REED Instruments R3510 User guide

Page 1
Long Stem pH Meter
Instruction Manual
R3510
1-877-849-2127 | [email protected] | www.reedinstruments.com
REED Instruments
Page 2
Introduction ................................................................................................ 3
Product Quality Safety
......................................................................................................... 4
Features Included Specications Instrument Description Display Description Operating Instructions
Power ON/OFF ...................................................................................... 9
Electrode Preparation and Usage Instructions
Testing with the pH Meter
Auto Power Off
pH Calibration Procedure ............................................................... 11-12
Troubleshooting Tips
Power On but No Display .................................................................... 13
Slow Response
Display Shows "- - -" and ATC Indicator
pH Value Fluctuates Rapidly
Battery Replacement ................................................................................ 14
Applications Accessories and Replacement Parts Product Care Product Warranty Product Disposal and Recycling
Product Support
........................................................................................... 3
...................................................................................................... 5
...................................................................................................... 5
............................................................................................. 6
............................................................................... 7
.................................................................................... 8
........................................................................... 9-12
...................................... 9
................................................................... 10
.................................................................................9-10
................................................................................ 13
.................................................................................... 13
............................................. 13
................................................................ 13
.............................................................................................. 15
....................................................... 15
............................................................................................ 16
..................................................................................... 17
.............................................................. 17
....................................................................................... 17
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 3
Introduction
Thank you for purchasing your REED R3510 Long Stem pH Meter. Please read the following instructions carefully before using your instrument. By following the steps outlined in this manual your meter will provide years of reliable service.
Product Quality
This product has been manufactured in an ISO9001 facility and has been calibrated during the manufacturing process to meet stated product specications. If a certicate of calibration is required please contact the nearest authorized REED distributor or authorized Service Center. Please note an additional fee for this service will apply.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 4
Safety
Never attempt to repair or modify your instrument. Dismantling your product, other than for the purpose of replacing batteries, may cause damage that will not be covered under the manufacturer's warranty. Servicing should only be provided by an authorized service center.
WARNING
• Remove and immediately recycle or dispose of used batteries according to local regulations and keep away from children. Do NOT dispose of batteries in household trash or incinerate.
• Even used batteries may cause severe injury or death.
• Call a local poison control center for treatment information.
• Identication of compatible battery type: (2) CR2032
• Nominal battery voltage: 3V.
• Non-rechargeable batteries are not to be recharged.
• Do not force discharge, recharge, disassemble, heat above 140°F (60°C) or incinerate. Doing so may result in injury due to venting, leakage or explosion resulting in chemical burns.
• Ensure the batteries are installed correctly according to polarity (+ and -).
• Do not mix old and new batteries, different brands or types of batteries, such as alkaline, carbon-zinc, or rechargeable batteries.
• Remove and immediately recycle or dispose of batteries from equipment not used for an extended period of time according to local regulations.
•
Always completely secure the battery compartment. If the battery compartment does not close securely, stop using the product, remove the batteries, and keep them away from children.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 5
Features
• Large, clear LCD display
• Long 4.7" (119.4 mm) pH probe
• Easy one-touch auto-calibration
• Data hold function
• Waterproof housing
• Protective cap keeps sensor moistened
• Low battery indicator and auto shut off
Included
• pH Meter
• Electrode Storage Bottle
• Batteries
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 6
Specifications
pH
Measuring Range: 0.00 to 14.00 pH Accuracy: ±0.2 pH Resolution: 0.01 pH
Probe Specifications
Reference: Single, Ag/AgCl Junction: Ceramic Electrolyte: Gel, 3.3M KCl Tip Shape: Round Probe Diameter: 0.5" (12mm) Probe Length: 5.11" (130mm)
General Specifications
pH Calibration: Yes (3-point) Display: LCD Data Hold: Yes LCD Size: 1.14 x 0.89" (29 x 22.5mm) Auto Shut-Off: Yes (after approximately 20 minutes / Off) Low Battery Indicator: Yes Power Supply: 2 x CR2032 Batteries Battery Life: 100 hours of continuous use Product Certications: CE, IP54 Operating Temperature: 32 to 122°F (0 to 50°C) Operating Humidity Range: 10 to 80% Storage Temperature: 14 to 122°F (-10 to 50°C) Storage Humidity Range: 10 to 80% Dimensions: 3.7 x 1.4 x 0.79" (94 x 36 x 20mm) Weight: 4.05oz (115g)
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 7
Instrument Description
1
2
3
4
5
1. LCD Display
2. HOLD Button
3. POWER Button
4. Electrode
5. Electrode Storage Bottle
6. CAL Button
6
1-877-849-2127 | [email protected] | www.reedinstruments.com
REED Instruments
Page 8
Display Description
4
1
2
1. Hold Status Indicator
2. pH Measurement Value
3. Battery Status Indicator
4. Calibration Mode Indicator
3
1-877-849-2127 | [email protected] | www.reedinstruments.com
REED Instruments
Page 9
Operating Instructions
Before use, soak the glass bulb in distilled water for a minimum of 30 minutes. In order to obtain the most accurate results possible, a pH meter should be calibrated at least once a day.
