Thank you for purchasing your REED R2815 Temperature Calibrator.
Please read the following instructions carefully before using your
instrument. By following the steps outlined in this manual your meter will
provide years of reliable service.
Product Quality
This product has been manufactured to meet the stated product
specications. If a certicate of calibration is required please contact the
nearest authorized REED distributor or authorized Service Center. Please
note an additional fee for this service will apply.
Safety
Never attempt to repair or modify your instrument. Dismantling your
product, other than to replace batteries, may cause damage that will not
be covered under the manufacturer's warranty. Servicing should only
be provided by an authorized service center. To avoid injury to the user
or damage to the instrument, please read the safety information below
before initial use:
• Do not operate the instrument around ammable or explosive gas,
vapor or dust.
• Never apply more than 30V between any two terminals, or between any
terminal and ground terminal.
Note: For optimal accuracy, allow the instrument to warm up 5 minutes
before operating. If the automatic reference-junction temperature
compensation of the instrument deviates from its designed accuracy,
contact an Authorized REED Service Center.
Display: 6-Digit LCD
Backlit Display: Yes
Kick Stand: Yes
Magnetic Hanger Compatible: Yes (R5900 sold separately)
Built-In Flashlight: Yes
Zero Adjustment Button: Yes
Power Supply: 3 AA Batteries
Battery Life: Approx. 20 hours
Cold Junction
Compensation: Yes
Auto shut off: Yes (up to 60 minutes/off)
Low Battery Indicator: Yes
Replaceable Test Leads: Yes
Product Certications: CE
Operating Temperature: 32 to 122°F (0 to 50°C)
Operating Humidity Range: 0 to 85%
Storage Temperature: 14 to 122°F (-10 to 50°C)
Dimensions: 7.5 x 3.5 x 2.1" (191 x 90 x 53mm)
Weight: 17.6oz (500g)
Press the Power button to turn the instrument on. To turn the instrument
off, press and hold the Power button.
Auto Power-Off
As a default, the instrument will automatically turn off after 5 minutes of
inactivity. To turn off this feature, refer to the "Enabling/Disabling Auto
Power OFF" section of this manual for additional details.
Backlight
Press the Backlight button to turn the LCD Backlight on or off.
Note: The backlight automatically turns off after approximately 60
seconds by default. To turn it back on, press the Backlight button. The
backlight illumination duration can be adjusted in the factory settings.
For more information, refer to the "Setting Backlight Time" section of this
manual for additional details.
The output terminal of the instrument can produce DC voltages or can
simulate thermocouple temperature set by the user.
To avoid electric shock, ensure that the voltage applied between the
calibrator terminals or between any terminal and ground does not
exceed the rated voltage specied for the calibrator. Do not use the
calibrator in situations where the terminal-to-ground voltage exceeds
30V peak.
Connecting Leads to Output Terminals
Connection method for all outputs:
1. Connect the black lead to the COM terminal and the red lead to the
V/Ω terminal of the output.
2. Attach the opposite ends of the leads to the input terminals of the
controlled device, ensuring that the terminal polarities are correct.
Red
Black
Outputting DC Voltage
1. Press the MEASURE/SOURCE button to switch to the output mode.
The SOURCE indicator will illuminate on the display. By default, the
calibrator will be in DC voltage mode with an initial measurement
range of 100mV. The "mV" symbol on the right side of the main
display will light up.
2. Press the RANGE button to select the desired measurement range
(options: 100mV, 1V).
3. Set the output value using the output setting buttons:
Note: Press the ZERO button to reset the output setting to the default
value (0).
Analog Output Resistance
1.Press the FUNC button to switch to the analog resistance function.
The "Ω" symbol on the right side of the main display will light up,
indicating resistance mode.
2.Press the RANGE button to select the desired measurement range
(400Ω / 4KΩ).
Note: For the 400Ω range, two excitation currents are available: 1mA
and 0.1mA. You can manually switch between them to improve accuracy
based on your needs.
