Thank you for purchasing your REED R2900 Dry Block Calibrator. Please
read the following instructions carefully before using your instrument. By
following the steps outlined in this manual your meter will provide years of
reliable service.
Product Quality
This product has been manufactured in an ISO9001 facility and has been
calibrated during the manufacturing process to meet the stated product
specications. If a certicate of calibration is required please contact the
nearest authorized REED distributor or authorized Service Center. Please
note an additional fee for this service will apply.
Safety
• Never attempt to repair or modify your instrument.
• Dismantling your instrument other than for the purpose of replacing
fuses may cause damage that will not be covered under the
manufacturer's warranty. Servicing should only be provided by an
authorized Service Center
• Use this instrument only for temperature calibration as specied in the
instruction manual.
• Operate indoors and in environments listed in the instruction manual.
• Inspect the instrument for damage before each use. Do not use if
damaged or malfunctioning.
• Handle the instrument and thermometer probes with care to prevent
drops, shocks, or overheating.
• Only trained personnel should operate this equipment.
• Before initial use, after transport, or after storage in humid conditions,
energize the instrument for 2 hours to ensure it meets IEC 1010-1 safety
requirements. Remove any moisture beforehand.
• Allow at least 6 inches (15 cm) of clearance around the instrument for
air circulation.
Temperature Range: 91 to 572°F (33 to 300°C)
Accuracy: ≤212°F (≤100°C): ±1.6°F (0.8°C)
212 to 392°F (100 to 200°C):
±3.2°F (1.6°C) 392 to 572°F
(200 to 300°C): ±5.6°F (2.8°C)
Stability: ≤212°F (≤100°C): ±0.2°F (0.1°C)
Resolution: 0.1°F (0.1°C)
Heating Time: Approx. 30 mins from 91 to 572°F
(33 to 300°C)
Cooling Time: Approx. 30 mins from 572 to 122°F
(300 to 100°C)
Well Depth: 3.54" (90mm )
Aperture Diameter: Ø3.5mm(2), Ø4.2mm, Ø5mm, Ø6.8mm
Display: Dual LED
Power Supply: 110VAC, 2.5A
Product Certications: CE
Operating Temperature: 41 to 95°F (5 to 35°C)
Storage Temperature: -4 to 140°F (-20 to 60°C)
Operating Humidity Range: 15-80%
Dimensions: 7.1 x 4.5 x 8.8" (180 x 114 x 233mm)
Weight: 6.6lbs (3kg)
212 to 392°F (100 to 200°C): ±0.4°F (0.2°C)
392 to 572°F (200 to 300°C):±0.8°F (0.4°C)
Note: The fan inside the calibrator has two speeds and runs continuously
when the unit is in operation. The fan runs slow for heating and
maintaining operation and runs fast for rapid cooling. The area around the
calibrator must be kept clear to allow adequate ventilation. The airow is
directed out the front and can be extremely hot.
Status Indicator Description
Note: Using any other parameter other than the ones described in this
manual will affect performance and could damage the calibrator.
AT
OUT
ALM1
ALM2
°F/°C
Auto-Tuning Parameter Adjustment (For Authorized Service
Centers Only)
Heating Output Power Indicator
Overload Alarm indicates that the Calibrator temperature
exceeds the set temperature. To set the alarm, refer to
Setting Overload Alarms 1 & 2 for details.
Note: When triggered the heating output power is turned off.
Overload Alarm indicates that the Calibrator temperature
exceeds the set temperature. To set the alarm, refer to
Setting Overload Alarm 1 & 2 for details.
Note: When triggered, the rapid cooling process will begin.
Temperature unit of measure.
The four-button keypad allows for easy set-point temperature. The
temperature set-point can be set to a resolution of 0.1°F or °C.
The SET button is used to conrm the selected value within
the parameter and to exit the set up mode and resume
SET
normal operation.
Note: At any time, you can press the SET button to exit the
Setup mode and resume normal operation.
The enter button is used to scroll through parameters.
The up arrow is used to increase temperature values and to
toggle between menu options within the selected parameter.
The down arrow is used to decrease temperature values and to
toggle between menu options within the selected parameter.