Power ON/OFF
Turn the meter on or off by pressing the "Power" button.
Electrode Preparation and Usage Instructions
Remove the electrode storage bottle loosened, pull the bottle off completely to expose the electrode.
Note: It's normal to nd white crystals on the cap or electrode assembly.
by rotating it counterclockwise. Once
Important: For pH meters, always ensure the storage bottle is lled with storage buffer solution. This keeps the electrode wet and in optimal condition for storage.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 10
Testing with the pH Meter
1.
Dip the exposed electrode into the test solution. Press the "Power" button
to turn on the meter. Gently stir the solution to achieve a
stable reading.
2. While in measurement mode, a small dot "•" will ash on the display,
indicating the meter is actively taking readings.
3. To freeze the current reading, press the button. The "Hold" icon will appear on the screen, and the small dot will stop ashing.
4. Press the button again to exit the hold mode.
5. Once nished, clean the electrode, replace the storage bottle, and store the meter in an environment between 0–50°C.
Auto Power Off
To preserve battery life, the meter is programmed to turn off after approximately 20 minutes of inactivity.
To disable the auto power-off function:
1. Before turning on the device, press and hold the and buttons simultaneously.
2. Keep holding the buttons until the letter "n" appears on the screen.
Once "n" is displayed, release the keys to exit and return to normal operation.
continued...
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 11
Note: Disabling sleep mode will reset after every power-off, so this process must be repeated each time the device is powered on.
pH Calibration Procedure
To obtain the most accurate results possible, a pH meter should be calibrated at least once a day.
If the pH meter is stored for an extended period of time, prior to calibration, condition your electrode in pH 4 buffer solution or distilled water for a minimum of 1 hour.
1. Ensure the pH meter is turned on and ready for use.
2.
Immerse the electrode into a buffer solution (choose from pH 4, 7, or 10).
3.
Start with the pH 7 buffer solution, as this is typically calibrated rst for accuracy. Follow with pH 4 or pH 10 to complete the calibration process.
4. Press the button to enter the pH calibration mode.
5. The "CAL" icon will briey appear on the screen, and the pH value of the buffer (4, 7, or 10) will be displayed.
Note: If there is an issue (e.g., incorrect buffer, damaged probe, or inability to detect the correct voltage range), the meter will exit calibration mode automatically after 10 seconds. The message "End" will briey appear on the screen before returning to normal operation.
continued...
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 12
Acceptable Voltage Ranges:
pH 4.00: 97mV to 250mV pH 7.00: -60mV to 60mV pH 10.00: -250mV to -97mV
6. If the probe correctly recognizes the buffer solution, the corresponding pH value (4, 7, or 10) will appear on the display within 2 seconds.
7. If your buffer solution has a slightly different pH (e.g., 7.01), press the button to adjust the value.
The adjustable calibration Point Range are as follows:
For pH 4.0: The adjustable range is 3.50 to 4.50. For pH 7.0: The adjustable range is 6.50 to 7.50. For pH 10.0: The adjustable range is 9.50 to 10.50.
8. Once the electrode provides a stable reading and no buttons are pressed, the meter will automatically save the calibration value and exit calibration mode.
9. Rinse the probe with de-ionized water or a suitable rinse solution (e.g., tap water) after each measurement to extend the life of the meter and ensure accurate readings.
10. Repeat the above steps for all three buffer solutions (pH 4, 7, and 10) to ensure accurate 3-point calibration.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 13
Troubleshooting Tips
Power On but No Display
1. Verify that the batteries are correctly inserted, making proper contact with the terminals, and are in the correct polarity.
2. If the issue persists, replace the old batteries with new ones and try again.
3. Remove the batteries for at least one minute, then reinsert them and power on the device.
Slow Response
To improve response time:
1.