To set the excitation current:
• Press the CONFIG button to enter the excitation current setting
menu. "OHM.IS" will appear in the secondary display area at the
lower right corner.
• The main display area will show "I-H" for 1mA or "I-L" for 0.1mA.
• Use the
• Press the ZERO button to save the setting.
• Press CONFIG button to exit the setting menu.
3.Set the output value using the output setting buttons:
• Change the digit to be adjusted using the
• Adjust the value using the buttons.
Note: If the excitation current does not match the selected range (400Ω), the
excitation current indicator on the right side of the display will flash. In this case,
return to the setting interface to adjust the excitation current appropriately.
buttons to select the excitation current.
buttons.
Analog Output Thermocouple
1.
Press the FUNC button to switch to the thermocouple (TC) function.
The "TC" indicator in the upper right corner of the display will light up.
2.Press the RANGE button to select the desired thermocouple type (R,
S, K, E, J, T, N, B, A, C, D).
3.If cold junction compensation is required, press RJ-ON to activate
the cold junction compensation function. If not, skip this step.
There are two types of cold junction compensation:
Automatic: The instrument uses its internal temperature sensor to detect
the ambient temperature automatically.
Manual: The user sets the cold junction compensation temperature
manually.
To set cold junction compensation:
• Press the CONFIG button to enter the cold junction compensation
settings menu. "RJST" will appear in the secondary display area at
the lower right corner.
• The main display will show "MANU" for manual compensation or
"AUTO" for automatic compensation.
• Use the
automatic compensation.
• Press the ZERO button to save the setting.
• If manual compensation is selected, you will be prompted to set the
manual temperature. The secondary display will show "RJVA," and
the main display will show the temperature to be set. Use the
to 50°C.
• Press the ZERO button to save the setting.
• Press CONFIG button to exit the setting menu.
Note: In automatic cold junction compensation, the current ambient
temperature will be displayed in the lower-left corner of the screen.
In manual cold junction compensation, the user-set compensation
temperature will be displayed, and the "M" symbol will light up.
4. Set the output value using the output setting buttons:
• Change the digit to be adjusted using the
• Adjust the value using the
buttons to switch between manual and
buttons to adjust the temperature within the range of -10°C
Press the FUNC button to switch to the analog thermal resistance
output function. The "RTD" indicator in the upper left corner of the
display will light up.
2. Press the RANGE button to select the desired thermal resistance
type (PT100, PT200, PT500, PT1000, Cu50).
For PT100 and Cu50 divisions, two excitation currents are available (1mA
and 0.1mA). The user can switch between them for more accurate thermal
resistance simulation.
To set the excitation current:
• Press the CONFIG button to enter the excitation current setting
interface. "OHMI.S" will appear in the secondary display area at the
lower right corner.
• The main display will show "I-H" for 1mA or "I-L" for 0.1mA.
• Use the
• Press the ZERO button to save the setting.
• Press CONFIG button to exit the setting menu.
3. Set the output value using the output setting buttons:
• Change the digit to be adjusted using the
• Adjust the value using the
Note: For PT100 and Cu50 divisions, if the selected excitation current
does not match the required value, the excitation current indicator on the
right side of the display will flash. You will need to re-enter the setting
interface to adjust the excitation current accordingly.
• Press the ZERO button to perform a reset operation in any
measurement range of the DC voltage, resistance, thermocouple, or
thermal resistance output functions.
•
This action will reset the output setting value to the default initial value,
providing a convenient way for users to quickly return to the starting point.
Measurement Mode
Warning: When using the calibrator in measurement mode, ensure
that all connections are made with the proper leads. The maximum
allowable voltage between any input terminal and ground is 60V peak.
Exceeding this limit may result in electric shock. Never apply a voltage
higher than this maximum to the terminal or ground.
If the measured value exceeds the allowable measurement range, the
"OL" (Overload) symbol will appear in the main display area, indicating
that the input is out of range.
