REED Instruments R5002, R5002-NIST User manual

Model
High Voltage Insulation Resistance Tester
R5002
www.GlobalTestSupply.com
Table of Contents
Safety ............................................................................................ 3
Features .........................................................................................4
Specications .............................................................................4-6
Instrument Description .............................................................7-10
Operating Instructions ............................................................11-18
AC/DC Voltage Measurement ................................................... 11
Internal Power Source Measurement ........................................ 12
Low Resistance Measurement & Continuity Check .................... 12
Measuring Insulation Resistance ............................................... 13
Auto-discharge ......................................................................... 14
Insulation Resistance Measurement Principle ............................ 14
Polarization Index and Dielectric Absorption Ratios ..............14-15
Use of Guard terminal ..........................................................15-16
Data Hold ................................................................................. 16
Peak Hold ............................................................................16-17
MAX/MIN Value ......................................................................... 17
Relative Value ............................................................................ 17
Emergency Stop ....................................................................... 18
Backlight ................................................................................... 18
Auto power off .......................................................................... 18
Battery Replacement ................................................................... 18
www.GlobalTestSupply.com
Safety
Complies with IEC61010 safety measurement requirement in pollution degree 2, overvoltage category (CAT. IV 600V) and double insulation
Do not apply more than 600V to this meter
Do not use this meter around explosive gas, vapor, or dust
Do not use this meter in a wet environment
When using the test leads, keep your ngers away from the lead contacts, behind the nger guards
Do not use this meter with any parts or cover removed
Do not come in contact the circuit under test while taking measurements
Do not use this meter if it is damaged or parts are exposed; look for cracks or missing plastic
Be careful when working above 30V rms
Discharge all circuits after measuring high voltage
Place test leads in their proper input terminals
Make sure all the test leads are rmly connected to the input terminals
Make sure the insulation resistance tester is turned off when opening the battery compartment
Do not change the battery when in a wet environment
Do not use or store this meter in high temperatures, humidity, explosive environments, and near strong magnetic elds
Dry this meter before storing if wet
When not in use, set the range selector switch to the “OFF” position and remove the test leads
Remove the batteries if the instrument is not used for a long period of time
www.GlobalTestSupply.com
Features
Insulation test range: 0.1MΩ to 60GΩ
Insulation test voltages: 500V, 1000V, 2500V, 5000V
AC/DC voltage: 0.5V to 600V
200mA Continuity
Resistance: 0.1Ω to 6kΩ
Auto-discharge function & voltage output warning function
Backlight level selector
Live circuit warning symbols plus audible warning
Live circuit detection prevents insulation test if >30V is detected
Auto-power off function & battery check
Time set measurement function
Polarization index measurement (PI)
Dielectric absorption ratio measurement (DAR)
Auto ranging with 6000 counts LCD display and bar-graph
MAX/MIN, PEAK, Relative Value, & Data Hold functions for AC/DC voltage measurements
6000 count LCD display with bar-graph
Specifications
Operating conditions: 32 to 104°F (0 to 40°C); 80% RH Storage conditions: 14 to 140°F (-10 to 460°C); <80% RH Sampling rate: 2.5 readings per second Fuse: 500mA/600V (6 x 32mm) ceramic 3AG fast blow Zero adjust: Automatic Auto power off: 20 min Power supply: 8 C Batteries or AC Adapter (included) Dimensions: 7.8 x 5.8 x 3.4" (198 x 148 x 86mm) Weight: 3lbs (1.5kg)
continued ...