Operation Instructions
Initial Start up
1. Place the calibrator on a at surface with at least 8" of free space
around the instrument.
2. Connect the included power cord and plug into a power outlet.
3. Turn the calibrator on by toggling the power switch to ON.
4. The fan should begin blowing air through the instrument.
5. After a brief 3-second self-test, the calibrator will begin normal
operation by displaying both the current target surface temperature
and the last registered set temperature.
Note: When the calibrator is rst powered ON, the default set temperature
is set to 50°C. Refer to “Setting the Temperature Set-Point” for details.
6.
The calibrator will now heat up to reach the registered set temperature.
Note: The top display (Green LED) will continue to update the current
temperature until the set temperature is reached.
It is recommended to select a set-point less than 140°F (60°C) and allow
the unit to cool down before turning it off.
Warning: Do not turn the instrument off at a temperature higher than
212°F (100°C). Lower the temperature below 212°F (100°C)
before powering OFF.
Selecting the Temperature Unit of Measure
1. To unlock, hold down both the SET and buttons simultaneously
and release when the display indicates ‘’At’’ & ‘’oFF’’ conrming that
the parameters are now unlocked.
2. Press the SET button for three seconds and release to enter the
setup mode as indicated by and
3. Press button once and the display will indicate and
which conrms the current temperature scale.
4. Use the and buttons to select between °C and °F which will
ash until conrmed.
5. Press the SET button to conrm selection.
6. Press the SET button again to exit set up mode and resume
normal operation.
Setting the Temperature Set-Point
1. Use the and buttons to increase or decrease the temperature
set-point value.
2. Press the button to increase the temperature value.
3. Press the button to decrease the temperature value.
Note: The values will adjust by 0.1° at every click and will adjust by 1°
if pressed and held.
4. Press the SET button to conrm the new temperature set-point.
5. Press the SET button again to exit set up mode and resume
normal operation.
The calibrator allows a user to lock/unlock the keypad to avoid unwanted
changes to the settings. Follow the steps below to lock or unlock the
keypad.
1. Press the button four times until the screen displays
& .
2. Use the and buttons to select between LoC1, LoC2 & oFF
which will ash until conrmed.
Note: LoC1 will lock all of the keypad buttons, while LoC2 will only
lock the button allowing the user to continue to modify/conrm the set
temperature if required.
3. Press the SET button to conrm selection.
4. Press the SET button again to exit set up mode and resume
normal operation.
Replacing the Fuses
The calibrator must be switched off before attempting to replace fuses.
To replace fuses:
1. Open the faulty fuse compartment by rotating it counter clockwise.
2. Gently slide out the fuse.
3. Replace with the appropriate fuse (see Specications below) and
tighten the compartment back into place.
Fuse Specifications
Fuse 1
Fuse 2
2.5A/250V
200mA/250V
This instrument is equipped with operator-accessible fuses. If a
fuse blows, it may be due to a power surge or failure of an internal
component. Replace the fuse once. If the fuse blows again, it is likely
caused by failure of an internal component, and you should contact an
Authorized REED Service Center.
REED Instruments guarantees this instrument to be free of defects in
material or workmanship for a period of one (1) year from date of shipment.
During the warranty period, REED Instruments will repair or replace, at no
charge, products or parts of a product that proves to be defective because
of improper material or workmanship, under normal use and maintenance.
REED Instruments total liability is limited to repair or replacement of the
product. REED Instruments shall not be liable for damages to goods,
property, or persons due to improper use or through attempts to utilize the
instrument under conditions which exceed the designed capabilities. In
order to begin the warranty service process, please contact us by phone
at 1-877-849-2127 or by email at [email protected] to discuss
the claim and determine the appropriate steps to process the warranty.
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling
your instrument. Your product contains electronic components and
must be disposed of separately from standard waste products.
If you have any questions on your product, please contact your authorized
REED distributor or REED Instruments Customer Service by phone at
1-877-849-2127 or by email at [email protected].
Please visit www.REEDInstruments.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice.
All rights reserved. Any unauthorized copying or reproduction of this
manual is strictly prohibited without prior written permission from
REED Instruments.