Clean the probe by immersing the electrode in tap water for 10–15 minutes.
2. After soaking, rinse the probe thoroughly with distilled water.
3. Alternatively, you can use a general-purpose electrode cleaner for more effective cleaning.
Display Shows "- - -" and ATC Indicator
This indicates that the pH reading is out of range. The solution being tested may be either too acidic or too alkaline for the device to measure.
pH Value Fluctuates Rapidly
Rapid uctuations are normal if the electrode is exposed to air. Ensure the electrode is properly immersed in the solution to get stable readings.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 14
Battery Replacement
The batteries must be replaced when the low battery icon appears on the LCD.
1. Ensure the meter is turned off before proceeding.
2. Using a coin, loosen the battery cover by turning it clockwise. Important: Do not remove or misplace the black O-ring.
3. Replace the old batteries with two new CR2032 button cells.
4. Conrm the batteries are securely in place with the correct polarity.
5. Secure the battery cover back into place and tighten it by turning it counterclockwise until it locks.
WARNING
occur • A swallowed button cell or coin battery can cause Internal Chemical Burns in as little as 2 hours • Keep new and used batteries OUT OF REACH OF CHILDREN • Seek immediate medical attention
if a battery is suspected to be swallowed or inserted inside any part of the body • For treatment information call: 1-800-498-8666.
INGESTION HAZARD • DEATH or serious injury can
Note: After replacing the batteries, it is necessary to re-calibrate the meter
for accurate readings. If you do not intend to use the instrument for a month or longer, remove the batteries to avoid damage caused by leakage. Do not leave batteries in unused instruments.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 15
Applications
• Water Conditioning
• General purpose pH measurement
• Aquariums
• Wastewater monitoring
• Beverage Production
Accessories and Replacement Parts
• R1404 Buffer Solution, 4.01 pH
• R1407 Buffer Solution, 7.00 pH
• R1410 Buffer Solution, 10.00 pH
• R1420 Electrode Storage Solution
• R1425 Electrode Cleaning Solution
• CA-52A Soft Carrying Case
Don't see your part listed here? For a complete list of all accessories and replacement parts visit your product page on www.REEDInstruments.com.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 16
Product Care
To keep your instrument in good working order we recommend the following:
• Store your product in a clean, dry place.
• Change the battery as needed.
• If your instrument isn't being used for a period of one month or longer
please remove the battery.
• Clean your product and accessories with biodegradable cleaner. Do not
spray the cleaner directly on the instrument. Use on external parts only.
•
Before rst use or after storage, you may notice white KCl crystals forming outside your electrode. This will not interfere with measurements. Simply rinse the electrode.
• Always keep the pH glass bulb wet by using the cap to protect and
store the electrode.
• Always rinse the pH electrode in de-ionized water or rinse solution
before next use.
• Never touch or rub the glass bulb.
• Make sure the electrode is clean. Between measurements, rinse the
electrode with de-ionized water. If the electrode has been exposed to a solvent immiscible with water, clean it with a solvent miscible with water e.g. ethanol or acetone and rinse carefully with water.
• Store the electrode carefully using storage solution (If not available
use pH4 solution).
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 17
Product Warranty
REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one (1) year from date of shipment. During the warranty period, REED Instruments will repair or replace, at no charge, products or parts of a product that proves to be defective because of improper material or workmanship, under normal use and maintenance. REED Instruments total liability is limited to repair or replacement of the product. REED Instruments shall not be liable for damages to goods, property, or persons due to improper use or through attempts to utilize the instrument under conditions which exceed the designed capabilities. In order to begin the warranty service process, please contact us by phone at 1-877-849-2127 or by email at [email protected] to discuss the claim and determine the appropriate steps to process the warranty.
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling your instrument. Your product contains electronic components and must be disposed of separately from standard waste products.
Product Support
If you have any questions on your product, please contact your authorized REED distributor or REED Instruments Customer Service by phone at 1-877-849-2127 or by email at [email protected].
Please visit www.REEDInstruments.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice. All rights reserved. Any unauthorized copying or reproduction of this manual is strictly prohibited without prior written permission from REED Instruments.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com
Page 18
TEST & MEASURE
WITH CONFIDENCE
CHECK OUT OUR LATEST PRODUCTS!
www.REEDInstruments.com
1-877-849-2127 | [email protected] | www.reedinstruments.com
REED Instruments
Loading...