www.GlobalTestSupply.com
Includes: One black alligator test lead, one green alligator test lead, one red alligator test lead, one red 2-prong alligator test lead, batteries, and a tool box
DC/AC Voltage Measurement
Range: 0.5-600.0V Resolution: 0.1V Accuracy: AC: +1.0% reading + 5 digits (40 to 60Hz)
+2.5% reading + 10 digits (61 to 400Hz) DC: +1.0% reading + 5 digits
Insulation Resistance Measurement
Rated voltage 500V 1000V 2500V 5000V
0.005-6MΩ 0.005-6MΩ 0.05-60MΩ 0.05-60MΩ
Measuring range (Auto-ranging)
Open circuit voltage
Rated current
Short-circuit current: Approx 1mA Accuracy: +2.5% reading + 15 digits (at 0.005 to 600.0MΩ);
+3% reading + 15 digits (at 0.61 to 6.00GΩ); +4% reading + 15 digits (at 6.1 to 60.0GΩ)
Voltage monitor range: 5 to 6000VDC (resolution 1V) Accuracy: +1.5% reading + 5 digits
6.01-60MΩ 6.01-60MΩ 60.1-600MΩ 60.1-600MΩ
60.1-600MΩ 60.1-600MΩ 0.61-6GΩ 0.61-6GΩ
0.61-6GΩ 0.61-6GΩ 6.1-60GΩ 6.1-60GΩ DC 500V
+20%, -0%
1-1.2mA (at
0.5MΩ load)
DC 1000V +20%, -0%
1-1.2mA (at 1MΩ load)
DC 2500V +20%,-0%
1-1.2mA (at
2.5MΩ load)
DC 5000V +20%,-0%
1-1.2mA (at 5MΩ load)
www.GlobalTestSupply.com
continued ...
Low Resistance Measurement & Continuity Check
Range Resolution Accuracy
Resistance (Auto-ranging)
Continuity Buzzer: ≤35Ω Short circuit Current: ≥200mA Open Circuit Test Voltage: ≥4.5V
0.1 to 600Ω 0.1Ω
6.01 to 6kΩ 0.001kΩ
+1.5% reading + 10 digits
+1.5% reading + 15 digits
For service on this or any other REED product or information on other REED products, contact REED Instruments at info@reedinstruments.com
www.GlobalTestSupply.com
Instrument Description
1. Insulation resistance test/lock button
2. Relative/Time-down button
3. Data hold button
4. Internal battery check/time-up button
5. MAX/MIN & Test button
6. LCD display
7. Power adaptor input terminal
8. EARTH: high resistance measurement input
9. V/Ω input terminal & GUARD: grounding protection input terminal
10. COM & high voltage line shielding input terminal
11. LINE: 500 to 5000V high voltage output terminal
12. PI/DAR, peak voltages measurement & time clear button
13. Backlight levels button
14. Function switch
15. Test leads
www.GlobalTestSupply.com
Input Terminals
Insulation Resistance Measurement
LINE: 500V to 5000V range insulation resistance measurement/high voltage line output terminal (LINE + COM: two plugs red test lead to one alligator clip)
COM: Return terminal for all measurements (LINE + COM: two prong red test lead)
GUARD: Grounding protection input terminal (black test led) EARTH: Insulation resistance measurement input terminal
(green test lead)
ACV/DCV/Low Resistance Measurements & Continuity Check
V/Ω: Input for voltage, low resistance measurements & continuity check (red test lead)
COM: Return terminal for all measurements (black test lead)
www.GlobalTestSupply.com
Front Panel
HOLD: Freezes current reading Backlight: Turns the backlight on, makes it brighter, and turns it off
Insulation Resistance Measurement
TEST/LOCK: Press and hold to test or lock on insulation resistance measurement
Time-up: Adjust timer up Time-down: Adjust timer down PI: Polarization index measurement DAR: Dielectric absorption ratio measurement STOP: Stop insulation resistance measurements
and high voltage output
AC/DC Measurements
MAX/MIN: MAX/MIN record PEAK: PEAK-MAX/PEAK-MIN capture REL: Relative measurement BATT: Check battery power
www.GlobalTestSupply.com
LCD Display
1. Auto-ranging indicator
2. Battery life indicator
3. High voltage output or >30V AC/DC voltage indicator
4. Indicator for automatically discharged
5. Continuity buzzer indicator
6. Analogue bar graph
7. Unit symbols
8. Function indicators
9. Clock
10. Timer 1 & Timer 2 indicators
11. Active timer symbol
12. Indicates for negative reading
13. AC/DC indicators
14. Polarization index
15. Dielectric absorption ratio
16. Secondary reading display
17. Check feature pass or fail
18. Primary reading display
www.GlobalTestSupply.com
10
Operating Instructions
After pressing the TEST button under insulation measurement, the wire will generate high voltage. Do not touch it to avoid electric shock.
AC/DC Voltage Measurement
To avoid electric shock, do not test if voltage to ground AC/DC is more than 600V, even if the line voltage is below 600V.