Connecting Leads to Input Terminals
Connection Method for DC Voltage, Thermocouple, Thermal
Resistance, and Resistance (2W) Measurement:
1. Connect the black lead to the COM terminal of the input.
2. Connect the red lead to the VΩ terminal of the input.
3. Attach the other ends of the two leads to the measurement terminals
of the device under test.
4. Ensure that the terminal polarity is correct to ensure
Warning: Disconnect Power Supply: Before connecting the calibrator
to the device under test, ensure that the device's power supply
is disconnected.
Avoid Damage and Injury: Incorrect operation or measurement can
potentially damage the instrument or cause injury. Exercise extreme
caution and follow proper procedures during all measurement operations
.
Measuring DC Voltage
1. Ensure that the measurement lead is disconnected from the device
under test.
2. Press the MEASURE/SOURCE button to switch to measurement
mode. The MEASURE indicator will light up on the display screen.
By default, the calibrator will be set to DC voltage function with a
measurement range of 100mV. The "mV" symbol will appear on the
right side of the main display area.
3. Press the RANGE button to select the desired measurement range
(100mV, 1V, 30V).
4. Connect the measurement lead to the device under test. The
measured voltage value will be displayed in the main display area.
Measuring Resistance
1. Ensure that the measurement lead is disconnected from the device
under test.
2. Press the FUNC button to switch to the resistance function. The "Ω"
symbol will light up on the display screen.
3. Press the RANGE button to select the desired measurement range
(500Ω or 5KΩ).
4. Set the resistance wire system:
• Press the CONFIG button to enter the wire system setting
interface. The secondary display area at the lower right corner will
show "Wire System"
• The main display area will show the available wire systems
(2W, 3W, 4W).
5. Connect the measurement lead to the measurement terminals of the
device under test. The measured resistance value will be displayed in
the main display area.
Measuring Thermocouples
1. Ensure that the measurement lead is disconnected from the device
under test.
2. Press the FUNC button to switch to the thermocouple function. The
"TC" symbol will light up on the display screen.
3. Press the RANGE button to select the required thermocouple type
(R, S, K, E, J, T, N, B, A, C, D).
4. If cold junction compensation is required, press the RJ-ON button
to enable this function. If cold junction compensation is not required,
skip this step.
Setting Cold Junction Compensation:
• Automatic Temperature Compensation: The instrument uses its
internal temperature sensor to detect the ambient
temperature automatically.
• Manual Temperature Compensation: The user sets the
compensation temperature manually.
To set the excitation current:
•
Press the CONFIG button to enter the cold junction compensation
setting interface. "RJST" will appear in the secondary display area at
the lower right corner, indicating the compensation method settings
• The main display will show "MANU" for manual temperature
compensation or "AUTO" for automatic temperature compensation.
• Use the buttons to select the desired excitation current.
Follow the steps below to access and change the default calibrator settings.
Accessing the Settings
1. Press and hold the Backlight button.
2. While holding the Backlight button, press the Power button to turn
on the instrument.
3. Once the settings interface appears, release the Backlight button.
Enabling/Disabling Auto Power OFF
1. Upon entering the settings interface, "APOF" will appear on the
display, indicating the automatic shutdown setting.
2. Use the buttons to adjust the automatic shutdown time
between 0 and 60 minutes. A setting of 0 disables the automatic
shutdown; other values represent the time in minutes after which the
instrument will shut down.
3. Press the ZERO button to conrm selection. The display will show
"SAVE," conrming that the new setting has been saved.
Note: You can turn the meter OFF and then back ON at any time to
resume normal operation.
Setting Backlight Time
1. Press the CONFIG button until "BLOF" appears on the display,
indicating the backlight time setting.
2. Use the buttons to adjust the backlight time setting
between 0 and 3600 seconds. A setting of 0 disables the automatic
backlight shutdown; other values represent the time in seconds after
which the backlight will turn off.
3. Press the ZERO button to conrm selection. The display will show
"SAVE," conrming that the new setting has been saved.
Note: You can turn the meter OFF and then back ON at any time to
resume normal operation.
