Indicator: Yes Jaw Opening: 0.67” (17mm) Overvoltage Category: CAT. II 600V, CAT. III 300V Product Certications: CE, ETL, RoHS Operating Temperature: 32 to 104°F (0 to 40°C) Storage Temperature: 14 to 122°F (-10 to 50°C) Dimensions: 6.9 x 2.4 x 1.3” (175 x 60 x 33.5mm) Weight: 6oz (170g) Optional accessories: Line Splitter (R5400)
Safety Test Lead Set (R1000) Fused Test Lead Set (R1020) Hard Shell Carrying Case (R9940) Small Soft Carrying Case (CA-52A)
Overload protection: 100A
• Accuracy guarantee scale: 5~100% of measuring range, <20 digit residual reading is allowed for 600mA open circuit.
• AC crest factor may reach 3.0 at 4,000 counts; for non-sinusoidal waveform, the error of crest factor increases with the following formula:
a. Add 3% when the crest factor is 1~2 b. Add 5% when the crest factor is 2~2.5 c. Add 7% when the crest factor is 2.5~3 Overload protection: 600V-PTC
5
www.GlobalTestSupply.com
INSTRUMENTS
Electrical Symbols
Product Description
1. Clamp Jaws
2. Clamp Trigger (Press the trigger to open the clamp)
3. Non Contact Voltage Indicator
4. Function Button (Change from ACV/DCV/Hz, Resistance Ω/Diode/Capacity, Current ACA/DCA, NCV, Off)
5. HOLD/Backlight Button (To keep the current measurement on the screen / Hold button for 2 seconds for backlight to turn on)
6. ZERO Button (Return DCA to zero, measure the relative value of capacity & voltage)
7. SELECT Button
8. LCD Display
9. Input Jack / COM Input Jack
6
www.GlobalTestSupply.com
INSTRUMENTS
LCD Description
7
www.GlobalTestSupply.com
INSTRUMENTS
Operating Instructions
Measuring AC/DC voltage/HZ
1. Select AC voltage and Hz or DC voltage.
2. Insert the red probe in the red jack (positive terminal) and black probe in the black jack (COM terminal).
3. Touch the red and black probe to item under test.
4. The value will appear on the LCD.
Note: Do not exceed 600V (AC/DC) as there is a risk of electric shock and potential damage to the instrument that may occur.
Measure Resistance/Diode/Circuit On/Off/Capacity
• Insert the red probe in the red jack (positive terminal) and black probe in the black jack (COM terminal).
• Connect the probe in parallel with the tested component to measure.
• The value will appear on the LCD.
Note: When measuring the voltage/capacity/diode range, do not exceed DC 60 V or AC 30V to avoid personal injury.
8
www.GlobalTestSupply.com
INSTRUMENTS
Measuring AC/DC Current
AC current
• Select your desired range for AC current (600mA~, 6,000mA~, 60A~). Open the clamp head and attach the wire on the hook. Ensure that the hooks are rmly closed. Value will display on the LCD.
DC current
• Press the SELECT button and enter in the DC current range (6,000mA, 60A). Press the ZERO button to bring the reading back to zero. Press it multiple times if pressing it once doesn’t work. Open the clamp head and attach the wire on the hook. Ensure that the hooks are rmly closed. Value will display on the LCD.
NOTE: Pull out the testing probe to avoid electric shock. The clamp should be in the same direction as the measured object in process of zeroing obtain accurate reading because of the product’s high sensitivity.
9
www.GlobalTestSupply.com
INSTRUMENTS
Measuring Non-Contact Field Measurement
To measure the detected AV voltage or magnetic eld close the clamp around the object to detect the motion. The analog quantity is “EF” when it is ≤ critical voltage and will display as “———“, different variations of buzzing will distinguish the intensity of detected eld.
Note: When the unit is in NVC measurement mode, pull out the testing probe to avoid electric shock.
Backlight
Press the HOLD button for 2 seconds to turn on the LCD backlight.
Automatically Turns Off
If the unit is left unused for 15 minutes it will automatically turn off to save power. Press the SELECT button in shutdown mode to power it on. The buzzer will make 5 sounds prompting the cancellation of the auto off function. Turn it off and then restart, the auto off function will be recovered. The buzzer will send 5 continuous alarming sounds, and then 1 long alarming sound before the product automatically shuts down. When the auto off function is canceled, it will send 5 alarming sounds every 15 minutes.
10
www.GlobalTestSupply.com
INSTRUMENTS
Functions
When pressing any function button or turning any switch the buzzer will make a “beep” noise. When the measuring voltage or current the buzzer will also send a continuous “beep” intermittent sound to warm the outrange. The functions are: AC/DC voltage >about 600V, mA AC/DC current >620mA (or 6200mA), A AC/DC large current >62A.
Battery Replacement
1. Ensure the power switch is set to “OFF” and pull the probe out of input jack.
2. Remove the battery cover with a screwdriver.
3. Install two new AAA batteries.
4. Replace the cover and screw in the battery cover.
For service on this or any other REED product or information on other REED products, contact REED Instruments at info@reedinstruments.com.
www.GlobalTestSupply.com
11