1. Set the rotary switch to the AC or DC position.
2. Connect the red test lead to the VΩ terminal and the black test lead
to the COM terminal.
3. Connect the test leads to the circuit under test.
4. Read the voltage value on the LCD.
5. When measuring DC voltage, if the red test lead is negative
voltage, “-” will appear on the display.
www.GlobalTestSupply.com
11
Internal Power Source Measurement
1. Disconnect the test leads.
2. Set the switch to DC.
3. Press the BATT button.
4. Read the voltage value
on the LCD.
5. After 5 seconds the meter
will auto-off or press the BATT button to turn off the battery voltage function.
Low Resistance Measurement & Continuity Check
Do not run this test unless AC/DC = 0V. Do not use this mode to check diodes.
1. Set the rotary switch to the Ω position.
2. Connect the red test lead
to the VΩ terminal and the black lead to the COM terminal.
3. Connect the tips of
the test leads to the circuit under test.
4. Read the resistance
value on the display.
5. When the resistance
of a circuit is less than 50Ω, the audible tone will sound.
www.GlobalTestSupply.com
12
Measuring Insulation Resistance
Prior to testing check the circuit with a high voltage detector to be sure that there is no electrical charge
It is highly recommended to wear insulated gloves while taking high voltage measurements
Do not make measurements with the battery compartment open
Do not take measurements during thunderstorms
Be sure to connect the black Earth Cord to the Earth terminal of the circuit under test
Measurements cannot be made if the “>30V” live circuit warning appears or if “AUTO DISCHARGE” appears on the LCD screen along with a warning buzzer
1. Check the amount of voltage that can be applied to the circuit
under test, and set the range switch to the desired insulation resistance range.
2. Connect the black Earth cord to the Earth terminal on the circuit
under test.
3. Touch circuit with the tip of the red line probe and press the
“PRESS TO TEST” button. The buzzer will sound intermittently during measurement.
4. The measured value will display on the LCD, and it is automatically
held after measurement.
5. At any output voltage, when the tested resistance is less than
10MΩ, the testing time cannot exceed 10 seconds continuously.
Note: Do not touch the circuit under test immediately after testing. Capacitance stored in the circuit may cause electric shock. Leave test leads connected to the circuit and never touch the circuit until the discharge is complete.
www.GlobalTestSupply.com
13
Auto-discharge
This function releases the capacitance stored in a circuit after testing. The LCD will ash “AUTO DISCHARGE” when the meter is in discharge condition and the live voltage monitor will display “O V”. This function will be released by removing test leads 2 sec or more before discharge is complete. Set the range switch to “OFF” position, and remove test leads from the instrument.
Insulation Resistance Measurement Principle
Insulation tests should only be performed on de-energized circuits.
Continuous Insulation Resistance Measurement
Press and turn the “PRESS TO TEST” button clockwise to perform a continuous insulation resistance measurement. After testing, turn the button counterclockwise to the initial position. A beep will sound and the reading will hold on the LCD for 10 seconds.
Timer Insulation Resistance Measurement
This is a function to automatically conduct a test at a set time (from 1 min to 15 min).
1. Press the Time-up and Time-down buttons to set the time from 1
min to 15 min.
2. To activate the timed insulation resistance measurement, press and
turn the “PRESS TO TEST” button clockwise till it locks.
3. After testing, turn the button counterclockwise to the initial
position. A beep will sound and the reading will hold in the LCD for 10 seconds.
Polarization Index and Dielectric Absorption Ratios
Polarization Index (PI) is the ratio of a 10-minute to 1-minute insulation resistance. Polarization index = resistance value measured after 10 min (Time2)/resistance value measured after 1 min (Time1). Dielectric Absorption Ratio (DAR) is the ratio of a 1-minute to 30-second insulation resistance. Dielectric Absorption Ratio = resistance value measured after 1 min (Time2)/resistance value measured after 30 sec (Time1).
continued ...
www.GlobalTestSupply.com
14
1. Press the PI/DAR button to select PI or DAR Measurement.
2. The LCD will display PI & Time1/Time2 or DAR & Time1/Time2.
3. To begin measuring, press and turn the “PRESS TO TEST” button
clockwise till it locks.