1. Press the CONFIG button until "LTOF" appears on the display,
indicating the ashlight time setting.
2. Use the buttons to set the desired ashlight time settings
between 0-30 minutes. A setting of 0 disables the automatic
ashlight shutdown; other values represent the time in minutes after
which the ashlight will turn off.
3. Press the ZERO button to conrm selection. The display will show
"SAVE," conrming that the new setting has been saved.
Note: You can turn the meter OFF and then back ON at any time to
resume normal operation.
Enabling/Disabling the Beeper Sound
1. Press the CONFIG button until "BEEP" appears on the display,
indicating the buzzer setting.
2. Use the buttons to toggle between ON (Enabled) or
OFF (Disabled).
3. Press the ZERO button to conrm selection. The display will show
"SAVE," conrming that the new setting has been saved.
Note: You can turn the meter OFF and then back ON at any time to
resume normal operation.
Selecting the Temperature Unit of Measure
1. Press the CONFIG button until "TEPU" appears on the display,
indicating the temperature unit setting.
2. Use the buttons to select between °C and °F.
3. Press the ZERO button to conrm selection. The display will show
"SAVE," conrming that the new setting has been saved.
Note: You can turn the meter OFF and then back ON at any time to
resume normal operation.
Warning: To prevent electric shock, always remove the test leads from
the calibrator before opening the battery compartment. Ensure the
battery door is securely closed before using the calibrator.
1. Turn off the calibrator and remove the test leads.
2. Use a athead screwdriver to turn the battery door screws a quarter
turn counterclockwise and remove the door.
3. Replace the 3 x AA batteries in the battery compartment.
4. Close the battery door and securely tighten the screws.
Applications
• Ideal for calibrating devices that use thermocouples as input signals
(e.g., meters, controllers, transmitters, data loggers,
and recorders)
• Essential for applications requiring precise thermocouple temperature
R2920 Surface Thermocouple Probe
R2930 Right Angle Thermocouple Probe
R2940 Air/Gas Thermocouple Probe
R2950 Immersion Thermocouple Probe
R2960 Needle Tip Thermocouple Probe
R2970 Type K Pipe Clamp Thermocouple Probe
R2980 Type K Air Oven/Freezer Thermocouple Probe
TP-01 Type K Beaded Wire Probe
CA-05A Soft Carrying Case
R9940 Hard Shell Carrying Case
R1000 Safety Test Lead SetW
R1020 Fused Test Lead Set
R5900 Magnetic Meter Strap
Don't see your part listed here? For a complete list of all accessories and
replacement parts visit your product page on www.REEDInstruments.com.
Product Care
To keep your instrument in good working order we recommend the following:
• Store your product in a clean, dry place.
• Change the battery as needed.
• If your instrument isn't being used for a period of one month or longer
please remove the battery.
• Clean your product and accessories with biodegradable cleaner. Do not
spray the cleaner directly on the instrument. Use on external parts only.
REED Instruments guarantees this instrument to be free of defects in
material or workmanship for a period of one (1) year from date of shipment.
During the warranty period, REED Instruments will repair or replace, at no
charge, products or parts of a product that proves to be defective because
of improper material or workmanship, under normal use and maintenance.
REED Instruments total liability is limited to repair or replacement of the
product. REED Instruments shall not be liable for damages to goods,
property, or persons due to improper use or through attempts to utilize the
instrument under conditions which exceed the designed capabilities. In
order to begin the warranty service process, please contact us by phone at
1-877-849-2127 or by email at [email protected] to discuss the
claim and determine the appropriate steps to process the warranty.
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling your
instrument. Your product contains electronic components and must be
disposed of separately from standard waste products.
Product Support
If you have any questions on your product, please contact your authorized
REED distributor or REED Instruments Customer Service by phone at
1-877-849-2127 or by email at [email protected].
Please visit www.REEDInstruments.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice.
All rights reserved. Any unauthorized copying or reproduction of
this manual is strictly prohibited without prior written permission
from REED Instruments.