4. After testing, turn the button to counterclockwise to the initial
position. A beep will sound and will display “PASS if the reading is PI/DAR>4, or “FAIL” if the reading is PI/DAR<1. The reading will hold for 10 seconds.
PI standard
PI value 2 to 4 (typically3): Considered as good insulation (older types) PI value 1 to 1.5: Not acceptable (older types) PI value 1: (Very high insulation resistance)
Modern type of (good) insulation systems PI value 1.0 or less: Fail Example: If measuring a B-class insulation material and the polarization
index is 1.1, the insulating material contains too much water or is seriously polluted. The polarization index is below the minimum acceptable value and you need to repair or replace parts.
DAR standard
DAR value 1.6 or more: Excellent DAR value 1.25 to 1.6: PASS DAR value <1.25 or less: Fail
Use of Guard terminal
When measuring a cable, the leakage current owing on the surface of cable jacket and the current owing inside the insulator combine and may cause error in value. To prevent this error, wind a conductive wire around the point where leakage current ows, then connect it to the Guard terminal as shown in the gure below.
www.GlobalTestSupply.com
continued ...
15
This removes the surface leakage resistance of the cable insulation so you only measure the volume resistance of the insulator. Make sure to use the Guard cord supplied with this instrument to connect the instrument to the Guard terminal.
Data Hold
This function freezes the reading on the display. Press the HOLD key momentarily to activate or to exit the hold function.
Peak Hold
The Peak Hold function captures the peak AC/DC Voltage from 0.5V to 600V, and can capture peaks from 10 to 100 milliseconds in duration.
1. Press the PEAK button. The symbol “PEAK” will appear on the
LCD.
2. Press the MAX/MIN button to activate the MAX/MIN recording
mode. The symbol “MAX & PEAK” will appear on the LCD. The screen will display and hold the maximum reading and will update only when a new “MAX” reading occurs.
3. Press the MAX/MIN button again and “MIN &PEAK” will appear on
the screen. The LCD will display and hold the minimum reading and update only when a new “MIN” reading occurs.
www.GlobalTestSupply.com
continued ...
16
4. Press the MAX/MIN button again, and a “MAX MIN PEAK” symbol
will appear on the LCD. The meter will display the present reading, but will continue to update and store the max and min readings.
5. To exit MAX/MIN-PEAK mode press the PEAK button or hold the
MAX/MIN button for 2 seconds.
MAX/MIN Value
Only for AC/DC voltage measurement
1. Press the MAX/MIN button to activate the MAX/MIN recording
mode. The symbol “MAX” will appear on the screen. The LCD will display and hold the maximum reading and will update only when a new “MAX” reading occurs.
2. Press the MAX/MIN button again and “MIN” will appear on the
screen. The meter will display and hold the minimum reading and will update only when a new “MIN” reading occurs .
3. Press the MAX/MIN button and “MAX MIN” will appear on the
screen. The meter will display the present reading, but will continue to update and store the max and min readings.
4. To exit MAX/MIN mode press and hold the MAX/MIN buttons for 2
seconds.
Relative Value
(Only for AC/DC voltage measurement) The relative measurement feature allows you to take AC/DC voltage measurements relative to a stored reference value. The displayed value is the difference between the reference value and the measured value.
1. Press the REL button to store the reading currently on display. The
“REL” symbol will appear on the display.
2. The display will now show the difference between the stored value
and the measured value.
3. Press the REL to return to normal operation.
www.GlobalTestSupply.com
17
Emergency Stop
If you are taking high-voltage insulation resistance measurement and the danger of a short-circuit arises, stop testing immediately by pressing the “TEST” button.
Backlight
Press the Backlight button to select the display backlight levels and to turn the backlight on and off. The backlight will automatically turn off in 60 seconds.
Auto power off
This meter enters Sleep Mode and blanks the display if left inactive for 20 minutes. To exit Sleep Mode, press any button or turn the meter ON again.
Battery Replacement
When the Battery symbol appears on the LCD screen, it is time to replace the batteries.
1. Turn the insulation resistance tester to the OFF position and
remove all connections from the terminals
2. Remove the screw from the battery compartment, and separate the
battery compartment from the case bottom
3. Replace with 8 new 1.5V (LR14) batteries
4. Reattach the case bottom and battery compartment, and reinstall
the screw
www.GlobalTestSupply.com
18
Loading...